Multivariate Chemometric Assisted Analysis of Metformin Hydrochloride, Gliclazide and Pioglitazone Hydrochloride in Bulk Drug and Dosage Forms

Radhika Bhaskar 1, Rahul Bhaskar 1,*, Mahendra K Sagar 1, Vipin Saini 1
PMCID: PMC3846061  PMID: 24312816

Abstract

Purpose: In this work a numerical method, based on the use of spectrophotometric data coupled to partial least squares (PLS) regression and net analyte preprocessing combined with classical least square (NAP/CLS) multivariate calibration, is reported for the simultaneous determination of metformin hydrochloride (MET), gliclazide (GLZ) and pioglitazone hydrochloride (PIO) in synthetic samples and combined commercial tablets. Methods: Spectra of MET, GLZ and PIO were recorded at concentrations within their linear ranges (5-25 µg/ml, 0.5-8 µg/ml and 0.5-3 µg/ml respectively) and were used to compute a total of 25 synthetic mixtures involving 15 calibration and 10 validation sets between wavelength range of 200 and 400 nm in 0.1N HCl. The suitability of the models was decided on the basis of root mean square error (RMSE) values of calibration and validation data. Results: The analytical performances of these chemometric methods were characterized by relative prediction errors and recovery studies (%) and were compared with each other. These two methods were successfully applied to pharmaceutical formulation, tablet, with no interference with excipients as indicated by the recovery study results. Mean recoveries of the commercial formulation set together with the figures of merit (calibration sensitivity, selectivity, limit of detection, limit of quantification etc.) were estimated. Conclusion: The proposed methods are simple, rapid and can be easily used as an alternative analysis tool in the quality control of drugs and formulation.

Keywords: Partial least-squares, Spectroscopy, Metformin, Gliclazide, Pioglitazone

Introduction

Oral ingestion has long been the most convenient and commonly employed route of drug delivery due to its ease of administration, least aseptic constraints and flexibility in the design of the dosage form. It is well known that modified release dosage forms may offer one or more advantages over immediate release formulations of the same drug. There are many ways to design modified release dosage forms for oral administration; from film coated pellets, tablets or capsules to more sophisticated and complicated delivery systems such as osmotically driven systems, systems controlled by ion exchange mechanism, systems using three dimensional printing technology and systems using electrostatic deposition technology. The design of modified release drug product is usually intended to optimize a therapeutic regimen by providing slow and continuous delivery of drug over the entire dosing interval whilst also providing greater patient compliance and convenience.1

MET, GLZ and PIO are active principles widely used and frequently combined in pharmaceutical preparation. All these three drugs are complimentary to each other. GLZ being an insulin secretagogue helps in insulin secretion from pancreas2whereas; insulin secreted under GLZ influence can be utilized by MET for its action. MET not only utilizes the insulin secreted under gliclazide influence but also converts from peripheral tissues.3 Drawbacks associated with GLZ are weight gain and hypoglycemia.4 This can easily be overcome by MET. PIO on the other hand is basically responsible for eliminating the problem of insulin resistance occurred on long term uses of sulphonyl ureas.5

For the treatment of diabetes mellitus the usual combination of drugs which are available in the market consists of MET and PIO and/or glipizide, or MET and GLZ but all the three drugs are not available in a single formulation. This additionseems to be aimed at improving the antidiabetic efficacy.

Pharmaceutical processing and formulation often introduce various interferants (chemicals other than drug/s under investigation) into the system. When performing quantification these interferants can disturb univariate analysis, but with multivariate analysis the quantification can still be performed. Several multivariate techniques of data analysis have been developed and used in the chemometric community by the researchers, out of which PLS and NAP/CLS methods are one of them.6 PLS regression is a supervisedmultivariate method with which quantitative analysis of multiple solid forms can be performed even if the differences between the spectra are minor.7 The method involves a calibration step in which the relation between spectra and component concentrations is estimated from a set of reference samples, and a prediction step in which the results of the calibration are used to estimate the component concentrations in an unknown sample spectrum.8 NAP/CLS is one of the methods under net analyte signal preprocessing (NAS). The NAS is the part of the signal which is directly related to the concentration predicted by the calibration model. In mathematical terms, it is the part of a spectrum which is orthogonal to the space spanned by the spectra of all analytes except one.9

Materials and Methods

Instrument, reagents and softwares

Elico SL 191 double beam UV-Visible Spectrophotometer, with 1 cm path length was used for the absorbance measurement. All the chemicals used were of analytical grade. Pure MET was obtained from Abhilasha Pharma Pvt. Ltd., Gujarat, GLZ was obtained from Kwality Pharmaceuticals, Amritsar and PIO from GMH Laboratories, Baddi.

The design expert 8.0.4 software and Matlab 7.5 with MVC1 toolbox were used for construction of binary mixtures and the statistical treatment of the data along application of various multivariate methods.

Preparation of standards

1mg/ml MET, GLZ and PIO stock solutions were prepared by dissolving accurately weighed amounts of finely powdered pure MET, GLZ and PIO in small quantity of methanol and the final volumes were made respectively with 0.1N HCl. Suitably diluted samples from each stock were utilized for λmax determination of individual component followed by serial dilution with 0.1N HCl to obtain the aliquots falling in linearity.

Standard solutions for multivariate calibration

The calibration and validation mixtures were prepared by mixing MET, GLZ and PIO solutions in different ratios varying in their individual linearity ranges viz. 0-25 µg/ml, 0-8 µg/ml, 0-3 µg/ml. The concentrations of combinations were decided by design expert 8.0.4 software under central composite design. Total 25 sets were prepared out of which 15 sets (Table 1) were utilized as calibration set whereas, the rest 10 served as validation sets (Table 2). All the mixtures were scanned at 220-299 nm range digitized at every 3 nm. The absorbance below 220 nm and above 299 nm was not taken under consideration due to too much of noise and diminished responses respectively.

Table 1. Calibration set composition.

Runs MET (µg/ml) GLZ (µg/ml) PIO (µg/ml)
C1 5 1 1
C2 25 6 0.75
C3 25 1 3
C4 5 6 3
C5 0 3.5 2
C6 25 3.5 2
C7 15 3 2
C8 15 6 2
C9 15 3.5 0
C10 15 3.5 3
C11 16 4 1
C12 19 0 1
C13 20 8 1
C14 20 4 1
C15 18.5 2 1.6

Table 2. Validation set composition.

Runs MET (µg/ml) GLZ (µg/ml) PIO (µg/ml)
V1 25 0 1.3
V2 4.5 2 1.5
V3 6 0.5 3
V4 25 7 2.5
V5 13 6 3
V6 10 8 0.5
V7 25 8 3
V8 5 8 3
V9 16 4 2
V10 10 6 1

Sample preparation

Commercial tablets of MET, GLZ and PIO were analyzed for accuracy. The tablets were processed by taking at least 10 tablets for each and finely crushed to powder in separate mortar-pestles. An equivalent amount of the obtained powder of each drug was weighed, dissolved in methanol, sonicated for 20 min, made up the volume with 0.1N HCl and filtered through a 0.5 µm membrane filter. The final concentrations and analyte ratios in each test solution lied within the corresponding calibration ranges. Each sample solution was prepared in triplicate and measured in random order.

Theory

PLS-1: To start working on PLS-1 using MATLAB, first a data matrix X and a concentration vector Y need to be identify against J sensors and I samples. Both X and Y is required for the calculation of singular value decomposition (SVD). On performing PLSSVD on X and Y matrix, the result will be further 3 matrixes i.e. the singular value matrix (S), the right singular value matrix (V), and the left singular value matrix (U). V matrix can also be termed as loading matrix which helps in the determination of score matrix (T), using the following equation:

X × V = T Eq.1

Reconstruction of original data matrix X is computed by using the preselected numbers of factors as:

X(estimated) = T × V' Eq.2

The predicted value of y can be stated as:

y(estimated) = x(estimated) × b Eq.3

Where, b is regression vector.10

Before finalizing the calibration data, to avoid over fitting, the optimum number of latent variables or factors (A) (figure1) should be selected by applying the cross validation method, leaving one sample at a time.11

Figure 1.

Figure 1

Plot of RMS(CV) vs factor number for calibration set prediction using cross validation of (a) MET PLS-1, (b) MET NAP/CLS, (c) GLZ PLS-1, (d) GLZ NAP/CLS, (e) PIO PLS-1, (f) PIO NAP/CLS

NAP/CLS12: In contrast to PLS-1, the concept of NAS based calibration utilizes the contribution of two types of analyte signals, Yk i.e. the analyte of interest and Y-k, signals developed by sources of variability. The virtual signals obtained are a sum of these two and can be presented as:

Y = Yk + Y-k Eq.4

For unit concentration of k the J×1 vector can be denoted as sk hence

Y = xksk' + Y-k Eq.5

Both sides of equations when multiplied with an appropriate filtering or preprocessing J×J matrix, named, MNAP which in turn is supposed to be orthogonal to Y-k, the eq.5 get converted to:

Y MNAP = xksk' MNAP Eq.6

Eq.6 can also be presented as:

Y$ = xk(sk$)' Eq.7

Where, Y$ is matrix of net analyte calibration spectra and sk$ is net sensitivity for analyte k.

The filtering matrix in eq.6 as mentioned above is orthogonal to Yk and can be calculated as

MNAP = L - (Y-k)pY-k Eq.8

Where, L is J×J unitary matrix and (Y-k) p is pseudo-inverse of Y-k. Pseudo-inverse of Y-k can be calculated by applying singular value decomposition (SVD) at factor A:

MNAP = [L - UU'] Eq.9

The applied filter MNAP removes all sources of variability except k. The new generated problem can be resolved by applying classical least square (CLS) method in combination with NAS and that leads to the generation of equation 10.

sk$ = (Yk$)'xk(xk'xk)-1 Eq.10

Hence unknown concentration xkis determined by:

xk = (sk$'sK$)-1sK$'yk$ Eq.11

The usual statistical parameters giving an indication of the quality of fit of all data are the root mean square difference (RMSECV), square of the correlation coefficient (R2) and relative error of prediction (REP%). The expressions of these parameters are:

graphic file with name apb-3-79-eq012-015.jpg

Where cact and cpred are the actual and predicted concentrations during the cross validation process, m is number of samples used in cross validation and validation.7 The goodness of data fit can be visualized in figure 2.

Figure 2.

Figure 2

Plots of actual vs predicted values for (a) MET PLS-1, (b) MET NAP/CLS, (c) GLZ PLS-1, (d) GLZ NAP/CLS, (e) PIO PLS-1, (f) PIO NAP/CLS

Along with the above said statistical formulae, another preferred method for assessing the relative accuracy of the studied models is the linear regression analysis of actual verses predicted data by comparing the results of the estimated slope and intercept with their ideal value of 1 and 0. If the point (1, 0) is inside the EJCR (elliptical joint confidence region) for cross validation data, it can be concluded that constant and proportional bias are absent (figure 3).

Figure 3.

Figure 3

Ellipticle joint confident region for slope and intercept corresponding to regressions of the actual vs predicted concentrations of (a) MET PLS-1, (b) MET NAP/CLS, (c) GLZ PLS-1, (d) GLZ NAP/CLS, (e) PIO PLS-1, (f) PIO NAP/CLS

Results and Discussion

UV-Vis spectra of MET, GLZ, PIO and mixture

Figure 4 shows the individual absorption spectra of MET, GLZ and PIO along with their mixture in 0.1N HCl between 200 and 300 nm.

Figure 4.

Figure 4

Overlay of MET, GLZ, PIO and Mixture.

PLS-1 and NAP/CLS Results

The statistical parameters obtained after applying PLS-1 and NAP/CLS to the spectrophotometric data of cross validation and validation are shown in Table 3. The results suggest that the present method is accurate in concern to the validation samples, as suggested by the low RMSE and REP value for this validation set.

Table 3. Statistical parameters for the optimized models.

Parameters

MET GLZ PIO

PLS-1 NAP/CLS PLS-1 NAP/CLS PLS-1 NAP/CLS
Calibration set results
No. of factors 6 6 9 11 8 10
Press 0.6673 0.4725 0.0502 0.1022 0.0087 0.0273
RMSE(µg/ml) 0.2110 0.1776 0.0579 0.0826 0.0241 0.0427
REP% 1.3271 1.1172 1.5817 2.2535 1.4876 2.6345
Slope 0.9996 1.0003 1.0000 1.0006 0.9987 0.9988
R2 0.9991 0.9994 0.9992 0.9984 0.9992 0.9976
Bias 0.0000 0.0000 0.0000 0.0000 0.0000 0.0000
Validation set results
Press 0.5330 0.4973 0.1179 0.2057 0.0455 0.0904
RMSE(µg/ml) 0.2309 0.2230 0.1086 0.1434 0.0674 0.0951
REP% 1.6549 1.5986 2.1939 2.8977 3.2416 4.5721
Slope 1.0010 1.0089 1.0018 1.0101 1.0193 0.9847
R2 0.9993 0.9993 0.9988 0.9979 0.9954 0.9884
Bias 0.0677 0.0137 -0.0186 0.0141 0.0112 0.0121
Figure of merits
LOD(µg/ml) 0.1861 0.1564 0.0510 0.0727 0.0213 0.0377
LOQ(µg/ml) 0.5639 0.4741 0.1546 0.2205 0.0646 0.1142
SEM 0.0563 0.0474 0.0154 0.0221 0.0064 0.0114

Analysis of commercial sample

Commercial mixture products were analyzed using the proposed spectrophotometric methods. Results are summarized in Table 4. As can be seen, satisfactory results were obtained by the proposed methods.

Table 4. Prediction results on recovery samples.

Commercial Sample (nominal content)
Metformin HCl* Gliclazide* Pioglitazone HCl*
PLS-1 NAP/CLS PLS-1 NAP/CLS PLS-I NAP/CLS
MET-500mg, GLZ-30mg, PIO-15mg 500.33 (2.08)
(100.06%)
501.33 (2.08)
(100.26%)
30.33 (2.08)
(101.11%)
29.00 (1.00)
(96.66%)
14.33 (2.08)
(95.55%)
16.66 (1.52)
(111.11%)
MET-500mg, GLZ-30mg, PIO-45mg 500.33 (2.51)
(100.06%)
496.66 (1.15)
(99.33%)
31.00 (2.00)
(103.33%)
29.33 (2.51)
(97.77%)
45.33 (2.08)
(100.74%)
45.00 (2.00)
(100.00%)
MET-500mg, GLZ-80mg, PIO-15mg 496.66 (1.52)
(99.33%)
500.00 (2.00)
(100.00%)
79.00 (2.00)
(98.75%)
82.66 (1.52)
(103.33%)
15.00 (2.00)
(100.00%)
14.33 (1.54)
(95.55%)
MET-500mg, GLZ-80mg, PIO-45mg 498.66 (3.05)
(99.73%)
498.33 (1.52)
(99.66%)
82.00 (1.00)
(102.50%)
79.33 (1.52)
(99.19%)
46.00 (1.73)
(102.22%)
44.33 (1.53)
(98.51%)
*The results are averages of three replicates and are given in mg per sample. ±S.D. is in parenthesis.

Conclusion

A comparative study with the use of PLS-1 and NAP/CLS for the separation and simultaneous estimation of MET, GLZ and PIO in a binary mixture has been accomplished, showing that this spectrophotometric method provides a good example of the high resolving power of these techniques. In other words, almost comparable results were obtained for these three drugs in both synthetic and commercial mixture. The results obtained confirm the suitability of the proposed method for accurate analysis of MET, GLZ and PIO in pharmaceutical preparations. These methods were applied directly to the commercial mixture preparations without previous treatment. In addition the proposed methods are suitable for application without interference of the excipients as well.

Conflict of Interest

There is no conflict of interest in this study.

References

  • 1.Abdul S, Poddar SS. A flexible technology for modified release of drugs: multi layered tablets. J Control Release. 2004;97:393–405. doi: 10.1016/j.jconrel.2004.03.034. [DOI] [PubMed] [Google Scholar]
  • 2.Campbellt DB, Laviellez R, Nathan C. The mode of action and clinical pharmacology of gliclazide: a review. Diabetic Res ClinPract. 1991;14:S2l–36. doi: 10.1016/0168-8227(91)90005-x. [DOI] [PubMed] [Google Scholar]
  • 3.Dunn CJ, Peters DH. Metformin.A review of its pharmacological properties and therapeutic use in non-insulin-dependent diabetes mellitus. Drugs. 1995;49(5):721–49. doi: 10.2165/00003495-199549050-00007. [DOI] [PubMed] [Google Scholar]
  • 4.Ziegler O, Drouin P. Hemobiological Properties of Gliclazide. J DiabGnnp. 1994;8:235–9. doi: 10.1016/1056-8727(94)90050-7. [DOI] [PubMed] [Google Scholar]
  • 5.Waugh J, Keating GM, Plosker GL, Easthope S, Robinson DM. Pioglitazone: A Review of its Use in Type 2 Diabetes Mellitus. Drugs. 2006;66(1):85–109. doi: 10.2165/00003495-200666010-00005. [DOI] [PubMed] [Google Scholar]
  • 6.Olivieria AC, Goicoechea HC, Inon FA. MVC1: an integrated MatLab toolbox for first-order multivariate calibration. ChemomIntell Lab Syst. 2004;73:189–97. [Google Scholar]
  • 7.Ni Y, Wang Y, Kokot S. Multicomponent kinetics spectrophotometric determination of pefloxacin and norfloxacine in pharmaceutical preparations and human plasma samples with the aid of chemometrics. SpectrochimActa A MolBiomolSpectrosc. 2008;70(5):1049–59. doi: 10.1016/j.saa.2007.10.013. [DOI] [PubMed] [Google Scholar]
  • 8.Culzoni MJ, Zan MMD, Robles JC, Mantovani VE, Goicoechea HC. Chemometrics-assisted UV-spectroscopic strategies for the determination of theophylline in syrup. J Pharm Biomed Anal. 2005;39(5):1068–74. doi: 10.1016/j.jpba.2005.06.009. [DOI] [PubMed] [Google Scholar]
  • 9.Goicoechea HC, Olivieria AC. A comparison of orthogonal signal correction and net analyte preprocessing methodsTheoretical and experimental studies. ChemometrIntell Lab Syst. 2001;56:73–81. [Google Scholar]
  • 10.Mark H, Workman J. Chemometrics in spectroscopy. 1st ed. London: Academic Press; 2007. [Google Scholar]
  • 11.Haaland DM, Thomas EV. Partial least square methods for spectral analyses:II Application to simulated and glass spectral data. Anal Chem. 1988;60:1193–202. [Google Scholar]
  • 12.Goicoechea HC, Olivieria AC. MULTIVAR- A program for multivariate calibration incorporating net analyte signal calculations. Trends Anal Chem. 2000;19:599–605. [Google Scholar]

Articles from Advanced Pharmaceutical Bulletin are provided here courtesy of Tabriz University of Medical Sciences