Surgical Removal of an Atrial Septal Occluder Device Embolized to the Main Pulmonary Artery

Emre Boysan 1, Omer Faruk Cicek 1,, Mustafa Cuneyt Cicek 1, Ziyaddin Hamurcu 1, Sami Gurkahraman 1
PMCID: PMC3967478  PMID: 24512412

Abstract

Percutaneous closure of atrial septal defects in adults has emerged as an alternative to surgery. We report a sequela of such closure in a 16-year-old boy: embolization of the atrial septal defect occluder into the main pulmonary artery when the patient experienced an episode of intense coughing immediately after device deployment. We removed the device surgically and closed the atrial septal defect in a standard manner, with an autologous pericardial patch.

Key words: Adolescent; device removal; foreign-body migration; heart septal defects, atrial/therapy; male; prosthesis implantation/adverse effects; pulmonary artery/surgery; septal occluder device/adverse effects


Percutaneous occlusion of atrial septal defect (ASD) in adults has emerged as an alternative to surgery.1 Percutaneous closure is associated with less surgical morbidity, avoidance of a scar, and shorter hospital stay.1 However, its increased use has brought recognition of several fairly frequent sequelae. One sequela associated with the procedure is embolization of the device into the pulmonary artery (PA).2 We describe the case of a teenager in whom this occurred.

Case Report

In September 2012, a 16-year-old boy was evaluated by transthoracic echocardiography (TTE) and found to have an ostium secundum ASD. Transesophageal echocardiography (TEE), performed in order to learn whether the defect was suitable for percutaneous closure, revealed that the ostium secundum ASD had a diameter of 20 mm and rim margins of 20 mm (superior), 11 mm (inferior), 14 mm (superior vena cava), 12 mm (inferior vena cava), and 4 mm (aortic). The patient was referred for percutaneous ASD closure with an Occlutech Figulla® Flex occluder device (Occlutech International AB; Helsingborg, Sweden).

Using standard techniques, we deployed a 24-mm device under fluoroscopic and TEE guidance. The postinterventional TEE showed accurate positioning of the device and no relevant shunting. In the recovery room after extubation, the patient began to cough repeatedly. Although he was hemodynamically stable, we performed a TTE after his coughing episode, in order to eliminate any doubts about the location of the device. This TTE showed absence of the device in the atrial septum; it was instead lodged in the main PA. Fluoroscopy confirmed this dislocation (Fig. 1), so emergency surgery was undertaken via a full median sternotomy through a small skin incision. After aortobicaval cannulation and aortic cross-clamping, we arrested the heart by antegrade administration of cold-blood cardioplegic solution. The right atrium was opened with the patient on total cardiopulmonary bypass, and the ostium secundum ASD was indeed seen to have a diameter of approximately 20 mm. The embolized occluder was palpable externally at the pulmonary trunk and was withdrawn from there via a small incision in the main PA (Fig. 2). The ASD was closed in a standard manner, with an autologous pericardial patch. The postoperative period was uneventful, and the follow-up echocardiograms showed no residual shunt or valvular regurgitation.

Fig. 1.

Fig. 1

Fluoroscopy shows the occluder device (arrow) in the main pulmonary artery.

Fig. 2.

Fig. 2

Intraoperative photograph shows the embolized occluder device (white arrow) during its extraction from the main pulmonary artery (black arrow).

Discussion

Both the design of the occluder device and the implantation technique have improved since King and colleagues3 reported the first percutaneous transcatheter closure of ASD in 1976. As the numbers of catheter-based interventions steadily increase, so do the numbers of reported complications. Chessa and colleagues2 reported that embolization or malposition was the most frequent sequela of this procedure. Devices usually embolize into the main PA, and most of these instances necessitate surgical intervention.

Mashman and associates4 reported 2 cases of septal occluder embolization to the PA. Both of those devices needed surgical retrieval. Lysitsas and associates5 reported the embolization of a septal occluder to the main PA after 2 years of implantation. Their patient had been admitted with symptoms of acute right-sided heart failure, and closure of the secundum ASD was performed, together with surgical retrieval of the device. Amanullah and co-authors6 reported that, from October 2002 through December 2010, device embolization occurred in 2 of 284 cases of device closure performed for secundum ASD. One occluder embolized into the right ventricle and the other into the ascending aorta. The devices were retrieved and the defects closed by surgical intervention.

Another way of removing embolized devices is percutaneous retrieval. Balbi and colleagues7 reported that they successfully retrieved a device from a left ventricle by using a modified snare technique. Then too, Chan and coworkers8 managed to retrieve an embolized device percutaneously, from a right ventricle. They suggested attempting percutaneous retrieval first, in an effort to save the patient from an unplanned emergency open-heart operation. On the other hand, we found no report of successful percutaneous retrieval of an embolized device from the PA, and we believe that percutaneous retrieval should be attempted only in rare and evidently suitable cases.

Inadequate rims (especially deficient aortic rims) and the use of a very oversized occluder device are the most common reasons for device embolization after an apparently successful deployment.9 In our patient, post interventional TEE showed perfect device positioning after release, so we did not expect clinical problems. We speculate that the patient's intensive coughing in the recovery room caused intra-abdominal pressure overload and consequent right atrial pressure overload, which dislocated the occluder. Our patient's aortic rim margin was 4 mm, sufficient to meet the criteria for percutaneous closure, but still short enough (in our judgment) to promote device dislocation. In addition, the grossly oversized occluder could have led to the erosion of this rim. Interventionalists tend to grossly oversize these devices to avoid their embolization, but the problem, paradoxically, can be worsened by this practice.10 Both of these factors are important in the management of patients undergoing percutaneous ASD occlusion.

Stricter selection criteria governing both the adequacy of the aortic rim and the size of the occluder might reduce the incidence of sequelae after percutaneous device closure of an ASD. We should never forget that surgical closure also achieves 100% closure with low morbidity and mortality rates,11 and the cosmetic results are excellent if the approach is minimally invasive.

References

  • 1.Losay J, Petit J, Lambert V, Esna G, Berthaux X, Brenot P, Angel C. Percutaneous closure with Amplatzer device is a safe and efficient alternative to surgery in adults with large atrial septal defects. Am Heart J. 2001;142(3):544–8. doi: 10.1067/mhj.2001.116757. [DOI] [PubMed] [Google Scholar]
  • 2.Chessa M, Carminati M, Butera G, Bini RM, Drago M, Rosti L. Early and late complications associated with transcatheter occlusion of secundum atrial septal defect. J Am Coll Cardiol. 2002;39(6):1061–5. doi: 10.1016/S0735-1097(02)01711-4. [DOI] [PubMed] [Google Scholar]
  • 3.King TD, Thompson SL, Steiner C, Mills NL. Secundum atrial septal defect. Nonoperative closure during cardiac catheterization. JAMA. 1976;235(23):2506–9. doi: 10.1001/jama.1976.03260490024013. [DOI] [PubMed] [Google Scholar]
  • 4.Mashman WE, King SB, Jacobs WC, Ballard WL. Two cases of late embolization of Amplatzer septal occluder devices to the pulmonary artery following closure of secundum atrial septal defects. Catheter Cardiovasc Interv. 2005;65(4):588–92. doi: 10.1002/ccd.20429. [DOI] [PubMed] [Google Scholar]
  • 5.Lysitsas DN, Wrigley B, Banerjee P, Glennon PE, Parmar JM, Shiu MF, Been M. Presentation of an embolised Amplatzer septal occluder to the main pulmonary artery 2 years after implantation. Int J Cardiol. 2009;131(3):e106–7. doi: 10.1016/j.ijcard.2007.07.072. [DOI] [PubMed] [Google Scholar]
  • 6.Amanullah MM, Siddiqui MT, Khan MZ, Atiq MA. Surgical rescue of embolized Amplatzer devices. J Card Surg. 2011;26(3):254–8. doi: 10.1111/j.1540-8191.2011.01249.x. [DOI] [PubMed] [Google Scholar]
  • 7.Balbi M, Pongiglione G, Bezante GP. Percutaneous rescue of left ventricular embolized Amplatzer septal occluder device. Catheter Cardiovasc Interv. 2008;72(4):559–62. doi: 10.1002/ccd.21679. [DOI] [PubMed] [Google Scholar]
  • 8.Chan KT, Cheng BC. Retrieval of an embolized Amplatzer septal occluder. Catheter Cardiovasc Interv. 2010;75(3):465–8. doi: 10.1002/ccd.22297. [DOI] [PubMed] [Google Scholar]
  • 9.Misra M, Sadiq A, Namboodiri N, Karunakaran J. The ‘aortic rim’ recount: embolization of interatrial septal occluder into the main pulmonary artery bifurcation after atrial septal defect closure. Interact Cardiovasc Thorac Surg. 2007;6(3):384–6. doi: 10.1510/icvts.2006.141044. [DOI] [PubMed] [Google Scholar]
  • 10.Amin Z, Hijazi ZM, Bass JL, Cheatham JP, Hellenbrand WE, Kleinman CS. Erosion of Amplatzer septal occluder device after closure of secundum atrial septal defects: review of registry of complications and recommendations to minimize future risk. Catheter Cardiovasc Interv. 2004;63(4):496–502. doi: 10.1002/ccd.20211. [DOI] [PubMed] [Google Scholar]
  • 11.Hopkins RA, Bert AA, Buchholz B, Guarino K, Meyers M. Surgical patch closure of atrial septal defects. Ann Thorac Surg. 2004;77(6):2144–50. doi: 10.1016/j.athoracsur.2003.10.105. [DOI] [PubMed] [Google Scholar]