Should the mechanical index be revised for ARFI imaging?

Charles C Church 1, Cecille Labuda 1, Kathryn Nightingale 2
PMCID: PMC3924964  NIHMSID: NIHMS450316  PMID: 24533174

Abstract

The mechanical index (MI) quantifies the likelihood that exposure to diagnostic ultrasound will produce an adverse biological effect by a nonthermal mechanism. The current formulation of the MI is based on inertial cavitation thresholds in two liquids, water and blood, as calculated by a formalism assuming very short pulse durations. Although tissue contains a high proportion of water, it is not a liquid but a viscoelastic solid. Further, acoustic radiation force impulse imaging employs high-intensity pulses up to several hundred acoustic periods long. The effect of these differences was studied in water, blood and five representative tissues.

Keywords: Acoustic Radiation Force Imaging, ARFI, bioeffects, inertial cavitation, mechanical index, MI

I. INTRODUCTION

In the United States (US), the acoustic output levels used in diagnostic ultrasound machines approved under the Food and Drug Administration’s (FDA) track 3 process have been limited by federal statute since 1976. The original guidelines were determined based upon the highest output levels existing at that time, and for which no bioeffects had been reported; these limits were not based on scientific evidence related to specific bioeffects due to ultrasound [1]. In 1992, these guidelines were relaxed with the implementation of two new safety indices, the thermal index (TI) and the mechanical index (MI) [2]. These metrics were derived through an effort to relate output guidelines to potential bioeffects. However, the upper limits for acoustic output (Ispta, MI) were also tied to the pre-existing limits, rather than to scientific evidence of bioeffects [1]. Since 1992, acoustic output levels have increased within the context of the newer guidelines [3]. Concurrently, new imaging technologies have been developed that employ unique beam sequences that often approach the upper bounds of current limits, including harmonic imaging modes [4], and acoustic radiation force-based elasticity imaging modes [5][6].

When the MI and TI were first implemented, consideration was given to the question of whether upper limits on acoustic output should be retained by the FDA, or if outputs should be determined via risk-benefit analysis based on the ALARA (as low as reasonably achievable) principle [7]. In 2008, the AIUM issued a consensus report on potential bioeffects of diagnostic ultrasound [1]. In this report, it was recommended that the FDA be encouraged to develop an open, scientifically valid process for assessing the benefits and risks of removing or modifying upward the current regulatory limits. It is widely recognized that many imaging modes may benefit from transient increases in both thermal and non-thermal parameters, particularly in cases where tissue overlying the beam focus is highly attenuating. This is especially true for ARFI imaging modes for which the depth of penetration is limited to 6 – 8 cm at the upper limit on the MI, 1.9. This study is part of an effort to assess the risk in vivo of the relatively long pulses necessary for ARFI imaging.

II. THE MECHANICAL INDEX

The MI is defined as the estimated peak rarefactional pressure in vivo, pr, divided by the square root of the center frequency, fc, of the acoustic beam. The expression is based on an analytical solution obtained by Apfel and Holland [8] for the radial motion of pre-existing air bubbles in water and blood exposed to a wave having a pulse length of a single acoustic period. Apfel and Holland determined the approximate acoustic pressure amplitude required to cause an optimally sized bubble (i.e., the bubble size having the lowest threshold) to undergo inertial cavitation, i.e., a large expansion followed by a rapid, violent collapse. Such a collapse can radiate damaging shock waves and cause the gas within the bubble to attain a very high temperature, 5000 K in the case of the MI, thereby producing large numbers of highly reactive free radicals.

In its current form, the MI uses only two of the many parameters that characterize an acoustic field: the peak rarefactional pressure and the center frequency. Further, the results upon which it is based are valid only for liquids. This work investigates the effects of pulse lengths much longer than the 14-period maximum reported earlier [9], and it also models several tissues using values for elasticity (or more precisely, the shear modulus) and viscosity determined experimentally rather than estimating these values as was done previously [10].

III. METHODS

A. Bubbles in Liquids

The acoustic response of bubbles in liquids was modeled using the Gilmore-Akulichev formulation for bubble dynamics; details of the general approach may be found in [9]:

(1R˙c)RR¨+32(1R˙3c)R˙2=(1+R˙c)H+(1R˙c)RcdHdt, (1)

where R, is the bubble radius, c is the calculated speed of sound in the liquid, H is the enthalpy of the liquid, and t is time; the single and double overdots indicate first and second derivatives with respect to time. Liquids are modeled using a Tait equation of state.

B. Bubbles in Tissues

The acoustic response of bubbles in tissues was modeled using a Keller-Miksis-like equation assuming a gas bubble in a linear Voigt viscoelastic solid [11]:

(1R˙c)RR¨+32(1R˙3c)R˙2=(1+R˙c)papIρ+Rρcddt[papI], (2)

where ρ is the density of the surrounding tissue,

papI=pg0(R0R)3κ2σRp0+PAg(t)[4G3R3(R3R03)+4μR˙R], (3)

σ is the surface tension, p0 is the hydrostatic pressure, PA is acoustic pressure amplitude, g(t) gives the shape of the driving pressure wave, G is modulus of rigidity (rigidity), µ is shear viscosity, and the subscript 0 indicates the initial value of a parameter.

C. Computational Approach

Thresholds for inertial cavitation are determined by performing numerical computations similar to the analytical work underlying the MI. In the work reported here, the acoustic frequency ranges from 0.5 to 10.0 MHz, and the pulse length is varied over the range 1 – 100 acoustic periods. The range of equilibrium bubble radii (R0) studied is 0.1–10.0 µm. Two criteria for the threshold acoustic pressure (Pt) are used: Rmax = 2R0 and Tmax = 5000 K (the criterion used for the MI).

In all cases, it is assumed that: (1) a spherical bubble is initially at rest (i.e., dR/dt = 0) in an infinite material, (2) there is no exchange of gas with the surrounding material, (3) the gas within the bubble pulsates adiabatically, and (4) the bubble radius is small compared to the wavelength. A fourth-order Runge–Kutta technique is used to solve the initial value problem. This formalism is quite similar to that forming the basis for the MI [8], although corrections for liquid compressibility are included, befitting the high pressure amplitudes studied in some cases.

The values for the physical parameters characterizing the materials surrounding the bubbles are given in Table I. The values for tissue were determined by experiments performed in the low- MHz frequency range by Madsen et al. [12] and Yang and Church [13].

Table I.

Material Properties.

Material Density Surface
Tension
Viscosity Rigidity
(kg/m3) (mN/m) (mPa-s) (MPa)
Water 1000 68.0 1.0 0.000
Blood 1050 56.0 5.0 0.000
Heart 1080 56.0 16.0 0.124
Kidney 1100 56.0 5.0 0.180
Liver 1100 56.0 9.0 0.040
Muscle 1100 56.0 7.0 0.450
Skin 1070 56.0 130.0 4.000

The method used to determine the threshold for inertial cavitation, Pt, is essentially the same as that used by Apfel and Holland [8]. For a given combination of parameters, the minimal pressure amplitude required to satisfy the cavitation threshold criterion is determined for each equilibrium radius in the range studied, and the global value for Pt, is then taken from the threshold for the bubble requiring the least acoustic pressure to satisfy the given threshold criterion; that bubble is described as being “optimally sized”. This procedure is repeated for each frequency, pulse length, and material studied.

IV. RESULTS AND DISCUSSION

A. Threshold as a Function of Bubble Radius

Representative results for cavitation threshold as a function of equilibrium bubble radius are given in Figs. 1(a) – (c) for the cases of air bubbles in blood, skeletal muscle and liver, respectively, for an acoustic frequency of 3.0 MHz. Results for the two threshold criteria, Rmax = 2R0 and Tmax = 5000 K, are shown in each figure. Figure 1 illustrates several observations common to most or all of the data sets: (1) the effect of increasing pulse length is to reduce the threshold for most, although not all, bubbles near to, or larger than, the optimally sized bubble, (2) for a given pulse length, bubble radius, and liquid, the threshold pressure for Tmax = 5000 K is always greater than that for Rmax = 2R0, (3) the threshold for either criterion increases with the material’s viscosity under equivalent exposure conditions, (4) for bubbles larger than the optimal size, there are usually several distinct local minima, the result of driven resonance responses, while (5) small bubbles usually exhibit few, and often no, resonance minima.

Figure 1.

Figure 1

Thresholds for inertial cavitation of air bubbles in (A) blood, (B) skeletal muscle and (C) liver at 3 MHz for both threshold criteria; the numbers in the legends indicate the pulse length.

B. Threshold as a Function of Acoustic Frequency

The frequency response was determined for each pulse length by fitting the threshold data to a power law with and without an intercept: Pt = A + Bfcn and Pt = Bfcn, respectively; the latter is that assumed for the MI [2]. Based on values of the correlation coefficient, the simple power law with A = 0 fits the data very well (0.953 < r2 < 1.000), although the power law with an intercept fits somewhat better (0.999 < r2 < 1.000).

Plots of the cavitation threshold vs. frequency are shown in Fig. 2. The computational data are indicated by the symbols, and the curves are the best fits to Pt = Bfcn. The values for the fits shown in Fig. 2 are given in Table II. Notice that the values of the coefficients for the temperature criterion are approximately twice those for the radial criterion, indicating that it takes a more violent collapse to achieve the former endpoint. Similarly, values for the exponent n, are also larger for the temperature criterion, although the differences are not as great. Significantly, the exponent determined for each material is larger than the value assumed for the MI, suggesting that this index may unnecessarily restrict output levels at higher frequencies.

Figure 2.

Figure 2

Thresholds for inertial cavitation of optimally sized air bubbles in each material for threshold criteria of (A) Rmax = 2R0 and (B) Tmax = 5000 K. Curves are the best fits of Pt = Bfcn to the numerical data (points).

Table II.

Parameters for Best Fits of Threshold to Frequency.

Material Rmax = 2R0 Tmax = 5000 K
B n B n
(MPa) (non) (MPa) (non)
Water 0.170 0.451 0.325 0.617
Blood 0.199 0.576 0.440 0.798
Heart 0.427 0.601 1.003 0.875
Kidney 0.382 0.374 0.686 0.644
Liver 0.275 0.599 0.651 0.866
Muscle 0.682 0.285 1.107 0.537
Skin 5.686 0.394 10.916 0.700

C. Threshold as a Function of Acoustic Pulse Length

Because the values of the dimensional thresholds are so different, it is convenient to normalize the threshold at each pulse length to the value at a pulse length of 1 period. The results for normalized threshold Pt,n, are shown in Fig. 3 for both criteria. The data are indicated by the symbols, and the curves are the best fits to Pt,n = Bfcn. The values of the parameters for the fits shown in Fig. 3 are given in Table III.

Figure 3.

Figure 3

Normalized thresholds for inertial cavitation in each material for threshold criteria of (A) Rmax = 2R0 and (B) Tmax = 5000 K at a pulse length of 100 periods. Curves are the best fits of Pt = Bfcn to the numerical data (points).

Table III.

Parameters for Best Fits of Normalized Threshold to Pulse Length.

Material Rmax = 2R0 Tmax = 5000 K
B n B n
(MPa) (non) (MPa) (non)
Water 0.271 0.231 0.811 0.070
Blood 0.572 0.281 0.95 0.027
Heart 0.821 0.112 1.000 0.000
Kidney 0.466 0.342 0.968 0.012
Liver 0.721 0.172 0.995 0.003
Muscle 0.505 0.314 0.975 0.013
Skin 0.941 0.033 1.000 0.000

D. The Final Formulation

Because the two sets of data are both well fit by the same equation, it is convenient to combine the two sets of fits into a single result. The parameters for the final equation, Pt,tis = Bfcn, are given in Table IV. Notice that the exponent on frequency is greater than the 0.5 assumed for the MI, with the values for the more intense threshold criterion being approximately 0.75. This supports the possibility that the frequency response of the MI may be safely modified for certain selected imaging modes requiring greater outputs, e.g., ARFI imaging.

Table IV.

Average Parameters for Best Fits of Normalized Threshold to Pulse Length.

Material Rmax = 2R0 Tmax = 5000 K
B n B n
(MPa) (non) (MPa) (non)
All 0.941 0.681 2.142 0.737
All Tissue 1.090 0.680 2.455 0.746
Soft Tissue 0.237 0.731 0.762 0.755

ACKNOWLEDGEMENTS

This work was supported in part by NIH grant R21EB013763 (CCC, CL) and R01EB002132 and R01CA142824 (KN).

REFERENCE

  • 1.Fowlkes JB, Abramowicz JS, Church CC, Holland C, Miller DL, O’Brien WD, Sanghvi N, Stratmeyer ME, Zachary JF. American Institute of Ultrasound in Medicine Consensus Report on Potential Bioeffects of Diagnostic Ultrasound: Executive Summary. J. Ultrasound Med. 2008;vol. 27:503–515. doi: 10.7863/jum.2008.27.4.503. [DOI] [PubMed] [Google Scholar]
  • 2.AIUM/NEMA, Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment. Washington, DC: American Institute of Ultrasound in Medicine / National Electrical Manufacturers Association; 1992. [Google Scholar]
  • 3.Martin K. The acoustic safety of new ultrasound technologies. Ultrasound. 2010 Jul;vol. 18:110–118. [Google Scholar]
  • 4.Kollmann C. New sonographic techniques for harmonic imaging--underlying physical principles. Eur. J. Radiol. 2007 Nov;vol. 64:164–172. doi: 10.1016/j.ejrad.2007.07.024. [DOI] [PubMed] [Google Scholar]
  • 5.Lazebnik R. Tissue Strain Analytics from Siemens Ultrasound. Siemens Medical Ultrasound Website. http://www.medical.siemens.com/webapp/wcs/stores/servlet/PSGenericDisplay~q_catalogId~e_-1~a_langId~e_-1~a_pageId~e_102707~a_storeId~e_10001.htm.
  • 6.Mendelson EB, Chen J-F, Karstaedt P. Assessing tissue stiffness may boost breast imaging specificity. Diagnostic Imaging. 2009:15–17. [Google Scholar]
  • 7.O’Brien WD, Abbott JG, Stratmeyer ME, Harris GR, Schafer ME, Siddiqi TA, Merritt CRB, Duck FA, Bendick PJ. Acoustic output upper limits proposition: should upper limits be retained? J. Ultrasound Med. 2002 Dec;vol. 21:1335–1341. doi: 10.7863/jum.2002.21.12.1335. [DOI] [PubMed] [Google Scholar]
  • 8.Apfel RE, Holland CK. Gauging the likelihood of cavitation from shortpulse, low-duty cycle diagnostic ultrasound. Ultrasound Med. Biol. 1991;vol. 17:179–185. doi: 10.1016/0301-5629(91)90125-g. [DOI] [PubMed] [Google Scholar]
  • 9.Church CC. Frequency, pulse length and the mechanical index. Acoust. Res. Lett. Online. 2005;vol. 6:162–168. [Google Scholar]
  • 10.Church CC, Yang X. A Theoretical Study of Gas Bubble Dynamics in Tissue. In: Atchley AA, Sparrow VM, Keolian RM, editors. in Innovations in Nonlinear Acoustics: 17th International Symposium on Nonlinear Acoustics. Melville, NY: American Institute of Physics; 2006. pp. 217–224. [Google Scholar]
  • 11.Yang X, Church CC. A model for the dynamics of gas bubbles in soft tissue. J. Acoust. Soc. Amer. 2005;Vol. 118:3595–3606. doi: 10.1121/1.2118307. [DOI] [PubMed] [Google Scholar]
  • 12.Madsen EL, Sathoff HJ, Zagzebski HJ. Ultrasonic shear wave properties of soft tissues and tissue-like materials. J. Acoust. Soc. Amer. 1983;Vol. 74:1346–1355. doi: 10.1121/1.390158. [DOI] [PubMed] [Google Scholar]
  • 13.Yang X, Church CC. A Simple Viscoelastic Model for Soft Tissues in the Frequency Range 6–20 MHz. IEEE Trans. UFFC. 2006 Aug;vol. 53:1404–1411. doi: 10.1109/tuffc.2006.1665097. [DOI] [PubMed] [Google Scholar]