Mapping chemical elements on the surface of orthodontic appliance by SEM-EDX

Marcin Mikulewicz 1,A,B,C,D,E,F,G,, Paulina Wołowiec 2,B,C,D,E,F, Izabela Michalak 2,B,C,D,E,F, Katarzyna Chojnacka 2,A,B,C,D,E,F, Wojciech Czopor 3,E,F, Adam Berniczei-Royko 4,D,E,F, Andras Vegh 5,D,E,F, Thomas Gedrange 6,D,E,F,G
PMCID: PMC4043564  PMID: 24857929

Abstract

Background

During orthodontic treatment, the various elements that constitute the fixed appliance undergo different processes. As a result of a change of the surface, elution/coverage of metals on the surface can be observed in the process of corrosion/passivation.

Material/Methods

Scanning electron microscopy with an energy-dispersive X-ray analytical system (SEM-EDX) was used to analyze the composition of stainless steel elements of orthodontic fixed appliances (before and after orthodontic treatment), to obtain the composition of the surface of the elements. The analyzed elements were: brackets (Victory Series APC PLUS 022, 3M Unitek, Monrovia, CA, USA); wires (0.017×0.025, 3M Unitek, Monrovia, CA, USA); and bands (37+, 3M Unitek, Monrovia, CA, USA).

Results

The results showed a decrease of chromium and iron contribution to the surface, with increase of oxygen content in used vs. new elements of the appliance.

Conclusions

Our results confirm the formation of oxides (passivation layer) on the surface of stainless steel as a result of the presence of the orthodontic appliance in patients’ oral cavities.

MeSH Keywords: Corrosion, Microscopy, Electron, Scanning, Orthodontic Appliances, Stainless Steel

Background

During treatment, fixed orthodontic appliances are placed in the oral cavity for around 2–3 years and are exposed to physical and chemical factors that may cause corrosion and passivation [1,2]. As a result of saliva flow, chewing, brushing, and friction between brackets and wires, as well as the effect of acidic drinks, mouthwash, toothpaste, and biofilm formation, the surfaces of the elements of the appliance are modified [3,4]. In stainless steel corrosion, the metals in the alloys undergo dissolution with the formation of ions, which are released into the surrounding environment (saliva) [5]. In passivation, oxides of metals are formed as a layer that protects the alloy from corrosion. As long as the passivation layer is intact, the corrosion does not progress. If the breakdown potential of an alloy is exceeded, the layer of oxides dissolves and corrosion proceeds [6].

Saliva (an electrolyte) provides an environment that supports the corrosion process. Additionally, organic acids produced by microorganisms, as well as enzymes, increase the progress of corrosion in the oral environment [7]. Considering the mechanisms of the process, the following types of corrosion can be distinguished: uniform, pitting, crevice, fretting, intergranular, galvanic, stress, and microbiological [810]. As a result of corrosion, metal ions (e.g., nickel and chromium) are released from the surface of orthodontic alloys, whose carcinogenic, mutagenic, cytotoxic, and allergenic impact can cause undesirable effects [1113]. These metals are bioaccumulated in tissues [14]. Apart from these biological results, another consequence of metal ion release from the alloy surface is damage to the surface (increase in roughness and the formation of craters and pits), which leads to weakening of the structure’s strength and may result in cracks or fractures of the elements [10,1517]. The developing anaerobic environment enables the growth of bacteria that reduce sulphates to sulphides, which is accompanied by the formation of dark, insoluble residues [18].

The elements contained in the alloys of stainless steel (e.g., chromium and nickel) are responsible for forming a passivation layer on the elements’ surfaces, consisting of oxides (mainly Cr2O3 and NiO) that protect them from corrosion [7].

The method for the analysis of the surface morphology of orthodontic materials often used in the available literature is scanning electron microscopy with an energy-dispersive X-ray analytical system [1821]. SEM-EDX analyses enable qualitative and quantitative assessments of particular elements in the alloys, and the results are proportional to the weight fractions of these elements in the alloy [22]. It is possible to indirectly estimate the ion release from the nickel, chromium, and iron of the orthodontic alloys [22].

Corrosion causes dissolution and consequent disappearance of metals from the surface of the alloy. Passivation means that metals on the alloy surfaces are covered with oxygen. Consequently, as a result of a change on the surface, elution/coverage of metals on the surface are observed in the process of corrosion/passivation, respectively. The aim of this study was to evaluate changes in the composition of the surfaces of elements for new and used orthodontic appliances by SEM-EDX.

Material and Methods

New and used (after 20 months of treatment) parts of fixed orthodontic appliances were evaluated. Wires, brackets, and bands (5 each in every element) were all made of stainless steel. All the elements were purchased from the following companies: brackets (Victory Series APC PLUS 022, 3M Unitek, Monrovia, CA, USA), wires (0.017×0.025, 3M Unitek, Monrovia, CA, USA), and bands (37+, 3M Unitek, Monrovia, CA, USA). Used elements (the same as the new ones) were collected from patients treated in the orthodontic practise. All parts were sent to a specialized laboratory for analysis.

Scanning Electron Microscopy (SEM-EDX)

The surfaces of the parts of fixed orthodontic appliances were evaluated by the SEM-EDX technique before and after orthodontic treatment. Samples were mounted appropriately and were subjected to X-ray microanalysis using Brücker AXS Quantax in conjunction with the software ESPRIT version 1.8.2. The ultrastructure of the examined material was analysed using a scanning electron microscope EVO LS15 ZEISS with an SE1 and CZ BSD detector, operating at EHT=10 kV. For each part of the fixed orthodontic appliances, 10 analyses (on both sides) were performed.

Statistical methods

The results were analyzed statistically by Statistica version 10.0 and descriptive statistics (means, standard deviations) are reported. Normality in the distribution of experimental results was assessed by the Shapiro-Wilk test. On this basis, a statistical test was selected, which was used to investigate the significance of differences between the groups. The differences between the groups were investigated using the Mann-Whitney U test, t-test, or Cochrane-Cox test. Results were considered significantly different when p<0.05 and p<0.1.

Results

Micrographs of the components of the orthodontic appliances before and after the orthodontic treatment (magnification 20 and 1000 times) are shown in Figure 1. Maps of the distribution of elements on the surface of the analysed elements magnified by 1000 are shown in Figure 2. Table 1 presents the atomic content of the elements on the examined surfaces. Pictures were taken and mapping of the atomic density of the elements was performed. During the analysis of the surface composition of the tested elements at a magnification of 1000, a decrease in the content of Cr and Fe was observed on the surface of the brackets, bands, and wires used (Table 1).

Figure 1.

Figure 1

The surface: (A, A′) – wires (new); (B, B′) – wires (used); (C, C′) bands – (new); (D, D′) – bands (used); (E, E′) – brackets (new); (F, F′) – brackets (used); ×20 (A–D) and ×1000 (A′–D′).

Figure 2.

Figure 2

Maps of distribution of elements on the surface – (A) wires (new); (B) wires (used); (C) bands (new); (D) bands (used); (E) brackets (new); (F) brackets (used), ×1000.

Table 1.

Atomic concentration of elements (% of all detected elements) on the surface of wires, bands and brackets, n=10.

Wire Bands Brackets
New Used p New Used p New Used p
Fe 68.87±1.03 58.83±9.09 <0.05* 64.64±0.99 59.31±7.70 <0.05* 70.11±3.00 68.39±4.52 ns
Cr 18.46±0.73 17.40±2.29 ns 17.25±0.35 16.56±1.71 ns 16.55±0.58 15.24±0.83 <0.05**
Ni 10.22±0.97 11.41±0.81 <0.05* 13.01±1.05 13.24±0.79 <0.05* 7.03±1.66 9.46±2.07 <0.05**
O <0.01 3.39±2.86 <0.05* <0.01 0.08±0.05 <0.05* 0.80±0.85 0.43±0.46 ns

Atomic concentration of other elements (Mn, Al, Ti, Na, Ca, Mg, K, C, Si, S, N, P, Cl) were below 1%.

*

U Mann-Whitney test;

**

Student’s t-test;

ns – not statistically significant.

Discussion

The surface composition of the wires, bands, and brackets changed after 20 months of orthodontic treatment. In all the elements, the percentage contribution of Fe and Cr decreased (in wires: Fe 15%, Cr 6%; in bands: Fe 8%, Cr 4%; in brackets: Fe 2.5%, Cr 8%) and Ni and O increased (in wires: Ni 12%, O from below LLD (lower limit of detection) to 3.4%; in bands: Ni 1%, O from below LLD to 0.08%; in brackets: Ni 35%, O decreased 46%). This could suggest solubilization of Fe and Cr and passivation by oxides.

Stainless steel, containing Fe and Cr, is classified as a corrosion-resistant alloy. However, according to published reports, 316L stainless steel (of which orthodontic brackets are made) possesses a corrosion potential (release of Ni). The resistance to corrosion of stainless steel is related to the Cr content, which spontaneously forms oxides and thus a passivation layer. Acidity and chlorides cause destruction of the passivation layer. The passivation layer, besides Cr, also contains other metals – Fe, Ni, and Mo [8].

The products of stainless steel passivation consist mainly of iron (48.82%), chromium (17.9%), nickel (4.73%), and oxygen (19.56%) [22]. These can be Fe2O3 (or hydrated forms), Fe3O4, FePO4, Fe(OH)3, or Cr(OH)3 [18,2325]. Similar results were obtained by Menezes et al. during a study of bracket surfaces made of stainless steel after 60-day incubation in a physiological saline solution [22]. Using a magnification of 1000, the atomic density mapping of the elements was analysed. Areas with the highest density of O were found, which suggests that oxides were present in the examined surfaces.

We observed that an organic layer containing Na, P, S, Cl, K, and Ca was present on the wires [26]. The research also showed that there was significantly more residue on the wire surface. In brackets, the largest amount of residue was observed between the wings, yet the wing surfaces, as well as the areas that adjoined the wires, were relatively free of biofilm. The accumulation of residue in the above-mentioned areas on the wires and bracket surfaces is a result of difficulties in access during tooth brushing.

Conclusions

The analysis of the surface composition showed a decrease in the content of chromium and iron on the surface of the components of fixed orthodontic appliances, which showed the contribution of passivation and corrosion.

Footnotes

Source of support: This research was financially supported by The National Centre for Research and Development in Poland (N R130006 10)

References

  • 1.Sfondrini MF, Cacciafesta V, Maffia E, et al. Chromium release from new stainless steel, recycled and nickel-free orthodontic brackets. Angle Orthod. 2009;79:361–67. doi: 10.2319/042108-223.1. [DOI] [PubMed] [Google Scholar]
  • 2.Fernandez-Minano E, Ortiz C, Vincente A, et al. Metallic ion content and damage to the DNA in oral mucosa cells of children with fixed orthodontic appliances. Biometals. 2011;24:935–41. doi: 10.1007/s10534-011-9448-z. [DOI] [PubMed] [Google Scholar]
  • 3.Harzer W, Schröter A, Gedrange T, Muschter F. Sensitivity of titanium brackets to the corrosive influence of fluoride-containing toothpaste and tea. Angle Orthod. 2001;71:318–23. doi: 10.1043/0003-3219(2001)071<0318:SOTBTT>2.0.CO;2. [DOI] [PubMed] [Google Scholar]
  • 4.Amini F, Rakhshan V, Mesgarzadeh N. Effects of long-term fixed orthodontic treatment on salivary nickel and chromium levels: a 1-year prospective cohort study. Biol Trace Elem Res. 2012;150:15–20. doi: 10.1007/s12011-012-9457-y. [DOI] [PubMed] [Google Scholar]
  • 5.Mikulewicz M, Chojnacka K, Wozniak B, Downarowicz P. Release of Metal Ions from Orthodontic Appliances: An In Vitro Study. Biol Trace Elem Res. 2012;146:272–80. doi: 10.1007/s12011-011-9233-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Kim H, Johnson JW. Corrosion of stainless steel, nickel-titanium, coated nickel-titanium, and titanium orthodontic wires. Angle Orthod. 1999;69:39–44. doi: 10.1043/0003-3219(1999)069<0039:COSSNT>2.3.CO;2. [DOI] [PubMed] [Google Scholar]
  • 7.Amini F, Borzabadi Farahani A, Jafari A, Rabbani M. In vivo study of metal content of oral mucosa cells in patients with and without fixed orthodontic appliances. Orthod Craniofac Res. 2008;11:51–56. doi: 10.1111/j.1601-6343.2008.00414.x. [DOI] [PubMed] [Google Scholar]
  • 8.Eliades T, Athanasiou AE. In Vivo Aging of Orthodontic Alloys: Implications for Corrosion Potential, Nickel Release, and Biocompatibility. Angle Orthod. 2002;72:222–37. doi: 10.1043/0003-3219(2002)072<0222:IVAOOA>2.0.CO;2. [DOI] [PubMed] [Google Scholar]
  • 9.House K, Sernetz F, Dymock D, et al. Corrosion of orthodontic appliances – should we care? Am J Orthod Dentofacial Orthop. 2008;133:584–92. doi: 10.1016/j.ajodo.2007.03.021. [DOI] [PubMed] [Google Scholar]
  • 10.Chaturvedi TP, Upadhayay SN. An overview of orthodontic material degradation in oral cavity. Indian J Dent Res. 2010;21:275–84. doi: 10.4103/0970-9290.66648. [DOI] [PubMed] [Google Scholar]
  • 11.Gedrange T, Kuhn UD, Walter B, et al. Effects of 28-day mechanical and chewing stress on content of bound and diffusible ions in muscles of mastication. Exp Toxicol Pathol. 2001;53:207–13. doi: 10.1078/0940-2993-00181. [DOI] [PubMed] [Google Scholar]
  • 12.Hafez HS, Selmi EMN, Eid FHK, et al. Cytotoxicity, genotoxicity, and metal release in patients with fixed orthodontic appliances: a longitudinal in-vivo study. Am J Orthod Dentofac Orthop. 2011;140:298–308. doi: 10.1016/j.ajodo.2010.05.025. [DOI] [PubMed] [Google Scholar]
  • 13.Mikulewicz M, Chojnacka K, Wołowiec P. Release of metal ions from fixed orthodontic appliance an in vitro study in continuous flow system. Angle Orthod. 2014;84:140–48. doi: 10.2319/113012-911.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Mikulewicz M, Chojnacka K, Zielinska A, Michalak I. Exposure to metals from orthodontic appliances by hair mineral analysis. Environ Toxicol Phar. 2011;32:10–16. doi: 10.1016/j.etap.2011.02.009. [DOI] [PubMed] [Google Scholar]
  • 15.Kao CT, Ding SJ, Chen YC, Huang TH. The anticorrosion ability of titanium nitride (TiN) plating on an orthodontic metal bracket and its biocompatibility. J Biomed Mater Res (Appl Biomater) 2002;63:786–92. doi: 10.1002/jbm.10484. [DOI] [PubMed] [Google Scholar]
  • 16.Kao CT, Ding SJ, He H, et al. Cytotoxicity of orthodontic wire corroded in fluoride solution in vitro. Angle Orthod. 2007;77:349–54. doi: 10.2319/0003-3219(2007)077[0349:COOWCI]2.0.CO;2. [DOI] [PubMed] [Google Scholar]
  • 17.Luft S, Keilig L, Jager A, Bourauel C. In-vitro evaluation of the corrosion behavior of orthodontic brackets. Orthod Craniofac Res. 2009;12:43–51. doi: 10.1111/j.1601-6343.2008.01436.x. [DOI] [PubMed] [Google Scholar]
  • 18.Bagatin CR, Ito IY, Andrucioli MCD, et al. Corrosion in Haas expanders with and without use of an antimicrobial agent: an in situ study. J Appl Oral Sci. 2011;19:662–67. doi: 10.1590/S1678-77572011000600020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Gedrange T, Mai R, Richter G, et al. X-ray microanalysis of elements in the masticatory muscle after paresis of the right masseter. J Dent Res. 2005;84:1026–30. doi: 10.1177/154405910508401111. [DOI] [PubMed] [Google Scholar]
  • 20.Fjeld M, Ogaard B. Scanning electron microscopic evaluation of enamel surfaces exposed to 3 orthodontic bonding systems. Am J Orthod Dentofacial Orthop. 2006;130:575–81. doi: 10.1016/j.ajodo.2006.07.002. [DOI] [PubMed] [Google Scholar]
  • 21.Amini F, Rakhsahan V, Pousti M, et al. Variations in surface roughness of seven orthodontic archwires: an SEM-profilometry study. Korean J Orthod. 2012;42:129–37. doi: 10.4041/kjod.2012.42.3.129. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Menezes LM, Souza RM, Dolci GS, Dedavid BA. Analysis of biodegradation of orthodontic brackets using scanning electron microscopy. Dental Press J Orthod. 2010;15:48–51. [Google Scholar]
  • 23.Escriva-Cerdan C, Blasco-Tamarit E, Garcia-Garcia DM, et al. Temperature Effect on the Corrosion Behaviour of Alloy 31 in polluted H3PO4 and Analysis of the Corrosion Products by Laser Raman Microscope Int. J Electrochem Sci. 2012;7:5754–64. [Google Scholar]
  • 24.Huttunen-Saarivirta E, Honkanen M, Lepisto T, et al. Microbiologically influenced corrosion (MIC) in stainless steel heat exchanger. App Surf Sci. 2012;258:6512–26. [Google Scholar]
  • 25.Kocijan A, Milosev I. The influence of complexing agent and proteins on the corrosion of stainless steels and their metal components. J Mater Sci-Mater Med. 2003;14:69–77. doi: 10.1023/a:1021505621388. [DOI] [PubMed] [Google Scholar]
  • 26.Edie JW, Andreasen GF, Zaytoun MP. Surface Corrosion of Nitinol and Stainless Steel Under Clinical Conditions. Angle Orthod. 1981;51:319–24. doi: 10.1043/0003-3219(1981)051<0319:SCONAS>2.0.CO;2. [DOI] [PubMed] [Google Scholar]

Articles from Medical Science Monitor : International Medical Journal of Experimental and Clinical Research are provided here courtesy of International Scientific Information, Inc.

RESOURCES