. 2013 Nov;51(11):3653–3657. doi: 10.1128/JCM.01517-13

The Aptima HPV Assay Fulfills the Cross-Sectional Clinical and Reproducibility Criteria of International Guidelines for Human Papillomavirus Test Requirements for Cervical Screening

D A M Heideman a,, A T Hesselink a, F J van Kemenade a, T Iftner b, J Berkhof c, F Topal a, D Agard a, C J L M Meijer a, P J F Snijders a
PMCID: PMC3889747  PMID: 23985912

Abstract

The Aptima HPV assay (Hologic Gen-Probe, San Diego, CA) is an FDA-approved assay for detecting human papillomavirus (HPV) E6/E7 mRNA from 14 high-risk HPV types. This study evaluated the clinical performance of the Aptima HPV assay for cervical intraepithelial neoplasia of grade 2 or worse (CIN2+), relative to the high-risk HPV GP5+/GP6+ PCR, in a cross-sectional clinical equivalence analysis using the noninferiority score test with cervical samples from population-based screening, i.e., 69 cervical scraping samples from women with CIN2+ and 843 from women without evidence of CIN2+. In addition, intralaboratory reproducibility over time and interlaboratory agreement of the Aptima HPV assay results were assessed with another set of 548 cervical samples. The Aptima HPV assay showed a clinical sensitivity for CIN2+ of 94.2% (95% confidence interval [CI], 85.5 to 97.8%) and a clinical specificity for CIN2+ of 94.5% (95% CI, 92.8 to 95.9%); by comparison, these figures were 97.1% (95% CI, 89.1 to 99.3%) (67/69 samples) and 93.6% (95% CI, 91.7 to 95.0%) (785/839 samples), respectively, for GP5+/GP6+ PCR. The clinical sensitivity and specificity of the Aptima HPV assay were noninferior to those of GP5+/GP6+ PCR (P = 0.039 and 0.00016, respectively). In addition, high reproducibility of the Aptima HPV assay, as reflected by the intralaboratory reproducibility over time of 96.0% (95% CI, 94.4 to 97.3%) (526/548 samples; kappa = 0.89) and interlaboratory agreement of 96.7% (95% CI, 95.4 to 98.1%) (531/548 samples; kappa = 0.91), was found. Altogether, these data show that the Aptima HPV assay meets the cross-sectional clinical and reproducibility criteria of the international guidelines for HPV test requirements for cervical screening. Longitudinal data are needed to ensure that the long-term negative predictive value of this mRNA assay is similar to those of validated HPV DNA tests.

INTRODUCTION

The Aptima HPV assay (Hologic Gen-Probe, San Diego, CA, USA) is an in vitro nucleic acid amplification test designed to detect human papillomavirus (HPV) E6/E7 mRNA from 14 high-risk HPV types (i.e., types 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, and 68) as a pool in cervical samples collected in preservative medium (1, 2). The Aptima HPV assay is based on target capture following cell lysis, with subsequent transcription-mediated amplification and probe hybridization protection for the detection of E6/E7 mRNA expression in one measurement. The assay can be run on the fully automated Panther and TIGRIS direct tube sampling (DTS) systems (Hologic Gen-Probe, San Diego, CA) or the semiautomated DTS system (3). The Aptima HPV assay has been approved by the FDA for testing of women ≥21 years of age whose Pap tests show atypical squamous cells of undetermined significance (ASC-US) and for screening of women ≥30 years of age as an adjunct to Pap testing (http://www.fda.gov/MedicalDevices/ProductsandMedicalProcedures/DeviceApprovalsandClearances/Recently-ApprovedDevices/ucm278520.htm).

There is widespread interest in HPV testing as a primary tool for cervical screening given its higher sensitivity versus cytology for the detection of high-grade cervical intraepithelial neoplasia and cervical cancer (i.e., cervical intraepithelial neoplasia of grade 2 or worse [CIN2+]), consequently offering better protection against cervical precancer/cancer (4, 5). It is imperative, however, that HPV assays used for primary cervical screening have proven sufficient clinical performance for the detection of CIN2+ to allow effective detection of women with clinically meaningful cervical disease while minimizing the detection of women with transient HPV infections, to reduce unnecessary follow-up procedures (6). Based on the available data from large prospective screening studies, specific clinical standards for HPV DNA tests for primary cervical screening have been formulated by an international consortium (7). These guidelines can be used to assess the clinical performance of a candidate HPV test, relative to one of the two prototype HPV tests with proven good clinical performance in cervical screening (i.e., high-risk HPV Hybrid Capture 2 [HC2] or GP5+/GP6+ PCR), by a cross-sectional clinical equivalence analysis with a well-defined sample series collected in a screening setting (6). Resulting noninferiority analyses for clinical sensitivity and specificity subsequently allow clinical validation of candidate assays for screening purposes without the need for a large prospective screening study. Furthermore, since screening tests should perform robustly and technically accurately, assessment of the candidate tests for intralaboratory reproducibility over time and interlaboratory agreement is also an intrinsic part of the international guidelines (7).

This study set out to clinically evaluate the Aptima HPV assay according to these international guidelines, with samples originating from a cervical screening cohort. The clinical performance of the Aptima HPV assay was determined relative to high-risk HPV GP5+/GP6+ PCR, and the intralaboratory reproducibility and interlaboratory agreement of the Aptima HPV assay were analyzed.

MATERIALS AND METHODS

Study population.

A total of 912 archived cervical samples collected in PreservCyt liquid-based cytology medium (Hologic, Marlborough, MA) were used. These scraping samples were collected during the course of routine population-based screening and a screening trial (trial NTR2606 at http://www.trialregister.nl/) in the Utrecht and North Holland regions of The Netherlands. These samples were stored at −80°C prior to use. The series included a set of samples for clinical sensitivity analysis, with 69 scraping samples from women (median age, 39 years [range, 30 to 60 years]) who had histologically confirmed CIN2+ (i.e., 32 CIN2 cases, 33 CIN3 cases, and 4 squamous cell carcinoma cases). These cases were considered to be representative of CIN2+ detected by combined HPV and cytology screening; as such, 19 (28%) had a high-risk HPV GP5+/GP6+ PCR-positive normal Pap smear and 50 (72%) an abnormal Pap smear either positive or negative in a high-risk HPV GP5+/GP6+ PCR assay, including 7 cases with borderline or mild dyskaryosis and 43 with greater than mild dyskaryosis. The series also included a set of samples for clinical specificity analysis, including 843 consecutive scraping samples from women (median age, 41 years [range, 31 to 60 years]) who had normal cytology findings and were without evidence of CIN2+ within a 2-year follow-up period (referred to as controls). Informed consent was obtained from all study participants, and this study followed the ethical guidelines of the medical center.

In addition, 548 PreservCyt samples, of which 155 were positive by the high-risk HPV HC2 assay (Qiagen, Hilden, Germany), were used for intralaboratory reproducibility and interlaboratory agreement analyses. Three portions of the original scraping samples were independently subjected to the Aptima HPV assay. The first two portions were tested within the same laboratory (VU University Medical Center, Amsterdam, The Netherlands), with an interval of 4 weeks, and the third was tested in a different laboratory (University of Tübingen, Tübingen, Germany). Different assay lots were used in the different laboratory settings.

Aptima HPV assay.

Cervical scraping samples were tested with the Aptima HPV assay according to the recommendations of the manufacturer (3). Briefly, a 1-ml aliquot of each PreservCyt sample was transferred to 2.9 ml of buffered detergent solution, and a 400-μl aliquot of the mixture was then tested on a semiautomated direct tube sampling (DTS) system (Gen-Probe). Assay results were interpreted on the basis of the signal/cutoff ratio for the analyte, and specimens with signal/cutoff ratios of ≥0.5 were considered positive.

GP5+/GP6+ PCR.

DNA was extracted from the PreservCyt samples using magnetic beads (Macherey-Nagel, Düren, Germany) on a Microlab STARlet robotic system (Hamilton Robotics, Martinsried, Germany), according to the manufacturer's instructions. DNA was subjected to GP5+/GP6+ PCR followed by enzyme immunoassay (EIA) using an oligonucleotide probe cocktail for 14 high-risk HPV types (i.e., types 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, and 68), as described previously (8). As a quality control for the presence of amplifiable DNA and the absence of PCR inhibitors in the isolated material, we performed a separate PCR for β-globin.

Genotyping assays.

Genotyping of EIA-positive GP5+/GP6+ PCR products was performed with a reverse line blot (RLB) assay as described previously (9). The RLB assay uses type-specific oligonucleotide probes to detect and to differentiate high-risk HPV types 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, and 68 and low-risk genotypes 6, 11, 26, 30, 32, 34, 40, 42, 43, 44, 53, 54, 55, 57, 61, 64, 67, 69, 70, 71 (CP8061), 72, 73, 81 (CP8304), 82 (MM4), IS39, 83 (MM7), 84 (MM8), 85, 86, Jc9710, and CP6108.

The linear array HPV genotyping test was performed with leftover DNA extracts according to the instructions of the manufacturer (Roche Molecular Diagnostics, Pleasanton, CA, USA). The linear array HPV genotyping test detects 37 mucosal HPV types, including high-risk genotypes 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, and 68 and low-risk genotypes 6, 11, 26, 40, 42, 53, 54, 55, 61, 62, 64, 66, 67, 69, 70, 71, 72, 73 (MM9), 81, 82 (MM4), 83 (MM7), 84 (MM8), 82 (IS39), and CP6108.

Statistical analysis.

The Aptima HPV assay analyses were performed in a blinded fashion, and results were then compared to GP5+/GP6+ PCR and histopathology/cytopathology data. Clinical sensitivity and specificity values of the Aptima HPV assay were compared with those of the GP5+/GP6+ PCR assay by noninferiority score testing using previously defined thresholds for noninferiority, i.e., relative sensitivity for CIN2+ of ≥90% and relative specificity for CIN2+ of ≥98% (7). For intralaboratory and interlaboratory agreement, a lower confidence limit of not less than 87% was used as the threshold (7). The level of agreement was determined by using the kappa statistic. R software was used for noninferiority score analysis. For other statistical computations, SPSS 15.0 software was used. The level of statistical significance was set at 0.05.

RESULTS

Clinical sensitivity and specificity analyses.

Valid Aptima HPV assay results were obtained for all 69 (100%) scraping samples from CIN2+ case subjects used for the sensitivity analysis and 839 (99.5%) of the scraping samples from control subjects without evidence of CIN2+ used for the specificity analysis. Of the 69 CIN2+ cases, 65 (94.2%) had positive Aptima HPV assay results (Table 1). This resulted in clinical sensitivity of the Aptima HPV assay for CIN2+ of 94.2% (95% confidence interval [CI], 85.5 to 97.8%). By comparison, the clinical sensitivity of GP5+/GP6+ PCR was 97.1% (95% CI, 89.1 to 99.3%) (67/69 cases). The four cases that did not score positive in the Aptima HPV assay included two CIN2 cases and one CIN3 case identified by a GP5+/GP6+ PCR-positive/cytology-negative screen test and one CIN3 case identified by a cytology-positive/GP5+/GP6+ PCR-negative screen test. The additional case that was negative by GP5+/GP6+ PCR but positive by the Aptima HPV assay involved a woman with CIN2 identified by abnormal cytology findings.

Table 1.

Aptima HPV assay findings among cervical scraping samples collected in population-based screening, stratified by case-control status

Sample type and Aptima HPV assay result No. with GP5+/GP6+ PCR result of:
Total no.
Negative Positive
Control
    Negative 771 22 793
    Positive 14 32 46
    Total 785 54 839
Case (CIN2+)
    Negative 1a 3 4
    Positive 1 64 65
    Total 2 67 69
a

Positive for HPV70 by the linear array HPV genotyping test.

Of the 839 controls without evidence of CIN2+, 46 (5.5%) were Aptima HPV assay positive (Table 1). This corresponds to a clinical specificity for CIN2+ of 94.5% (95% CI, 92.8 to 95.9%). The CIN2+ specificity of GP5+/GP6+ PCR in this series was 93.6% (95% CI, 91.7 to 95.0%) (785/839 controls). Both clinical sensitivity and specificity values of the Aptima HPV assay were noninferior to those of the reference assay, i.e., relative clinical sensitivity for CIN2+ of 0.97 (95% CI, 0.90 to 1.04; P = 0.039) and relative clinical specificity for CIN2+ of 1.01 (95% CI, 0.99 to 1.04; P = 0.00016).

Discrepancy analysis.

For discrepancy analysis of samples with discordant Aptima HPV assay versus GP5+/GP6+ PCR outcomes (i.e., 4 cases and 36 controls), we first performed genotyping of the GP5+/GP6+ PCR products of GP5+/GP6+ PCR-positive/Aptima HPV assay-negative samples using the reverse line blot genotyping method. In addition, to verify the GP5+/GP6+ PCR genotyping results and to further evaluate the samples that were Aptima HPV assay positive but GP5+/GP6+ PCR negative, leftover DNA extracts of all samples with discrepant results were subjected to the linear array HPV genotyping test. Among the case series (Table 2), all 3 Aptima HPV assay-negative but GP5+/GP6+ PCR-positive specimens showed genotypes in the linear array HPV genotyping test that were fully concordant with genotypes found by genotyping the GP5+/GP6+ PCR products (i.e., HPV16 [n = 1], HPV66 [n = 1], and HPV39/HPV58/HPV66 [n = 1]). E6/E7 mRNAs of all of these types are targets in the Aptima HPV assay. The one Aptima HPV assay-positive but GP5+/GP6+ PCR-negative case showed HPV52 in the linear array HPV genotyping test, and the single case that was negative in both the Aptima HPV assay and GP5+/GP6+ PCR showed HPV70 in the linear array HPV genotyping test.

Table 2.

Discrepancy analysis among GP5+/GP6+ PCR-positive and Aptima HPV assay-negative cases and GP5+/GP6+ PCR-negative and Aptima HPV assay-positive cases by linear array HPV genotyping testing and genotyping of GP5+/GP6+ PCR products

Discrepant case no. Results for:
HPV type(s) found by genotyping of:
GP5+/GP6+ PCR Aptima HPV assay Linear array HPV GP5+/GP6+ PCR products
1 Positive Negative 39, 58, 66 39, 58, 66
2 Positive Negative 66 66
3 Positive Negative 16 16
4 Negative Positive 52 NDa
a

ND, not done.

Among the controls (Table 3), 22 were Aptima HPV assay negative and GP5+/GP6+ PCR positive and 14 were Aptima HPV assay positive and GP5+/GP6+ PCR negative. Of the 22 Aptima HPV assay-negative but GP5+/GP6+ PCR-positive specimens, the linear array HPV genotyping test found 16 samples (73%) positive for one or more high-risk HPV types, 2 were positive for only low-risk HPV types, and 4 were HPV negative by the linear array HPV genotyping test. The linear array HPV genotyping test and RLB assay showed concordant data (i.e., full matches in genotypes detected) for 8 controls, compatible data (i.e., at least one genotype in common) for 5 controls, and discordant data (i.e., no match in genotypes found) for 9 controls (including the 4 that were negative in the linear array HPV genotyping test). Of the 14 Aptima HPV assay-positive but GP5+/GP6+ PCR-negative specimens, 9 were negative in the linear array HPV genotyping test, 4 were positive for types that could have been detected by GP5+/GP6+ PCR (i.e., HPV31 [n = 1], HPV39 [n = 1], HPV45 [n = 1], and HPV59 [n = 1]), and 1 was positive for a genotype that is not targeted by the Aptima HPV assay and GP5+/GP6+ PCR (i.e., HPV62).

Table 3.

Discrepancy analysis among GP5+/GP6+ PCR-positive and Aptima HPV assay-negative controls and GP5+/GP6+ PCR-negative and Aptima HPV assay-positive controls by linear array HPV genotyping testing and genotyping of GP5+/GP6+ PCR products

Discrepant control no. Results for:
HPV type(s) found by genotyping of:
GP5+/GP6+ PCR Aptima HPV assay Linear array HPV GP5+/GP6+ PCR productsa
1 Positive Negative 16 16
2 Positive Negative 31 31
3 Positive Negative 39 39
4 Positive Negative 51 51
5 Positive Negative 56, 70 56, 70
6 Positive Negative 58 58
7 Positive Negative 59 59
8 Positive Negative 66 66
9 Positive Negative 56, 73 56
10 Positive Negative 18, 59, 83 18
11 Positive Negative 35, 54, 62 35
12 Positive Negative 39, 40, 53, 59 39, 40
13 Positive Negative 45, 53 45
14 Positive Negative 16, 53, 54 33
15 Positive Negative 31, 58 51
16 Positive Negative 61 16
17 Positive Negative 61 16
18 Positive Negative HPVCP6108, 6, 51, 84, 52 45
19 Positive Negative Negative 31
20 Positive Negative Negative 59
21 Positive Negative Negative 66
22 Positive Negative Negative 66
23 Negative Positive Negative ND
24 Negative Positive Negative ND
25 Negative Positive Negative ND
26 Negative Positive Negative ND
27 Negative Positive Negative ND
28 Negative Positive Negative ND
29 Negative Positive Negative ND
30 Negative Positive Negative ND
31 Negative Positive Negative ND
32 Negative Positive 31 ND
33 Negative Positive 39, 67 ND
34 Negative Positive 45 ND
35 Negative Positive 59 ND
36 Negative Positive 62 ND
a

ND, not done.

Intralaboratory reproducibility and interlaboratory agreement analyses.

The intralaboratory reproducibility and interlaboratory agreement were assessed with another set of 548 scraping samples, of which 155 (28.2%) were high-risk HPV positive by HC2. Valid Aptima HPV assay results were obtained for 542 (98.9%), 529 (96.5%), and 548 (100%) of the specimens in laboratory 1/run 1, laboratory 1/run 2, and laboratory 2, respectively. After retesting of the specimens with initial invalid results, all samples generated valid test results, implying that the internal control at initial testing was indicating assay failure rather than sample failure. The Aptima HPV assay displayed high reproducibility, as reflected by intralaboratory reproducibility over time of 96.0% (95% CI, 94.4 to 97.3%) (526/548 samples; kappa = 0.89) (Table 4) and interlaboratory agreement of 96.7% (95% CI, 95.4 to 98.1%) (531/548 samples; kappa = 0.91) (Table 5).

Table 4.

Intralaboratory reproducibility over timea

Aptima HPV assay run 1 result No. with run 2 result of:
Total no.
Negative Positive
Negative 400 13 413
Positive 9 126 135
Total 409 139 548
a

The overall HPV test agreement was 96.0% (95% CI, 94.4 to 97.3%) (526/548 samples), with a kappa value of 0.89.

Table 5.

Interlaboratory agreementa

Aptima HPV assay laboratory 1 (run 1) result No. with laboratory 2 result of:
Total no.
Negative Positive
Negative 410 3 413
Positive 14 121 135
Total 424 124 548
a

Laboratory 1 was VU University Medical Center (Amsterdam, The Netherlands), and laboratory 2 was University Hospital Tübingen (Tübingen, Germany). The overall HPV test agreement was 96.7% (95% CI, 95.4 to 98.1%) (531/548 samples), with a kappa value of 0.91. Similar overall agreement was observed when data from laboratory 1, run 2, were compared with data from laboratory 2, i.e., 95.1% (95% CI, 93.3 to 96.6%) (521/548 samples), with a kappa value of 0.87.

DISCUSSION

In this study, we compared the clinical performance of the Aptima HPV assay with that of GP5+/GP6+ PCR in a cohort of screening participants. The clinical sensitivity and specificity for CIN2+ of the Aptima HPV assay were noninferior to those of GP5+/GP6+ PCR using the predetermined thresholds of 90% and 98%, respectively, as set out by an international consortium (7). Furthermore, the assay displayed sufficient intralaboratory reproducibility over time and interlaboratory agreement, with both lower confidence limits of percent agreement being much higher than 87% and the corresponding kappa values being greater than 0.5, in line with the proposed guidelines (7).

In the current study, the reference test was the GP5+/GP6+ PCR. Both GP5+/GP6+ PCR and HC2 have been clinically validated in randomized trials (7), and, like HC2, the GP5+/GP6+ PCR can be used for cross-sectional clinical validation of candidate HPV tests (10, 11). Collectively, our data indicate that the Aptima HPV assay meets the clinical and reproducibility criteria for HPV test requirements for primary cervical screening. It should be noted, however, that it has been stated that the clinical equivalence criteria based on relative cross-sectional accuracy are for HPV DNA tests and may not necessarily be valid for other molecular markers, such as overexpressed proteins or viral RNA (4). Since no long-term natural history data exist for such biomolecules, longitudinal data on the low-risk period after a negative Aptima HPV assay result are desirable to define the optimal screening interval for this assay.

Our findings are fully in line with those of other studies that compared the clinical performance of the Aptima HPV assay with that of a variety of commercially available HPV assays for ASC-US/low-grade squamous intraepithelial lesion triage or primary screening (1225). The current study adds because it addresses all aspects of the guidelines, including intralaboratory reproducibility over time and interlaboratory agreement of the Aptima HPV assay.

Among the case samples, one specimen was negative with both the Aptima HPV assay and GP5+/GP6+ PCR. The linear array HPV genotyping test revealed HPV70, a genotype that is not targeted by either assay. In addition, four discordant results between the Aptima HPV assay and GP5+/GP6+ PCR were seen in the case series (two CIN2 cases and two CIN3 cases). These mostly included cytologically negative for intraepithelial lesion or malignancy samples (3/4 cases), and the findings were not particularly related to a specific HPV type. The controls that were not detected by the Aptima HPV assay but were positive in the GP5+/GP6+ PCR assay may be explained by the absence of detectable mRNA while HPV DNA is measurable. The GP5+/GP6+ PCR and linear array HPV genotyping test revealed a lower level of agreement in genotyping results among discordant control samples, compared to case samples. This suggests that these controls harbor lower viral loads, which in part may also underlie the negative assay results with the Aptima HPV assay. Previous studies have shown that lower viral loads are less likely to coincide with detectable E6/E7 mRNA levels (25). Most of the controls that were detected by the Aptima HPV assay over the GP5+/GP6+ PCR were also negative by the linear array HPV genotyping test, which may reflect higher viral mRNA loads than DNA loads. One instance of discrepant findings for the Aptima HPV assay versus GP5+/GP6+ PCR among controls may be explained by cross-reactivity of the Aptima HPV assay with a nontargeted HPV type (e.g., HPV62). In total, our study indicates that the inherent diversity in assay chemistry and related possible minor differences at the analytical level between the Aptima HPV assay and GP5+/GP6+ PCR have no significant effects on their clinical performances.

In conclusion, our data indicate that the Aptima HPV assay, in a cross-sectional setting, performs clinically comparably to GP5+/GP6+ PCR and meets the cross-sectional criteria for clinical sensitivity and specificity for CIN2+, intralaboratory reproducibility over time, and interlaboratory agreement of the international guidelines for HPV test requirements for primary cervical screening (7). Longitudinal data are needed to ensure long-term negative predictive values similar to those for the validated HPV DNA tests GP5+/GP6+ PCR and HC2.

ACKNOWLEDGMENTS

This research was supported by an unrestricted grant from Gen-Probe Inc. The source of funding did not have any influence on the design and analysis of the results.

C.J.L.M.M., P.J.F.S., and D.A.M.H. are shareholders of Self-screen BV, a spin-off company of VU University Medical Center. C.J.L.M.M. has received speaker's fees from Roche, Qiagen, GSK, and Merck. J.B. has received speaker's fees from Qiagen. T.I. has received speaker's fees from Roche and Hologic, and the University Hospital of Tübingen has received an unrestricted research grant from Gen-Probe. All other authors declare that they have no conflicts of interest.

Footnotes

Published ahead of print 28 August 2013

REFERENCES

  • 1. Dockter J, Schroder A, Eaton B, Wang A, Sikhamsay N, Morales L, Giachetti C. 2009. Analytical characterization of the APTIMA HPV assay. J. Clin. Virol. 45(Suppl 1):S39–S47 [DOI] [PubMed] [Google Scholar]
  • 2. Getman D, Aiyer A, Dockter J, Giachetti C, Zhang F, Ginocchio CC. 2009. Efficiency of the APTIMA HPV assay for detection of HPV RNA and DNA targets. J. Clin. Virol. 45(Suppl 1):S49–S54 [DOI] [PubMed] [Google Scholar]
  • 3.Hologic Gen-Probe 2011. APTIMA HPV assay package insert. Hologic Gen-Probe Inc., San Diego, CA [Google Scholar]
  • 4. Arbyn M, Ronco G, Anttila A, Meijer CJ, Poljak M, Ogilvie G, Koliopoulos G, Naucler P, Sankaranarayanan R, Peto J. 2012. Evidence regarding human papillomavirus testing in secondary prevention of cervical cancer. Vaccine 30(Suppl 5):F88–F99 [DOI] [PubMed] [Google Scholar]
  • 5. Murphy J, Kennedy EB, Dunn S, McLachlin CM, Fung Kee Fung M, Gzik D, Shier M, Paszat L. 2012. HPV testing in primary cervical screening: a systematic review and meta-analysis. J. Obstet. Gynaecol. Can. 34:443–452 [DOI] [PubMed] [Google Scholar]
  • 6. Meijer CJ, Berkhof J, Heideman DA, Hesselink AT, Snijders PJ. 2009. Validation of high-risk HPV tests for primary cervical screening. J. Clin. Virol. 46(Suppl 3):S1–S4 [DOI] [PubMed] [Google Scholar]
  • 7. Meijer CJ, Berkhof J, Castle PE, Hesselink AT, Franco EL, Ronco G, Arbyn M, Bosch FX, Cuzick J, Dillner J, Heideman DA, Snijders PJ. 2009. Guidelines for human papillomavirus DNA test requirements for primary cervical cancer screening in women 30 years and older. Int. J. Cancer 124:516–520 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Snijders PJF, Van den Brule AJC, Jacobs MV, Pol RP, Meijer CJLM. 2005. HPV DNA detection and typing in cervical scrapes. Methods Mol. Med. 119:101–114 [DOI] [PubMed] [Google Scholar]
  • 9. van den Brule AJ, Pol R, Fransen-Daalmeijer N, Schouls LM, Meijer CJ, Snijders PJ. 2002. GP5+/6+ PCR followed by reverse line blot analysis enables rapid and high-throughput identification of human papillomavirus genotypes. J. Clin. Microbiol. 40:779–787 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Hesselink AT, Meijer CJLM, Poljak M, Berkhof J, van Kemenade FJ, van der Salm ML, Bogaarts M, Snijders PJF, Heideman DAM. 2013. Clinical validation of the Abbott RealTime High Risk HPV assay according to the guidelines for human papillomavirus DNA test requirements for cervical screening. J. Clin. Microbiol. 51:2409–2410 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Hesselink AT, Heideman DA, Berkhof J, Topal F, Pol RP, Meijer CJ, Snijders PJ. 2010. Comparison of the clinical performance of PapilloCheck human papillomavirus detection with that of the GP5+/6+-PCR-enzyme immunoassay in population-based cervical screening. J. Clin. Microbiol. 48:797–801 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Szarewski A, Mesher D, Cadman L, Austin J, Ashdown-Barr L, Ho L, Terry G, Liddle S, Young M, Stoler M, McCarthy J, Wright C, Bergeron C, Soutter WP, Lyons D, Cuzick J. 2012. Comparison of seven tests for high-grade cervical intraepithelial neoplasia in women with abnormal smears: the Predictors 2 study. J. Clin. Microbiol. 50:1867–1873 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Clad A, Reuschenbach M, Weinschenk J, Grote R, Rahmsdorf J, Freudenberg N. 2011. Performance of the Aptima high-risk human papillomavirus mRNA assay in a referral population in comparison with Hybrid Capture 2 and cytology. J. Clin. Microbiol. 49:1071–1076 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Nieves L, Enerson CL, Belinson S, Brainard J, Chiesa-Vottero A, Nagore N, Booth C, Pérez AG, Chávez-Avilés MN, Belinson J. 2013. Primary cervical cancer screening and triage using an mRNA human papillomavirus assay and visual inspection. Int. J. Gynecol. Cancer 23:513–518 [DOI] [PubMed] [Google Scholar]
  • 15. Stoler MH, Wright TC, Jr, Cuzick J, Dockter J, Reid JL, Getman D, Giachetti C. 2013. APTIMA HPV assay performance in women with atypical squamous cells of undetermined significance cytology results. Am. J. Obstet. Gynecol. 208:144.e1–144.e8 [DOI] [PubMed] [Google Scholar]
  • 16. Arbyn M, Roelens J, Cuschieri K, Cuzick J, Szarewski A, Ratnam S, Reuschenbach M, Belinson S, Belinson JL, Monsonego J. 2013. The APTIMA HPV assay versus the Hybrid Capture 2 test in triage of women with ASC-US or LSIL cervical cytology: a meta-analysis of the diagnostic accuracy. Int. J. Cancer 132:101–108 [DOI] [PubMed] [Google Scholar]
  • 17. Szarewski A, Ambroisine L, Cadman L, Austin J, Ho L, Terry G, Liddle S, Dina R, McCarthy J, Buckley H, Bergeron C, Soutter P, Lyons D, Cuzick J. 2008. Comparison of predictors for high-grade cervical intraepithelial neoplasia in women with abnormal smears. Cancer Epidemiol. Biomarkers Prev. 17:3033–3042 [DOI] [PubMed] [Google Scholar]
  • 18. Monsonego J, Hudgens MG, Zerat L, Zerat JC, Syrjänen K, Halfon P, Ruiz F, Smith JS. 2011. Evaluation of oncogenic human papillomavirus RNA and DNA tests with liquid-based cytology in primary cervical cancer screening: the FASE study. Int. J. Cancer 129:691–701 [DOI] [PubMed] [Google Scholar]
  • 19. Wu R, Belinson SE, Du H, Na W, Qu X, Wu R, Liu Y, Wang C, Zhou Y, Zhang L, Belinson JL. 2010. Human papillomavirus messenger RNA assay for cervical cancer screening: the Shenzhen Cervical Cancer Screening Trial I. Int. J. Gynecol. Cancer 20:1411–1414 [DOI] [PubMed] [Google Scholar]
  • 20. Monsonego J, Hudgens MG, Zerat L, Zerat JC, Syrjänen K, Smith JS. 2012. Risk assessment and clinical impact of liquid-based cytology, oncogenic human papillomavirus (HPV) DNA and mRNA testing in primary cervical cancer screening (the FASE study). Gynecol. Oncol. 125:175–180 [DOI] [PubMed] [Google Scholar]
  • 21. Ratnam S, Coutlee F, Fontaine D, Bentley J, Escott N, Ghatage P, Gadag V, Holloway G, Bartellas E, Kum N, Giede C, Lear A. 2011. Aptima HPV E6/E7 mRNA test is as sensitive as Hybrid Capture 2 Assay but more specific at detecting cervical precancer and cancer. J. Clin. Microbiol. 49:557–564 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Dockter J, Schroder A, Hill C, Guzenski L, Monsonego J, Giachetti C. 2009. Clinical performance of the APTIMA HPV Assay for the detection of high-risk HPV and high-grade cervical lesions. J. Clin. Virol. 45(Suppl 1):S55–S61 [DOI] [PubMed] [Google Scholar]
  • 23. Mesher D, Szarewski A, Cadman L, Austin J, Ashdown-Barr L, Ho L, Terry G, Young M, Stoler M, Bergeron C, McCarthy J, Wright C, Liddle S, Soutter WP, Lyons D, Cuzick J. 2013. Comparison of human papillomavirus testing strategies for triage of women referred with low-grade cytological abnormalities. Eur. J. Cancer 49:2179–2186 [DOI] [PubMed] [Google Scholar]
  • 24. Cuzick J, Cadman L, Mesher D, Austin J, Ashdown-Barr L, Ho L, Terry G, Liddle S, Wright C, Lyons D, Szarewski A. 2013. Comparing the performance of six human papillomavirus tests in a screening population. Br. J. Cancer 108:908–913 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Broccolo F, Fusetti L, Rosini S, Caraceni D, Zappacosta R, Ciccocioppo L, Matteoli B, Halfon P, Malnati MS, Ceccherini-Nelli L. 2013. Comparison of oncogenic HPV type-specific viral DNA load and E6/E7 mRNA detection in cervical samples: results from a multicenter study. J. Med. Virol. 85:472–482 [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)