Complexity of Continuous Glucose Monitoring Data in Critically Ill Patients: Continuous Glucose Monitoring Devices, Sensor Locations, and Detrended Fluctuation Analysis Methods

Matthew Signal 1, Felicity Thomas 1, Geoffrey M Shaw 2, J Geoffrey Chase 1
PMCID: PMC3876327  PMID: 24351175

Abstract

Background

Critically ill patients often experience high levels of insulin resistance and stress-induced hyperglycemia, which may negatively impact outcomes. However, evidence surrounding the causes of negative outcomes remains inconclusive. Continuous glucose monitoring (CGM) devices allow researchers to investigate glucose complexity, using detrended fluctuation analysis (DFA), to determine whether it is associated with negative outcomes. The aim of this study was to investigate the effects of CGM device type/calibration and CGM sensor location on results from DFA.

Methods

This study uses CGM data from critically ill patients who were each monitored concurrently using Medtronic iPro2s on the thigh and abdomen and a Medtronic Guardian REAL-Time on the abdomen. This allowed interdevice/calibration type and intersensor site variation to be assessed. Detrended fluctuation analysis is a technique that has previously been used to determine the complexity of CGM data in critically ill patients. Two variants of DFA, monofractal and multifractal, were used to assess the complexity of sensor glucose data as well as the precalibration raw sensor current. Monofractal DFA produces a scaling exponent (H), where H is inversely related to complexity. The results of multifractal DFA are presented graphically by the multifractal spectrum.

Results

From the 10 patients recruited, 26 CGM devices produced data suitable for analysis. The values of H from abdominal iPro2 data were 0.10 (0.03–0.20) higher than those from Guardian REAL-Time data, indicating consistently lower complexities in iPro2 data. However, repeating the analysis on the raw sensor current showed little or no difference in complexity. Sensor site had little effect on the scaling exponents in this data set. Finally, multifractal DFA revealed no significant associations between the multifractal spectrums and CGM device type/calibration or sensor location.

Conclusions

Monofractal DFA results are dependent on the device/calibration used to obtain CGM data, but sensor location has little impact. Future studies of glucose complexity should consider the findings presented here when designing their investigations.

Keywords: continuous glucose monitoring, detrended fluctuation, detrended fluctuation analysis, fractal, sensorl

Introduction

Critically ill patients often experience high levels of insulin resistance17 and stress-induced hyperglycemia, which may negatively impact outcomes.13,8,9 The 2001 landmark study by Van den Berghe and coauthors9 presented the first data showing that lower blood glucose (BG) levels in critically ill patients were associated with improved outcomes. Many subsequent studies tried to replicate those results, with some showing reduced mortality,10,11 others failing to match the results,1214 and many seeing no difference.15 Since then, it has been determined that glycemic variability also plays a very important role.1621

With Food and Drug Administration approval of the first commercially available continuous glucose monitoring (CGM) device in 2004, glycemic dynamics could be captured at a much higher resolution compared with conventional BG monitoring. More recently, CGM devices have allowed researchers to test the latest hypothesis, that glucose complexity could also be associated with mortality in critically ill patients.22,23 Glucose complexity, in a very simplistic view, is a measure of the “fuzziness” of a glucose trace.

The current hypothesis is that a healthy glucose regulatory system will be highly reactive to disturbances and make many small adjustments to keep glucose concentration within a normal range.22 Conversely, a failing glucose regulatory system will be “sluggish,” and the glucose profile should appear smooth with very few high-frequency adjustments. To date, there have been two studies that have investigated glucose complexity in critically ill patients.22,23 Both studies used detrended fluctuation analysis (DFA) to quantify glucose complexity, reporting an association between the results of DFA and mortality.

Outside of glycemia research, DFA has been widely used to quantify the scaling and correlation properties of other nonstationary, physiological time series. For example, DFA has been applied to interbreath interval of human respiration, interbeat interval of human heartbeat, and interstride interval of human stride to differentiate between healthy and pathological conditions.2430 The aim of this study was to extend the knowledge of glucose complexity in critically ill adults by investigating the effects of CGM device type/calibration and CGM sensor location on DFA results.

Methods

Patients

This study uses CGM data from patients admitted to the Christchurch Hospital intensive care unit (ICU), a mixed surgical and medical ICU, between June and November 2012. Inclusion criteria for the study were an expected duration of ICU admission longer than 5 days and two consecutive BG measurements >144mg/dl, indicating the need for glycemic control using the stochastic targeted (STAR) protocol.31 The STAR protocol is a model-based glycemic control protocol that modulates insulin and nutrition to control BG concentration. Written and signed consent was obtained from the patient or next of kin if the patient was unable to consent. This study and use of data were approved by the Upper South A Regional Ethics Committee, New Zealand.

Continuous Glucose Monitoring

Each participant in the study was monitored using three Medtronic CGM devices (Medtronic Diabetes, Northridge, CA) for a period of up to 6 days. Two sensors were located on the patient’s abdomen, one of which was connected to a Medtronic Guardian REAL-Time monitor and the other was connected to a Medtronic iPro2 recorder. The third sensor was located on the patient’s thigh and was connected to a second Medtronic iPro2 recorder. This configuration allowed comparison of results between different devices and sensor locations within each subject. Importantly, all sensors in the study had the same Medtronic Enlite glucose sensor technology. It should be noted that these CGM devices and sensors were not designed for use in the ICU and were being used off label in this study.

One significant difference between the Guardian and iPro2 CGM devices is the calibration algorithm. Calibration algorithms use independent calibration BG measurements to convert the raw sensor current (ISIG) into a series of sensor glucose (SG) values for the user. The iPro2 device stores the sensor signal information internally, and it is retrospectively calibrated after the device has been removed from the patient. Retrospective calibration allows the calibration algorithm to use information both before and after the time point of interest to obtain an optimal calibration to each reference point. However, the Guardian CGM device displays a glucose value in real time, and the calibration algorithm can only use prior data for calibration. Otherwise, all sensor technology was identical.

Calibration BG measurements were obtained by specifically trained ICU nurses at least three times per day, per manufacturer guidelines. A blood sample was drawn from the patient’s arterial line, and a blood gas analyzer (BGA) was used to determine the glucose concentration. The value from the BGA was immediately entered into the Guardian REAL-Time device and then recorded for retrospective calibration of the iPro2 devices. In addition to BGA measurements, glucometer BG measurements using arterial blood samples were done every 1–3 h for glycemic control. These measurements were independent of CGM and could therefore be used to assess CGM performance.

Glucose Complexity

Glucose complexity was quantified in this study using DFA, which has been widely used to quantify the scaling and correlation properties of nonstationary, physiological time series. For a self-similar time series (X), the scale invariant structure can be described by X(ct) = cHX(t), where the power law exponent (H) describes the scaling and is used to quantify complexity. In some cases, the scaling properties of a time series are associated with changes in physiology and may be useful to help better understand certain illnesses or medical conditions.26,28

In terms of glucose complexity, the current hypothesis is that a healthy glucose regulatory system will make many small adjustments to keep glucose concentration within a healthy range, and the high complexity is characterized by a low value of H. In contrast, a failing glucose regulatory system will be “sluggish” and appear smooth with low complexity, characterized by high values of H.

Two modes of DFA are available depending on the properties of the signal or time series. Monofractal DFA is used when the scaling properties of the time series can be quantified by a single power law exponent, which is independent of time and space. However, a limitation of monofractal DFA is that real-world physiological signals often do not exhibit simple monofractal scaling behavior over the entire time period.28,32 In these cases, multiple scaling exponents are required to fully characterize the correlation properties of the signal, and multifractal DFA should be employed.

The results of a multifractal DFA are typically displayed in a multifractal spectrum (see Appendix for details). The width, shape, and location of the multifractal spectrum can all be used to give important information about the relationship between the time series and the physiological phenomenon being studied. This capability has been illustrated in previous studies that have used multifractal DFA to help differentiate between healthy and pathological conditions, such as heart disease.28

An example of two spectrums produced using multifractal DFA are shown in Figure 1. The spectrum plotted with red dots was produced from a signal with monofractal scaling properties, and the spectrum plotted with blue crosses was produced from a signal with multifractal scaling properties, data for both of which were provided by Ihlen.32

Figure 1.

Figure 1.

Example of multifractal spectrum that is produced from multifractal DFA. Note the monofractal signal produces a very narrow spectrum, indicating monofractal scaling is present and monofractal DFA is sufficient to characterize the scaling and correlation properties of the signal.

Note the spectrum for the monofractal signal is very narrow compared with the multifractal signal, indicating the scaling exponents are almost independent of time and space. Thus a single H from monofractal DFA is sufficient to characterize the scaling and correlation properties of the signal.

This study uses both monofractal DFA and multifractal DFA implemented in MATLAB (Mathworks, Natick, MA) based on the descriptions provided by Ihlen32 and Kantelhardt and coauthors.33 A thorough discussion of both methods can be found elsewhere (for monofractal DFA 27 and multifractal DFA33). However, a general implementation for both methods is summarized in Appendix with examples.

Analysis

This study uses DFA to investigate the glucose complexity of critically ill patients who were monitored by three simultaneous CGM devices during their ICU stay. Specifically, it investigates

  1. Whether CGM device type/calibration or CGM sensor location affects DFA results and

  2. Whether monofractal or multifractal DFA is more appropriate for CGM signals, given the use of monofractal DFA as a discriminator by Lundelin and coauthors22 and Brunner and coauthors.23

Each patient enrolled in the study had three CGM devices monitoring glucose levels for up to 6 days. The warm-up period for these devices is 1–2 h,34,35 but due to off-label use here, the first 12 h of SG data were excluded to ensure that the devices were performing properly during the period of interest. Data sets with less than 500 SG measurements were excluded, and SG data sets with less than 1000 SG measurements were analyzed with increased care to ensure robust results, as this value is a recommended minimum.32 The SG data were analyzed using both monofractal DFA and multifractal DFA to determine whether the analysis method has an impact on results.

In particular,

  • Data from the two iPro2 CGM devices, one on the thigh and one on the abdomen, were compared to assess sensor location effects independent of technology;

  • Data from the iPro2 on the abdomen were compared with data from the Guardian REAL-Time located on the opposite side of the patient’s abdomen to assess the impact of CGM device type/calibration on DFA results; and

  • Analyses were repeated using the ISIG, which removed any effects induced by device calibration.

Monofractal DFA results are presented in a table, as the result from each analysis is a single scaling exponent, H. Results from the multifractal DFA are presented in figures containing a plot of the multifractal spectrum. When comparing results, similar values of H from the monofractal analysis and spectrums of similar shape/position from the multifractal analysis indicate little or no difference in the scaling properties of the time series.

Numerical results are presented as median (25th–75th percentile) where applicable. The Wilcoxon signed rank test was used to determine statistical significance when comparing intersensor site or interdevice type results. This test was used because the results are not independent or normally distributed.

Results

During the study period, 10 patients were recruited and consented to participate in the study. Table 1 shows the cohort demographics as well as measures of BG control for these patients. The results show that 81% of BG measurements were within the 72–144 mg/dl euglycemic range and no BG measurements were recorded below 54 mg/dl.

Table 1.

Cohort Demographics and Blood Glucose Control Resultsa

Demographics
Patients 10
Age (years) 51 (39–64)
Sex (M/F) 5/5
APACHE II 24 (17–27)
APACHE III 85 (52–99)
SAPS II 52 (30–59)
Length of stay (days) 20 (10–33)
Outcome (L/D) 6/4
Diabetes (None/T1/T2) 10/0/0
Blood glucose control
Time between BG (h) 1.5 (0.9–2.3)
Median (IQR) BG (mg/dl) 124 (112–137)
Percentage BG in 72–144 mg/dl band 81%
Percentage BG < 72 mg/dl 0.14%
Percentage BG < 54 mg/dl 0%
a

Displayed as median (interquartile range) where applicable. M/F, male/female; SAPS, simplified acute physiology score; APACHE, acute physiology and chronic health evaluation; L/D, lived/ died; T1, type 1 diabetes; T2, type 2 diabetes; IQR, interquartile range.

Of the 30 CGM sensors used during the study, 26 sensors from 9 patients produced enough data for use in the DFA analysis. The four excluded data sets contained less than 500 measurements and were considered too short for analysis. Three of the excluded data sets were from a short-stay patient eliminating all their CGM data, and the other excluded data set was from a sensor with adhesion issues that was dislodged early in the monitoring period.

Monofractal Detrended Fluctuation Analysis

The results from the monofractal DFA are shown in Table 2. The top half shows the results of DFA when analyzing SG data from the CGM devices. It is important to reiterate that the Guardian and iPro2 devices use different calibration algorithms with the same sensor technology. Thus, these results include any effects of calibration on DFA results.

Table 2.

Results from Monofractal Detrended Fluctuation Analysis of Sensor Glucose and ISIG Data Over Cohort

Analyzing calibrated SG data
CGM device type (both in abdomen)
Guardian iPro2 P value
Number of data sets 9 8
Scaling exponent (H) 1.43 (1.37–1.48) 1.56 (1.46–1.60)
Difference in H (iPro2 - Guardian) 0.10 (0.03–0.20) 0.08a
Sensor location (both iPro2)
Abdomen Thigh P value
Number of data sets 8 9
Scaling exponent (H) 1.56 (1.46–1.60) 1.52 (1.50–1.61)
Difference in H (thigh - abdomen) 0.04 (-0.06–0.11) 0.64a
Analyzing pre-calibration ISIG data
CGM device type (both in abdomen)
Guardian iPro2 P value
Number of data sets 9 8
Scaling exponent (H) 1.42 (1.34–1.52) 1.54 (1.37–1.60)
Difference in H (iPro2 - Guardian) 0.06 (-0.04–0.10) 0.53a
Sensor location (both iPro2)
Abdomen Thigh P value
Number of data sets 8 9
Scaling exponent (H) 1.54 (1.37–1.60) 1.51 (1.47–1.60)
Difference in H (thigh - abdomen) 0.04 (-0.03–0.09) 0.38a
a

Wilcoxon signed rank test.

In Table 2, the “CGM device type” section shows the results when comparing CGM device type. Across the cohort, the retrospectively calibrated iPro2 reported higher scaling exponents of 1.56 (1.46–1.60) compared with 1.43 (1.37–1.48) for the Guardian REAL-Time device. Furthermore, when comparing the two different abdominal devices monitoring the same individual, the H values for iPro2 data were 0.10 (0.03–0.11) higher than the H values for Guardian data (p = .08).

The “Sensor location” section in Table 2 shows the results comparing a sensor inserted in the abdomen with a sensor inserted in the thigh of the same CGM device type (iPro2). There is no significant difference in the H values for different sensor locations (p = .64). Furthermore, performance at both sites relative to independent BG measurements was similar, at 11.8% mean absolute relative difference (MARD) in the abdomen and 12.4% MARD in the thigh.

The bottom half of Table 2 shows results from a similar analysis conducted on ISIG, which removes the effects of CGM device calibration. The data sets are stratified into the same groups as the top half of Table 2 based on CGM type and sensor location. When the effects of calibration are removed, the results show no significant differences between the groups, in any of the subanalyses.

Multifractal Detrended Fluctuation Analysis

Monofractal DFA revealed that all 26 CGM data sets have H values between 1.2 and 1.8, as shown in Table 2. Thus, for the multifractal DFA the integration step [Equation (1) in Appendix] was omitted, and the resulting H values from the analyses were adjusted by +1.32

Figure 2 shows the multifractal spectrums for all SG and ISIG data sets used in the analyses. The two subplots on the left side were created using SG data, comparing CGM device type (top) and sensor location (bottom). Equivalent plots on the right-hand side of Figure 2 were created using ISIG data. There were no significant or obvious associations between sensor site or CGM device type and the shape, width, or location of the multifractal spectrums.

Figure 2.

Figure 2.

Multifractal spectrums comparing CGM device types and sensor locations. The plots on the left were created using SG data, and the plots on the right were created using ISIG data.

Figure 3 shows four examples of how monofractal and multifractal DFA can give contradicting results and why the results must be interpreted with care. In all four cases, monofractal DFA of the two SG data sets resulted in the same scaling exponent. However, the multifractal DFA resulted in significantly different spectrums for each of the paired data sets. Furthermore, the width of the spectrums suggests the scaling properties of the time series are multifractal and that multifractal DFA is a more appropriate analysis technique.28,32

Figure 3.

Figure 3.

Multifractal spectrum comparison for data sets that had the same scaling exponent from monofractal DFA.

Figure 4 shows two examples of SG data from patients in this study and the subsequent results of multifractal DFA. Figure 4A shows SG data from three sensors that report similar glucose traces throughout the monitoring period, but the multifractal DFA produces three quite different multifractal spectrums. In contrast, Figure 4B shows SG data from two sensors that do not track each other very well, and despite the discrepancies in SG data, the multifractal DFA resulted in two multifractal spectrums that are almost identical.

Figure 4.

Figure 4

(A) This example shows good agreement between SG data for each of the three CGM devices, but the multifractal spectrums for each data set are quite different. (B) This example shows average agreement between SG data for two CGM devices, but the multifractal spectrums for each data set overlap.

Discussion

The aim of this study was to investigate the effects of CGM device type/calibration and CGM sensor location on DFA results in critically ill patients. Due to the configuration of CGM devices and sensor locations in this study, there was a unique opportunity to study the effects of these parameters on DFA results.

Monofractal Detrended Fluctuation Analysis

The effects of CGM device type on monofractal DFA results revealed an important trend. The two sensors located in the patient’s abdomen were identical, but one CGM device was an iPro2 and one was a Guardian REAL-Time. H values from the iPro2 data were consistently higher than H values from the Guardian device. This outcome is most likely due to the calibration algorithms used to process the raw sensor data and estimate the underlying BG concentration. The retrospective calibration utilized by the iPro2 device potentially has a higher degree of smoothing/ filtering that leads to lower complexity of the output CGM trace. Due to privacy restrictions placed on the calibration algorithms, this could not be fully confirmed by comparing the algorithms directly. However, an analysis of the raw precalibration ISIG data from the sensors showed no significant difference between the H values from each device. Thus, the difference in H observed when assessing SG data can be attributed to the calibration algorithm and/or potentially the CGM hardware used.

Sensor location (thigh iPro2 data versus abdomen iPro2 data) had little effect on the scaling exponent determined through monofractal DFA, and the range of H values for each sensor location were similar. Furthermore, comparing the results from two sensors monitoring the same individual showed no consistent trend of one sensor having a higher H value than the other. Repeating the analysis using the raw sensor ISIG to eliminate any calibration effects gave similar results, and the scaling exponent was not dependent on location. These results suggest that CGM sensor location, at least thigh versus abdomen, should not have a significant effect on the results of a study using monofractal DFA.

Multifractal Detrended Fluctuation Analysis

The two previous studies of DFA in ICU patients only used monofractal DFA to assess glucose complexity,22,23 but this study also tested multifractal DFA. Analyses of SG data using multifractal DFA in this study were unable to associate the shape, width, or position of the multifractal spectrum with CGM device type or sensor location. Furthermore, overlaying all of the multifractal spectrums from all sensors in a single figure, similar to Figure 2, showed no clear trends to differentiate between CGM device types or sensor locations. The ISIG results showed essentially the same results.

However, one very important finding from the multifractal DFA was the width of the multifractal spectrums. Each spectrum was spread across a wide range of exponents, which suggests the scaling of SG data is multifractal, and not monofractal, by nature. Thus, future studies that investigate glucose complexity using CGM data should test for multifractal scaling and consider multifractal DFA in their analyses.

Another interesting finding from the multifractal analysis is depicted in the subplots of Figure 3. In these four cases, monofractal DFA of both data sets produced the same H value, but their multifractal spectrums were clearly and significantly different. These differences between monofractal and multifractal results are important, as they could skew the interpretation of results and lead to incorrect conclusions from studies of this nature. These results reiterate the need for a thorough analysis and highlight why care must be taken when interpreting results.

Figure 4 illustrates an important secondary observation from this study, that complexity analysis provides different information to other metrics such as mean glucose or glycemic variability. Figure 4A shows three SG traces that track each other well and report similar overall glycaemia. Although the mean and variability of the traces are similar, their multifractal spectrums are very different, indicating each SG trace has unique scaling properties. Conversely, Figure 4B shows two SG traces that have quite different dynamics and do not track each other very well, but the multifractal spectrums for each sensor show very similar scaling properties.

Limitations

There are several limitations of this study that need to be addressed. First, the data set used here contains 26 sets of CGM data from a relatively small cohort of critically ill patients, compared with the two previous studies of DFA in critically ill patients. Despite this limitation, the analysis did highlight several important aspects that should be considered in future studies, including the effects of sensor location and CGM device/calibration on DFA results. Equally, the total number of sensors available for assessing the impact of calibration and location ensure novel results that are robust.

Second, the patients in this study all had stress-induced hyperglycemia due to the severity of their illness rather than pre-existing diabetes. A similar study investigating patients with diabetes in the ICU may yield different results, leading to different conclusions. Thus, the findings presented here are not necessarily generalizable to all patient types.

Third, the sensors and devices used in this study were different from those used in the two previous studies of DFA in critically ill patients. Therefore, a direct comparison between all three studies is not appropriate. The devices in those studies were no longer available when these study data were collected, and a newer generation of technology was used. Thus, the study conditions that would allow an exact comparison could not be replicated. However, any potential differences due to advancing technology would, again, indicate the impact of technology on DFA assessment of physiology.

Fourth, the quality of results from DFA depends on the quality of CGM data. While an MARD of ~12% is considered good performance by current CGM device standards, it cannot be guaranteed that the device has accurately and reliably recorded only true physiological glycemic dynamics without interferences or noise. Fortunately, sensor technologies are constantly evolving, and their accuracy improving. It is also possible that clinical factors, such as certain drugs and/or illnesses, may have an impact on sensor performance. However, the impact of such clinical factors were mitigated in this study by strictly comparing SG data from devices monitoring the same individual, ensuring sensors were exposed to the same monitoring conditions.

Conclusion

The results of this study revealed three key findings that should be considered when analyzing glucose complexity in critically ill adults:

  1. Monofractal DFA results were sensitive to the type of CGM device used to collect the glucose data and the calibration, in particular. Data from the iPro2 CGM device gave consistently higher DFA results compared with data from the Guardian REAL-Time CGM device.

  2. Sensor location (abdomen versus thigh) had no significant effect on DFA results.

  3. Multifractal DFA results were not always consistent with monofractal DFA results. Furthermore, the width of the multifractal spectrums suggests that multifractal DFA should be considered in future glucose complexity studies.

Further investigations of glucose complexity are required before solid conclusions can be drawn. However, this study clearly highlights where care should be taken in the design of those studies and the analysis of the results.

Glossary

(BG)

blood glucose

(BGA)

blood gas analyzer

(CGM)

continuous glucose monitoring

(DFA)

detrended fluctuation analysis

(ICU)

intensive care unit

(MARD)

mean absolute relative difference

(SG)

sensor glucose

Appendix: Detrended Fluctuation Analysis Implementation

First, the time series (x) is summed and mean subtracted using Equation (1)

Y(i)=k=1i[xkx],i=1,,N. (1)

Second, the profile Y(i) is divided into Ns nonoverlapping segments of equal length s, and the trend for each segment is approximated using a least-square fit, as shown in Figure A1 for segment sizes of 32, 64, and 128 SG measurements.

Figure A1.

Figure A1.

Three examples showing segmented SG data with linear regression lines. The segment size increases from top to bottom. The variance between each regression line and the corresponding SG data in that segment is calculated using Equation (2).

The variance between the time series and the least-square fit for each segment v is calculated using Equation (2)

F2(v,s)=12i=1S(Y[(v1)s+i]yv(i))2. (2)

Third, the qth-order fluctuation function is calculated using Equation (3), where a monofractal DFA is obtained by holding q = 2. Essentially, Equation (3) gives a qth-order root mean square of the variances calculated using Equation (2). For positive q, large deviations from the fitted trend will have more influence on the fluctuation function than small deviations, and for negative values of q, small deviations will have a larger influence on the fluctuation function. The degree to which the fluctuation function is influenced by q is determined by the magnitude of q.

Fq(S)={1Nsv=1Ns[F2(v,s)]q/2}1/q. (3)

The scaling behavior of the fluctuation functions is illustrated by analyzing a plot of log (Fq(s)) versus log (s) for each q. For a scale-invariant series, the relationship is linear and the slope represents power law exponent H as in Figure A2 (left). For a multifractal time series, the scaling exponent will change for different values of q, as shown in Figure A2 (right).

Figure A2.

Figure A2.

(Left) Example plot of log(F) versus log(s) for a single value of q, where the slope of the linear regression line is scaling exponent H. (Right) Example of how H changes for different values of q, for a multifractal time series.

For the case of q = 0, Equation (3) cannot be employed, so a logarithmic averaging procedure is used instead (Equation 4):

F0(S)=exp{12Nsv=1Nslog[F2(v,s)]}. (4)

Multifractal DFA performs best when the signal being analyzed is noise-like rather than a random walk. To determine the type of signal a monofractal DFA can be run prior to multifractal DFA. If power law exponent H is between 0.2 and 0.8, then the time series is noise-like, and multifractal DFA can be run without modification. However, if H is between 1.2 and 1.8, then the time series is more like a random walk. For random walk signals, the integrating process [Equation (1)] can be skipped, and +1 should be added to the power law exponents determined in the multifractal analysis.32

To aid the interpretation of the multifractal scaling properties of a time series, the mass exponent (t(q)), q-order singularity exponent (h(q)) and q-order singularity dimension (D(q)) are calculated using

τ(q)=q.H(q)1, (5)
h(q)=τ(q), (6)
D(q)=q.h(q)τ(q), (7)

A plot of t(q) versus q (Figure A3, left) or h(q) versus q (Figure A3, right) can be used to determine the degree of multifractal scaling in a time series. A plot of D(q) versus h(q) displays the mulitfractal spectrum. The width, shape, and location of the multifractal spectrum can all be used to give important information about the relationship between the time series and the physiological phenomenon being studied.

Figure A3.

Figure A3.

(Left) plot Plot of t(q) versus. q and (right) h(q) versus. q, either of which can be used to interpret the scaling properties of a multifractal time series.

Funding

This work was supported by the Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand.

References

  • 1.Capes SE, Hunt D, Malmberg K, Gerstein HC. Stress hyperglycaemia and increased risk of death after myocardial infarction in patients with and without diabetes: a systematic overview. Lancet. 2000;355(9206):773–778. doi: 10.1016/S0140-6736(99)08415-9. [DOI] [PubMed] [Google Scholar]
  • 2.Finney SJ, Zekveld C, Elia A, Evans TW. Glucose control and mortality in critically ill patients. JAMA. 2003;290(15):2041–2047. doi: 10.1001/jama.290.15.2041. [DOI] [PubMed] [Google Scholar]
  • 3.Krinsley JS. Association between hyperglycemia and increased hospital mortality in a heterogeneous population of critically ill patients. Mayo Clin Proc. 2003;78(12):1471–1478. doi: 10.4065/78.12.1471. [DOI] [PubMed] [Google Scholar]
  • 4.McCowen KC, Malhotra A, Bistrian BR. Stress-induced hyperglycemia. Crit Care Clin. 2001;17(1):107–124. doi: 10.1016/s0749-0704(05)70154-8. [DOI] [PubMed] [Google Scholar]
  • 5.Mizock BA. Alterations in fuel metabolism in critical illness: hyperglycaemia. Best Pract Res Clin Endocrinol Metab. 2001;15(4):533–551. doi: 10.1053/beem.2001.0168. [DOI] [PubMed] [Google Scholar]
  • 6.Umpierrez GE, Isaacs SD, Bazargan N, You X, Thaler LM, Kitabchi AE. Hyperglycemia: an independent marker of in-hospital mortality in patients with undiagnosed diabetes. J Clin Endocrinol Metab. 2002;87(3):978–982. doi: 10.1210/jcem.87.3.8341. [DOI] [PubMed] [Google Scholar]
  • 7.Van den Berghe G, Wouters PJ, Bouillon R, Weekers F, Verwaest C, Schetz M, Vlasselaers D, Ferdinande P, Lauwers P. Outcome benefit of intensive insulin therapy in the critically ill: Insulin dose versus glycemic control. Crit Care Med. 2003;31(2):359–366. doi: 10.1097/01.CCM.0000045568.12881.10. [DOI] [PubMed] [Google Scholar]
  • 8.Bistrian BR. Hyperglycemia and infection: which is the chicken and which is the egg? JPEN J Parenter Enteral Nutr. 2001;25(4):180–181. doi: 10.1177/0148607101025004180. [DOI] [PubMed] [Google Scholar]
  • 9.Van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, Vlasselaers D, Ferdinande P, Lauwers P, Bouillon R. Intensive insulin therapy in the critically ill patients. N Engl J Med. 2001;345(19):1359–1367. doi: 10.1056/NEJMoa011300. [DOI] [PubMed] [Google Scholar]
  • 10.Chase JG, Shaw G, Le Compte A, Lonergan T, Willacy M, Wong XW, Lin J, Lotz T, Lee D, Hann C. Implementation and evaluation of the SPRINT protocol for tight glycaemic control in critically ill patients: a clinical practice change. Crit Care. 2008;12(2):R49. doi: 10.1186/cc6868. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Krinsley JS. Effect of an intensive glucose management protocol on the mortality of critically ill adult patients. Mayo Clin Proc. 2004;79(8):992–1000. doi: 10.4065/79.8.992. [DOI] [PubMed] [Google Scholar]
  • 12.Brunkhorst FM, Engel C, Bloos F, Meier-Hellmann A, Ragaller M, Weiler N, Moerer O, Gruendling M, Oppert M, Grond S, Olthoff D, Jaschinski U, John S, Rossaint R, Welte T, Schaefer M, Kern P, Kuhnt E, Kiehntopf M, Hartog C, Natanson C, Loeffler M, Reinhart K. German Competence Network Sepsis (SepNet). Intensive insulin therapy and pentastarch resuscitation in severe sepsis. N Engl J Med. 2008;358(2):125–139. doi: 10.1056/NEJMoa070716. [DOI] [PubMed] [Google Scholar]
  • 13.Finfer S, Delaney A. Tight glycemic control in critically ill adults. JAMA. 2008;300(8):963–965. doi: 10.1001/jama.300.8.963. [DOI] [PubMed] [Google Scholar]
  • 14.Preiser JC, Devos P, Ruiz-Santana S, Mélot C, Annane D, Groeneveld J, Iapichino G, Leverve X, Nitenberg G, Singer P, Wernerman J, Joannidis M, Stecher A, Chioléro R. A prospective randomised multi-centre controlled trial on tight glucose control by intensive insulin therapy in adult intensive care units: the Glucontrol study. Intensive Care Med. 2009;35(10):1738–1748. doi: 10.1007/s00134-009-1585-2. [DOI] [PubMed] [Google Scholar]
  • 15.Griesdale DE, de Souza RJ, van Dam RM, Heyland DK, Cook DJ, Malhotra A, Dhaliwal R, Henderson WR, Chittock DR, Finfer S, Talmor D. Intensive insulin therapy and mortality among critically ill patients: a meta-analysis including NICE-SUGAR study data. CMAJ. 2009;180(8):821–827. doi: 10.1503/cmaj.090206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Egi M, Bellomo R, Stachowski E, French CJ, Hart G. Variability of blood glucose concentration and short-term mortality in critically ill patients. Anesthesiology. 2006;105(2):244–252. doi: 10.1097/00000542-200608000-00006. [DOI] [PubMed] [Google Scholar]
  • 17.Krinsley JS. Glycemic variability: a strong independent predictor of mortality in critically ill patients. Crit Care Med. 2008;36(11):3008–3013. doi: 10.1097/CCM.0b013e31818b38d2. [DOI] [PubMed] [Google Scholar]
  • 18.Hermanides J, Vriesendorp TM, Bosman RJ, Zandstra DF, Hoekstra JB, Devries JH. Glucose variability is associated with intensive care unit mortality. Crit Care Med. 2010;38(3):838–842. doi: 10.1097/CCM.0b013e3181cc4be9. [DOI] [PubMed] [Google Scholar]
  • 19.Pidcoke HF, Wanek SM, Rohleder LS, Holcomb JB, Wolf SE, Wade CE. Glucose variability is associated with high mortality after severe burn. J Trauma. 2009;67(5):990–995. doi: 10.1097/TA.0b013e3181baef4b. [DOI] [PubMed] [Google Scholar]
  • 20.Kaneki M, Sakai M, Shimizu N, Chang K. Is normalized mean blood glucose level good enough for the intensive care unit?--glycemic variability as a new independent predictor of mortality. Crit Care Med. 2008;36(11):3104–3106. doi: 10.1097/CCM.0b013e31818c38ab. [DOI] [PubMed] [Google Scholar]
  • 21.Ali NA, O’Brien JM, Jr, Dungan K, Phillips G, Marsh CB, Lemeshow S, Connors AF, Jr, Preiser JC. Glucose variability and mortality in patients with sepsis. Crit Care Med. 2008;36(8):2316–2321. doi: 10.1097/CCM.0b013e3181810378. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Lundelin K, Vigil L, Bua S, Gomez-Mestre I, Honrubia T, Varela M. Differences in complexity of glycemic profile in survivors and nonsurvivors in an intensive care unit: a pilot study. Crit Care Med. 2010;38(3):849–854. doi: 10.1097/CCM.0b013e3181ce49cf. [DOI] [PubMed] [Google Scholar]
  • 23.Brunner R, Adelsmayr G, Herkner H, Madl C, Holzinger U. Glycemic variability and glucose complexity in critically ill patients: a retrospective analysis of continuous glucose monitoring data. Crit Care. 2012;16(5):R175. doi: 10.1186/cc11657. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Lee JM, Kim DJ, Kim IY, Suk Park K, Kim SI. Nonlinear-analysis of human sleep EEG using detrended fluctuation analysis. Med Eng Phys. 2004;26(9):773–776. doi: 10.1016/j.medengphy.2004.07.002. [DOI] [PubMed] [Google Scholar]
  • 25.Penzel T, Kantelhardt JW, Grote L, Peter JH, Bunde A. Comparison of detrended fluctuation analysis and spectral analysis for heart rate variability in sleep and sleep apnea. IEEE Trans Biomed Eng. 2003;50(10):1143–1151. doi: 10.1109/TBME.2003.817636. [DOI] [PubMed] [Google Scholar]
  • 26.Peng CK, Mietus JE, Liu Y, Lee C, Hausdorff JM, Stanley HE, Goldberger AL, Lipsitz LA. Quantifying fractal dynamics of human respiration: age and gender effects. Ann Biomed Eng. 2002;30(5):683–692. doi: 10.1114/1.1481053. [DOI] [PubMed] [Google Scholar]
  • 27.Eke A, Herman P, Kocsis L, Kozak LR. Fractal characterization of complexity in temporal physiological signals. Physiol Meas. 2002;23(1):R1–R38. doi: 10.1088/0967-3334/23/1/201. [DOI] [PubMed] [Google Scholar]
  • 28.Goldberger AL, Amaral LA, Hausdorff JM, Ivanov PCh, Peng CK, Stanley HE. Fractal dynamics in physiology: alterations with disease and aging. Proc Natl Acad Sci U S A. 2002;99(Suppl 1):2466–2472. doi: 10.1073/pnas.012579499. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Hausdorff JM, Purdon PL, Peng CK, Ladin Z, Wei JY, Goldberger AL. Fractal dynamics of human gait: stability of long-range correlations in stride interval fluctuations. J Appl Physiol (1985) 1996;80(5):1448–1457. doi: 10.1152/jappl.1996.80.5.1448. [DOI] [PubMed] [Google Scholar]
  • 30.Peng CK, Havlin S, Hausdorff JM, Mietus JE, Stanley HE, Goldberger AL. Fractal mechanisms and heart rate dynamics. Long-range correlations and their breakdown with disease. J Electrocardiol. 1995;28(Suppl:59-65) doi: 10.1016/s0022-0736(95)80017-4. [DOI] [PubMed] [Google Scholar]
  • 31.Evans A, Le Compte A, Tan CS, Ward L, Steel J, Pretty CG, Penning S, Suhaimi F, Shaw GM, Desaive T, Chase JG. Stochastic targeted (STAR) glycemic control: design, safety, and performance. J Diabetes Sci Technol. 2012;6(1):102–115. doi: 10.1177/193229681200600113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Ihlen EA. Introduction to multifractal detrended fluctuation analysis in matlab. Front Physiol. 2012;3:141. doi: 10.3389/fphys.2012.00141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Kantelhardt JW, Zschiegner SA, Koscielny-Bunde E, Havlin S, Bunde A, Stanley HE. Multifractal detrended fluctuation analysis of nonstationary time series,”. Physica A. 2002;316:87–114. [Google Scholar]
  • 34.Medtronic Minimed. iPro2 user guide. 2010.
  • 35.Medtronic Minimed. Guardian REAL-Time continuous glucose monitoring system user guide. 2006.

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES