Assuring the quality of next-generation sequencing in clinical laboratory practice

Amy S Gargis 1, Lisa Kalman 1, Meredith W Berry 2, David P Bick 3, David P Dimmock 3, Tina Hambuch 4, Fei Lu 2, Elaine Lyon 5,6, Karl V Voelkerding 5,6, Barbara A Zehnbauer 1, Richa Agarwala 7, Sarah F Bennett 8, Bin Chen 1, Ephrem L H Chin 9,33, John G Compton 10, Soma Das 11, Daniel H Farkas 12, Matthew J Ferber 13, Birgit H Funke 14,15, Manohar R Furtado 16, Lilia M Ganova-Raeva 17, Ute Geigenmüller 18,33, Sandra J Gunselman 19,33, Madhuri R Hegde 9, Philip L F Johnson 20, Andrew Kasarskis 21,33, Shashikant Kulkarni 22, Thomas Lenk 23, C S Jonathan Liu 24, Megan Manion 24, Teri A Manolio 25, Elaine R Mardis 26, Jason D Merker 27, Mangalathu S Rajeevan 28, Martin G Reese 29, Heidi L Rehm 14,15, Birgitte B Simen 30, Joanne M Yeakley 31,33, Justin M Zook 32, Ira M Lubin 1
PMCID: PMC3827024  NIHMSID: NIHMS525968  PMID: 23138292

To the Editor:

We direct your readers’ attention to the principles and guidelines (Supplementary Guidelines) developed by the Next-generation Sequencing: Standardization of Clinical Testing (Nex-StoCT) workgroup. These guidelines represent initial steps to ensure that results from tests based on next-generation sequencing (NGS) are reliable and useful for clinical decision making. The US Centers for Disease Control and Prevention (CDC) convened this national workgroup, which collaborated to define platform-independent approaches for establishing technical process elements of a quality management system (QMS) to assure the analytical validity and compliance of NGS tests with existing regulatory and professional quality standards. The workgroup identified and addressed gaps in quality practices that could compromise the quality of both clinical laboratory services and translational efforts needed to advance the implementation and utility of NGS in clinical settings.

The workgroup was composed of experts with knowledge of and experience with NGS and included clinical laboratory directors, clinicians, platform and software developers and informaticians, as well as individuals actively engaged in NGS guideline development from accreditation bodies and professional organizations. Representatives from US government agencies also participated.

These guidelines address four topics that are components of quality management in a clinical environment: (i) test validation, (ii) quality control (QC) procedures to assure and maintain accurate test results, (iii) the independent assessment of test performance through proficiency testing (PT) or alternative approaches and (iv) reference materials (RMs). Discussions were limited to the analytic and informatics processes required for accurate variant calling. The workgroup did not address how variants are prioritized, interpreted or reported.

The workgroup recommendations are summarized in Table 1. Although the workgroup focused on detection of DNA sequence variations associated with heritable human disorders, many of the principles and recommendations described are also relevant to the application of NGS to other areas of laboratory medicine, including the diagnosis, prognosis and treatment of cancer and infectious-disease testing.

Table 1.

Selected workgroup recommendations for establishing NGS test systems for clinical use

Requirements for
test establishment
Objective NGS-specific recommendationsa
Validation Document reliability of the platform, test, and informatics pipeline before testing of patient specimens
  • Platform validation: establish that the system provides reliable sequence analysis across the genomic regions targeted by the test.

  • Test validation: establish that the system correctly identifies disease-associated (and other) variants in targeted regions of the genome (Supplementary Guidelines, section 4).

  • Informatics pipeline validation: establish that the algorithm(s) reliably analyze platform data to produce an accurate sequence.

  • Establish and validate alternate methods (for example, Sanger sequencing) to derive high-quality sequence data for problematic genomic regions.

Quality control Document reliability of the sequence analysis during patient testing
  • Utilize a combination of QC materials, both intrinsic and/or spiked in, that mimic genomic complexity and the types of mutations the test is designed to detect.

  • During patient testing, quality metrics (for example, quality scores, depth and uniformity of coverage, mapping quality, GC bias and transition/transversion ratio) should be assessed and compared to those established during validation.

  • Clinically actionable findings should be confirmed by independent analysis using an alternate method.

Proficiency Testing The independent assessment of test performance
  • PT challenges should target the analysis of both disease-associated and naturally occurring sequence variations across the genomic regions targeted by the test to measure the reliability of sequence analysis.

  • Electronic sequence files may permit a comparison of alignment and variant calling methods across laboratories but will require additional consideration of platform differences.

  • PT programs should consider the different genomic regions targeted by each recipient laboratory’s assays to properly compare interlaboratory performance.

Reference Materials The use of materials for quality management of the analytical phase of testing
  • RMs with both naturally occurring and disease-associated sequence variations are needed for test validation, QC procedures and the independent assessment of test performance.

  • Synthetic DNA and electronic reference data files may serve as RMs for rare or challenging sequence variations.

  • Efforts should be undertaken to establish a suitable NGS RM and the sequence of the RM should be refined as the technology changes. Such a RM should be annotated to indicate regions of high and low sequence reliability.

a

See Supplementary Guidelines for complete recommendations. RM, reference material.

Validation is the process of establishing analytical performance specifications for a clinical test system developed in house to confirm that the system is suitable for its intended use1. During the validation process, the laboratory must demonstrate that the assay functions as expected and provides reliable results. The workgroup considered the requirements of the Clinical Laboratory Improvement Amendments (CLIA) and provided recommendations for validation of clinical NGS tests. The validation process can be divided into three stages: platform, test-specific and informatics pipeline validation (Supplementary Fig. 1). Platform validation provides evidence that the assay can deliver reliable sequence data within the regions of the genome targeted for analysis. Test-specific validation demonstrates that the assay can detect clinically important sequence variations for the intended application. Validation of the informatics pipeline establishes the software settings necessary to ensure that the test can reliably provide accurate sequence data and detect variations. Although each stage of validation is considered separately, they are interdependent. Validation requires application-specific considerations for whether the test targets a gene panel, the exome or the whole genome, as well as the types of sequence variations that are detected.

In the United States, diagnostic tests that are provided to clinical laboratories are regulated by the US Food and Drug Administration (FDA). To date, no NGS technologies have been approved or cleared by the FDA. These tests are developed in house as laboratory-developed tests and are regulated under CLIA2. The CLIA regulations define evaluation of analytical reliability and limitations and require laboratories to establish specifications for performance characteristics for each test system developed in house. The characteristics evaluated to establish the analytical validity of test results include accuracy, precision, analytical sensitivity, analytical specificity, reportable range, reference range or reference intervals, and other relevant performance metrics. Laboratories in the United States and other countries may also comply with the QMS standards described in ISO 15189 (ref. 3). The performance characteristics defined in CLIA2 and professional guidance documents4 (ACMG standards, guidelines, and policies available at; CAP NGS checklist available to subscribers at http://www.cap.org/) do not readily translate to NGS testing practices owing to the complexity of the technology and the informatics analyses required for large-scale genome analyses. Therefore, the workgroup adapted the definitions of these performance characteristics to better fit the use of NGS in the clinical laboratory (Table 2). A comparison between the CLIA definitions and those developed by the workgroup is presented in Supplementary Table 1, and the unique metrics for NGS that laboratories should establish and monitor to assure high-quality analytical results are presented in Supplementary Table 2. For example, the depth of coverage, or the number of independent reads assessed when making a base call, is a crucial metric for establishing the accuracy, analytical sensitivity and analytical specificity of an NGS test. Owing to the high costs and extensive data analyses required for these tests, it is challenging to establish the precision (e.g., repeatability of testing replicates) of an NGS assay by determining concordance of sequencing results among a large number of samples. The workgroup proposed an alternative approach in which additional metrics such as the depth and uniformity of sequencing coverage would be incorporated to limit the number of samples required to establish precision. The workgroup redefined ‘reportable range’ and ‘reference range’ in terms of the qualitative nature of these DNA test results. This guideline classifies reportable range as the region of the genome from which sequence of an acceptable quality can be derived by the laboratory test, and reference range as the spectrum of nonpathogenic base changes observed in a population. Sequence variations outside this spectrum could be disease associated, but further investigation may be necessary to confirm disease association. The determination of reference range is problematic because the spectrum of sequence variations that can be defined as ‘normal’ or deleterious will vary across individuals and populations.

Table 2.

Workgroup definitions of CLIA performance characteristics for NGS

Performance characteristic Workgroup established definition for NGS applicationsa
Accuracy The degree of agreement between the nucleic acid sequences derived from the assay and a reference sequence.
Precision The degree to which repeated sequence analyses give the same resultrepeatability (within-run precision) and reproducibility (between-run precision).
Analytical sensitivity The likelihood that the assay will detect the targeted sequence variations, if present.
Analytical specificity The probability that the assay will not detect a sequence variation when none are present (the false positive rate is a useful measurement for sequencing assays).
Reportable range The region of the genome in which sequence of an acceptable quality can be derived by the laboratory test.
Reference range Reportable sequence variations the assay can detect that are expected to occur in an unaffected population.
a

These definitions may be applied to both NGS and Sanger sequencing. A more detailed comparison between the CLIA definitions and those developed by the workgroup is presented in Supplementary Table 1.

QC procedures monitor whether the components of an assay—including the reagents, specimen processing, instrumentation and data processing (the informatics pipeline)—are functioning properly and delivering accurate results during testing of patient samples. The workgroup considered NGS-specific QC metrics that are useful for monitoring the performance of the assay, including: depth of coverage and uniformity of coverage, quality scores for base calling and alignment, allelic read percentage, strand bias, GC bias and decline in signal intensity (Supplementary Table 2).

Proficiency testing and external quality assessment programs provide a formal mechanism for comparing inter-laboratory test performance and can help to identify analytical and interpretive errors and problems with QC, instrument calibration and assay design. Laboratories are encouraged to subscribe when such programs are available. PT programs typically provide several ‘blinded’ samples (PT challenge) to participating laboratories for analysis. The participants subsequently report their results, an interpretation and a brief description of their assay methods to the PT program. Data are compared among participating laboratories to assess inter-laboratory test performance, and a cumulative summary is shared with participants. At present, no formal PT programs exist for NGS, and development of a program faces four fundamental obstacles: (i) the absence of a defined scope for the challenge (which region(s) of the genome will be targeted); (ii) the absence of well-characterized PT materials suitable for a range of applications (for example, analysis of gene panels, the exome and the whole genome); (iii) the absence of standard metrics for use as comparators among participating laboratories and (iv) cost and time commitments for participants and the PT provider. To address these issues, the workgroup developed principles and guidelines for combining a formal PT challenge (when available) with an alternative assessment process, such as inter-laboratory exchange of previously characterized samples (this may include genomic DNA (gDNA) and/or electronic data files). This combined approach could reduce expenses and provide a flexible means to assess inter-laboratory test performance.

Reference materials are used during test validation, QC and PT to establish and monitor the quality of clinical laboratory tests1,4. RMs are homogeneous and stable, and they have the particular property being measured, such as the presence of disease-associated sequence variations. Many different types of samples can be used as RMs for NGS, including characterized DNA derived from human cell lines or patient specimens, synthetic DNA or electronic data. The advantages and disadvantages of each of these RMs are described in Supplementary Table 3. Characterization of RMs for NGS is complicated by the size of the genomic region(s) targeted for analysis. Laboratories now use gDNA obtained from a variety of sources, including human cell lines characterized in the 1000 Genomes Project5, to develop and validate NGS assays. However, the large spectrum of disease-associated sequence variations cannot be represented in one or any practical number of gDNA RMs. The workgroup recommended that one or more RMs that contain a combination of disease-associated and naturally occurring sequence variations should be evaluated for use in assay validation, QC and PT. The workgroup suggested that electronic reference data files containing real or simulated sequence data can be used for establishing and monitoring the performance characteristics of the NGS informatics pipeline. This approach is useful because combinations of DNA sequence variations can be engineered into these files to assess the capability of the informatics pipeline to make accurate variant calls. The CDC’s Genetic Testing Reference Materials Coordination Program (http://wwwn.cdc.gov/dls/genetics/RMMaterials), the US National Institutes of Health and the US National Institute for Standards and Technology are developing and characterizing RMs for NGS to meet the many needs of clinical laboratories.

Sequencing technologies are evolving rapidly, and although the technical details may change, quality management requirements for test system validation, QC and PT will remain. The workgroup identified areas in which additional data collection and analysis are needed to assure the quality of clinical NGS (Supplementary Table 4). To our knowledge, the principles and guidance developed by the workgroup are the first efforts to establish consensus among a diverse group of stakeholders. These recommendations will further inform the deliberations of regulatory agencies, professional societies and accrediting organizations that are considering the application of QMS standards and guidelines to NGS for clinical applications. These guidelines will be updated and expanded as NGS technologies evolve. Information about the continued activities of the working group is available at http://www.cdc.gov/osels/lspppo/Genetic_Testing_Quality_Practices/Nex-StoCT.html. We encourage collaborations and ongoing discussions among laboratories, clinicians, manufacturers, service providers, software developers, professional organizations and government agencies to ensure the quality of clinical NGS tests.

Supplementary Material

Supplementary Guidelines

ACKNOWLEDGMENTS

The authors thank E. Mansfield, N. Aziz, P. Keller, J. Kaufman, S. Ghosh, R. Leach, J. Robinson, A. Felsenfeld, M. Salit and E. Highsmith for their help and support. The findings and conclusions in this report are those of the authors and do not necessarily represent the views of the CDC or the US Agency for Toxic Substances and Disease Registry. The identification of certain commercial equipment, instruments or materials in this document does not imply recommendation or endorsement by the US National Institute of Standards and Technology, nor does it imply that the products identified are necessarily the best available for the purpose. The research was supported in part by an appointment to A.S.G. to the Research Participation Program at the CDC administered by the Oak Ridge Institute for Science and Education through an interagency agreement between the US Department of Energy and the CDC.

Footnotes

Note: Supplementary information is available at http://www.nature.com/doifinder/10.1038/nbt.2403.

COMPETING FINANCIAL INTERESTS

The authors declare competing financial interests: details are available at http://www.nature.com/doifinder/10.1038/nbt.2403.

References

  • 1.Clinical and Laboratory Standards Institute (CLSI) Quality management system: a model for laboratory services; approved guideline 4th edn. (Document no. GP26–A4, Vol. 31, No. 15) Clinical and Laboratory Standards Institute; 2011. [Google Scholar]
  • 2.Centers for Medicare and Medicaid Services. Clinical laboratory improvement amendments of 1988 (Part 493) 1443–1495. US Department of Health and Human Services; 1988. [Google Scholar]
  • 3.International Organization for Standardization. ISO 15189 Medical laboratories—Particular requirements for quality and competence. International Organization for Standardization; 2007. [Google Scholar]
  • 4.Clinical and Laboratory Standards Institute (CLSI) Molecular Methods for Clinical Genetics and Oncology Testing ; Approved Guideline-Third Edition CLSI (Document MM01-A3, Vol 32, No. 7) Clinical and Laboratory Standards Institute; 2012. [Google Scholar]
  • 5.Durbin RM, et al. Nature. 2010;467:1061–1073. doi: 10.1038/nature09534. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Guidelines

RESOURCES