Abstract
Abstracta
Introduction
Noninvasive continuous positive airway pressure (CPAP) is recognized as an effective treatment for severe airway obstruction in young children. The aim of the present study was to compare a clinical setting with a physiological setting of noninvasive CPAP in infants with nocturnal alveolar hypoventilation due to severe upper airway obstruction (UAO) or bronchopulmonary dysplasia (BPD).
Methods
The breathing pattern and respiratory muscle output of all consecutive infants due to start CPAP in our noninvasive ventilation unit were retrospectively analysed. CPAP set on clinical noninvasive parameters (clinical CPAP) was compared to CPAP set on the normalization or the maximal reduction of the oesophageal pressure (Poes) and transdiaphragmatic pressure (Pdi) swings (physiological CPAP). Expiratory gastric pressure (Pgas) swing was measured.
Results
The data of 12 infants (mean age 10 ± 8 mo) with UAO (n = 7) or BPD (n = 5) were gathered. The mean clinical CPAP (8 ± 2 cmH2O) was associated with a significant decrease in Poes and Pdi swings. Indeed, Poes swing decreased from 31 ± 15 cmH2O during spontaneous breathing to 21 ± 10 cmH2O during CPAP (P < 0.05). The mean physiological CPAP level was 2 ± 2 cmH2O higher than the mean clinical CPAP level and was associated with a significantly greater improvement in all indices of respiratory effort (Poes swing 11 ± 5 cm H2O; P < 0.05 compared to clinical CPAP). Expiratory abdominal activity was present during the clinical CPAP and decreased during physiological CPAP.
Conclusions
A physiological setting of noninvasive CPAP, based on the recording of Poes and Pgas, is superior to a clinical setting, based on clinical noninvasive parameters. Expiratory abdominal activity was present during spontaneous breathing and decreased in the physiological CPAP setting.
Keywords: Airway obstruction, Continuous positive airway pressure, Oesophageal pressure, Expiratory abdominal activity, Infant
Introduction
Noninvasive continuous positive airway pressure (CPAP) is recognized as an effective treatment for severe upper airway obstruction (UAO) in young children. Indeed, the maintenance of airway patency throughout the entire breathing cycle by means of CPAP has been shown to be associated with an unloading of the respiratory muscles, an improvement of breathing pattern and gas exchange [1-6].
In adults and adolescents with obstructive sleep apnoea (OSA), the titration of CPAP is based on the polysomnographic disappearance of apnoea, hypopnoea, respiratory effort-related arousal and snoring, as recommended by the American Academy of Sleep Medicine (AASM) [7,8]. The diseases responsible for airway obstruction in infants differ from those of older children and adults with the predominance of anatomical abnormalities of the upper airways such as laryngomalacia or tracheomalacia, Pierre Robin syndrome or other maxillofacial malformations [3,6].
Some other diseases involving the lower respiratory tract, such as chronic lung diseases of prematurity, also called bronchopulmonary dysplasia (BPD), are associated with lung diseases and predominantly peripheral airway obstruction, which may be severe and cause intrinsic positive end-expiratory pressure (PEEPi). Because of these differences in pathophysiology and the lack of guidelines for this age group, the titration of CPAP in infants is generally based on clinical parameters such as the disappearance of the stridor and retractions, the decrease in respiratory and heart rates and the normalization of gas exchange [2,3]. To facilitate the acclimatization of the infant with CPAP, the initial level is usually set at 4 cmH2O, followed by a gradual increase of the CPAP level until the best clinical efficacy and comfort are obtained.
Importantly, small changes in the CPAP level may have significant clinical consequences. Indeed, the minimal airway diameter is the most critical because, according to Poiseuille's law, the resistance increases with an exponent 4 of the radius. This has been observed in infants with acute viral bronchiolitis, in whom a 2 cmH2O change of CPAP level was associated with a significant change in the work of breathing and breathing pattern [9]. In such a homogeneous group of infants, the optimal CPAP level, determined by monitoring the oesophageal (Poes) and gastric pressure (Pgas), was 7 cmH2O for all the patients; however, it is not known if this level is also appropriate for infants with UAO and BPD, in whom CPAP could counteract the PEEPi.
The aim of the present study was to compare two settings of CPAP: a first setting based on noninvasive clinical parameters and a second setting based on the invasive recording of Poes and Pgas in a group of infants with severe UAO or BPD.
Materials and methods
Patients
All the data of consecutive infants less than 24 months old with severe UAO or BPD evaluated for noninvasive CPAP in a noninvasive ventilation unit of a paediatric university hospital were retrospectively analysed. All the patients with UAO had a laryngotracheal endoscopy under general anaesthesia. The UAO persisted in all the UAO patients despite endoscopic resection of the aryepiglottic folds and antireflux treatment using proton pump inhibitors. All the UAO and BPD patients were hypoxaemic (pulse oximetry (SpO2) less than 90% for more than 10 consecutive min and/or more than 10% of sleep time) and hypercapnic (transcutaneous carbon dioxide pressure (PtcCO2) above 50 mmHg for more than 10 consecutive min and/or more than 10% of sleep time) during an overnight sleep study [6]. Exclusion criteria were severe mental retardation and important midfacial deformity precluding the tolerance of a nasal mask. The study was approved by the Institutional Review Board of the French learned society for respiratory medicine, Société de Pneumologie de Langue Française (CEPRO 2012-031), and all the parents gave informed consent for the CPAP evaluation of their children.
CPAP equipment
Because of the lack of adequate nasal interfaces for infants, all the patients were equipped with custom-made nasal masks manufactured by the prosthetists of the maxillofacial department of our hospital [10,11]. The CPAP devices used for the study were the Synchrony or Harmony (Philips Respironics, Murrysville, PA, USA), S9 or VPAP IV (ResMed, Bella Vista, Australia) and ICON (Fisher & Paykel Healthcare, Auckland, NZ) with an integrated humidification system. A commercial simple line circuit, recommended by the manufacturer, was connected to the nasal mask via an exhalation valve (Plateau Exhalation Valve; Philips Respironics). Oxygen therapy was not used during the study.
Measurements
Airway pressure (Paw) was measured with a differential pressure transducer (MP45 Low Pressure Transducer model ± 100 cmH2O; Validyne Engineering Corp, Northridge, CA, USA) on the Plateau Exhalation Valve during CPAP treatment. We were not able to measure airflow during spontaneous breathing, because the infants did not tolerate the mask alone with the pneumotachograph due to the increase in dead space and during CPAP because of the constant airflow.
Poes and Pgas were measured using a 2.1-mm external diameter catheter mounted pressure transducer system with two integrated pressure transducers (Gaeltec, Dunvegan, Isle of Skye, UK) inserted pernasally after careful local anaesthesia (lidocaine 2%; AstraZeneca, Rueil-Malmaison, France) [12,13]. After calibration, the catheter was advanced gently until the distal tip was in the stomach and the proximal pressure transducer was in the middle portion of the oesophagus. Appropriate placement of the Poes transducer was assessed using the occlusion technique [14]. Adequate placement of the Pgas transducer was ascertained by gentle manual pressure on the patient's abdomen to observe fluctuations in Pgas, which should be absent on the Poes trace. Transdiaphragmatic pressure (Pdi) was obtained by online subtracting of the Poes signal from the Pgas signal. All the signals were digitized at 128 Hz and sampled for analysis using an analogical/numeric acquisition system (MP100 System; BIOPAC Systems, Inc, Goleta, CA, USA), run on a PC computer (Elonex, Gennevilliers, France) and displayed using AcqKnowledge software (BIOPAC Systems, Inc). SpO2, heart rate and respiratory rate (fR) were recorded.
Protocol
No sedation was administered. The procedure was performed with the presence of at least one parent (generally the mother) and started with the placement of the oesogastric catheter at the time of a natural daytime nap. After 10 min of quiet breathing (ideally sleep), a 2-min baseline recording (spontaneous breathing) was taken. Next, the oesogastric recording was blinded to the investigator, and CPAP was started with an initial CPAP level at 4 cmH2O followed by a 1 cmH2O increase every 5 min until maximal clinical improvement. The clinical CPAP level considered optimal was the level associated with the disappearance of the stridor or retractions, a SpO2 greater than 95% on room air, the normalization or greatest decrease in heart rate and fR and the best comfort as assessed by the infant's remaining or falling asleep. After 10 min of quiet breathing at this CPAP level, a 2-min recording (clinical CPAP) was performed. Thereafter the CPAP level was decreased to the baseline value of 4 cmH2O for 5 min. After this period, the oesogastric recording was revealed to the investigator. The CPAP level was then adjusted on the normalization or the maximal decrease in the Poes swing, independently of the clinical parameters. After 10 min of quiet breathing at this CPAP level, a 2-min recording (physiological CPAP) was performed. Next, the CPAP level was increased by 1 cmH2O. A 2-min recording (physiological CPAP+1) was performed after 10 min of quiet breathing. The long-term CPAP use and clinical outcome were assessed in all patients.
Data analysis
fR and inspiratory time/total respiratory cycle time (TI/TTOT) ratio were estimated from the Poes trace. Poes and Pdi swings and the oesophageal pressure time product (PTPoes) and diaphragmatic pressure-time product (PTPdi) per breath and per minute were measured and calculated as previously described [12,15-17]. Normally, PTPoes is measured as the area subtended by Poes and the chest wall static recoil pressure-time curve over the TI, taking into account the PEEPi, as described by Sassoon et al. [15]. As it was impossible to measure airflow in these infants, owing to the increase in dead space when a face mask was added or during CPAP because of the continuous positive airflow, we could not measure the real TI. For the calculation of the PTPoes/breath, the onset of inspiration was thus defined as the deflection on the Poes trace and the end of the inspiration as the peak of Pdi [18]. The PTPdi/breath was obtained by measuring the area under the Pdi signal from the onset of its positive deflection to its return to baseline. Both PTPoes and PTPdi were also expressed per minute by multiplying the pressure-time products per breath by the fR (PTPoes/min and PTPdi/min) [15]. The Pgas swing during expiration was measured as a marker of expiratory abdominal activity. On the Pgas trace, we measured the decrease from the maximal end-expiratory level to the minimal value [19]. The presence of a positive Pgas swing during expiration was considered abnormal [20].
Statistical analysis
Data are given as means ± SD. Comparisons between the baseline and different CPAP conditions were made using (1) one-way or one-factor repeated-measures analysis of variance (ANOVA) if data were normally distributed or (2) Friedman repeated-measures ANOVA on ranks if data were not normally distributed. When a significant difference was observed, pairwise comparisons were performed using the Holm-Sidak test. The CPAP levels were compared using a t-test if data were normally distributed or a Mann-Whitney U test if data were not normally distributed. A P value less than 0.05 was considered statistically significant.
Results
Twelve infants (mean post natal age 10 ± 8 mo, mean weight 6.6 ± 2.3 kg) were enrolled in the study (Table 1). Seven infants had severe UAO, and five infants had BPD. Spontaneous breathing pattern showed high values of fR, TI/TTOT ratio and respiratory effort, as shown by the increase in the Poes and Pgas swings, PTPoes/breath and PTPdi/breath, and PTPoes/min and PTPdi/min (Tables 2 and 3 and Figures 1 and 2). The mean fR tended to be slightly higher for the BPD patients (54 ± 6 breaths/min) than in the UAO patients (51 ± 12 breaths/min; P = 0.612) and their TI/TTOT ratio was significantly lower (0.4 ± 0.1 vs 0.6 ± 0.1 for the UAO patients; P = 0.003). All the markers of respiratory effort, except the Pgas swing during expiration (7 ± 4 vs 1 ± 2 for the UAO patients; P = 0.03), were statistically comparable.
Table 1.
Patients | Gender | Age (mo) | Weight (kg) | Underlying disease |
---|---|---|---|---|
1 | Girl | 2 | 3.6 | Pierre Robin syndrome |
2 | Boy | 2 | 5.0 | Prader-Willi syndrome |
3 | Girl | 10 | 6.7 | Obstructive sleep apnoea |
4 | Boy | 2 | 4.5 | Laryngomalacia and Down syndrome |
5 | Boy | 3 | 5.1 | Laryngomalacia |
6 | Boy | 4 | 5.2 | Laryngomalacia |
7 | Girl | 13 | 9.0 | Laryngomalacia |
8 | Boy | 5 | 4.4 | Bronchopulmonary dysplasia |
9 | Boy | 14 | 8.3 | Bronchopulmonary dysplasia |
10 | Girl | 20 | 8.5 | Bronchopulmonary dysplasia |
11 | Boy | 22 | 9.0 | Bronchopulmonary dysplasia |
12 | Boy | 22 | 10.3 | Bronchopulmonary dysplasia |
Table 2.
Spontaneous breathing | Clinical CPAP | Physiological CPAP | Physiological CPAP+1 cmH2O | |
---|---|---|---|---|
Respiratory parameters | ||||
fR (breaths/min) | 49 ± 14 | 44 ± 21 | 38 ± 16b | 36 ± 18 |
TI/TTOT | 0.5 ± 0.1 | 0.4 ± 0.1 | 0.4 ± 0.1c | 0.4 ± 0.1 |
Respiratory muscle output | ||||
Poes swing (cmH2O) | 31 ± 15 | 21 ± 10d | 11 ± 5b | 14 ± 6 |
Pdi swing (cmH2O) | 26 ± 8 | 20 ± 8d | 12 ± 6b | 15 ± 7 |
Pgas swing (cmH2O) | 4 ± 2 | 3 ± 1 | 2 ± 1c | 2 ± 1 |
Expiratory Pgas swing (cmH2O) | 4 ± 4 | 3 ± 5 | 1 ± 1 | 2 ± 2 |
PTPoes/breath (cmH2O/s) | 12 ± 6 | 9 ± 6 | 6 ± 5b | 9 ± 9 |
PTPdi/breath (cmH2O/s) | 11 ± 4 | 9 ± 6 | 6 ± 4b | 8 ± 6 |
PTPoes/min (cmH2O/s/min) | 571 ± 340 | 359 ± 227 | 183 ± 119e | 244 ± 161 |
PTPdi/min (cmH2O/s/min) | 452 ± 146 | 350 ± 164 | 201 ± 108b | 239 ± 173 |
CPAP level (cmH2O) | NA | 8 ± 2 | 10 ± 2 | 11 ± 2 |
aCPAP: continuous positive airway pressure, fR: respiratory rate, NA: not applicable, Poes: oesophageal pressure, Pdi: transdiaphragmatic pressure, Pgas: gastric pressure, PTP: pressure-time product, TI/TTOT: inspiratory time/total respiratory cycle time ratio. bP < 0.05 between physiological CPAP and spontaneous breathing or clinical CPAP. cP < 0.05 between spontaneous breathing and physiological CPAP. dP < 0.05 between spontaneous breathing and clinical CPAP. eP < 0.05 between physiological CPAP and spontaneous breathing, physiological CPAP or physiological CPAP+1 cmH2O.
Table 3.
UAO (n = 7) | Spontaneous breathing | Clinical CPAP | Physiological CPAP | Physiological CPAP+1 cmH2O |
---|---|---|---|---|
Respiratory parameters | ||||
fR (breaths/min) | 46 ± 16b | 36 ± 19 | 34 ± 17 | 32 ± 20 |
TI/TTOT | 0.6 ± 0.1b | 0.5 ± 0.1 | 0.4 ± 0.1 | 0.4 ± 0.1 |
Respiratory muscle output | ||||
Poes swing (cmH2O) | 30 ± 17b | 18 ± 10 | 12 ± 6 | 16 ± 7 |
Pdi swing (cmH2O) | 25 ± 9b | 20 ± 10c | 14 ± 7 | 17 ± 8 |
Pgas swing (cmH2O) | 5 ± 2b | 3 ± 1 | 3 ± 1 | 3 ± 1 |
Expiratory Pgas swing (cmH2O) | 1 ± 2 | 1 ± 1 | 1 ± 1 | 2 ± 2 |
PTPoes/breath (cmH2O/s) | 13 ± 7 | 10 ± 8 | 7 ± 6 | 11 ± 10 |
PTPdi/breath (cmH2O/s) | 12 ± 5 | 11 ± 7 | 7 ± 5 | 10 ± 7 |
PTPoes/min (cmH2O/s/min) | 599 ± 415b | 291 ± 162 | 201 ± 150 | 271 ± 188 |
PTPdi/min (cmH2O/s/min) | 459 ± 155b | 332 ± 161c | 212 ± 135 | 262 ± 193 |
CPAP level (cmH2O) | NA | 9 ± 1 | 11 ± 2 | 12 ± 2 |
BPD (n = 5) | Spontaneous Breathing | Clinical CPAP | Physiological CPAP | Physiological CPAP+1 cmH2O |
Respiratory parameters | ||||
fR (breaths/min) | 54 ± 6 | 56 ± 18 | 45 ± 13 | 46 ± 11 |
TI/TTOT | 0.4 ± 0.0 | 0.4 ± 0.0 | 0.4 ± 0.1 | 0.4 ± 0.1 |
Respiratory muscle output | ||||
Poes swing (cmH2O) | 32 ± 11d | 25 ± 11c | 10 ± 2 | 11 ± 2 |
Pdi swing (cmH2O) | 27 ± 9d | 20 ± 5c | 11 ± 4 | 11 ± 1 |
Pgas swing (cmH2O) | 2 ± 1 | 2 ± 1 | 2 ± 1 | 2 ± 0 |
Expiratory Pgas swing (cmH2O) | 7 ± 4 | 7 ± 6 | 1 ± 1 | 1 ± 1 |
PTPoes/breath (cmH2O/s) | 10 ± 4d | 9 ± 4c | 4 ± 2 | 4 ± 2 |
PTPdi/breath (cmH2O/s) | 9 ± 4 | 7 ± 3 | 5 ± 2 | 3 ± 0 |
PTPoes/min (cmH2O/s/min) | 526 ± 201d | 468 ± 290c | 156 ± 43 | 181 ± 43 |
PTPdi/min (cmH2O/s/min) | 442 ± 153 | 379 ± 183 | 183 ± 50 | 156 ± 0 |
CPAP level (cmH2O) | NA | 7 ± 1 | 9 ± 1 | 10 ± 1 |
aBPD: bronchopulmonary dysplasia, CPAP: continuous positive airway pressure, fR: respiratory rate, NA: not applicable, Poes: oesophageal pressure, Pdi: transdiaphragmatic pressure, Pgas: gastric pressure, PTP: pressure time product, TI/TTOT: inspiratory time/total respiratory cycle time ratio, UAO: upper airway obstruction. bP < 0.05 between spontaneous breathing and clinical CPAP or physiological CPAP or physiological CPAP+1 cmH2O. cP < 0.05 between clinical CPAP and physiological CPAP. dP < 0.05 between spontaneous breathing and physiological CPAP or physiological CPAP+1 cmH2O.
The mean clinical CPAP level was 8 ± 2 cmH2O (range, 6 to 10 cmH2O) (Figure 3). This level was associated with a significant decrease in Poes and Pdi swings (Table 2). Of note, the mean clinical CPAP level was significantly higher for the UAO patients (9 ± 1 cmH2O) than for the BPD patients (7 ± 1 cmH2O) (P = 0.006) (Table 3).
The mean physiological CPAP level was 2 ± 2 cmH2O higher than the mean clinical CPAP level and was associated with a significantly greater improvement in fR, TI/TTOT ratio and all the indices of respiratory effort as compared to spontaneous breathing (Table 2). Interestingly, the mean physiological CPAP level was not significantly different between the two groups (with a mean level of CPAP of 11 ± 2 cmH2O and 9 ± 1 cmH2O for the UAO and BPD patients, respectively; P = 0.12). The physiological CPAP+1 was associated with a nonsignificant increase in the Poes swing and the PTPoes/min, together with agitation and arousals in all the patients.
During spontaneous breathing, one of the five UAO patients and all four BPD patients had a Pgas swing during expiration (Pgas was not available for three patients due to technical reasons) (Figure 4). During the clinical CPAP setting, three of the seven UAO patients and all five BPD patients had a Pgas swing during expiration. During the physiological CPAP setting, this expiratory abdominal activity disappeared in one of the five BPD patients but persisted or increased in two UAO patients (Figure 4). During the physiological CPAP+1 setting, the expiratory abdominal activity increased in these two UAO patients, slightly reappeared in two other UAO patients and increased moderately in the two BPD patients (Pgas was not available in the other three BPD patients).
The physiological CPAP level was chosen for the patients. The clinical tolerance of this level was excellent, with all patients sleeping more than 6 h per night with CPAP and the normalization of nocturnal gas exchange, defined by a minimal SpO2 greater than 95% and a maximal PtcCO2 less than 50 mmHg. All the patients were successfully discharged to home on long-term nasal CPAP, and all were weaned from CPAP after a median period of 10 mo (range, 6 mo to 3 yr). CPAP was associated with an improvement in weight in all the patients. Three patients could be weaned from nasogastric nutritional support after 1 to 2 mo, and, in the other nine patients, CPAP was associated with an at least 1 SD weight gain after 1 to 3 mo. None of the patients required a tracheotomy, and no deaths occurred.
Discussion
This study shows that a CPAP setting based on clinical noninvasive parameters is associated with an improvement in breathing pattern and respiratory effort in infants with severe UAO or BPD, but this improvement is significantly greater with an invasive physiological setting based on the monitoring of Poes and Pgas tracings.
We evaluated two groups of patients. The first group had severe UAO, in whom noninvasive CPAP has proven to be efficacious [3,6]. The second group comprised patients with severe BPD. We acknowledge that noninvasive CPAP is not a classic treatment for these infants, but the patients evaluated for CPAP in the present study represent a highly selected group of patients with extremely severe lung disease [21,22], as documented by the tremendous increase in work of breathing and alveolar hypoventilation. These patients are expected to have a combination of lower airway obstruction and parenchymal lung disease. The lower airway obstruction causes hyperinflation and PEEPi, which is effectively counteracted by CPAP, unloading the respiratory muscles and improving breathing pattern and gas exchange [23]. Despite an older age, the mean fR tended to be higher and the mean TI/TTOT ratio lower in the BPD patients than in the UAO patients, which may be explained by the predominantly lower airway obstruction in BPD patients. This may also explain the high prevalence of abnormal expiratory abdominal activity during spontaneous breathing in BPD patients.
The mean difference between the clinical CPAP and physiological CPAP levels was 2 ± 2 cmH2O. A 1 cmH2O increase in this physiological CPAP level was not tolerated by the infants with clinical arousal and an increase in expiratory abdominal effort, as assessed on the Pgas tracing. This narrow therapeutic CPAP range is similar to our findings in infants hospitalized for acute viral bronchiolitis [9]. These observations underline the sensitivity of these infants to small changes in the CPAP level.
Our findings regarding the expiratory Pgas recording are particularly interesting. Expiratory abdominal activity can be due to hyperinflation induced by CPAP overtitration, as shown by studies in healthy awake subjects in whom high levels of CPAP induced an expiratory abdominal activity [24-26]. In our group of patients, however, our hypothesis is that this expiratory abdominal activity was the consequence of persistent airway obstruction during spontaneous breathing and CPAP. Indeed, four of nine of our patients had expiratory abdominal activity during spontaneous breathing, which persisted during the clinical CPAP setting, as observed by Lofaso et al. [19]. Indeed, these authors monitored Poes and Pgas with airflow during nasal CPAP titration in 12 adult patients with OSA and observed that 6 patients presented expiratory abdominal activity during spontaneous breathing. This expiratory abdominal activity persisted at the CPAP level associated with the disappearance of apnoea and disappeared only when a higher CPAP level was reached [19]. The importance of this CPAP undertitration has recently been highlighted by Seo et al. [27]. These authors analysed the expiratory abdominal muscle activity by means of surface electrodes in 81 patients with OSA and periodic limb movements (PLMs) during CPAP titration [27]. PLMs persisted in 72 patients while the AASM-defined hypopnoea were eliminated [7]. These subjects still showed a pattern of flow limitation and the presence of cyclic alternative pattern during non-rapid eye movement sleep. The disappearance of flow limitation required a higher CPAP pressure and was associated with the disappearance of PLMs. When comparing the CPAP pressure necessary to eliminating AASM-defined hypopnoea and the pressure required to eliminating PLMs, Seo et al. found a mean difference of 1.47 ± 1.96 cmH2O. Forty-four patients needed 2 cmH2O or greater CPAP increase for the disappearance of flow limitation and PLM. The clinical relevance of this high-target CPAP titration seems important, as PLMs may be associated with evidence of electroencephalographic changes of arousal and with changes in the plethysmographic curve obtained from a finger oximeter, which indicates sympathetic stimulation [28]. Of note, the mean CPAP difference in the present study was similar to that observed in adult patients [27].
Our data underline the importance of expiratory abdominal activity for optimal titration of CPAP in infants, but an oesogastric pressure recording is not available in most centres. Surface electromyography of the abdominal muscles is probably a promising tool and worth being developed and validated for use in young infants. In the meantime, it has to be noted that the clinical CPAP level was associated with an improvement in breathing pattern and respiratory effort, as well as excellent clinical tolerance, with the majority of the patients falling asleep. A first recommendation may thus be to set CPAP at the level associated with the disappearance of the stridor, SpO2 greater than 95%, the greatest decrease in heart rate and fR, and the best comfort as assessed by the infant falling asleep. As the mean difference between the clinical CPAP level and the physiological CPAP level was about 2 cmH2O, we recommend increasing the clinical CPAP level by 1 to 2 cmH2O if this level remains associated with the disappearance of the stridor, a SpO2 greater than 95%, the greatest decrease in heart rate and fR and the best comfort as assessed by the infant's remaining or falling asleep.
Our study has several limitations. We were unable to measure airflow and thus could not accurately report the mechanical inspiratory time during spontaneous breathing. Indeed, the increase in dead space following the application of the nasal mask resulted in an increase in respiratory effort and dyspnoea associated with a fall in SpO2. Also, the occurrence of a continuous high airflow during CPAP precluded the measurement of the patient's own airflow. We did not monitor gas exchange in all the infants during the study period. All the infants had severe hypoxaemia and hypercapnia during sleep, which was alleviated by nasal CPAP as previously reported by our group in a comparable group of patients [3,5,6]. Despite the fact that the infants fell asleep with the nasal CPAP, we did not perform an overnight recording. This is a limitation, as the most efficient pressure level may change with the sleep stage. Different CPAP devices were used. CPAP devices, as is true of other home ventilators, evolve very rapidly. As CPAP is the most simple ventilator mode, delivering only a constant positive pressure, the choice of the device is mainly a question of ergonometry, prescriber or patient preference. Indeed, we did not use more recent CPAP modes, such as bilevel, auto-adjusted and BiPAP Bi-Flex modes, which are not adapted for infants as we have shown previously [5]. We acknowledge that a small difference may be observed between the CPAP level set on the device and the Paw measured by the pressure transducer. In the present study, however, the CPAP level reported was the Paw and not the level given by the CPAP device. We did not perform a polysomnography and are thus unable to confirm the association between persistent respiratory effort and cortical arousal.
Conclusions
This study shows that a physiological setting of noninvasive CPAP, based on the recording of Poes and Pgas, may be superior to a clinical setting based on clinical noninvasive parameters. Expiratory abdominal activity is worth detecting, underlining the importance of the development and validation of noninvasive, reliable and accurate markers of abdominal activity for an optimal setting of noninvasive CPAP in infants.
Key messages
• In routine care, CPAP level should be adjusted on the disappearance of the stridor, retraction and/or dyspnoea, a pulse oximetry greater than 95%, the greatest decrease in heart and respiratory rates, and the best comfort as assessed by the infant's falling asleep.
• A setting of CPAP based on the recording of Poes and Pgas may be superior to a clinical setting.
• Expiratory abdominal activity may be helpful for an optimal titration of CPAP in infants.
Abbreviations
AASM: American Academy of Sleep Medicine; BPD: bronchopulmonary dysplasia; CPAP: continuous positive airway pressure; fR: respiratory rate; OSA: obstructive sleep apnoea; Paw: airway pressure; Pdi: transdiaphragmatic pressure; PEEPi: intrinsic positive end-expiratory pressure; Pgas: gastric pressure; PLM: periodic limb movement; Poes: oesophageal pressure; PtcCO2: transcutaneous carbon dioxide pressure; PTPoes: oesophageal pressure-time product; PTPdi: diaphragmatic pressure time product; SpO2: pulse oximetry; TI/TTOT: inspiratory time/total respiratory cycle time; UAO: upper airway obstruction
Competing interests
The authors declare that they have no competing interests.
Authors' contributions
SK performed the analysis of tracings and the redaction of the manuscript. AR and SA helped with recording of the patients, analysis of the data and redaction of the manuscript. NL contributed to the recruitment of the patients, analysis of tracings and redaction of the manuscript. AP performed the nasal masks and contributed to the analysis of tracings and redaction of the manuscript. BF contributed to the study concept and design, recruitment of the patients, analysis of tracings and redaction of the manuscript. All authors read and approved the final manuscript.
Contributor Information
Sonia Khirani, Email: sonia_khirani@yahoo.fr.
Adriana Ramirez, Email: a.ramirez@adepassistance.fr.
Sabrina Aloui, Email: aloui.sabrina1@gmail.com.
Nicolas Leboulanger, Email: nicolas.leboulanger@trs.aphp.fr.
Arnaud Picard, Email: a.picard@trs.aphp.fr.
Brigitte Fauroux, Email: brigitte.fauroux@trs.aphp.fr.
Acknowledgements
The research of Brigitte Fauroux is supported by the Association Française contre les Myopathies (AFM), Vaincre la Mucoviscidose (VLM), Assistance Publique-Hôpitaux de Paris, Inserm, Université Pierre et Marie Curie-Paris 6, ADEP Assistance, S2A Santé, and IP Santé Domicile.
References
- Guilleminault C, Nino-Murcia G, Heldt G, Baldwin R, Hutchinson D. Alternative treatment to tracheostomy in obstructive sleep apnea syndrome: nasal continuous positive airway pressure in young children. Pediatrics. 1986;17:797–802. [PubMed] [Google Scholar]
- Guilleminault C, Pelayo R, Clerk A, Leger D, Bocian RC. Home nasal continuous positive airway pressure in infants with sleep-disordered breathing. J Pediatr. 1995;17:905–912. doi: 10.1016/S0022-3476(95)70026-9. [DOI] [PubMed] [Google Scholar]
- Fauroux B, Pigeot J, Polkey MI, Roger G, Boulé M, Clément A, Lofaso F. Chronic stridor caused by laryngomalacia in children: work of breathing and effects of noninvasive ventilatory assistance. Am J Respir Crit Care Med. 2001;17:1874–1878. doi: 10.1164/ajrccm.164.10.2012141. [DOI] [PubMed] [Google Scholar]
- Waters WA, Everett FM, Bruderer JW, Sullivan CE. Obstructive sleep apnea: the use of nasal CPAP in 80 children. Am J Respir Crit Care Med. 1995;17:780–785. doi: 10.1164/ajrccm.152.2.7633742. [DOI] [PubMed] [Google Scholar]
- Essouri S, Nicot F, Clément A, Garabedian EN, Roger G, Lofaso F, Fauroux B. Noninvasive positive pressure ventilation in infants with upper airway obstruction: comparison of continuous and bilevel positive pressure. Intensive Care Med. 2005;17:574–580. doi: 10.1007/s00134-005-2568-6. [DOI] [PubMed] [Google Scholar]
- Leboulanger N, Picard A, Soupre V, Aubertin G, Denoyelle F, Galliani E, Roger G, Garabedian EN, Fauroux B. Physiological and clinical benefits of noninvasive respiratory support in infants with Pierre Robin sequence. Pediatrics. 2010;17:e1056–e1063. doi: 10.1542/peds.2010-0856. [DOI] [PubMed] [Google Scholar]
- Iber C, Ancoli-Israel S, Chesson AL, Jr, Quan SF. AASM Manual for the Scoring of Sleep and Associated Events: Rules, Terminology and Technical Specifications. Westchester, IL: American Academy of Sleep Medicine; 2007. for the American Academy of Sleep Medicine. [Google Scholar]
- Kushida CA, Chediak A, Berry RB, Brown LK, Gozal D, Iber C, Parthasarathy S, Quan SF, Rowley JA. Positive Airway Pressure Titration Task Force of the American Academy of Sleep Medicine: Clinical guidelines for the manual titration of positive airway pressure in patients with obstructive sleep apnea. J Clin Sleep Med. 2008;17:157–171. [PMC free article] [PubMed] [Google Scholar]
- Essouri S, Durand P, Chevret L, Balu L, Devictor D, Fauroux B, Tissières P. Optimal level of nasal continuous positive airway pressure in severe viral bronchiolitis. Intensive Care Med. 2011;17:2002–2007. doi: 10.1007/s00134-011-2372-4. [DOI] [PubMed] [Google Scholar]
- Fauroux B, Lavis JF, Nicot F, Picard A, Boelle PY, Clément A, Vazquez MP. Facial side effects during noninvasive positive pressure ventilation in children. Intensive Care Med. 2005;17:965–969. doi: 10.1007/s00134-005-2669-2. [DOI] [PubMed] [Google Scholar]
- Ramirez A, Delord V, Khirani S, Leroux K, Cassier S, Kadlub N, Aubertin A, Picard A, Fauroux B. Interfaces for long term noninvasive positive pressure ventilation in children. Intensive Care Med. 2012;17:655–662. doi: 10.1007/s00134-012-2516-1. [DOI] [PubMed] [Google Scholar]
- Fauroux B, Pigeot J, Polkey MI, Isabey D, Clément A, Lofaso F. In vivo physiological comparison of two ventilators used for domiciliary ventilation in children with cystic fibrosis. Crit Care Med. 2001;17:2097–2105. doi: 10.1097/00003246-200111000-00009. [DOI] [PubMed] [Google Scholar]
- Stell IM, Tompkins S, Lovell AT, Goldstone JC, Moxham J. An in vivo comparison of a catheter mounted pressure transducer system with conventional balloon catheters. Eur Respir J. 1999;17:1158–1163. doi: 10.1034/j.1399-3003.1999.13e36.x. [DOI] [PubMed] [Google Scholar]
- Baydur A, Behrakis PK, Zin WA, Jaeger MJ, Milic-Emili JA. A simple method for assessing the validity of the esophageal balloon technique. Am Rev Respir Dis. 1982;17:788–791. doi: 10.1164/arrd.1982.126.5.788. [DOI] [PubMed] [Google Scholar]
- Sassoon CSH, Light RW, Lodia R, Sieck GC, Mahutte CK. Pressure-time product during continuous positive airway pressure, pressure support ventilation and T-piece during weaning from mechanical ventilation. Am Rev Respir Dis. 1991;17:459–475. doi: 10.1164/ajrccm/143.3.469. [DOI] [PubMed] [Google Scholar]
- Field S, Grassino A, Sanci S. Respiratory muscle oxygen consumption estimated by the diaphragm pressure-time index. J Appl Physiol. 1984;17:44–51. doi: 10.1152/jappl.1984.57.1.44. [DOI] [PubMed] [Google Scholar]
- Barnard PA, Levine S. Critique on application of diaphragmatic time-tension index to spontaneously breathing humans. J Appl Physiol. 1986;17:1067–1072. doi: 10.1152/jappl.1986.60.3.1067. [DOI] [PubMed] [Google Scholar]
- Parthasarathy S, Jubran A, Tobin MJ. Assessment of neural inspiratory time in ventilator-supported patients. Am J Respir Crit Care Med. 2000;17:546–552. doi: 10.1164/ajrccm.162.2.9901024. [DOI] [PubMed] [Google Scholar]
- Lofaso F, d'Ortho M, Fodil F, Delclaux C, Harf A, Lorino A. Abdominal muscle activity in sleep apnea during continuous positive airway pressure titration. Chest. 2001;17:390–396. doi: 10.1378/chest.120.2.390. [DOI] [PubMed] [Google Scholar]
- Campbell EJM, Green JH. The variations in intra-abdominal pressure and the activity of the abdominal muscles during breathing; a study in men. J Physiol. 1953;17:282–290. doi: 10.1113/jphysiol.1953.sp004999. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tiddens HA, Hofhuis W, Casotti V, Hop WC, Hulsmann AR, de Jongste JC. Airway dimensions in bronchopulmonary dysplasia: implications for airflow obstruction. Pediatr Pulmonol. 2008;17:1206–1213. doi: 10.1002/ppul.20928. [DOI] [PubMed] [Google Scholar]
- Hilgendorff A, Reiss I, Gortner L, Schüler D, Weber K, Lindemann H. Impact of airway obstruction on lung function in very preterm infants at term. Pediatr Crit Care Med. 2008;17:629–635. doi: 10.1097/PCC.0b013e31818d17c8. [DOI] [PubMed] [Google Scholar]
- Giovannini-Chami L, Khirani S, Thouvenin G, Ramirez A, Fauroux B. Work of breathing to optimize noninvasive ventilation in bronchiolitis obliterans. Intensive Care Med. 2012;17:722–724. doi: 10.1007/s00134-012-2469-4. [DOI] [PubMed] [Google Scholar]
- Agostoni E. Diaphragm activity and thoracoabdominal mechanics during positive pressure breathing. J Appl Physiol. 1962;17:215–220. doi: 10.1152/jappl.1962.17.2.215. [DOI] [PubMed] [Google Scholar]
- Alex CG, Aronson RM, Onal E, Lopata M. Effects of continuous positive airway pressure on upper airway and respiratory muscle activity. J Appl Physiol. 1987;17:2026–2030. doi: 10.1152/jappl.1987.62.5.2026. [DOI] [PubMed] [Google Scholar]
- Layon J, Banner MJ, Jaeger MJ, Peterson CV, Gallagher TJ, Modell JH. Continuous positive airway pressure and expiratory positive airway pressure increase functional residual capacity equivalently. Chest. 1986;17:517–521. doi: 10.1378/chest.89.4.517. [DOI] [PubMed] [Google Scholar]
- Seo WH, Guilleminault C. Periodic leg movement, nasal CPAP, and expiratory muscles. Chest. 2012;17:111–118. doi: 10.1378/chest.11-1563. [DOI] [PubMed] [Google Scholar]
- Sforza E, Nicolas A, Lavigne G, Gosselin A, Petit D, Montplaisir J. EEG and cardiac activation during periodic leg movements in sleep: support for a hierarchy of arousal responses. Neurology. 1999;17:786–791. doi: 10.1212/WNL.52.4.786. [DOI] [PubMed] [Google Scholar]