Abstract
Background: Orthodontic tooth movement results from application of forces to teeth. Elastics in orthodontics have been used both intra-orally and extra- orally to a great effect. Their use, combined with good patient co-operation provides the clinician with the ability to correct both anteroposterior and vertical discrepancies. Force decay over a period of time is a major problem in the clinical usage of latex elastics and synthetic elastomers. This loss of force makes it difficult for the clinician to determine the actual force transmitted to the dentition. It’s the intent of the clinician to maintain optimal force values over desired period of time. The majority of the orthodontic elastics on the market are latex elastics. Since the early 1990s, synthetic products have been offered in the market for latex-sensitive patients and are sold as nonlatex elastics. There is limited information on the risk that latex elastics may pose to patients. Some have estimated that 0.12–6% of the general population and 6.2% of dental professionals have hypersensitivity to latex protein. There are some reported cases of adverse reactions to latex in the orthodontic population but these are very limited to date. Although the risk is not yet clear, it would still be inadvisable to prescribe latex elastics to a patient with a known latex allergy. To compare the in-vitro performance of latex and non latex elastics. Materials & Methods: Samples of 0.25 inch, latex and non latex elastics (light, medium, heavy elastics) were obtained from three manufacturers (Forestadent, GAC, Glenroe) and a sample size of ten elastics per group was tested. The properties tested included cross sectional area, internal diameter, initial force generated by the elastics, breaking force and the force relaxation for the different types of elastics. Force relaxation testing involved stretching the elastics to three times marketed internal diameter (19.05 mm) and measuring force level at intervals over a period of 48 hours. The data were analyzed with student independent – t test, analysis of variance and the Tukey – HSD test at p <0.05 level of significance. Results: Non latex elastics had greater cross sectional area than latex elastics in all types of elastics. Forestadent heavy elastics had grater cross sectional area than GAC and Glenroe. There was no statistically significant difference in the internal diameter in between all type of elastics. Forestadent non latex elastics had greater breaking force compared to GAC and Glenroe elastics. Forces generated by the elastics decreased over 48 hours to an average load approximating 65-75% of the manufacturer’s values. Force degradation was greater in non latex elastics compared to latex elastics. Conclusion: The results of the study demonstrated that the clinical choice of elastics should be based on the patient’s medical history and the specific mechanical properties of the type of elastic. How to cite the article: Kamisetty SK, Nimagadda C, Begam MP, Nalamotu R, Srivastav T, Shwetha GS. Elasticity in Elastics-An in-vitro study. J Int Oral Health 2014;6(2):96-105
Key words: : Artificial saliva, elasticity, force degradation, force relaxation, latex elastics, non latex elastic
Introduction
“Force is the medicine in Orthodontics” - Sheldon Friel. Orthodontic tooth movement results from application of forces to teeth. Materials used to move teeth include arch wire loops, coil springs, elastics and synthetic elastomers, etc. Rubber and its derivatives and synthetic elastomers of polyurethane nature have been a reliable method of delivering forces in fixed appliances. Elastomer is a general term, which encompasses materials that after substantial deformation rapidly return to their original dimensions.
Natural rubber is an elastomer, but not all elastomers can be called rubber. The word rubber synergises with natural or tree rubber which is a hydrocarbon polymer of isoprene units. 1 The synthetic rubbers which have been developed posses different chemical structures but resemble tree rubber in many physical properties. Synthetic rubber polymers used for Orthodontic purpose is usually polyurethane rubber. They can be synthesized by extending polyester or a polyether glycol or polyhydrocarbon diol with a diisocyante. Depending upon the end use a variety of means of processing and synthesizing may be employed. 2 Increased incidence of latex allergic reactions was the reason for the increase in the use of non latex products within the orthodontic specialty. Hence, the assessment of the material properties of non latex elastics becomes increasingly important clinically.
Force decay over a period of time is a major problem in the clinical usage of latex elastics and synthetic elastomers.
Figure 2: Glenroe intra oral elastics.
This loss of force makes it difficult for the clinician to determine the actual force transmitted to the dentition. It is the intent of the clinician to maintain optimal force values over desired period of time. Hence knowledge of force decay of elastics and synthetic elastomers will help to
determine their clinical usage.
The purpose of this study was to compare invitro performance of latex and non-latex elastics and to measure the internal diameter, cross sectional area, initial forces generated, force relaxation of elastics held at constant extension for 48 hours and the breaking force for the different types of elastics.
Materials and Methods
Samples of latex and non-latex, non-colored Orthodontic elastics were obtained from three manufacturers - Forestadent (Pforzheim, Germany), Glenroe (Bradenton, Florida) and GAC International (Islandia, NY). (Figure 1 - 3 ). The elastics were reported to be of 0.25-inch internal diameter (ID) and of light, medium and heavy forces. The samples obtained were well within shelf life. They were refrigerated in plastic covers provided by the manufacturers and kept away from sunlight to prevent any
Figure 1: Forestadent intra oral elastics.
Figure 3: GAC intra oral elastics.
deterioration. All testing was conducted on 10 samples of each elastic type, thus there were 180 elastics tested for each specific mechanical test. All elastics were tested as intact loops.
Methodology
The internal diameter and cross sectional area of each elastic was measured with the use of a Mitutoyo thickness gauge (Model 7309, Mitutoyo, Aurora, III) ( Figure 4 ). Measurements were made at four locations on 10 elastics. The mean cross-sectional area and internal diameter were calculated and used to assess the uniformity of the elastic's morphology. The following formula was used to calculate the cross sectional area.
Figure 4: Mitutoyo Digital Vernier caliper.
Cross sectional area = width × thickness
Breaking force test was performed with a Universal Testing Machine. Two stainless steel pins of 1mm were mounted on two self cure acrylic blocks of dimension 35 × 9 × 9 mm, which were fixed on the upper and lower jaws of the testing machine. The elastics were engaged on the hooks and extended at a crosshead speed of 1 mm/min until failure, at which time the peak failure force and extension were recorded. The forces generated at the extension of twice and thrice the internal diameter of the elastics was also recorded in order to compare initial force values at this stretched distance.
The 48 hours load relaxation mechanical testing was conducted with a Universal Testing Machine. For load relaxation mechanical testing the elastics were mounted between two stainless steel pins of 1mm on acrylic board. The pins were set apart at a fixed distance of 19.05 mm, so
as to stretch the elastics to three times the marketed internal diameter. The acrylic board was stored in artificial saliva and the temperature of the artificial saliva was regulated at 37 o C using a submersible water heater and thermostat to simulate the oral environment.
The ingredients of the artificial saliva were as follows: 1.3 g/l potassium chloride, 0.1 g/l sodium chloride, 0.05 g/l magnesium chloride, 0.1 g/l calcium chloride, 2.5× 10 ˉ5 g/l sodium fluoride, 0.035 g/l potassium dihydrogen phosphate and 0.162 g/l zinc sulphate. The pH value was 7.0. 2 Forces generated by the elastics were recorded immediately after they were placed in the apparatus at 1, 6, 12, 24, 48 hours. (Figure 5A , 5B , 5C ) with a pair of
Figure 5A: Custom-made apparatus for 48-hour load relaxation tests- superior view.
Figure 5B: Custom-made apparatus for 48-hour load relaxation tests- lateral view.
Figure 5C: Ph 7 for the artificial saliva.
tweezers, each elastic was transferred to the Instron universal testing machine from the acrylic measuring board at specific time interval as mentioned above.The resultant force at different immersion times were recorded in the personal computer.
From each specimen, the percentage of force relaxation (%R) was obtained as follows
%R = 100 x Fo - Ft
Where Fo: Initial force
Ft: Force at that particular time 1, 6, 12, 24, 48 hrs).
Percentage of load remaining (%LR) was obtained as follows
% LR = 100 - %R
Where %R: Percentage of force relaxation (1, 6, 12, 24, 48 hrs).
Statistical Methods
The cross sectional area, internal diameter, breaking force and load relaxation data were analyzed. Mean and standard deviation was calculated to determine the statistical significance of the differences in between different types of elastics. Mean and SD of the force values generated at 2 and 3 times extensions were calculated and descriptively compared with the manufacturers specified force values. Comparison was made between latex and non latex elastics with student t-test. Comparison was also made between light, medium, heavy elastics and between manufacturers with ANOVA and Tukey- HSD for multiple comparisons. All the statistical analyses were conducted with the package SPSS/pc+. (statistical package for social science, version 11.0 and systat 8.0).
Results
The results showed that non latex elastics had greater cross sectional area compared to latex elastics except in GAC light elastics where there was no significant difference. The heavy elastics had larger cross sectional area than the medium elastics and the medium elastics had larger cross sectional area than the light elastics. Between manufacturers, Forestadent heavy elastics had greater cross sectional area compared to GAC and Glenroe but in light and medium elastics, Glenroe had greater cross sectional area compared to Forestadent and GAC There was no statistical significant difference in the internal diameter between the various types of elastics ( Table 1 )The results of breaking force comparison showed that non latex elastics had greater breaking force than latex elastics of all
Table 1: Variability in elastic cross sectional areas between latex and non latex of 3 manufacturers.
ELASTIC TYPE | LIGHT | MEDIUM | HEAVY | ||||||||||
Mean | S.D | Mean diff | Sig | Mean | S.D | Mean diff | Sig | Mean | S.D | Mean diff | Sig | ||
FORESTADENT | Latex | 0.703 | 0.031 | -0.32 | 0.000 | 1.07 | 0.02 | -0.085 | 0.000 | 1.62 | 0.03 | -0.08 | 0.000 |
Non latex | 1.02 | 0.06 | 1.16 | 0.02 | 1.7 | 0.02 | |||||||
GAC | Latex | 0.62 | 0.02 | -0.015 | 0.15 | 0.81 | 0.03 | -0.17 | 0.000 | 1.18 | 0.03 | -0.089 | 0.000 |
Non latex | 0.63 | 0.021 | 0.98 | 0.1 | 1.27 | 0.04 | |||||||
GLENROE | Latex | 0.95 | 0.012 | -0.128 | 0.005 | 1.15 | 0.02 | -0.045 | 0.002 | 1.2 | 0.03 | -0.06 | 0.000 |
P<0.01 = significant
Table shows that the cross sectional area of non-latex elastics (light, medium, heavy) were significantly greater than latex elastics in all three manufacturers, except in GAC light elastics where there was no statistical difference (p>0.05) between latex and non latex elastics. Student independent t test was used to find the significance between the two (latex, non latex) groups.
types from all manufacturers except GAC and Glenroe medium elastics where there was no statistically significant difference in breaking force between latex and non latex elastics. Heavy elastics had greater breaking force than medium and medium showed greater breaking force than
light for all types of elastics. Between manufacturers it can be related as Forestadent had greater breaking force than GAC and GAC had greater breaking force than Glenroe. ( Table 2 ).
Table 2: Comparison of breaking force between latex and non latex elastics.
ELASTIC TYPE | LIGHT | MEDIUM | HEAVY | ||||||||||
Mean | S.D | Mean diff | Sig | Mean | S.D | Mean diff | Sig | Mean | S.D | Mean diff | Sig | ||
FORESTADENT | Latex | 1822 | 42.61 | -89.8 | 0.000 | 2391 | 126.4 | -193 | 0.000 | 3663 | 79.59 | -215.2 | 0.000 |
Non latex | 1911 | 46.3 | 2584 | 56.29 | 3878 | 60.43 | |||||||
GAC | Latex | 1163 | 43.08 | -68.6 | 0.000 | 2357 | 38.05 | -1.9 | 0.917 | 3424 | 65.52 | -86.1 | 0.003 |
Non latex | 1232 | 30.97 | 2359 | 42.21 | 3510 | 41.87 | |||||||
GLENROE | Latex | 1414 | 36.08 | -210.4 | 0.000 | 1861 | 47.45 | -42.6 | 0.053 | 2220 | 61.55 | -264.5 | 0.000 |
Non latex | 0.000 | 56.26 | 1904 | 44.64 | 2485 | 61.69 |
P<0.05 = significant
Table shows that there were no statistically significant difference in breaking force in between latex and non latex GAC and Glenroe medium elastics. In the rest of the groups, non latex elastics had significantly greater breaking force than latex elastics. Student independent t- test was used to find significance.
When force degradation was compared Forestadent (light, medium and heavy), GAC (heavy), and Glenroe (heavy) non-latex elastics lost more force than latex elastics. There
was no statistically significant difference in force degradation between latex and non latex elastics in GAC light (at 1,6,12, and 48 hours) GAC medium (at 12 and 24 hrs) and Glenroe light elastics (at 6, 24, and 48 hours). At 24hrs GAC light latex elastics showed more force degradation than non-latex elastics. At 1 hour, 6 hrs and 48hrs the GAC medium non latex elastics showed more force degradation than latex elastics. At 1hour and 12 hrs the Glenroe light non latex elastics showed more force degradation than latex elastics Glenroe medium latex elastics showed more force degradation than medium non latex elastics. The results showed that the force degradation for all manufacturers latex and non-latex heavy elastics showed maximum force degradation than medium and medium showed more force degradation than light in 1, 6, 12, 24 & 48 hours. Between manufacturers it can be related as Forestadent had lesser force degradation than Glenroe and Glenroe had lesser force degradation than GAC. (Table 3 , 4 )
Table 3: Load relaxation comparison between different elastics.
ELASTIC TYPE | LOAD AT 1 HOUR | LOAD AT 6 HOUR | LOAD AT 12 HOUR | LOAD AT 24 HOUR | LOAD AT 48 HOUR | ||||||
Mean diff | sig. | Mean diff | sig. | Mean diff | sig. | Mean diff | sig. | Mean diff | sig. | ||
FORESTADENT LATEX | LIGHT vs MEDIUM | -37.74 | 0.000 | -37.13 | 0.000 | -36.78 | 0.000 | -34.54 | 0.000 | -31.46 | 0.000 |
LIGHT vs HEAVY | -108.36 | 0.000 | -103.1 | 0.000 | -96.15 | 0.000 | -90.06 | 0.000 | -86.47 | 0.000 | |
LIGHT vs MEDIUM | -70.61 | 0.000 | -65.16 | 0.000 | -59.37 | 0.000 | -55.51 | 0.000 | -55.01 | 0.000 | |
FORESTADENT NON LATEX | LIGHT vs MEDIUM | -33.21 | 0.000 | -33.18 | 0.000 | -32.56 | 0.000 | -31.43 | 0.000 | -30.11 | 0.000 |
LIGHT vs HEAVY | -126.95 | 0.000 | -119.29 | 0.000 | -117.28 | 0.000 | -113.96 | 0.000 | -107 | 0.000 | |
LIGHT vs MEDIUM | -93.74 | 0.000 | -86.11 | 0.000 | -84.72 | 0.000 | -82.52 | 0.000 | -76.88 | 0.000 | |
GAC LATEX | LIGHT vs MEDIUM | -20.56 | 0.000 | -17.55 | 0.000 | -17.22 | 0.000 | -14.65 | 0.000 | -14.51 | 0.000 |
LIGHT vs HEAVY | -50.96 | 0.000 | -46.94 | 0.000 | -44.18 | 0.000 | -39.97 | 0.000 | -40.06 | 0.000 | |
LIGHT vs MEDIUM | -30.4 | 0.000 | -29.38 | 0.000 | -26.16 | 0.000 | -25.31 | 0.000 | -25.55 | 0.000 | |
GAC NON LATEX | LIGHT vs MEDIUM | -23.11 | 0.000 | -21.17 | 0.000 | -19.18 | 0.000 | -17.79 | 0.000 | -18.91 | 0.000 |
LIGHT vs HEAVY | -58.81 | 0.000 | -54.04 | 0.000 | -50.35 | 0.000 | -48.36 | 0.000 | -46.78 | 0.000 | |
LIGHT vs MEDIUM | -35.7 | 0.000 | -32.86 | 0.000 | -31.17 | 0.000 | -30.56 | 0.000 | -27.87 | 0.000 | |
GLENROE LATEX | LIGHT vs MEDIUM | -28.4 | 0.000 | -27.51 | 0.000 | -26.58 | 0.000 | -24.73 | 0.000 | -23.98 | 0.000 |
LIGHT vs HEAVY | -46.96 | 0.000 | -43.48 | 0.000 | -41.35 | 0.000 | -40.39 | 0.000 | -37.48 | 0.000 | |
LIGHT vs MEDIUM | -18.56 | 0.000 | -15.97 | 0.000 | -14.77 | 0.000 | -15.66 | 0.000 | -13.5 | 0.000 | |
GLENROE NON LATEX | LIGHT vs MEDIUM | -20.21 | 0.000 | -21.9 | 0.000 | -21.05 | 0.000 | -18.79 | 0.000 | -18.21 | 0.000 |
LIGHT vs HEAVY | -54.77 | 0.000 | -51.85 | 0.000 | -48.18 | 0.000 | -45.24 | 0.000 | -42.39 | 0.000 | |
LIGHT vs MEDIUM | -34.55 | 0.000 | -29.95 | 0.000 | -27.12 | 0.000 | -26.44 | 0.000 | -24.18 | 0.000 |
P<0.05 = significant
Table shows that Heavy elastics had greater force loss compared to both Medium and Light elastics and Medium elastics had greater force loss than light elastics. Tukey-HSD test was done to find significance.
Table 4: Load relaxation comparison between manufacturers.
ELASTIC TYPE | LOAD AT 1 HOUR | LOAD AT 6 HOUR | LOAD AT 12 HOUR | LOAD AT 24 HOUR | LOAD AT 48 HOUR | ||||||
Mean diff | sig. | Mean diff | sig. | Mean diff | sig. | Mean diff | sig. | Mean diff | sig. | ||
LATEX LIGHT | Forestadent vs Glenroe | 6.62 | 0.000 | 5.97 | 0.000 | 5.77 | 0.000 | 5.04 | 0.000 | 5.78 | 0.000 |
Forestadent vs GAC | 8.99 | 0.000 | 7.01 | 0.000 | 6.87 | 0.000 | 5.6 | 0.000 | 7.21 | 0.000 | |
Glenroe vs GAC | 2.36 | 0.003 | 1.03 | 0.284 | 1.1 | 0.341 | 0.55 | 0.595 | 1.42 | 0.031 | |
LATEX MEDIUM | Forestadent vs Glenroe | 15.97 | 0.000 | 15.6 | 0.000 | 15.97 | 0.000 | 14.85 | 0.000 | 13.26 | 0.000 |
Forestadent vs GAC | 26.17 | 0.000 | 26.6 | 0.000 | 26.43 | 0.000 | 25.49 | 0.000 | 24.16 | 0.000 | |
Glenroe vs GAC | 10.2 | 0.000 | 10.99 | 0.000 | 10.46 | 0.000 | 10.63 | 0.000 | 10.9 | 0.000 | |
LATEX HEAVY | Forestadent Vs Glenroe | 68.02 | 0.000 | 65.61 | 0.000 | 60.56 | 0.000 | 54.7 | 0.000 | 54.77 | 0.000 |
Forestadent vs GAC | 66.38 | 0.000 | 63.18 | 0.000 | 58.83 | 0.000 | 55.69 | 0.000 | 53.62 | 0.000 | |
Glenroe vs GAC | -1.64 | 0.185 | -2.43 | 0.003 | -1.73 | 0.131 | 0.986 | 0.431 | -1.15 | 0.139 | |
NON LATEX LIGHT | Forestadent vs Glenroe | 11.78 | 0.000 | 12.02 | 0.000 | 10.56 | 0.000 | 9.7 | 0.000 | 10.43 | 0.000 |
Forestadent vs GAC | 16.18 | 0.000 | 15.03 | 0.000 | 13.63 | 0.000 | 12.95 | 0.000 | 13.02 | 0.000 | |
Glenroe vs GAC | 4.4 | 0.000 | 3 | 0.002 | 3.07 | 0.000 | 3.25 | 0.000 | 2.59 | 0.000 | |
NON LATEX MEDIUM | Forestadent vs Glenroe | 24.78 | 0.000 | 23.32 | 0.000 | 22.07 | 0.000 | 22.34 | 0.000 | 22.32 | 0.000 |
Forestadent vs GAC | 26.27 | 0.000 | 27.06 | 0.000 | 27 | 0.000 | 26.59 | 0.000 | 24.22 | 0.000 | |
Glenroe vs GAC | 1.49 | 0.140 | 3.73 | 0.000 | 4.93 | 0.000 | 4.25 | 0.000 | 1.89 | 0.014 | |
NON LATEX HEAVY | Forestadent vs Glenroe | 83.96 | 0.000 | 79.47 | 0.000 | 79.66 | 0.000 | 78.42 | 0.000 | 75.03 | 0.000 |
Forestadent vs GAC | 84.32 | 0.000 | 80.27 | 0.000 | 80.56 | 0.000 | 78.55 | 0.000 | 73.23 | 0.000 | |
Glenroe vs GAC | 0.355 | 0.921 | 0.8 | 0.742 | 0.89 | 0.405 | 0.128 | 0.985 | -1.79 | 0.042 |
P<0.05 = significant
Table shows that force degradation is less in Forestadent when compared to Glenroe and GAC. When compared between Glenroe and GAC, force degradation is less in Glenroe except in latex heavy elastics where GAC had less force degradation than Glenroe. Tukey-HSD test was used to find the significance level.
Percentage of load remaining was significantly greater in latex elastics than non-latex elastics. When compared between light, medium and heavy elastics the percentage of load remaining was significantly greater in light elastics. When compared between manufacturers the percentage of load remaining was significantly greater in Forestadent than Glenroe and GAC. Force degradation within the first hour was in the range of 14% - 21%. At 48 hrs around 26% - 35% of the force was lost. When stretched to 2 times the internal diameter the elastics generated forces lower than or equal to the manufacturers specified values, where as when the elastics stretched to 3 times the internal diameter, the forces generated were larger than or equal to the manufacturers specified values. ( Table 5 )
Table 5: Comparison of mean experimental loads with 3 I.D load values.
ELASTIC TYPE | % LOAD AT 1 HR | % LOAD AT 6 HR | % LOAD AT 12 HR | % LOAD AT 24 HR | % LOAD AT 48 HR |
Forestadent Latex Light | 86 | 81 | 78 | 75 | 74 |
Forestadent Latex Medium | 84 | 80 | 78 | 74 | 71 |
Forestadent Latex Heavy | 82 | 77 | 73 | 69 | 67 |
Forestadent Non Latex Light | 84 | 79 | 75 | 71 | 70 |
Forestadent Non Latex Medium | 82 | 78 | 75 | 72 | 70 |
Forestadent Non Latex Heavy | 80 | 75 | 73 | 70 | 67 |
GAC Latex Light | 85 | 81 | 78 | 76 | 73 |
GAC Latex Medium | 82 | 76 | 74 | 70 | 68 |
GAC Latex Heavy | 79 | 75 | 71 | 68 | 66 |
GAC Non Latex Light | 83 | 78 | 75 | 72 | 70 |
GAC Non Latex Medium | 82 | 76 | 72 | 69 | 68 |
GAC Non Latex Heavy | 79 | 74 | 70 | 67 | 65 |
Glenroe Latex Light | 84 | 79 | 76 | 74 | 72 |
Glenroe Latex Medium | 83 | 79 | 76 | 73 | 71 |
Glenroe Latex Heavy | 80 | 75 | 72 | 70 | 67 |
Glenroe Non Latex Light | 82 | 76 | 73 | 71 | 68 |
Glenroe Non Latex Medium | 81 | 78 | 75 | 71 | 68 |
Glenroe Non Latex Heavy | 81 | 76 | 72 | 68 | 65 |
Table shows the comparison of forces generated by the elastic when stretched to 19.05mm at 0 hour with the force generated by elastics at 1hour, 6hours, 12hours, 24hours and 48hours. % of load remaining is greater in latex compared to non latex elastics. When compared between manufacturers, the % of load remaining was significantly greater in Forestadent elastics than Glenroe and GAC.
Discussion
Australian standards for latex orthodontic elastic bands which state that the breaking strength of the elastics must be greater than 150 Kpa and the extensions at which failures occur must be at least 750% of the resting ID. Thus
according to the Standards Association of Australia all of
the elastics in the study were strong enough and extendable enough to meet clinical requirements. 3 , 4
Cross-sectional area showed that there was wide variation in the morphology of elastics of the same manufacturer and same elastic type. The variation in cross sectional area could have clinical ramifications like varied forces being applied by the same type of elastics when stretched for the same distance. The heavy elastics had larger cross-sectional areas than medium elastics followed by light elastics. This is in concurrence with the study by Russell et al. 5 But there was no statistically significant difference in the internal diameter between different type of elastics.
The present study revealed that when stretched to 2 times the inner diameter, the elastics generated forces equal to or less than that of the manufacturers specified force values. Whereas, when stretched to 3 times the inner diameter, the elastics generated forces which were equal to or higher than the manufacturers specified force values. These findings were in concurrence with the results of Bales, Russell et al and Kanchana and Godfrey, who reported that the force exerted at the manufacturers recommended extension of 3 times the inner diameter was greater than stated by the manufacturer and that extensions of only 2 times the inner diameter gave clinically appropriate force levels. 6 - 9
In the breaking force comparisons between different elastic forces, the heavy elastics showed greater breaking forces than medium and medium showed greater breaking forces than light for different elastic types. This is in concurrence with the study by Russell et al. 5 Breaking force of non latex elastics were significantly greater than latex elastics in all groups except in GAC medium elastics where there was no statistical significant difference between the latex and nonlatex elastics. In the clinical context, this study shows that there are very less chances for the elastics to break during their application in the mouth. In the force degradation comparisons of elastics used in this study, the non-latex elastics lost more force than the latex elastics for different types of elastics of all manufacturers. This was in agreement with the studies of Andreasen and Bishara, Wong and Russell et al. Berman 3 , 8 , 10 The percentage force degradation values were in agreement with recent studies by Kanchana and Godfrey, Kersyand Russell et al. 3 , 6 , 11 The probable reason for the force degradation values in this study to be low could be attributed to the more accurate measuring methods in the tests and advances in the manufacturing process. There was little difference in the force degradation values mentioned in the study by Russell et al 5 for GAC elastics, which were used in this study. This could be attributed to the different force values for light, medium and heavy elastics in this study.
Liu et al 12 stated that the normal range of clinical use during talking and chewing is between 20mm and 50mm. Hence in our study, for load relaxation mechanical testing the elastics were mounted between two stainless steel pins of 1mm on acrylic board; the pins were set apart at a fixed distance of 19.05mm, so as to stretch the elastics to three times the marketed internal diameter. 2 The force degradation measurements were recorded by a custom built set up used in the study by Tong Wang. 13 The media in which elastics have been tested vary considerably. As early as Paulich noted that the initial force decay of elastics depended on the environment in which elastics were tested. Overall, elastics and non-latex chains have been tested at room temperature and at 37 o C, in dry and moist air, in distilled water, normal saline and artificial saliva. The elastics in this study were tested in artificial saliva at 37 o C, which was maintained by a thermostat. 14 , 15 The findings of Ash and Nikolai 16 were that saliva, especially in an in-vivo environment, had a significant effect. Ferriter evaluated the effect of pH on the force-degradation rates of chain elastics and stated that the decay rate of orthodontic polyurethane chain elastics is inversely proportional to the oral pH, with a corollary that basic ph levels are most hostile to polyurethane chain elastics. 17
In the force degradation comparisons between different elastics, all manufacturers latex and non-latex heavy elastics showed more force degradation than medium and medium showed more force degradation than light in 1 hour, 6 hours, 12 hours, 24 hours and 48 hours except Glenroe non-latex light elastics which lost more force than medium. This result was in agreement with the study by Russell et al 5 and Andreasen and Bishara. 18 The greater force degradation of heavy elastic as compared to medium and medium elastics as compared to light could be attributed to the increased cross sectional area of the heavy elastics. The increased surface area in turn increases the possibility of the elastics contact with air or water leading to increased degradation.
In comparisons between elastics of different manufacturers Forestadent elastics showed less force degradation than GAC and GAC showed less force degradation than Glenroe in all types of elastics at 1, 6, 12, 24 & 48 hours. This showed that Forestadent elastics were better in
delivering force over a period of 48 hours. The percentage of force degradation of different elastics at 1, 6, 12, 24 & 48 hours is almost in agreement with the values given by Kanchana and Godfrey 6 and Mckersy 11 in their study. The mild variation could be because of the different manufacturers’ and stretches used in their study.
This study shows that clinically, latex elastics are better in force delivery over a period of 48 hours. Non-latex elastics were found to lose more force than latex elastics. Hence non-latex elastics are to be used in situations where the patient is allergic to latex products. It is better to choose elastics which deliver a higher load; around 25% to 35% more load than that desired for the clinical situation. Clinically, the initial forces generated will be used for overcoming frictional forces of the wire and the bracket. The forces generated by the elastic on loading are not
completely transferred onto the tooth; moreover, the maximum force degradation takes place within the first hour.
It was observed that within the first 24 hours, around 25% – 33% of force was lost. In the next 24 hours, only around 1% – 3% of force was lost in all the elastics. In comparison with the force levels at the end of 24 hours and 48 hours, a difference of only 1% - 3% was seen. This implies that force was relatively stable between the first day and the next day and only a negligible amount was lost. Hence changing elastics at the end of two days produces almost the same amount of force as it would if it was changed in one day.
Conclusion
Non latex elastics had greater cross sectional area than latex elastics in all types of elastics (Forestadent light, medium, heavy, GAC medium, heavy, Glenroe light, medium, heavy) except GAC light elastics where there was no statistical difference between latex and non latex elastics.
Over a period of 48 hours, there was a decrease in the loads generated by all elastics. The amount of force that was retained at the end of one day and two days was not significantly different for both latex and non latex elastics. However, the latex elastics retained larger loads than the non latex elastics and Forestadent elastics retained larger load than GAC and Glenroe. Force degradation was higher in the heavy elastics when compared with the medium and light elastics.
Latex elastics are the preferred choice, except when patients exhibit allergic manifestations to latex. Further research is needed with the aim of producing elastics with identical internal diameter, cross sectional area and force degradation parameters. Manufacturer induced variations in the dimensions of these elastics has to be minimized to produce uniform force levels irrespective of the brand the clinician chooses for his practice. In our study, force degradation was evaluated for a period of 2 days, hence further studies are needed to evaluate the amount of force
that is retained over a period of time and also to check whether a clinically desirable amount of force is delivered. This would be useful in determining when the elastic should be changed in clinical situations. Clinical studies are necessary to assess the different elastic types’ behaviour in terms of force degradation and stability in oral environment.
Footnotes
Source of Support: Nil
Conflict of Interest: None
Contributor Information
Supradeep Kumar Kamisetty, Department of Orthodontics & Dentofacial Orthopaedics, St. Joseph Dental College, Eluru, Andhra Pradesh, India.
Chakrapani Nimagadda, Department of Orthodontics & Dentofacial Orthopaedics, St. Joseph Dental College, Eluru, Andhra Pradesh, India.
Madhoom Ponnachi Begam, Chennai,Tamilnadu, India.
Raghuveer Nalamotu, Department of Orthodontics & Dentofacial Orthopaedics, St. Joseph Dental College, Eluru, Andhra Pradesh, India.
Trilok Srivastav, Department of Orthodontics & Dentofacial Orthopedics, People’s Dental Academy, Bhopal, Madhya Pradesh, India.
Shwetha GS, Department of Orthdontics & Dentofacial Orthopedics, K.L.E institute of Dental Sciences, Bangalore, Karnataka, India.
References
- 1.AK Wong. Orthodontic elastic materials. Angle Orthod. 1976;46:196–205. doi: 10.1043/0003-3219(1976)046<0196:OEM>2.0.CO;2. [DOI] [PubMed] [Google Scholar]
- 2.EJ Neiburger. A Case of Possible Latex Allergy. J Clin Orthod. 1991;25:559–560. [PubMed] [Google Scholar]
- 3.ADA council on scientific affairs. The dental team and latex hypersensitivity. J Am Dent Assoc. 1999;130:257–264. doi: 10.14219/jada.archive.1999.0177. [DOI] [PubMed] [Google Scholar]
- 4.CC Liu, JC Wataha, RG Craig. The effect of repeated stretching on the force decay and compliance of vulcanized cis-polyisoprene orthodontic elastics. Dent Mater. 1993;9:37–40. doi: 10.1016/0109-5641(93)90103-w. [DOI] [PubMed] [Google Scholar]
- 5.KA Russell, AD Milne, RA Kiianna, JM Lee. In vitro assessment of the mechanical properties of latex and non-latex orthodontic elastics. Am J Orthod Dentofacial Orthop. 2001;120:36–44. doi: 10.1067/mod.2001.114642. [DOI] [PubMed] [Google Scholar]
- 6.P Kanchana, K Godfrey. Calibration of force extension and force degradation characteristics of orthodontic latex elastic. Am J Orthod Dentofac Orthop. 2000;118:280–287. doi: 10.1067/mod.2000.104493. [DOI] [PubMed] [Google Scholar]
- 7.WR Bell. A study of applied force as related to the use of elastics and coil springs. Angle Orthod. 1951;21:151–154. doi: 10.1043/0003-3219(1951)021<0151:ASOAFA>2.0.CO;2. [DOI] [PubMed] [Google Scholar]
- 8.S Bishara, GF Andreasen. A comparison of time related forces between plastic alastiks and latex elastics. Angle Orthod. 1970;40:319–328. doi: 10.1043/0003-3219(1970)040<0319:ACOTRF>2.0.CO;2. [DOI] [PubMed] [Google Scholar]
- 9.WP Rock, HJ Wilson, SE Fisher. Force reduction of orthodontic elastomeric chains after one month in the mouth. Br J Orthod. 1986;13:147–150. doi: 10.1179/bjo.13.3.147. [DOI] [PubMed] [Google Scholar]
- 10.R Berman. The Application of Elastic Forces. J Clin Orthod. 1969;3:404–409. [PubMed] [Google Scholar]
- 11.Ml Kersey, K Glover, G Heo, D Raboud, PW Major. Comparison of Dynamic and Static Testing of Latex and Nonlatex Orthodontic Elastics. Angle Orthod. 2003;73(2):181–186. doi: 10.1043/0003-3219(2003)73<181:ACODAS>2.0.CO;2. [DOI] [PubMed] [Google Scholar]
- 12.MT O'Reilly, DJ Rinchuse, J Close. Class II elastics and extractions and temporomandibular disorders: A longitudinal prospective study. Am J Orthod Dentofacial Orthop. 1993;103(5):459–463. doi: 10.1016/S0889-5406(05)81797-4. [DOI] [PubMed] [Google Scholar]
- 13.T Wang, G Zhou, X Tan, Y Dong. Evaluation of Force Degradation Characteristics of Orthodontic Latex Elastics in Vitro and In Vivo. Angle Orthod. 2007;77(4):688–693. doi: 10.2319/022306-76. [DOI] [PubMed] [Google Scholar]
- 14.RJ Egolf, EA Beole, HS Upasaw. Factors associated with orthodontic patient compliance with intraoral elastic and headgear wear. Am J Orthod Dentofacial Orthop. 1990;97(4):336–348. doi: 10.1016/0889-5406(90)70106-M. [DOI] [PubMed] [Google Scholar]
- 15.JP Ferriter, Jr Meyers CE, L Lorton. The effect of hydrogen ion concentration on the force-degradation rate of orthodontic polyurethane chain elastics. Am J Orthod Dentofacial Orthop. 1990;98(5):404–410. doi: 10.1016/S0889-5406(05)81648-8. [DOI] [PubMed] [Google Scholar]
- 16.J Ash, R Nikolai. Relaxation of orthodontic elastic chains in vitro and in vivo. JDent Res. 1978;57:685–690. doi: 10.1177/00220345780570050301. [DOI] [PubMed] [Google Scholar]
- 17.DM Killiany, J Duplessis. Relaxation of Elastomeric Chains. J Clin Orthod. 1985;19:592–593. [PubMed] [Google Scholar]
- 18.GF Andreasen, S Bishara. Comparison of Alastic chain with elastics involved with intra-arch molar to molar forces. Angle Orthod. 1970;40:151–158. doi: 10.1043/0003-3219(1970)040<0151:COACWE>2.0.CO;2. [DOI] [PubMed] [Google Scholar]