Abstract
Ultra-low coverage sequencing (ULCS) is one of the most promising strategies for sequencing based clinical application. These clinical applications, especially prenatal diagnosis, have a strict requirement of turn-around-time; therefore, the application of ULCS is restricted by current high throughput sequencing platforms. Recently, the emergence of rapid sequencing platforms, such as MiSeq and Ion Proton, brings ULCS strategy into a new era. The comparison of their performance could shed lights on their potential application in large-scale clinic trials. In this study, we performed ULCS (<0.1X coverage) on both MiSeq and Ion Proton platforms for 18 spontaneous abortion fetuses carrying aneuploidy and compared their performance on different levels. Overall basic data and GC bias showed no significant difference between these two platforms. We also found the sex and aneuploidy detection indicated 100% sensitivity and 100% specificity on both platforms. Our study generated essential data from these two rapid sequencing platforms, which provides useful reference for later research and potentially accelerates the clinical applications of ULCS.
Introduction
Ultra-low coverage sequencing (ULCS) is one of the most promising clinical strategies, such as prenatal diagnosis, spontaneous abortion (SA) reason tracking and etc. Although only 0.1X coverage is sequenced for each sample, the data provided efficient genomic information for copy number variation (CNV) detection. Therefore, ULCS strategy is widely used in clinic trials for large-segment CNV detection, such as invasive [1]/noninvasive [2], [3] prenatal diagnosis, Spontaneous Abortion (SA) reason tracking [4]. Unlike karyotyping analysis, ULCS could be applied without cell culture, which is laborious and prone to failure for specific samples, such as SA tissues [5]. In the clinical practice, most of these sequence-based diagnoses have a strict turn-around-time (TAT) requirement, especially for prenatal diagnosis. However, it takes several days for current high throughput platforms to finish a sequencing run; for example, HiSeq 2000 will take approximately 11 days to generate 600 Gbp [6]. Recently, the emergence of rapid sequencing platform, such as MiSeq and Ion Proton, will bring the clinical application of ULCS into a new era.
Targeting clinical laboratories and diagnostic market, in 2011 Illumina released MiSeq, a lower throughput instrument with shorter TAT. With sequencing-by-synthesis and fluorescently labeled reversible-terminator nucleotides technologies, MiSeq currently generates 10 Gbp paired-end 150 bp reads. Benefit from the improved optical system, MiSeq can finish a sequence run within 27-hours [6]. In the same year, Ion Torrent released their Personal Genome Machine (PGM). With the development of integrated semiconductor device, PGM can generate millions of sequential reads by detecting the proton releasing during the base incorporation without any optical systems. By upgrading the semiconductor device, the newly released Ion Proton can generate up to 10 Gbp of sequencing data (with average length of 100 bp) in 4 hours. Recent studies have already showed their feasibility for ULCS clinical application [4]. However, there is a lack of comparison between these two platforms using the same sample set.
In this study, we performed ULCS (<0.1X) to 18 SA fetal tissue samples on both MiSeq and Ion Proton platforms. Also, we compared their performance on GC bias at different level. Afterwards, we investigated sex distinguish and aneuploidy detection performance of both platforms for clinical application. Our research is the first study to generate ULCS data for both platforms using the same sample set, which provides a useful reference of platform selection in the context of clinical especially prenatal diagnosis.
Results
Sample Recruitment and Data Generation
To compare the performance between MiSeq and Ion Proton for ULCS, we selected 18 SA fetal samples from a previous study [4]. These fetuses carried 18 different aneuploidies. Their karyotypes are obtained by comparative genomic hybridization and confirmed by FISH. DNA extraction were performed following standard Phenol-Chloroform extraction protocol [4] and quantified with the Quant-iT dsDNA HS Assay Kit (Invitrogen). Considering the amount of input DNA and the PCR process are the most important bias-related factors [7], we used equal amount of input DNA (50 ng) and same numbers of PCR cycles (10 cycles) for library preparation before sequencing on both platforms to minimize the influence to the GC bias comparison (Materials and Methods).
On MiSeq platform, we performed 150 bp paired-end sequencing following the standard manufacture protocol [4]. After a 27-hours sequence run, we generated 4.58 million 150 bp paired-end reads in total for this study, 38.15 Mbp sequence base in average for each sample (Table 1 & Table S1A). 86.35% of these reads could be mapped to the reference genome (GRCh37, UCSC release hg19). After removal of the PCR duplications, ∼214 K paired reads for each sample were obtained, resulting 84.24% effective (unique non-duplicated) reads for the following analysis. On Ion Proton platform, we generated 39.33 million single end reads with a median length of 111 bp. The read length distribution (Figure S1) showed a peak around 127 bp. 87.10% of these single end reads could be mapped to the human genome uniquely. Finally, we obtained 1.70 million unique non-duplicated mapped reads per sample on Ion Proton, which is 7.97 fold more than the data yield on MiSeq.
Table 1. Sequencing data statistics.
MiSeq | Ion Proton | |||||||
Raw | 90 K** | Raw | 90 K | |||||
AVERAGE | S.D.* | AVERAGE | S.D. | AVERAGE | S.D. | AVERAGE | S.D. | |
Reads Number | 254,360 | 36,617 | 180,000 | – | 2,184,945 | 611,149 | 90,000 | – |
Unique Map Reads | 219,624 | 32,237 | 155,513 | 4,128 | 1,901,086 | 526,637 | 78,400 | 1,342 |
Unique Map Rate(%) | 86.35 | 2.27 | 86.40 | 2.29 | 87.10 | 1.49 | 87.11 | 1.49 |
Duplication Rate(%) | 0.08 | 0.01 | 0.07 | 0.01 | 10.10 | 0.64 | 0.62 | 0.19 |
Unique non-duplication Reads | 214,288 | 31,740 | 151,731 | 4,241 | 1,707,684 | 467,757 | 77,916 | 1,336 |
Unique non-duplication Rate(%) | 84.24 | 2.33 | 84.30 | 2.36 | 78.30 | 1.46 | 86.57 | 1.49 |
*S.D. represents for Standard Deviation.
**90K paired end reads.
Therefore, to conduct a reasonable comparison, we randomly drew 90K raw reads from each sample and generated basic statistics of those reads on both platforms (Table 1 & Table S1B). In this equivalent data set, the average duplication rate of Ion Proton data was 0.62%, which was 0.07% on MiSeq. But benefited from lower sequence error rate, it show a higher mapped rate on Ion Proton platform; therefore, Ion Proton left 2.27% more unique non-duplicated reads than MiSeq for further analysis.
Characteristics of Sequence Bias
Coverage evenness is one of the main factors affects CNV detection, and CNV detection is the most important application of ULCS strategy. Therefore, it’s meaningful to evaluate the coverage evenness in different level. Firstly, we investigated the coefficient of variation (CV) of unique reads percentage (UR%) on each autosome from the 90K reads data set (Materials and Methods). This CV, which is also employed in a previous study [8], could effectively describe the relative variance of chromosomal UR%, and overall GC bias. The CVs on both MiSeq and Ion Proton were lower than 15% on each chromosome (Figure 1A). The smaller CVs indicated better repeatability for application based from large-scale ULCS, such as aneuploidy screening [8]. The Wilcoxon signed rank test (p-value = 0.3986) indicated there were no significant difference of the autosomal CVs between MiSeq and Ion Proton. Also, we found autosomes with lower (such as chr13) and higher (such as chr19) GC content showed larger CV on both platforms, which is consistent to a previous study [8]. In the consistency analysis, the chromosomal UR% showed high linearity (R2 = 0.9909) between MiSeq and Ion Proton (Figure 2) with slope and intercept of 0.9753 and 0.0047 respectively. These numbers implied that these two platforms would display no significant difference on aneuploidy detection.
Considering ULCS would be applied for micro-deletion/duplication syndromes detection, we evaluated the relationship between coverage evenness across GC content at local 1 Mbp window level. The relative depth (RD) of each 1 Mbp observation window was calculated (Materials and Methods).We found most of average RDs for each GC content category fell within the range between 0.8 and 1.20 on both platforms (Figure 1B & Figure S2A/B & Figure S3), which indicates good genome evenness across GC content on both platforms. Besides, we discovered that these two platforms showed similar RD distribution across different GC levels (Wilcoxon signed rank test, p-value = 0.2663). These curves showed reduced RD in GC-poor (GC<38%) regions but increased RD in GC-enrich regions. These analyses implied that current GC bias correction strategies [3] could be employed on these two platforms to detect micro-deletion/duplication syndromes.
Aneuploidy Detection Power using ULCS
As mentioned, aneuploidies detection is one of the clinical applications of ULCS. To detect aneuploidies among of 18 samples, we employed a Z-score method as described in a previous study [4]. Z-scores enable us to compare the UR% among different chromosomes in a standardized manner; even the chromosomal size and mappability are different. For example, using Z-score method, we can compare UR-chr1 (mean 8.43/8.55 and SD 0.25/0.21) with UR-chr22 (mean 1.43/1.46 and SD 0.17/0.12) effectively, and detect aneuploidies with the same cutoff. Using the equivalent data set, all of the aneuploidies are accurately detected (Figure 3A and Table 2), which indicated 100% sensitivity and 100% specificity for both platforms. This result implied a comparable detection power of these two platforms, presenting their feasibility for clinical applications. However, we realized that it is an incomplete conclusion based on only 18 positive samples; indeed, a larger sample size would be more statistically meaningful.
Table 2. Aneuploidy Detection results using UCLS.
Sample | MiSeq | Ion Proton | Standard Karyotype | ||
Detection | Consistency* | Detection | Consistency* | ||
SAA01 | 47,XX +2 | Yes | 47,XX +2 | Yes | 47,XX +2 |
SAA02 | 47,XY +3 | Yes | 47,XY +3 | Yes | 47,XY +3 |
SAA03 | 47,XY +4 | Yes | 47,XY +4 | Yes | 47,XY +4 |
SAA04 | 47,XX +5 | Yes | 47,XX +5 | Yes | 47,XX +5 |
SAA05 | 47,XX +6 | Yes | 47,XX +6 | Yes | 47,XX +6 |
SAA06 | 47,XY +7 | Yes | 47,XY +7 | Yes | 47,XY +7 |
SAA07 | 47,XY +8 | Yes | 47,XY +8 | Yes | 47,XY +8 |
SAA08 | 47,XX +9 | Yes | 47,XX +9 | Yes | 47,XX +9 |
SAA09 | 47,XY +11 | Yes | 47,XY +11 | Yes | 47,XY +11 |
SAA10 | 47,XY +10 | Yes | 47,XY +10 | Yes | 47,XY +10 |
SAA11 | 47,XX +12 | Yes | 47,XX +12 | Yes | 47,XX +12 |
SAA12 | 47,XX +13 | Yes | 47,XX +13 | Yes | 47,XX +13 |
SAB01 | 47,XY +14 | Yes | 47,XY +14 | Yes | 47,XY +14 |
SAB02 | 47,XX +17 | Yes | 47,XX +17 | Yes | 47,XX +17 |
SAB03 | 47,XY +18 | Yes | 47,XY +18 | Yes | 47,XY +18 |
SAB04 | 47,XY +20 | Yes | 47,XY +20 | Yes | 47,XY +20 |
SAB05 | 47,XY +21 | Yes | 47,XY +21 | Yes | 47,XY +21 |
SAB06 | 47,XY +22 | Yes | 47,XY +22 | Yes | 47,XY +22 |
*Consistency: We obtain the consistency by comparing the detection result with the standard karyotype.
Interestingly, in the aneuploidy detection, we also found that chromosomal GC content was one of the main reasons affecting the significance of our detection. The curve in Figure 3A indicated trisomy of chromosomes with lower or higher GC content trend to have smaller Z-score than those with average GC content. For instance, the Z-score of chr7 (GC = 40.75%) for SAA06 (47 XY, +7) is 26.90 for MiSeq and 22.75 for Ion Proton, while chr22 (GC = 47.99%) for SAB06 (47 XY, +22) have smaller Z-score of 5.63 and 5.61 for MiSeq and Ion Proton respectively. This phenomenon consistent with the chromosomal CV evaluation.
Sex Chromosome
Theoretically, the fetal sex can be detected by the UR-Y% easily. However, unexpected aligned read from other genomic region, mostly induced by sequencing error or miss alignment, might lead to unexpected high UR-Y% and false determination of sex. Using the sequence reads from female fetus, we found the unexpected aligned reads mainly enriched in two region on chrY (chrY: 13,104,553–13,748,577 and chrY: 58,819,361–59,373,566) (Figure S4). To achieve higher accuracy, we removed these two regions from the calculation of UR-Y%. Afterwards, we employed a Gaussian finite mixture model fitted by E–M algorithm for sex distinguish (Materials and Methods). In this analysis, these 18 samples were successfully classified into two clusters (Figure 3B). Both platforms showed 100% accuracy for the sex. However, there were still some slight differences between these two platforms. In the female clusters, UR-X% and UR-Y% of MiSeq data (SD = 0.1371 and 0.0049 respectively) showed more narrow and centralized distribution than those of Ion Proton (SD = 0.0985 and 0.0250 respectively), while in the male cluster, UR-X% and UR-Y% of MiSeq data (SD = 0.2082 and 0.0245 respectively) was also more narrow and centralized than the Ion Proton data (SD = 0.1745 and 0.0228 respectively). This phenomenon may relate to the sequence reads length difference between these two platforms. In a previous study, we found the longer sequence reads will improve the accuracy of sex distinguish [4].
Discussion
In this study, we sequenced 18 SA fetuses on both MiSeq and Ion Proton platforms. By comparing the basic statistics on the equivalent read data set, we found these two platforms showed similar performance. In chromosomal CV analysis, both MiSeq and Ion Proton showed CV lower than 15% on all chromosomes, which was good for following aneuploidy detection. Based on the equivalent data set, we accurately detected the aneuploidies and sex with 100% sensitivity and specificity at high confidence. Moreover, the consistency analysis indicated high comparability between these two platforms.
In the bias analysis, the curve between RD and GC content revealed GC bias on both MiSeq and Ion Proton platforms. Currently, researches attributed this bias to the PCR process during library preparation, assuming more PCR cycles can potentially lead to significant GC bias [7]. Therefore, in this study, same PCR cycles were recruited in the library preparation for both MiSeq and Ion Proton. As expected, both platforms showed similar GC bias, which is consistent with former research and indirectly supports that PCR cycles number is the main factor inducing GC bias. It is conceivable that the surface chemistry processes before sequencing, such as bridge PCR for MiSeq and emulsion PCR for Ion Proton, and the sequencing methods, in the case of ULCS, had contributed less significant GC bias than the PCR step during library preparation. Therefore, methods to reduce the PCR bias during library preparation could potentially improve platform performance for the clinical application of ULCS strategy.
In clinical practice, throughput, cost and TAT are the critical factors restricted their application. Comparison of these two platforms is shown (Table 3). Currently, in each paired-end 150 bp run, Illumina MiSeq needs 24–27 hours to generate up to ∼17 million paired reads, while Ion Proton can generate up to ∼80 million long reads using 3–4 hours with PI chips and 200 bp kit. The shorter TAT of Ion Proton, which enables us to finish the detection within 1∼2 working days, may bring benefits to applications with strict TAT requirement such as prenatal diagnosis. However, the higher throughput will bring both advantage and disadvantages to Ion Proton platform. On one hand, higher throughput will enable us to generate more data for each sample. On the other hand, we need to improve the library preparation for higher coverage sequencing. For example, in raw data for these 18 samples, the duplication rate of Ion Proton platform was as high as 10.10% in average, which decreased the percentage of effective reads. Several optimizations on input DNA amount, enzyme and number of PCR cycles need to be addressed in the future. In the cost aspect, for each gigabyte sequence data, MiSeq and Ion Proton need $150 and $100 respectively (Table 3). Assuming that we need 1 million raw reads for ULCS strategy, MiSeq and Ion Proton will cost $46–53 and $14–15 for each patient/sample respectively. The lower cost of Ion Proton will show advantage on clinical application. The dramatic decrease of TAT and cost make sequencing technology comparable or even superior to some existing approaches, such as comparative genomic hybridization.
Table 3. Main output comparisons of current benchtop sequencing platforms for ULCS strategy.
Platform | Reads length | #Base production | #Reads production | Cost | Run time | #Sample per runc | Cost per sample |
MiSeq | Paired-end 150 bp | Up to 5.1 Gbp | 17Ma | $750[12] | 24–27 hours | 14–16 | $46–53 |
Ion Proton | ∼200 bp | Up to 10 Gbp | 80M | $1000b | 3–4 hours | 70–75 | $14–15 |
paired end reads.
Only include the PI chip ($699), reagent for template ($166) and sequencing ($135), pricing from Invitrogen US territory website (http://www.invitrogen.com/, accessed 2 August 2013).
Assuming that we need 1M raw reads per sample, and 10% pooling variance.
To sum up, in this study, we generated efficient and essential raw data for MiSeq and Ion Proton using ULCS for 18 samples carrying rare mutations. Also, our comparisons on data production, GC-bias, sex distinguish, TAT and cost provided important reference for following researches. Our evaluation revealed the feasibility for clinical application of ULCS on MiSeq and Ion Proton platforms. Our study has demonstrated the capability of Ultra-Low Coverage Sequencing strategy in clinical applications, more samples are required to make a more comprehensive conclusion.
Materials and Methods
Sample Collection and DNA Extraction
In this study, our spontaneous abortions samples were collected between Jan. 2008 and Jan. 2011 from the Institute of Reproduction and Stem Cell Engineering of Central South University, China, and used in former study [4]. CGH and FISH are recruited to obtain and confirm the fetal karyotype. Informed written consent was obtained from each participant, and this study was approved by the Institutional Review Board of BGI and the Institute of Reproduction and Stem Cell Engineering of Central South University.
MiSeq: Libraries Preparation, Sequencing and Mapping
In brief, for library preparation, we sheared the gDNA into 150–200 bp fragments, performed end repair, “A”-overhangs, and adapter ligation. For more details, we purified the DNA fragments using the Agencourt AMPure Kit (Beckman) and amplified it through 10-cycles PCR with multiplex primers. Index tags were induced during this process. After purification of the PCR products, we used Agilent Bioanalyzer DNA 1000 Kit (Agilent Technologies) to confirm the size distribution of the libraries, and quantified them by real-time quantitative PCR. We then pooled the libraries in equal molecular amounts and sequenced with 150 paired-end cycles on the Illumina MiSeq following the standard protocol. The data has been uploaded to NCBI SRA database (SRA116521).
After all, sequencing reads were mapped to the human genome reference (GRCh37, UCSC release hg19) using SOAP2 [9], and only unique non-duplicated reads were used for the follow-up aneuploidy analysis.
Proton: Libraries Preparation, Sequencing and Mapping
In library preparation, we first treated the fragment DNA with T4 DNA polymerase, T4 phosphonucleotide kinase and the Klenow fragment of Escherichia coli DNA polymerase for end repair. Then we ligated the sequence adapter (with 10-bps barcode) to the repaired DNA, and performed nick repair. Afterwards, we amplified the size selected DNA through a 10-cylce PCR. After PCR, we confirmed the library size distribution and quantified the libraries using Agilent Bioanalyzer 2100.
For template preparation procedure and sequencing, we followed the Ion Proton manufacturer’s standard protocol. We pooled the libraries at equal moles, and conducted the template preparation using the Ion PI Template OT2 200 Kit on the Ion OneTouch™ 2 System. According to the user’s guide, we recovered and enriched the Ion Sphere Particles (ISPs), and used the Ion Sphere TM Quality Control assay with the Qubit 2.0 Fluorometer for quality assurance. Afterwards, we loaded the qualified ISPs with template into the Ion PI chips for sequencing. On the Ion Proton System, we used Ion PI Sequencing 200 Kit at 400 flows following the manufacturer’s instruction. Also, the data has been uploaded to NCBI SRA database (SRA116521).
The sequencing reads were mapped to the human reference genome (GRCh37, UCSC release hg19) using Tmap [10]. After removal of PCR duplication, the rest uniquely mapped reads were used for the following analysis.
Aneuploidy Detection
To detect the aneuploidies, we employed the Z-score method in this study. Firstly, we calculated the unique reads percentage (UR%) for each chromosome (labeled as subscript , ) of each sample (labeled as subscript , ). The UR%, termed as , was defined as the ratio between the count of unique reads from chromosome and all autosomes. Secondly, the mean value () and the standard deviation () of the UR% of chromosome were calculated. For an affected chromosome , samples unaffected for that chromosome served as the control set. The Z-score was calculated for each chromosome as follows:
A chromosome with a Z-score≥3 was classified as a trisomy, while the chromosome with a Z-score≤−3 classified as a monosomy. All unaffected chromosomes would present −3<Z-score<3.
Sex Distinguishing
For sex distinguishing, we recruited mclust (http://www.stat.washington.edu/mclust/), a package of Gaussian finite mixture model fitted by EM algorithm [11] in R. For more details, UR-X% and UR-Y% are calculated and given to mclust (modelNames = “VVI”, G = 2) as two-dimension raw input. After the iteration, our samples would be divided into 2 clusters, meanwhile, parameters such as mean, SD and etc. would be estimated using maximum-likelihood method. The sex of these two clusters would be determined by their average UR-X% and UR-Y%; in brief, the male cluster would show smaller UR-X% but higher UR-Y% comparing with female cluster.
Supporting Information
Acknowledgments
We sincerely thank our colleagues at the BGI-Shenzhen for sequencing, and foundations that support our study. Also, we appreciate the help from Wanting Chen (chenwanting@genomics.cn) for the manuscript writing preparation.
Funding Statement
The study was funded by ShenZhen Engineering Laboratory for Clinical molecular diagnostic, Shenzhen Engineering Laboratory for Birth Defects Screening, and China National GeneBank-Shenzhen. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Dan S, Chen F, Choy KW, Jiang F, Lin J, et al. (2012) Prenatal detection of aneuploidy and imbalanced chromosomal arrangements by massively parallel sequencing. PLoS One 7: e27835. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Dan S, Wang W, Ren J, Li Y, Hu H, et al. (2012) Clinical application of massively parallel sequencing-based prenatal noninvasive fetal trisomy test for trisomies 21 and 18 in 11,105 pregnancies with mixed risk factors. Prenat Diagn 32: 1225–1232. [DOI] [PubMed] [Google Scholar]
- 3. Chen S, Lau TK, Zhang C, Xu C, Xu Z, et al. (2013) A method for noninvasive detection of fetal large deletions/duplications by low coverage massively parallel sequencing. Prenat Diagn 33: 584–590. [DOI] [PubMed] [Google Scholar]
- 4. Xie W, Tan Y, Li X, Lin G, Jiang H, et al. (2013) Rapid detection of aneuploidies on a benchtop sequencing platform. Prenat Diagn 33: 232–237. [DOI] [PubMed] [Google Scholar]
- 5. Lebedev IN, Ostroverkhova NV, Nikitina TV, Sukhanova NN, Nazarenko SA (2004) Features of chromosomal abnormalities in spontaneous abortion cell culture failures detected by interphase FISH analysis. Eur J Hum Genet 12: 513–520. [DOI] [PubMed] [Google Scholar]
- 6. Quail MA, Smith M, Coupland P, Otto TD, Harris SR, et al. (2012) A tale of three next generation sequencing platforms: comparison of Ion Torrent, Pacific Biosciences and Illumina MiSeq sequencers. BMC Genomics 13: 341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Aird D, Ross MG, Chen WS, Danielsson M, Fennell T, et al. (2011) Analyzing and minimizing PCR amplification bias in Illumina sequencing libraries. Genome Biol 12: R18. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Jiang F, Ren J, Chen F, Zhou Y, Xie J, et al. (2012) Noninvasive Fetal Trisomy (NIFTY) test: an advanced noninvasive prenatal diagnosis methodology for fetal autosomal and sex chromosomal aneuploidies. BMC Med Genomics 5: 57. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Li R, Yu C, Li Y, Lam TW, Yiu SM, et al. (2009) SOAP2: an improved ultrafast tool for short read alignment. Bioinformatics 25: 1966–1967. [DOI] [PubMed] [Google Scholar]
- 10. Cartwright DA, Troggio M, Velasco R, Gutin A (2007) Genetic mapping in the presence of genotyping errors. Genetics 176: 2521–2527. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Yeung KY, Fraley C, Murua A, Raftery AE, Ruzzo WL (2001) Model-based clustering and data transformations for gene expression data. Bioinformatics 17: 977–987. [DOI] [PubMed] [Google Scholar]
- 12. Loman NJ, Misra RV, Dallman TJ, Constantinidou C, Gharbia SE, et al. (2012) Performance comparison of benchtop high-throughput sequencing platforms. Nat Biotechnol 30: 434–439. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.