Published in final edited form as: J Cereb Blood Flow Metab. 2000 Aug;20(8):1225–1236. doi: 10.1097/00004647-200008000-00008

Quantitative Measurements of Cerebral Blood Oxygen Saturation Using Magnetic Resonance Imaging

Hongyu An *, Weili Lin *,
PMCID: PMC4096835  NIHMSID: NIHMS598300  PMID: 10950383

Summary

A quantitative estimate of cerebral blood oxygen saturation is of critical importance in the investigation of cerebrovascular disease because of the fact that it could potentially provide information on tissue viability in vivo. In the current study, a multi-echo gradient and spin echo magnetic resonance imaging sequence was used to acquire images from eight normal volunteer subjects. All images were acquired on a Siemens 1.5T Symphony whole-body scanner (Siemens, Erlangen, Germany). A theoretical signal model, which describes the signal dephasing phenomena in the presence of deoxyhemoglobin, was used for postprocessing of the acquired images and obtaining a quantitative measurement of cerebral blood oxygen saturation in vivo. With a region-of-interest analysis, a mean cerebral blood oxygen saturation of 58.4% ± 1.8% was obtained in the brain parenchyma from all volunteers. It is in excellent agreement with the known cerebral blood oxygen saturation under normal physiologic conditions in humans. Although further studies are needed to overcome some of the confounding factors affecting the estimates of cerebral blood oxygen saturation, these preliminary results are encouraging and should open a new avenue for the noninvasive investigation of cerebral oxygen metabolism under different pathophysiologic conditions using a magnetic resonance imaging approach.

Keywords: BOLD, Cerebral blood oxygen saturation, Magnetic resonance imaging


Clinical outcomes from a hypoxic or ischemic insult are a function of the magnitude and duration of diminished oxygen delivery to the brain (Heiss et al., 1983; Jones et al., 1981; Plum et al., 1986). The pathophysiology of altered oxygen delivery and oxygen demand can be viewed as a highly dynamic process (Grotta et al., 1986; Heiss et al., 1994). These fundamental insights have broad implications for a wide variety of disorders that involve or may involve abnormal cerebral oxygen delivery and oxygen demand, such as ischemic stroke, cerebral trauma, and shock. The most direct (and therefore optimal) measure of the balance between oxygen supply and demand is a measure of the cerebrovenous oxygen content, CvO2 (Siesjo, 1978), which represents the remaining oxygen dissolved in plasma and the residual saturation of venous hemoglobin. Because there is a minimal barrier to oxygen diffusion, CvO2 reflects the amount of delivered oxygen that is not consumed by the brain. This relationship is maintained even under clinically relevant low flow conditions (Sharan et al., 1989). For these reasons, many investigators have developed methods for measuring the CvO2, or the partial pressure of oxygen in venous blood (PvO2), from within the venous sinuses or jugular system (Cruz et al., 1990; Cruz, 1988, 1993; Bullock et al., 1993). However, these methods are invasive in nature and carry certain risks preventing them from being widely used in a daily clinical setting.

With the advent of fast imaging techniques and the improved understanding of the biophysical basis of blood-oxygen level dependent (BOLD) (Ogawa and Lee, 1990a; Ogawa et al., 1990b,c, 1993a,b) contrast mechanisms, magnetic resonance imaging (MRI) now has the potential for studying the pathophysiology of disordered brain oxygen metabolism. Specifically, it has been demonstrated that when the concentration of deoxyhemoglobin is increased, a decrease in T2 and T2* is anticipated, resulting in a decrease in MR signal intensity in T2- and T2*-weighted images and vice versa. With animal models many investigators demonstrated that BOLD effects can be used to monitor the changes of oxygen saturation in vivo under pathophysiologic conditions, such as hypoxia (Turner et al., 1991; Prielmeier et al., 1994; Jezzard et al., 1994; Hoppel et al., 1993; Rostrup et al., 1995; Kennan et al., 1997; Lin et al., 1998a), hyper- and hypocapnia (Jezzard et al., 1994; Davis et al., 1998; Lin et al., 1999), hemodilution (Lin et al., 1998b), and ischemia (De Crespigny et al., 1992; Ono et al., 1997). However, all of the above studies only focus on relative measurements of cerebral blood oxygen saturation (Y) and little attention has been given as to how an absolute measurement of Y can be obtained with the BOLD effects.

Recently, with ex vivo blood samples and in a range of physiologically relevant oxygen saturation, both Wright et al. (1991) and Foltz et al. (1999) demonstrated that a calibration curve could be obtained between T2 of the blood and its oxygen saturation. Subsequently, this experimentally determined calibration curve was used for in vivo experiments. A positive correlation was shown when comparing the MR estimated oxygen saturation at the level of the descending aorta with those obtained from blood gas analysis of the blood samples obtained through an intraarterial catheter placed within the descending aorta. However, only results from large vessels were available, making it difficult to assess its ability in obtaining blood oxygen saturation within the capillary, particularly in the brain parenchyma. Conversely, several theoretical models have been proposed to characterize the signal behavior in the presence of local magnetic field inhomogeneities, such as deoxyhemoglobin and contrast agent (Majumdar, 1991; Majumdar et al., 1991; Muller et al., 1991; Yablonskiy and Haacke, 1994; Kennan et al., 1994; Stables et al., 1998; Van Zijl et al., 1998); they can be separated into two different categories based on the underlying sources, which contribute to the signal loss observed in MR images. As the signal model proposed by Van Zijl et al. (1998), only signal loss within the intravascular space was considered, whereas signal loss outside of the intravascular space was ignored. The major advantage of this approach is that it is insensitive to the effects of magnetic field variation induced by sources other than deoxyhemoglobin. However, the main drawback is that the amount of deoxyhemoglobin-induced signal loss relies largely on the available blood volume. Because it has been demonstrated that the normal cerebral blood volume (CBV) in the brain parenchyma ranges between 2% to 5% (Grubb et al., 1973; Sakai et al., 1985; Perlmutter et al., 1987; Brooks et al., 1985; Leggett and Williams, 1991; Lin et al., 1997), it may lead to poor signal-to-noise for this approach. In contrast, the signal model proposed by Yablonskiy and Haacke (1994) focuses on deoxyhemoglobin-induced signal loss outside of the intravascular space, which could potentially be more easily detected with a gradient echo imaging approach. Therefore, in this study, a multi-echo gradient and spin echo sequence was developed to acquire images from normal volunteers. These images were subsequently used to obtain quantitative estimates of cerebral blood oxygen saturation in vivo based upon the theoretical model proposed by Yablonskiy and Haacke (1994).

Theory

For a set of randomly orientated cylinders, which contain paramagnetic particles, a theoretical model was proposed by Yablonskiy and Haacke (1994) to characterize the MR signal dephasing phenomena in the static dephasing regime. It states that the measured MR signal (the effects of spin-spin relaxation (T2) are ignored in the following equations) can be characterized as

S¯(t)=ρ·(1-λ)·exp[-λ·f¯c(δω·t)] (1)

and

f¯c(δω·t)=13·01(2+u)·1-u·1-J0(32·δω·t·u)u2·du (2)

where ρ is the spin density, λ is the volume fraction of the cylinders, J0(x) is the zeroth order Bessel function, and δω is the characteristic frequency that corresponds to the nuclear precession frequency in the equatorial field. In addition to the assumption of a random orientation for the cylinders, two major assumptions are made for this signal model. First, signal loss resulting from diffusion effects is ignored. Second, the signal loss induced by the paramagnetic particles within the cylinders is also neglected. The implications of these assumptions in the estimates of quantitative blood oxygen saturation in vivo are addressed in detail in the Discussion.

In the case of brain parenchyma, capillary vessels form a complicated interconnecting network of long cylinders, compared with their radii, with random orientations. Therefore, the theoretical model proposed by Yablonskiy and Haacke (1994) for a set of randomly orientated cylinders could be used to characterize the signal loss induced by the deoxyhemoglobin, the paramagnetic particles, outside of the intravascular space. If one assumes that the arterial blood is fully oxygenated then λ represents the venous blood volume and δω is the deoxyhemoglobin induced frequency shift which is defined as (Yablonskiy and Haacke, 1994)

δω=γ·43·π·Δχ0·Hct·(1-Y)·B0 (3)

where γ is the gyromagnetic ratio, which is equal to 2.68 × 108 rad/s/Tesla, Hct is the fractional hematocrit, B0 is the main magnetic field strength, Y is the fractional cerebral blood oxygen saturation, and Δχ0 is the susceptibility difference between fully oxygenated and fully deoxygenated blood that has been measured to be 0.18 ppm per unit Hct (Weisskoff and Kiihne, 1992). To obtain a quantitative measurement of Y, δω must be extracted from Eq. 1. Unfortunately, Eq. 1 can not be solved analytically. Two asymptotic forms, namely a short time scale and a long time scale expression, are used to approximate Eq. 1 (Yablonskiy and Haacke, 1994).

For the short time scale, where δω · |t| ≤ 1.5, the asymptotic form can be written as

S¯s(t)=ρ·(1-λ)·exp[-0.3·λ·(δω·t)2]. (4)

The subscript s denotes the short time scale. In contrast, for the long time scale, where δω · |t| > 1.5, the MR signal can be approximated by the following asymptotic form

S¯l(t)=ρ·(1-λ)·exp[-R2·(t-tc)]. (5)

where the subscript l denotes the long time scale and tc, the critical time, is defined as tc = 1/δω. Here R2(=R2-R2,whereR2=1/T2andR2=1/T2) is the relaxation rate that characterizes the signal loss caused by local susceptibilities and can be written as

R2=λ·δω (6)

Therefore, when R2′ and λ are known, an estimate of δω can be obtained from Eq. 6 that, in turn, provides a quantitative measurement of Y through Eq. 3.

MATERIALS AND METHODS

Magnetic resonance imaging experiments

Magnetic resonance imaging protocols were approved by the Human Study Committee at Washington University. In total, 8 normal healthy volunteer subjects were studied and written informed consent was obtained from all subjects.

All images were acquired on a Siemens 1.5T Symphony whole-body scanner (Siemens Medical Systems, Erlangen, Germany) with a gradient strength of 22 mT/m. A four-channel head array coil was used to receive the signal, whereas the body coil was used for transmitting. A two-dimensional, multi-echo gradient and spin echo sequence (Fig. 1) was used to acquire images and subsequently used to extract R2, R2′, and λ. In total, 21 echoes with an echo spacing of 4.96 milliseconds were acquired. The spin echo occurred at the eleventh echo and 10 echoes were placed symmetrically on each side of the spin echo. The echo times ranged from 63.18 milliseconds to 162.38 milliseconds with the spin echo occurring at 112.78 milliseconds. The imaging parameters were as follows: TR = 1000 milliseconds; field-of-view = 160 × 256 mm2; a matrix size of 160 × 256 resulting in an inplane resolution of 1 × 1 mm2; slice thickness = 7.5 mm; and two acquisitions to improve the signal-to-noise ratio. Although multislice acquisition could be achieved with the proposed method, only one slice was acquired in this study so that the potential crosstalk, signal interference between two adjacent slices because of an imperfect rf slice profile, could be minimized.

FIG. 1.

FIG. 1

Sequence diagram of the multi-echo gradient and spin echo. GSS, GPE, and GR represent the slice select, phase encoding, and frequency encoding, respectively. Analog-to-digital converter (ADC) indicates where images are acquired.

Estimations of R2, R2′, λ, and Y

All images were transferred to a Sun Workstation for postprocessing. To improve the signal-to-noise ratio, images were first collapsed to a matrix size of 128 × 128 before any postprocessing. Subsequently, R2 was extracted using

R2=ln(S(TE-τ)S(TE+τ))/2τ (7)

where TE is the echo time for the spin echo and τ is the time interval between the spin echo and the acquired gradient echoes (Fig. 1). By so doing, it is possible to obtain R2 estimates without worrying about the potential confounding factors associated with the differences in rf slice profiles between the 90° and 180° pulses (Yablonskiy and Haacke, 1997). To improve the accuracy of the R2 estimates, instead of relying on one pair of gradient echo images symmetrically placed about the spin echo, five pairs of gradient echo images (the 1–5th and the 17–21th gradient echoes) were used (Fig. 3A, unfilled squares) for the estimates of R2. Although it is possible to use more than five pairs of gradient echo images, the accuracy of the R2 estimates for the tissues of interest, such as gray matter and white matter, could be compromised if those with too short of a τ are included. A more detailed discussion for the choice of five gradient echo pairs is given in the Appendix.

FIG. 3.

FIG. 3

Experimentally measured signal is plotted as a function of TE (A) and the measured magnetic resonance signal after removing the R2 effect (B). Unfilled squares indicate the echoes used for the R2 estimates. In contrast, the filled box indicates the echoes used for the R2′ estimates, whereas the unfilled box represents the echoes used to fit a second order polynomial function.

After R2 was obtained, the effect of R2 was removed from the original signal so that Eqs. 4 and 5 could be used to describe the MR signal behavior in the presence of local magnetic field inhomogeneities. To better describe how R2′ and λ were estimated from the acquired images by the multi-echo gradient and spin echo sequence (Fig. 1), the t in Eqs. 4 and 5 was replaced by ΔTEi, which represents the time interval between the echo time of the ith gradient echo and the echo time of the spin echo was 1 ≤ i ≤ 21.

By taking the natural logarithm of both sides of Eqs. 4 and 5, the authors found

ln(S¯s(ΔTEi))=a+[-0.3·λ·(δω·ΔTEi)2] (8)
ln(S¯l(ΔTEi))=a+[-R2c·(ΔTEi-tc)]=-R2c·ΔTEi+b (9)

where a = ln(ρ(1 − λ)) and b = a + R2′tc. Notice that ln((sTEi)) has a quadratic dependence on ΔTEi, while ln(lTEi)) is linearly dependent on ΔTEi with a slope equal to R2′. Therefore, R2′ can be obtained with a linear least squares curve fitting of the MR signal of the gradient echoes as long as δω · |ΔTEi| is greater than 1.5. In contrast, the signal from the central 9 echoes including the spin echo and 4 gradient echoes on each side of the spin echo (7th–15th echoes, the unfilled box in Fig. 3B) was fitted to a second order polynomial function. This was done to minimize the effects of noise in the estimate of MR signal at the spin echo SsTE = 0)) so that a more accurate estimate of λ could be obtained through the following equation (Yablonskiy, 1998).

λ=ln(S¯l(ΔTE=0)/S¯s(ΔTE=0)). (10)

After obtaining R2′ and λ, δω can then be estimated from Eq. 6 and, finally, an absolute estimate of Y can be obtained through Eq. 3. In this study, a constant Hct of 0.42 was used for all subjects and a small vessel Hct to large vessel Hct ratio of 0.85 (Eichling et al., 1975) was used to convert the large vessel Hct to small vessel Hct.

Region of interest analysis

To obtain an estimate of R2, R2′, and λ in the brain parenchyma, an ROI that encompassed both hemispheres was used. However, six ROIs that equally divided each hemisphere into three ROIs were defined for the measurements of cerebral blood oxygen saturation of each volunteer. Both mean and standard deviation of R2, R2′, λ, and Y were recorded so that a comparison between subjects could be made.

Error analysis

An error analysis was performed to investigate how the presence of noise in the experimentally acquired MR images propagated through R2, R2′ and λ estimations to the final Y maps. Detailed descriptions of the error analysis for all estimated parameters including R2, R2′, λ, and Y are given in the Appendix.

RESULTS

Representative images obtained from the proposed sequence (Fig. 1), including the first echo (Fig. 2A), the 11th echo (spin echo, Fig. 2B), and the last echo (Fig. 2C) from one subject are shown in Fig. 2. Although the first and the last echoes have exactly the same R2′ effects, a substantial decrease in signal-to-noise ratio is observed for the last echo when compared with the first echo. This is not surprising because the last echo has a larger T2 effect, leading to a reduction in MR signal, which in turn, decreases signal-to-noise ratio. In addition, the time interval between the first echo (last echo) and the spin echo is 49.6 milliseconds for the multi-echo gradient and spin sequence, which is analogous to a gradient echo sequence with a TE of 49.6 milliseconds. As a result, tissues with short T2*, such as subcutaneous fat, will not be visible in the first and the last gradient echo images. In contrast, the signal of subcutaneous fat (arrow) is recovered in the spin echo image because of the signal refocusing achieved by the 180° pulse.

FIG. 2.

FIG. 2

Anatomic images from the first echo (A), spin echo (B) and the last echo (C) obtained from the multi-echo gradient and spin echo sequence are shown.

Magnetic resonance signal as a function of TE is shown in Fig. 3A from a 4 × 4 region of one subject. The 21 points correspond to the signal from the 21 echoes acquired. Although the MR signal is anticipated to decay exponentially with two time constants, T2 and T2′, it is not immediately apparent owing to the fact that T2′ is expected to be small and T2 will dominate the signal decay. After removing the effects of T2, the MR signal as a function of echo time is shown in Fig. 3B. As anticipated, the signal curve is symmetrical about the spin echo with the highest signal observed around the spin echo. The changes of MR signal are minimal when ΔTE is small, whereas a more rapidly decaying signal is observed for larger ΔTE. This is in agreement with theoretical expectations as described by Eqs. 4 and 5.

The estimated R2 (Fig. 4A), R2* (Fig. 4B), R2′ (Fig. 4C), blood volume fraction (Fig. 4D), and Y (Fig. 4E) maps from subject 8 are shown in Fig. 4 through the parameter estimation approach outlined in Materials and Methods. As anticipated, the CSF has a lower R2 when compared to the brain parenchyma and a rather uniform R2 map is observed for the brain parenchyma. In contrast, a slightly elevated R2* (arrow), R2′ (arrow) and blood volume fraction (arrow) is observed at the frontal area when compared with other regions of the brain parenchyma. Finally, although some spatial variations are also observed in the Y map (Fig. 4E), a rather uniform cerebral blood oxygen saturation ranging between 55.7% to 57.9% (Table 1, subject 8) in the brain parenchyma is obtained. Regions marked as red represent where the parameter estimations failed to estimate Y. The reasons for this are addressed in the Discussion.

FIG. 4.

FIG. 4

R2 (A), R2* (B), R2′ (C), volume fraction (D), and color coded oxygen saturation Y (E) are shown. Color bar indicates the corresponding oxygen saturation from 0% to 100%.

TABLE 1.

Cerebral blood oxygen saturation measurements from volunteer studies

Subject Y (%)
1 2 3 4 5 6 7 8
Region 1 58.2 ± 7.3 58.7 ± 9.3 60.2 ± 5.4 55.6 ± 5.9 59.7 ± 7.9 59.9 ± 3.6 54.1 ± 4.9 55.7 ± 4.9
Region 2 57.5 ± 4.5 58.6 ± 5.0 58.8 ± 4.7 59.1 ± 11.2 65.2 ± 8.4 60.1 ± 4.3 52.2 ± 7.1 57.9 ± 7.2
Region 3 58.1 ± 3.7 56.5 ± 3.7 60.9 ± 7.4 55.1 ± 12.9 58.8 ± 6.5 61.2 ± 5.0 57.2 ± 8.1 57.2 ± 9.1
Region 4 55.5 ± 4.8 65.5 ± 8.8 58.1 ± 4.3 56.6 ± 5.2 55.3 ± 4.5 63.4 ± 3.7 55.1 ± 4.6 57.0 ± 5.2
Region 5 55.7 ± 5.3 61.1 ± 5.2 56.7 ± 4.3 62.8 ± 8.9 59.4 ± 5.8 63.5 ± 6.4 58.8 ± 7.4 57.3 ± 8.6
Region 6 57.1 ± 3.6 56.0 ± 4.0 53.7 ± 8.1 67.9 ± 9.3 57.6 ± 5.9 61.7 ± 2.8 54.1 ± 6.7 56.5 ± 7.3

Mean ± SD 57.5 ± 3.6 59.4 ± 6.4 58.1 ± 5.9 59.5 ± 9.3 59.3 ± 6.6 61.6 ± 4.5 55.3 ± 6.6 56.9 ± 7.2

The measured R2, R2′, and λ in the brain parenchyma from all subjects are as follows: R2 = 10.5 ± 1.69 Hz; R2′ = 6.7 ± 3.92 Hz; and λ = 16.1 ± 7.43%. In addition, a summary of the estimated Y values from all volunteers is shown in Table 1 for all six regions, respectively. The estimated Y ranges between 55.3% to 61.6% in all subjects with a group mean of 58.4% ± 1.8%.

Finally, to investigate whether the observed spatial variation in the Y is caused by the noise in the acquired images or is indicative of physiologic variations, the calculated error of Y based on an error propagation analysis (Appendix) is compared with the experimentally measured intersubject variability shown in Table 1. An overall intersubject variability of 6.6% was obtained from the Y maps of all subjects. In contrast, an error of 5.24% is anticipated based on the error analysis (Eq. A12). Because the experimentally measured intersubject variability is only slightly higher than that obtained from the error analysis, the current results suggested that the observed spatial variation in the Y maps is likely to be dominated by noise with relatively little effects from physiologic variation.

DISCUSSION

It is well established that the brain extracts roughly 40% to 50% of available oxygen from the arterial blood to maintain neuronal activities under normal physiologic conditions in humans. With PET and continuous inhalation of 15O2, Yamauchi et al. (1996) studied 10 normal volunteer subjects and an oxygen extraction fraction (OEF) of 42.6 ± 5.1% was obtained from different cortical areas. In addition, similar results were reported by Nakane et al. (1998) with continuous infusion of H215O. In total, 9 patients who exhibited no neurologic deficits and no apparent lesions on computed tomography were investigated. They showed that the OEF ranged between 40% to 44% in different cortical areas. To compare our estimates of cerebral blood oxygen saturation (Y) with those reported OEF values through PET measurements, it is necessary to convert the normal OEF values to the cerebral blood oxygen saturation.

Under normal physiologic conditions, arterial blood can be assumed to be fully saturated (100%) with an oxygen content, CaO2, on the order of 19.2 vol % (Cooney, 1976). Assuming that the normal OEF is 42.6% based on the results obtained from Yamauchi et al. (1996), the CvO2 can be calculated through the relation of OEF = (CaO2-CvO2)/CaO2 that yields a CvO2 of 10.9 vol %. Subsequently, the CvO2 can be converted to oxygen saturation via the oxygen dissociation curve and the Hill equation (Cooney, 1976) through

Y=(P/P0)2.71+(P/P0)2.7

where P is the partial pressure of oxygen and P0 is the partial pressure for 50% saturation that is 27.2 mm Hg for PO2 = 95 mm Hg, PCO2 = 40 mm Hg, pH = 7.4 and a body temperature of 37°C. A cerebral blood oxygen saturation of 56.5% is therefore obtained based on the reported OEF in the literature. This is in excellent agreement with our results. A mean Y of 58.4% ± 1.8% was obtained from all subjects, implying that 41.6% ± 1.8% of available oxygen is extracted for maintaining normal neuronal activities from the arterial blood under normal physiologic conditions in humans. Our findings are encouraging and demonstrate the feasibility of using an MR imaging approach to obtain an estimate of in vivo cerebral blood oxygen saturation noninvasively.

However, the observed large intersubject variability in the estimates of Y from all subjects deserves further comment. As shown in Results, the experimentally measured intersubject variability (6.6%) is slightly higher than that obtained from the error analysis (5.24%). This suggests that the observed Y variations in the brain parenchyma are predominately derived from the presence of noise in MR images rather than physiologic variation. This is perhaps not surprising because a balance between cerebral blood flow and oxygen metabolism is normally preserved under normal physiologic conditions. The ratio of the two parameters (or OEF) has shown relatively little regional variation in the brain (Derdeyn and Powers, 1997). Our results are in agreement with those obtained from PET. However, given the large anticipated errors in our Y estimates, it is necessary to further improve the signal-to-noise ratios of the acquired images before the application of the proposed method in a clinical setting.

Model assumptions

Several assumptions are made for the signal model proposed by Yablonskiy and Haacke (1994) and three of them are particularly relevant to our study. The potential implications of the three assumptions on the measurements of Y are addressed below. First, the blood vessels are assumed to have a random orientation with respect to the static magnetic field so that a statistical approach can be used to derive the signal equation (Eq. 1). This was justified because the main thrust of our study was to obtain a quantitative measurement of cerebral blood oxygen saturation in the brain parenchyma. Given the large voxel size of 2 × 2 × 7.5 mm3 (after filtering the images to a matrix size of 128 × 128) and the fact that capillaries are very small in size (~10 μm) and randomly distributed in the brain parenchyma, this assumption should be fulfilled. However, it is not applicable to large vessels, such as pial veins, which could potentially occupy a substantial fraction of one voxel. This may explain some of the regions that are marked in red in Fig. 4E, in which the presence of large vessels is anticipated. As a result, the employed signal model fails to provide an estimate of Y in these voxels.

Second, only signal loss induced by the paramagnetic particles, deoxyhemoglobin, and outside of the cylinders is considered in this signal model. No intravascular effects are considered. This assumption could potentially impact on the estimates of Y under pathophysiologic conditions in which an elevated rCBV is anticipated (perhaps in regions of high grade tumors; Aroner et al., 1994) or if a voxel consists largely of blood. Functional MRI study is a typical example of the latter scenario in which the signal changes during a task-driven functional MRI study are primarily contributed by signal within the intravascular space of pial veins at a 1.5T system (Haacke et al., 1994; Boxerman et al., 1995; Song et al., 1996). Therefore, the proposed method may not be suitable for functional MRI studies without the appropriate modifications necessary to take into account the signal contribution in the intravascular space.

Nevertheless, normal volunteer subjects under a resting condition were studied. In addition, it is well known that normal cerebral blood volume ranges between 2% to 5% in the brain parenchyma (Grubb et al., 1973; Sakai et al., 1985; Perlmutter et al., 1987; Brooks et al., 1985; Leggett and Williams, 1991; Lin et al., 1997). Accordingly, the effects from large vessels, which do not form a random orientation or contribute largely to the signal in one voxel, are likely to be associated with specific anatomical locations and have minimal effects on the estimates of Y in the brain parenchyma. In addition, only images from the odd echoes (Fig. 1) were acquired and used for the estimates of Y. It is likely to further reduce the signal contributions from the flowing spins in the intravascular space because they will not be completely refocused by the odd echoes, resulting in a reduction in signal for the flowing spins.

Finally, signal dephasing induced by proton diffusion was also ignored in the signal model. To better address the effects of proton diffusion in the estimates of Y, the effects of diffusion in the intravascular space and extravascular space were considered separately. With a theoretical signal model, Kennan et al. (1994) have shown that the effects of diffusion in the intravascular space are both sequence and perturber size dependent. Based on simulation, they demonstrate that the effects of diffusion on MR signal are maximal when the size of the perturbers is on the order of 10μm, roughly the size of a capillary, for a spin echo sequence. Therefore, without taking into account the effects of diffusion in the intravascular space, errors in the Y estimates could occur. However, to what extent diffusion affects the Y estimates is weighted by the rCBV. Therefore, analogous to the previously arguments, the effects of diffusion in the intravascular space should be negligible owing to the rather small rCBV in the brain parenchyma.

In contrast, Kiselev and Posse (1999) have recently proposed an analytical model that appropriately takes into account the effects of diffusion in the extravascular compartment. With a computer simulation, their results suggest that the additional signal dephasing induced by diffusion in the extravascular space will cause the spin echo to occur earlier than expected, resulting in inaccurate measures of R2 with the proposed approach (Eq. 7) in our study. Subsequently, an overestimation of blood volume fraction was expected, leading to an underestimation of Y. To minimize the effects of diffusion in the extravascular space, a second order polynomial function was used to fit the signal around the spin echo including the spin echo (unfilled box in Fig. 3B). By so doing, the effects of diffusion were likely to be reduced and a more accurate estimate of the signal of the spin echo could be obtained.

Effects of sequence timing

As mentioned in the Theory section, the signal model proposed by Yablonskiy and Haacke (1994) can not be solved analytically, and two asymptotic forms were used to approximate the solutions. The limit to separate the two asymptotic forms is the product of δω and |ΔTEi|. When δω*|ΔTEi| is less than 1.5, the MR signal without R2 effects can be approximated by the short time scale expression, whereas the long time scale expression needs to be used when δω*|ΔTEi| is greater than 1.5. The latter requirement is particularly important for obtaining an accurate estimate of R2′ because a linear least squares curve fitting is used to fit the acquired MR signal and the slope of this fitted line provides an estimate of R2′. To ensure that the MR signal used for the curve fitting meets the requirement, a priori knowledge of δω is needed, which in turn, requires a known Y. Obviously, Y is the parameter that will be estimated and yet it is unknown before the MR imaging experiments. Therefore, it is of importance to determine the upper limit of the oxygen saturation that can be accurately estimated with the used sequence timing of the multi-echo gradient and spin echo sequence in our study.

Four MR images acquired with a ΔTEi of 34.72 milliseconds, 39.68 milliseconds, 44.64 milliseconds, and 49.6 milliseconds (filled box in Fig. 3B) were used for the linear curve fitting to obtain an estimate of R2′. Under this condition, the product of δω*|ΔTEi| will always be greater than 1.5 for all four images when Y is less than 60%. It fulfills the requirement for the utilization of the long time scale expression to obtain an estimate of R2′, which in turn, allows us to obtain an accurate estimate of Y. Conversely, when Y is greater than 60%, some of the four images used for the linear curve fitting will lie within the short time scale regime leading to inaccuracy in the estimates of R2′. This is because the signal within the short time scale regime will no longer be linearly dependent on ΔTEi, but rather quadratically dependent on ΔTEi. As a result, an underestimation of Y can occur when Y is greater than 60%. The degree of underestimation in Y is highly dependent on the true Y values, with larger underestimation for higher Y. Although a quantitative estimate of the degree of underestimation for Y values greater than 60% can be obtained with a Monte Carlo simulation, it was not done because a cerebral blood oxygen saturation of 56.5% was reported from normal volunteer subjects (Yamauchi et al., 1996). Therefore, the employed sequence timing should allow us to obtain an accurate estimate of Y in normal volunteers. However, modifications by increasing ΔTEi will be needed when cerebral blood oxygen saturation greater than 60% is anticipated such as under hyperemic or hypercapnic conditions. Further studies will be needed to assess the accuracy of the proposed methods in obtaining quantitative estimates of Y under these pathophysiologic conditions.

Effects of other paramagnetic sources and magnetic field variations

It has been demonstrated by many investigators with MR and PET that the total cerebral blood volume including arterial and venous blood pools is approximately 2% to 3% and 4% to 5% for the white matter and gray matter, respectively (Grubb et al., 1973; Sakai et al., 1985; Perlmutter et al., 1987; Brooks et al., 1985; Leggett and Williams, 1991, Kuppusamy, 1996). Although a wide range of the ratio of cerebral venous blood volume to the total cerebral blood volume has been reported (0.75 to 0.85) in the literature (Tomita et al., 1978; Tomita, 1988), Pollard et al. (1996a,b) recently demonstrated that a ratio of 0.75 would be a reasonable choice under normal physiologic conditions. Thus, a venous blood volume fraction of 1.5% to 2.3% and 3% to 3.8% is anticipated for white matter and gray matter, respectively.

However, a λ = 16.1 ± 7.43% was obtained from all subjects in our study. This is clearly much higher than the reported venous blood volume fraction. In addition, an elevated R2′ and λ in the frontal areas where air–tissue interface effects are present is also observed in some volunteer subjects. These observed elevations of R2′ and λ are likely because paramagnetic sources other than deoxyhemoglobin are present in vivo and can induce an additional signal loss in the MR images. These sources include other paramagnetic particles and magnetic field variations. The former case contains, for example, blood products and calcified plaques. Because normal volunteer subjects were studied, it is unlikely that these paramagnetic sources existed. In contrast, the latter case could potentially affect our estimates of Y. There are two potential situations in which magnetic field variations can be present: the local susceptibility differences between tissues, such as air–tissue interfaces, and a nonuniform static magnetic field across the slice thickness. In both cases, the additional signal decay can be characterized by an additional sinc function multiplied into Eq. 1 (Hardy et al., 1990), leading to a more rapid signal decay when compared with the condition in which deoxyhemoglobin is the only paramagnetic source. As a result, an overestimation of R2′ is anticipated. In addition, because the effects of magnetic field variation will be completely eliminated by the application of a 180° refocusing pulse, the signal in a spin echo image remain unaffected. Consequently, an overestimation of λ is also anticipated. This is due to the fact that the numerator in Eq. 10 is increased because of increased R2′, Eq. 9, whereas the denominator will remain unchanged. As a result, an overestimation of R2′ and λ is likely to be present in our results.

Because the main thrust of this study was to obtain a quantitative estimate of in vivo blood oxygen saturation, the observed overestimation of R2′ and λ needs to be discussed in the context of Y. As shown in Eq. 6, δω, and thus Y, is obtained by taking the ratio of R2′ and λ. With the additional signal loss induced by the presence of magnetic field variation, R2′ and λ were expected to be overestimated. If one can assume that the overestimated R2′ and λ are similar in magnitude to the first order, then the magnetic field variation will have minimal effects on Y. Based on the calculation of the correlation coefficient between R2′ and λ (Appendix, Eq. A13), it was found to be 0.94, indicating that the two parameters have a highly linear relation. This may explain the observed a rather uniform Y in the entire brain parenchyma in Fig. 4E despite the elevated R2′ and λ in the frontal regions where air–tissue interface is present. More importantly, a group mean of 58.4% ± 1.8% was obtained for the MR estimated Y from all volunteers, comparing favorably with that reported in the literature. Nonetheless, to obtain a more accurate estimate of Y and λ, local magnetic field variations will need to be taken into account. Several different approaches to minimize the effects of magnetic field variations are currently under investigation and it is beyond the scope of the current study.

Ratio of small vessel to large vessel Hct

To obtain a quantitative measurement of cerebral blood oxygen saturation, a known Hct is required (Eq. 3). Although it is possible to obtain a direct measurement of the large vessel Hct from each volunteer, it is not done because the large vessel Hct has been known to be rather stable in humans. It ranges between 0.4–0.45 depending on the gender (Braunwald et al., 1987). Therefore, a constant large vessel Hct of 0.42 was assumed in this study for all volunteer subjects. However, this large vessel Hct needs to be converted to the small vessel Hct before the estimation of Y through Eq. 3. A wide range of the ratio of small vessel cerebral Hct (cHct) to large vessel Hct (cHct/Hct = 0.62–0.92) has been reported by different investigators in different species including humans (Cremer and Seville, 1983; Everett et al., 1956; Levin and Ausman, 1969; Larsen and Lassen, 1964; Oldendorf et al., 1965). Despite these discrepancies in the reported cHct/Hct ratios, a general consensus of cHct/Hct = 0.85 has been used by PET (Eichling et al., 1975). Therefore, a cHct/Hct of 0.85 was used in the current study to convert the large vessel Hct to cHct.

With a constant Hct used in the current study, errors may have occurred because of the potential variation in Hct between subjects and the uncertainty in cHct/Hct ratio. Specifically, as shown in Eq. 3, δω is directly proportional to Hct. Consequently, errors induced by the inaccuracy in Hct will contribute directly to the estimates of Y. Nevertheless, the errors induced by the former case are a global effect and may be easily overcome with a direct measurement of Hct from each subject. Therefore, the potential variation in Hct between subjects should not affect the proposed method in obtaining a quantitative measurement of cerebral blood oxygen saturation. In contrast, the potential errors induced by the latter factor are likely to be small under normal physiologic conditions and are beyond the scope of this study.

In conclusion, because the discovery of BOLD contrast, MR has shown promise in obtaining quantitative estimates of blood oxygen saturation in vivo. However, technical difficulties in the estimates of parameters, such as R2′ and λ, have hampered the success in obtaining in vivo blood oxygen saturation. The goal of the current study was to demonstrate that an absolute estimate of blood oxygen saturation could be obtained noninvasively with an MR imaging approach. Despite the simplifications of the signal model for the capillary network in an organ as complex as the brain and some confounding factors that still needed to be resolved, a mean cerebral blood oxygen saturation of 58.4% ± 1.8% was obtained from all subjects. This is in excellent agreement with the known cerebral blood oxygen saturation under normal physiologic conditions. Although more systematic validations in animal models will be needed before the application of the proposed method in a clinical setup, our preliminary results are encouraging and underscore the possibility of using an MR imaging approach for obtaining an absolute estimate of cerebral blood oxygen saturation in vivo.

Acknowledgments

This work was supported in part by a NIH grant NS37312.

The authors are grateful to Dr. E. Mark Haacke for his valuable input and participation in several discussions on the theoretical aspects of this work.

Abbreviations used

BOLD

blood oxygen level-dependent

CaO2

arterial oxygen content

CvO2

cerebrovenous oxygen content

CBV

cerebral blood volume

rCBV

regional cerebral blood volume

CT

computed tomography

HCT

fractional hematocrit

MRI

magnetic resonance imaging

OEF

oxygen extraction fraction

PET

positron emission tomography

PvO2

partial pressure of oxygen invenous blood

Y

fractional cerebral blood oxygen saturation

APPENDIX

Error analyses

The noise in the measured MR signal propagates through the estimation of R2, R2′, and λ to the final Y measurement. First, let us define the measured parameters as:

R^2=R2±σR^2 [A1]
R^2=R2±σR^2 [A2]
λ^=λ±σλ^ [A3]
Y^=Y±σY^ [A4]

where R2, R2′, λ, and Y represent the individual parameters without the presence of noise. To calculate the variance of Ŷ, the variance of 2 and then R^2 and λ̂ should be calculated beforehand.

As mentioned in Materials and Methods, 2 is found from Eq. 7. Therefore, the variance in the R2 map can be obtained through

σR^22=(1SNRS(TE-τ)2+1SNRS(TE+τ)2)/(2τ)2 [A5]

where SNRS(TE−τ) and SNRS(TE+τ) are the signal-to-noise ratios of the two gradient echoes occurring at TE−τ and TE+τ, respectively. In the current study, five pairs of the gradient echoes on each side of the spin echo (1st–5th and 17th–21st echo) were used for the estimation of R2. Therefore the variance of 2 can be rewritten as:

σR^22=j(1SNRS(TE-τj)2+1SNRS(TE+τj)2)j(2τj)2 [A6]

where j = 1 to M and M is the number of gradient echo pairs used. Based on the sequence timing of the multi-echo gradient and spin echo sequence and the experimentally measured signal-to-noise ratio of the gradient echo images, the standard deviations of 2 was obtained as σ2 = 0.068 Hz.

The parameter R^2 was obtained by first removing the effects of R2 from the signal, followed by a least squares linear curve fitting. The variance of R^2 can be calculated by taking into account the estimation errors propagated from the estimate of 2 (Press et al., 1992) as

σR^22=αΔ

where

α=iN1σi2σi2=(1+TEij2τj)21SNRi2+TEi2·σR^22 [A7]

and

Δ=α·S(TE2)-S2(TE)S(TE)=iNTEσi2S(TE2)=iNTE2σi2 [A8]

where i = 1 … N and N = 4 because the signal in the first 4 gradient echoes were used for the linear least squares curve fitting, and j = 1 to M and M is the number of gradient echo pairs used for the estimates of R2. Substituting the signal-to-noise ratios of the experimentally acquired images and the variance of R2 estimates through Eq. A6, an estimate of the standard deviation of R^2 yields σR^2=0.536Hz.

The estimates of the errors for the intercept yield σS¯^l(ΔTE=0)=0.0366 and σS¯^s(ΔTE=0)=0.0032 where l refers to the long time scale and s to the short time scale component, respectively. Subsequently, the variance of λ̂ can be calculated as

σλ^2=σS¯^l(ΔTE=0)2+σS¯^s(ΔTE=0)2-2·rS¯^l(ΔTE=0)S¯^s(ΔTE=0)·σS¯^l(ΔTE=0)·σS¯^s(ΔTE=0) [A9]

where rS¯^l(ΔTE=0)S¯^s(ΔTE=0) is the correlation coefficient between S¯^l(ΔTE=0) and S¯^s(ΔTE=0) because the noise will be correlated in this case. The correlation coefficient is obtained from

rS¯^l(ΔTE=0)S¯^s(ΔTE=0)=σS¯^l(ΔTE=0)S¯^s(ΔTE=0)2σS¯^l(ΔTE=0)·σS¯^s(ΔTE=0) [A10]

From Eq. A10, rS¯^l(ΔTE=0)S¯^s(ΔTE=0)=0.89 and thus σλ̂ = 0.0338 from Eq. A9.

Finally, as mentioned previously, Y can be obtained through Eq. 3:

Y^=1-14/3π·γ·Δχ0·Hct·B0·R^2λ^ [A11]

The variance of Ŷ is determined from:

σY^2=(c/λ)2σR^22+(c·R2/λ2)2σλ^2+2·c/λ·(-c·R^2/λ2)·rR^2λ^·σR^2·σλ^=(c/λ)2σR^22+(c·δω/λ)2σλ^2+2·c/λ·(-c·δω/λ)·rR^2λ^·σR^2·σλ^ [A12]

where

c=14/3π·γ·Δχ0·Hct·B0

and rR^2λ^ is the correlation coefficient between R^2 and λ̂ which can be obtained through the following equation:

rR^2λ^=σR^2λ^2σR^2·σλ^ [A13]

By substituting the results from Eqs. A7 and A9,

rR^2λ^=0.94.

Finally, the experimentally measured R^2 (= 6.7 Hz) and λ (= 16.1%) from all subjects were substituted into Eq. A12, and the standard deviation of Ŷ was then found to be σŶ = 5.24%.

Choice of number of gradient echo pairs to calculate R2

The choice of five pairs of gradient echoes (unfilled squares, Fig. 3A) to obtain an estimate of 2 is justified for the following reasons. First, although it is possible to obtain 2 estimates with only a single pair of gradient echo images, the precision of the 2 can be substantially improved by increasing the number of pairs used. However, the accuracy of 2 estimates relies largely on the time interval (2τ) between the echoes utilized. From our calculation, an optimized time interval (2τ) for the estimates of 2 is 2τ ≈ 0.76T2 where the T2 value used corresponds to the tissue of interest. For gray matter and white matter, T2 has been estimated to be 100 milliseconds and 80 milliseconds (Haacke et al., 1999), respectively. Therefore, to obtain accurate estimates of T2 (R2) for both gray matter and white matter, the time interval between the two gradient echoes should be approximately 76 milliseconds and 61 milliseconds, respectively. The time intervals of the five gradient echoes used in the current study were 99.2 milliseconds, 89.28 milliseconds, 79.36 milliseconds, 69.44 milliseconds and 59.52 milliseconds, which are close to the optimized time intervals.

References

  1. Aroner HJ, Gazit IE, Louis DN, et al. Cerebral blood volume maps of gliomas: comparison with tumor grade and histologic findings. Radiology. 1994;191:41–51. doi: 10.1148/radiology.191.1.8134596. [DOI] [PubMed] [Google Scholar]
  2. Boxerman JL, Hamberg LM, Rosen BR, Weiskoff RM. MR Contrast Due to Intravascular Magnetic Susceptibility Perturbations. Magn Reson Med. 1995;34:555–566. doi: 10.1002/mrm.1910340412. [DOI] [PubMed] [Google Scholar]
  3. Braunwald E, Isselbacher KJ, Petersdorf RG, Wilson JD, Martin JB, Fauci AS. Harrison’s principles of internal medicine. New York, NY: McGraw-Hill Book Company; 1987. [Google Scholar]
  4. Brooks DJ, Beaney RP, Leenden KL, Marshall J, Thomas DJ, Jones T. Regional cerebral oxygen utilization, blood flow and blood volume in benign intracranial hypertension studied by PET. Neurology. 1985;35:1030–1034. doi: 10.1212/wnl.35.7.1030. [DOI] [PubMed] [Google Scholar]
  5. Bullock R, Stewart L, Rafferty C, Teasdale GM. Continuous monitoring of jugular bulb oxygen saturation and the effect of drugs acting on cerebral metabolism. Acta Neurochir Suppl (Wein) 1993;59:113–118. doi: 10.1007/978-3-7091-9302-0_20. [DOI] [PubMed] [Google Scholar]
  6. Cooney DO. Biomedical engineering principles: an introduction to fluid, heat, and mass transport processes. New York, NY: Marcel Dekker Inc; 1976. pp. 356–362. [Google Scholar]
  7. Cremer JE, Seville MP. Regional brain blood flow, blood volume, and haematocrit values in the adult rat. J Cereb Blood Flow Metab. 1983;3:254–256. doi: 10.1038/jcbfm.1983.35. [DOI] [PubMed] [Google Scholar]
  8. Cruz J, Miner ME, Allen SJ, Alves WM, Gennarelli TA. Continuous monitoring of cerebral oxygenation in acute brain injury: injection of mannitol during hyperventilation. J Neurosurg. 1990;73:725–730. doi: 10.3171/jns.1990.73.5.0725. [DOI] [PubMed] [Google Scholar]
  9. Cruz J. Jugular venous oximetry. Acta Neurochir Suppl (Wein) 1993;59:86–90. [Google Scholar]
  10. Cruz J. Continuous versus serial global cerebral hemometabolic monitoring: applications in acute brain trauma. Acta Neurochir Suppl (Wein) 1988;42:35–39. doi: 10.1007/978-3-7091-8975-7_8. [DOI] [PubMed] [Google Scholar]
  11. De Crespigny AJ, Wendland MF, Derugin N, Kozniewska E, Moseley ME. Real-time observation of transient focal ischemia and hyperemia in cat brain. Magn Reson Med. 1992;27:391–397. doi: 10.1002/mrm.1910270220. [DOI] [PubMed] [Google Scholar]
  12. Davis TL, Kwong KK, Weisskoff RM, Rosen BR. Calibrated functional MRI: mapping the dynamics of oxidative metabolism. Proc Natl Acad Sci U S A. 1998;95:1834–1839. doi: 10.1073/pnas.95.4.1834. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Derdeyn CP, Powers WJ. Metabolic studies using PET in stroke investigation. In: Welch KMA, Caplan LR, Reis DJ, Siesjo BK, Weir B, editors. Primer on cerebrovascular diseases. San Diego, CA: Academic Press; 1997. pp. 644–650. [Google Scholar]
  14. Eichling JO, Raichle ME, Grubb RL, Larson KB, Ter-Pogossian MM. In vivo determination of cerebral blood volume with radioactive oxygen-15 in the monkey. Circ Res. 1975;37:707–714. doi: 10.1161/01.res.37.6.707. [DOI] [PubMed] [Google Scholar]
  15. Everett NB, Simmons B, Lasher EP. Distribution of blood (Fe59) and plasma (I131) volumes of rats determined by liquid nitrogen freezing. Circ Res. 1956;4:419–424. doi: 10.1161/01.res.4.4.419. [DOI] [PubMed] [Google Scholar]
  16. Foltz WD, Merchant N, Downar E, Stainsby JA, Wright GA. Coronary venous oximetry using mRI. Magn Reson Med. 1999;42:837–848. doi: 10.1002/(sici)1522-2594(199911)42:5<837::aid-mrm3>3.0.co;2-8. [DOI] [PubMed] [Google Scholar]
  17. Grotta JC. Emerging stroke therapies. Neurology. 1986;6(3):285–292. doi: 10.1055/s-2008-1041472. [DOI] [PubMed] [Google Scholar]
  18. Grubb RL, Phelps ME, Ter-Pogossian MM. Regional blood volume in humans: x-ray fluorescence studies. Arch Neurol. 1973;28:31–44. doi: 10.1001/archneur.1973.00490190056006. [DOI] [PubMed] [Google Scholar]
  19. Haacke EM, Brown RW, Thompson MR, Venkatesan R. Magnetic resonance imaging—physical principles and sequence design. New York, NY: John Wiley and Sons; 1999. [Google Scholar]
  20. Haacke EM, Hopkins A, Lai S, Buckley P, Friedman L, Meltzer H, Hedera P, Friedland R, Klein S, Thompson L, et al. 2D and 3D high resolution gradient echo functional imaging of the brain: venous contributions to signal in motor cortex studies. NMR Biomed. 1994;7:54–62. doi: 10.1002/nbm.1940070109. [DOI] [PubMed] [Google Scholar]
  21. Hardy PA, Kucharczyk W, Henkelman RM. Cause of signal loss in MR images of old hemorrhagic lesions. Radiology. 1990;174:549–555. doi: 10.1148/radiology.174.2.2296664. [DOI] [PubMed] [Google Scholar]
  22. Heiss WD, Rosner G. Functional recovery of cortical neurons as related to degree and duration of ischemia. Ann Neurol. 1983;14:294–301. doi: 10.1002/ana.410140307. [DOI] [PubMed] [Google Scholar]
  23. Heiss WD, Graf R, Wienhard K, Lottgen J, Saito R, Fujita T, Rosner G, Wagner R. Dynamic penumbra demonstrated by sequential multitracer PET after middle cerebral artery occlusion in cats. J Cereb Blood Flow. 1994;14:892–902. doi: 10.1038/jcbfm.1994.120. [DOI] [PubMed] [Google Scholar]
  24. Hoppel BE, Weisskoff RM, Thulborn KR, Moore JB, Kwong KK, Rosen BR. Measurement of regional blood oxygenation and cerebral hemodynamics. Magn Reson Med. 1993;30:715–723. doi: 10.1002/mrm.1910300609. [DOI] [PubMed] [Google Scholar]
  25. Jezzard P, Heineman F, Taylor J, DesPres D, Wen H, Balaban RS, Turner R. Comparison of EPI gradient-echo contrast changes in cat brain caused by respiratory challenges with direct simultaneous evaluation of cerebral oxygenation via a cranial window. NMR in Biomed. 1994;7:35–44. doi: 10.1002/nbm.1940070107. [DOI] [PubMed] [Google Scholar]
  26. Jones TH, Morawetz RB, Crowell RM. Thresholds of focal ischemia in awake monkeys. J Neurosurg. 1981;54:773–782. doi: 10.3171/jns.1981.54.6.0773. [DOI] [PubMed] [Google Scholar]
  27. Kennan RP, Zhong J, Gore JC. Intravascular susceptibility contrast mechanism in tissues. Magn Reson Med. 1994;31:9–21. doi: 10.1002/mrm.1910310103. [DOI] [PubMed] [Google Scholar]
  28. Kennan RP, Scanley BE, Gore JC. Physiological basis for BOLD MR signal changes due to hypoxia/hyperoxia: separation of blood volume and magnetic susceptibility effects. Magn Reson Med. 1997;37:353–356. doi: 10.1002/mrm.1910370621. [DOI] [PubMed] [Google Scholar]
  29. Kiselev VG, Posse S. Analytical model of susceptibility-induced MR signal dephasing: effect of diffusion in a microvascular network. Magn Reson Med. 1999;41:499–509. doi: 10.1002/(sici)1522-2594(199903)41:3<499::aid-mrm12>3.0.co;2-o. [DOI] [PubMed] [Google Scholar]
  30. Kuppusamy K, Lin W, Cizek G, Haacke EM. A quantitative assessment of in vivo regional cerebral blood volume using high resolution 3D T1 weighted pre and post injection of a contrast agent. Radiology. 1996;201:106–112. doi: 10.1148/radiology.201.1.8816529. [DOI] [PubMed] [Google Scholar]
  31. Larsen OA, Lassen N. Cerebral hematocrit in normal man. J Appl Physiol. 1964;19:571–574. doi: 10.1152/jappl.1964.19.4.571. [DOI] [PubMed] [Google Scholar]
  32. Levin VA, Ausman JL. Relationship of peripheral venous hematocrit to brain hematocrit. J Appl Physiol. 1969;26:433–437. doi: 10.1152/jappl.1969.26.4.433. [DOI] [PubMed] [Google Scholar]
  33. Leggett RW, Williams LR. Suggested reference values for regional blood volumes in humans. Health Phys. 1991;60:139–154. doi: 10.1097/00004032-199102000-00001. [DOI] [PubMed] [Google Scholar]
  34. Lin W, Paczynski RP, Kuppusamy K, Hsu CY, Haacke EM. Quantitative measurements of regional cerebral blood volume using MRI in rats: effects of arterial carbon dioxide tension and mannitol. Magn Reson Med. 1997;38:420–428. doi: 10.1002/mrm.1910380311. [DOI] [PubMed] [Google Scholar]
  35. Lin W, Paczynski RP, Celik A, Kuppusamy K, Hsu CY, Powers WJ. Experimental hypoxemic hypoxia: changes in R2* of brain parenchyma accurately reflect the combined effects of changes in arterial and cerebral venous oxygen saturation. Magn Reson Med. 1998a;39:474–481. doi: 10.1002/mrm.1910390318. [DOI] [PubMed] [Google Scholar]
  36. Lin W, Paczynski RP, Celik A, Hsu CY, Powers WJ. Effects of acute normovolemic hemodilution on T2*-weighted images of rat brain. Magn Reson Med. 1998b;40:857–864. doi: 10.1002/mrm.1910400611. [DOI] [PubMed] [Google Scholar]
  37. Lin W, Celik A, Paczynski RP, Hsu CY, Powers WJ. Quantitative magnetic resonance imaging in experimental hyper-capnia: improvement in the relation between changes in brain R2* and the oxygen saturation of venous blood after correction for changes in cerebral blood volume. J Cereb Blood Flow Metab. 1999;19:853–862. doi: 10.1097/00004647-199908000-00004. [DOI] [PubMed] [Google Scholar]
  38. Majumdar S. Quantitative study of the susceptibility difference between trabecular bone and bone marrow: computer simulation. Magn Reson Med. 1991;22:101–110. doi: 10.1002/mrm.1910220111. [DOI] [PubMed] [Google Scholar]
  39. Majumdar S, Thomasson D, Shimakawa A, Genant HK. Quantitation of the susceptibility difference between trabecular bone and bone marrow: experimental studies. Magn Reson Med. 1991;22:111–127. doi: 10.1002/mrm.1910220112. [DOI] [PubMed] [Google Scholar]
  40. Muller RN, Gillis P, Moiny F, Roch A. Transverse relaxivity of particulate MRI contrast media: from theories to experiment. Magn Reson Med. 1991;22:178–182. doi: 10.1002/mrm.1910220203. [DOI] [PubMed] [Google Scholar]
  41. Nakane H, Ibayashi S, Fujil K, Sadoshima S, Irie K, Kitazono T, Fujishima M. Cerebral blood flow and metabolism in patients with silent brain infarction: occult misery perfusion in the cerebral cortex. J Neurol Neurosurg Psychiatry. 1998;65:317–321. doi: 10.1136/jnnp.65.3.317. [DOI] [PMC free article] [PubMed] [Google Scholar]
  42. Ogawa S, Lee TM. Magnetic resonance imaging of blood vessels at high field: in vivo and in vitro measurements and image simulation. Magn Reson Med. 1990a;16:9–18. doi: 10.1002/mrm.1910160103. [DOI] [PubMed] [Google Scholar]
  43. Ogawa S, Lee TM, Kay AR, Tank DW. Brain magnetic resonance imaging with contrast dependent on blood oxygenation. Proc Natl Acad Sci U S A. 1990b;87:9698–9872. doi: 10.1073/pnas.87.24.9868. [DOI] [PMC free article] [PubMed] [Google Scholar]
  44. Ogawa S, Lee TM, Nayak AS, Glynn P. Oxygenation-sensitive contrast in magnetic resonance image of rodent brain at high magnetic fields. Magn Reson Med. 1990c;14:68–78. doi: 10.1002/mrm.1910140108. [DOI] [PubMed] [Google Scholar]
  45. Ogawa S, Lee TM, Barrere B. The sensitivity of magnetic resonance image signals of a rat brain to changes in the cerebral venous blood oxygenation. Magn Reson Med. 1993a;29:205–210. doi: 10.1002/mrm.1910290208. [DOI] [PubMed] [Google Scholar]
  46. Ogawa S, Menon RS, Tank DW, Kim SG, Markle H, Ellermann JM, Ugurbil K. Functional brain mapping by blood oxygenation level-dependent contrast magnetic resonance imaging. Biophysical J. 1993b;64:803–812. doi: 10.1016/S0006-3495(93)81441-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  47. Oldendrof WH, Kitano M, Shimizu S, Oldendorf S. Hematocrit of the human cranial blood pool. Circ Res. 1965;17:532–539. doi: 10.1161/01.res.17.6.532. [DOI] [PubMed] [Google Scholar]
  48. Ono Y, Morikawa S, Inubushi T, Shimizu H, Yoshimoto T. T2*-weighted magnetic resonance imaging of cerebrovascular reactivity in rat reversible focal cerebral ischemia. Brain Res. 1997;744:207–215. doi: 10.1016/S0006-8993(96)01079-7. [DOI] [PubMed] [Google Scholar]
  49. Perlmutter JS, Powers WJ, Herscovitch P, Fox PT, Raichle ME. Regional asymmetries of cerebral blood flow, blood volume, and oxygen utilization and extraction in normal volunteers. J Cereb Blood Flow Metab. 1987;7:64–67. doi: 10.1038/jcbfm.1987.9. [DOI] [PubMed] [Google Scholar]
  50. Plum F, Pulsinelli WA. Cerebral metabolism and hypoxic-ischemic brain injury. In: Asbury A, McKhann G, McDonald A, editors. Diseases of the central nervous system. Philadelphia: W. B. Saunders; 1986. pp. 1086–1101. [Google Scholar]
  51. Pollard V, Prough DS, DeMelo AE, Deyo DJ, Uchida T, Stoddart HF. Validation in volunteers of a near infrared spectroscope monitoring brain oxygenation in vivo. Anesth Analg. 1996a;82(2):269–277. doi: 10.1097/00000539-199602000-00010. [DOI] [PubMed] [Google Scholar]
  52. Pollard V, Prough DS, DeMelo AE, Deyo DJ, Uchida T, Widman R. The influence of carbon dioxide and body position on near-infrared spectroscopic assessment of cerebral hemoglobin oxygen saturation. Anesth Analg. 1996b;82(2):278–287. doi: 10.1097/00000539-199602000-00011. [DOI] [PubMed] [Google Scholar]
  53. Press W, Teukolsky SA, Vetterling WT, Flannery BP. Numerical recipes in C, the art of scientific computing. 2. Cambridge, England: Cambridge University Press; 1992. [Google Scholar]
  54. Prielmeier F, Nagatomo Y, Frahm J. Cerebral blood oxygenation in rat brain during hypoxic hypoxia: quantitative MRI of effective transverse relaxation rates. Magn Reson Med. 1994;31:678–681. doi: 10.1002/mrm.1910310615. [DOI] [PubMed] [Google Scholar]
  55. Rostrup E, Larsson HBW, Toft PB, Garde K, Henriksen O. Signal changes in gradient echo images of human brain induced by hypo- and Hyperoxia. NMR in Biomed. 1995;8:41–47. doi: 10.1002/nbm.1940080109. [DOI] [PubMed] [Google Scholar]
  56. Sakai F, Nakazawa K, Tazaki Y, Ishii K, Hino H, Igarashi H, Kanda T. Regional cerebral blood volume and hematocrit measured in normal human volunteers by single photon emission computed tomography. J Cereb Blood Flow Metab. 1985;5:207–213. doi: 10.1038/jcbfm.1985.27. [DOI] [PubMed] [Google Scholar]
  57. Sharan M, Jones MD, Koehler RC, Traystman RJ, Popel AS. A compartmental model for oxygen transport in brain microcirculation. Ann Biomed Eng. 1989;17:13–38. doi: 10.1007/BF02364271. [DOI] [PubMed] [Google Scholar]
  58. Siesjo BK. Brain energy metabolism. New York, NY: John Wiley and Sons; 1978. [Google Scholar]
  59. Song AW, Wong EC, Tan SG, Hyde JS. Diffusion weighted fMRI at 1.5T. Magn Reson Med. 1996;35:155–158. doi: 10.1002/mrm.1910350204. [DOI] [PubMed] [Google Scholar]
  60. Stables LA, Kennan RP, Gore JC. Asymmetric spin-echo imaging of magnetically inhomogeneous system: theory, experiment, and numerical studies. Magn Reson Med. 1998;40:432–442. doi: 10.1002/mrm.1910400314. [DOI] [PubMed] [Google Scholar]
  61. Tomita M, Gotoh F, Aamano T, Tanahashi N, Kobari M, Shinohara T, Mihara B. Transfer function through regional cerebral cortes evaluated by a photoelectric method. Am J Physiol. 1978;235:H56–63. doi: 10.1152/ajpheart.1983.245.3.H385. [DOI] [PubMed] [Google Scholar]
  62. Tomita M. Significance of cerebral blood volume. In: Tomita M, Sawada T, Naritomi H, Heiss WD, editors. Cerebral hyperemia and ischemia: from the standpoint of cerebral blood volume. New York, NY: Elsevier Science; 1988. pp. 3–30. [Google Scholar]
  63. Turner R, Le Bihan D, Moonen CTW, Despres D, Frank J. Echo-planar time course MRI of cat brain oxygenation changes. Magn Reson Med. 1991;22:159–166. doi: 10.1002/mrm.1910220117. [DOI] [PubMed] [Google Scholar]
  64. Van Zijl PC, Eleff SM, Ulatowski JA, Oja JM, Ulug AM, Traystman RJ, Kauppinen RA. Quantitative assessment of blood flow, blood volume and blood oxygenation effects in functional magnetic resonance imaging. Nat Med. 1998;4(2):159–167. doi: 10.1038/nm0298-159. [DOI] [PubMed] [Google Scholar]
  65. Weisskoff RM, Kiihne S. MRI susceptometry: image-based measurement of absolute susceptibility of MR contrast agents and human blood. Magn Reson Med. 1992;24:375–383. doi: 10.1002/mrm.1910240219. [DOI] [PubMed] [Google Scholar]
  66. Wright GA, Hu BS, Macovski A. Estimation oxygen saturation of blood in vivo with MR imaging at 1.5T. J Magn Reson Imaging. 1991;1:275–283. doi: 10.1002/jmri.1880010303. [DOI] [PubMed] [Google Scholar]
  67. Yablonskiy DA, Haacke EM. Theory of NMR signal behavior in magnetically inhomogeneous tissues: the static dephasing regime. Magn Reson Med. 1994;32:749–763. doi: 10.1002/mrm.1910320610. [DOI] [PubMed] [Google Scholar]
  68. Yablonskiy DA, Haacke EM. An MRI method for measuring T2 in the presence of static and RF magnetic field inhomogeneities. Magn Reson Med. 1997;37(6):872–876. doi: 10.1002/mrm.1910370611. [DOI] [PubMed] [Google Scholar]
  69. Yablonskiy DA. Quantitation of intrinsic magnetic susceptibility-related effects in a tissue matrix. Phantom study. Magn Reson Med. 1998;39:417–428. doi: 10.1002/mrm.1910390312. [DOI] [PubMed] [Google Scholar]
  70. Yamauchi H, Fukuyama H, Nagahama Y, Nabatame H, Nakamura K, Yamamoto Y, Yonekura Y, Konishi J, Kumura J. Evidence of misery perfusion and risk for recurrent stroke in major cerebral arterial occlusive diseases from PET. J Neurol Neurosurg Psychiatry. 1996;61:18–25. doi: 10.1136/jnnp.61.1.18. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES