Abstract
The ability to quantify and visualize sub-micron-scale oscillatory motions of objects in three-dimensions has a wide range of application in acoustics, materials sciences and medical imaging. Here we demonstrate that volumetric snapshots of rapid periodic motion can be captured using optical coherence tomography with sub-nanometer-scale motion sensitivity and microsecond-scale temporal resolution. This technique, termed OCT vibrography, was applied to generate time-resolved volumetric vibrographs of a miniature drum driven acoustically at several kilohertz.
Pressure-driven oscillations with nanometer-scale amplitudes at acoustic frequencies are found in a variety of physical and biological measurement systems. For instance, miniature resonators have been used for radiation pressure cooling [1], the sensing of molecules [2] and the high-precision weighing of single cells [3]. Moreover, acoustic vibrations have been used for dynamic mechanical analysis and rheology and for photoacoustic imaging [4] and elastography. In all of these systems, measurements of the acoustic motion on the surface or within the object are essential. Although the identification of the intrinsic parameters, such as the resonance frequency, is sufficient for some cases, other applications require the actual amplitude and phase information, and such applications can greatly benefit from a volumetric imaging technique capable of providing spatial graphs—“vibrographs”—of the sample vibration.
Optical interferometry is ideally suited for precise measurements of oscillatory motion. Laser Doppler velocimetry and stroboscopic holography have been used to measure sub-micron-scale vibrations at frequencies up to MHz [5]. However, these techniques are limited to surface measurements. Optical coherence tomography (OCT) offers the potential for capturing motion at various depths in scattering samples. Phase-sensitive OCT with sub-nanometer amplitude sensitivity has been used for elastography [6-8], vibration-amplitude mapping [9], and phase microscopy of static or slowly moving samples [10, 11].
Here, we demonstrate a new technique capable of providing reconstructed volumetric snapshots of rapidly moving samples. This technique, which we call OCT vibrography, is based on the synchronized acquisition of phase information that is directly related to the vibration of the sample. After post-processing, three-dimensional (3D) snapshots of an object vibrating at acoustic frequencies in response to applied stimuli could be obtained. We show that OCT vibrography using a wavelength-swept laser source can achieve sub-nanometer amplitude sensitivity by minimizing mechanical noise associated with laser tuning and beam scanning. For a proof-of-concept experiment, we applied OCT vibrography to the visualization of the acoustic resonance modes of a membrane.
Figure 1 shows the schematic of our experimental setup used to measure vibrations. The OCT console was an optical frequency domain imaging system based on a wavelength-swept source [12]. The sample was a glass plate mounted on a piezoelectric transducer (PZT; AE0203D08F, Thorlabs). Sinusoidal waveforms from a function generator (33120A, Agilent) were used to drive the PZT actuator. A multi-functional board (DAQ; PCI 6115, National Instruments) was used to acquire interference data at 10 MS/s and control the polygon filter of the laser source to generate a 15-20 kHz A-line rate. The tuning range of the laser was 125 nm with a mean wavelength of 1280 nm. The DAQ board also produced an analog output (16 bit) to control a two-axis galvanometer beam scanner (6210H, Cambridge Technologies) and provided a TTL trigger signal to the function generator. By generating all control signals from the internal time-base clock (20 MHz) of the DAQ board, we achieved synchronization among data acquisition, beam scanning, and sample actuation.
Fig. 1.
(Color online) (a) Schematic of a system. Sig: signal generator; B: galvanometer beam scanner. (b) Schematic of synchronized data acquisition. A number of A-lines, NA, are acquired during each cycle of oscillation, while the scanner (B) is parked at a spatial location for the duration of a user-controlled number of motion cycles (Sig). An arrangement of the data with respect to time generates vibrographs at each motion phase.
The frequency of the driving sinusoidal waveform (Sig) was set at an integer fraction of the A-line rate (15-20 kHz), so that an integer number of A-lines, NA, were acquired during each cycle of sample oscillation. NA also represents the number of motion phases captured in a cycle. The beam scan was synchronized to the oscillation so that after a predetermined number of cycles, NB, the beam was stepped to the next lateral location using the galvanometer beam scanner (B). The lowest possible NA and NB are 2 and 1, respectively, where two motion phases are captured per cycle (Nyquist limit) and the beam is moved after the completion of each vibration cycle. However, NA>6 is highly desirable to accurately reconstruct the sinusoidal motion in terms of amplitude and phase. Figure 1b illustrates a case for NA=8 and NB=2. When the scans over the regions of interest in the sample were completed, the A-line profiles acquired at the same phase of the oscillation (φ=0, 2π/NA, … or 2π(NA−1)/NA) were grouped to produce “snapshots” [13] or vibrographs of the sample.
The intensity and phase of the interference signals from a specific location in the external and internal surfaces of the sample were measured during the oscillatory motion. The phase angle, φ, can be expressed as φ(t) = Δφ sin(2πft) + φn, where Δφ = 4π/λ*δz is the phase amplitude corresponding to the vibration amplitude δz (in nm), f is the vibration frequency, and φn is the intrinsic phase noise given by <δ φn>2=1/(2*SNR), where SNR is the signal-to-noise ratio for the intensity of the interference signal in a single A-line [14]. When the vibration amplitude is determined from N A-lines, the amplitude sensitivity is given by λ/4π (2*SNR*N)-1/2, where N=NA for single-cycle measurement, and N=NA*NB for averaging over NB cycles [6-9]. Spatial averaging over multiple points in a region where the motion is uniform can further enhance the sensitivity. Spatial averaging over multiple points in a region where the motion is uniform can further enhance the sensitivity.
To reach the SNR-limited sensitivity, it was found to be essential to isolate the measurement system from room acoustic noise and mechanical noises, such as the vibration of the rotating polygon filter of the laser, using vibration absorbing platforms and pads made with low resonant frequency (15 Hz) viscoelastic sorbothane polymer. We also found it necessary to reduce the amplitude jitter (1-5 nm in amplitude) from the galvanometer scanner [11]. The noise from the scanner was due to its finite frequency response. We canceled out this deterministic noise in real time by numerically subtracting a pre-recorded trace of the scanner's step response measured from a stationary sample. We estimate the total residual mechanical noise to be <20 pm.
To test the sensitivity of the OCT system, we imaged a sample vibrating at 1.5 kHz with an A-line rate of 15 kHz. The amplitude of vibration was measured to be 7 nm, as confirmed with a laser Doppler velocimeter. We measured the phase angle of the OCT pixel corresponding to the top surface of the glass plate while the beam was fixed. The SNR of the pixel was 30 dB. Figure 2a-inset shows a typical time trace of the phase angle. The spectrum in Fig. 2a shows the Fourier domain trace of the measured phase angles (N=2880; NA=10, NB=288), showing a sample vibration peak at 1.5 kHz and a noise floor of 46 pm, close to the theoretical limit of 42 pm. The SNR was varied using a graded neutral density filter to explore the relationship between the SNR and sensitivity. Then, we measured the average of the random phase variation in a group of A-lines to calculate the amplitude sensitivity. Figure 2b shows the vibration sensitivity at SNR levels ranging from 20 to 40 dB for different numbers (N=1 to 1000) of A-lines averaged. In all cases, SNR-limited theoretical sensitivity was obtained. At SNR=40 and N=1000 (integration time, 0.667 s), the sensitivity was 26 pm.
Fig. 2.
(Color online) (a) Inset is the representative time trace of the measured phase angle. Fourier domain spectrum of the PZT vibration of 288 cycles. The peak points to the vibration frequency of 1.5 kHz. (b) Sensitivity measurement of OCT vibrography. Circles, experimental data for N=1, 10, 100, and 1000. Error bars, s.d. Lines, theoretical curves for SNR-limited sensitivity.
We next used the vibrography system to capture 3D snapshots of an acoustically driven drum head consisting of a 200- μm-thick latex membrane stretched over and glued to a 5-mm diameter metal tube (Fig. 3a). Sinusoidal signals from the function generator were applied to a loudspeaker, and the radiated sound was used to evoke the vibration of the drumhead. The OCT beam was scanned over the entire membrane (NB=1), covering 512 by 256 XY spatial points. The SNR at the drum surface was approximately 40 dB. By sweeping the sound frequency, we found a resonance at 800 Hz for the fundamental resonance mode, the (0,1) mode, where the maximum displacement was observed. Figure 3b shows snapshots of the membrane taken at two opposite vibration phases, φ=0 and π, respectively, at f=800 Hz (NA=20; A-line rate was 16 kHz). The cutaway view reveals the homogeneous vibration across the full thickness of the latex membrane (Fig. 3c). A movie created from the snapshots shows the Bessel profile of the vibration mode with a 3D spatial resolution of approximately 10 μm and a motion-phase resolution of 2π/20 (Supplementary Movie 1). It took 163.8 s to acquire the full 3D data set (512 by 256 by 400 pixels in XYZ and 20 motion phases).
Fig. 3.
(Color online) (a) Schematic of setup with a latex drum. (b) Vibrography snapshot images (0 and π motion phases) of the vibrating drum at the fundamental frequency of 800 Hz. (c) A cutaway image showing the motion in a cross-section of the membrane.
As the sound frequency was increased, we observed higher-order vibration modes. Figure 4a shows snapshots of the motion at 1.78 kHz (NA=9), when the resonant excitation of the second-order radial mode, the (0,2) mode, became evident. The ratio of the resonance frequencies between the first and second radial modes was close to 2.3, as expected from acoustic theory [15]. Figure 4b shows a 3D contour plot, which highlights the characteristic amplitude pattern of the second-order radial mode. A movie created from the snapshots show the profile of the vibration mode with a motion-phase resolution of 2π/9 (Supplementary Movie 2). The data acquisition time was 73.6 s.
Fig. 4.
(Color online) (a) Snapshot vibrographs of the drumhead (0 and π motion phases) at 1.78 kHz. At this frequency, the second-order mode was predominantly produced. Scale bar, 1 mm. (b) A 3D contour graph.
We have described a synchronized OCT system capable of visualizing the acoustic vibrations. The maximum frequency was approximately 3 kHz, only limited by the finite A-line rate (20 kHz) of the current system. With a higher A-line rate, the range could be extended by an order of magnitude. The lower limit of the frequency range is primarily determined by considering the overall data acquisition time, as the acquisition time is proportional to the inverse of the frequency times the number of spatial points in XY. For frequencies below 100 Hz, the data acquisition and scanning sequence shown in Fig. 1b may no longer be optimal, and a modified sequence could reduce the data acquisition time (e.g. NA=8 is sufficient to obtain a sub-nm sensitivity at SNR of >30 dB; see Fig. 2b). Time synchronization and mechanical noise reduction enabled us to obtain the sub-nanometer displacement sensitivity. We expect that OCT vibrography can be used to visualize the vibrational modes of optomechanical devices for fundamental and applied studies. OCT vibrography may allow a new approach in dynamic rheometry. Another area of application is evaluating the function of the tympanic membrane and ossicles in response to sound, as well as the inner ear [16].
Supplementary Material
Acknowledgments
We thank J. R. Rosowski and J. T. Cheng for discussion. This work was funded by the National Institutes of Health (P41EB015903, U54CA143837).
References
- 1.Kleckner D, Bouwmeester D. Sub-kelvin optical cooling of a micromechanical resonator. Nature. 2006;444:75–78. doi: 10.1038/nature05231. [DOI] [PubMed] [Google Scholar]
- 2.Kippenberg TJ, Vahala KJ. Cavity opto-mechanics. Opt Express. 2007;15:17172–17205. doi: 10.1364/oe.15.017172. [DOI] [PubMed] [Google Scholar]
- 3.Burg TP, Godin M, Knudsen SM, Shen W, Carlson G, Foster JS, Babcock K, Manalis SR. Weighing of biomolecules, single cells and single nanoparticles in fluid. Nature. 2007;446:1066–1069. doi: 10.1038/nature05741. [DOI] [PubMed] [Google Scholar]
- 4.Xu MH, Wang LHV. Photoacoustic imaging in biomedicine. Rev Sci Instrum. 2006;77 [Google Scholar]
- 5.Cheng JT, Aarnisalo AA, Harrington E, Hernandez-Montes MD, Furlong C, Merchant SN, Rosowski JJ. Motion of the surface of the human tympanic membrane measured with stroboscopic holography. Hearing Research. 2010;263:66–77. doi: 10.1016/j.heares.2009.12.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Liang X, Orescanin M, Toohey KS, Insana MF, Boppart SA. Acoustomotive optical coherence elastography for measuring material mechanical properties. Opt Lett. 2009;34:2894–2896. doi: 10.1364/OL.34.002894. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Kennedy BF, Liang X, Adie SG, Gerstmann DK, Quirk BC, Boppart SA, Sampson DD. In vivo three-dimensional optical coherence elastography. Optics Express. 2011;19:6623–6634. doi: 10.1364/OE.19.006623. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Chhetri RK, Kozek KA, Johnston-Peck AC, Tracy JB, Oldenburg AL. Imaging three-dimensional rotational diffusion of plasmon resonant gold nanorods using polarization-sensitive optical coherence tomography. Phys Rev E. 2011;83:040903. doi: 10.1103/PhysRevE.83.040903. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Wang RK, Nuttall AL. Phase-sensitive optical coherence tomography imaging of the tissue motion within the organ of Corti at a subnanometer scale: a preliminary study. Journal of Biomedical Optics. 2010;15:056005. doi: 10.1117/1.3486543. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Choma MA, Ellerbee AK, Yang CH, Creazzo TL, Izatt JA. Spectral-domain phase microscopy. Opt Lett. 2005;30:1162–1164. doi: 10.1364/ol.30.001162. [DOI] [PubMed] [Google Scholar]
- 11.Joo C, Akkin T, Cense B, Park BH, de Boer JE. Spectral-domain optical coherence phase microscopy for quantitative phase-contrast imaging. Opt Lett. 2005;30:2131–2133. doi: 10.1364/ol.30.002131. [DOI] [PubMed] [Google Scholar]
- 12.Yun SH, Tearney GJ, de Boer JF, Iftimia N, Bouma BE. High-speed optical frequency-domain imaging. Optics Express. 2003;11:2953–2963. doi: 10.1364/oe.11.002953. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Chang EW, Kobler JB, Yun SH. Triggered optical coherence tomography for capturing rapid periodic motion. Scientific Reports. 2011;1:48. doi: 10.1038/srep00048. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Vakoc BJ, Yun SH, de Boer JF, Tearney GJ, Bouma BE. Phase-resolved optical frequency domain imaging. Opt Express. 2005;13:5483–5493. doi: 10.1364/opex.13.005483. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Christian RS, Davis RE, Tubis A, Anderson CA, Mills RI, Rossing TD. Effects of air loading on timpani membrane vibrations. Journal of the Acoustical Society of America. 1984;76:1336–1345. [Google Scholar]
- 16.Chen FY, Zha DJ, Fridberger A, Zheng JF, Choudhury N, Jacques SL, Wang RK, Shi XR, Nuttall AL. A differentially amplified motion in the ear for near-threshold sound detection. Nat Neurosci. 2011;14:770–774. doi: 10.1038/nn.2827. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.