Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2014 Dec 1.
Published in final edited form as: J Magn Reson. 2013 Oct 31;237:169–174. doi: 10.1016/j.jmr.2013.10.013

Low-field MRI can be more sensitive than high-field MRI

Aaron M Coffey 1,2, Milton Truong 1, Eduard Y Chekmenev 1,2,3,
PMCID: PMC3897717  NIHMSID: NIHMS537250  PMID: 24239701

Abstract

MRI signal-to-noise ratio (SNR) is the key factor for image quality. Conventionally, SNR is proportional to nuclear spin polarization, which scales linearly with magnetic field strength. Yet ever-stronger magnets present numerous technical and financial limitations. Low-field MRI can mitigate these constraints with equivalent SNR from non-equilibrium ‘hyperpolarization’ schemes, which increase polarization by orders of magnitude independently of the magnetic field. Here, theory and experimental validation demonstrate that combination of field independent polarization (e.g. hyperpolarization) with frequency optimized MRI detection coils (i.e. multi-turn coils using the maximum allowed conductor length) results in low-field MRI sensitivity approaching and even rivaling that of high-field MRI. Four read-out frequencies were tested using samples with identical numbers of 1H and 13C spins. Experimental SNRs at 0.0475 T were ∼40% of those obtained at 4.7 T. Conservatively, theoretical SNRs at 0.0475 T 1.13-fold higher than 4.7 T were possible despite an ∼100-fold lower detection frequency, indicating feasibility of high-sensitivity MRI without technically challenging, expensive high-field magnets. The data at 4.7 T and 0.0475 T was obtained from different spectrometers with different RF probes. The SNR comparison between the two field strengths accounted for many differences in parameters such as system noise figures and variations in the probe detection coils including Q factors and coil diameters.

Keywords: hyperpolarization, MRI, NMR, low-field MRI, Litz wire, 13C

1. Introduction

The superlative factor governing MRI image quality is the signal-to-noise ratio (SNR). High-field MRI using superconductive magnets has revolutionized medical diagnostics, with ever-increasing magnetic fields fostering sensitivity improvements via higher SNR. Yet low magnetic field strengths offer many attractive advantages such as reduced magnet size and cost, greater subject safety due to lower absorption of radio-frequency energy, and negligible subject induced magnetic field inhomogeneities [1]. These low-field MRI advantages can potentially be truly transformative, permitting performance of the entire MRI exam in seconds [2] provided sufficient SNR is available. The SNR of conventional, higher field detection is a complex equation of nuclear spin polarization, detection frequency, and other factors [3] arising from Faraday inductive detection. Nuclear spin polarization is a key factor contributing to this SNR. It is a relative measure of nuclear spin alignment with the applied magnetic field B0. Equilibrium nuclear spin polarization, which is only 10-6 -10-5 at conditions of human body temperature and B0 of several Tesla, scales linearly with B0 and, therefore, SNR decreases at low field. However, non-equilibrium ‘hyperpolarization’ schemes make polarization independent of the detection field, providing a unique opportunity for high SNR and image quality at low field.

Hyperpolarization techniques temporarily increase polarization by several orders of magnitude to unity, or 100, referred to as the hyperpolarized state. These techniques include dissolution-Dynamic Nuclear Polarization (DNP) [4], Para-Hydrogen and Synthesis Allow Dramatically Enhanced Nuclear Alignment (PASADENA) [5], Spin Exchange Optical Pumping (SEOP) [6] and others. Regardless of the hyperpolarization approach used, the main goal in the context of biomedical applications is preparation of exogenous contrast agents with high polarization to enable molecular imaging of relatively dilute spin systems otherwise not amenable by conventional MRI. Examples of such contrast agents include hyperpolarized noble gases for lung imaging and 13C-labeled metabolites. The latter can be out of balance due to abnormal metabolism, and therefore act as reporters or biomarkers of diseases including those of cancer. 13C-pyruvate contrast agent reporting on elevated rate of glycolysis in cancer is one such example already in clinical trial [7] due to recognized status as a promising molecular imaging agent for prostate cancer.

While low-field MRI has been shown useful for hyperpolarized states of noble gases in lung imaging [8] and much progress has been made for utilizing hyperpolarized contrast agents in molecular imaging [4, 7], the B0 field independent nature of polarization in hyperpolarized contrast agents has not been fully taken advantage of. Maximizing imaging detection sensitivity as a function of both detection field B0 [9] and frequency ω0 still remains a challenge. Prior systematic efforts to develop a theoretical SNR foundation for MRI neglected optimization of the Faraday induction coils to the detection frequency ω0 [3, 9-11]. We recently demonstrated 13C hyperpolarized signals can be nearly field independent [12]. Here, we present a general theoretical description of hyperpolarized MRI sensitivity in the form of SNR as a function of detection frequency ω0 and experimental validation at four frequencies: 0.5 MHz, 2.0 MHz, 50 MHz, and 200 MHz. It is concluded that low-field MRI can be significantly more sensitive than high-field MRI for detection of hyperpolarized spin states. This is contrary to conventional wisdom that high-field MRI is always more sensitive.

2. Materials and Methods

Sample phantoms and preparation of nuclear spin polarizations

1H and 13C spectroscopic and imaging comparisons utilized two spherical phantoms of sodium 1-13C-acetate. The phantom for 1H studies was 1.0 g of sodium 1-13C-acetate (product #279293, Isotec-Sigma-Aldrich) dissolved in 99.8% D2O resulting in 2.8 mL total volume. A larger sample for 13C studies consisted of 5.18 g of sodium 1-13C-acetate dissolved in 99.8% D2O resulting in 17.5 mL total volume. High-field data were acquired on a 4.7 T Varian small animal MRI scanner with a multi-nuclear RF probe (Doty Scientific, Columbia, SC). Low-field data were collected on a 0.0475 T spectrometer (Magritek, Wellington, New Zealand) equipped with a custom gradient coil insert (Magritek) and in-house developed H-X and X-H radiofrequency (RF) probes, where the X channel was tuned and optimized to the 13C resonance frequency. The in-house probes consisted of inner ω0 optimized solenoid detection coils and outer saddle excitation coils for multi-nuclear experiments. The details are provided in Supporting Information.

Prior to spectroscopic or MR imaging at 0.0475 T as shown in Figs. 1 and 2, the sample was prepolarized [13]. Prepolarization for detection at 0.0475 T was necessary to (i) simulate the condition of the hyperpolarized state (when nuclear spin polarization significantly exceeds the Boltzmann distribution), and (ii) to obtain nuclear spin polarization close to 4.7 T Boltzmann equilibrium levels for comparison of detected SNR. For 1H studies, the sample was prepolarized at 9.4 T for > 30 seconds, and detection at 0.0475 T occurred following an ∼5 second transfer delay. At detection, the resulting 1H polarization was P = (1.050±0.016)*10-5. Polarization level was calculated by comparison of the prepolarized NMR signal intensity with that of the thermally polarized sample. Conditions simulating hyperpolarized 1H detection were explored in addition to 13C detection because of the potential advantages of indirect proton MRI of hyperpolarized contrast agents in vivo [14]. Similarly for 13C studies, the sample was prepolarized at 7.0 T for > 5 minutes, and detected at 0.0475 T following an ∼5 second transfer delay. 13C polarization was calculated as P = (4.70±0.02)*10-6 by comparing prepolarized NMR signal intensity with that of the thermally polarized sample.

Fig 1.

Fig 1

13C and 1H MRI of sodium 1-13C-acetate. 4.7 T acquisition used Boltzmann 13C P = 4.06*10-6 and 1H P = 1.61*10-5, and 0.0475 T used approximately the same polarization levels, 13C P = 4.70*10-6 and 1H P = 1.05*10-5. All measurements used a spherical phantom: for 1H 1.0 g sodium 1-13C-acetate in 99.8% D2O with 2.8 mL total volume and for 13C 5.18 g sodium 1-13C-acetate in 99.8% D2O with 17.5 mL total volume. All acquisition and processing parameters were identical except 13C excitation pulse angle α. No image extrapolation or zero filling was used.

Fig 2.

Fig 2

13C and 1H NMR spectroscopy of sodium 1-13C-acetate. 4.7 T acquisition used Boltzmann 13C P = 4.06*10-6 and 1H P = 1.61*10-5, and 0.0475 T used approximately the same polarization levels, 13C P = 4.70*10-6 and 1H P = 1.05* 10-5. All measurements used a spherical phantom: for 1H 1.0 g sodium 1-13C-acetate in 99.8% D2O with 2.8 mL total volume and for 13C 5.18 g sodium 1-13C-acetate in 99.8% D2O with 17.5 mL total volume. All NMR acquisition and processing parameters were identical.

Spectroscopic results (Fig. 2) used identical acquisition parameters on the two MRI systems: square RF excitation pulses with calibrated τ90°, 1 k complex acquisition points, spectral width of 2 kHz, and 500 ms acquisition time. Imaging (Fig. 1) was similarly performed with identical parameters with the exception of 13C RF excitation pulse angle α. On the 4.7 T scanner, images were acquired with Varian's 2D balanced FSSFP sequence. At 0.0475 T, Magritek's fast 2D gradient echo sequence was used. For 1H on both systems RF excitation pulse angle α = 18°, spectral width was 10 kHz, and acquisition time was 6.4 ms per line of k-space. 13C imaging parameters were spectral width of 5 kHz, 6.4 ms acquisition time, pulse angle α = 90° at 4.7 T, and α = 18° at 0.0475 T. For the latter the reduced angle was necessary to avoid consuming too much polarization during gradient echo imaging acquisition of k-space. 1H imaging in-plane resolution was 0.375×0.375 mm2, (field of view = 24×24 mm2), and 13C was 2.5×2.5 mm2 (field of view = 80×80 mm2) respectively. The resulting 1H and 13C images had 64×64 and 32×32 imaging matrices, and they are presented without any extrapolation or any further manipulation.

3. Results and Discussion

Seminal work by Hoult [3, 15] described the SNR for Faraday inductive detection of the MRI signal in RF coils as

Ψrms=12|ε|V=KB1VSω0μNPN2[4FkΔf(TCζRC+TSRS)]1/2. (1)

Eq. (1) relates the magnitude of the electromotive force induced in the RF coil |ε|, noise V, oscillating RF field homogeneity over the subject K, oscillating RF field strength per unit current over the subject B1, subject volume VS, detection frequency ω0, nuclear magnetic moment μN, nuclear spin polarization P, number of spins N, preamplifier noise figure F, Boltzmann's constant k, receiver bandwidth Δf, RF coil temperature TC, proximity effect factor ζ, RF coil resistance RC, subject temperature TS, and equivalent subject resistance RS = RI + RE, the sum of subject inductive and dielectric losses respectively [16, 17], also known as body noise. With the RF coil of a fixed geometry (e.g. same wire length, number of turns, etc.), polarization P induced by B0, and negligible body noise from the subject with respect to RF coil noise (i.e. conditions of TCζ RCTSRS), then by Eq. (1) for a range of detection frequencies the SNRω07/4, corresponding to the common situation of high resolution NMR [18].

However, if the polarization is endowed by hyperpolarization produced independently of the detection magnetic field, the SNRω07/4 dependence becomes SNRω03/4, because one ω0 is eliminated. Additionally, if the RF detection coil is optimized by using a maximum conductor length l consonant with ω0 of up to λ/10 to create turns (i.e. forming multi-turn inductors while avoiding challenges such as onset of elevated electromagnetic radiation losses owing to increasing efficiency as antennas) [19], then SNR dependence on detection frequency ω0 becomes more complex to evaluate. The rule l = λ/10 for maximum, continuous conductor length finds alternative expression in terms of the detection frequency ω0 as l = παc/5ω0, where αc is the conductor's wave propagation velocity. Provided the coil diameter dC is fixed, lowering ω0 then results in greater conductor length l and hence more turns n, variables common to both B1 and RC. Thus, the resonance frequency defines the number of turns n and the conductor length l using this approach. This commonality represents an underlying frequency dependency which may be consolidated; see Appendix A. for details pertaining to modification of Eq. (1) using l = παc / 5ω0 to redefine B1 and RC in terms of the resonance frequency. With B1 and RC in Eq. (1) expressed as consolidated functions of ω0, then for the condition of coil noise dominance, or TCζRCTSRS, SNR for the hyperpolarized state becomes

Ψrms,hp=1525/4KVSμNPN[μrρ(TC)]1/4(αcdWFkTCΔfζ)1/2μ03/4ω01/4dC2 (2)

for wire diameter dW, coil diameter dC, permeability of free space μ0, relative conductor permeability μr, conductor resistivity ρ, and velocity factor α correcting the speed of light c to conductor wave propagation velocity. Eq. (2) states SNR is a function of coil geometry, basic conductor properties, and fundamental physical constants. It can be recast to the condition of non-hyperpolarized (i.e. thermal) polarization for ω0 optimized RF coils by substituting polarization induced by B0 into Eq. (2), or P = constω0 (see Appendix A). Consequently, Eq. (2)'s ω01/4 dependence becomes ω05/4. The ω05/4 dependence differs from conventional ω07/4 dependence [3] owing to an additional cancellation factor of ω01/2 obtained from interplay of gains and losses associated with the terms B1 and RC under the condition of l = λ/10 = παc / 5ω0 . While ω05/4 dependence clearly retains the favorability of conventional MR detection at higher fields, SNRω01/4 For hyperpolarized spin states produced independently of the detection magnetic field represents very weak frequency dependence.

As SNRQ [3, 11] low-field detection may offer a favorable arena for SNR improvements on the basis of Q. The inductor electrical quality factor Q is expressed as Q = ω0L /RC, where L and RC are the coil inductance and resistance respectively. While Qω01/2 for TCζRCTSRS conventionally [20] (i.e. sample losses are negligible), the frequency optimized coil (where l = παc / 5ω0) has a more complex dependence of Q on frequency ω0 because Qω01/2L and L itself exhibits frequency dependence due to change in the number of turns and other coil geometric factors. Nevertheless, a detailed analysis of experimental Q values is provided in the Supporting Information. To summarize, the Q values of the tuned RF circuits were 28 (0.5 MHz), 90 (50 MHz), 62 (2.0 MHz), 69 (200 MHz).

As illustrated by Eq. (2), low-field MRI affords unique opportunities to significantly increase overall SNR when RF coil noise dominates for detection frequencies below 10 MHz [15], or TCζRCTSRS. Low-field magnets allow solenoid geometry for the RF detection coil [8], unlike the high field's volume coils— birdcage, Alderman-Grant, and others— which permits 2.6-fold more efficient use of wire to form inductive loops resulting in 3.1-fold greater overall SNR [3]. 100-fold frequency differences result in only 3.2-fold SNR difference. Thus, coil geometry alone nearly compensates a 100-fold decrease in detection frequency for hyperpolarized spin states, 3.1 versus 3.2. Other wire-related factors present opportunities for further SNR gains.

Low field affords SNR challenges and opportunities related to the RF coil conductor. The proximity effect ζ between neighboring turns in multi-turn RF coils increases resistance RC with concomitant SNR loss. However, this loss can be minimized through unique winding geometries as exemplified by crystal radio coils where improved turn density and hence B1 promote higher SNR. Coil resistance RC can simply be directly minimized for higher SNR through lowering coil temperature TC and using Litz wire [21] or superconducting wire. Albeit constructing superconducting or reduced temperature TC RF detection coils is a technological feat, using Litz wire is straightforward. Coil resistance reduction by a practical Litz wire factor of 0.44 due to reduced ζ and skin effect losses [21], for example, leads to SNR gain by 1.5 fold. Consequently, low-field SNR for hyperpolarized states better than high-field SNR is possible despite ζ loss related challenges on the order of 1.3 for RF coils with equally spaced wire and a length-to-diameter ratio of two [22]. We note that Litz wire does not provide benefits at the frequencies of high-field MRI coils. Nevertheless, MR detection at 100-fold lower frequency should yield 1.13-fold greater SNR compared to high-frequency detection (details given in Supporting Information, Table S6).

The estimated 1.13-fold SNR gain is very conservative. The afore-mentioned superconducting low-field RF coils can provide further SNR gains of as much as an order of magnitude [20]. Additionally, the hyperpolarized state's SNRω01/4 stems from using lossy conductors (RC ≠ 0) with frequency dependent resistance, whereas superconducting RF coils (RC = 0) have SNR nominally independent of detection frequency ω0, ignoring subject losses RS [9]. Subject losses are negligible in the low-frequency regime for clinical scale RF coils [1], but clinical high-field MRI at B0 ≥ 1.5 T is well known for body noise being greater than coil noise, i.e. TSRSTCζRC [20]. Also, while dielectric losses RE are commonly mitigated by various techniques, subject inductive losses RIω02 [3, 11, 15, 23-26] are inescapable. Thus, subject noise limits the overall SNR achievable at both high field and low field, but proves more favorable at lower frequencies. We note that not fulfilling the condition of coil noise dominance, or TCζRCTSRS, invalidates Eq. (2), but fully accounting for RS in the context of ω0 optimized coils is very complex and outside the main scope of this work. It should be stressed, however, that fundamental subject noise barriers hinder high-field detection [15] of hyperpolarized contrast agents.

Experimental validation of Eq. (2) analyzed 1H and 13C SNR for similar 1H and 13C spin polarization P of identical samples on 4.7 T Varian and 0.0475 T MRI scanners. Figs. 1 and 2 demonstrate imaging and spectroscopic detection using frequency optimized RF coils with similar coil sizes suitable for in vivo animal studies. After accounting for experimental limitations and imperfections when working with two different MRI systems (Supporting Information, Table S4), then by Eq. (2) the theoretical ratio of SNRs at 0.0475 T and 4.7 T magnetic field strengths (i.e. SNRs0.0475T/SNR4.7T) was 45% for 1H and 46% for 13C respectively. Using spectroscopic acquisition, Fig. 2, the experimentally determined ratio of SNRs at 0.0475 T and 4.7 T was 41±1% for 1H and 40±1% for 13C. This reflects good quantitative agreement with theory despite nearly 100-fold difference in field and frequency. While SNR0.0475T/SNR4.7T of 113% was expected theoretically for detection of 1H and 13C nuclei, experimental limitations as described in the Supporting Information significantly hampered 1H and 13C detection sensitivity at 0.0475 T. In particular, deviation from ideal Litz wire selection and RF coil coupling to other components due to a severe space limitation of an 89-mm I.D. gradient insert bore reduced achievable maximum SNR by a factor of 2.

The presented work is motivated by low-field MRI of 13C hyperpolarized contrast agents through direct detection or indirectly through protons. Proton detection takes advantage of long-lived hyperpolarization stored on 13C carboxyl carbons with the added benefits of increased sensitivity and lower gradient power requirements [7] in vitro [27] and in vivo [14]. The above theoretical and experimental SNR comparison of low and high field detection was conducted under the condition of RF coil noise dominance rather than subject/body noise dominance. The latter is in fact typical for high-field MRI of large animals and humans [20]. Subject noise disproportionately penalizes high field [20] but not low-field inductive detection [1]. As a result, this framework of RF coil optimization through maximized use of conductor length (i.e. approaching l = λ /10 = παc / 5ω0) will likely favor hyperpolarized contrast agent low-field detection at clinical scale even more.

4. Conclusion

To summarize, a theoretical basis for SNR of hyperpolarized contrast agents as a function of detection frequency is described and validated experimentally. Low-field MRI can indeed be more sensitive for hyperpolarized contrast agents. Moreover, hyperpolarized low-field MRI in combination with cryogenically cooled RF coils [20, 28] can significantly surpass the sensitivity of hyperpolarized high-field MRI, which contradicts the ‘conventional wisdom’ of high-field MRI SNR superiority. In conclusion, low-field hyperpolarized MRI has the potential to revolutionize molecular imaging by providing better quality images, allowing sub-minute examinations at a significantly reduced cost, and perhaps solving one of the greatest challenges in Radiology: advancing human health while reducing the costs [29] of emerging great technologies [30].

Supplementary Material

01

Fig 3.

Fig 3

MRI instrumentation. A) Schematic representation and alignment of the RF coils and shield to the B0 magnet, and B) scale comparison of the 4.7 T and 0.0475 T MRI scanners. The dimensions provided are in meters.

Highlights.

  • SNR theory of hyperpolarized MR as a function of B0 and frequency optimized RF coil

  • Low-field MRI sensitivity approaches and even rivals that of high-field MRI

  • RF coil development for low-field MR

  • Multi-nuclear imaging and spectroscopy of hyperpolarization at low B0 field

Acknowledgments

We thank Professor Boyd M. Goodson for his comments on the writing of the manuscript and gratefully acknowledge funding support from NIH R25 CA136440, 3R00 CA134749, DOD CDMRP W81XWH-12-1-0159/BC112431.

Appendix A. Derivations of SNR equations

The SNR for MRI has been described in seminal work by Hoult [3, 15] for Faraday inductive detection of the MRI signal in radiofrequency (RF) coils as

Ψrms=12|ε|V=KB1VSω0μNPN2[4FkΔf(TCζRC+TSRS)]1/2 (A1)

where the equation variables are as previously defined for Eq. (1). With the RF coil of a fixed geometry— e.g. same wire length, number of turns, etc.— polarization P induced by B0, and negligible sample noise with respect to RF coil noise, i.e. conditions of TCζRCTSRS , then by Eq. (A1) the SNRω07/4. However, if polarization P is endowed by hyperpolarization independent of B0 and the RF detection coil is specifically optimized for ω0, i.e. using the maximum allowed conductor length or using l = λ/10, SNR dependence on detection frequency ω0 becomes more complex due to B1 and RC frequency dependency.

An optimal RF coil's maximum allowed conductor length is l = λ/10 when maintaining current phase across the coil. The detection frequency ω0 explicitly defines this length in lieu of wavelength λ, since αc = (λω0)/(2π), as

l=παc5ω0. (A2)

Velocity factor α corrects the speed of light c to the conductor's wave propagation velocity. The RF field B1 generated by conductor unit current J finds similar re-expression. In the quasi-static limit Biot-Savart law defines B1 [3] generated by n loops of radius a as [31]

B1(z)J=μ0n2a2(z2+a2)3/2 (A3)

for distance z from the RF coil center and permeability of free space μ0. Albeit direct calculation of B1 homogeneity factor K is feasible [32], since K ∼ 1 typically and coil diameter dC = 2a, B1 over the sample is approximately B1 at the coil center (z = 0), or

B1(0)J=μ0(ndC). (A4)

With coil length lCl for n > 1, n = (llC)/πdCl/πdC and B1 is written as

B1=μ0αc5ω0dc2. (A5)

The inverse proportionality in B1 with respect to ω0 in Eq. (A5) is driven by the change in the number of turns n, Eq. (A4) under the condition of fixed coil diameter dC and the condition imposed by Eq. (A2).

The current J reduces exponentially to 1/e of its conductor surface value at a distance known as the skin depth δ, which varies with ω0 as

δ=(2ρ(TC)μrμ0ω0)1/2 (A6)

for relative permeability μr and temperature dependent resistivity ρ. J flows through an effective cross-sectional areal product of δ and conductor circumference, or A = πdWδ, for wire diameter dW. When constructed of a single, solid conductor, coil resistance RC is expressed using Eq. (A2) and Eq. (A6) as

Rc=ρlA=ρlπdWδ=(αc5dW)(μrμ0ρ(TC)2ω0)1/2. (A7)

When Eqs. (A5) and (A7) are substituted into Eq. (A1), Eq. (A1) becomes for conditions TCζRCTSRS

Ψrms,hp=1525/4KVSμNPN[μrρ(TC)]1/4(αcdWFkTCΔfς)1/2μ03/4ω01/4dC2. (A8)

The proportionality to detection frequency of ω01/4 Stems from using lossy conductors, Eq. (A7).

With polarization P endowed by B0, i.e. the non-hyperpolarized condition, polarization is written as

P=(I+1)ω03kTS. (A9)

After substitution of Eq. (A9) and nuclear magnetic moment μNγћI, Eq. (A8) becomes

Ψrms=13525/4KVSγ2I(I+1)NkTS[μrρ(TC)]1/4(αcdWFkTCΔfζ)1/2μ03/4ω05/4dC2 (A10)

for gyromagnetic ratio γ, Planck's constant ћ, and nuclear spin number I.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Hayden ME, Bidinosti CP, Chapple EM. Specific absorption rates and signal-to-noise ratio limitations for MRI in very-low magnetic fields. Concept Magnetic Res A. 2012;40A:281–294. [Google Scholar]
  • 2.Imai H, Kimura A, Hori Y, Iguchi S, Kitao T, Okubo E, Ito T, Matsuzaki T, Fujiwara H. Hyperpolarized 129Xe lung MRI in spontaneously breathing mice with respiratory gated fast imaging and its application to pulmonary functional imaging. NMR Biomed. 2011;24:1343–1352. doi: 10.1002/nbm.1697. [DOI] [PubMed] [Google Scholar]
  • 3.Hoult DI, Richards RE. The signal-to-noise ratio of the nuclear magnetic resonance experiment. J Magn Reson. 1976;24:71–85. doi: 10.1016/j.jmr.2011.09.018. [DOI] [PubMed] [Google Scholar]
  • 4.Golman K, in't Zandt R, Thaning M. Real-time metabolic imaging. Proc Natl Acad Sci U S A. 2006;103:11270–11275. doi: 10.1073/pnas.0601319103. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Bowers CR, Weitekamp DP. Para-Hydrogen and Synthesis Allow Dramatically Enhanced Nuclear Alignment. J Am Chem Soc. 1987;109:5541–5542. [Google Scholar]
  • 6.Walker TG, Happer W. Spin-exchange optical pumping of noble-gas nuclei. Rev Mod Phys. 1997;69:629–642. [Google Scholar]
  • 7.Kurhanewicz J, Vigneron DB, Brindle K, Chekmenev EY, Comment A, Cunningham CH, DeBerardinis RJ, Green GG, Leach MO, Rajan SS, Rizi RR, Ross BD, Warren WS, Malloy CR. Analysis of Cancer Metabolism by Imaging Hyperpolarized Nuclei: Prospects for Translation to Clinical Research Neoplasia. 2011;13:81–97. doi: 10.1593/neo.101102. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Tsai LL, Mair RW, Rosen MS, Patz S, Walsworth RL. An open-access, very-low-field MRI system for posture-dependent He-3 human lung imaging. J Magn Reson. 2008;193:274–285. doi: 10.1016/j.jmr.2008.05.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Myers W, Slichter D, Hatridge M, Busch S, Mössle M, McDermott R, Trabesinger A, Clarke J. Calculated signal-to-noise ratio of MRI detected with SQUIDs and Faraday detectors in fields from 10μT to 1.5T. J Magn Reson. 2007;186:182–192. doi: 10.1016/j.jmr.2007.02.007. [DOI] [PubMed] [Google Scholar]
  • 10.Carias MF, Dominguez-Viqueira W, Santyr GE. Improving Signal-to-Noise Ratio of Hyperpolarized Noble Gas MR Imaging at 73.5 mT Using Multiturn Litz Wire Radiofrequency Receive Coils. Concept Magnetic Res B. 2011;39B:37–42. [Google Scholar]
  • 11.Dominguez-Viqueira W, Berger W, Parra-Robles J, Santyr GE. Litz Wire Radiofrequency Receive Coils for Hyperpolarized Noble Gas MR Imaging of Rodent Lungs at 73.5 mT. Concept Magnetic Res B. 2010;37B:75–85. [Google Scholar]
  • 12.Coffey AM, Shchepin RV, Wilkens K, Waddell KW, Chekmenev EY. A Large Volume Double Channel 1H-X RF Probe for Hyperpolarized Magnetic Resonance at 0.0475 Tesla. J Magn Reson. 2012;220:94–101. doi: 10.1016/j.jmr.2012.04.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Pound RV. Nuclear Spin Relaxation Times in Single Crystals of LiF. Physical Review. 1951;81:156–156. [Google Scholar]
  • 14.Mishkovsky M, Cheng T, Comment A, Gruetter R. Localized in vivo hyperpolarization transfer sequences. Magn Reson Med. 2012;68:349–352. doi: 10.1002/mrm.23231. [DOI] [PubMed] [Google Scholar]
  • 15.Hoult DI, Lauterbur PC. The sensitivity of the zeumatographic experiment involving human samples. J Magn Reson. 1979;34:425–433. [Google Scholar]
  • 16.Gadian DG, Robinson FNH. Radiofrequency losses in NMR experiments on electrically conducting samples. J Magn Reson. 1979;34:449–455. [Google Scholar]
  • 17.Redpath TW, Hutchison JMS. Estimating patient dielectric losses in NMR images. Magn Reson Imaging. 1984;2:295–300. doi: 10.1016/0730-725x(84)90195-4. [DOI] [PubMed] [Google Scholar]
  • 18.Minard KR, Wind RA. Solenoidal microcoil design - Part II: Optimizing winding parameters for maximum signal-to-noise performance. Concept Magnetic Res. 2001;13:190–210. [Google Scholar]
  • 19.Terman FE. Radio Engineers's Handbook. 1943 [Google Scholar]
  • 20.Darrasse L, Ginefri JC. Perspectives with cryogenic RF probes in biomedical MRI. Biochimie. 2003;85:915–937. doi: 10.1016/j.biochi.2003.09.016. [DOI] [PubMed] [Google Scholar]
  • 21.Lotfi AW, Lee FC. A high frequency model for Litz wire for switch-mode magnetics. Industry Applications Society Annual Meeting, Conference Record of the 1993 IEEE; 1993. pp. 1169–1175. [Google Scholar]
  • 22.Medhurst RG. H. F. Resistance and Self-Capacitance of Single-Layer Solenoids. Wireless Engineer. 1947 Feb;Mar;:35–43. 80-92. [Google Scholar]
  • 23.Marrocco BJD, Drost DJ, Prato FS. An optimized head coil design for MR imaging at 0.15-T. Magn Reson Med. 1987;5:143–159. doi: 10.1002/mrm.1910050206. [DOI] [PubMed] [Google Scholar]
  • 24.Bidinosti CP, Chapple EM, Hayden ME. The sphere in a uniform RF field - Revisited. Concept Magnetic Res B. 2007;31B:191–202. [Google Scholar]
  • 25.Gilbert KM, Scholl TJ, Chronik BA. RF coil loading measurements between 1 and 50 MHz to guide field-cycled MRI system design. Concept Magnetic Res B. 2008;33B:177–191. [Google Scholar]
  • 26.Blasiak B, Volotovskyy V, Deng C, Tomanek B. An optimized solenoidal head radiofrequency coil for low-field magnetic resonance imaging. Magn Reson Imaging. 2009;27:1302–1308. doi: 10.1016/j.mri.2009.05.018. [DOI] [PubMed] [Google Scholar]
  • 27.Chekmenev EY, Norton VA, Weitekamp DP, Bhattacharya P. Hyperpolarized 1H NMR Employing Low gamma Nucleus for Spin Polarization Storage. J Am Chem Soc. 2009;131:3164–3165. doi: 10.1021/ja809634u. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Resmer F, Seton HC, Hutchison JMS. Cryogenic receive coil and low noise preamplifier for MRI at 0.01 T. J Magn Reson. 2010;203:57–65. doi: 10.1016/j.jmr.2009.11.021. [DOI] [PubMed] [Google Scholar]
  • 29.Matlashov AN, Schultz LJ, Espy MA, Kraus RH, Savukov IM, Volegov PL, Wurden CJ. SQUIDs vs. Induction Coils for Ultra-Low Field Nuclear Magnetic Resonance: Experimental and Simulation Comparison. IEEE Transactions on Appiled Superconductivity. 2011;21:465–468. doi: 10.1109/TASC.2010.2089402. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Lee VS. MRI: From science to society. J Magn Reson Imaging. 2013;37:753–760. doi: 10.1002/jmri.24044. [DOI] [PubMed] [Google Scholar]
  • 31.Jin J. Electromagnetic analysis and design in magnetic resonance imaging. CRC PressI Llc. 1999 [Google Scholar]
  • 32.Minard KR, Wind RA. Solenoidal micro coil design. Part I: Optimizing RF homogeneity and coil dimensions. Concept Magnetic Res. 2001;13:128–142. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

01

RESOURCES