Abstract
Objective
The creation of consistenly transmural lesions with epicardial ablation on the beating heart has represented a significant challenge for current technology. This study examined the chronic performance of the AtriCure Coolrail™ device, an internally cooled, bipolar radiofrequency ablation device designed for off-pump, epicardial ablation. It also examined the reliability of using acute intraoperative conduction delay to evaluate lesion integrity.
Methods
Seven swine underwent median sternotomy. The right atrial appendage and inferior vena cava were isolated with a bipolar radiofrequency clamp. Linear ablation lines were created between these structures with the AtriCure Coolrail™. Paced activation maps were recorded with epicardial patch electrodes acutely before and after ablation and after surviving the animals for four weeks. The conduction time across the linear ablation was calculated from these maps. The lesions were histologically evaluated with trichrome staining.
Results
Only 76% of cross sections of Coolrail™ lesions were transmural, and only one of twelve ablation lines was transmural in every cross section examined. Mapping data were available in five of the animals. Significant conduction delay was present after the creation of each line of ablation acutely; however, after 4 weeks, conduction time returned to pre-ablation values, demonstrating lack of transmurality.
Conclusion
The AtriCure Coolrail™ failed to reliably create transmural lesions. While it was able to create acute conduction delay, its failure to transmurally ablate the atrial myocardium left gaps along the length of the lesion which resulted in neither chronic conduction block nor delay across any line of ablation.
The Cox-Maze procedure, introduced in 1987, was the first successful surgical treatment of atrial fibrillation. It involved creating incisions in both atria to interrupt the reentrant circuits that were thought to be responsible for atrial fibrillation. In long term follow-up from our institution, the Cox-Maze III procedure (CMP) achieved over 90% success in curing symptomatic atrial fibrillation1. Unfortunately, the cut-and-sew procedure was technically difficult and time-consuming. Over the past decade, alternative energy sources have been used to replace the incisions of the CMP. These have included radiofrequency energy, laser, cryoablation, microwave, and high frequency ultrasound2-5. While simpler and less invasive than the CMP, these ablation-based procedures often still involved cardiopulmonary bypass. In order to develop a less invasive, beating heart approach, devices have been developed to perform epicardial ablation to circumvent the need for cardiopulmonary bypass. However, despite the introduction of a variety of energy sources and the use of supplemental irrigation and/or suction, no device has yet been able to create reliable continuous transmural lesions6. This study had two specific aims. First, the chronic performance of a novel internally cooled radiofrequency device designed for epicardial atrial ablation, the AtriCure Coolrail™, was examined. This device has been used clinically to create linear lesions but has not undergone previous chronic experimental evaluation. Second, this experiment was designed to evaluate the reliability of using intraoperative conduction delay and block as a means to predict the chronic transmurality of radiofrequency lesions.
Because of the failure of previous technologies on the beating heart, a method to predict transmurality of lesions intraoperatively is needed to evaluate the efficacy of off-pump ablation. Conduction delay or conduction block across a line of ablation has been one of the main tools used to evaluate the success of ablations in the electrophysiology suite and has recently been proposed as a method to assess whether a lesion has resulted in a transmural scar in the operating room7, 8. This study tested the efficacy of this technique to predict chronic conduction block in a porcine model.
Methods
Seven Hanford miniature swine weighing 50kg to 70kg were used in this study. All animals received humane care in compliance with the Guide for the Care and Use of Laboratory Animals, (National Academy Press, Washington, DC). Each animal was premedicated with Telazol, Ketamine, and Xylazine, intubated, anesthetized with isoflourane, and monitored continuously throughout the procedure with ECG and invasive arterial pressure recordings. Perioperative antibiotics were given during the initial surgery and consisted of Ceftiotur (2mg/kg). Antibiotics were continued for three days postoperatively.
Each pig had a set of right and left atrial ablations performed through a median sternotomy. After creation of a pericardial sling, a set of three molded silicone plaques with a total of 252 unipolar electrodes were placed onto the epicardial surface to obtain epicardial activation maps. The electrode templates were constructed from a form-fitting silicon elastomer (Specialty Silicone Fabricators,Paso Robles, Calif) that fit snugly on the entire atrial epicardium and contained 0.5-mm diameter silver electrodes (Pacific Wire & Cable, Inc, Santa Ana, Calif). The interelectrode distance was 5 mm. These plaques were secured with Rommel tourniquets to allow for consistent placement before and after the lesion set creation. After a baseline ACT was measured, a heparin bolus (100 units/kg) was given intravenously and ACT was maintained above 350 seconds until after all ablations were completed.
Pacing thresholds were obtained on either side of the planned lines of ablation as shown in Figure 1. Activation maps during normal sinus rhythm and paced activation (twice threshold, 120 bpm) were recorded from identical sites before and immediately after completion of the lesion set.
The lesion set is shown in Figure 1. The atrial appendages were isolated with a single application of the AtriCure Synergy™ device, a bipolar radiofrequency clamp device with dual electrodes on each jaw of the clamp. The device was used per the manufacturer's instructions. Ablations around the inferior vena cava and the left pulmonary veins were also created with single applications of the Synergy device. On the right atrium, a line of ablation was created with multiple applications of the AtriCure Coolrail™ device to connect the ablation at the base of the right atrial appendage to the ablation around the IVC. This epicardial radiofrequency ablation device has already been described in another publication from our laboratory (Figure 2)9. Briefly, the device is a handheld pen-like device that has two hollow, 3cm long electrodes through which sterile water is circulated during ablation to reduce tissue-electrode temperature. An analogous line on the left atrium was created with multiple applications of the Coolrail™ device between the base of the left atrial appendage and the left pulmonary veins. Initially, application of the Coolrail™ was planned for 50 seconds with an overlap of 0.5 cm per application based on dosimetry data previously published9. After perforation and subsequent death due to intraoperative hemorrhage in the first pig with this ablation time, the protocol was changed to a 40 second ablation with only minimal overlap of prior ablation lines.
After completion of the ablation protocol and acquisition of the acute post-ablation activation maps, hemostasis was secured. The pericardium was closed and the sternum was reapproximated. After the operation, the animal was closely monitored for at least 48 hours and survived for 28 days. Each animal received 325mg of aspirin daily beginning on the first post-operative day. The animals received buprenorphine (0.005-0.01mg/kg IM) for post-operative analgesia.
At 28 days, the animals had a re-operative sternotomy utilizing the same anesthesia protocol used for the initial operation. The pericardium was reopened, and the atria were carefully dissected free of adhesions. The epicardial electrode plaques were repositioned and activation maps were obtained during normal sinus rhythm and during paced rhythm from the same sites as during the initial operation.
Once the chronic post-ablation maps were obtained, the animals were euthanized humanely with a concentrated potassium chloride solution. The aorta was cross clamped and the hearts were perfused with 60ml of 1% 2,3,5-triphenyl-tetrazolium chloride solution through the coronary circulation via infusion into the root of the aorta. Then, the hearts were removed en bloc and examined grossly for any evidence of intra-atrial thrombus formation. They were then placed in 1% 2,3,5-triphenyl-tetrazolium chloride solution and incubated at room temperature for 45 minutes to aid in visualization of the ablations. Each lesion was sectioned at 5mm intervals perpendicular to the long axis of the ablation. The sections were fixed in formalin, molded in paraffin, and stained with Gomori's trichome. The sections were microscopically examined to assess transmurality and measure lesion depth and tissue thickness.
Analysis of mapping data were done offline using custom designed software as described previously10. Unipolar electrogram data were recorded at a gain of 250 and a frequency response of 0.5 to 500 Hz. The data were digitized at 1000 Hz. Activation times were picked automatically by selecting the peak derivative of voltage over time from the electrogram following paced activation. Automatic activation time was manually reviewed and edited for accuracy. Only the activation maps on the right side of the atria were analyzed because the left atrium in the Hanford miniature swine was too narrow in relation to the Coolrail™ ablation line, and the spatial resolution of the mapping system was not fine enough to map over the remaining viable tissue of the left atrium.
Statistical Analysis
All data are expressed as mean ± standard deviation. Comparison of conduction time across the line of ablation was done with a Student's t-test.
Results
Operative Results
The first animal died intra-operatively during the initial procedure from a perforation caused by the Coolrail™ device. In the initial protocol, applications of the device were for 50 seconds, and there was at least 0.5cm of overlap between applications to ensure a continuous linear ablation. Given the perforation, the protocol was modified as noted in the previous section, and subsequently, the remaining six animals survived the operation. There were no further perforations caused by the device after adaptation of the new protocol. There were no postoperative complications related to ablation. None of the animals showed any signs of cerebrovascular accidents or other signs of thromboembolic events.
Histology
Gross visual inspection of ablations revealed no intra-atrial thrombi.
The device performance for the Coolrail™ is summarized in Figure 3. Sectioning of the 12 connecting lesions created by the Coolrail™ yielded 72 total cross-sections. Of these, 76% (55/72) were transmural. Only one of the twelve connecting lesions created in the study had transmural ablations at every cross-section examined. This was one of the lines connecting the base of the right atrial appendage to the inferior vena cava. At least one (range 1-4) cross-section in each of the other eleven lines demonstrated failure to achieve transmural ablation. In two cases, there was an outright gap in the ablation line at the point of overlap between applications of the Coolrail device. These gaps were excluded from analysis as they represented tissue that was not ablated, rather than failed device performance.
Device performance varied with tissue depth (Figure 3). In tissue less than 4 mm thick, 97% (33/34) of all cross sections examined were transmurally ablated. In tissue 4 mm thick or greater, 58% (22/38) of cross sections were transmurally ablated. Average lesion depth for the study was 3.7 ± 1.5 mm, and tissue thickness was 4.4 ± 2.2 mm. Average lesion width was 8.6 ± 2.5 mm.
Electrophysiology
Complete mapping data, including pre-ablation, acute post-ablation, and chronic post-ablation activation maps, were available for analysis in five animals. The pericardium of the sixth animal was too tightly adhered to the atrial epicardium to allow for safe dissection and subsequent mapping.
A combination of conduction delay and change in activation sequence was used to evaluate for conduction block on acute and chronic maps (Figure 4). Conduction block was defined by a change in the activation sequence around the observed line of block, while conduction delay was defined as delay in conduction without a change in the sequence of activation. Four of five activation maps demonstrated acute conduction block after the ablation line was performed. The other showed only conduction delay without block (from 13 ms pre-ablation to 38 ms post-ablation) through the line of ablation. Interestingly, this map correlated with the sole line of ablation with completely transmural cross-sections in histological analysis. Conduction time in the chronic setting for this right atrium decreased to 28ms.
Conduction time across the Coolrail™ line of ablation was significantly increased after ablation in the initial operation. Conduction time before ablation averaged 14.4 ± 5.6 ms. This increased significantly to 36.8 ± 15.1 ms acutely post-ablation (p=0.007). Four weeks after ablation, epicardial mapping demonstrated a return to baseline conduction times. The chronic post-ablation conduction time averaged 19.4 ± 6.8 ms, and was not statistically different from pre-ablation values (p=0.242). The conduction time for all five animals with complete maps is plotted in Figure 5. In each case, conduction delay across the Coolrail™ line was increased in the acute post-ablation map but fell back towards pre-ablation values in the chronic epicardial map at four weeks.
Discussion
In the last decade, the Cox-Maze procedure has been modified to make it more accessible and less invasive. Devices have been used to replace the incisions of the Cox-Maze III with lines of ablation to make it easier and faster to perform. At our institution, the use of bipolar radiofrequency clamps has decreased the operative time by more than half3. Moreover, these bipolar devices have been shown experimentally to create reliable transmural lesions11, 12. The bipolar RF based Cox-Maze IV had achieved over 80% freedom from AF off antiarrhythmic drugs at 6 months13. However, these devices still require cardiopulmonary bypass and small incisions in the atria to perform the entire Cox-Maze lesion set.
Less extensive lesion sets have been introduced to decrease the invasiveness of these procedures so that they can be performed without cardiopulmonary bypass. The most common are limited to pulmonary vein isolation alone and occasionally involve surgical ablation of the ganglionated plexi14, 15. These operations are attractive because they can be done with minimal access techniques and without cardiopulmonary bypass, avoiding potential morbidity. Unfortunately, the results of these new procedures have been less than satisfactory, particularly in patients with long-standing, persistent AF, where the success rates can be less than 50%14. This is not surprising, and similar results have been demonstrated in the electrophysiology laboratory with pulmonary vein isolation alone in patients with long-standing atrial fibrillation16.
In order to improve results, more extensive left atrial lesion sets have been proposed17. Unfortunately, these have required creating linear lesions on the beating heart. To answer this need, devices have been recently introduced to perform epicardial, beating heart ablation on the atria6. However, experience in our laboratory and others has shown that it is difficult to perform consistently transmural lesions on the beating heart9, 18. We have shown that even small gaps (<1mm) in a linear lesion can result in a lack of conduction block19. Therefore, it is vital that any new device be tested experimentally prior to widespread clinical use. The AtriCure Coolrail™ device was introduced specifically to make linear lesions on the beating heart. It was not independently tested experimentally prior to clinical use. The first goal of this study was to examine whether the device was capable of creating consistent transmural ablation in a chronic animal model. This study simulated the proposed clinical use of the device by making connecting lesions to create lines of conduction block.
The Coolrail™ was not capable of creating a long linear lesion in this chronic study. The performance of the Coolrail™ has already been described by this laboratory in an acute animal study9. None of the twelve lines created with the device created a pathologically and/or electrically transmural lesion. The only ablation line that had all histological sections transmural had no evidence of conduction block by epicardial mapping. While every cross section examined in the line of ablation was transmural, the pattern of conduction in the chronic activation map did not demonstrate acute conduction block, indicating that there must have been a gap or gaps along the ablation line.
These chronic results confirm the acute results found after staining with 1% 2,3,5-triphenyl-tetrazolium chloride solution. Although this study was not designed specifically to compare acute histology with chronic histology, comparing the chronic histological data to previously acquired acute data, there was no statistically significant difference in depth of tissue penetration (p=0.147) or likelihood of transmural ablation (p=0.621) as determined by acute staining with 1% 2,3,5-triphenyl-tetrazolium chloride solution in our prior study versus chronic histological evaluation with trichrome staining.
The inability of this device and of other unidirectional devices we have tested to create reliably transmural lesions emphasizes the importance of developing an intraoperative method to test for lesion integrity6. Various groups have proposed using conduction block or delay as a way to verify lesion transmurality7, 8. Unfortunately, this study also demonstrated that acute conduction delay or block did not predict chronic transmurality when used to assess epicardial, radiofrequency ablation in the porcine atria. Histological analysis demonstrated that the majority of lines created in this study with the Coolrail™ contained at least one non-transmural gap. These gaps were not readily identifiable based on activation sequence mapping. In each case where a gap was seen clearly in histology, the activation map demonstrated a pattern consistent with acute conduction block. The wavefront of electrical activation traveled from the right atrial free wall around the anterior surface of the atrium to activate the atrial tissue on the opposite side of the line of ablation (See Figure 4, acute post-ablation). Figure 5 demonstrates that this significantly increased the conduction time between the two selected sites. In the chronic setting, however, conduction through non-transmural gaps was clearly identified, and conduction times returned to baseline (Figures 4 and 5).
A shortcoming of acute intraoperative activation mapping is demonstrated in this study. When gaps were present histologically, no evidence of this was found acutely with activation mapping. The phenomenon of chronic recovery of conduction is well described in the electrophysiology literature. Catheter-based pulmonary vein isolation for atrial fibrillation is based on identifying and ablating conduction gaps until the pulmonary veins demonstrate entrance and exit block16. While every effort is made at the initial ablation to isolate the pulmonary veins, patients can present with recurrent AF, and repeat electrophysiological mapping often reveals recovery of conduction from the majority of pulmonary vines20-22. This recovery seen clinically is similar to the recovery of conduction in this study. There is evidence of acute injury and recovery at the tissue level. One echocardiography-based study found swelling in the ablated crista terminalis that resolved over the course of four weeks23. This observation likely explains the acute conduction block, as localized tissue swelling and inflammation may lead to acute stunning and electrical block. However, as the tissue recovered, electrical activity was restored. Further evidence in the electrophysiology literature suggests that this recovery of activation may not take an extended period of time. Observing the results of pulmonary vein isolation over the course of an hour revealed recovery of conduction in a significant number of patients24.
This study underscores the importance of using a reliable, rigorously investigated energy source when performing surgical ablation until a reliable method of intraoperative lesion evaluation is developed. The practice at our institution is to use devices that have been shown to reliably create transmural ablation, such as bipolar radiofrequency clamps.11 In addition, exit block from the pulmonary veins is confirmed by pacing during surgical ablation. While this cannot reliably predict long-term conduction block, it does address failed ablation acutely and repeat ablations are performed until exit block is achieved.
The failure of conduction delay as a means to predict chronic transmurality has important implications for surgical practice. In catheter-based procedures, failure to achieve permanent conduction block with ablation can be remedied by a relatively low morbidity redo procedure. Thus, an imperfect method to predict lesion transmurality carries with it less risk to the patient. When considering surgical ablation, a redo operation carries a higher morbidity. Lesion formation needs to be as complete as possible to help patients avoid the risk of failure. The Coolrail™ is unreliable in this regard.
Study Limitations
The animal model used in this study may not perfectly simulate conditions encounted clinically. The average thickness of atrial tissue encountered in this study was 4.4 ± 2.2 mm. In some areas of the diseased human atrial, myocardial thickness can exceed 10 mm23. Transmural ablation in the diseased human atria may be harder to accomplish.
During the course of the study, one animal had a perforation that led to intraoperative death. This resulted from a longer ablation in an area of overlap. The study protocol was modified to a shorter ablation, and ablations were not overlapped. It is important to note that the clinical use of this device often involves overlapped ablations of long duration. Thus, caution should be exercised when applying this device in this manner in patients.
Acknowledgements
The research was supported in part by National Institutes of Health grants 5R01HL32257, RO1 HL085113, T32HL07776, and a grant to Dr. Damiano from AtriCure, Inc. (West Chester, Ohio). We also gratefully acknowledge the technical assistance of Diane Toeniskoetter and Naomi Still.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Prasad SM, Maniar HS, Camillo CJ, et al. The Cox maze III procedure for atrial fibrillation: long-term efficacy in patients undergoing lone versus concomitant procedures. J Thorac Cardiovasc Surg. 2003;126:1822–8. doi: 10.1016/s0022-5223(03)01287-x. [DOI] [PubMed] [Google Scholar]
- 2.Banerjee A, Singh S, Tempe DK. Intraoperative endocardial ablation of chronic atrial fibrillation along with mitral valve surgery using high frequency ultrasound with a ball-tipped harmonic scalpel probe. Indian Heart J. 2004;56:178–80. [PubMed] [Google Scholar]
- 3.Gaynor SL, Diodato MD, Prasad SM, et al. A prospective, single-center clinical trial of a modified Cox maze procedure with bipolar radiofrequency ablation. J Thorac Cardiovasc Surg. 2004;128:535–42. doi: 10.1016/j.jtcvs.2004.02.044. [DOI] [PubMed] [Google Scholar]
- 4.Ad N, Cox JL. The Maze procedure for the treatment of atrial fibrillation: a minimally invasive approach. J Card Surg. 2004;19:196–200. doi: 10.1111/j.0886-0440.2004.4036_1.x. [DOI] [PubMed] [Google Scholar]
- 5.Knaut M, Spitzer SG, Karolyi L, et al. Intraoperative microwave ablation for curative treatment of atrial fibrillation in open heart surgery--the MICRO-STAF and MICRO-PASS pilot trial. MICROwave Application in Surgical treatment of Atrial Fibrillation. MICROwave Application for the Treatment of Atrial Fibrillation in Bypass-Surgery. Thorac Cardiovasc Surg. 1999;47(Suppl 3):379–84. doi: 10.1055/s-2007-1013205. [DOI] [PubMed] [Google Scholar]
- 6.Schuessler RB, Lee AM, Melby SJ, et al. Animal studies of epicardial atrial ablation. Heart Rhythm. 2009;6:S41–S45. doi: 10.1016/j.hrthm.2009.07.028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Pak HN, Hwang C, Lim HE, Kim JS, Kim YH. Hybrid epicardial and endocardial ablation of persistent or permanent atrial fibrillation: a new approach for difficult cases. J Cardiovasc Electrophysiol. 2007;18:917–23. doi: 10.1111/j.1540-8167.2007.00882.x. [DOI] [PubMed] [Google Scholar]
- 8.Edgerton JR, Jackman WM, Mack MJ. A new epicardial lesion set for minimal access left atrial maze: the Dallas lesion set. Ann Thorac Surg. 2009;88:1655–7. doi: 10.1016/j.athoracsur.2009.05.046. [DOI] [PubMed] [Google Scholar]
- 9.Lee AM, Aziz A, Sakamoto S, Schuessler RB, Damiano RJ., Jr Epicardial ablation on the beating heart: limited efficacy of a novel, cooled radiofrquency ablation device. Innovations: Technol Techniq Cardiothorac Vascu Surg. 2009;4:86–92. doi: 10.1097/IMI.0b013e3181a348a2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Schuessler RB, Kay MW, Melby SJ, Branham BH, Boineau JP, Damiano RJ., Jr. Spatial and temporal stability of the dominant frequency of activation in human atrial fibrillation. J Electrocardiol. 2006;39:S7–12. doi: 10.1016/j.jelectrocard.2006.04.009. [DOI] [PubMed] [Google Scholar]
- 11.Melby SJ, Gaynor SL, Lubahn JG, et al. Efficacy and safety of right and left atrial ablations on the beating heart with irrigated bipolar radiofrequency energy: a long-term animal study. J Thorac Cardiovasc Surg. 2006;132:853–60. doi: 10.1016/j.jtcvs.2006.05.048. [DOI] [PubMed] [Google Scholar]
- 12.Prasad SM, Maniar HS, Diodato MD, Schuessler RB, Damiano RJ., Jr. Physiological consequences of bipolar radiofrequency energy on the atria and pulmonary veins: a chronic animal study. Ann Thorac Surg. 2003;76:836–41. doi: 10.1016/s0003-4975(03)00716-1. discussion 841-2. [DOI] [PubMed] [Google Scholar]
- 13.Voeller RK, Bailey MS, Zierer A, et al. Isolating the entire posterior left atrium improves surgical outcomes after the Cox maze procedure. J Thorac Cardiovasc Surg. 2008;135:870–7. doi: 10.1016/j.jtcvs.2007.10.063. [DOI] [PubMed] [Google Scholar]
- 14.Edgerton JR, McClelland JH, Duke D, et al. Minimally invasive surgical ablation of atrial fibrillation: six-month results. J Thorac Cardiovasc Surg. 2009;138:109–13. doi: 10.1016/j.jtcvs.2008.09.080. discussion 114. [DOI] [PubMed] [Google Scholar]
- 15.Wolf RK, Schneeberger EW, Osterday R, et al. Video-assisted bilateral pulmonary vein isolation and left atrial appendage exclusion for atrial fibrillation. J Thorac Cardiovasc Surg. 2005;130:797–802. doi: 10.1016/j.jtcvs.2005.03.041. [DOI] [PubMed] [Google Scholar]
- 16.Calkins H, Brugada J, Packer DL, et al. HRS/EHRA/ECAS expert Consensus Statement on catheter and surgical ablation of atrial fibrillation: recommendations for personnel, policy, procedures and follow-up. A report of the Heart Rhythm Society (HRS) Task Force on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2007;4:816–61. doi: 10.1016/j.hrthm.2007.04.005. [DOI] [PubMed] [Google Scholar]
- 17.Edgerton JR, Edgerton ZJ, Weaver T, et al. Minimally invasive pulmonary vein isolation and partial autonomic denervation for surgical treatment of atrial fibrillation. Ann Thorac Surg. 2008;86:35–8. doi: 10.1016/j.athoracsur.2008.03.071. discussion 39. [DOI] [PubMed] [Google Scholar]
- 18.Melby SJ, Zierer A, Kaiser SP, Schuessler RB, Damiano RJ., Jr. Epicardial microwave ablation on the beating heart for atrial fibrillation: the dependency of lesion depth on cardiac output. J Thorac Cardiovasc Surg. 2006;132:355–60. doi: 10.1016/j.jtcvs.2006.02.008. [DOI] [PubMed] [Google Scholar]
- 19.Melby SJ, Lee AM, Zierer A, et al. Atrial fibrillation propagates through gaps in ablation lines: implications for ablative treatment of atrial fibrillation. Heart Rhythm. 2008;5:1296–301. doi: 10.1016/j.hrthm.2008.06.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Nanthakumar K, Plumb VJ, Epstein AE, Veenhuyzen GD, Link D, Kay GN. Resumption of electrical conduction in previously isolated pulmonary veins: rationale for a different strategy? Circulation. 2004;109:1226–9. doi: 10.1161/01.CIR.0000121423.78120.49. [DOI] [PubMed] [Google Scholar]
- 21.Cappato R, Negroni S, Pecora D, et al. Prospective assessment of late conduction recurrence across radiofrequency lesions producing electrical disconnection at the pulmonary vein ostium in patients with atrial fibrillation. Circulation. 2003;108:1599–604. doi: 10.1161/01.CIR.0000091081.19465.F1. [DOI] [PubMed] [Google Scholar]
- 22.Gerstenfeld EP, Callans DJ, Dixit S, Zado E, Marchlinski FE. Incidence and location of focal atrial fibrillation triggers in patients undergoing repeat pulmonary vein isolation: implications for ablation strategies. J Cardiovasc Electrophysiol. 2003;14:685–90. doi: 10.1046/j.1540-8167.2003.03013.x. [DOI] [PubMed] [Google Scholar]
- 23.Ren JF, Marchlinski FE, Callans DJ, Zado ES. Echocardiographic lesion characteristics associated with successful ablation of inappropriate sinus tachycardia. J Cardiovasc Electrophysiol. 2001;12:814–8. doi: 10.1046/j.1540-8167.2001.00814.x. [DOI] [PubMed] [Google Scholar]
- 24.Wang XH, Liu X, Sun YM, et al. Early identification and treatment of PV re-connections: role of observation time and impact on clinical results of atrial fibrillation ablation. Europace. 2007;9:481–6. doi: 10.1093/europace/eum101. [DOI] [PubMed] [Google Scholar]
- 25.Saremi F, Channual S, Krishnan S, Gurudevan SV, Narula J, Abolhoda A. Bachmann Bundle and its arterial supply: imaging with multidetector CT--implications for interatrial conduction abnormalities and arrhythmias. Radiology. 2008;248:447–57. doi: 10.1148/radiol.2482071908. [DOI] [PubMed] [Google Scholar]