Abstract
Purpose
Blood perfusion is a well-known factor that complicates accurate control of heating during hyperthermia treatments of cancer. Since blood perfusion varies as a function of time, temperature and location, determination of appropriate power deposition pattern from multiple antenna array Hyperthermia systems and heterogeneous tissues is a difficult control problem. Therefore, we investigate the applicability of a real-time eigenvalue model reduction (virtual source - VS) reduced-order controller for hyperthermic treatments of tissue with nonlinearly varying perfusion.
Methods
We impose a piecewise linear approximation to a set of heat pulses, each consisting of a 1-min heat-up, followed by a 2-min cool-down. The controller is designed for feedback from magnetic resonance temperature images (MRTI) obtained after each iteration of heat pulses to adjust the projected optimal setting of antenna phase and magnitude for selective tumor heating. Simulated temperature patterns with additive Gaussian noise with a standard deviation of 1.0°C and zero mean were used as a surrogate for MRTI. Robustness tests were conducted numerically for a patient's right leg placed at the middle of a water bolus surrounded by a 10-antenna applicator driven at 150 MHz. Robustness tests included added discrepancies in perfusion, electrical and thermal properties, and patient model simplifications.
Results
The controller improved selective tumor heating after an average of 4-9 iterative adjustments of power and phase, and fulfilled satisfactory therapeutic outcomes with approximately 75% of tumor volumes heated to temperatures >43°C while maintaining about 93% of healthy tissue volume < 41°C. Adequate sarcoma heating was realized by using only 2 to 3 VSs rather than a much larger number of control signals for all 10 antennas, which reduced the convergence time to only 4 to 9% of the original value.
Conclusions
Using a piecewise linear approximation to a set of heat pulses in a VS reduced-order controller, the proposed algorithm greatly improves the efficiency of hyperthermic treatment of leg sarcomas while accommodating practical nonlinear variation of tissue properties such as perfusion.
1. Introduction
The therapeutic outcome of hyperthermia is known to be significantly affected by blood perfusion [1], which is usually a nonlinear function of temperature [2]. A controller that works for constant perfusion is not guaranteed to converge for nonlinear perfusion cases. Moreover, theoretically the required steering and re-focusing time for a feedback controller to selectively heat the tumor is proportional to the square of the number of the antennas [3]. Heating during this period is ineffective, and this time expenditure could be intolerably long, for systems with a large number of antennas. Quantitatively, it would require 432 minutes for a BSD-2000 Sigma-Eye 3-ring 12-paired antenna array, if each single learning-iteration takes 3 minutes. These problems are addressed in literature [4] [5] [3]. However, in general, those controllers still require M2 steps for model identification when M antennas are used. Recently, Cheng et. al. [6] developed a model reduction method using virtual source (VS) approach that was validated by steady state temperature optimization when perfusion is either constant or nonlinearly varying. Therefore, we were motivated to design a real-time reduced-order feedback controller incorporating this model-reduction method for a hyperthermic treatment with nonlinear blood perfusion.
2. Methods
2.1. Nonlinear blood perfusion models for tissues and tumor
The hyperthermia community commonly incorporates the Pennes bio-heat transfer equations [7] with some empirically curve-fitted perfusion relations [8] to simulate the bio-heat transfer process.
<1> |
One such fitted perfusion model was developed using a half-Gaussian curve to fit Song's experiment results [2] to mimic blood vessel dilation to dissipate excessive heat. The following curve was employed to simulate this phenomenon [9] [5] [6].
<2.1> |
For tumor perfusion dynamics, some experimental studies [10] [2] reported that tumor perfusion tends to decrease with increase of temperature beyond a certain critical temperature, probably because tumor vessels are more sensitive to temperature and thus are damaged as temperature increases beyond a mild hyperthermic range [11]. Therefore we have chosen to mimic perfusion dynamics in tumor by a decreasing half-Gaussian curve.
<2.2> |
2.2. The design of the proposed feedback controller
A minimum-norm least-squares-errors-based (MNLSE) feedback algorithm [3] is incorporated with a model-reduction of VS for steady state temperature optimization [6]; therefore, the feedback correction is limited to a smaller VS subspace at each iteration. This way, we can potentially shorten the lengthy M2-step model-identification process (M-antenna applicator) by using a smaller number of N (< M) virtual antennas in fewer steps of N2 (≪M2.) Then, a reduced-order VS feedback controller was developed to operate in this reduced subspace. The VS feedback controller starts from an initial assumption of the reduced patient model, which is then iteratively updated using thermal imaging feedback. Accordingly, the system matrices defining the patient model are updated, and the power is adjusted based on the difference between the thermal images and expected temperature distribution at each iterative session. Concurrent with this update, the temperature distribution is steered to maximize heating at the target location.
2.3. Setup for numerical validation
The feasibility and robustness of the proposed VS-based reduced-order feedback controller was numerically investigated on a human lower leg dataset heated in a 10-antenna mini annular phased array (MAPA) applicator. The MAPA is coherently driven at 150 MHz. For impedance matching and thermal cooling, the antennas are coupled to the patient via a bolus filled with de-ionized water. Numerical patient and applicator models were constructed as shown in Fig. 1 and then imported into the electromagnetic simulator-HFSS (Ansoft Corporation (Pittsburgh, PA)).
The corresponding power deposition patterns generated by these EM fields were used as source terms for an in-house finite difference temperature solver. A uniform temperature of 37°C was used as the initial and boundary condition. Tissue property values used were adapted from literature [12] [13] [14] [15] [16] and are summarized in Table 1. Parameters for the nonlinearly varying tissue perfusion are given in Table-2.
Table-1. Nominal Property Values (assumed for 150 MHz).
Relative electric permittivity, εR,
|
Electrical conductivity, σ (S/m) | Density, ρ (kg/m3) | Specific heat, Ct (J/kg/°K) | Thermal conductivity, k (W/m/°C) | Blood perfusion, wb (kg/m3/s) | ||
---|---|---|---|---|---|---|---|
Blood | - | - | - | 3770 | - | - | |
Bone cortical | 14.4 | 0.07 | 1700 | 1300 | 2.3 | 0.0 | |
Bone marrow | 6.1 | 0.024 | 1700 | 1300 | 0.58 | 0.1 | |
Fat | 5.8 | 0.037 | 900 | 2987 | 0.33 | 1.1 | |
Muscle | 62.2 | 0.727 | 1050 | 3639 | 0.50 | 3.6 | |
Tumor | 68 | 0.9 | 1050 | 3639 | 0.50 | 1.8 | |
Water | 78 | 0.002 | 996 | 4178 | 0.61 | - |
Permittivity of vacuum, ε0 = 8.854*10−12 Farad/m and permeability of vacuum, μ0 = 4*π*10−7 Henry/m. The permeabilities of all tissues were assumed to be equal to that of vacuum.
Table-2. Parameter Values For Nonlinear Perfusion Models.
Perfusion in Normal Condition | Perfusion Increase | Threshold Temperature | Saturation Parameter | |
---|---|---|---|---|
w1 | w2 | Tcrit | s | |
Fat | 0.77 | 0.66 | 45 | 12 |
Muscle | 0.90 | 5.40 | 45 | 12 |
Tumor | 0.90 | 1.80 | 43 | 12 |
2.4. Feasibility studies
A set of practical perturbations to these assumed initial tissue properties were used to validate the controller robustness. Specifically, we simulated the robustness of the controller to variations in electrical and thermal property values, including blood perfusion, which can vary significantly between different patients and tissue sites [12] [13] [14]. The deviations in the initial model investigated in this work are listed in Table 3.
Table-3. Property Deviations for Lower Leg Tissues.
Relative electric permittivity, εR,
|
Electrical conductivity, σ (S/m) | Density, ρ (kg/m3) | Specific heat, (J/kg/°K) | Thermal Ct conductivity (W/m/°C) | Blood perfusion, k wb (kg/m3/s) | ||
---|---|---|---|---|---|---|---|
Blood | - | - | - | +50% | - | - | |
Bone cortical | +50% | −50% | −50% | +50% | +50% | +0.0% | |
Bone marrow | +50% | +50% | −50% | −50% | −50% | +50% | |
Fat | +50% | −50% | +50% | −50% | −50% | +50% | |
Muscle | −50% | −50% | -50% | +50% | +50% | −50% | |
Tumor | −50% | +50% | +50% | −50% | −50% | +50% |
Also, we used the simplified model preserving exact patient leg and tumor geometry of that in Fig. 1, but assuming all property values are the same as pure muscle, except those in the tumor [3]. Treatment preparation is greatly simplified if the proposed algorithm remains robust to this easy-to-build simplified model. Moreover, MRTI comes with unavoidable measurement noise, and therefore Gaussian white noise with a standard deviation (STD) of 1.0°C was added to the simulated temperature patterns from the MRTI to determine robustness of the controller in the presence of noise.
In this study, the following numerical simulations were performed for a lower leg tumor model with nonlinear variation of perfusion: the reduced-order approximate tumor model with piecewise constant perfusion was reconstructed within the reduced subspace spanned by the first 2 to 5 VS basis vectors of the three variants: 1) lower leg model, 2) lower leg model with property deviations as given in Table 3, and 3) pure-muscle lower leg model with property deviations.
3. Results and Discussion
The proposed real-time MRTI-guided VS reduced-order feedback controller accomplished successful selective tumor heating, even when patient blood perfusion values change nonlinearly with temperature as shown in Figs 3-5. Fig. 4 shows that an optimal heating vector was achieved by iterative feedback refinement producing a steady state heating distribution with ∼75% of the tumor volume heated to > 41 °C, accompanied by only ∼7% of the healthy volume > 41 °C. This steady state condition was reached in ∼N2 steps, when the first N VS basis vectors from the initial erroneous model were used. Moreover, as illustrated in Fig. 5, the heating vector at each correction step was iteratively adjusted by the proposed controller, based on the MNLSE synthesized approximate linear model spanned by the first 2-5 VS basis vectors (determined by initial standard or erroneous models). This controlled correction is seen to converge.
As is evident from the numerical simulations, the proposed controller successfully accomplishes satisfactory therapeutic outcomes. The reduced-order controller also reduces the convergence time for the simulated lower-leg sarcoma hyperthermia treatment in a 10-antenna array to only 4-9% of the time required without the virtual source model reduction approach.
Acknowledgments
The authors would like to thank Dr. Paolo F. Maccarini and Ansoft Corporation (Pittsburgh, PA) for facilitating use of HFSS finite element analysis software. This work was supported by grant NCI CA42745.
References
- 1.Cheng KS, Roemer RB. Blood Perfusion and Thermal Conduction Effects in Gaussian Beam, Minimum Time Single-Pulse Thermal Therapies. Medical Physics. 2005;32(2):311–317. doi: 10.1118/1.1835591. [DOI] [PubMed] [Google Scholar]
- 2.Song CW, Lokshina A, Rhee JG, Patten M, Levitt SH. Implication of blood flow in hyperthermia treatment of tumors. IEEE Transactions on Biomedical Engineering. 1984;31:9–16. doi: 10.1109/TBME.1984.325364. [DOI] [PubMed] [Google Scholar]
- 3.Cheng KS, Stakhursky V, Stauffer PR, Dewhirst MW, Das SK. Online feedback focusing algorithm for hyperthermia cancer treatment. International Journal of Hyperthermia. 2007;23(7):1–16. doi: 10.1080/02656730701678877. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Kohler T, Maass P, Wust P, Seebass M. A fast algorithm to find optimal controls of multiantenna applicators in regional hyperthermia. Physics in Medicine and Biology. 2001;46(9):2503–2514. doi: 10.1088/0031-9155/46/9/318. [DOI] [PubMed] [Google Scholar]
- 5.Kowalski ME, Jin JM. A temperature-based feedback control system for electromagnetic phased-array hyperthermia: Theory and simulation. Physics in Medicine and Biology. 2003;48(5):633–651. doi: 10.1088/0031-9155/48/5/306. [DOI] [PubMed] [Google Scholar]
- 6.Cheng KS, Stakhursky V, Craciunescu OI, Stauffer PR, Dewhirst MW, Das SK. Fast temperature optimization of multi-source hyperthermia applicators with reduced order modeling of virtual sources. Physics in Medicine and Biology. 2008;53(6):1619–1635. doi: 10.1088/0031-9155/53/6/008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Pennes HH. Analysis of tissue and arterial blood temperatures in the resting human forearm. Journal of Applied Physiology. 1948;1:93–122. doi: 10.1152/jappl.1948.1.2.93. [DOI] [PubMed] [Google Scholar]
- 8.Tompkins DT, Vanderby R, Klein SA, Beckman WA, Steeves DM, Frey DM, Palival BR. Temperature-dependent versus constant rate blood perfusion modeling in ferromagnetic thermoseed hyperthermia. International Journal of Hyperthermia. 1994;10:517–536. doi: 10.3109/02656739409009355. [DOI] [PubMed] [Google Scholar]
- 9.Lang J, Erdmann B, Seebass M. Impact of nonlinear heat transfer on temperature control in regional hyperthermia. IEEE Trans Biomed Eng. 1999;46(9):1129–1138. doi: 10.1109/10.784145. [DOI] [PubMed] [Google Scholar]
- 10.Muller-Klieser W, Vaupel P. Effect of hyperthermia on tumor blood flow. Biorheology. 1984;21(4):529–538. doi: 10.3233/bir-1984-21412. [DOI] [PubMed] [Google Scholar]
- 11.Song CW. Effect of local hyperthermia in blood flow and microenvironment: A review. Cancer Rev. 1984;44(Suppl):4721s–4730s. [PubMed] [Google Scholar]
- 12.Gabriel C, Gabriel S, Corthout E. The dielectric properties of biological tissues. I. Literature survey. Physics in Medicine and Biology. 1996;41(11):2231–2249. doi: 10.1088/0031-9155/41/11/001. [DOI] [PubMed] [Google Scholar]
- 13.Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues. II. Measurements in the frequency range 10 Hz to 20 GHz. Physics in Medicine and Biology. 1996;41(11):2251–2269. doi: 10.1088/0031-9155/41/11/002. [DOI] [PubMed] [Google Scholar]
- 14.Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues. III. Parametric models for the dielectric spectrum of tissues. Physics in Medicine and Biology. 1996;41(11):2271–2293. doi: 10.1088/0031-9155/41/11/003. [DOI] [PubMed] [Google Scholar]
- 15.Hayt WH, Jr, Buck JA. Engineering electromagnetics. McGraw-Hill Higher Education. 2006 [Google Scholar]
- 16.Van Den Berg CAT, Van De Kamer JB, De Leeuw AAC, Jeukens CRLPN, Raaymakers BW, Van Vulpen M, Lagendijk JJW. Towards patient specific thermal modelling of the prostate. Physics in Medicine and Biology. 2006;51(4):809–825. doi: 10.1088/0031-9155/51/4/004. [DOI] [PubMed] [Google Scholar]