Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2014 Dec 1.
Published in final edited form as: Neuroimage. 2013 Jun 22;83:10.1016/j.neuroimage.2013.06.047. doi: 10.1016/j.neuroimage.2013.06.047

Direct Visualization of Short Transverse Relaxation Time Component (ViSTa)

Se-Hong Oh 1, Michel Bilello 1, Matthew Schindler 2, Clyde E Markowitz 2, John A Detre 1,2, Jongho Lee 1
PMCID: PMC3815972  NIHMSID: NIHMS511125  PMID: 23796545

Abstract

White matter of the brain has been demonstrated to have multiple relaxation components. Among them, the short transverse relaxation time component (T2 < 40 ms; T2* < 25 ms at 3T) has been suggested to originate from myelin water whereas long transverse relaxation time components have been associated with axonal and/or interstitial water. In myelin water imaging, T2 or T2* signal decay is measured to estimate myelin water fraction based on T2 or T2* differences among the water components. This method has been demonstrated to be sensitive to demyelination in the brain but suffers from low SNR and image artifacts originating from ill-conditioned multi-exponential fitting. In this study, a novel approach that selectively acquires short transverse relaxation time signal is proposed. The method utilizes a double inversion RF pair to suppress a range of long T1 signal. This suppression leaves short T2* signal, which has been suggested to have short T1, as the primary source of the image. The experimental results confirms that after suppression of long T1 signals, the image is dominated by short T2* in the range of myelin water, allowing us to directly visualize the short transverse relaxation time component in the brain. Compared to conventional myelin water imaging, this new method of direct visualization of short relaxation time component (ViSTa) provides high quality images. When applied to multiple sclerosis patients, chronic lesions show significantly reduced signal intensity in ViSTa images suggesting sensitivity to demyelination.

Keywords: background suppression, multiple water components in white matter, myelin water imaging, T1 of myelin water, T1 filter

Introduction

Myelin sheaths derived from oligodendrocytes provide electrical insulation surrounding axons to expedite signal conduction in the healthy brain. In certain inflammatory neurological disorders such as multiple sclerosis (MS), the myelin sheaths degenerate and cause disabilities. Failure of normal myelination also occurs in certain developmental disorders known as leukodystrophies. Restoration of oligodendrocytes and axonal myelination is a therapeutic target for both MS and leukodystriphies (Keough and Yong, 2013). Accordingly, the development of myelin or myelin-related biomarkers has important clinical implications for the diagnosis and prognosis of such diseases.

Several MRI methods including myelin water imaging (Mackay et al., 1994), magnetization transfer imaging (Balaban and Ceckler, 1992; Gass et al., 1994), and diffusion tensor imaging (Basser et al., 1994; Werring et al., 1999) have been proposed as potential biomarkers for myelin. Among them, myelin water imaging (MWI) measures signals from water molecules between neighboring myelin layers. It has been suggested that this fraction of water has shorter transverse relaxation time constants (T2 and T2*) than other water inside of axons or in the extracellular space outside of fibers (Du et al., 2007b; Mackay et al., 1994; Menon and Allen, 1991; van Gelderen et al., 2012; Vasilescu et al., 1978). In a few studies, not only transverse relaxation but also longitudinal relaxation time constant (T1) has been suggested to vary among the components (Deoni et al., 2008; Does and Gore, 2002; Helms and Hagberg, 2009; Koenig et al., 1990; Labadie et al., 2013; Lancaster et al., 2002; Stanisz et al., 2005). Similarly to the transverse relaxation time constants, the shortest T1 component was assumed to have originated from myelin water.

In conventional MWI, the transverse relaxation is measured by multi-echo gradient echo (GRE) or spin echo (SE) sequences. Since the multiple water components have different relaxation times, the measured signal shows a multi-exponential decay. This decay is fitted with exponential decay basis functions to estimate the distribution of transverse relaxation time constants (Whittall and MacKay, 1989). Then the fraction of the short transverse relaxation time component or myelin water, which is approximately in the range of T2 < 40 ms and T2* < 25 ms at 3 T (Hwang et al., 2010; Kolind et al., 2009; Oh et al., 2006), is calculated relative to the total water to generate a myelin water fraction (MWF) map. However, the fitting process is ill-conditioned and the resulting MWF map is sensitive to noise and artifacts (Cover, 2008; Reiter et al., 2009; Whittall and MacKay, 1989).

Here we propose a novel approach of selectively acquiring the short transverse relaxation time component in the brain. This method suppresses long T1 signals such that the short T1 signal dominates the image. We demonstrate that after long T1 suppression, the remaining signals have T2* in the range of myelin water. The method is applied to MS patients and reveals that chronic lesions with demyelination have significant signal reduction compared to surrounding white matter. This new approach is referred to as direct Visualization of Short Transverse relaxation time component (ViSTa).

Methods

Double inversion RF pulses for long T1 water suppression

An inversion recovery sequence using a single inversion RF pulse is commonly used to suppress a small range of T1 signals. On the other hand, a pair of inversion RF pulse can be designed to suppress a wide range of T1 (Duijn et al., 1992; Duyn et al., 2001). In this double inversion sequence, the transverse magnetization at an echo time (TE) after a 90° excitation can be written as follows:

Mxy=M0(1(2(2eTD/T1)eTI1/T1)eTI2/T1)eTE/T2 (Eq. 1)

where the three intervals (TI1, TI2, and TD) are the time between the RF pulses as shown in Figure 1A. These intervals are adjusted to suppress the signal from the desired T1 components.

Figure 1.

Figure 1

(A) Double inversion RF pulse sequence for direct visualization of short transverse relaxation time component imaging (ViSTa). The timing (TI1 = 560 ms, TI2 = 220 ms, and TD = 380 ms) was optimized to have maximum suppression in water signal in the range of 750 ≤ T1 ≤ 5000 ms. (B) Transverse magnetization as a function of T1 in the ViSTa sequence. The double inversion pulse enabled signal suppression over a wide range of long T1.

In our method, the three intervals were chosen to have a large signal in short T1 and a small signal in long T1 such that the measured signal was dominated by the short T1 component. Specifically, we optimized for the parameters that result in larger than 40% of fully relaxed magnetization (M0) for T1 ≤ 200 ms (for a short T1 component in white matter), 0.3 % or less for 750 ≤ T1 ≤ 2000 ms (for long T1 components in white matter and gray matter), and 0.5% or less for 2000 ms ≤ T1 ≤ 5000 ms (for CSF). The strict suppression of long T1 components was, in part, because the T1 of myelin water at 3 T is not well characterized (see Discussion). As a result, the sequence parameters were chosen to have minimal contribution from long T1 signals while keeping a relatively large signal from the short T1 component. To find the parameters that satisfy this condition, a parameter search was performed on TI1, TI2, and TD over the range of 10 ms to 1000 ms in the step size of 10 ms. The optimal inversion timing that satisfied the condition was TI1 = 560 ms, TI2 = 220 ms, and TD = 380 ms. The resulting transverse magnetization is shown in Figure 1B. These parameters suppress signals not only from long T1 components in white matter (750 to 1000 ms at 3 T), but also from gray matter (1300 to 1800 ms at 3 T) and CSF (~ 4000 ms at 3 T). This suppression helps to avoid bright regions and artifacts in the image. Similarly, fat can be suppressed using a fat saturation pulse before the RF excitation at the cost of slight signal reduction from the magnetization transfer (MT) effect. Thus, all static signals except the short T1 signal were suppressed. When these parameters are used, the signal from the short T1 component may also be attenuated if T1 is longer than 100 ms. Hence, the method may show sensitivity to both the fraction and T1 of the short T1 component.

To demonstrate that the suppression of long T1 components resulted in short T2* signal in the range of myelin water, the double inversion sequence was acquired with multi-echo readout and the resulting signal decay curves were compared with those from conventional GRE. A MWF map was generated from the GRE data and compared to a ViSTa image.

All healthy volunteers and patients participated in the study provided written consent approved by an institutional review board.

MRI scans in healthy volunteers

Data were collected from 5 healthy subjects (mean age = 32 ± 3 years). A 3 T MRI scanner (Trio, Siemens, Erlangen, Germany) was used with a 32 channel head array receiver. The scan started with a localizer, followed by a region-of-interest shimming over the brain area. Three acquisitions were performed as follows:

  • Scan 1) A double inversion prepared multi-echo GRE sequence (ViSTa sequence): single slice, FOV = 220 mm2, resolution = 1.38 × 1.38 mm2, slice thickness = 3 mm, TR = 1160 ms, TI1 = 560 ms, TI2 = 220 ms, TD = 380 ms, TE = 1.95 ms to 84.1 ms (echo spacing = 2.65 ms), number of echoes = 32, flip angle = 90° (duration = 1 ms; sinc), matrix size = 160 × 160, bandwidth = 710 Hz/pixel, and scan time = 3.1 min. A non-selective hyperbolic secant pulse (duration = 10.24 ms and bandwidth = 1 kHz) was used for the inversion pulses. The sequence was repeated five times (15.5 min) and the images were averaged to improve SNR. No fat saturation was applied. In three subjects, additional data were acquired with fat saturation (no average) to estimate the signal reduction from the MT effect.

  • Scan 2) A multi-echo GRE sequence for GRE-MWI: An RF-spoiled GRE sequence with no double inversion RF pulses with the same scan parameters as Scan 1 except TR (= 97 ms) and flip angle (= 28°). The data was collected 12 times and averaged. The total scan time was 3.1 min.

  • Scan 3) A ViSTa sequence using a single echo with a low readout bandwidth: the same sequence structure and parameters as in Scan 1 except number of echoes = 1, TE = 6.36 ms, and bandwidth = 100 Hz/pixel. The total scan time was 3.1 min (no average). A fat saturation pulse was applied to remove bright fat signal.

Data from Scan 1 and Scan 2 were used to compare signal characteristics of ViSTa and GRE. Data from Scan 2 and Scan 3 were used to compare the spatial distributions of short T2* and image quality between ViSTa and GRE.

Data analysis

To characterize the multi-component signal decays in the ViSTa (Scan 1) and GRE data (Scan 2), ROI analysis was performed at five manually drawn ROIs: genu, splenium, major forceps, minor forceps and internal capsule. In each ROI, a mean decay curve was generated by averaging signals within the ROI. The distribution of the multiple components was estimated by fitting the mean decay curve with multiple exponential T2* decays (Whittall and MacKay, 1989), given by:

yi=k=1Mskexp(tiT2k),i=1,2,,N (Eq. 2)

where ti is an echo time, yi is the measured decay at ti, N is the total number of echoes, and sk and T2k* are the relative amplitude and T2* for each component respectively. M (= 120) is the number of T2* decay components in the model. T2k* is spaced logarithmically from 2.77 ms to 300 ms (Lenz et al., 2011). To estimate the relative amplitude of each component, a nonnegative least-square (NNLS) fitting method, widely used in MWI, was applied (Whittall and MacKay, 1989). Regularization, which is to find sk that minimizes χ2+μk=1Msk2 where μ is positive and χ2 is a residual sum of squares with the constraint of 1.02χmin2χ21.025χmin2, was tested but was not applied in the ROI analysis because SNR was very high after ROI averaging (ROI-averaged first echo SNR was over 420) and the estimated myelin water fraction was very similar (7.2% vs. 7.0% in GRE and 95.4 % vs. 95.1% in ViSTa for non-regularized vs. regularization). The regularization was applied to generate a MWF map in GRE data where the NNLS fitting was performed in individual voxels.

In the ViSTa images, the signal level was low and, therefore, non-zero mean noise distribution of the magnitude images resulted in a constant offset in the decay curves (Henkelman, 1985). This offset biases the T2* estimation and, therefore, was subtracted from the data before the fitting process (Gambarota et al., 2001; Ghugre et al., 2005). The offset was estimated by averaging a background noise area outside of the head.

The NNLS fitting resulted in a distribution of T2* components. T2* components between 3 ms and 25 ms were regarded as myelin water (Hwang et al., 2010). The MWF was calculated as a ratio between sum of the myelin water components and sum of all components.

In the ViSTa data, the fraction of the short T2* signal relative to total water signal was estimated in each ROI. The signal intensity of the short T2* component at TE = 0 ms was extrapolated using the short T2* components (i.e. 3 ms < T2* < 25 ms) estimated in the NNLS fitting results in the ViSTa data. The total water signal was estimated from the GRE data (Scan 2) by extrapolating the signal intensity at TE = 0 ms using all components in the NNLS fitting results. Since the measured GRE signal is total water signal multiplied by (1 - e-TR/T1)·sinθ/(1 - e-TR/T1·cosθ), the estimated signal intensity at TE = 0 ms was divided by 0.25 assuming T1 to be 800 ms. The ratio of short T2* signal to total water signal at TE = 0 was then calculated and compared with the MWF of GRE-MWI. The same receiver gain was used for both ViSTa and GRE scans for this quantification.

In addition to the ROI analysis, the GRE data were processed using the regularized NNLS method in voxel-by-voxel basis to generate a MWF map. In the GRE-MWF map, the mean MWF in each ROI was calculated and compared with the MWF calculated in the ROI analysis.

The GRE-MWF map was compared with both a single-echo ViSTa image from Scan 3, and a normalized ViSTa image. The normalization of ViSTa was performed as follows: First, the single-echo ViSTa image was divided by the first echo GRE image. The resulting image was multiplied by a magnitude-thresholded mask generated from the GRE image to avoid noise amplification. Then, a scaling factor was multiplied to compensate for T1- and T2*-weighting in GRE and ViSTa. This scaling factor was calculated using nominal T1 and T2* values. In GRE, nominal T1 and T2* were assumed to be 800 ms and 40 ms respectively. In ViSTa, nominal T1 and T2* were 118 ms and 10 ms respectively and the signal was assumed to originate fully from the short T2* component. The resulting scaling factor was (GRE T1-weight × GRE T2*-weight) / (ViSTa T1-weight × ViSTa T2*-weight) = (0.25 × 0.95) / (0.69 × 0.53) = 0.65. Since the normalization was performed using nominal values, the resulting image has limited accuracy in quantification (see Discussion). In this study, the primary purpose of the normalization was to remove coil sensitivity from the ViSTa images so that the spatial distribution of short T2* signal in the brain can be compared between ViSTa and GRE-MWF.

MRI scans in multiple sclerosis patients

Images were acquired from clinically diagnosed MS patients to demonstrate signal characteristics of ViSTa in MS lesions. Four patients were scanned on a clinical 3T MRI scanner (Trio, Siemens) that was different from the scanner used for the healthy volunteers. A 12 channel phased-array head coil (Matrix, Siemens) was used for data acquisition. The ViSTa sequence was acquired in addition to a routine clinical MS protocol. The sequence was acquired after magnetization prepared rapid acquisition gradient echo (MPRAGE; sagittal orientation), flow-attenuated inversion recovery (FLAIR; axial orientation) and proton density-weighted (PDW; axial orientation) scans, prior to contrast administration. The resolutions were 1 × 1 × 1 mm3 for MPRAGE, 0.86 × 0.86 × 3 mm3 for FLAIR, and 0.34 × 0.34 × 3 mm3 for PDW. A single slice that clearly contained lesions in FLAIR and PDW images was chosen as the imaging slice for the ViSTa scan. A post-contrast MPRAGE data was acquired following contrast agent injection (Gd-DTPA). ViSTa scan was acquired with slightly different scan parameters in each patient. In Patient 4 (shown in Figure 5), the parameters were as follows: resolution = 1.38 × 1.38 mm2, number of echoes = 1 echo, TE = 6.12 ms, flip angle for excitation = 90°, slice thickness = 3 mm, and scan time = 3.1 min. The same double inversion RF timing used in the healthy volunteer scans was applied. Data were acquired three times to improve SNR. A GRE scan of matching parameters with 32 echoes was also acquired but data showed in poor quality MWI, most likely due to poor shim.

Figure 5.

Figure 5

FLAIR (A), PDW (B), post-contrast MPRAGE (C), and single-echo ViSTa (D) images from an MS patient. Chronic lesions (arrows) show significant signal reduction in the ViSTa image.

Results

The images acquired with GRE and ViSTa are shown in Figure 2. The display scale is different between the two scans and the maximum range of the first echo image was twice that of the other echo images. The GRE images show significant signal drop in the frontal lobe area at the last echo due to field inhomogeneity (Fig. 2C). The first echo ViSTa image reveals that white matter has high signal intensity whereas gray matter areas show substantially reduced signal (Fig. 2D). CSF also shows substantial signal reduction but has residual signal potentially originating from CSF flow (Stadlbauer et al., 2010). Arteries and fat are also highlighted in the ViSTa images because of inflow of blood and short T1 of fat. In the last echo (Fig. 2F), white matter signal reaches the noise level.

Figure 2.

Figure 2

(A) – (C) The first, 16th and last echo images from GRE. (D) – (F) The same three echo images from ViSTa. Flow artifacts in the phase encoding direction (anterior-posterior) is observed in the ViSTa images.

The measured signal decays and T2* distributions in each ROI are shown in Figure 3 (genu – blue, splenium – red, major forceps – green, minor forceps – cyan, and internal capsule – magenta). Compared to the decay curves from the GRE scan (Fig. 3A), the ViSTa signals decay much more rapidly (Fig. 3C). This is also clearly demonstrated in the T2* distribution from the NNLS fitting (Fig. 3B vs. 3D) confirming that the primary ViSTa signal has shorter T2* than the GRE signals. The T2* distributions from the GRE data (Fig. 3B) show a small fraction of the short T2* component (7.2 ± 1.2 %; an average of all subjects in splenium, major forceps and internal capsule ROIs hereafter unless noted) similar to a recent study (Lenz et al., 2011). In contrast, the ViSTa results demonstrate that the primary signals are from the short T2* component (95.4 ± 4.1 %) in the range of myelin water (3 ms < T2* < 25 ms). These results confirm that the suppression of long T1 primarily leaves short T2* signal in the range of myelin water.

Figure 3.

Figure 3

Signal decay measurements in GRE (A) and ViSTa (C). NNLS multi-exponential fitted results in GRE (B) and ViSTa (D). The lines are color coded as follows: Genu – blue, splenium – red, major forceps – green, minor forceps – cyan, and internal capsule – magenta. The signals in ViSTa decay much rapidly compared to those in GRE. The resulting T2* distributions (D) suggest that the primary component in ViSTA has T2* less than 25 ms. On the other hand, the GRE results (C) show the signal dominance from a long T2* component (T2* > 25 ms).

Compared to other ROIs, the signals from genu and minor forceps decay faster in the GRE data (Figs. 3A). This rapid T2* decay is from the field inhomogeneity effects where intravoxel dephasing induces additional signal loss. As a result, the T2* distributions show relatively shorter T2* with no myelin water peak (Fig. 3B). This has been observed in the previous studies (Hwang and Du, 2009; Hwang et al., 2010). In a few subjects, the primary T2* peak became less than 25 ms miscategorizing the signal as myelin water. Since this decay does not reflect intrinsic T2* decay of white matter, the genu and minor forceps were excluded from the ROI averages (Table 1). In the ViSTa data, on the other hand, expedited decay is not clearly observed in the two ROIs, however, in a few subjects, the decay resulted in two short T2* peaks (both within 25 ms). This may reflect spatially varying field inhomogeneity effects across an ROI and/or inflow artifact from neighboring vessels. The signal fractions of these ROIs are similar to the other ROIs and listed in Table 1. However, they are excluded from the ROI averages for a comparison with GRE data.

Table 1.

List of the short T2* fractions at ROIs (mean ± standard deviation).

Fraction of short T2* component
GRE MWF (Scan 2) ViSTa (Scan1)
Signal averaged in ROI Pixel-by-pixel estimation Relative to ViSTa signal Relative to total water signal
Genu - - (100 ± 0.0%) 5.8 ± 0.4%
Splenium 8.0 ± 0.8 % 8.9 ± 1.8 % 91.4 ± 5.6% 5.4 ± 0.3%
Major forceps 5.8 ± 0.7 % 5.8 ± 0.8% 95.2 ± 3.4% 5.2 ± 0.4%
Minor forceps - - (99.7 ± 0.6%) 5.0 ± 0.4%
Internal capsule 7.7 ± 2.1 % 7.9 ± 2.0 % 99.6 ± 0.4% 5.9 ± 0.2%
Mean 7.2 ± 1.2 % 7.5 ± 1.6 % 95.4 ± 4.1% 5.5 ± 0.4%

When the fraction of short T2* signal in ViSTa relative to total water signal was calculated, the results yielded on average 5.5 ± 0.4 % (the same result were obtained when averaged over splenium, major forceps, and internal capsule ROIs and when averaged over all ROIs; Table 1). This suggests that overall the short T2* signal fraction in ViSTa is slightly smaller than the myelin water fraction measured in the GRE data (7.2 % ± 1.2 % when signals were averaged over an ROI first and estimated and 7.5 ± 1.6 % when MWF values were averaged over ROIs from a MWF map; Table 1) (see Discussion).

In Figure 4, the MWF maps from the GRE data are compared to the original and normalized ViSTa images from Scan 3 (the same 3.1 min. scan). Both GRE-MWF and normalized ViSTa images reveal higher intensity in splenium, internal capsule and optic radiation areas than neighboring white matter. The GRE-MWF images (Figs. 4A and 4D) show either no MWF or excessive MWF in the frontal lobe area due to the field inhomogeneity which has been reported in previous studies (Hwang et al., 2011; Hwang et al., 2010). When comparing remaining brain areas, the ViSTa images still show superior image quality with much less speckle artifact. This is also true when comparing the ViSTa images to SE-based MWI in the literature (Kolind et al., 2009). Compared to the conventional MWI, ViSTa has limited accuracy in quantification due to T1-weighting and other factors (see Discussion for apparent MWF in ViSTa). Despite this limitation, ViSTa may provide clinically useful information in detecting demyelination (Figure 5).

Figure 4.

Figure 4

GRE-MWF (A and D), ViSTa images (B and E), and ViSTa images after normalization (C and F) from two subjects. The ViSTa images (Scan 3 using a single echo with low readout bandwidth) show higher image quality than the GRE-MWF images. Both of them are acquired in 3.1 min. each. Yellow arrows indicate enhanced CSF after normalization.

Figure 5 shows FLAIR, PDW, post-contrast MPRAGE, and ViSTa images from an MS patient (Patient 4). The two large lesions (arrows) did not show signal enhancement in the post-contrast MPRAGE image (Fig. 5C) suggesting they are, so-called, T1 black holes with demyelination and axonal loss. In these lesions, the ViSTa image shows significantly reduced signal levels successfully delineating the lesions (Fig. 5D). Venous in-flow is noticeable in the ViSTa image. The other three patients also showed T1 black hole lesions and the corresponding areas in ViSTa had signal reduction (see supplementary figure for additional patient data).

When the effect of fat saturation was evaluated by comparing the first echo signal intensity with and without fat saturation, the signal dropped by 12.5% (three subjects, averaged over all ROIs). This may originate from the magnetization transfer effects. Fat saturation improves the visualization of the short T2* signal by suppressing bright fat signal (Fig. 2D vs. Fig. 4B) but is optional and can be removed for better quantification.

Discussion

In this work, a new method that selectively acquires short T2* signal by suppressing long T1 signal is demonstrated. The experimental results show that the remaining signal is dominated by the short T2* signal (on average, 95.4%), confirming that the suppression of long T1 leaves primarily short T2* signal. This observation can be explained by a multiple water component model in white matter that includes a short T1 and T2* component and long T1 and T2* component(s) (Deoni et al., 2008; Does and Gore, 2002; Helms and Hagberg, 2009; Koenig et al., 1990; Labadie et al., 2013; Lancaster et al., 2002; Stanisz et al., 2005). The observed signal decay after the long T1 suppression shows T2* values in the range of myelin water indicating that the remaining signal is from myelin water. The fraction of the short T2* signal over total water signal (on average 5.5 ± 0.4%) in ViSTa is in a similar range to the myelin water fraction in GRE-MWI (7.2 ± 1.2%), though the slightly reduced fraction suggests that there may be additional T1-weighted signal loss from the double inversion. The patient data further support the origin of ViSTa signal as myelin water. In the MS lesions, the FLAIR, PDW, and MPRAGE images suggest that the average water signal, dominated by long T1 components, has increased T1, T2, and proton density signal compared to surrounding white matter. However, the signal in the ViSTa image was decreased. This cannot be explained by the changes in the long T1 components because increases in proton density and T2 of the long T1 components would cause an increase in ViSTa signal. The increase in T1 in long T1 components may reduce the signal depending on the T1 value. However, the measured signal relative to total water (5.5%) is substantially larger than the expected signal in the range of T1 > 750 ms (0.5% or less). Hence, the changes in the long T1 components alone cannot explain the signal reduction in MS lesions. On the other hand, these lesions are expected to have severe demyelination that will result in signal reduction in ViSTa due to the loss of myelin water. Hence, our data support a multiple water component model that has a short T1 and T2* component and long T1 and T2* components. More importantly, T2* characteristics and MS patient results suggest that the short T2* signal in ViSTa is from myelin water.

In the ViSTa sequence, the T1 relaxation of the short T2* component (or myelin water) is an important design parameter for sequence optimization. However, measuring the actual T1 is challenging due to the small fraction (~ 10%) and the short relaxation time (T2* < 25 ms at 3 T). Previous studies have suggested that myelin water has shorter T1 than that of the other water components (Deoni et al., 2008; Does and Gore, 2002; Helms and Hagberg, 2009; Koenig et al., 1990; Labadie et al., 2013; Lancaster et al., 2002; Stanisz et al., 2005). The estimated value ranges from 118 ms at 3 T (Labadie et al., 2013) to at 390 ms at 1.5 T (Deoni et al., 2008). Some of these values were estimated using multi-parametric fitting, which is known to be sensitive to a model under investigation (e.g. two pool vs. three pool), initial values, and search ranges. Recently, Labadie et al., used T1 relaxography and suggested T1 of myelin water as 118 ms (the median value of all measurements) at 3 T (Labadie et al., 2013). At this T1 value, the myelin water signal in ViSTa will have 69% of the fully relaxed myelin water signal (Eq. 1) and ViSTa signal will have a myelin water fraction of 5.0% relative to total water signal which is calculated as the multiplication of the mean MWF (7.2%) in GRE-MWI and the signal reduction factor (0.69). This estimation is very close to our measurement in ViSTa (5.5%). Further research is necessary to accurately estimate the T1 of myelin water value. Once the T1 is known, a better sequence timing can be designed to improve SNR and quantification in ViSTa.

Other potential origins of the smaller short T2* fraction observed in ViSTa are cross-relaxation and chemical exchange between myelin water and macromolecules. Alternatively, a relatively short TR (= 97 ms) used for GRE may have increased myelin water fraction in GRE-MWI (Du et al., 2007b). Further research is necessary for a complete understanding of ViSTa signal.

Because of the confounding factors such as T1-weighting and exchange, ViSTa may have limited accuracy in quantifying myelin water fraction. Furthermore, the single-echo ViSTa acquisition used a relatively long TE (= 6.36 ms), and the resulting T2*-weighting was corrected by nominal T2*. This nominal correction can be an additional source of error in quantification. When the estimated fractions were compared between multi-echo ViSTa results and single-echo ViSTa results (all subjects and all ROIs), the difference was relatively small (mean fraction was 5.5 % in multi-echo data, and 6.0 % in single-echo data). This error can be further reduced by using a spiral or radial sequence with a short TE (< 1 ms). Because of these confounding factors, the fraction measured in our experiment may be referred to as apparent myelin water fraction (aMWF) to distinguish it from conventional MWF. Despite this limitation, ViSTa can provide qualitative myelin water images with no post-processing, and generate high quality images with demonstrated sensitivity to MS lesions. These features make ViSTa appealing for clinical applications where myelination is being assessed. Note that conventional MWF has also been reported to be sensitive to field strength, coil, parameters in sequence and post-processing (Guo et al., 2012; Kolind et al., 2009).

In the original ViSTa images (Figs. 4B and 4E), no post-processing was performed and the images have coil sensitivity weighting. This can be removed using a coil sensitivity map or a proton density image. In our data (Figs. 4C and 4F), the correction was performed by dividing the first echo GRE images that has slight T1-weighting. This weighting resulted in unwanted CSF enhancement in Figs 4C and 4F (arrows). The CSF enhancement can be removed by using a small flip angle (~ 5°) in GRE.

Conventional myelin water imaging has been shown to provide good sensitivity to demyelination (Laule et al., 2008; Laule et al., 2004; Mackay et al., 1994). It has also been suggested to differentiate demyelination from inflammation or edema, which is challenging with conventional MRI (Laule et al., 2007; Stanisz et al., 2004). Thus, MWI may have important diagnostic value in differentiating inflammatory lesions with and without demyelination. ViSTa may share the same advantage because inflammation, which is characterized by increased extracellular volume fraction, has been shown to produce limited changes in myelin water fraction and myelin water T2 distribution (Stanisz et al., 2004). Inflammation induced myelin water T1 change has not yet been studied and needs further investigation. If confirmed, the high image quality of ViSTa will be advantageous in detecting demyelinated lesions and distinguishing them from other types of lesions.

In our study, ViSTa was compared to GRE-MWI. The GRE- MWI requires less SAR and provides larger volume coverage than SE- MWI. However, the GRE-based method suffers from sensitivity to B0 field inhomogeneity (e.g. frontal lobe area in Fig. 4A). Techniques have been proposed to compensate for the field inhomogeneity effects (Hwang and Du, 2009; Hwang et al., 2010). Unlike GRE-based MWI, ViSTa does not require multiple echoes and generates myelin water images even near regions of field inhomogeneity.

The future extension of ViSTa for 3D coverage can be achieved by replacing the 2D single-echo ViSTa readout to a 3D multi-shot EPI, 3D interleaved stack-of-spiral, or 3D radial readout. Radial and spiral trajectories have the additional advantage of acquiring the center of k-space prior to T2* decay and, therefore, may provide higher SNR and improved quantification.

In a few studies, an inversion RF pulse prepared ultra-short echo time acquisition sequence (IR-UTE) was suggested to acquire short T2* signals in white matter (Du et al., 2007a; Waldman et al., 2003; Wilhelm et al., 2012). The inversion time was optimized to suppress long T1 signals in white matter. Additionally, two echoes (TE1 = 8 ~ 80 μs; TE2 = 1.2 ~ 11.08 ms) were acquired to further reduce long T2* signal contribution by subtracting the second echo from the first echo. Recently, Wilhelm et al. (Wilhelm et al., 2012) have demonstrated that when the duration of RF (~ 20 μs) and TE2 (= 1.2 ms) are short, the signal originates primarily from myelin protons (T2* < 1 ms) that exist inside the myelin and are different from myelin water protons. On the other hand, ViSTa uses a relatively long RF (1 ms) and TE (1.95 ~ 6.36 ms) and, therefore, has limited contribution from myelin proton signal.

Recently, several other new approaches that utilize magnetic susceptibility characteristics of myelin have been proposed to assess the integrity of myelin (Cohen-Adad et al., 2012; Duyn et al., 2007; Lee et al., 2010; Lee et al., 2011; Li et al., 2012; Liu, 2010; Miller et al., 2010; Wilhelm et al., 2012). These approaches, in conjunction with diffusion and MT imaging, may provide complementary information to ViSTa. The combination of these modalities may help us to better understand the biophysics of white matter and provide clinically useful information.

Conclusions

In this paper, we proposed a new approach to directly acquire short T2* signal. The method utilizes a double inversion RF scheme to suppress long T1 water signals. After the suppression, a short T2* component in the range of myelin water dominated the image. The method provides high quality images and successfully detected MS demyelinated lesions.

Supplementary Material

01

Highlights.

  • A new approach that selectively acquire short T2* is demonstrated.

  • Signal characteristics suggest that the primary signal is from myelin water.

  • MS lesions show signal reduction demonstrating sensitivity to demyelination.

Acknowledgments

This work was supported by NIH EB015893 and a Penn Medicine Comprehensive Neuroscience Center Pilot Grant.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  1. Balaban RS, Ceckler TL. Magnetization transfer contrast in magnetic resonance imaging. Magn Reson Q. 1992;8:116–137. [PubMed] [Google Scholar]
  2. Basser PJ, Mattiello J, LeBihan D. MR diffusion tensor spectroscopy and imaging. Biophys J. 1994;66:259–267. doi: 10.1016/S0006-3495(94)80775-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Cohen-Adad J, Polimeni J, Helmer K, Benner T, McNab J, Wald L, Rosen B, Mainero C. T2* mapping and B0 orientation-dependence at 7T reveal cyto- and myeloarchitecture organization of the human cortex. NeuroImage. 2012;60:1006–1014. doi: 10.1016/j.neuroimage.2012.01.053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Cover KS. A robust and reliable method for detecting signals of interest in multiexponential decays. Rev Sci Instrum. 2008;79:055106.1–055106.11. doi: 10.1063/1.2930799. [DOI] [PubMed] [Google Scholar]
  5. Deoni SCL, Rutt BK, Arun T, Pierpaoli C, Jones DK. Gleaning multicomponent T1 and T2 information from steady-state imaging data. Magn Reson Med. 2008;60:1372–1387. doi: 10.1002/mrm.21704. [DOI] [PubMed] [Google Scholar]
  6. Does MD, Gore JC. Compartmental study of T1 and T2 in rat brain and trigeminal nerve in vivo. Magn Reson Med. 2002;47:274–283. doi: 10.1002/mrm.10060. [DOI] [PubMed] [Google Scholar]
  7. Du J, Bydder M, T AM, Shimakawa A, Hinks S, Bydder GM. Ultrashort TE (UTE) imaging and T2 quantification of short T2 components in brain white matter. Proceedings of the 15th Annual Meeting of ISMRM; Berlin, Germany. 2007a. p. 3368. [Google Scholar]
  8. Du YP, Chu R, Hwang D, Brown MS, Kleinschmidt-DeMasters BK, Singel D, Simon JH. Fast multislice mapping of the myelin water fraction using multicompartment analysis of T2* decay at 3T: a preliminary postmortem study. Magn Reson Med. 2007b;58:865–870. doi: 10.1002/mrm.21409. [DOI] [PubMed] [Google Scholar]
  9. Duijn J, Matson G, Maudsley A, Hugg J, Weiner M. Human brain infarction: proton MR spectroscopy. Radiology. 1992;183:711–718. doi: 10.1148/radiology.183.3.1584925. [DOI] [PubMed] [Google Scholar]
  10. Duyn JH, Tan CX, Gelderen P, Yongbi MN. High-sensitivity single-shot perfusion-weighted fMRI†. Magn Reson Med. 2001;46:88–94. doi: 10.1002/mrm.1163. [DOI] [PubMed] [Google Scholar]
  11. Duyn JH, van Gelderen P, Li TQ, de Zwart JA, Koretsky AP, Fukunaga M. High-field MRI of brain cortical substructure based on signal phase. Proc Natl Acad Sci USA. 2007;104:11796–11801. doi: 10.1073/pnas.0610821104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Gambarota G, Cairns BE, Berde CB, Mulkern RV. Osmotic effects on the T2 relaxation decay of in vivo muscle. Magn Reson Med. 2001;46:592–599. doi: 10.1002/mrm.1232. [DOI] [PubMed] [Google Scholar]
  13. Gass A, Barker G, Kidd D, Thorpe J, MacManus D, Brennan A, Tofts P, Thompson A, McDonald W, Miller D. Correlation of magnetization transfer ration with clinical disability in multiple sclerosis. Ann Neurol. 1994;36:62–67. doi: 10.1002/ana.410360113. [DOI] [PubMed] [Google Scholar]
  14. Ghugre NR, Enriquez CM, Coates TD, Nelson MD, Wood JC. Improved R2* measurements in myocardial iron overload. J Magn Reson Imag. 2005;23:9–16. doi: 10.1002/jmri.20467. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Guo J, Ji Q, Reddick WE. Multi-slice myelin water imaging for practical clinical applications at 3.0 T. Magn Reson Med. 2012 doi: 10.1002/mrm.24527. available online. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Helms G, Hagberg GE. In vivo quantification of the bound pool T1 in human white matter using the binary spin bath model of progressive magnetization transfer saturation. Phys Med Biol. 2009;54:N529–N540. doi: 10.1088/0031-9155/54/23/N01. [DOI] [PubMed] [Google Scholar]
  17. Henkelman RM. Measurement of signal intensities in the presence of noise in MR images. Med Phys. 1985;12:232–233. doi: 10.1118/1.595711. [DOI] [PubMed] [Google Scholar]
  18. Hwang D, Chung H, Nam Y, Du YP, Jang U. Robust mapping of the myelin water fraction in the presence of noise: Synergic combination of anisotropic diffusion filter and spatially regularized nonnegative least squares algorithm. J Magn Reson Imag. 2011;34:189–195. doi: 10.1002/jmri.22586. [DOI] [PubMed] [Google Scholar]
  19. Hwang D, Du YP. Improved myelin water quantification using spatially regularized non-negative least squares algorithm. J Magn Reson Imag. 2009;30:203–208. doi: 10.1002/jmri.21783. [DOI] [PubMed] [Google Scholar]
  20. Hwang D, Kim D, Du Y. In vivo multi-slice mapping of myelin water content using T2* decay. NeuroImage. 2010;52:198–204. doi: 10.1016/j.neuroimage.2010.04.023. [DOI] [PubMed] [Google Scholar]
  21. Keough MB, Yong VW. Remyelination therapy for multiple sclerosis. Neurotherapeutics. 2013;10:44–54. doi: 10.1007/s13311-012-0152-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Koenig SH, Brown RD, 3rd, Spiller M, Lundbom N. Relaxometry of brain: why white matter appears bright in MRI. Magn Reson Med. 1990;14:482–495. doi: 10.1002/mrm.1910140306. [DOI] [PubMed] [Google Scholar]
  23. Kolind SH, Mädler B, Fischer S, Li DKB, MacKay AL. Myelin water imaging: implementation and development at 3.0 T and comparison to 1.5 T measurements. Magn Reson Med. 2009;62:106–115. doi: 10.1002/mrm.21966. [DOI] [PubMed] [Google Scholar]
  24. Labadie C, Lee J-H, Rooney WD, Jarchow S, Aubert-Frécon M, Springer CS, Möller HE. Myelin water mapping by spatially regularized longitudinal relaxographic imaging at high magnetic fields. Magnetic Resonance in Medicine. 2013 doi: 10.1002/mrm.24670. [DOI] [PubMed] [Google Scholar]
  25. Lancaster JL, Andrews T, Hardies LJ, Dodd S, Fox PT. Three-pool model of white matter. J Magn Reson Imaging. 2002;17:1–10. doi: 10.1002/jmri.10230. [DOI] [PubMed] [Google Scholar]
  26. Laule C, Kozlowski P, Leung E, Li DKB, MacKay AL, Moore G. Myelin water imaging of multiple sclerosis at 7 T: correlations with histopathology. NeuroImage. 2008;40:1575–1580. doi: 10.1016/j.neuroimage.2007.12.008. [DOI] [PubMed] [Google Scholar]
  27. Laule C, Vavasour I, Moore G, Oger J, Li D, Paty D, MacKay A. Water content and myelin water fraction in multiple sclerosis. J Neurol. 2004;251:284–293. doi: 10.1007/s00415-004-0306-6. [DOI] [PubMed] [Google Scholar]
  28. Laule C, Vavasour IM, Kolind SH, Li DKB, Traboulsee TL, Moore G, MacKay AL. Magnetic resonance imaging of myelin. Neurotherapeutics. 2007;4:460–484. doi: 10.1016/j.nurt.2007.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Lee J, Shmueli K, Fukunaga M, van Gelderen P, Merkle H, Silva AC, Duyn JH. Sensitivity of MRI resonance frequency to the orientation of brain tissue microstructure. Proc Natl Acad Sci U S A. 2010;107:5130–5135. doi: 10.1073/pnas.0910222107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Lee J, van Gelderen P, Kuo L, Merkle H, Silva AC, Duyn JH. T2*-based fiber orientation mapping. NeuroImage. 2011;57:225–234. doi: 10.1016/j.neuroimage.2011.04.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Lenz C, Klarhöfer M, Scheffler K. Feasibility of in vivo myelin water imaging using 3D multigradient-echo pulse sequences. Magn Reson Med. 2011;68:523–528. doi: 10.1002/mrm.23241. [DOI] [PubMed] [Google Scholar]
  32. Li X, Vikram DS, Lim IAL, Jones CK, Farrell JAD, van Zijl P. Mapping Magnetic Susceptibility Anisotropies of White Matter in vivo in the Human Brain at 7 Tesla. NeuroImage. 2012;62:314–330. doi: 10.1016/j.neuroimage.2012.04.042. [DOI] [PMC free article] [PubMed] [Google Scholar]
  33. Liu C. Susceptibility tensor imaging. Magn Reson Med. 2010;63:1471–1477. doi: 10.1002/mrm.22482. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Mackay A, Whittall K, Adler J, Li D, Paty D, Graeb D. In vivo visualization of myelin water in brain by magnetic resonance. Magn Reson Med. 1994;31:673–677. doi: 10.1002/mrm.1910310614. [DOI] [PubMed] [Google Scholar]
  35. Menon R, Allen P. Application of continuous relaxation time distributions to the fitting of data from model systmes and excised tissue. Magn Reson Med. 1991;20:214–227. doi: 10.1002/mrm.1910200205. [DOI] [PubMed] [Google Scholar]
  36. Miller KL, Smith SM, Jezzard P. Asymmetries of the balanced SSFP profile. Part II: White matter. Magn Reson Med. 2010;63:396–406. doi: 10.1002/mrm.22249. [DOI] [PubMed] [Google Scholar]
  37. Oh J, Han ET, Pelletier D, Nelson SJ. Measurement of in vivo multi-component T2 relaxation times for brain tissue using multi-slice T2 prep at 1.5 and 3 T. Magn Reson Imaging. 2006;24:33–43. doi: 10.1016/j.mri.2005.10.016. [DOI] [PubMed] [Google Scholar]
  38. Reiter DA, Lin P-C, Fishbein KW, Spencer RG. Multicomponent T2 relaxation analysis in cartilage. Magn Reson Med. 2009;61:803–809. doi: 10.1002/mrm.21926. [DOI] [PMC free article] [PubMed] [Google Scholar]
  39. Stadlbauer A, Salomonowitz E, van der Riet W, Buchfelder M, Ganslandt O. Insight into the patterns of cerebrospinal fluid flow in the human ventricular system using MR velocity mapping. NeuroImage. 2010;51:42–52. doi: 10.1016/j.neuroimage.2010.01.110. [DOI] [PubMed] [Google Scholar]
  40. Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, Henkelman RM. T1, T2 relaxation and magnetization transfer in tissue at 3T. Magn Reson Med. 2005;54:507–512. doi: 10.1002/mrm.20605. [DOI] [PubMed] [Google Scholar]
  41. Stanisz GJ, Webb S, Munro CA, Pun T, Midha R. MR properties of excised neural tissue following experimentally induced inflammation. Magnetic Resonance in Medicine. 2004;51:473–479. doi: 10.1002/mrm.20008. [DOI] [PubMed] [Google Scholar]
  42. van Gelderen P, de Zwart JA, Lee J, Sati P, Reich DS, Duyn JH. Nonexponential T2* decay in white matter. Magn Reson Med. 2012;67:110–117. doi: 10.1002/mrm.22990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Vasilescu V, Katona E, Simplaceanu V, Demco D. Water compartments in the myelinated nerve. III. Pulsed NMR result. Experientia. 1978;34:1443–1444. doi: 10.1007/BF01932339. [DOI] [PubMed] [Google Scholar]
  44. Waldman A, Rees JH, Brock CS, Robson MD, Gatehouse PD, Bydder GM. MRI of the brain with ultra-short echo-time pulse sequences. Neuroradiology. 2003;45:887–892. doi: 10.1007/s00234-003-1076-z. [DOI] [PubMed] [Google Scholar]
  45. Werring DJ, Clark CA, Barker GJ, Thompson AJ, Miller DH. Diffusion tensor imaging of lesions and normal-appearing white matter in multiple sclerosis. Neurology. 1999;52:1626–1626. doi: 10.1212/wnl.52.8.1626. [DOI] [PubMed] [Google Scholar]
  46. Whittall KP, MacKay AL. Quantitative interpretation of NMR relaxation data. JMR. 1989;84:134–152. [Google Scholar]
  47. Wilhelm MJ, Ong HH, Wehrli SL, Li C, Tsai PH, Hackney DB, Wehrli FW. Direct magnetic resonance detection of myelin and prospects for quantitative imaging of myelin density. PNAS. 2012;109:9605–9610. doi: 10.1073/pnas.1115107109. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

01

RESOURCES