. 2014 May 5;9(5):e96661. doi: 10.1371/journal.pone.0096661

A Simplified Method to Measure Choroidal Thickness Using Adaptive Compensation in Enhanced Depth Imaging Optical Coherence Tomography

Preeti Gupta 1,2, Elizabeth Sidhartha 1,2, Michael J A Girard 1,3, Jean Martial Mari 4, Tien-Yin Wong 1,2,5,6, Ching-Yu Cheng 1,2,5,6,*
Editor: Yingfeng Zheng7
PMCID: PMC4010516  PMID: 24797674

Abstract

Purpose

To evaluate a simplified method to measure choroidal thickness (CT) using commercially available enhanced depth imaging (EDI) spectral domain optical coherence tomography (SD-OCT).

Methods

We measured CT in 31 subjects without ocular diseases using Spectralis EDI SD-OCT. The choroid-scleral interface of the acquired images was first enhanced using a post-processing compensation algorithm. The enhanced images were then analysed using Photoshop. Two graders independently graded the images to assess inter-grader reliability. One grader re-graded the images after 2 weeks to determine intra-grader reliability. Statistical analysis was performed using intra-class correlation coefficient (ICC) and Bland-Altman plot analyses.

Results

Using adaptive compensation both the intra-grader reliability (ICC: 0.95 to 0.97) and inter-grader reliability (ICC: 0.93 to 0.97) were perfect for all five locations of CT. However, with the conventional technique of manual CT measurements using built-in callipers provided with the Heidelberg explorer software, the intra- (ICC: 0.87 to 0.94) and inter-grader reliability (ICC: 0.90 to 0.93) for all the measured locations is lower. Using adaptive compensation, the mean differences (95% limits of agreement) for intra- and inter-grader sub-foveal CT measurements were −1.3 (−3.33 to 30.8) µm and −1.2 (−36.6 to 34.2) µm, respectively.

Conclusions

The measurement of CT obtained from EDI SD-OCT using our simplified method was highly reliable and efficient. Our method is an easy and practical approach to improve the quality of choroidal images and the precision of CT measurement.

Introduction

The choroid is important to support retinal and visual function as it supplies nutrients and oxygen to retinal pigment epithelial (RPE) cells and photoreceptors [1]. Therefore, the choroid may play a role in the pathophysiology of many vision threatening retinal diseases such as age-related macular degeneration [2], [3], polypoidal choroidal vasculopathy [4], [5], central serous chorioretinopathy [6], [7], Vogt-Koyanagi-Harada [8] disease and myopic macular degeneration [9][12]. To elucidate the mechanisms through which the choroid affects these retinal diseases, quantitative assessment of choroidal characteristics such as choroidal thickness (CT) is required.

However, it was not trivial to image CT because of its posterior location and pigments in the RPE layer [13], until the advent of spectral domain optical coherence tomography (SD-OCT) with enhanced depth imaging (EDI). EDI SD-OCT has provided many new insights into choroidal qualitative morphology. However, to date a notable disparity exists between the CT measurements obtained in different studies. Such variations could be due to unavailability of a standardized and simple to use measurement method and absence of built-in automated software to measure CT in most of the commercially available OCT machines [14]. At present, most of the studies perform the measurements manually by using the in-built caliper system provided by the machine, which is prone to measurement errors. The manual method is also time- and effort-consuming, making it unfeasible especially when dealing with large population data.

As a result, while many studies have published the distribution of CT in patients with retinal diseases and normal controls [15][19], none of these studies have clearly described the method used in detailed to measure CT. Therefore, the reliability of the CT measurement methods in most of the papers is unknown. In this paper, we describe a simple, semi-automated, time-efficient method to measure CT using images acquired by EDI SD-OCT and available compensation algorithms. We aimed to assess the reliability of this new CT measurement technique in a sample of healthy eyes.

Materials and Methods

Study Subjects and Design

Data for this analysis were derived from the Singapore Malay Eye Study (SiMES), a population-based cross-sectional study of eye diseases in Malay adults, age ranged from 40–80 years living in Singapore. Details of the study design, sampling plan, and methodology have been reported elsewhere [20]. In brief, participants recruited in the current study underwent standardized and detailed ophthalmic examination, including Spectralis EDI SD-OCT imaging (see next section). The study was approved by the Institutional Review Board of Singapore Eye Research Institute. It followed the tenants of the Declaration of Helsinki and written informed consent was obtained from the subjects after explanation of the nature and possible consequences of the study.

Spectralis images of 31 subjects were randomly selected using a random number generated in Stata (College par, Texas, USA). Choroidal images were only selected from right eye of each subject as our measurement technique can be applied uniformly between right and left eyes. Therefore only the right eyes of the subjects were evaluated. In addition, selecting only one eye from individuals would help to avoid inter-eye correlation issue in statistical analysis. “Normal fundus” was defined as free of any macular or vitroretinal diseases on the basis of clinical fundus examination by experienced Ophthalmologist and the results of OCT imaging. Exclusion criteria for the normal participants included: best corrected LogMAR VA >0.3, evidence of macular or vitroretinal diseases, previous retinal or refractive surgery, past history of intraocular surgery, or clinical features compatible with a diagnosis of glaucoma suspect or glaucoma.

To evaluate the intra-grader and inter-grader reliability, 31 Spectralis images were randomly selected for the initial grading phase. Grader A and grader B, masked to subject characteristics and clinical diagnosis, independently graded these images to assess inter-grader reliability. In addition, grader A repeated the measurements after 2 weeks to assess intra-grader reliability. Both graders assessed the same sets of training images before commencing the grading task.

EDI SD-OCT Imaging

CT was obtained using Spectralis SD-OCT with EDI modality (Wavelength: 870 nm; Heidelberg Engineering, Heidelberg, Germany) after pupil dilation using tropicamide 1% and phenylephrine hydrochloride 2.5%. Subjects’ keratometry readings and the refraction data were entered into the software program to estimate optical magnification and, therefore, to allow for more accurate comparisons across individuals. A single experienced examiner masked to the clinical diagnosis of the subject performed the EDI-OCT examination. Seven sections, each comprising 100 averaged scans (using the automatic averaging and eye tracking features of the proprietary device), were obtained in an angle of 5°–30° rectangle centered at the fovea. The horizontal section passing through the center of the fovea was selected for analysis. For each subject only the right eye was chosen for subsequent analysis.

Measurement of Choroidal Thickness

The two major steps in our CT semi-automatic measurement protocol are: (A) post processing of images by adaptive compensation technique and (B) quantitative measurement of CT using Photoshop software. On average, measurement of CT at multiple locations (1.5 and 3 mm nasal and temporal to the fovea) requires approximately 1 minute per image.

(A) Adaptive compensation technique

An accurate evaluation of the CT with EDI-OCT mainly relies on how well one can delineate the choroid-scleral interface (CSI), which anatomically represents the junction between the choroid and the sclera, and is a principal landmark for quantitative measurements of choroid. However, at present identification of CSI is highly variable as there is no algorithm available in Spectralis for its automatic detection. Therefore, in order to accurately determine CT, once the EDI-OCT image was obtained, the CSI was enhanced using a novel post-processing compensation algorithm [21] which greatly improved the detection of CSI by correcting the deleterious effects of light attenuation. In brief, this novel adaptive compensation algorithm improved the ability to detect and visualize the CSI. First, it removes noise over-amplification at high depth and shadow artifacts casted by blood vessels (thus decreasing the intra-layer contrast of the choroid). Second, it improves the visibility of posterior choroid boundary, by significantly increasing the inter-layer contrast across the CSI ( Figure 1 ).

Figure 1. Illustration of the raw versus the compensated image.

Figure 1

(A) Raw OCT image of a healthy subject. The choroid-scleral interface is only partially visible. Note the presence of blood vessels shadows as indicated by red asterisks. (B) Adaptive compensation was applied to the raw image in order to remove blood vessels shadows, enhance contrast and improve visibility of the choroid-scleral interface (more uniform).

(B) Measurement of CT using photoshop software

The enhanced images were measured in Photoshop CS6 extended (Adobe Systems Incorporated, San Jose, California). CT at sub-foveal, 1.5 mm and 3 mm nasal and temporal from the fovea were measured. Figure 2 shows a flow diagram summarizing the CT measurement protocol. The detailed explanation of the steps involved in quantification of CT measurement using Photoshop is provided in Appendix S1.

Figure 2. Flow diagram of the measurement protocol.

Figure 2

Statistical Analysis

Statistical analysis was performed using MedCalc Version 12.6 (MedCalc Software, Ostend, Belgium) and SPSS Version 20.0 (SPSS, Inc., Chicago, IL, USA). Shapiro-Wilk tests were used to check the normality of CT data at various locations. The reliability of the CT measurement was assessed using intra- and inter-grader agreements between two independent graders, measured by the absolute agreement model of the intra-class correlation coefficient (ICC) [22]. ICC value of 0.81–1.00 indicates almost perfect agreement. Values of less than 0.40 indicate poor to fair agreement. Bland Altman plot analyses [23], [24] were performed to see if there is any proportional bias between measurements. A two-tailed paired sample t-test was used to analyze differences between means in CT by location.

Results

Images from 31 eyes of 31 participants (aged 64.4±7.4 years) were included in the analysis ( Table 1 ). The mean (standard deviation) sub-foveal CT in this study was 230.37 (66.66) µm (average of 1st and 2nd measurements of grader A). CT measured by the two graders is summarized in Table 2 . The mean sub-foveal CT measured by grader A for the 1st and 2nd measurements was 229.74 (65.12) µm and 231 (69.14) µm, respectively, and was 232.19 (67.89) µm by grader B.

Table 1. Baseline characteristics of study subjects.

Characteristics Mean (SD)
Age, years 64.35 (7.42)
Gender, male 11 (35.48)
Axial length, mm 23.33 (0.88)
Average choroidal thickness,* µm
Sub-foveal 230.37 (66.66)
Nasal, 1.5 mm 214.33 (74.34)
Nasal, 3 mm 185.77 (69.96)
Temporal, 1.5 mm 224.53 (57.27)
Temporal, 3 mm 229.74 (52.03)

Data are expressed as mean (SD) except for gender, which is expressed as number (%).

*The average of 1st and 2nd measurement of Grader A.

Table 2. Summary of choroidal thickness measurements at various locations.

Location Grader A, 1st Measurement Grader A, 2nd Measurement Grader B Measurement
Sub-foveal 229.74 (65.12) 231.00(69.14) 232.19 (67.89)
Nasal, 1.5 mm 213.84 (75.12) 214.84(75.49) 221.52(76.08)
Nasal, 3 mm 191.13 (72.79) 180.42(68.25) 187.55(71.06)
Temporal, 1.5 mm 223.94 (56.33) 225.13(59.24) 231.65 (60.82)
Temporal, 3 mm 228.68 (50.64) 230.81(54.57) 230.71 (53.80)

Data are mean (SD).

Using adaptive compensation both the intra-grader reliability (ICC: 0.95 to 0.97) and inter-grader reliability (ICC: 0.93 to 0.97) were perfect for all five locations of CT ( Table 3 ). However, with the conventional technique of manual CT measurements using built-in callipers provided with the Heidelberg explorer software, the intra- (ICC: 0.87 to 0.94) and inter-grader reliability (ICC: 0.90 to 0.93) for all the measured locations is lower ( Table 4 ). Using adaptive compensation, the Bland Altman analysis of intra-grader reliability for sub-foveal CT measurement showed 95% LOA of −33.3 to 30.8 with a mean difference of −1.3 µm ( Figure 3 ). No significant systemic (except at nasal 3 mm, p = 0.003) and proportional bias was detected in intra-grader CT measurements at all locations. The Bland Altman analysis of inter-grader reliability for sub-foveal CT measurement showed 95% LOA of −36.6 to 34.2 with a mean difference of −1.2 µm ( Figure 4 ). No significant proportional bias was observed in the inter-grader CT measurements at all locations. Nonetheless, a significant systemic bias at both 1.5 and 3 mm nasal locations was found in the inter-grader CT measurement comparison (p = 0.001). However, this could be due to thinnest CT at nasal locations, making it more prone to systemic bias in CT measurements.

Table 3. Intra- and inter-grader agreements for the choroidal thickness measurement at 5 horizontal locations using adaptive compensation.

Assessment of proportional biasd
Locations, CT measurement ICC (95% CI) Mean difference(95% LOA) P Valuec Pearson’s correlation coefficient, r P Value
Intra-graderReliabilitya Sub-foveal 0.97 (0.94 to 0.98) −1.3 (−33.3 to 30.8) 0.672 0.248 0.179
Nasal, 1.5 mm 0.95 (0.90 to 0.97) −1.0 (−48.0 to 46.0) 0.818 −0.160 0.933
Nasal, 3 mm 0.95 (0.87 to 0.98) 10.7 (−25.3 to 46.7) 0.003 0.249 0.176
Temporal, 1.5 mm 0.96 (0.92 to 0.98) −1.2 (−31.9 to 29.5) 0.674 −0.187 0.313
Temporal, 3 mm 0.95 (0.90 to 0.97) −2.1 (−33.5 to 29.2) 0.464 −0.248 0.178
Inter-graderReliabilityb Sub-foveal 0.96 (0.93 to 0.98) −1.2 (−36.6 to 34.2) 0.715 0.070 0.708
Nasal, 1.5 mm 0.97 (0.92 to 0.99) −9.0 (−34.4 to 16.4) 0.001 −0.127 0.495
Nasal, 3 mm 0.97 (0.89 to 0.98) −9.8 (−37.4 to 17.8) 0.001 −0.230 0.212
Temporal, 1.5 mm 0.93 (0.86 to 0.96) −6.5 (−48.3 to 35.3) 0.099 −0.075 0.687
Temporal, 3 mm 0.94 (0.88 to 0.97) 0.1 (−35.5 to 35.7) 0.977 0.043 0.818

LOA, Limits of Agreement; ICC, Intraclass Correlation Coefficient; CI, Confidence Interval.

a

Mean difference was determined from the1st time measurement minus the 2nd time measurement.

b

Mean difference was determined from Grader A measurement minus Grader B measurement.

c

P value of one sample t-tests (comparing between mean difference and zero value) to indicate presence of systemic bias.

d

Using Pearson’s correlation coefficients of regression line.

Table 4. Intra- and inter-grader agreements for the choroidal thickness measurement at 5 horizontal locations using Spectralis SD-OCT with conventional manual technique.

Locations, CT measurement ICC (95% CI) Mean difference (95% LOA)
Intra-grader Reliabilitya Sub-foveal 0.94 (0.87to 0.97) 9.7 (−30.3 to 49.7)
Nasal, 1.5 mm 0.92 (0.85 to 0.96) 2.9 (−41.1 to 54.9)
Nasal, 3 mm 0.92 (0.85 to 0.96) 4.1 (−36.1 to 44.3)
Temporal, 1.5 mm 0.87 (0.75 to 0.93) −7.0 (−52.1 to 38.1)
Temporal, 3 mm 0.88 (0.77 to 0.94) −1.9 (−44.9 to 48.6)
Inter-grader Reliabilityb Sub-foveal 0.90 (0.60 to 0.96) −19.1 (−63.7 to 25.6)
Nasal, 1.5 mm 0.93 (0.68 to 0.97) 18.3 (−59.7 to 23.2)
Nasal, 3 mm 0.91 (0.67 to 0.96) −15 (−54.3 to 24.3)
Temporal, 1.5 mm 0.91 (0.72 to 0.96) −12.7 (−49.8 to 24.4)
Temporal, 3 mm 0.90 (0.58 to 0.96) −15.7 (−51.3 to 19.9)

LOA, Limits of Agreement; ICC, Intraclass Correlation Coefficient; CI, Confidence Interval.

a

Mean difference was determined from the1st time measurement minus the 2nd time measurement.

b

Mean difference was determined from Grader A measurement minus Grader B measurement.

Figure 3. Bland Altman plot of intra-grader reliability of sub-foveal choroidal thickness (CT) measurement.

Figure 3

The difference was calculated by the 1st measurement minus the 2nd measurement. Pink dashed line represents regression line of difference between 1st and 2nd measurements.

Figure 4. Bland Altman plots of inter-grader reliability of sub-foveal choroidal thickness (CT) measurement.

Figure 4

The difference was calculated by the grader A measurement minus the grader B measurement. Pink dashed line represents regression line of difference between the two graders measurements.

In addition, the choroid intra-layer contrast (a measure of shadow presence when high) and the CSI inter-layer contrast (a measure of boundary visibility when high) were computed for all images (n = 31) before and after applying adaptive compensation (as in [25]). We found that the intra-layer contrast significantly decreased from 0.84±0.07 to 0.60±0.07 (p<0.001; t-test), whereas the inter-layer contrast significantly increased from 0.50±0.14 to 0.90±0.10 (p<0.001; paired t-test) after applying adaptive compensation ( Table 5 ).

Table 5. Intra- and inter-layer contrast (n = 31 subjects).

Mean (SD) P-Value
Intra-layer Contrast Standard (without compensation) 0.84 (0.07) <0.001
Adaptive Compensation 0.60 (0.07)
Inter-layer Contrast Standard (without compensation) 0.50 (0.14) <0.001
Adaptive Compensation 0.90 (0.10)

Discussion

Despite significant advances in imaging technology there are considerable variations in CT measurements across clinical studies. In this report, we described a simple, semi-automated method using adaptive compensation to measure CT from images acquired by EDI SD-OCT. The results showed that the method permits a highly reproducible tool for assessing CT from EDI SD-OCT images. Our measurement method is also simple and requires little time to perform (on average ∼1 minute per image). With these features, our method may have great potential for use in both clinical and population-based studies which involve large number of images.

There are considerable differences in CT measurements across studies. Even though advances in OCT technology have reduced acquisition time, adequate visualization of choroid is still lacking. There are several challenges in imaging the choroid. First, the choroid is located behind the RPE and Bruch’s membrane, making it less accessible than the retina and more difficult to be visualized. Second, the pigmentation in the RPE and the choroid itself decreases the signal intensity. Lastly, unlike retinal imaging [26], wherein the RPE and internal limiting membrane are thin and clearly identifiable, the transition zone between the choroid and sclera (i.e. CSI) has blurred border and is broader, making CSI difficult to identify.

However, it is not always easy to distinguish CSI on acquired images, which may be related to scan quality or anatomic variation. Some choroidal scans have a distinct hypo-reflective line corresponding to the supra-choroidal space, but often this line can be indistinct leading to measurement error [27]. At present, identification of CSI is highly variable. To overcome these challenges, in this study we enhanced choroidal-scleral junction using a post-processing adaptive compensation algorithm [21]. Measurement of CT was performed in the post-processed images using Photoshop (Adobe Systems Incorporated, San Jose, California), a readily available and easy-to-use program.

Adaptive compensation provides two major improvements. First, the intra-layer contrast (of the choroid) significantly decreased after applying adaptive compensation, indicating successful shadow correction within the choroid. Second, the inter-layer contrast (across the CSI) significantly increased after applying adaptive compensation, indicating better visualization of CSI and therefore allowing for precise measurement of CT. These results are consistent with a previous study on standard compensation [25] and indicate significant improvements in image quality. Also, a study by Lin et al. established outer choroidal contrast as a valid quantifiable measure of choroidal image quality and demonstrated that inverted SD-OCT imaging optimizes visualization of CSI and choroidal vessels through improved outer choroid contrast [28].

With improved intra- and inter-layer contrast, the compensated images are likely more accurate than the non-compensated images as the deleterious effects of OCT light attenuation have been corrected, thus making the CSI more visible. The post-compensation images are more representative of the eye tissue architecture, since better estimates of the ocular tissues optical properties (e.g. reflectivity) are provided. This has been formally demonstrated in the original article by Girard et al. [25], and in the following studies [21], [29], [30]. Lastly, although adaptive compensation is useful in all images, it is likely to be more useful when the CSI is poorly visible, e.g. in images where light attenuation is strong ( Figure 1 ), emphasizing further the need for compensation for more accurate thickness measurement of the choroid.

Adaptive compensation achieves high intra-grader (ICC: 0.95 to 0.97) and inter-grader (ICC: 0.93 to 0.97) repeatability in CT measurements compared to conventional method (intra-grader ICC: 0.87 to 0.94, inter-grader ICC: 0.90 to 0.93), suggesting the use of adaptive compensation to improve the visualization of CSI and to obtain more reliable CT measurements. However, the high repeatability can be explained by the standardization and strict adherence to a rigorous grading protocol in the present study. In addition, both graders underwent the same training set for standardization purpose before embarking on the actual grading task. The present results are in line in the studies by Ikuno et al. [31] and Yamashita et al. [32] who reported an inter-grader ICC of 0.97 and 0.94, respectively, for sub-foveal CT measurements. However, in these two previous studies the intra-grader repeatability was not evaluated, and it is not clear how the choroid images, where CSI could not be clearly visualized, were processed.

But the present study has its own limitations, as the measurement of CT was subjective in nature, and was therefore subject to measurement bias. However, our method may be less prone to observer error because the CSI was enhanced by adaptive compensation to give better visibility. An automated, and thus may be more objective, method of measuring CT would be of potential interest to facilitate, speed-up and render operator-independent such analyses.

There are few recent studies on automatic choroidal segmentation in OCT images. Zhang et al. proposed an automatic segmentation algorithm for the choroidal vessels using Cirrus OCT in 24 normal subjects. But their aim was to quantify choroidal vasculature thickness and choroicapillaries equivalent thickness rather than the CT [33]. Likewise, Torzicky et al. [34] and Duan et al. [35] developed the automatic algorithms to detect the boundary between the choroid and sclera based on polarization sensitive OCT which are not commercially available. Although, Tian et al. [36] in 2013 proposed an automatic algorithm that could segment choroid in commercially available Spectralis OCT, but their algorithm was tested only on 20 EDI OCT images and need to be validated on more images to prove the robustness of algorithm before its application in clinical studies. While we are preparing the manuscript, a commercially available CT measurement algorithm in swept-source OCT (Topcon Corp., Tokyo, Japan) is being made available. However, in view of lack of supporting agreement studies on the algorithm and manual choroidal segmentation (in both normal and diseased eyes), the accuracy and reliability of automated CT analysis by using swept-source OCT is yet to be established for use in clinical settings.

In conclusion, we described a simplified, semi-automated and practical method “adaptive compensation followed by using Photoshop” that gives excellent intra- and inter-grader reliability (ICC>0.93) to quantify CT in the EDI SD-OCT images. This method has great potential usage in both clinical and population based studies, as reliable and accurate measurements of CT from EDI-OCT images are essential in distinguishing clinically significant change of CT and assisting in risk-profiling for various posterior segment diseases.

Supporting Information

Appendix S1

Detailed steps to quantify choroidal thickness using Photoshop.

(DOCX)

Funding Statement

This study was supported by a grant (R760/44/2010) from National Medical Research Council, Singapore. Image post-processing techniques were supported by the Ministry of Education, Academic Research Funds, Tier 1, Singapore (MJAG). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1. Linsenmeier RA, Padnick-Silver L (2000) Metabolic dependence of photoreceptors on the choroid in the normal and detached retina. Invest Ophthalmol Vis Sci 41: 3117–3123. [PubMed] [Google Scholar]
  • 2. Grossniklaus HE, Green WR (2004) Choroidal neovascularization. Am J Ophthalmol 137: 496–503. [DOI] [PubMed] [Google Scholar]
  • 3. Koizumi H, Yamagishi T, Yamazaki T, Kawasaki R, Kinoshita S (2011) Subfoveal choroidal thickness in typical age-related macular degeneration and polypoidal choroidal vasculopathy. Graefes Arch Clin Exp Ophthalmol 249: 1123–1128. [DOI] [PubMed] [Google Scholar]
  • 4. Gomi F, Tano Y (2008) Polypoidal choroidal vasculopathy and treatments. Curr Opin Ophthalmol 19: 208–212. [DOI] [PubMed] [Google Scholar]
  • 5. Chung SE, Kang SW, Lee JH, Kim YT (2011) Choroidal thickness in polypoidal choroidal vasculopathy and exudative age-related macular degeneration. Ophthalmology 118: 840–845. [DOI] [PubMed] [Google Scholar]
  • 6. Gupta B, Mohamed MD (2011) Photodynamic therapy for variant central serous chorioretinopathy: efficacy and side effects. Ophthalmologica 225: 207–210. [DOI] [PubMed] [Google Scholar]
  • 7. Imamura Y, Fujiwara T, Margolis R, Spaide RF (2009) Enhanced depth imaging optical coherence tomography of the choroid in central serous chorioretinopathy. Retina 29: 1469–1473. [DOI] [PubMed] [Google Scholar]
  • 8. Fong AH, Li KK, Wong D (2011) Choroidal evaluation using enhanced depth imaging spectral-domain optical coherence tomography in Vogt-Koyanagi-Harada disease. Retina 31: 502–509. [DOI] [PubMed] [Google Scholar]
  • 9. Fujiwara T, Imamura Y, Margolis R, Slakter JS, Spaide RF (2009) Enhanced depth imaging optical coherence tomography of the choroid in highly myopic eyes. Am J Ophthalmol 148: 445–450. [DOI] [PubMed] [Google Scholar]
  • 10. Ikuno Y, Tano Y (2009) Retinal and choroidal biometry in highly myopic eyes with spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci 50: 3876–3880. [DOI] [PubMed] [Google Scholar]
  • 11. Fitzgerald ME, Wildsoet CF, Reiner A (2002) Temporal relationship of choroidal blood flow and thickness changes during recovery from form deprivation myopia in chicks. Exp Eye Res 74: 561–570. [DOI] [PubMed] [Google Scholar]
  • 12.Cheung CM, Loh BK, Li X, Mathur R, Wong E, et al. (2013) Choroidal thickness and risk characteristics of eyes with myopic choroidal neovascularization. Acta Ophthalmol. [DOI] [PubMed]
  • 13. Povazay B, Hermann B, Unterhuber A, Hofer B, Sattmann H, et al. (2007) Three-dimensional optical coherence tomography at 1050 nm versus 800 nm in retinal pathologies: enhanced performance and choroidal penetration in cataract patients. J Biomed Opt 12: 041211. [DOI] [PubMed] [Google Scholar]
  • 14.Kahn KN, McKibbin M, Kahn RS (2011) Variability in subfoveal choroidal thickness measurements. Invest Ophthalmol Vis Sci 52: 7221; author reply 7221. [DOI] [PubMed]
  • 15. Spaide RF, Koizumi H, Pozzoni MC (2008) Enhanced depth imaging spectral-domain optical coherence tomography. Am J Ophthalmol 146: 496–500. [DOI] [PubMed] [Google Scholar]
  • 16.Manjunath V, Taha M, Fujimoto JG, Duker JS (2010) Choroidal thickness in normal eyes measured using Cirrus HD optical coherence tomography. Am J Ophthalmol 150: 325–329 e321. [DOI] [PMC free article] [PubMed]
  • 17. Maul EA, Friedman DS, Chang DS, Boland MV, Ramulu PY, et al. (2011) Choroidal thickness measured by spectral domain optical coherence tomography: factors affecting thickness in glaucoma patients. Ophthalmology 118: 1571–1579. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Ho J, Branchini L, Regatieri C, Krishnan C, Fujimoto JG, et al. (2011) Analysis of normal peripapillary choroidal thickness via spectral domain optical coherence tomography. Ophthalmology 118: 2001–2007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Shin JW, Shin YU, Lee BR (2012) Choroidal thickness and volume mapping by a six radial scan protocol on spectral-domain optical coherence tomography. Ophthalmology 119: 1017–1023. [DOI] [PubMed] [Google Scholar]
  • 20. Rosman M, Zheng Y, Wong W, Lamoureux E, Saw SM, et al. (2012) Singapore Malay Eye Study: rationale and methodology of 6-year follow-up study (SiMES-2). Clin Experiment Ophthalmol 40: 557–568. [DOI] [PubMed] [Google Scholar]
  • 21. Mari JM, Strouthidis NG, Park SC, Girard MJ (2013) Enhancement of lamina cribrosa visibility in optical coherence tomography images using adaptive compensation. Invest Ophthalmol Vis Sci 54: 2238–2247. [DOI] [PubMed] [Google Scholar]
  • 22. Fleiss JL, Cohen J (1973) The Equivalence of Weighted Kappa and the Intraclass Correlation Coefficient as Measures of Reliability. Educational and Psychological Measurement 33: 613–619. [Google Scholar]
  • 23. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1: 307–310. [PubMed] [Google Scholar]
  • 24. Bland JM, Altman DG (2012) Agreed statistics: measurement method comparison. Anesthesiology 116: 182–185. [DOI] [PubMed] [Google Scholar]
  • 25. Girard MJ, Strouthidis NG, Ethier CR, Mari JM (2011) Shadow removal and contrast enhancement in optical coherence tomography images of the human optic nerve head. Invest Ophthalmol Vis Sci 52: 7738–7748. [DOI] [PubMed] [Google Scholar]
  • 26. Hee MR, Puliafito CA, Duker JS, Reichel E, Coker JG, et al. (1998) Topography of diabetic macular edema with optical coherence tomography. Ophthalmology 105: 360–370. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Yiu G, Pecen P, Sarin N, Chiu SJ, Farsiu S, et al. (2014) Characterization of the choroid-scleral junction and suprachoroidal layer in healthy individuals on enhanced-depth imaging optical coherence tomography. JAMA Ophthalmol 132: 174–181. [DOI] [PubMed] [Google Scholar]
  • 28. Lin P, Mettu PS, Pomerleau DL, Chiu SJ, Maldonado R, et al. (2012) Image inversion spectral-domain optical coherence tomography optimizes choroidal thickness and detail through improved contrast. Invest Ophthalmol Vis Sci 53: 1874–1882. [DOI] [PubMed] [Google Scholar]
  • 29. Foin N, Mari JM, Davies JE, Di Mario C, Girard MJ (2013) Imaging of coronary artery plaques using contrast-enhanced optical coherence tomography. Eur Heart J Cardiovasc Imaging 14: 85. [DOI] [PubMed] [Google Scholar]
  • 30. Foin N, Mari JM, Nijjer S, Sen S, Petraco R, et al. (2013) Intracoronary imaging using attenuation-compensated optical coherence tomography allows better visualisation of coronary artery diseases. Cardiovasc Revasc Med 14: 139–143. [DOI] [PubMed] [Google Scholar]
  • 31. Ikuno Y, Maruko I, Yasuno Y, Miura M, Sekiryu T, et al. (2011) Reproducibility of retinal and choroidal thickness measurements in enhanced depth imaging and high-penetration optical coherence tomography. Invest Ophthalmol Vis Sci 52: 5536–5540. [DOI] [PubMed] [Google Scholar]
  • 32. Yamashita T, Yamashita T, Shirasawa M, Arimura N, Terasaki H, et al. (2012) Repeatability and reproducibility of subfoveal choroidal thickness in normal eyes of Japanese using different SD-OCT devices. Invest Ophthalmol Vis Sci 53: 1102–1107. [DOI] [PubMed] [Google Scholar]
  • 33. Zhang L, Lee K, Niemeijer M, Mullins RF, Sonka M, et al. (2012) Automated segmentation of the choroid from clinical SD-OCT. Invest Ophthalmol Vis Sci 53: 7510–7519. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Torzicky T, Pircher M, Zotter S, Bonesi M, Gotzinger E, et al. (2012) Automated measurement of choroidal thickness in the human eye by polarization sensitive optical coherence tomography. Opt Express 20: 7564–7574. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Duan L, Yamanari M, Yasuno Y (2012) Automated phase retardation oriented segmentation of chorio-scleral interface by polarization sensitive optical coherence tomography. Opt Express 20: 3353–3366. [DOI] [PubMed] [Google Scholar]
  • 36. Tian J, Marziliano P, Baskaran M, Tun TA, Aung T (2012) Automatic measurements of choroidal thickness in EDI-OCT images. Conf Proc IEEE Eng Med Biol Soc 2012: 5360–5363. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Appendix S1

Detailed steps to quantify choroidal thickness using Photoshop.

(DOCX)


Articles from PLoS ONE are provided here courtesy of PLOS