Abstract
A gradient reversed-phase high performance liquid chromatography (HPLC) method was developed for the assay of cetrorelix acetate, a synthetic decapeptide with gonadotropin-releasing hormone (GnRH) antagonistic activity used in infertility treatment. The HPLC method, which is used to determine cetrorelix in bulk and pharmaceutical dosage forms, was validated per ICH guidelines. The chromatographic separation was achieved on a C18 reversed-phase column using acetonitrile, water and trifluoroacetic acid (TFA) as mobile phase and wavelength was set at 275 nm. The calibration curve was linear (r2 = 0.999) over cetrorelix concentrations ranging from 62.50 to 12.50 μg/mL (n = 6). The limits of detection (LOD) and quantification (LOQ) were calculated from the peak-to-noise ratio as 15.6 and 62.5 μg/mL, respectively. The method had an accuracy of > 97% and intra- and inter-day RSD of < 0.3% and < 1.6%, respectively and was validated with excellent specificity, sensitivity, and stability. The validated method was successfully applied for determination of cetrorelix in bulk and pharmaceutical dosage forms.
Key Words: Cetrorelix acetate, HPLC, Assay, UV detection, Formulation
Introduction
The decapeptide Cetrorelix ([Ac-D-Nal -(p-Cl)-D-Phe -D-Pal -Ser - Thy -D-Cit -Leu -Arg -Pro -D-Ala –NH2 ], Figure 1), is a third-generation Gn-RH antagonist that is currently mainly used in infertility treatment (for the prevention of premature ovulation in patients undergoing controlled ovarian stimulation) and is suitable for the treatment of benign prostatic hypertrophy and sexual hormone-dependent tumors (1, 2, 3, 4 and 5).
Cetrorelix has been analyzed by HPLC using fluorescence, mass spectrometry, UV-detectoion and combinations of UV-detection and detection of radioactivity, for quantification in biological fluids, liposome and dry powder inhalation (1, 6, 7, 8, 9, 10 and 11). Fluorescence detection and mass spectrometry thechniquse suffer from not being commonly available in all analysis laboratories (12-13). Furthermore mass spectrometry method may have the highest sensitivity, but the determination process is complex (14-15). A HPLC method has been reported by Irngartinger et al. for determination of Cetrorelix in dry powder inhalation whit UV-detection but validation procedure has not been described (10).
Unlike small organic molecules whose chromatographic behavior is described by a finite partitioning equilibrium between the stationary phase and the mobile phase, peptides typically do not exhibit such an effect. Instead, they exhibit an adsorption phenomenon in which the polypeptide is adsorbed onto the stationary phase and elutes only when the solvent strength of the mobile phase is sufficient to compete with the hydrophobic forces keeping it there (16-17). The sensitivity of peptide retention to subtle changes in the organic phase concentration makes isocratic elution difficult because the organic phase concentration must be maintained very precisely (18-19). Furthermore, Small peptides like cetrorelix (MW: 1431) appear to chromatograph by a hybrid of partitioning and adsorption mechanisms and it make the chromatographic process more complicated (16). For this reasons, gradient elution of small peptides like cetrorelix, even with shallow gradients, is preferred, since it results in much sharper peaks than isocratic elution and isocratic elution is rarely used for peptide separations (16-20). Grohganz et al. reported an isocratic HPLC method for determination of cetrorelix in liposome formulation whit UV detection (9, 1 and 11). In this method cetrorelix was separated from excipients during sample preparation and due to the use of isocratic system cetrorelix was detected in short retention time (RT: 3.5 min) (9, 1 and 11). Because of the risk of interference between cetrorelix and excipients in very short run time, this method can not be used for assay of cetrorelix in its injection formulation.
In addition, to the best of our knowledge no official report for the determination of cetrorelix in bulk material and parenteral formulation has ever been published in any pharmacopoeia. Thus, it would be of particular interest to develop and validate a simple, precise, specific, accurate, and stable gradient RP-HPLC method with UV-detector according to ICH guidelines for the determination and routine analysis of cetrorelix in pure and pharmaceutical formulations (21).
Experimental
Chemicals
Cetrorelix for injection (Cetrotide®) were prepared from Merck serono, Canada. Each vial of Cetrotide® contains 0.25 mg of cetrorelix as cetrorelix acetate and 54.80 mg of mannitol. The cetrorelix acetate was purchased from Henan New-Sensation Chemical Co., Ltd. ((Mainland, China). HPLC grade acetonitrile and TFA were from Merck (Hohenbrunn, Germany). The other reagents were of analytical grade.
Standard preparation
Standard stock solutions of cetrorelix acetate (1.0 mg/mL) was prepared by direct weighing of standard substance with subsequent dissolution in deionize water. The standards for the calibration curve were prepared in volumetric flasks (5mL) using standard stock solutions by serial dilution to yield concentrations of 1250, 500, 250, 125 and 62.5 μg/mL (cetrorelix as acetate).
Quality control preparation
A second weighing of cetrorelix acetate independent of the stock was used for the preparation of the quality control samples (QCs). This QC stock solution (1 mg/mL) was prepared in deionized water. QC samples were prepared in volumetric flasks (5ml) using QC stock solutions by serial dilution to yield concentrations of 500, 250 and 125 μg/ml (cetrorelix as acetate).
Assay preparation
Vials of cetrorelix for Injection (Cetrotide® 0.25 mg) were separately dissolved in 1 mL water for injection to obtain a concentration of 250 μg/mL. The drug concentration of the resulting sample solution was determined by HPLC using the calibration curve of standard solution. All determinations were conducted in triplicate.
Chromatographic conditions
The liquid chromatographic system consisted of an Agilent 1200 Series HPLC System equipped with autosampler (G1329A), UV detector (G1314B), degasser (G1379B), and binary pump (G1312A) (GenTech Scientific, NY,USA). The HPLC parameters were controlled by ChemStation Software Rev.B.03.01. Analysis was carried out at 275 nm with a Lichrospher® C18, 250 × 4.60 mm, 5μm column using gradient elution of two solutions, A (0.1% (v/v) TFA in water) and B (0.1% (v/v TFA in acetonitrile). Both mobile phases were degassed by vacuum for 15 min. The following gradient program was applied: 90% A for 5min, from 90% A to 70% B in 15 min, 70% B for 10 min, from 70% B to 90% A in 5 min and 90 % A for 30 min. The flow rate was 1 mL/min, and total run time was 65 min. The injection volume was 20 μL.
Method validation
Specificity
The selectivity of the methods was evaluated by comparing the chromatograms of blank samples (deionized water and aqueous solution of mannitol, 54.8 mg/mL) versus standard samples.
Linearity
Six calibration standards (1250–62.5 μg/mL) were prepared covering the expected range. The standards were prepared in deionize water. The data for peak area versus drug concentration were treated by linear regression analysis.
Sensitivity
The sensitivity of the analytical technique was expressed as the LOQ, which is the minimum concentration of cetrorelix that can be quantitatively determined with the peak-to-noise ratio at least 10:1, and the LOD as a peak-to-noise ratio at 3:1. The LOQ is accepted if the analyte peak response is identifiable, discrete and reproducible with a precision of 2% and accuracy of 95–115% (21, 22).
Precision and accuracy
The intra-day and inter-day precision were determined by analyzing three replicates of quality control samples at concentrations of 125, 250 and 500 μg/mL (cetrorelix as acetate) on the same day and three times on three days, respectively. The precision was evaluated by the relative standard deviation (RSD) and the acceptable range of RSD was no more than 2% (21, 22). The accuracy was assessed by the methodological recovery. The recovery of the method was calculated by comparing the determined concentration of QC samples to the theoretical concentrations.
Stability
Stability of the QC samples were evaluated after short-term storage (at 25 °C for 24 h), after long-term storage (at – 20 °C for 1 week) and after going through freeze-and-thaw cycles (from −20 oC to room temperature for every 24 h). The stability of standard stock solution was also evaluated.
Statistical analysis
The data were submitted to statistical analysis using Excel® software.
Results and Discussion
HPLC method development
Choice of detection wavelength
Cetrorelix peak was monitored at different wavelengths (230, 265 and 275 nm) and a compromise wavelength of 275 nm was selected as the optimum wavelength for HPLC analysis because it maximizes the peak symmetry of cetrorelix, while giving a flat baseline and minimum signal of TFA (Figure 2). Furthermore the ultraviolet spectra of aqueous solution of cetrorelix showed the maximum absorption wavelength at 275 nm (Figure 3).
Choice of gradient program for mobile phase
Separation of small molecules involves continuous partitioning of the molecules between the mobile phase and the hydrophobic stationary phase. Peptides, however, are too large to partition into the hydrophobic phase; they adsorb to the hydrophobic surface after entering the column and remain adsorbed until the concentration of organic phase reaches the critical concentration necessary to cause desorption (18-19). Because peptides diffuse slowly, RP-HPLC results in broader peaks than obtained with small molecules (16). In addition, Small peptides have a hybrid separation mechanism. They desorb more quickly with changes in organic modifier concentration than small molecules which partition, however they desorb more gradually than proteins. For this reasons, gradient elution is generally preferred for small peptide separations (16, 19, 20 and 23).
Cetrorelix is a small peptide (MW: 1431) with hybrid separation mechanism. The sensitivity of cetrorelix elution to the change in acetonitrile concentration per unit time is illustrated in Figure 4. Large changes occur in the retention time of cetrorelix and base line flatness with relatively small changes in the acetonitrile concentration per unit time. The sensitivity of cetrorelix retention to subtle changes in the acetonitrile concentration makes isocratic elution difficult because the organic modifier concentration must be maintained very precisely.
TFA sets the eluent pH and interacts with the peptide to enhance the separation. It is normally used at concentrations of about 0.1% (v/v). TFA concentrations up to 0.5% have been useful in solubilizing larger or more hydrophobic proteins and lower concentrations are occasionally used for tryptic digest separations (24). Using TFA concentration lower than 0.1 % in mobile phase will disturb the peak shape for cetrorelix (Figure 5).
In the current study in addition to the percentages of acetonitrile and concentration of the TFA, effects of re-equilibration time, gradient delay time and ramp time on the retention time and symmetry of peak were investigated. One factor at the time was varied while the others retained at a constant value.
Validation of the method
Specificity
The peak purity was ensured by comparing the chromatogram of the cetrorelix standard samples with that of blank samples. The HPLC chromatograms were recorded for standard samples (aqueous cetrorelix solution) and assay samples (aqueous cetrorelix solution with mannitol) and their comparison revealed no peaks in the vicinity of cetrorelix retention time (around 17.7 minutes). No signs of interference with mannitol were detected in cetrorelix for injection formulation (Figure 6).
Linearity
The linearity of the HPLC method used for cetrorelix assay was evaluated by injecting standard concentrations of cetrorelix drug substance. The plot was linear over the concentration range of 62.5 -1250 μg/mL yielding a regression equation Y= 5.2 X – 125.4 with a coefficient of correlation of 0.999 and with confidence intervals at p = 0.05 (Figure 7).
Limit of quantitation and detection
The LOQ was found to be 62.5 μg/mL with signal to noise ratio of about 10. The recovery of quantitation level was 110% with an RSD less than 1.4 %, which met the validation criteria for recovery and precision.
Precision
The results obtained for inter- and intraday precisions are presented in Table 1. Method precision has a RSD below 0.27% for Intraday and 1.60 % for interday precision, which comply with the acceptance criteria proposed (RSD < 2%) (21 and 22).
Table 1.
Theoretical concentration |
Calculated concentration
(mean±S.D., n=3) |
Precision
(R.S.D.) (%) |
Accuracy
Recovery % |
---|---|---|---|
Intraday | |||
125 | 121.6 ± 0.11 | 0.11 | 96.85 |
250 | 242.56 ± 0.40 | 0.17 | 97.03 |
500 | 485.13 ± 1.28 | 0.26 | 97.03 |
Interday | |||
125 | 121.28 ± 1.90 | 1.56 | 97.02 |
250 | 243.46 ± 2.97 | 1.21 | 97.38 |
500 | 488.56 ± 5.57 | 1.14 | 98.69 |
Accuracy
Accuracy was determined by evaluating the recovery of analyte. The percent recovery between theoretical (Ctheo) and calculated (Ccalc) concentration was derived by the following equation:
Recovery (%) = Ccalc/Ctheo× 100
The accuracy of the QC samples ranged from 97 to 99%, indicating excellent accuracy of the proposed HPLC method (Table 1).
Stability
Short-term stability: QC samples were kept at room temperature for 24 h and analyzed. The accuracy for samples ranged from 95 to 98 % after short term stability testing.
Long-term stability: QC samples were kept at – 20 oC for one week. These samples were thawed at room temperature and analyzed. The accuracy for samples ranged from 95 to 99 % after long term stability testing.
Freeze and thaw cycles: QC samples were prepared and frozen at −20 ◦C for 24 h. The samples were thawed at room temperature and analyzed. The samples were refrozen for 24 h under the same conditions. The freeze–thaw cycle was repeated two more times, and then analyzed on the third cycle. The accuracy for samples ranged from 94 to 99% after freeze thaw stability testing.
Standard stock solution stability: The standard stock solutions were found to be stable for one week when refrigerated at +4 oC. The concentration on comparison with freshly prepared standard after the storage was 98.70%.
Application of the method for the analysis of cetrorelix for injection formulation (Cetrotide ® 0.25 mg)
The method was applied to assay cetrorelix in 10 samples of cetrorelix for injection formulation (Cetrotide® 0.25 mg). Peak areas of cetrorelix were measured and amount of drug was calculated from the respective calibration plots. The average assays (n = 10) was 101.52 ± 2.62.
Conclusion
The developed RP-HPLC method for determination of cetrorelix is simple, precise, accurate, linear, and sensitive. This method was validated based on ICH guideline (21).The measured signal was shown to be precise, accurate and linear over the concentration range tested (62.5–1250 μg/mL) with a correlation coefficient better than 0.999. The developed method showed no interference with the formulation excipient. Therefore, this method can be used for the routine determination of cetrorelix in pure and pharmaceutical formulations.
Acknowledgment
The authors would like to thank the personnel of medicinal chemistry research lab, Shahid Beheshti School of Pharmacy for their sincere assistance to conduct this research.
References
- 1.Grohganz H, Tho I, Brandl M. Development and in-vitro evaluation of a liposome based implant formulation for the decapeptide cetrorelix. Eur. J. Pharm. Biopharm. 2005;59:439–448. doi: 10.1016/j.ejpb.2004.10.005. [DOI] [PubMed] [Google Scholar]
- 2.Engel J, Reissmann Th, Cetrorelix . a potent LHRH-antagonist: Chemistry pharmacology and clinical data. In: In: Xu XJ, Ye YH, Tam JP., editors. Peptides Biology and Chemistry. New York: Kluwer Academic Press; 1996. pp. 142–146. [Google Scholar]
- 3.Reissmann T, Schally AV, Bouchard P, Riethmuller H, Engel J. The LHRH antagonist Cetrorelix. Hum. Reprod. Update. 2000;6:322–331. doi: 10.1093/humupd/6.4.322. [DOI] [PubMed] [Google Scholar]
- 4.Schneider A, Lang A, Naumann W. Fluorescence Spectroscopic Determination of the Critical Aggregation Concentration of the GnRH Antagonists Cetrorelix, Teverelix and Ozarelix. J. Fluoresc. 2010;20:1233–1240. doi: 10.1007/s10895-010-0674-5. [DOI] [PubMed] [Google Scholar]
- 5.Reissmann TH, Engel J, Kutscher B, Bernd M, Hilgard P, Peukert M, Szelenyi I, Reichert S, Gonzales-Barcena D, Nieschiag E, Comaro-Schally AM, Schally AV. Cetrorelix. Drug of the Future. 1994;19:228–237. [Google Scholar]
- 6.Raffel HH, Locher M, Borbe HO. High-performance liquid chromatographic assay for the determination of the decapeptide cetrorelix, a novel luteinizing hormone-releasing hormone antagonist, in human plasma. J. Chromatogr. B. 1994;653:102–105. doi: 10.1016/0378-4347(93)e0145-g. [DOI] [PubMed] [Google Scholar]
- 7.Niwa M, Enomoto K, Yamashita K. Measurement of the novel decapeptide cetrorelix in human plasma and urine by liquid chromatography–electrospray ionization mass spectrometry. J. Chromatogr. B. 1999;729:245–253. doi: 10.1016/s0378-4347(99)00168-1. [DOI] [PubMed] [Google Scholar]
- 8.Schwahn M, Schupke H, Gasparic A, Krone D, Peter G, Hempel R, Kronbach T, Locher M, Jahn W, Engel J. Disposition and metabolism of cetrorelix, a potent luteinizing hormone-releasing hormone antagonist, in rats and dogs. Drug. Meta. Dipos. 1999;28:10–20. [PubMed] [Google Scholar]
- 9.Grohganz H, Schläfli O, Rischer M, Brandl M. Development and validation of a HPLC method for routine quantification of the decapeptide Cetrorelix in lip. osome dispersions. J. Pharma. Ceut. Biomed. 2004;34:963–969. doi: 10.1016/j.jpba.2003.11.010. [DOI] [PubMed] [Google Scholar]
- 10.Irngartinger M, Camuglia V, Damm M, Goede J, Frijlink HW. Pulmonary delivery of therapeutic peptides via dry powder inhalation: effects of micronisation and manufacturing. J. Pharma. Ceut. Biomed. 2004;58:7–14. doi: 10.1016/j.ejpb.2004.03.016. [DOI] [PubMed] [Google Scholar]
- 11.Grohganz H, Rischer M, Brandl M. Adsorption of the decapeptide Cetrorelix depends both on the composition of dissolution medium and the type of solid surface. Europ. J. Pharm. Sci. 2004;21:191–196. doi: 10.1016/j.ejps.2003.10.008. [DOI] [PubMed] [Google Scholar]
- 12.Sobhani H, Shafaati A, Nafissi-Varcheh N, Aboofazeli R. A Reversed Phase High Performance Liquid Chromatographic Method for Determination of Rapamycin. Iran. J. Pharm. Res. 2013;12:77–81. [PMC free article] [PubMed] [Google Scholar]
- 13.Gharia T, Kobarfard F, Mortazavi SA. Development of a Simple RP-HPLC-UV Method for Determination of Azithromycin in Bulk and Pharmaceutical Dosage forms as an Alternative to the USP Method. Iran. J. Pharm. Res. 2013;12:57–63. [PMC free article] [PubMed] [Google Scholar]
- 14.Larry Campbell J, Yves Le, Blanc JC. Peptide and protein drug analysis by MS: challenges and opportunities for the discovery environment. Bioanalysis. 2011;3:645–657. doi: 10.4155/bio.11.31. [DOI] [PubMed] [Google Scholar]
- 15.Tamvakopoulos C. Mass spectrometry for the quantification of bioactive peptides in biological fluids. Mass. Spectrom. Rev. 2007;26:389–402. doi: 10.1002/mas.20120. [DOI] [PubMed] [Google Scholar]
- 16.Carr D. The Handbook of Analysis and Purification of Peptides and Proteins by Reversed Phase HPLC. 3nd ed. California : Grace Vydac; pp. 4–26. [Google Scholar]
- 17.Colin TMant, Yuxin Chen, Zhe Yan, Traian V Popa, James M Kovacs, Janine B Mills, Brian P Tripet, Robert S. Hodges. In: GB Fildes., editor. Vol. 386. HPLC Analysis and Purification of Peptides Peptide Characterization and Application Protocols. Humana Press Inc: New Jersey; 2007. pp. 3–57. [Google Scholar]
- 18.Aguilar MI, editor. HPLC of Peptides and Proteins: Methods and Protocols (Methods in Molecular Biology) Vol. 3. Humana Press ; 2004. pp. 103–256. [Google Scholar]
- 19.Shyamala , Vishnu Priya P, Anjali Devi N. HPLC Method Development for Proteins and Polypeptides. IJSIT. 2013;2:266–276. [Google Scholar]
- 20.Vergote V, Burvenich C, Van de Wiele C, De Spiegeleer B. Quality specifications for peptide drugs: a regulatory-pharmaceutical approach. J. Pept. Sci. 2009;15:697–710. doi: 10.1002/psc.1167. [DOI] [PubMed] [Google Scholar]
- 21. Validation of analytical procedures: text andmethodology. Proceedings of the International Conference on Harmonization (ICH ’96); Geneva, Switzerland: International Federation of Pharmaceutical Manufactures and Associations (IFPMA); 1996. Methodology Q2 (R1) [Google Scholar]
- 22.Chan CC, Lam H, Lee YC, Zhang ZM, editors. Analytical Method Validation and Instrument Performance Verification. New Jersey: John Wiley and Sons; (2004. pp. 11–85. [Google Scholar]
- 23.Ford JC, Smith JA. Synthetic Peptide Purification by Application of Linear Solvent Strength Gradient Theory. J. Chrom. 1989;483:131–143. [Google Scholar]
- 24.Guo D, Mant CT, Hodges RS. Effects of Ion-Pairing Reagents on the Prediction of Peptide Retention in Reversed-Phase High Performance Liquid Chromatography. J. Chrom. 1987;386:205–222. doi: 10.1016/s0021-9673(01)94598-4. [DOI] [PubMed] [Google Scholar]