Abstract
Assessment of the sentinel lymph node (SLN) in patients with early stage breast cancer is vital in selecting the appropriate surgical approach. However, the existing methods, including methylene blue and nuclides, possess low efficiency and effectiveness in mapping SLNs, and to a certain extent exert side effects during application. Indocyanine green (ICG), as a fluorescent dye, has been proved reliable usage in SLN detection by several other groups. In this paper, we introduce a novel surgical navigation system to detect SLN with ICG. This system contains two charge-coupled devices (CCD) to simultaneously capture real-time color and fluorescent video images through two different bands. During surgery, surgeons only need to follow the fluorescence display. In addition, the system saves data automatically during surgery enabling surgeons to find the registration point easily according to image recognition algorithms. To test our system, 5 mice and 10 rabbits were used for the preclinical setting and 22 breast cancer patients were utilized for the clinical evaluation in our experiments. The detection rate was 100% and an average of 2.7 SLNs was found in 22 patients. Our results show that the usage of our surgical navigation system with ICG to detect SLNs in breast cancer patients is technically feasible.
Introduction
A sentinel lymph node (SLN) is the first lymph node (LN) or a group of LNs draining from the breast [1]. Since SLN mapping, as introduced for the management of breast cancer by Giuliano et al., is currently thought to be the standard for staging clinically negative axilla [2-4], sentinel lymph node biopsy (SLNB) for breast cancer is a promising surgical technique to avoid unnecessary axillary lymph node (ALN) dissection, leading to the improvement of post-operative quality of life [5]. Presently, three detection reagents are used for detecting SLNs in clinical settings: (1) blue dye, which is widely used due to its inexpensiveness, but has limited ability to visualize afferent lymphatic vessels and SLNs [6]; (2) radioactive colloids, which require a physician specializing in nuclear medicine, to use a handheld gamma counter, making SLN localization difficult; and (3) indocyanine green (ICG), which gives a fluorescent signal has been used by several other groups for locating SLNs in breast cancer to give a reliable higher signal-to-background ratio (SBR) [7-20], and in addition cheap and of low toxicity [7,21,22]. The fluorescent dyes, such as ICG, with a high SBR and detection depth in real-time observation enable superior SLN detection compared to the other two reagents. The only concern of the drug is allergic reaction occasionally [23].
Since ICG emits near-infrared (NIR) fluorescence, undetectable by human eyes, an appropriate intraoperative detection system is required to produce real-time images with a high SBR. Current systems, in breast cancer research, use a single-camera for NIR imaging , requiring the shadowless surgical lights to be turned off to detect SLNs, and then turned on to resume the surgery. This on/off switching of shadowless light is inconvenient and extends the duration of the operation. A system with three-cameras has advantages in multispectral imaging and superior background noise reduction [2,6,24-28]. However, the complexity of the system in reducing background noise increases the number of operation complication during surgery [26-29]. Therefore, an easily operated system that provides real-time images with a high SBR is needed.
In this study, we proposed a new surgical navigation system with two cameras for high SBR real-time imaging. To ensure convenience and practicality for clinical application, our surgical navigation system was designed to localize the SLN by providing an automatic real-time color and fluorescent image display recording during surgery. We assessed ICG as a fluorescent dye for SLN detection. In order to identify an optimized dose/concentration of ICG for future clinical use, we performed a preclinical test, in which serial ICG dilutions are evaluated in SLN resection experiments in nude mice and rabbits. Subsequently, clinical studies were further conducted in breast cancer patients to evaluate the feasibility of the surgical navigation system in detecting SLNs. Our study provides evidence that our surgical navigation system could provide surgeons with real-time images to accurately locate and resect SLNs during surgery.
Materials and Methods
Surgical navigation system
The prototype surgical navigation system (Figure 1a) used in this study was developed by the Institute of Automation, Chinese Academy of Sciences (CASIA). The system had two charge-coupled device (CCD) cameras: an electron-multiplying CCD (EMCCD) was used to collect NIR fluorescent images (ProEM 1024B Excelon, Princeton Instrument, USA); a color CCD was used to collect visible color images (Pilot piA1400-17gc, Basler, Germany), in which two beams of light, separated by a beam splitter cube (NT49-683, Edmund Optics, USA), were received by the camera that then produced the video images on an external monitor. A fluorescence filter (wavelength 810nm-870nm) was placed in front of the EMCCD camera, while a visible light filter (wavelength 400nm-650nm) was put in front of the color CCD camera. The F-mount flange distance of the Nikon lens (Nikon Nikkor 70-300mm, f/4.5-5.6G) was 46.5mm and was chosen for this design for clear imaging. The hardware parts are shown in Figure 1b.
Illumination was provided by a light emitting diode (LED) (NIR light source, center wavelength 760nm, maximum power 20W) for fluorochrome excitation and a 150W halogen lamp (KL1500LCD, SCHOTT, Germany) for white light color imaging. The white light was also coupled with a fiber optic bundle and homogenized by a beam expander. According to the separation of the light spectrum, it was ensured that when the shadowless surgical light was closed, the doctors could see the operative field via a halogen light. Under these conditions, the visible light did not affect the fluorescence imaging results.
Based on the hardware system, the surgical navigation system control software was developed in two modes. One was the real-time video imaging mode, in which the color and fluorescence real-time videos were displayed separately. The other was the camera capture mode. After setting the parameters of the two cameras, one click could complete the simultaneous acquisition of the two camera images. The user interface (Figure 1c) provided complete control of the system including the camera EM mode, exposure time, image display, image overlay and image archiving. In the camera capture mode, after a set of images was acquired, the fluorescence and color images were displayed (Figure 1d). All pictures and videos could be acquired automatically during the surgery. The image processing function was designed in the software. The registration point of white light and fluorescent images was automatically calculated by the similarity matching algorithm. Then, these two images converged according to the results in the software. This feature allowed doctors to clearly locate the lesion.
Imaging Agents
ICG was purchased from the Yichuang Pharmaceutical Limited Liability Company (Dandong, China). To prevent exposure to the sunlight and fluorescence bleaching, the ICG was stored at 4°C. Two hours prior to surgery, 25 mg ICG was dissolved in 5 ml of water to yield a concentration of 5 mg/ml, which was an optimal concentration based on our preclinical study.
Lymphatic mapping in animal models
The Institutional Animal Care and Use Committee of Shantou University Medical College approved all animal studies. Five nude mice (nu/nu, Vital River Laboratory Animal Technology Co., Ltd., Beijing, China) and 10 rabbits (New Zealand white rabbits, Experimental Animal Center of Guangdong, Foshan, China) were used to map the lymphatic vessels and lymph nodes. In the mouse study, female mice were anesthetized by an injection of a 0.2mL mixture of ketamine, xylene and sterile distilled water at a ratio of 7:3:4. Fluorescent images were acquired with the imaging system after subcutaneous injection of 0.1mL ICG on the right side of the second mammary pad. After gently massaging the surrounding breast tissue to develop the lymph channels the skin was cut and the SLNs were removed.
Similarly, female New Zealand rabbits were anesthetized with an injection of sodium pentobarbital (30 mg/kg intravenously) through the ear vein and placed in a prone position on a fixation bed. After the rabbits were sedated (typically 4–5 min), fluorescent images were then acquired after subcutaneous injection of 0.1 ml ICG around the areola. A varied concentration of ICG was administered into the second armpit areola to locate the SLN and to quantitate the light intensity. The light intensity statistics were performed using the Prism 5.0 (GraphPad-Prism) computer program.
Patient characteristics and surgical procedure
Twenty-two breast cancer patients, ranging from 32 to 68 years of age (median age of 49 years) with early stage breast cancer were admitted to the Breast Center of the Cancer Hospital of Shantou University Medical College and were enrolled in the study. Of those, twelve (54.5%) were premenopausal women and ten were postmenopausal (45.5%). All patients had a tumor size less than 5 cm (i.e., T1-2N0M0) and negative lymph nodes, and were eligible for sentinel lymph node biopsy (SLNB). Informed consent was given formally to all patients before surgery. All subjects gave written informed consent after the experimental procedures were fully explained. This study was approved by the Institutional Review Board (IRB) of the Cancer Hospital of Shantou University Medical College and performed in accordance with the ethical standards of the Declaration of Helsinki.
Before surgery, the surgical navigation system was moved above the operating field and prepared for imaging where the NIR light source and halogen fiber were covered by sterile sheets. In this setting, the lens of the camera was situated approximately 60 cm above the patient. Preoperative and intraoperative pictures are shown in Figures 1e and f. The injection concentration of ICG was 5mg/ml, which was based on a published paper and our pre-clinical trials [30]. ICG was subcutaneously injected into 2-4 points of the areola. With continuous massage for 5-10 minutes, the lymph vessels connecting to the injection point were visualized, using the surgical navigation system, along with a light spot showing SLN on the fluorescent image window. With the assistance of the video showing the maximum gray value, the surgeons could trace the SLN according to the fluorescent image. SLNs detected were then removed under the navigation of real-time NIR fluorescence imaging. The excised SLNs were also examined by NIR imaging, and then sent in for pathological examination.
Results
Surgical navigation system test
The spatial resolution, field of view and working distance were the main parameters tested after finishing the prototype of the surgical navigation system. The detailed specifications of the system are shown in Table 1. In a practice setting, the distance between the LED light source and surgical area was around 10-20cm; if the distance was beyond 20cm, the fluorescent image became indistinguishable. The results of the standard television card tests with this system are shown in Figure S1.
Table 1. Specifications of the surgical navigation system.
System Features | Performance parameters |
---|---|
Chip area | EMCCD 1.3’’; COLOR CCD 2/3’’ |
Lens focal length | 70-300mm |
Working distance | 600mm |
Surgical field of view | 79mmW*79mmH to 125mmW*125mmH |
Sensor resolution (spatial) | 1024*1024 imaging pixels |
13 x 13 μm pixels | |
13.3 x 13.3 mm imaging area (optically centered) | |
Sensor resolution (temporal) | 8.5fps(full frame),16.7fps(binning 2Xver) |
Image windows | Visible, fluorescent, overlay |
Optimized dose/concentration of ICG in vivo
SLNs could be detected as early as 3 to 5 minutes after injection and the enabling time for surgery was 10 minutes after injection (Figures 2a to e). The peak of intensity at a concentration of 5mg/ml and dose of 0.1ml of the ICG solution appeared 90 minutes after injection (Figure 2a). Based on our experiments, we chose three concentrations (0.025mg/ml, 1mg/ml, and 5mg/ml) for a feasibility test. In preclinical studies, strong light intensity was visible in the EP tube at a concentration of 0.025mg/ml in vitro, but could not be detected in vivo. The 1mg/ml concentration was suitable for animal detection, but sufficient for clinical usage. Finally a concentration of 5mg/ml was recommended for clinical use due to its strong light intensity and long duration.
Detection of SLN by ICG in nude mice in vivo
To determine the feasibility of the system in a preclinical setting, 0.1mL of a 1mg/ml ICG solution was injected into the armpit of a mouse (Figure 3). The fluorescent and visible images are shown in Figures 3a and b. Using the software alignment operation, the fluorescent image was overlaid with the visible image (Figures 3c and f). Finally, the light-emitting tissue was confirmed as the SLN (Figures 3d and e) by pathological examination.
In vivo detection of SLN by ICG in rabbits
Five experiments were conducted on rabbits to ensure that SLNs could be located by using ICG. During the surgical experiment, real-time videos were shown on the monitor (Figures 4a and b). Figure 4c is the overlay of the pseudocolor fluorescent signal on top of the color image. Following excision of the SLN, a fluorescent image was taken (Figure 4d). The visible image and the merged image are shown in Figures 4e and f. The video of the fluorescence for the entire surgical procedure is shown in Video S1.
Detection of SLN in breast cancer patients by ICG
Clinical statistical results are shown in Table 2. Five minutes after injection, the draining lymph vessels and SLNs were visualized on the fluorescent image window, using the color and overlay images (Figure 5). According to the light-emitting position, an incision was executed with a scalpel. Then the light-emitting region could be targeted with the assistance of our system. The surgeons were able to identify the SLN according to the fluorescent image (Figure 5d) and accurately locate it on the color image (Figure 5e). The fluorescent and color images were overlaid together by using our software (Figures 5f and i). After removal of the light-emitting region (Figure 5h), the SLN kept lighting up (Figure 5g). The surgery video is shown in Video S2.
Table 2. Characteristics of 22 patients who underwent ICG-guided SLNB.
Study no. | Age | Menstrual state | Tumor location | Tumor size (cm) | cTNM | pTNM | ICG dose (ml) | SLN no. | SLN metastasis | ALN no. | ALN metastasis | Surgery |
|
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | 43 | pre | Right upper outer | 3×1.9 | T2N1M0 | T2N1M0 | 2 | 2 | 1/2 | 11 | 1/11 | MRM +SLNB | |
2 | 46 | pre | Left lower inner | 1.2×0.8 | T1N0M0 | T1N0M0 | 1 | 1 | 0/1 | 32 | 0/32 | BCS +ALND+SLNB | |
3 | 50 | pre | Left upper outer | 0.5×0.3 | T1N0M0 | T1N0M0 | 1 | 3 | 0/3 | 21 | 0/21 | MRM +SLNB | |
4 | 68 | post | Left upper outer | 2.9×2.4 | T2N1M0 | T2N0M0 | 0.5 | 3 | 0/3 | 21 | 0/21 | MRM +SLNB | |
5 | 38 | pre | Left lower outer | 1.8×0.6 | T1N0M0 | T1N0M0 | 1 | 1 | 0/1 | 18 | 0/18 | MRM +SLNB | |
6 | 48 | pre | Right upper outer | 4.0×2.8 | T2N1M0 | T2N0M0 | 1 | 6 | 0/6 | 30 | 0/30 | MRM +SLNB | |
7 | 66 | post | Left upper inner | 4.2×3.7 | T2N1M0 | T1N0M0 | 1 | 2 | 0/2 | 16 | 0/16 | MRM +SLNB | |
8 | 47 | pre | Left lower outer | 5.7×1.6 | T3N1M0 | T3N2M0 | 1 | 3 | 1/3 | 25 | 4/25 | MRM +SLNB | |
9 | 62 | post | Left lower borderline | 0.9×0.8 | T1N0M0 | T1N0M0 | 1 | 3 | 0/3 | 13 | 0/13 | MRM +SLNB | |
10 | 55 | post | Right upper inner | 4.0×4.0 | T2N1M0 | T2N0M0 | 1 | 4 | 0/4 | 9 | 0/9 | MRM +SLNB | |
11 | 59 | post | Left upper inner | 3.0×2.2 | T2N1M0 | T2N3M0 | 1 | 1 | 1/1 | 17 | 11/17 | MRM +SLNB | |
12 | 67 | post | Right upper outer | 1.8×1.6 | T1N0M0 | T1N1M0 | 1 | 1 | 0/1 | 12 | 2/12 | BCS +ALND+SLNB | |
13 | 32 | pre | Right upper borderline | 3.6×2.9 | T2N1M0 | T2N0M0 | 1 | 4 | 0/4 | 12 | 0/12 | MRM +SLNB | |
14 | 40 | pre | Left middle | 1.8×1.2 | T1N1M0 | T2N0M0 | 2 | 4 | 0/4 | 10 | 0/10 | MRM +SLNB | |
15 | 45 | pre | Left lower outer | 1.5×0.8 | T1N0M0 | T1N0M0 | 2 | 4 | 0/4 | 15 | 0/15 | BCS +ALND+SLNB | |
16 | 57 | post | Left lower outer | 2.0×1.0 | T2N0M0 | T1N0M0 | 2 | 3 | 0/3 | 11 | 0/11 | MRM +SLNB | |
17 | 33 | pre | Right lower | 2.3×1.2 | T2N1M0 | T2N1M0 | 2 | 4 | 1/4 | 20 | 0/20 | MRM +SLNB | |
18 | 63 | post | Left upper outer | 2.4×2.6 | T2N1M0 | T2N0M0 | 2 | 3 | 0/3 | 5 | 0/5 | MRM +SLNB | |
19 | 53 | post | Right upper outer | 2.5×1.7 | T2N1M0 | T2N1M0 | 2 | 2 | 1/2 | 14 | 0/14 | MRM +SLNB | |
20 | 33 | pre | Left upper outer | 3.1×1.6 | T2N0M0 | T2N1M0 | 2 | 1 | 1/1 | 6 | 0/6 | MRM +SLNB | |
21 | 50 | post | Left upper | 5.3×4.5 | T3N1M0 | T3N3M0 | 2 | 2 | 2/2 | 19 | 9/19 | MRM +SLNB | |
22 | 43 | pre | Left upper outer | 1.6×1.2 | T1N0M0 | T1N0M0 | 2 | 2 | 0/2 | 24 | 0/24 | MRM +SLNB |
ALN axillary lymph node, ALND axillary lymph node dissection, SLNB sentinel lymph node biopsy, MRM modified radical mastectomy, BCS breast conserving surgery
SLNB using the surgical navigation system was performed on all patients and all resected LNs were evaluated by pathological examination. One or more SLNs were found in all 22 patients (100%). The total number of identified SLNs was 59, or 2.7 per patient (range 1-6). All 59 SLNs detected by the surgical navigation system gave a fluorescent signal and the pathology results confirmed they were LN tissues (detection rate=100%). All patients then received axillary dissection with a total of 361 LNs removed. The mean number of LNs removed was 16.4 per patient (range 9-32).
Eight out of 59 SLNs (13.6%) contained metastases, while 27 out of 361 LNs (7.5%) showed metastases. One patient possessed to have pathological LN metastases, but not SLN metastases, whereas pathological examination of another three patients revealed metastases, but not in the ALN. The slice in Figure 6a is from the patients who did not have SLN metastases. The slice in Figure 6b is from a patient whose SLN was diagnosed with metastases. No side effects were reported after the injection of ICG.
Discussion
Radiology approaches such as X-rays, computed tomography (CT), magnetic resonance imaging (MRI), positron emission tomography (PET) and single photon emission computed tomography (SPECT) have been considered to assist surgical procedures, but most of them were non-applicable for intraoperative surgery. In contrast, the fluorescence imaging approach offers superior application of non-radiation and high resolution and sensitivity, compared with radiological imaging visual inspection and palpation during surgery [31]. So, for the intraoperative application, the instrument with fluorescence imaging would be provided with the following standards: high SBR real-time imaging of the surgical procedure, easy operation, non-radiation and non-physical contact to patients.
Our surgical navigation system fulfilled the above criteria. Compared to other current intraoperative systems, ours not only kept the advantage of easy operation in a single-camera system with improved quality of imaging and convenience of operation during surgery, but also ensured the core function of high SBR real-time imaging in three-camera systems. In comparison of new approach named goggle system [32], our system takes the advantage of high image resolution and low temporal noise. Therefore, with the guidance of our system, the surgeons could accurately and rapidly locate the SLN with the high quality visible images and the fluorescent images during the surgical procedure.
Since the detection rate was 100% and an average of 2.7 SLNs was detected in all patients in our study, this demonstrated that SLN detection with our system was practical and applicable, especially for the early stage breast cancer patients. It was reported that, a combination of radioactive colloid and blue dye was used for SLN mapping, which ensured the identification rate reaching 95%-97%, while each detection method had certain obvious shortcomings resulting in significantly lower identification rates by themselves [8,10,33-38]. With the usage of ICG in our study, the detection rate was 100%, which was almost the same as or even better than using the combination method. In future work, we plan a large sample study using our system to specifically compare ICG and blue dyes in terms of testifying their sensitivity and specificity during the surgical procedure [8].
To explore the optimized injection dosage and time of ICG for system detection, a series of trials was performed in our preclinical study. The results showed that, the SLN could be clearly visualized by using 1mg/ml ICG solution. However, the light intensity and duration were not enough for surgical purposes. Pre-clinical trials showed that the 5mg/ml ICG solution had strong fluorescence light intensity and the duration was long enough for 2.5 hours. The surgery of SLN dissection usually took 20-30 minutes, so that the fluorescence of ICG at 5mg/ml could be adequate for the whole process. Therefore, for clinical studies, we chose the concentration of 5mg/ml ICG solution. In future research, we will develop a surgical navigation system with improved sensitivity, which can detect weaker fluorescent signal and perform experiments that clarify the lowest concentration of the solution suitable for clinical use.
Although our surgical navigation system proved feasible in detecting SLN in breast cancer research with the usage of ICG, it also took the potential of application in other clinical areas, such as cervical cancer SLN detection studies [39].
In the future, with the aid of our surgical navigation system, we will try to dissect the orthotopic breast tumor by using the targeted NIR probe, which can distinguish the margins between the tumor and normal tissues, and guide the surgical resection appropriately [40,41].
Conclusions
In this paper, to solve the problem of positioning the SLN in early stage breast cancer research, we have developed a surgical navigation system and presented an efficient method for detecting SLN during surgeries with the advantage of real-time tracing of lymph flow and the one-step procedure. Accurate navigation and reliable treatment results will aid surgeons with better judgment during surgery. Our approach delivers valuable information and provides a useful method that facilitates more detailed exploration for surgical navigation research.
Supporting Information
Acknowledgments
The authors thank Adjunct Assistant Professor Dr. Karen M. von Deneen from the University of Florida, Dr. Yang Du and Dr. Zhenhua Hu for their modifications of this paper.
Funding Statement
This paper is supported by the National Basic Research Program of China (973 Program) under Grant No. 2011CB707700, the National Natural Science Foundation of China under Grant No. 81027002, and the Chinese Academy of Sciences Visiting Professorship for Senior International Scientists under Grant No. 2012T1G0036. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Denninghoff VC, Kahn AG, Falco J, Curutchet HP, Elsner B (2004) Sentinel lymph node: detection of micrometastases of melanoma in a molecular study. Mol Diagn 8: 253-258. doi: 10.2165/00066982-200408040-00007. PubMed: 15887981. [DOI] [PubMed] [Google Scholar]
- 2. Mieog JSD, Troyan SL, Hutteman M, Donohoe KJ, Vorst JR et al. (2011) Toward Optimization of Imaging System and Lymphatic Tracer for Near-Infrared Fluorescent Sentinel Lymph Node Mapping in Breast. Cancer - Annals of Surgical Oncology 18: 2483-2491. doi: 10.1245/s10434-011-1566-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Giuliano AE, Hunt KK, Ballman KV, Beitsch PD, Whitworth PW et al. (2011) Axillary Dissection vs No Axillary Dissection in Women With Invasive Breast Cancer and Sentinel Node Metastasis: A Randomized. Clinical Trials - Archives of Surgery 146: 980-980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Valsecchi ME, Silbermins D, de Rosa N, Wong SL, Lyman GH (2011) Lymphatic Mapping and Sentinel Lymph Node Biopsy in Patients With Melanoma: A Meta-Analysis. J Clin Oncol 29: 1479-1487. doi: 10.1200/JCO.2010.33.1884. PubMed: 21383281. [DOI] [PubMed] [Google Scholar]
- 5. Kootstra J, Hoekstra-Weebers JE, Rietman H, de Vries J, Baas P et al. (2008) Quality of life after sentinel lymph node biopsy or axillary lymph node dissection in stage I/II breast cancer patients: a prospective longitudinal study. Ann Surg Oncol 15: 2533-2541. doi: 10.1245/s10434-008-9996-9. PubMed: 18597146. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Hutteman M, Mieog JSD, Vorst JR, Liefers GJ, Putter H et al. (2011) Randomized, double-blind comparison of indocyanine green with or without albumin premixing for near-infrared fluorescence imaging of sentinel lymph nodes in breast cancer patients. Breast Cancer Res Treat 127: 163-170. doi: 10.1007/s10549-011-1419-0. PubMed: 21360075. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Hirche C, Murawa D, Mohr Z, Kneif S, Hünerbein M (2010) ICG fluorescence-guided sentinel node biopsy for axillary nodal staging in breast cancer. Breast Cancer Res Treat 121: 373-378. doi: 10.1007/s10549-010-0760-z. PubMed: 20140704. [DOI] [PubMed] [Google Scholar]
- 8. Sugie T, Sawada T, Tagaya N, Kinoshita T, Yamagami K et al. (2013) Comparison of the indocyanine green fluorescence and blue dye methods in detection of sentinel lymph nodes in early-stage breast cancer. Ann Surg Oncol 20: 2213-2218. doi: 10.1245/s10434-013-2890-0. PubMed: 23429938. [DOI] [PubMed] [Google Scholar]
- 9. Motomura K, Inaji H, Komoike Y, Kasugai T, Noguchi S et al. (1999) Sentinel node biopsy guided by indocyanine green dye in breast cancer patients. Jpn J Clin Oncol 29: 604-607. doi: 10.1093/jjco/29.12.604. PubMed: 10721942. [DOI] [PubMed] [Google Scholar]
- 10. Motomura K, Inaji H, Komoike Y, Hasegawa Y, Kasugai T et al. (2001) Combination technique is superior to dye alone in identification of the sentinel node in breast cancer patients. J Surg Oncol 76: 95-99. doi: 10.1002/1096-9098(200102)76:2. PubMed: 11223834. [DOI] [PubMed] [Google Scholar]
- 11. Kitai T, Inomoto T, Miwa M, Shikayama T (2005) Fluorescence navigation with indocyanine green for detecting sentinel lymph nodes in breast cancer. Breast Cancer 12: 211-215. doi: 10.2325/jbcs.12.211. PubMed: 16110291. [DOI] [PubMed] [Google Scholar]
- 12. Tagaya N, Yamazaki R, Nakagawa A, Abe A, Hamada K et al. (2008) Intraoperative identification of sentinel lymph nodes by near-infrared fluorescence imaging in patients with breast cancer. Am J Surg 195: 850-853. doi: 10.1016/j.amjsurg.2007.02.032. PubMed: 18353274. [DOI] [PubMed] [Google Scholar]
- 13. Sugie T, Kassim KA, Takeuchi M, Hashimoto T, Yamagami K et al. (2010) A novel method for sentinel lymph node biopsy by indocyanine green fluorescence technique in breast cancer. Cancers (Basel) 2: 713-720. doi: 10.3390/cancers2020713. PubMed: 24281090. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Ogawa Y, Ikeda K, Ogisawa K, Tokunaga S, Fukushima H et al. (2013) Outcome of sentinel lymph node biopsy in breast cancer using dye alone: a single center review with a median follow-up of 5 years. Surg Today. [DOI] [PubMed] [Google Scholar]
- 15. Yamamoto S, Maeda N, Yoshimura K, Oka M (2013) Intraoperative detection of sentinel lymph nodes in breast cancer patients using ultrasonography-guided direct indocyanine green dye-marking by real-time virtual sonography constructed with three-dimensional computed tomography-lymphography. Breast, 22: 933–7. PubMed: 23726129. [DOI] [PubMed] [Google Scholar]
- 16. Schaafsma BE, Verbeek FPR, Rietbergen DDD, van der Hiel B, van der Vorst JR et al. (2013) Clinical trial of combined radio- and fluorescence-guided sentinel lymph node biopsy in breast cancer. Br J Surg 100: 1037-1044. doi: 10.1002/bjs.9159. PubMed: 23696463. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Matsui A, Winer JH, Laurence RG, Frangioni JV (2011) Predicting the survival of experimental ischaemic small bowel using intraoperative near-infrared fluorescence angiography. Br J Surg 98: 1725-1734. doi: 10.1002/bjs.7698. PubMed: 21953541. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Sevick-Muraca EM (2012) Translation of Near-Infrared Fluorescence Imaging Technologies: Emerging Clinical Applications. Annual Review of Medicine, Vol 63 63: 217-231. [DOI] [PubMed] [Google Scholar]
- 19. Gioux S, Choi HS, Frangioni JV (2010) Image-guided surgery using invisible near-infrared light: fundamentals of clinical translation. Mol Imaging 9: 237-255. PubMed: 20868625. [PMC free article] [PubMed] [Google Scholar]
- 20. Vahrmeijer AL, Hutteman M, van der Vorst JR, van de Velde CJ, Frangioni JV (2013) Image-guided cancer surgery using near-infrared fluorescence. Nat Rev Clin Oncol 10: 507-518. doi: 10.1038/nrclinonc.2013.123. PubMed: 23881033. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Alford R, Simpson HM, Duberman J, Hill GC, Ogawa M et al. (2009) Toxicity of Organic Fluorophores Used in Molecular Imaging: Literature Review. Mol Imaging 8: 341-354. PubMed: 20003892. [PubMed] [Google Scholar]
- 22. Gilmore DM, Khullar OV, Gioux S, Stockdale A, Frangioni JV et al. (2013) Effective low-dose escalation of indocyanine green for near-infrared fluorescent sentinel lymph node mapping in melanoma. Ann Surg Oncol 20: 2357-2363. doi: 10.1245/s10434-013-2905-x. PubMed: 23440551. [DOI] [PubMed] [Google Scholar]
- 23. Su Z, Ye P, Teng Y, Zhang L, Shu X (2012) Adverse reaction in patients with drug allergy history after simultaneous intravenous fundus fluorescein angiography and indocyanine green angiography. J Ocul Pharmacol Ther 28: 410-413. doi: 10.1089/jop.2011.0221. PubMed: 22372690. [DOI] [PubMed] [Google Scholar]
- 24. Themelis G, Yoo JS, Soh K-S, Schulz R, Ntziachristos V (2009) Real-time intraoperative fluorescence imaging system using light-absorption correction. J Biomed Opt 14: 064012. doi: 10.1117/1.3259362. PubMed: 20059250. [DOI] [PubMed] [Google Scholar]
- 25. Gibbs-Strauss SL, Rosenberg M, Clough BL, Troyan SL, Frangioni JV (2009) First-in-human clinical trials of imaging devices: an example from optical imaging. Conf Proc IEEE Eng Med Biol Soc 2009: 2001-2004. PubMed: 19964033. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Hutteman M, Choi HS, Mieog JS, van der Vorst JR, Ashitate Y et al. (2011) Clinical translation of ex vivo sentinel lymph node mapping for colorectal cancer using invisible near-infrared fluorescence light. Ann Surg Oncol 18: 1006-1014. doi: 10.1245/s10434-010-1426-0. PubMed: 21080086. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Lee BT, Hutteman M, Gioux S, Stockdale A, Lin SJ et al. (2010) The FLARE Intraoperative Near-Infrared Fluorescence Imaging System: A First-in-Human Clinical Trial in Perforator Flap Breast Reconstruction. Plast Reconstr Surg 126: 1472-1481. doi: 10.1097/PRS.0b013e3181f059c7. PubMed: 21042103. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Crane LMA, Themelis G, Arts HJG, Buddingh KT, Brouwers AH et al. (2011) Intraoperative near-infrared fluorescence imaging for sentinel lymph node detection in vulvar cancer: First clinical results. Gynecol Oncol 120: 291-295. doi: 10.1016/j.ygyno.2010.10.009. PubMed: 21056907. [DOI] [PubMed] [Google Scholar]
- 29. Pleijhuis RG, Langhout GC, Helfrich W, Themelis G, Sarantopoulos A et al. (2011) Near-infrared fluorescence (NIRF) imaging in breast-conserving surgery: Assessing intraoperative techniques in tissue-simulating breast phantoms. Eur J Surg Oncol 37: 32-39. doi: 10.1016/j.ejso.2011.03.127. PubMed: 21106329. [DOI] [PubMed] [Google Scholar]
- 30. Imai K, Minamiya Y, Saito H, Nakagawa T, Ito M et al. (2013) Detection of pleural lymph flow using indocyanine green fluorescence imaging in non-small cell lung cancer surgery: a preliminary study. Surg Today 43: 249-254. doi: 10.1007/s00595-012-0237-2. PubMed: 22729459. [DOI] [PubMed] [Google Scholar]
- 31. van Dam GM, Themelis G, Crane LMA, Harlaar NJ, Pleijhuis RG et al. (2011) Intraoperative tumor-specific fluorescence imaging in ovarian cancer by folate receptor-α targeting: first in-human results. Nat Med 17: 1315-1319. doi: 10.1038/nm.2472. PubMed: 21926976. [DOI] [PubMed] [Google Scholar]
- 32. Liu Y, Njuguna R, Matthews T, Akers WJ, Sudlow GP et al. (2013) Near-infrared fluorescence goggle system with complementary metal–oxide–semiconductor imaging sensor and see-through display. J Biomed Opt 18: 101303-101303. doi: 10.1117/1.JBO.18.10.101303. PubMed: 23728180. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Goyal A, Newcombe RG, Chhabra A, Mansel RE (2006) Factors affecting failed localisation and false-negative rates of sentinel node biopsy in breast cancer--results of the ALMANAC validation phase. Breast Cancer Res Treat 99: 203-208. doi: 10.1007/s10549-006-9192-1. PubMed: 16541308. [DOI] [PubMed] [Google Scholar]
- 34. Zavagno G, De Salvo GL, Scalco G, Bozza F, Barutta L et al. (2008) A Randomized clinical trial on sentinel lymph node biopsy versus axillary lymph node dissection in breast cancer: results of the Sentinella/GIVOM trial. Ann Surg 247: 207-213. doi: 10.1097/SLA.0b013e31812e6a73. PubMed: 18216523. [DOI] [PubMed] [Google Scholar]
- 35. Straver ME, Meijnen P, van Tienhoven G, van de Velde CJ, Mansel RE et al. (2010) Sentinel node identification rate and nodal involvement in the EORTC 10981-22023 AMAROS trial. Ann Surg Oncol 17: 1854-1861. doi: 10.1245/s10434-010-0945-z. PubMed: 20300966. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Krag DN, Anderson SJ, Julian TB, Brown AM, Harlow SP et al. (2010) Sentinel-lymph-node resection compared with conventional axillary-lymph-node dissection in clinically node-negative patients with breast cancer: overall survival findings from the NSABP B-32 randomised phase 3 trial. Lancet Oncol 11: 927-933. doi: 10.1016/S1470-2045(10)70207-2. PubMed: 20863759. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. van der Vorst JR, Schaafsma BE, Verbeek FP, Hutteman M, Mieog JS et al. (2012) Randomized comparison of near-infrared fluorescence imaging using indocyanine green and 99(m) technetium with or without patent blue for the sentinel lymph node procedure in breast cancer patients. Ann Surg Oncol 19: 4104-4111. doi: 10.1245/s10434-012-2466-4. PubMed: 22752379. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Schaafsma BE, Verbeek FP, Rietbergen DD, van der Hiel B, van der Vorst JR et al. (2013) Clinical trial of combined radio- and fluorescence-guided sentinel lymph node biopsy in breast cancer. Br J Surg 100: 1037-1044. doi: 10.1002/bjs.9159. PubMed: 23696463. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Crane LM, Themelis G, Pleijhuis RG, Harlaar NJ, Sarantopoulos A et al. (2011) Intraoperative multispectral fluorescence imaging for the detection of the sentinel lymph node in cervical cancer: a novel concept. Mol Imaging Biol 13: 1043-1049. doi: 10.1007/s11307-010-0425-7. PubMed: 20835767. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Nguyen QT, Olson ES, Aguilera TA, Jiang T, Scadeng M et al. (2010) Surgery with molecular fluorescence imaging using activatable cell-penetrating peptides decreases residual cancer and improves survival. Proc Natl Acad Sci U S A 107: 4317-4322. doi: 10.1073/pnas.0910261107. PubMed: 20160097. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Savariar EN, Felsen CN, Nashi N, Jiang T, Ellies LG et al. (2013) Real-time in vivo molecular detection of primary tumors and metastases with ratiometric activatable cell-penetrating peptides. Cancer Res 73: 855-864. doi: 10.1158/0008-5472.CAN-12-2969. PubMed: 23188503. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.