. 1973 Feb;25(2):284–286. doi: 10.1128/am.25.2.284-286.1973

Auxotab—a Device for Identifying Enteric Bacteria

D L Rhoden 1, K M Tomfohrde 1, P B Smith 1, A Balows 1
PMCID: PMC380788  PMID: 4571660

Abstract

A multitest system called the Auxotab that uses ten dehydrated reagents on a paper card has been evaluated with 417 known stock cultures of Enterobacteriaceae. In double-blind studies with the Auxotab, 87% of the strains tested were correctly identified. Results of this study indicate that there is a need for modification of the product in regard to ease of handling, time required for use, and accuracy of identification of enteric bacteria.