Skip to main content
The British Journal of Radiology logoLink to The British Journal of Radiology
. 2014 Mar 19;87(1036):20130534. doi: 10.1259/bjr.20130534

Comparison of therapeutic efficacy and safety of radiofrequency ablation of hepatocellular carcinomas between internally cooled 15-G and 17-G single electrodes

H J Park 1,,2, M W Lee 1,, K D Song 1, D I Cha 1, H Rhim 1, T W Kang 1, S Lim 1, H K Lim 1
PMCID: PMC4067013  PMID: 24646182

Abstract

Objective:

To compare the ablation volume, local tumour progression rate and complication rate of radiofrequency ablation (RFA) for small hepatocellular carcinomas (HCCs) using 15-G and 17-G single electrodes.

Methods:

This retrospective study was approved by the institutional review board and informed consent was waived. We reviewed percutaneous RFA cases for HCCs using 15-G or 17-G electrodes without multiple overlapping ablations. A total of 36 pairs of HCCs matched according to tumour size and active tip length were included. We compared ablation volume and complication rate between the two electrode groups. Cumulative local tumour progression rates were estimated using the Kaplan–Meier method and compared using the log-rank test.

Results:

Tumour size and ablation time were not significantly different between the 15-G and 17-G groups (p = 0.661 and p = 0.793, respectively). However, ablation volume in the 15-G electrode group was larger than that in the 17-G group (14.4 ± 5.4 cm3 vs 8.7 ± 2.5 cm3; p < 0.001). No statistical difference in complication rates between the two electrode groups was found. The 10- and 20-month local tumour progression rates were not significantly different between the two groups (2.8% and 5.6% vs 11.1% and 19.3%; p = 0.166).

Conclusion:

Ablation volume by the 15-G electrode was larger than that by the 17-G electrode. However, local tumour progression rate and complication rate were not significantly different between the two electrode groups.

Advances in knowledge:

RFA of HCC using a 15-G electrode is useful to create larger ablation volumes than a 17-G electrode.


Radiofrequency ablation (RFA) is now considered to be one of the curative treatment modalities for the management of small hepatocellular carcinomas (HCCs).15 There are various types of electrodes for RFA, including internally cooled electrodes, multi-tined electrodes and perfusion electrodes.6,7 Among them, an internally cooled electrode is frequently used because it is simple and easy to use. The ablation zone created by an internally cooled electrode is usually cylindrical in shape along the longitudinal axis of the electrode. Hence, the horizontal diameter of the ablation zone is relatively smaller than the longitudinal diameter. Therefore, an internally cooled electrode frequently requires overlapping ablations to achieve a sufficient ablative margin for large tumours (i.e. size >2.5 cm). However, it is sometimes technically difficult to achieve a sufficient ablative margin under ultrasonography guidance since both the boundary of the index tumour and the active tip of the electrode are obscured by the echogenic zone generated by the previous ablation cycle. Therefore, it is ideal to achieve a large ablation zone using a single electrode without multiple overlapping ablations.

Electrodes with a larger diameter have a larger contact surface with the surrounding tissue than smaller ones and thereby have a higher active electric field.810 This in turn delivers a larger amount of radiofrequency energy and thus produces a greater amount of heat around the active tip. Consequently, large electrodes create large ablation zones. However, to our knowledge, commercially available internally cooled electrodes are not large and are almost exclusively 17-G.11,12

Recently, larger sized 15-G electrodes have been introduced and used for RFA of liver tumours in our institution. We have found that the 15-G electrode created a larger ablation volume than did the 17-G electrode, while the complication rate differed little. However, to our knowledge, there are no comparative data in the literature regarding the therapeutic efficacy and complication rate after RFA between 15-G and 17-G single electrodes. Hence, the purpose of this study was to evaluate and compare the ablation volume, local tumour progression rate and complication rate after RFA of HCCs using 15-G and 17-G single electrodes.

METHODS AND MATERIALS

Study population

This retrospective study was approved by the institutional review board of our hospital, and the requirement for informed consent was waived. From October 2011 to November 2012, a total of 417 patients with 440 HCCs were treated with RFA in the radiology department of the Samsung Medical Center, Seoul, Republic of Korea. Study populations in this study are summarized as a flowchart in Figure 1. The inclusion criteria were as follows: (a) patients with HCCs treated by RFA and (b) patients who met the Milan criteria (a single HCC, 5 cm or smaller, or less than three HCCs, each 3 cm or smaller) and (c) RFA performed by a single ablation with a single electrode. Exclusion criteria were as follows: (a) patients having local tumour progression after previous RFA or chemoembolization, (b) patients not having immediate follow-up CT or MRI after RFA, (c) patients not having at least 12-month follow-up CT or MRI. There were 63 patients whose HCCs were ablated with a 15-G electrode, and 53 patients whose HCCs were ablated with a 17-G electrode. To conduct a matched-pair analysis, we attempted to match each case in the 15-G and 17-G groups, according to the same active tip length of the electrode and tumour size (within 2 cm) since these can affect the therapeutic outcomes of RFA. This resulted in a total of 36 matched pairs of HCCs. Finally, the 15-G electrode group consisted of 34 patients with 36 HCCs [28 males and 6 females (mean age, 61.4 years; range, 46–81 years)], whereas the 17-G electrode group consisted of 35 patients with 36 HCCs [28 males and 7 females (mean age, 60.7 years; range, 38–76 years)]. The diagnosis of the HCC was based on either typical imaging features on CT or MR images3 (n = 62, 89.9%) or on histological confirmation by percutaneous biopsy (n = 7, 10.1%). The size of the HCCs was defined as the longest diameter of the tumour on ultrasonography images at the time of RFA procedures.

Figure 1.

Figure 1.

The flowchart depicts the process of selection of patients with radiofrequency ablation (RFA) for hepatocellular carcinomas. TACE, transcatheter arterial chemoembolization.

Radiofrequency ablation procedure

Before the RFA procedure, written informed consent was obtained from all patients. All RFA procedures were performed by one of three interventional radiologists (HR, HKL and MWL with 13, 13 and 8 years' experience, respectively, in the field of RFA of HCC, and each having >700 RFA cases of experience at the beginning of this study) under ultrasonography guidance (Logiq® E9; GE Healthcare, Milwaukee, WI). For local anaesthesia, 2% lidocaine hydrochloride (Lidocaine HCI INJ; Huons Co., Hwaseong, Republic of Korea) was injected at the puncture site. For pain control, 50 mg of pethidine hydrochloride (Samsung Pharmaceutical, Seoul, Republic of Korea) mixed with 50 ml of 5% dextrose in water was dripped intravenously. In addition, 100 μg of fentanyl citrate (Fentanyl Citrate Gu Ju INJ; GUJU Pharma, Seoul, Republic of Korea) mixed with 20 ml of normal saline was also injected intermittently one-quarter (25 μg of fentanyl citrate) at a time. Most patients were managed with conscious sedation while monitored anaesthesia care (MAC) was used in only 15.9% (11/69) of patients. MAC was preferred for index tumours located near portal vein branches, for which a patient is expected to have severe pain during the procedure.

We used 15-G or 17-G internally cooled length-adjustable electrodes (Proteus® RF Electrode; STARmed, Gyeonggi-do, Republic of Korea) with a 200-W RF generator (VIVA RF System; STARmed). In general, a 15-G electrode had a thicker calibre and rarely bent in the liver during insertion because it is stiffer than a 17-G electrode (Figure 2). The type of electrodes and the length of the active tip were chosen based on size, location and geometry of the tumour as well as the preference of the radiologists. If needed, artificial ascites was introduced before the RFA procedure to improve the sonic window or to reduce the level of collateral thermal injury to the adjacent organ.13,14 In addition, fusion imaging (volume navigation, Logiq E9) was applied during the RFA procedure whenever needed to improve detection and localization of HCCs.15

Figure 2.

Figure 2.

Photographs of 15-G (a) and 17-G (b) Proteus® electrode shafts (STARmed, Gyeonggi-do, Republic of Korea) with adjustable active tip under a compressor. To compare the stiffness of the two electrodes, we evaluated how well the electrodes bend by the same force using a compressor. With similar kilogram-force applied, the 15-G electrode bent less than the 17-G electrode, indicating that it is stiffer than the 17-G electrode. Arrows indicate the direction of force.

Although the protocol for radiofrequency energy deposition followed the manufacturer's instructions, ablation time was decided at the operator's discretion based on tumour size, extent of echogenic zone on ultrasonography images during RFA and the patient's condition. If the index tumour was located adjacent to a dangerous structure and thus susceptible to collateral damage or the ablative margin seemed >5–10 mm during RFA and thus large enough to cover the entire tumour, operators did not perform aggressive treatment. In these cases, the operators terminated the RFA procedure even though the ablation time was less than 12 min, which is the manufacturer's recommended duration time. On the contrary, if the size of the echogenic zone was smaller than expected and overlapping ablation was infeasible owing to an unfavourable location of the tumour, RFA was maintained for more than 12 min. At the end of the RFA procedures, the electrode path was cauterized to prevent tract seeding or bleeding.

Assessment of ablation zones

Therapeutic outcomes of both electrodes were evaluated. The diameter and volume of the ablation zones (non-enhancing area on CT images) created after using the 15-G and 17-G electrodes were evaluated in random order by a board-certified abdominal radiologist (HJP, with 7 years' clinical experience) who was blinded to the type of electrode used for RFA (15 G vs 17 G), size of HCCs and radiological reports of immediate post-RFA CT scans. Cases where the index tumours were located within 1 cm from the liver capsule or vessels >3 mm in diameter were excluded in this measurement, since the size and shape of the ablation zones are affected by these structures.16 Portal venous phase images of immediate post-RFA CT were used for quantitative analysis of the ablation zone. For volume measurement, maximum (Dmx) and minimum (Dmi) diameters were measured on an axial scan with the largest ablation area, whereas the vertical diameter (Dv) was measured from the most cranial to the most caudal border of the ablation zone. The volume of the ablation zone was calculated using the following formula: π (Dv × Dmx × Dmi)/6.10 The surrounding areas with benign periablational enhancement were not included in the measurement.

Follow-up after radiofrequency ablation and analysis of therapeutic efficacy

Liver CT or MRI was performed within 12 h following the RFA procedures to evaluate therapeutic efficacy and complications according to a standardization paper of image-guided tumour ablation.17 Technical success was defined as the index tumour being treated according to our treatment protocol and covered completely by the ablation zone, which means complete elimination of enhancing areas within the entire tumour on immediate follow-up liver CT or MRI. Technique effectiveness was defined by the absence of marginal nodular enhancements within the ablation zone on 1-month follow-up CT scans. If the RFA was considered technically effective on the follow-up CT scan performed 1 month later, liver CT was repeated every 3–4 months. Major complication was defined as any occurrence of events threatening the patient's life, leading to substantial disability and morbidity or demanding prolonged hospitalization. All other complications were considered minor. Local tumour progression was defined as the appearance of foci of untreated disease in tumours during follow-up that were previously considered to be completely ablated.

Statistical analysis

Continuous variables such as age, size of HCCs, ablation time, ablation volume and follow-up period between the 15-G and 17-G electrode groups were tested using Student's t-test or the Mann–Whitney test according to the normality assumption. Categorical variables such as sex, Child–Pugh classification, liver cirrhosis, aetiology of liver disease, previous HCC treatment, tumour location, fusion imaging guidance, use of artificial ascites and active tip length in each electrode were tested using Fisher's exact test or the χ2 test. Technical success, technique effectiveness and complication rate were compared using Fisher's exact test. Cumulative local tumour progression rates were estimated using the Kaplan–Meier method and compared using the log-rank test. A p-value <0.05 was considered statistically significant. All statistical analyses were performed with statistical software package IBM SPSS® v. 21.0 (SPSS Inc., Chicago, IL).

RESULTS

Baseline characteristics of the study population

The baseline patient characteristics of the 15-G and 17-G electrode groups are summarized in Table 1. There were no statistical differences between the two groups except previous treatment history of HCC [transcatheter arterial chemoembolization (TACE) (n = 11), TACE + RFA (n = 5), resection + TACE + RFA (n = 3), resection (n = 1), resection + TACE (n = 2), and no history (n = 12) in the 15-G group vs TACE (n = 3), TACE + RFA (n = 1), resection (n = 3), resection + TACE (n = 1), RFA (n = 1), and no history (n = 26) in the 17-G group]. Tumour size and active tip length of electrodes were also not significantly different between the two groups (Table 2).

Table 1.

Baseline characteristics of 69 patients with 72 hepatocellular carcinomas (HCCs)

Characteristics 15-G electrode (n = 34) 17-G electrode (n = 35) p-value                                  
Age (years)                                  
 Mean ± SD 61.4 ± 8.6 60.7 ± 9.5 0.848a                                  
 Range 46–81 38–76                                    
Sex: male/female 28/6 28/7 0.772b                                  
Child-Pugh: A/B 33/1 35/0 1.000c                                  
LC: yes/no 22/12 20/15 0.429c                                  
Aetiology of liver disease     0.479b                                  
 HBV 26 23                                    
 HCV 3 5                                    
 Alcoholic 0 1                                    
 Idiopathic 5 6                                    
Previous HCC treatment: yes/no 22/12 8/27 0.001b                                  

HBV, hepatitis B virus; HCV, hepatitis C virus; LC, liver cirrhosis; SD, standard deviation.

a

Mann–Whitney test.

b

χ2 test.

c

Fisher's exact test.

Table 2.

Characteristics of 72 hepatocellular carcinomas (HCCs) and treatment outcomes after radiofrequency ablation

Characteristics 15-G electrode (n = 36) 17-G electrode (n = 36) p-value
Size of HCCs (cm)
 Mean ± SD 1.4 ± 0.2 1.3 ± 0.2 0.661a
 Range 1.0–2.0 1.0–2.0  
Tumour locationb     0.789c
 Subcapsular location 7d 9  
 Perivascular location 3d 0  
 Others 27 27  
Fusion imaging guidance: yes/no 16/20 19/17 0.479c
Active tip length of electrodes
 2.0 cm/2.5 cm/3.0 cm 4/20/12 4/20/12 1.000c
Ablation time (minutes) 8.7 ± 3.5 (3–18) 8.4 ± 2.5 (3–12) 0.793a
Artificial ascites: yes/no 9/27 5/31 0.234c
Ablation zonee
 Volume (cm3) 14.4 ± 5.4 (7.6–9.4) 8.7 ± 2.5 (4.5–16.4) <0.001a
 Dmx (cm) 3.3 ± 0.5 (2.6–4.5) 2.9 ± 0.5 (2.1–3.8) 0.008f
 Dmi (cm) 2.5 ± 0.3 (2–3.2) 2.2 ± 0.3 (1.6–2.8) <0.001f
 Dv (cm) 3.2 ± 0.5 (2.4–4.0) 2.6 ± 0.5 (1.4–3.4) <0.001f
Follow-up (months)
 Mean ± SD 15.9 ± 2.4 16.4 ± 2.9 0.424a
 Range 12.6–20.2 12.1–20.6  
Technical success (%) 36/36 (100) 36/36 (100)  
Technical effectiveness (%) 36/36 (100) 36/36 (100)  
Complications (%) 2/36 (5.6) 3/36 (8.3) 1.000g
 Major complications (%) 1/36 (2.8) 0/36 (0) 1.000g
 Minor complications (%) 1/36 (2.8) 3/36 (8.3) 0.614g

Dmi, minimum diameter of largest transverse ablation zone; Dmx, maximum diameter of largest transverse ablation zone; Dv, length from cranial to caudal border of ablation zone.

Values are presented as means ± SD where applicable.

Numbers in parentheses are ranges.

a

Mann–Whitney test.

b

Subcapsular or perivascular location was defined when the tumour was abutting the liver capsule or hepatic vessels >3 mm in diameter, respectively.

c

χ2 test.

d

One case was located in both the subcapsular and perivascular areas.

e

In terms of ablation zone, 9 out of 36 HCCs using the 15-G electrode and 9 out of 36 HCCs using the 17-G electrode were analysed after excluding lesions with subcapsular or perivascular location.

f

Student's t-test.

g

Fisher's exact test.

Ablation zones

Characteristics of the ablation zone and treatment outcomes are summarized in Table 2. 9 (25%) HCCs from each group were included for analysis of size of ablation zone, considering the distance from the liver capsule or hepatic vessels around the index tumour, which can affect the shape and size of the ablation zone. Although the total ablation times between the 15-G and 17-G electrode groups were not significantly different (8.7 ± 3.5 vs 8.4 ± 2.5 min; p = 0.793), three measured diameters and volume of ablation zone in the 15-G electrode group were significantly greater than those in the 17-G electrode group (Figures 3 and 4).

Figure 3.

Figure 3.

Single hepatocellular carcinoma (HCC) treated with radiofrequency ablation (RFA) with a 15-G electrode in a 42-year-old female. (a) A single nodular HCC (arrow) measuring 2 cm was noted in Segment VI of the liver on the hepatobiliary phase of the liver MR image (repetition time/echo time, 3.1/1.5 ms; flip angle, 10°; matrix size, 256 × 256; bandwidth, 724.1 Hz per pixel; section thickness, 2 mm). (b) Axial portal phase CT image obtained immediately after RFA using a 15-G electrode with a 2.5-cm active tip for 10 min shows a large ablation zone (arrowheads) with no residual enhancing foci, suggestive of technical success. The measured ablation volume was 25.3 cm3. (c) Coronal portal phase CT image obtained immediately after RFA also shows a large round ablation zone (arrowheads).

Figure 4.

Figure 4.

Single hepatocellular carcinoma (HCC) treated with radiofrequency ablation (RFA) with a 17-G electrode in a 61-year-old male. (a) A small HCC (arrow) measuring 1.2 cm was noted in Segment VIII of the liver on the hepatobiliary phase of the liver MR image (repetition time/echo time, 3.1/1.5 ms; flip angle, 10°; matrix size, 256 × 256; bandwidth, 724.1 Hz per pixel; section thickness, 2 mm). (b) Axial portal phase CT image obtained immediately after RFA using a 17-G electrode with a 2.5-cm tip for 10 min shows a relatively small ablation zone (arrowheads), with ablation volume of 10.3 cm3. However, the tumour was covered with sufficient ablative margin, indicating technical success. (c) Coronal portal phase CT image obtained immediately after RFA shows a relatively small ovoid ablation zone (arrowheads).

Therapeutic outcome

Technical success was achieved in all RFA procedures in a single treatment session. Technique effectiveness rates were also 100% in both groups. The major complication rate between the 15-G and 17-G groups was not significantly different (2.8%, 1/36 vs 0%, 0/36; p = 1.000). One patient in the 15-G group developed extensive portal vein thrombosis, which was classified as a major complication because the patient was hospitalized for up to 10 days after RFA, which was longer than initially expected. On retrospective review, the index tumour was abutting a portal vein branch. The minor complication rate was not significantly different between the two groups (2.8%, 1/36 vs 8.3%, 3/36; p = 0.614). They included one case with a small amount of haemorrhage in the 15-G group and pleural effusion (n = 1), haemobilia (n = 1) and subsegmental infarction (n = 1) in the 17-G group. None of those minor complications had clinical significance and subsequently required no treatment. The 10- and 20-month local tumour progression rates were not significantly different between the two groups (2.8% and 5.6% vs 11.1% and 19.3%; p = 0.166) (Figure 5).

Figure 5.

Figure 5.

Cumulative local tumour progression rate of 15-G and 17-G electrode groups after percutaneous radiofrequency ablation. mo, months.

DISCUSSION

In this study, the 15-G electrode created a larger ablation volume than the 17-G electrode although the ablation time was similar. Both the therapeutic efficacy and complication rate after RFA were not significantly different between the two electrode groups. However, the local tumour progression rate had a tendency to be lower in the 15-G group than in the 17-G group, although it did not reach statistical significance. If a larger study population was included, the local tumour progression rate between the two groups may have reached statistical significance. Based on the results of this study, it is likely that RFA using a 15-G electrode shows better performance than that using a 17-G electrode.

The results of this study are in close agreement with an old study by Goldberg et al.8 They evaluated the parameters affecting the size and distribution of an ablation zone produced by radiofrequency electrodes and found that the extent of coagulation necrosis increased from 0.7 to 1.8 cm as the electrode diameter increased from 24 G to 12 G. Our results are also in line with recent studies using separable clustered electrodes, in which 15-G electrodes showed a larger ablation zone than 17-G electrodes under the same conditions.10,18 In the present study, there was no significant difference in ablation time between the 15-G and the 17-G groups (8.7 ± 3.5 min vs 8.4 ± 2.5 min; p= 0.793). This implies that a 15-G electrode can make a similar size ablation zone with shorter ablation time than a 17-G electrode. In other words, RFA using a 15-G electrode can reduce the overall ablation time. Another advantage of the 15-G electrode is that it is stiffer than the 17-G electrode (Figure 2). In general, the electrode path is affected by a stiff cirrhotic liver and the breathing motions of patients. Therefore, a stiff electrode would be useful for accurate electrode placement or repositioning during RFA procedures.

The complication rate did not differ between the 15-G group and the 17-G group in this study. However, portal vein thrombosis developed in the 15-G electrode group as a major complication after ablating an HCC abutting a portal vein branch. According to a recent study in which 15-G separable clustered electrodes were used,10 portal vein thrombosis developed in 3 (21.4%) out of 14 pigs. Although the exact reason is not known, a higher radiofrequency energy deposition of the 15-G electrode would be related to this complication. Since swine liver is thinner than human liver, the ablation zone is likely to contact a portal vein branch easily, thus it is susceptible to portal vein thrombosis. Therefore, human data with large series are needed to verify whether portal vein thrombosis occurs more frequently with 15-G electrodes or 17-G electrodes. Nevertheless, caution would be necessary for RFA of HCCs abutting a portal vein with a 15-G electrode. On the contrary, given that the heat-sink effect is one of the limitations of RFA for tumours located close to portal veins,19 a 15-G electrode may overcome this limitation with its higher ablation performance. Further study is warranted to assess whether a 15-G electrode is useful to overcome the heat-sink effect.

Until now, the diameter of commercially available electrodes has ranged from 17 G to 14 G.6,7 However, to our knowledge, 17-G electrodes have been exclusively used for internally cooled electrodes. Although we do not know the exact reason why 17-G electrodes are most widely used for RFA of liver tumour, based on our results, we believe that the 15-G electrode can be an alternative to a conventional 17-G electrode for RFA of liver tumours with high performance and an acceptable complication rate.

Our study has several limitations. First, this study suffers from a selection bias since this was a retrospective study. The type of electrode used was determined by the radiologists who performed the RFA procedures depending on their preference, tumour size and patient condition. Although baseline characteristic of HCCs and RFA procedures were not different between the two groups, the lack of standardization for choosing an electrode may be a limitation of this study. Second, ablation time for a single ablation in most cases was shorter than the recommended 12 min. Since tumour sizes were small in both groups (1.4 ± 0.2 cm in the 15-G group, 1.3 ± 0.2 cm in the 17-G group), RFA was terminated before 12 min if the echogenic zone was considered large enough to achieve a sufficient ablative margin. Since most patients under conscious sedation have some degree of pain from prolonged RFA procedures, fulfilling the 12 min seems unnecessary if the echogenic zone is large enough. Furthermore, this issue may not affect our results significantly because the total ablation time was not different between the two electrode groups. Third, although patients with a follow-up period of at least 12 months after RFA were included in this study, the follow-up period was relatively short to determine the long-term therapeutic efficacy. However, instead of long-term therapeutic outcomes of RFA, we aimed to compare the size of the ablation zone, short-term therapeutic efficacy and the safety of RFA of HCCs using 15-G and 17-G single electrodes.

CONCLUSION

In conclusion, the therapeutic efficacy and complication rate were not significantly different between the two groups. However, RFA with a 15-G single electrode created a larger ablation volume than a 17-G electrode under similar ablation conditions. In addition, the 15-G group had a tendency for a lower local tumour progression rate than the 17-G group, although it did not reach statistical significance.

FUNDING

This study was supported by a Samsung Medical Center grant [#GFO1130071].

REFERENCES

  • 1.Forner A, Llovet JM, Bruix J. Hepatocellular carcinoma. Lancet 2012; 379: 1245–55. doi: 10.1016/S0140-6736(11)61347-0 [DOI] [PubMed] [Google Scholar]
  • 2.European Association for the Study of the Liver; European Organisation for Research and Treatment of Cancer. EASL-EORTC clinical practice guidelines: management of hepatocellular carcinoma. J Hepatol 2012; 56: 908–43. doi: 10.1016/j.jhep.2011.12.001 [DOI] [PubMed] [Google Scholar]
  • 3.Bruix J, Sherman M; American Association for the Study of Liver D. Management of hepatocellular carcinoma: an update. Hepatology 2011; 53: 1020–2. doi: 10.1002/hep.24199 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Shiina S, Tateishi R, Arano T, Uchino K, Enooku K, Nakagawa H, et al. Radiofrequency ablation for hepatocellular carcinoma: 10-year outcome and prognostic factors. Am J Gastroenterol 2012; 107: 569–77; quiz 78. doi: 10.1038/ajg.2011.425 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Kim YS, Lim HK, Rhim H, Lee MW, Choi D, Lee WJ, et al. Ten-year outcomes of percutaneous radiofrequency ablation as first-line therapy of early hepatocellular carcinoma: analysis of prognostic factors. J Hepatol 2013; 58: 89–97. doi: 10.1016/j.jhep.2012.09.020 [DOI] [PubMed] [Google Scholar]
  • 6.McGhana JP, Dodd GD 3rd. Radiofrequency ablation of the liver: current status. AJR Am J Roentgenol 2001; 176: 3–16. doi: 10.2214/ajr.176.1.1760003 [DOI] [PubMed] [Google Scholar]
  • 7.Denys AL, De Baere T, Kuoch V, Dupas B, Chevallier P, Madoff DC, et al. Radio-frequency tissue ablation of the liver: in vivo and ex vivo experiments with four different systems. Eur Radiol 2003; 13: 2346–52. doi: 10.1007/s00330-003-1970-0 [DOI] [PubMed] [Google Scholar]
  • 8.Goldberg SN, Gazelle GS, Dawson SL, Rittman WJ, Mueller PR, Rosenthal DI. Tissue ablation with radiofrequency: effect of probe size, gauge, duration, and temperature on lesion volume. Acad Radiol 1995; 2: 399–404. [DOI] [PubMed] [Google Scholar]
  • 9.Goldberg SN, Gazelle GS, Halpern EF, Rittman WJ, Mueller PR, Rosenthal DI. Radiofrequency tissue ablation: importance of local temperature along the electrode tip exposure in determining lesion shape and size. Acad Radiol 1996; 3: 212–18. [DOI] [PubMed] [Google Scholar]
  • 10.Lee ES, Lee JM, Kim KW, Lee IJ, Han JK, Choi BI. Evaluation of the in vivo efficiency and safety of hepatic radiofrequency ablation using a 15-g octopus(R) in pig liver. Korean J Radiol 2013; 14: 194–201. doi: 10.3348/kjr.2013.14.2.194 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Minami Y, Kudo M. Radiofrequency ablation of hepatocellular carcinoma: a literature review. Int J Hepatol 2011; 2011: 104685. doi: 10.4061/2011/104685 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Livraghi T, Meloni F, Di Stasi M, Rolle E, Solbiati L, Tinelli C, et al. Sustained complete response and complications rates after radiofrequency ablation of very early hepatocellular carcinoma in cirrhosis: is resection still the treatment of choice? Hepatology 2008; 47: 82–9. [DOI] [PubMed] [Google Scholar]
  • 13.Song I, Rhim H, Lim HK, Kim YS, Choi D. Percutaneous radiofrequency ablation of hepatocellular carcinoma abutting the diaphragm and gastrointestinal tracts with the use of artificial ascites: safety and technical efficacy in 143 patients. Eur Radiol 2009; 19: 2630–40. [DOI] [PubMed] [Google Scholar]
  • 14.Kim YJ, Lee MW, Park HS. Small hepatocellular carcinomas: ultrasonography guided percutaneous radiofrequency ablation. Abdom Imaging 2013; 38: 98–111. doi: 10.1007/s00261-012-9883-5 [DOI] [PubMed] [Google Scholar]
  • 15.Lee MW, Rhim H, Cha DI, Kim YJ, Choi D, Kim YS, et al. Percutaneous radiofrequency ablation of hepatocellular carcinoma: fusion imaging guidance for management of lesions with poor conspicuity at conventional sonography. AJR Am J Roentgenol 2012; 198: 1438–44. doi: 10.2214/AJR.11.7568 [DOI] [PubMed] [Google Scholar]
  • 16.Park MJ, Kim YS, Rhim H, Lim HK, Lee MW, Choi D. A comparison of US-guided percutaneous radiofrequency ablation of medium-sized hepatocellular carcinoma with a cluster electrode or a single electrode with a multiple overlapping ablation technique. J Vasc Interv Radiol 2011; 22: 771–9. [DOI] [PubMed] [Google Scholar]
  • 17.Goldberg SN, Grassi CJ, Cardella JF, Charboneau JW, Dodd GD 3rd, Dupuy DE, et al. Image-guided tumor ablation: standardization of terminology and reporting criteria. Radiology 2005; 235: 728–39. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Lee ES, Lee JM, Kim WS, Choi SH, Joo I, Kim M, et al. Multiple-electrode radiofrequency ablations using Octopus® electrodes in an in vivo porcine liver model. Br J Radiol 2012; 85: e609–15. doi: 10.1259/bjr/61619687 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Lu DS, Raman SS, Limanond P, Aziz D, Economou J, Busuttil R, et al. Influence of large peritumoral vessels on outcome of radiofrequency ablation of liver tumors. J Vasc Interv Radiol 2003; 14: 1267–74. [DOI] [PubMed] [Google Scholar]

Articles from The British Journal of Radiology are provided here courtesy of Oxford University Press

RESOURCES