Kinetic Study of the Alkaline Degradation of Oseltamivir Phosphate and Valacyclovir Hydrochloride using Validated Stability Indicating HPLC

Ramzia I Al-Bagary 1, Asmaa A El-Zaher 1,, Fahima A Morsy 2, Mai M Fouad 2
PMCID: PMC4051789  PMID: 24932100

Abstract

Aqueous alkaline degradation was performed for oseltamivir phosphate (OP) and valacyclovir hydrochloride (VA). Isocratic stability indicating the use of high-performance liquid chromatography (HPLC) was presented for each drug in the presence of its degradation product. The separations were performed using the Nucleosil ODS column and a mobile phase consisting of phosphate buffer (pH = 7), acetonitrile, and methanol 50:25:25 (v/v/v) for OP. For VA separation, a Nucleosil CN column using phosphate buffer (pH = 7) and methanol 85:15 (v/v) was used as a mobile phase. Ultraviolet detection at 210 nm and 254 nm was used for OP and VA, respectively. The method showed high sensitivity concerning linearity, accuracy, and precision over the range 1–250 μg mL−1 for both drugs. The proposed method was used to determine the drug in its pharmaceutical formulation and to investigate the degradation kinetics of each drug’s alkaline-stressed samples. The reactions were found to follow a first-order reaction. The activation energy could also be estimated. International Conference on Harmonisation guidelines were adopted for method validation.

Keywords: oseltamivir phosphate, valacyclovir hydrochloride, degradation kinetic, stability indicating high-performance liquid chromatography

Introduction

Oseltamivir phosphate (OP) (Fig. 1A) is an ester-type prodrug of a neuraminidase inhibitor, which was developed for the treatment of the A and B strains of the influenza virus. OP is a white crystalline solid that has the chemical name (3R,4R,5S)-4-acetylamino-5-amino-3(1-ethylpropoxy)-1-cyclohexene-1-carboxylic acid, ethyl ester, phosphate (1:1). Its chemical formula is C16H28 N2O4.PO4. The molecular weight is 410.4 m for OP salt. It attracted much attention after the 2009 flu pandemic occurred, as it was a global outbreak of a new strain of the H1N1 influenza virus, often referred to colloquially as the “swine flu.” The detected virus contains a combination of genes from swine, avian (bird), and human influenza viruses.1 People in at-risk groups have to be treated with antivirals (oseltamivir or zanamivir) as soon as possible after the first experienced flu symptoms.2,3 Therapy with a neuraminidase inhibitor is considered especially important for patients with underlying risk factors, including pregnancy4 and those with severe or progressive clinical illness.5 Chemically, valacyclovir (VA) is a l-valine-2-[(2-amino-1, 6-dihydro-6-oxo-9-hipurin-9yl) methoxy] ethyl ester (Fig. 1B). It is a (l-valyl ester) prodrug of acyclovir. It is an antiherpes simplex virus type 1, 2, and varicella zoster virus.6 VA acts specifically by inhibiting the deoxyribonucleic acid polymerase of the virus.7,8 VA is converted rapidly into acyclovir after its oral administration via first-pass metabolism.9 The oral bioavailability of VA is relatively higher than that of acyclovir.10,11

Figure 1.

Figure 1

Chemical structure of oseltamivir phosphate (A) and valacyclovir (B).

A literature survey regarding oseltamivir analysis revealed that there are several methods, which are based on different techniques. Such as bioassay and capillary electrophoresis,12 cyclic voltammetry,13 spectrophotometry,14,15 fluorimetry,16 Fourier transform infrared spectrometry.17 To date, few liquid chromatographic methods with ultraviolet (UV) detection,1821 photodiode array,22 fluorescence detection,22 or with mass spectrometry (MS) detection have been reported.2327

A literature review of VA revealed that few analytical methods based on spectrophotometry,2830 high-performance liquid chromatography (HPLC) with UV detection,21,3134 HPLC with MS detection,35,36 and micellar electrokinetic chromatography37 were reported. The objective of the proposed methods is to develop simple and accurate methods for the estimation of OP and VA hydrochloride in bulk and in pharmaceutical dosage forms by reversed phase (RP)-HPLC. The method aimed to include kinetic studies, which are important for the quality control of pharmaceutical products. To our knowledge, the suggested method represents the first kinetics study of OP degradation and VA hydrochloride using HPLC. In addition, the method used for the separation of OP is more efficient and accurate when compared with its official HPLC method,21 as indicated by the presented statistical data.

Experimental

Instruments

A liquid chromatograph consisting of an Agilent HPLC instrument quaternary pump (Model G 1311 A pump, Agilent 1200 series; Agilent Technologies, Santa Clara, CA, USA) connected with a UV detector (Model G1314B, Agilent 1200series; Agilent Technologies). The injector was a manual Rheodyne® injector (Model 7725I; IDEX Health & Science LLC, Rohnert Park, CA, USA) equipped with a 20 μL injector loop (Agilent Technologies). The chromatographic conditions for OP and VA included two stationary phases, which were Nucleosil® 300–5 C18 (Sigma-Aldrich, St Louis, MO, USA) and Macherey-Nagel EC 250/4 Nucleosil® 120–7 CN (Macherey-Nagel GmbH & Co. KG, Düren, Germany), respectively. The mobile phases used for OP and VA consisted of phosphate buffer (pH = 7), acetonitrile, methanol 50:25:25(v/v) (A), and phosphate buffer (pH = 7) and methanol 85:15(v/v) (B), respectively. Each mobile phase was filtered through a 0.45 μm Millipore membrane filter (EMD Millipore, Billerica, MA, USA) and was degassed for 30 minutes in an ultrasonic bath prior to use. UV detection was done at 210 nm and 254 nm for OP and VA, respectively. The system operated at ambient temperature. The flow rate was isocratic at 1.0 mL/minute−1 and 0.8 mL/minute−1 for OP and VA, respectively. The samples were filtered through a 0.45 μm membrane filter and were injected by the aid of a 25 μL Hamilton® analytical syringe (Sigma-Aldrich).

Materials and reagents

Reference OP, with a molecular weight of 410.4 m (CAS number 204255–11–8) and VA (CAS number 124832–27–5), as well as powder were kindly supplied by Nile Company for Pharmaceuticals and Chemical Industries SAE (Cairo, Egypt) and GlaxoSmithKline (Brentford, UK), respectively. The purity of OP and VA was checked and found to be 99.8% w/w and 98.4% w/w, respectively. Pharmaceutical dosage forms (Taminil® tablets) manufactured by the Nile Company for Pharmaceuticals and Chemical Industries SAE, batch number 19419, claim to contain 75 mg of OP, while Valtrex® tablets manufactured by GlaxoSmithKline (batch number 5286A) claim claimed to contain 500 mg of VA; these were purchased from the local market.

All chemicals used were of analytical grade, and deionized water was of HPLC grade. Hydrochloric acid, methanol, acetonitrile for HPLC, monobasic potassium phosphate, anhydrous dibasic sodium phosphate, and sodium hydroxide were obtained from El-Nasr Pharmaceutical Chemicals Co. (ADWIC; Cairo, Egypt).

Standard solutions

OP stock solution (5 mg/mL−1) was prepared in methanol (OP stock solution). VA hydrochloride stock solution (5 mg/mL−1) was prepared in the phosphate buffer (pH = 7) (VA stock solution). A working standard solution of OP (A) (1 mg/mL−1) was prepared from an OP stock solution in methanol. A working standard solution of OP (B) (0.1 mg/mL−1) was prepared from the OP stock solution in methanol. A working standard solution of VA (A) (1 mg/mL−1) was prepared from a VA stock solution in the phosphate buffer (pH = 7). A working standard solution of VA (B) (0.1 mg/mL−1) was prepared from a stock solution in phosphate buffer (pH = 7).

General procedure

Accelerated alkaline degradation of OP

From the OP working standard solution (A), the aliquot equivalent to OP (10 mg) was transferred into a measuring flask (100 mL), then 10 mL of 0.1 M NaOH was added; the flask was then completed to volume with methanol. Complete hydrolysis was achieved at room temperature after 3 hours (180 min.), as investigated by liquid chromatography using the chromatographic conditions described under Instruments section.

Accelerated alkaline degradation of VA

From the VA working standard solution (A), 50 mL was transferred into a measuring flask (100 mL), and the flask was then completed to volume with 0.1 M NaOH. Complete hydrolysis was achieved at room temperature after 2.5 hours (150 min.), as investigated by liquid chromatography using the chromatographic conditions described under Instruments section.

Linearity

Volumes equivalent to (1–250 μg) from the standard solutions and the working standard solution (A) and (B) of OP and VA, respectively, were transferred separately into two series of measuring flasks (10 mL); they were then completed to volume with methanol and phosphate buffer (pH = 7) for OP and VA, respectively. Twenty μL of each of the previously described solutions were injected into the liquid chromatograph using the chromatographic conditions described under Instruments section. The corresponding peak areas were used to construct the calibration curves using area under peaks (AUPs) versus the corresponding concentrations in μg/mL−1; two regression equations were calculated for OP and VA, respectively.

Assay of OP in pharmaceutical formulation

The powder content of 20 capsules was evacuated, and this content was weighed. A quantity equivalent to OP (25 mg) was accurately weighed and transferred into a volumetric flask (25 mL) and methanol (10 mL) was added. The flask was then sonicated for 5 minutes, and its volume was completed with that of methanol. The above solution was filtered, and aliquots from the filtrate were taken and diluted with methanol to obtain a concentration range of 1–250 μg/mL−1. The procedure was then followed as described under the Linearity section.

Assay of VA in pharmaceutical formulations

Twenty tablets were weighed and finely powdered. A quantity of the powdered tablets equivalent to VA (25 mg) was accurately weighed and transferred into a volumetric flask (25 mL), and phosphate buffer (pH = 7) (10 mL) was added. The flask was then sonicated for 5 minutes and the volume was completed with phosphate buffer (pH = 7). The above solution was filtered, and aliquots from the filtrate were taken and diluted with phosphate buffer (pH = 7) to obtain a concentration range of 1–250 μg/mL−1. The procedure was then followed as described under the Linearity section.

Method validation

Method validation was carried out in accordance with the International Conference on Harmonisation guidelines.38

Kinetic studies

Studying the kinetic order of the reaction of OP

From the OP stock solution, the aliquot equivalent to OP (100 mg) was transferred into a measuring flask (50 mL), and then 0.1 M of NaOH (10 mL) was added and the volume was completed with methanol. A total of 5 mL sample solutions were taken at 30-minute intervals, placed into measuring flasks (10 mL), neutralized with 0.1 M HCl (1 mL), and the volume was completed with methanol. An aliquot equivalent to initial concentration (C0 = 40 μg/mL−1) was transferred into a measuring flask (5 mL) and completed to volume with methanol. The solution was injected into the liquid chromatograph using the chromatographic conditions described under Instruments section. The concentration of OP was calculated, and the log percentage of the remaining concentration was plotted against time.

Studying the kinetic order of the reaction of VA

From the VA working standard solution (A), the aliquot equivalent to VA (50 mg) was transferred into a measuring flask (100 mL) and completed to volume with 0.1 M of NaOH. Then, 4 mL sample solutions were taken every 15 minutes, placed into measuring flasks (10 mL), neutralized with 2 mL of 0.1 M HCl, and the volume was then completed with the phosphate buffer (pH = 7). Aliquot equivalent to the initial concentration (C0 = 100 μg/mL−1) were transferred into a measuring flask (5 mL) and completed to volume with the phosphate buffer (pH = 7). The solution was injected into the liquid chromatograph using the chromatographic conditions described under Instruments section. The concentration of VA was calculated, and the log percentage of the remaining concentration was plotted against time.

Studying the effect of NaOH concentration and temperature on the reaction rate of OP and VA

Into two separate series of measuring flasks (5 mL), aliquots equivalent to OP or VA (1 mg) were transferred, and 1 mL volume of NaOH (0.1 M, 0.2 M, and 0.3 M) were added and completed to the mark with methanol and phosphate buffer (pH = 7) for OP and VA, respectively. These solutions were transferred into clean, dry, conical flasks and refluxed in a thermostatically controlled water bath at 40 °C, 50 °C, and 60 °C for OP and at 30 °C, 40 °C, and 50 °C for VA. From each flask, 1 mL of the sample solutions were taken into a measuring flask (5 mL) every 10 minutes for OP, or after every 5 minutes for VA; the samples were then neutralized with 0.2 mL, 0.4 mL, and 0.6 mL of 0.1 M HCl, respectively. The volume was then completed with the respective solvent. Further dilution was done to obtain solutions with concentrations of 40 μg/mL−1 and 100 μg/mL−1 of OP and VA, respectively. Each solution was injected into the liquid chromatograph, as described under Instruments section. The log percentage of the remaining concentration against time was constructed for the different molarities of NaOH, and the rates constants and t1/2 were calculated.

Studying the effect of acidic, oxidizing, and photolytic conditions on the stability of OP and VA

Aliquots of OP or VA equivalent to 10 mg were transferred separately into two separate series of volumetric flasks (100 mL). To the first flask of each series, 0.1 M of HCl (10 mL) were added; the solutions were left for 4 hours then neutralized. To the second flask of each series, 5 mL of 50% H2O2 were added; the solution was left for 4 hours. The last flask of each series was completed to volume, as directed, and it was subjected to a UV lamp for 24 hours. The flasks were completed to volume with methanol or phosphate buffer (pH = 7) for OP and VA, respectively, in all cases. After the mentioned time, 20 μL of the solutions were injected into the liquid chromatograph using the chromatographic conditions described under Instruments section.

Results and Discussion

A simple isocratic HPLC method was developed for the determination of OP and VA in the presence their degradation products without prior separation. A trial for accelerated degradation was done for both drugs. Aqueous alkaline degradation was found to affect the drugs’ stability, and it led to 100% degradation at 3 hours and 2.5 hours for OP and VA, respectively (Fig. 2).

Figure 2.

Figure 2

The chromatograms of the degradation products of OP (A) and VA (B).

High-performance liquid chromatographic developments

To optimize the HPLC assay parameters, the type of columns used, its dimensions, its mobile phase conditions, and its choice of detection wavelength were carefully investigated. Different types of stationary phases, columns with different dimensions, and particle sizes were tried. It was found that the Nucleosil® 300–5 C18 column (250 mm × 4.6 mm) with a particle size of 5 μm and Macherey-Nagel EC 250/4 Nucleosil® 120–7 CN with a particle size of 7 μm gave the most suitable resolution for OP and VA, respectively. The mobile phase was chosen after several trials to reach the optimum stationary/mobile phase matching. Several mobile phase compositions with different buffer pH values were employed. The optimum separation for OP or VA, as well as the corresponding degradation product were obtained with a mobile phase consisting of phosphate buffer (pH = 7), methanol, and acetonitrile 50:25:25 (v/v), or phosphate buffer (pH = 7) and methanol 85:15 (v/v), respectively. The retention times were 3.3 ± 0.15 minutes for OP and 2.8 ± 0.15 minutes for the OP degradation product, or 6 ± 0.15 minutes for VA and 3.6 ± 0.15 minutes for the VA degradation product (Fig. 3). System suitability tests were used to ensure adequate performance of the chromatographic system. Items were evaluated for three replicate injections of the drugs at concentrations of 80 μg/mL−1 and 100 μg/mL−1 for OP and VA, respectively. Results were of the acceptable limits (Table 1).

Figure 3.

Figure 3

Representative chromatograms of OP and its degradation product (A) and VA and its degradation product (B).

Table 1.

System suitability results for the proposed HPLC method.

PARAMETER OP OP DEGRADATE VA VA DEGRADATE
aN 1986 2102 6784 10989
bR 2.61 10.54
cK 1.48 0.968 1.63 0.63
dα 1.529 2.587
eT 1.69 1.65 0.94 1.17
a

Notes: Number of theoretical plates.

b

Resolution.

c

Capacity factor.

d

Selectivity factor.

e

Tailing factor.

The chromatographic separation in this work allowed for the complete baseline separation of OP from its degradation product in less than 5 min., and of VA from its degradation product in less than 8 minutes.

HPLC validation

Linearity and range

Linear relationships were obtained between the AUPs at the selected wavelengths (210 nm and 254 nm for OP and VA, respectively). The linearity of the calibration curves and adherence of the system to Beer’s law were verified by the high correlation coefficient. The analytical data for the calibration curves are summarized in Table 2.

Table 2.

Assay parameters and methods validation for the proposed HPLC method.

PARAMETER OP VA
Retention time (min.) 3.3 ± 0.15 6 ± 0.15
Wavelength of detection, nm 210 254
Calibration range, μg ml−1 1–250 1–250
Regression equation y = 31.37x−31.7400 y = 49.71x + 8.4311
Correlation coefficient (r) 0.9998 0.9999
aSb 0.244 0.200
bSa 0.369 0.267
cLOD 0.055 0.29
dLOQ 0.165 0.879
Confidence limit of the slope 31.37 ± 0.002 49.7 ± 0.010
Confidence limit of the intercept 31.74 ± 0.007 8.43 ± 0.001
Standard error of the estimation 51.9457615 43.73
Intra day: Mean of concentrations (μg/ml) n=3 79.68 99.87 150.97 102.40 116.98 142.
eRSD,% 0.680 1.580 0.808 0.217 0.165 0.221
Inter day: Mean of concentrations (μg/ml) n=3 79.44 98.74 151.52 101.96 118.75 142.
RSD,% 1.11 0.95 0.37 1.07 1.30 1.18
Results: Drug in dosage form,% 96.31121 ±0.506964 100.1937 ± 0.215629
Drug added,% 100.2444 ± 1.63194 100.2633 ± 0.362817
a

Notes: Standard deviation of the slope.

b

Standard deviation of the intercept.

c

Limit of detection.

d

Limit of quantification.

e

Relative standard deviation. LOD and LOQ calculated by styx’s.

Specificity

The specificity of separation can be demonstrated by the resolution of the two components (2.61, 10.54) for OP and VA separation from their degradation products under the experimental condition respectively. Good peaks, a clear baseline, and the absence of interference are demonstrated in Figure 3. The specificity of the proposed method was shown by the analysis of laboratory-prepared mixtures containing known amounts of the standard drugs at 25%–100%, which were fortified to the degradation samples. The percentage recovery rate at each level was in good agreement, irrespective of the percent of the added degradation (Table 3). Moreover; the specificity of the method was ascertained by successful application to Taminil® tablets for OP and Valtrex® tablets for VA, without any peak distortions or variations that resulted from the excipients. The small relative standard deviation percentage (0.526% for OP, 0.215% for VA) indicates that the methods were specific (Table 2).

Table 3.

Accuracy determination of OP and VA and their degradation by the proposed HPLC method.

DRUG% / DEGRADATION% CONC. FOUND CONC. RECOVERY%
OP VA OP VA OP VA OP VA
100/37.5 100/100 40 50 39.8 49.88 99.5 99.76
100/33.3 100/40 75 50 75.6 50.88 100.8 101.76
100/26.6 100/25 60 80 59.4 80.61 99 100.76
100/18.7 100/50 80 60 79.04 59.96 98.8 99.93
100/38.8 100/66.6 90 15 90.9 15.08 101 100.53
100/11.1 100/33.3 90 15 90 15.06 100 100.40
Mean 99.85 100.52
SD ±0.916 ±0.711

Accuracy

The standard addition technique was used to ascertain the accuracy of the results (Table 2). The accuracy of the drug substances was ascertained by comparing the results of the proposed method with an official method21 for OP and the reported HPLC method34 for VA. Results indicate a nonsignificant difference, proving the method’s accuracy (Table 4).

Table 4.

Statistical analysis of the results obtained by applying the proposed HPLC methods.

STATISTICAL TERMS OP VA
REFERENCE METHOD (21) HPLC METHOD REFERENCE METHOD (34) HPLC METHOD
Mean 100.375 100.345 99.452 99.997
S.D.± 1.368 0.727 1.134 0.583
S.E.± 0.559 0.297 0.463 0.238
R.S.D,% 1.363 0.725 1.141 0.583
N 6 6 6 6
V 1.872 0.529 1.287 0.340
t(1.81) 0.008 1.047
F(5.05) 3.541 3.780

Precision

Precision was determined by applying the procedures to three concentrations for three replicates. The precision of the suggested methods was also expressed in terms of relative standard deviation of the interday and intraday analysis (Table 2).

Robustness

The robustness of the HPLC method was investigated via an analysis of samples under a variety of experimental conditions, such as small changes in pH (6.0–7.5), or changing the column using a 250 mm × 4.6 mm id, C18 Lichrosorb 10 m analytical column for OP, and using the column 250 × 4 id Waters Spherisorb 5 μ CN for VA. The effect on retention time and peak parameters was studied. It was found that the method was robust when the column and the mobile phase pH were varied. During these investigations, the retention times were modified; however, the areas and peak symmetries were conserved. Changes in instruments or personnel did not alter the results, which indicated the ruggedness of the proposed method.

Limits of detections and quantifications

The detection and quantification limits were calculated based on the standard deviation of the response and the slope of the calibration curves, as follows:

LOD=3.3×σ/S, (1)
LOQ=10×σ/S, (2)

where σ is the standard deviation of the response and S is the slope of the regression line of the calibration curve, indicating the sensitivity of the method.

Stability study

Treatment of both drugs with acid, an oxidizing agent—namely, hydrogen peroxide—and, finally, exposure to light for a fixed period of time (as mentioned under the Studying the effect of acidic, oxidizing, and photolytic conditions on the stability of OP and VA section) resulted in nearly no change of the drugs’ chromatogram, as shown in Figure 4.

Figure 4.

Figure 4

Chromatogram representing results of acid induced degradation of OP (a1) and VA (b1), oxidizing agent-induced degradation of OP (a2) and VA (b2), UV-induced degradation of OP (a3) and VA (b3).

Kinetics study of the degradation

Treatment of OP and VA with alkali resulted in the gradual decomposition of both drugs. The linear relationship (Fig. 5) between the percentage log of the remaining concentration against time indicated first-order degradation. Since the hydrolysis was performed in a large excess of NaOH (0.1 mol/L−1), it followed a pseudo first-order reaction rate.39 The term is used when two reactants are involved in the reaction, but one of them is in such a large excess (NaOH) that any change in the concentration is negligible compared with the change in the concentration of the other reactant (drug).

Figure 5.

Figure 5

First order plot of hydrolysis of OP and VA (4 mg%) with o.1 M NaOH.

Different parameters that affect the rate of the reaction were studied. The temperature dependence and effect of base strength on OP and VA degradation was studied by conducting the reaction at different temperatures using different base strengths (Fig. 6). At each temperature, the rate constant and t1/2 were calculated. It was concluded that as the temperature increased, the rate of hydrolysis also increased, with a decrease noted in the t1/2 (Table 5). Also, the energy of activation was determined by calculating the rate constant from the following equation:40

Log K2/K1=Ea/2.303R*T2T1/T2T1 (3)

Figure 6.

Figure 6

First order plot of the hydrolysis of OP and VA (10 mg%) with o.1 M NaOH at different temperature.

Table 5.

Kinetic data of the stress alkaline hydrolysis OP and VA using the proposed HPLC method.

STRENGTH OF NAOH TEMPERATURE (°C) K(H−1) T1/2 (H)
OP VA OP VA OP VA
0.1M Room temp. Room temp. 0.016 0.062 42.987 11.145
40 30 0.029 0.058 24.073 11.036
50 40 0.041 0.074 17.001 9.403
60 50 0.0578 0.253 11.988 2.736
0.2 M 40 --- 0.062 --- 11.145 ---
50 --- 0.124 --- 5.572 ---
60 --- 0.168 --- 4.122 ---
0.3 M 40 --- 0.081 --- 8.597 ---
50 --- 0.120 --- 5.787 ---
60 --- 0.210 --- 3.307 ---

Where “Ea” is the activation energy, “T1” and “T2” are the two temperatures in Kelvin, “R” is the gas constant, and “k1” and “k2” are the rate constants at the two temperatures used. The calculated “Ea” was found to be 30.013 k joule mol−1 and 8.79 k joule mol−1 for OP and VA, respectively. Another factor that affects the rate of the reaction is the base strength of NaOH. Different base strengths were used to study its effect upon the hydrolysis, and it was found that increasing the base strength results in a uniform increase in the rate constant and a decrease in the t1/2 for OP. It is worth mentioning that attempts to use higher base strengths for VA result in a spontaneous degradation, revealing high instability of the drug towards higher base strengths that was also confirmed by its law value of the activation energy.

Conclusion

The proposed HPLC methods provide simple, sensitive, and specific methods suitable for the quantitative analysis of both drugs in the presence of their degradation products for the routine quality control analysis of OP and VA, either in their pure form or in available pharmaceutical dosage forms, with no interference from the excipients or the degradation product. The reaction kinetic of the degradation was found to be a pseudo first-order reaction under the experiment’s basic stress conditions. Other stress conditions did not affect the drugs investigated. The methods proved that the selectivity, accuracy, and simple mobile phases used provide simple and economic applications, and they reflect suitability for quality control laboratories.

Footnotes

Author Contributions

Conceived and designed the experiments: RIB, AAZ, FAM. Analyzed the data: RIB, AAZ, FAM, MMF. Wrote the first draft of the manuscript: AAZ, MMF. Contributed to the writing of the manuscript: FAM, MMF. Agree with manuscript results and conclusions: RIB, AAZ, FAM. Jointly developed the structure and arguments for the paper: RIB, AAZ, MMF. Made critical revisions and approved final version: RIB, AAZ, FAM. All authors reviewed and approved of the final manuscript.

DISCLOSURES AND ETHICS

As a requirement of publication the authors have provided signed confirmation of their compliance with ethical and legal obligations including but not limited to compliance with ICMJE authorship and competing interests guidelines, that the article is neither under consideration for publication nor published elsewhere, of their compliance with legal and ethical guidelines concerning human and animal research participants (if applicable), and that permission has been obtained for reproduction of any copyrighted material. This article was subject to blind, independent, expert peer review. The reviewers reported no competing interests.

ACADEMIC EDITOR: Gabor Patonay, Editor in Chief

FUNDING: This work was financially supported by the research committee in National Organization for Drug Control and Research.

COMPETING INTERESTS: Authors disclose no potential conflicts of interest

REFERENCES

  • 1.Writing Committee of the WHO Consultation on Clinical Aspects of Pandemic (H1N1) 2009 Influenza Clinical Aspects of Pandemic 2009 Influenza A (H1N1) Virus Infection. N Engl J Med. 2010;362:1908–17. doi: 10.1056/NEJMra1000449. [DOI] [PubMed] [Google Scholar]
  • 2.Novel Swine-Origin Influenza A (H1N1) Virus Investigation Team Emergence of a novel swine origin influenza A (H1N1) virus in humans. N Engl J Med. 2009;360:2605–15. doi: 10.1056/NEJMoa0903810. [DOI] [PubMed] [Google Scholar]
  • 3.Itoh Y, Shinya K, Kiso M, et al. In vitro and in vivo characterization of new swine origin H1N1 influenza viruses. Nature. 2009;460:1021–5. doi: 10.1038/nature08260. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Lim ML, Chong CY, Tee WS, Lim WY, Chee JJ. Influenza A/H1N1 infection in pregnancy—an Asian perspective. BJOG. 2009;2010;117:551–6. doi: 10.1111/j.1471-0528.2010.02522.x. [DOI] [PubMed] [Google Scholar]
  • 5.WHO guidelines for pharmacological management of pandemic (H1N1) 2009 influenza and other influenza viruses. Geneva: World Health Organization; Feb, 2010. [last accessed March 7, 2012]. http://www.who.int/csr/resources/publications/swineflu/h1n1_use_antivirals_20090820/en/index.html. [PubMed] [Google Scholar]
  • 6.Balimane PV, Tamai I, Guo A, et al. Biochem Biophys Res Commun. 1998;250:246–51. doi: 10.1006/bbrc.1998.9298. [DOI] [PubMed] [Google Scholar]
  • 7.Beauchamp LM, Krenitsky TA. Drugs Future. 1993;18:619–28. [Google Scholar]
  • 8.Beauchamp LM, Orr GF, De Miranda P, TC TC Burnette, Krenitsky TA. Antiviral Chem Chemother. 1992;3:157–64. [Google Scholar]
  • 9.Landowski CP, Sun D, Foster DR, et al. J Pharmacol Exp Ther. 2003;306:778–86. doi: 10.1124/jpet.103.051011. [DOI] [PubMed] [Google Scholar]
  • 10.Thomsen AE, Christensen MS, Bagger MA, Steffansen B. Eur J Pharm Sci. 2004;23:319–25. doi: 10.1016/j.ejps.2004.08.005. [DOI] [PubMed] [Google Scholar]
  • 11.Anand BS, Katragadda S, Mitra AK. J Pharmacol Exp Ther. 2004;311:659–67. doi: 10.1124/jpet.104.069997. [DOI] [PubMed] [Google Scholar]
  • 12.Bosch ME, Ojeda CB, Sánchez AJR, Rojas FS. Analytical methodologies for the determination of oseltamivir. Res J Pharm Biol Chem Sci. 2010;1:368–76. [Google Scholar]
  • 13.Ivic MLA, Petrovic SD, Mijin DZ, Drljevic-Duric KM. The qualitative determination of oseltamivir phosphate in tamiflu® capsule by cyclic voltammetry. Hem Ind. 2011;65:87–91. [Google Scholar]
  • 14.Shanmukha KJV, Somarouthu P, Medaboina G, Bala SC. Application of potassium permanganate to the spectrophotometric determination of oseltamivir phosphate in bulk and capsules. Asian J Pharm Clin Res. 2012;5(2):18–22. [Google Scholar]
  • 15.Thatte AA, Kadam RJ, Pramila T, Bhoi UA, Deshpande KB. Development, validation and application of uv spectrophotometric method for the determination of oseltamivir phosphate in bulk and pharmaceutical dosage. International Journal of ChemTech Research. 2011;3(2):569–73. [Google Scholar]
  • 16.Walash MI, Belal F, El-Enany N, El-Maghrabey MH. Spectrofluorimetric determination of oseltamivir phosphate through derivatization with o-phthalaldehyde. Application to pharmaceutical preparations with a preliminary study on spiked plasma samples. Luminescence. 2012;27(6):511–8. doi: 10.1002/bio.1385. [DOI] [PubMed] [Google Scholar]
  • 17.Bunaciu AA, Nita S, Fleschin S, Aydoğmuş Z, Aboul-Enein HY. Fourier transform infrared spectrophotometry method used for oseltamivir determination in pharmaceutical formulations. GU J Sci. 2012;25(3):631–4. [Google Scholar]
  • 18.Bahrami G, Mohammadi B, Kiani A. Determination of oseltamivir carboxylic acid in human serum by solid phase extraction and high performance liquid chromatography with UV detection. J Chromatogr B. 2008;864:1–2. 38–42. doi: 10.1016/j.jchromb.2008.01.048. [DOI] [PubMed] [Google Scholar]
  • 19.Joseph-Charles J, Geneste C, Laborde-Kummer E, Gheyouche R, Boudis H, Dubost JP. Development and validation of a rapid HPLC method for the determination of oseltamivir phosphate in Tamiflu® and generic versions. J Pharm Biomed Anal. 2007;44(4):1008–13. doi: 10.1016/j.jpba.2007.04.002. [DOI] [PubMed] [Google Scholar]
  • 20.Ameti MA, Slavkovska J, Starkoska K, Arsova-Sarafinovska Z. A simple isocratic RP-HPLC method for quality control of oseltamivir capsules. Macedonian Journal of Chemistry and Chemical Engineering. 2012;31(2):205–15. [Google Scholar]
  • 21.The United States Pharmacopeia (USP36), National Formulary (NF31) The United States Pharmacopeial Convention Inc; Rockville, MD: 2005. [Google Scholar]
  • 22.Aydogmus Z, Caglar S, Toker S. RP-HPLC method for determination of oseltamivir phosphate in capsules and spiked plasma. Anal Lett. 2010;43:2200–9. [Google Scholar]
  • 23.Heinig K, Buchel F. Sensitive determination of oseltamivir and oseltamivir carboxylate in plasma, urine, cerebrospinal fluid and brain by liquid chromatography–tandem mass spectrometry. J Chromatogr B. 2008;876(1):129–36. doi: 10.1016/j.jchromb.2008.10.037. [DOI] [PubMed] [Google Scholar]
  • 24.Kromdijk W, Rosing H, van den Broek MP, Beijnen JH, Huitema AD. Quantitative determination of oseltamivir and oseltamivir carboxylate in human fluoride EDTA plasma including the ex vivo stability using high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2012:891–2. 57–63. doi: 10.1016/j.jchromb.2012.02.026. [DOI] [PubMed] [Google Scholar]
  • 25.Junwal M, Sahu A, Handa T, Shah RP, Singh S. ICH guidance in practice: degradation behaviour of oseltamivir phosphate under stress conditions. J Pharm Biomed Anal. 2012;62:48–60. doi: 10.1016/j.jpba.2012.01.001. [DOI] [PubMed] [Google Scholar]
  • 26.Berendsen BJA, Wegh RS, Essers ML, Stolker AAM, weigel S. Quantitative trace analysis of a broad range of antiviral drugs in poultry muscle using column-switch liquid chromatography coupled to tandem mass spectrometry. Anal bioanal Chem. 2012;402:1611–23. doi: 10.1007/s00216-011-5581-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Hu ZY, Laizure SC, Meibohm B, Herring VL, Parker RB. Simple and sensitive assay for quantification of oseltamivir and its active metabolite oseltamivir carboxylate in human plasma using high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry: Improved applicability to pharmacokinetic study. J Pharm Biomed Anal. 2012;72:245–50. doi: 10.1016/j.jpba.2012.08.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Ramakrishna VS, Lakshmi PBS, kumar DR, babu CR. Spectrophotometric determination of Valacyclovir HCl through oxidative coupling reaction in bulk and its pharmaceutical preparations. International Journal of ChemTech Research. 2012;4(1):138–42. [Google Scholar]
  • 29.Ramakrishna VS, kumar DR, Rao NVNM, Babu CR. Method development and determination of valacyclovir HCl in pharmaceutical dosage forms by visible spectrophotometry. International Journal of ChemTech Research. 2012;4(3):1009–14. [Google Scholar]
  • 30.Reddy JS, Ahmed MASM, Chakravarth IE, Prabhavathi K. Spectrophotometric estimation of valacyclovir in pharmaceutical preparations. J Chem Pharm Res. 2011;3(4):773–6. [Google Scholar]
  • 31.Ganesh M, Bahagiyalakshmi M, Hemalatha P, Rao CVN, Jung HT, Rajasekar K. RP-HPLC estimation of valacyclovir HCl in tablet formulation. Asian J Chem. 2011;23(3):1317–20. [Google Scholar]
  • 32.Rao KS, Sunil M. Stability-indicating liquid chromatographic method for valacyclovir. International Journal of ChemTech Research. 2009;1(3):702–8. [Google Scholar]
  • 33.Sultana Y, Agarwal NK, Safia Khanam S. Development and validation of stability indicating RP-HPLC method for estimation of valacyclovir in pharmaceutical dosage forms. International J Pharm Clin Res. 2013;5(1):7–12. [Google Scholar]
  • 34.Rasool SK, Naik DV, Babu DP, Nalluri BN. RP-HPLC method for the estimation of valacyclovir in bulk and pharmaceutical formulations. Int J Pharm Pharm Sci. 2012;4(1):214–8. [Google Scholar]
  • 35.Sasanya JJ, Abd-Alla AM, Parker AG, Cannavan A. Analysis of the antiviral drugs acyclovir and valacyclovir-hydrochloride in tsetse flies (glossina pallidipes) using LC-MSMS. J Chromatogr B Analyt Technol Biomed Life. 2010;878(26):2384–90. doi: 10.1016/j.jchromb.2010.07.008. [DOI] [PubMed] [Google Scholar]
  • 36.Yadav M, Upadhyay V, Singhal P, Goswami S, Shrivastav PS. Stability evaluation and sensitive determination of antiviral drug, valacyclovir and its metabolite acyclovir in human plasma by a rapid liquid chromatography-tandem mass spectrometry method. J Chromatogr B Analyt Technol Biomed Life Sci. 2009;877:8–9. 680–8. doi: 10.1016/j.jchromb.2009.01.042. [DOI] [PubMed] [Google Scholar]
  • 37.Al Azzam KM, Saad B, Makahleah A, Aboul-Enein HY, Elbashir AA. Assay and stability-indicating micellar electrokinetic chromatography method for the simultaneous determination of valacyclovir, acyclovir and their major impurity guanine in pharmaceutical formulations. Biomed Chromatogr. 2010;24(5):535–43. doi: 10.1002/bmc.1323. [DOI] [PubMed] [Google Scholar]
  • 38.Guidance for Industry: Q2B Validation of Analytical Procedures: Methodology, U.S. Department of Health and Human Services, Food and Drug Administration Centre for Drug Evaluation and Research (CDER), Centre for Biologics Evaluation and Research (CBER) Nov, 1996. http://www.fda.gov/cder/guidance/index.htm.
  • 39.Florence AT, Attwood D. Physical Principles of Pharmacy. 5th ed. London, UK: Pharmaceutical Press; 2011. [Google Scholar]
  • 40.Martin A, Swarbrick J, Cammarata A. Physical Pharmacy. 3rd ed. Philadelphia, USA: Lea and Febiger; 1983. [Google Scholar]

Articles from Analytical Chemistry Insights are provided here courtesy of SAGE Publications

RESOURCES