Reliability of the Good Balance System® for Postural Sway Measurement in Poststroke Patients

Hyungeun Ha 1, Kihun Cho 2, Wanhee Lee 1,*
PMCID: PMC3927022  PMID: 24567689

Abstract

[Purpose] The purpose of this study was to examine test-retest reliability of the Good Balance system® for measurement of postural sway in poststroke patients. [Subjects] Sixty chronic stroke patients (40 men and 20 women; age 63.08 years; stroke duration 16.45 months) participated in this study. [Methods] Postural sway was evaluated using a force platform system (Good Balance system, Metitur Oy, Jyvaskyla, Finland). Two examiners measured postural sway for all participants during two separate testing sessions. The second measurement was performed one week after the first measurement. Intraclass correlation coefficients [ICC(2,1)] were used for estimation of reliability. [Results] The ICC (95% CI) for intra-examiner reliability was good to very good, ranging from 0.69 to 0.93 (0.53–0.96), and the ICC for inter-examiner reliability was good to very good, ranging from 0.85 to 0.98 (0.77–0.99). [Conclusion] The results of the current study indicated that the intra- and inter-examiner reliability of the Good Balance system® for measurement of postural sway was good to very good. Therefore, we suggest that measurement of postural sway using the Good Balance system® would be useful for clinical assessment in poststroke patients.

Key words: Postural sway, Reliability, Stroke

INTRODUCTION

Poor postural control during ambulation or activities of daily living (ADL) is a common disability after stroke1). Asymmetric weight bearing in a standing position has been shown to have a negative relationship with independence in performance of ADL2). A previous study demonstrated that humans required three elements in order to control functional standing balance: maintenance of the standing position in static circumstances, maintenance of standing while experiencing internally produced perturbations associated with movements of their extremities, and maintenance of standing while experiencing externally produced perturbations3). Postural control is essential for independence in performance of ambulation and ADL in poststroke patients; therefore, accurate evaluation of the postural control system is important4).

In general, traditional balance measures, such as the berg balance scale5), timed up and go test6), and functional reaching test7), are most commonly used as assessment tools for evaluation of postural control ability in stroke patients. However, these tools cannot be regarded as quantitative and objective methods of measurement8). Recently, in the clinic or laboratory, computerized measurement and feedback systems have been used for evaluation of both static and dynamic balance control abilities9, 10). Among the various types of equipment used for measurement of postural control, force platform technology can provide a method for quantifying an individual’s postural control ability by measuring the center of pressure11). In particular, the Good Balance system® consists of an equilateral triangular force platform connected to a computer through a three-channel amplifier with an analog to digital (A/D) converter. The following variables are calculated in the Good Balance system®: the extent of mediolateral (ML) movement of the center of pressure (COP) (X movement), the extent of anteroposterior (AP) movement of the COP (Y movement), and the mean values for all of the measurement points in relation to the midline of the platform (lateral displacement).

Despite use of the Good Balance system® in many studies as equipment for measurement of postural control ability10, 12, 13), there is no evidence regarding the reliability of the Good Balance system® in stroke patients. The reliability of measurement using the Good Balance system® need to be determined before it can be used as a potential predictor of functional performance. Thus, the purpose of this study was to examine the test-retest reliability of the Good Balance system® for measurement of postural sway in poststroke patients. We hypothesized that postural sway while maintaining postural control in a standing posture has adequate reliability for research and clinical use in stroke patients.

SUBJECTS AND METHODS

This single-group repeated-measures design involved a baseline measurement session and a follow-up session one week later. Subjects were recruited from patients admitted for rehabilitation following stroke. Subjects were included if they had (1) hemiparesis resulting from stroke for more than six months; (2) were able to understand and follow simple verbal instructions (Korean version of the Mini-Mental State Examination score > 21); (3) had no known musculoskeletal conditions that would affect the ability to stand safely; and (4) had no serious visual impairment or hearing disorder. Potential subjects were excluded if they had cardiovascular or other conditions affecting their balance and were unable to provide informed consent. All subjects underwent conventional rehabilitation programs consisting of physical and occupational therapy during the duration of hospitalization. General characteristics of the subjects were obtained from their medical records. Six of the 66 potential subjects were excluded because they met the exclusion criteria. Finally, 60 subjects were included in this study. We explained the objective and requirements of our study to all participants, and they voluntarily signed informed consent forms. Ethical approval for the study was granted by the Sahmyook University institutional review board.

The Good Balance system® (Good Balance system, Metitur Oy, Jyvaskyla, Finland) was used for measurement of postural sway velocity and velocity moment of subjects in the standing posture. On the basis of these coordinate values for x and y, the following parameters were calculated: (1) mean speed of the movement of the COP in the AP direction (mm/s); (2) mean speed of the movement of the COP in the ML direction (mm/s), and (3) mean velocity moment (mm2/s). Postural sway velocity moment is defined as the average horizontal area covered by movement of the center (AP and ML direction) of force per second. The force platform, which was an equilateral triangle (800 mm), was connected to a three-channel DC amplifier. Signals from the amplifier were converted into digital form using a 12-byte converter (sampling frequency=50 Hz) and stored on the hard disk of a personal computer.

To measure postural sway, a subject stood on the force plate with their legs spread at shoulder width and then looked at a number on a monitor for 30 seconds. In order to measure postural sway, the subjects were asked to stand quietly in a comfortable upright position on the force plate while looking straight ahead. All tests were performed with shoes removed. According to the user manual, a distance of between 1 to 3 m is suitable for accurate measurements. In our study, the distance between the subject and the monitor was set 1.5 m. The following instruction was communicated to the subjects in order to ask them to move their bodies as little as possible: “Please try your best to stand without swaying.”

Two physical therapists participated as examiners in the reliability analysis based on previous evidence14). Examiners acquired data from all participants during two separate testing sessions (sessions one and two). Three repeats of each measurement sessions were performed, and the average was used in each session. A rest period (three minutes) was provided between measurements in order to prevent fatigue. After the first measurement was performed, the second measurement was performed one week later. Testing sessions were held at the same time of day for each participant.

Intraclass correlation coefficients [ICC(2,1)] were used to determine the test-retest reliability for postural sway between the two measurement sessions. ICC values less than or equal to 0.20 were considered poor; 0.21 to 0.40, 0.41 to 0.60, 0.61 to 0.80, and 0.81 to 1.00 were considered to be fair, moderate, good, and very good values, respectively15). The paired t-test was performed in order to test for any systematic differences between sessions. Results were considered significant at p<0.05, and statistical analyses were performed using SPSS ver. 12.0 (SPSS for Windows; SPSS Inc, Chicago, IL, USA).

RESULTS

A summary of the general characteristics of the subjects who completed the measurements successfully is shown in Table 1. A summary of the results for intra-examiner reliability of postural sway for the two sessions is shown for the two examiners in Table 2. The ICC (95% CI) for all measures was good to very good, ranging from 0.69 to 0.93 (0.53–0.96). A summary of the results for inter-examiner reliability of postural sway for the two sessions between examiners is shown in Table 3. The ICC (95% CI) was good to very good for all measures, ranging from 0.85 to 0.98 (0.77–0.99).

Table 1. General characteristics of the subjects.

Variables Male (n=40) Female (n=20) Overall (n=60)
Paretic side Left/Right (%) 11/29 (27.5/72.5) 5/15 (25/75) 16/44 (26.2/72.1)
Etiology Infarction/Hemorrhage (%) 24/16 (60/40) 13/7 (65/35) 37/23 (60.7/37.7)
Age (years) 63.1±6.2 63.0±6.7 63.1±6.3
Height (cm) 165.3±6.8 157.4±4.2 162.7±7.1
Weight (kg) 62.3±8.0 56.3±5.6 60.3±7.8
Onset duration (months) 16.6±3.0 16.3±2.4 16.4±2.8
BBS (scores) 38.8±4.8 39.3±3.5 38.9±4.4
TUG (sec) 24.1±3.8 24.1±3.7 24.1±3.7
MMSE-K (scores) 25.7±2.8 26.1±2.8 25.9±2.8

Values are expressed as n (%) or as the mean±SD. BBS, Berg balance scale; TUG, Timed up and go test; MMSE-K, mini mental state examination-Korean version

Table 2. Intra-examiner reliability of postural sway (n=60).

Variables Session 1 Session 2 ICC 95% CI
E1 Velocity (mm/s) AP 8.5±4.4 8.0±4.6 0.87 0.79–0.92
ML 13.1±4.5 12.7±4.9 0.69 0.53–0.80
Velocity moment (mm2) 36.8±21.0 37.1±21.0 0.87 0.79–0.92
E2 Velocity (mm/s) AP 8.8±4.1 8.4±4.1 0.93 0.89–0.96
ML 13.3±4.1 13.0±4.5 0.78 0.66–0.86
Velocity moment (mm2) 37.5±20.2 36.5±19.2 0.93 0.89–0.95

Values are expressed as the mean±SD. E, examiner; AP, anteroposterior; ML, mediolateral; ICC, intraclass correlation coefficient; 95% CI, 95% confidence interval

Table 3. Inter-examiner reliability of postural sway (n=60).

Variables 1st test 2nd test
ICC 95% CI ICC 95% CI
Velocity (mm/s) AP 0.97 0.95–0.98 0.92 0.88–0.95
ML 0.92 0.87–0.95 0.85 0.77–0.91
Velocity moment (mm2) 0.96 0.94–0.98 0.96 0.93–0.97

AP, anteroposterior; ML, mediolateral; ICC, intraclass correlation coefficient; 95% CI, 95% confidence interval

DISCUSSION

This study was conducted in order to determine the suitability of the Good Balance system® for measurement of postural sway in poststroke patients. Our main results demonstrate good to very good inter- and intra-reliability of the Good Balance system® for estimating a stroke patient’s postural sway measurement.

Postural sway measurements using a force plate record vertical strength and can be used to explain postural control ability using parameters such as postural sway velocity and velocity moment16). The amount of postural sway is usually divided into the anteroposterior direction and the mediolateral direction17). In general, postural sway is known to increase with age, and the frequency of falls increases as sway increases18). In addition, postural sway is 1–1.5 times higher in stroke patients than in elderly people19). The ability to maintain body balance against externally produced perturbations is achieved by the postural control system. Thus, impairment in this system could lead to increased postural sway as observed in stroke patients20). In other words, accurate evaluation for postural sway in stroke patients is important for successful rehabilitation and return to the home and community.

Many previous studies targeting elderly people have reported good reliability of postural sway measured using a force platform. For example, Lafond et al.21) assessed the intra-session reliability of COP measurement of postural steadiness in elderly. They reported that mean velocity was the most reliable measure of COP. In addition, Benvenuti et al.22) found good reliability (ICC=0.74) based on two consecutive trials for COP mean velocity. However, the reliability of postural sway outcome in stroke patients has been addressed in only a few studies, and the results were not consistent. Liston et al.4) investigated reliability of measures obtained from 20 stroke patients using the Balance Master and reported its reliability for both movement path (ICC=0.84) and time (ICC=0.88). In contrast, Helbostad et al.23) reported that two repeated measures gave adequate repeatability for elderly people but not for stroke patients when they were tested with their eyes open. Because these findings were obtained from relatively small sample size (12 and 23 stroke patients), the results of these studies are not unequivocal. Therefore, in the current study, we investigated the reliability in a larger sample size (60 stroke patients), and the results demonstrated good to very good intra- and inter-examiner reliability of the Good Balance system® for measurement of postural sway. Therefore, we suggest that measurement of postural sway using the Good Balance system® would be useful for clinical assessment in poststroke patients.

The current study showed that the measurements of postural sway obtained from the Good Balance system® were reliable in the intra- and inter-reliability analysis, but there are some limitations. In the current study, we determined that the Good Balance system® was reliable equipment for measurement of postural sway. However, the findings of previous studies targeting stroke patients were not consistent; therefore, there is still the potential for controversy. Thus, we suggest that further study on the reliability of postural sway measurements using a force plate targeting stroke patients is needed. Another limitation of the current study is the small number of measurements for each session (only three measurements). Conduct of a larger number of trials might contribute to the reliability of postural sway measurement. Thus, conduct of additional studies on this issue is required.

REFERENCES

  • 1.Jørgensen L, Engstad T, Jacobsen BK: Higher incidence of falls in long-term stroke survivors than in population controls: depressive symptoms predict falls after stroke. Stroke, 2002, 33: 542–547 [DOI] [PubMed] [Google Scholar]
  • 2.Marigold DS, Eng JJ, Tokuno CD, et al. : Contribution of muscle strength and integration of afferent input to postural instability in persons with stroke. Neurorehabil Neural Repair, 2004, 18: 222–229 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Berg K: Measuring balance in the elderly: a review. Physiother Can, 1989, 41: 240–246 [Google Scholar]
  • 4.Liston RA, Brouwer BJ: Reliability and validity of measures obtained from stroke patients using the Balance Master. Arch Phys Med Rehabil, 1996, 77: 425–430 [DOI] [PubMed] [Google Scholar]
  • 5.Blum L, Korner-Bitensky N: Usefulness of the Berg Balance Scale in stroke rehabilitation: a systematic review. Phys Ther, 2008, 88: 559–566 [DOI] [PubMed] [Google Scholar]
  • 6.Ng SS, Hui-Chan CW: The timed up & go test: its reliability and association with lower-limb impairments and locomotor capacities in people with chronic stroke. Arch Phys Med Rehabil, 2005, 86: 1641–1647 [DOI] [PubMed] [Google Scholar]
  • 7.Katz-Leurer M, Fisher I, Neeb M, et al. : Reliability and validity of the modified functional reach test at the sub-acute stage post-stroke. Disabil Rehabil, 2009, 31: 243–248 [DOI] [PubMed] [Google Scholar]
  • 8.Hatch J, Gill-Body KM, Portney LG: Determinants of balance confidence in community-dwelling elderly people. Phys Ther, 2003, 83: 1072–1079 [PubMed] [Google Scholar]
  • 9.Chen IC, Cheng PT, Chen CL, et al. : Effects of balance training on hemiplegic stroke patients. Chang Gung Med J, 2002, 25: 583–590 [PubMed] [Google Scholar]
  • 10.Cho KH, Lee KJ, Song CH: Virtual-reality balance training with a video-game system improves dynamic balance in chronic stroke patients. Tohoku J Exp Med, 2012, 228: 69–74 [DOI] [PubMed] [Google Scholar]
  • 11.Barclay-Goddard R, Stevenson T, Poluha W, et al. : Force platform feedback for standing balance training after stroke. Cochrane Database Syst Rev, 2004, CD004129 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Sihvonen SE, Sipila S, Era PA: Changes in postural balance in frail elderly women during a 4-week visual feedback training: a randomized controlled trial. Gerontology, 2004, 50: 87–95 [DOI] [PubMed] [Google Scholar]
  • 13.Sihvonen S, Sipila S, Taskinen S, et al. : Fall incidence in frail older women after individualized visual feedback-based balance training. Gerontology, 2004, 50: 411–416 [DOI] [PubMed] [Google Scholar]
  • 14.Cho KH, Lee HJ, Lee WH: Reliability of rehabilitative ultrasound imaging for the medial gastrocnemius muscle in poststroke patients. Clin Physiol Funct Imaging, 2013, Jun: 12.Epub ahead of print. [DOI] [PubMed] [Google Scholar]
  • 15.Mota P, Pascoal AG, Sancho F, et al. : Test-retest and intrarater reliability of 2-dimensional ultrasound measurements of distance between rectus abdominis in women. J Orthop Sports Phys Ther, 2012, 42: 940–946 [DOI] [PubMed] [Google Scholar]
  • 16.Johansson R, Magnusson M: Human postural dynamics. Crit Rev Biomed Eng, 1991, 18: 413–437 [PubMed] [Google Scholar]
  • 17.Piirtola M, Era P: Force platform measurements as predictors of falls among older people − a review. Gerontology, 2006, 52: 1–16 [DOI] [PubMed] [Google Scholar]
  • 18.Fernie GR, Gryfe CI, Holliday PJ, et al. : The relationship of postural sway in standing to the incidence of falls in geriatric subjects. Age Ageing, 1982, 11: 11–16 [DOI] [PubMed] [Google Scholar]
  • 19.de Haart M, Geurts AC, Huidekoper SC, et al. : Recovery of standing balance in postacute stroke patients: a rehabilitation cohort study. Arch Phys Med Rehabil, 2004, 85: 886–895 [DOI] [PubMed] [Google Scholar]
  • 20.Connell LA, Lincoln NB, Radford KA: Somatosensory impairment after stroke: frequency of different deficits and their recovery. Clin Rehabil, 2008, 22: 758–767 [DOI] [PubMed] [Google Scholar]
  • 21.Lafond D, Corriveau H, Hebert R, et al. : Intrasession reliability of center of pressure measures of postural steadiness in healthy elderly people. Arch Phys Med Rehabil, 2004, 85: 896–901 [DOI] [PubMed] [Google Scholar]
  • 22.Benvenuti F, Mecacci R, Gineprari I, et al. : Kinematic characteristics of standing disequilibrium: reliability and validity of a posturographic protocol. Arch Phys Med Rehabil, 1999, 80: 278–287 [DOI] [PubMed] [Google Scholar]
  • 23.Helbostad JL, Askim T, Moe-Nilssen R: Short-term repeatability of body sway during quiet standing in people with hemiparesis and in frail older adults. Arch Phys Med Rehabil, 2004, 85: 993–999 [DOI] [PubMed] [Google Scholar]

Articles from Journal of Physical Therapy Science are provided here courtesy of Society of Physical Therapy Science