Faster In and Faster Out: Accelerating Insulin Absorption and Action by Insulin Infusion Site Warming

Eda Cengiz 1,, Stuart A Weinzimer 1, Jennifer L Sherr 1, Eileen M Tichy 1, Lori Carria 1, Darryll Cappiello 1, Amy Steffen 1, William V Tamborlane 1
PMCID: PMC3887414  PMID: 24367934

Abstract

Objective: This study was undertaken to investigate the effect of an insulin infusion site warming device, the InsuPatch40 (IP40) (InsuLine Medical Ltd., Petach-Tikvah, Israel), on insulin aspart pharmacodynamics (PD) and pharmacokinetics (PK) in adolescents with type 1 diabetes.

Subjects and Methods: Seventeen subjects with type 1 diabetes (age, 15±1 years; hemoglobin A1c, 7.5±0.2% [58±2.2 mmol/mol]) underwent two euglycemic clamps performed on separate mornings with and without IP40 activation with warming temperature at 40°C. On both days, the basal infusion was suspended, and glucose levels were maintained between 90 and 100 mg/dL by a variable rate dextrose infusion for up to 5 h after a 0.2 U/kg bolus of insulin aspart.

Results: Time to peak insulin action and time to half-maximal action occurred earlier with a greater early glucodynamic effect (area under the curve [AUC] for glucose infusion rate from 0 to 30 min) with IP40 than without the IP40, whereas the AUC for the time–action profile and the peak action did not differ with and without infusion site warming. PK parameters were in agreement with PD parameters, namely, a significantly earlier time to reach the maximum increment in insulin concentrations and greater early bioavailability (AUC for the change in insulin concentration from 0 to 30 min) with the IP40. The tail of the plasma insulin response curve was also shortened with infusion site warming, with the time to reach baseline insulin concentration occurring significantly earlier (P=0.04).

Conclusions: Our data demonstrate that skin warming around the infusion site to 40°C with the IP40 is an effective means to accelerate absorption and action of rapid-acting insulin. These improvements in time–action responses have the potential to enhance the performance of open- and closed-loop insulin delivery systems.

Introduction

The time–action profiles of subcutaneous injections of bolus doses of insulin aspart, glulisine, and lispro are more rapid than that of regular insulin but not rapid enough to replicate meal-related metabolic responses in individuals without diabetes. Pharmacokinetics (PK) studies using the glucose clamp technique demonstrated that plasma levels of all three rapid-acting insulin analogs peak within 60 min after bolus doses within the 0.15–0.2 U/kg range. However, the metabolic action of these insulins does not peak until approximately 120 min after subcutaneous injection, and the duration of insulin action often exceeds 5 h.1–3 These pharmacodynamics (PD) characteristics of premeal bolus doses of rapid-acting insulin analogs contribute to both early postprandial hyperglycemia due to delays in the onset of insulin action and late postmeal hypoglycemia due to the prolonged duration of action in patients with type 1 diabetes. Some of the problems that result from delays in insulin absorption can be overcome by administering premeal bolus doses 15–30 min before eating, but compliance with early dosing is often difficult to maintain.

Problems with the relatively slow time–action profiles of subcutaneously administered insulin analogs are exaggerated in closed-loop systems that use external insulin pumps owing, in part, to the delay between the start of the meal and increases in interstitial glucose concentrations that trigger automatic increases in insulin delivery, resulting in large postprandial glycemic excursions. Even more important, with closed-loop insulin delivery, is that the “meal bolus” is stretched out over more than 2 h as the algorithm responds to changes in interstitial glucose concentrations after a meal.4 Thus, a major focus of current research in artificial pancreas development has been on methods to accelerate the absorption of insulin from the subcutaneous space.

It has long been known that increasing blood flow to an injection site by warming the skin increases the rate of insulin absorption.5–7 A novel insulin pump, an insulin infusion site warming device, the InsuPatch (InsuLine Medical Ltd., Petach-Tikvah, Israel), was designed on this principle.8 We have recently shown that using this device to heat the skin surface to 38.5°C reduces the time to reach maximal action of a standard 0.2 U/kg bolus dose of insulin aspart from approximately 125 min to approximately 90 min in adolescents with type 1 diabetes.1 The objective of this randomized, crossover study was to test the effect of this device at a higher temperature of 40°C because studies indicate that blood flow to the infusion site could be increased further by warming the skin to the higher temperature.5–7,9,10

Subjects and Methods

Subjects

The study was conducted in adolescents with type 1 diabetes who attended the Yale Children's Diabetes Clinic (New Haven, CT). Eligibility criteria included a clinical diagnosis of type 1 diabetes for ≥1 year, age between 12 and 18 years, continuous subcutaneous insulin infusion therapy for at least 3 months, hemoglobin A1c (HbA1c) <10.0% (86 mmol/mol), body mass index <95% for age and sex to eliminate the effect of insulin resistance due to obesity as a confounding factor, and the ability to comprehend written and spoken English. Subjects were excluded for any other medical disease aside from type 1 diabetes or for treated hypothyroidism, use of medications that might affect glycemic control, pregnancy or breast-feeding, not consistently using barrier methods or abstinence as contraception, or any other condition that in the judgment of the investigators would interfere with the subject's or parent's ability to provide informed consent or the investigator's ability to perform the study. The Yale University Human Investigation Committee approved the study.

At the initial enrollment visit, the risks and benefits of the study were explained. Informed consent was obtained from the parents or subjects who were 18 years of age, and informed assent was obtained from the subjects <18 years of age. History and physical examinations were performed, and the HbA1c level was measured.

InsuPatch at 40°C

In this device, the warming element is incorporated into the adhesive layer of the insulin infusion set and connected by wires to batteries that are carried in a modified insulin pump carrying case. The warming element contains a temperature sensor and controller unit that maintains the skin warming at a fixed temperature. In these experiments, the skin warming temperature was set at 40°C. When activated around the time of a bolus injection, the device applies controlled heat around the insulin infusion site that, in turn, increases blood flow into the area.

Procedures

Five-hour euglycemic clamps were conducted on two separate occasions for each subject separated by <8 weeks. One clamp was performed with the InsuPatch at 40°C (IP40) activated and the other without IP40 activation in random order, allowing the subjects to act as their own control. On each occasion, subjects were admitted on the afternoon prior to the euglycemic clamp procedure. A new insulin infusion set was inserted into the subcutaneous tissue of the anterior abdominal wall, and an intravenous catheter was placed for overnight blood sampling. Blood glucose levels were measured hourly during the night, and the basal insulin infusion was adjusted as needed to achieve glucose levels between 80 and 120 mg/dL at the start of the euglycemic clamp procedure the next morning.

All subjects fasted overnight and throughout the 5-h clamp study. A second intravenous catheter was placed on the contralateral arm for infusion of exogenous glucose on the morning of the clamp study; all subjects received insulin aspart at 0.2 U/kg of body of weight bolus with or without IP40 activation at the start of the study. During IP40 studies, IP40 was activated to warm the insulin infusion site for 15 min prior to the bolus and was kept on for 60 min after administration of the bolus. In all studies, the basal infusion of insulin via the insulin pump was suspended after the bolus was given.

A variable rate of 20% dextrose was infused and adjusted every 5 min based on bedside measurements of plasma glucose to maintain blood glucose levels from 90 to 100 mg/dL during the study for a period of 5 h. The serum insulin levels were measured by serum samples that were collected at 10-min intervals for the first 90 min, at 15-min intervals during 91–180 min, and then every 30 min during 181–300 min of the clamp study.

Biochemical methods

HbA1c was measured by the DCA Vantage® analyzer (Siemens Medical Equipment, Malvern, PA), plasma glucose by the YSI glucose analyzer (Yellow Springs Instrument Life Sciences Inc., Yellow Springs, OH), and plasma insulin by the Mercodia iso-insulin enzyme-linked immunosorbent assay test, with a reported cross-reactivity of 80% with insulin aspart (MercodiaAB, Uppsala, Sweden).

Statistical analyses

The exogenous glucose infusion rate (GIR) was analyzed every 10 min and adjusted for changes in the glucose space, as previously described.11 The PD parameters that were calculated for each clamp study included the maximum glucose infusion rate (GIRmax), time to maximum glucose infusion rate (TGIRmax), and time to half-maximum GIR (T50%GIRmax). In addition to total area under the curve (AUC) for GIR from 0 to 300 min (AUCGIR 0–300min), partial AUCGIR values from 0 to 30 min (AUCGIR 0–30min) were also analyzed to determine early glucodynamic action.

Increments in plasma insulin levels measured from baseline at each time point throughout the clamp study were used for analysis to adjust for within subject differences in baseline insulin levels. Mean increment (ΔCins), maximum increment (ΔCins max), total AUC for ΔCins (AUCΔCins 0–300min), and the AUCΔCins 0–30min (an index of early insulin exposure) were calculated. The time to reach maximum increments (TΔCmax), time to half-maximum increments (TΔC50%max), and the time to 50% reduction from maximum increment (T50% ΔCfall) in plasma insulin were calculated as measures of insulin PK. Time to reach baseline insulin concentrations (TΔCins0), defined as ΔCins <10 μU/mL, is used to determine the completion of insulin aspart bolus bioavailability.

Statistical comparisons were performed using GraphPad Prism® version 5.0 software (GraphPad Software Inc., La Jolla, CA). Paired t tests were used where applicable to compare plasma glucose level, insulin concentration, and glucose clamp data, which are reported as mean±SEM values.

Results

Seventeen of the 20 adolescents (six female, 14 male) who enrolled in the study completed both clamp studies. Two subjects were withdrawn from the study because they were started on medications that might affect metabolic responses between the first and the second clamp, and another patient who consented was subsequently withdrawn because of a history of polycystic ovary syndrome. The PK data were not available in one subject because of problems with processing samples for the insulin assay. Thus, data from 17 subjects were analyzed for the comparison of PD parameters, and data from 16 subjects were analyzed for PK measures. The mean age was 15±1 years, mean body mass index was 21±0.5 kg/m2, and mean HbA1c was 7.5±0.2% (58±2.2 mmol/mol). Mean plasma glucose concentrations did not differ between the two studies during the 5-h clamps (100±1 mg/dL with IP40 and 101±1 mg/dL without IP40, P=0.9).

As illustrated in Figure 1 and Table 1, IP40 activation shifted the time–action curve of insulin aspart to the left, with an earlier TGIRmax (P=0.0004) and T50%GIRmax (P=0.0005). Although the total AUCGIR 0–300min and GIRmax did not differ significantly between the two studies, AUCGIR 0–30min and mean GIR0–30min were greater with IP40 activation (P=0.04 and 0.02, respectively).

FIG. 1.

FIG. 1.

Average glucose infusion rate (GIR) values by time required to maintain euglycemia under glucose clamp conditions with (closed squares, solid line) and without (open circles, dotted line) activation of the InsuPatch at 40°C (IP40). Data are mean±SEM values.

Table 1.

Pharmacokinetic and Pharmacodynamic Measures With and Without Activation of the InsuPatch at 40°C

Parameter No InsuPatch With InsuPatch P
Pharmacodynamic (n=17)
 Mean GIR (mg/kg/min)
4.0±0.3
4.0±0.4
0.9
 GIRmax (mg/kg/min)
10.2±0.9
10.0±0.7
0.7
TGIRmax (min)
111±7
77±5
0.0004
T50%GIRmax (min)
54±4
33±3
0.0005
 Mean GIR0–30min (mg/kg/min)
1.2±0.3
2.4±0.3
0.02
 AUCGIR 0–30min (mg/kg)
35±09
65±10
0.04
 AUCGIR 0–300min (mg/kg)
1,328±119
1,399±109
0.5
Pharmacokinetic (n=16)
 Mean ΔCins (μU/mL)
41±4
47±6
0.2
 ΔCmax (μU/mL)
96±8
121±15
0.08
TΔCmax (min)
73±7
43±3
0.002
TΔC50% max (min)
31±2
20±2
0.0002
 AUCΔCins 0–30min (μU·min/mL)
726±93
1,066±155
0.02
 AUCΔCins 0–300min (μU·min/mL)
11,330±1,256
12,440±1,576
0.3
TΔCins0 200±10 183±11 0.04

Data are mean±SEM values. Plasma insulin concentrations as a measure of insulin pharmacokinetics are calculated as increments above baseline (ΔC). Values of P<0.05 are considered significant.

AUC, area under the curve; C, concentration; GIR, glucose infusion rate; ins, insulin; max, maximum; T, time.

The mean baseline plasma insulin concentrations were similar between the two studies (27±3 μU/mL vs. 28±4 μU/mL with and without IP40, respectively, P=0.2). The increments in plasma insulin that were achieved after the 0.2 U/kg bolus of insulin aspart are shown in Figure 2. As shown in Table 1, the mean ΔCins and the ΔCmax levels were higher with IP40 activation, but the differences did not achieve statistical significance (P=0.2 and 0.08, respectively). The ΔC50%max values were also similar for both conditions (P=0.8). On the other hand, the TΔCmax and TΔC50%max values in plasma insulin occurred significantly earlier with IP40 activation (P=0.002 and 0.0002, respectively). In agreement with the aforementioned PK measures, T50%ΔCfall was significantly earlier with IP40 (114±7 min vs. 137±11 min, P=0.04). Although the overall bioavailability of insulin, as indicated by AUCΔCins 0–300min was not significantly different on non-IP40 and IP40 days, the early bioavailability of insulin represented by AUCΔCins 0–30min was 47% greater with IP40 activation versus without IP40 (P=0.02). Moreover, the TΔCins0, defined as ΔCins <10 μU/mL, was almost 20 min earlier during clamps with IP40 (P=0.04). All subjects tolerated the IP40 well without any site irritation or discomfort.

FIG. 2.

FIG. 2.

Increments in insulin plasma concentration (ΔCins) from baseline during euglycemic clamps with (squares) and without (circles) activation of the InsuPatch at 40°C (IP40) following bolus administration of insulin aspart at 0.2 U/kg. Data are mean±SEM values.

Discussion

Our previous study established that warming the skin surface around an insulin infusion set to 38.5°C with the InsuPatch (IP38.5) was effective in accelerating the time–action profile of a standard bolus dose of insulin aspart.1 This finding was confirmed and extended in the present study with the use of IP40. As expected, in comparison with control studies without site warming, IP40 resulted in a much faster absorption and action of the insulin aspart bolus, especially during the first 30 min following administration. Moreover, the TΔCmax of insulin action with IP40 (77±5 min) occurred even earlier than we previously observed with IP38.5 (90±6 min).1

It is particularly noteworthy that the PK analyses demonstrated that skin warming with the IP40 promoted more rapid elimination of the rapid-acting insulin bolus from the circulation. Despite higher peak increments in plasma insulin levels with infusion site warming, plasma insulin concentrations returned to baseline values significantly earlier during IP40 active clamp studies, an outcome that we had not observed in our previous IP38.5 study.1 Accelerating the disappearance of the insulin bolus has been an important aim for ultra-fast-acting insulin development because it offers a means to reduce the risk of late postprandial hypoglycemia, especially after the last meal of the day.

We took a conservative approach in our first study by warming the skin to only 38.5°C because this might increase battery life in future clinical use devices, even though warming the skin to 40°C has been shown to enhance the magnitude of increase in local blood flow compared with the lower temperature.7,10 We anticipated that skin warming to 40°C would be well tolerated by the patients because devices that warm the skin surface to 42°C have been approved by the Food and Drug Administration. Moreover, previously published studies investigating temperature threshold for skin burn injury have shown that no signs of thermal injury were observed when skin was exposed to temperatures up to 43°C for a prolonged period of time.12–15 The concern that skin warming might reduce the bioavailability of the insulin bolus by denaturing some of the insulin in the subcutaneous depot was not supported by our findings: overall PK (mean ΔCins and ΔCmax) and PD (AUCGIR 0–300min and GIRmax) measures were similar between conditions, and, as mentioned, the early PK and PD properties were improved with site warming.

There are a few limitations of this study. The study was nonblinded for both the investigator and the subject for practicality. The dose of insulin for our study was 0.2 U/kg for all subjects, and even though variability in PK/PD depending on insulin dose has been shown for regular insulin, it is possible that the acceleration of insulin action could vary based on the amount of insulin dose delivered, with possibly even faster action for lower insulin doses.2,3,16 We excluded subjects with potential insulin resistance due to obesity and polycystic ovary disease to limit confounding factors for insulin action.17,18 The negative impact of adiposity to insulin absorption could have been an additional confounding factor as shown in tracer absorption studies performed in subjects without diabetes using regular insulin.19

The effectiveness of IP40 compares very favorably with other approaches aimed at increasing the rate of absorption and action of subcutaneously injected insulin. Changing the excipients that are used to keep insulin in solution has been successful in accelerating the rate of absorption of regular human insulin and insulin aspart, as has pretreating the infusion site with recombinant human hyaluronidase to break down barriers to fluid dispersion in the subcutaneous space.20–23 Alternative insulin delivery methods such as inhaled insulin and intradermal delivery with microneedles have been shown to confer ultra-fast-acting profiles during meal and insulin clamp studies.24–26 Further studies are needed to determine whether the time–action profiles of rapid-acting insulin analogs can be accelerated even more by combining site warming with one or another of these alternative approaches.

Conclusions

The infusion site warming appears to be a well-tolerated, safe, and effective method to speed the absorption of subcutaneous insulin boluses, with potential for use in both open- and closed-loop insulin delivery.

Acknowledgments

This work was made possible by grants from the JDRF (JDRF 17-2010-426) and the National Institutes of Health Clinical and Translational Science Award (TR000142) that supported the staff of the Yale-New Haven Hospital Research Unit. We thank Yale Pediatric Diabetes Clinic & Research Center Team members, our patients, and their families.

Author Disclosure Statement

W.V.T. is a consultant for Medtronic Diabetes, Novo Nordisk, Sanofi, Unomedical, and Insuline Medical. S.A.W. is a consultant for Animas, Becton Dickinson, and Insuline Medical, has received in-kind grant support from Medtronic, and serves as a speaker for Eli Lilly. E.C., J.L.S., E.M.T., D.C., A.S., and L.C. have no disclosures.

References

  • 1.Cengiz E, Weinzimer SA, Sherr JL, Cengiz E, Tichy E, Martin M, Carria L, Steffen A, Tamborlane WV: Acceleration of insulin pharmacodynamic profile by a novel insulin infusion site warming device. Pediatr Diabetes 2012;14:168–173 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Swan KL, Dziura JD, Steil GM, Voskanyan GR, Sikes KA, Steffen AT, Martin ML, Tamborlane WV, Weinzimer SA: Effect of age of infusion site and type of rapid-acting analog on pharmacodynamic parameters of insulin boluses in youth with type 1 diabetes receiving insulin pump therapy. Diabetes Care 2009;32:240–244 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Swan KL, Weinzimer SA, Dziura JD, Swan KL, Weinzimer SA, Dziura JD, Steil GM, Voskanyan GR, Steffen AT, Martin ML, Tamborlane WV: Effect of puberty on the pharmacodynamic and pharmacokinetic properties of insulin pump therapy in youth with type 1 diabetes. Diabetes Care 2008;31:44–46 [DOI] [PubMed] [Google Scholar]
  • 4.Weinzimer SA, Steil GM, Swan KL, Dziura J, Kurtz N, Tamborlane WV: Fully automated closed-loop insulin delivery versus semiautomated hybrid control in pediatric patients with type 1 diabetes using an artificial pancreas. Diabetes Care 2008;31:934–939 [DOI] [PubMed] [Google Scholar]
  • 5.Berger M, Cuppers HJ, Hegner H, Jorgens V, Berchtold P: Absorption kinetics and biologic effects of subcutaneously injected insulin preparations. Diabetes Care 1982;5:77–91 [DOI] [PubMed] [Google Scholar]
  • 6.Koivisto VA, Felig P: Alterations in insulin absorption and in blood glucose control associated with varying insulin injection sites in diabetic patients. Ann Intern Med 1980;92:59–61 [DOI] [PubMed] [Google Scholar]
  • 7.Sindelka G, Heinemann L, Berger M, Frenck W, Chantelau E: Effect of insulin concentration, subcutaneous fat thickness and skin temperature on subcutaneous insulin absorption in healthy subjects. Diabetologia 1994;37:377–380 [DOI] [PubMed] [Google Scholar]
  • 8.Raz I, Weiss R, Yegorchikov Y, Bitton G, Nagar R, Pesach B: Effect of a local heating device on insulin and glucose pharmacokinetic profiles in an open-label, randomized, two-period, one-way crossover study in patients with type 1 diabetes using continuous subcutaneous insulin infusion. Clin Ther 2009;31:980–987 [DOI] [PubMed] [Google Scholar]
  • 9.Petrofsky J, Paluso D, Anderson D, Swan K, Yim JE, Murugesan V, Chindam T, Goraksh N, Alshammari F, Lee H, Trivedi M, Hudlikar AN, Katrak V: The contribution of skin blood flow in warming the skin after the application of local heat; the duality of the Pennes heat equation. Med Eng Physics 2011;33:325–329 [DOI] [PubMed] [Google Scholar]
  • 10.Charkoudian N: Skin blood flow in adult human thermoregulation: how it works, when it does not, and why. Mayo Clin Proc 2003;78:603–612 [DOI] [PubMed] [Google Scholar]
  • 11.DeFronzo RA, Tobin JD, Andres R: Glucose clamp technique: a method for quantifying insulin secretion and resistance. Am J Physiol 1979;237:E214–E223 [DOI] [PubMed] [Google Scholar]
  • 12.Greenhalgh DG, Lawless MB, Chew BB, Crone WA, Fein ME, Palmieri TL: Temperature threshold for burn injury: an oximeter safety study. J Burn Care Rehabil 2004;25:411–415 [DOI] [PubMed] [Google Scholar]
  • 13.Siekmann H: Recommended maximum temperatures for touchable surfaces. Appl Ergonom 1990;21:69–73 [DOI] [PubMed] [Google Scholar]
  • 14.Diller KR, Hayes LJ, Blake GK: Analysis of alternate models for simulating thermal burns. J Burn Care Rehabil 1991;12:177–189 [DOI] [PubMed] [Google Scholar]
  • 15.Diller KR: Analysis of burns caused by long-term exposure to a heating pad. J Burn Care Rehabil 1991;12:214–217 [DOI] [PubMed] [Google Scholar]
  • 16.Heinemann L: Variability of insulin absorption and insulin action. Diabetes Technol Ther 2002;4:673–682 [DOI] [PubMed] [Google Scholar]
  • 17.Codner E, Soto N, Lopez P, Trejo L, Avila A, Eyzaguirre FC, Iniguez G, Cassorla F: Diagnostic criteria for polycystic ovary syndrome and ovarian morphology in women with type 1 diabetes mellitus. J Clin Endocrinol Metab 2006;91:2250–2256 [DOI] [PubMed] [Google Scholar]
  • 18.Purnell JQ, Zinman B, Brunzell JD: The effect of excess weight gain with intensive diabetes mellitus treatment on cardiovascular disease risk factors and atherosclerosis in type 1 diabetes mellitus: results from the Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications Study (DCCT/EDIC) study. Circulation 2013;127:180–187 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Vora JP, Burch A, Peters JR, Owens DR: Relationship between absorption of radiolabeled soluble insulin, subcutaneous blood flow, and anthropometry. Diabetes Care 1992;15:1484–1493 [DOI] [PubMed] [Google Scholar]
  • 20.Heinemann L, Hompesch M, Flacke F, Simms P, Pohl R, Albus K, Pfützner A, Steiner S: Reduction of postprandial glycemic excursions in patients with type 1 diabetes: a novel human insulin formulation versus a rapid-acting insulin analog and regular human insulin. J Diabetes Sci Technol 2011;5:681–686 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Hompesch M, Muchmore DB, Morrow L, Ludington E, Vaughn DE: Improved postprandial glycemic control in patients with type 2 diabetes from subcutaneous injection of insulin lispro with hyaluronidase. Diabetes Technol Ther 2012;14:218–224 [DOI] [PubMed] [Google Scholar]
  • 22.Hompesch M, Muchmore DB, Morrow L, Vaughn DE: Accelerated insulin pharmacokinetics and improved postprandial glycemic control in patients with type 1 diabetes after coadministration of prandial insulins with hyaluronidase. Diabetes Care 2011;34:666–668 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Morrow L, Muchmore DB, Hompesch M, Ludington EA, Vaughn DE: Comparative pharmacokinetics and insulin action for three rapid-acting insulin analogs injected subcutaneously with and without hyaluronidase. Diabetes Care 2013;36:273–275 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Pettis RJ, Hirsch L, Kapitza C, Nosek L, Hövelmann U, Kurth HJ, Sutter DE, Harvey NG, Heinemann L: Microneedle-based intradermal versus subcutaneous administration of regular human insulin or insulin lispro: pharmacokinetics and postprandial glycemic excursions in patients with type 1 diabetes. Diabetes Technol Ther 2011;13:443–450 [DOI] [PubMed] [Google Scholar]
  • 25.Boss AH, Petrucci R, Lorber D: Coverage of prandial insulin requirements by means of an ultra-rapid-acting inhaled insulin. J Diabetes Sci Technol 2012;6:773–779 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.McVey E, Hirsch L, Sutter DE, Kapitza C, Dellweg S, Clair J, Rebrin K, Judge K, Pettis RJ: Pharmacokinetics and postprandial glycemic excursions following insulin lispro delivered by intradermal microneedle or subcutaneous infusion. J Diabetes Sci Technol 2012;6:743–754 [DOI] [PMC free article] [PubMed] [Google Scholar]