Abstract
Purpose
To implement and validate a previously proposed ultra-short echo time (UTE) method for measuring collagen-bound and pore water concentrations in bone based on their T2 differences.
Methods
Clinically compatible UTE image sequences for quantitative T2-based bound and pore water imaging in bone were implemented and validated on a 3T human scanner and a 4.7T small bore system. Bound and pore water images were generating using T2-selective adiabatic pulses. In both cases, the magnetization preparation was integrated into a 3D UTE acquisition, with 16 radial spokes acquired per preparation. Images were acquired from human cadaveric femoral mid-shafts, from which isolated bone samples were subsequently extracted for non-imaging analysis using T2 spectroscopic measurements.
Results
A strong correlation was found between imaging-derived concentrations of bound and pore water and those determined from the isolated bone samples.
Conclusion
These studies demonstrate the translation of previously developed approaches for distinguishing bound and pore water from human cortical bone using practical human MRI constraints of gradient performance and RF power deposition.
Keywords: bone, UTE, adiabatic, bound water, pore water
INTRODUCTION
Recent studies showed that 1H NMR signals with short T2 (≈ 400 µs) correspond primarily to collagen-bound water and those with longer T2 components (1 ms − 1 s) correspond primarily to pore water (1–4) These bound and pore water measures correlate to mechanical properties of bone, including yield stress, peak stress, and pre-yield or elastic toughness (5–8). In particular, bones with a greater concentration of bound water and a lower concentration of pore water have higher values of peak stress, yield stress, and pre-yield toughness. These results suggest that appropriate MRI methods that robustly distinguish and quantitatively measure bound- and pore-water concentrations in cortical bone may offer a viable methodology for predicting fracture risk. In particular, they can assess the contribution of the bone tissue to fracture resistance in addition to the structural contribution already provided by conventional MRI or X-ray computed tomography. This is potentially quite useful since clinical assessment of areal bone mineral density by dual-energy X-ray absorptiometry does not necessarily capture all the deleterious effects of aging and certain diseases (type 2 diabetes, chronic kidney disease) on fracture risk (9,10).
Bi-exponential analysis of T2* signal decays has shown correlations between the fitted components amplitudes and the bound and pore water concentrations (7,11–13). This approach requires noise-sensitive non-linear regression and may be limited at high static field strengths by the similarity of T2* of bound and pore water (3,4,14). An alternative approach uses T2-selective adiabatic radiofrequency (RF) pulses over a broad enough resonance bandwidth to effectively distinguish bound- and pore-water signals (14). Specifically, playing two consecutive broad-bandwidth adiabatic full passage pulses will drive short T2 magnetization (bound water) to saturation while rotating long T2 magnetization (pore water) through 360°, leaving it essentially unaffected. This approach is referred to as the Double Adiabatic Full Passage (DAFP). To image bound water, a similar approach uses one adiabatic full passage pulse followed by an appropriate delay to invert and null pore water magnetization while the bound water magnetization experiences a saturation-recovery process. This approach is referred to as the Adiabatic Inversion Recovery (AIR). Presented here are demonstrations and validations of DAFP and AIR methods of imaging bound- and pore-water of human cadaver bones, using clinically practical parameters, on both a 4.7T small-bore and a 3.0T human system.
METHODS
The Vanderbilt Donor Program supplied human femurs from 3 cadaveric donors, two males and one female, mean age 77 years. Mid-shaft sections of each bone were cut to ≈ 80 mm in length. Images of the femur mid-shafts along with a CuSO4-doped 10% H2O/90% D2O phantom (in a 10 mm NMR tube adjacent to the bone) were acquired using the DAFP and AIR sequences, detailed below, with 96 × 96 × 96 mm3 field of view and a nominal isotropic resolution of 1.5 mm. After imaging, cylindrical cortical bone samples (4–9 mm length, 6 mm diameter) were cored from four radial locations near the middle of the mid-shaft. These samples, along with a long-T2 water sample of known volume, were used to provide reference values of bound and pore water concentrations using a previously described CPMG protocol (14) at 4.7 T. Imaging at 3T was performed on a Philips Ingenia (Best, NL) using their dStream Knee 8ch receive coil and the body coil for signal transmission. Measurements at 4.7 T were performed on an Agilent Direct Drive (Santa Clara, CA) using an in-house built 63-mm diameter low-proton birdcage style coil for whole bone imaging and an in-house built 20 mm diameter low-proton loop-gap style RF coil for CPMG measurements on isolated bone specimen (15).
Pulse Sequences
Figure 1 shows sequence diagrams of the DAFP and AIR sequences. In all cases, the following sequence parameters were used: radial sampling of k-space with 83 points at 250 kHz receiver bandwidth, acquisition time per spoke = 332 µs; a post-acquisition spoiler gradient 1.74 ms duration and 31 mT/m amplitude; repetition time per spoke (TRA) = 3.18 ms; number of spokes per shot (NS) = 16; total number of spokes = 8192; RF excitation pulse width = 115 µs. The radial spokes were distributed evenly over the k-space sphere (16). A variable flip angle schedule was used for excitations in order to generate approximately constant transverse magnetization for all 16 spokes (17), with an initial prescribed flip angle, θ1 = 12.5°, and effective total flip angle, θE = 60° (i.e., longitudinal magnetization is reduced by cos(θE) by the combination of all 16 excitations). In all cases, the effective echo time (TE), as measured from the center of the RF excitation pulse to the start of acquisition, was 105.5 µs (4.7T) and 127.5 µs (3.0T).
Figure 1.
The 3D-UTE pulse sequence used. The PREP pulse is a double HS8 pulse for DAFP and a single HS8 pulse for AIR. The time delay between the end of the preparation pulse and the start of data acquisition is TD. The effective inversion-recovery time TI = TD + TRA × NS/2, where NS radial spokes are acquired with period TRA during every TR period.
The AIR sequence used a sequence repetition time (TR) = 300 ms, TI = 90ms/85ms (4.7T/3T), and a 10-ms duration, 3.5 kHz bandwidth, 8th ordered hyperbolic secant (HS8) pulse (18) as the preparation pulse. The DAFP sequence used TR = 400 ms, TD = 5 ms, and two consecutive HS8 pulses (20 ms total duration). The maximum gradient amplitudes and slew rates of the human system were also used on the 4.7T. The TR values for each sequence were dictated by FDA-defined RF power deposition limits on the 3.0T scanner. On the 4.7T system, one excitation provided sufficient signal, resulting in scan times of ≈ 3 ½ min and ≈ 2 ½ min for DAFP and AIR, respectively. On the 3.0T system, lower signal-to-noise ratio (SNR) dictated 4 averaged excitations (≈ 13 ½ m) for DAFP and 6 averaged excitations (≈ 20 ½ m) for AIR to achieve SNR comparable to 4.7T.
In addition, a conventional UTE (CUTE) image was acquired for each bone at 3.0T and 4.7T, and at 4.7T a B1 map was also acquired. The CUTE acquisition used TR/TE = 2.5 ms/62.5 µs and a 25 µs duration, 6° flip excitation pulse. The B1 mapping was performed by the Bloch-Siegert method (19) with a multi-slice spin echo acquisition. Ten axial slices (3 mm thick/5 mm gap) spanned the length of the bone. The B1 measured in the water phantom of each slice was used to determine the actual flip angle seen in each slice for analysis of AIR and DAFP data (see below). Variation of |B1| within the slice was independently determined to be < 2.5% for the coil used on the 4.7T. On the 3.0T, the body RF coil was used for transmission and was independently determined to vary in |B1| by < 4.5% over the entire bone volume, so no B1 mapping was necessary.
Data Analysis
All data were analyzed using MATLAB (Natick, MA). Images were reconstructed using standard trajectory mapping, density compensation, and gridding methods (20). Bound and Pore water concentrations were computed on a voxel-by-voxel basis, as described below, then regions of interest (ROIs) were defined at the approximate locations from which the cylindrical bone samples were taken. The signal equations for DAFP and AIR measurements in cortical bone are shown below (note the correction in Eq [1], compared with a previous report (14)):
[1] |
and
[2] |
where α is the inversion efficiency of the AFP pulse, β is the signal loss due to relaxation-induced blurring, S0 is proportional to water concentration, and superscripts pw and bw indicate pore water and bound water, respectively. Replacing pw or bw with ref, provides the signal equations for the reference marker for each sequence. At 4.7T, previously obtained values were used for inversion efficiency (α), R1, and R*2 of bound water, pore water, and the reference marker (14): αbw/pw/ref = 0.09/−0.78/−0.83, 1/R1bw/pw/ref = 357 ms/551 ms/13 ms, 1/R2* bw/pw/ref = 290 µs/1280 µs/13 ms. At 3.0T, R1pw was estimated from one bone using a saturation-recovery fast spin echo acquisition, and R1bw was estimated to change similarly from 4.7T as did R1pw. R*2 values at 3.0T were used as measured by Du et al. for ex vivo human cortical bone (12). Because R2 values were assumed to be nearly B0 independent, the same α values were used at 3.0T as were previously measured at 4.7T. A summary of parameter values used at 3.0T were αbw/pw/ref = 0.09/−0.78/−0.83, 1/R1bw/pw/ref = 290 ms/450 ms/10 ms, 1/R2* bw/pw/ref = 350 µs/2600 µs/10 ms. The blurring-induced signal loss values were empirically estimated by simulating the effect of blurring using the known bone geometry for each bone (see Appendix for more details). Individual β values were found for each ROI and bone, but mean estimates used to create images were βbw/pw= 0.77/0.97 at 3.0 T and 0.74/0.93 at 4.7 T. In both cases, βref was defined = 1.0. Thus, given the observed bone signals SDAFP and SAIR, the equilibrium signals, S0pw and S0bw were computed from each bone ROI or voxel using Eq [1] and [2]. These relative measures of proton density were then converted into absolute units of mol 1H/Lbone by comparison to corresponding values of S0ref, which were known to reflect 111.1 mol 1H/LH20.
The non-imaging data from the extracted cortical bone samples were analyzed by fitting CPMG echo amplitudes to a broad range of decaying exponential functions by non-negative least squares criteria subject to a minimum curvature constraint, resulting in a T2-spectrum for each sample (3,21). The integrated T2 spectrum amplitude over various domains provided signal amplitude measures for bound water (100 µs < T2 < 1 ms), pore water (1 ms < T2 < 1 s) and reference sample (T2 > 1 s). The bound and pore water signals amplitudes were then converted into units of mol 1H/Lbone by comparison with the reference signal amplitude and known volumes of the bone and reference samples, and the known proton concentration of water, as above.
RESULTS
Figure 2 shows approximately the same slice taken from 3D bound- and pore-water images of one bone at 3T and 4.7T. The gray scale images are CUTE images; color overlaid images are the bound or pore water concentration maps generated from the respective method. The DAFP image shows consistently a higher concentration of pore water in the posterior section of the femur, which agrees with previous findings (6,22), and in general there is an apparent negative correlation between the spatial distribution of bound and pore water, as expected. The signal to noise ratio (SNR) of DAFP/AIR images were ~27/~22 at 4.7T and ~26/~28 at 3.0T, defined as , where µS and µN are mean signal in a region of cortical bone and background noise, respectively. (At 3.0T, the background of the AIR images showed significant signal from the foam used to hold the bone samples, so for this µN measure one scan was repeated with a larger FOV but equal voxel size and receiver bandwidth.) The RF coil used at 3.0T is suitable for wrist and lower leg imaging, so these SNR values should be predictive of in vivo scans of the radius and tibia.
Figure 2.
Imaging results from 3 T (top) and 4.7 T (bottom) of the DAFP and AIR sequences showing three cardinal planes of pore and bound water maps. Note the negative correlation between bound and pore water throughout the bone volume and the higher concentration of pore water in the posterior section of the femur.
Figure 3 shows a representative T2 spectrum from an extracted cortical bone sample, with the bound water, pore water, and water marker signals labeled. Figure 4 shows linear correlations between bound/pore water concentrations measures from the extracted samples and those from the AIR and DAFP images at approximate locations of the extracted bone samples (shown by red squares on inset image). Coefficients of determination for pore water concentrations were r2 = 0.41 at 3T and r2 = 0.94 at 4.7T; for bound water concentrations they were r2 = 0.76 at 3T and r2 = 0.55 at 4.7T.
Figure 3.
A representative T2 spectrum from the CPMG measurements of the cored samples of cortical bone showing signals from bound water, pore water, and the water marker. The amount of bound and pore water was converted into units of mol 1H/L based on the known size and concentration of the water marker.
Figure 4.
Concentrations from CPMG measurements versus DAFP and AIR results at approximate ROI locations from 3 T and 4.7T images of a) bound water and b) pore water. Both 3 T and 4.7 T imaging measurements showed strong linear correlations with CPMG measurements.
DISCUSSION
The magnetization preparations used in the AIR and DAFP pulse sequences were previously demonstrated to effectively distinguish bound and pore water signals in isolated human cortical bone samples (14). Presented here is the translation of these methods into clinically practical MRI protocols, and the quantitative evaluation of these MRI protocols on human cadaver bones at 3.0 and 4.7 T. The results suggest that the AIR and DAFP methods are effective for quantitative imaging of bound and pore water, respectively, but there are numerous factors that may affect their performance and utility.
First, in contrast to previous non-localized studies of isolated bone samples, the imaging protocols presented here required accelerated acquisition to achieve scan times amenable to human studies. Power deposition from the AFP pulses set the lower limit on TR, so additional acceleration was achieved by acquiring NS =16 radial spokes in k-space per TR period, similar to a conventional MP-RAGE protocol (23). The 3D radial trajectory sampled the k-space origin with every radial spoke, so accurate quantitation of image intensity required a variable excitation flip angle schedule that generated approximately the same amplitude of transverse magnetization for each spoke (17). Increasing NS requires reduced flip angles, resulting in , so the choice of NS depends on the SNR and scan time requirements. The choice of NS may also affect signal accuracy, because each spoke experiences slightly different magnetization preparation. For the AIR sequence, each of the NS spokes is acquired at a different TI and, therefore, includes a varying amount of non-nulled pore water signal. As long as the net pore water signal across the NS spokes is zero, this is not a problem, but increasing NS will likely result in greater net pore water signal. For the DAFP sequence, the recovery of bound water magnetization will grow with each spoke, so NS × TRA should be kept small compared to T1 of bound water.
In addition accelerated acquisition, practical use of the AIR and DAFP protocols depends upon having good estimates of a number of parameters in the signal equations, Eq [1] and [2]. As done here for scans on the 4.7T, it is relatively quick and easy to map B1, thereby providing good estimates of θ1 and θE on a case-by-case basis. However, estimates of bound- and pore-water relaxation rates may not be readily acquired during a clinical protocol, so good population estimates are needed. The values used here and in a previous study (14) of a small sample of cadaver bones have been sufficient to demonstrate efficacy of the AIR and DAFP methods, but it is likely that errors in these values underlie the systematic deviations between the imaging and CPMG measures seen in Fig 4. Given the parameters used in this work, an error of 10% in R1 results in a 5/8% error of DAFP/AIR signal, while a 10% error in R2* gives a 1/4% error of DAFP/AIR signal. Further, it may not be suitable to describe R1 and R2* with scalar values. In particular, pore water likely consists of a relatively broad spectrum of T1 values due to the variation in pore sizes within the bone (2,3), which likely explains why TI must be empirically set to null the net pore water magnetization rather that by calculation from the estimated R1pw (14).
Two parameters that are known but require special attention for accurate AIR and DAFP measures are TE and receiver bandwidth. Although it is common to define TE from the end of the RF excitation pulse, the effect of relaxation during the RF pulse must be incorporated to ensure accurate measures. For hard pulse 3D UTE, as used here, transverse relaxation can be effectively accounted for by measuring TE from the middle of the RF pulse rather than the end (24). Accounting for transverse relaxation during the acquisition is a somewhat more complicated problem. Because the bound water T2* is similar to the acquisition duration (332 µs), its relatively broad point spread function results in an underestimation of bound water signal compared to signal from the long T2 water reference (25, 26). In the present studies, as noted in the METHODS, the bound and pore water signal losses was empirically estimated, which resulted in the βbw/pw/ref = 0.77/0.97/1.0 at 3.0 T and 0.74/0.93/1.0 at 4.7 T. See the Appendix for calculation details.
CONCLUSION
These studies demonstrate the translation of previously developed approaches for distinguishing bound and pore water from human cortical bone. The methods, referred to AIR and DAFP here, were implemented as part of 3D UTE pulse sequences, subject to the practical human MRI constraints of gradient performance and RF power deposition. The results showed good correlation between these imaging measures of bound and pore water and those determined by previously established non-localized CPMG measures.
ACKNOWLEDGEMENTS
This work was supported by NIH grant EB001744.
Grant Support: NIH EB014308
APPENDIX
Signal Amplitude Correction for Blurring
Because the relaxation times of bound and pore water are similar to the acquisition duration, it is necessary to account for the effect transverse relaxation during the acquisition on image signal amplitude. Relaxation during acquisition broadens the image-domain point spread function, which can blur a significant amount of signal out of voxel or ROI, and ramp sampling exacerbates the problem because the signal decays more rapidly in k-space. For the 3D bone imaging in this paper, the samples are roughly invariant in the direction of the long axis of the bone, so the blurring effect can be neglected in that direction. Also, because the k-space sampling is radial, the point spread function can be solved in 1D (25,26), then applied in the 2D plane corresponding to the axial view of the bone (Fig 2).
For a known image geometry, T2*(s), and k-space trajectory, the signal attenuation can be numerically estimated as follows: i) a masked 2D bone image, s(r) (bone signal = 1, all other signal equals 0) is Fourier transformed to produce the k-space signal, S(k); ii), the effect of T2* decay during acquisition is imparted by multiplying S(k) by H(k), derived below; iii) the resulting apodized signal is inverse Fourier transformed to produce a blurred image, sb(r); and iv) the signal loss term, β, is then computed on a voxel-by-voxel basis as β = sb(r)/s(r).
The apodizing function, H(k), is derived for a 2D radial acquisition as follows. The signal decay during acquisition as a function of time is
[A1] |
and k is a function of t by the relationship:
[A2] |
where γ is the gyromagnetic ratio and g(t) is the gradient waveform. Assume that g(t) increases linearly at constant slew rate up to max gradient amplitude, G, at time t = t0, then,
[A3] |
Let , then
[A4] |
Substituting [A4] into [A1] give the apodizing function in k-space
[A5] |
REFERENCES
- 1.Fantazzini P, Brown RJS, Borgia GC. Bone tissue and porous media: common features and differences studied by NMR relaxation. Magn Reson Img. 2003;21(3–4):227–234. doi: 10.1016/s0730-725x(03)00129-2. [DOI] [PubMed] [Google Scholar]
- 2.Ni Q, Nyman JS, Wang X, Santos ADL, Nicolella DP. Assessment of water distribution changes in human cortical bone by nuclear magnetic resonance. Meas Science Technol. 2007;18(3):715–723. [Google Scholar]
- 3.Horch RA, Nyman JS, Gochberg DF, Dortch RD, Does MD. Characterization of 1H NMR signal in human cortical bone for magnetic resonance imaging. Magn Reson Med. 2010;64(3):680–687. doi: 10.1002/mrm.22459. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Ong HH, Wright AC, Wehrli FW. Deuterium nuclear magnetic resonance unambiguously quantifies pore and collagen-bound water in cortical bone. J Bone Miner Res. 2012;27(12):2573–2581. doi: 10.1002/jbmr.1709. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Nyman JS, Ni Q, Nicolella DP, Wang X. Measurements of mobile and bound water by nuclear magnetic resonance correlate with mechanical properties of bone. Bone. 2008;42(1):193–199. doi: 10.1016/j.bone.2007.09.049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Fernández-Seara Ma, Wehrli SL, Takahashi M, Wehrli FW. Water content measured by proton-deuteron exchange NMR predicts bone mineral density and mechanical properties. J Bone Miner Res. 2004;19(2):289–296. doi: 10.1359/JBMR.0301227. [DOI] [PubMed] [Google Scholar]
- 7.Bae WC, Chen PC, Chung CB, Masuda K, D’Lima D, Du J. Quantitative ultrashort echo time (UTE) MRI of human cortical bone: correlation with porosity and biomechanical properties. J Bone Miner Res. 2012;27(4):848–857. doi: 10.1002/jbmr.1535. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Horch RA, Gochberg DF, Nyman JS, Does MD. Non-invasive predictors of human cortical bone mechanical properties: T(2)-discriminated H NMR compared with high resolution X-ray. PloS one. 2011;6(1):e16359. doi: 10.1371/journal.pone.0016359. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kanis JA, Johnell O, Oden A, Dawson A, De Laet C, Jonsson B. Ten Year Probabilities of Osteoporotic Fractures According to BMD and Diagnostic Thresholds. Osteoporosis Int. 2001;12(12):989–995. doi: 10.1007/s001980170006. [DOI] [PubMed] [Google Scholar]
- 10.Vestergaard P. Discrepancies in bone mineral density and fracture risk in patients with type 1 and type 2 diabetes - a meta-analysis. Osteopros Int. 2007;18(4):427–444. doi: 10.1007/s00198-006-0253-4. [DOI] [PubMed] [Google Scholar]
- 11.Diaz E, Chung CB, Bae WC, Statum S, Znamirowski R, Bydder GM, Du J. Ultrashort echo time spectroscopic imaging (UTESI): an efficient method for quantifying bound and free water. NMR Biomed. 2012;25(1):161–168. doi: 10.1002/nbm.1728. [DOI] [PubMed] [Google Scholar]
- 12.Du J, Hermida JC, Diaz E, Corbeil J, Znamirowski R, D’Lima DD, Bydder GM. Assessment of cortical bone with clinical and ultrashort echo time sequences. Magn Reson Med. 2012 doi: 10.1002/mrm.24497. in press available online, doi:10.1002/mrm.24497. [DOI] [PubMed] [Google Scholar]
- 13.Du J, Diaz E, Carl M, Bae W, Chung CB, Bydder GM. Ultrashort echo time imaging with bicomponent analysis. Magn Reson Med. 2012;67(3):645–649. doi: 10.1002/mrm.23047. [DOI] [PubMed] [Google Scholar]
- 14.Horch RA, Gochberg DF, Nyman JS, Does MD. Clinically compatible MRI strategies for discriminating bound and pore water in cortical bone. Magn Reson Med. 2012;68(6):1774–1784. doi: 10.1002/mrm.24186. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Horch RA, Wilkens K, Gochberg DF, Does MD. RF coil considerations for short-T2 MRI. Magn Reson Med. 2010;64(6):1652–1657. doi: 10.1002/mrm.22558. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Wong STS, Roos MS. A strategy for sampling on a sphere applied to 3D selective RF pulse design. Magn Reson Med. 1994;32(6):778–784. doi: 10.1002/mrm.1910320614. [DOI] [PubMed] [Google Scholar]
- 17.Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic Resonance Imaging - Physical Principles and Sequence Design. 1999:463–465. [Google Scholar]
- 18.Garwood M. Improved Performance of Frequency-Swept Pulses Using Offset-Independent Adiabaticity. J Magn Reson. 1996;120:133–137. [Google Scholar]
- 19.Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by Bloch-Siegert shift. Magn Reson Med. 2010;63(5):1315–1322. doi: 10.1002/mrm.22357. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Zwart NR, Johnson KO, Pipe JG. Efficient sample density estimation by combining gridding and an optimized kernel. Magn Reson Med. 2012;67(3):701–710. doi: 10.1002/mrm.23041. [DOI] [PubMed] [Google Scholar]
- 21.Whittall KP, MacKay AL. Quantitative interpretation of NMR relaxation data. J Magn Reson. 1989;84(1):134–152. [Google Scholar]
- 22.Currey JD, Brear K, Zioupos P. The Effects of Ageing and Changes in Mineral Content in Degrading the Toughness of Human Femora. J Biomech. 1996;29(2):257–260. doi: 10.1016/0021-9290(95)00048-8. [DOI] [PubMed] [Google Scholar]
- 23.Mugler JP, Brookeman JR. Three-dimensional magnetization-prepared rapid gradient-echo imaging (3D MP RAGE) Magn Reson Med. 1990;15(1):152–157. doi: 10.1002/mrm.1910150117. [DOI] [PubMed] [Google Scholar]
- 24.Robson MD, Gatehouse PD. Consequences of T2 relaxation during half-pulse slice selection for ultrashort TE imaging. Magn Reson Med. 2010;64(2):610–615. doi: 10.1002/mrm.22323. [DOI] [PubMed] [Google Scholar]
- 25.Rahmer J, Bornert P, Groen J, Bos C. Three-Dimensional Radial Ultrashort Echo-Time Imaging with T2 Adapted Sampling. Magn Reson Med. 2006;55(5):1075–1082. doi: 10.1002/mrm.20868. [DOI] [PubMed] [Google Scholar]
- 26.Carl M, Chiang J-TA. Investigations of the origin of phase differences seen with ultrashort TE imaging of short T2 meniscal tissue. Magn Reson Med. 2012;67(4):991–1003. doi: 10.1002/mrm.23075. [DOI] [PubMed] [Google Scholar]