Abstract
In the last 20 years, the use of positron emission tomography (PET) has grown dramatically because of its oncological applications, and PET facilities are now easily accessible. At the same time, various groups have explored the specific advantages of PET in heart disease and demonstrated the major diagnostic and prognostic role of quantitation in cardiac PET. Nowadays, different approaches for the measurement of myocardial blood flow (MBF) have been developed and implemented in user-friendly programs. There is large evidence that MBF at rest and under stress together with the calculation of coronary flow reserve are able to improve the detection and prognostication of coronary artery disease. Moreover, quantitative PET makes possible to assess the presence of microvascular dysfunction, which is involved in various cardiac diseases, including the early stages of coronary atherosclerosis, hypertrophic and dilated cardiomyopathy, and hypertensive heart disease. Therefore, it is probably time to consider the routine use of quantitative cardiac PET and to work for defining its place in the clinical scenario of modern cardiology.
1. Introduction
At the beginning of its clinical use in the late eighties of the last century, positron emission tomography (PET) was mainly proposed for neurological and cardiological indications. At that time, 18F-labelled deoxyglucose (FDG) was regarded as an effective tracer for studying brain glucose consumption and for identifying viable hibernating myocardial tissue in patients with congestive heart failure of ischemic etiology [1]. Effective myocardial perfusion tracers such as 13N-labelled ammonia, 15O-labelled water, and Rubidium-82 (82Rb), a potassium-analogue, were also available [2–6]. The intrinsic advantages of PET over single-photon emission computed tomography (SPECT) were particularly attractive for heart imaging, because the possibility to perform an effective attenuation correction permitted to overcome a main limitation of SPECT perfusion imaging [3, 4, 6]. Similarly, the metabolic imaging with 18FDG had no counterpart within SPECT tracers [7]. However, in spite of several comparative studies that demonstrated the superior diagnostic performance of PET over SPECT for perfusion imaging and the wide literature about viability using flow-metabolism imaging with any perfusion tracer plus 18FDG, the clinical use of PET in cardiology did not achieve the expected dissemination. Several practical considerations hampered its use: first of all, naturally, the need of on-site cyclotron for tracer production. Even the availability of a generator-produced perfusion tracer like 82Rb could not overcome the problem, because the exceedingly high costs did not justify its purchase in the absence of a very high patient throughput. Finally, in many countries the reimbursement issue remained unsolved.
In the following years, the appearance of gated SPECT reinforced the role of traditional nuclear cardiology and improved its diagnostic and prognostic capabilities [8, 9]. Simultaneously, the exponential growth of oncological indications limited the availability of PET scanners for other purposes. Therefore, PET did not reach a widespread clinical use for heart disease. On the other hand, the intrinsic quantitative capabilities of PET were applied in a large number of research studies that explored their value in a broad spectrum of clinical conditions.
Presently, the wide diffusion of PET facilities (often with on-site cyclotron) supported by oncology, the enhanced availability of 82Rb generators, and the imminent marketing of 18F-labelled perfusion tracers have renewed the interest for the issue of a generalized use of PET in cardiology. As a proof of this, the relative merits of PET and SPECT and the problems related to the incorporation of PET in the clinical practice have been most recently reviewed [10–12]. However, the most important reason to support the widespread use of cardiac PET is still the same that had been identified 20 years ago by M.M. Ter-Pogossian:
“The (…) utility of PET in cardiovascular research and in clinical cardiology results from its quantitative power and applicability to assessment of the kinetics in tissue of physiologic labeled substrates and tracers. (…) Because of its quantitative power (…) PET provides information delineating biochemical and physiological processes that is not available by any other means.” [13].
Therefore, it would be desirable to get this opportunity to orient the renewed interest for cardiac PET in the direction of a quantitative approach to data interpretation, in order to exploit the full and most specific potential of this methodology. Aim of this paper is to describe the added value and the clinical role of quantitation in cardiac PET to support the expansion of this imaging modality and the routine application of PET measurements in the clinical practice.
2. Technical Considerations in Quantitative Cardiac PET
To be used for quantitation, PET must be of optimal quality. Therefore, all the technical requirements of high-quality PET imaging must be fulfilled. Although PET has the advantage of an inherently effective attenuation correction, which is nowadays usually performed with the computed tomography (CT) images acquired using a PET/CT scanner [14], several factors may degrade PET images and accordingly affect the accuracy of quantitative measurements. The most important are related to the intrinsic nature of positron emission, such as the positron range, that is, the distance that a positron travels before undergoing annihilation, and the circumstance that the two emitted photons are not perfectly collinear; then there are scatter and random coincidences; various algorithms have been implemented in the reconstruction programs to correct whenever possible these interferences [15–20]. Attenuation correction itself can cause artifacts if not properly performed, for instance because of patient motion, and there are various methods to deal with the possible misalignment problems [21–23]. Other technical improvements of PET that favorably influence its quantitative capabilities are 3D acquisition, the availability of better detector materials, the time-of-flight technology, the option to acquire dynamic sequences in list mode instead of using frames of predefined duration, the implementation of electrocardiographic, and possibly of respiratory gating [24–27]. The image reconstruction has been improved by the use of iterative algorithms; although their advantage over the standard filtered back projection for the issue of quantitation is not yet definitively established [28, 29].
3. Tracers for Quantitative Cardiac PET
Both perfusion and metabolic tracers are available for quantitative cardiac PET. For the measurement of myocardial blood flow (MBF) at rest and under stress and for the calculation of the coronary flow reserve (CFR), which is the ratio of stress over resting MBF, three tracers have been extensively used and evaluated: 15O-water, 13N-ammonia, and 82Rb. More limited is the experience with 11C-acetate: because of its peculiar kinetics, this tracer allows the simultaneous measurement of MBF and of myocardial oxygen consumption, and apparently is superior to 13N-ammonia in image quality and statistical accuracy of quantitative data [30–33]. However, 11C requires an on-site cyclotron, and particular care in target handling and tracer synthesis because of its gaseous nature, so that its use in the clinical practice has never got wide acceptance [34]. An additional perfusion tracer labeled with 18F is under advanced clinical evaluation and will be likely available in the near future [35]. For quantitative imaging of glucose metabolism there is 18FDG [36, 37]. Metabolic and neuronal tracers labeled with 11C have been used as well, but their availability is limited and their role in clinical imaging is still undefined [37, 38].
The features of the commonly employed perfusion tracers are summarized in Table 1. None of them can be regarded as definitely perfect. There are no doubts that in terms of extraction a diffusible tracer such as 15O-water is ideal [39–44]. However, the very short half-life and the immediate washout from myocardial tissue preclude the achievement of qualitatively acceptable images for visual assessment, so that the study evaluation is exclusively based on the MBF measurements [25, 43, 44]. Furthermore, not all on-site cyclotrons are able to produce 15O-water. Conversely, the production and handling of 13N-ammonia are much simpler and this can be regarded as the most reasonable tracer for all centers with an on-site cyclotron [25, 34]. The combination of visually evaluable images for an easy region-of-interest definition, quite high myocardial extraction, defined tracer kinetics, and wide clinical experience make 13N-ammonia probably the best routine tracer for quantitative cardiac PET to this point available [25, 43–52]. The simple use of myocardial distribution images for the qualitative comparison of rest and stress tracer uptake, although employed with good results in the clinical practice, does not make the most of the full potential of this tracer [53]. As regards 82Rb, its main advantages are that it is a generator-produced tracer and there is a large clinical experience on its use [43, 44]. On the other hand, the image quality is lower than that of 13N-ammonia both because of the shorter half-life and of the high energy of the positrons, with consequent long path from site of emission to site of annihilation [43]. Furthermore, the extraction fraction is suboptimal and the kinetics model less established [43, 44]. Nevertheless, many recent papers clearly indicate that 82Rb quantitative PET is a well feasible and reliable technique [54–58]. The present availability of 82Rb generators produced by vendors with worldwide distribution is an additional prerequisite for the more extensive use of this methodology [25, 59, 60].
Table 1.
Tracer | Patient throughput | Availability | Image quality | Background | Quantitation |
---|---|---|---|---|---|
H2 15O | High | Low | Very poor | NA | Optimal |
13NH3 | Medium | Medium | Fair to good | Quite high | Good |
82Rb | High | High | Fair to good | Acceptable | Fair |
18F-agents | Low | High | High | Low | Good |
4. Quantitative Measurements in Cardiac PET
The largest experience in cardiac quantitative PET is in the measurement of MBF. There are a large number of different approaches, mainly based on compartmental models of tracer kinetics [61]. In general, they require a dynamic acquisition starting at the moment of tracer injection, which is then prolonged until the phase of tracer uptake in the myocardium is completed. By the identification of volumes of interest drawn on the ventricular cavities and on the myocardial wall of the left ventricle, it is possible to obtain time activity curves depicting the input of the tracer and its accumulation within the myocardial tissue. These time activity curves are then properly fitted and the related parameters calculated, obtaining the tracer input function and the tracer amount within the myocardium. According to the chosen compartmental model, these values are included in the equations for the calculation of the kinetic parameter that best depicts MBF. The equations often include appropriate correction factors for other variables, such as activity spill over from blood into the myocardium, as well as for the partial volume effect and the system resolution [62–64]. The technical details of these procedures go beyond the aim of this paper, and can be found in several methodological papers and outstanding reviews [43, 44, 59]. Just as an example, the algorithm proposed by Lortie et al. for MBF calculation using 82Rb will be shortly described [58]. They use for describing 82Rb kinetics a one-compartment model:
(1) |
where C a(t) and C m(t) are the concentrations of 82Rb in arterial blood and myocardium, respectively. To address spill over and partial volume effect, a correction is introduced:
(2) |
where C PET is the measured concentration of 82Rb, F LV a number between 0 and 1 and (1 − F LV) an estimate of partial volume recovery coefficient [64]. After that the parameters are estimated using a weighted least-squares method, K 1 values are converted to MBF taking into account the flow-dependent 82Rb extraction:
(3) |
where a and b are correction factors derived from comparison with 13N-ammonia MBF using a weighted least-squares algorithm.
Many studies have compared the reliability of the different models used for the quantitation of MBF and have shown that there are methodological inequalities that must be taken into account if the results of diverse laboratories have to be compared [49, 61, 65]. What is most important for the clinician, however, is that nowadays there is a large choice of programs for performing in a user-friendly way this kind of procedures [66–70]. Although they have not reached the diffusion and the implementation degree of the gated SPECT analysis programs, they are nevertheless easily accessible. Some of them have been already implemented in PET scanners by vendors, others can be purchased from software producers, and finally some have been developed in academic institutions and either commercialized or made freely available on the net. Yet, the results they achieve, also because of the different models that can be used, are not always exchangeable. However, various studies demonstrate that the reproducibility of MBF calculated using the same method is excellent [66–75].
As regards glucose consumption, the most practiced approach is the graphical analysis as proposed by Patlak [76]. Traditionally, 18FDG imaging has been performed with a visual (or semiquantitative) evaluation of tracer uptake after glucose loading for the detection of increased uptake in regions with perfusion defects (flow/metabolism mismatch) and this method has been considered a noninvasive gold standard for the identification of viable myocardium within dysfunctional ventricular segments [1, 7]. More recently, however, several studies have shown that the measurement of the glucose metabolic rate under more physiologic fasting conditions has an important meaning in different clinical settings [77, 78]. Therefore, it can be expected that quantification of 18FDG will be increasingly used.
5. MBF and CFR in Various Clinical Settings
Aim of the following section is to report the many studies that have investigated MBF and CFR in health and disease. Most of these studies have been performed on small patient populations. However, their results are consistent in showing that quantitative measurements are highly effective for detecting significant differences related to the explored condition. First of all, the behavior of MBF and CFR in normal subjects has been examined. Although minor differences exist, mainly related to the tracer and the calculation model, the values appear quite homogeneous, with baseline mean resting values ranging from 0.8 to 1.2 mL/min/g tissue and mean stress values (under coronary vasodilator, adenosine, or dipyridamole) ranging from 2.7 to 4.6 mL/min/g tissue [41, 47, 79–85]. Accordingly, the mean CFR ranges from 2.9 to 4.4 mL/min/g tissue [41, 47, 79–85]. The baseline values are influenced by the oxygen demand and thus the correction of the resting MBF for the rate pressure product has been proposed in order to obtain a reliable CFR [47, 71, 85–87]. Age influences MBF through the changes in cardiac work [85]. On the other hand, a reduction in hyperemic MBF has been registered with ageing [88]. These changes cause a decrease in CFR in older subjects [89]. As regards gender, some reports indicate higher MBF values at rest and during hyperemia in women, with a possible relationship with the better blood lipid profile [87, 90]. In another most recent report, it has been observed that among normal subjects those with potential risk factors such as familial history of coronary artery disease (CAD) and slight abnormalities in blood lipids have lower MBF values than “true” normals, that is, subjects with completely negative familial history and perfectly regular blood profile [91].
A series of studies has explored MBF in relationship with various conditions that could affect it. In syndrome X, considered as a paradigm of potential ischemia in the absence of obstruction of the epicardial vessels, the various studies have not demonstrated a clear-cut difference in stress MBF values between patients and healthy controls: although a quite large proportion of syndrome X patients have a reduced CFR, there is an overlap with the normal subjects and no relationship could be demonstrated between CFR abnormalities and symptoms or ischemia-like electrocardiographic changes [80, 92–95]. In the broader category of patients with typical angina and normal coronary angiograms, the large majority has an abnormal CFR, that is, caused both by increased resting MBF and decreased stress MBF [80, 96]. More recently, a relationship between the presence of CFR abnormality and worse risk factor profile has been reported in these subjects [97].
Other studies have examined the behavior of MBF and CFR in patients with various risk factors for CAD. In hypertensive patients, another group in whom ischemic symptoms are frequently reported without detectable coronary angiographic abnormalities, MBF has been found decreased even in subjects without signs of left ventricular hypertrophy, and improvements have been demonstrated after antihypertensive therapy [98, 99]. In particular, the improvements may precede the antihypertensive effect and are partly related to the drug category [99, 100]. On the other hand, in patients with left ventricular hypertrophy, lower MBF and CFR have been found as compared with age-matched controls, and the difference is larger in those with primary hypertrophy (hypertrophic cardiomyopathy-HCM) than in those with secondary hypertrophy (caused by hypertension or by aortic stenosis); in this last group there is still a quite good hyperemic response, but this is partly offset by the high resting MBF [101, 102]. Another very interesting finding is that even in young patients fulfilling the definition of borderline hypertension there is a CFR reduction as compared to age-matched controls: this would imply that an impaired vasodilator capacity can be a most precocious sign of hypertension-related myocardial damage [103].
In diabetic patients, it has been demonstrated that there is a reduction in CFR in young subjects without other evidence of cardiac involvement and with no or just minimal signs of vasculopathy in other organs [104]. In noninsulin dependent diabetics, male sex, and poor glycemic control, but not age and type of therapy, have been found related to the degree of microvascular impairment [105]. By adding to the vasodilator stress the cold pressor test (CPT), which explores the endothelial reactivity to sympathetic stimulation, Di Carli et al. observed that together with a reduced vasodilator reserve, diabetic patients with neuropathy have an abnormal response to CPT, in agreement with other signs of sympathetic nerve dysfunction [106]. In a larger patient population, the abnormal response to CPT has been registered in approximately one third of asymptomatic diabetics and associated with higher baseline MBF [107]. In premenopausal women, those with diabetes have a reduced CFR and an abnormal CPT response compared with healthy controls, and their values are similar to those observed in postmenopausal women, showing one mechanism through which diabetes abolishes the sex differential in CAD [108]. Apparently, both endothelium-dependent and -independent coronary vasodilator function abnormalities are related to the chronic hyperglycemia and not to the type of diabetes (insulin-deficient versus insulin-resistant) [109]. However, other authors have registered an impaired CPT response in insulin-resistant patients [110]. Moreover, an increasing severity of microvascular dysfunction has been observed in patients with type 2 diabetes, starting from those with insulin resistance but without glucose intolerance, to those with isolated impaired glucose tolerance, normotensive diabetes, and hypertension plus diabetes [111].
The capability of different treatments to correct MBF abnormalities in diabetic has been tested. Bengel et al. have demonstrated that the improved glycemic control using an insulinotropic agent does not influence the MBF values at rest, under adenosine and during CPT [112]. Conversely, in diabetic patients with CAD, insulin infusion improves resting and adenosine MBF both in ischemic and nonischemic regions [113]. In the long term, the achievement of euglycemia is associated with a significant improvement of the MBF response to CPT [114]. The same group has been able to demonstrate a slower progression of the coronary vessel structural alterations present at baseline in diabetics, after an effective glucose-lowering treatment prolonged for several months [115, 116].
In patients with familial hypercholesterolemia and familial combined hyperlipidemia, there is a CFR impairment, which in the latter case is also related to the phenotype [117, 118]. In asymptomatic patients with hypercholesterolemia, a reduced CFR has been as well observed, independently of the presence of CAD and directly related to the low-density lipoprotein levels [119–121]. As observed for diabetes, treatment is able to reverse the MBF abnormalities in these patients. Even short-term conditioning together with a low-fat diet has a favorable effect [122]. Specific lipid lowering therapy with statins improves adenosine MBF in otherwise healthy subjects with hypercholesterolemia only if they had reduced values before treatment [123]. A positive effect was as well registered in patients with familial hypercholesterolemia [124]. In CAD patients there is a delay in the CFR improvement as compared to the achievement of blood lipid normalization [125]. The CFR increase is detected mainly in the regions with abnormal reserve before treatment [126, 127]. However, in patients with diffuse atherosclerosis with angiographically normal coronary arteries, a disseminated MBF improvement has been registered [128]. Similarly, in a group of patients with angina, but without significant obstructions at coronary angiography, all affected by mild to moderate hypercholesterolemia, the baseline CFR significantly improves after 6 months of lipid lowering therapy; interestingly, there is also a reduction in anginal symptoms [129].
In otherwise healthy smokers, Campisi et al. have demonstrated preserved CFR under dipyridamole, but an abnormal response during CPT, which is more severe in case of more prolonged smoking habit [130]. Starting from the hypothesis that increased oxidative stress could be one of the mechanisms through which smoke causes vascular dysfunction, Kaufmann et al. have demonstrated a CFR improvement in smokers submitted to intravenous vitamin C administration [131]. The simple cessation of smoking restores a normal response to CPT in otherwise healthy subjects [132].
There are other factors that have been found to affect the MBF and CFR. For instance, Schindler et al. have observed a decreased response to CPT in patients with coronary risk factors and chronic inflammation detected by high C-reactive protein [133]. The same authors have registered an abnormal response to CPT in overweight and obese subjects and a decreased vasodilator CFR in the latter group only, suggesting that the endothelium dependent vasomotor function is more precociously affected than the total vasodilator capacity in patients with increased weight [134]. In postmenopausal women, long-term hormone replacement therapy has been found to improve the response to CPT only in those without associated coronary risk factors, whilst in those with a significant risk profile the reduced CFR under dipyridamole at baseline is not influenced by estrogens [135].
In summary, it appears clear that quantitative cardiac PET is a precious and reliable tool to investigate pathophysiologic changes in heart disease and to evaluate the possibility to prevent or correct them before they reach an otherwise detectable clinical expression.
6. MBF and CFR in Patients with CAD
As regards the behavior of MBF and CFR in patients with CAD, there is a demonstrated relationship between the degree of coronary artery stenosis and the impairment of stress MBF and CFR, which is present independently of the employed perfusion tracer, as shown by studies performed with 15O-water, 13N-ammonia and more recently with 82Rb [136–140]. Another most interesting finding is that CFR in CAD patients is reduced as compared to healthy controls also in territories without epicardial stenosis [139]. Similarly, it has been demonstrated that the MBF and CFR in territories subtended by patent coronary artery by-pass grafts are lower than in normal controls, and this independently of the type of conduit (saphenous vein or internal mammary artery grafts) [141]. Conversely, it has been found that in case of coronary artery stenting there is a good early recovery of MBF and CFR, with values that are no significantly different from those of normal territories [142].
In territories with stenosis and perfusion defects in qualitative 82Rb PET, the MBF is abnormally reduced; however, low values are registered also in territories with angiographic stenosis but without apparent perfusion defects; moreover an inverse relationship between coronary stenosis and CFR has been registered [143]. It has been as well demonstrated that there is a relationship between stress MBF (and CFR) and the presence of symptoms and electrocardiographic changes [144]. There is a good agreement between the CFR values measured using quantitative PET and the flow velocity reserve measured using Doppler and transesophageal echocardiography [145]. PET can be useful to establish the effectiveness of coronary collateral circulation [146–148]. The expected reduction of MBF as compared to normal territories has been found to be much more severe in case of regional dysfunction, suggesting a protective role of well-developed collateral circulation [148].
7. The Additional Value of Quantitative Perfusion PET
Previous studies comparing the visual assessment of myocardial perfusion using PET and SPECT had demonstrated the superiority of the former, chiefly because of the better quality of PET images and the sounder attenuation correction methodology [44, 53, 149, 150]. This superiority alone, however, would hardly justify the use of a more demanding, more expensive, and less widely available methodology, although some cost-effectiveness studies suggest that the higher accuracy of PET could compensate for the higher expenses [151, 152]. In particular, studying a wide patient population with matched pretest probability of CAD, Merhige et al. could demonstrate more than 50% reduction in invasive angiography and revascularization, 30% cost savings, and similarly excellent 1 year outcomes in patients evaluated with PET versus those examined with SPECT [153]. The scenario could become completely different if the added value of quantitative PET were taken into account. The advantage of MBF and CFR quantitation by PET over the simple evaluation of the relative tracer uptake, which may be performed either by PET or SPECT, must be examined separately in the various heart diseases.
8. Diagnosis of CAD
The current concept of diagnosis in patients with suspect CAD is to optimize the technique choice on the basis of the pretest probability. In general, therefore, a noninvasive imaging methodology finds its role in patients with intermediate pretest likelihood, because those with very low probability do not require imaging and those with high probability should directly proceed to invasive evaluation with coronary angiography [157]. The specific contribution of quantitative cardiac perfusion PET in this scenario is to help identifying the total ischemic burden in case of multivessel CAD and to pick patients with balanced three-vessel CAD, who are frequently missed using the relative tracer uptake, as it is performed in myocardial perfusion SPECT and even in qualitative perfusion PET [158–163]. By using previously identified stress MBF and CFR thresholds, Johnson and Gould have been able to construct a model that offers a high diagnostic reliability for an objective assessment of CAD burden of the single patient [144].
What is probably more important and certainly more innovative is the possibility to use quantitative perfusion PET to detect patients with preclinical CAD. As previously detailed, MBF and CFR have been found abnormal in subjects with various risk factor profiles. This circumstance and the fact that in many cases MBF and CFR monitoring may identify the degree of response to risk factor correction would support the use of quantitative PET for orienting primary prevention of CAD [164].
9. Evaluation of CAD Severity
As already mentioned, the quantification of MBF and CFR increases the capability to detect multivessel coronary artery disease. More generally, there are sample data indicating that the hemodynamic meaning of a coronary stenosis is not easily predictable and that there is a wide variability of flow abnormalities for any given stenosis pattern [137, 138, 165]. Therefore, the quantification of MBF and CFR is much more reliable to predict the true myocardial perfusion than the relative tracer uptake in SPECT imaging.
10. CAD Prognostication
Independently of the definition of the CAD extent, a most important issue in current cardiology is to perform a correct and reliable prognostication of these patients. The prognostic value of myocardial perfusion PET has been already demonstrated using the simple visual assessment of myocardial uptake, thus with the same approach currently employed for SPECT perfusion tracers [166]. In particular, the superiority of a PET-based prognostication had been recognized even in patients with apparently normal perfusion pattern in 201Tl myocardial perfusion scintigraphy and in 99mTc-sestamibi SPECT [167, 168]. The addition of gating during PET acquisition further improves the prognostic performance of 82Rb PET [169]. The abnormal MBF and CFR response to CPT has been shown to have prognostic implications [170]. With the more commonly used vasodilator stress, subsequent studies have shown that the quantitative assessment of MBF and of CFR has a clear prognostic meaning [171–174]. Most recently, a very large study has convincingly demonstrated that the global MFR has an incremental value for prognostication over all other established variables that can be derived from gated PET with 82Rb [175]. Patients in the lowest tertile of CFR have a significantly worse prognosis than the other subjects, independently of perfusion pattern and functional response to vasodilator stress. Conversely, patients with CFR in the highest tertile have a good prognosis even in case of perfusion defects and suboptimal functional response [175].
11. Detection of Myocardial Microvascular Disease
Another major contribution of quantitative myocardial PET has been to open the way to the assessment of myocardial microvascular disease. There is a wide range of cardiac conditions that imply, in a more or less central position for their physiopathology, the presence of a dysfunction in coronary microcirculation. According to a proposed classification, they can be grouped in four categories: type A includes coronary microvascular dysfunction occurring in the absence of obstructive epicardial coronary artery disease and other myocardial diseases; type B is the coronary microvascular dysfunction occurring in cardiomyopathies; type C covers coronary microvascular dysfunction occurring in the presence of obstructive CAD; finally, type D comprises iatrogenic coronary microvascular dysfunction [176]. The types A, C, and D, however, are someway related to CAD, because type A can be considered the equivalent of the early abnormalities in preclinical coronary artery disease, type C includes the reported abnormalities in territories not related to stenotic vessels, and type D the abnormalities in MBF that have been registered in patients submitted to percutaneous revascularization procedures. A very particular case of type D could be considered the cardiac allograft vasculopathy in heart-transplanted patients, because the process involves both the epicardial and the microvascular coronary system. In these subjects, recent data indicate that quantitative PET is able to identify (and potentially follow up) the cardiac allograft vasculopathy [177]. On the contrary, type B includes various different heart diseases in which the presence of microvascular dysfunction has been convincingly demonstrated mainly by quantitative perfusion PET.
The largest number of studies involves HCM patients. In these subjects, symptoms and signs that can mimic coronary artery disease are often registered, but mainly in the absence of any abnormality of the epicardial vessels [178, 179]. Conversely, the presence of microvascular impairment with reduced MBF under vasodilatation and decreased CFR has been demonstrated by several reports [180–182]. Moreover, various studies indicate that the microvascular involvement is a mainstay of HCM physiopathology and not just a late consequence of the increased wall thickness [183–186]. Accordingly and most importantly, the detection of microvascular dysfunction has major prognostic implications in these patients [187, 188]. The presence of microvascular impairment has been also demonstrated in patients with Anderson-Fabry disease, in whom cardiac involvement is a frequent complication [189, 190]. Similar findings have been observed in patients with idiopathic dilated cardiomyopathy (DCM) [191–195]. The prognostic implications of reduced stress MBF are present in DCM patients as well [196]. Other potential uses of quantitative PET in the patients with heart failure have been proposed [197, 198]. Although shown in a very preliminary study on a small patient population, the quantitative analysis of MBF appears promising to evaluate the effectiveness of medical treatment [199]. Finally, it has been demonstrated that in patients with hypertensive cardiomyopathy there is an impairment of microvascular function, which can explain the presence of anginal symptoms and is not homogeneously related with the degree of ventricular hypertrophy [200].
12. Hybrid Imaging
The current standard of PET devices is to have an integrated CT scanner that allows attenuation correction and facilitates the anatomical interpretation of PET findings. This offers the advantage of permitting a simultaneous evaluation of myocardial perfusion and coronary vessel morphology by means of CT angiography, as has been demonstrated in several articles [196–207]. Other studies show that a CT aimed to estimate the calcium score can be used for attenuation correction in quantitative PET and therefore the two clinical data can be usefully combined [203, 208]. Only part of these studies, however, specifically evaluate the integration of PET quantitative perfusion parameters and CT results [204, 206, 207]. The main interest of this particular combination is to elucidate the complex relationship between coronary vessel abnormalities (beyond lumen obstruction and thus including wall atherosclerosis) and myocardial perfusion [209, 210]. Naturally these advantages must be weighed against the higher technical, logistical, and dosimetric burden of hybrid imaging [211].
13. Dosimetric Data
The use of PET implies the administration of radioactive tracers. Therefore, a comment on the risks related to ionizing radiations is mandatory. The available data indicate that PET offers always a favorable dosimetric profile as compared to myocardial perfusion SPECT (Table 2) [154–156]. Among PET tracers, however, the use of 82Rb implies a much higher dose than 15O-water or 13N-ammonia. In case of hybrid imaging, the additional dose related to the CT procedure must be taken into account [211]. The most recent advances in this field, nevertheless, allow a significant reduction of the radiation dose for the CT and therefore justify the growing trend to consider hybrid imaging as a most desirable approach for perfusion PET [212, 213].
Table 2.
Technique | Dose, mSv |
---|---|
Thallium-201 SPECT | 25.1–31.4 |
Technetium-99m-sestamibi SPECT | 10.6–12 |
82Rb PET |
13.5–16 |
13N-ammonia PET | 2.4 |
15O-water PET | 2.5 |
CT angiography (64 MSCT) | 8.4–21.4 |
Coronary angiography | 2.3–22.7 |
14. Comparison with Other Imaging Modalities
There are no doubts that PET is the most established technique for MBF and CFR measurement. Given the importance of these quantitative parameters, there is a growing interest for the possibility to obtain them using other imaging modalities that are already highly effective for the assessment of heart disease. In particular, the capability of magnetic resonance imaging to achieve reliable MBF using a first-pass technique during gadolinium injection has been demonstrated [214]. Although it requires quite complex computational approaches, this is probably the most effective alternative to PET, with the major advantage of not requiring ionizing radiation, but with other limitations related to the general contraindications for MRI in some patient groups, such as subjects with pacemakers or with renal disease [215]. Conversely, the feasibility of MBF quantitation using CT angiography has been more recently demonstrated, but dosimetric problems appear a major hindrance to the widespread use of this modality [216, 217].
15. Future Developments and Challenges
The most interesting coming development for cardiac PET is the possibility to use an 18F labeled tracer for perfusion imaging. The compound 18F-flurpiridaz is already under advanced evaluation and phase 3 studies are ongoing [35, 218, 219]. The image quality is excellent and the results of the preliminary reports of phase 1 and phase 2 suggest that this tracer could be a major advance in perfusion PET [35, 218, 219]. The prolonged half-life of 18F opens the way for centers without on-site cyclotron to purchase flurpiridaz from a vendor, and therefore increases the number of potential users. It makes as well possible to inject the tracer during dynamic exercise using the standard equipment (either bicycle ergometer or treadmill), with the advantage of a physiologic stimulation instead of a pharmacologic one. On the other hand, this last procedure would preclude the quantitation of MBF, because of the impossibility to acquire a dynamic PET during tracer injection. Preliminary data suggest that a quantitative assessment of MBF using flurpiridaz is feasible and reliable [220–222]. Therefore, the availability of this tracer will open new perspectives, but the difficult choice between performing exercise PET and obtaining MBF quantitation has to be made.
The tracer problems that flurpiridaz could partly overcome are not the sole hurdles to the definitive acceptance of quantitative PET in the clinical routine. Most of these obstacles, however, are in general related to the disadvantages of PET versus SPECT for the clinical practice and are also influenced by the heterogeneous logistic, regulatory, and economic situations in the different countries [11, 25]. For instance, in USA the cardiologists are frequently directly involved in the execution of imaging techniques using radioactive tracers; this has driven a major growth in the number of nuclear cardiology exams and could increase the clinical demand of more effective approaches [223]. On the other hand, the large number of office-based practices could be an obstacle to the widespread use of PET, because this technique is certainly more demanding than SPECT and requires large investments [11]. Conversely, in Europe the execution of nuclear cardiology is usually limited to the nuclear medicine specialist, and most exams are performed in hospitals [224]. This could adversely influence the progress of cardiac PET because the interest for cardiology in nuclear medicine departments is often scarce, and the pressure of the clinical demand more limited [223]. However, the circumstance that PET scanners are already available in most nuclear medicine laboratories would facilitate the choice of shifting cardiology from SPECT to PET [25].
The specific problems concerning the quantitative approach are the more complex acquisition procedure, the greater computational burden, and the need of dedicated processing software. List mode acquisition capabilities for dynamic studies, however, are not a major problem in most modern PET scanners. The implementation of reliable programs for image processing and subsequent MBF quantitation directly in the PET scanner software by the producers would be highly desirable, and would drastically reduce the difficulties encountered so far, making the whole procedure more easily acceptable for the physician. Unfortunately, the demand is still too limited for obliging the vendors to make this choice. Ongoing studies, which compare the available programs and models in order to identify the most reliable, could be helpful for reducing the current uncertainty about the ideal approach.
16. Final Remarks
The usefulness of quantitative PET in heart disease has been convincingly proven. However, every new technique must find its place in the clinical scenario through the demonstration of its cost/effectiveness and of its superiority over already established methodologies. In case of quantitative PET, large prospective studies focused on defining its relative merits in a direct comparison with other approaches are still lacking. The problem involves more in general the role of noninvasive diagnostic techniques and some important trials are trying to clarify this issue, but reaching definitive evidence will still take time, and unfortunately the chosen approach for PET evaluation does not include quantitative measurements [225, 226]. On the other hand, there are no doubts that the technical and clinical prerequisites for a widespread use of quantitative PET have already been fulfilled. What is probably most important nowadays is that the cardiologic community becomes aware of the potential advantages of quantitative PET and start considering its use. The nuclear medicine physician, conversely, must improve the availability and reliability of quantitative cardiac PET, so that demand and offer may match and increase together. If always performed having in mind the patient benefit, this dialogue will certainly lead in a reasonable time to the realistic definition of the true role of quantitative cardiac PET.
References
- 1.Tillisch J, Brunken R, Marshall R, et al. Reversibility of cardial wall-motion abnormalities predicted by positron tomography. New England Journal of Medicine. 1986;314(14):884–888. doi: 10.1056/NEJM198604033141405. [DOI] [PubMed] [Google Scholar]
- 2.Gould KL, Goldstein RA, Mullani NA, et al. Noninvasive assessment of coronary stenoses by myocardial perfusion imaging during pharmacologic coronary vasodilation. VIII. Clinical feasibility of positron cardiac imaging without a cyclotron using generator-produced rubidium-82. Journal of the American College of Cardiology. 1986;7(4):775–789. doi: 10.1016/s0735-1097(86)80336-9. [DOI] [PubMed] [Google Scholar]
- 3.Stewart RE, Schwaiger M, Molina E, et al. Comparison of rubidium-82 positron emission tomography and thallium-201 SPECT imaging for detection of coronary artery disease. American Journal of Cardiology. 1991;67(16):1303–1310. doi: 10.1016/0002-9149(91)90456-u. [DOI] [PubMed] [Google Scholar]
- 4.Tamaki N, Yonekura Y, Senda M, et al. Value and limitation of stress thallium-201 single photon emission computed tomography: comparison with nitrogen-13 ammonia positron tomography. Journal of Nuclear Medicine. 1988;29(7):1181–1188. [PubMed] [Google Scholar]
- 5.Walsh MN, Bergmann SR, Steele RL, et al. Delineation of impaired regional myocardial perfusion by positron emission tomography with H2 15O. Circulation. 1988;78(3):612–620. doi: 10.1161/01.cir.78.3.612. [DOI] [PubMed] [Google Scholar]
- 6.Demer LL, Gould KL, Goldstein RA, et al. Assessment of coronary artery disease severity by positron emmison tomography. Comparison with quantitative arteriography in 193 patients. Circulation. 1989;79(4):825–835. doi: 10.1161/01.cir.79.4.825. [DOI] [PubMed] [Google Scholar]
- 7.Brunken R, Schwaiger M, Grover-McKay M. Positron emission tomography detects tissue metabolic activity in myocardial segments with persistent thallium perfusion defects. Journal of the American College of Cardiology. 1987;10(3):557–567. doi: 10.1016/s0735-1097(87)80198-5. [DOI] [PubMed] [Google Scholar]
- 8.Smanio PEP, Watson DD, Segalla DL, Vinson EL, Smith WH, Beller GA. Value of gating of technetium-99m sestamibi single-photon emission computed tomographic imaging. Journal of the American College of Cardiology. 1997;30(7):1687–1692. doi: 10.1016/s0735-1097(97)00363-x. [DOI] [PubMed] [Google Scholar]
- 9.Sciagrà R. The expanding role of left ventricular functional assessment using gated myocardial perfusion SPECT: the supporting actor is stealing the scene. European Journal of Nuclear Medicine and Molecular Imaging. 2007;34(7):1107–1122. doi: 10.1007/s00259-007-0405-6. [DOI] [PubMed] [Google Scholar]
- 10.Bateman TM. Advantages and disadvantages of PET and SPECT in a busy clinical practice. Journal of Nuclear Cardiology. 2012;19:S3–S11. doi: 10.1007/s12350-011-9490-9. [DOI] [PubMed] [Google Scholar]
- 11.Heller GV. Practical issues regarding the incorporation of PET into a busy SPECT practice. Journal of Nuclear Cardiology. 2012;19:S12–S18. doi: 10.1007/s12350-011-9493-6. [DOI] [PubMed] [Google Scholar]
- 12.Garcia EV. Physical attributes, limitations, and future potential for PET and SPECT. Journal of Nuclear Cardiology. 2012;19:S19–S29. doi: 10.1007/s12350-011-9488-3. [DOI] [PubMed] [Google Scholar]
- 13.Ter-Pogossian MM. Instrumentation for cardiac positron emission tomography: background and historical perspective. In: Bergmann SR, Sobel BE, editors. Positron Emission Tomography of the Heart. Mount Kisco, NY, USA: Futura Publishing; 1992. pp. 1–12. [Google Scholar]
- 14.Beyer T, Townsend DW, Brun T, et al. A combined PET/CT scanner for clinical oncology. Journal of Nuclear Medicine. 2000;41(8):1369–1379. [PubMed] [Google Scholar]
- 15.DeBenedetti S, Cowan CE, Konneker WR, Primakoff H. On the angular distribution of two-photon annihilation radiation. Physical Review. 1950;77(2):205–212. [Google Scholar]
- 16.Phelps ME, Hoffman EJ, Huang SC, Ter Pogossian MM. Effect of positron range on spatial resolution. Journal of Nuclear Medicine. 1975;16(7):649–652. [PubMed] [Google Scholar]
- 17.Bergstrom M, Eriksson L, Bohm C. Correction for scattered radiation in a ring detector positron camera by integral transformation of the projections. Journal of Computer Assisted Tomography. 1983;7(1):42–50. doi: 10.1097/00004728-198302000-00008. [DOI] [PubMed] [Google Scholar]
- 18.Hoffman EJ, Huang SC, Phelps ME, Kuhl DE. Quantitation in positron emission computed tomography: 4. Effect of accidental coincidences. Journal of Computer Assisted Tomography. 1981;5(3):391–400. doi: 10.1097/00004728-198106000-00015. [DOI] [PubMed] [Google Scholar]
- 19.Ollinger JM. Detector efficiency and Compton scatter in fully 3D PET. IEEE Transactions on Nuclear Science. 1995;42(4):1168–1173. [Google Scholar]
- 20.Daube-Witherspoon ME, Karp JS, Casey ME, et al. PET performance measurements using the NEMA NU 2-2001 standard. Journal of Nuclear Medicine. 2002;43(10):1398–1409. [PubMed] [Google Scholar]
- 21.Lautamäki R, Brown TLY, Merrill J, Bengel FM. CT-based attenuation correction in 82Rb-myocardial perfusion PET-CT: incidence of misalignment and effect on regional tracer distribution. European Journal of Nuclear Medicine and Molecular Imaging. 2008;35(2):305–310. doi: 10.1007/s00259-007-0607-y. [DOI] [PubMed] [Google Scholar]
- 22.Gould KL, Pan T, Loghin C, Johnson NP, Guha A, Sdringola S. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections. Journal of Nuclear Medicine. 2007;48(7):1112–1121. doi: 10.2967/jnumed.107.039792. [DOI] [PubMed] [Google Scholar]
- 23.Martinez-Möller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from misaligned CT in cardiac perfusion PET/CT studies: frequency, effects, and potential solutions. Journal of Nuclear Medicine. 2007;48(2):188–193. [PubMed] [Google Scholar]
- 24.Martinez-Möller A, Zikic D, Botnar RM, et al. Dual cardiac-respiratory gated PET: implementation and results from a feasibility study. European Journal of Nuclear Medicine and Molecular Imaging. 2007;34(9):1447–1454. doi: 10.1007/s00259-007-0374-9. [DOI] [PubMed] [Google Scholar]
- 25.Le Guludec D, Lautamäki R, Knuuti J, Bax JJ, Bengel FM. Present and future of clinical cardiovascular PET imaging in Europe—a position statement by the European Council of Nuclear Cardiology (ECNC) European Journal of Nuclear Medicine and Molecular Imaging. 2008;35(9):1709–1724. doi: 10.1007/s00259-008-0859-1. [DOI] [PubMed] [Google Scholar]
- 26.Schepis T, Gaemperli O, Treyer V, et al. Absolute quantification of myocardial blood flow with 13N- ammonia and 3-dimensional PET. Journal of Nuclear Medicine. 2007;48(11):1783–1789. doi: 10.2967/jnumed.107.044099. [DOI] [PubMed] [Google Scholar]
- 27.Conti M. State of the art and challenges of time-of-flight PET. Physica Medica. 2009;25(1):1–11. doi: 10.1016/j.ejmp.2008.10.001. [DOI] [PubMed] [Google Scholar]
- 28.Chen GP, Branch KR, Alessio AM, et al. Effect of reconstruction algorithms on myocardial blood flow measurement with 13N-ammonia PET. Journal of Nuclear Medicine. 2007;48(8):1259–1265. doi: 10.2967/jnumed.106.038232. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Hove JD, Rasmussen R, Freiberg J, Holm S, Kelbæk H, Kofoed KE. Clinical evaluation of iterative reconstruction (ordered-subset expectation maximization) in dynamic positron emission tomography: quantitative effects on kinetic modeling with N-13 ammonia in healthy subjects. Journal of Nuclear Cardiology. 2008;15(4):530–534. doi: 10.1016/j.nuclcard.2008.02.034. [DOI] [PubMed] [Google Scholar]
- 30.Gropler RJ, Siegel BA, Geltman EM. Myocardial uptake of carbon-11-acetate as an indirect estimate of regional myocardial blood flow. Journal of Nuclear Medicine. 1991;32(2):245–251. [PubMed] [Google Scholar]
- 31.Van Den Hoff J, Burchert W, Wolpers HG, Meyer GJ, Hundeshagen H. A kinetic model for cardiac PET with [1-carbon-11]-acetate. Journal of Nuclear Medicine. 1996;37(3):521–529. [PubMed] [Google Scholar]
- 32.Van den Hoff J, Burchert W, Börner AR, et al. [1-11C]acetate as a quantitative perfusion tracer in myocardial PET. Journal of Nuclear Medicine. 2001;42(8):1174–1182. [PubMed] [Google Scholar]
- 33.Timmer SAJ, Lubberink M, Germans T, et al. Potential of [11C]acetate for measuring myocardial blood flow: studies in normal subjects and patients with hypertrophic cardiomyopathy. Journal of Nuclear Cardiology. 2010;17(2):264–275. doi: 10.1007/s12350-009-9181-y. [DOI] [PubMed] [Google Scholar]
- 34.Hur MG, Park JH, Kim IJ, et al. Design and evaluation of a carbon-11 gas target having auxiliary cooling fins in the target cavity for the stable production of carbon-11. Nuclear Instruments and Methods in Physics Research, Section A. 2011;655:108–110. [Google Scholar]
- 35.Maddahi J. Properties of an ideal PET perfusion tracer: new PET tracer cases and data. Journal of Nuclear Cardiology. 2012;19:S30–S37. doi: 10.1007/s12350-011-9491-8. [DOI] [PubMed] [Google Scholar]
- 36.Ratib O, Phelps ME, Huang SC. Positron tomography with deoxyglucose for estimating local myocardial glucose metabolism. Journal of Nuclear Medicine. 1982;23(7):577–586. [PubMed] [Google Scholar]
- 37.Kudo T. Metabolic imaging using PET. European Journal of Nuclear Medicine and Molecular Imaging. 2007;34(supplement 1):S49–S61. doi: 10.1007/s00259-007-0440-3. [DOI] [PubMed] [Google Scholar]
- 38.Melon P, Schwaiger M. Imaging of metabolism and autonomic innervation of the heart by positron emission tomography. European Journal of Nuclear Medicine. 1992;19(6):453–464. doi: 10.1007/BF00177375. [DOI] [PubMed] [Google Scholar]
- 39.Bergmann SR, Fox KAA, Rand AL. Quantification of regional myocardial blood flow in vivo with H2 15O. Circulation. 1984;70(4):724–733. doi: 10.1161/01.cir.70.4.724. [DOI] [PubMed] [Google Scholar]
- 40.Iida H, Kanno I, Takahashi A, et al. Measurement of absolute myocardial blood flow with H2 15O and dynamic positron-emission tomography. Strategy for quantification in relation to the partial-volume effect. Circulation. 1988;78(1):104–115. doi: 10.1161/01.cir.78.1.104. [DOI] [PubMed] [Google Scholar]
- 41.Bergmann SR, Herrero P, Markham J, Weinheimer CJ, Walsh MN. Noninvasive quantitation of myocardial blood flow in human subjects with oxygen-15-labeled water and positron emission tomography. Journal of the American College of Cardiology. 1989;14(3):639–652. doi: 10.1016/0735-1097(89)90105-8. [DOI] [PubMed] [Google Scholar]
- 42.Hermansen F, Rosen SD, Fath-Ordoubadi F, et al. Measurement of myocardial blood flow with oxygen-15 labelled water: comparison of different administration protocols. European Journal of Nuclear Medicine. 1998;25(7):751–759. doi: 10.1007/s002590050279. [DOI] [PubMed] [Google Scholar]
- 43.Kaufmann PA, Camici PG. Myocardial blood flow measurement by PET: technical aspects and clinical applications. Journal of Nuclear Medicine. 2005;46:75–88. [PubMed] [Google Scholar]
- 44.Machac J. Cardiac positron emission tomography imaging. Seminars in Nuclear Medicine. 2005;35(1):17–36. doi: 10.1053/j.semnuclmed.2004.09.002. [DOI] [PubMed] [Google Scholar]
- 45.Bellina CR, Parodi O, Camici P, et al. Simultaneous in vitro and in vivo validation of nitrogen-13-ammonia for the assessment of regional myocardial blood flow. Journal of Nuclear Medicine. 1990;31(8):1335–1343. [PubMed] [Google Scholar]
- 46.Hutchins GD, Schwaiger M, Rosenspire KC, Krivokapich J, Schelbert H, Kuhl DE. Noninvasive quantification of regional blood flow in the human heart using N-13 ammonia and dynamic positron emission tomographic imaging. Journal of the American College of Cardiology. 1990;15(5):1032–1042. doi: 10.1016/0735-1097(90)90237-j. [DOI] [PubMed] [Google Scholar]
- 47.Krivokapich J, Smith GT, Huang SC, et al. 13N ammonia myocardial imaging at rest and with exercise in normal volunteers. Quantification of absolute myocardial perfusion with dynamic positron emission tomography. Circulation. 1989;80(5):1328–1337. doi: 10.1161/01.cir.80.5.1328. [DOI] [PubMed] [Google Scholar]
- 48.Muzik O, Beanlands RSB, Hutchins GD, Mangner TJ, Nguyen N, Schwaiger M. Validation of nitrogen-13-ammonia tracer kinetic model for quantification of myocardial blood flow using PET. Journal of Nuclear Medicine. 1993;34(1):83–91. [PubMed] [Google Scholar]
- 49.Choi Y, Huang SC, Hawkins RA, et al. A simplified method for quantification of myocardial blood flow using nitrogen-13-ammonia and dynamic PET. Journal of Nuclear Medicine. 1993;34(3):488–497. [PubMed] [Google Scholar]
- 50.DeGrado TR, Hanson MW, Turkington TG, et al. Estimation of myocardial blood flow for longitudinal studies with13N-labeled ammonia and positron emission tomography. Journal of Nuclear Cardiology. 1996;3(6):494–507. doi: 10.1016/s1071-3581(96)90059-8. [DOI] [PubMed] [Google Scholar]
- 51.Herrero P, Markham J, Shelton ME, Weinheimer CJ, Bergmann SR. Noninvasive quantification of regional myocardial perfusion with rubidium-82 and positron emission tomography. Exploration of a mathematical model. Circulation. 1990;82(4):1377–1386. doi: 10.1161/01.cir.82.4.1377. [DOI] [PubMed] [Google Scholar]
- 52.Nitzsche EU, Choi Y, Czernin J, Hoh CK, Huang SC, Schelbert HR. Noninvasive quantification of myocardial blood flow in humans: a direct comparison of the [13N] ammonia and the [15O]water techniques. Circulation. 1996;93(11):2000–2006. doi: 10.1161/01.cir.93.11.2000. [DOI] [PubMed] [Google Scholar]
- 53.Husmann L, Wiegand M, Valenta I, et al. Diagnostic accuracy of myocardial perfusion imaging with single photon emission computed tomography and positron emission tomography: a comparison with coronary angiography. International Journal of Cardiovascular Imaging. 2008;24(5):511–518. doi: 10.1007/s10554-007-9288-7. [DOI] [PubMed] [Google Scholar]
- 54.Herrero P, Markham J, Shelton ME, Bergmann SR. Implementation and evaluation of a two-compartment model for quantification of myocardial perfusion with rubidium-82 and positron emission tomography. Circulation Research. 1992;70(3):496–507. doi: 10.1161/01.res.70.3.496. [DOI] [PubMed] [Google Scholar]
- 55.Yoshida K, Mullani N, Gould KL. Coronary flow and flow reserve by PET simplified for clinical applications using rubidium-82 or nitrogen-13-ammonia. Journal of Nuclear Medicine. 1996;37(10):1701–1712. [PubMed] [Google Scholar]
- 56.Lin JW, Sciacca RR, Chou RL, Laine AF, Bergmann SR. Quantification of myocardial perfusion in human subjects using 82Rb and wavelet-based noise reduction. Journal of Nuclear Medicine. 2001;42(2):201–208. [PubMed] [Google Scholar]
- 57.El Fakhri G, Sitek A, Guérin B, Kijewski MF, Di Carli MF, Moore SC. Quantitative dynamic cardiac 82Rb PET using generalized factor and compartment analyses. Journal of Nuclear Medicine. 2005;46(8):1264–1271. [PubMed] [Google Scholar]
- 58.Lortie M, Beanlands RSB, Yoshinaga K, Klein R, DaSilva JN, DeKemp RA. Quantification of myocardial blood flow with 82Rb dynamic PET imaging. European Journal of Nuclear Medicine and Molecular Imaging. 2007;34(11):1765–1774. doi: 10.1007/s00259-007-0478-2. [DOI] [PubMed] [Google Scholar]
- 59.Klein R, Beanlands RSB, DeKemp RA. Quantification of myocardial blood flow and flow reserve: technical aspects. Journal of Nuclear Cardiology. 2010;17(4):555–570. doi: 10.1007/s12350-010-9256-9. [DOI] [PubMed] [Google Scholar]
- 60.Groves AM, Speechly-Dick ME, Dickson JC, et al. Cardiac 82rubidium PET/CT: initial European experience. European Journal of Nuclear Medicine and Molecular Imaging. 2007;34(12):1965–1972. doi: 10.1007/s00259-007-0537-8. [DOI] [PubMed] [Google Scholar]
- 61.Choi Y, Huang SC, Hawkins RA, et al. Quantification of myocardial blood flow using 13N-ammonia and PET: comparison of tracer models. Journal of Nuclear Medicine. 1999;40(6):1045–1055. [PubMed] [Google Scholar]
- 62.Hoffman EJ, Huang SC, Plummer D, Phelps ME. Quantitation in positron emission computed tomography: 6. Effect of nonuniform resolution. Journal of Computer Assisted Tomography. 1982;6(5):987–999. doi: 10.1097/00004728-198210000-00022. [DOI] [PubMed] [Google Scholar]
- 63.Hove JD, Gambhir SS, Kofoed KF, Freiberg J, Kelbæk H. Quantitation of the regional blood flow in the interventricular septum using positron emission tomography and nitrogen-13 ammonia. European Journal of Nuclear Medicine and Molecular Imaging. 2003;30(1):109–116. doi: 10.1007/s00259-002-1014-z. [DOI] [PubMed] [Google Scholar]
- 64.Hutchins GD, Caraher JM, Raylman RR. A region of interest strategy for minimizing resolution distortions in quantitative myocardial PET studies. Journal of Nuclear Medicine. 1992;33(6):1243–1250. [PubMed] [Google Scholar]
- 65.Coxson PG. Consequences of using a simplified kinetic model for dynamic PET data. Journal of Nuclear Medicine. 1997;38(4):660–667. [PubMed] [Google Scholar]
- 66.Klein R, Renaud JM, Ziadi MC, et al. Intra-and inter-operator repeatability of myocardial blood flow and myocardial flow reserve measurements using rubidium-82 pet and a highly automated analysis program. Journal of Nuclear Cardiology. 2010;17(4):600–616. doi: 10.1007/s12350-010-9225-3. [DOI] [PubMed] [Google Scholar]
- 67.Nesterov SV, Han C, Mäki M, et al. Myocardial perfusion quantification with 15O-labeled water PET: high reproducibility of the new cardiac analysis software (CarimasTM) European Journal of Nuclear Medicine and Molecular Imaging. 2009;36:1594–1602. doi: 10.1007/s00259-009-1143-8. [DOI] [PubMed] [Google Scholar]
- 68.Knešaurek K, Zhang Z. Repeatability of regional myocardial blood flow calculation in 82Rb PET imaging. BMC Medical Physics. 2009;9(1, article 2) doi: 10.1186/1756-6649-9-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.El Fakhri G, Kardan A, Sitek A, et al. Reproducibility and accuracy of quantitative myocardial blood flow assessment with 82Rb PET: comparison with 13N-ammonia PET. Journal of Nuclear Medicine. 2009;50(7):1062–1071. doi: 10.2967/jnumed.104.007831. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Slomka PJ, Alexanderson E, Jácome R, et al. Comparison of clinical tools for measurements of regional stress and rest myocardial blood flow assessed with 13N-ammonia PET/CT. Journal of Nuclear Medicine. 2012;53:171–181. doi: 10.2967/jnumed.111.095398. [DOI] [PubMed] [Google Scholar]
- 71.Sawada S, Muzik O, Beanlands RSB, Wolfe E, Hutchins GD, Schwaiger M. Interobserver and interstudy variability of myocardial blood flow and flow-reserve measurements with nitrogen 13 ammonia-labeled positron emission tomography. Journal of Nuclear Cardiology. 1995;2(5):413–422. doi: 10.1016/s1071-3581(05)80029-7. [DOI] [PubMed] [Google Scholar]
- 72.Nagamachi S, Czernin J, Kim AS, et al. Reproducibility of measurements of regional resting and hyperemic myocardial blood flow assessed with PET. Journal of Nuclear Medicine. 1996;37(10):1626–1631. [PubMed] [Google Scholar]
- 73.Kaufmann PA, Gnecchi-Ruscone T, Yap JT, Rimoldi O, Camici PG. Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET. Journal of Nuclear Medicine. 1999;40(11):1848–1856. [PubMed] [Google Scholar]
- 74.Jagathesan R, Kaufmann PA, Rosen SD, et al. Assessment of the long-term reproducibility of baseline and dobutamine-induced myocardial blood flow in patients with stable coronary artery disease. Journal of Nuclear Medicine. 2005;46(2):212–219. [PubMed] [Google Scholar]
- 75.Manabe O, Yoshinaga K, Katoh C, Naya M, DeKemp RA, Tamaki N. Repeatability of rest and hyperemic myocardial blood flow measurements with82Rb dynamic PET. Journal of Nuclear Medicine. 2009;50(1):68–71. doi: 10.2967/jnumed.108.055673. [DOI] [PubMed] [Google Scholar]
- 76.Patlak CS, Blasberg RG, Fenstermacher JD. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. Journal of Cerebral Blood Flow and Metabolism. 1983;3(1):1–7. doi: 10.1038/jcbfm.1983.1. [DOI] [PubMed] [Google Scholar]
- 77.Morita K, Katoh C, Yoshinaga K, et al. Quantitative analysis of myocardial glucose utilization in patients with left ventricular dysfunction by means of 18F-FDG dynamic positron tomography and three-compartment analysis. European Journal of Nuclear Medicine and Molecular Imaging. 2005;32(7):806–812. doi: 10.1007/s00259-004-1743-2. [DOI] [PubMed] [Google Scholar]
- 78.Masci PG, Marinelli M, Piacenti M, et al. Myocardial structural, perfusion, and metabolic correlates of left bundle branch block mechanical derangement in patients with dilated cardiomyopathy: a tagged cardiac magnetic resonance and positron emission tomography study. Circulation: Cardiovascular Imaging. 2010;3(4):482–490. doi: 10.1161/CIRCIMAGING.109.934638. [DOI] [PubMed] [Google Scholar]
- 79.Araujo LI, Lammertsma AA, Rhodes CG, et al. Noninvasive quantification of regional myocardial blood flow in coronary artery disease with oxygen-15-labeled carbon dioxide inhalation and positron emission tomography. Circulation. 1991;83(3):875–885. doi: 10.1161/01.cir.83.3.875. [DOI] [PubMed] [Google Scholar]
- 80.Geltman EW, Henes G, Senneff MJ, Sobel BE, Bergmann SR. Increased myocardial perfusion at rest and diminished perfusion reserve in patients with angina angiographically normal coronary arteries. Journal of the American College of Cardiology. 1990;16(3):586–595. doi: 10.1016/0735-1097(90)90347-r. [DOI] [PubMed] [Google Scholar]
- 81.Sambuceti G, Parodi O, Marcassa C, et al. Alteration in regulation of myocardial blood flow in one-vessel coronary artery disease determined by positron emission tomography. American Journal of Cardiology. 1993;72(7):538–543. doi: 10.1016/0002-9149(93)90348-g. [DOI] [PubMed] [Google Scholar]
- 82.Chan SY, Brunken RC, Czernin J, et al. Comparison of maximal myocardial blood flow during adenosine infusion with that of intravenous dipyridamole in normal men. Journal of the American College of Cardiology. 1992;20(4):979–984. doi: 10.1016/0735-1097(92)90201-w. [DOI] [PubMed] [Google Scholar]
- 83.Merlet P, Mazoyer B, Hittinger L, et al. Assessment of coronary reserve in man: comparison between positron emission tomography with oxygen-15-labeled water and intracoronary Doppler technique. Journal of Nuclear Medicine. 1993;34(11):1899–1904. [PubMed] [Google Scholar]
- 84.Muzik O, Beanlands R, Wolfe E, Hutchins GD, Schwaiger M. Automated region definition for cardiac nitrogen-13-ammonia PET imaging. Journal of Nuclear Medicine. 1993;34(2):336–344. [PubMed] [Google Scholar]
- 85.Czernin J, Muller P, Chan S, et al. Influence of age and hemodynamics on myocardial blood flow and flow reserve. Circulation. 1993;88(1):62–69. doi: 10.1161/01.cir.88.1.62. [DOI] [PubMed] [Google Scholar]
- 86.Krivokapich J, Czernin J, Schelbert HR. Dobutamine positron emission tomography: absolute quantitation of rest and dobutamine myocardial blood flow and correlation with cardiac work and percent diameter stenosis in patients with and without coronary artery disease. Journal of the American College of Cardiology. 1996;28(3):565–572. doi: 10.1016/0735-1097(96)00205-7. [DOI] [PubMed] [Google Scholar]
- 87.Chareonthaitawee P, Kaufmann PA, Rimoldi O, Camici PG. Heterogeneity of resting and hyperemic myocardial blood flow in healthy humans. Cardiovascular Research. 2001;50(1):151–161. doi: 10.1016/s0008-6363(01)00202-4. [DOI] [PubMed] [Google Scholar]
- 88.Senneff MJ, Geltman EM, Bergmann SR, Hartman J. Noninvasive delineation of the effects of moderate aging on myocardial perfusion. Journal of Nuclear Medicine. 1991;32(11):2037–2042. [PubMed] [Google Scholar]
- 89.Camici PG, Gropler RJ, Jones T, et al. The impact of myocardial blood flow quantitation with PET on the understanding of cardiac diseases. European Heart Journal. 1996;17(1):25–34. doi: 10.1093/oxfordjournals.eurheartj.a014687. [DOI] [PubMed] [Google Scholar]
- 90.Duvernoy CS, Meyer C, Seifert-Klauss V, et al. Gender differences in myocardial blood flow dynamics lipid profile and hemodynamic effects. Journal of the American College of Cardiology. 1999;33(2):463–470. doi: 10.1016/s0735-1097(98)00575-0. [DOI] [PubMed] [Google Scholar]
- 91.Sdringola S, Johnson NP, Kirkeeide RL, Cid E, Gould KL. Impact of unexpected factors on quantitative myocardial perfusion and coronary flow reserve in young, asymptomatic volunteers. JACC: Cardiovascular Imaging. 2011;4(4):402–412. doi: 10.1016/j.jcmg.2011.02.008. [DOI] [PubMed] [Google Scholar]
- 92.Cannon RO, Camici PG, Epstein SE. Pathophysiological dilemma of syndrome X. Circulation. 1992;85(3):883–892. doi: 10.1161/01.cir.85.3.883. [DOI] [PubMed] [Google Scholar]
- 93.Camici PG, Gistri R, Lorenzoni R, et al. Coronary reserve and exercise ECG in patients with chest pain and normal coronary angiograms. Circulation. 1992;86(1):179–186. doi: 10.1161/01.cir.86.1.179. [DOI] [PubMed] [Google Scholar]
- 94.Rosen SD, Uren NG, Kaski JC, Tousoulis D, Davies GJ, Camici PG. Coronary vasodilator reserve, pain perception, and sex in patients with syndrome X. Circulation. 1994;90(1):50–60. doi: 10.1161/01.cir.90.1.50. [DOI] [PubMed] [Google Scholar]
- 95.Bøttcher M, Bøtker HE, Sonne H, Nielsen TT, Czernin J. Endothelium-dependent and -independent perfusion reserve and the effect of L-arginine on myocardial perfusion in patients with syndrome X. Circulation. 1999;99(14):1795–1801. doi: 10.1161/01.cir.99.14.1795. [DOI] [PubMed] [Google Scholar]
- 96.Graf S, Khorsand A, Gwechenberger M, et al. Myocardial perfusion in patients with typical chest pain and normal angiogram. European Journal of Clinical Investigation. 2006;36(5):326–332. doi: 10.1111/j.1365-2362.2006.01635.x. [DOI] [PubMed] [Google Scholar]
- 97.Graf S, Khorsand A, Gwechenberger M, et al. Typical chest pain and normal coronary angiogram: cardiac risk factor analysis versus PET for detection of microvascular disease. Journal of Nuclear Medicine. 2007;48(2):175–181. [PubMed] [Google Scholar]
- 98.Gistri R, Ebert AG, Palombo C, Marabotti C, Choudhury L, Camici PG. Effect of blood pressure lowering on coronary vasodilator reserve in arterial hypertension. Cardiovascular Drugs and Therapy. 1994;8(1):169–171. doi: 10.1007/BF00877107. [DOI] [PubMed] [Google Scholar]
- 99.Higuchi T, Abletshauser C, Nekolla SG, Schwaiger M, Bengel FM. Effect of the angiotensin receptor blocker valsartan on coronary microvascular flow reserve in moderately hypertensive patients with stable coronary artery disease. Microcirculation. 2007;14(8):805–812. doi: 10.1080/10739680701410827. [DOI] [PubMed] [Google Scholar]
- 100.Naya M, Tsukamoto T, Morita K, et al. Olmesartan, but not amlodipine, improves endothelium-dependent coronary dilation in hypertensive patients. Journal of the American College of Cardiology. 2007;50(12):1144–1149. doi: 10.1016/j.jacc.2007.06.013. [DOI] [PubMed] [Google Scholar]
- 101.Choudhury L, Rosen SD, Patel D, Nihoyannopoulos P, Camici PG. Coronary vasodilator reserve in primary and secondary left ventricular hypertrophy. A study with positron emission tomography. European Heart Journal. 1997;18(1):108–116. doi: 10.1093/oxfordjournals.eurheartj.a015090. [DOI] [PubMed] [Google Scholar]
- 102.Olsen MH, Wachtell K, Meyer C, et al. Association between vascular dysfunction and reduced myocardial flow reserve in patients with hypertension: a LIFE substudy. Journal of Human Hypertension. 2004;18(6):445–452. doi: 10.1038/sj.jhh.1001716. [DOI] [PubMed] [Google Scholar]
- 103.Laine H, Raitakari OT, Niinikoski H, et al. Early impairment of coronary flow reserve in young men with borderline hypertension. Journal of the American College of Cardiology. 1998;32(1):147–153. doi: 10.1016/s0735-1097(98)00222-8. [DOI] [PubMed] [Google Scholar]
- 104.Pitkänen OP, Nuutila P, Raitakari OT, et al. Coronary flow reserve is reduced in young men with IDDM. Diabetes. 1998;47(2):248–254. doi: 10.2337/diab.47.2.248. [DOI] [PubMed] [Google Scholar]
- 105.Yokoyama I, Momomura SI, Ohtake T, et al. Reduced myocardial flow reserve in non-insulin-dependent diabetes mellitus. Journal of the American College of Cardiology. 1997;30(6):1472–1477. doi: 10.1016/s0735-1097(97)00327-6. [DOI] [PubMed] [Google Scholar]
- 106.Di Carli MF, Bianco-Batlles D, Landa ME, et al. Effects of autonomic neuropathy on coronary blood flow in patients with diabetes mellitus. Circulation. 1999;100(8):813–819. doi: 10.1161/01.cir.100.8.813. [DOI] [PubMed] [Google Scholar]
- 107.Momose M, Abletshauser C, Neverve J, et al. Dysregulation of coronary microvascular reactivity in asymptomatic patients with type 2 diabetes mellitus. European Journal of Nuclear Medicine. 2002;29(12):1675–1679. doi: 10.1007/s00259-002-0977-0. [DOI] [PubMed] [Google Scholar]
- 108.Di Carli MF, Afonso L, Campisi R, et al. Coronary vascular dysfunction in premenopausal women with diabetes mellitus. American Heart Journal. 2002;144(4):711–718. [PubMed] [Google Scholar]
- 109.Di Carli MF, Janisse J, Grunberger G, Ager J. Role of chronic hyperglycemia in the pathogenesis of coronary microvascular dysfunction in diabetes. Journal of the American College of Cardiology. 2003;41(8):1387–1393. doi: 10.1016/s0735-1097(03)00166-9. [DOI] [PubMed] [Google Scholar]
- 110.Quiñones MJ, Hernandez-Pampaloni M, Schelbert H, et al. Coronary vasomotor abnormalities in insulin-resistant individuals. Annals of Internal Medicine. 2004;140(9):700–708. doi: 10.7326/0003-4819-140-9-200405040-00009. [DOI] [PubMed] [Google Scholar]
- 111.Prior JO, Quiñones MJ, Hernandez-Pampaloni M, et al. Coronary circulatory dysfunction in insulin resistance, impaired glucose tolerance, and type 2 diabetes mellitus. Circulation. 2005;111(18):2291–2298. doi: 10.1161/01.CIR.0000164232.62768.51. [DOI] [PubMed] [Google Scholar]
- 112.Bengel FM, Abletshausert C, Neverve J, et al. Effects of nateglinide on myocardial microvascular reactivity in Type 2 diabetes mellitus—a randomized study using positron emission tomography. Diabetic Medicine. 2005;22(2):158–163. doi: 10.1111/j.1464-5491.2004.01371.x. [DOI] [PubMed] [Google Scholar]
- 113.Lautamäki R, Airaksinen KEJ, Seppänen M, et al. Insulin improves myocardial blood flow in patients with type 2 diabetes and coronary artery disease. Diabetes. 2006;55(2):511–516. doi: 10.2337/diabetes.55.02.06.db05-1023. [DOI] [PubMed] [Google Scholar]
- 114.Schindler TH, Facta AD, Prior JO, et al. Improvement in coronary vascular dysfunction produced with euglycaemic control in patients with type 2 diabetes. Heart. 2007;93(3):345–349. doi: 10.1136/hrt.2006.094128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Schindler TH, Facta AD, Prior JO, et al. Structural alterations of the coronary arterial wall are associated with myocardial flow heterogeneity in type 2 diabetes mellitus. European Journal of Nuclear Medicine and Molecular Imaging. 2009;36(2):219–229. doi: 10.1007/s00259-008-0885-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 116.Schindler TH, Cadenas J, Facta AD, et al. Improvement in coronary endothelial function is independently associated with a slowed progression of coronary artery calcification in type 2 diabetes mellitus. European Heart Journal. 2009;30(24):3064–3073. doi: 10.1093/eurheartj/ehp482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 117.Pitkänen OP, Raitakari OT, Niinikoski H, et al. Coronary flow reserve is impaired in young men with familial hypercholesterolemia. Journal of the American College of Cardiology. 1996;28:1705–1711. doi: 10.1016/S0735-1097(96)00376-2. [DOI] [PubMed] [Google Scholar]
- 118.Pitkänen OP, Nuutila P, Raitakari OT, et al. Coronary flow reserve in young men with familial combined hyperlipidemia. Circulation. 1999;99(13):1678–1684. doi: 10.1161/01.cir.99.13.1678. [DOI] [PubMed] [Google Scholar]
- 119.Yokoyama I, Ohtake T, Momomura SI, Nishikawa J, Sasaki Y, Omata M. Reduced coronary flow reserve in hypercholesterolemic patients without overt coronary stenosis. Circulation. 1996;94(12):3232–3238. doi: 10.1161/01.cir.94.12.3232. [DOI] [PubMed] [Google Scholar]
- 120.Dayanikli F, Grambow D, Muzik O, Mosca L, Rubenfire M, Schwaiger M. Early detection of abnormal coronary flow reserve in asymptomatic men at high risk for coronary artery disease using positron emission tomography. Circulation. 1994;90(2):808–817. doi: 10.1161/01.cir.90.2.808. [DOI] [PubMed] [Google Scholar]
- 121.Kaufmann PA, Gnecchi-Ruscone T, Schäfers KP, Lüscher TF, Camici PG. Low density lipoprotein cholesterol and coronary microvascular dysfunction in hypercholesterolemia. Journal of the American College of Cardiology. 2000;36(1):103–109. doi: 10.1016/s0735-1097(00)00697-5. [DOI] [PubMed] [Google Scholar]
- 122.Czernin J, Barnard RJ, Sun KT, et al. Effect of short-term cardiovascular conditioning and low-fat diet on myocardial blood flow and flow reserve. Circulation. 1995;92(2):197–204. doi: 10.1161/01.cir.92.2.197. [DOI] [PubMed] [Google Scholar]
- 123.Janatuinen T, Laaksonen R, Vesalainen R, et al. Effect of lipid-lowering therapy with pravastatin on myocardial blood flow in young mildly hypercholesterolemic adults. Journal of Cardiovascular Pharmacology. 2001;38(4):561–568. doi: 10.1097/00005344-200110000-00008. [DOI] [PubMed] [Google Scholar]
- 124.Yokoyama I, Yonekura K, Inoue Y, Ohtomo K, Nagai R. Long-term effect of simvastatin on the improvement of impaired myocardial flow reserve in patients with familial hypercholesterolemia without gender variance. Journal of Nuclear Cardiology. 2001;8(4):445–451. doi: 10.1067/mnc.2001.115517. [DOI] [PubMed] [Google Scholar]
- 125.Guethlin M, Kasel AM, Coppenrath K, Ziegler S, Delius W, Schwaiger M. Delayed response of myocardial flow reserve to lipid-lowering therapy with fluvastatin. Circulation. 1999;99(4):475–481. doi: 10.1161/01.cir.99.4.475. [DOI] [PubMed] [Google Scholar]
- 126.Huggins GS, Pasternak RC, Alpert NM, Fischman AJ, Gewirtz H. Effects of short-term treatment of hyperlipidemia on coronary vasodilator function and myocardial perfusion in regions having substantial impairment of baseline dilator reverse. Circulation. 1998;98(13):1291–1296. doi: 10.1161/01.cir.98.13.1291. [DOI] [PubMed] [Google Scholar]
- 127.Wielepp P, Baller D, Gleichmann U, Pulawski E, Horstkotte D, Burchert W. Beneficial effects of atorvastatin on myocardial regions with initially low vasodilatory capacity at various stages of coronary artery disease. European Journal of Nuclear Medicine and Molecular Imaging. 2005;32(12):1371–1377. doi: 10.1007/s00259-005-1828-6. [DOI] [PubMed] [Google Scholar]
- 128.Yokoyama I, Momomura SI, Ohtake T, et al. Improvement of impaired myocardial vasodilatation due to diffuse coronary atherosclerosis in hypercholesterolemics after lipid-lowering therapy. Circulation. 1999;100(2):117–122. doi: 10.1161/01.cir.100.2.117. [DOI] [PubMed] [Google Scholar]
- 129.Baller D, Notohamiprodjo G, Gleichmann U, Holzinger J, Weise R, Lehmann J. Improvement in coronary flow reserve determined by positron emission tomography after 6 months of cholesterol-lowering therapy in patients with early stages of coronary atherosclerosis. Circulation. 1999;99(22):2871–2875. doi: 10.1161/01.cir.99.22.2871. [DOI] [PubMed] [Google Scholar]
- 130.Campisi R, Czernin J, Schöder H, et al. Effects of long-term smoking on myocardial blood flow, coronary vasomotion, and vasodilator capacity. Circulation. 1998;98(2):119–125. doi: 10.1161/01.cir.98.2.119. [DOI] [PubMed] [Google Scholar]
- 131.Kaufmann PA, Gnecchi-Ruscone T, di Terlizzi M, Schäfers KP, Lüscher TF, Camici PG. Coronary heart desease in smokers: vitamin C restores coronary microcirculatory function. Circulation. 2000;102(11):1233–1238. doi: 10.1161/01.cir.102.11.1233. [DOI] [PubMed] [Google Scholar]
- 132.Morita K, Tsukamoto T, Naya M, et al. Smoking cessation normalizes coronary endothelial vasomotor response assessed with 15O-water and PET in healthy young smokers. Journal of Nuclear Medicine. 2006;47(12):1914–1920. [PubMed] [Google Scholar]
- 133.Schindler TH, Nitzsche EU, Olschewski M, et al. Chronic inflammation and impaired coronary vasoreactivity in patients with coronary risk factors. Circulation. 2004;110(9):1069–1075. doi: 10.1161/01.CIR.0000140264.56496.76. [DOI] [PubMed] [Google Scholar]
- 134.Schindler TH, Cardenas J, Prior JO, et al. Relationship between increasing body weight, insulin resistance, inflammation, adipocytokine leptin, and coronary circulatory function. Journal of the American College of Cardiology. 2006;47(6):1188–1195. doi: 10.1016/j.jacc.2005.10.062. [DOI] [PubMed] [Google Scholar]
- 135.Campisi R, Nathan L, Pampaloni MH, et al. Noninvasive assessment of coronary microcirculatory function in postmenopausal women and effects of short-term and long-term estrogen administration. Circulation. 2002;105(4):425–430. doi: 10.1161/hc0402.102860. [DOI] [PubMed] [Google Scholar]
- 136.Uren NG, Melin JA, De Bruyne B, Wijns W, Baudhuin T, Camici PG. Relation between myocardial blood flow and the severity of coronary- artery stenosis. The New England Journal of Medicine. 1994;330(25):1782–1788. doi: 10.1056/NEJM199406233302503. [DOI] [PubMed] [Google Scholar]
- 137.Di Carli M, Czernin J, Hoh CK, et al. Relation among stenosis severity, myocardial blood flow, and flow reserve in patients with coronary artery disease. Circulation. 1995;91(7):1944–1951. doi: 10.1161/01.cir.91.7.1944. [DOI] [PubMed] [Google Scholar]
- 138.Goldstein RA, Kirkeeide RL, Demer LL. Relation between geometric dimensions of coronary artery stenoses and myocardial perfusion reserve in man. Journal of Clinical Investigation. 1987;79(5):1473–1478. doi: 10.1172/JCI112976. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 139.Beanlands RS, Muzik O, Melon P, et al. Noninvasive quantification of regional myocardial flow reserve in patients with coronary atherosclerosis using nitrogen-13 ammonia positron emission tomography. Determination of extent of altered vascular reactivity. Journal of the American College of Cardiology. 1995;26(6):1465–1475. doi: 10.1016/0735-1097(95)00359-2. [DOI] [PubMed] [Google Scholar]
- 140.Anagnostopoulos C, Almonacid A, El Fakhri G, et al. Quantitative relationship between coronary vasodilator reserve assessed by 82Rb PET imaging and coronary artery stenosis severity. European Journal of Nuclear Medicine and Molecular Imaging. 2008;35(9):1593–1601. doi: 10.1007/s00259-008-0793-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 141.Campisi R, Czernin J, Karpman HL, Schelbert HR. Coronary vasodilatory capacity and flow reserve in normal myocardium supplied by bypass grafts late after surgery. American Journal of Cardiology. 1997;80(1):27–31. doi: 10.1016/s0002-9149(97)00278-6. [DOI] [PubMed] [Google Scholar]
- 142.Kosa I, Blasini R, Schneider-Eicke J, et al. Early recovery of coronary flow reserve after stent implantation as assessed by positron emission tomography. Journal of the American College of Cardiology. 1999;34(4):1036–1041. doi: 10.1016/s0735-1097(99)00336-8. [DOI] [PubMed] [Google Scholar]
- 143.Yoshinaga K, Katoh C, Manabe O, et al. Incremental diagnostic value of regional myocardial blood flow quantification over relative perfusion imaging with generator-produced rubidium-82 PET. Circulation Journal. 2011;75:2628–2634. doi: 10.1253/circj.cj-11-0502. [DOI] [PubMed] [Google Scholar]
- 144.Johnson NP, Gould KL. Physiological basis for angina and ST-segment change PET-verified thresholds of quantitative stress myocardial perfusion and coronary flow reserve. JACC: Cardiovascular Imaging. 2011;4:990–998. doi: 10.1016/j.jcmg.2011.06.015. [DOI] [PubMed] [Google Scholar]
- 145.Radvan J, Marwick TH, Williams MJ, Camici PG. Evaluation of the extent and timing of the coronary hyperemic response to dipyridamole: a study with transesophageal echocardiography and positron emission tomography with oxygen 15 water. Journal of the American Society of Echocardiography. 1995;8(6):864–873. doi: 10.1016/s0894-7317(05)80010-0. [DOI] [PubMed] [Google Scholar]
- 146.Demer LL, Gould KL, Goldstein RA, Kirkeeide RL. Noninvasive assessment of coronary collaterals in man by PET perfusion imaging. Journal of Nuclear Medicine. 1990;31(3):259–270. [PubMed] [Google Scholar]
- 147.McFalls EO, Araujo LI, Lammertsma A, et al. Vasodilator reserve in collateral-dependent myocardium as measured by positron emission tomography. European Heart Journal. 1993;14(3):336–343. doi: 10.1093/eurheartj/14.3.336. [DOI] [PubMed] [Google Scholar]
- 148.Vanoverschelde JLJ, Wijns W, Depre C, et al. Mechanisms of chronic regional postischemic dysfunction in humans: new insights from the study of noninfarcted collateral-dependent myocardium. Circulation. 1993;87(5):1513–1523. doi: 10.1161/01.cir.87.5.1513. [DOI] [PubMed] [Google Scholar]
- 149.Di Carli MF, Hachamovitch R. New technology for noninvasive evaluation of coronary artery disease. Circulation. 2007;115(11):1464–1480. doi: 10.1161/CIRCULATIONAHA.106.629808. [DOI] [PubMed] [Google Scholar]
- 150.Bateman TM, Heller GV, McGhie AI, et al. Diagnostic accuracy of rest/stress ECG-gated Rb-82 myocardial perfusion PET: comparison with ECG-gated Tc-99m sestamibi SPECT. Journal of Nuclear Cardiology. 2006;13(1):24–33. doi: 10.1016/j.nuclcard.2005.12.004. [DOI] [PubMed] [Google Scholar]
- 151.Gould KL, Goldstein RA, Mullani NA. Economic analysis of clinical positron emission tomography of the heart with rubidium-82. Journal of Nuclear Medicine. 1989;30(5):707–717. [PubMed] [Google Scholar]
- 152.Patterson RE, Eisner RL, Horowitz SF. Comparison of cost-effectiveness and utility of exercise ECG, single photon emission computed tomography, positron emission tomography, and coronary angiography for diagnosis of coronary artery disease. Circulation. 1995;91(1):54–65. doi: 10.1161/01.cir.91.1.54. [DOI] [PubMed] [Google Scholar]
- 153.Merhige ME, Breen WJ, Shelton V, Houston T, D’Arcy BJ, Perna AF. Impact of myocardial perfusion imaging with PET and 82Rb on downstream invasive procedure utilization, costs, and outcomes in coronary disease management. Journal of Nuclear Medicine. 2007;48(7):1069–1076. doi: 10.2967/jnumed.106.038323. [DOI] [PubMed] [Google Scholar]
- 154.Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ. Radiation dose to patients from cardiac diagnostic imaging. Circulation. 2007;116(11):1290–1305. doi: 10.1161/CIRCULATIONAHA.107.688101. [DOI] [PubMed] [Google Scholar]
- 155.Achenbach S, Anders K, Kalender WA. Dual-source cardiac computed tomography: image quality and dose considerations. European Radiology. 2008;18(6):1188–1198. doi: 10.1007/s00330-008-0883-3. [DOI] [PubMed] [Google Scholar]
- 156.Thompson RC, Cullom SJ. Issues regarding radiation dosage of cardiac nuclear and radiography procedures. Journal of Nuclear Cardiology. 2006;13(1):19–23. doi: 10.1016/j.nuclcard.2005.11.004. [DOI] [PubMed] [Google Scholar]
- 157.Hendel RC, Berman DS, Di Carli MF, et al. ACCF/ASNC/ACR/AHA/ASE/SCCT/SCMR/SNM 2009 Appropriate Use Criteria for Cardiac Radionuclide Imaging. A Report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the American Society of Nuclear Cardiology, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the Society of Cardiovascular Computed Tomography. Journal of the American College of Cardiology. 2009;53(23):2201–2229. doi: 10.1016/j.jacc.2009.02.013. [DOI] [PubMed] [Google Scholar]
- 158.Beller GA. Underestimation of coronary artery disease with SPECT perfusion imaging. Journal of Nuclear Cardiology. 2008;15(2):151–153. doi: 10.1016/j.nuclcard.2008.01.012. [DOI] [PubMed] [Google Scholar]
- 159.Berman DS, Kang X, Slomka PJ, et al. Underestimation of extent of ischemia by gated SPECT myocardial perfusion imaging in patients with left main coronary artery disease. Journal of Nuclear Cardiology. 2007;14(4):521–528. doi: 10.1016/j.nuclcard.2007.05.008. [DOI] [PubMed] [Google Scholar]
- 160.Yoshinaga K, Katoh C, Noriyasu K, et al. Reduction of coronary flow reserve in areas with and without ischemia on stress perfusion imaging in patients with coronary artery disease: a study using oxygen 15-labeled water PET. Journal of Nuclear Cardiology. 2003;10(3):275–283. doi: 10.1016/s1071-3581(02)43243-6. [DOI] [PubMed] [Google Scholar]
- 161.Parkash R, deKemp RA, Ruddy TD, et al. Potential utility of rubidium 82 PET quantification in patients with 3-vessel coronary artery disease. Journal of Nuclear Cardiology. 2004;11(4):440–449. doi: 10.1016/j.nuclcard.2004.04.005. [DOI] [PubMed] [Google Scholar]
- 162.Seth R, deKemp RA, Hart B, et al. Quantification of absolute perfusion reserve using 82Rb positron emission tomography defines greater extent of disease in three vessel coronary atherosclerosis. Journal of the American College of Cardiology. 2001;37, article 387A [Google Scholar]
- 163.Hajjiri MM, Leavitt MB, Zheng H, Spooner AE, Fischman AJ, Gewirtz H. Comparison of positron emission tomography measurement of adenosine-stimulated absolute myocardial blood flow versus relative myocardial tracer content for physiological assessment of coronary artery stenosis severity and location. JACC: Cardiovascular Imaging. 2009;2(6):751–758. doi: 10.1016/j.jcmg.2009.04.004. [DOI] [PubMed] [Google Scholar]
- 164.Pearson TA. New tools for coronary risk assessment: what are their advantages and limitations? Circulation. 2002;105(7):886–892. doi: 10.1161/hc0702.103727. [DOI] [PubMed] [Google Scholar]
- 165.Gould KL. Does coronary flow trump coronary anatomy? JACC: Cardiovascular Imaging. 2009;2(8):1009–1023. doi: 10.1016/j.jcmg.2009.06.004. [DOI] [PubMed] [Google Scholar]
- 166.Marwick TH, Shan K, Patel S, Go RT, Lauer MS. Incremental value of rubidium-82 positron emission tomography for prognostic assessment of known or suspected coronary artery disease. American Journal of Cardiology. 1997;80(7):865–870. doi: 10.1016/s0002-9149(97)00537-7. [DOI] [PubMed] [Google Scholar]
- 167.MacIntyre WJ, Go RT, King JL, et al. Clinical outcome of cardiac patients with negative thallium-201 SPECT and positive rubidium-82 PET myocardial perfusion imaging. Journal of Nuclear Medicine. 1993;34(3):400–404. [PubMed] [Google Scholar]
- 168.Yoshinaga K, Chow BJW, Williams K, et al. What is the prognostic value of myocardial perfusion imaging using rubidium-82 positron emission tomography? Journal of the American College of Cardiology. 2006;48(5):1029–1039. doi: 10.1016/j.jacc.2006.06.025. [DOI] [PubMed] [Google Scholar]
- 169.Dorbala S, Hachamovitch R, Curillova Z, et al. Incremental prognostic value of gated Rb-82 positron emission tomography myocardial perfusion imaging over clinical variables and rest LVEF. JACC: Cardiovascular Imaging. 2009;2(7):846–854. doi: 10.1016/j.jcmg.2009.04.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 170.Schindler TH, Nitzsche EU, Schelbert HR, et al. Positron emission tomography-measured abnormal responses of myocardial blood flow to sympathetic stimulation are associated with the risk of developing cardiovascular events. Journal of the American College of Cardiology. 2005;45(9):1505–1512. doi: 10.1016/j.jacc.2005.01.040. [DOI] [PubMed] [Google Scholar]
- 171.Tio RA, Dabeshlim A, Siebelink HMJ, et al. Comparison between the prognostic value of left ventricular function and myocardial perfusion reserve in patients with ischemic heart disease. Journal of Nuclear Medicine. 2009;50(2):214–219. doi: 10.2967/jnumed.108.054395. [DOI] [PubMed] [Google Scholar]
- 172.Herzog BA, Husmann L, Valenta I, et al. Long-term prognostic value of 13N-ammonia myocardial perfusion positron emission tomography. Added value of coronary flow reserve. Journal of the American College of Cardiology. 2009;54(2):150–156. doi: 10.1016/j.jacc.2009.02.069. [DOI] [PubMed] [Google Scholar]
- 173.Fukushima K, Javadi MS, Higuchi T, et al. Prediction of short-term cardiovascular events using quantification of global myocardial flow reserve in patients referred for clinical 82Rb PET perfusion imaging. Journal of Nuclear Medicine. 2011;52(5):726–732. doi: 10.2967/jnumed.110.081828. [DOI] [PubMed] [Google Scholar]
- 174.Ziadi MC, Dekemp RA, Williams KA, et al. Impaired myocardial flow reserve on rubidium-82 positron emission tomography imaging predicts adverse outcomes in patients assessed for myocardial ischemia. Journal of the American College of Cardiology. 2011;58(7):740–748. doi: 10.1016/j.jacc.2011.01.065. [DOI] [PubMed] [Google Scholar]
- 175.Murthy VL, Naya M, Foster CR, et al. Improved cardiac risk assessment with noninvasive measures of coronary flow reserve. Circulation. 2011;124:2215–2224. doi: 10.1161/CIRCULATIONAHA.111.050427. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 176.Camici PG, Crea F. Coronary microvascular dysfunction. New England Journal of Medicine. 2007;356(8):830–840. doi: 10.1056/NEJMra061889. [DOI] [PubMed] [Google Scholar]
- 177.Wu YW, Chen YH, Wang SS, et al. PET assessment of myocardial perfusion reserve inversely correlates with intravascular ultrasound findings in angiographically normal cardiac transplant recipients. Journal of Nuclear Medicine. 2010;51(6):906–912. doi: 10.2967/jnumed.109.073833. [DOI] [PubMed] [Google Scholar]
- 178.Maron BJ, Epstein SE, Roberts WC. Hypertrophic cardiomyopathy and transmural myocardial infarction without significant atherosclerosis of the extramural coronary arteries. American Journal of Cardiology. 1979;43(6):1086–1102. doi: 10.1016/0002-9149(79)90139-5. [DOI] [PubMed] [Google Scholar]
- 179.Pasternac A, Noble J, Streulens Y. Pathophysiology of chest pain in patients with cardiomyopathies and normal coronary arteries. Circulation. 1982;65(4):778–789. doi: 10.1161/01.cir.65.4.778. [DOI] [PubMed] [Google Scholar]
- 180.Camici P, Chiriatti G, Lorenzoni R, et al. Coronary vasodilation is impaired in both hypertrophied and nonhypertrophied myocardium of patients with hypertrophic cardiomyopathy: a study with nitrogen-13 ammonia and positron emission tomography. Journal of the American College of Cardiology. 1991;17(4):879–886. doi: 10.1016/0735-1097(91)90869-b. [DOI] [PubMed] [Google Scholar]
- 181.Choudhury L, Elliott P, Rimoldi O, et al. Transmural myocardial blood flow distribution in hypertrophic cardiomyopathy and effect of treatment. Basic Research in Cardiology. 1999;94(1):49–59. doi: 10.1007/s003950050126. [DOI] [PubMed] [Google Scholar]
- 182.Maron BJ, Wolfson JK, Epstein SE, Roberts WC. Intramural (“small vessel”) coronary artery disease in hypertrophic cardiomyopathy. Journal of the American College of Cardiology. 1986;8(3):545–557. doi: 10.1016/s0735-1097(86)80181-4. [DOI] [PubMed] [Google Scholar]
- 183.Knaapen P, Germans T, Camici PG, et al. Determinants of coronary microvascular dysfunction in symptomatic hypertrophic cardiomyopathy. American Journal of Physiology. 2008;294(2):H986–H993. doi: 10.1152/ajpheart.00233.2007. [DOI] [PubMed] [Google Scholar]
- 184.Sotgia B, Sciagrà R, Olivotto I, et al. Spatial relationship between coronary microvascular dysfunction and delayed contrast enhancement in patients with hypertrophic cardiomyopathy. Journal of Nuclear Medicine. 2008;49(7):1090–1096. doi: 10.2967/jnumed.107.050138. [DOI] [PubMed] [Google Scholar]
- 185.Sciagrà R, Sotgia B, Olivotto I, et al. Relationship between atrial fibrillation and blunted hyperemic myocardial blood flow in patients with hypertrophic cardiomyopathy. Journal of Nuclear Cardiology. 2009;16(1):92–96. doi: 10.1007/s12350-008-9005-5. [DOI] [PubMed] [Google Scholar]
- 186.Olivotto I, Girolami F, Sciagrà R, et al. Microvascular function is selectively impaired in patients with hypertrophic cardiomyopathy and sarcomere myofilament gene mutations. Journal of the American College of Cardiology. 2011;58:839–848. doi: 10.1016/j.jacc.2011.05.018. [DOI] [PubMed] [Google Scholar]
- 187.Cecchi F, Olivotto I, Gistri R, Lorenzoni R, Chiriatti G, Camici PG. Coronary microvascular dysfunction and prognosis in hypertrophic cardiomyopathy. The New England Journal of Medicine. 2003;349(11):1027–1035. doi: 10.1056/NEJMoa025050. [DOI] [PubMed] [Google Scholar]
- 188.Olivotto I, Cecchi F, Gistri R, et al. Relevance of coronary microvascular flow impairment to long-term remodeling and systolic dysfunction in hypertrophic cardiomyopathy. Journal of the American College of Cardiology. 2006;47(5):1043–1048. doi: 10.1016/j.jacc.2005.10.050. [DOI] [PubMed] [Google Scholar]
- 189.Elliott PM, Kindler H, Shah JS, et al. Coronary microvascular dysfunction in male patients with Anderson-Fabry disease and the effect of treatment with α galactosidase A. Heart. 2006;92(3):357–360. doi: 10.1136/hrt.2004.054015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 190.Tomberli B, Fornaro A, Sciagrà R, et al. Coronary microvascular dysfunction as an early feature of cardiac involvement in male and female patients with Fabry disease. European Journal of Heart Failure. 2012;11(supplement 1):p. S6. doi: 10.1093/eurjhf/hft104. [DOI] [PubMed] [Google Scholar]
- 191.Weiss MB, Ellis K, Sciacca RR, Johnson LL, Schmidt DH, Cannon PJ. Myocardial blood flow in congestive and hypertrophic cardiomyopathy. Relationship to peak wall stress and mean velocity of circumferential fiber shortening. Circulation. 1976;54(3):484–494. doi: 10.1161/01.cir.54.3.484. [DOI] [PubMed] [Google Scholar]
- 192.Neglia D, Parodi O, Gallopin M, et al. Myocardial blood flow response to pacing tachycardia and to dipyridamole infusion in patients with dilated cardiomyopathy without overt heart failure: a quantitative assessment by positron emission tomography. Circulation. 1995;92(4):796–804. doi: 10.1161/01.cir.92.4.796. [DOI] [PubMed] [Google Scholar]
- 193.Opherk D, Schwarz F, Mall G. Coronary dilatory capacity in idiopathic dilated cardiomyopathy: analysis of 16 patients. American Journal of Cardiology. 1983;51(10):1657–1662. doi: 10.1016/0002-9149(83)90205-9. [DOI] [PubMed] [Google Scholar]
- 194.Canetti M, Akhter MW, Lerman A, et al. Evaluation of myocardial blood flow reserve in patients with chronic congestive heart failure due to idiopathic dilated cardiomyopathy. American Journal of Cardiology. 2003;92(10):1246–1249. doi: 10.1016/j.amjcard.2003.08.002. [DOI] [PubMed] [Google Scholar]
- 195.Stolen KQ, Kemppainen J, Kalliokoski KK, et al. Myocardial perfusion reserve and peripheral endothelial function in patients with idiopathic dilated cardiomyopathy. American Journal of Cardiology. 2004;93(1):64–68. doi: 10.1016/j.amjcard.2003.08.074. [DOI] [PubMed] [Google Scholar]
- 196.Neglia D, Michelassi C, Giovanna Trivieri M, et al. Prognostic role of myocardial blood flow impairment in idiopathic left ventricular dysfunction. Circulation. 2002;105(2):186–193. doi: 10.1161/hc0202.102119. [DOI] [PubMed] [Google Scholar]
- 197.Gewirtz H. Cardiac PET: a versatile, quantitative measurement tool for heart failure management. JACC: Cardiovascular Imaging. 2011;4(3):292–302. doi: 10.1016/j.jcmg.2010.12.006. [DOI] [PubMed] [Google Scholar]
- 198.Shah P, Choi BG, Mazhari R. Positron emission tomography for the evaluation and treatment of cardiomyopathy. Annals of the New York Academy of Sciences. 2011;1228(1):137–149. doi: 10.1111/j.1749-6632.2011.06017.x. [DOI] [PubMed] [Google Scholar]
- 199.Neglia D, De Maria R, Masi S, et al. Effects of long-term treatment with carvedilol on myocardial blood flow in idiopathic dilated cardiomyopathy. Heart. 2007;93(7):808–813. doi: 10.1136/hrt.2006.095208. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 200.Gimelli A, Schneider-Eicke J, Neglia D, et al. Homogeneously reduced versus regionally impaired myocardial blood flow in hypertensive patients: two different patterns of myocardial perfusion associated with degree of hypertrophy. Journal of the American College of Cardiology. 1998;31(2):366–373. doi: 10.1016/s0735-1097(97)00503-2. [DOI] [PubMed] [Google Scholar]
- 201.Namdar M, Hany TF, Koepfli P, et al. Integrated PET/CT for the assessment of coronary artery disease: a feasibility study. Journal of Nuclear Medicine. 2005;46(6):930–935. [PubMed] [Google Scholar]
- 202.Di Carli MF, Dorbala S, Curillova Z, et al. Relationship between CT coronary angiography and stress perfusion imaging in patients with suspected ischemic heart disease assessed by integrated PET-CT imaging. Journal of Nuclear Cardiology. 2007;14(6):799–809. doi: 10.1016/j.nuclcard.2007.07.012. [DOI] [PubMed] [Google Scholar]
- 203.Schenker MP, Dorbala S, Hong ECT, et al. Interrelation of coronary calcification, myocardial ischemia, and outcomes in patients with intermediate likelihood of coronary artery disease: a combined positron emission tomography/computed tomography study. Circulation. 2008;117(13):1693–1700. doi: 10.1161/CIRCULATIONAHA.107.717512. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 204.Santana CA, Folks RD, Garcia EV, et al. Quantitative 82Rb PET/CT: development and validation of myocardial perfusion database. Journal of Nuclear Medicine. 2007;48(7):1122–1128. doi: 10.2967/jnumed.107.039750. [DOI] [PubMed] [Google Scholar]
- 205.Sampson UK, Dorbala S, Limaye A, Kwong R, Di Carli MF. Diagnostic accuracy of rubidium-82 myocardial perfusion imaging with hybrid positron emission tomography/computed tomography in the detection of coronary artery disease. Journal of the American College of Cardiology. 2007;49(10):1052–1058. doi: 10.1016/j.jacc.2006.12.015. [DOI] [PubMed] [Google Scholar]
- 206.Kajander S, Joutsiniemi E, Saraste M, et al. Cardiac positron emission tomography/computed tomography imaging accurately detects anatomically and functionally significant coronary artery disease. Circulation. 2010;122(6):603–613. doi: 10.1161/CIRCULATIONAHA.109.915009. [DOI] [PubMed] [Google Scholar]
- 207.Naya M, Murthy VL, Blankstein R, et al. Quantitative relationship between the extent and morphology of coronary atherosclerotic plaque and downstream myocardial perfusion. Journal of the American College of Cardiology. 2011;58:1807–1816. doi: 10.1016/j.jacc.2011.06.051. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 208.Burkhard N, Herzog BA, Husmann L, et al. Coronary calcium score scans for attenuation correction of quantitative PET/CT 13N-ammonia myocardial perfusion imaging. European Journal of Nuclear Medicine and Molecular Imaging. 2010;37(3):517–521. doi: 10.1007/s00259-009-1271-1. [DOI] [PubMed] [Google Scholar]
- 209.Knuuti J. Are Coronary plaque characteristics on computed tomography angiography associated with myocardial perfusion? Journal of the American College of Cardiology. 2011;58:1817–1818. doi: 10.1016/j.jacc.2011.05.059. [DOI] [PubMed] [Google Scholar]
- 210.Liga R, Marini C, Coceani M, et al. Structural abnormalities of the coronary arterial wall in addition to luminal narrowing affect myocardial blood flow reserve. Journal of Nuclear Medicine. 2011;52:1704–1712. doi: 10.2967/jnumed.111.091009. [DOI] [PubMed] [Google Scholar]
- 211.Naya M, Di Carli MF. Myocardial perfusion PET/CT to evaluate known and suspected coronary artery disease. Quarterly Journal of Nuclear Medicine and Molecular Imaging. 2010;54(2):145–156. [PubMed] [Google Scholar]
- 212.Javadi M, Mahesh M, McBride G, et al. Lowering radiation dose for integrated assessment of coronary morphology and physiology: first experience with step-and-shoot CT angiography in a rubidium 82 PET-CT protocol. Journal of Nuclear Cardiology. 2008;15(6):783–790. doi: 10.1007/BF03007359. [DOI] [PubMed] [Google Scholar]
- 213.Kajander S, Ukkonen H, Sipilä H, Teräs M, Knuuti J. Low radiation dose imaging of myocardial perfusion and coronary angiography with a hybrid PET/CT scanner. Clinical Physiology and Functional Imaging. 2009;29(1):81–88. doi: 10.1111/j.1475-097X.2008.00838.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 214.Lee DC, Johnson NP. Quantification of absolute myocardial blood flow by magnetic resonance perfusion imaging. JACC: Cardiovascular Imaging. 2009;2(6):761–770. doi: 10.1016/j.jcmg.2009.04.003. [DOI] [PubMed] [Google Scholar]
- 215.Dill T. Contraindications to magnetic resonance imaging. Heart. 2008;94(7):943–948. doi: 10.1136/hrt.2007.125039. [DOI] [PubMed] [Google Scholar]
- 216.Nakauchi Y, Iwanaga Y, Ikuta S, et al. Quantitative myocardial perfusion analysis using multi-row detector CT in acute myocardial infarction. Heart. 2012;98:566–572. doi: 10.1136/heartjnl-2011-300915. [DOI] [PubMed] [Google Scholar]
- 217.Beller GA. Recent advances and future trends in multimodality cardiac imaging. Heart Lung and Circulation. 2010;19(3):193–209. doi: 10.1016/j.hlc.2009.11.003. [DOI] [PubMed] [Google Scholar]
- 218.Yu M, Nekolla SG, Schwaiger M, Robinson SP. The next generation of cardiac positron emission tomography imaging agents: discovery of flurpiridaz F-18 for detection of coronary disease. Seminars in Nuclear Medicine. 2011;41(4):305–313. doi: 10.1053/j.semnuclmed.2011.02.004. [DOI] [PubMed] [Google Scholar]
- 219.Berman DS, Germano G, Slomka PJ. Improvement in PET myocardial perfusion image quality and quantification with flurpiridaz F 18. Journal of Nuclear Cardiology. 2012;19:S38–S45. doi: 10.1007/s12350-011-9487-4. [DOI] [PubMed] [Google Scholar]
- 220.Nekolla SG, Reder S, Saraste A, et al. Evaluation of the novel myocardial perfusion positron-emission tomography tracer 18F-BMS-747158-02: comparison to 13N-ammonia and validation with microspheres in a pig model. Circulation. 2009;119(17):2333–2342. doi: 10.1161/CIRCULATIONAHA.108.797761. [DOI] [PubMed] [Google Scholar]
- 221.Sherif HM, Nekolla SG, Saraste A, et al. Simplified quantification of myocardial flow reserve with flurpiridaz F 18: validation with microspheres in a pig model. Journal of Nuclear Medicine. 2011;52(4):617–624. doi: 10.2967/jnumed.110.083196. [DOI] [PubMed] [Google Scholar]
- 222.Maddahi J, Huang S, Truong D, et al. Preliminary results of absolute quantification of rest and stress myocardial blood flow with flurpiridaz F-18 PET in normal and coronary artery disease patients in a single-center study. Journal of Nuclear Cardiology. 2010;17, article 743 [Google Scholar]
- 223.Vermeeren PC. Nuclear cardiology in Europe: nuclear medicine physicians tend to build walls around their skills and thereby limit the growth of the field. Journal of Nuclear Cardiology. 1999;6(2):241–244. doi: 10.1016/s1071-3581(99)90085-5. [DOI] [PubMed] [Google Scholar]
- 224.Marcassa C, Bischof Delaloye A, Cuocolo A, et al. The regulatory background of nuclear cardiology in Europe: a survey by the European Council of Nuclear Cardiology. European Journal of Nuclear Medicine and Molecular Imaging. 2006;33(12):1508–1512. doi: 10.1007/s00259-006-0211-6. [DOI] [PubMed] [Google Scholar]
- 225.Hachamovitch R, Johnson JR, Hlatky MA, et al. The study of myocardial perfusion and coronary anatomy imaging roles in CAD (SPARC): design, rationale, and baseline patient characteristics of a prospective, multicenter observational registry comparing PET, SPECT, and CTA for resource utilization and clinical outcomes. Journal of Nuclear Cardiology. 2009;16(6):935–948. doi: 10.1007/s12350-009-9140-7. [DOI] [PubMed] [Google Scholar]
- 226.Hachamovitch R, Nutter B, Hlatky MA, SPARC Investigators, et al. Patient management after noninvasive cardiac imaging results from SPARC (Study of myocardial perfusion and coronary anatomy imaging roles in coronary artery disease) Journal of the American College of Cardiology. 2012;59:462–474. doi: 10.1016/j.jacc.2011.09.066. [DOI] [PubMed] [Google Scholar]