Abstract
Existing devices for early infant male circumcision (EIMC) have inherent limitations. We evaluated the newly developed AccuCirc device by circumcising 151 clinically well, full-term male infants with birth-weight ≥ 2.5 kg within the first 10 days of life from a convenience sample in two hospitals in Botswana. No major adverse events were observed. There was one local infection, five cases of minor bleeding and one case of moderate bleeding. In three cases the device made only partial incisions that were completed immediately by the provider without complications. Parental satisfaction was high: > 96% of mothers stated they would circumcise a future son. The pre-assembled, sterile AccuCirc kit has the potential to overcome obstacles related to supply chain management and on-site instrument disinfection that can pose challenges in resource-limited settings. In our study the AccuCirc was safe and it should be considered for programmatic EIMC in resource-limited settings.
BACKGROUND
Randomized trials have provided compelling evidence that male circumcision (MC) reduces men’s risk of heterosexually-acquired HIV-1 infection by about 60% [1–3] and it has been shown to be a cost-saving intervention in sub-Saharan Africa [4, 5]. Although there have been no prospective randomized trials specifically demonstrating a protective effect of MC against HIV when done before adulthood, observational studies and models suggest that the cost-benefit and safety appear to be greatest when the procedure is done in early infancy [6–9]. The World Health Organization (WHO) recommends MC as a component of comprehensive HIV prevention programming and identifies neonatal circumcision as a lower-risk procedure with fewer complications than circumcision performed at older ages. WHO recommends that “countries should consider how to promote neonatal circumcision in a safe, culturally acceptable and sustainable manner” [8].
Specialized devices are required for safe early infant male circumcision (EIMC, performed ≤ 60 days of life [10]), and it is important to provide programs with options for devices that suit local needs so that this evidence-based HIV prevention strategy can be optimized and successfully implemented in routine public health settings. In 2010, WHO published the Manual for Early Infant Male Circumcision under Local Anaesthesia [10], which included the Mogen, Gomco and Plastibell devices. At the time of publication, the AccuCirc was a new EIMC device (approved by the United States Food and Drug Administration in 2007) and it had not yet undergone field evaluation to assess its performance in resource-limited settings.
Although published reports and our own studies have shown a very low rate of serious adverse events associated with EIMC [11–15], we believe the adverse event rate could be decreased. There are rare but serious potential complications associated with all three devices currently on the WHO prequalified list of EIMC devices: 1) use of the Mogen clamp can result in partial or total amputation of the glans penis or removal of too little foreskin [10, 13, 16–18]; 2) migration of the Plastibell can result in necrosis of the glans and other injuries, and risk is increased if the incorrect size “bell” is utilized [10, 14, 19, 20], and 3) mismatching the sizes of the separate pieces of the Gomco clamp can result in laceration of the glans penis [10, 16]. Sequelae of these rare adverse outcomes may be worse in settings with limited resources for treating such complications.
The potential complications of these three devices are inherent in their design. Although training and increased experience of the providers can reduce the risk of complications, the design of the devices themselves will always present the potential for adverse events. The AccuCirc has several safety features that make an adverse event resulting from the device itself extremely unlikely. It is a onetime use device that has a built-in shield to protect the glans from laceration and it incorporates a retractable blade that cannot be reused. We therefore conducted a single-arm field-evaluation of the AccuCirc device in Botswana to assess safety and parental satisfaction.
METHODS
Informed Consent, Safety Monitoring, and Role of the Funding Source
Written informed consent was obtained from the mothers or legal guardians (if mothers were age <21 years, the age of majority in Botswana) before neonates were enrolled. The single-arm study was approved by the Botswana Ministry of Health’s Health Research and Development Committee (HRDC) and by Brigham and Women’s Hospital Institutional Review Board (IRB). An independent Data Safety Monitoring Committee was responsible for safety monitoring. This trial is registered with ClinicalTrials.gov with the number NCT00971958. The sponsors of the study had no role in study design, data collection, data analysis, data interpretation, or writing of the report. The corresponding author had full access to all the data in the study and had final responsibility for the decision to submit for publication. The AccuCirc devices used in this evaluation were donated by Clinical Innovations, Murray, UT, USA.
Subjects
Mothers and infants were enrolled between February and September 2011 at two hospitals in southeastern Botswana: one in the capital city, Gaborone, and one in the village of Molepolole. A convenience sample of eligible mothers and infants were identified and recruited by study staff in maternity wards in the respective hospitals. Maternal eligibility criteria were Botswana citizenship, non-incarceration, and ability to follow-up at seven days and six weeks after the procedure. Infant eligibility criteria were male gender, gestational age ≥37 weeks, age ≤10 days of life, birth-weight ≥2,500 grams, no evidence of neonatal sepsis or other illness requiring hospitalization, no family history of bleeding disorder, and no genital or other abnormality that is a contraindication to EIMC. The maternal HIV status was not one of the inclusion/exclusion criteria and thus was not recorded for the purposes of this study.
AccuCirc Device, Kit and Procedure
The AccuCirc comes in a sterile pack that contains the device itself, a surgical pen, drape, two hemostats, three betadine swabs, xeroform dressing, gauze and cleansing wipe and all items are disposable. The device itself consists of 1) a flexible foreskin probe and shielding ring and 2) a single-action clamp that contains a circular blade. It is available in two sizes: 1.1 cm and 1.3 cm (penile diameter). Complete manufacturer instructions are available on line [21]. A surgical pen-mark is made on the foreskin at the level of the corona. The physiologic adhesions between the foreskin and the glans are removed using the flexible foreskin probe. With the use of hemostats, the shielding ring is then inserted into the preputial space, between the foreskin and the glans. The adjustable foreskin holder is deployed, and using a hemostat, the tissue is aligned according to the pen-mark within the inner and outer rings. The clamp is applied and activated to deliver a circumferential, hemostatic crush while simultaneously incising the foreskin. After 5 minutes of hemostatic pressure, the clamp is released and the wound is dressed (Figure 1).
Figure 1.

A) a surgical pen-mark (indicated by ---) is made on the foreskin at the level of the corona. Adhesions are removed using the flexible foreskin probe. B) with the use of hemostats, the shielding ring is then inserted into the preputial space, between the foreskin and the glans. C) the adjustable foreskin holder is deployed, and D) using a hemostat, the tissue is aligned according to the pen-mark within the inner and outer rings. E) the clamp is applied and F) activated to deliver a circumferential, hemostatic crush while simultaneously incising the foreskin. G) after 5 minutes of pressure, the clamp is released and the wound is dressed.
Study Intervention
All circumcisions were performed by a study physician (NN, RA or RP) using clean techniques (sterile kit and new, non-sterile gloves, in a clean clinical setting but not a surgical theater) [12, 22]. The study physicians, who had no significant prior experience in performing EIMC with the AccuCirc device, were trained by an AccuCirc-experienced physician prior to study initiation and only performed procedures independently after competency was demonstrated.
Study staff performed a physical examination to exclude infants with abnormalities precluding circumcision. Infants had approximately one gram of eutectic mixture of local anesthetic (EMLA) applied to the outer foreskin and shaft of the penis and covered with an occlusive dressing (the pros and cons of various local anesthetics approved for use for EIMC have been previously described [23] (it should be noted EMLA should not be used on the mucosal/inner surface of the foreskin because of higher absorption possibly leading to systemic toxicity); EMLA was wiped off prior to the procedure. During the procedure, infants were given concentrated sucrose solution either by gloved finger or by syringe to augment the anesthetic effect of EMLA, as has been previously described [24–26]. Infants’ upper bodies and arms were swaddled, and legs restrained by an assistant. Infants had the genital area, lower abdomen and upper legs cleansed with betadine immediately prior to EIMC.
After circumcision, infants were checked for post-procedure bleeding or other immediate complications. Mothers were asked to wait with the infant for two hours following the procedure prior to leaving the facility to ensure there was no excessive bleeding and to ensure that the infant had urinated. Mothers were given Vaseline for wound care and written post-procedure care instructions that included a phone number to reach a study physician 24 hours a day, seven days a week, in case of any questions or concerns.
Follow-up and Evaluation
Post-procedure follow-up visits were scheduled at seven days and six weeks. At these visits, parents were asked about complications and child outpatient medical cards were also reviewed for evidence of visits to non-study clinics or providers for adverse events potentially related to the procedure. The infants also had a physical exam, including inspection of the circumcision site, by one of the three study physicians.
Outcomes
The primary study outcomes were adverse events and parental satisfaction. We also measured the time taken to perform the procedure from the infant’s diaper being removed until the time it was replaced, and recorded the number of AccuCirc devices opened so that correct sizing was achieved.
Adverse events that were potentially related to EIMC were defined prior to study initiation and categorized as bleeding, infection, structural and other. Minor adverse events included bleeding requiring only the application of pressure beyond the immediate post-procedure period, or infection limited to the circumcision site. Moderate adverse events included intervention for bleeding other than pressure (e.g. suturing), or soft-tissue infection spreading beyond the penis (lower abdominal wall, upper legs). Major adverse events included the need for a separate clinic visit or hospitalization for bleeding from the circumcision site; need for IV fluids or blood products; systemic infection/sepsis; removal of too much or incorrect tissue; removal of too little tissue requiring repeat procedure; structural injury to the glans, urethra or shaft; or subsequent problems with urination.
To assess parental satisfaction we used a structured questionnaire that was administered at the six-week follow-up visit. “Highly or completely” was defined as a score of 90% or greater on a visual analogue scale from 0 – 100%. Parents were also asked if they would be likely to choose EIMC for a future son.
Sample Size
This was a pilot study aiming to explore the performance of the AccuCirc in resource-limited settings. This study was not powered to detect a statistically significant difference in major adverse events when comparing the AccuCirc with other devices (previously reported rates of major complications following EIMC are as low as 0.2% in well-resourced settings) [11, 12, 15].
Statistical Analysis
Frequency tables for each of the primary outcomes present proportions, means, medians, standard deviations, and interquartile ranges as appropriate. All statistical analyses were performed with SAS software version 9.3 (SAS Institute, Cary, NC).
RESULTS
Study Participants
We circumcised 151 infants at a median age of 2 days (IQR: 2–4 days) and a median birth weight of 3.13 kg (IQR: 2.95–3.42 kg). On average, study participants were born at an estimated gestational age of 39.4 weeks. Three additional infants were enrolled, but did not undergo circumcision: two experienced a local reaction to the topical anesthesia we previously described [23] which precluded continuing with the procedure, and one was found, after the foreskin was retracted but prior to circumcision, to have glandular hypospadias, which made him ineligible for the procedure. We opened 155 kits and four infants required a different size: ultimately we used 117 size 1.1 cm and 34 size 1.3 cm clamps. The average time required to complete a single procedure was 16.5 minutes (SD: 6.7 minutes).
Of the 151 infants circumcised, 150 completed the 7-day follow-up visit; one child did not return in person for any follow-up, but his parents were contacted by telephone and reported no adverse events or complications from the procedure. One hundred and thirty seven infants returned for the 6-week follow-up visit.
Adverse Events
We observed ten adverse events (6.6%), all of which were classified as either minor (n=9) or moderate (n=1) (Table 1). Five cases of minor bleeding occurred (all resolved with brief, local pressure). One case of local infection was diagnosed by a non-study physician and not confirmed by the study team; the infant was prescribed oral amoxicillin and had no further sequelae. In three cases the AccuCirc device made only a partial incision (cut approximately 70% of circumference) requiring manual completion by the study physician. In these three instances, the EIMC was completed as a single procedure: immediately upon realizing that the cut was incomplete, the provider used sterile surgical scissors to incise along the hemostatic crush-line made by the AccuCirc device without further complications. The only moderate adverse event involved prolonged bleeding (90 minutes) immediately following the procedure in an infant who had inadvertently not received Botswana standard of care, prophylactic vitamin K at birth. Bleeding resolved approximately 30 minutes after administration of 2 mg phytomenadione (vitamin K1) intramuscularly. A complete case report describing the event has been previously published [27].
Table 1.
Adverse events and parentala satisfaction following NMC among N=150 infants circumcised with the AccuCirc device from February to September 2011 in southeastern Botswana.
| Outcome (total n with data) | N (%) |
|---|---|
| Adverse Events (n=150) | |
| Bleeding | 6 (4.0) |
| Mild | 5 (3.3) |
| Moderate | 1 (0.7) |
| Severe | 0 (0.0) |
| Local infectionb | 1 (0.7) |
| Structural | 0 (0.0) |
| Other (incomplete incision)c | 3 (2.0) |
|
| |
| Parentala Satisfaction at 6 weeks | |
| Completely or highly satisfied with procedure (n=137) | 125 (91.2) |
| Would recommend highly or completely EIMC (n=133) | 127 (95.5) |
| Would circumcise another son in the future (Y/N) (n=137) | 132 (96.4) |
Only mothers (no fathers) responding
Not confirmed by study team; treated with oral antibiotics
Device made only a partial incision requiring manual completion by the provider.
Parental Satisfaction
Of 137 infants who returned for the 6-week follow-up visit, 91% of mothers reported being highly or completely satisfied with the procedure; 97% stated they would highly or completely recommend the procedure to a friend or relative. More than 96% stated that they would circumcise another son should they have one in the future (Table 1).
DISCUSSION
In this single-arm field evaluation of the AccuCirc device for EIMC in Botswana, we observed no major complications. Minor bleeding (3.3%), which is expected in any surgical procedure, was similar to our previous study of EIMC using the Mogen clamp (3.2%)[14]. Although 13 children were not seen for their second scheduled follow-up visit, they were seen by study staff for their first visit. Although complications from EIMC are rare, many will occur immediately following (bleeding, soft-tissue injury) or within a few days of the procedure (infection), and in evaluating the safety of the AccuCirc, these complications were our main focus. Longer-term follow-up of issues such as unsatisfactory cosmetic result or meatal stenosis (that may be independent of which device is used for EIMC) are beyond the scope of this pilot study but do deserve systematic investigation. Because serious adverse events are very rare after the immediate post-EIMC period, parental education as to what should prompt need for follow-up (further bleeding, any fever, abnormal discoloration or malodor of wound) should ensure optimal care. In our study design, mothers were asked to wait with the infant for two hours following the procedure to ensure there was no excessive bleeding and to ensure that the infant had urinated. Mothers were provided with written instructions along with a phone number, operational 24 hours per day and seven days per week, for the staff physicians in case of participant questions or concerns. This may be a model that could be replicated for scale up of EIMC programs. Parental satisfaction with the procedure was high and > 96% of parents reported being very likely or likely to circumcise another son in the future.
Although the device failed to completely incise the circumference of the foreskin in three instances, no harm resulted to the infants. As the applied pressure is determined by the device mechanics and not by the clinician deploying it, the device may need revision if this is found to occur in larger studies. Alternatively, it could be that the clinician must make sure that the foreskin is evenly distributed along the foreskin holder prior to deploying the blade so as to ensure an even incision.
The potential for incomplete incisions would limit the use of the device to facilities where standard surgical instruments, such as scissors, would be readily available until the issue has been fully resolved.
Ideal device characteristics for scale-up of EIMC programs differ from those needed in adult MC programs. The 60-day age limit for EIMC is based on safety data [10]. Furthermore, as babies get older and heavier, they are more likely to bleed following the procedure and this could negate the sutureless nature of the procedure. Because of the limited window in which the procedure can be performed, the procedure needs to be offered in a sustained fashion (rather than as a mass campaign) and near to where infants are born or receive health care (rather than recruiting men from a large catchment area to come to a central location where dozens of procedures can be done per day). EIMC programs therefore require device characteristics suited to rural maternity clinics or home-based service provision, where procedures will likely be performed by non-physicians and where moderate or major complications must be minimized. Our findings suggest the safety characteristics of the AccuCirc device may be well-suited for EIMC.
The AccuCirc device could also simplify supply chain management and eliminate the need for reusable instrument inventory and sterilization at remote sites because it comes in a sterile pack containing necessary supplies together with the device [28]. In contrast, use of the Plastibell places an additional burden on supply chain management because all six sizes must be continually stocked, and use of the Mogen and Gomco clamps require reliable and timely instrument disinfection processes.
We would like to emphasize that the most important criteria as EIMC is operationalized is that the infant must be clinically well before undergoing the procedure. Unfortunately neonatal mortality continues to be unacceptably high in many parts of the world and although the EIMC procedure itself is not likely to result in major morbidity or mortality, it is important that such events not be erroneously ascribed to EIMC as the procedure is scaled-up. We believe that each program will have to contextually customize inclusion criteria, although being clinically well should always be a criterion. Context-specific programmatic determinations will have to be made regarding timing of EIMC for those infants who fail to meet programmatic inclusion criteria at birth (e.g. ill infants, low birth weight, prematurity).
In summary, we safely and acceptably circumcised 151 infants in Botswana with the AccuCirc device. If additional field studies confirm our findings [29], the AccuCirc should be considered for WHO pre-qualification to increase options for safe and sustainable scale-up of EIMC in resource-limited settings.
Acknowledgments
Source of Funding: Grant U2GPS000941; NIH grant K23 AI084579
Footnotes
Conflicts of Interest: None
Contributor Information
Rebeca M Plank, Division of Infectious Diseases, Brigham and Women’s Hospital, Department of Medicine.
Kathleen E Wirth, Harvard School of Public Health.
Nnamdi O Ndubuka, Northern Inter-Tribal Health Authority, Prince Albert, Saskatchewan, Canada.
Rasak Abdullahi, Botswana-Harvard School of Public Health AIDS Initiative Partnership for HIV Research and Education, Gaborone, Botswana.
Maggie Nkgau, Botswana Ministry of Health.
Chiapo Lesetedi, Botswana Ministry of Health.
Kathleen M Powis, Departments of Internal Medicine and Pediatrics, Massachusetts General Hospital.
Mompati Mmalane, Botswana-Harvard School of Public Health AIDS Initiative Partnership for HIV Research and Education, Gaborone, Botswana.
Joseph Makhema, Botswana-Harvard School of Public Health AIDS Initiative Partnership for HIV Research and Education, Gaborone, Botswana.
Roger Shapiro, Department of Immunology and Infectious Diseases, Harvard School of Public Health, Division of Infectious Diseases, Beth Israel Deaconess Medical Center.
Shahin Lockman, Division of Infectious Diseases, Brigham and Women’s Hospital; Department of Immunology and Infectious Diseases, Harvard School of Public Health.
References
- 1.Auvert B, et al. Randomized, controlled intervention trial of male circumcision for reduction of HIV infection risk: the ANRS 1265 Trial. PLoS Med. 2005;2(11):e298. doi: 10.1371/journal.pmed.0020298. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Bailey RC, et al. Male circumcision for HIV prevention in young men in Kisumu, Kenya: a randomised controlled trial. Lancet. 2007;369(9562):643–56. doi: 10.1016/S0140-6736(07)60312-2. [DOI] [PubMed] [Google Scholar]
- 3.Gray RH, et al. Male circumcision for HIV prevention in men in Rakai, Uganda: a randomised trial. Lancet. 2007;369(9562):657–66. doi: 10.1016/S0140-6736(07)60313-4. [DOI] [PubMed] [Google Scholar]
- 4.White RG, et al. Male circumcision for HIV prevention in sub-Saharan Africa: who, what and when? AIDS. 2008;22(14):1841–50. doi: 10.1097/QAD.0b013e32830e0137. [DOI] [PubMed] [Google Scholar]
- 5.Njeuhmeli E, et al. Voluntary medical male circumcision: modeling the impact and cost of expanding male circumcision for HIV prevention in eastern and southern Africa. PLoS Med. 8(11):e1001132. doi: 10.1371/journal.pmed.1001132. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Binagwaho A, et al. Male circumcision at different ages in Rwanda: a cost-effectiveness study. PLoS Med. 2010;7(1):e1000211. doi: 10.1371/journal.pmed.1000211. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Weiss HA, et al. Complications of circumcision in male neonates, infants and children: a systematic review. BMC Urol. 2010;10:2. doi: 10.1186/1471-2490-10-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.WHO. New Data on Male Circumcision and HIV Prevention: Policy and Programme Implications. 2007 Available from: http://www.who.int/hiv/mediacentre/MCrecommendations_en.pdf.
- 9.UNAIDS. Neonatal and child male circumcision: a global review. UNAIDS; Geneva: 2010. [Google Scholar]
- 10.WHO. Manual for early infant male circumcision under local anaesthesia. WHO Document Production Services; Geneva, Switzerland: 2011. [Google Scholar]
- 11.Wiswell TE, Geschke DW. Risks from circumcision during the first month of life compared with those for uncircumcised boys. Pediatrics. 1989;83(6):1011–5. [PubMed] [Google Scholar]
- 12.Ben Chaim J, et al. Complications of circumcision in Israel: a one year multicenter survey. Isr Med Assoc J. 2005;7(6):368–70. [PubMed] [Google Scholar]
- 13.Young MR, et al. Safety of over Twelve Hundred Infant Male Circumcisions Using the Mogen Clamp in Kenya. PLoS One. 2012;7(10):e47395. doi: 10.1371/journal.pone.0047395. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Plank RM, et al. J Acquir Immune Defic Syndr. 2013. A Randomized Trial of Mogen Clamp versus Plastibell for Neonatal Male Circumcision in Botswana. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Christakis DA, et al. A trade-off analysis of routine newborn circumcision. Pediatrics. 2000;105(1 Pt 3):246–9. [PubMed] [Google Scholar]
- 16.USFDA; M.D. United States Food and Drug Administration, editor. Potential for Injury from Circumcision Clamps. Silver Spring, MD; 2000. [Google Scholar]
- 17.Sherman J, et al. Circumcision: successful glanular reconstruction and survival following traumatic amputation. J Urol. 1996;156(2 Pt 2):842–4. [PubMed] [Google Scholar]
- 18.Patel HI, et al. Genitourinary injuries in the newborn. J Pediatr Surg. 2001;36(1):235–9. doi: 10.1053/jpsu.2001.20062. [DOI] [PubMed] [Google Scholar]
- 19.Bode CO, Ikhisemojie S, Ademuyiwa AO. Penile injuries from proximal migration of the Plastibell circumcision ring. J Pediatr Urol. 2010;6(1):23–7. doi: 10.1016/j.jpurol.2009.05.011. [DOI] [PubMed] [Google Scholar]
- 20.Gee WF, Ansell JS. Neonatal circumcision: a ten-year overview: with comparison of the Gomco clamp and the Plastibell device. Pediatrics. 1976;58(6):824–7. [PubMed] [Google Scholar]
- 21.Clinical Innovations. AccuCirc Single Action Circumcision Device: Instructions for Use. Murray; Utah: 2010. [Google Scholar]
- 22.Naimer SA, Trattner A. Are sterile conditions essential for all forms of cutaneous surgery? The case of ritual neonatal circumcision. J Cutan Med Surg. 2000;4(4):177–80. doi: 10.1177/120347540000400401. [DOI] [PubMed] [Google Scholar]
- 23.Plank RM, et al. Loss of anatomical landmarks with eutectic mixture of local anesthetic cream for neonatal male circumcision. J Pediatr Urol. 2013;9(1):e86–90. doi: 10.1016/j.jpurol.2012.09.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Taddio A. Pain management for neonatal circumcision. Paediatr Drugs. 2001;3(2):101–11. doi: 10.2165/00128072-200103020-00003. [DOI] [PubMed] [Google Scholar]
- 25.Brady-Fryer B, Wiebe N, Lander JA. Pain relief for neonatal circumcision. Cochrane Database Syst Rev. 2004;(4):CD004217. doi: 10.1002/14651858.CD004217.pub2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Blass EM, Hoffmeyer LB. Sucrose as an analgesic for newborn infants. Pediatrics. 1991;87(2):215–8. [PubMed] [Google Scholar]
- 27.Plank RM, et al. Vitamin K Deficiency Bleeding and Early Infant Male Circumcision in Africa. Obstet Gynecol. 2013;122(2 PART 2):503–505. doi: 10.1097/AOG.0b013e31828b2f5c. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Arasaratnam A, Humphreys G. Emerging economies drive frugal innovation. Bull World Health Organ. 2013;91:6–7. doi: 10.2471/BLT.13.020113. [DOI] [PMC free article] [PubMed] [Google Scholar]