Abstract
Objectives:
The purpose of this study was to investigate the relationship between compression force and hardness values in ultrasound elastography.
Methods:
Ultrasound elastography was performed using an elastography phantom, comprising inclusions with different elasticities and echogenicities. The compression force was set to approximately 100 gw (light force) and approximately 500 gw (heavy force). The elasticity index (EI) of the inclusion was measured. The EI was a relative hardness value of a structure within an elastographic image. Similarly, the EI of the background was measured as a reference. The elasticity ratio (ER) was calculated as the EI of the inclusion divided by the EI of the reference.
Results:
The hardness of the phantom could be discerned with both the EI and ER, regardless of the compression force. The EI and ER with heavy force tended to be higher than those with light force, but the difference was not significant. A strong correlation was observed between the EI and ER of soft structures, whereas the correlation between the EI and ER of hard structures was weak, and the ER values varied widely.
Conclusions:
The EI offers potential as a good indicator for assessing the hardness.
Keywords: ultrasonography, elastography
Introduction
Ultrasound elastography is an imaging technique that allows non-invasive assessment of tissue hardness. The technique has been used to assess the hardness of various normal and abnormal tissues,1–6 and to differentiate malignant from benign lesions in the breast, prostate, thyroid, lymph nodes and salivary glands.7–24
Ultrasound elastography uses either strain or shear wave velocity.25,26 In strain elastography, tissue strain induced by compression with the probe is used as an indicator of hardness. Strain is smaller in a hard than in a soft tissue and is calculated by comparing echo signals obtained before and after compression. The hardness of tissues or lesions is visually and qualitatively assessed on the basis of the strain distribution and pattern of elastographic images.1,6–10,12,15–17,19–24 Semi-quantitative analysis, in which strain is represented in numerical values, has recently been used as a more objective method.1–5,9–11,13,17 However, whether these values can be effectively used for clinical assessment remains unclear because the image quality of strain elastography, which is performed manually, depends strongly on the imaging procedure.14,18 To the best of our knowledge, the effects of compression force on hardness values have not yet been reported.
The purpose of this study was to investigate the relationship between compression force and hardness values with an elastography phantom. Comparisons were performed using directly measured hardness values and their normalized values. Intra- and interoperator agreements were also evaluated to assess the reproducibility of the measured values.
Methods and Materials
Ultrasound scanner and phantom
Ultrasound elastography was performed using a Logiq S8 (GE Healthcare, Milwaukee, WI) with a 10-MHz linear probe (ML6-15-D; GE Healthcare). An elastography phantom (ELPT-002; OST, Chiba, Japan) was used, comprising five inclusions with different elasticities and echogenicities, mounted in a background material of lower elasticity. The five inclusions simulated the Tsukuba elasticity score7 (Figure 1a). Three of the inclusions consisting of homogeneous materials and corresponding to Tsukuba elasticity score of 1, 2 and 4 were selected for elastography. We selected inclusions with homogeneous elasticities and echogenicities for concise analysis. The Score 1 inclusion (S1) was as soft as the background, the Score 2 inclusion (S2) was slightly harder than the background, and the Score 4 inclusion (S4) was harder than the background.
Scanning protocol
Elastographic images were obtained by repetitive compression with the probe. We set the compression force to approximately 100 gw (light force) and approximately 500 gw (heavy force). To keep the force constant, operators trained before the examination with the phantom placed on a scale. The examination was performed three times with each force by three operators.
Elastographic images were displayed in real time as colour-coded images within a region of interest (ROI) placed on B-mode images (Figure 1b). The colour indicated the relative hardness of structures within the ROI, red, yellow/green and blue represented soft, moderately hard and hard, respectively. We placed the 30 × 28 mm ROI on the B-mode image with a field of view of 50 × 50 mm. The centre of the ROI was placed on the inclusion. The surface of the phantom was not included in the ROI, and a space of 2 mm was left because it is known that a band of overestimation of hardness on the surface of the phantom when in contact with the probe is seen on elastographic images.27 In this study, a hard (blue) band approximately 4–5 mm thick was seen on the surface of the phantom. B-mode images were displayed beside elastographic images, and the quality bar and quality graph were also displayed on the screen. These indicators showed the appropriateness of the compression force, as evaluated on the basis of echo signal similarity obtained before and after compression. During the examination, we monitored the indicators to keep the compression force as constant as possible.
Image measurement
Elastographic images were stored as motion images. From these, we selected high-quality images that lasted for 3 s. On replayed motion images, the elasticity index (EI) of the inclusion was measured at 0.1-s intervals, and the mean was calculated (Figure 2). The EI was a relative hardness value of a structure within an elastographic image. A mean EI of all structures within an elastographic image was defined as 1. A value of 0–0.09 was assigned to soft structures and 1–6 to hard structures. The EI was measured with a circular ROI of 8 mm in diameter placed on the inclusion. Similarly, the EI of the background was measured as a reference at areas above, below and lateral to the inclusion for normalization of the EI of the inclusion. Because the EI is a simple relative value obtained directly with ultrasound systems, normalization using the EI of a reference is considered effective for obtaining stable and constant results. A circular ROI of 4 mm in diameter was placed 2 mm away from the inclusion. The ROI lateral to the inclusion was placed on the side close to the centre of the phantom. For example, the ROI lateral to the S1 was placed on the side close to the S2. The elasticity ratio (ER) was calculated as the EI of the inclusion divided by the EI of the reference, and the same comparisons were performed as follows.
Data analysis
The Kruskal–Wallis and Steel–Dwass tests were used to compare three inclusions for the EI and ER with each force. The (Wilcoxon) signed-ranks test was used to compare the EI and ER of the light force to those of the heavy force. The Friedman and Steel–Dwass tests were used to compare the ER calculated with the three references. Spearman's rank correlation was calculated between the EI and ER of each force.
Intraclass correlation coefficient was calculated for intra- and interoperator agreement. An intraclass correlation coefficient of 0–0.20 indicates poor agreement, 0.21–0.40 indicates fair agreement, 0.41–0.60 indicates moderate agreement, 0.61–0.80 indicates good agreement and 0.81–1.00 indicates excellent agreement. Values of p < 0.05 were considered significant.
Results
EIs are summarized in Table 1. With both compression forces, EIs showed significant differences among the three inclusions (Kruskal–Wallis test, p < 0.001; Steel–Dwass test, p < 0.001). ERs calculated with the three references were brought together because the differences among them were not significant. The details are described below. ERs are also summarized in Table 1. With both forces, ERs showed significant differences among the three inclusions (Kruskal–Wallis test, p < 0.001; Steel–Dwass test, p < 0.001). EIs and ERs tended to be higher with heavy force than with light force, but the difference was significant only in ERs of S1 [(Wilcoxon) signed-ranks test, p < 0.01; Table 1]. ERs calculated with the three references above, below and lateral to the inclusion are summarized in Table 2. ERs calculated with different reference areas and different forces were different. ERs calculated with references below the inclusion showed the highest values, except for the ER of S2 with heavy force. The Friedman test revealed significant differences among ERs, except for the ERs of S2 with light force (p < 0.02). However, the Steel–Dwass test revealed a significant difference only in the ERs of S4 with light force between the references above and below the inclusion and between the references below and lateral to the inclusion (p < 0.001). Strong correlations were seen between the EI and ER of S1 and S2, regardless of the compression force (Figure 3 and Table 3). S4 showed a weak correlation and ER values varied widely.
Table 1.
Compression force | Elasticity index |
Elasticity ratio |
||||
---|---|---|---|---|---|---|
S1 | S2 | S4 | S1 | S2 | S4 | |
Light | 1.44 (1.13–1.55) | 2.77 (2.41–2.91) | 5.14 (4.93–5.20) | 1.35 (1.16–1.69) | 3.51 (3.01–3.85) | 7.41 (6.73–8.42) |
Heavy | 1.60 (1.21–2.14) | 2.79 (2.39–3.25) | 5.28 (4.98–5.59) | 1.73 (1.27–2.51) | 3.57 (3.02–4.19) | 7.55 (7.01–8.53) |
S1, score 1 inclusion; S2, score 2 inclusion; S4, score 4 inclusion.
Table 2.
Compression force | S1 |
S2 |
S4 |
||||||
---|---|---|---|---|---|---|---|---|---|
Above | Below | Lateral | Above | Below | Lateral | Above | Below | Lateral | |
Light | 1.20 (1.10–1.53) | 1.64 (1.19–1.82) | 1.45 (1.24–1.67) | 3.51 (3.04–3.84) | 3.63 (3.04–3.85) | 3.37 (3.04–3.76) | 6.67 (6.13–7.30) | 8.69 (8.12–9.54) | 7.34 (6.83–7.74) |
Heavy | 1.59 (1.09–2.03) | 1.85 (1.34–2.49) | 1.81 (1.35–2.60) | 3.60 (2.92–4.07) | 3.42 (2.98–4.07) | 3.75 (3.17–4.51) | 7.55 (6.94–7.94) | 8.25 (7.53–8.53) | 7.38 (6.97–8.90) |
S1, score 1 inclusion; S2, score 2 inclusion; S4, score 4 inclusion.
Table 3.
Compression force | S1 | S2 | S4 |
---|---|---|---|
Light | 0.90 (p < 0.001) | 0.93 (p < 0.001) | 0.27 (p = 0.049) |
Heavy | 0.95 (p < 0.001) | 0.94 (p < 0.001) | 0.57 (p < 0.001) |
S1, score 1 inclusion; S2, score 2 inclusion; S4, score 4 inclusion.
Intra- and interoperator agreement was good or excellent for both the EI and ER and for both compression forces (Table 4).
Table 4.
ICC | Compression force | Elasticity index |
Elasticity ratio |
||||
---|---|---|---|---|---|---|---|
Operator 1 | Operator 2 | Operator 3 | Operator 1 | Operator 2 | Operator 3 | ||
Intra-operator | Light | 0.99 | 0.95 | 0.97 | 0.95 | 0.96 | 0.95 |
Heavy | 0.98 | 0.93 | 0.74 | 0.94 | 0.93 | 0.77 | |
Interoperator | Light | 0.98 | 0.98 | ||||
Heavy | 0.89 | 0.89 |
ICC, intraclass correlation coefficient.
Discussion
A light compression force is recommended to obtain appropriate elastographic images because a high pressure force makes the association between pressure and strain become non-proportional.7,28 We performed elastography with not only light but also heavy forces. With both forces we were able to discern the hardness of the phantom using the hardness values. Hardness values tended to be higher with heavy than with light force. Krouskop et al29 performed elastography with breast tissue samples and also observed an increase in the elastic modulus with strain. Hardness can, therefore, be discerned with heavy force but may be overestimated. Generally, malignant lesions are harder than benign lesions.9–11,13,14,17,18,29 Heavy compression forces may lead to false-positive results. The frequency of probe movement also influences image quality. Havre et al30 reported that the best quality was achieved in the range of 80–120 cycles min−1.
In the assessment of hardness on elastographic images, visual evaluation of the hardness pattern and semi-quantitative analysis with numerical hardness values, for which the hardness ratio between the lesion and surrounding tissue is generally calculated, are used. Some studies have compared the diagnostic accuracy of these two methods and reported that the semi-quantitative analysis was equal or superior to visual evaluation.9,10,17 Franchi-Abella et al27 stated on semi-quantitative analysis that the type of system, position of the ROI in the surrounding tissue and hardness of the lesion change the hardness ratio, which is not proportional to the theoretical ratio in most cases. We used both EI and ER as hardness values and were able to discern hardness of the phantom with both. In the calculation of hardness ratio, it is reasonable that the ROI of the reference is placed at a depth similar to that of the target to avoid stress decay;1,11,14,18,27 however, this is not always possible for anatomical reasons.1 In this study, the ERs were calculated with the three references, above, below and lateral to the inclusion, and little difference was seen among them. The ERs calculated with reference below the inclusion tended to show the highest values.
Regardless of the compression force, strong correlations were seen between the EIs and ERs of the soft inclusions. However, the correlation between the EIs and ERs of the hard inclusions was weak, and the ER values varied widely. This might be because the hard inclusion changed the strain of the background. Strain varies not only with the elasticity of the structure but also with the hardness and position of the surrounding structures and measured EIs are affected by these complex factors. Moreover, although the ER is a normalized value, which is calculated as the EI of the inclusion divided by the EI of the reference, ER showed a correlation with EI. This means that normalization with the surrounding structure is ineffective for obtaining a stable measurement value for the hardness of lesions or tissues. Special attention is necessary for determining diagnostic criteria with the ER.
Intra- and interoperator agreement was good for both the EI and the ER and for both compression forces. These results must be owing to careful monitoring of the compression force to keep it as constant as possible during examinations.
Some limitations of the present study need to be addressed. The first limitation was that we did not know the true elasticity of the phantom. We therefore were unable to verify how well the EI and the ER corresponded to theoretical values. Furthermore, the phantom used in this study simulated the Tsukuba elasticity score, which was proposed for breast disease.7 The elasticity and configuration may differ markedly from other lesions and tissues. Second, hardness values vary between systems, and the Tsukuba elasticity score had been determined with a different system. Further studies with other systems should therefore be undertaken. Third, the compression force was applied manually, therefore it was not constant. Finally, our results are preliminary and need to be confirmed in clinical studies. To date, many clinical studies have used ERs as diagnostic criteria.1–5,9–11,13,14,17,18 Based on our results, the EI is also worth studying.
In conclusion, the hardness of the phantom could be discerned with both the EI and the ER, regardless of the compression force. EIs and ERs with heavy force tended to be higher than those with light force. The difference, however, was not significant. A strong correlation was observed between the EI and ER of soft structures, whereas the correlation between the EI and ER of hard structures was weak, and ER values varied widely. According to these results, the EI offers potential as a good indicator for assessing the hardness of lesions or tissues. It is expected that the EI may be used solely in the evaluation of the soft tissues. Additional studies including clinical studies are needed to improve the stability and quantitative ability of the value.
References
- 1.Drakonaki EE, Allen GM, Wilson DJ. Real-time ultrasound elastography of the normal Achilles tendon: reproducibility and pattern description. Clin Radiol 2009; 64: 1196–202. doi: 10.1016/j.crad.2009.08.006 [DOI] [PubMed] [Google Scholar]
- 2.Niitsu M, Michizaki A, Endo A, Takei H, Yanagisawa O. Muscle hardness measurement by using ultrasound elastography: a feasibility study. Acta Radiol 2011; 52: 99–105. doi: 10.1258/ar.2010.100190 [DOI] [PubMed] [Google Scholar]
- 3.Yanagisawa O, Niitsu M, Kurihara T, Fukubayashi T. Evaluation of human muscle hardness after dynamic exercise with ultrasound real-time tissue elastography: a feasibility study. Clin Radiol 2011; 66: 815–19. doi: 10.1016/j.crad.2011.03.012 [DOI] [PubMed] [Google Scholar]
- 4.Ariji Y, Gotoh A, Hiraiwa Y, Kise Y, Nakayama M, Nishiyama W, et al. Sonographic elastography for evaluation of masseter muscle hardness. Oral Radiol 2013; 29: 64–9. [Google Scholar]
- 5.Gotoh A, Ariji Y, Hasegawa T, Nakayama M, Kise Y, Matsuoka M, et al. Sonographic elastography for assessing changes in masseter muscle elasticity after low-level static contraction. Oral Radiol 2013; 29: 140–5. [Google Scholar]
- 6.Bhatia KS, Rasalkar DD, Lee YP, Wong KT, King AD, Yuen YH, et al. Real-time qualitative ultrasound elastography of miscellaneous non-nodal neck masses: applications and limitations. Ultrasound Med Biol 2010; 36: 1644–52. doi: 10.1016/j.ultrasmedbio.2010.07.010 [DOI] [PubMed] [Google Scholar]
- 7.Itoh A, Ueno E, Tohno E, Kamma H, Takahashi H, Shiina T, et al. Breast disease: clinical application of US elastography for diagnosis. Radiology 2006; 239: 341–50. doi: 10.1148/radiol.2391041676 [DOI] [PubMed] [Google Scholar]
- 8.Schaefer FK, Heer I, Schaefer PJ, Mundhenke C, Osterholz S, Order BM, et al. Breast ultrasound elastography—results of 193 breast lesions in a prospective study with histopathologic correlation. Eur J Radiol 2011; 77: 450–6. doi: 10.1016/j.ejrad.2009.08.026 [DOI] [PubMed] [Google Scholar]
- 9.Zhao QL, Ruan LT, Zhang H, Yin YM, Duan SX. Diagnosis of solid breast lesions by elastography 5-point score and strain ratio method. Eur J Radiol 2012; 81: 3245–9. doi: 10.1016/j.ejrad.2012.06.004 [DOI] [PubMed] [Google Scholar]
- 10.Yerli H, Yilmaz T, Kaskati T, Gulay H. Qualitative and semiquantitative evaluations of solid breast lesions by sonoelastography. J Ultrasound Med 2011; 30: 179–86. [DOI] [PubMed] [Google Scholar]
- 11.Cho N, Moon WK, Kim HY, Chang JM, Park SH, Lyou CY. Sonoelastographic strain index for differentiation of benign and malignant nonpalpable breast masses. J Ultrasound Med 2010; 29: 1–7. [DOI] [PubMed] [Google Scholar]
- 12.Pallwein L, Mitterberger M, Struve P, Pinggera G, Horninger W, Bartsch G, et al. Real-time elastography for detecting prostate cancer: preliminary experience. BJU Int 2007; 100: 42–6. doi: 10.1111/j.1464-410X.2007.06851.x [DOI] [PubMed] [Google Scholar]
- 13.Zhang Y, Tang J, Li YM, Fei X, Lv FQ, He EH, et al. Differentiation of prostate cancer from benign lesions using strain index of transrectal real-time tissue elastography. Eur J Radiol 2012; 81: 857–62. doi: 10.1016/j.ejrad.2011.02.037 [DOI] [PubMed] [Google Scholar]
- 14.Lyshchik A, Higashi T, Asato R, Tanaka S, Ito J, Mai JJ, et al. Thyroid gland tumor diagnosis at US elastography. Radiology 2005; 237: 202–11. doi: 10.1148/radiol.2363041248 [DOI] [PubMed] [Google Scholar]
- 15.Rago T, Santini F, Scutari M, Pinchera A, Vitti P. Elastography: new developments in ultrasound for predicting malignancy in thyroid nodules. J Clin Endocrinol Metab 2007; 92: 2917–22. doi: 10.1210/jc.2007-0641 [DOI] [PubMed] [Google Scholar]
- 16.Yerli H, Yilmaz T, Oztop I. Clinical importance of diastolic sonoelastographic scoring in the management of thyroid nodules. AJNR Am J Neuroradiol 2013; 34: E27–30. doi: 10.3174/ajnr.A2751 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Wang H, Brylka D, Sun LN, Lin YQ, Sui GQ, Gao J. Comparison of strain ratio with elastography score system in differentiating malignant from benign thyroid nodules. Clin Imaging 2013; 37: 50–5. doi: 10.1016/j.clinimag.2012.04.003 [DOI] [PubMed] [Google Scholar]
- 18.Lyshchik A, Higashi T, Asato R, Tanaka S, Ito J, Hiraoka M, et al. Cervical lymph node metastases: diagnosis at sonoelastography—initial experience. Radiology 2007; 243: 258–67. doi: 10.1148/radiol.2431052032 [DOI] [PubMed] [Google Scholar]
- 19.Alam F, Naito K, Horiguchi J, Fukuda H, Tachikake T, Ito K. Accuracy of sonographic elastography in the differential diagnosis of enlarged cervical lymph nodes: comparison with conventional B-mode sonography. AJR Am J Roentgenol 2008; 191: 604–10. doi: 10.2214/AJR.07.3401 [DOI] [PubMed] [Google Scholar]
- 20.Hinz T, Hoeller T, Wenzel J, Bieber T, Schmid-Wendtner MH. Real-time tissue elastography as promising diagnostic tool for diagnosis of lymph node metastases in patients with malignant melanoma: a prospective single-center experience. Dermatology 2013; 226: 81–90. doi: 10.1159/000346942 [DOI] [PubMed] [Google Scholar]
- 21.Bhatia KS, Rasalkar DD, Lee YP, Wong KT, King AD, Yuen HY, et al. Evaluation of real-time qualitative sonoelastography of focal lesions in the parotid and submandibular glands: applications and limitations. Eur Radiol 2010; 20: 1958–64. doi: 10.1007/s00330-010-1756-0 [DOI] [PubMed] [Google Scholar]
- 22.Klintworth N, Mantsopoulos K, Zenk J, Psychogios G, Iro H, Bozzato A. Sonoelastography of parotid gland tumours: initial experience and identification of characteristic patterns. Eur Radiol 2012; 22: 947–56. doi: 10.1007/s00330-011-2344-7 [DOI] [PubMed] [Google Scholar]
- 23.Celebi I, Mahmutoglu AS. Early results of real-time qualitative sonoelastography in the evaluation of parotid gland masses: a study with histopathological correlation. Acta Radiol 2013; 54: 35–41. doi: 10.1258/ar.2012.120405 [DOI] [PubMed] [Google Scholar]
- 24.Yerli H, Eski E, Korucuk E, Kaskati T, Agildere AM. Sonoelastographic qualitative analysis for management of salivary gland masses. J Ultrasound Med 2012; 31: 1083–9. [DOI] [PubMed] [Google Scholar]
- 25.Bhatia KS, Lee YY, Yuen EH, Ahuja AT. Ultrasound elastography in the head and neck. Part I. Basic principles and practical aspects. Cancer Imaging 2013; 13: 253–9. doi: 10.1102/1470-7330.2013.0026 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Drakonaki EE, Allen GM, Wilson DJ. Ultrasound elastography for musculoskeletal applications. Br J Radiol 2012; 85: 1435–45. doi: 10.1259/bjr/93042867 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Franchi-Abella S, Elie C, Correas JM. Ultrasound elastography: advantages, limitations and artefacts of the different techniques from a study on a phantom. Diagn Interv Imaging 2013; 94: 497–501. doi: 10.1016/j.diii.2013.01.024 [DOI] [PubMed] [Google Scholar]
- 28.Varghese T, Ophir J, Krouskop TA. Nonlinear stress-strain relationships in tissue and their effect on the contrast-to-noise ratio in elastograms. Ultrasound Med Biol 2000; 26: 839–51. [DOI] [PubMed] [Google Scholar]
- 29.Krouskop TA, Wheeler TM, Kallel F, Garra BS, Hall T. Elastic moduli of breast and prostate tissues under compression. Ultrason Imaging 1998; 20: 260–74. [DOI] [PubMed] [Google Scholar]
- 30.Havre RF, Elde E, Gilja OH, Odegaard S, Eide GE, Matre K, et al. Freehand real-time elastography: impact of scanning parameters on image quality and in vitro intra- and interobserver validations. Ultrasound Med Biol 2008; 34: 1638–50. [DOI] [PubMed] [Google Scholar]