Skip to main content
. Author manuscript; available in PMC: 2014 Dec 24.
Published in final edited form as: Circulation. 2013 Dec 24;128(25):2799–2809. doi: 10.1161/CIRCULATIONAHA.113.005697

Table 2.

Sources of EMI and their potential effects

Type of field Environment Source Potential effects Warranted Precautions
Damage to
CIEDs Circuitry
Pacemaker
inhibition
Noise
Tracking
Asynchronous
pacing
Noise
reversion
Communication
failure
Current
induction
Arrhythmia
Detection/ICD
shocks
Inhibition of
Tachyarrhythmia
Detection
Sensing and Pacing
Threshold Changes
Power-on-reset Rate Adaptive
Sensor Activation
Electromagnetic Daily life Digital music players9,10 Avoid close proximity (<30 cm) with generator.
Stereo Speaker and headphones11 (magnetic field strength ≥10 gauss) Avoid close proximity with generator.
Neodymium-iron-boron magnets incorporated into toys, jewelry, and clothes12,13 Avoid close proximity with generator.
Metal detectors14, 15 Alert security personnel regarding the implanted device.
Electronic Article Surveillance Devices1619 Patients should not linger near surveillance devices to minimize the duration of EMI that might be detected by the CIEDs.
Cellular Phones2026 Avoid close proximity with generator.
Hand held home phones27 Avoid close proximity with generator.
Tasers2830
Household leakage current from improper grounding3134 Avoid contact with improperly grounded equipment.
Automotive engine11 Avoid close proximity of ignition system, engine, and alternator with generator.
Work and industry Arc welding equipment, electric motors, and degaussing coils35 Advise employer regarding implanted medical device and consider environmental testing by a trained EMI consultant (list of consultants can be provided by the CIEDs manufacturer). Maintain a minimal distance of 2 ft (61 cm) from the welding arc and cables or large motors. Do not exceed tested currents with the welding equipment, wear insulated gloves while operating electrical equipment, and verify that electrical equipment is properly grounded.
Aircraft cockpits36
RF ID equipment37
High voltage power coils38,39 Assessment of the threshold safe distance for EMI and follow up device interrogations should be arranged.
Medical MRI40 MRI should be performed at experienced centers utilizing a strict protocol for patient monitoring and CIEDs programming (refer to Table 3).
Electrosurgery41, 42 No intervention or magnet application is needed for sub-umbilical procedures. For procedures above the umbilicus, programming or magnet application should be performed to activate asynchronous pacing for pacemaker dependent patients as well as deactivation of tachyarrhythmia therapies for ICD recipients. The ground electrode to direct the current path away from the CIEDs.
LVAD4345
Cardioversion4648 CIEDs should be programed to a bipolar sensing configuration and should be excluded from the current path by use of an antero-posterior patches/paddles position, placed >6 inches away from the generator.48 Immediate access to a CIEDs programmer is recommended.
Dental equipment4951
Radiofrequency catheter ablation5255 Deactivate tachyarrhythmia therapies and activate asynchronous pacing for pacemaker dependent patients.
Spinal cord56, 57 and deep brain stimulators58, 59 Program bipolar sensing configuration. Test for interactions intraoperatively and when modifying programming.
Gastroscopy, endoscopy, and colonoscopy60 No anticipated effects in the absence of electrosurgery.
Capsule endoscopy6164 Tachyarrhythmia therapies should be turned off prior to procedure
Transcutaneous electrical nerve stimulation:6567 Magnet application or programming to an asynchronous mode (VOO/DOO) is recommended in pacemaker dependent patients. ICD therapies should be disabled during the treatment.
Telemetry monitors68 Consider deactivation of minute ventilation sensors.
Electroconvulsive therapy69, 70 Availability of a magnet or CIEDs programming equipment is advised.
Ionizing Radiation Radiotherapy7175 CIEDs should be interrogated within 24 hours of each treatment when using a direct beam to the chest or high-energy photon radiation for pacemaker dependent patients.43
Programming the CIEDs to a higher pacing rate than baseline allows rapid recognition of power-on-reset or device malfunction when onsite interrogation is unavailable.
Acoustic Radiation Lithotripsy7679 Patients should be continuously monitored. Also, synchronization of the hydraulic shocks to the R wave should be performed. Dual chamber pacemakers should be reprogrammed to VVI mode. ICD tachyarrhythmia therapies should be deactivated during treatment. CIEDs should be interrogated post procedure.