Skip to main content
PLOS One logoLink to PLOS One
. 2014 Mar 3;9(3):e90690. doi: 10.1371/journal.pone.0090690

Gold Nanorods as a Contrast Agent for Doppler Optical Coherence Tomography

Bo Wang 1,2,3, Larry Kagemann 1,2, Joel S Schuman 1,2,*,¤, Hiroshi Ishikawa 1,2, Richard A Bilonick 1,4, Yun Ling 1,4, Ian A Sigal 1,2, Zach Nadler 1, Andrew Francis 1, Michelle G Sandrian 1,2, Gadi Wollstein 1
Editor: Sanjoy Bhattacharya5
PMCID: PMC3940929  PMID: 24595044

Abstract

Purpose

To investigate gold nanorods (GNRs) as a contrast agent to enhance Doppler optical coherence tomography (OCT) imaging of the intrascleral aqueous humor outflow.

Methods

A serial dilution of GNRs was scanned with a spectral-domain OCT device (Bioptigen, Durham, NC) to visualize Doppler signal. Doppler measurements using GNRs were validated using a controlled flow system. To demonstrate an application of GNR enhanced Doppler, porcine eyes were perfused at constant pressure with mock aqueous alone or 1.0×1012 GNR/mL mixed with mock aqueous. Twelve Doppler and volumetric SD-OCT scans were obtained from the limbus in a radial fashion incremented by 30°, forming a circular scan pattern. Volumetric flow was computed by integrating flow inside non-connected vessels throughout all 12 scans around the limbus.

Results

At the GNR concentration of 0.7×1012 GNRs/mL, Doppler signal was present through the entire depth of the testing tube without substantial attenuation. A well-defined laminar flow profile was observed for Doppler images of GNRs flowing through the glass capillary tube. The Doppler OCT measured flow profile was not statistically different from the expected flow profile based upon an autoregressive moving average model, with an error of −0.025 to 0.037 mm/s (p = 0.6435). Cross-sectional slices demonstrated the ability to view anterior chamber outflow ex-vivo using GNR-enhanced Doppler OCT. Doppler volumetric flow measurements were comparable to flow recorded by the perfusion system.

Conclusions

GNRs created a measureable Doppler signal within otherwise silent flow fields in OCT Doppler scans. Practical application of this technique was confirmed in a constant pressure ex-vivo aqueous humor outflow model in porcine eyes.

Introduction

The ability to image flow, such as using Doppler ultrasound and laser flowtometry, has impacted our understanding of physiology [1] and medical management [2], [3]. However, the ability to image low volume flow fields with existing technology is still limited. It would be advantageous if these flow fields could be made visible, facilitating measurement, without perturbation. Doppler ultrasound is limited by its resolution, which dependent on the frequency of the ultrasound pulse. Even high frequency ultrasound can only achieve lateral and axial resolution of ∼50 µm [4]. Meanwhile, laser flowtometry lacks the ability to measure depth information regarding the outflow vessels of interest [5].

Optical coherence tomography (OCT) is an imaging technique that creates high-resolution structural scans of optically reflecting tissues using low-coherence interferometry [6]. Advances in acquisition time makes OCT a practical device for detecting flows [7]. OCT's ability to detect flow with high spatial resolution had been demonstrated in a variety of tissues, including the human retina [8], skin [9], and the heart [10].

As with other Doppler imaging technologies, only fluids containing scattering media at the wavelength of the OCT light source can exhibit Doppler shifts. In some instances physiologically relevant flows, such as that of the aqueous humor, do not show a distinct Doppler signal. Aqueous humor has particular relevance to glaucoma because the balance between its production and outflow determined intraocular pressure (IOP). Elevated IOP had been demonstrated to be the main risk factor for the development of glaucoma, an optic neuropathy that is a leading cause of irreversible blindness worldwide [11]. While not useful for assessing IOP, the addition of GNRs to the outflow pathway may allow visualization of Doppler signal within the normally silent aqueous outflow pathways. The ability to image aqueous outflow could improve our understanding of outflow patterns and glaucoma pathophysiology.

Nanoparticles have been increasingly used as a structural [12] and functional [13], [14] contrast agent in branches of medicine. We have previously demonstrated the ability to enhance structural OCT signals using intravitreal GNR injections [15]. However, despite the use of GNRs [12], [15], [16] and other contrast agents [17][19] to enhance structural OCT imaging, little work had been done to explore their ability to enhance functional measurements such as Doppler. This manuscript describes the use of GNRs as a contrast agent for Doppler OCT that allows ex-vivo imaging of the outflow pathway of porcine eyes.

Methods

Titration

A concentrated polyelectrolyte sodium salt-coated plasmon resonant GNR with a localized surface plasmon resonance of 850 nm and aspect ratio of 4 (∼40 nm×10 nm) were created using a seed-mediated, surfactant directed wet chemical synthesis (Nanorods, Germantown, MD). The polystyrene sulfonate (PSS) coating was used to neutralize the positively charged surfactants used in GNR synthesis. [15] The GNRs were titrated with deionized water to 5 different concentrations: 0.3, 0.4, 0.5, 0.7 and 1.0×1012 GNRs/mL. The addition of water or mock aqueous does not alter the spectrophotometry profile of the GNRs (Figure S1). Doppler signal was subjectively assessed in a glass capillary tube of radius 0.32 mm using the different dilutions. The tube size was chosen to maximize our ability to see Doppler flow throughout a region that would be larger than any aqueous channel.

OCT scanning

Doppler scans (2×3×2 mm, 700×20×1024 pixels) were acquired using spectral-domain (SD-) OCT (Bioptigen, Durham, NC) while the fluid was stationary within the capillary tube. The OCT had a 100 nm bandwidth light source centered on 870 nm, yielding a theoretical axial resolution of 3.5 µm. 12 Doppler frames were acquired with an acquisition rate of 28 kHz, resulting in a Nyquist limit of 4.62 mm/s normal to the scanning beam. The conversion of OCT Doppler signal to angle-corrected velocity is specified in the supporting materials.

Validation

A GNR solution mixed with deionized water was scanned as it flowed through the capillary tube. The glass tube was suspended 45 degrees to the direction of the OCT scanning beam. The GNR solution was pumped at a volumetric rate of 0.56 µL/s with a constant pressure perfusion system and was given 10 minutes to reach a steady state. The rate of perfusion was recorded using a DATAQ acquisition unit (DATAQ Instruments, Akron, OH) and Doppler scans (2×0.4×2 mm, 700×20×1024 pixels) were acquired using SD-OCT. The flow profile recorded by Doppler OCT was compared to a theoretical flow profile assuming laminar flow given a volumetric flow rate of 0.56 µL/s. The two measurements were compared using an autoregressive-moving-average (ARMA) model. A negative log difference was used to reduce the expected autocorrelation between flow values in adjacent pixels.

Ex-vivo porcine model

We used ethically sourced ex-vivo porcine eyes from the local abattoir (Thoma's Meat Market, 748 Dinnerbell Road, Saxonburg, PA 16056). Permission was obtained from the abattoir to use the animal parts for research. Therefore, no IACUC approval was required for these experiments. Two cohorts of ex-vivo porcine eyes were warmed to 40°C and perfused in the anterior chamber with (1) Barany's mock aqueous [20] alone (control; n = 6) or (2) 1×1012 GNRs/mL solution with mock aqueous (n = 2). The eyes were scanned within 16 hours of death. A mix of GNR and mock aqueous was tested using spectrophotometry to insure that the functional properties of the nanorods were not altered.

A 27-gauge perfusion needle was inserted through the cornea, between the lens and the iris into the posterior chamber (Figure 1). The site of insertion of the needle was to the nasal side for all eyes. A trial eye demonstrated that injecting GNR solution alone was not sufficient to image the outflow system because the mixture of the injected GNR with aqueous humor decreased GNR concentration. Therefore, a 30-gauge needle was inserted to the temporal side to withdraw aqueous humor during initial transfer of GNRs solution. GNR solutions flowed under constant pressure for 10 minutes to allow outflow to stabilize to a constant (Figure S2). Doppler scans were taken over a 4×4×2 mm area (700×20×1024 pixels) around the anterior chamber of the eye. Twelve Doppler (700×20×1024 pixels) and volumetric scans (one per clock hour) were taken around the limbus with a scan size of 4×4×2 mm. Scans were oriented in a radial fashion incremented by 30° from scan to scan.

Figure 1. Experimental setting for ex-vivo imaging of anterior chamber outflow in porcine eyes.

Figure 1

A needle attached to a constant pressure perfusion system, whose flow is measured by the flow recorder, is inserted between the lens and iris in the posterior chamber. 12 OCT scans (red box) were taken along the limbus of the eye(red arrow). Dashed lines shows the equipment indicated by the boxes.

Volumetric flow rates were calculated by integrating all flow in a circular pattern around the limbus. Individual non-connected vessels were identified on 3D structural scans. The angle between the scanning beam and vessel was computed from a vector formed by vessel centroids as seen on structural images. The angle was used to calculate the angle-corrected velocity. The volumetric flow through the vessel was determined by integrating the angle-corrected velocity and over the region of the vessel. Total volumetric flow was calculated as the sum of observed flow around the limbus by a masked observer.

Results

Titration

Doppler signal was observed in the testing tube across all GNRs dilutions (Figure 2). A concentration of 1.0×1012 GNRs/mL yielded an optimal Doppler signal with only a moderate increase in Doppler noise as the A-scan penetrates further in the solution, and accordingly this concentration was chosen for all further testing. All stationary Doppler scans showed a systematic alternating pattern in the Doppler signal that was attributed to room or galvonometer vibrations during the scanning procedure.

Figure 2. Determining optimal GNR concentration.

Figure 2

(A–E) Doppler OCT cross-sectional images of non-flowing gold nanorods (GNR) at (A) 0.3, (B) 0.4, (C) 0.5, (D) 0.7 and (E) 1.0×1012 GNRs/mL. The grayscale mapping represents the detected flow towards or away from the OCT scanning beam, centered on 0 mm/s (gray). (F) GNR enhanced saline flowing in a glass tube with a concentration of 1.0×1012 GNRs/mL. The color map represents flow towards (red), away (blue) and stationary (black) relative to the OCT scanning beam. (D–F) shows the ability to get Doppler signal with a tube radius of up to 0.32 mm. The red circle corresponds to the outline of the inner surface of the tube. Blue and red corresponded to flow away and towards the probe, respectively. The area of hyper reflective convex arch within F is an imaging artifact.

Validation

OCT Doppler cross-sectional scans of testing tube with constant flow of 0.56 µL/s with 1.0×1012 GNRs/mL solution showed laminar flow (Figure 3A). A theoretical velocity profile within the tube was calculated based upon the flow measured by the perfusion system. The difference between velocity profile assessed by Doppler OCT and the theoretical flow profile given a constant flow of 0.56 µL/s showed no statistically significant difference. The difference was centered at 0.192±0.498 mm/s with an ARMA intercept ranging between −0.025 to 0.037 mm/s (p = 0.6435). A difference between the theoretical and recorded of flow rate whose confidence interval contains 0 indicates that OCT was able to properly measure GNR-enhanced Doppler.

Figure 3. Assessing Doppler velocity measurement.

Figure 3

Angle corrected velocity profile acquired within a glass capillary tube containing 1.0×1012 gold nanorods/mL with the computed velocity profile overlain in red.

Ex-vivo porcine model

Spectrophotometry reading of GNRs mixed with mock aqueous showed characteristic peaks at 850 nm and 500 nm, similar to the profile obtained from the innate GNR. Using mock aqueous alone, the Doppler signal in the pig anterior chamber was not measurable (Figure 4A). All vessels perfused with mock aqueous appeared dark on the structural scan, showing little reflectivity. The addition of 1.0×1012 GNRs/mL solution increased the contrast within the aqueous outflow vessels and showed clear Doppler signals (Figure 4E). Velocity profile of the Doppler scan showed laminar flow within the vessels (Figure 4F). The volumetric flow rate measured by the reservoir system was 3.42 and 7.02 µl/min in each eye. The rate computed by SD-OCT Doppler for the two eyes was 4.41 and 7.31 µl/min, respectively.

Figure 4. Ex-vivo Doppler imaging.

Figure 4

Structural and Doppler OCT scans of the outflow pathway in porcine eyes perfused with (A–B) mock aqueous (control, Aq) and (C–D) 1×1012 gold nanorods/mL with mock aqueous (GNR+Aq). (E) Merged structural and Doppler scan corresponding to the location in the green box. Red and blue corresponds to flow away and towards the probe, respectively. (F) Velocity profile of the Doppler scans in the location corresponding to the red dotted lines.

Discussion

This study demonstrated the first use of GNRs for enhancing functional ophthalmic imaging using OCT. An optimal concentration of GNRs was identified to maximize penetration without losing signal in a system with fixed flow. The validity of the flow measurement using GNR enhanced Doppler OCT was confirmed in an experiment with known flow parameters. Finally, the use of GNRs and Doppler OCT was demonstrated ex-vivo in porcine eyes, providing the first Doppler OCT scans of intrascleral aqueous outflow.

As OCT scanning beam traverse through a medium, scattering and absorption reduced the signal-to-noise ratio of the system. The tradeoff between signal strength and depth of penetration played an important factor in the optimal GNR concentration. Low concentrations of GNR resulted in poor Doppler imaging because it did not induce enough Doppler signal after scattering loss, creating the inhomogeneous Doppler signal (Figure 2A–C). Increasing GNR concentration improved the quality of Doppler OCT. However, structural scanning experienced lower depth of penetration due to increased scattering from the solution. The optimal concentration of GNRs for enhanced Doppler signal without the creation of obstructive shadows was between 0.7×1012–1.0×1012 GNRs/ml (Figure 2).

We used a controlled flow model to validate the GNR enhanced Doppler signals measurements. Comparison between the computed theoretical velocity profile and Doppler OCT velocity profile showed a normal distribution difference centered around zero, confirming the accuracy of Doppler flow measurements.

We demonstrated GNRs' potential to enhance Doppler OCT signals by using the ex-vivo porcine model (Figure 4). Due to the small size of vessels present in the anterior chamber outflow pathway, we used a 1.0×1012 GNRs/mL solution. We expect the solution to mix with the remaining aqueous humor remaining in the anterior chamber, reducing its concentration. Even if the concentration remains unchanged, we still expect strong Doppler signal within the outflow vessels because the vessels are smaller than the glass capillary and thus experience less signal loss with depth. The experiment was carried out on cadaveric eyes, whose distal outflow pathway may have been blocked at certain sections due to coagulation or other factors. This may explain why Doppler signal was seen only in some outflow vessels but not others.

By integrating flow over the Doppler scans around the limbus, we were able to calculate volumetric aqueous flow in the eyes. The integration of the Doppler scans showed a slight overestimation of the total flow in both eyes. As seen in Figure 5, the directions of the vessels were found by fitting a line through the centroid of the vessels from adjacent scans. The angle between the OCT scanning beam and the line through the vessel centroid was used to determine the angle-adjusted flow rate. The overestimation might be due to the nearly perpendicular angle (60–80°) between the scanning beam and the aqueous outflow vessels. Therefore, even small differences in incident angle could have large effects on the flow calculations. Moreover, the vessels were traced in the 3D structural scans to determine their pathway. Highly tortuous vessels that showed as disconnected sections in the 3D scan could lead to an overestimation of the total flow if the same tortuous vessels is sampled multiple times. Furthermore, we cannot rule out that the Doppler scans did not account for the uveoscleral flow, which might further expand the reported difference total flow and the Doppler computed flow. However, this work provided proof of concept that GNRs could eventually be used to examine aqueous outflow.

Figure 5. Determining volumetric flow.

Figure 5

Adjacent B-scan images containing the vessels of interest are identified and marked (dashed white line) to determine the centroid. A line (white line) was fitted through the centroid to determine the angle of the vessel, θ, with respect to the OCT imaging beam (blue arrow). This allows us to integrate the flow within the Doppler information (red) and correct for the angle of flow with respect to the OCT scanning beam.

While the experiments were conducted ex-vivo, a number of hurdles remain in the in-vivo use of GNRs. Current methods of inserting GNRs into the eye are invasive. A non-invasive delivery vehicle would help make GNRs more useful for scanning the outflow system in-vivo. Theoretically, the GNRs were small enough to pass through the cornea [16]. However, GNR eye drops were not used in this experiment because they would limit our ability to control the concentration within the eye. Another potential limitation of the GNR was the inflammatory response they induce in-vivo [15]. Further in-vivo studies are necessary to determine optimal coating for the GNRs to prevent such an inflammatory response. In addition, it is unknown how the excess GNRs are cleared, such as through macrophage induced phagocytosis, and in-vivo studies are necessary to insure that they do not permanently impede vision.

Improvements in Doppler imaging will also permit new possibilities for anterior segment imaging. Specifically, the use of swept-source lasers will permit superior imaging to the deeper regions of the angle, including schlemn's canal. Furthermore, advances in Doppler imaging, such as dual beam Doppler OCT [17], [18], permit Doppler measurements without the dependence on incidence angle of single beam systems.

To our knowledge, this represented the first use of OCT to visualize aqueous outflow. Previous studies on aqueous outflow primarily involved the use of fluoroscein [19]. However, fluorescein could only measure aqueous loss indirectly through decreasing fluorescein concentration. GNR enhanced OCT Doppler scanning might lead to increased understanding of how outflow changes in glaucomatous eyes. Moreover, the ability to relate structural findings with functional outflow measurements recorded simultaneously might be a useful tool in ocular research.

Conclusion

This study demonstrated the use of gold nanorods (GNRs) for enhancing Doppler OCT. Doppler OCT measurements were validated in experimental control flow and demonstrated ex-vivo in intrascleral aqueous pathway in porcine eyes.

Supporting Information

Table S1

Laminar Flow.

(DOCX)

Figure S1

Spectrophotometry of GNR in various media. Spectrophotometry of (1) mock aqueous humor (mock AH) alone, (2) 2×1012 gold nanorods/ml (GNR), (3) GNR mixed with distilled water and (4) GNR mixed with mock aqueous humor.

(TIFF)

Figure S2

GNR flow during imaging. Volume remaining in the reservoir over time for eye 1. A linear slope indicates a constant flow rate.

(TIFF)

Funding Statement

This study was supported in part by the American Health Assistance Foundation (AHAF); National Institute of Health R21-019092, R01- 013178, T32-008208, P30-008098; Eye and Ear Foundation (Pittsburgh, PA); Research to Prevent Blindness (New York, NY). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1. Ernst H, Hahn EG, Balzer T, Schlief R, Heyder N (1996) Color Doppler ultrasound of liver lesions: Signal enhancement after intravenous injection of the ultrasound contrast agent Levovist. Journal of Clinical Ultrasound 24: 31–35. [DOI] [PubMed] [Google Scholar]
  • 2. Dirnagl U, Kaplan B, Jacewicz M, Pulsinelli W (1989) Continuous measurement of cerebral cortical blood flow by laser-Doppler flowmetry in a rat stroke model. Journal of Cerebral Blood Flow & Metabolism 9: 589–596. [DOI] [PubMed] [Google Scholar]
  • 3. Aaslid R, Markwalder T-M, Nornes H (1982) Noninvasive transcranial Doppler ultrasound recording of flow velocity in basal cerebral arteries. Journal of Neurosurgery 57: 769–774. [DOI] [PubMed] [Google Scholar]
  • 4. Goertz DE, Christopher DA, Yu JL, Kerbel RS, Burns PN, et al. (2000) High-frequency color flow imaging of the microcirculation. Ultrasound in Medicine & Biology 26: 63–71. [DOI] [PubMed] [Google Scholar]
  • 5. Rajan V, Varghese B, van Leeuwen TG, Steenbergen W (2009) Review of methodological developments in laser Doppler flowmetry. Lasers Med Sci 24: 269–283. [DOI] [PubMed] [Google Scholar]
  • 6. Huang D, Swanson EA, Lin CP, Schuman JS, Stinson WG, et al. (1991) Optical coherence tomography. Science 254: 1178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Izatt JA, Kulkarni MD, Yazdanfar S, Barton JK, Welch AJ (1997) In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography. Optics Letters 22: 1439–1441. [DOI] [PubMed] [Google Scholar]
  • 8. Leitgeb R, Schmetterer L, Drexler W, Fercher A, Zawadzki R, et al. (2003) Real-time assessment of retinal blood flow with ultrafast acquisition by color Doppler Fourier domain optical coherence tomography. Opt Express 11: 3116–3121. [DOI] [PubMed] [Google Scholar]
  • 9. Zhao Y, Chen Z, Saxer C, Xiang S, de Boer JF, et al. (2000) Phase-resolved optical coherence tomography and optical Doppler tomography for imaging blood flow in human skin with fast scanning speed and high velocity sensitivity. Opt Lett 25: 114–116. [DOI] [PubMed] [Google Scholar]
  • 10.Yazdanfar S, Kulkarni MD, Izatt JA (1998) High resolution imaging of in vivo cardiac dynamics using color Doppler optical coherence tomography. Advances in Optical Imaging and Photon Migration. [DOI] [PubMed]
  • 11. Quigley HA (1996) Number of people with glaucoma worldwide. British Journal of Ophthalmology 80: 389–393. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Oldenburg AL, Hansen MN, Ralston TS, Wei A, Boppart SA (2009) Imaging gold nanorods in excised human breast carcinoma by spectroscopic optical coherence tomography. J Mater Chem 19: 6407. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Mornet S, Vasseur S, Grasset F, Duguet E (2004) Magnetic nanoparticle design for medical diagnosis and therapy. J Mater Chem 14: 2161. [Google Scholar]
  • 14. Nahrendorf M, Zhang H, Hembrador S, Panizzi P, Sosnovik DE, et al. (2008) Nanoparticle PET-CT Imaging of Macrophages in Inflammatory Atherosclerosis. Circulation 117: 379–387. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Sandrian MG, Wollstein G, Schuman JS, Bilonick RA, Ling Y, et al.. (2012) Inflammatory response to intravitreal injection of gold nanorods. Br J Ophthalmol. [DOI] [PMC free article] [PubMed]
  • 16. Oldenburg AL, Hansen MN, Zweifel DA, Wei A, Boppart SA (2006) Plasmon-resonant gold nanorods as low backscattering albedo contrast agents for optical coherence tomography. Optics Express 14: 6724–6738. [DOI] [PubMed] [Google Scholar]
  • 17. Ehlers JP, Gupta PK, Farsiu S, Maldonado R, Kim T, et al. (2010) Evaluation of Contrast Agents for Enhanced Visualization in Optical Coherence Tomography. IOVS 51: 6614–6619. [DOI] [PubMed] [Google Scholar]
  • 18. Lee TM, Oldenburg AL, Sitafalwalla S, Marks DL, Luo W, et al. (2003) Engineered microsphere contrast agents for optical coherence tomography. Opt Lett 28: 1546–1548. [DOI] [PubMed] [Google Scholar]
  • 19. Chen J, Saeki F, Wiley BJ, Cang H, Cobb MJ, et al. (2005) Gold Nanocages: Bioconjugation and Their Potential Use as Optical Imaging Contrast Agents. Nano Lett 5: 473–477. [DOI] [PubMed] [Google Scholar]
  • 20. Bárány EH (1964) Simultaneous Measurement of Changing Intraocular Pressure and Outflow Facility in the Vervet Monkey by Constant Pressure Infusion. IOVS 3: 135–143. [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1

Laminar Flow.

(DOCX)

Figure S1

Spectrophotometry of GNR in various media. Spectrophotometry of (1) mock aqueous humor (mock AH) alone, (2) 2×1012 gold nanorods/ml (GNR), (3) GNR mixed with distilled water and (4) GNR mixed with mock aqueous humor.

(TIFF)

Figure S2

GNR flow during imaging. Volume remaining in the reservoir over time for eye 1. A linear slope indicates a constant flow rate.

(TIFF)


Articles from PLoS ONE are provided here courtesy of PLOS

RESOURCES