. 2014 Apr 14;9(4):e94627. doi: 10.1371/journal.pone.0094627

Comparison of M.I.C.E. and Etest with CLSI Agar Dilution for Antimicrobial Susceptibility Testing against Oxacillin-Resistant Staphylococcus spp

Eloiza H Campana 1,*, Cecilia G Carvalhaes 1, Bruna Nonato 1, Antonia M de O Machado 2, Ana C Gales 1
Editor: Herman Tse3
PMCID: PMC3986405  PMID: 24732297

Abstract

Objective

The main objective of this study was to comparatively evaluate the performance of M.I.C.E. and Etest methodologies to that of agar dilution for determining the antimicrobial susceptibility profile of oxacillin-resistant Staphylococcus spp.

Methods

A total of 100 oxacillin-resistant Staphylococcus spp. isolates were collected from hospitalized patients at a teaching hospital. Antimicrobial susceptibility testing for vancomycin, teicoplanin and linezolid was performed using the reference CLSI agar dilution method (2009), Etest and M.I.C.E. methodologies. The MIC values were interpreted according to CLSI susceptibility breakpoints and compared by regression analysis.

Results

In general, the essential agreement (±1-log2) between M.I.C.E. and CLSI agar dilution was 93.0%, 84.0% and 77.0% for linezolid, teicoplanin and vancomycin, respectively. Essential agreement rates between M.I.C.E. and Etest were excellent (>90.0%) for all antibiotics tested. Both strips (M.I.C.E. and Etest) yielded two very major errors for linezolid. Unacceptable minor rates were observed for teicoplanin against CoNS and for vancomycin against S. aureus.

Conclusions

According to our results, linezolid and teicoplanin MICs against all staphylococci and S. aureus, respectively, were more accurately predicted by M.I.C.E. strips. However, the Etest showed better performance than M.I.C.E. for predicting vancomycin MICs against all staphylococci. Thus, microbiologists must be aware of the different performance of commercially available gradient strips against staphylococci.

Introduction

Determination of vancomycin minimum inhibitory concentration (MIC) has been of crucial importance to guide antimicrobial therapy against staphylococcal infections since intermediate resistance to vancomycin has not been accurately detected by disc diffusion [1]. In addition, treatment failures have been reported when vancomycin is prescribed for treatment of oxacillin-resistant Staphylococcus aureus (ORSA) infections, especially for strains exhibiting vancomycin MICs, ≥ 2 µg/mL. Studies have suggested that vancomycin treatment success rates are indirectly proportional to vancomycin MICs, i.e., vancomycin treatment success decreases as the MIC of the ORSA strains increases [2]. In this manner, alternative therapeutic drugs such as linezolid have been considered for treatment of ORSA infections. Linezolid has activity against clinically significant gram-positive cocci, including ORSA and coagulase-negative Staphylococcus (CoNS). The Clinical Laboratory Standards Institute (CLSI) document M100-S22 [3] recommends the determination of the linezolid MIC for isolates categorized as non-susceptible by disc-diffusion methodology. Therefore, the Oxoid M.I.C.Evaluator™ (M.I.C.E.) (Thermo Fisher Scientific, Basingstoke, UK) methodology represents a rapid alternative for determining vancomycin and linezolid MICs.

In a previous study, we had compared the vancomycin MICs determined by M.I.C.E. with those obtained by CLSI broth microdilution (BMD) and observed that the vancomycin MICs values determined by M.I.C.E. were higher than those obtained by BMD [4]. A similar finding was latter reported by Rennie et al. [5]. We thought these results could have resulted from the different techniques, gradient agar diffusion vs. BMD, employed. The clinical relevance of determining vancomycin MIC by a reliable technique motivated us to perform the current study, where the performance of M.I.C.E. was comparatively evaluated with those of CLSI agar dilution and Etest (AB bioMérieux, Marcy l'Étoile, France).

Materials and Methods

Ethical Statement

Ethical approval was not required because the study was conducted as part of surveillance control management. Written informed consent was not required because patients received routine clinical care, and there were no additional specimens collected or study-specific interventions. Patient records/information was anonymized and de-identified prior to analysis.

Bacterial Strains

A total of 100 clinical oxacillin-resistant Staphylococcus spp. isolates (50 S. aureus and 50 coagulase-negative staphylococci; CoNS) were collected from hospitalized patients at a Brazilian teaching hospital located in the city of São Paulo. All bacterial isolates were recovered from blood culture. Only a single isolate per patient was evaluated. S. haemolyticus (58%) was the most frequent specie among CoNS, followed by S. hominis (30%) and S. epidermidis (12%). No Staphylococcus lugdunensis isolate was identified in this collection. Confirmation of species identification was performed by MALDI-TOF mass spectrometry (Bruker Daltonics, Bremen, Germany) after confirming that the cultures were pure and possessed identical colony morphologies.

Antimicrobial Susceptibility Testing (AST)

AST was performed against vancomycin, teicoplanin and linezolid by M.I.C.E., Etest (gradient methods), and agar dilution (reference method) according to CLSI guidelines or respective manufacturers [3], [6]. Antimicrobial powders were purchased from Sigma-Aldrich (St. Louis, MO, USA). The Etest and M.I.C.E. MICs were determined as the value at which the elliptical growth margin intersected the strips, except for linezolid. For this antimicrobial agent, different manufacturer's recommendations have been established for MIC reading, i.e. 90% vs. 80% of the growth inhibition by Etest and M.I.C.E., respectively. MICs were rounded up to the next higher twofold dilution for comparison purposes. Quality control of susceptibility testing was performed by testing S. aureus ATCC 29213 and Enterococcus faecium ATCC 29212 with results within the CLSI expected ranges for all antimicrobials and AST methods tested. The MICs were read by three independent observers with no discordant MICs readings.

Statistical Analysis

The results of the MICs obtained by agar dilution, Etest and M.I.C.E. techniques were analyzed and compared by regression analysis using the Statistical Package for the Social Sciences, version 17.0 (SPSS Inc., Chicago, IL).

Essential agreement was defined when the result of the MICs obtained by Etest or M.I.C.E. ranged ±1-log2 dilution of those obtained by agar dilution (reference method). Differences of ≥2-log2 dilutions were considered as discordant results. Categorical agreement was defined as test results within the same susceptibility category. Errors were ranked as follows: very major errors, false susceptible results by gradient methods; major errors, false resistant results produced by gradient methods; and minor errors, intermediate by reference method and susceptible or resistant by the gradient methods or intermediate by the gradient methods and susceptible or resistant by the reference method. Acceptable error levels were ≤1.5% for very major errors, ≤3% for major errors and 10% for minor errors as recommended [7].

Results

In comparison to agar dilution, M.I.C.E. methodology yielded essential and categorical agreement rates of 77.0%/83.0%, 84.0%/87.0%, and 93.0%/98.0% for vancomycin, teicoplanin, and linezolid, respectively. When these rates were analyzed according to the staphylococcal species, lower essential and categorical agreement vancomycin rates were observed for S. aureus than CoNS isolates (66.0%/68.0% and 88.0%/98.0%, respectively), as shown in Table 1. In contrast, for teicoplanin, lower essential and categorical agreement rates were observed for CoNS (80.0%/74.0%) than S. aureus (88.0%/100%). When testing linezolid, a variation in the essential and categorical agreement rates was also observed according to staphylococcal species (98.0%/96.0% and 88.0%/100% for S. aureus and CoNS, respectively). Essential and categorical agreement rates for teicoplanin against CoNS varied according to CoNS species. These rates were lower for S. haemolythicus than those observed for S. epidermidis and S. hominis. Most discordant results led to occurrence of minor errors (16.0% and 12.0% for vancomycin and teicoplanin against S. aureus and CoNS, respectively). One (2.0%) major error and two (4.0%) very major errors were observed for vancomycin and linezolid against S. aureus, respectively.

Table 1. Essential and categorical agreement rates between gradient diffusion tests (M.I.C.E. and Etest) against Staphylococcus spp.

Diffusion Test M.I.C.E. Etest
Essential Categorical Essential Categorical
Vancomycin
S. aureus 66.0% 68.0% 92.0% 98.0%
S. epidermidis 83% 100,0% 100% 83%
S. haemolythicus 90% 97,0% 97% 97%
S. hominis 87% 100,0% 100% 93%
CoNS 88.0% 98.0% 98.0% 94.0%
Generala 77.0% 83.0% 95.0% 96.0%
Teicoplanin
S. aureus 88.0% 100% 82.0% 100%
S. epidermidis 100% 100% 100% 83%
S. haemolythicus 76% 66% 72% 72%
S. hominis 80% 80% 67% 80%
CoNS 80.0% 74.0% 74.0% 76.0%
Generala 84.0% 87.0% 78.0% 88.0%
Linezolid
S. aureus 98.0% 96.0% 66.0% 96.0%
S. epidermidis 100% 100% 83% 100%
S. haemolythicus 90% 100% 72% 100%
S. hominis 80% 100% 47% 100%
CoNS 88.0% 100% 66.0% 100%
Generala 93.0% 98.0% 66.0% 98.0%

a. S. aureus and CoNS species.

The essential and categorical agreement rates obtained by Etest compared to agar dilution were 95.0%/96.0%, 78.0%/88.0%, and 66.0%/98.0% for vancomycin, teicoplanin, and linezolid, respectively. A trend for lower linezolid MICs was observed with Etest against both species analyzed. According to staphylococcal species no significant variation in the essential and categorical agreement rates were observed for vancomycin (92.0%/98.0% and 98.0%/94.0% for S. aureus and CoNS, respectively). However, lower essential and categorical agreement rates were observed for CoNS (74.0%/76.0%) than S. aureus (82.0%/100%) for teicoplanin. The majority of discordant results led to occurrence of minor errors (11.0% and 6.0% for teicoplanin and vancomycin, respectively). One (2.0%) major error against CoNS and two (4.0%) very major errors against S. aureus were observed for teicoplanin and linezolid, respectively.

An excellent concordance was observed between M.I.C.E. and Etest MIC results, which yielded essential agreement rates of 94.0%, 100.0% and 93.0% for vancomycin, teicoplanin, and linezolid, respectively. However, an unacceptable minor error rate (18.0%) was detected for vancomycin due to a trend of even higher MIC values by M.I.C.E. strips.

Discussion

We observed a trend for higher vancomycin MICs for both strips, Etest and M.I.C.E. This finding is corroborated by previous studies performed under CLSI recommendations independent of methodology tested, BMD or agar dilution [4], [5]. In general, vancomycin MICs determined by M.I.C.E. showed a 1-log2 dilution higher than those of Etest [4], [5]. It is important to notice that vancomycin CLSI susceptibility breakpoints for S. aureus (≤2 µg/mL) are 1-log2 dilution lower than those of CoNS (≤4 µg/mL). It could be one of the reasons for the better performance of M.I.C.E. strips against CoNS.

To the best of our knowledge, only five studies evaluating M.I.C.E. methodology have been published to date [4], [5], [8][10]. Only three of them evaluated Staphylococcus spp., as shown in Table 2. Our results are in agreement with those studies already published, except for those published by Mushtaq et al. [8]. These distinct findings could be due to the medium employed for AST, since Mushtaq et al. [8] tested Iso-Sensitest agar according to the British Society for Antimicrobial Chemotherapy (BSAC) guidelines. The distinct set of Staphylococcus isolates tested could also have contributed to the different results.

Table 2. Essential agreement rates between gradient diffusion tests (M.I.C.E. and Etest) and reference techniques (agar and broth microdilution) against Staphylococcus spp.

Method Agar dilution Broth microdilution
Guideline CLSIa BSACb CLSIa CLSIa
Antimicrobial This study Mushtaq et al. 2010 Carvalhaes et al. 2011 Rennie et al. 2012
M.I.C.E. Etest M.I.C.E. Etest M.I.C.E. Etest M.I.C.E. Etest
Vancomycin 77.0% 95.0% 95.6% 96.2% 72.5% 90.0% 61.0% 84.0%
Teicoplanin 84.0% 78.0% - - - - - -
Linezolid 93.0% 66.0% 99.3% 99.3% 97.5% 100% 74.0% 52.0%
N of isolates 100 154 40 157

a. CLSI, Clinical Laboratory Standards Institute;

b. BSAC, British Society for Antimicrobial Chemotherapy.

The present study was the only one to evaluate the performance of M.I.C.E. for testing teicoplanin against Staphylococcus spp. Despite of not observing a trend for higher teicoplanin MIC values, the categorical agreement for both strips (M.I.C.E. and Etest) compared to agar dilution against CoNS was lower than 80%. When the M.I.C.E. results for teicoplanin against CoNS were evaluated, the categorical agreement rate (74.0%) remained lower than that of essential agreement (80.0%). The breakpoints for interpreting the category of susceptibility to teicoplanin against CoNS vary between CLSI (susceptible ≤ 8 µg/mL; resistant ≥ 32 µg/mL) and EUCAST (susceptible ≤ 4 µg/mL; resistant > 4 µg/mL). This fact could justify the high rates of minor (24.0%) and major errors (14.0%) obtained by interpreting this compound by these respective guidelines, and reemphasize the need for revision of the current teicoplanin breakpoints against CoNS.

Different results have been reported for linezolid by different authors. It might be due to the use of distinct reading criteria (80% vs 90% of growth inhibition for M.I.C.E. and Etest, respectively). In the present study, a trend for lower linezolid MICs was noticed for Etest, in agreement with a previous report [5]. In general, linezolid MICs determined by Etest showed a 1-log2 dilution higher than those of agar dilution; however, this trend did not result in changes in the categorization of the isolates.

If M.I.C.E. results were interpreted according to the European Committee on Antimicrobial Susceptibility Testing (EUCAST) [11] breakpoints, no variation in the categorical agreement rates would be observed for vancomycin and linezolid against S. aureus and CoNS, as shown in Table 3. However, the categorical agreement rate for teicoplanin against S. aureus would drop from 100% to 68.0%, while it would increase from 74.0% to 80.0% against CoNS. In addition, by applying the EUCAST breakpoints, the error rates would be unacceptable for vancomycin and teicoplanin against S. aureus and for teicoplanin against CoNS. These different rates could be consequent to the absence of an intermediate category by EUCAST.

Table 3. Essential and categorical agreement rates by applying CLSI and EUCAST breakpoints against S. aureus and CoNS.

Specie S. aureus CoNS
Guideline CLSIa EUCASTb CLSIa EUCASTb
Antimicrobial Categorical Agreement Minor Errors Major Errors Very Major Errors Categorical Agreement Minor Errors Major Errors Very Major Errors Categorical Agreement Minor Errors Major Errors Very Major Errors Categorical Agreement Minor Errors Major Errors Very Major Errors
M.I.C.E. Vancomycin 68.0% 30.0% 2.0% 0% 68.0% - 32.0% 0% 98.0% 2.0% 0% 0% 98.0% - 2.0% 0%
Teicoplanin 100% 0% 0% 0,0% 68.0% - 28.0% 4.0% 74.0% 24.0% 0% 2% 80.0% - 14.0% 6.0%
Linezolid 96.0% 0% 0% 4.0% 96.0% - 0% 4.0% 100% 0% 0% 0% 100% - 0% 0%
Etest Vancomycin 98.0% 2.0% 0% 0% 98.0% - 2.0% 0% 94.0% 6.0% 0% 0% 98.0% - 2.0% 0%
Teicoplanin 100% 0% 0% 0% 62.0% - 34.0% 4.0% 76.0% 22.0% 2% 0% 72.0% - 22.0% 6.0%
Linezolid 96.0% 0% 0% 4.0% 96.0% - 0% 4.0% 100% 0% 0% 0% 100% - 0% 0%
Susceptibility profile S% I% R% S% I% R% S% I% R% S% I% R%
Vancomycin 100% 0% 0% 100% - 0% 100% 0% 0% 100% - 0%
Teicoplanin 100% 0% 0% 96% - 4% 72% 12% 16% 58% - 42%
Linezolid 96% 0% 4% 96% - 4% 100% 0% 0% 100% - 0%

CLSI, Clinical Laboratory Standards Institute. Breakpoints: Vancomicyn: S. aureus S≤2, I 4-8, R≥16, CoNS S≤4, I 8-16, R≥32; Teicoplanin: S≤8, I 16, R≥32; Linezolid: S≤4, R≥8.

EUCAST, European Committee on Antimicrobial Susceptibility Testing. Breakpoints: Vancomicyn: S. aureus S≤2, R>2, CoNS S≤4, R>4; Teicoplanin: S. aureus S≤2, R>2, CoNS S≤4, R>4; Linezolid: S≤4, R>4.

According to our results, linezolid and teicoplanin MICs against all staphylococci and S. aureus, respectively, were more accurately predicted by M.I.C.E. strips. However, the Etest showed better performance than M.I.C.E. for predicting vancomycin MICs against all staphylococci. Thus, microbiologists must be aware of the different performance of commercially available gradient strips against staphylococci.

Acknowledgments

We would like to thank Dr. Danilo Elias Xavier, Dr. Raquel Girardello and the staff of Laboratório Central do Hospital São Paulo for valuable help in the development of this work.

Funding Statement

This study was partially supported by Thermo Fisher Scientific, São Paulo, Brazil. ACG is a researcher from the National Council for Science and Technological Development (CNPq), Ministry of Science and Technology, Brazil (Process number: 307816/2009-5). No additional external funding received for this study. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1. Tenover FC, Lancaster MV, Hill BC, Steward CD, Stocker SA, et al. (1998) Characterization of staphylococci with reduced susceptibility to vancomycin and other glycopeptides. J Clin Microbiol. 36: 1020–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Sakoulas G, Moise-Broder PA, Schentag J, Forrest A, Moellering RC Jr, et al. (2004) Relationship of MIC and bactericidal activity to efficacy of vancomycin for treatment of methicillin-resistant Staphylococcus aureus bacteremia. J Clin Microbiol. 42(6): 2398–402. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Clinical Laboratory Standards Institute (2012) Performance standards for antimicrobial susceptibility testing: twenty-second Informational Supplement. M100–S22. CLSI, Wayne, PA.
  • 4. Carvalhaes CG, Campana EH, Barbosa PP, Paula AM, Gales AC (2011) Comment on: Performance of the Oxoid M.I.C. Evaluator strips compared with the Etest and BSAC agar dilution. J. Antimicrob. Chemother. 66: 1192–1193. [DOI] [PubMed] [Google Scholar]
  • 5. Rennie RP, Turnbull L, Brosnikoff C, Cloke J (2012) First Comprehensive Evaluation of the M.I.C. Evaluator Device for Aerobic Gram-positive and Gram-negative Bacterial Species compared to Etest and CLSI Broth Micro-dilution. J Clin Microbiol. 50(4): 1147–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Clinical and Laboratory Standards Institute (2009) Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically. Approved guideline M7-A8. CLSI, Wayne, PA.
  • 7.National Committee for Clinical Laboratory Standards (2000) Development of in vitro susceptibility testing criteria and quality control parameters. Document M23-A2. National Committee for Clinical Laboratory Standards, Wayne, PA.
  • 8. Mushtaq S, Warner M, Cloke J, Mariya A-S, Livermore DM (2010) Performance of the Oxoid M.I.C. Evaluator strips compared with the Etest assay and BSAC agar dilution. J. Antimicrob. Chemother. 65: 1702–1711. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Rennie RP, Turnbull L, Brosnikoff C, Cloke J (2012) First comprehensive evaluation of the M.I.C. evaluator device compared to Etest and CLSI reference dilution methods for antimicrobial susceptibility testing of clinical strains of anaerobes and other fastidious bacterial species. J Clin Microbiol. 50(4): 1153–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Tristram SG (2008) A comparison of Etest, M.I.C.Evaluator strips and CLSI broth microdilution for determining β-lactam antimicrobial susceptibility in Haemophilus influenzae . J Antimicrob Chemother. 62(6): 1464–6. [DOI] [PubMed] [Google Scholar]
  • 11.European Committee on Antimicrobial Susceptibility Testing (2013) Breakpoint tables for interpretation of MICs and zone diameters. Version 3.1.