The predicted prevalence of false positive blood cultures due to hyperactive neonatal blood cells in a radiometric detection system was confirmed. Suppression of this blood background radioactivity in the system was achieved by using a hypertonic medium containing 10% sucrose. The radiometric system produced accurate results as fast as the conventional blood culturing method, saved labor and minimized the recovery of extraneous contaminants.
Full text
PDFSelected References
These references are in PubMed. This may not be the complete list of references from this article.
- Clymo A. B. Distribution of coagulase-negative staphylococci from newborns. J Clin Pathol. 1971 Nov;24(8):770–770. doi: 10.1136/jcp.24.8.770-b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cocchi P., Mori S., Becattini A. N.B.T. tests in premature infants. Lancet. 1969 Dec 27;2(7635):1426–1427. doi: 10.1016/s0140-6736(69)90970-2. [DOI] [PubMed] [Google Scholar]
- Coen R., Grush O., Kauder E. Studies of bactericidal activity and metabolism of the leukocyte in full-term neonates. J Pediatr. 1969 Sep;75(3):400–406. doi: 10.1016/s0022-3476(69)80265-9. [DOI] [PubMed] [Google Scholar]
- DeBlanc H. J., Jr, DeLand F., Wagner H. N., Jr Automated radiometric detection of bacteria in 2,967 blood cultures. Appl Microbiol. 1971 Nov;22(5):846–849. doi: 10.1128/am.22.5.846-849.1971. [DOI] [PMC free article] [PubMed] [Google Scholar]
- DeLand F., Wagner H. N., Jr Automated radiometric detection of bacterial growth in blood cultures. J Lab Clin Med. 1970 Mar;75(3):529–534. [PubMed] [Google Scholar]
- Humbert J. R., Kurtz M. L., Hathaway W. E. Increased reduction of nitrobule tetrazolium by neutrophils of newborn infants. Pediatrics. 1970 Jan;45(1):125–128. [PubMed] [Google Scholar]
- Park B. H., Holmes B., Good R. A. Metabolic activities in leukocytes of newborn infants. J Pediatr. 1970 Feb;76(2):237–241. doi: 10.1016/s0022-3476(70)80168-8. [DOI] [PubMed] [Google Scholar]
- Rosner R. Comparison of a blood culture system containing liquoid and sucrose with systems containing either reagent alone. Appl Microbiol. 1970 Feb;19(2):281–282. doi: 10.1128/am.19.2.281-282.1970. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Waters J. R. Sensitivity of the 14 CO 2 radiometric method for bacterial detection. Appl Microbiol. 1972 Jan;23(1):198–199. doi: 10.1128/am.23.1.198-199.1972. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zwarun A. A. Effect of osmotic stabilizers on 14 CO 2 production by bacteria and blood. Appl Microbiol. 1973 Apr;25(4):589–591. doi: 10.1128/am.25.4.589-591.1973. [DOI] [PMC free article] [PubMed] [Google Scholar]