Abstract
Whether failure in unicompartmental knee arthroplasty (UKA) is related to implant design remains unclear. We hypothesize that current available UKAs fit within 2 mm. Forty-eight CTs of cadaveric knees were compared to current available UKA brands. Overall no-fit compared to at least one component within 2 mm is high (91.7%) and worse for males (100%) compared to females (83.3%). Good fit was observed for the medial but not for the lateral tibia plateau. Seven males (29.2%) had larger dimensions of more than 2 mm. For the widest UKA brand, 12 (57%) males and 2 females (8. 3%) had lateral femoral condyles 3 mm larger. Current UKA's in our sample population fit less on the lateral tibia and on femoral condyles.
Keywords: knee arthroplasty, component fit, condylar dimensions, unicompartmental knee arthroplasty, medial and lateral condyle
Introduction
Unicompartmental knee arthroplasty (UKA) has good long-term survivorship and is gaining popularity (15,17). Despite excellent clinical results (5,13,14,17), early failure rates of UKA remain in the 5% range (6,8,11,18). Whether this is related to surgical technique, patient selection or implant design needs clarification. Some cohorts, such as obese patients have shown higher failure rates with specific, narrow implants (2).
We wonder whether femoral condyles and tibial plateaus are sufficiently covered with current available UKA designs. We hypothesize all current available components of modern UKA fit within 2 or 3 mm medial and lateral femoral condyles and medial and lateral tibia plateaus.
Materials and Methods
Height, weight, and knee dimension sizes of a sample of cadaveric subjects were easured (Table I). All measurements were made 5 mm below the deepest point of the articular surface of the medial and lateral tibial plateau. The sample was composed of 24 female and 24 male cadavers, showing no significant arthritic changes. No details on age or race were available. We computed the mean and standard deviation of sample characteristics and performed t-tests to test for differences between male and female subjects. Using paired ttests we also compared lateral and medial knee dimensions and tested for differences in the whole sample, among males only, and among females only. To examine how well the component dimensions defined two cutoffs for determining whether components would adequately fit the subject knee, a 2 mm no-fit was defined as a UKA component dimension at least 2 mm smaller than the measured knee dimension.
Table I. Sample description and comparison of male to female measures.
| Total | Males | Females | P-Value Comparing Males to Females | |
|---|---|---|---|---|
| Height (cm) | 168.7(10.3) | 173.6(9.1) | 163.4(8.9) | p<0.001 |
| Weight (kg) | 73.8(21.9) | 74.9(22.2) | 72.6(20.6) | p=0.715 |
| Medial Tibia AP Length | 5.06(0.46) | 5.37(0.38) | 4.75(0.29) | p<0.001 |
| Medial Tibia ML Width | 3.04(0.32) | 3.27(0.25) | 2.82(0.19) | p<0.001 |
| Lateral Tibia AP Length | 4.74(0.46) | 5.03(0.34) | 4.45(0.38) | p<0.001 |
| Lateral Tibia ML Width | 3.21(0.32) | 3.41(0.26) | 3.01(0.23) | p<0.001 |
| Medial Condyle AP Length | 5.73(0.45) | 6.01(0.33) | 5.45(0.37) | p<0.001 |
| Medial Condylar ML Width | 2.61(0.29) | 2.80(0.23) | 2.43(0.22) | p<0.001 |
| Lateral Condyle AP Length | 6.23(0.51) | 6.55(0.35) | 5.92(0.45) | p<0.001 |
| Lateral Condylar ML Width | 2.85(0.33) | 3.09(0.25) | 2.61(0.19) | p<0.001 |
| Med/Fem Art. Surface AP Length | 4.84(0.41) | 5.04(0.35) | 4.65(0.38) | p=0.001 |
| Lat/Fem Art. Surface AP Length | 4.46(0.47) | 4.71(0.41) | 4.22(0.41) | p<0.001 |
| Lateral Tibia AP/ML Ratio | 1.48(0.09) | 1.48(0.11) | 1.48(0.08) | p=0.869 |
| Medial Tibia AP/ML Ratio | 1.67(0.09) | 1.64(0.10) | 1.69(0.08) | p=0.093 |
| Lateral Condyle AP/ML Ratio | 2.2(0.17) | 2.13(0.17) | 2.27(0.12) | p=0.002 |
| Medial Condyle AP/ML Ratio | 2.21(0.18) | 2.16(0.19) | 2.25(0.18) | p=0.080 |
We computed the proportion of subjects with at least one knee dimension 2mm larger than at least one brand of UKA component. We computed the proportion of the sample with knee dimensions at least 2 mm larger and compared measurements to current available UKA systems (Table II & III). We focused on the two largest models (DePuy HP and Zimmer HF). The DePuy HP have the longest and widest tibial and Zimmer HF have the widest femoral condyle dimensions. Of the subjects whose knees are at least 2 mm larger than the DePuy and Zimmer components we determined how often it was the medial or lateral condyle ML width which was too wide or how often the tibial AP medial and lateral lengths were too long, and how often the tibial ML medial and lateral widths were too wide (Table IV).
Table II.
| a: Tibial implant dimensions (cm) of available UKAs | ||||||
|---|---|---|---|---|---|---|
| Size 1 | Size 2 | Size 3 | Size 4 | Size 5 | Size 6 | |
| AP - ML (Ratio) | AP - ML (Ratio) | AP - ML (Ratio) | AP - ML (Ratio) | AP - ML (Ratio) | AP - ML (Ratio) | |
| [1] Stryker Triathlon | 4.1-2.3(1.78) | 4.4-2.5(1.76) | 4.7-2.7(1.74) | 5.0-2.9(1.72) | 5.3-3.1(1.71) | 5.6-3.3(1.70) |
| [2] Depuy Sigma® High Performance Partial Knee System | 4.2-2.4(1.75) | 4.5-2.6(1.73) | 4.8-2.8(1.71) | 5.1-3.0(1.70) | 5.4-3.2(1.69) | 5.7-3.4(1.68) |
| [3] Zimmer Unicompartmental High Flex Knee System | 4.1-2.3(1.78) | 4.4-2.5(1.76) | 4.7-2.7(1.74) | 5.0-2.9(1.72) | 5.3-3.1(1.71) | 5.6-3.3(1.70) |
| [4] Smith and Nephew Journey | 3.8-2.4(1.58) | 4.2-2.5(1.68) | 4.6-2.7(1.70) | 4.9-2.9(1.69) | 5.2-3.0(1.73) | 5.5-3.2(1.72) |
| [5] Biomet Oxford | 3.8-2.6(1.46) | 4.1-2.6(1.58) | 4.4-2.8(1.57) | 4.7-3.0(1.57) | 5-3.2(1.56) | 5.3-3.4(1.56) |
| [6] Wright Advance | 4.0-2.4(1.67) | 4.4-2.6(1.69) | 4.9-2.9(1.69) | 5.4-3.3(1.64) | ||
| b: Femoral implant dimensions of available UKAs | ||
|---|---|---|
| Femoral ML Width | ||
| Min (cm) | Max (cm) | |
| [1] Stryker Triathlon | 1.9 | 2.4 |
| [2] Depuy Sigma® High Performance Partial Knee System | 1.8 | 2.5 |
| [3] Zimmer Unicompartmental High Flex Knee System | 2.1 | 2.6 |
| [4] Smith and Nephew Journey | 1.8 | 2.5 |
| [5] Biomet Oxford | 1.9 | 2.3 |
| [6] Wright Advance | 1.9 | 2.2 |
Table III. Proportion of sample with 2 mm-fits of components1 to knee dimensions.
| Total | Males | Females | ||||
|---|---|---|---|---|---|---|
| 2mm-Fit N(%) | 2mm-No Fit N(%) | 2mm-Fit N(%) | 2mm-No Fit N(%) | 2mm-Fit N(%) | 2mm-No Fit N(%) | |
| Any component2 | 4 (8.3) | 44 (91.7) | 0 (0.0) | 24 (100) | 4 (16.7) | 20 (83.3) |
| Depuy3 | 13 (27.1) | 35 (72.9) | 0 (0.0) | 24 (100) | 13 (54.2) | 11 (45.8) |
| Zimmer4 | 20 (41.7) | 28 (58.3) | 2 (8.3) | 22 (91.7) | 18 (75.0) | 6 (25.0) |
2mm-Fit defined as subject knee dimension lies within +/- 2mm of component dimension, 2mm-No Fit defined as subject knee dimension larger than component dimension +2mm.
Any component displays number and proportion of sample with at least one knee dimension larger than at least one of the available component sizes.
Depuy has longest Tibial AP Length and Tibial ML Width.
Zimmer has widest Femoral condylar ML Width.
Table IV. Dimensions where Depuy and Zimmer components are 2mm smaller than subject measures.
| Depuy Dimensions 2 mm smaller than subject measure1 | Zimmer Dimensions 2 mm larger than subject measure 2 | |||
|---|---|---|---|---|
| Males N(%) | Females N(%) | Males N(%) | Females N(%) | |
| Total | 22 (100.0) | 11 (45.8) | 22 (91.7) | 6 (25.0) |
| Femoral condyle3 ML Width | ||||
| Medial | 17 (70.8) | 4 (16.7) | 16 (66.7) | 2 (8.3) |
| Lateral | 23 (95.8) | 10 (41.7) | 21 (87.5) | 5 (20.8) |
| Tibia AP Length | ||||
| Medial | 2 (8.3) | 0 (0.0) | 3 (12.5) | 0 (0.0) |
| Lateral | 0 (0.0) | 0 (0.0) | 1 (4.2) | 0 (0.0) |
| Tibia ML Width | ||||
| Medial | 3 (12.5) | 0(0.0) | 6 (25.0) | 0 (0.0) |
| Lateral | 7 (29.2) | 0(0.0) | 12 (50.0) | 1 (4.2) |
Depuy has longest Tibial AP Length and Tibial ML Width
Zimmer has widest Condylar ML Width
AP length of condylar component part not published – sole comparison for this part of the UKA component is the ML width
Results
The sample was composed of 24 female and 24 male cadavers. Height, weight and dimensions for medial, lateral femoral and tibial plateau are displayed in Table I. Differences of each measurement are included. While AP/ML ratios were similar between males and females on the lateral tibial plateau, there were differences for the medial tibial plateau and lateral femoral condyle. Among the total sample, lateral femoral condyle AP length was significantly longer than the medial condyle. The same relationship was also true for ML condyle width. Tibial lateral AP length was significantly shorter than the medial length while the lateral tibia ML width was longer than medial tibia ML width. Medial and lateral femoral condyle AP/ML ratios were similar while the medial tibia AP/ML ratio was significantly longer than the lateral ratio. The same relationships were true among males and females only.
Eighteen knees were measured by two observers describing observer agreement for the knee dimensions with means for coefficient of variation between 0.01 and 0.22. Our small coefficient of variations indicate low variability with respect to the size of the mean.
Current available UKA brands have different dimensions, different AP/ML ratios and vary between sizes. Tibial components AP lengths ranged from 3.8 cm to 5.7 cm and tibia ML width ranged from 2.3 cm to 3.4 cm, corresponding to AP/ML ratios ranging between 1.46 and 1.78 (Table III). Femoral ML width ranged from 1.8 cm to 2.6 cm. DePuy HP had the largest tibial component (AP length = 5.7 cm, ML width = 3.4 cm) while Zimmer HF had the largest femoral ML width (2.6 cm). All female tibial knee dimensions fall within +/- 2 mm of the range of UKA component dimensions for all UKA brands. Two knees exceeded the ML width dimensional range for all brands of UKA components while one exceeded all AP length dimensional ranges.
UKA dimensions are smaller than measured sizes and overall no-fit compared to any component within 2 mm is high (91.7%) and worse for males (100%) compared to females (83.3%) (Table IV). Comparing the UKA brand with the longest AP dimensions good fit was observed for the medial tibial plateau (Table IV) : two male knees (8.3%) had longer AP dimension of 2 mm. For the widest ML UKA brand three male knees (12.5%) had wider ML dimensions of 2 mm. For the lateral tibial plateau, AP fit within 2 mm was good and only 1 male knee (4.6%) exceeded 2 mm. The ML fit for the widest UKA brand showed 7 male knees (29.2%) had larger dimensions of more than 2 mm.
Coverage of the medial and lateral femoral condyle is worse. Femoral condyle ML widths were at least 2 mm longer than the widest femoral component for 16 (66.7%) of male medial condyles, 21 (87.5%) of male lateral condyles, 2 (8.3%) of female medial condyles, and 5 (20.8%) of female lateral condyles.
Discussion
As more encouraging long-term results after UKA are published, indications are extended to younger and heavier patients. Weight and activity have been discussed controversially in the literature and some implants reported no higher failure rate in heavier patients in resurfacing Marmor implants (4,5). Similar results were published for the Miller-Gallante UKA (1,9,15,16,22) but not with inset all-poly tibiae or narrow implants in patients with a BMI above 32 (2). We wondered whether, besides implant design and surgical technique, the dimensions of femoral and tibial implants could play a role and how well current available implants fit.
We hypothesized that all currently available components of modern UKA fit within 2 mm.
For the medial tibial plateau overall implant fit was better for females. Looking at the implant with the largest ML width, 12.5% had wider tibial plateaus by 2 mm. This describes the best case scenario using the largest implant in respect to AP length and ML width. We did not compute the coverage for smaller implants, which would result in worse coverage and higher percentages of patients with larger medial or lateral tibial plateaus. However, these numbers show that specifically for larger males, improvement of tibial coverage is desirable for both tibial plateaus but even more so on the lateral side.
The poor fit of femoral components should be cautiously interpreted. Both condyles have a different geometry with the medial femoral condyle being more curved and the lateral condyle being more straight. Given these geometric differences the implants have to be narrower, otherwise surgeons would not be able to rotate and place the components along the specific curvature of both condyles. Asymmetric components fit better on the medial side and worse laterally. A symmetric femoral component fits better onto the lateral condyle. No femoral component is designed for the lateral condyle with an anterior radius twice that of the medial side and being shorter anteriorly compared to the medial condyle (12,20).
Our measurements are comparable to other publications which studied tibial fit. Servien et al (19) measured tibial CT after 17 medial and 18 lateral UKA's. Average AP length was 50.8 mm for the medial and 47.2 mm for lateral plateau with a ML width of 28.8 medially and 29.3 on the lateral side. They calculated an AP/ML ratio of 1.8 medial and 1.6 lateral. Our measurements are similar, but our ML widths are wider. We measured CTs of normal non-arthritic knees 5 mm below the lowest point of medial and lateral tibial plateau, which may be slightly higher compared to Servien et al (19) measurements, since they measured the dimension just below the implant of CT. Servien et al sample was not well balanced, since 31 of 37 subjects were females (19). Servien et al observed that some tibial components matched the medial tibial plateau better than others and vice versa the lateral tibial condyle. Authors felt that better coverage could be a success factor. Surgeons tend to avoid medial overhang and therefore downsize the tibial implant by compromising anteroposterior coverage.
For the Korean population, smaller values are described in one publication for the medial tibia only : 47.1 mm for AP dimension (male 49.8 mm and female 47.1 mm) and 24.8 mm for ML dimension (26.1 mm for male and 23.5 mm for females) (21). Our measurements are similar for the female AP lengths, but not for the rest : our male AP measurements are 6 mm longer and our ML dimensions 6 mm wider (males 7 mm wider and females 5 mm). Surendan et al (21) concluded that for the Korean population UKA brands tend to oversize in the ML dimension. Fitzpatrick et al concluded that even theoretical optimized implants could not cover more than 76% of the exposed cortical rim (7). Insufficient tibial coverage may induce tibial plateau collapse (3,10) if forces are transmitted to cancellous bone. The described differences of various components are important and surgeons may benefit from knowing specific sizes in regard to AP length and ML width for medial, lateral femoral and tibial condyles.
Our study describes limitations of current available unicompartmental implants for both, medial and lateral, femoral and tibial condyles. The worst fit was observed for the lateral tibial plateau and for both femoral condyles. A different tibial implant, rounder with a lower AP/ML ratio would improve lateral tibial plateau fit. Design improvements for medial and lateral femoral condyles are difficult and limited due to the different geometry of medial and lateral femoral condyles. Surgeons should be aware of the consequences of using an asymmetric versus a symmetric femoral component for either medial or lateral femoral condyle and the consequences of poor fit. Surgeons should use implant sizes and their potential to improve implant fit for their different patients.
Footnotes
Wolfgang Fitz, MD is member of the Scientific Advisory Board of Conformis Inc, Bedford, MA, USA, he holds stocks or stock options of Conformis Inc., Bedford, MA and he receives royalties for intellectual property. Drs. Bliss and Losina have no related conflict with this study.
References
- 1.Argenson JN, Chevrol-Benkeddache Y, Aubaniac JM. Modern unicompartmental knee arthroplasty with cement : a three to ten-year follow-up study. J Bone Joint Surg. 2002;84-A:2235–2239. [PubMed] [Google Scholar]
- 2.Berend KR, Lombardi AV, Jr, Mallory TH, Adams JB, Groseth KL. Early failure of minimally invasive unicompartmental knee arthroplasty is associated with obesity. Clin Orthop Relat Res. 2005;440:60–66. doi: 10.1097/01.blo.0000187062.65691.e3. [DOI] [PubMed] [Google Scholar]
- 3.Bohm I, Landsiedl F. Revision surgery after failed unicompartmental knee arthroplasty : a study of 35 cases. J Arthroplasty. 2000;15:982–989. doi: 10.1054/arth.2000.9841. [DOI] [PubMed] [Google Scholar]
- 4.Cartier P, Cheaib S. Unicondylar knee arthroplasty. 2-10 years of follow-up evaluation. J Arthroplasty. 1987;2:157–162. doi: 10.1016/s0883-5403(87)80023-2. [DOI] [PubMed] [Google Scholar]
- 5.Cartier P, Sanouiller JL, Grelsamer RP. Unicompartmental knee arthroplasty surgery. 10-year minimum follow-up period. J Arthroplasty. 1996;11:782–788. doi: 10.1016/s0883-5403(96)80177-x. [DOI] [PubMed] [Google Scholar]
- 6.Dervin GF, Carruthers C, Feibel RJ, et al. Initial experience with the Oxford unicompartmental knee arthroplasty. J Arthroplasty. 2011;26:192–197. doi: 10.1016/j.arth.2010.02.007. [DOI] [PubMed] [Google Scholar]
- 7.Fitzpatrick C, Fitzpatrick D, Lee J, Auger D. Statistical design of unicompartmental tibial implants and comparison with current devices. Knee. 2007;14:138–144. doi: 10.1016/j.knee.2006.11.005. [DOI] [PubMed] [Google Scholar]
- 8.Gioe TJ, Killeen KK, Hoeffel DP, et al. Analysis of unicompartmental knee arthroplasty in a community-based implant registry. Clin Orthop Relat Res. 2003;416:111–119. doi: 10.1097/01.blo.0000093004.90435.d1. [DOI] [PubMed] [Google Scholar]
- 9.Lustig S, Parratte S, Magnussen RA, Argenson JN, Neyret P. Lateral unicompartmental knee arthroplasty relieves pain and improves function in posttraumatic osteoarthritis. Clin Orthop Relat Res. 2012;470:69–76. doi: 10.1007/s11999-011-1963-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.McAuley JP, Engh GA, Ammeen DJ. Revision of failed unicompartmental knee arthroplasty. Clin Orthop Relat Res. 2001;392:279–282. doi: 10.1097/00003086-200111000-00036. [DOI] [PubMed] [Google Scholar]
- 11.McGovern TF, Ammeen DJ, Collier JP, Currier BH, Engh GA. Rapid polyethylene failure of unicondylar tibial components sterilized with gamma irradiation in air and implanted after a long shelf life. J Bone Joint Surg. 2002;84-A:901–906. doi: 10.2106/00004623-200206000-00002. [DOI] [PubMed] [Google Scholar]
- 12.Mensch JS, Amstutz HC. Knee morphology as a guide to knee replacement. Clin Orthop Relat Res. 1975;112:231–241. [PubMed] [Google Scholar]
- 13.Newman J, Pydisetty RV, Ackroyd C. Unicompartmental or total knee replacement : the 15-year results of a prospective randomised controlled trial. J Bone Joint Surg. 2009;91-B:52–57. doi: 10.1302/0301-620X.91B1.20899. [DOI] [PubMed] [Google Scholar]
- 14.Patil S, Colwell CW, Jr, Ezzet KA, D'Lima DD. Can normal knee kinematics be restored with unicompartmental knee replacement ? J Bone Joint Surg. 2005;87-A:332–338. doi: 10.2106/JBJS.C.01467. [DOI] [PubMed] [Google Scholar]
- 15.Pennington DW, Swienckowski JJ, Lutes WB, Drake GN. Unicompartmental knee arthroplasty in patients sixty years of age or younger. J Bone Joint Surg. 2003;85-A:1968–1973. doi: 10.2106/00004623-200310000-00016. [DOI] [PubMed] [Google Scholar]
- 16.Pennington DW, Swienckowski JJ, Lutes WB, Drake GN. Lateral unicompartmental knee arthroplasty : survivorship and technical considerations at an average follow-up of 12.4 years. J Arthroplasty. 2006;21:13–17. doi: 10.1016/j.arth.2004.11.021. [DOI] [PubMed] [Google Scholar]
- 17.Price AJ, Dodd CA, Svard UG, Murray DW. Oxford medial unicompartmental knee arthroplasty in patients younger and older than 60 years of age. J Bone Joint Surg. 2005;87-B:1488–1492. doi: 10.1302/0301-620X.87B11.16324. [DOI] [PubMed] [Google Scholar]
- 18.Robertsson O, Knutson K, Lewold S, Lidgren L. The Swedish Knee Arthroplasty Register 1975-1997 : an update with special emphasis on 41,223 knees operated on in 1988-1997. Acta Orthop Scand. 2001;72:503–513. doi: 10.1080/000164701753532853. [DOI] [PubMed] [Google Scholar]
- 19.Servien E, Saffarini M, Lustig S, Chomel S, Neyret P. Lateral versus medial tibial plateau : morphometric analysis and adaptability with current tibial component design. Knee Surg Sports Traumatol Arthrosc. 2008;16:1141–1145. doi: 10.1007/s00167-008-0620-0. [DOI] [PubMed] [Google Scholar]
- 20.Shinno N. Statico-dynamic analysis of movement of the knee. IV. Functional significance of the menisci in the movement of the knee. Tokushima J Exp Med. 1961;8:189–202. [PubMed] [Google Scholar]
- 21.Surendran S, Kwak DS, Lee UY, et al. Anthropometry of the medial tibial condyle to design the tibial component for unicondylar knee arthroplasty for the Korean population. Knee Surg Sports Traumatol Arthrosc. 2007;15:436–442. doi: 10.1007/s00167-006-0188-5. [DOI] [PubMed] [Google Scholar]
- 22.Swienckowski JJ, Pennington DW. Unicompartmental knee arthroplasty in patients sixty years of age or younger. J Bone Joint Surg. 2004;86-A(suppl 1):131–142. doi: 10.2106/00004623-200409001-00004. [DOI] [PubMed] [Google Scholar]