. 2010 Oct 15;299(6):H1790–H1794. doi: 10.1152/ajpheart.00370.2010

A novel approach to in vivo mitral valve stress analysis

Chun Xu 1, Clay J Brinster 2, Arminder S Jassar 1,2, Mathieu Vergnat 1, Thomas J Eperjesi 1, Robert C Gorman 1,2, Joseph H Gorman 3rd 1,2, Benjamin M Jackson 2,
PMCID: PMC3774186  PMID: 20952665

Abstract

Three-dimensional (3-D) echocardiography allows the generation of anatomically correct and time-resolved geometric mitral valve (MV) models. However, as imaged in vivo, the MV assumes its systolic geometric configuration only when loaded. Customarily, finite element analysis (FEA) is used to predict material stress and strain fields rendered by applying a load on an initially unloaded model. Therefore, this study endeavors to provide a framework for the application of in vivo MV geometry and FEA to MV physiology, pathophysiology, and surgical repair. We hypothesize that in vivo MV geometry can be reasonably used as a surrogate for the unloaded valve in computational (FEA) simulations, yielding reasonable and meaningful stress and strain magnitudes and distributions. Three experiments were undertaken to demonstrate that the MV leaflets are relatively nondeformed during systolic loading: 1) leaflet strain in vivo was measured using sonomicrometry in an ovine model, 2) hybrid models of normal human MVs as constructed using transesophageal real-time 3-D echocardiography (rt-3DE) were repeatedly loaded using FEA, and 3) serial rt-3DE images of normal human MVs were used to construct models at end diastole and end isovolumic contraction to detect any deformation during isovolumic contraction. The average linear strain associated with isovolumic contraction was 0.02 ± 0.01, measured in vivo with sonomicrometry. Repeated loading of the hybrid normal human MV demonstrated little change in stress or geometry: peak von Mises stress changed by <4% at all locations on the anterior and posterior leaflets. Finally, the in vivo human MV deformed minimally during isovolumic contraction, as measured by the mean absolute difference calculated over the surfaces of both leaflets between serial MV models: 0.53 ± 0.19 mm. FEA modeling of MV models derived from in vivo high-resolution truly 3-D imaging is reasonable and useful for stress prediction in MV pathologies and repairs.

Keywords: finite element analysis, cardiac surgery, cardiac physiology


mitral valve (MV) disease is common and mortal. MV surgery— repair and replacement—as well as interventional procedures are commonly exercised treatment options. Finite element analysis (FEA) has been applied to the MV frequently over the last 15 years. Novel imaging techniques have allowed the generation of anatomically correct and time-resolved geometric models of the MV and its supporting apparatus. However, as imaged in vivo, the MV assumes its systolic geometric configuration only when loaded. Customarily, FEA is used to predict material stress and strain fields rendered by applying a load on an initially unloaded model. Therefore, this study endeavors to provide a framework for the application of in vivo MV geometry and FEA to human MV physiology, pathophysiology, and repair. We hypothesize that in vivo MV geometry can be reasonably used as a surrogate for the unloaded valve in computational (FEA) simulations, yielding reasonable and meaningful stress and strain magnitudes and distributions.

MATERIALS AND METHODS

MV leaflet strain by sonomicrometry.

In four normal sheep, sonomicrometry array localization was used to image a rectangular array of four transducers implanted on the anterior leaflet throughout the cardiac cycle. The details of surgical implantation on cardiopulmonary bypass and the methods of sonomicrometry array localization have been previously described (5). Linear and areal strains were calculated by dividing the rectangular array into two triangles, and the enclosed areas were computed at each time point throughout the cardiac cycle. Isovolumic contraction was taken from end diastole (ED, defined at the peak of QRS complex) to end isovolumic contraction (EIVC, defined as the first time point that the aortic blood pressure increases).

Repeated loading of hybrid human MV.

A hybrid model of the normal human MV at midsystole was constructed from 10 healthy adult subjects imaged using real-time three-dimensional (3-D) echocardiography (rt-3DE). These experiments have been previously described (17). The hybrid model was selected so as to represent the “average” or characteristic normal human MV. This anatomically correct hybrid MV geometric model was loaded repeatedly—with the deformed leaflet geometry following the first loading providing the initial unloaded geometry for the second loading step—to assess the strain and deformation expected under initial and subsequent pressure load applications. A total of 8,340 and 4,073 nodes and 16,292 and 7,683 triangular elements represented the anterior leaflet and posterior leaflet, respectively. The triangulated leaflets surfaces were imported into a commercial finite element program (Abaqus/Explicit 6.3, HKS, Pawtucket, RI) to quantify predicted regional stress distributions on the MV leaflets. Leaflet tissue was assumed to be orthotropic and linearly elastic, with Young's modulus determined from excised porcine MV tissue data (8). Thin shell elements (type S3R) were used. Specific anterior and posterior leaflet properties are indicated in Table 1. All freedom of motion except in-plane rotation was fixed for the annular nodes. The coaptation area between anterior and posterior leaflets was defined as an interface pair to prevent the nodes along coapting surfaces from passing through each other. A small degree of sliding was allowed between two leaflets. Chordae tendineae were represented by strings connecting the papillary muscle tips to the insertion points on the leaflets and modeled by a tension-only stress element (element type T3D2). The nodes of the chordal attachment points on the leaflets were fixed with all degrees of freedom except in-plane motion. Systolic loading was accomplished via the application of an 80-mmHg pressure gradient across the MV. Stress, strain, and displacement were recorded as output variables.

Table 1.

Mitral valve material properties used in FEA model

Parameter Anterior Leaflet Posterior Leaflet Primary Chordae Secondary Chordae
Thickness, mm 1.4 1.1
Cross-sectional area, mm2 0.4 0.7
Ecirc, Pa 6.20E +06 2.35E +06 2.20E +07 2.20E +07
Erad, Pa 2.10E +06 1.887E +6
Poisson's ratio 0.45 0.45 0.45 0.45
Density, kg/m3 1.04E +04 1.04E +04 1.04E +04 1.04E +04

FEA, finite element analysis; Ecirc, circumferential modulus of elasticity; Erad, radial modulus of elasticity.

Deformation in human MV under isovolumic in vivo loading.

Models of three normal human MV were constructed from rt-3DE, as described previously (17). The models generated at ED and EIVC were rotated and translated in Matlab (The Mathworks, Natick, MA) to minimize the sum of squares difference between leaflet surfaces in each of the n = 3 paired valves; the aligned models were then analyzed to assess deformation under an in vivo pressure load during isovolumic contraction.

Statistics.

All results are presented as means ± SD unless otherwise indicated. Geometric comparison of whole MV leaflet models was accomplished once two models were optimally aligned by calculating a mean absolute difference in z-coordinate over the surfaces of valve leaflets [i.e., a standard deviation between i = 1 to m MV models compared at the number of points over the surface of the anterior and posterior leaflets (nAL&PL)]:

σ=AL&PLi(z¯zi)2mnAL&PL

All human experiments were approved by the Institutional Review Board, and informed consent was obtained before echocardiography. For all ovine experiments, the animals were cared for in compliance with the Guide for the Care and Use of Laboratory Animals (National Institutes of Health, Revised 2010), and the experimental protocols were approved by the Institutional Animal Care and Use Committee.

RESULTS

MV leaflet strain by sonomicrometry.

There was minimal linear and areal strain through the loading associated with isovolumic contraction: the average linear strain was 0.02 ± 0.01, and average areal strain was 0.04 ± 0.02. Figure 1 demonstrates the linear strain through the cardiac cycle, along with the synchronized aortic blood pressure, in a single representative animal.

Fig. 1.

Fig. 1.

Linear strain according to anterior leaflet sonomicrometry array. Linear strain of a quadrangular array of sonomicrometry transducers on the anterior mitral leaflet of a normal ovine mitral valve (MV). Electrocardiogram and aortic blood pressure (ABP) traces are also displayed to allow identification of relevant time points in the cardiac cycle.

Repeated loading of hybrid human MV.

With pressure loading, the hybrid anatomically correct normal human MV experienced peak von Mises stress of 0.63 MPa. Subsequently, the deformed valve leaflets (with persistent boundary conditions) were reloaded (from a hypothetical stress-free state): peak von Mises stress generated was 0.65 MPa. In addition, stress distribution was qualitatively similar in the second, compared with the first, loading event, as demonstrated in Fig. 2. Stress magnitude with the second loading step at all locations on mitral leaflets was within 4% of initial loading.

Fig. 2.

Fig. 2.

Von Mises stress (A and B) and principal strain (C and D) on the MV predicted by finite element analysis during repeated loading. A and C: initial loading of the normal human hybrid MV model. B and D: subsequent reloading of the deformed valve, demonstrating little difference in the magnitude or distribution of pressure-derived mechanical stress or principal strain. Stress magnitude predicted in the second loading step was withing 4% of stress magnitude predicted in the initial loading step at all MV leaflet elements. Max, maximum; Min, minimum.

Similarly, as also demonstrated in Fig. 2, strain distribution and peak strain magnitude were similar with the second loading step. Finally, the predicted deformation was small with the repeat loading step: 1.2-mm peak deformation, confirming the relatively constant geometric conformation of the MV leaflets with reloading.

Deformation in human MV under isovolumic in vivo loading.

Figure 3 compares a single human MV geometric model generated at ED and EIVC, aligned post hoc to allow for translational and rotational displacement during isovolumic contraction; the other n = 2 cases demonstrated similar invariance in valve geometry during isovolumic contraction. The mean difference between corresponding points on the leaflet surfaces was just 0.53 ± 0.19 mm. The supplemental video (posted with the online version of this article) details the minimal deformation in the conformation of the MV leaflets in the course of isovolumic contraction.

Fig. 3.

Fig. 3.

Three-dimensional model of MV at end diastole and end isovolumic contraction. Normal human MV model in 2 views generated from real-time 3-dimensional echocardiography by transesophageal echocardiography at end diastole (white spheres) and end isovolumic contraction (shaded dark gray) demonstrated minimal deformation during the loading of isovolumic contraction.

DISCUSSION

One might ask what is the utility of current advanced and complicated FEA models of the MV. Some have used MV FEA to ask what might happen to the in vivo valve if a particular repair or intervention is performed. In general, the approach has been to use highly idealized and simplistic geometric models (though often rigorously complicated material properties and fluid-structure formulations) to analyze ex vivo or anatomically idealized MVs to assess physiology (8); to analyze standardized pathological valves to predict stress distribution and, potentially, the integrity and failure behavior of repair techniques; to “evaluate proposed surgical repairs” using idealized computational models (1, 3, 9, 11, 15, 16, 18, 19), including models that attempt integration of fluid-structure interactions (4, 13); or to evaluate pathological alterations on the function of idealized computational models (2, 10, 12). Alternatively, realistic geometric models of the MV have been rarely used, for instance, to assess the relative contributions of primary and secondary chordae to structural integrity (14). Finally and more conventionally, FEA has been used to test and predict failure of devices for MV repair or replacement (21). Table 2 summarizes these varied approaches and summarizes their respective methods and scientific utility.

Table 2.

Applications of FEA models of mitral valve

Approach Utility and Features References Using Approach
Analysis of in vivo valves ∙ Applicable clinically to individual human patients 14, 20
∙ Potentially useful for prediction of patient outcome and guiding therapy
Prediction of repair integrity and longevity ∙ Potentially useful for design of repair techniques 1, 3, 4, 9, 11, 13, 15, 16, 18, 19
∙ Not generally clinically useful in an individual patient
∙ Standardized anatomy; not individualized image-determined model
Testing of device performance ∙ Conventional industrial use of FEA 21
∙ For failure prediction
∙ For design refinement
∙ Useful for prosthetic and bioprosthetic valves and for annuloplasty rings
Physiology of normal or pathological valve ∙ Theoretical physiological study 2, 10, 12
∙ Analytic (idealized) geometric model or ex vivo anatomic model
Other ∙ Inverse determination of material properties from in vivo invasive imaging 6, 7

These attempts at MV modeling are almost uniformly based on idealized or standardized geometries. In contrast, and more recently, Votta et al. (20) have concerned themselves with complicated, anatomically correct analyses of MV stress prediction but have not addressed the question of the rational basis for supposing that FEA of an already-loaded geometric representation of the MV apparatus will yield realistic predictive values of local leaflet stresses and strains. In fact, though that group from Politecnico di Milano imaged the leaflets, annulus, chordae, and papillary muscles through systole and though they modeled the annulus dynamically in time from rt-3DE images (and incorporated a dynamic annular geometry in their model of progressive loading), they made no attempt to compare their loaded ED MV leaflet geometry with the imaged MV leaflet geometry at EIVC.

In the current work, the first set of experiments, MV leaflet strain by sonomicrometry, demonstrates that leaflet strain is minimal during isovolumic contraction and hence during pressure loading of the left ventricle. Hence, we expect that the deformation of the valve and its leaflets is small during loading, so that MV models generated by noninvasive imaging (e.g., cardiac MRI, cardiac computed tomography, or rt-3DE) of the pressure-loaded valve may from a geometric standpoint be appropriately used in FEA work as unloaded valve geometries.

The second set of experiments, Repeated loading of hybrid human MV, uses a rigorously defined (high-resolution 3-D imaging-derived model) MV geometry and FEA. The midsystolic valve is repeatedly loaded, with the deformed leaflet geometry of the first loading providing the initial (unloaded or stress-free) geometry for the second loading. Using the loaded geometry of the MV as the input for a physiological (left ventricular pressure load) FEA (second loading event) provides very similar (within 4% at all locations over both leaflet surfaces) stress results as does loading the initial MV model (first loading event). Hence, reloading (in FEA) a MV model generated from imaging a physiologically (in vivo) loaded valve is not unreasonable and is expected to provide meaningful stress and strain measures.

In the final experiments, Deformation in human MV under isovolumic in vivo loading, there was minimal deformation in the valve geometry during pressure loading. Because left ventricular volume is constant during isovolumic contraction, the structures (the base of the ventricle, the papillary muscles, and the fibrous trigone of the heart) constraining the valve are relatively static. Therefore, deformations in leaflet geometry during this portion of the cardiac cycle are most likely due to pressure loading. That minimal deformation confirms the hypothesis that the loaded in vivo MV geometry can be reasonably used as a surrogate for the unloaded valve in computational (FEA) simulations.

Others' measurement of MV leaflet material properties can be compared with the results herein presented. In results similar to our demonstration, using sonomicrometry, that mitral leaflet surface strain during loading is minimal relative to the large-scale anatomic and geometric changes in surface geometry (shape), Krishnamurthy et al. (6) found unexpectedly high coefficients of elasticity of the ovine anterior MV leaflet in a study using biplane marker angiography and FEA solutions. With their value of 11 MPa for the radial modulus of elasticity, a peak systolic pressure of 80 mmHg, and a mitral leaflet with a local radius of curvature of 1 cm and a thickness of 1 mm, one would expect only a maximal strain of 0.01; this value is very close to our measured leaflet strain. (The circumferential modulus of elasticity measured in the Krishnamurthy work was greater, 43 MPa, but is of the same order of magnitude.)

Drawbacks of the current study include a less-than-comprehensive MV model: 1) whereas the leaflet surface profile is accurately determined by rt-3DE, chordae tendinae and papillary muscles were not reliably imaged and so their incorporation in the model is, at best, heuristically motivated; and 2) the material properties model used is far from as comprehensive or, presumably, realistic as others previously published. The material properties—the linear orthotropic leaflet Young's moduli were used—are simplified; however, the small strains measured in both the sonomicrometry and FEA repeated loading experiments would seem to justify using linear material properties. In fact, Krishnamurthy et al. (7) found that the anterior leaflet of the MV demonstrates a fairly linear stress-strain relationship during isovolumic relaxation. Furthermore, these potential drawbacks are irrelevant to the stated aims and robust conclusions of the study: the goal was not to create a religiously accurate MV FEA model, but rather to justify the FEA approach for analysis and stress prediction of in vivo, noninvasively imaged MV models. (This situation is most often of clinical relevance, in humans who cannot or will not undergo invasive imaging, experimental valve loading experiments, leaflet strain measurements, or ex vivo valvular analyses.) As such, the current experiments justify the application of any MV FEA model, in particular the elegant and more fully developed and validated models of Kunzelman et al. (13) or Prot et al. (14).

Therefore, given the results of the current experiments, a rational approach to in vivo MV stress analysis would incorporate realistic, anatomically resolved, empiric material properties of leaflets and chordae, in vivo high-resolution truly 3-D imaging for geometric model determination, and FEA modeling. In particular, future experiments, incorporating more realistic material properties, will hopefully assess the relative contributions of material properties, pressure load, and geometry in the in vivo human MV.

GRANTS

This research project was supported by National Heart, Lung, Blood Institute Grants HL-63954 and HL-73021. R. C. Gorman and J. H. Gorman are supported by individual established investigator awards from the American Heart Association (Dallas, TX).

DISCLOSURES

No conflicts of interest, financial or otherwise, are declared by the author(s).

Supplementary Material

Supplemental Video

REFERENCES

  • 1. Avanzani A. A computational procedure for prediction of structural effects of edge-to-edge repair on mitral valve. J Biomech Eng 130: 031015, 2008. [DOI] [PubMed] [Google Scholar]
  • 2. Cochran RP, Kunzelman KS. Effect of papillary muscle position on mitral valve function: relationship to homografts. Ann Thorac Surg 66: S155–S161, 1998. [DOI] [PubMed] [Google Scholar]
  • 3. Dal Pan F, Donzella G, Fucci C, Schreiber M. Structural effects of an innovative surgical technique to repair heart valve defects. J Biomech 38: 2460–2471, 2004. [DOI] [PubMed] [Google Scholar]
  • 4. Einstein DR, Kunzelman KS, Reinhall PG, Nicosia MA, Cochran RP. Non-linear fluid-coupled computational model of the mitral valve. J Heart Valve Dis 14: 376–385, 2005. [PubMed] [Google Scholar]
  • 5. Gorman JH, 3rd, Gupta KB, Streicher JT, Gorman RC, Jackson BM, Ratcliffe MB, Bogen DK, Edmunds LH., Jr Dynamic three-dimensional imaging of the mitral valve and left ventricle by rapid sonomicrometry array localization. J Thorac Cardiovasc Surg 112: 712–726, 1996. [DOI] [PubMed] [Google Scholar]
  • 6. Krishnamurthy G, Ennis DB, Itoh A, Bothe W, Swanson JC, Karlsson M, Kuhl E, Miller DC, Ingels NB., Jr Material properties of the ovine mitral valve anterior leaflet in vivo from inverse finite element analysis. Am J Physiol Heart Circ Physiol 295: H1141–H1149, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Krishnamurthy G, Itoh A, Bothe W, Swanson JC, Kuhl E, Karlsson M, Craig Miller D, Ingels NB., Jr Stress-strain behavior of mitral valve leaflets in the beating ovine heart. J Biomech 42: 1909–1916, 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Kunzelman KS, Cochran RP, Chuong C, Ring WS, Verrier ED, Eberhart RD. Finite element analysis of the mitral valve. J Heart Valve Dis 2: 326–340, 1993. [PubMed] [Google Scholar]
  • 9. Kunzelman KS, Reimink MS, Cochran RP. Variations in annuloplasty ring and sizer dimensions may alter outcome in mitral valve repair. J Card Surg 12: 322–329, 1997. [DOI] [PubMed] [Google Scholar]
  • 10. Kunzelman KS, Reimink MS, Cochran RP. Annular dilatation increases stress in the mitral valve and delays coaptation: a finite element computer model. Cardiovasc Surg 5: 427–434, 1997. [DOI] [PubMed] [Google Scholar]
  • 11. Kunzelman KS, Reimink MS, Cochran RP. Flexible versus rigid ring annuloplasty for mitral valve annular dilatation: a finite element model. J Heart Valve Dis 7: 108–116, 1998. [PubMed] [Google Scholar]
  • 12. Kunzelman KS, Quick DW, Cochran RP. Altered collagen concentration in mitral valve leaflets: biochemical and finite element analysis. Ann Thorac Surg 66: S198–S205, 1998. [DOI] [PubMed] [Google Scholar]
  • 13. Kunzelman KS, Einstein DR, Cochran RP. Fluid-structure interaction models of the mitral valve: function in normal and pathological states. Philos Trans R Soc Lond B Biol Sci 362: 1393–1406, 2007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Prot V, Haaverstad R, Skallerud B. Finite element analysis of the mitral apparatus: annulus shape effect and chordal force distribution. Biomech Model Mechanobiol 8: 43–55, 2009. [DOI] [PubMed] [Google Scholar]
  • 15. Reimink MS, Kunzelman KS, Cochran RP. The effect of chordal replacement suture length on function and stresses in repaired mitral valves: a finite element study. J Heart Valve Dis 5: 365–375, 1996. [PubMed] [Google Scholar]
  • 16. Reimink MS, Kunzelman KS, Verrier ED, Cochran RP. The effect of anterior chordal replacement on mitral valve function and stresses. A finite element study. ASAIO J 41: M754–M762, 1995. [DOI] [PubMed] [Google Scholar]
  • 17. Ryan LP, Jackson BM, Eperjesi TJ, St Plappert TJ, John-Sutton M, Gorman RC, Gorman JH., 3rd A methodology for assessing human mitral leaflet curvature using real-time 3-dimensional echocardiography. J Thorac Cardiovasc Surg 136: 726–734, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Schievano S, Kunzelman KS, Nicosia MA, Cochran RP, Einstein DR, Khambadkone S, Bonhoeffer P. Percutaneous mitral valve dilatation: single balloon versus double balloon. A finite element study. J Heart Valve Dis 18: 28–34, 2009. [PubMed] [Google Scholar]
  • 19. Votta E, Maisano F, Bolling SF, Alfieri O, Montevecchi FM, Redaelli A. The Geoform disease-specific annuloplasty system: a finite element study. Ann Thorac Surg 84: 92–101, 2007. [DOI] [PubMed] [Google Scholar]
  • 20. Votta E, Caiani E, Veronesi F, Soncini M, Montevecchi FM, Redaelli A. Mitral valve finite-element modeling from ultrasound data: a pilot study for a new approach to understand mitral function and clinical scenarios. Philos Transact A Math Phys Eng Sci 366: 3411–3434, 2008. [DOI] [PubMed] [Google Scholar]
  • 21. Wieting DW, Eberhardt AC, Reul H, Breznock EM, Schreck SG, Chandler JG. Strut fracture mechanism of the Bjork-Shiley convexo-concave heart valve. J Heart Valve Dis 8: 206–217, 1999. [PubMed] [Google Scholar]