Abstract
In the last decade, high intensity focused ultrasound (HIFU) has gained popularity as a minimally-invasive and non-invasive therapeutic tool for treatment of cancers, arrhythmias, and other medical conditions. HIFU therapy is often guided by magnetic resonance imaging (MRI), which provides anatomical images for therapeutic device placement, temperature maps for treatment guidance, and post-operative evaluation of the region of interest. While piezoelectric transducers are dominantly used for MR-guided HIFU, capacitive micromachined ultrasonic transducers (CMUTs) show competitive advantages such as ease of fabrication, integration with electronics, improved efficiency, and reduction of self-heating. In this paper, we will show our first results of an unfocused CMUT transducer monitored by MR-temperature maps. This 2.51 mm by 2.32 mm, unfocused CMUT heated a HIFU phantom by 14°C in 2.5 min. This temperature rise was successfully monitored by MR thermometry in a 3.0 T General Electric scanner.
Index Terms: CMUTS, HIFU, MR-guidance, MRI, temperature mapping
I. Introduction
High intensity focused ultrasound (HIFU) has gained popularity in medical procedures because it can be focused onto a region of interest without harming intervening tissues. Focusing allows ultrasound to be applied non-invasively or minimally-invasively, with low morbidity and mortality [1]. HIFU is being investigated for neurological treatments of the central nervous system [2], [3], arrythmia treatments in the heart [4]–[7], drug delivery [8], and coagulation for deep internal bleeding [9], [10]. HIFU cancer treatments of the bone, brain, breast, liver, kidneys, prostate, and uterine fibroids are receiving increased interest; progress in these treatments is reviewed in [11].
Magnetic resonance imaging (MRI) is one of the dominant imaging modalities for guidance of HIFU cancer treatments; MR-guided HIFU systems are used extensively to treat uterine fibroids [12]–[16]. These table-top systems also have demonstrated the feasibility of HIFU in the pig liver [17], [18]. MR images provide three essential functions: (1) anatomical image guidance for placement of the HIFU probe, (2) real-time temperature monitoring using proton resonance frequency shift to target the location and duration of the therapy [19]–[21], and (3) post-process functional assessment to evaluate the degree of tissue necrosis [22]. The second function, temperature monitoring, is critical because local perfusion is variable and cannot be predicted [23].
Other imaging modalities do not accurately monitor the thermal dose or identify the necrosed tissue. While ultrasound imaging shows an increasing hyperechogenic area during treatment, this region does not correspond with ablated tissue and often obscures the tumor [24]. Also, ultrasound cannot delineate remaining viable tumor [25]. Contrast enhanced computer tomography (CT) can identify the region of tumor necrosis. However, this area cannot be imaged repeatedly or in real-time during the treatment.
Although quartz crystals with converging lenses were originally used in the 1950’s for HIFU [26], compound piezoelectric transducers have been dominant for the past 50 years. Compared to quartz crystals, piezoelectric materials are easily formed into spherical shapes, which allow mechanical focusing and eliminate coupling through a lens [27]. Since the 1950’s, many research groups have been optimizing and customizing compound piezoelectricbased materials for high power applications [28], [29]. Using these materials, both single-element, unfocused [30], [31] and mechanically-focused [27], [32] piezoelectric transducers have been used for therapeutics in the heart, liver, and kidney, to name a few. Transducer arrays have also been developed for dynamic electronic focusing of therapy [33]–[36].
Though piezoelectric transducers have been traditionally used for HIFU, capacitive micromachined ultrasonic transducers (CMUTs) have become highly competitive with regard to fabrication and performance. Because CMUTs are fabricated using a silicon micromachining process with sub-micron accuracy, they can be easily fabricated for a broad range of frequencies and sizes [37]. In addition, CMUTs are easily integrated with electronics, either monolithically or through flip-chip bonding [37]. In terms of performance, CMUTs have wide bandwidth in immersion and a high efficiency [38], [39]. In addition to these well-known benefits, CMUTs do not suffer from self-heating effects because they are fabricated from highly, thermally-conductive silicon. This makes CMUTs well suited for high power and continuous wave (CW) applications like HIFU [40], [41].
Our goal is to develop non-invasive and minimally-invasive HIFU CMUTs for applying therapy under MR-guidance. The configuration of these focused transducers will be determined by the anatomy of the region of interest and the constraints of the particular application. One example application we have discussed in previous conference publications is an 8-element, annular array for HIFU ablation of lower abdominal cancers [40], [41]. Other applications include intravascular ablations, such as cardiac ablation of arrhythmias [42].
While a focused transducer is our ultimate goal, in this first study, we used an unfocused test transducer to demonstrate that a CMUT could heat tissue-like material and that this heating could be monitored by MRI. Using an unfocused transducer is more challenging than a focused transducer for several reasons. First, without focusing, heating tissues to high temperatures in short amounts of time is more difficult since the power density is not as concentrated. Thus, an unfocused CMUT needs to be efficient to heat tissue. Second, imaging close to a transducer is difficult if the transducer materials introduce artifacts into the image. In this study, we show that CMUTs demonstrate a good efficiency necessary for heating and produce minimal artifacts, which enables MR-temperature monitoring very close to the transducer. This paper is a continuation our conference proceeding [43].
II. Methods
The CMUT used for this experiment was fabricated using a high-temperature, direct fusion-bonding process described in [44]. Cells formed from this process consisted of a silicon dioxide cavity and electrically conductive silicon membrane. Individual cells (Fig. 1) were patterned into a 2.51 mm by 2.32 mm transducer. Because the silicon is highly electrically conductive, metal is not used on the membrane surfaces, which reduces electromigration effects during high power operation. Only small aluminum pads are needed for wire bonding the CMUT to a printed circuit board (PCB). In addition, silicon is very thermally conductive, which reduces self-heating effects that degrade transducer performance.
Fig. 1.
Cross section of one cell of the CMUT transducer designed for HIFU operation at 2.5 MHz. Individual cells were fabricated in a close-packed structure to form a 2.51 mm by 2.32 mm transducer.
CMUTs are also MR-compatible. Silicon (χ=-6.52×10−6) and silicon dioxide (χ=−16.3×10−6) are level II compatible and produce no artifact; they have a susceptibility similar to tissue (χ=−7×10−6 to -11X10−6) and water (χ=−9.05×10−6) [45]. While aluminum (χ=20.7×10−6) is not level II compatible [45], because the CMUT is fabricated with conductive silicon membranes, no aluminum is used on the surface of the cells. The total volume of aluminum used in the bonding pads (150 × 150 µm and a thickness of 0.3 µm) is 2.0×10−14 m2 or 5.6X10−9 kg, a negligible amount.
Because the test equipment was not MR-compatible, long lengths of cable were required to drive the CMUT in the 3.0 T magnet. The equipment was located on a cart in the MR control room. An arbitrary waveform generator (Agilent 33250A, Agilent, Palo Alto, CA), provided a CW, 2.5 MHz-sinusoidal signal, which was then amplified by a 3 W amplifier (ENI 403LA, Electronic Navigation Industries, Richardson, TX). A power meter (HP E4419B, Hewlett Packard, Palo Alto, CA) was used to monitor the electrical power output from the amplifier. The AC signal was superimposed on the DC voltage at the bias T, which consisted of a DC blocking capacitor at the AC input port and a resistor, used for short circuit protection, at the DC voltage port. A bias voltage of 172 V was provided by a high voltage DC supply (SRS PS310, Stanford Research Systems, Stanford, CA). The combined signal was filtered using a custom-designed, 5 MHz low pass filter and connected through the panel in the control room; this grounded the drive system to the screen room. This low pass filter suppressed the high frequency noise components of the ENI amplifier, which creates noise at the nuclear resonance frequency and reduces the signal to noise ratio (SNR) of the MR image (Fig. 2).
Fig. 2.
Electrical setup in the magnet room and control room of the MR suite. The test equipment is not MR-compatible, so must be left inside the MR control room. Only a MR-compatible cable, an air core inductor, and the CMUT can be safely used inside the magnet.
All the materials inside the magnet room and 3.0 T magnet (126 MHz proton resonance frequency) needed to be MR-compatible and non-ferrous. A MR-compatible mini-coax (Belden Cable, St. Louis, MO) was used to connect the CMUT to the control panel inside the magnet room. We used an air-core toroid of approximately 32 µH to cancel the reactive component of the CMUT and to match the circuit as closely to 50 ohms as possible (Fig. 2). This makes it more efficient for the amplifier to drive the load.
Before the experiment, we characterized the CMUT and circuit outside the magnet room because our equipment was not MR-compatible. All these measurements allowed us to calculate the power gain of the device during the experiment in the 3.0 T magnet. First, we measured the impedance at node C and B (Fig. 3) to use in power calculations. At 2.5 MHz, the low pass filter impedance was negligible, so nodes A and C have roughly the same impedance. We also characterized the voltage gain of the circuit from node A to B to be 6.25. This allowed us to monitor the CMUT during operation by observing the voltage at node A.
Fig. 3.
Simplified electrical circuit used for driving the HIFU CMUT. The bias T superimposed the 172 V DC bias and 2.5 MHz CW sinusoidal signal used to drive the CMUT. The inductor was used to tune the CMUT so that the reactive component of the impedance is zero at 2.5 MHz. The inductor also provides a voltage gain of 6.25.
To avoid charging and breaking the actual device, we measured the output pressure of an identical device under the desired DC and AC driving voltage conditions using an calibrated piezoelectric hydrophone with 400 µm aperture (PZT Z44 0400 hydrophone, Onda Corporation, Sunnyvale, CA). The device was first submerged in soybean oil (speed of sound of 1430 m/s and density of 930 kg/m3) for electrical isolation. The hydrophone was placed 2 cm from the surface of the transducer and positioned laterally to find the largest pressure. Care was taken to make sure the hydrophone was properly aligned with the transducer face to minimize the errors. We accounted for the frequency response of the hydrophone [46], attenuation of the soybean oil (0.43 dB/cm at 2.5 MHz [47]), and diffraction [48] to calculate the pressure at the surface of the transducer. We did not account for nonlinear propagation and harmonics, which gives a conservative estimate of the output pressure and thus the power of the transducer. The measured output pressure and the impedance of the soybean oil (Z=1.45 MRayl) were used to calculate the average power density at the face of the transducer. Multiplying by the area of the device (5.8 mm2) gave us the total acoustic output power.
We only studied one drive level for the CMUT in this experiment, corresponding to 172 V DC voltage (60% of collapse) and 250 Vpp AC voltage. In previous work, we found that a DC voltage of 60%, while applying as large of an AC voltage as possible, produced the optimum output pressure [49]. The acoustic power in not linear with respect to the applied voltage because the electrostatic force exerted on the membrane is dependent on the square of the distance between the ground and top electrodes. The maximum output pressure is related to the balance between the DC operating point and the AC voltage applied. At higher DC voltages, the force and pressure per volt increases nonlinearly as the collapse voltage is reached. However, as the DC voltage approaches the collapse voltage, the amount of AC voltage that can be applied before the membrane collapses is smaller [49].
For MR imaging, the CMUT was placed in the 3.0 T GE scanner (Lucas Center, Stanford, CA) and 10 drops of soybean oil were dribbled on the surface for electrical insulation. A polyethylene bag isolated the oil layer and CMUT from the coupling gel (MediChoice, Mechanicsville, VA), which coupled the ultrasound energy into the 4-inch diameter, tissue-mimicking, HIFU phantom (ATS Laboratories, Inc., Bridgeport, CT). The opaque phantom has an acoustic attenuation of 0.503 dB/m/K, a speed of sound of 1538 m/s, a heat capacity of 3500 J/kg/K, and a thermal conductivity of 0.5 W/m/k. Since the phantom is water based, the proton resonance frequency properties will resemble that of soft tissue. The coupling gel was applied directly from the bottle and not degassed; however, when placing the HIFU phantom on top of the oil and coupling gel, we exerted force to minimize the amount of coupling fluids between the phantom and the transducer and also to minimize the air pockets. Two imaging planes, coronal and sagittal, were measured in two different experiments using a 5 inch GE surface coil placed around the HIFU phantom (Fig. 4).
Fig. 4.
Sagittal (a) and coronal (b) views of the setup used for MR-temperature mapping of the HIFU CMUT.
Before the heating experiment, we measured the susceptibility artifact of the CMUT to verify that it did not interfere with measurements of the heated region. To do this, we added a second phantom underneath the CMUT to identify the bottom surface of the printed circuit board on the image. We imaged the CMUT, with no applied voltage, in the sagittal plane using a fast gradient echo sequence with a transmit repetition time/transmit echo time (TR/TE) of 28.7/19.1 ms, the same sequence used for the thermal imaging.
During the heating, we used a fast gradient echo sequence with a TR/TE of 17.4/13.5 ms for the coronal plane and 28.7/19.1 ms for the sagittal plane. The magnitude, phase, real, and imaginary results were saved from the scan. One hundred images per series were captured for 7.5 and 12.5 minutes for the coronal and sagittal planes, respectively. During this imaging time, we applied 2.5 min and 5 min of therapy at 2.5 MHz for the coronal and sagittal planes, respectively. Each image was 256 by 256 pixels, with a spatial resolution of 0.7 mm in the coronal images and 0.9 mm in the sagittal images.
The proton resonance frequency shift due to temperature was observed as a change in phase, Δϕ, that we calculated by subtracting successive images from the first image. The change in temperature, ΔT, was then calculated using the following equation [50]:
| (1) |
where the proton resonance shift α is 0.01 ppm/°C, the proton resonance frequency γ is 26734 rad/sec/G, the magnetic field strength B0 is 3.0 T, and transmit echo time (TE) is given by the scan parameters mentioned previously. The temperature coefficient α is stable for temperatures under focused ultrasound application [51]. To account for baseline drift of the MR image over time, we averaged the phase difference in a 1.5 cm by 1.5 cm region of the phantom far from the heated region; this phase difference was then subtracted from Δϕ before the temperature was calculated. The temperature error was determined by calculating the standard deviation of the temperature map before heating. This standard deviation was calculated in a 1.5 cm by 1.5 cm area that encompassed the region of interest where heating would occur.
The Cumulative Equivalent Minutes at 43°C (CEM43) was then calculated assuming a temperature rise from 37°C. The CEM43 is given by the following equation [52]:
| (2) |
where R = 0.25 if T(t) < 43 and R = 0.5 if T(t) > 43.
This assumes that heating of a room temperature, tissue-mimicking phantom requires roughly the same amount of power as heating a 37°C phantom. Heating tissue is reliant on the attenuation and speed of sound of the tissue; between 20°C and 37°C, the speed of sound and attenuation shows a change of less than 10% [53], [54]. The attenuation increases rapidly at temperatures greater than 50°C because of protein denaturation. This increase in attenuation will cause an even greater degree of heating. This means that by starting at a lower temperature, we are effectively underestimating our heating and necrosis threshhold.
III. Results and Discussion
We first measured the susceptibility artifact of the device with no applied power. We found that the artifact extended at most 1 mm from the surface of the CMUT (Fig. 5). Given that we measured the coronal slice 3–6 mm away from the device surface and that the heating areas in both slices extend far past 1 mm, our temperature maps in those locations are not impacted by susceptibility artifacts.
Fig. 5.
Measurement in the sagittal plane o of the artifact caused by the CMUT.
Next, we evaluated the performance of the CMUT by calculating efficiencies and also observing the temperature rise of the phantom under MR temperature guidance. To evaluate the effectiveness of the tuning circuit, we calculated the electrical efficiency, defined as the electrical power delivered to the CMUT compared to the electrical power supplied by the amplifier. We found the electrical efficiency by measuring the applied voltage at node A (Fig. 2) and calculating the delivered power to the CMUT using the impedance measured at nodes A and B. With normal care in calibration, it is possible to measure impedance with errors of less than 1%. Though 2.5 W was measured from the output of the amplifier, only 1.19 W of electrical power was delivered to the CMUT, which is an electrical efficiency of 48%. The cable, connections, and inductor have parasitics, which cause this loss.
From the hydrophone measurement, under 172 Vbias and 250 Vpp AC voltage, the CMUT showed a surface output pressure of 1.3 MPa peak to peak and an acoustic output power of 0.8 W in immersion. Since 1.19 W of electrical power was supplied to the CMUT, the electrical to acoustical efficiency of the CMUT is thus 68%. Though the hydrophone is calibrated by the manufacturer (Onda Corporation, Sunnyvale, CA), there are still errors associated with the accuracy of the acoustic pressure measurement, and thus the accuracy of the calculated acoustic output power. Though we took care to align the hydrophone with the transducer, tilt and positioning errors will cause the measured output pressure to be less than the maximum. This will cause us to underestimate our efficiency. However, because the transducer is unfocused, its beamwidth is large at 2 cm compared to the size of the hydrophone, so the error will be minimal.
Already, this unfocused CMUT transducer was efficient enough to produce a maximum temperature rise in the HIFU phantom of 14.6°C in the coronal plane after 2.5 min of heating (Fig. 6 a). In measurements of the sagittal plane, 18.6°C temperature rise was achieved after 5 min of heating (Fig. 6 b). Approximately 200 voxels were in the heated area of each image. The standard deviation of the temperature maps, which indicates the uncertainty and error in the temperature measurement, was 0.7°C and 0.8°C in the coronal and sagittal planes, respectively. This error is only 5% compared to the maximum temperature rise.
Fig. 6.
Coronal (a) and sagittal (b) temperature maps at the end of the heating time period, at 2.5 and 5 min respectively. Calculated CEM43 for the corononal (c) and sagittal (d) images, which shows that some areas reach 128 equivalent minutes at the end of the heat. This is enough to necrose heat tissue.
These temperature rises are reasonable for unfocused ultrasound; a change in temperature of over 13°C would raise the body temperature above 50°C, which causes coagulative necrosis [55], [56]. Using the CEM43, we quantified the degree of necrosis as a function of temperature and time. The maximum CEM43 reached by any pixel in the treatment region was 252 and 6230 equivalent minutes in the coronal and sagittal planes, respectively. Looking only at the pixels that reached the 128 equivalent minute [57] threshold for myocardial necrosis (Fig. 6c,d), we found that this device could potentially necrose cardiac tissue to a depth of approximately 1 cm, discounting the effects of perfusion. Because the HIFU phantom does not simulate perfusion, the heating and temperature rise in vivo will be less than what we measured in the phantom. To deliver more power and overcome perfusion to necrose tissue in shorter amounts of time, we will develop a focused transducer design.
IV. Conclusion
We have demonstrated that CMUTs can be used to heat a HIFU phantom with unfocused ultrasound by 14–18°C over a 2.5–5 min time period. This CMUT demonstrated an MR-artifact of less than 1 mm radially from its surface and was successfully monitored using MR-temperature maps. Though the temperature rise in an unperfused, HIFU phantom from this unfocused transducer is relatively small, the heating produced a CEM43 of over 128 min near the transducer. This is the threshold necrosis value for cardiac tissue. These first experimental results and MR-temperature maps are promising for the further development of CMUTs in ultrasonic therapeutics.
In the future, we aim to build a ring annular focused transducer for cancer ablation, which will enable greater heating in shorter periods of time, even with perfused tissue. We also plan to improve our electrical boards and connections to reduce radio frequency noise and induction of eddy currents. We can do this by reducing the electrical loops in our circuit and designing a new tuning circuit that does not require such a large inductor.
ACKNOWLEDGMENT
This research is supported by NIH R01 CA77677 and R01 CA 121163.
Contributor Information
Serena H. Wong, Email: shwong@stanford.edu.
Kim Butts-Pauly, Email: kbpauly@stanford.edu.
Butrus T. Khuri-Yakub, Email: khuriyakub@stanford.edu.
References
- 1.ter Haar G. Ultrasound focal beam surgery. Ultrasound in Medicine and Biology. 1995;25(5):1089–1100. doi: 10.1016/0301-5629(95)02010-1. [DOI] [PubMed] [Google Scholar]
- 2.Fry WJ, Mosberg WH, Barnard JW, Fry FJ. Ultrasonics lesions in the central nervous system with ultrasound. Journal of Neurosurgery. 1954;11:471–478. doi: 10.3171/jns.1954.11.5.0471. [DOI] [PubMed] [Google Scholar]
- 3.Fry WJ, Mosberg WH, Barnard JW, Fry FJ. Production of focal destructive lesions in the central nervous system with ultrasound. Science. 1955;122:517–518. doi: 10.3171/jns.1954.11.5.0471. [DOI] [PubMed] [Google Scholar]
- 4.Zimmer JE, Hynynen K, Marcus F, He DS. The feasibility of using ultrasound for cardiac ablation. 1993:1203–1206. doi: 10.1109/10.412655. [DOI] [PubMed] [Google Scholar]
- 5.Gentry KL, Smith SW. Integrated catheter for 3-d intracardiac echocardiography and ultrasound ablation. doi: 10.1109/tuffc.2004.1320739. [DOI] [PubMed] [Google Scholar]
- 6.Lesh MD, Kalman JM, Karch MR. Use of intracardiac echocardiography during electrphysiologic evaluation and therapy of atrial arrhthmias. Journal of Cardiovascular Electrophysiology. 1988;9(Suppl.):S40–S47. [PubMed] [Google Scholar]
- 7.Kluiwstra JU, Zhang Y, VanBaren P, Strickberger SA, Ebbini ES, Cain CA. Ultrasound phased arrays for noninvasive myocardial ablation: initial studies. 1995 [Google Scholar]
- 8.Curra FP, Crum LA. Therapeutic ultrasound: Surgery and drug delivery. Acoustical Science and Technology. 2003;24(6):343–348. [Google Scholar]
- 9.Vaezy S, Martin R, Yaziji H, Kaczkowski P, Keilman G, Carter S, Caps M, Chi EY, Bailey M, Crum L. Hemostasis of punctured blood vessels using high-intensity focused ultrasound. Ultrasound in Medicine and Biology. 1998;24:903910. doi: 10.1016/s0301-5629(98)00050-7. [DOI] [PubMed] [Google Scholar]
- 10.Zderica V, Keshavarzia A, Noble ML, Pauna M, Shararb SR, Cruma LA, Martina RW, Vaezy S. Hemorrhage control in arteries using high intensity focused ultrasound: A survival study. Ultrasonics. 2006;44:4653. doi: 10.1016/j.ultras.2005.08.002. [DOI] [PubMed] [Google Scholar]
- 11.Kennedy JE, ter Haar G, Cranston D. High intensity focused ultrasound: surgery of the future? 2003;76:590–599. doi: 10.1259/bjr/17150274. [DOI] [PubMed] [Google Scholar]
- 12.Jolesz FA, Hynynen K. Magnetic resonance image-guided focused ultrasound surgery. Cancer J. 2002 [PubMed] [Google Scholar]
- 13.Jolesz FA, Hynynen K, McDannold N, Freundlich D, Kopelman D. Noninvasive thermal ablation of hepatocellular carcinoma by using magnetic resonance imaging-guided focused ultrasound. Gastroenterology. 2004;127 doi: 10.1053/j.gastro.2004.09.056. [DOI] [PubMed] [Google Scholar]
- 14.Tempany CM, Stewart EA, McDannold N, Quade BJ, Jolesz FA, Hynynen K. MR imaging-guided focused ultrasound surgery of uterine leiomyomas: A feasibility study. Radiology. 2003;226(3):897–905. doi: 10.1148/radiol.2271020395. [DOI] [PubMed] [Google Scholar]
- 15.Hynynen K, Pomeroy O, Smith DN, Huber PE, McDannold NJ, Kettenbach J, Baum J, Singer S, Jolesz FA. Mr imaging-guided focused ultrasound surgery of fibroadenomas in the breast: A feasibility study. Radiology. 2001;219(1):176–185. doi: 10.1148/radiology.219.1.r01ap02176. [DOI] [PubMed] [Google Scholar]
- 16.Hynynen K, Colucci V, Chung A, Jolesz F. Noninvasive arterial occlusion using MRI-guided focused ultrasound. Ultrasound Med Biol. 1996;22(8):1071–1077. doi: 10.1016/s0301-5629(96)00143-3. [DOI] [PubMed] [Google Scholar]
- 17.Jolesz FA, Hynynen K, McDannold N, Freundlich D, Kopelman D. Noninvasive thermal ablation of hepatocellular carcinoma by using magnetic resonance imaging-guided focused ultrasound. Gastroenterology. 2004;127(5 Suppl 1):S242–S247. doi: 10.1053/j.gastro.2004.09.056. [DOI] [PubMed] [Google Scholar]
- 18.Daum DR, Smith NB, King R, Hynynen K. In vivo demonstration of noninvasive thermal surgery of the liver and kidney using an ultrasonic phased array. Ultrasound Med Biol. 1999;25(7):1087–1098. doi: 10.1016/s0301-5629(99)00053-8. [DOI] [PubMed] [Google Scholar]
- 19.Cline HE, Hynynen K, Hardy CJ, Watkins RD, Schenck JF, Jolesz JF. Mr temperature mapping of focused ultrasound surgery. Magnetic Resonance in Medicine. 1994;31(6):628–636. doi: 10.1002/mrm.1910310608. [DOI] [PubMed] [Google Scholar]
- 20.Rieke V, Vigen KK, Sommer G, Daniel BL, Pauly JM, Butts K. Referenceless prf shift thermometry. Magnetic Resonance Medicine. 2004;5131(6):628–636. doi: 10.1002/mrm.20090. [DOI] [PubMed] [Google Scholar]
- 21.Butts K, Rieder SJ, Ehman RL, Thompson RM, Jack CR. Interleaved echo planar imaging on a standard MRI system. Magnetic Resonance in Medicine. 1994;31(1) doi: 10.1002/mrm.1910310111. [DOI] [PubMed] [Google Scholar]
- 22.Cline HE, Schenck JF, Hynynen K, Watkins RD, Souza SP, Jolesz FA. Mr-guided focused ultrasound surgery. Jurnal of Computer Assisted Tomography. 1992;16:956–965. doi: 10.1097/00004728-199211000-00024. [DOI] [PubMed] [Google Scholar]
- 23.Lewin JS, Petersilge CA, Hatem SF, Duerk JL, Lenz G, Clampitt ME, Williams ML, Kaczynski KR, Lanzieri CF, Wise AL, Haaga JR. Interactive mr imaging-guided biopsy and aspiration with a modified clinical c-arm system. AJR American Jouranl Roentgenology. 1998;170(6):1593–1601. doi: 10.2214/ajr.170.6.9609180. [DOI] [PubMed] [Google Scholar]
- 24.Vogl TG. Laser-induced thermotherapy: Do we really need mr thermometry? European Radiolology. 2002;12(1):5–6. doi: 10.1007/s003300101117. [DOI] [PubMed] [Google Scholar]
- 25.Rossi S, Di Stasi M, Buscarini E, Quaretti P, Garbagnati F, Squassante L, Paties CT, Silverman DE, Buscarini L. Percutaneous rf interstitial thermal ablation in the treatment of hepatic cancer. 1996;167(3):759–768. doi: 10.2214/ajr.167.3.8751696. [DOI] [PubMed] [Google Scholar]
- 26.Fry WJ. Precision high intensiy focused ultrasound machines for surgery. American Journal of Physics in medicine. 1958;37:152–156. [PubMed] [Google Scholar]
- 27.Rivens IH, Clark RL, ter Haar GR. Design of focused ultrasound surgery transducers. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control. 1996;43(6):1023–1031. [Google Scholar]
- 28.Zhang S, Xia R, Lebrun L, Anderson D, Shrout TR. Piezoelectric materials for high power, high temperature applications. Materials Letters. 2005;59(27) [Google Scholar]
- 29.Chapelon JY, Cathignol D, Cain C, Ebbini E, Kluiwstra JU, Sapozhnikov OA, Fleury G, Berriet R, Chupin L, Guey JL. New piezoelectric transducers for therapeutic ultrasound. Ultrasound in Medicine and Biology. 2000;26(1):153–139. doi: 10.1016/s0301-5629(99)00120-9. [DOI] [PubMed] [Google Scholar]
- 30.Zimmer JE, Hynynen K, He DS, Marcus F. The feasibility of using ultrasound for cardiac ablation. IEEE Transactions on Biomedical Engineering. 42(9) doi: 10.1109/10.412655. [DOI] [PubMed] [Google Scholar]
- 31.Deardorff DL, Diederich CJ. Ultrasound applicators with internal water-cooling for high-powered interstitial thermal therapy. IEEE Trans Biomed Eng. 2000;47(10):1356–65. doi: 10.1109/10.871409. [DOI] [PubMed] [Google Scholar]
- 32.Brentnall MD, Martin RW, Vaezy S, Kaczkowski P, Forster F F, Crum L. A new high intensity focused ultrasound applicator for surgical applications. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control. 2001;48 doi: 10.1109/58.895907. [DOI] [PubMed] [Google Scholar]
- 33.Ebbini E, Bischof JC, Coad JE. Lesion formation and visualization using dual-mode ultrasound phased arrays. 2001:1351–1354. [Google Scholar]
- 34.Wan H, VanBaren P, Ebbini ES, Cain CA. Ultrasound surgery: Comparison of strategies using phased array systems. IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control. 1996;43(6):1085–1098. [Google Scholar]
- 35.Sanghvi Shih-jeh Wu N, Seip R, Tan JS, Frizzell L, Kouzmanoff J. Ultrasound phased arrays for prostate treatment. Journal of the Acoustical Society of America. 2001;109(6):3055–3064. doi: 10.1121/1.1373444. [DOI] [PubMed] [Google Scholar]
- 36.Gentry KL, Smith SW. Integrated catheter for 3-d intracardiac echocardiography and ultrasound ablation. IEEE Trans. Ultrason., Ferroelect., Freq. Contr. 2004;51:800–808. doi: 10.1109/tuffc.2004.1320739. [DOI] [PubMed] [Google Scholar]
- 37.Ergun AS, Huang Y, Zhuang X, Oralkan O, Yarahoglu GG, Khuri-Yakub BT. Capacitive micromachined ultrasonic transducers: fabrication technology. Ultrasonics, Ferroelectrics and Frequency Control, IEEE Transactions on. 2005;52(12):2242–2258. doi: 10.1109/tuffc.2005.1563267. [DOI] [PubMed] [Google Scholar]
- 38.Ladabaum I, Jin XC, Soh HT, Atalar A, Khuri-Yakub BT. Surface micromachined capacitive ultrasonic transducers. IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control. 1998;45(3):678–690. doi: 10.1109/58.677612. [DOI] [PubMed] [Google Scholar]
- 39.Yaralioglu GG, Ergun AS, Bayram B, Hggstrm E, Khuri-Yakub BT. Calculation and measurement of electromechanical coupling coefficient of capacitive micromachined ultrasonic transducers. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control. 2003;50(4):449–456. doi: 10.1109/tuffc.2003.1197968. [DOI] [PubMed] [Google Scholar]
- 40.Wong SH, Ergun AS, Yaralioglu GG, Oralkan, Liu E, Kupnik M, Butts RK, Khuri-Yakub BT. Capacitive micromachined ultrasonic transducers for high intensity focused ablation of upper abdominal tumors. 2006:841–844. [Google Scholar]
- 41.Wong SH, Ergun AS, Yaralioglu GG, Oralkan, Kupnik M, Butts RK, Khuri-Yakub BT. Design of HIFU CMUT arrays for treatment of liver and renal cancer. 2007;911:54–60. [Google Scholar]
- 42.Wong SH, Wygant IO, Yeh DT, Zhuang X, Bayram B, Kupnik, O.Oralkan M, Ergun AS, Yaralioglu GG, Khuri-Yakub BT. Capacitive micromachined ultrasonic transducer arrays for integrated diagnostic/therapeutic catheters. 2006;ume 829:395–399. [Google Scholar]
- 43.Wong SH, Watkins RD, Kupnik M, Butts-Pauly K, Khuri-Yakub BT. Advantages of capacitive micromachined ultrasonics transducers (cmuts) for high intensity focused ultrasound (hifu) 2007 [Google Scholar]
- 44.Huang Y, Haeggstrom E, Bayram B, Zhuang X, Ergun AS, Cheng CH, Khuri-Yakub BT. Fabricating capactive micromachined ultrasonics transducers with wafer-bonding technology. Microelectromechanical Systems. 2003;12(2):128–137. [Google Scholar]
- 45.Schenck JF. The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds. Medical Physics. 1996;23(6) doi: 10.1118/1.597854. [DOI] [PubMed] [Google Scholar]
- 46. http://www.ondacorp.com/. [Google Scholar]
- 47.Oralkan O, Ergun AS, Johnson JA, Karaman M, Utkan Demirci, Kaviani K, Lee TH, Khuri-Yakub BT. Capacitive micromachined ultrasonics transdicer: Next-generation arrays for acoustic imaging? IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control. 2002;49(11):1596–1610. doi: 10.1109/tuffc.2002.1049742. [DOI] [PubMed] [Google Scholar]
- 48.Kino G. Acoustic waves: devices, imaging, and analog signal processing. Prentice-Hall; 1987. [Google Scholar]
- 49.Yaralioglu GG, Bayram B, Khuri-Yakub BT. Finite element analysis of cmuts: Conventional vs. collapse operation modes. Proceeding of the IEEE Ultrasonics Symposium. 2006:586–589. [Google Scholar]
- 50.Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K, Suzuki Y. A precise and fast temperature mapping using water proton chemical shift. Magnetic Resonance Medicine. 1995;34:814823. doi: 10.1002/mrm.1910340606. [DOI] [PubMed] [Google Scholar]
- 51.Wu Tao, Kurtis R Kendell, Joel P Felmlee, Bradley D Lewis, Richard L Ehman. Reliability of water proton chemical shift temperature calibration for focused ultrasound ablation therapy. Medical Physics. 2000;27(1):221–224. doi: 10.1118/1.598864. [DOI] [PubMed] [Google Scholar]
- 52.Sapareto SA, Dewey WC. Thermal dose determination in cancer therapy. International Journal of Radiology Oncolology Biology Physics. 1984;10:787. doi: 10.1016/0360-3016(84)90379-1. [DOI] [PubMed] [Google Scholar]
- 53.Bamher J, Hill C. Ultrasonic attenuation and propagation speed in mammalian tissues as a function of temperature. Ultrasound in Med. and Biol. 1979;5:149–157. doi: 10.1016/0301-5629(79)90083-8. [DOI] [PubMed] [Google Scholar]
- 54.Techavipoo U, Varghese T, Chen Q, Stiles TA, Zagzebski JA, GFrank GR. Temperature dependence of ultrasonic propagation speed and attenuation in excised canine liver tissue measured using transmitted and reflected pulses. Journal of Acoustical America. 2004;115(6):2859–2865. doi: 10.1121/1.1738453. [DOI] [PubMed] [Google Scholar]
- 55.Seegenschmiedt MH, Fessenden P, Vernon CC. Principles and Practices of Thermoradiotherapy and Thermochemotherapy. Berlin,Germany: Springer-Verlag; 1995. [Google Scholar]
- 56.Foster FS, Henkelman RM, Plewes DB. Quantifying tissue damage due to focused ultrasound heating observed by MRI. Magnetic Resonance in Medicine. 1999;41:321–328. doi: 10.1002/(sici)1522-2594(199902)41:2<321::aid-mrm16>3.0.co;2-9. [DOI] [PubMed] [Google Scholar]
- 57.Haemmerich D, Webster JG, Mahvi DM. Thermal dose versus isotherm as lesion boundary estimator for cardiac hepatic radiofrequency ablation. 2003:134–136. [Google Scholar]