As a result of the improved survival of patients with heart failure (HF) and the overall rise in the prevalence of HF,1 the number of patients in advanced (Stage D) HF continues to increase, thus exceeding the limited availability of donor organs by a wide margin.2 Initially used primarily as a bridge to heart transplantation, mechanical circulatory support is now increasingly offered as a destination therapy to patients with advanced HF in clinical deterioration who are not transplant candidates. Improvement in survival to 80% at 1 year post-implant3 has steadily followed the development of new technologies such as the continuous-flow pump, which now encompasses 99% of LVADs,3 and improvements in patient and device management. Far from being a panacea, mechanical circulatory support is still fraught with challenges. Among them, post-LVAD right ventricular failure (RVF) is a major cause of morbidity and mortality.
Despite (1) overall improved outcomes and lower rates of RVF with the use of the newer, continuous-flow LVADs over pulsatile-flow devices4 and (2) development of clinical prediction scores to facilitate preoperative identification of patients at risk for RVF after implantation,5-9 RVF still occurs in 13%-40% of continuous-flow device.10 LVAD function relies heavily on right ventricular (RV) function for adequate preload. Severe RVF can lead to systemic hypoperfusion, multi-organ failure, and death, but also to prolonged or recurrent hospitalization and poor quality of life even in less extreme cases. LVAD recipients who develop RVF have poor outcomes,11-13 including higher perioperative, short-, and long-term mortality14-16 and reduced survival to transplantation.11, 13-15, 17-21 RVF has also been associated with higher risk of bleeding, renal failure, and hypotension,13, 14, 20 and longer intensive care unit and hospital stays.11, 14
In the absence of durable and safe options for RV support, preoperative characterization of RV function and appropriate patient selection is of paramount importance in improving outcomes post-LVAD. In this article, we review the current evidence for preoperative assessment of RV function and RVF risk, with focus on cardiac imaging and specifically echocardiography. To put the discussion in context, we also briefly review the definitions, incidence, and patho-physiology of RVF. Finally, we discuss challenges and suggest directions for future research.
Definition and Incidence of Post-LVAD Right Ventricular Failure
Although RVF is common after LVAD implantation, the precise incidence estimates vary from 9.4% to 44%, depending on the definition of RVF, the characteristics of the study population, and the type of LVAD (Table 1).6, 7, 9-11, 13, 14, 16, 17, 22-33 INTERMACS defines RVF as (1) need of an RVAD, or (2) requirement of inhaled nitric oxide or inotropic therapy for more than 1 week any time after LVAD implantation in the presence of symptoms and signs of persistent RV dysfunction, such as central venous pressure >18 mmHg with a cardiac index <2.3 L/min/m2 in the absence of elevated left atrial or pulmonary capillary wedge pressure (>18 mmHg), cardiac tamponade, ventricular arrhythmias, or pneumothorax.34 Most studies have used a variation of this definition combining clinical findings and hemodynamics (Table 1). Severe RVF requiring RVAD has been reported in 9.4% to 23.4% of patients,8, 22, 23 whereas definitions incorporating need for inotropes yield estimates ranging from 20.2% to 40%.6, 7, 13, 14, 16
Table 1.
Incidence of Right Ventricular Failure After Left Ventricular Assist Device Implantation
Study | Patients | Single or Multicenter | LVAD Type | RVF Definition | RVF Rate |
---|---|---|---|---|---|
Fukamachi et al (1999)22 | N=100, BTT 97%, DT 2%, mean age 52.5, male 86% | Single | Pulsatile 100% | Requiring RVAD support | 11.0% |
Kavarana et al (2002)11 | N=69, mean age 56.2, male 86% | Single | Pulsatile 100% | Requiring inotropic support for ≥ 14 days, or RVAD support | 30.4% |
Ochiai et al (2002)23 | N=245, BTT 98%, 2% DT, mean age 64, male 85% | Single | Pulsatile 100% | Requiring RVAD support | 9.4% |
Dang et al (2006)14 | N=108, BTT 100%*, mean age 51, male 93% | Single | Pulsatile 100% | Requiring inotropic support or pulmonary vasodilators for ≥ 14 days, or RVAD support | 38.9% |
Puwanant et al (2008)24 | N=33, BTT 67%, DT 21%, BTR 12%, mean age 54 | Single | Continuous 55%, Pulsatile 45% | Requiring inotropic support or pulmonary vasodilators for > 14 days | 33.0% |
Potapov et al (2008)25 | N=54, mean age 52, male 91% | Single | Pulsatile 100% | Two of the following criteria in the first 48h: MAP ≤55 mmHg; CVP>16 mm Hg; MvO2 ≤55%; CI <2 L/min/m2; inotropic support score >20 units; or need for RVAD | 16.7% |
Matthews et al (2008)6 | N=197, BTT 94%, DT 6%, mean age 50, male 79% | Single | Continuous 15%, Pulsatile 85% | Requiring RVAD or ECMO support; or inotrope support for >14 days; or inhaled NC for ≥48 h; or hospital discharge with an intravenous inotrope | 34.5% |
Fitzpatrick et al (2008)7 | N=266, mean age 52, 65% male | Single | Continuous 4%, Pulsatile 96% | Requiring BiVAD placement | 37.0% |
Drakos et al (2010)5 | N=175, BTT 58%, DT 42%, mean age 57, male 83% | Single | Continuous 14%, Pulsatile 86% | Requiring RVAD support, or inhaled nitric oxide for ≥48 hours, or inotropic support for > 14 days | 44.0% |
Kormos et al (2010)13 | N=484, BTT 100%, mean age 52, male 78% | Multicenter | Continuous 100% | Requiring inotropic support ≥ 14 days or starting ≥14 days post-implantation; or RVAD support | 20.2% |
Hennig et al (2011)26 | N=40, BTT 100%, mean age 55, male 95% | Single | Continuous 77%, Pulsatile 23% | Two of the following criteria during the 1st week after surgery: MAP ≤55 mmHg, CVP ≥16 mmHg, MvO2 ≤55%, CI <2 L/min/m2, inotropic support score >20 units, or requiring RVAD support | 35.0% |
Baumwol et al (2011)17 | N=40, BTT 83%, DT 17%, mean age 52, male 91% | Single | Continuous 100% | Requiring inotropic support for >14 days, inhaled NO > 48 hours or sildenafil/iloprost on cessation of iNO, or right-sided mechanical support | 32.5% |
Topilsky et al (2011)27 | N=83, BTT 33%, DT 67%, mean age 63, male 81% | Single | Continuous 100% | Requiring RVAD, or inotropic support>7 days post-operatively | 27.7% |
Kukucka et al (2011)28 | N=115, mean age 54, male 90% | Single | Continuous 100% | 2 of the following: MAP <55 mmHg; CVP>16 mm Hg; MvO2 <55%; CI <2 L/min/m2; inotropic support score >20 units; or need for an RVAD | 13.0% |
Kato et al (2012)29 | N=111, mean age 56, male 78% | Single | Continuous 71%, Pulsatile 29% | Requiring inotropic support for >14 days or restarting inotropes after 14 days, inhaled NO >48 hours, or RVAD | 31.5% |
Grant et al (2012)10 | N=117, BTT 67%, DT 33%, mean age 58, male 79% | Single | Continuous 100% | Requiring inotropic support for >14 days, or RVAD support | 40.0% |
Kukucka et al (2012)30 | N=122, mean age 55, male 88% | Single | Continuous 100% | Two of the following criteria in the absence of cardiac tamponade: < 48 hours post-surgery: MAP<55 mm Hg; CVP>16 mm Hg; MvO2 < 55%; CI < 2 liters/min/m2; inotropic support score >20 units | 12.3% |
Raina et al (2013)31 | N=55, mean age 54, male 71% | Single | Continuous 93%, Pulsatile 7% | Requiring inotropic support for ≥14 days, or RVAD support | 29.0% |
Kato et al (2013)32 | N=68, mean age 63, male 90% | Single | Continuous 100% | Requiring inotropic support or pulmonary vasodilators at 14 days post-operatively, or RVAD support | 35.3% |
Vivo et al (2013)33 | N=109, BTT 49%, DT 49%, BTD 2%, mean age 54, male 77% | Single | Continuous 100% | Requiring inotropic support for ≥14 days, or RVAD support | 22.9% |
Atluri et al (2013)9 | N=218, mean age 56, male 81% | Single | Continuous 100% | Requiring BiVAD placement | 23.4% |
BiVAD: biventricular assist device; BTD: bridge to decision; BTT: bridge to transplant; BTR: bridge to recovery; CI: cardiac index; CVP: central venous pressure; DT: destination therapy; ECMO: extracorporeal membrane oxygenation; LVAD: left ventricular assist device; MAP: mean arterial pressure; MvO2: mixed venous oxygen saturation; NO: nitric oxide; RVAD: right ventricular assist device; RVF: right ventricular failure.
Not explicitly defined; however, bridge-to-transplantation rate was calculated with the entire cohort in the denominator. For studies without BTT vs. DT information, indication was not reported in the original publication.
Although the need for placement of an RVAD might be viewed as an indicator of more “definitive” RVF, the outcomes of patients presenting with RVF, defined as need for prolonged inotropic or other support besides RVAD, are equally unfavorable (Supplemental Table 1). Therefore, it would be important to include need for prolonged (or delayed) inotrope use when defining RVF. In this respect, following the current INTERMACS scheme for RVF severity might be reasonable. This scheme suggests that RVF be reported as (1) severe, if RVAD was implanted; (2) moderate, when inotropes or intravenous or inhaled pulmonary vasodilators were used for longer than 1 week at any time after LVAD implantation; and (3) mild, when a combination of ≥ 2 symptoms and signs are present (central venous pressure >18 mmHg, cardiac index CI < 2.3 L/min/m2, ascites, moderate to severe peripheral edema, or echocardiographic or physical examination evidence of elevated central venous pressure) but without need for RVAD or prolonged inotropic and/or vasodilator support. The latter category would not be reported in a dichotomous definition of RVF. These more granular definitions incorporate prolonged use of inotropes, which is clearly associated with adverse outcomes, and facilitate comparison between studies and reporting the effect of RVF on outcomes according to RVF severity.
Of note, no large study to date has prespecified a time frame for RVF (e.g. 30 days or 90 days). This would be important for consistency, which would facilitate a more valid comparison of rates and predictive modeling. In a recent large series,35 the time to unplanned RVAD insertion ranged from 1 to 66 days after LVAD surgery, and only 4 out of 44 patients required RVAD support later than 10 days post LVAD. Other groups have reported similar findings.13 However, data on the incidence and impact of RVF after the 90-day window are limited. Finally, quantitative RV function has not been incorporated into RVF definitions to date.
Pathophysiology of Post-LVAD Right Ventricular Failure
The pathophysiology of RVF post-LVAD is not well understood. Implantation of an LVAD promotes rapid and dramatic changes, as cardiac output is largely restored and left ventricular (LV) filling pressures are relieved. This in turn leads to an increase in RV preload and decrease in RV afterload.36 The immediate decrease in RV afterload, however, is mostly the result of reduced LV filling pressures,37, 38 and the observed immediate reduction in pulmonary vascular resistance (PVR) is primarily the result of increased cardiac output.37 Reversal of structural changes in the pulmonary vasculature (previously considered “fixed” pulmonary hypertension) and further reduction in PVR may also occur, but take more time to realize.39, 40 In the interim, the acute rise in venous return in conjunction with perioperative transfusion and intravenous fluid administration, may further strain the RV, thus increasing wall stress and exacerbating tricuspid regurgitation (TR).41
Alterations in hemodynamics only partially explain changes in RV function post LVAD. The left and right ventricles are interdependent, by virtue of the interventricular septum, interlacing muscle fibers, and the pericardium.42 The effects of LVAD support on the septum and RV function have been controversial. In animal studies, LVAD support induces ischemia of the septum with decreased thickening43 and altered septal geometry.44 However, in other animal models, septal shift did not affect RV function.45 It is difficult though to extrapolate findings from healthy animals to humans with HF. In a study of 76 continuous-flow LVAD recipients, persistent leftward bowing of the septum 30 days post-implant was associated with significantly worse outcomes at 90 days.46 In an in silico cardiovascular-respiratory system model, increased LVAD pump speeds appear to reduce septal contribution to RV and LV ejection and increase RV work.47 Finally, patients with more severe RV dysfunction may be more prone to LV suction with higher LVAD pump speeds.48
Challenges in RVF Risk Prediction
Although preoperative assessment of RV function with contemporary imaging has a strong potential to improve RVF risk prediction, there are four major challenges. First, a host of intra-operative events may impair RV function, including increased pulmonary vascular resistance (PVR) secondary to suboptimal ventilation and alveolar hypoxia, hyperinflation, or atelectasis; acidosis causing ischemic injury; direct RV ischemia by injuring prior bypass grafts; and ischemia due to prolonged cardiopulmonary bypass, bleeding, transfusions, or shock.13, 49 Second, varying practices in concomitant procedures during LVAD implantation may also alter the course of RV function. For example, there is growing evidence that concomitant tricuspid valve repair (or replacement) for severe TR results in improved RV function post-implant.50 Third, there is wide inter-institutional variation in the use and duration of inotropes and the threshold for RVAD support after LVAD surgery. For example, extended (≥14 days) inotropic support after LVAD surgery ranged from <15% 13, 33 to >30% 10 in recent continuous-flow LVAD series. Fourth, the population of LVAD recipients is a moving target.2 As post-LVAD outcomes continue to improve, there is a clear trend towards more elective implantations in less sick patients over time 3 with destination therapy becoming rapidly the most frequent indication.3 Also, as LVAD technology evolves, new target populations may be identified.
Despite the aforementioned challenges and potential confounders, preoperative RV function has been identified consistently as a predictor of post-implant RVF. Several studies have considered preoperative clinical variables and cardiac imaging (i.e. echocardiography) as a means to better risk stratify LVAD recipients for RVF. In the following section we review the available evidence, primarily focusing on echocardiography.
Risk Prediction Models and Risk Factors for Right Ventricular Failure
Clinical Scores
Current RVF prediction scores rely on HF severity, laboratory evidence of end-organ damage, right heart hemodynamics, and history of cardiac procedures as surrogates of preoperative RV dysfunction (Table 2).6-9, 13, 16 Importantly, these models have been derived from retrospective studies in patients with mostly pulsatile-flow LVADs and thus do not fully represent the current LVAD population. The predictors considered vary significantly across studies, as do definitions of RVF. Of note, preoperative risk factors for worse post-LVAD outcomes are also risk factors for RVF; i.e., clinical HF status, high RV filling pressures, and low RV stroke work index (RVSWI).
Table 2.
Clinical Risk Prediction Scores for Right Ventricular Failure in Left Ventricular Assist Device Recipients
Study | Patients | RVF Definition & Rate | Multivariable Predictors | Echocardiographic RV Parameters Considered |
---|---|---|---|---|
Michigan RV Failure Risk Score6 (2008) | - 197 LVADs - 28 continuous-flow - 94% BTT |
- Need for RVAD - Need for inotropes - RVF rate: 35% |
- Preoperative vasopressors (4 pts) - AST ≥80 IU/L (2 pts) - Bilirubin ≥2.0 mg/dL (2.5 pts) - Creatinine ≥2.3 mg/dL (3 pts) |
- RV systolic function (visual semi-quantitative) - TR (visual semi-quantitative) |
Penn RVAD Risk Score7 (2008) | - 266 LVADs - 6 continuous-flow - BTT vs. DT not reported |
- Need for RVAD - Included ITT RVAD - RVF rate: 37% |
- Cardiac index ≤2.2 L/min/m2 - RVSWI ≤0.25 mmHg × L/m2 - Severe RV dysfunction - Creatinine ≥1.9 mg/dL - Prior cardiac surgery - Systolic BP ≤96 mmHg |
- RV systolic function (visual semi-quantitative) |
Utah RV Risk Score16 (2010) | - 175 LVADs - 25 continuous-flow - 58% BTT, 42% DT |
- Need for RVAD - Need for inotropes - Need for inhaled NO - RVF rate: 44% |
- DT indication (3.5 pts) - IABP (4 pts) - PVR (1-4 pts) - Inotrope dependency (2.5 pts) - Obesity (2 pts) - ACEI or ARB use (−2.5 points) - β-blocker use (2 pts) |
- Right atrial area |
Kormos13 (2010) | - 484 LVADs - All continuous flow - BTT 100% |
- Need RVAD - Need for inotropes - RVF Rate: 20.2% |
- CVP/PCWP>0.63 (OR 2.3) - Need for preoperative ventilator support (OR 5.5) - BUN>39 mg/dL (OR 2.1) |
- None |
Pittsburgh Decision Tree8 (2012) | - 183 LVADs - 40 continuous-flow - BTT vs. DT not reported |
- Need for RVAD - RVF rate: 15% |
Age, heart rate, transpulmonary gradient; right atrial pressure; INR, white blood cell count, ALT, number of inotropic agents | - None |
CRITT9 (2013) | - 167 LVADs, all continuous-flow - 51 BiVADs - BTT vs. DT not reported |
- Need for BiVAD - RVF rate: 23% |
- CVP >15 mmHg (C) - Severe RV dysfunction (R) - Preoperative intubation (I) - Severe TR (T) - Heart rate >100 (tachycardia [T]) |
- RV systolic function (visual semi-quantitative) - Severe TR (visual semi-quantitative) |
ACEI: angiotensin-converting enzyme inhibitor; ALT: alanine aminotransferase; ARB: angiotensin receptor blocker; AST: aspartate aminotransferase; BiVAD: biventricular assist device; BP: blood pressure; BTT: bridge-to-transplantation; CVP: central venous pressure; DT: destination therapy; IABP: intra-aortic balloon pump; INR: international normalized ratio; ITT: intension-to-treat; LVAD: left ventricular assist device; NO: nitric oxide; PVR: pulmonary vascular resistance; RV: right ventricle; RVAD: right ventricular assist device; RVF: right ventricular failure; RVSWI: right ventricular stroke work index; TR: tricuspid regurgitation
The application of RVF prediction scores in clinical practice has been limited, partially because of modest score performance, and only the Michigan score has seen some uptake in research and practice. Developed in 197 LVAD recipients (94% bridge to transplant; continuous-flow device in 28), of whom 68 (35%) developed RVF,6 the score incorporates preoperative vasopressor requirements, liver enzyme elevations, and creatinine levels as predictors of RVF. The c statistic in the deviation cohort was 0.73.6 However, in a retrospective study investigating the incremental value of quantitative RV parameters,10 the Michigan score had only modest discrimination (c=0.66) for RVF prediction. In a small validation study in continuous-flow LVAD recipients, no score predicted the need for RVAD.51 Our group has retrospectively assessed RVF scores in 69 continuous-flow LVAD recipients and observed that, although the Michigan score offered the highest discrimination (c=0.64), all scores had modest performance and low predictive values.52 Of note, despite several studies investigating various echocardio-graphic RV function parameters, no score has incorporated quantitative RV parameters to date.
Hemodynamic Parameters
Invasive Assessment of RV Systolic Function
Although RV ejection fraction can be derived by fluoroscopy, this method is rarely used in practice. Alternatively, Right-sided hemodynamics allows inference of RV function. RVSWI, defined as (mean pulmonary artery pressure – mean right atrial pressure)/stroke volume index, has been considered in several studies,11, 13, 22, 23, 25 but has not been uniformly shown to predict RVF in multivariate analyses.9, 10, 13, 23 Potential drawbacks of RVSWI are (1) being a derivative of multiple measures, and thus more prone to error, and (2) dependence on preload.53 Also, the use of serial RVSWI for follow-up after LVAD implantation is limited.
Pulmonary Vascular Resistance and Pulmonary Artery Systolic Pressure
Although long-term unloading of the LV can lead to clinically relevant reductions of PVR and potentially restore candidacy for transplantation,39 elevated preoperative PVR has been identified as an independent predictor of RVF16 though not consistently.10 It is interesting that in several studies preoperative PVR did not emerge as a predictor of RVF.6, 8, 13 This may be related to the fact that PVR is highly dependent on cardiac output and is therefore dramatically affected by LVAD-induced increases in cardiac output. Low pulmonary artery systolic pressure (PASP) has been considered a surrogate of severe RV dysfunction in advanced HF and has been associated with post-LVAD RVF in early studies.22, 23 However, this is probably seen only in the extreme end of RV dysfunction. In contemporary studies, low PASP was either a weak predictor6 or did not confer useful predictive information.10, 13, 16 This is probably because (1) PASP is a function of both PVR and RV systolic function and thus various combinations of these parameters can yield similar PASP values, blunting discrimination for RV dysfunction; and (2) LVAD support is now offered to a HF population with a smaller proportion of patients with extreme RV dysfunction from the outset and hence the predictive ability of low PASP is less relevant.
Echocardiography
Accurate determination of RV function with transthoracic echocardiography has proven to be challenging, secondary to the retrosternal position of the RV and its complex geometry. The RV inflow and outflow tracts are out of plane and difficult to image simultaneously. Post-implant changes and device-related artifacts further limit post-LVAD visualization and measurements. Despite these challenges, echocardiography is still the most available, feasible, affordable, and safe imaging modality for advanced HF patients. However, RV contractility is often subjectively classified into normal or mildly, moderately, or severely reduced. This semiquantitative approach has poor reproducibility, accuracy, and sensitivity to detect clinically relevant changes. The American Society of Echocardiography (ASE) has recently provided guidelines for objective, quantitative assessment of RV function.54 In Table 3, we summarize the available data on the predictive value of quantitative RV parameters in LVAD populations for RVF prediction.10, 24, 25, 27-29, 31-33
Table 3.
Quantitative Right-Sided Parameters Considered in Studies Evaluating Echocardiography for Prediction of Right Ventricular Failure
Study | Patients | Echocardiographic RV Parameters Considered | Univariate Echo Predictors | Multivariate Echo Predictors |
---|---|---|---|---|
Potapov et al (2008)25 | N=54, mean age 52, male 91% | RV short/long axis ratio, RV EDD, RA diameter, RV EF, TR | RV short/long axis ratio, RV EDD, RV EF, TR, LV EDD | - RV short/long axis ratio - TR grade III-IV |
Puwanant et al (2008)24 | N=33, mean age 54, BTT 67%, DT 21%, BTR 12%, | RV FAC, RV MPI, TAPSE, RA volume, TR, RV systolic pressure | TAPSE, RV systolic pressure | - Multivariate analysis not performed |
Kukucka et al (2011)28* | N=115, mean age 54, male 90% | R/L EDD ratio, RV areas, RV FAC RV volumes, RV EF, TAPSE | R/L EDD ratio | - Multivariate analysis not performed |
Topilsky et al (2011)27 | N=83, BTT 33%, DT 67%, mean age 63, male 81% | Qualitative RV function, TR velocity, TRDc, RV ejection time, TR vena contracta width, TAPSE, MPI | TRDc, MPI, TR-RV ejection time | - LVEDD, TRDc |
Grant et al (2012)10 | N=117, mean age 58, male 79%, BTT 67%, DT 33% | RV FAC, RV dysfunction (visual semi-quantitative), TAPSE, R/L ratio, TR, RV free wall strain | RV dysfunction (visual semi-quantitative), RV free wall strain | - RV free wall strain |
Kato et al (2012)29 | N=111, mean age 56, male 78% | RV FAC, TAPSE | LV EDD, LV ESD, LV EF, LA diameter to LV EDD (ratio) | - LVEDD - LVEF - LA diameter to LV EDD |
Raina et al (2013)31 | N=55, mean age 54, male 71%, BTT 65%, DT 35% | RV dimensions and areas, R/L EDD ratio, RV FAC, TAPSE, RA pressure, RV outflow tract VTI | RVFAC, RA pressure, LA volume | - None |
Kato et al (2013)32 | N=68, mean age 63, 90% male | RV EDD, RV FAC, TAPSE, TDI S’, TDI E’, RV E/E’, RV global strain | RVFAC, TAPSE, RV E/E’, RV global strain | - Multivariate analysis not performed |
Vivo et al (2013)33 | N=109, BTT 49%, DT 49%, BT 2%, mean age 54, male 77% | RV diameter, RV area, RV FAC, qualitative RV size and function, TAPSE, TDI S’, RV CI, TR velocity and severity, RA pressure, RV/LV diameter ratio, PASP | RV/LV diameter, RVSWI | - Increased RV/LV diameter ratio |
BTR: bridge-to-recovery; BTT: bridge-to-transplantation; CI: cardiac index; DT: destination therapy; E’: early diastolic tricuspid annular velocity; EDD: end-diastolic diameter; E/E’: tricuspid early inflow velocity to early diastolic annular velocity ratio; EF: ejection fraction; ESD: end-systolic diameter; FAC: fractional area change; LA: left atrial; LV: left ventricular; MPI: myocardial performance index; PASP: pulmonary artery systolic pressure; RA: right atrial; RV: right ventricular; RVSWI: right ventricular stroke work index; S’: tricuspid annular systolic velocity; TAPSE: tricuspid annular plane systolic excursion; TDI: tissue Doppler imaging; TR: tricuspid regurgitation; TRDc: duration of TR corrected for heart rate; VTI: velocity time integral.
Transesophageal echocardiography.
For studies without BTT vs. DT information, indication was not reported in the original publication.
M-mode Parameters
Normal RV function is highly dependent on longitudinal shortening.54 A simple approach to study RV longitudinal function is to place an M-mode cursor on the tricuspid annulus in an apical RV view and measure longitudinal annular displacement. Tricuspid annular plane systolic excursion (TAPSE) should be ≥1.6 cm in adults.54 Low TAPSE has been associated with worse survival in pulmonary arterial hypertension.55, 56 However, TAPSE is a regional, linear parameter for a complex structure and is dependent on loading conditions and insonation angle (Figure 1). Although TAPSE has been reported to predict RVF in a small early study,24 this finding has not been confirmed in subsequent LVAD cohorts.10, 28, 29, 31
Figure 1.
Right ventricular fractional area change (RVFAC) estimation in (A) a 73-year old male with ischemic cardiomyopathy who has become inotrope dependent in the past few months and (B) a 43-year old female with nonischemic cardiomyopathy awaiting heart transplantation who required an intra-aortic balloon pump for stabilization. Both patients received a Heart Mate II device 10 days later as destination therapy and bridge-to-transplant, respectively. Patient A was weaned from inotropes on postoperative day 5, whereas patient B required inotropic support for 38 days. Note the heavy trabeculation and the strong echo from the defibrillator lead in B.
2-Dimensional Parameters
RV fractional area change (RVFAC) is a feasible quantitative alternative to “eyeballing” RV systolic function and correlates well with cardiac magnetic resonance imaging measurements.57, 58 A RVFAC <35% is considered abnormal.54 A >10% reduction in RVFAC at 1 month was associated with worse quality of life and poor exercise capacity in patients with an LVAD.59 However, the reproducibility and prognostic value of RVFAC in LVAD patients is uncertain, as the available studies report suboptimal quality of assessments and fair correlation coefficients for observer variability (0.50-0.76).59, 60 Heavy trabeculation and pacemaker or defibrillator lead artifacts are common challenges when evaluating RVFAC in patients with advanced HF (Figure 2). In a small retrospective study that included planned RV support in the definition of RVF,31 lower RVFAC predicted RVF. However, RVFAC did not predict RVF in another larger study29 and was only marginally or not predictive in other retrospective studies.10, 24, 28
Figure 2.
Tricuspid annular systolic plane excursion (TAPSE) in the same patients as in Figure 1. Patient A exhibits a very low TAPSE (0.7 cm), indicative of severe RV dysfunction, despite the mildly abnormal RVFAC. In contrast, patient B has a low-normal TAPSE (1.6 cm) despite severely depressed RVFAC. However, as evident from the elevated heart rate (105 bpm), patient B is in hemodynamic stress, which is accompanied by exaggerated translational motion of the heart, thus overestimating TAPSE. The use of color M-mode, although not necessary for TAPSE estimation, facilitates detection of directional changes in motion.
The size of the RV, mainly in the form of RV/LV diameter ratio, is surrogate of RV contractility and in some studies demonstrated strong association with RVF.28, 33 Other studies, however, failed to confirm this finding.10, 31 The main challenge with the RV/LV diameter ratio is the standardization of the acquisition plane and level (basal, mid, apical). In a study specifically assessing reproducibility of the RV/LV diameter ratio,29 the intra- and inter-observer interclass correlation coefficients were 0.57 and 0.61, respectively.
Doppler and Tissue Doppler Based Parameters
Although severe TR has been identified as a strong predictor of RVF in some studies,9, 17, 25 other, large studies did not find such an association.6, 10 Shorter TR duration, presumably a marker of early equalization of right ventricular and atrial pressures, was associated with worse outcomes in a cohort of continuous-flow LVAD recipients.27 However, TR has not been uniformly evaluated in studies investigating RVF predictors,13, 16 including echocardiography-focused studies.32 Recently, tricuspid annular dilation assessed by transesophageal echocardiography was identified as a predictor of mid-term post-LVAD survival among patients without preoperative severe TR.30
Despite the strong association with RV dysfunction in the general HF population, TR may not a suitable marker of RV function in contemporary LVAD candidates. First, the majority of these patients have previously received an implantable defibrillator with or without cardiac resynchronization capability.61 Pacemaker and defibrillator leads cause or worsen TR in a considerable proportion of recipients.62 Although the clinical implications of lead-induced TR are still unclear and severe TR is uncommon, this effect nevertheless blunts the value of TR as a RV function surrogate. Second, a growing number of LVAD recipients undergo simultaneous procedures for severe TR, thus confounding the use of TR as a predictor of RVF.
Tissue Doppler imaging (TDI) is an attractive alternative to TAPSE since myocardial velocities are easy to obtain and reproduce. Systolic velocity of the tricuspid annulus reflects longitudinal RV function. However, velocities depend on insonation angle and loading conditions (Figure 3). Also, translational motion of the heart and 'tethering' by adjacent diseased myocardial segments can produce velocities that are not representative of the performance of the interrogated segment. In a study of 68 LVAD recipients, systolic tricuspid annular velocity did not predict RVF.32
Figure 3.
Tricuspid annular systolic velocity (S’) by tissue Doppler imaging in the same patients as in Figure 1. Similar to tricuspid annular systolic plane excursion (TAPSE), longitudinal velocities are depended on load, translational motion, and insonation angle. Patient A exhibits abnormal S’ (<10 cm/s), whereas patient B has a low-normal S’ (11 cm/s), despite discordant RVFAC results.
The myocardial performance index (MPI), defined as the ratio of isovolumic time divided by the ejection time,54 is considered an indicator of global RV function. A RV MPI >0.40 by pulsed-wave Doppler or >0.55 by tissue Doppler indicates RV dysfunction.54 Longer isovolumic phases and shorter ejection time signify worse ventricular performance as expressed by a higher MPI. Although the MPI is fairly independent of heart rate and geometry, it is highly sensitive to loading conditions and less reliable with elevated right atrial pressures.54 Data are largely lacking, but in a small study the RV MPI was not a predictor of RVF post implant.24
Right Ventricular Mechanics
Strain expresses myocardial shortening or lengthening (in the longitudinal and circumferential directions), or thickening or thinning (in the radial direction), whereas strain rate is the rate of deformation over time. Strain and strain rate, collectively termed “ventricular mechanics”, reflect myocardial performance and provide a more direct assessment of myocardial contractility over ejection fraction or linear measurements. These indices may be calculated either from tissue velocity measurements (TDI) or from tracking of unique ultrasound speckles from frame-to-frame 2D (B-mode) images. Strain and strain rate are independent of ventricular morphology and are angle-independent when obtained by speckle tracking. However, strain measurements are intrinsically load-dependent. ASE does not currently recommend TDI-based regional RV mechanics because of high variability, lack of standardization, and paucity of normative data.54
Speckle tracking represents a promising alternative, due to (1) automated myocardial delineation and tracking and (2) insights into regional myocardial performance (Figure 4).54, 63, 64 The literature reporting on RV function using speckle tracking is growing. In a retrospective study of 117 patients undergoing LVAD implantation, RV free wall longitudinal strain by velocity vector imaging (a variant of speckle tracking using standard DICOM images) predicted RVF with 76% specificity and 68% sensitivity at a cutoff of -9.6%.10 In this study, RVF (defined as need for RVAD or inotropic agents for >14 days) was encountered in 47 patients (40%). RV free wall strain was a predictor of RVF in models including the Michigan score and had incremental predictive value, increasing the C statistic from 0.66 to 0.77 (P<0.01) and improving net reclassification.10 In another study of 68 patients undergoing elective LVAD surgery, RV longitudinal strain by speckle tracking was significantly impaired preoperatively (–12.6% ± 3.3% vs. –16.2% ± 4.3%, P<0.001) in 24 patients (35.3%) who experienced RVF by 14 days.32 Other groups have recently reported similar findings. 65, 66
Figure 4.
Right ventricular (RV) longitudinal strain by speckle tracking in the same patients as in Figure 1. In patient A (ischemic etiology), the interventricular septum demonstrates very low peak strain (normal values: −14±4% to −22±5% from base to apex)63 suggestive of severe systolic dysfunction, whereas the RV free wall demonstrates preserved systolic function, especially at the basal and mid segments (lower reference values: −18% to −20%).54, 64 In patient B (nonischemic etiology), the very low peak strain values across all segments and the very low global strain (−4.3%) suggest severe global RV systolic dysfunction. Note also the fragmented systolic activation in the linearized map (lower left panel) in patient B as compared to the more homogenous activation in patient A.
Although ASE does not currently recommend speckle-tracking parameters as a clinical tool secondary to cross-platform standardization issues and lack of definitive data, we believe that global and free-wall longitudinal RV mechanics (strain and strain rate), preferably assessed by speckle tracking, are probably the most promising approach for RVF prediction. Strain and especially strain rate are less sensitive to loading conditions and represent systolic RV function reasonably well.67 Also, the consistency of findings on the predictive value of longitudinal RV mechanics for RVF across different research groups is promising.10, 32, 65, 66 Of note, post-LVAD assessment of RV mechanics is feasible,32 hence facilitating tracking of RV function, and adds information incremental to that obtained from preoperative imaging.
Other Imaging
Multi-detector computed tomography (MDCT) has been successfully used to visualize the LVAD cannulae and detect post-operative complications.68 In LVAD recipients, RV ejection fraction by MDCT was highly reproducible and correlated well with RVFAC.60 However, data with MDCT in LVAD patients are limited. Also, the need for nephrotoxic contrast precludes routine use of MDCT in LVAD candidates, who often have concomitant kidney disease.
Conclusions and Future Directions
Right ventricular failure remains a challenge in the era of continuous-flow LVADs. Beyond the inherent challenges in the assessment of RV function, intraoperative events, concomitant surgical procedures, and post-operative changes in pulmonary hemodynamics and device settings may confound the ability of preoperative assessment of RV function to predict RVF. Additional challenges include evolving mechanical circulatory support technology and shifts in the target population. However, the full potential of contemporary imaging, especially echocardiography, has not been fully utilized in the preoperative assessment of LVAD candidates. In this direction, we believe that the following issues should be a priority.
Standardization of Echocardiographic Protocols
Preoperative RV function is still assessed subjectively in many patients. Therefore, an important first step would be the standardization of echocardiographic protocols before and after device implantation across LVAD centers. This would be greatly facilitated if the INTERMACS registry included a dedicated echocardiographic module; the currently required data are rudimentary (semiquantitative valvular regurgitation, LVEF, LVEDD). A common module for all centers could include quantitative data on LV and RV structure and function and a standard set of Doppler-based hemodynamics. A pilot RV mechanics module for centers able to contribute mechanics data could be launched also and expanded over time.
Refinement of RVF Definition and Serial Assessment of RV Function
Currently, the definition of RVF does not include any parameters of RV function. One way to incorporate RV systolic function parameters would be to require several echocardiographic criteria for RVF, e.g. the presence of 2 out of 4 standard predefined RV dysfunction criteria (low TAPSE, low RVFAC, low tricuspid s’, or high RIMP – using the ASE cut-offs) in addition to hemodynamic criteria and signs/symptoms. This could increase the specificity to overcome the challenges associated with single parameters as discussed above. A single parameter might be misleading, but if multiple parameters point to the same direction, then chances are that “true” RV dysfunction is highly likely. In fact, in delayed RVF cases when central lines have been already removed, the RV dysfunction criteria might be used in lieu of hemodynamics. These parameters are feasible to measure in the postoperative period.32 In the same sense, quantitative RV systolic function parameters can be used to track RV function over time postoperatively and potentially track the effect of afterload-reducing agents on RV function. Also, longer-term serial assessment of RV function would be important, since pulmonary hemodynamics change over time post LVAD placement.40
Merging Research and Practice
The entire field is moving fast. Therefore, it is important for research and practice to converge. For example, it would be important for every center to acquire their own experience with echocardiographic parameters and communicate the results and outcomes with the LVAD team. From a practical perspective, we believe that based on the current data it would be reasonable to cautiously re-evaluate candidacy for LVAD – and perhaps consider alternatives – for patients with an absolute value of RV free wall or global strain of <10%, especially in conjunction with clinical surrogates of RV dysfunction.10
In the same spirit, leveraging INTERMACS would be the fastest avenue forward. The derivation and validation of predictive models based on clinical and echocardiographic variables in the INTERMACS database would (1) leverage existing infrastructure; (2) ensure multi-center input and applicability of results; and (3) allow for regular updating of selection algorithms as both technology and target population evolve.
Supplementary Material
Acknowledgments
Sources of Funding
This work has been partially supported by (1) the Emory University Research Committee and the Atlanta Clinical & Translational Science Institute (PHS Grant UL1 RR025008 from the Clinical and Translational Science Award program, National Institutes of Health, National Center for Research Resources) and (2) an American Heart Association Scientist Development Grant (13SDG15960001)
Footnotes
Disclosures
None.
References
- 1.Roger VL, Go AS, Lloyd-Jones DM, Benjamin EJ, Berry JD, Borden WB, Bravata DM, Dai S, Ford ES, Fox CS, Fullerton HJ, Gillespie C, Hailpern SM, Heit JA, Howard VJ, Kissela BM, Kittner SJ, Lackland DT, Lichtman JH, Lisabeth LD, Makuc DM, Marcus GM, Marelli A, Matchar DB, Moy CS, Mozaffarian D, Mussolino ME, Nichol G, Paynter NP, Soliman EZ, Sorlie PD, Sotoodehnia N, Turan TN, Virani SS, Wong ND, Woo D, Turner MB, American Heart Association Statistics C. Stroke Statistics S Heart disease and stroke statistics--2012 update: a report from the American Heart Association. Circulation. 2012;125:e2–e220. doi: 10.1161/CIR.0b013e31823ac046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Miller LW, Guglin M. Patient selection for ventricular assist devices: a moving target. J Am Coll Cardiol. 2013;61:1209–1221. doi: 10.1016/j.jacc.2012.08.1029. [DOI] [PubMed] [Google Scholar]
- 3.Kirklin JK, Naftel DC, Kormos RL, Stevenson LW, Pagani FD, Miller MA, Timothy Baldwin J, Young JB. Fifth INTERMACS annual report: risk factor analysis from more than 6,000 mechanical circulatory support patients. J Heart Lung Transplant. 2013;32:141–156. doi: 10.1016/j.healun.2012.12.004. [DOI] [PubMed] [Google Scholar]
- 4.Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, Sun B, Tatooles AJ, Delgado RM, 3rd, Long JW, Wozniak TC, Ghumman W, Farrar DJ, Frazier OH, HeartMate Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009;361:2241–2251. doi: 10.1056/NEJMoa0909938. [DOI] [PubMed] [Google Scholar]
- 5.Drakos SG, Janicki L, Horne BD, Kfoury AG, Reid BB, Clayson S, Horton K, Haddad F, Li DY, Renlund DG, Fisher PW. Risk factors predictive of right ventricular failure after left ventricular assist device implantation. Am J Cardiol. 2010;105:1030–1035. doi: 10.1016/j.amjcard.2009.11.026. [DOI] [PubMed] [Google Scholar]
- 6.Matthews JC, Koelling TM, Pagani FD, Aaronson KD. The right ventricular failure risk score a pre-operative tool for assessing the risk of right ventricular failure in left ventricular assist device candidates. J Am Coll Cardiol. 2008;51:2163–2172. doi: 10.1016/j.jacc.2008.03.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Fitzpatrick JR, Frederick JR, Hsu VM, Kozin ED, O'Hara ML, Howell E, Dougherty D, McCormick RC, Laporte CA, Cohen JE, Southerland KW, Howard JL, Jessup ML, Morris RJ, Acker MA, Woo YJ. Risk score derived from pre-operative data analysis predicts the need for biventricular mechanical circulatory support. J Heart Lung Transplant. 2008;27:1286–1292. doi: 10.1016/j.healun.2008.09.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Wang Y, Simon MA, Bonde P, Harris BU, Teuteberg JJ, Kormos RL, Antaki JF. Decision tree for adjuvant right ventricular support in patients receiving a left ventricular assist device. J Heart Lung Transplant. 2012;31:140–149. doi: 10.1016/j.healun.2011.11.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Atluri P, Goldstone AB, Fairman AS, Macarthur JW, Shudo Y, Cohen JE, Acker AL, Hiesinger W, Howard JL, Acker MA, Woo YJ. Predicting Right Ventricular Failure in the Modern, Continuous Flow Left Ventricular Assist Device Era. Ann Thorac Surg. 2013 doi: 10.1016/j.athoracsur.2013.03.099. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Grant AD, Smedira NG, Starling RC, Marwick TH. Independent and incremental role of quantitative right ventricular evaluation for the prediction of right ventricular failure after left ventricular assist device implantation. J Am Coll Cardiol. 2012;60:521–528. doi: 10.1016/j.jacc.2012.02.073. [DOI] [PubMed] [Google Scholar]
- 11.Kavarana MN, Pessin-Minsley MS, Urtecho J, Catanese KA, Flannery M, Oz MC, Naka Y. Right ventricular dysfunction and organ failure in left ventricular assist device recipients: a continuing problem. Ann Thorac Surg. 2002;73:745–750. doi: 10.1016/s0003-4975(01)03406-3. [DOI] [PubMed] [Google Scholar]
- 12.Genovese EA, Dew MA, Teuteberg JJ, Simon MA, Kay J, Siegenthaler MP, Bhama JK, Bermudez CA, Lockard KL, Winowich S, Kormos RL. Incidence and patterns of adverse event onset during the first 60 days after ventricular assist device implantation. Ann Thorac Surg. 2009;88:1162–1170. doi: 10.1016/j.athoracsur.2009.06.028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Kormos RL, Teuteberg JJ, Pagani FD, Russell SD, John R, Miller LW, Massey T, Milano CA, Moazami N, Sundareswaran KS, Farrar DJ, HeartMate IICI Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on outcomes. J Thorac Cardiovasc Surg. 2010;139:1316–1324. doi: 10.1016/j.jtcvs.2009.11.020. [DOI] [PubMed] [Google Scholar]
- 14.Dang NC, Topkara VK, Mercando M, Kay J, Kruger KH, Aboodi MS, Oz MC, Naka Y. Right heart failure after left ventricular assist device implantation in patients with chronic congestive heart failure. J Heart Lung Transplant. 2006;25:1–6. doi: 10.1016/j.healun.2005.07.008. [DOI] [PubMed] [Google Scholar]
- 15.Morgan JA, John R, Lee BJ, Oz MC, Naka Y. Is severe right ventricular failure in left ventricular assist device recipients a risk factor for unsuccessful bridging to transplant and post-transplant mortality. Ann Thorac Surg. 2004;77:859–863. doi: 10.1016/j.athoracsur.2003.09.048. [DOI] [PubMed] [Google Scholar]
- 16.Drakos SG, Janicki L, Horne BD, Kfoury AG, Reid BB, Clayson S, Horton K, Haddad F, Li DY, Renlund DG, Fisher PW. Risk factors predictive of right ventricular failure after left ventricular assist device implantation. Am J Cardiol. 2010;105:1030–1035. doi: 10.1016/j.amjcard.2009.11.026. [DOI] [PubMed] [Google Scholar]
- 17.Baumwol J, Macdonald PS, Keogh AM, Kotlyar E, Spratt P, Jansz P, Hayward CS. Right heart failure and “failure to thrive” after left ventricular assist device: clinical predictors and outcomes. J Heart Lung Transplant. 2011;30:888–895. doi: 10.1016/j.healun.2011.03.006. [DOI] [PubMed] [Google Scholar]
- 18.Holman WL, Kormos RL, Naftel DC, Miller MA, Pagani FD, Blume E, Cleeton T, Koenig SC, Edwards L, Kirklin JK. Predictors of death and transplant in patients with a mechanical circulatory support device: a multi-institutional study. J Heart Lung Transplant. 2009;28:44–50. doi: 10.1016/j.healun.2008.10.011. [DOI] [PubMed] [Google Scholar]
- 19.Lietz K, Long JW, Kfoury AG, Slaughter MS, Silver MA, Milano CA, Rogers JG, Naka Y, Mancini D, Miller LW. Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era: implications for patient selection. Circulation. 2007;116:497–505. doi: 10.1161/CIRCULATIONAHA.107.691972. [DOI] [PubMed] [Google Scholar]
- 20.Schenk S, McCarthy PM, Blackstone EH, Feng J, Starling RC, Navia JL, Zhou L, Hoercher KJ, Smedira NG, Fukamachi K. Duration of inotropic support after left ventricular assist device implantation: risk factors and impact on outcome. J Thorac Cardiovasc Surg. 2006;131:447–454. doi: 10.1016/j.jtcvs.2005.09.031. [DOI] [PubMed] [Google Scholar]
- 21.Deng MC, Loebe M, El-Banayosy A, Gronda E, Jansen PG, Vigano M, Wieselthaler GM, Reichart B, Vitali E, Pavie A, Mesana T, Loisance DY, Wheeldon DR, Portner PM. Mechanical circulatory support for advanced heart failure: effect of patient selection on outcome. Circulation. 2001;103:231–237. doi: 10.1161/01.cir.103.2.231. [DOI] [PubMed] [Google Scholar]
- 22.Fukamachi K, McCarthy PM, Smedira NG, Vargo RL, Starling RC, Young JB. Preoperative risk factors for right ventricular failure after implantable left ventricular assist device insertion. Ann Thorac Surg. 1999;68:2181–2184. doi: 10.1016/s0003-4975(99)00753-5. [DOI] [PubMed] [Google Scholar]
- 23.Ochiai Y, McCarthy PM, Smedira NG, Banbury MK, Navia JL, Feng J, Hsu AP, Yeager ML, Buda T, Hoercher KJ, Howard MW, Takagaki M, Doi K, Fukamachi K. Predictors of severe right ventricular failure after implantable left ventricular assist device insertion: analysis of 245 patients. Circulation. 2002;106:I198–202. [PubMed] [Google Scholar]
- 24.Puwanant S, Hamilton KK, Klodell CT, Hill JA, Schofield RS, Cleeton TS, Pauly DF, Aranda JM., Jr. Tricuspid annular motion as a predictor of severe right ventricular failure after left ventricular assist device implantation. J Heart Lung Transplant. 2008;27:1102–1107. doi: 10.1016/j.healun.2008.07.022. [DOI] [PubMed] [Google Scholar]
- 25.Potapov EV, Stepanenko A, Dandel M, Kukucka M, Lehmkuhl HB, Weng Y, Hennig F, Krabatsch T, Hetzer R. Tricuspid incompetence and geometry of the right ventricle as predictors of right ventricular function after implantation of a left ventricular assist device. J Heart Lung Transplant. 2008;27:1275–1281. doi: 10.1016/j.healun.2008.08.012. [DOI] [PubMed] [Google Scholar]
- 26.Hennig F, Stepanenko AV, Lehmkuhl HB, Kukucka M, Dandel M, Krabatsch T, Hetzer R, Potapov EV. Neurohumoral and inflammatory markers for prediction of right ventricular failure after implantation of a left ventricular assist device. Gen Thorac Cardiovasc Surg. 2011;59:19–24. doi: 10.1007/s11748-010-0669-9. [DOI] [PubMed] [Google Scholar]
- 27.Topilsky Y, Oh JK, Shah DK, Boilson BA, Schirger JA, Kushwaha SS, Pereira NL, Park SJ. Echocardiographic predictors of adverse outcomes after continuous left ventricular assist device implantation. JACC Cardiovasc Imaging. 2011;4:211–222. doi: 10.1016/j.jcmg.2010.10.012. [DOI] [PubMed] [Google Scholar]
- 28.Kukucka M, Stepanenko A, Potapov E, Krabatsch T, Redlin M, Mladenow A, Kuppe H, Hetzer R, Habazettl H. Right-to-left ventricular end-diastolic diameter ratio and prediction of right ventricular failure with continuous-flow left ventricular assist devices. J Heart Lung Transplant. 2011;30:64–69. doi: 10.1016/j.healun.2010.09.006. [DOI] [PubMed] [Google Scholar]
- 29.Kato TS, Farr M, Schulze PC, Maurer M, Shahzad K, Iwata S, Homma S, Jorde U, Takayama H, Naka Y, Gillam L, Mancini D. Usefulness of two-dimensional echocardiographic parameters of the left side of the heart to predict right ventricular failure after left ventricular assist device implantation. Am J Cardiol. 2012;109:246–251. doi: 10.1016/j.amjcard.2011.08.040. [DOI] [PubMed] [Google Scholar]
- 30.Kukucka M, Stepanenko A, Potapov E, Krabatsch T, Kuppe H, Habazettl H. Impact of tricuspid valve annulus dilation on mid-term survival after implantation of a left ventricular assist device. J Heart Lung Transplant. 2012;31:967–971. doi: 10.1016/j.healun.2012.06.003. [DOI] [PubMed] [Google Scholar]
- 31.Raina A, Seetha Rammohan HR, Gertz ZM, Rame JE, Woo YJ, Kirkpatrick JN. Postoperative right ventricular failure after left ventricular assist device placement is predicted by preoperative echocardiographic structural, hemodynamic, and functional parameters. J Card Fail. 2013;19:16–24. doi: 10.1016/j.cardfail.2012.11.001. [DOI] [PubMed] [Google Scholar]
- 32.Kato TS, Jiang J, Schulze PC, Jorde U, Uriel N, Kitada S, Takayama H, Naka Y, Mancini D, Gillam L. Serial Echocardiography Using Tissue Doppler and Speckle Tracking Imaging to Monitor Right Ventricular Failure Before and After Left Ventricular Assist Device Surgery. JACC: Heart Failure. 2013;1:216–222. doi: 10.1016/j.jchf.2013.02.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Vivo RP, Cordero-Reyes AM, Qamar U, Garikipati S, Trevino AR, Aldeiri M, Loebe M, Bruckner BA, Torre-Amione G, Bhimaraj A, Trachtenberg BH, Estep JD. Increased right-to-left ventricle diameter ratio is a strong predictor of right ventricular failure after left ventricular assist device. J Heart Lung Transplant. 2013;32:792–799. doi: 10.1016/j.healun.2013.05.016. [DOI] [PubMed] [Google Scholar]
- 34.Kirklin JK, Naftel DC, Stevenson LW, Kormos RL, Pagani FD, Miller MA, Ulisney K, Young JB. INTERMACS database for durable devices for circulatory support: first annual report. J Heart Lung Transplant. 2008;27:1065–1072. doi: 10.1016/j.healun.2008.07.021. [DOI] [PubMed] [Google Scholar]
- 35.Takeda K, Naka Y, Yang JA, Uriel N, Colombo PC, Jorde UP, Takayama H. Outcome of unplanned right ventricular assist device support for severe right heart failure after implantable left ventricular assist device insertion. J Heart Lung Transplant. 2013 doi: 10.1016/j.healun.2013.06.025. [DOI] [PubMed] [Google Scholar]
- 36.Rich JD. Right ventricular failure in patients with left ventricular assist devices. Cardiol Clin. 2012;30:291–302. doi: 10.1016/j.ccl.2012.03.008. [DOI] [PubMed] [Google Scholar]
- 37.Zimpfer D, Zrunek P, Roethy W, Czerny M, Schima H, Huber L, Grimm M, Rajek A, Wolner E, Wieselthaler G. Left ventricular assist devices decrease fixed pulmonary hypertension in cardiac transplant candidates. J Thorac Cardiovasc Surg. 2007;133:689–695. doi: 10.1016/j.jtcvs.2006.08.104. [DOI] [PubMed] [Google Scholar]
- 38.Miller LW, Pagani FD, Russell SD, John R, Boyle AJ, Aaronson KD, Conte JV, Naka Y, Mancini D, Delgado RM, MacGillivray TE, Farrar DJ, Frazier OH, HeartMate IICI Use of a continuous-flow device in patients awaiting heart transplantation. N Engl J Med. 2007;357:885–896. doi: 10.1056/NEJMoa067758. [DOI] [PubMed] [Google Scholar]
- 39.Mikus E, Stepanenko A, Krabatsch T, Loforte A, Dandel M, Lehmkuhl HB, Hetzer R, Potapov EV. Reversibility of fixed pulmonary hypertension in left ventricular assist device support recipients. Eur J Cardiothorac Surg. 2011;40:971–977. doi: 10.1016/j.ejcts.2011.01.019. [DOI] [PubMed] [Google Scholar]
- 40.Kutty RS, Parameshwar J, Lewis C, Catarino PA, Sudarshan CD, Jenkins DP, Dunning JJ, Tsui SS. Use of centrifugal left ventricular assist device as a bridge to candidacy in severe heart failure with secondary pulmonary hypertension. Eur J Cardiothorac Surg. 2013;43:1237–1242. doi: 10.1093/ejcts/ezs678. [DOI] [PubMed] [Google Scholar]
- 41.Holman WL, Bourge RC, Fan P, Kirklin JK, Pacifico AD, Nanda NC. Influence of left ventricular assist on valvular regurgitation. Circulation. 1993;88:II309–318. [PubMed] [Google Scholar]
- 42.Dell'Italia LJ. Anatomy and physiology of the right ventricle. Cardiol Clin. 2012;30:167–187. doi: 10.1016/j.ccl.2012.03.009. [DOI] [PubMed] [Google Scholar]
- 43.Daly RC, Chandrasekaran K, Cavarocchi NC, Tajik AJ, Schaff HV. Ischemia of the interventricular septum. A mechanism of right ventricular failure during mechanical left ventricular assist. J Thorac Cardiovasc Surg. 1992;103:1186–1191. [PubMed] [Google Scholar]
- 44.Moon MR, DeAnda A, Castro LJ, Daughters GT, 2nd, Ingels NB Jr., Miller DC. Effects of mechanical left ventricular support on right ventricular diastolic function. J Heart Lung Transplant. 1997;16:398–407. [PubMed] [Google Scholar]
- 45.Hendry PJ, Ascah KJ, Rajagopalan K, Calvin JE. Does septal position affect right ventricular function during left ventricular assist in an experimental porcine model? Circulation. 1994;90:II353–358. [PubMed] [Google Scholar]
- 46.Topilsky Y, Hasin T, Oh JK, Borgeson DD, Boilson BA, Schirger JA, Clavell AL, Frantz RP, Tsutsui R, Liu M, Maltais S, Kushwaha SS, Pereira NL, Park SJ. Echocardiographic variables after left ventricular assist device implantation associated with adverse outcome. Circ Cardiovasc Imaging. 2011;4:648–661. doi: 10.1161/CIRCIMAGING.111.965335. [DOI] [PubMed] [Google Scholar]
- 47.Ramachandran DP, Luo C, Ma TS, Clark JW. Modeling study of the failing heart and its interaction with an implantable rotary blood pump. Conf Proc IEEE Eng Med Biol Soc. 2011;2011:2403–2409. doi: 10.1109/IEMBS.2011.6090670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Hayward CS, Salamonsen R, Keogh AM, Woodard J, Ayre P, Prichard R, Walker R, Kotlyar E, Macdonald PS, Jansz P, Spratt P. Effect of alteration in pump speed on pump output and left ventricular filling with continuous-flow left ventricular assist device. ASAIO J. 2011;57:495–500. doi: 10.1097/MAT.0b013e318233b112. [DOI] [PubMed] [Google Scholar]
- 49.Marzec LN, Ambardekar AV. Preoperative Evaluation and Perioperative Management of Right Ventricular Failure After Left Ventricular Assist Device Implantation. Semin Cardiothorac Vasc Anesth. 2013 doi: 10.1177/1089253213488246. [DOI] [PubMed] [Google Scholar]
- 50.Piacentino V, 3rd, Ganapathi AM, Stafford-Smith M, Hsieh MK, Patel CB, Simeone AA, Rogers JG, Milano CA. Utility of concomitant tricuspid valve procedures for patients undergoing implantation of a continuous-flow left ventricular device. J Thorac Cardiovasc Surg. 2012;144:1217–1221. doi: 10.1016/j.jtcvs.2012.07.064. [DOI] [PubMed] [Google Scholar]
- 51.Pettinari M, Jacobs S, Rega F, Verbelen T, Droogne W, Meyns B. Are right ventricular risk scores useful? Eur J Cardiothorac Surg. 2012;42:621–626. doi: 10.1093/ejcts/ezs104. [DOI] [PubMed] [Google Scholar]
- 52.Kalogeropoulos AP, Siwamogsatham S, Weinberger JF, Kelkar AA, Georgiopoulou VV, Pekarek A, Wittersheim KM, Gupta D, Cole RT, Laskar SR, Nguyen DQ, Sims DB, Butler J, Vega JD. Clinical Scores and Echocardiography for Right Ventricular Failure Risk Prediction after Implantation of Continuous-Flow Left Ventricular Assist Devices. J Heart Lung Transplant. 2013;32:S273. [Google Scholar]
- 53.Reed CE, Dorman BH, Spinale FG. Assessment of right ventricular contractile performance after pulmonary resection. Ann Thorac Surg. 1993;56:426–431. doi: 10.1016/0003-4975(93)90874-h. discussion 431-422. [DOI] [PubMed] [Google Scholar]
- 54.Rudski LG, Lai WW, Afilalo J, Hua L, Handschumacher MD, Chandrasekaran K, Solomon SD, Louie EK, Schiller NB. Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography endorsed by the European Association of Echocardiography, a registered branch of the European Society of Cardiology, and the Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2010;23:685–713. doi: 10.1016/j.echo.2010.05.010. quiz 786-688. [DOI] [PubMed] [Google Scholar]
- 55.Forfia PR, Fisher MR, Mathai SC, Housten-Harris T, Hemnes AR, Borlaug BA, Chamera E, Corretti MC, Champion HC, Abraham TP, Girgis RE, Hassoun PM. Tricuspid annular displacement predicts survival in pulmonary hypertension. Am J Respir Crit Care Med. 2006;174:1034–1041. doi: 10.1164/rccm.200604-547OC. [DOI] [PubMed] [Google Scholar]
- 56.Corciova FC, Arsenescu-Georgescu C. Prognostic factors in pulmonary hypertension. Maedica (Buchar) 2012;7:30–37. [PMC free article] [PubMed] [Google Scholar]
- 57.Lai WW, Gauvreau K, Rivera ES, Saleeb S, Powell AJ, Geva T. Accuracy of guideline recommendations for two-dimensional quantification of the right ventricle by echocardiography. Int J Cardiovasc Imaging. 2008;24:691–698. doi: 10.1007/s10554-008-9314-4. [DOI] [PubMed] [Google Scholar]
- 58.Anavekar NS, Gerson D, Skali H, Kwong RY, Yucel EK, Solomon SD. Two-dimensional assessment of right ventricular function: an echocardiographic-MRI correlative study. Echocardiography. 2007;24:452–456. doi: 10.1111/j.1540-8175.2007.00424.x. [DOI] [PubMed] [Google Scholar]
- 59.Lam KM, Ennis S, O'Driscoll G, Solis JM, Macgillivray T, Picard MH. Observations from non-invasive measures of right heart hemodynamics in left ventricular assist device patients. J Am Soc Echocardiogr. 2009;22:1055–1062. doi: 10.1016/j.echo.2009.06.006. [DOI] [PubMed] [Google Scholar]
- 60.Garcia-Alvarez A, Fernandez-Friera L, Lau JF, Sawit ST, Mirelis JG, Castillo JG, Pinney S, Anyanwu AC, Fuster V, Sanz J, Garcia MJ. Evaluation of right ventricular function and post-operative findings using cardiac computed tomography in patients with left ventricular assist devices. J Heart Lung Transplant. 2011;30:896–903. doi: 10.1016/j.healun.2011.03.009. [DOI] [PubMed] [Google Scholar]
- 61.Kalogeropoulos AP, Georgiopoulou VV, Giamouzis G, Smith AL, Agha SA, Waheed S, Laskar S, Puskas J, Dunbar S, Vega D, Levy WC, Butler J. Utility of the Seattle Heart Failure Model in patients with advanced heart failure. J Am Coll Cardiol. 2009;53:334–342. doi: 10.1016/j.jacc.2008.10.023. [DOI] [PubMed] [Google Scholar]
- 62.Al-Mohaissen MA, Chan KL. Prevalence and mechanism of tricuspid regurgitation following implantation of endocardial leads for pacemaker or cardioverter-defibrillator. J Am Soc Echocardiogr. 2012;25:245–252. doi: 10.1016/j.echo.2011.11.020. [DOI] [PubMed] [Google Scholar]
- 63.Mor-Avi V, Lang RM, Badano LP, Belohlavek M, Cardim NM, Derumeaux G, Galderisi M, Marwick T, Nagueh SF, Sengupta PP, Sicari R, Smiseth OA, Smulevitz B, Takeuchi M, Thomas JD, Vannan M, Voigt JU, Zamorano JL. Current and evolving echocardiographic techniques for the quantitative evaluation of cardiac mechanics: ASE/EAE consensus statement on methodology and indications endorsed by the Japanese Society of Echocardiography. J Am Soc Echocardiogr. 2011;24:277–313. doi: 10.1016/j.echo.2011.01.015. [DOI] [PubMed] [Google Scholar]
- 64.Kalogeropoulos AP, Georgiopoulou VV, Howell S, Pernetz MA, Fisher MR, Lerakis S, Martin RP. Evaluation of right intraventricular dyssynchrony by two-dimensional strain echocardiography in patients with pulmonary arterial hypertension. J Am Soc Echocardiogr. 2008;21:1028–1034. doi: 10.1016/j.echo.2008.05.005. [DOI] [PubMed] [Google Scholar]
- 65.Cameli M, Lisi M, Righini FM, Focardi M, Lunghetti S, Bernazzali S, Marchetti L, Biagioli B, Galderisi M, Maccherini M, Sani G, Mondillo S. Speckle tracking echocardiography as a new technique to evaluate right ventricular function in patients with left ventricular assist device therapy. J Heart Lung Transplant. 2013;32:424–430. doi: 10.1016/j.healun.2012.12.010. [DOI] [PubMed] [Google Scholar]
- 66.Kalogeropoulos A, Georgiopoulou V, Siwamogsatham S, Sims D, Kelkar A, Wittersheim K, Pekarek A, Pernetz M, Nguyen D, Smith A, Vega J, Butler J. Echocardiography for Right Ventricular Failure Prediction After Implantation of Left Ventricular Assist Devices: Preliminary Results From a Prospective Cohort Study. American Heart Association Scientific Sessions. 2013 [Google Scholar]
- 67.Jamal F, Bergerot C, Argaud L, Loufouat J, Ovize M. Longitudinal strain quantitates regional right ventricular contractile function. Am J Physiol Heart Circ Physiol. 2003;285:H2842–2847. doi: 10.1152/ajpheart.00218.2003. [DOI] [PubMed] [Google Scholar]
- 68.Raman SV, Sahu A, Merchant AZ, Louis LBt, Firstenberg MS, Sun B. Noninvasive assessment of left ventricular assist devices with cardiovascular computed tomography and impact on management. J Heart Lung Transplant. 2010;29:79–85. doi: 10.1016/j.healun.2009.06.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.