Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2015 Jul 1.
Published in final edited form as: J Magn Reson Imaging. 2013 Oct 11;40(1):13–25. doi: 10.1002/jmri.24333

Clinical Performance of Contrast Enhanced Abdominal Pediatric MRI with Fast Combined Parallel Imaging Compressed Sensing Reconstruction

Tao Zhang 1,*, Shilpy Chowdhury 2,*, Michael Lustig 3, Richard A Barth 2, Marcus T Alley 2, Thomas Grafendorfer 4, Paul D Calderon 5, Fraser JL Robb 6, John M Pauly 1, Shreyas S Vasanawala 2
PMCID: PMC3984374  NIHMSID: NIHMS510828  PMID: 24127123

Abstract

Purpose

To deploy clinically, a combined parallel imaging compressed sensing method with coil compression that achieves a rapid image reconstruction, and assess its clinical performance in contrast-enhanced abdominal pediatric MRI.

Materials and Methods

With IRB approval and informed patient consent/assent, 29 consecutive pediatric patients were recruited. Dynamic contrast-enhanced MRI was acquired on a 3T scanner using a dedicated 32-channel pediatric coil and a 3D SPGR sequence, with pseudo-random undersampling at a high acceleration (R=7.2). Undersampled data were reconstructed with three methods: a traditional parallel imaging method and a combined parallel imaging compressed sensing method with and without coil compression. The three sets of images were evaluated independently and blindly by two radiologists at one siting, for overall image quality and delineation of anatomical structures. Wilcoxon tests were performed to test the hypothesis that there was no significant difference in the evaluations, and inter-observer agreement was analyzed.

Results

Fast reconstruction with coil compression did not deteriorate image quality. The mean score of structural delineation of the fast reconstruction was 4.1 on a 5-point scale, significantly better (P<0.05) than traditional parallel imaging (mean score 3.1). Fair to substantial inter-observer agreement was reached in structural delineation assessment.

Conclusion

A fast combined parallel imaging compressed sensing method is feasible in a pediatric clinical setting. Preliminary results suggest it may improve structural delineation over parallel imaging.

Keywords: Contrast-enhanced abdominal pediatric MRI, Coil compression, Parallel imaging, Compressed sensing

INTRODUCTION

There are several challenges in pediatric body MRI: the inherently small anatomy requires an MR acquisition with a very high spatial resolution; the nature of rapid hemodynamics in pediatric patients requires imaging quickly for contrast-enhanced MRI (1-3); and fast breathing and voluntary motion cause several artifacts, so anesthesia is usually needed (4). Thus, imaging speed is critical to more widespread use of pediatric abdominal MRI.

To significantly accelerate MR acquisition, two types of approaches are usually considered. The first type is known as parallel imaging (PI) (5,6). PI uses a set of phased array coils (7) with different coil sensitivities for collecting data. The coil sensitivities are used to accelerate MR data acquisition by undersampling k-space and to reconstruct the missing data. The second type is compressed sensing (CS), which has been developed more recently (8). CS exploits the data redundancy (also known as sparsity) in MR images (9), which essentially requires less data to be acquired. A nonlinear optimization method is often used to reconstruct the undersampled data. Furthermore, the combination of PI and CS can achieve even faster imaging (10-12). Compared to the traditional Fourier transform used for Cartesian imaging at the Nyquist sampling rate, the complexity and computation time of a PI and CS reconstruction is dramatically increased (13,14). Most PI and combined parallel imaging compressed sensing methods have reconstruction time proportional to as much as the third order of the number of coils. Dedicated high-density coils can push the limits of acceleration factors (15-17). However, the challenge of the long image reconstruction time is even more significant. The lengthy reconstruction (usually varying from several minutes up to a few hours) is not practical in a clinical setting, where image quality must be confirmed prior to either awakening a child from anesthesia or releasing a patient from the imaging suite. This is particularly true for contrast-enhanced imaging.

Coil compression (CC) has been proposed to reduce the computation for large coil arrays (18-20). The original data from many coils can be combined into few virtual coils, on which the reconstruction is performed. As a result, the reconstruction time can be significantly shortened because of the reduced number of virtual coils (20). While this may enable the application of PI and CS with high acceleration in a clinical setting, it may also alter the reconstructed images, effecting clinical performance.

In this work, we focus on the clinical deployment of a combined parallel imaging compressed sensing technique with rapid image reconstruction and the assessment of the resulting image quality. We implement a combined parallel imaging compressed sensing method, namely L1-SPIRiT (21), with a CC (20) algorithm to enable fast image reconstruction. We investigate the clinical performance of the combined parallel imaging compressed sensing technique with coil compression in contrast-enhanced abdominal pediatric MRI.

MATERIALS AND METHODS

Patient Recruitment

With institutional review board approval and informed patient consent and/or assent, 29 consecutive patients (14 males and 15 females) referred for contrast-enhanced abdominal MRI under anesthesia at our institution with concern of inflammation (3 cases) or for tumor surveillance (26 cases) were recruited from September 2010 to May 2011. Patient demographics are summarized in Table 1, with ages ranging from 3 months to 9 years (mean, 3.79 years).

Table 1.

Demographics of patients, acquisition parameters, reconstruction times and clinical indications.

Subject number Age (years) Gender Slice Thickness (mm) S/I FOV (cm) In-plane Matrix L1-SPIRiT Reconstruction Time (s) CC-L1-SPIRiT Reconstruction Time (s)
1 2.75 M 1 26 320×320 267 34
2 3.92 M 1 30 320×320 242 33
3 0.67 F 1 24 320×320 276 37
4 5.5 F 1 30 320×320 263 33
5 5.75 F 1 32 320×320 285 36
6 6.42 F 1 28 320×320 364 48
7 4.5 F 1 28 320×320 263 37
8 3.33 F 1 28 320×320 289 40
9 2.42 M 0.8 28 288×288 329 45
10 5.33 M 0.8 30 288×288 307 40
11 4.25 F 1 26 288×288 330 44
12 3.67 M 0.8 24 288×288 261 34
13 3.75 M 0.8 28 288×288 356 41
14 3.25 M 1 24 288×288 254 39
15 0.25 F 1 20 320×320 359 41
16 9 M 1 32 320×320 336 39
17 2.25 M 1 28 320×320 321 43
18 1.08 F 1 26 320×320 282 45
19 3.75 F 1 28 320×320 209 31
20 3.08 M 1 34 320×320 203 31
21 0.92 F 1 26 320×320 202 28
22 3.92 M 1 28 320×320 258 35
23 5 M 0.8 28 320×320 342 38
24 1.25 F 1 28 320×320 213 30
25 6 M 1 30 320×320 214 30
26 2.5 F 1 28 320×320 214 31
27 1.92 F 1 28 320×320 302 34
28 6.92 M 1 32 320×320 228 33
29 6.58 F 1 30 320×320 211 31

Image Acquisition

All imaging was performed on a 3T MR750 MRI scanner (GE Healthcare, Waukesha, WI, USA) with a dedicated 32-channel pediatric custom phased array coil that consisted of a pair of 4-by-4 array coils, one positioned posterior and the other anterior to the abdomen (15). A 3D spoiled gradient echo sequence with intermittent fat suppression was modified to incorporate variable density Poisson-disc k-space sampling (22), which facilitates compressed sensing by ensuring relatively incoherent aliasing artifacts from k-space undersampling. Prescribed acquisition parameters were echo time (TE) 0.7-1.09 ms, repetition time (TR) 3.6-4.4 ms, bandwidth (BW) 128 kHz, slice thickness 0.8-1mm, FOV 24-32 cm, in-plane matrix 288×288 or 320×320 (shown in Table 1), and a total acceleration factor of 7.2. The average acquisition time for each temporal phase was 22.9 seconds (range, 19.5-27.6 seconds). One pre-contrast and up to four post-contrast acquisitions were performed. Single dose gadobenate dimeglumine was diluted as necessary in saline to ensure a volume of at least 10 mL and power injected intravenously at 1mL/sec rate for all but one patient. The remaining one patient was injected with a single dose gadofosveset trisodium, again diluted in saline to a volume of 10 mL and power injected at 1mL/sec rate.

Image Reconstruction

For each subject, three image reconstructions of the arterial phase data were performed. The first reconstruction was a traditional parallel imaging reconstruction with the commercially available Autocalibrating Reconstruction for Cartesian imaging (ARC) algorithm (23). The second reconstruction was L1-SPIRiT reconstruction, which is a combination of autocalibrating parallel imaging and compressed sensing. The third reconstruction was the coil compressed L1-SPIRiT (CC-L1-SPIRiT), which is the combination of L1-SPIRiT with coil compression (CC) (19).

For the third reconstruction, the original 32 coils were combined into 6 virtual coils before L1-SPIRiT reconstruction was performed. In the second and third reconstructions, L1-SPIRiT was performed using a projection onto convex set (POCS) algorithm (21) and a fixed 50 iterations were set as the point of convergence. In this iterative reconstruction, a soft-thresholding operation in the Wavelet domain was carried out. The Wavelet threshold was chosen empirically: for L1-SPIRiT, the threshold was initialized with 0.05 and gradually decreased to 0.0002 as the number of iterations increased; for CC-L1-SPIRiT, the threshold was fixed at 0.0025 for all iterations. Note that the image quality of L1-SPIRiT heavily depends on the chosen Wavelet threshold (14). An inaccurate threshold would introduce artifacts in the reconstructed images, such as image blurring or synthetic appearance. When these image artifacts appeared in the reconstruction with the default parameter, a case-specific parameter tuning was performed. Prior to the L1-SPIRiT reconstruction, the data were normalized such that the maximum value in a zero-filled reconstruction was one in order to guarantee that the wavelet thresholding parameter was independent of the scaling of the data.

Image Evaluation

Two board-certified pediatric radiologists (reader 1 and 2 with 8 and 23 years of clinical experience with pediatric MR imaging respectively) who were blinded to the types of reconstructions and patient history/diagnoses independently assessed image qualitatively. First, images were assessed for their overall image quality including qualitative signal-to-noise ratio (SNR), image contrast, synthetic appearance, coherent structural artifacts and image blurring. The scoring criteria are listed in Table 2. Here, synthetic appearance refers to the subjective somewhat “cartoon-like” artificial appearance, sometimes seen in compressed sensing reconstructions. Based on the overall image quality scores, the reconstructed images were considered to be diagnostically acceptable or non-acceptable (shown in Table 2). Then the quality of delineation of several anatomical structures (renal arteries, liver, portal veins, hepatic veins, pancreas, adrenals and spine) that are routinely assessed on these abdominal MRI exams was graded on the pre-defined criteria listed in Table 3. The reconstructions were first presented to the radiologists individually in a blinded fashion and randomized order, and then presented in pairs for direct side-by-side comparison, again in blinded randomized order. For side-by-side paired comparison, the radiologists were also asked for preference of the delineation of anatomical structures based on the criteria shown in Table 4. When present, the quality of delineation of pathologic lesions was also evaluated for paired comparison. The delay time between the reading sessions was more than one month.

Table 2.

Scoring Criteria for Individual Assessments of Overall Image Quality*

Score Qualitative SNR Image Contrast Synthetic Appearance Coherent Structural Artifact Image Blurring
1 Non-diagnostic Flat (Very little contrast between arterial structures and background) Severe synthetic appearance Non-diagnostic Blurring of all structures
2 Somewhat limiting Discrete (No gradation) Moderate synthetic appearance Significantly limiting artifact Blurring of most structures
3 Adequate for most structures Adequate (Gradation can be seen just in low intensity or high intensity objects) Mild synthetic appearance Mildly limiting artifact Blurring of few structures
4 More than adequate for most structures Outstanding (Gradation can be seen in both low intensity and high intensity objects) No synthetic appearance Minimal limiting artifact No blurring of any structures
5 More than adequate for all structures - - No Artifact -
*

Images with individual assessment score equal or greater than 3 are considered to be diagnostically acceptable, and less than 3 are considered to be diagnostically non-acceptable.

Table 3.

Scoring Criteria for Delineation of Anatomical Structures

Score Renal Arteries Liver Portal Veins Hepatic Veins Pancreas Adrenals Spine
1 (Non-diagnostic) Poorly seen Lesions larger than 2 cm can be assessed Central portal vein can not be assessed Central hepatic vein can not be assessed Heterogeneous signals and contours not defined Adrenals can not be assessed Vertebrae from disc can not be separated
2 (Limited) Seen but not confident for stenosis assessment Lesions larger than 5mm can not be assessed Left or right portal vein can not be assessed Left, right and middle hepatic veins can be seen Homogenous signals or contours defined Margins and mass of adrenals can be seen but not confident Most of the vertebrae or disc can be seen
3 (Diagnostic) Confident in the margins of renal artery and accessory renal arteries Some lesions larger than 5mm can be assessed Both left and right portal vein can be assessed Branch hepatic veins can be seen Homogenous signals and contours defined margins and mass of adrenals can be seen Most of the vertebrae and disc can be seen
4 (Good) First order branches of renal arteries are sharply delineated All lesions larger than 5mm can be assessed Some second order branches can be assessed Branch hepatic veins can be seen sharply Homogenous signals; contours defined; Pancreatic duct partly seen Most of the margins are clearly defined All the vertebrae and disc can be seen
5 (Excellent) Up to second order branches of renal arteries are sharply delineated Lesions larger than 2mm can be assessed All second order branches can be assessed Branch hepatic veins to within 1 cm of capsule can be seen Can see entire pancreatic duct (from tail to ampulla) All the margins are well defined All the vertebrae and disc can be seen; growth plate sharply defined
6 (Outstanding) - - Vessels to within 1 mm of capsule can be assessed - - - -

Table 4.

Scoring Criteria for Paired Side-by-side Assessments of Quality of Structural Delineation*

Score Quality of Structural Delineation (Left vs. Right)
-3 Structures seen on the left but not on the right
-2 Better delineated on the left than on the right
-1 Aesthetic improvement on the left
0 Equivalent
1 Aesthetic improvement on the right
2 Better delineated on the right than on the left
3 Structures seen on the right but not no the left
*

One reconstruction was presented on the left side of the screen and the other on the right side in a randomized blinded fashion for comparisons. For quality of structural delineation, structures evaluated were renal artery and branches, liver, portal vein and branches, hepatic vein and branches, pancreatic duct, adrenal, spine, and lesion. Structures on both sides were evaluated unless absent.

Statistical Analysis of Overall Image Quality

The assessed images were first divided into two groups based on the scores: diagnostically acceptable and non-acceptable. The percentage of the number of cases that were diagnostically acceptable was calculated for each reconstruction. Then, a paired Wilcoxon test was performed to assess the null hypothesis that there was no significant difference in image quality between ARC and L1-SPIRiT, ARC and CC-L1-SPIRiT, and L1-SPIRiT and CC-L1-SPIRiT, when assessed individually, for qualitative SNR, image contrast, synthetic appearance, coherent structural artifacts and image blurring.

Statistical Analysis of Delineation of Specific Abdominal Anatomical Structures

Individual assessments

A paired Wilcoxon test was performed to assess the null hypothesis that there was no significant difference in image quality between ARC and L1-SPIRiT, ARC and CC-L1-SPIRiT, and L1-SPIRiT and CC-L1-SPIRiT, when images were assessed individually for anatomical structures (renal arteries, liver, hepatic veins, portal veins, pancreas, adrenals, and spine). For the seven tests of anatomical structural delineation, a Benjamini-Hochberg multiple comparison correction was also applied to adjust to the statistical significance level.

Paired comparisons

For evaluations with images shown side-by-side in pairs, a paired Wilcoxon test was performed to assess the null hypothesis that there was no significant preference in quality of structural delineation between ARC and L1-SPIRiT, ARC and CC-L1-SPIRiT, and L1-SPIRiT and CC-L1-SPIRiT.

In all the aforementioned analysis, a P value of 0.05 was used as a criterion of statistical significance. Inter-observer agreements between the two readers for all qualitative image assessments and delineation of anatomical structures were analyzed using bi-rater weighted kappa coefficients. The weighted kappa coefficients were interpreted as almost perfect (0.8-1), substantial (0.6-0.8), moderate (0.4-0.6), fair (0.2-0.4), slight (0-0.2) and poor (<0).

RESULTS

Reconstruction time for each acquisition (each temporal phase) with CC-L1-SPIRiT ranged from 28-48 seconds (mean, 36 seconds) compared to 202-364 seconds with L1-SPIRiT (mean, 275 seconds), for nearly an eight-fold reduction in image reconstruction time. This reconstruction time of less than one minute is acceptable for clinical use, even when resulting images have to be inspected to confirm adequate image quality prior to releasing the patient. The reconstruction time for each case is shown in Tab. 1. Figure 1 shows a representative results in a 6-year-old female referred for abdominal pain and concern for inflammatory bowel disease, and highlights the overall improvement in image quality with a compressed sensing reconstruction.

Figure 1.

Figure 1

Example of ARC (a), L1-SPIRiT (b), and CC-L1-SPIRiT (c) reconstructions of a 6-year-old female. Note delineation of hepatic vein branch (black arrow), portal vein (white arrow), pancreatic duct (large dashed arrow), bile duct (small dashed arrow), and cortex of left iliac bone (short white arrow).

Individual Assessment Of Overall Image Quality

SNR

Results for all qualitative image assessments are detailed in Table 5 and 6. For reader 1, all cases had diagnostically acceptable qualitative SNR with L1-SPIRiT and CC-L1-SPIRiT reconstructions. For reader 2, 27/29 and 28/29 cases had diagnostically acceptable SNR for L1-SPIRiT and CC-L1-SPIRiT, respectively. With ARC reconstruction, non-acceptable SNR was present in 13/29 cases for reader 1 and 11/29 cases for reader 2.

Table 5.

Individual Assessments of Overall Image Quality: the proportion with diagnostically acceptable image quality for each reconstruction is shown (29 cases).

Overall Image Quality Number of cases (proportions) Reader 1 Number of cases (proportions) Reader 2

ARC L1-SPIRiT CC-L1-SPIRiT ARC L1-SPIRiT CC-L1-SPIRiT
Qualitative SNR 16 (55%) 29 (100%) 29 (100%) 18 (62%) 27 (93%) 28 (97%)
Image contrast 29 (100%) 29 (100%) 29 (100%) 24 (83%) 26 (90%) 27 (93%)
No Synthetic appearance 29 (100%) 26 (90%) 24 (83%) 29 (100%) 29 (100%) 28 (100%)
No Structural coherent artifacts 29 (100%) 29 (100%) 29 (100%) 29 (100%) 27 (93%) 29 (100%)
No Image blurring 28 (97%) 28 (97%) 28 (97%) 17 (59%) 24 (83%) 26 (90%)
Table 6.

Mean scores of qualitative image assessments and delineation of anatomical structures, and the paired Wilcoxon ranked sum test between ARC and L1-SPIRiT, ARC and CC-L1-SPIRiT, and L1-SPIRiT and CC-L1-SPIRiT (when reconstructed images were assessed individually)*

Mean Scores (Reader1/Reader2) P values of the Wilcoxon test (Reader 1/Reader2)
ARC L1-SPIRIT CC-L1-SPIRiT ARC vs L1-SPIRiT ARC vs CC-L1-SPIRiT L1-SPIRiT vs CC-L1-SPIRiT
Qualitative Image Assessments
Qualitative SNR 2.7/2.7 4.9/3.8 5.0/4.0 <0.05/<0.05 <0.05/<0.05 1.0/0.14
Coherent Structural Artifacts 5.0/3.8 4.6/4.2 4.7/4.3 <0.05/<0.05 <0.05/<0.05 0.79/0.46
Image Contrast 3.9/3.0 4.0/3.4 3.9/3.6 0.35/<0.05 1.0/<0.05 0.35/0.34
Image Blurring 3.9/2.5 3.8/2.9 3.8/2.9 <0.05/<0.05 0.15/<0.05 0.42/0.77
Synthetic Appearance 4.0/3.9 3.5/3.6 3.1/3.6 <0.05/<0.05 <0.05/<0.05 <0.05/1.0
Delineation of Anatomical Structures
Renal Artery 4.0/3.6 4.7/3.9 4.8/4.1 <0.05/0.06 <0.05/<0.05 0.34/0.17
Liver 3.3/3.2 4.7/3.7 4.6/3.8 <0.05/<0.05 <0.05/<0.05 0.07/0.62
Portal Veins 5.0/4.3 5.8/4.9 5.9/4.9 <0.05/<0.05 <0.05/<0.05 0.34/1.0
Hepatic Veins 3.2/2.1 4.3/3.1 4.3/3.2 <0.05/<0.05 <0.05/<0.05 1.0/0.49
Pancreas 2.8/2.5 3.8/3.6 3.8/3.6 <0.05/<0.05 <0.05/<0.05 0.82/1.0
Adrenals 3.2/3.2 4.4/3.7 4.5/4.0 <0.05/<0.05 <0.05/<0.05 0.81/0.11
Spine 2.9/2.8 4.8/4.1 4.8/4.0 <0.05/<0.05 <0.05/<0.05 1.0/0.53
*

P<0.05 was considered as a statistically significant level (shown in bold)

Image Contrast

All cases had diagnostically acceptable image contrast with ARC, L1-SPIRiT and CC-L1-SPIRiT reconstructions for reader 1. For reader 2, the proportion of cases with significant degradation of image contrast progressively decreased from ARC reconstruction to L1-SPIRiT and finally CC-L1-SPIRiT, with at least 90% of cases having acceptable image contrast for all CS reconstructions.

Synthetic Appearance

None of the cases with ARC had a significant synthetic appearance for both reader 1 and reader 2. However, for reader 1, a significant synthetic appearance was seen in three cases with L1-SPIRiT and five cases with CC-L1-SPIRiT. Similarly for reader 2, synthetic appearance was present in one case with CC-L1-SPIRiT reconstruction. Figure 2(b) shows an example of synthetic appearance in L1-SPIRiT reconstruction. However, the synthetic appearance in the reconstructed images depends significantly on the reconstruction parameters (e.g., Wavelet threshold). By adjusting the reconstruction parameters, synthetic appearance can be removed as shown in Figure 2(c).

Figure 2.

Figure 2

Example of ARC reconstruction (a), L1-SPIRiT reconstruction with the default reconstruction parameters (b), L1-SPIRiT reconstruction with adjusted reconstruction parameters (c), and CC-L1-SPIRiT reconstruction with the default reconstruction parameters (d) of a 6-year-old female. Note delineation of pancreatic duct (arrow). L1-SPIRiT reconstruction with the default reconstruction parameters has a synthetic appearance, which is reflected by the circled cartoon-like almost “noiseless” liver. The synthetic appearance was removed with adjusted reconstruction parameters.

Coherent Structural Artifacts

For reader 1, all 29/29 cases had either no or minimal structural coherent artifacts with ARC, L1-SPIRiT and CC-L1-SPIRiT reconstructions. Since all three reconstructions were acquired using Poisson disc pseudo-random k-space sampling, coherent structural artifacts were neither present nor expected. Similarly, for reader 2, all 29/29 cases had no or minimal structural coherent artifacts with ARC and CC-L1-SPIRiT reconstructions. However, structural artifacts were present in two cases with L1-SPIRiT.

Image Blurring

For reader 1, each of the three reconstructions had one case with a non-diagnostic degree of blurring, which was influenced by motion. For reader 2, the proportion of cases with a significant degree of blurring decreased progressively from ARC (41%) to L1-SPIRiT (17%) and then CC-L1-SPIRiT (10%) reconstruction.

Aside from the analysis that was focused on the proportion of cases with a diagnostically acceptable image quality, comparisons of qualitative image assessments between ARC and L1-SPIRiT, ARC and CC-L1-SPIRiT, and L1-SPIRiT and CC-L1-SPIRiT, when assessed individually are detailed in Table 6. Figures 3 shows representative results of blinded randomized scoring of overall qualitative image assessments. From Table 6, we can see that both L1-SPIRiT and CC-L1-SPIRiT have a statistically significant difference compared with ARC when assessed individually (paired Wilcoxon rank sum test, P<0.05) in qualitative SNR, coherent structural artifacts and synthetic appearance for both readers. Similar statistically significant difference is also achieved in image contrast and image blurring for reader 2, but not for reader 1. L1-SPIRiT and CC-L1-SPIRiT do not have a statistically significant difference in overall image quality for both readers except in synthetic appearance for reader 1.

Figure 3.

Figure 3

Representative results of overall image quality assessments for ARC, L1-SPIRiT and CC-L1-SPIRiT reconstructions for Reader 1 and Reader 2: (a) Qualitative SNR; (b) Image contrast; (c) Synthetic appearance; (d) Coherent structural artifacts; and (e) Image blurring. Each color bar represents the percentage of number of the cases with the same score. The mean score of each reconstruction is shown on top of the color bar.

Delineation Of Specific Abdominal Anatomical Structures

Individual Assessment

Delineation for all anatomical structures except the renal artery was superior with both L1-SPIRiT and CC-L1-SPIRiT reconstructions compared with ARC (paired Wilcoxon rank sum test, P<0.05) for both readers. However, no statistically significant difference was found between L1-SPIRiT and CC-L1-SPIRiT reconstructions for delineation of all anatomical structures for reader 1 or reader 2, as detailed in Table 6. Figure 4 shows representative results of blinded randomized order scoring of delineation of renal arteries, liver, portal veins, hepatic veins, pancreas, adrenals, and spine with ARC, L1-SPIRiT, and CC-L1-SPIRiT reconstructions for both readers. After the Benjamini-Hochberg multiple comparison correction, statistical significance was still achieved to reject the null hypothesis that there was no significant difference in image quality between ARC and L1-SPIRiT, and ARC and CC-L1-SPIRiT for all anatomical structures except the renal artery.

Figure 4.

Figure 4

Representative results of the delineation of anatomical structures for ARC, L1-SPIRiT and CC-L1-SPIRiT reconstructions for Reader 1 and Reader 2: (a) Renal artery; (b) liver; (c) Portal vein; (d) Hepatic vein; (e) pancreas; (f) adrenals and (g) spine. Each color bar represents the percentage of number of the cases with the same score. The mean score of each reconstruction is shown on top of the color bar.

Paired Comparisons

Delineation of all anatomical structures was superior with L1-SPIRiT and CC-L1-SPIRiT compared with ARC for each reader, as summarized in Table 7. Further, for reader 1, there was no clear preference for L1-SPIRiT or CC-L1-SPIRiT when compared against each other for most structures. Reader 2 preferred CC-L1-SPIRiT for some anatomical structures (renal artery, portal vein, pancreas, adrenals and spine) and had no strong preference between L1-SPIRiT and CC-L1-SPIRiT in the other anatomical structures. The Wilcoxon test results are also shown in Table 7. The comparisons of all seven structures were combined and the results were illustrated using bar graphs in Fig. 5. Again, CC-L1-SPIRiT had superior or equivalent structural delineation compared to ARC for all comparisons and both readers (Fig. 5 (c, d)).

Table 7.

Paired assessment of delineation of anatomical structures. The number of cases with preference for ARC, L1-SPIRiT, or CC-L1-SPIRiT when compared in pairs is shown for each anatomical structure using the criteria in Table 4. Note cells with zero cases are left blank in the table.*

score ARC vs L1-SPIRiT ARC vs CC-L1-SPIRiT L1-SPIRiT vs CC-L1-SPIRiT
-3 -2 -1 0 1 2 3 -3 -2 -1 0 1 2 3 -3 -2 -1 0 1 2 3
Renal Artery
Reader 1 7 12 10 7 16 6 26 3
P < 0.05 P < 0.05 P = 0.15
Reader 2 1 16 9 3 12 15 2 23 5 1
P < 0.05 P < 0.05 P < 0.05
Portal Vein
Reader 1 2 9 18 5 23 1 6 18 5
P < 0.05 P < 0.05 P = 0.80
Reader 2 1 11 10 6 1 3 22 4 20 8 1
P < 0.05 P < 0.05 P < 0.05
Hepatic Vein
Reader 1 2 5 22 2 6 20 1 1 5 19 4
P < 0.05 P < 0.05 P = 0.44
Reader 2 2 7 9 8 3 4 10 14 1 1 22 5 1
P < 0.05 P < 0.05 P = 0.07
Pancreas
Reader 1 16 5 8 15 4 8 2 1 22 5 1
P < 0.05 P < 0.05 P = 0.07
Reader 2 1 2 9 11 5 1 8 16 4 1 18 9 2
P < 0.05 P < 0.05 P < 0.05
Adrenals
Reader 1 6 15 8 4 14 11 3 21 5
P < 0.05 P < 0.05 P = 0.53
Reader 2 22 5 1 1 14 12 3 22 6 1
P < 0.05 P < 0.05 P < 0.05
Spine
Reader 1 28 1 28 1 2 9 16 2
P < 0.05 P < 0.05 P < 0.05
Reader 2 2 5 22 6 22 1 1 20 7 1
P < 0.05 P < 0.05 P < 0.05
Lesions
Reader 1 8 16 1 5 18 4 19 1
P < 0.05 P < 0.05 P = 0.23
Reader 2 10 11 3 8 11 5 21 3
P < 0.05 P < 0.05 P = 0.15
*

P<0.05 was considered as a statistically significant level (shown in bold)

Figure 5.

Figure 5

Representative results of the paired assessment of structural delineation based on the scoring criteria in Tab. 4. Bar graphs were generated based on the comparisons of all seven anatomical structures. The scores of the comparisons (-3, -2, -1, 0, 1, 2, 3) were represented by (⋙, ≫, >, =, <, ≪, ⋘) respectively in the bar graph. For example, “ARC > L1-SPIRiT” means that there was aesthetic improvement in ARC than L1-SPIRiT. The comparison results between ARC and L1-SPIRiT are shown in (a) and (b) for reader 1 and reader 2 respectively. The comparison results between ARC and CC-L1-SPIRiT are shown in (c) and (d). The comparison results between L1-SPIRiT and CC-L1-SPIRiT are shown in (e) and (f).

In our study cohort, 24 patients had pathological lesions. For reader 1, delineation of lesions was superior with L1-SPIRiT reconstruction compared with ARC in all cases, and was superior in 23/24 cases with CC-L1-SPIRiT compared with ARC. For reader 2, delineation of lesions was superior in 14/24 cases with L1-SPIRiT compared with ARC and in 16/24 cases with CC-L1-SPIRiT compared with ARC. For reader 2, the difference between L1-SPIRiT and ARC as well as between CC-L1-SPIRiT and ARC for delineation of lesions again showed a significant improvement with compressed sensing, but not as strongly as with reader 1.

Inter-observer Agreement

Overall inter-observer agreement for image qualitative assessments and delineation of specific anatomical structures are shown in Table 8. Two readers had fair to substantial agreement for most delineation of specific anatomical structures with ARC, L1-SPIRiT and CC-L1-SPIRiT. However, the agreement on overall image quality assessments was not as good as structural delineation because the scoring criteria for overall image quality were more subjective.

Table 8.

Inter-observer agreement results using weighted kappa coefficients between Reader 1 and Reader 2 for overall image quality assessments and delineation of anatomical structures for ARC, L1-SPIRiT and CC-L1-SPIRiT.*

ARC L1-SPIRiT CC-L1-SPIRiT
Overall Image Quality
Qualitative SNR Fair Slight Slight
Coherent Structural Artifacts Slight Poor Slight
Image Contrast Poor Poor Poor
Image Blurring Slight Fair Fair
Synthetic Appearance Slight Poor Slight
Delineation of Anatomical Structures
Renal Artery Fair Moderate Fair
Liver Moderate Fair Slight
Portal Vein Moderate Slight Poor
Hepatic Vein Fair Fair Moderate
Pancreas Fair Substantial Substantial
Adrenals Substantial Moderate Substantial
Spine Substantial Slight Slight
*

Interpretation of weighted kappa coefficient (kw) used: kw<0, poor; 0<=kw<0.2, slight; 0.2<=kw<0.4, fair; 0.4<=kw<0.6, moderate; 0.6<=kw<0.8, substantial; 0.8<=kw<1, almost perfect.

DISCUSSION

This work addresses some of the challenges of pediatric MR imaging by leveraging a compressed sensing approach to double the speed of contrast-enhanced abdominal MRI. The study showed that compressed sensing, when combined with a fast image reconstruction method, is feasible in a pediatric clinical setting. Delineation of anatomical structures that are routinely evaluated in a clinical setting shows improved quality with a compressed sensing reconstruction, and further, coil compression does not adversely affect image quality. The results of individual overall image quality assessment suggest that a compressed sensing reconstruction will yield diagnostically acceptable SNR that is unlikely to be achieved by a parallel imaging alone. Image contrast degradation, image blurring and coherent structural artifacts are unlikely to be seen in a compressed sensing reconstruction. The results of side-by-side comparison show that compressed sensing reconstructions are likely to provide superior structural delineation than traditional parallel imaging and that accelerated image reconstruction via coil compression can be accomplished with no significant image degradation.

In this study, the average reconstruction time for combined parallel imaging compressed sensing method without coil compression is approximately 275 seconds. In a typical contrast-enhanced abdominal imaging setting, one pre-contrast and up to four post-contrast acquisitions are usually acquired. As a result, the whole image reconstruction takes up to approximately 23 minutes. This is not practical in a clinical setting where image quality needs to be confirmed before patients are released. However, when coil compression is applied, the entire reconstruction time is reduced to less than 3 minutes. If receiving coils with higher channel counts (more than 32) are used, the reconstruction time will increase even further. However, arguably for the same acquisition, the number of virtual coils needed may not increase dramatically. Thus, a practical reconstruction time will be maintained.

The combined parallel imaging compressed sensing reconstructions (L1-SPIRiT and CC-L1-SPIRiT) implemented in this work were parameter-dependent. The parameters used in the study were chosen empirically and were fixed for all the cases. Synthetic appearance may appear in the image due to the fixed non-optimal reconstruction parameters for specific cases. However, it can be removed by properly adjusting the reconstruction parameters. In general, the reconstruction parameters applied in this study provided at least 90% of reconstructions without obvious synthetic appearance. Reconstructions with an automatic parameter selection may completely remove the synthetic appearance in the reconstructed images, but are beyond the scope of this manuscript.

Overall, we did not find much difference in image quality between L1-SPIRiT (compressed sensing) and CC-L1-SPIRiT (compressed sensing and coil compression) reconstructions, though images obtained with CC-L1-SPIRiT showed superior quality for some overall image qualitative assessments and delineation of anatomical structures, compared with L1-SPIRiT. However, CC-L1-SPIRiT increased the synthetic appearance of images while decreasing image blurring. This image assessment is subjective, and the individual radiologist preference for this synthetic appearance may vary, particularly with increasing exposure to and clinical experience with these types of images.

In this study, quantitative assessment of image quality was not performed. Calculation of parameters such as SNR in the context of parallel imaging is technically difficult and usually not feasible clinically; further, these complexities are even more challenging when compressed sensing is introduced. Thus, subjective image quality and delineation of abdominal anatomical structures were thoroughly evaluated in a clinical setting. However, arguably these subjective assessments, along with diagnostic accuracy, are ultimately more relevant to clinical adoption of an imaging method.

The qualitative assessments in the study were subjective: different readers may have different interpretation of the scoring criteria of the image assessment, especially for overall image quality. For example, because a variable density sampling pattern was used, the resulting image resolution of a parallel imaging reconstruction compared to compressed sensing reconstruction was not clearly predictable, and radiologists’ perception of image blurring and resolution is subjective. Unfortunately, to understand the impact of compressed sensing reconstruction on image resolution, phantom studies may not be applicable because the phantoms are usually much sparser than in vivo images. Fully sampled arterial phase in vivo datasets could be more ideal, but are impossible to acquire for contrast-enhanced abdominal MRI due to contrast passage during the long time (>2 minutes) to acquire a high spatial resolution fully-sampled dataset. However, with clearly defined criteria, the evaluation of delineation of specific anatomical structures is arguably more practical to assess these images. From Figure 3, Figure 4 and Table 8, we can see a much better inter-observer agreement for the delineation of anatomical structures, where the scoring criteria is defined less subjectively, than for overall image quality assessment.

The study cohort was a heterogeneous patient population, which included patients with abdominal tumors (hepatocellular carcinoma, Wilms tumor, benign liver tumors), a ventricular mass, and solid-organ transplants. The benefit of including patients with multiple and diverse clinical conditions is the assessment of the feasibility of using CC-L1-SPIRiT in various clinical circumstances. However, further validation should be directed towards specific patient population with specific diseases.

One limitation of our study was that the acceleration factor was fixed in this study. At this high acceleration of 7.2, compressed sensing reconstructions clearly improved image quality and anatomical structures were better delineated. This lends evidence that this approach to imaging may be acceptable for clinical examinations in a pediatric setting. However, we did not perform the study at lower and higher accelerations. Therefore, further studies should be performed using different acceleration factors, enabling determination of the factor at which the reconstruction deteriorates to an unacceptable degree.

Anatomical structures routinely evaluated on abdominal MRI exams were clearly delineated with compressed sensing reconstructions. In addition, pathological lesions, like bone lesions and abdominal tumors were well delineated with CC-L1-SPIRiT compared with ARC. However, given the small sample sizes of patients with any specific lesion, our study does not address the diagnostic accuracy of using a compressed sensing and coil compression for any particular imaging indication.

Acknowledgments

The authors thank Prof. Brian Hargreaves for valuable discussions.

Grant Sponsors: This work was supported by NIH grants R01 EB009690, P41 EB015891, John & Tashia Morgridge Faculty Scholars Fund, and GE Healthcare.

References

  • 1.Olsen OE. Imaging of abdominal tumours: CT or MRI? Pediatr Radiol. 2008;38(Suppl 3):452–458. doi: 10.1007/s00247-008-0846-5. [DOI] [PubMed] [Google Scholar]
  • 2.Darge K, Anupindi SA, Jaramillo D. MR imaging of the abdomen and pelvis in infants, children, and adolescents. Radiology. 2011;261:12–29. doi: 10.1148/radiol.11101922. [DOI] [PubMed] [Google Scholar]
  • 3.Vasanawala SS, Lustig M. Advances in pediatric body MRI. Pediatr Radiol. 2011;41(Suppl 2):549–554. doi: 10.1007/s00247-011-2103-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Sury MR, Smith JH. Deep sedation and minimal anesthesia. Paediatr Anaesth. 2008;18:18–24. doi: 10.1111/j.1460-9592.2007.02379.x. [DOI] [PubMed] [Google Scholar]
  • 5.Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952–962. [PubMed] [Google Scholar]
  • 6.Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA) Magn Reson Med. 2002;47:1202–1210. doi: 10.1002/mrm.10171. [DOI] [PubMed] [Google Scholar]
  • 7.Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med. 1990;16:192–225. doi: 10.1002/mrm.1910160203. [DOI] [PubMed] [Google Scholar]
  • 8.Donoho D. Compressed sensing. IEEE Trans Inf Theory. 2006;52:1289–1306. [Google Scholar]
  • 9.Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182–1195. doi: 10.1002/mrm.21391. [DOI] [PubMed] [Google Scholar]
  • 10.Liang D, Liu B, Wang J, Ying L. Accelerating SENSE using compressed sensing. Magn Reson Med. 2009;62:1574–1584. doi: 10.1002/mrm.22161. [DOI] [PubMed] [Google Scholar]
  • 11.Vasanawala SS, Alley MT, Hargreaves BA, Barth RA, Pauly JM, Lustig M. Improved pediatric MR imaging with compressed sensing. Radiology. 2010;256:607–616. doi: 10.1148/radiol.10091218. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Vasanawala SS, Murphy MJ, Alley MT, et al. Practical parallel imaging compressed sensing MRI: Summary of two years of experience in accelerating body MRI of pediatric patients. Proceedings of IEEE Int Symp Biomed Imag; Chicago. 2011. pp. 1039–1043. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Brau AC, Beatty PJ, Skare S, Bammer R. Comparison of reconstruction accuracy and efficiency among autocalibrating data-driven parallel imaging methods. Magn Reson Med. 2008;59:382–395. doi: 10.1002/mrm.21481. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Murphy M, Alley M, Demmel J, Keutzer K, Vasanawala S, Lustig M. Fast L1-SPIRiT compressed sensing parallel imaging MRI: scalable parallel implementation and clinically feasible runtime. IEEE Trans Med Imaging. 2012;31:1250–1262. doi: 10.1109/TMI.2012.2188039. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Vasanawala S, Grafendorfer T, Calderon P, et al. Millimeter isotropic resolution volumetric pediatric abdominal MRI with a dedicated 32 channel phased array coil. Proceedings of the 19th Annual Meeting of ISMRM; Montreal. 2011. p. 161. [Google Scholar]
  • 16.McDougall MP, Wright SM. 64-channel array coil for single echo acquisition magnetic resonance imaging. Magn Reson Med. 2005;54:386–392. doi: 10.1002/mrm.20568. [DOI] [PubMed] [Google Scholar]
  • 17.Schmitt M, Potthast A, Sosnovik DE, et al. A 128-channel receive-only cardiac coil for highly accelerated cardiac MRI at 3 Telsa. Magn Reson Med. 2008;59:1431–1439. doi: 10.1002/mrm.21598. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Buehrer M, Pruessmann KP, Boesiger P, Kozerke S. Array compression for MRI with large coil arrays. Magn Reson Med. 2007;57:1131–1139. doi: 10.1002/mrm.21237. [DOI] [PubMed] [Google Scholar]
  • 19.Huang F, Vijayakumar S, Li Y, Hertel S, Duensing GR. A software channel compression technique for faster reconstruction with many channels. Magn Recon Imaging. 2008;26:133–141. doi: 10.1016/j.mri.2007.04.010. [DOI] [PubMed] [Google Scholar]
  • 20.Zhang T, Pauly JM, Vasanawala SS, Lustig M. Coil compression for accelerated imaging with Cartesian sampling. Magn Reson Med. 2013;69:571–582. doi: 10.1002/mrm.24267. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Lustig M, Pauly JM. SPIRiT: iterative self-consistent parallel imaging reconstruction from arbitrary k-space. Magn Reson Med. 2010;64:457–471. doi: 10.1002/mrm.22428. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Alley MT, Murphy MJ, Keutzer K, et al. Improved time-resolved, 3D phase contrast imaging through variable Poisson sampling and partial respiratory triggering. Proceedings of the 19th Annual Meeting of ISMRM; Montreal. 2011. p. 1218. [Google Scholar]
  • 23.Beatty P, Brau A, Chang S, et al. A method for autocalibrating 2-D accelerated volumetric parallel imaging with clinically practical reconstruction times. Proceedings of the 15th Annual Meeting of ISMRM; Berlin. 2007. p. 1749. [Google Scholar]

RESOURCES