TABLE 2.

The summary of the limiters’ performances.

Performances Diode limiter Commercial limiter BPT-based limiter
IL (dB) 20 MHz –6.0 –6.6 –6.3
60 MHz –6.8 –7.7 –7.4
100 MHz –8.0 –8.0 –7.7
THD (dB) 20 MHz –72.4 –74.1 –77.3
60 MHz –59.9 –59.9 –64.9
100 MHz –69.5 –66.3 –74.6
TNF (dB) 20 MHz 11 10.4 9.6
60 MHz 10.7 10.3 8.2
100 MHz 10.4 10.7 6.6
Bandwidth (MHz) 131 147 135
RT (ns) 89 82 43

* The IL, THD, and TNF values were measured at 50 mVp-p. RT is the time caused by negative pulse.