Abstract
Optical coherence tomography (OCT) is an interferometry-based imaging modality that generates high-resolution cross-sectional images of the retina. Circumpapillary retinal nerve fiber layer (cpRNFL) and optic nerve head assessments are the mainstay of glaucomatous structural measurements in OCT. However, because these measurements are not always available or precise, it would be useful to have another reliable indicator. The macula has been suggested as an alternative scanning location for glaucoma diagnosis. Using time-domain (TD-) OCT, macular measurements have shown to provide good glaucoma diagnostic capabilities. With the adoption of spectral-domain OCT, which allows a higher image resolution than TD-OCT, segmentation of inner macular layers becomes possible. These layers are specifically prone to glaucomatous damage and thickness measurements show a comparable performance to that of glaucomatous cpRNFL measurements. The role of macular measurements for detection of glaucoma progression is still under investigation. More sophisticated measurement and analysis tools that can amplify the advantages of macular measurements are expected. For example, improvement of image quality would allow better visualization, development of various scanning modes would optimize macular measurements, and further refining of the analytical algorithm would provide more accurate segmentation. With these achievements, macular measurement can be an important surrogate for glaucomatous structural assessment.
Introduction
Glaucoma is an optic neuropathy that is characterized by progressive loss of the retinal ganglion cells (RGCs) and their axons in the retinal nerve fiber layer (RNFL), thinning of the neuroretinal rim in the optic nerve head (ONH), and visual field (VF) deficit. During clinical evaluation, examining the ONH and surrounding area, focusing on features suggesting neural tissue loss, assesses the glaucomatous structural changes. Complicating this assessment is the high variability of the ONH size and shape even among healthy subjects: a wide range of optic disc and cup sizes, variable size and configuration of the blood vessels, variable angle of penetration into the eyeball of the optic nerve (titled disc), and peripapillary changes, such as atrophy and others. Taking into consideration this marked variability of ONH size and shape, it is not surprising that the ability of detecting glaucoma, especially at an early stage of the disease, varies considerably among clinicians. In the last two decades, a few ocular imaging devices have been introduced to allow micron scale quantification of ocular structures. Optical coherence tomography (OCT) is an interferometry-based ocular imaging modality that generates high-resolution cross-sectional images of the retina. This device has been shown to be valuable in the diagnosis and monitoring of retinal diseases and glaucoma.1–7 Clinical utility of OCT in glaucoma is predominantly based on assessment of the circumpapillary RNFL8–12 because it allows a thorough sample of all retinal axons as they approach the ONH. However, the variability of ONH size and shape described above, along with pathologic features, such as large peripapillary atrophy, papilledema, and others, might affect the reliability of the circumpapillary RNFL measurements. Zimmer et al. first suggested imaging of the macula as a potential location for glaucoma evaluation.13 In this manuscript, we review the clinical significance, previous works, and future direction of macular imaging by OCT in glaucoma.
Why is macular imaging by OCT meaningful in glaucoma?
Compared to the ONH, the macula is a relatively simple structure that is devoid of large vessels. It has multiple cellular and plexiform layers and a central depression (fovea) devoid of RGCs. The macula offers several potential physiological and anatomical advantages for glaucoma evaluation. Firstly, the retinal nerve fiber is composed of the RGC axons, and therefore assessment of the RGCs may be a more direct method for measuring glaucomatous damage than circumpapillary RNFL thickness. Additionally, the macula is the only place in the retina where more than a single RGC body exists in the ganglion cell layer. Because the cell body is substantially larger than the soma of the cell, this might improve the ability to detect damage to these cells.13–16 Furthermore, within the macula, more than 50% of the RGCs of the entire retina are located; thus scanning of the macula allows sampling of the majority of the RGCs. The macula shape, more specifically the RGC layer, is generally less variable among healthy individuals than other diagnostically important structures, such as the RNFL and ONH. Sensitivity of the RGCs could potentially be higher than that of the RNFL because changes in this layer would more likely be the result of a pathologic process rather than of normal variation.17
Although glaucomatous macular changes are difficult to detect clinically, OCT allows accurate quantitative assessment. Imaging of this region is easier and less prone to eye motion artifacts because, unlike ONH imaging, macular imaging does not require eccentric fixation. Eccentric fixation is known to affect OCT measurements.18 Recent development of spectral domain (SD-) OCT technology allows image acquisition at faster speeds than were possible with conventional time domain (TD-) OCT. As a result, more scans can be acquired and combined to create 3D macular images.19 Due to the relatively slow acquisition rate of TD-OCT, the conventional scanning pattern used for the macula was six evenly distributed radial scans with interpolation between neighboring scans. The interpolation mostly affected the outer part of the macula, where the scans are further apart, which is also the location of most glaucomatous damage. SD-OCT scans either employ a raster scan pattern or a denser collection of radial scans with less interpolation than what was employed for TD-OCT images.
Macula thickness performance with TD-OCT
Several studies evaluated the diagnostic capability of the macular thickness.1–7 Some reports compared the diagnostic abilities of macular thickness, RNFL thickness, and ONH parameters. Using an early version of TD-OCT, Giovannini et al. showed that volumetric analysis of the macula correlated significantly with glaucoma status.2 Greenfield et al. reported that macular thickness changes correlated with changes in visual function and circumpapillary RNFL in glaucoma and may be a surrogate indicator of RGC loss.3 Wollstein et al. reported that macular thickness, as measured by OCT, was capable of detecting glaucomatous damage and corresponded with the circumpapillary RNFL thickness; however, circumpapillary RNFL thickness had a higher sensitivity and specificity for detection of VF abnormalities.1 Leung et al. also showed that circumpapillary RNFL thickness outperformed both total macular and macular NFL thickness in terms of glaucoma detection and visual function correlation.6 Medeiros et al. reported that RNFL and ONH measurements had better discriminating performance than macular measurements when assessed by a later version of TD-OCT.7 Study outcomes comparing RNFL and macular thickness measurements are summarized as Table 1. These studies indicated that although macular thickness has good glaucoma diagnostic capabilities, circumpapillary RNFL measurement performance is superior. One possible explanation might be due to the use of the total macular thickness, which contains all retinal layers. Some of the retinal layers are not involved in the glaucomatous process and thus may reduce the sensitivity and specificity of diagnosis. This leads to concrete efforts in further segmenting the macula to allow quantification of individual retinal layers.
Table 1.
Authors | OCT device | Main finding |
---|---|---|
Wollstein et al | Prototype OCT | RNFL thickness had higher sensitivity and specificity for the detection of visual field abnormalities than macular thickness |
Leung et al | Stratus OCT | RNFL thickness outperformed both total macular and macular NFL thickness in terms of glaucoma detection and visual function correlation |
Medeiros et al | Stratus OCT | The RNFL parameter inferior thickness was significantly better than the macular thickness parameter |
Na et al | Cirrus OCT | RNFL thickness measurements were generally superior to those of macular thickness |
Seong et al | RTVue OCT | Macular inner retinal thickness showed glaucoma discrimination ability comparable to that of RNFL thickness in patients with early visual field defects In eyes with advanced or peripheral defect, RNFL measurement showed a better glaucoma diagnostic ability than did MIRL measurement |
Huang et al | RTVue OCT | Average RNFL thickness is the optimal individual OCT parameter to detect perimetric glaucoma Simultaneous evaluation on disc morphology, RNFL, and macular inner retinal thickness can improve the diagnostic accuracy in diagnosing glaucoma. |
Nakatani et al | Topcon 3D OCT | For the diagnosis of early glaucoma, macular parameters had high discriminating power and high reproducibility comparable with RNFL parameters |
Segmentation of macular layers
The macula is composed of multiple layers organized from innermost to outermost: inner limiting, nerve fiber, RGC, inner plexiform, inner nuclear, outer plexiform, outer nuclear, and the retinal pigment epithelial (RPE) layers. Glaucoma primarily affects the axon and body of RGCs, which constitutes the inner retina.
Differentiation of individual retinal layers by ocular imaging requires a high level of resolution and an advanced segmentation algorithm. Ishikawa et al. developed a customized segmentation algorithm for macular segmentation using TD-OCT images.16 They demonstrated comparable glaucoma diagnostic capability of the inner retinal layers with circumpapillary RNFL. Moreover, inner retinal layer thickness was statistically significantly better in discriminating between healthy and glaucomatous eyes than total macular thickness. Wang et al. measured the RGC layer thickness in glaucomatous eyes with an SD-OCT device, using a computer-aided manual segmentation procedure.20 They reported that it was feasible to obtain local measurements of RGC thickness that corresponded to functional findings.
Manufacturers of SD-OCT devices incorporated automatic macular segmentation analysis into their operating system. Ganglion cell complex (GCC) was designated for thickness measurements from the internal limiting membrane to the inner nuclear layer, which is composed of RGCs, along with their axons and dendrites. Other manufacturers segmented the inner retina into the ganglion cell inner plexiform layer (GCIPL). Cho et al. examined the relationship between VF mean sensitivity (MS) and macular GCC thickness, and reported a similar level of correlation as with circumpapillary RNFL thickness.21 Rao et al. also demonstrated that the strongest structure-function association among the various macular measurements using SD-OCT was found to be the inner retinal thickness.22
Assessing the diagnostic performance of GCC thickness, Seong et al. reported that GCC thickness was comparable to circumpapillary RNFL thickness in terms of glaucoma diagnostic capability.23 Sakamoto et al. compared macular RNFL images obtained by 3D SD-OCT with those obtained by color and red free fundus photography. They found that more macular RNFL defects were detected on 3D SD-OCT images than on color fundus photographs.24 Kim et al. classified glaucoma into three disease severity groups based on VF’s mean deviation. They reported that the macular GCC and circumpapillary RNFL thicknesses showed similar diagnostic performances in detecting early, moderate, and severe glaucoma.25 Kotera et al. showed that the mean macular inner retinal thickness was significantly thinner in suspected glaucoma and preperimetric glaucomatous eyes than in the healthy eyes, while mean total retinal and macular NFL thicknesses were not.26 This report might point out the utility of macular inner retinal measurements as an early indicator of glaucomatous change.
Assessing covariability associated with macular inner retinal thickness, Kim et al. showed that thin GCC thickness correlated with older age and longer axial length.27 Similarly, Mwanza et al. reported that thinner GCIPL was associated with thinner RNFL, older age, longer ocular axial length, and male sex.28
Taken together, published studies indicated that the segmented macular inner retinal layer thickness performs better than the total macular thickness and similar to (but not better than) the circumpapillary RNFL thickness in glaucoma diagnosis.
Clinical application
In clinical practice, situations where glaucomatous changes were detected in the circumpapillary RNFL without corresponding damage in the macula or vice versa have been encountered. Na et al. classified cases according to whether they were better diagnosed by macular thickness or circumpapillary RNFL measurement.29 Overall, more eyes were diagnosed by circumpapillary RNFL than by macular measurements. Eyes with exclusive macular damage tended to have larger ONH sizes than eyes that had solely abnormal circumpapillary RNFL measurements. It was suggested that this difference in diagnostic capability was due to measuring circumpapillary RNFL thickness in a fixed-sized circumpapillary diameter (3.4 mm), which brought the sampling circle closer to the disc margin in a larger ONH. Since RNFL thickness measured close to the disc margin tends to be thicker, this might mask early glaucomatous structural abnormality.
The papillomacular bundle has been shown to be resistant to glaucomatous damage until the end stage of the disease. Thus, measurement of the posterior pole macular thickness may be a strategy to measure advanced glaucomatous function. However, progressive RNFL loss is not easy to detect in advanced glaucoma because most of the RNFL is already lost at this stage. Sung et al. reported that more than half of their advanced glaucoma participants (VF mean deviation < −10 dB) could not be evaluated by photographic assessment of the ONH and RNFL because these structures exhibited advanced glaucomatous abnormalities that precluded clinical assessment of structural changes.30 The participants showed a much higher rate of progression in macular thickness than with circumpapillary RNFL thickness and a better agreement with progression by VF.
Glaucoma is typically described as causing a focal abnormality, especially in early stages of the disease. In order to exploit this in a fashion similar to the one employed by the glaucoma hemifield test in VF, Um et al. categorized the posterior-pole macular area into five regions in each hemifield.31 The difference between corresponding locations was compared to the difference between these locations in healthy eyes. The asymmetry in hemifield macular thickness showed better glaucoma diagnostic capability than the average circumpapillary RNFL thickness in early glaucomatous eyes.
Progression detection
Clinical utility of ocular imaging devices could be further amplified by detecting disease progression in glaucoma. Most OCT progression studies conducted so far were confined to circumpapillary RNFL measurements and only a few studies evaluated macular thickness measurements. Medeiros et al. reported that circumpapillary RNFL thickness outperformed macular thickness and optic disc parameters in detecting glaucoma progression using TD-OCT.32 Needless to say, measurement repeatability is of paramount importance in progression detection. Mwanza et al recently demonstrated higher measurement reproducibility of macular ganglion cell layer thickness by use of SD OCT.17 Recently, Na et al reported the glaucoma progression detection capability of macular ganglion cell layer thickness. According to their results, ganglion cell layer thickness showed similar sensitivity to RNFL or total macular thickness in terms of agreement with progression determined by optic disc/RNFL photographic or VF assessment.33
The denser scanning and improved measurements reproducibility offered by SD-OCT hold promise to improve detection of structural progression. However, more studies to confirm this hypothesis are yet to be published.
Limitation
In the mean time, a limitation of using macular thickness measurements for glaucoma diagnosis should be also considered. One of the factors that may affect macular measurements for glaucoma detection and the relationship with functional tests is the spatial summation within central vision. Within the macula, the spatial summation becomes a significant factor affecting the structure-function relationship, which subsequently affects diagnostic performance as many studies use visual field defects as the definition of glaucoma.
Future directions
To enhance the clinical utility of macular measurements obtained by OCT in glaucoma diagnosis, image quality itself must be improved because the analytic outcome depends heavily on image quality. Reduced image quality substantially affects the segmentation of the retinal layers and reliable and repeatable measurements. In particular, segmentation of inner and outer macular layers demands high quality image. Enhanced image quality facilitates precise segmentation of various cellular layers of the macula. By recent development of SD-OCT technology, image resolution has been improved. By use of SD-OCT, Mwanza et al showed the test-retest standard deviation of GCIPL thickness was 1.43 μm. There has been an effort made to improve image quality using various techniques. Nakano et al suggested that speckle noise was the primary artifact in OCT images, and a speckle noise-reducing technique was adopted by using eye-tracking and averaging, which allowed clearer visualization and measurement of the macular ganglion cell layer.34 Development of various applications of scanning modes to optimize macular measurements is warranted, as is further refining of the analytical algorithm for accurate segmentation.
Conclusion
Circumpapillary RNFL and ONH assessments are the mainstay of glaucomatous structural measurements. However, because these measurements cannot always be obtained, it would be useful to have another reliable indicator for glaucomatous structural assessment by OCT. Several studies show that the macular inner retinal layer acts comparably to circumpapillary RNFL in glaucoma diagnosis. Investigations are ongoing into its use for detection of glaucoma progression. More sophisticated measurement and analysis tools that can amplify the advantages of macular measurements are expected. With these achievements, macular measurement can be an important clinical surrogate for glaucomatous structural assessment.
Acknowledgments
Funding
This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.
Footnotes
Competing Interest
Kyung Rim Sung; none, Gadi Wollstein; none, Nae Rae Kim; none, Jung Hwa Na; none, Jessica E. Nevins; none, Chan Yun Kim, MD, PhD; none, Joel S. Schuman has intellectual property licensed by Massachusetts Institute of Technology to Carl Zeiss Meditec.
Authorship contribution
Kyung Rim Sung; conception and design, acquisition of data, analysis and interpretation of data, drafting the article, critical revision, final approval of the version
Gadi Wollstein; drafting the article, critical revision, final approval of the version
Na Rae Kim, MD; drafting the article, critical revision
Jung Hwa Na, MD; drafting the article
Jessica E. Nevins; critical revision
Chan Yun Kim, MD, PhD; drafting the article, critical revision
Joel S. Schuman, MD; critical revision, final approval of the version
The Corresponding Author has the right to grant on behalf of all authors and does grant on behalf of all authors, an exclusive licence on a worldwide basis to the BMJ Publishing Group Ltd and its Licensees to permit this article (if accepted) to be published in BJO editions and any other BMJPGL products to exploit all subsidiary rights, as set out in our licence.
References
- 1.Wollstein G, Schuman JS, Price LL, et al. Optical coherence tomography (OCT) macular and peripapillary retinal nerve fiber layer measurements and automated visual fields. Am J Ophthalmol. 2004;138:218–225. doi: 10.1016/j.ajo.2004.03.019. [DOI] [PubMed] [Google Scholar]
- 2.Giovannini A, Amato G, Mariotti C. The macular thickness and volume in glaucoma: an analysis in normal and glaucomatous eyes using OCT. Acta Ophthalmol Scand Suppl. 2002;236:34–36. doi: 10.1034/j.1600-0420.80.s236.44.x. [DOI] [PubMed] [Google Scholar]
- 3.Greenfield DS, Bagga H, Knighton RW. Macular thickness changes in glaucomatous optic neuropathy detected using optical coherence tomography. Arch Ophthalmol. 2003;121:41–46. doi: 10.1001/archopht.121.1.41. [DOI] [PubMed] [Google Scholar]
- 4.Guedes V, Schuman JS, Hertzmark E, et al. Optical coherence tomography measurement of macular and nerve fiber layer thickness in normal and glaucomatous human eyes. Ophthalmology. 2003;110:177–189. doi: 10.1016/s0161-6420(02)01564-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Lederer DE, Schuman JS, Hertzmark E, et al. Analysis of macular volume in normal and glaucomatous eyes using optical coherence tomography. Am J Ophthalmol. 2003;135:838–843. doi: 10.1016/s0002-9394(02)02277-8. [DOI] [PubMed] [Google Scholar]
- 6.Leung CK, Chan WM, Yung WH, et al. Comparison of macular and peripapillary measurements for the detection of glaucoma: an optical coherence tomography study. Ophthalmology. 2005;112:391–400. doi: 10.1016/j.ophtha.2004.10.020. [DOI] [PubMed] [Google Scholar]
- 7.Medeiros FA, Zangwill LM, Bowd C, et al. Evaluation of retinal nerve fiber layer, optic nerve head, and macular thickness measurements for glaucoma detection using optical coherence tomography. Am J Ophthalmol. 2005;139:44–55. doi: 10.1016/j.ajo.2004.08.069. [DOI] [PubMed] [Google Scholar]
- 8.Burgansky-Eliash Z, Wollstein G, Chu T, et al. Optical coherence tomography machine learning classifiers for glaucoma detection: a preliminary study. Invest Ophthalmol Vis Sci. 2005;46:4147–52. doi: 10.1167/iovs.05-0366. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Sung KR, Kim DY, Park SB, et al. Comparison of retinal nerve fiber layer thickness measured by Cirrus HD and Stratus optical coherence tomography. Ophthalmology. 2009;116:1264–70. doi: 10.1016/j.ophtha.2008.12.045. [DOI] [PubMed] [Google Scholar]
- 10.Park SB, Sung KR, Kang SY, et al. Comparison of glaucoma diagnostic Capabilities of Cirrus HD and Stratus optical coherence tomography. Arch Ophthalmol. 2009;127:1603–9. doi: 10.1001/archophthalmol.2009.296. [DOI] [PubMed] [Google Scholar]
- 11.Cho JW, Sung KR, Hong JT, et al. Detection of Glaucoma by Spectral Domain- Scanning Laser Ophthalmoscopy/Optical Coherence Tomography (SD-SLO/OCT) and Time Domain Optical Coherence Tomography. J Glaucoma. 2011;20:15–20. doi: 10.1097/IJG.0b013e3181d1d332. [DOI] [PubMed] [Google Scholar]
- 12.Shin CJ, Sung KR, Um TW, et al. Comparison of Retinal Nerve Fiber Layer Thickness Measurements Calculated by the Optic Nerve Head Map (NHM4) and RNFL 3.45 Modes of Spectral-domain Optical Coherence Tomography (OCT) (RTVue-100) Br J Ophthalmol. 2010;94:763–7. doi: 10.1136/bjo.2009.166314. [DOI] [PubMed] [Google Scholar]
- 13.Zeimer R, Asrani S, Zou S, et al. Quantitative detection of glaucomatous damage at the posterior pole by retinal thickness mapping. A pilot study Ophthalmology. 1998;105:224–31. doi: 10.1016/s0161-6420(98)92743-9. [DOI] [PubMed] [Google Scholar]
- 14.Ojima T, Tanabe T, Hangai M, et al. Measurement of retinal nerve fiber layer thickness and macular volume for glaucoma detection using optical coherence tomography. Jpn J Ophthalmol. 2007;51:197–203. doi: 10.1007/s10384-006-0433-y. [DOI] [PubMed] [Google Scholar]
- 15.Tan O, Li G, Lu AT, et al. Mapping of macular substructures with optical coherence tomography for glaucoma diagnosis. Ophthalmology. 2008;115:949–56. doi: 10.1016/j.ophtha.2007.08.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Ishikawa H, Stein DM, Wollstein G, et al. Macular segmentation with optical coherence tomography. Invest Ophthalmol Vis Sci. 2005;46:2012–7. doi: 10.1167/iovs.04-0335. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Mwanza JC, Oakley JD, Budenz DL, et al. Macular ganglion cell-inner plexiform layer: automated detection and thickness reproducibility with spectral domain-optical coherence tomography in glaucoma. Invest Ophthalmol Vis Sci. 2011;52:8323–9. doi: 10.1167/iovs.11-7962. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Campbell RJ, Coupland SG, Buhrmann RR, et al. Effect of eccentric and inconsistent fixation on retinal optical coherence tomography measures. Arch Ophthalmol. 2007;125:624–7. doi: 10.1001/archopht.125.5.624. [DOI] [PubMed] [Google Scholar]
- 19.Wojtkowski M, Bajraszewski T, Gorczyńska I, et al. Ophthalmic imaging by spectral optical coherence tomography. Am J Ophthalmol. 2004;138:412–9. doi: 10.1016/j.ajo.2004.04.049. [DOI] [PubMed] [Google Scholar]
- 20.Wang M, Hood DC, Cho JS, et al. Measurement of local retinal ganglion cell layer thickness in patients with glaucoma using frequency-domain optical coherence tomography. Arch Ophthalmol. 2009;127:875–81. doi: 10.1001/archophthalmol.2009.145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Cho JW, Sung KR, Lee S, et al. Relationship between visual field sensitivity and macular ganglion cell complex thickness as measured by spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51:6401–7. doi: 10.1167/iovs.09-5035. [DOI] [PubMed] [Google Scholar]
- 22.Rao HL, Zangwill LM, Weinreb RN, et al. Structure-function relationship in glaucoma using spectral-domain optical coherence tomography. Arch Ophthalmol. 2011;129:864–71. doi: 10.1001/archophthalmol.2011.145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Seong M, Sung KR, Choi EH, et al. Macular and peripapillary retinal nerve fiber layer measurements by spectral domain optical coherence tomography in normal-tension glaucoma. Invest Ophthalmol Vis Sci. 2010;51:1446–52. doi: 10.1167/iovs.09-4258. [DOI] [PubMed] [Google Scholar]
- 24.Sakamoto A, Hangai M, Nukada M, et al. Three-dimensional imaging of the macular retinal nerve fiber layer in glaucoma with spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2010;51:5062–70. doi: 10.1167/iovs.09-4954. [DOI] [PubMed] [Google Scholar]
- 25.Kim NR, Lee ES, Seong GJ, et al. Structure-function relationship and diagnostic value of macular ganglion cell complex measurement using Fourier-domain OCT in glaucoma. Invest Ophthalmol Vis Sci. 2010;51:4646–51. doi: 10.1167/iovs.09-5053. [DOI] [PubMed] [Google Scholar]
- 26.Kotera Y, Hangai M, Hirose F, et al. Three-dimensional imaging of macular inner structures in glaucoma by using spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2011;52:1412–21. doi: 10.1167/iovs.10-5572. [DOI] [PubMed] [Google Scholar]
- 27.Kim NR, Kim JH, Lee J, et al. Determinants of perimacular inner retinal layer thickness in normal eyes measured by Fourier-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2011;52:3413–8. doi: 10.1167/iovs.10-6648. [DOI] [PubMed] [Google Scholar]
- 28.Mwanza JC, Durbin MK, Budenz DL, et al. Profile and predictors of normal ganglion cell-inner plexiform layer thickness measured with frequency-domain optical coherence tomography. Invest Ophthalmol Vis Sci. 2011;52:7872–9. doi: 10.1167/iovs.11-7896. [DOI] [PubMed] [Google Scholar]
- 29.Na JH, Sung KR, Baek S, et al. Macular and retinal nerve fiber layer thickness: which is more helpful in the diagnosis of glaucoma? Invest Ophthalmol Vis Sci. 2011;52:8094–101. doi: 10.1167/iovs.11-7833. [DOI] [PubMed] [Google Scholar]
- 30.Sung KR, Sun JH, Na JH, et al. Progression Detection Capability of Macular Thickness in Advanced Glaucomatous Eyes. Ophthalmology. 2011;119:308–13. doi: 10.1016/j.ophtha.2011.08.022. [DOI] [PubMed] [Google Scholar]
- 31.Um TW, Sung KR, Wollstein G, et al. Asymmetry in hemifield macular thickness as an early indicator of glaucomatous change. Invest Ophthalmol Vis Sci. 2012;53:1139–44. doi: 10.1167/iovs.11-8373. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Medeiros FA, Zangwill LM, Alencar LM, et al. Detection of glaucoma progression with stratus OCT retinal nerve fiber layer, optic nerve head, and macular thickness measurements. Invest Ophthalmol Vis Sci. 2009;50:5741–8. doi: 10.1167/iovs.09-3715. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Na JH, Sung KR, Baek S, et al. Detection of Glaucoma Progression by Assessment of Segmented Macular Thickness Data Obtained Using Spectral Domain Optical Coherence Tomography. Invest Ophthalmol Vis Sci. 2012 May 3; doi: 10.1167/iovs.11-9369. [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
- 34.Nakano N, Hangai M, Nakanishi H, et al. Macular ganglion cell layer imaging in preperimetric glaucoma with speckle noise-reduced spectral domain optical coherence tomography. Ophthalmology. 2011;118:2414–26. doi: 10.1016/j.ophtha.2011.06.015. [DOI] [PubMed] [Google Scholar]