Skip to main content
Iranian Journal of Pharmaceutical Research : IJPR logoLink to Iranian Journal of Pharmaceutical Research : IJPR
. 2013 Autumn;12(4):645–657.

Performance Evaluation of a Novel Potentiometric Membrane Sensor for Determination of Atorvastatin in Pharmaceutical Preparations

Farhad Ahmadi a,b,*, Nasim Asaadi b
PMCID: PMC3920707  PMID: 24523744

Abstract

A novel potentiometric ion-selective PVC membrane sensor for analysis of atorvastatin (AT) in pharmaceutical preparations based on atorvastatin-(tetraphenyl borate), (AT-(TPB)2) as sensing element, tetraphenyl borate as additive and tris-2-ethyl-hexyl phosphate (TOP) as plasticizer solvent was prepared. The electrode shows a good Nernestian response over the concentration range of 0.09–5586 μg mL-1of AT with slope of 30.1±0.1 mV/decade and limit of detection0.056μg mL-1.The response time of sensor is fats (less than 10 sec) and could be used for about one month in the pH range of 4.5–8.0. The electrode exhibit good selectivity for the AT in the presence of large amount of co-drugs and inorganic cations. The method is precise and accurate with mean relative standard deviation of <2%.Atorvastatin is determined successfully in several tablets by the proposed membrane.

Key Words: Atorvastatin, PVC membrane, Potentiometric sensor, Analysis of pharmaceutical preparation

Introduction

Atorvastatin, R-(R*,R*)]-2-(4-fluorophenyl)- β,δ-dihydroxy-5-(1-methylethyl)-3-phenyl- 4-[(phenylamino) carbonyl]-1H-pyrrole-1- heptanoic acid (pKa = 4.46, log P = 6.36), is a synthetic lipid-lowering agent (1, 2). Atorvastatin (AT) is an inhibitor of 3-hydroxy- 3-methylglutaryl-coenzyme A (HMG-CoA) reductase. This enzyme catalyzes the conversion of HMG-CoA to mevalonate, an early and rate-limiting step in cholesterol biosynthesis (3). It is applied for reducing of LDL-cholesterol, apolipoprotein B, and triglycerides and increasing of HDL-cholesterol in the treatment of hyperlipidaemias (4-7). Also, AT is indicated to diminish the risk of myocardial infarction (8), stroke (9) which decrease the risk for revascularization procedures (10) and angina (11). AT is rapidly absorbed after oral administration, however, due to pre-systematic clearance in the gastro-intestinal mucosa and metabolism in the liver, its absolute bioavailability is approximately 12%, and low plasma concentration is followed after taking the drug (5, 12). There are many reported methods for the determination of AT alone (13-20) or in combination with other drugs especially with amlodipine in pharmaceutical dosage forms or individually in biological fluids. Existing analytical methods for AT included: spectrophotometric methods (21, 22), HPLC-UV (23, 24), FT-Raman spectroscopy (25), automated enzyme inhibition assay method using radioactivity detection (26), UPLC (27) and also HPLC equipped with mass spectrometry detection (13-15). Novakora and coworkers, reported the HPLC methods for determination of AT in various fields of applications, including bio-analytical assays, pharmaceutical assays and environmental applications (28). Although the chromatographic methods are more sensitive and specific, but these devices requires sophisticated and expensive instruments, a specific column for each determination and special care with reagents before its injection into the chromatographic system (filtration, extraction and degassing) (23, 29-31). Furthermore, automated enzyme inhibition assay method need to long sample pretreatingtimes (26) and FT-Raman spectroscopy (25) has a small calibration range. Without all of these drawbacks and presenting high analytical ranges and reproducibility’s, minimal interference from associated and related species, direct application to turbid and colored drug solutions without any pre-treatment and low cost and easy operation of the potentiometric instrumentation, direct potentiometry using conventional ion selective electrodes (ISEs) is nowadays a highly desirable alternative for the analysis of pharmaceutical compounds (32-34). In the present work, a plastic membrane electrode is prepared based on incorporation of an ion pair complex of sodium tetraphenylborate (NaTPB) anion with AT in a PVC matrix plasticized with tris-2-ethyl-hexyl phosphate (TOP). The electrode was applied for analysis a range of AT concentrations in the region 0.089–5586 μg mL-1 using the Nernstain equation.

Experimental

Reagents and materials

The pure form of atorvastatin (C33H35FN2O5, gfw=558.7) used in the present work was supplied by Bakhtar Bioshimi Pharmaceutical Company (Kermanshah, Iran). All chemicals used where of analytical or pharmaceutical grade and solutions were prepared in double distilled water. Dibutyl phthalate (DBP), dioctyl phthalate (DOP), tri-butyl phosphate (TBP), tris-2-ethyl-hexyl phosphate (TOP), 2-nitrophenyloctyl ether (2-NPOE), NaTPB, high relative molecular weight PVC, and tetra-hydrofuran (THF) were obtained from Fluka, or Merck and used as received. 0.0558 gr of AT was dissolved in 1.0 mL MeOH and 0.5 mL of HCl 0.01 M to form the AT(HCl)2 and then diluted to 10 mL by adding double-distilled water. Then the ion pair was prepared from aqueous medium by adding 5.0 mL of saturated NaTPB solution to 10 mL of10-2M of AT solution. The resulting ionpair precipitate was filtered, washed thoroughly with distilled water and dried under N2 atmosphere at room temperature. Stock solution of AT (5586000 ng mL-1) was prepared by dissolving an appropriate amount of the AT in MeOH-water (20-80%). Other dilute solutions (0.00558–5586 μg mL-1) were prepared by serial dilution and both the pH and ionic strength were kept constant at 5 and 0.01M, respectively.

Apparatus

Potentiometric measurements were carried out at 25 ± 0.1 °C on a multichannel digital potentiometer pH/ion analyzer (model pH 162). The membranes were equilibrated for 1 day in 10-3M AT solution and the potentials were measured using PVC matrix membranes in conjunction with Ag/AgCl electrodes by setting up following electrochemical cell assembly:

The FT-IR measurement was carried out using Shimadzu (IR Prestige-21). The elemental analysis was performed using Heraeus CHN elemental Analyzer. The pH values of solutions were adjusted employing a Metrohm model 827 using a combined glass electrode. The chromatographic analysis was carried out according to our previous work (22) using a KNAURE instrument equipped with a power supplier and a UV detector. The HPLC was controlled by EZ-Chrome Elite software. The separation was performed on a Eurospher 100-5C18 column (240 × 4.0 mm i.d.). The mobile phase, consisted of MeOH–acetonitrile–0.05 M phosphate buffer (pH 3.0, 40:35:25 V/V%). The column was placed at an oven at 40 °C. The flow rate was 1.0 mL min–1, and the detection wavelength was 237 nm. The injection volume was 20 μL and the run time was 7.0 min.

Electrode preparation

The membrane composition was studied by varying the percentages (w/w %) of the ion pair, polyvinyl chloride (PVC) and plasticizer, until the optimum composition that exhibits the best performance characteristics was obtained (35- 37). The membranes were prepared by dissolving the required amount of the AT-(TPB)2, PVC and TOP, in 2.0 mL of THF. The solution mixture was poured into a 5.0 cm Petri dish and left to dry in air. To obtain the uniform membrane thickness, the amount of THF was kept constant in all preparations, and its evaporation was fixed for 4 h. A 15 mm diameter disk was cut out from the prepared membrane and glued using PVC–THF paste to the polished end of a plastic tube. The electrode body was filled with internal solution. The electrode is preconditioned by soaking in 558.6 μg mL-1 of AT solution.

Selectivity

The potentiometric selectivity coefficients (Kpot A.B ) of the proposed sensor towards different substances were determined according to previous reported work (38) by a separate solution method using the following equation (I) (39):

-log(kA.Bpot)=E1-E22.303RTZAF+(1-ZAZB)logaA          (I)

where KA.BPOTis the potentiometric selectivity coefficient, E1 is the potential measured in 10-3 M AT solution, E2 is the potential measured in 10-3 M interfering solution, ZA and ZB are the charges of AT and interfering ion, respectively, aA is the activity of the drug and 2.303 RTZAFrepresents the calibration slope of the investigated sensor (mV/concentration decade).

Construction of calibration graph

The conventional plastic membrane electrode was immersed in solutions of different concentrations of AT covering the concentration range 0.0055–5586 μg mL-1 of AT. The cell potential was recorded for each solution with constant stirring at 25 ± 0.1 °C and plotted against log[AT]. The slope of the calibration graph was calculated using Nernestain equation (II):

E=E1SE0+ 2.303 RTZF logAT             (II)

Where R is the gas constant, F is the Faraday equivalent and z is the charge of the analyte. The term EISE 0is a constant which is the sumof all invariants in the system.

Analysis of AT in lipitor, amostat, and biotor plus tablets

The tablet sample preparations were performed in accordance of work of Hussein et al., (40). A homogenized powder was prepared from five accurately weighed of each Lipitor (10, 20, 40 mg) –contains atorvastatin (10, 20, 40 mg), Amostat (5/20, mg) contains–amlodipine/ atorvastatin (5/20 mg), and Biotor Plus (5/20 mg) contains–amlodipine/atorvastatin (5/20 mg) tablets. An appropriate amount of each tablet powder was transferred into a 10 mL Erlenmeyer flask. Dissolution of the drug was assisted by means of a magnetic stirrer and by addition of 2.0 mL of MeOH. The mixture was then filtered and made up to the mark with acetate buffer (pH = 5.0, 0.01 M) in a 50 mL volumetric flask. In accordance of the label for each tablet (10, 20, 40 mg AT per tablet, and5/20 mg amlodipine/ atorvastatinper tablet), the amount of AT should be 55860 ng mL1M in volumetric flask. The AT content was determined by the proposed membrane-selective electrode, using the calibration method.

Results and Discussions

The proposed ion pair structure of AT-TPB

AT is a white to off–white crystalline powder with pKa= 4.46. It is insoluble in aqueous solutions with pH ≤ 4.0, while it is very slightly soluble in distilled water and freely soluble in methanol. For preparation of ionpair we found out that the AT is soluble in mixture of MeOH-0.01 M HCl solution (66:33%). The AT can behave as cation in acidic medium, due to presence of amid and carboxylic moiety (41). The composition of the precipitate was investigated by elemental analysis. The calculated percentage of C, H, and N for the AT-(TPB)2 (with mole ratio 1:2 was 81.16%, 6.42% and 2.34%, while the found percentages amounted to 80.5%, 6.21%, 2.26%, respectively. The results indicated the formation of a 1:2 ion pair (AT:TPB). Also the FT-IR measurements revealed that the –CO–NH– and –COOH moiety can be protonated and formed the ion pair with NaTPB. According to the calculated elemental analysis data and the FT-IR measurements for the ion-associate complex we proposed the follow structure for resulted 1:2 AT-(TPB)2 ionpair (Figure 1).

Figure 1.

Figure 1

Suggested structure of atorvastatin-tetraphenylborate (AT-(TPB)2) ionpair

The evaluation of proposed ionpair as ionophor in PVC membrane

In the next experiments AT-(TPB)2 was used as an ionophor in construction of PVC-membrane electrode for AT. The potential response of constructed electrode for AT and several drugs are shown in Figure 2. As it is seen, except for the AT, for all other drugs, the slopes of the corresponding potential-pM plots are much lower than the expected Nernstian slope of 30 mV per decade for the 1:2 AT-(TPB)2 ion pair (42).

Figure 2.

Figure 2

Electrode response for AT prepared with AT-(TPB)2 ionpair as ionophore, the membrane contain of PVC:NaTPB:ionopher: TOP with 35:1.0:1.5:65%, respectively, the pH=4.0

Investigation of membrane composition

The proposed membrane sensor generates stable potential response in aqueous solutions containing AT after conditioning for 24 h in a 558.6 μg mL-1AT solution. Table 1 shows the data obtained with membranes having various ratios of different constituents. It is well known that some important features of the PVC-based membranes, such as the nature and amount of ionophor, the properties of the plasticizer, the plasticizer/PVC ratio and, especially, the amount of additive used, significantly influence the sensitivity and selectivity of the ion-selective electrodes (35, 36, 42). From the data, TOP was found to be the optimum available plasticizer for the PVC membrane sensor. It plasticizes the membrane, dissolves the ion-association complex and adjusts both of the membranes permittivity and ion-exchanger sites mobility to give highest possible selectivity and sensitivity (43). Other plasticizers such as TBP, NPOE, DBP and DOP failed in dissolving the ionpair and thus gave noisy responses. As can be seen from Table 1, by increasing amount of ion pair in membranes, the slopes were increased. Using 4% of AT-(TPB)2 and 65% of TOP in the membrane electrode displays Nernstian slope towards AT (membrane NO.9). As it is observed, (membrane NO. 12) the response of electrode to AT is affected only by the ionpair. Also, the sensitivity of the electrode in the absence of additive is poor (no.4 with a slope 27.2 mV/decade). Addition of additives (NaTPB) improves the sensitivity of the AT sensor. Table 1 revealed that addition of NaTPB as a suitable additive to the membrane increases the sensitivity of the sensor from 27.2 (NO.4) to 30.1 mV/decade (NO. 9). It is well established that the presence of lipophilic negatively charged additives improves the potentiometric behavior of certain selective electrodes not only by reducing the ohmic resistance and improving the response behavior and selectivity, but also, in cases where the extraction capability of the ionophor is poor, by enhancing the sensitivity of the membrane electrode (44). However, the best performance was obtained with membrane NO.9. The optimum response of the electrode was tested after conditioning for different periods of time in 558.6 μg mL-1 AT. The slope obtained using 24 h of conditioning was closer to the theoretical slope expected on the basis of the Nernstian equation. Longer conditioning times produced no further improvements in the response. The proposed PVC membrane electrode was used over a period of 1.0 month without any significant change in potential. During these times, the detection limit and the slope of the electrode remained almost constant. Subsequently, the electrochemical behavior of the electrode gradually deteriorated. This would be due to aging effect and leaching of the ion pair, NaTPB and the solvent mediator from the membrane into the solution by time.

Table 1.

Optimization of membrane ingredients

NO Percentage(W/W) of various components in membrane
L.R (μg mL -1 ) RT (s)
PVC AT-(TPB) 2 NaTPB NPOE DOP TBP TOP DBP slope
1 33.0 1.0 - 66.0 - - - 39.5 16.76 - 5586 20
2 33.0 1.0 - - 66.0 - - 22.2 223.4 – 5586 35
3 33.0 1.0 - - - 66.0 - 19.3 223.4 – 5586 40
4 33.0 1.0 - - - - 66.0 27.2 3.9 – 5586 18
5 33.0 1.0 - - - - - 66.0 20.5 28 – 5586 25
6 33.5 1.0 0.5 - - - 65 28.5 0.44– 5586 18
7 35 1.0 1.5 - - - 65 27.9 2.8 – 5586 18
8 36.0 0.5 2.0 - - - 62 20.1 5.6 – 5586 20
9 30.0 4.0 1.0 - - - 65.0 30.1 0.09 – 5586 10
10 30.0 3.0 0.0 - - - 67.0 28.8 0.11 – 5586 12
11 30.0 3.0 1.0 - - - 66.0 29.4 0.11 – 5586 10
12 30 0.0 1.0 - - - 69 12 1117 – 5586 50

The effect of pH on electrode response

The pH dependence of the electrode potential was tested over the pH range 2.5-11 for 2234, 1.4 and 0.14 μg mL-1 AT solutions. The pH was adjusted with NaOH or HCl solution. The potential was independent of pH in the range 4.5-8for three concentrations (Figure 3). In acidic pH the precipitation of AT was observed and the potential of the electrode moved to negative potentials with increasing in pH of the working solution. Over the pH 8.0, the potential decayed. In addition, the potentials displayed by the electrode were noisy as the ionpair may decompose and leaked in to the solution.

Figure 3.

Figure 3

Effect of pH of the test solution on the electrode potential reading, (●) 2234, (¡) 1.4and (▲) 0.14 μg mL-1AT solutions, the composition of membrane was like to the membrane NO.9

Effect of temperature of the test solution

Calibration plot (E versus pAT) was constructed in the test solution temperature range 25–60 °C for proposed membrane. For the determination of the isothermal coefficients (dE°/dt) of the electrode, the standard electrode potentials () of the cell at different temperatures were obtained from the calibration plots as the intercepts at pAT = 0. The slope, usable concentration range, and the standard potential () of the electrode at each temperature are given in Table 2. For the determination of the isothermal coefficient (dE°/dt) of the cell, the standard electrode potentials , is plotted versus t-25 (Figure 4), where t is the temperature of the test solution (°C). A straight-line plot is obtained according to the following equation (III):

Table 2.

Performance characteristics of proposed electrode at different temperatures

Temperature ( ° C) Slope (mV/decade) Usable concentration range (pAT) E ° (mV)
25 30.1 5-2 -32.5
30 30.1 5-2 -32.8
35 30 5-2 -32
40 30 5-2 -32.8
45 29.8 4.8-2 -34.4
50 29.5 4.8-2 -35.5
60 29 4.5-2 -37

Figure 4.

Figure 4

Variation of ˚Eelec with temperature

E°= E°(25) + (dE°/dt)(t-25) (III)

The slope of the straight line obtained represents the isothermal coefficient of 0.00018 V/°C for proposed electrode (45). This low value of isothermal temperature coefficient revealed that the electrode has good thermal stability within the studied temperature range (25-60 °C).

Effect of internal reference solution

The influence of the concentration of the internal solution was investigated for different concentrations of the AT solutions (55.8, 558.6, and 5586 μg mL-1). It was found that, by changing the concentration of the internal reference solution, the slope of the Nernstian plots remains nearly constant. Therefore, 558.6 μg mL-1concentration of the reference solution was quite appropriate for smooth functioning of the proposed sensor.

Response time

Response time is an important factor for any potentiometric sensor. In fact response time is the time needed for the electrode to reach a stable. In this study, response time was recorded by changing the AT activity; over an activity range of 0.111-5586 μg mL-1. The resulting data depicted in Figure 5, show that the time required to achieve a steady potential after successive immersion of a series of AT drugis ≤10 s.

Figure 5.

Figure 5

Dynamic response of the PVC membrane electrode based on AT-(TPB)2 for step change in activities of AT

Selectivity

The selectivity of the membrane in the presence of some inorganic cations, co-formulated drug (Amlodipine), other statin drugs, was investigated. The selectivity coefficients of electrode which determined by the separate solution method using Eq. (I) are presented in Table 3. The −log(KA.Bpot=πr2) values of the electrode show a high selectivity of the sensor for the AT. The selectivity of membrane is mainly dependence to how much matching is present between the locations of the lipophilic sites in the two competing species in the bathing solution side and those present in the receptor of the ion exchanger. The inorganic cations do not interfere owing to the differences in ionic size. In the case of other statin drugs, the high selectivity is mainly attributed to the difference in polarity and lipophilic character of their molecules relative to AT.

Table 3.

Selectivity coefficients −log(KA.Bpot) of various interference with the AT-selective electrode

Interfering compounds −log(
KA.Bpot
)
Interfering compounds −log(
KA.Bpot
)
Carbidopa 4.6 Na+ 4.5
Acetaminophen 3.5 K+ 4.8
Amoxicillin 3.6 NH4+ 4.4
Amlodipine 4.2 Cd2+ 4.2
Levodopa 4.7 Cu2+ 2.8
Minoxidil 3.5 Ag+ 3.5
Hydrocortisone 4.1 Mg2+ 4.1
Levostatine 3.8 Fe2+ 4.2
Ca2+ 3.4 Fe3+ 3.1

Effect of MeOH on membrane performance

As the AT is insoluble in aqueous solutions with pH ≤ 4.0 and is very slightly soluble in distilled water, therefore, the sample preparation was performed in 20:80 %V/V of MeOH: acetate buffer, (see section 2.6). Therefore, the effect of MeOH on performance of the membrane was also studied in partially non-aqueous medium using water–methanol mixture. The membrane worked satisfactorily up to a maximum 20% (v/v) content of methanol. In these mixtures, the working concentration range and slope remained unaltered, however, above a 20% non-aqueous content the slope decreased appreciably and reliable measurements could not be obtained. Also, the membranes were destroyed due to leaching of the ionophore from the PVC matrix.

Calibration curve, linearity range, detection limit, and reproducibility

The choice of optimum membrane composition and measurement conditions must be resulted inacceptable calibration curve and a high degree of linearity over a usable range of concentrations. The best performance was exhibited by membrane containing the PVC:AT-(TPB)2: NaTPB: TOP in the ratio 30: 4.0: 1.0: 65, respectively. This membrane showed a nearly Nernestian response with slope of 30.1 ± 0.1 mV/decade and a linear concentration range 0.09–5586 μg mL-1 (Figure 6). The detection limit, defined as the cross section of the two extrapolated linear segments of the calibration curve, was0.056 μg mL-1. Five replicate determinations at three different concentration levels (0.5, 50, 5000 ng mL-1) were carried out using the proposed electrode to test the precision of the method. The relative standard deviations (RSD) were found 2.18%, 1.98, and 1.54%, respectively, indicating reasonable repeatability and reproducibility of the selected method.

Figure 6.

Figure 6

Calibration curve of AT (μg mL-1) obtained with proposed electrode at optimum membrane composition and measurement conditions

Analytical applications

The investigated electrode was proved to be useful in the potentiometric determination of AT in different tablets such as Lipitor (10, 20, 40 mg)–contains atorvastatin (10, 20, 40 mg), Amostat (5/20 mg)–contain amlodipine/atorvastatin (5/20 mg), and Biotor Plus (5/20 mg)–contain amlodipine/atorvastatin (5/20 mg). Collective results are given in Table 4. From the results, it is evident that the present electrode is very useful as a potentiometric sensor for the determination of AT in pharmaceutical preparations. It must be mentioned that the electrode should be calibrated every day before measurements. If any deterioration in the slope is observed, the electrode must be replaced with another new one. To compare the proposed method to a reported method, AT in three Amostat, Biotor Plusand Lipitor tablets were assayed by HPLC method. Statistical comparison of the results of the proposed and reported methods (Table 4) was performed with regard to accuracy and precision using the t- and F-ratio tests. At 95% confidence level, the calculated t- and F-values did not exceed the critical values, indicating that there is no significant difference between the proposed method and HPLC method with regard to accuracy and precision.

Table 4.

determination of AT content in five types of tablets

Tablet Name Proposed sensor (n=4)
HPLC (n=4)
Add (mg/tablet) Found (mg/tablet) ±SD Recovery% Found (mg/tablet) ±SD
Lipitor (10) 0.0 9.85±0.18 98.5 9.94±0.12
2.0 11.83±0.24 98.5 12.03±0.18
5.0 14.80±0.25 98.6 15.12±0.21
at- test (95%; 3.18)
aF-test(95% ; 9.28)
0.83
2.25
Lipitor (20) 0.0 19.95±0.39 99.7 20.1±0.25
2.0 22.05±0.45 100.0 22.14±0.24
5.0 25.14±0.40 100.6 24.97±0.34
at- test (95%; 3.18)
aF-test(95% ; 9.28)
0.65
2.43
Lipitor (40) 0.0 40.2±0.77 100.5 39.87±0.41
2.0 41.89±0.70 99.7 42.13±0.50
5.0 44.8±0.80 97.7 45.10±0.65
at- test (95%; 3.18)
aF-test(95% ; 9.28)
0.76
3.53
Amostat (5/20) 0.0 19.75±0.35 98.7 19.89±0.24
2.0 22.1±0.49 100.4 21.98±0.27
5.0 24.85±0.41 99.4 25.12±0.31
at- test (95%; 3.18)
aF-test(95% ; 9.28)
0.66
2.13
Biotor Plus (5/20) 0.0 19.9±0.37 99.5 20.13±0.27
2.0 21.87±0.47 22.4 22.05±0.32
5.0 24.75±0.41 99.0 25.09±0.28
at- test (95%; 3.18)
aF-test(95% ; 9.28)
3.10
1.88

Conclusion

We founded that, this plasticized PVC membrane containing AT-(TPB)2 has been shown good sensitivity, reproducibility, selectivity and better life time. Also, it is very easy to prepare and has a wide linear dynamic range and low detection limit. The electrode works in a pH range of 4.5– 8, which makes it useful for measurements of AT in neutral and physiological pH. High selectivity, low detection limit, and rapid response make this electrode suitable for measuring the concentration of AT in pharmaceutical preparations, without any preconcentration or pretreatment steps. Table 5 compared the performance characteristics of the proposed electrode with other previously reported works for analysis of AT. Although, this work has higher limit of detection (LOD) and lower linear range (L.R) in comparison with other methods such as LC-MS or HPLC-UV, but it is more selective, easy to performance, economic, simple and does not need sample preparation.

Table 5.

the comparison of proposed sensor with other analytical methods for analysis of AT

Method Name Sample preparation L.R (μg mL -1 ) LOD (μg mL -1 ) Ref.
Derivative Spectrophotometry -- 10–35 1.75 [46]
Spectrofluorimetry -- 1.5–4.0 0.012 [47]
LC-MS/MS Protein Precipitation 0.0005–0.15 -- [48]
capillary electrophoresis -- 1.0–100 60 [18]
LC-MS Solid phase extraction 0.0002–0.03 0.00006 [49]
HPLC-UV Liquid-liquid-liquid microextraction 0.001–0.5 0.0004 [50]
Proposed sensor -- 0.09–5586 0.056

Acknowledgments

We gratefully acknowledge Vice Chancellor for Research and Technology, Kermanshah University of Medical Sciences for financial support. This work was performed in partial fulfillment of the requirements for a Pharm. D Nasim Assadi in Faculty of Pharmacy, Kermanshah University of Medical Sciences, Kermanshah-Iran.

References

  • 1.Lea AP, Mc Tavish D, Atorvastatin A review of its pharmacology and therapeutic potential in the management of hyperlipidaemias. Drugs. 1997;53:828–847. doi: 10.2165/00003495-199753050-00011. [DOI] [PubMed] [Google Scholar]
  • 2.Ness GC, Chambers CM, Lopez D. Atorvastatin action involves diminished recovery of hepatic HMG-CoA reductase activity. J. Lipid. 1998;39:75–84. [PubMed] [Google Scholar]
  • 3.Conde K, Pineda G, Newton RS, Fernandez ML. Hypocholesterolemic effects of 3-hydroxy-3-methylglutaryl coenzyme A (HMG-CoA) reductase inhibitors in the guinea pig: Atorvastatin versus simvastatin. Biochem. Pharmacol. 1999;58:1209–1219. doi: 10.1016/s0006-2952(99)00203-8. [DOI] [PubMed] [Google Scholar]
  • 4.Malek Mohamammd M, Fahimi F, Fakharian A, Karimi Gamishan M, Sistanizad M, Fayyazi N, Khalilzadeh S. Methacoline Challenge test as an Evaluator of Response to Statins in Bronchial Hyperresponsiveness. Iran. J. Pharm. Res. 2012;11:969–977. [PMC free article] [PubMed] [Google Scholar]
  • 5.Nawrocki JW, Weiss SR, Davidson MH, Sprecher DL, Schwartz SL, Lupien PJ, Jones PH, Haber HE, Black DM. Reduction of LDL cholesterol by 25% to 60% in patients with primary hypercholesterolemia by atorvastatin, a new HMG-CoA reductase inhibitor. Arterioscler. Thromb. Vasc. Biol. 1995;15:678–682. doi: 10.1161/01.atv.15.5.678. [DOI] [PubMed] [Google Scholar]
  • 6.Jones P, Kafonek S, LauroraI , Hunninghake D. Comparative dose efficacy study of atorvastatin versus simvastatin, pravastatin, lovastatin, and fluvastatin in patients with hypercholesterolemia (the CURVES study) Am. J. Cardiol. 1998;81:582–587. doi: 10.1016/s0002-9149(97)00965-x. [DOI] [PubMed] [Google Scholar]
  • 7.Black DM, Bakker-Arkema RG, Nawrocki JW. An overview of the clinical safety profile of atorvastatin (lipitor), a new HMG-CoA reductase inhibitor. Arch. Intern. Med. 1998;158:577–584. doi: 10.1001/archinte.158.6.577. [DOI] [PubMed] [Google Scholar]
  • 8.Cheng X, Liao YH, Zhang J, Li B, Ge H, Yuan J, Wang M, Lu B, Liu Y, Cheng Y. Effects of Atorvastatin on Th polarization in patients with acute myocardial infarction. J. Eur. Heart. Failure. 2005;7:1099–1104. doi: 10.1016/j.ejheart.2005.01.020. [DOI] [PubMed] [Google Scholar]
  • 9.Sever PS, Poulter NR, Dahlöf B, Wedel H. Different time course for prevention of coronary and stroke events by atorvastatin in the Anglo-scandinavian cardiac outcomes trial–lipid-lowering arm (ASCOT-LLA) Am. J. Cardiol. 2005;96:39–44. doi: 10.1016/j.amjcard.2005.06.025. [DOI] [PubMed] [Google Scholar]
  • 10.Newman C, Tsai J, Szarek M, Luo D, Gibson E. Comparative safety of atorvastatin 80 mg versus 10 mg derived from analysis of 49 completed trials in 14, 236 patients. Am. J. Cardial. 2006;97:61–67. doi: 10.1016/j.amjcard.2005.07.108. [DOI] [PubMed] [Google Scholar]
  • 11.Tousoulis D, Antoniades C, Katsi V, Bosinakou E, Kotsopoulou M, Tsioufis C, Stefanadis C. The impact of early administration of low-dose atorvastatin treatment on inflammatory process, in patients with unstable angina and low cholesterol level. Int. J. Cardiol. 2006;109:48–52. doi: 10.1016/j.ijcard.2005.05.055. [DOI] [PubMed] [Google Scholar]
  • 12.Bakker-Arkema RG, Davidson MH, Goldstein RJ, Isaacsohn RJ, Davignon SL, Weiss SR, Keilson LM, Brown WV, Miller VT, Shurzinske LJ, M BlackD. Efficacy and safety of a new HMG-CoA reductase inhibitor, atorvastatin, in patients with hypertriglyceridemia. J. Am. Med. Assoc. 1996;275:128–133. [PubMed] [Google Scholar]
  • 13.Miao XS, Metcalfe CD. Determination of cholesterol-lowering statin drugs in aqueous samples using liquid chromatography–electrospray ionization tandem mass spectrometry. J. Chromatogr. A. 2003;998:133–141. doi: 10.1016/s0021-9673(03)00645-9. [DOI] [PubMed] [Google Scholar]
  • 14.Bullen WW, Miller RA, Hayes RN. Development and validation of a high-performance liquid chromatography tandem mass spectrometry assay for atorvastatin, ortho-hydroxyatorvastatin, and para-hydroxy atorvastatin in human, dog, and rat plasma. J. Am. Soc. Mass Spectrom. 1999;10:55–66. doi: 10.1016/S1044-0305(98)00118-4. [DOI] [PubMed] [Google Scholar]
  • 15.Jemal M, Ouyang Z, Chen BC, Teitz D. Quantitation of the acid and lactone forms of atorvastatin and its biotransformation products in human serum by high-performance liquid chromatography with electrospray tandem mass spectrometry. Rapid Commun. Mass Spectrom. 1999;13:1003–1015. doi: 10.1002/(SICI)1097-0231(19990615)13:11<1003::AID-RCM597>3.0.CO;2-L. [DOI] [PubMed] [Google Scholar]
  • 16.Hai-Rong S, Zhong-Dong L, Ming-Kang Z. HPLC assay and pharmacokinetic study of atorvastatin in beagle dogs after oral administration of atorvastatin self-microemulsifying drug delivery system. Die Pharmazie. 2006;61:18–20. [PubMed] [Google Scholar]
  • 17.Bahrami G, Mohammadi B, Mirzaeei S, Kiani A. Determination of atorvastatin in human serum by reversed-phase high-performance liquid chromatography with UV detection. J. Chromatogr. B. 2005;826:41–45. doi: 10.1016/j.jchromb.2005.08.008. [DOI] [PubMed] [Google Scholar]
  • 18.AlShehri MM. A validated capillary electrophoresis method for simultaneous determination of ezetimibe and atorvastatin in pharmaceutical formulations. Saudi Pharm. J. 2012;20:143–148. doi: 10.1016/j.jsps.2011.09.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Erturk S, Aktas ES, Ersoy L, Fıcıcıoglulu S. An HPLC method for the determination of atorvastatin and its impurities in bulk drug and tablets. J. Pharm. Biomed. Anal. 2003;33:1017–1023. doi: 10.1016/s0731-7085(03)00408-4. [DOI] [PubMed] [Google Scholar]
  • 20.Erturk S, Onal A, Cetin S M. Analytical methods for the quantitative determination of 3-hydroxy-3-methylglutaryl coenzyme a reductase inhibitors in biological samples. J. Chromatogr. B. 2003;793:193–205. doi: 10.1016/s1570-0232(03)00314-3. [DOI] [PubMed] [Google Scholar]
  • 21.Darwish HW, Hassan SA, Salem MY, El-Zeiny BA. Three different spectrophotometric methods manipulating ratio spectra for determination of binary mixture of Amlodipine and Atorvastatin. Spectrochim. Acta A. 2011;83:140–148. doi: 10.1016/j.saa.2011.08.005. [DOI] [PubMed] [Google Scholar]
  • 22.Ahmadi F, Ghordoie-Milan R. Spectrophotometric determination of atorvastatin and amlodipine by H-point standard addition method with simultaneous addition of both analytes innonaqueous solution. Collec. Czech. Chem. Commun. 2011;76:193–205. [Google Scholar]
  • 23.Farahani H, Norouzi P, Beheshti A, Sobhi HR, Dinarvand R, Ganjali MR. Quantitation of atorvastatin in human plasma using directly suspended acceptor droplet in liquid–liquid–liquid microextraction and high-performance liquid chromatography-ultraviolet detection. Talanta. 2009;80:1001–1006. doi: 10.1016/j.talanta.2009.08.033. [DOI] [PubMed] [Google Scholar]
  • 24.Mohammadi A, Rezanour N, Ansari Dogaheh M, Ghorbani Bidkorbeh F, Hashem M, Walker RB. A stability-indicating high performance liquid chromatographic (HPLC) assay for the simultaneous determination of atorvastatin and amlodipine in commercial tablets. J. Chromatogr. B. 2007;846:215–221. doi: 10.1016/j.jchromb.2006.09.007. [DOI] [PubMed] [Google Scholar]
  • 25.Skorda D, Kontoyannis CG. Identification and quantitative determination of atorvastatin calcium polymorph in tablets using FT-Raman spectroscopy. Talanta. 2008;74:1066–1070. doi: 10.1016/j.talanta.2007.07.030. [DOI] [PubMed] [Google Scholar]
  • 26.Valesky RJ, LiuL G, Musson D, Zhao JJ. Automated enzyme inhibition assay method for the determination of atorvastatin-derived HMG-CoA reductase inhibitors in human plasma using radioactivity detection. J. Pharmacol. Toxicol. Methods. 2008;57:61–69. doi: 10.1016/j.vascn.2007.06.002. [DOI] [PubMed] [Google Scholar]
  • 27.Kadav AA, Vora DN. Stability indicating UPLC method for simultaneous determination of atorvastatin, fenofibrate and their degradation products in tablets. J. Pharmaceut. Biomed. Anal. 2008;48:120–126. doi: 10.1016/j.jpba.2008.05.018. [DOI] [PubMed] [Google Scholar]
  • 28.Novakova L, Satınsky D, Solich P. HPLC methods for the determination of simvastatin and atorvastatin. Trend Anal. Chem. 2008;27:352–367. [Google Scholar]
  • 29.Ahmadi F, Ahmadi J, Rahimi-Nasrabadi M. Computational approaches to design a molecular imprinted polymer for highselective extraction of 3,4-methylenedioxymethamphetamine from plasma. J. Chromatogr. A. 2011;1218:7739–7747. doi: 10.1016/j.chroma.2011.08.020. [DOI] [PubMed] [Google Scholar]
  • 30.Gholivand MB, Khodadadian M, Ahmadi F. Computer aided-molecular design and synthesis of a high selective molecularlyimprinted polymer for solid-phase extraction of furosemide from human plasma. Anal. Chim. Acta. 2010;658:225–232. doi: 10.1016/j.aca.2009.11.019. [DOI] [PubMed] [Google Scholar]
  • 31.Ahmadi F, Asgharloo H, Sadeghi S, Gharehbagh-Aghababa V, Adibi H. Post-derivatization procedure for determination of hippuric acid after extraction by an automated micro solid phase extraction system and monitoring by gas chromatography. J. Chromatogr. B. 2009;877:2945–2951. doi: 10.1016/j.jchromb.2009.06.036. [DOI] [PubMed] [Google Scholar]
  • 32.Alizadeh N, Mehdipour R. Drug-selective electrode for ketamine determination in pharmaceutical preparations and electrochemical study of drug with BSA. J. Pharm. Biomed. Anal. 2002;30:725–731. doi: 10.1016/s0731-7085(02)00350-3. [DOI] [PubMed] [Google Scholar]
  • 33.Arvand M, Vejdani M, Moghimi M. Construction and performance characterization of an ion selective electrode for potentiometric determination of atenolol in pharmaceutical preparations. Desalination. 2008;225:176–184. [Google Scholar]
  • 34.Hussien EM, Abdel-Gawada FM, Issa YM. Ion-selective electrodes for determination of fluoxetine in capsules and in biological fluids. Biochem. Engineer. J. 2011;53:210–215. [Google Scholar]
  • 35.Ganjali1 MR, Gholivand MB, Rahimi-Nasrabadi M, Maddah B, Salavati-Niasari M, Ahmadi F. Synthesis of a new octadentates schiff’s base and its application in construction of a highly selective and sensitive lanthanum (III) membrane sensor. Sensor Lett. 2006;4:356–363. [Google Scholar]
  • 36.Gholivand MB, Ahmadi F, Rafiee E. A novel Al (III)-selective electrochemical sensor based on N,N’-bis (salicylidene)-1,2-phenylenediamine complexes. Electroanalysis. 2006;16:1620–1626. [Google Scholar]
  • 37.Gholivand MB, Rahimi-Nasrabadi M, Ganjali MR, Salavati-Niasari M. Highly selective and sensitive copper membrane electrode based on a new synthesized Schiff base. Talanta. 2007;73:553–560. doi: 10.1016/j.talanta.2007.04.010. [DOI] [PubMed] [Google Scholar]
  • 38.Khalil S, Kelzieh A, Ibrahim SA. Ion-selective electrode for the determination of prazosin in tablets. J. Pharm. Biomed. Anal. 2003;33:825–829. doi: 10.1016/s0731-7085(03)00337-6. [DOI] [PubMed] [Google Scholar]
  • 39.Cosofret VV, Buck RP. Evaluation of the selectivity coefficients in pharmaceutical analysis. Analyst . 1984;109:1321–1328. doi: 10.1039/an9840901321. [DOI] [PubMed] [Google Scholar]
  • 40.Hussiena EM, Abdel-Gawada FM, Issa YM. Ion-selective electrodes for determination of fluoxetine in capsules and inbiological fluids. Biochem. Engin. J. 2011;53:10–215. [Google Scholar]
  • 41.Rahimi-Nasrabadi M, Ahmadi F, Pourmortazavi SM, Ganjali MR, Alizadeh K. Conductometric study of complex formations between some substituted pyrimidines and some metal ions in acetonitrile and the determination of thermodynamic parameters. J. Mol. Liq. 2009;144:97–101. [Google Scholar]
  • 42.El-Ragehy NA, El-Kosasy AM, Abbas SS, El-Khatee SZ. Polymeric membrane electrodes for selective determination of the central nervous system acting drugs fluphenazine hydrochloride and nortriptyline hydrochloride. Anal. Chim. Acta. 2000;418:93–100. [Google Scholar]
  • 43.Bakker E, Buhlmann PH, Pretsch E. Carrier-based ion-selective electrodes and bulk optodes 1 general characteristics. Chem. Rev. 1997;97:3083–3132. doi: 10.1021/cr940394a. [DOI] [PubMed] [Google Scholar]
  • 44.Gupta KC, D’Arc MJ. Effect of concentration of ion exchanger, plasticizer and molecular weight of cyanocopolymers on selectivity and sensitivity of Cu (II) ion selective electrodes. Anal. Chim. Acta. 2001;437:199–216. doi: 10.1016/s0039-9140(00)00479-3. [DOI] [PubMed] [Google Scholar]
  • 45.Khalil S, Kelzieh A, Ibrahim SA. Ion-selective electrode for the determination of prazosin in tablets. J. Pharmaceut. Biomed. Anal. 2003;33:825–829. doi: 10.1016/s0731-7085(03)00337-6. [DOI] [PubMed] [Google Scholar]
  • 46.Baghdady YZ, Al-Ghobashy MA, Abdel-Aleem AE, Weshahy SA. Spectrophotometric and TLC-densitometric methods for the simultaneous determination of Ezetimibe and Atorvastatin calcium. J. Adv. Res. 2013;4:51–59. doi: 10.1016/j.jare.2012.01.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Sharaf El-Din MMK, Salama FMM, Nassar MWI, Attia KAL, Kaddah MMY. Validated spectrofluorimetric method for the determination of atorvastatinin pharmaceutical preparations. J. Pharm. Anal. 2012;2:200–205. doi: 10.1016/j.jpha.2012.01.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Polagania SR, Pillib NR, Gajulab R, Gandu V. Simultaneous determination of atorvastatin, metformin and glimepiride in human plasma by LC–MS/MS and its application to a human pharmacokinetic study. J. Pharm. Anal. 2013;3:9–19. doi: 10.1016/j.jpha.2012.09.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Hermann M, Christensen H, Reubsaet JLE. Determination of atorvastatin and metabolites in human plasma with solid-phase extraction followed by LC–tandem MS. Anal. Bioanal. Chem. 2005;382:1242–1249. doi: 10.1007/s00216-005-3266-5. [DOI] [PubMed] [Google Scholar]
  • 50.Farahani H, Norouzi P, Beheshti A, Sobhi HR, Dinarvand R, Ganjali MR. Quantitation of atorvastatin in human plasma using directly suspended acceptor droplet in liquid–liquid–liquid microextraction and high-performance liquid chromatography-ultraviolet detection. Talanta. 2009;80:1001–1006. doi: 10.1016/j.talanta.2009.08.033. [DOI] [PubMed] [Google Scholar]

Articles from Iranian Journal of Pharmaceutical Research : IJPR are provided here courtesy of Brieflands

RESOURCES