Abstract
Purpose
The Sophisticated Harmonic Artifact Reduction for Phase data (SHARP) method has been proposed for the removal of background field in MRI phase data. It relies on the spherical mean value (SMV) property of harmonic functions, and its accuracy depends on the radius of the sphere used for computing the SMV and truncation threshold needed for deconvolution. The goal of this work is to develop an alternative SMV based background field removal method with reduced dependences on these parameters.
Methods
The proposed background field removal method (termed iterative SMV or iSMV) consists of applying the SMV operation repeatedly on the field map and it was validated in a phantom and in vivo brain data of five healthy volunteers.
Results
The iSMV method demonstrates accurate background field removal in the phantom. Compared to SHARP, the iSMV method shows a significantly reduced dependence on the SMV radius both in phantom and in human data. Because a smaller radius can be chosen, the iSMV method allows retaining a larger part of the region of interest, compared to SHARP.
Conclusion
The iSMV method is an effective background field removal method with a reduced dependence on method parameters.
Keywords: MRI, harmonic functions, spherical mean value, background field removal
INTRODUCTION
There has been increasing interest in the use of MRI signal phase and quantitative susceptibility mapping (QSM) (1-13) for the study of tissue magnetic properties (11,14-24). The MRI signal phase reflects the total magnetic field experienced by water proton spins, which is the superposition of the local magnetic field, generated by susceptibility sources within the chosen region of interest (ROI), and the background field, generated by susceptibility sources outside the ROI. A precise estimation of the local field map is essential for obtaining an accurate map of tissue magnetic susceptibility. Therefore, the background field must be removed through a separate background removal procedure. Three commonly used methods are the High-Pass Filter (HPF) method (25), the Projection onto Dipole Fields (PDF) method (26), and the Sophisticated Harmonic Artifact Reduction for Phase Data (SHARP) method (18).
The background field corresponds to the magnetic field measured on a uniform phantom whose shape is identical to the ROI. Consequently, it is a harmonic function, i.e., a solution to the Laplace equation (9,27,28). The HPF method exploits the fact that such a background field tends to be very smooth by removing the low spatial frequency components. It is effective and has the advantage of being very straightforward to implement but, since the Fourier basis functions are not solutions to the Laplace equation, the HPF method may give rise to noticeable artifacts (26). Additionally, the size of the low-pass filter must be chosen carefully, since it affects the final local field. The PDF method fits the total field using a set of dipole functions centered at locations outside the ROI, such that the fitted field is, by construction, a harmonic function. It is more computationally expensive and accuracy of the local field suffers near the boundary of the ROI, because the assumed orthogonality between the dipole field originating inside and those outside the ROI is only approximate. The SHARP method relies on the fact that a harmonic field satisfies the following property. For any given sphere at any given location within the ROI, the mean value of a harmonic function over that sphere is equal to the function value at the center of that sphere (27). It is referred to in the following as the Spherical Mean Value (SMV) property. The SHARP method is straightforward to implement and computationally efficient. However, in practice, the local field obtained with the SHARP method depends on the radius used in the SMV operation and on the threshold needed for the subsequent deconvolution (see below). Additionally, since the SMV cannot be computed for voxels that are less than a radius away from the ROI border, the local field is only available on a modified ROI that is smaller than the original ROI.
Here, we propose an alternative SMV based background removal method that overcomes these shortcomings by repeatedly applying the SMV operation on the field map. The proposed method, named iterative Spherical Mean Value (iSMV), is validated in both phantom experiment and in vivo.
THEORY
For a given ROI denoted by R, let F0 be the total field map, which can be estimated from MRI phase data after proper spatial unwrapping. The total field can be separated into two components: F0 = Fnh + Fh, with Fnh the non-harmonic component or the local field, and Fh the harmonic component or background field. Both the SHARP method and the proposed iSMV method utilize the SMV property of harmonic functions. In this section, we include the definition of the SMV operation, a short review of the SHARP method, and the theory of the proposed iSMV method.
The SMV operation
We use B(x0, r) to denote a sphere with radius r centered at a location x0 within region R. For a given function f defined on R and a given ragius r, the SMVr at a point x0 is given by
(1) |
where V is the volume of the ball B(x0, r). The SMVr operation is only defined for locations x0 for which the ball B(x0, r) lies completely within R. If f is harmonic in R, then, according to the SMV property, f(x0) = (SMVrf)(x0).
If the SMV operation is applied to every point x in R, we get the function (SMVrf)(x). This SMV operation is simply an average filter and can be efficiently implemented using Fourier Transform. Note that for points near the boundary of R, the SMVr operation cannot be faithfully calculated since it would require data outside R, which we assume is not available. This is, for instance, the case when the ROI is chosen such that only air is present outside the ROI, where MRI phase measurements are either not possible or unreliable. We thus distinguish two non-overlapping regions within R: 1) Rr, defined as the region where the SMVr operation does not require data outside R, and 2) R′, defined as the region whose voxels are less than a distance r away from the boundary of R. Note that Rr becomes smaller as r increases.
The SHARP method
The SHARP method calculates the local field using a three-step process, consisting of a convolution, a point-to-point multiplication with a mask Mr, and a deconvolution, i.e.,
(2) |
Here, the convolution of the total field F0 with the (1 – SMVr) kernel removes the harmonic background field, the point-to-point multiplication with Mr sets the value of all voxels outside of Rr to zero, and the deconvolution (1 – SMVr)–1 is computed in Fourier space using a point-to-point division with the Fourier Transform of the (1 – SMVr) kernel. This kernel vanishes at k = 0, such that truncation is needed before applying the division to avoid noise amplification (18,28). In principle, the effect of an average filter cannot be reverted unless careful treatment of the boundary values of Fnh is carried out. This thresholded k-space division leads to loss of information and accuracy. Additionally, results are available only on the smaller ROI Rr, where the SMVr can be faithfully calculated. Overall, the local field generated by the SHARP method depends on the chosen radius r and on the chosen truncation threshold for the deconvolution process (28).
The iSMV method
For functions f defined on the ROI R, we define the operator Sr as follows
(3) |
Let us now fix the radius r and make the assumption that the background field Fh is known on the border R′. Define a modified field map F̃0 that is equal to Fh on the border R′ and equal to F0 on the inner (cropped) region Rr. Define the function as the result of applying the operator Sr an infinite number of times on F̃0. The function satisfies the following properties:
It is equal to Fh on the border R′ by definition, and
- For all , such that satisfies the SMV property for radius r everywhere inside Rr. Therefore, must be a harmonic function itself (29). Since the solution to the Dirichlet problem of finding a harmonic function inside a given region that satisfies a given boundary condition is unique for a closed ROI (30), it is found that . The local field on Rr can then be calculated as
The proposed method depends now only on the choice of the radius r.(4)
METHODS
SHARP Implementation
SHARP was reported to achieve the best result with a radius equal to 6 to 7 voxels (28). In this paper, we chose to compute iSMV and SHARP with a 1 voxel radius and a 6 voxels radius to investigate the radius dependences of both methods. For SHARP, we used the recommended values for the truncation threshold: 0.15 for the 1 voxel radius and 0.0225 for the 6 voxels radius (28).
iSMV Implementation
In order to compute the local field using the iSMV method, knowledge of the background field at the boundary of the ROI is required. In practice, the local field is often much smaller than the background field. Therefore, in this study, it was assumed that the local field was zero at the boundary and, consequently, the background field was equal to the total (measured) field at the boundary. If other prior information about the boundary values is known, it can be easily incorporated in the method. The operator Sr was implemented by first performing the SMVr over the entire field of view (FOV), which can be done efficiently in Fourier space, followed by the replacement of the boundary values (those within the region R′) with the corresponding boundary values of the total field F0. The repeated application of the Sr operator implemented in this fashion was halted once the relative error between the norms of two successive iterations was smaller than 5 × 10–5. The visual interpretation of the algorithm is shown in Figure 1.
Since iSMV is an iterative method, it is important to use an accurate SMVr kernel for the convolution. In the construction of the SMVr kernel, each voxel was weighted according to the volume of the intersection of that voxel with the sphere with radius r.
Phantom experiments
A water phantom was constructed by placing two vials of 0.025% and 0.016% Gadolinium solutions in a container filled with water. This phantom was scanned using a 3D multi-gradient echo sequence on a 1.5-T scanner (GE Excite HD) with a 5-inch surface coil for signal reception. The imaging parameters were TE/△TE/#TE/TR/FA/BW/FOV/matrix size = 3.1ms/ 3.1ms/ 10/52.3ms/15°/±62.50kHz/24×24×9.6 cm3/256×256×80. The scan was repeated after the two vials were removed and the equivalent volume of water was added to replace the removed volume. The second scan was used to obtain the reference background field, to which the computed background fields from the first scan were compared. The ROI was segmented by signal thresholding of the magnitude images. The background field was estimated in the data from the first scan using the SHARP and iSMV methods, using two different radii (1 voxel and 6 voxels). The accuracies of both methods were evaluated by calculating the relative error of the computed local field with respect to the reference (true) local field. The relative error was defined as the norm of the difference between the estimated field and the reference field divided by the norm of the reference field, evaluated on the cropped region of interest Rr:
(5) |
with F the field for which the error was calculated (either the local or the background field).
In vivo Human Brain
Five healthy volunteers were scanned using a 3D multi-gradient echo sequence on a 3T scanner (GE Excite HD), under approval by the Institutional Review Board. Imaging parameters were TE/△TE/#TE/TR/FA/BW/FOV/matrix size=5ms/5ms/8/50ms/20°/±62.50kHz/21.75×24×12.8 cm3/232×256×64. A visual comparison was performed between the estimated local fields obtained by SHARP and iSMV using two different radii (1 voxel and 6 voxels).
RESULTS
Figure 2 shows the results obtained in the phantom experiment. The local field estimated using SHARP with a radius of 6 voxels is noticeably different from that using a radius of 1 voxel, whereas the iSMV estimated local fields are consistent between the two radii. The relative error of the local field was substantially larger for SHARP using a 1 voxel radius, compared to SHARP using a 6 voxel radius and iSMV using a 1 or 6 voxel radius, all of which resulted in similar errors.
Figure 3 presents a comparison between the SHARP and iSMV results in a healthy subject using a 1 voxel radius and a 6 voxel radius. For the local fields estimated using the SHARP method (top row), a noticeable difference was seen between the two radii. Using SHARP with a 6 voxel radius as a reference, the relative error (Eq. 5) between the two SHARP results was 78%. The local fields estimated using iSMV method (bottom row) produced consistent results: using iSMV with a 6 voxel radius as a reference, the relative error (Eq. 5) between the two iSMV results was 9%.
Over all subjects, the relative error between the 1 and 6 voxel radius iSMV results was 11%±2%. Using the 6 voxel radius SHARP as the reference, the relative errors of the 1 voxel radius SHARP, 1 voxel radius iSMV, and 6 voxel radius iSMV were 78%±2%, 17%±2%, and 17%±3% respectively.
Computation times for the in vivo brain data were 3s for both the 1 voxel radius and the 6 voxel radius SHARP, 105s for the 1 voxel radius iSMV, and 76s for the 6 voxel radius iSMV. The algorithm was implemented with MATLAB 2009a (MathWorks, Natick, MA, USA) on an Intel® Xeon W3530 @ 2.80GHz with 8GB RAM.
Figure 4 shows a comparison between SHARP using a 6 voxel radius and iSMV using a 1 voxel radius in a second healthy subject in three orthogonal planes. A high degree of similarity between the two estimated local fields was observed, but the iSMV estimate was available on a larger ROI.
DISCUSSION
In this paper, an iterative spherical mean value (iSMV) method is described for the removal of the background field in MRI. Our phantom and human data show that the background fields estimated using iSMV are similar to those obtained with the SHARP method, but are much less dependent on the choice of radius when performing the SMV operation. In both the phantom and the in vivo human brain, significant changes in the SHARP results were observed when a different radius of the sphere was used. Generally, a larger radius increases the accuracy of the SHARP estimates but comes at the cost of a further cropping of the ROI for which the local field can be computed (denoted by Rr in the current work). Consequently, SHARP inherently has a tradeoff between accuracy of the result and loss of voxels near the boundary. With iSMV, however, an estimate of the local field can now be obtained for an ROI that is only slightly smaller than the original ROI. A limitation of this work is that the relative error of the in vivo data is computed using the SHARP method with a 6 voxels radius as reference, since the true local field is not available. This makes it difficult to assess the accuracy of the iSMV results on the human data, compared to the experiments, where the true local field is available.
The iSMV method fundamentally treats the problem of separating harmonic and non-harmonic components as a boundary value problem for estimating a background field in a closed region. In this work, it was assumed that, at the boundary, the background field was equal to the total field. In many instances, the chosen ROI will be at or close to the air-tissue interface (~9 ppm susceptibility difference w.r.t. water, ~2 ppm for bone), which induces a magnetic field that is up to an order of magnitude larger than those generated by local sources of susceptibility, such as venous blood (~ 0.3 ppm), cortex grey matter (~ 0.05 ppm), substantia nigra (~0.12 ppm), and globus pallidus (~0.2 ppm) (5,8,31). For brain imaging on healthy subjects, the background field caused by the air-tissue boundary and bone dominates the local field generated by brain tissue due to their susceptibility difference. For patients with hemorrhage close to the skull, it may happen that the local field is similar in magnitude compared to the background field. Unless additional information is available about the local field at the ROI boundary, this can lead to an erroneous estimation of the local field near the boundary.
Compared to iSMV, the SHARP method uses the same SMV property of harmonic functions, but does not specify an assumed ROI boundary condition explicitly. Additionally, the SHARP method utilizes a thresholded k-space division to undo the effect of the SMV operation on the local field, and thus introduces another parameter that needs to be chosen carefully, since it influences the final result. The HPF method exploits the fact that the sought-after harmonic fields are generally smooth and it does not use a boundary condition. Additionally, the local field estimate obtained with HPF does not need the choice of an ROI and is available over the entire FOV. The obtained background field is not necessarily harmonic and the size of the low pass filter is an additional parameter to be optimized. Both SHARP and HPF are computationally more efficient than the initial implementation of iSMV in this work, though improvements may be expected by borrowing methods used in boundary value problems. Compared to iSMV, the PDF method retains data of the entire ROI (26), but relies on the approximate orthogonality between local and background field dipoles. The estimated background field is a harmonic function but it does not use a boundary condition. Its computational burden is lower than iSMV but higher than SHARP and HPF. In this note of limited scope, we have focused on developing an alternative SMV based method with a reduced dependence of the SMV radius on the final result. A comparison of all these methods using rigorous mathematical analyses and experimental evaluations is a natural continuation of this work in future research.
CONCLUSION
We have demonstrated that the iterative spherical mean value (iSMV) method is accurate for background field removal in MRI. When compared to SHARP, the dependence of the estimated local field on the radius used for SMV is strongly reduced, and allows the estimation of the local field on a larger region of interest.
ACKNOWLEDGEMENTS
This work was supported in part by NIH grants R01NS072370, R01EB013443, R21DK090690, R21CA152275, and 1R43EB015293.
REFERENCES
- 1.Wharton S, Bowtell R. Whole-brain susceptibility mapping at high field: A comparison of multiple- and single-orientation methods. NeuroImage. 2010;53(2):515–525. doi: 10.1016/j.neuroimage.2010.06.070. [DOI] [PubMed] [Google Scholar]
- 2.Wharton S, Schäfer A, Bowtell R. Susceptibility mapping in the human brain using threshold-based k-space division. Magnetic Resonance in Medicine. 2010;63(5):1292–1304. doi: 10.1002/mrm.22334. [DOI] [PubMed] [Google Scholar]
- 3.de Rochefort L, Liu T, Kressler B, Liu J, Spincemaille P, Lebon V, Wu J, Wang Y. Quantitative susceptibility map reconstruction from MR phase data using bayesian regularization: Validation and application to brain imaging. Magnetic Resonance in Medicine. 2010;63(1):194–206. doi: 10.1002/mrm.22187. [DOI] [PubMed] [Google Scholar]
- 4.de Rochefort L, Brown R, Prince MR, Wang Y. Quantitative MR susceptibility mapping using piece-wise constant regularized inversion of the magnetic field. Magnetic Resonance in Medicine. 2008;60(4):1003–1009. doi: 10.1002/mrm.21710. [DOI] [PubMed] [Google Scholar]
- 5.Liu T, Liu J, de Rochefort L, Spincemaille P, Khalidov I, Ledoux JR, Wang Y. Morphology enabled dipole inversion (MEDI) from a single-angle acquisition: Comparison with COSMOS in human brain imaging. Magnetic Resonance in Medicine. 2011;66(3):777–783. doi: 10.1002/mrm.22816. [DOI] [PubMed] [Google Scholar]
- 6.Liu T, Spincemaille P, de Rochefort L, Kressler B, Wang Y. Calculation of susceptibility through multiple orientation sampling (COSMOS): A method for conditioning the inverse problem from measured magnetic field map to susceptibility source image in MRI. Magnetic Resonance in Medicine. 2009;61(1):196–204. doi: 10.1002/mrm.21828. [DOI] [PubMed] [Google Scholar]
- 7.Liu T, Xu W, Spincemaille P, Avestimehr AS, Wang Y. Accuracy of the Morphology Enabled Dipole Inversion (MEDI) Algorithm for Quantitative Susceptibility Mapping in MRI. Medical Imaging, IEEE Transactions on. 2012;31(3):816–824. doi: 10.1109/TMI.2011.2182523. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Liu J, Liu T, de Rochefort L, Ledoux J, Khalidov I, Chen W, Tsiouris AJ, Wisnieff C, Spincemaille P, Prince MR, Wang Y. Morphology enabled dipole inversion for quantitative susceptibility mapping using structural consistency between the magnitude image and the susceptibility map. NeuroImage. 2012;59(3):2560–2568. doi: 10.1016/j.neuroimage.2011.08.082. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Li L, Leigh JS. Quantifying arbitrary magnetic susceptibility distributions with MR. Magnetic Resonance in Medicine. 2004;51(5):1077–1082. doi: 10.1002/mrm.20054. [DOI] [PubMed] [Google Scholar]
- 10.Deistung A, Schäfer A, Schweser F, Biedermann U, Turner R, Reichenbach JR. Toward in vivo histology: A comparison of quantitative susceptibility mapping (QSM) with magnitude-, phase-, and R2*-imaging at ultra-high magnetic field strength. NeuroImage. 2013;65(0):299–314. doi: 10.1016/j.neuroimage.2012.09.055. [DOI] [PubMed] [Google Scholar]
- 11.Schweser F, Sommer K, Deistung A, Reichenbach JR. Quantitative susceptibility mapping for investigating subtle susceptibility variations in the human brain. NeuroImage. 2012;62(3):2083–2100. doi: 10.1016/j.neuroimage.2012.05.067. [DOI] [PubMed] [Google Scholar]
- 12.Marques JP, Maddage R, Mlynarik V, Gruetter R. On the origin of the MR image phase contrast: An in vivo MR microscopy study of the rat brain at 14.1 T. NeuroImage. 2009;46(2):345–352. doi: 10.1016/j.neuroimage.2009.02.023. [DOI] [PubMed] [Google Scholar]
- 13.Marques J, Bowtell R. Application of a Fourier‐based method for rapid calculation of field inhomogeneity due to spatial variation of magnetic susceptibility. Concepts in Magnetic Resonance Part B: Magnetic Resonance Engineering. 2005;25(1):65–78. [Google Scholar]
- 14.Liu T, Spincemaille P, de Rochefort L, Wong R, Prince M, Wang Y. Unambiguous identification of superparamagnetic iron oxide particles through quantitative susceptibility mapping of the nonlinear response to magnetic fields. Magnetic Resonance Imaging. 2010;28(9):1383–1389. doi: 10.1016/j.mri.2010.06.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Yao B, Li T- Q, Gelderen Pv, Shmueli K, de Zwart JA, Duyn JH. Susceptibility contrast in high field MRI of human brain as a function of tissue iron content. NeuroImage. 2009;44(4):1259–1266. doi: 10.1016/j.neuroimage.2008.10.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Li W, Wu B, Liu C. Quantitative susceptibility mapping of human brain reflects spatial variation in tissue composition. NeuroImage. 2011;55(4):1645–1656. doi: 10.1016/j.neuroimage.2010.11.088. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Langkammer C, Schweser F, Krebs N, Deistung A, Goessler W, Scheurer E, Sommer K, Reishofer G, Yen K, Fazekas F, Ropele S, Reichenbach JR. Quantitative susceptibility mapping (QSM) as a means to measure brain iron? A post mortem validation study. NeuroImage. 2012;62(3):1593–1599. doi: 10.1016/j.neuroimage.2012.05.049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Schweser F, Deistung A, Lehr BW, Reichenbach JR. Quantitative imaging of intrinsic magnetic tissue properties using MRI signal phase: An approach to in vivo brain iron metabolism? NeuroImage. 2011;54(4):2789–2807. doi: 10.1016/j.neuroimage.2010.10.070. [DOI] [PubMed] [Google Scholar]
- 19.Wong R, Jian S, Yi W. Probing sepsis and sepsis-like conditions using untargeted SPIO nanoparticles. Conf Proc IEEE Eng Med Biol Soc, Buenos Aires. 2010:3053–3056. doi: 10.1109/IEMBS.2010.5626123. [DOI] [PubMed] [Google Scholar]
- 20.Bilgic B, Pfefferbaum A, Rohlfing T, Sullivan EV, Adalsteinsson E. MRI estimates of brain iron concentration in normal aging using quantitative susceptibility mapping. NeuroImage. 2012;59(3):2625–2635. doi: 10.1016/j.neuroimage.2011.08.077. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Wang Y, de Rochefort L, Liu T, Kressler B. Magnetic source MRI: A new quantitative imaging of magnetic biomarkers. Conf Proc IEEE Eng Med Biol Soc, Minnesota. 2009:53–56. doi: 10.1109/IEMBS.2009.5335128. [DOI] [PubMed] [Google Scholar]
- 22.de Rochefort L, Nguyen T, Brown R, Spincemaille P, Choi G, Weinsaft J, Prince MR, Wang Y. In vivo quantification of contrast agent concentration using the induced magnetic field for time-resolved arterial input function measurement with MRI. Med Phys. 2008;35(12):5328–5339. doi: 10.1118/1.3002309. [DOI] [PubMed] [Google Scholar]
- 23.Liu T, Surapaneni K, Lou M, Cheng L, Spincemaille P, Wang Y. Cerebral Microbleeds: Burden Assessment by Using Quantitative Susceptibility Mapping. Radiology. 2012;262(1):269–278. doi: 10.1148/radiol.11110251. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Deistung A, Schweser F, Wiestler B, Abello M, Roethke M, Sahm F, Wick W, Nagel AM, Heiland S, Schlemmer H- P, Bendszus M, Reichenbach JR, Radbruch A. Quantitative Susceptibility Mapping Differentiates between Blood Depositions and Calcifications in Patients with Glioblastoma. PLoS ONE. 2013;8(3):e57924. doi: 10.1371/journal.pone.0057924. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Haacke EM, Tang J, Neelavalli J, Cheng YCN. Susceptibility mapping as a means to visualize veins and quantify oxygen saturation. Journal of Magnetic Resonance Imaging. 2010;32(3):663–676. doi: 10.1002/jmri.22276. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Liu T, Khalidov I, de Rochefort L, Spincemaille P, Liu J, Tsiouris AJ, Wang Y. A novel background field removal method for MRI using projection onto dipole fields (PDF). NMR in Biomedicine. 2011;24(9):1129–1136. doi: 10.1002/nbm.1670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Li L. The general application of the spherical mean value method for image noise reduction. Conf Proc IEEE Eng Med Biol Soc, Philadelphia. 2002:185–186. [Google Scholar]
- 28.Ferdinand Schweser KS, Atterbury Marie, Deistung Andreas, Lehr Berengar Wendel, Reichenbach Jürgen R. On the Impact of Regularization & Kernel Type on SHARP-Corrected GRE Phase Images. Conf Proc ISMRM, Montréal. 2011:2667. [Google Scholar]
- 29.Burckel RB. A strong converse to Gauss's mean-value theorem. The American Mathematical Monthly. 1980;87:819–820. [Google Scholar]
- 30.Hedberg L. Spectral synthesis in sobolev spaces, and uniqueness of solutions of the Dirichlet problem. Acta Mathematica. 1981;147(1):237–264. [Google Scholar]
- 31.Jain V, Abdulmalik O, Propert KJ, Wehrli FW. Investigating the magnetic susceptibility properties of fresh human blood for noninvasive oxygen saturation quantification. Magnetic Resonance in Medicine. 2012;68(3):863–867. doi: 10.1002/mrm.23282. [DOI] [PMC free article] [PubMed] [Google Scholar]