A New Approach for Scatter Removal and Attenuation Compensation from SPECT/CT Images

Shabnam Oloomi 1, Hadi Noori Eskandari 2, Seyed Rasoul Zakavi 3, Peter Knoll 4, Faraz Kalantari 5, Mohsen Hajizadeh Saffar 1,*
PMCID: PMC3909630  PMID: 24494071

Abstract

Objective(s): In SPECT, the sinogram contains scatter and lack of attenuated counts that degrade the reconstructed image quality and quantity. Many techniques for attenuation and scatter correction have been proposed. An acceptable method of correction is to incorporate effects into an iterative statistical reconstruction. Here, we propose new Maximum Likelihood Expectation Maximization (MLEM) formula to correct scattering and attenuating photons during reconstruction.

Materials and Methods: In this work, scatters are estimated through Klein-Nishina formula in all iterations and CT images are used for accurate attenuation correction. Reconstructed images resulted from different MLEM reconstruction formula have been compared considering profile agreement, contrast, mean square error, signal-to-noise ratio, contrast-to-noise ratio and computational time.

Results: The proposed formula has a good profile agreement, increased contrast, signal-to-noise (SNR) & contrast-to-noise ratio (CNR), computational time and decreased mean square error (MSE) compared with uncorrected images and/or images from conventional formula.

Conclusion: In conclusion, by applying the proposed formula we were able to correct attenuation and scatter via MLEM and improve the image quality, which is a necessary step for both qualitative and quantitative SPECT images.

Key Words: Attenuation correction, MLEM, Scatter correction, SPECT

Introduction

Single photon emission computed tomography (SPECT) is an imaging modality used to visualize the biological uptake and distribution of an applied radiopharmaceutical. The physical effects of attenuated and scattered photons have to be taken into account to improve the image quality. For non-uniform attenuators (for example in the human thorax), the generation of a patient attenuation map is necessary. Several attenuation correction methods have been reported and are used in clinical studies. Transmission computed tomography with an external gamma-ray source has been widely used in nuclear medicine for cardiac SPECT studies but is limited only to 99mTc or 201Tl studies. The images obtained by this method can be used only for attenuation correction but not for anatomical orientation. In recent years, multi-modality imaging using techniques from two different modalities (PET or SPECT and x-ray CT) was developed. Attenuation coefficient maps generated from X-ray CT images have several advantages such as shorter acquisition time and improved image quality obtained by even low-dose CT scanning protocols which also enables the anatomic orientation.

Over the last two decades, intensive efforts have been made to compensate for the scatter effect in SPECT in order to improve the quantitative and qualitative accuracy of the reconstructed images (1,2) . A class of widely used scatter compensation methods is based on the estimation of the scatter component in the photopeak projection data and subsequent subtraction or deconvolution of the scatter contribution from the measured projection data. Scatter compensation methods in this class are fast and simple, but increase the noise in the reconstructed images. Another class which is a promising approach for scatter compensation, consists of reconstruction-based scatter compensation methods (RBSC) (3) resulting in images with both less bias and reduced variance as compared with subtraction-based scatter compensation methods (4-6).

RBSC methods are based on modeling the scattered photons in projection–back projection processes. Several techniques have been developed for calculating the scatter, one of them which was used in this study is based upon the integration of the Klein–Nishina formula in non-uniform media (3, 7, 8).

In this work, we introduce new approach for both attenuation and scatter correction during reconstruction using the MLEM approach. To obtain a proposed iterative formula for reconstruction of the SPECT images, we utilize the necessary conditions to optimally maximize the likelihood function. Then, its importance is evaluated in inhomogeneous media of digital and experimental phantoms. Finally, the contrast, mean square error (MSE), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and computational time of the newly developed algorithm were compared with the conventional MLEM ones.

Methods and Materials

Proposed MLEM iterative formula

The projections acquired in different angles around the object of interest can be used to reconstruct trans-axial slice images through analytical or iterative methods containing algebraic and statistical methods. Most commonly used iterative reconstruction method is maximum likelihood expectation maximization (MLEM) or ordered subset expectation maximization (OSEM), a faster implementation of MLEM algorithm (9-11). In this section, a proposed MLEM iterative formula for reconstruction of the SPECT image is described. Considering the following assumptions:

n for Number of detection bin, equal to detector pixels × projection number

m for Number of image pixels, K for number of iterations

fi j=1,2,…,m are the pixel values of the image, proportional to the number of radionuclide activity in pixel k

g i i=1,2,…,m are the sinogram measured data from ith detection bin in gamma camera

a ij i=1,2,...,n and j =1,2,...,m are the elements of the system matrix or detection probability of emitted photons from aij i=1,2,…,n pixel j of the subject to be detected in ith detection bin in gamma camera.

aijμ=aije-kjrikμk are the attenuated system matrix elements,

r ik and µk for i=1,2,...,n and kԑj=1,2,...m are the length and attenuation coefficient of pixels of number k, which are along the direction of pixel j to detection bin i

SCi i=1,2,…,n are the appeared scatter photons in ith detector of gamma camera.

As gi contains primary and scattered photons, so in each detector we have:

Egi=(AF)i+SCi Equation 1

Where E(gi) is the mathematical expectancy of Poisson variable gi. The variable gi have a possibility function such as:

Pgi,SCi+k=1maikμfk=SCi+kj=1maikμfkgigiexp(-SCi-kj=1maikμfk Equation 2

Here, we estimate the values of for k=1,2,...,m using the logarithm of maximum likelihood (LML) function, that is:

Lf=i=1ngiLnSCi+kj=1maikμfk-SCi-kj=1maikμfk-Lngi Equation 3

Now, we obtain such that this vector maximizes the LML function. The Vector satisfies Equation 4 which is a necessary condition to maximize the LML function:

Lfj=giβij+aijμSCi+kj=1maikμfk-βij-aijμ=0 Equation 4

j=1,2,, m

Applying in equation 4, we can obtain the following MLEM iterative formula for reconstruction of the SPECT images:

fjK+1=fjKi=1naijμi=1n(giβij+aijμSCi+kj=1maikμfkK-βij) Equation 5

 j=1,2,,mK=0, 1, 2,  j=1,2,,m K=0,1,2,

Where , j=1,2,...,m is the initial guess for the activity of pixels of SPECT image. It should be note that if SCi=0 then =0 and the proposed MLEM iterative formula is converted to the MLEM formula.

fjK+1=fjKi=1naijμi=1n(giaijμkj=1maikμfkK) Equation 6

Phantom studies

A digital NCAT and NEMA image quality phantoms were used in this work.

A) Digital phantom

The NURBS-based cardiac-torso (NCAT) phantom was used to assess the performance of our suggested formula in realistic conditions (12). It can produces attenuation and an activity distribution map of the organs with user defined data (Figure 1). In activity distribution map, 99mTc activity ratio was 100, 100, 40, 22, 6 and 6 in myocardium, gallbladder, liver, spleen, lung and background respectively (13). The attenuation map was generated for photon energy of 140 keV. The phantom dimension was 40×40×20 cm3 that was digitized into 128×128×64 voxels. The Monte Carlo simulation program was based on a published paper by Kalantari et al (11).

Figure 1.

Figure 1

Three trans-axial slices of NCAT phantom; Activity distribution (top row) and their corresponding attenuation map (bottom row)

B) Experimental phantom: image quality phantom

A GE‘s Infineon Hawkeye SPECT/CT scanner was used to acquire row projection data from the NEMA image quality phantom (Figure 2). The phantom has a roughly elliptical shape and contains six fillable spheres of varying sizes with inner diameters of 10, 13, 17, 22, 28 and 37 mm (14). The background and 4 hot spheres of the phantom (inner diameters of 10, 13, 17, 22 mm) were filled with 99mTc, with an activity/concentration ratio of 8:1, the cold spheres activity was zero. So, the activity map of this phantom contains uniform background, 4 hot spheres, 2 cold spheres, and an absorber in the center. Projection data of the phantom were measured with 64×64 pixels, from 0-360° with 6 degree increments.

Figure 2.

Figure 2

An image quality phantom (up); attenuation map (middle) and its activity distribution (down

Attenuation and scatter correction

The CT based attenuation correction was performed in this work for the NEMA image quality phantom study (15-17), using bilinear method energy mapping (18-20).

Correction for attenuation in NCAT phantom was based on the attenuation map, generated during simulation.

Scatter distribution of a specified pixel element j from neighbor pixels, along a particular ray-of-view, bin i, for any projection was calculated by Klein-Nishina formula (21).

Image reconstruction methods

In this work, we applied five different image reconstruction algorithms based on the Maximum Likelihood Expectation Maximization (MLEM) algorithm:

  • Reconstruction of row projections data via MLEM without any correction. They are named "Reference" when reconstructing projections consist of only primary photons and "Un_Cor" when reconstructing projections consist of primary and scatter photons.

  • Reconstruction with attenuation correction of row data (Equation 6), using specific attenuation map and producing aijμ as attenuated system matrix elements. They are named "Att_Cor".

  • Reconstruction with attenuation and scatter correction of row data via conventional formula (Equation 7) and named "Conventional" (22).

fjK+1=fjKi=1naiji=1n(aijgikj=1maikfkK+SCi), Equation 7

i=1,2,, n j=1,2,, m           

Where SCi is the estimated scatter projection. In this study, SCj was updated during the iterations.

  • Reconstruction with attenuation and scatter correction of row data via our proposed MLEM iterative formula with inter-slice scatter estimation (New1, Equation 5).

  • Reconstruction with attenuation and scatter correction of row data via our proposed MLEM iterative formula using 3 slices scatter estimation (New3, Equation 5).

For both, the NCAT simulation and the image quality phantom measurement a 360° SPECT acquisition with 60 different projection angles were used for reconstruction.

Evaluation parameters

Six different parameters were applied to compare the different images in this work. They are profile agreement, contrast, mean square error (MSE), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and computational time used for image reconstruction.

Profile agreement

Horizontal profiles will show the activity distribution and are used to evaluate the agreement between the proposed MLEM method-corrected slice images (NEW1 and NEW3) and the reference image in NCAT phantom.

Furthermore, horizontal profiles are used to evaluate the agreement between the corrected slice images of proposed MLEM method (NEW1 and NEW3) and conventional MLEM formula in both NCAT and image quality phantoms.

Contrast

The contrast was calculated as described by Wieczorek (9) with the following formula:

Ctr=N2n2N1n1-1 Equation 8

Where N2 and N1 are the sum of pixel values in the particular hot and background region, respectively, and ni is the number of pixel elements.

Figure 3.

Figure 3

Region of interest in different slices, the regions of interest in the first row were used to calculate the signal and the regions of interest in the second row were used to calculate noise

Mean squared error (MSE)

To evaluate the similarity between each reconstructed image (P) and reference image (T), MSE was defined as the average of the square differences(23).

MSE=1nxnyx=1nxy=1ny(Px,y-Tx,y2 Equation 9

Where nx and ny are the number of image pixels in each row and column, respectively.

Signal-to-noise ratio (SNR)

Here, the SNR is defined as the ratio of mean signal to standard deviation of the background. Different region of interest (ROI) were selected along the heart wall, left part of the heart, liver and the biggest hot sphere in 26th, 33th, 44th slices of NCAT phantom and in the image quality phantom, respectively. Another ROI in the background was selected for all slices. These ROIs were used for different quantitative measurements of this study.

Contrast-to-Noise Ratio

The contrast-to-noise ratio (CNR) is defined as follows (24).

CNR=Contrast Recovery CoecientCoefficient Of Variation Equation 10

Where Contrast Recovery Coefficient=ml-mbmbC-1

and ml and mb are the mean lesion (hot) and background activity, C is the real contrast in the phantom and coefficient of variation is defined as the ratio of the standard deviation to the mean.

Computational time

The reconstruction time per iteration for different methods was determined using the MATLAB software package.

Results

Activity distribution and profile agreement

Figure 4, from top to bottom, shows tomographic images of the 26th, 33th and 44th slice of NCAT and one slice of image quality phantom, respectively.

Figure 4.

Figure 4

Images resulted from different reconstruction methods: a)Reference, b)Un_Cor, c)Att_Cor, d) Conventional, e) New1, f) New3

Reconstructed images resulted from different MLEM reconstruction methods have been shown in different columns of Figure 4. The images from left to right are reference (for simulated slices only), Un_Cor, Att_Cor, Conventional, New1, and New3.

The horizontal profiles passing through 26th row of each slice image (Figure 5) show a good agreement between profiles of the corrected images resulted from the new MLEM formula and the results from conventional MLEM. The misplacement of events has decreased significantly by scatter correction and the contrast enhanced according to Table 1. Figure 5 and Table 1 show that the image reconstructed without scatter compensation had more over-estimation of counts and loss of contrast due to the presence of scatter in comparison with Un_Cor and Att_Cor.

Figure 5.

Figure 5

Horizontal profile of 26th, 33th and 44th slices of NCAT and image quality phantoms

Table 1.

Contrast of images resulted from different reconstruction methods

Contrast Reference No_cor Att_cor Conventional New1 slice New3 slices
Image quality phantom ---------- 2.8 4.0 3.9 5.1 5.5
NCAT 26 6.8 5.6 6.1 6.0 6.3 6.2
NCAT 33 6.1 4.9 5.6 5.5 5.9 6.0
NCAT 44 8.1 6.7 6.3 6.3 6.8 6.6

Mean squared error (MSE)

MSE between the reference image and other reconstructed images shows the same behavior in all slices (Figure 6).

Figure 6.

Figure 6

Mean square error of the images against different reconstructed methods

The lower the value of the MSE the better is the similarity of the reconstructed image with the reference image. According to this, the differences between the reference and Un_Cor images are the greatest and also the differences between the reference and the scatter and attenuated corrected images are the smallest.

Signal_to_noise (SNR) and contrast-to-noise ratio (CNR)

Signal_to_noise ratio of NCAT phantom based on simulated data in Figure 7 and contrast-to- noise ratio of NCAT and image quality phantom in Figure 8 show that our new approaches (New1 and New3) have comparable accuracy with the conventional formula.

Figure 7.

Figure 7

Signal-to-noise ratio of the images against different reconstructed methods

Figure 8.

Figure 8

Contrast-to-noise ratio of the images against different reconstructed methods

CNR of Un_Cor, Att_Cor, Conventional, New 1 and New 3 are 127.00, 206.65, 222.28, 208.92 and 201.82 respectively.

Computation time

Computation time required for reconstruction of one slice image (64×64 pixels) from a set of 64 projections in the last iteration is 0.029, 0.118 and 0.123 s for Conventional, New1 and New3 respectively. This means that using new formula, the computation time increases by a factor of 4, but it is still in a reasonable time for clinical purpose.

Discussion

In this study, we introduce a novel reconstruction formula implementing attenuation and scatter correction of row data with inter-slice scatter estimation (namely New 1 and New 3, Equation 5).

The new algorithm was tested by a simulation of the NCAT phantom and the SPECT acquisition of the NEMA image quality phantom.

As follows, six different parameters were measured in the final images to quantify and compare the results of the different reconstruction algorithms.

  • Profile agreement

  • Contrast

  • Mean Square Error (MSE)

  • Signal-to-Noise ratio (SNR)

  • Contrast-to-Noise ratio (CNR)

  • Computational time necessary for image reconstruction

The applicability of the presented algorithm is shown by well agreement of horizontal profiles (Figure 5).

The scatter and attenuation compensated images show a better contrast (with the mean of 25% increase) than the uncorrected images allowing a better delineation of the lesions in the scatter and attenuation-compensated images. This is in well agreement with the trials demonstrated that scatter and attenuation can increase the contrast in SPECT studies (11). Also, in comparison with the reference slice, the images resulted from the presented new method show a slightly better contrast.

The MSE values were reduced by 52.9% to 94.3% in the corrected images as compared with the reference images (Figure 6) which is in agreement with other studies (11, 25). MSE was reduced from 0.4 in the Un_Cor to 0.09, 0.04, 0.05 and 0.02 in the Att_Cor, Conventional, New1 and New3, respectively as compared with the reference images (in the slice 44). The same behavior is observed in other slices. The obtained results also point to the importance of scatter and attenuation correction together during reconstruction of images, compared with attenuation correction only without scatter correction. Signal-to-noise study, based on the simulated data, shows that all the scatter and attenuation corrected images have up to 67.1% higher SNR in comparison with the Att_Cor slice images (Figure 7).

Contrast_to_noise (CNR) study also shows that all the scatter and attenuation corrected images have up to 52.5% higher CNR in comparison with Att_Cor images (Figure 8). This is in agreement with several trials demonstrated that scatter correction (26, 27), attenuation correction (11) and both correction applied together (28, 29) can significantly increase the CNR. Due to more accurate scatter estimation in New3, the resulting images show the best CNR among other corrected images.

Conclusion

The proposed formula for incorporating scatter and attenuation during MLEM, enables to remove scatter and compensate attenuation as a necessary step for quantitative SPECT images. The new mathematical method presented in this study increases the contrast, SNR, and CNR of the images and decreases the MSE in comparison with Un_Cor and Att_Cor images.

Acknowledgment

This work is a part of Ph.D. thesis done in Mashhad University of Medical Sciences, Mashhad, Iran. We would like to thank research vice-president of Mashhad University of Medical Sciences and medical physics research center for the financial support.

References

  • 1.Buvat I, Rodriguez-Villafuerte M, Todd-Pokropek A, Benali H, Di Paola R. Comparative assessment of nine scatter correction methods based on spectral analysis using Monte Carlo simulations. J Nucl Med . 1995;36:1476–1488. [PubMed] [Google Scholar]
  • 2.Zaidi H. Scatter modelling and correction strategies in fully 3-D PET. Nucl Med Commun . 2001;22:1181. doi: 10.1097/00006231-200111000-00003. [DOI] [PubMed] [Google Scholar]
  • 3.Bai C, Zeng GL, Gullberg GT. A slice-by-slice blurring model and kernel evaluation using the Klein-Nishina formula for 3D scatter compensation in parallel and converging beam SPECT. Phys Med Biol . 2000;45:1275. doi: 10.1088/0031-9155/45/5/314. [DOI] [PubMed] [Google Scholar]
  • 4.Kadrmas DJ, Frey EC, Karimi SS, Tsui BMW. Fast implementations of reconstruction-based scatter compensation in fully 3D SPECT image reconstruction. Phys Med Biol . 1999;43:857. doi: 10.1088/0031-9155/43/4/014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Kulkarni S, Khurd P, Zhou L, Gindi G. Rapid optimization of SPECT scatter correction using model LROC observers. IEEE Nucl Sci Symp Conf Rec . 2007;5:3986–3993. doi: 10.1109/NSSMIC.2007.4436989. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.King MA, Glick SJ, Pretorius PH, Wells RG, Gifford HC, Narayanan M, et al. Amsterdam, the Netherlands: Elsevier Academic Press; 2004. Attenuation, scatter, and spatial resolution compensation in SPECT; pp. 74–89. In: Wernick MN, Aarsvold JN, editors. Emission Tomography: The Fundamentals of SPECT and PET. [Google Scholar]
  • 7.Jonsson C, Larsson SA. A spatially varying Compton scatter correction for SPECT utilizing the integral Klein-Nishina cross section. Phys Med Biol . 2001;46:1767. doi: 10.1088/0031-9155/46/7/303. [DOI] [PubMed] [Google Scholar]
  • 8.Floyd C, Jaszczak R, Harris C, Coleman R. Energy and spatial distribution of multiple order Compton scatter in SPECT: a Monte Carlo investigation. Phys Med Biol . 2000;29:1217. doi: 10.1088/0031-9155/29/10/005. [DOI] [PubMed] [Google Scholar]
  • 9.Wieczorek H. The image quality of FBP and MLEM reconstruction. Phys Med Biol . 2010;55:3161. doi: 10.1088/0031-9155/55/11/012. [DOI] [PubMed] [Google Scholar]
  • 10.Slomka PJ, Patton JA, Berman DS, Germano G. Advances in technical aspects of myocardial perfusion SPECT imaging. J Nucl Cardiol . 2009;16:255–276. doi: 10.1007/s12350-009-9052-6. [DOI] [PubMed] [Google Scholar]
  • 11.Kalantari F, Rajabi H, Saghar M. Quantification and reduction of attenuation related artifacts in SPET by applying attenuation model during iterative image reconstruction: a Monte Carlo study. Hell J Nucl Med . 2011;14:278. [PubMed] [Google Scholar]
  • 12.Segars W, Mahesh M, Beck T, Frey E, Tsui B. Realistic CT simulation using the 4D XCAT phantom. Med Phys . 2008;35:3800. doi: 10.1118/1.2955743. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Wackers FJ, Berman DS, Maddahi J, Watson DD, Beller GA, Strauss HW, et al. Technetium-99m-hexakis-2-methoxy-iso-butyl-isonitrile: human biodistribution, dosimetry, safety and preliminary comparison to thallium-201 for myocardial perfusion imaging. J Nucl Med . 1989;30:301–311. [PubMed] [Google Scholar]
  • 14.Knoll P, Kotalova D, Köchle G, Kuzelka I, Minear G, Mirzaei S, et al. Comparison of advanced iterative reconstruction methods for SPECT/CT. Z Med Phys . 2012;22:58–69. doi: 10.1016/j.zemedi.2011.04.007. [DOI] [PubMed] [Google Scholar]
  • 15.Utsunomiya D, Tomiguchi S, Shiraishi S, Yamada K, Honda T, Kawanaka K, et al. Initial experience with X-ray CT based attenuation correction in myocardial perfusion SPECT imaging using a combined SPECT/CT system. Ann Nucl Med . 2005;19:485–489. doi: 10.1007/BF02985576. [DOI] [PubMed] [Google Scholar]
  • 16.Kinahan PE, Hasegawa BH, Beyer T, editors X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med . 2003;33:166–179. doi: 10.1053/snuc.2003.127307. [DOI] [PubMed] [Google Scholar]
  • 17.Buck AK, Nekolla S, Ziegler S, Beer A, Krause BJ, Herrmann K, et al. Spect/Ct. J Nucl Med . 2008;49:1305–1319. doi: 10.2967/jnumed.107.050195. [DOI] [PubMed] [Google Scholar]
  • 18.Seo Y, Wong KH, Sun M, Franc BL, Hawkins RA, Hasegawa BH. Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system. J Nucl Med . 2005;46:868–877. [PubMed] [Google Scholar]
  • 19.Teimourian B, Ay M, Zafarghandi MS, Ghafarian P, Ghadiri H, Zaidi H. A novel energy mapping approach for CT-based attenuation correction in PET. Med Phys. 2012;39 doi: 10.1118/1.3694108. [DOI] [PubMed] [Google Scholar]
  • 20.Kheruka S, Naithani U, Maurya A, Painuly N, Aggarwal L, Gambhir S. A study to improve the image quality in low-dose computed tomography (SPECT) using filtration. Indian J Nucl Med . 2011;26:14. doi: 10.4103/0972-3919.84595. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Hajizadeh M, Oloomi S, P knoll, H Taleshi. A new approach to scatter correction in SPECT images based on Klein_Nishina equation. Iran J Nucl Med . 2013;21.1:19–25. [Google Scholar]
  • 22.Hutton BF, Buvat I, Beekman FJ. Review and current status of SPECT scatter correction. Phys Med Biol . 2011;56:R85. doi: 10.1088/0031-9155/56/14/R01. [DOI] [PubMed] [Google Scholar]
  • 23.Konik AB. Evaluation of attenuation and scatter correction requirements in small animal PET and SPECT imaging: The University of Iowa; 2010. [Google Scholar]
  • 24.Van Holen R, Vandenberghe S, Staelens S, De Beenhouwer J, Lemahieu I. Fast 3D iterative image reconstruction for SPECT with rotating slat collimators. Phys Med Biol . 2009;54:715. doi: 10.1088/0031-9155/54/3/016. [DOI] [PubMed] [Google Scholar]
  • 25.Ärlig Å, Gustafsson A, Jacobsson L, Ljungberg M, Wikkelsö C. Attenuation correction in quantitative SPECT of cerebral blood flow: a Monte Carlo study. Phys Med Biol . 2000;45:3847. doi: 10.1088/0031-9155/45/12/324. [DOI] [PubMed] [Google Scholar]
  • 26.Xiao J, de Wit TC, Staelens SG, Beekman FJ. Evaluation of 3D Monte Carlo–Based Scatter Correction for 99mTc Cardiac Perfusion SPECT. J Nucl Med . 2006;47:1662–1669. [PubMed] [Google Scholar]
  • 27.Gustafsson A, Ärlig Å, Jacobsson L, Ljungberg M, Wikkelsö C. Dual-window scatter correction and energy window setting in cerebral blood flow SPECT: a Monte Carlo study. Phys Med Biol . 2000;45 doi: 10.1088/0031-9155/45/11/323. [DOI] [PubMed] [Google Scholar]
  • 28.Lazaro D, El Bitar Z, Breton V, Hill D, Buvat I. Fully 3D Monte Carlo reconstruction in SPECT: a feasibility study. Phys Med Biol . 2005;50:3739. doi: 10.1088/0031-9155/50/16/006. [DOI] [PubMed] [Google Scholar]
  • 29.Beekman FJ, de Jong HWAM, van Geloven S. Efficient fully 3-D iterative SPECT reconstruction with Monte Carlo-based scatter compensation. IEEE Trans Med Imaging . 2002;21:867–877. doi: 10.1109/TMI.2002.803130. [DOI] [PubMed] [Google Scholar]
  • 30.Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51:921–928. doi: 10.2967/jnumed.109.071571. [DOI] [PubMed] [Google Scholar]

Articles from Iranian Journal of Basic Medical Sciences are provided here courtesy of Mashhad University of Medical Sciences