Noninvasive positive pressure ventilation in subjects with stable COPD: a randomized trial

Surya P Bhatt 1, Michael W Peterson 2, Jeffrey S Wilson 1, Lakshmi Durairaj 1,
PMCID: PMC3842217  PMID: 24293994

Abstract

Background

The use of domiciliary noninvasive positive pressure ventilation (NPPV) in stable chronic obstructive pulmonary disease (COPD) with chronic hypercapnic respiratory failure has yielded variable effects on survival, quality of life, and dyspnea. We hypothesized that use of NPPV in stable COPD and partial pressure of carbon dioxide (PaCO2) <52 mmHg might result in improvement in quality of life and dyspnea.

Methods

Thirty patients with stable COPD (forced expiratory volume in the first second <50% predicted and PaCO2 <52 mmHg) were prospectively randomized to receive domiciliary NPPV (bilevel positive airway pressure, 15/5 cm H2O) or usual therapy for 6 months. Measurements were made at baseline, 6 weeks, 3 months, and 6 months. Primary outcomes were quality of life as assessed by the Chronic Respiratory Disease Questionnaire (CRQ), and dyspnea as measured by the Transitional Dyspnea Index (TDI).

Results

Fifteen subjects in the NPPV arm and 12 controls completed all the study visits. At 6 weeks and 3 months, the NPPV arm showed significant improvement in TDI total score. However, this effect persisted only in the TDI-Task at 6 months (P=0.03). NPPV use was associated with a small improvement in the CRQ-Mastery domain (0.6 versus −0.1, P=0.04). The arterial partial pressure of oxygen (PaO2) in the control arm worsened over the period of the study, whereas it remained stable in the NPPV arm (change −7.2 mmHg versus +2.1 mmHg, respectively, P=0.02).

Conclusion

NPPV resulted in a small improvement in quality of life indices in stable COPD patients with PaCO2 <52 mmHg. Future larger studies will clarify the role of NPPV in this stable subgroup of patients with COPD.

Keywords: noninvasive, ventilation, chronic obstructive pulmonary disease, normocapnia

Introduction

Chronic obstructive pulmonary disease (COPD) is associated with a progressive decline in lung function and resultant chronic respiratory failure. This leads to significant rates of hospitalization, disability, and death.1 COPD is also associated with sleep-disordered breathing and nocturnal hypoventilation, resulting in nocturnal and daytime hypoxemia and hypercapnia.2,3 These, in turn, can contribute to fatigue, dyspnea, and impaired quality of life. Nocturnal noninvasive positive pressure ventilation (NPPV) can theoretically rest overloaded respiratory muscles, prevent nocturnal hypoventilation, and reset central respiratory drive in patients with hypercapnia. The use of NPPV in late-stage COPD appears logical to change this inexorable course, to alleviate symptoms, and to improve quality of life. While NPPV has a definite role in the management of acute hypercapnic respiratory failure,4 its role in the management of late-stage stable COPD is controversial. Multiple studies have shown no survival benefit for nocturnal NPPV in chronic stable hypercapnic COPD patients.510 The effects on quality of life and dyspnea scores in such patients have been inconsistent.811 The variable efficacy might have been due to late administration of the intervention in the course of disease, in the presence of advanced hypercapnic respiratory failure. We sought to assess the effect of applying NPPV on indices of health-related quality of life in patients with relative normocapnia. We hypothesized that patients with stable severe COPD and partial pressure of carbon dioxide (PaCO2) <52 mmHg have improved quality of life and reduced dyspnea with the use of NPPV.

Materials and methods

We conducted a prospective, randomized controlled study.12 Subjects were screened from the pulmonary clinic at the University of Iowa Hospital. The protocol was approved by the institutional review board (20010044) and informed consent was obtained from each subject.

Patients

Subjects with stable COPD, defined as a forced expiratory volume in the first second (FEV1)/forced vital capacity (FVC) ratio of <0.70, with at least a 10 pack-year smoking history and PaCO2 <52 mmHg were eligible.13 Stable COPD was defined as no exacerbations in the 4 weeks prior to initiation of the study. Only patients with a low clinical probability of having obstructive sleep apnea as assessed using the Berlin Questionnaire were included. This was done to remove the possible confounding effect of sleep apnea.14 Those with diagnoses of congestive heart failure, obstructive sleep apnea, chronic respiratory conditions other than COPD, age <35 years, any disease limiting life expectancy to less than 2 years, active malignancy other than nonmelanotic skin cancer in the previous 2 years, and with any anatomic variation or disease process that precluded wearing a nasal mask were excluded. Subjects could be active or ex smokers, but had to have a stable smoking status for at least 6 months prior to study initiation. Patients could also have undergone pulmonary rehabilitation or be in a maintenance program, but they could not initiate a pulmonary rehabilitation program during the 6-month observation period.

Subjects were randomized to receive or not receive domiciliary NPPV at 15/5 cm H2O. Subjects were randomized in a 1:1 ratio to receive NPPV versus control treatment using a random number generator. Randomization was done using opaque sealed envelopes which were opened during screening visits. For this study, we defined hypercapnia as ≥52 mmHg based on the criteria used for reimbursement for noninvasive ventilation by the Centers for Medicare and Medicaid Services.15 Once subjects were randomized to the NPPV arm, they were initiated on NPPV at home by a respiratory therapist. Subjects were fitted with a nasal-oral contour comfort gel full face mask (Respironics Inc, Murrysville, PA, USA) or nasal pillows (Nellcor Puritan-Bennett Inc, Pleasanton, CA, USA) as needed for optimal comfort and minimal leak. NPPV was initiated at home using a BiPAP® Synchrony ventilator (Respironics Inc) and titrated to final pressures of 15/5 cm H2O within a week. Subjects were instructed to use the ventilator all night or for at least 6 hours every night for the next 6 months. Respiratory therapists called the subjects every day during the first week to ensure optimal usage and to troubleshoot any complications. The therapists also made a home visit to check the use of domiciliary NPPV periodically during the first week and as needed thereafter. All subjects were provided with heated humidification.

A complete history and physical examination was undertaken, with a posterolateral chest radiograph, 12-lead electrocardiogram, lung physiology (spirometry, arterial blood gases on room air), and respiratory muscle strength (negative inspiratory force [NIF]) at baseline. We measured exercise capacity by the six-minute walk test.16 Dyspnea was assessed using the Baseline Dyspnea Index and the Transitional Dyspnea Index (TDI).17 We included a global assessment of health-related quality of life developed for patients with chronic lung disease, ie, the Chronic Respiratory Disease Questionnaire (CRQ).18 Sleep quality was assessed using the Pittsburgh Sleep Quality Index.19 Subjects were reassessed at 6 weeks, 3 months, and 6 months after randomization. The primary outcome was change in the CRQ domains and change in TDI. The minimal clinically important difference in CRQ was defined as a change in the score by 0.5 units,18 and that in TDI by one unit.17 Secondary outcomes were change in NIF, arterial partial pressure of oxygen (PaO2), FEV1, and six-minute walk test. Ventilator use data were downloaded using Encore software (Respironics Inc) at each follow-up visit.

Side effects and compliance

At each visit, subjects on NPPV were asked about side effects, such as dryness of eyes, sinus congestion, skin abrasions, and epistaxis. Compliance was assessed by both patient-reported hours of use at each visit and machine-downloaded data for hours of use. Data obtained from the machine were used for analyses.

Statistical analysis

Results are expressed as the mean (standard deviation) for normally distributed data and median with interquartile range for nonuniformly distributed data. The baseline characteristics of subjects in the NPPV and control arms were compared using the Student’s t-test for continuous variables and the Mann–Whitney U test as appropriate, and the chi-square test for categorical variables. Differences within groups at the end of 6 months were compared with baseline using the paired t-test and between groups using the independent t-test. Repeated-measures analysis of variance was used to assess significant changes over time, and the independent t-test was used to assess where any change occurred between groups.

Distribution of data was assessed using Mauchly’s test of sphericity, and Greenhouse-Geisser correction was used where sphericity could not be assumed. Relationships between outcomes and change from baseline to 6 months in variables that could affect outcomes were assessed using Spearman’s correlation. A P-value<0.05 was considered to be statistically significant. Multivariate analyses for outcomes after adjusting for baseline PaCO2 and NIF were done and showed no change in the results, and hence are not reported. We did not make any sample size justification because this was a pilot study in relatively normocapnic subjects. All analyses were performed using Statistical Package for the Social Sciences version 11.5 software (SPSS Inc., Chicago, IL, USA).

Results

Thirty subjects with COPD were recruited. Twenty-seven completed all assessments, ie, 15 in the NPPV group and 12 in the control group (Figure 1). The baseline characteristics of subjects who were able to complete the study are shown in Table 1. There were a higher proportion of males in the control group. The two groups were well matched for age, body mass index, disease severity, and baseline indices of dyspnea, functional capacity, quality of life, and sleep indices. At 6 weeks and 3 months, NPPV resulted in a significant improvement in TDI in the NPPV arm (data in Figure 2, Table S1). However, this effect persisted only in the TDI-Task at 6 months (−0.8 versus −1.8, P=0.03). Table 2 shows the effects of NPPV at the end of 6 months. NPPV did not result in improvement in global quality of life as assessed by the total CRQ score. Subjects using NPPV did experience a small improvement in the mastery domain of the CRQ, a difference in the realm of being clinically significant (0.6 versus −0.1, P=0.04, Figure 2). When assessed by repeated measures over time, only the improvement in the mastery domain of CRQ was sustained (data in Table S1).

Figure 1.

Figure 1

Randomization and follow-up of patients in intervention and control arms.

Abbreviations: COPD, chronic obstructive pulmonary disease; FEV1, forced expiratory volume in the first second; FVC, forced vital capacity; PaCO2, partial pressure of carbon dioxide in arterial blood; NPPV, noninvasive positive pressure ventilation.

Table 1.

Comparison of baseline characteristics in cases and controls

Variable NPPV (n=15) Controls (n=12)
Age (years) 70 (66–73) 68 (65–78)
Sex (% male) 8 (53) 12 (100)*
BMI (kg/m2) 24.8 (2.8) 24.8 (4.8)
Pack-years of smoking 59 (29) 61 (30)
FEV1 %pred 30.3 (7) 29.6 (7.4)
FVC %pred 68.8 (14.1) 59.4 (8)
NIF (cm H2O) −48.9 (14.5) −63.2 (23.9)
PaCO2 (mmHg) 42.4 (5.6) 41.9 (4.6)
PaO2 (mmHg) 65.1 (13.3) 73.9 (9.9)
6MWD (m) 284 (101) 321 (91)
CRQ score
Dyspnea 3.8 (1) 3.7 (1)
Fatigue 4.3 (1) 4.1 (1)
Emotional function 5.5 (0.7) 5.3 (1.1)
Mastery 5 (1.1) 5.3 (1.4)
Dyspnea
TDI-Function 1.9 (0.7) 1.5 (0.7)
TDI-Task 1.3 (0.5) 1.4 (0.7)
TDI-Effort 1.3 (0.5) 1.3 (0.5)
TDI total 4.5 (1.6) 4.2 (1.5)
PSQI 3.7 (3) 6.1 (3.2)

Notes: All results are expressed as the mean (standard deviation) except for median (interquartile range).

*

P<0.05.

Abbreviations: FEV1, forced expiratory volume in the first second; %pred, percent predicted; FVC, forced vital capacity; NIF, negative inspiratory force; NPPV, noninvasive positive pressure ventilation; PaCO2, partial pressure of carbon dioxide in arterial blood; PaO2, partial pressure of oxygen in arterial blood; 6MWD, six-minute walk distance; CRQ, Chronic Respiratory Disease Questionnaire; TDI, Transitional Dyspnea Index at baseline; PSQI, Pittsburgh Sleep Quality Index; BMI, body mass index.

Figure 2.

Figure 2

Change in dyspnea measured by TDI and quality of life (CRQ-Mastery domain) over time in intervention and control arms. Dots represent control arm; squares represent intervention arm.

Notes: *P<0.05 and **P<0.01.

Abbreviations: TDI, Transitional Dyspnea Index; CRQ, Chronic Respiratory Disease Questionnaire; NPPV, noninvasive positive pressure ventilation.

Table 2.

Comparison of outcomes between intervention and control arms at 6 months

NPPV (n=15)
Controls (n=12)
P-value
Baseline 6 months Difference Baseline 6 months Difference
FEV1 %pred 30.3 (7) 31.2 (11.9) 0.9 29.6 (7.4) 31 (7.8) 1.4 0.84
FVC %pred 68.8 (14.1) 71.9 (16.7) 3.1 59.4 (8) 60.8 (30.2) 1.3 0.74
NIF (cm H2O) −48.9 (14.5) −55.9 (20.3) 7 −63.2 (23.9) −63 (21.9) 0.3 0.14
PaCO2 (mmHg) 42.4 (5.6) 44.3 (8.5) 1.9 41.9 (4.6) 42.9 (4.5) 1 0.68
PaO2 (mmHg)* 65.1 (13.3) 67.2 (14.4) 2.1 73.9 (9.9) 66.8 (9.9) −7.2 0.02
6MWD (m) 284 (101) 280 (142) −14 321 (91) 304 (129) −17 0.73
CRQ score
Dyspnea 3.8 (1) 4.1 (1.2) 0.3 3.7 (1) 3.4 (0.9) −0.3 0.26
Fatigue 4.3 (1) 4.6 (1.3) 0.4 4.1 (1) 4.2 (0.9) 0 0.36
Emotional function 5.5 (0.7) 5.6 (1.1) 0.2 5.3 (1.1) 5.3 (1.3) 0 0.46
Mastery* 5 (1.1) 5.6 (1) 0.6 5.3 (1.4) 5.2 (1.2) −0.1 0.04
Dyspnea
TDI-Function 1.9 (0.7) 0 (0.8) −1.9 1.5 (0.7) −0.2 (0.7) −1.7 0.63
TDI-Task* 1.3 (0.5) 0.5 (0.7) −0.8 1.4 (0.7) −0.3 (0.7) −1.8 0.03
TDI-Effort 1.3 (0.5) 0.5 (0.7) −0.8 1.3 (0.5) −0.3 (0.7) −1.6 0.06
TDI total 4.5 (1.6) 0.9 (2) −3.4 4.2 (1.5) −0.5 (2.3) −4.7 0.35
PSQI 3.7 (3) 3.4 (2.3) 0.2 6.1 (3.3) 5.7 (3.2) 0.4 0.77

Notes: All results are expressed as the mean (standard deviation).

*

P<0.05.

Abbreviations: FEV1, forced expiratory volume in the first second; %pred, percent predicted; FVC, forced vital capacity; NIF, negative inspiratory force; PaCO2, partial pressure of carbon dioxide in arterial blood; PaO2, partial pressure of oxygen in arterial blood; 6MWD, six-minute walk distance; CRQ, Chronic Respiratory disease Questionnaire; TDI, Transitional Dyspnea Index at baseline; PSQI, Pittsburgh Sleep Quality Index.

The oxygenation status of controls worsened over the period of the study (−7.2 mmHg), whereas NPPV resulted in conservation of oxygen levels (+2.1 mmHg, P=0.02). There was a trend towards improvement in NIF (delta 7 with NPPV versus delta 0.3 in controls), but this did not reach statistical significance (P=0.14). There was no change in PaCO2. There was also no improvement in six-minute walk distance. NPPV did not result in a change in FEV1. There was no difference in the number of exacerbations in either group during the period of the study, with one exacerbation in each group.

Compliance and side effects

The average duration of NPPV use per night was about 3.1 hours (Table 3). More than 40% used it for more than 4 hours per night. More than half (8/15) used it for more than 4 hours in the first week (initial adopters). Of those using NPPV for more than 4 hours per night in the first week, 75% used NPPV for a mean of more than 4 hours per night for the entirety of the study. Eighty-eight percent of initial adopters expressed interest in continuing use of NPPV beyond the duration of the study. Even though there were some adverse effects from using NPPV, these were minimal (Table 4). There was no change in the quality of sleep (Table 2). Overall, 53% of subjects indicated a willingness to continue using NPPV beyond the period of the study. Interestingly, the improvement in CRQ-Mastery correlated significantly with the average number of hours of NPPV use (Spearman’s r=−0.83, P<0.001, Figure 3). More side effects were reported by subjects in the NPPV arm, although this did not preclude them from continued use of the intervention (Table 4).

Table 3.

Compliance with noninvasive positive pressure ventilation

Days used 80.6 (69.1)
% days used 42 (36)
Number of hours used (average per night), patient-reported 3.9 (3.4)
Number of hours used (average per night), machine-reported 3.1 (3.3)
% use greater than 4 hours per night 40%
Initial compliance (usage >4 hours/night in first week) 8 (53%)
Number indicating willingness to continue using device (%) 8 (53)

Note: All numbers are expressed as the mean (standard deviation).

Table 4.

Side effects

NPPV (n=15) Controls (n=12)
Dryness of eyes 1 (1) 1 (0)
Sinus congestion 5 (2) 0
Nose bleeds 1 (0) 1 (0)
Discomfort 5 (3) 0
Skin breaks 1 (1) 0

Notes: All numbers expressed as total number experiencing symptoms any time during course of study; numbers in brackets indicate those who had persistent symptoms at 6 months.

Abbreviation: NPPV, noninvasive positive pressure ventilation.

Figure 3.

Figure 3

Correlation between change in CRQ-Mastery domain and average numbers of hours of use of noninvasive positive pressure ventilation.

Abbreviation: CRQ, Chronic Respiratory Disease Questionnaire.

Discussion

We showed that there was a small improvement in some quality of life indices in patients with stable COPD administered NPPV for 6 months. There was a significant improvement in CRQ-Mastery and TDI-Task over time with application of NPPV. There was also a beneficial effect on PaO2.

The role of domiciliary nocturnal NPPV in the management of chronic stable COPD is controversial. The current literature is conflicting, but seems to favor its use in patients with hypercapnia, and with high intensity, targeting lowering PaCO2 levels. While multiple retrospective observational studies have shown a survival benefit for nocturnal NPPV in chronic stable hypercapnic COPD patients,57,20 these results have not been supported in randomized controlled trials.810 These, and other smaller studies, also found inconsistent effects on sleep quality, quality of life, and dyspnea scores in hypercapnic COPD patients using NPPV.811,2123 Most of these trials attempted to use varying intensity of NPPV to reduce daytime hypercapnia and improve gas exchange, and seemed to indicate that adequate lowering of PaCO2 and nocturnal hypoventilation was associated with the most benefits.24 It is possible that the inconsistent effects of NPPV were due to usage of low levels of pressure. Dreher et al showed in a short-duration study that high-intensity NPPV with controlled pressure can significantly reduce nocturnal PaCO2, improve dyspnea scores, lung function tests, and quality of life.25 Based on these, Medicare and Medicaid Services recommended a cutoff of ≥52 mmHg, in addition to other criteria, for initiating domiciliary nocturnal NPPV in stable hypercapnic patients.15 In contrast with these recommendations, we hypothesized that the inconsistent effects seen in previous studies might have been due to the intervention being late in the course of disease.

In patients without hypercapnia, NPPV cannot be targeted toward PaCO2, but might still benefit patients by lowering the work of breathing. Although ours is the first study to target this group specifically, other studies did include subjects who had similar PaCO2 levels (Table 5). To our knowledge, no study has compared the effect of NPPV in normocapnic COPD with controls not undergoing another intervention such as rehabilitation and domiciliary oxygen that could confound the results.

Table 5.

Comparison of randomized studies of domiciliary NPPV in chronic stable COPD without hypercapnea

n Duration of follow-up NPPV intensity (cm H2O) PaCO2 (mmHg) Quality of life Dyspnea Sleep Exacerbations Survival
Garrod et al11 (versus rehabilitation) Cases (n=23) Controls (n=22) 8 weeks 16/4 44.2 CRQ improved CRQ dyspnea = NS
Casanova et al8 (versus domiciliary O2) 20 (24) One year 12/4 50.7 Improvement in psychomotor coordination Borg score improved NS NS
Lin et al23 12 (crossover) 2 weeks 12/– 50.5 Sleep worsened
Strumpf et al21 7 (crossover) 3 months 15/– 50 NS NS
Current study 15 (12) 6 months 15/5 42 CRQ-Mastery improved TDI-Task improved NS NS

Abbreviations: COPD, chronic obstructive pulmonary disease; NPPV, noninvasive positive pressure ventilation; PaCO2, partial pressure of carbon dioxide in arterial blood; CRQ, Chronic Respiratory Disease Questionnaire; O2, oxygen; TDI, Transitional Dyspnea Index; NS, no significant change.

NPPV use was associated with some minor benefits in our study in the domain of mastery, and in preservation of PaO2. This supports the results seen in two other short-duration studies aiming to recruit hypercapnic patients, but with PaCO2 levels comparable with our inclusion criteria.21,23 The difference in oxygenation levels was a surprising finding, and is most likely due to NPPV leading to redistribution of airflow toward areas of better perfusion, resulting in improvement in ventilation/perfusion matching.26 Results from short-duration trials cannot be generalized because there can be a significant placebo effect from the mask,27 and there can also be significant problems with long-term compliance.28 A longer-duration randomized study by Casanova et al did not have PaCO2 level as an inclusion criterion but had a mean PaCO2 of 50.7 mmHg. In this study, there was a marginal improvement in Borg dyspnea score and in psychomotor coordination.8 Only the study by Garrod et al had PaCO2 levels comparable with those seen in our study. While they did show an improvement in exercise capacity, oxygenation, and CRQ, the study compared the effect of adding NPPV to an exercise program with exercise alone.11 We showed a change in some quality of life indices, but these were limited by the small number of subjects. In addition to baseline PaCO2 and improvement in hypoventilation, the other major factor that can affect study results is the intensity of NPPV used. Both these comparable randomized controlled studies used inspiratory and expiratory pressures similar to those used in the current study.8,11

Our study has a number of strengths. We studied a group of patients who were not significantly hypercapnic, which is the usual target group for this intervention. We excluded patients with sleep apnea because improvement in sleep can potentially result in improvement in quality of life scores, independent of COPD-related improvement. We meticulously assessed side effects resulting from use of the NPPV mask. There were also some limitations to our study. We did not have a sham arm, and some of the results could be due to a placebo effect. The control subjects were all male, and while we are not aware of any gender differences in response to NPPV, this may have played a role. This study was underpowered to assess end points such as mortality or exacerbation frequency. However, those were not the main aims of our study. Compliance was low, probably due to lack of titration of pressures to each subject, lack of a run-in period, and the nature of the intervention. However, this is not unique to our study.11

In conclusion, use of NPPV is associated with minor improvements in PaO2 and the subdomains of quality of life and dyspnea indices. Whether attainment of the minimal clinically important difference of 0.5 for the mastery domain is worth the cost and inconvenience of NPPV can only be determined by each patient based on cost and tolerance. Nevertheless, this study found a significant difference in this domain. Based on this small study, NPPV has some benefit in COPD patients with PaCO2 <52 mmHg. Better compliance may be associated with improved outcome benefits in this group. Future large studies or meta-analyses in this group will clarify the utility of this intervention.

Supplementary material

Table S1.

Comparison between groups at each time point and over all time points (repeated-measures analysis of variance)

Baseline
6 weeks
3 months
6 months
P-value
NPPV Controls NPPV Controls NPPV Controls NPPV Controls
FEV1 %pred 30.3 (7) 29.6 (7.4) 32.2 (11.8) 31.5 (6.4) 31.2 (11.9) 31 (7.8) 0.81
FVC %pred 68.8 (14.1) 59.4 (8) 74.2 (19.2) 62.8 (8.8) 71.9 (16.7) 60.8 (30.2) 0.97
NIF (cm H2O) −48.9 (14.5) −63.2 (23.9) −59.7 (24.1) −64.5 (23.9) −55.9 (20.3) −63 (21.9) 0.16
PaCO2 (mmHg) 42.4 (5.6) 41.9 (4.6) 44.1 (7) 42.7 (4) 44.3 (8.5) 42.9 (4.5) 0.92
PaO2 (mmHg) 65.1 (13.3) 73.9 (9.9) 62.1 (15.2) 66.7 (10.5) 67.2 (14.4) 66.8 (9.9) 0.12
6MWD (m) 284 (101) 321 (91) 312 (110) 320 (94) 295 (101) 335 (90) 280 (142) 304 (129) 0.50
CRQ score
Dyspnea 3.8 (1) 3.7 (1) 4.1 (0.9) 3.7 (1.4) 3.9 (0.8) 3.6 (1.3) 4.1 (1.2) 3.4 (0.9) 0.64
Fatigue 4.3 (1) 4.1 (1) 4.5 (0.9) 3.9 (1.2) 4.4 (1) 4.1 (1.1) 4.6 (1.3) 4.2 (0.9) 0.39
Emotional function 5.5 (0.7) 5.3 (1.1) 5.6 (0.7) 5.1 (1.1) 5.5 (0.9) 5.2 (1.3) 5.6 (1.1) 5.3 (1.3) 0.37
Mastery* 5 (1.1) 5.3 (1.4) 5.5 (1) 5 (1.6) 5.6 (1) 5.2 (1.2) 5.6 (1) 5.2 (1.2) 0.047
Dyspnea
TDI-Function 1.9 (0.7) 1.5 (0.7) 0.6 (0.9)* −0.3 (1.1) 0.4 (0.8)* −0.2 (0.4) 0 (0.8) −0.2 (0.7) 0.31
TDI-Task 1.3 (0.5) 1.4 (0.7) 0.3 (0.5) −0.3 (0.6) 0.4 (0.5)* −0.3 (0.9) 0.5 (0.7)* −0.3 (0.7) 0.07
TDI-Effort 1.3 (0.5) 1.3 (0.5) 0.3 (0.5)* −0.3 (0.6) 0.4 (0.5)* −0.1 (0.7) 0.5 (0.7)* −0.3 (0.7) 0.13
TDI total 4.5 (1.6) 4.2 (1.5) 1.3 (1.9)** −0.8 (1.9) 1.3 (1.7)* −0.3 (0.9) 0.9 (2) −0.5 (2.3) 0.28

Notes: All results expressed as the mean (standard deviation).

*

P<0.05 and

**

P<0.01.

Abbreviations: FEV1, forced expiratory volume in the first second; %pred, percent predicted; FVC, forced vital capacity; NIF, negative inspiratory force; NPPV, noninvasive positive pressure ventilation; PaCO2, partial pressure of carbon dioxide in arterial blood; PaO2, partial pressure of oxygen in arterial blood; 6MWD, six-minute walk distance; CRQ, Chronic respiratory Disease Questionnaire; TDI, Transitional Dyspnea Index at baseline.

Acknowledgments

We extend our gratitude to the study subjects, to Charles Dayton for help with the institutional review board approval process, and to Bridget Zimmerman for statistical assistance.

Author contributions

SPB analyzed and interpreted the data, and wrote the manuscript. MWP and JSW contributed to the study conception and design, data collection, and revised the manuscript for important intellectual content. LD contributed to the study design, recruitment, data collection, data analyses, and interpretation and writing the manuscript. LD has full access to all data in the study and acts as the guarantor for accuracy of the manuscript.

Disclosure

The study was sponsored by Respironics Inc, which had no input in the design, data, or writing of the paper. Otherwise none of the authors have any conflicts of interest to declare.

References

  • 1.Deaths from chronic obstructive pulmonary disease – United States, 2000–2005. MMWR Morb Mortal Wkly Rep. 2008;57(45):1229–1232. [No authors listed] [PubMed] [Google Scholar]
  • 2.Catterall JR, Douglas NJ, Calverley PM, et al. Transient hypoxemia during sleep in chronic obstructive pulmonary disease is not a sleep apnea syndrome. Am Rev Respir Dis. 1983;128(1):24–29. doi: 10.1164/arrd.1983.128.1.24. [DOI] [PubMed] [Google Scholar]
  • 3.O’Donoghue FJ, Catcheside PG, Ellis EE, et al. Sleep hypoventilation in hypercapnic chronic obstructive pulmonary disease: prevalence and associated factors. Eur Respir J. 2003;21(6):977–984. doi: 10.1183/09031936.03.00066802. [DOI] [PubMed] [Google Scholar]
  • 4.Keenan SP, Mehta S. Noninvasive ventilation for patients presenting with acute respiratory failure: the randomized controlled trials. Respir Care. 2009;54(1):116–126. [PubMed] [Google Scholar]
  • 5.Jones SE, Packham S, Hebden M, Smith AP. Domiciliary nocturnal intermittent positive pressure ventilation in patients with respiratory failure due to severe COPD: long-term follow up and effect on survival. Thorax. 1998;53(6):495–498. doi: 10.1136/thx.53.6.495. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Budweiser S, Hitzl AP, Jorres RA, et al. Impact of noninvasive home ventilation on long-term survival in chronic hypercapnic COPD: a prospective observational study. Int J Clin Pract. 2007;61(9):1516–1522. doi: 10.1111/j.1742-1241.2007.01427.x. [DOI] [PubMed] [Google Scholar]
  • 7.Windisch W, Kostic S, Dreher M, Virchow JC, Jr, Sorichter S. Outcome of patients with stable COPD receiving controlled noninvasive positive pressure ventilation aimed at a maximal reduction of Pa(CO2) Chest. 2005;128(2):657–662. doi: 10.1378/chest.128.2.657. [DOI] [PubMed] [Google Scholar]
  • 8.Casanova C, Celli BR, Tost L, et al. Long-term controlled trial of nocturnal nasal positive pressure ventilation in patients with severe COPD. Chest. 2000;118(6):1582–1590. doi: 10.1378/chest.118.6.1582. [DOI] [PubMed] [Google Scholar]
  • 9.Clini E, Sturani C, Rossi A, et al. The Italian multicentre study on non-invasive ventilation in chronic obstructive pulmonary disease patients. Eur Respir J. 2002;20(3):529–538. doi: 10.1183/09031936.02.02162001. [DOI] [PubMed] [Google Scholar]
  • 10.McEvoy RD, Pierce RJ, Hillman D, et al. Nocturnal non-invasive nasal ventilation in stable hypercapnic COPD: a randomised controlled trial. Thorax. 2009;64(7):561–566. doi: 10.1136/thx.2008.108274. [DOI] [PubMed] [Google Scholar]
  • 11.Garrod R, Mikelsons C, Paul EA, Wedzicha JA. Randomized controlled trial of domiciliary noninvasive positive pressure ventilation and physical training in severe chronic obstructive pulmonary disease. Am J Respir Crit Care Med. 2000;162(4 Pt 1):1335–1341. doi: 10.1164/ajrccm.162.4.9912029. [DOI] [PubMed] [Google Scholar]
  • 12.University of Iowa Trial of Non-invasive Ventilation for Stable COPD Available from: http://clinicaltrials.gov/show/NCT01722773. NLM identifier: NCT01722773Accessed October 24, 2013
  • 13.American Thoracic Society/European Respiratory Society Task Force Standards for the Diagnosis and Management of Patients with COPD. Version 1.2 New York, NY: American Thoracic Society; 2004Available from: http://www.thoracic.org/clinical/copd-guidelines/Accessed October 18, 2013 [Google Scholar]
  • 14.Netzer NC, Stoohs RA, Netzer CM, Clark K, Strohl KP. Using the Berlin Questionnaire to identify patients at risk for the sleep apnea syndrome. Ann Intern Med. 1999;131(7):485–491. doi: 10.7326/0003-4819-131-7-199910050-00002. [DOI] [PubMed] [Google Scholar]
  • 15.US Department of Health and Human Services, Center for Medicare and Medicaid Services . Medicare Coverage Issues Manual §609. Baltimore, MD: CMS; 2002. Durable medical equipment reference list. [Google Scholar]
  • 16.ATS Committee on Proficiency Standards for Clinical Pulmonary Function Laboratories ATS statement: guidelines for the six-minute walk test. Am J Respir Crit Care Med. 2002;166(1):111–117. doi: 10.1164/ajrccm.166.1.at1102. [DOI] [PubMed] [Google Scholar]
  • 17.Witek TJ, Jr, Mahler DA. Minimal important difference of the transition dyspnoea index in a multinational clinical trial. Eur Respir J. 2003;21(2):267–272. doi: 10.1183/09031936.03.00068503a. [DOI] [PubMed] [Google Scholar]
  • 18.Guyatt GH, Berman LB, Townsend M, Pugsley SO, Chambers LW. A measure of quality of life for clinical trials in chronic lung disease. Thorax. 1987;42(10):773–778. doi: 10.1136/thx.42.10.773. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Buysse DJ, Reynolds CF, 3rd, Monk TH, Berman SR, Kupfer DJ. The Pittsburgh Sleep Quality Index: a new instrument for psychiatric practice and research. Psychiatry Res. 1989;28(2):193–213. doi: 10.1016/0165-1781(89)90047-4. [DOI] [PubMed] [Google Scholar]
  • 20.Windisch W, Haenel M, Storre JH, Dreher M. High-intensity non-invasive positive pressure ventilation for stable hypercapnic COPD. Int J Med Sci. 2009;6(2):72–76. doi: 10.7150/ijms.6.72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Strumpf DA, Millman RP, Carlisle CC, et al. Nocturnal positive-pressure ventilation via nasal mask in patients with severe chronic obstructive pulmonary disease. Am Rev Respir Dis. 1991;144(6):1234–1239. doi: 10.1164/ajrccm/144.6.1234. [DOI] [PubMed] [Google Scholar]
  • 22.Meecham Jones DJ, Paul EA, Jones PW, Wedzicha JA. Nasal pressure support ventilation plus oxygen compared with oxygen therapy alone in hypercapnic COPD. Am J Respir Crit Care Med. 1995;152(2):538–544. doi: 10.1164/ajrccm.152.2.7633704. [DOI] [PubMed] [Google Scholar]
  • 23.Lin CC. Comparison between nocturnal nasal positive pressure ventilation combined with oxygen therapy and oxygen monotherapy in patients with severe COPD. Am J Respir Crit Care Med. 1996;154(2 Pt 1):353–358. doi: 10.1164/ajrccm.154.2.8756806. [DOI] [PubMed] [Google Scholar]
  • 24.Wijkstra PJ. Non-invasive positive pressure ventilation (NIPPV) in stable patients with chronic obstructive pulmonary disease (COPD) Respir Med. 2003;97(10):1086–1093. doi: 10.1016/s0954-6111(03)00163-x. [DOI] [PubMed] [Google Scholar]
  • 25.Dreher M, Storre JH, Schmoor C, Windisch W. High-intensity versus low-intensity non-invasive ventilation in patients with stable hypercapnic COPD: a randomised crossover trial. Thorax. 2010;65(4):303–308. doi: 10.1136/thx.2009.124263. [DOI] [PubMed] [Google Scholar]
  • 26.De Backer L, Vos W, Dieriks B, et al. The effects of long-term noninvasive ventilation in hypercapnic COPD patients: a randomized controlled pilot study. Int J Chron Obstruct Pulmon Dis. 2011;6:615–624. doi: 10.2147/COPD.S22823. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Jenkinson C, Davies RJ, Mullins R, Stradling JR. Comparison of therapeutic and subtherapeutic nasal continuous positive airway pressure for obstructive sleep apnoea: a randomised prospective parallel trial. Lancet. 1999;353(9170):2100–2105. doi: 10.1016/S0140-6736(98)10532-9. [DOI] [PubMed] [Google Scholar]
  • 28.Criner GJ, Brennan K, Travaline JM, Kreimer D. Efficacy and compliance with noninvasive positive pressure ventilation in patients with chronic respiratory failure. Chest. 1999;116(3):667–675. doi: 10.1378/chest.116.3.667. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1.

Comparison between groups at each time point and over all time points (repeated-measures analysis of variance)

Baseline
6 weeks
3 months
6 months
P-value
NPPV Controls NPPV Controls NPPV Controls NPPV Controls
FEV1 %pred 30.3 (7) 29.6 (7.4) 32.2 (11.8) 31.5 (6.4) 31.2 (11.9) 31 (7.8) 0.81
FVC %pred 68.8 (14.1) 59.4 (8) 74.2 (19.2) 62.8 (8.8) 71.9 (16.7) 60.8 (30.2) 0.97
NIF (cm H2O) −48.9 (14.5) −63.2 (23.9) −59.7 (24.1) −64.5 (23.9) −55.9 (20.3) −63 (21.9) 0.16
PaCO2 (mmHg) 42.4 (5.6) 41.9 (4.6) 44.1 (7) 42.7 (4) 44.3 (8.5) 42.9 (4.5) 0.92
PaO2 (mmHg) 65.1 (13.3) 73.9 (9.9) 62.1 (15.2) 66.7 (10.5) 67.2 (14.4) 66.8 (9.9) 0.12
6MWD (m) 284 (101) 321 (91) 312 (110) 320 (94) 295 (101) 335 (90) 280 (142) 304 (129) 0.50
CRQ score
Dyspnea 3.8 (1) 3.7 (1) 4.1 (0.9) 3.7 (1.4) 3.9 (0.8) 3.6 (1.3) 4.1 (1.2) 3.4 (0.9) 0.64
Fatigue 4.3 (1) 4.1 (1) 4.5 (0.9) 3.9 (1.2) 4.4 (1) 4.1 (1.1) 4.6 (1.3) 4.2 (0.9) 0.39
Emotional function 5.5 (0.7) 5.3 (1.1) 5.6 (0.7) 5.1 (1.1) 5.5 (0.9) 5.2 (1.3) 5.6 (1.1) 5.3 (1.3) 0.37
Mastery* 5 (1.1) 5.3 (1.4) 5.5 (1) 5 (1.6) 5.6 (1) 5.2 (1.2) 5.6 (1) 5.2 (1.2) 0.047
Dyspnea
TDI-Function 1.9 (0.7) 1.5 (0.7) 0.6 (0.9)* −0.3 (1.1) 0.4 (0.8)* −0.2 (0.4) 0 (0.8) −0.2 (0.7) 0.31
TDI-Task 1.3 (0.5) 1.4 (0.7) 0.3 (0.5) −0.3 (0.6) 0.4 (0.5)* −0.3 (0.9) 0.5 (0.7)* −0.3 (0.7) 0.07
TDI-Effort 1.3 (0.5) 1.3 (0.5) 0.3 (0.5)* −0.3 (0.6) 0.4 (0.5)* −0.1 (0.7) 0.5 (0.7)* −0.3 (0.7) 0.13
TDI total 4.5 (1.6) 4.2 (1.5) 1.3 (1.9)** −0.8 (1.9) 1.3 (1.7)* −0.3 (0.9) 0.9 (2) −0.5 (2.3) 0.28

Notes: All results expressed as the mean (standard deviation).

*

P<0.05 and

**

P<0.01.

Abbreviations: FEV1, forced expiratory volume in the first second; %pred, percent predicted; FVC, forced vital capacity; NIF, negative inspiratory force; NPPV, noninvasive positive pressure ventilation; PaCO2, partial pressure of carbon dioxide in arterial blood; PaO2, partial pressure of oxygen in arterial blood; 6MWD, six-minute walk distance; CRQ, Chronic respiratory Disease Questionnaire; TDI, Transitional Dyspnea Index at baseline.