Pakistan Journal of Medical Sciences logoLink to Pakistan Journal of Medical Sciences
. 2014 Jan-Feb;30(1):116–121. doi: 10.12669/pjms.301.3635

Restoration of femoral offset, rotation centers, limbs length equality of Chinese total hip arthroplasty patients

Yu-ping Liu 1, Yue-dong Hao 2
PMCID: PMC3955554  PMID: 24639843

Abstract

Objective: To explore the restoration of femoral offset, rotation centers, limbs length equality of Chinese total hip arthroplasty patients with careful preoperative surgical planning, the appropriate prosthesis and skillful manipulation combined with a variety of verification tests during the operation.

Methods: There were 92 hips (from 92 patients) surgery was performed by the same surgeon using the posterlateral approach by careful preoperative surgical planning. Appropriate prosthesis was chosen determining the reasonable femur osteotomy location, skillful manipulation and paying attention to every detail combined with a variety of verification tests and preoperative measurements during the operation. We evaluated the offset and rotation centers of the healthy (not performed) side and the operated side, the preoperative and postoperative limbs length discrepancy and analyzed the change of femoral offset, rotation centers and limbs length discrepancy of THA patients by self-control.

Results: We found that the preoperative and postoperative femoral offset was basically not changed, the postoperative rotation centers had a tendency to the medial and inferior of the original rotation centers, the limbs length discrepancy and Harris Hip Score (HHS) were improved much more than before.

Conclusions: Careful preoperative surgical planning, the appropriate prosthesis and skillful manipulation combined with a variety of verification tests during the operation is significantly correlated to the remarkable radiological and clinical results of THA patients.

Key Words: Femoral offset, Limbs length, Rotation centers, Total hip arthoplasty

METHODS

Patients: From January 2002 to December 2007, a total of 92 primary THA patients with unilateral hip disease aged from 23 to 78 were performed by the same surgeon, using the Modular total hip prostheses (Link corporation, Germany, the cemented prostheses with SP-II or Classic, the cementless prostheses with Rebed and hybrid prostheses of these three). All patients were operated on using a posterlateral approach in each surgery with the anterior joint capsule and gluteus mdius muscle being preserved. The classification of the diseases of the 92 patients included femoral neck fracture, degenerative hip disease, and femur head necrosis and so on. All the patients were only performed in one limb. No Developmental Dysplasia Of The Hip or other congenital hip disease patients were included in these series 92 patients. (Table-I).

Table-I.

The patients and prosthesis in our study

Original diagnosis Sex Side Age Prosthesis
Osteoarthritis 33
Rheumatoid arthritis 1
Fracture neck of femur 25
Femoral head necrosis 33
Female 49
Male 43
Left 42
Right 50
Min 23
Max 78
Mean 56
Cementless 51
Cemented 27
Hybrid 14

The Ethics Committee of Tengzhou Central People’s Hospital approved the study. Informed written consent was obtained from each patient or from their relative if the patient was incapable of providing consent.

Treatments: All patients received a preoperative intravenous injection of the antibiotic Ceftriaxone. General or spinal anaesthesia were used in these patients.

We used a marker (10cm) to ensure to observe the results of the plain X-rays accurate effectively. Before operation, we strictly observed the plain film using the corporation’s templates combined with the maker to determine the sizes and types of the intraoperative prosthesis and fixed the accurate Osteotomy line of femoral neck and defined the actual rotation centers. We operated carefully and cautiously and applied the method of patellar to patellar comparative test8, Shuck test, Drop-kick test and other methods to ensure the limbs length equality during the THA operations. Plain anteroposterior and lateral radiographs were obtained on the first postoperative day. After operation we gave these patients clear guidance of rehabilitation and physical training.

Radiological measurement: All the plain films were taken according to the method that Lindgren9 introduced. The pre- and postoperative plain X-rays were studied and observed by the same doctor. The femoral offset was measured as the distance from the center of rotation of the femoral head to the long axis of the femoral shaft. The rotation centers X-value was the distance from rotation centers to the tangent line of the teardrop inner margin.10 The rotation centers Y-value was the distance from rotation centers to the line of bilateral teardrop. The LLD of postoperative and preoperative was the difference from the peak of the greater trochanter to the line between bilateral tear-drop. (Fig. 1, 2, 3 and 4).

Fig.1.

Fig.1

1, a the maker, the actual length is 10 cm 2 b is the line between bilateral tear-drop 3.The LLD is the difference of the length between c and d 4,X is the value X of rotation centers 5,Y is the value Y of rotation centers, the Length Y=d 6,O is the length of femoral offset

Fig.2.

Fig.2

A femoral neck fracture Patient who was performed THA with hybrid prostheses. We used the marker (10cm) preoperative and used the marker or the metal or ceramic femoral head postoperative as the reference

Fig.3.

Fig.3

A bilateral femoral head necrosis Patient, whose left hip is very gross, but the right side is basically good and the femoral head did not depressed

Fig.4.

Fig.4

The patient was performed THA with Rebed prostheses

Follow-up: All patients had immediate postoperative anteroposterior (AP) and lateral radiographs of the hip. We took repeat radiographs yearly until the latest follow-up. No patients were lost to follow-up. The minimum follow-up was 49 months (mean, 64 months; range, 49–88 months). AP radiographs were taken with the legs positioned in 158 internal rotation with the coccyx centered 2cm above the pubic symphysis. Three of us (JAG, CAB, JM) measured all radiographs for native and reconstructed femoral offset, acetabular inclination, and acetabular ante version from the standardized radiographs using MATLAB1 (The Math-Works Inc, Natick, MA). Any interobserver difference between measurements was noted and re-measured. All patients were followed up, for from 26 months to 84 months. The statistical analysis for comparison was performed using SPSS software version 18.0 for Windows. Statistical significance of difference was determined by paired Student’s t test. A value of P<0.05 was considered significant.

RESULTS

Radiological results

Femoral offset: We found that the preoperation femoral offset was 3.689±0.7538cm and the postoperation (Operated side) was 3.539±0.7271cm, p>0.05, the femoral offset was basically not changed.

Rotation centers X, Y: We found that the preoperation X value of rotation centers of control side was 4.14±0.801 cm, the preoperation Y value of rotation centers of control side was 2.381±0.7646 cm and the X value of rotation centers of operated side was 3.81±0.659cm, the Y value of rotation centers of operated side was 2.218±0.6979 cm the X and Y values of rotation centers decreased, p<0.05 as shown in Table-II.

Table-II.

The change of pre- and postoperative femoral offset, X, Y value of rotation centre, LLD and HHS score

Preoperative (n=92) Postoperative (n=92) Statistical significance
Femoral offset 3.689±0.7538cm 3.539±0.7271cm NS
X value of ration center 4.14±0.801cm 3.81±0.659cm P<0.0001
Y value of ration center 2.381±0.7646cm 2.218±0.6979cm P=0.015
HSS 52±6.64 91±4.17 P<0.0001
LLD 0.646±0.7620 cm 1.364±1.2849cm P<0.0001

Data are mean ± SD, Statistical significance of difference was determined by paired Student’s t test,

A value of P<0.05 was considered significant

The discrepancy of two limbs: Before operation, we valued the LLD of all cases, the LLD of these cases were from 0 to 5.5cm. 33 0f these cases had LLD less than 0.6 cm, 25 of them with LLD more than 0.6cm but less than 1.2cm while 8 of these cases had LLD more than 1.2cm but less than 1.8cm and there were 26 cases whose LLD were more than 1.8cm. After operation, we found that the cases with LLD less than 0.6cm were 54, the cases with LLD more than 0.6cm but less than 1.2cm were 19, the cases whose LLD was more than 1.2cm but less than 1.8cm were 16 and only 3 cases had LLD over 1.8cm. The Average value of LLD was from 1.344 cm and decreased to 0.464 cm as shown in Fig.5.

Fig.5.

Fig.5

LLD, limbs length discrepancy, 1 column, LLD < 0.6cm 2 column, 1.2 cm >LLD >0.6 cm 3 column, 1.2<LLD<1.8cm 4 column, LLD > 1.8cm

Clinical results and Complications: We followed up these patients form 26 months to 84 months (mean 49 months), the mean HHS of these patients were improved from 52 (31-70) to 92.3(85-96) at the latest follow-ups, no patients required a crutch to support walking. The life of these patients had a qualitative leap. We did not find any case with infection joint dislocation or any other complications.

DISCUSSION

For total hip arthroplasty (THA), the challenge was to obtain optimal function of the reconstruction hip and to correct the femoral offset, decrease any limbs length discrepancy, and guarantee the centers of rotation of the hip joint.

The preservation of femoral offset was an important element of THA for reproducing the biomechanics of the hip joint after total hip arthroplasty, since it reduced the load transferred to the cup and enabled better joint stability. Inadequate femoral offset could lead to joint instability, high joint reaction forces, increased polyethylene wear, and decreased range of motion.1-3,11,12 For this reason, decreased femoral offset reduced abductor muscle strength, increased incidence of limp. An increase in the joint reaction force may also lead to a higher rate of wear because of decreased femoral offset. Devane PA et al.6 had confirmed that reduction of normal femoral offset leads to an increase in PE wear. A failure to reproduce femoral offset in THA could result in limp, fatigue, impingement, and recurrent subluxation and dislocation.7

After total hip arthroplasty (THA), the femoral offset was generally reduced. But our findings that the preoperative femoral offset was 3.689± 0.7646cm and postoperative (Operated side) was 3.539± 0.7271cm p>0.05, the femoral offset was basically not changed. This may be the reasons or evidences of our better clinical results.

Leg length discrepancy is also an important functional parameter that is related to success in THA. LLD was a well-known complication of THA4,5,13-18, Leg length inequality can contribute to low back pain, sciatic nerve palsy, ipsilateral knee pain, and abnormal force transmission across the hip joint,14,19-23 and may contribute to hip instability, aseptic prosthesis loosening and compromised cardiopulmonary function.5 Leg length inequality was also a primary cause for malpractice liability lawsuits after THA in the United States.13,14,21

Williamson and Reckling23 reported that 27% of patients required heel, Love and Wright also reported limbs lengthening of greater than 1.5 cm in 18% of the 40 patients in their reports. Several methods using pins, rulers, and calipers have been described for intraoperative correction of limbs-length inequality.24-26 Typically, measurement of the distance between two reference points marked on the pelvis and femur has been performed, methods using the anterior superior iliac spine or iliac wing as a reference.

Measurement of limbs length and offset with a maker appears to be reliable, therefore eliminated the need to revise the magnification of the plain radiographs, Usually, the surgeon addressed the equalization of limbs length and restoration of offset by preoperative physical examination, manual and/or computer-assisted template, and/or the use of mechanical tools, pins and tape measures.24-26 In this study, we still observed data by preoperative plain film examination with the guidance of the marker, and used template of the Link corporation as a preoperative measurement tool to observe limbs length, offset and X, Y value of rotation centers and to determine the type and size of the prostheses.

In fact, all observations of the planning parameters preoperative are not fully accurate because of the magnification effect, anatomic conditions, or possible defective execution. The marker is a relatively accurate and reliable index that can be employed easily for preoperative plain film measurement during THA. So the accuracy and reliability of our observation was with relative high confidence.

Preoperative templating may be a way of restoring the required medial offset and correcting the LLD. We never denied Preoperative planning for THA enables selection of the appropriate length for the prosthetic neck and eventually the type of implants to utilize. Jasty et al17 had used a caliper in association with preoperative templating to measure limbs length and found that 16% of the patients had limbs length inequality after surgery Woolson et al13 used preoperative templating and found 86% of patients LLD less than 6mm. Several authors have described various methods to obtain correction of LLD18,27 with a view to restore the normal geometry of the hip during the total hip arthroplasty, Although these methods had many merits and achieved relative excellent results , there were still limitations of these technique for example, Patient's anaesthesia condition, whether Patient's body posture changed intraoperative and so on.

We used the methods from Woolson et al13 and Jasty et al17 and found that 80 percent of the patients had a postoperative LLD less than 1.2 cm, and 59% had a LLD within 6 mm. The results were not as good as theirs but we thought our results were more reliable, because first, we had performed total hip arthroplasty just only in one limbs, and the normal limbs as a reference. Second, we used marker and made our observation more accurate. Third, sometimes we had to increase the length of the neck intraoperative and accordingly to lengthen the femoral offset to maintain the hip stability. This may be a factor of our results not being so excellent.

The rotation centers had a minute lower medial shift and we did not find enough literature discussion on it. We couldn’t conclude whether the deflection of rotation centers affects the longevity of the prosthesis, but most surgeons think it is better to optimize on the reconstruction of hip8,15,28 because minute medial shift of rotation centers can reduce the gravitational force arm which in turn reduces the load transferred to the cup and accordingly decreases the PE wear.

There were many factors which could influence the terminal results of THA. For example, Age, femoral head size, length of follow up, femoral ante version, patient gender, weight, and activity level can affect PE wear. So our study does, however, have several limitations. Our follow-up is relatively short and therefore we are unable to guarantee the long-term advantage of the results, the number of patients was not large enough and so on. Maybe it can only reveal a part of the mechanism of total hip arthroplasty.

ACKNOWLEDGMENT

The authors wish to thank Dr. Yong-gang Zhou of the Department of Joint Surgery, PLA general hospital, for his help of the radiography images and extensive efforts in assembling and editing this manuscript in this study.

References

  • 1.McGrory BJ, Morrey BF, Cahalan TD, An KN, Cabanela ME. Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. J Bone Joint Surg Br. 1995;77(6):865–869. PMid:7593096. [PubMed] [Google Scholar]
  • 2.Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH. Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res. 2001;388:125–134. doi: 10.1097/00003086-200107000-00019. http://dx.doi.org/10.1097/00003086-200107000-00019. [DOI] [PubMed] [Google Scholar]
  • 3.Downing ND, Clark DI, Hutchinson JW, Colclough K, Howard PW. Hip abductor strength following total hip arthroplasty: a prospective comparison of the posterior and lateral approach in 100 patients. Acta Orthop Scand. 2001;72(3):215–220. doi: 10.1080/00016470152846501. http://dx.doi.org/10.1080/00016470152846501. [DOI] [PubMed] [Google Scholar]
  • 4.Abraham WD, Dimon JH, 3rd Leg length discrepancy in total hip arthroplasty. Orthop Clin North Am. 1992;23:201–209. PMid:1570134. [PubMed] [Google Scholar]
  • 5.Gurney B, Mermier C, Robergs R, Gibson A, Rivero D. Effects of limb-length discrepancy on gait economy and lower-extremity muscle activity in older adults. J Bone Joint Surg Am. 2001;83-A(6):907–915. doi: 10.2106/00004623-200106000-00013. PMid:11407800. [DOI] [PubMed] [Google Scholar]
  • 6.Devane PA, Horne JG. Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res. 1999;369:59–72. doi: 10.1097/00003086-199912000-00007. http://dx.doi.org/10.1097/00003086-199912000-00007. [DOI] [PubMed] [Google Scholar]
  • 7.Barrack RL. Factors influencing polyethylene wear in total joint arthroplasty. Orthopedics. 1998;21(9):937–940. doi: 10.3928/0147-7447-19980901-05. PMid:9769031. [DOI] [PubMed] [Google Scholar]
  • 8.Charles MN, Bourne RB, Davey JR, Greenwald AS, Morrey BF, Rorabeck CH. Soft-tissue balancing of the hip: the role of femoral offset restoration. Instr Course Lect. 2005;54:131–141. PMid:15948440. [PubMed] [Google Scholar]
  • 9.Lindgren JU, Rysavy J. Restoration of femoral offset during hip replacement. A radiographic cadaver study. Acta Orthop Scand. 1992;63(4):407–410. doi: 10.3109/17453679209154755. http://dx.doi.org/10.3109/17453679209154755. [DOI] [PubMed] [Google Scholar]
  • 10.Karachalios T, Hartofilakidis G, Zacharakis N, Tsekoura M. A 12- to 18-year radiographic follow-up study of Charnley low-friction arthroplasty. The role of the center of rotation. Clin Orthop Relat Res. 1993;296:140–147. PMid:8222417. [PubMed] [Google Scholar]
  • 11.Asayama I, Naito M, Fujisawa M, Kambe T. Relationship between radiographic measurements of reconstructed hip joint position and the Trendelenburg sign. J Arthroplasty. 2002;17(6):747–751. doi: 10.1054/arth.2002.33552. http://dx.doi.org/10.1054/arth.2002.33552. [DOI] [PubMed] [Google Scholar]
  • 12.Amstutz HC, Sakai DN. Total joint replacement for ankylosed hips. Indications, technique, and preliminary results. J Bone Joint Surg Am. 1975;57(5):619–625. PMid:1150702. [PubMed] [Google Scholar]
  • 13.Woolson ST, Hartford JM, Sawyer A. Results of a method of leg-length equalization for patients undergoing primary total hip replacement. J Arthroplasty. 1999;14(2):159–164. doi: 10.1016/s0883-5403(99)90119-5. http://dx.doi.org/10.1016/S0883-5403(99)90119-5. [DOI] [PubMed] [Google Scholar]
  • 14.Bose WJ. Accurate limb-length equalization during total hip arthroplasty. Orthopedics. 2000;23(5):433–436. PMid:10825109. [PubMed] [Google Scholar]
  • 15.Bourne RB, Rorabeck CH. Soft tissue balancing: the hip. J Arthroplasty. 2002;17(4 Suppl 1):17–22. doi: 10.1054/arth.2002.33263. http://dx.doi.org/10.1054/arth.2002.33263. [DOI] [PubMed] [Google Scholar]
  • 16.Hoikka V, Santavirta S, Eskola A, Paavilainen T, Wirta J, Lindholm TS. Methodology for restoring functional leg length in revision total hip arthroplasty. J Arthroplasty. 1991;6(3):189–193. doi: 10.1016/s0883-5403(06)80163-4. http://dx.doi.org/10.1016/S0883-5403(06)80163-4. [DOI] [PubMed] [Google Scholar]
  • 17.Jasty M, Webster W, Harris W. Management of limb length inequality during total hip replacement. Clin Orthop Relat Res. 1996;333:165–71. PMid:8981892. [PubMed] [Google Scholar]
  • 18.Ranawat CS, Rao RR, Rodriguez JA, Bhende HS. Correction of limb-length inequality during total hip arthroplasty. J Arthroplasty. 2001;16(6):715–720. doi: 10.1054/arth.2001.24442. http://dx.doi.org/10.1054/arth.2001.24442. [DOI] [PubMed] [Google Scholar]
  • 19.Friberg O. Clinical symptoms and biomechanics of lumbar spine and hip joint in leg length inequality. Spine (Phila Pa 1976) 1983;8(6):643–651. doi: 10.1097/00007632-198309000-00010. http://dx.doi.org/10.1097/00007632-198309000-00010. [DOI] [PubMed] [Google Scholar]
  • 20.Giles LG, Taylor JR. Low-back pain associated with leg length inequality. Spine (Phila Pa 1976) 1981;6(5):510–521. doi: 10.1097/00007632-198109000-00014. http://dx.doi.org/10.1097/00007632-198109000-00014. [DOI] [PubMed] [Google Scholar]
  • 21.Hofmann AA, Skrzynski MC. Leg-length inequality and nerve palsy in total hip arthroplasty: a lawyer awaits! . Orthopedics. 2000;23(9):943–4. doi: 10.3928/0147-7447-20000901-20. PMid:11003095 . [DOI] [PubMed] [Google Scholar]
  • 22.Edeen J, Sharkey PF, Alexander AH. Clinical significance of leg-length inequality after total hip arthroplasty. Am J Orthop (Belle Mead NJ) 1995;24(4):347–351. PMid:7788314. [PubMed] [Google Scholar]
  • 23.Williamson JA, Reckling FW. Limb length discrepancy and related problems following total hip joint replacement. Clin Orthop Relat Res. 1978;134:135–138. PMid:729230. [PubMed] [Google Scholar]
  • 24.McGee HM, Scott JH. A simple method of obtaining equal leg length in total hip arthroplasty. Clin Orthop Relat Res. 1985;194:269–270. PMid:3978924. [PubMed] [Google Scholar]
  • 25.Naito M, Ogata K, Asayama I. Intraoperative limb length measurement in total hip arthroplasty. Int Orthop. 1999;23(1):31–33. doi: 10.1007/s002640050298. http://dx.doi.org/10.1007/s002640050298. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Shiramizu K, Naito M, Shitama T, Nakamura Y, Shitama H. L-shaped caliper for limb length measurement during total hip arthroplasty. J Bone Joint Surg Br. 2004;86(7):966–969. doi: 10.1302/0301-620x.86b7.14587. http://dx.doi.org/10.1302/0301-620X.86B7.14587. [DOI] [PubMed] [Google Scholar]
  • 27.Huddleston HD. An accurate method for measuring leg length and hip offset in hip arthroplasty. Orthopedics. 1997;20(4):331–332. doi: 10.3928/0147-7447-19970401-10. PMid:9127867. [DOI] [PubMed] [Google Scholar]
  • 28.D'Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW Jr. The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am. 2000;82(3):315–321. doi: 10.2106/00004623-200003000-00003. PMid:10724224. [DOI] [PubMed] [Google Scholar]

Articles from Pakistan Journal of Medical Sciences are provided here courtesy of Professional Medical Publications