Functional Validation and Comparison Framework for EIT Lung Imaging

Bartłomiej Grychtol 1,2,#, Gunnar Elke 3,#, Patrick Meybohm 4, Norbert Weiler 3, Inéz Frerichs 3, Andy Adler 5,*
Editor: Francesco Pappalardo6
PMCID: PMC4128601  PMID: 25110887

Abstract

Introduction

Electrical impedance tomography (EIT) is an emerging clinical tool for monitoring ventilation distribution in mechanically ventilated patients, for which many image reconstruction algorithms have been suggested. We propose an experimental framework to assess such algorithms with respect to their ability to correctly represent well-defined physiological changes. We defined a set of clinically relevant ventilation conditions and induced them experimentally in 8 pigs by controlling three ventilator settings (tidal volume, positive end-expiratory pressure and the fraction of inspired oxygen). In this way, large and discrete shifts in global and regional lung air content were elicited.

Methods

We use the framework to compare twelve 2D EIT reconstruction algorithms, including backprojection (the original and still most frequently used algorithm), GREIT (a more recent consensus algorithm for lung imaging), truncated singular value decomposition (TSVD), several variants of the one-step Gauss-Newton approach and two iterative algorithms. We consider the effects of using a 3D finite element model, assuming non-uniform background conductivity, noise modeling, reconstructing for electrode movement, total variation (TV) reconstruction, robust error norms, smoothing priors, and using difference vs. normalized difference data.

Results and Conclusions

Our results indicate that, while variation in appearance of images reconstructed from the same data is not negligible, clinically relevant parameters do not vary considerably among the advanced algorithms. Among the analysed algorithms, several advanced algorithms perform well, while some others are significantly worse. Given its vintage and ad-hoc formulation backprojection works surprisingly well, supporting the validity of previous studies in lung EIT.

Introduction

Electrical impedance tomography (EIT) has been proposed as a tool to monitor and guide ventilator therapy in mechanically ventilated patients [1][4]. Medical interest in EIT is driven by the discovery of the injurious effects of artificial ventilation [5] and, thus, the importance of lung protective ventilation strategies [6], [7]. No consensus exists as to how optimal ventilator settings (i.e. least harmful and still securing proper gas exchange) should be chosen in an individual patient [8]. However, there is a clear need for continuous assessment of regional lung ventilation. No medical examination technique is available at the bedside allowing this type of monitoring. With its inherent advantages of the radiation-free measuring principle, good time resolution, portability and ability to assess rapid changes in regional lung air content, EIT promises to fulfil many of the requested criteria for such patient monitoring.

A growing body of research seeks to develop and evaluate EIT-based parameters to guide ventilator therapy [1]. Most of these studies used versions of the original back-projection algorithm of [9] for image reconstruction; for recent examples see e.g. [10][17]. Newer types of image reconstruction procedures allow theoretical advantages [18][21]; however, these algorithms have not been systematically compared against each other, and often have not been tested in vivo. This paucity of systematic validation and comparison of EIT algorithms hampers its entrance into clinical practice.

The ability of EIT to measure regional distribution of ventilation has been validated against several established high-resolution imaging modalities like X-ray computed tomography [22], [23], electron beam computed tomography [24], single photon emission computed tomography [25], [26] and positron emission tomography [27].

Such validation of EIT to anatomical references is important, but exhibits certain drawbacks. Most such studies have not used anatomical models for EIT reconstruction (typically using circles on a Inline graphic pixel grid) making comparison with morphologically accurate images problematic. Also, the inherent low resolution of EIT creates large partial volume effects [28], while the scan rate of EIT is much higher (Inline graphic50 frames/s for some devices) than those of the reference radiological techniques. The most important drawback of such studies, however, is that anatomical validation is primarily sensitive to morphological accuracy, and thus less sensitive to the functional performance of EIT imaging. We believe that a functional imaging technique should also have a functional validation methodology. Thus, we propose a more functional approach for validation and comparison of EIT images obtained with different reconstruction algorithms where the ability to provide clinically relevant information is tested directly. Therefore, the aim of our study was to develop a framework to validate and compare EIT images by testing their ability to correctly reflect clinically significant changes in regional lung ventilation (or lack thereof). The backbone of our framework is an experimental EIT data set acquired in mechanically ventilated pigs. By manipulating three clinically relevant ventilator settings – tidal volume (VT), peek end-expiratory pressure (PEEP) and the OInline graphic fraction in inspired gas (FI o 2) – we were able to achieve different regional ventilation distributions following well defined large and discrete shifts in global and regional lung air content.

We used the proposed framework to compare the performance of a number of popular reconstruction algorithms, including GREIT [29] and backprojection. We sampled the analysed algorithms mainly from the large class of sensitivity-based methods available in the EIDORS suite [30]. In doing so, we studied the effect of individual design choices in assembling such algorithms. Other approaches, such as those based on a Bayesian framework (e.g. [31]), monotonicity (e.g. [32]), level sets (e.g. [33], [34]), factorisation (e.g. [35], or the D-Bar method (e.g. [36]) were not evaluated, as they have seen little evaluation for thoracic EIT data (but see [37], [38]. However, we make the data and software publicly available to allow extending the comparison to a wider range of present and future algorithms.

Methods

Ethics statement

After approval by the Committee for Animal Care of the Christian-Albrechts University Kiel (Permit Number: V742-72241.121-39), the study was conducted in compliance with the recommendations in the Guide for the Care and Use of Laboratory Animals of the National Institutes of Health.

Evaluation framework

Since the potential role of EIT in ventilation therapy is to detect changes in the regional distribution of ventilation, our proposed validation framework is explicitly designed to test its ability to do so. An overview of the framework is presented in Fig. 1.

Figure 1. Overview of the proposed methodology.

Figure 1

Peep – positive end-expiratory pressure; FI o 2 - fraction of oxygen in inspired gas; V – air volume; t – time; VT – tidal volume; CoV - centre of ventilation; FEM – finite element model.

First, we conducted an experiment in healthy pigs ventilated mechanically to produce shifts in ventilation distribution while keeping VT constant. Ventilation with pure O2 is known to induce absorption atelectasis in the dependent, i.e. posterior in supine animals, lung regions [39], [40], while Peep leads to a redistribution of inspired VT with more gas entering the dependent lung regions [41]. These well-known effects of changing Peep and FI o 2 settings allowed the definition of distinct changes in global and regional ventilation, which are summarized in Table 1.

Table 1. Expected effects of changes in ventilation therapy.

Intervention Tidal volume Ventilation distribution
Increase in Peep no change increased ventilation in dependent lung areas
100% FI o 2 no change decreased ventilation in dependent lung areas
Return to baseline no change return to baseline

Peep, positive end-expiratory pressure; FI o 2, fraction of O2 in inspired gas.

Second, EIT-derived measures of VT and distribution of ventilation along the gravitational axis were calculated for each analysed algorithm for each animal and condition. These values were used to test each algorithm by comparing against the expected findings, or physiological references (Table 1). For example, VT is a known volume change in the lung. While the distribution of VT changes with Peep, the volume is fixed. Thus, using this physiological constant, we can test that an EIT algorithm which is equally sensitive to volume across the image will, correctly, identify VT as constant, while an EIT algorithm which has spatially varying volume response, may see varying VT. Similarly, the known movement of VT toward dependent lung with increases in Peep can be used to test the sensitivity of EIT algorithms to movement within the image. The complete list of physiological references is given in Table 4.

Table 4. Performance of different image reconstruction algorithms during mechanical ventilation with constant tidal volume and variable end-expiratory pressure and O2 fraction.

Expected finding Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic
VInline graphic is independent of Peep
VInline graphic Inline graphic  =  VInline graphic Inline graphic 0.224 0.268 0.512 0.838 0.024* 0.008* 0.000* 0.225 0.247 0.815 0.018* 0.044*
VInline graphic Inline graphic  =  VInline graphic Inline graphic 0.022* 0.027* 0.090 0.247 0.008* 0.014* 0.000* 0.026* 0.030* 0.245 0.010* 0.013*
VInline graphic is independent of FI o 2
VInline graphic Inline graphic  =  VInline graphic Inline graphic 0.010* 0.009* 0.000* 0.010* 0.033* 0.025* 0.000* 0.009* 0.035* 0.030* 0.010* 0.004*
VInline graphic Inline graphic  =  VInline graphic Inline graphic 0.002* 0.002* 0.010* 0.005* 0.001* 0.001* 0.000* 0.002* 0.001* 0.042* 0.002* 0.001*
VInline graphic is reproducible
VInline graphic Inline graphic  =  VInline graphic Inline graphic 0.003* 0.003* 0.001* 0.012* 0.002* 0.001* 0.000* 0.002* 0.003* 0.007* 0.001* 0.001*
VInline graphic Inline graphic  =  VInline graphic Inline graphic 0.000* 0.000* 0.001* 0.002* 0.001* 0.001* 0.000* 0.000* 0.001* 0.002* 0.001* 0.000*
CoV is Peep dependent
CoVInline graphic Inline graphic CoVInline graphic 0.088 0.088 0.136 0.114 0.060 0.037* 0.121 0.085 0.075 0.083 0.043* 0.064
CoVInline graphic Inline graphic CoVInline graphic 0.034* 0.033* 0.044* 0.052 0.026* 0.017* 0.053 0.033* 0.035* 0.022* 0.019* 0.023*
CoV is FI o 2 dependent
CoVInline graphic Inline graphic CoVInline graphic 0.012* 0.012* 0.020* 0.020* 0.010* 0.010* 0.035* 0.011* 0.016* 0.009* 0.011* 0.010*
CoVInline graphic Inline graphic CoVInline graphic 0.659 0.679 0.689 0.713 0.586 0.639 0.751 0.662 0.664 0.735 0.604 0.683
CoV is reproducible
CoVInline graphic  =  CoVInline graphic 0.059 0.064 0.238 0.013* 0.061 0.023* 0.009* 0.065 0.107 0.026* 0.032* 0.007*
CoVInline graphic  =  CoVInline graphic 0.009* 0.009* 0.010* 0.000* 0.028* 0.016* 0.001* 0.010* 0.042* 0.006* 0.017* 0.002*

The expected physiological findings regarding the magnitude and distribution of ventilation are used as reference states. Inline graphic-values of the Student's Inline graphic-test are reported. For each test, Inline graphic (the number of animals).

E, Measures by EIT; VInline graphic, tidal volume; Peep (P), positive end-expiratory pressure; ZEEP (Z), zero end-expiratory pressure; FI o 2, fraction of OInline graphic in inspired gas; 21, FI o 2 equal to 21%; 100, FI o 2 equal to 100%; indices 1 and 2 identify the first and the second measurements at identical ZEEP or Peep levels; CoV, center of ventilation; Inline graphic hypothesis confirmed at Inline graphic.

Experimental protocol

The study was performed at the University Medical Center of Schleswig-Holstein, Campus Kiel in Germany. The experiments were carried out on eight anesthetized pigs (body weight (bw): Inline graphic kg, meanInline graphicstandard deviation (SD)) in supine position. The animals were at first sedated with azaperon (8 mg/kg bw). Anaesthesia was achieved by a continuous intravenous infusion of propofol (6 to 12 mg/kg bw per hour) and sufentanile (10 Inline graphicg/kg bw per hour). Subsequently, the animals were intubated and connected to a ventilator (Siemens Servo 900 C ventilator, Siemens-Elema, Solna, Sweden). Haemodynamic as well as ventilatory parameters including heart rate, partial pressure of COInline graphic (Pco 2) in respired gas, arterial O2 saturation (Sa o 2), airway pressures and respiratory system compliance were continuously monitored using the S/5 anaesthesia monitoring system with a gas-density compensated module (M-CAIOV, Datex Ohmeda, Helsinki, Finland). All animals were ventilated in a volume-controlled mode with a constant VT, respiratory rate (20 breaths/min) and inspiration-to-expiration ratio of 1∶2 in order to maintain normocapnia (end-tidal Pco 2 35–45 mmHg). If required, vecuronium bromide (0.1 mg/kg bw) was administered for muscle paralysis to suppress spontaneous breathing activity and thus avoid disturbance of the volume-controlled ventilation measurements.

During the examination period of 60 min, the animals were ventilated with different combinations of end-expiratory pressures and FI o 2 (Table 2). FI o 2 was consecutively changed from 21% to 100% and back to 21% and at each FI o 2, the animals were at first ventilated with zero end-expiratory pressure (ZEEP) and then with Peep of 5 cmH2O. In each condition, one EIT measurement was performed meaning that a total of six measurements was obtained in each animal.

Table 2. Lung function and ventilation parameters Inline graphic SD during study period.
Stage I II III IV V VI
Time period, min 0–15 15–20 20–35 35–40 40–55 55–60
Peep, cmH2O 0 5 0 5 0 5
FI o 2, % 21 21 100 100 21 21
VInline graphic, ml/kg bw 10.1Inline graphic1.6 9.9Inline graphic1.7 10.3Inline graphic1.8 10.0Inline graphic1.7 9.9Inline graphic1.5 9.9Inline graphic1.7
PInline graphic, cmH2O 21Inline graphic6 23Inline graphic5 20Inline graphic4 23Inline graphic5 20Inline graphic5 23Inline graphic5
PInline graphic, cmH2O 16Inline graphic5 20Inline graphic4 15Inline graphic4 19Inline graphic4 16Inline graphic4 20Inline graphic4
Crs, ml/cmH2O 26.4Inline graphic4.2 27.0Inline graphic4.5 28.4Inline graphic4.8 25.7Inline graphic4.0 28.2Inline graphic4.2 27.1Inline graphic4.0
Sa o 2, % 95Inline graphic3 97Inline graphic3 98Inline graphic4 98Inline graphic2 97Inline graphic2 96Inline graphic3

Peep, positive end-expiratory pressure; FI o 2, fraction of O2 in inspired gas; VInline graphic, tidal volume; bw, body weight; Ppeak, peak airway pressure; Pplat, inspiratory plateau airway pressure; Crs, respiratory system compliance; Sa o 2, saturation of O2.

EIT examinations were performed using the Goe-MF II EIT device (CareFusion, Höchberg, Germany). Sixteen self-adhesive electrodes (Blue Sensor BR-50-K, Ambu, Bad Nauheim, Germany) were placed on the chest circumference in one transverse plane lying approximately at the level of the 6th intercostal space. Electrical currents (50 kHz, 5 mArms) were applied through adjacent pairs of electrodes in a rotating mode. After each current injection, the resulting potential differences were measured by the remaining electrode pairs. EIT data were acquired at a frame rate of 13 images/s during 60 s time intervals.

EIT algorithms

We review EIT image reconstruction algorithms and develop a taxonomy of approaches. A representative sample of algorithm choices are selected for evaluation.

EIT image reconstruction

The recovery of information about internal conductivity distribution from surface voltage measurements is a severely ill-posed non-linear inverse problem. Time difference EIT (TD-EIT), which compares two sets of measurements, is much less sensitive to uncertainties in electrode placement, thorax geometry and some hardware errors (such as gain variability between channels), than algorithms that attempt to reconstruct the absolute conductivity distribution [42]. This paper considers only TD-EIT algorithms. Specifically, TD-EIT seeks to reconstruct an image of the change in conductivity (Inline graphic) between EIT measurements Inline graphic and Inline graphic, where Inline graphic represents the current frame of EIT measurements, while Inline graphic is the reference measurement frame, typically calculated by averaging over periods when conductivity distribution in the region of interest is stable. We use the notation (Inline graphic) for time difference data and (Inline graphic) for the conductivity change image (or model) Inline graphic. Depending on the image reconstruction approach, the image may be represented as a pixel grid, or on a finite element model (FEM) discretization. Of the many TD-EIT algorithms proposed (see references of [20], [29]) we consider only those that are applicable to the most common case for which clinical and experimental thoracic EIT measurements have been made. Thus, we assume: 1) TD-EIT measurements; 2) placement of 16 EIT electrodes in a single plane around the thorax; 3) measurement performed using adjacent stimulation and measurement (the Sheffield protocol); and 4) reconstruction of images onto a circular domain, since this is a common capability of all algorithms. We organize the space of EIT algorithms as in Fig. 2; first, algorithms are classified on the features used in the forward problem (and sensitivity matrix) and then on the inverse solution. Two main classes of image reconstruction procedures were chosen for analysis, based on backprojection, and sensitivity matrix (FEM) based solutions; the latter category is subdivided into regularization and optimization based methods. We review the formulation of each approach in the following subsections. Clearly, the possible combinations of algorithm variants are very large, and thus difficult to test. Instead, we define a "baseline" algorithm, Inline graphic which represents the simplest approach, and consider modifications from it. For each variation, we define a default case (which is part of Inline graphic) and a variant case, which is tested in an algorithm (Table 3).

Figure 2. Taxonomy of direct EIT reconstruction algorithms, classified in terms of the selection of forward and inverse model parameters.

Figure 2

Table 3. Algorithm variations considered. For each tested algorithm (horizontal row), the indicated variations from baseline (Inline graphic) are made.
Algorithm Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic Inline graphic
2D FEM (B)
NOSER prior (B)
3D FEM (F1)
Diff. data (F2)
Lung Inline graphic (F3)
Elec. move. (F4)
Noise weight (I1)
Robust errors (I2)
HPF prior (I3)
TSVD (I4)
Total variation (I5)
Sheffield backproj

Backprojection sensitivity model

The backprojection algorithm was developed [43] for the original Sheffield EIT system, and has seen numerous variants and improvements. Using an analogy to X-ray CT backprojection, normalized measurements are "backprojected" onto the equipotential region in the image, and the image is subsequently filtered. Improvements to account for the diffuse nature of electrical current propagation in the body were subsequently made; a good mathematical characterization is given by Santosa and Vogelius [44]. While many variants of Sheffield backprojection (Inline graphic) exist, clinical and experimental EIT has largely used only the one which was distributed in the Sheffield Mark I EIT system [45]. In discussions with the algorithm authors, it appears that the exact formulation of this algorithm has been lost. In order to accurately represent this algorithm, we obtained permission to reverse engineer Inline graphic from a Sheffield Mk I device, as described in [29].

Numerical sensitivity model

The forward model, Inline graphic, in EIT simulates the measured data vector Inline graphic for a given vector of model parameters Inline graphic, as Inline graphic. The data vector, Inline graphic, represents each measured voltage for each applied current pattern in a data frame; for the Sheffield stimulation protocol on 16 electrodes, this yields Inline graphic measurements per frame, of which Inline graphic are independent, due to reciprocity [46]. Inline graphic is typically based on a finite element model (FEM) and each parameter of the conductivity distribution is usually modelled in Inline graphic as the piecewise constant conductivity on a tetrahedral element. The forward model is used to calculate a Jacobian, Inline graphic, or sensitivity matrix, which describes the sensitivity of each data element to each model parameter.

graphic file with name pone.0103045.e154.jpg (1)

where the FEM estimate depends on the background conductivity, Inline graphic, around which the conductivity changes Inline graphic occur. The following section discusses various approaches to numerical (forward) models (Inline graphic), considering the baseline approach and its variants.

Model dimension: 2D FEM / 3D FEM

The choice of model dimension affects the relative magnitude of measured voltage in regions further from the current stimulations. Real voltage measurements occur in three dimensions, and are thus mismatched to 2D models. Early EIT algorithms used 2D FEM to reduce computational time; however, many recent algorithms continue to use 2D. One consequence is errors in the reconstructed position of objects because sensitivity falls off faster with distance in 3D than 2D.

The baseline model uses a circular 2D FEM (Fig. 3A), and the variant uses a cylindrical 3D FEM of height equal to the medium diameter, and the electrode plane in the centre (Fig. 3B). FEM models are constructed using Netgen [47] by specifying the region geometry and circular electrode contact regions.

Figure 3. Finite element models used.

Figure 3

Electrode nodes are indicated in green. A: 2D circular uniform FEM (Inline graphic) B: 3D cylindrical uniform FEM (Inline graphic) C: 3D cylindrical FEM with lung regions (Inline graphic) D: 2D circular uniform FEM with electrode movement (Inline graphic) (with arrows showing representative electrode movement).

Difference data: normalized diff. data / diff. data

TD-EIT defines difference data Inline graphic in terms of the current, Inline graphic, and reference, Inline graphic measurements. The majority of experimental and clinical algorithms (including Inline graphic) have used normalized difference imaging, where data Inline graphic are defined as Inline graphic. If difference imaging (without normalization) is used, then data Inline graphic are defined as Inline graphic. By normalizing, small measurement values are scaled up, and have a larger impact on the reconstructed images.

The baseline model uses normalized difference data, while the variant model uses difference data (without normalization).

Conductivity background: homogeneous Inline graphic / lung Inline graphic

TD-EIT assumes conductivity changes occur with respect to a reference conductivity, Inline graphic, which defines the sensitivity matrix, Inline graphic. The most common choice for EIT algorithms is a homogeneous value for Inline graphic, which is clearly a poor model for the thorax, in which the lungs are far less conductive than other tissues. The use of a homogeneous thorax conductivity distorts the position and amplitude of reconstructed contrasts [48].

The baseline model uses a homogeneous Inline graphic (Fig. 3B), while the variant model defines lung tissue with a relative conductivity of Inline graphic of that of other tissues [42]. The lung region is defined from CT images of pigs of similar size to those used in our experiments (Fig. 3C). We evaluate the effect of lung Inline graphic in a 3D FEM, since a full dimensional model is required to get more accurate current flow in the thorax.

Model electrode movement: static model / elec. move

One key difficulty with EIT measurements is due to the position uncertainty of the electrodes, especially for thoracic measurements, in which the body surface moves during breathing and posture change. Several approaches have been proposed to reduce the effect of electrode movements on the reconstructed EIT conductivity changes. We select the algorithm of [21], [49] based on using an augmented Inline graphic, sensitive to impedance changes, electrode movement, and shape deformation in the model. This approach allows estimation of the shape changes, which we do not use here; instead, by allowing calculation of movement, certain artefacts in the conductivity image may be reduced. This approach can be shown to be equivalent to formulating electrode movement as a structured noise term which is added to the measurement noise variance, Inline graphic. Following [49],

graphic file with name pone.0103045.e180.jpg (2)

where Inline graphic and Inline graphic represent the original and updated estimates of data noise variance, respectively. Inline graphic is the movement sensitivity matrix (change in data for each electrode coordinate change) and Inline graphic is a prior estimate of the covariance of electrode movement.

The baseline model assumes electrode position to be fixed, while the variant model updates noise estimates using (2).

Regularized EIT reconstruction

The class of linear regularized EIT reconstruction algorithms seeks a solution Inline graphic which minimizes the expression

graphic file with name pone.0103045.e186.jpg (3)

The first term, Inline graphic is the data error term, the mismatch between the model and the measured data weighted by a matrix Inline graphic, which models the measurement error on each data channel. The data error term has a norm Inline graphic. If a Gaussian model of data errors is used (Inline graphic), the data weighting matrix, Inline graphic, is related to the channel noise covariance, Inline graphic, by Inline graphic. This formulation is also used to incorporate the structured noise due to electrode movement, Inline graphic in (2). The second term, Inline graphic is a regularization term designed to address the ill-conditioning intrinsic to EIT; it serves as a penalty function which "discourages" unlikely (but otherwise feasible) solutions. It calculates the mismatch between the Inline graphic and prior constraints on "likely" models. Inline graphic is the prior mean and has always been set to zero for TD-EIT, since conductivity is as likely to increase as decrease. Inline graphic implements a penalty for large changes or non-smooth conductivities. The model error term has a norm Inline graphic. Inline graphic controls the relative strength of data and model error terms, and is typically called the regularization hyperparameter. As Inline graphic increases, the solution matches more closely to the model, and becomes increasingly smooth. In order to adequately compare different algorithms, functionally equivalent values of Inline graphic are chosen. Many strategies to choose an appropriate value have been presented (see references of [50]). We choose the "Noise Figure" approach [51], in which Inline graphic is chosen such that the ratio of values of signal to noise ratio (SNR) between Inline graphic and Inline graphic is required to be constant across algorithms. Since we have no control of the parameters of the SBP algorithm, all other algorithms are normalized to its value.

The most commonly studied case is for quadratic norms Inline graphic, which has the advantage of modelling Gaussian error, and yielding a closed form, linear solution:

graphic file with name pone.0103045.e207.jpg (4)

EIT algorithms of this form are known as one-step Gauss Newton solvers (GN) [52]. One important advantage is that image reconstruction can be expressed as matrix multiplication by Inline graphic; since Inline graphic may be pre-calculated, reconstruction is rapid and may be implemented in real-time. If either Inline graphic or Inline graphic are not quadratic, image reconstruction must be formulated iteratively.

In general, regularized EIT image reconstruction techniques are popular due to their flexibility to represent arbitrary body geometry, measurement configurations, and mathematical models of the conductivity distribution. Many papers have used the basic regularized formulation, and have explored the choices of parameter values. In this paper, the baseline reconstruction algorithm, Inline graphic, follows an early widely used regularized approach, NOSER [52], which makes the following parameter choices: it is quadratic, Inline graphic, data noise is modelled to be identical on each channel Inline graphic, and the regularization matrix, Inline graphic, is diagonal such that, Inline graphic, which implies an assumption that inter-element correlations are zero.

The following section discusses various approaches to regularized image reconstruction (I), considering the baseline approach and its variants.

Data noise model (Inline graphic): uniform / weighted

The most common approach is to "trust" all EIT data equally, and thus to model each channel with equal, uniform noise (Inline graphic). It is not necessary to set the actual noise intensity, since the data to model noise trade-off is set by the hyperparameter, Inline graphic. In practice, however, EIT noise can vary considerably between data channels, most likely due to variability in electrode contact quality. Such variations in data quality can be addressed by decreasing Inline graphic on channels with larger noise. One practical approach to setting an appropriate Inline graphic is to calculate the reciprocity error [53] (the difference between data values that should be equal by reciprocity [46]) and to use this value to scale Inline graphic. Clearly, modelling data noise would be expected to be more useful for lower quality data; we consider the EIT data in this study to be representative of relatively good measurements.

The baseline model sets Inline graphic, and the variant model sets Inline graphic using the reciprocity error approach [53], setting the parameter Inline graphic. This parameter choice implied data were on average weighted at Inline graphic of their weighting in the baseline model.

Data norm (Inline graphic): Gaussian, Inline graphic /robust, Inline graphic

As mentioned, a Gaussian (Inline graphic) model and data norms allow calculation of a linear matrix inverse. One disadvantage of this formulation is that it emphasizes larger errors (since the effect of an error is squared). One approach to address such errors is the use of a Inline graphic data norm. While such as approach has been relatively recently proposed in medical EIT [54], it has been widely used in the geophysical EIT literature, and is referred to as "robust error norms".

The baseline model uses a Gaussian data norm Inline graphic, while the variant uses Inline graphic using the formulation of [54].

Prior model (Inline graphic): diagonal / smoothing

The regularization matrix, Inline graphic, is chosen to penalize unlikely solutions. The NOSER [52] regularization matrix is diagonal, and thus penalizes large image amplitudes. Several studies have suggested that a better prior model is a smooth, rather than simply low amplitude distribution [51], [55], [56]. To achieve such smoothing, Inline graphic is designed to penalize non-smooth Inline graphic, imposing a filter across nearby model elements. One key difference between an amplitude and a smoothing filter is its behaviour as a function of model element density. As density increases, amplitude penalties tend to show a more "speckled" pattern of noise, since each image element is independent in this case.

The baseline model sets Inline graphic to penalize image amplitude [52], and the variant model uses the spatial high pass filter [51], in which Inline graphic is calculated to apply Gaussian filter where spatial wavelengths below 10% of the medium diameter are penalized.

Prior model (Inline graphic): diagonal / truncated SVD

The function of regularization may be understood via the singular value decomposition (SVD) of the Jacobian, Inline graphic, as the pseudo-inverse of Inline graphic, with the elimination of small singular values. Here, Inline graphic and Inline graphic are orthonormal matrices and Inline graphic is diagonal. Thus

graphic file with name pone.0103045.e246.jpg (5)

where

graphic file with name pone.0103045.e247.jpg (6)

The threshold Inline graphic functions like a regularization parameter. We choose its value to enforce the noise figure constraint similar to the other algorithms. Inline graphic forms a linear inverse and may be represented in the form of (4) with the values Inline graphic (uniform independent noise), a very large value of Inline graphic, and Inline graphic equal to Inline graphic with its first Inline graphic rows set to zero [57].

The baseline model sets Inline graphic to penalize image amplitude [52], and the variant model uses the truncated SVD to set Inline graphic as indicated above.

Model norm (Inline graphic): Gaussian / Total Variation

One disadvantage of a Gaussian (Inline graphic) model norm is that it blurs edges in images, which is physiologically unrealistic, since organ boundaries are anatomically well-defined. This blurring can be addressed using the Total Variation (TV) formulation [58], with Inline graphic. While TV produces apparently sharper images, there is some debate as to whether such images can give the appearance of features where none exist, especially in the context of EIT data errors.

The baseline model uses a Gaussian model Inline graphic, while the variant implements TV Inline graphic using the formulation of [54], [58].

Optimization based EIT reconstruction

One recent EIT reconstruction algorithm formulation is GREIT [29]. This algorithm proposes an approach to image reconstruction designed to implement a set of requirements (or figures of merit) on which a group of experts reached consensus. The parameters of a linear reconstruction matrix, Inline graphic, are chosen to best satisfy the listed requirements. This approach is similar in many ways to regularized image reconstruction; however it is the reconstruction matrix, Inline graphic, rather than the image, Inline graphic, which is the target in the objective function formulation. Thus, we classify GREIT as an "optimization-based EIT reconstruction". The GREIT approach provides a number of different parameters which can be selected for an image; however, in this paper we choose the circular model reconstruction matrix evaluated in [29].

EIT algorithms evaluated

In order to test the large variety of possible combinations of EIT reconstruction algorithms, we evaluate a baseline algorithm, Inline graphic, and modifications of it in a single variation. The list of considered algorithms is summarized in Table 3. We thus consider Inline graphic and nine variations, as well as SBP and GREIT, for a total of twelve algorithms.

Data analysis

Image reconstruction

Each EIT algorithm identified in Table 3 was implemented using EIDORS version 3.6 [30] and used to reconstruct real conductivity change in each EIT frame in each series of EIT scans, representing 60 seconds of recording for each animal and condition. The reference data set, Inline graphic was chosen to be an average of data scans at end-expiration. Images reconstructed onto an FEM were interpolated onto a Inline graphic pixel grid (Inline graphic and Inline graphic create a Inline graphic pixel image directly).

Modern algorithms can be tuned to control the trade-off between resolution and noise performance, generally via a "hyperparameter". To fairly compare algorithms, it is essential that the individual hyperparameters are set to equivalent values using an objective measure. Several such measures have been proposed [50]. In this study we chose to tune all algorithms such that the noise figure (NF) measure, first proposed for EIT by [51], for changes half the model radius away from the centre was Inline graphic. This particular value was chosen as it matched the performance of the backprojection algorithm, which does not have a tunable hyperparameter.

Ventilation pattern evaluation

From each sequence of EIT data, an average end-inspiratory and end-expiratory data scan was calculated and used for subsequent analysis. The average data scan was calculated by, first, identifying candidate end-inspiratory (end-expiratory) time points from the maximum (minimum) values in the average time course of EIT data. These candidate points were reviewed by a human operator in order to identify and, if necessary, reject physiologically infeasible time points. An example of identified (and rejected) events is shown in Fig. 4. The remaining points were averaged to create a single data scan, one for end-inspiration and one for end-expiration. Using these data, functional EIT scans were generated showing the distribution of local VT by plotting the average tidal differences between the end-inspiratory and end-expiratory values in corresponding pixel locations.

Lung areas (AL) in images were identified as regions with an EIT ventilation signal above Inline graphic of the maximum found in the image. This approach was recommended e.g. in [59]. These were aggregated over all 6 recordings to provide a single lung region of interest (ROI) for each animal and algorithm. The size and shape of the lung ROI varies for different algorithms, since some tend to "squeeze" image contrasts toward the image centre. For each measurement, we calculate VT by summing all pixel values within the lung ROI; we characterized the distribution of ventilation along the dorsoventral axis within the lung ROI as centre of ventilation (CoV), calculated analogously to centre of gravity, such that CoV Inline graphic in the image centre and positive in non-dependent (ventral) lung regions; Inline graphic. The CoV values depend on the position of the lung ROI within the EIT image. A decrease in CoV reflects a shift of ventilation distribution towards the dependent lung.

Statistical analysis

Statistical tests are performed as follows: to test for inequality (Inline graphic), we use the one-tailed Inline graphic-test to test rejection of the null hypothesis, Inline graphic; to test for equality (Inline graphic), we use the two-tailed Inline graphic-test to test rejection of the null hypothesis, Inline graphic for VT, and Inline graphic for CoV. Thus we consider EIT-derived VT values equal when their means differ by less than 10%. We treat mean CoV values equal when they differ by less than 0.03, which corresponds to half a pixel height in a Inline graphic image. For each test a Inline graphic value is calculated, and Inline graphic is considered significant.

The Inline graphic-test requires an assumption of a normal distribution of data. We investigated this assumption using two approaches. First, the distribution of each set of calculated parameters was tested against the normal distribution using the Kolmogorov-Smirnov normality test, using Inline graphic. Next, data were analysed using the same protocol, but with the Wilcoxon signed rank test instead of the Inline graphic-test.

Results

Images were reconstructed for all animals for each case and algorithm. Fig. 5 shows representative images of tidal change VT at Peep and ZEEP in three animals under 21% FI o 2.

Sample image and identified end-inspiratory (blue) and end-expiratory (red) events.

Sample image and identified end-inspiratory (blue) and end-expiratory (red) events.

Left: Image of average tidal volume change Right: Average EIT signal (arbitrary units) vs time (s) showing identified events.

Figure 5. EIT images for all algorithms from three representative animals.

Figure 5

For each animal, images of VT at ZEEP and Peep are shown individually normalized to the maximum amplitude in each image (Blue: decrease in conductivity, Red: increase in conductivity). Each column shows images for a different algorithm. Peep, positive end-expiratory pressure; ZEEP, zero end-expiratory pressure; VT tidal volume.

Qualitative assessment

All algorithms successfully reconstructed the lungs as a central object with conductivity changing with ventilation (shown in blue as decrease in conductivity between expiration and inspiration). Images for different algorithms vary qualitatively in the shape of the lung region and the type and amount of artefacts in the images (Fig. 5). We qualitatively identify five types of artefacts: 1) boundary artefacts, especially near the electrodes, represented as image contrasts at the image edge; 2) speckle, showing high spatial frequency contrasts especially visible at the lung boundary (especially Inline graphic, Inline graphic); 3) ringing, showing inverse contrast near larger image contrasts (ie. red regions adjacent to blue lungs) (Inline graphic, Inline graphic, Inline graphic); 4) spatial distortions, which disturb the shape, visible especially in the shape of the lung regions (Inline graphic); and 5) streak artefacts, showing contrasting lines radially projecting toward the electrodes (Inline graphic, Inline graphic, Inline graphic). The presence of such artefacts disturbs the analysis of the images in several ways, primarily by introducing noise in the images, and by disturbing the analysis of the regional distribution of ventilation.

Images reconstructed with the baseline algorithm Inline graphic show the lungs as a single object. Although the images are generally smooth, boundaries of individual triangular finite elements on which the images are reconstructed prior to rasterization are clearly visible giving the appearance of implausible structures. This is true of all algorithms beside Inline graphic and Inline graphic which reconstruct directly onto a pixel grid. Artefacts are clearly visible on the periphery near the electrodes (mostly in red).

Modifications of the numerical sensitivity model (algorithms Inline graphic to Inline graphic) generally have an obvious and pronounced impact on the reconstructed EIT images. Only images obtained with Inline graphic, which differs from Inline graphic only in that a 3D forward model is used, do not show appreciable difference with respect to Inline graphic. Reconstructions of difference data with Inline graphic, as opposed to normalized difference data as in Inline graphic, show much more artefacts than the other algorithms. However, although the lung shape appears distorted, it shows clear separation between the two lungs. The replacement of homogeneous Inline graphic with one that includes lung tissue contrast in Inline graphic produces images with clearly separated lungs but slightly more artefacts than Inline graphic. Reconstructions with electrode movement Inline graphic show virtually no boundary artefacts. However, the lungs appear slightly smaller and more round in comparison to Inline graphic.

Modifications of the parameters of regularized reconstruction also dramatically affect the appearance of EIT images. Noise weighting based on reciprocity in Inline graphic produced images with less artefacts than Inline graphic, but more than Inline graphic. However, tidal changes seem to be pushed together and toward the centre resulting in comparatively smaller lungs. Inline graphic also offers limited reduction in boundary artefacts. In contrast, images reconstructed with Inline graphic show increased artefacts compared with Inline graphic; the lungs have a speckled appearance. As expected, Inline graphic produced images with sharp boundaries. However, the resulting lung shape does not look plausible. Perhaps surprisingly, the use of the robust data norm with Inline graphic in Inline graphic seems to offer no advantage in terms of boundary artefacts over Inline graphic in our data set, but it reconstructs the lungs as bigger and, in some cases, better separated. Images produced with Inline graphic and Inline graphic stand out for their smoothness and lack of boundary artefacts. Inline graphic is most comparable with Inline graphic in terms of the shape of the lung region, although the lungs appear slightly smaller and less separable. Images obtained with Inline graphic are similar to those obtained with Inline graphic, but the lungs appear even smaller while their shape is less smooth. It also seems to be distorted – tidal changes are pushed toward the centre and streak artefacts pointing outwards seem to show implausible structures.

Functional assessment

The changes in ventilation pattern apparent in the reconstructed images were tested against physiological references, listed in Table 4, where Inline graphic values of the corresponding statistical tests are also reported. The Inline graphic-values are based on the Inline graphic-test, which assumes parameter values are normally distributed. We justify this assumption as follows. First, all parameter values pass the Kolmogorov-Smirnov normality test (at Inline graphic). Second, we recalculated all Inline graphic values using the Wilcoxon signed rank test, which showed substantially similar results; compared to the values in Table 4, the ordering of algorithms by Inline graphic-value was close to identical for all parameter values.

The independence of VT from airway pressure was confirmed for most algorithms at FI o 2 of 100% (Inline graphic in Table 4). Notable exceptions are: Inline graphic, Inline graphic and, surprisingly, Inline graphic. However, at FI o 2 of 21% VT was less stable and only five algorithms passed the independence test (Inline graphic, Inline graphic, Inline graphic, Inline graphic and Inline graphic). Nonetheless, VT was independent of FI o 2 for all algorithms. Repeated measurements at the same settings showed very little variation in VT, which was reproducible for all algorithms.

At 21% FI o 2, no significant decrease in CoV associated with the introduction of Peep was detected, save by the Inline graphic and Inline graphic algorithms. In contrast, at 100% FI o 2, the distribution of ventilation was significantly more skewed towards ventral lung regions at ZEEP as compared to Peep. This is reflected in the decrease in CoV associated with the introduction of Peep, which was detected by all algorithms but Inline graphic and Inline graphic where it narrowly escaped significance. Accordingly, we found that the increase in FI o 2 from 21% to 100% produced for all algorithms a significant increase in CoV, i.e. a shift of ventilation distribution towards ventral regions, at ZEEP but not at Peep. Repeated EIT recordings with the same ventilator settings produced equivalent values of CoV for all algorithms at Peep. At ZEEP, CoV was reproducible at Inline graphic only for six algorithms (Inline graphic, Inline graphic, Inline graphic, Inline graphic, Inline graphic and Inline graphic), although only for Inline graphic and Inline graphic did Inline graphic exceed Inline graphic.

Discussion

EIT shows significant promise as a technique to monitor regional changes in lung ventilation, and to improve patient outcomes by informing ventilator therapy [4]. Since EIT is a challenging mathematical problem, many advanced reconstruction procedures have been developed, each offering potentially useful advances in terms of image quality or robustness against data noise. However, such advanced reconstruction schemes have seen little application to clinical or experimental EIT research, with the exception of GREIT [29] algorithm, which has seen some recent use.

In this paper, we propose a framework for functional validation and comparison of lung EIT images. This framework is then used to evaluate whether some advanced algorithms for EIT image reconstruction improve its ability to resolve the distribution of lung ventilation. As proposed, our validation focuses on EIT's ability to represent two key parameters, VT and CoV, in accordance with known physiology. These are currently the most important for the proposed clinical applications of EIT; they have been applied in recent work by e.g. [60][62]. Additionally, our framework permits straightforward addition of new parameters by an extension of the evaluation software. Rather than using simulations or a saline phantom (e.g. [63]), our approach is more clinically orientated in that we use measures based on an in vivo experimental model. The key innovation is that we directly test reconstructed images against physiological "knowns". In particular, we test for the presence of significant changes in the reconstructed images as a result of experimental interventions with known physiological consequences. In the current study we do not assess the magnitude of these changes, although the presented framework could be extended in the future to include such assessment, once a number of technical difficulties are overcome. These include the use of normalized difference data, which hinders quantitative comparison of image pixel values to expected conductivity changes in experimental data; and the use of incorrect model geometry (discussed further below) resulting in spatially distorted images.

We consider algorithms based on a sensitivity model of EIT, which can be expressed in terms of a direct minimization of an error expression (eqn. 3). This includes many possible "ingredients": the dimension and background conductivity of the FEM; electrode movement modelling and reconstruction; time-difference vs. normalized time-difference measurement; data noise model; data and model norms; assumptions about the conductivity distribution imposed through the prior model; and method of calculating the pseudo-inverse of the sensitivity matrix. We do not consider all available algorithms; direct inversion methods (such as D-bar), those based on a Bayesian framework or non-linear sensitivity models (e.g. level set, monotonicity, or factorization) were not evaluated. We have designed our evaluation software to make it straightforward to add such comparisons as implementations of these algorithms become available. Even with the nine factors considered, it would not be possible to evaluate the full set of possible algorithms; instead we define a baseline algorithm and variants of it, yielding twelve algorithms, which were evaluated based on two approaches. First, representative images of ventilation were studied visually to assess the relative amount and character of artefacts as well as the shape of the region showing tidal changes. Second, measures of tidal changes and their distribution in the reconstructed images were tested against known physiological behaviour in experimental data.

To fairly compare algorithms we chose their hyperparameters, which control the trade-off between resolution and noise performance, such that the value of the noise figure (NF) metric [51] for a specific change in of Inline graphic. Still, because the NF of algorithms varies across the image, sometimes greatly, this approach to comparison is not completely fair for conductivity changes elsewhere in the domain. Moreover, this high value of NF means that for some of the tested algorithms the amount of regularization was insufficient to fully appreciate the nature of the prior.

Based on the results in Table 4, algorithms which exhibit a better Inline graphic-value are interpreted as offering an improved correspondence to known physiological "ground-truths", and are thus better. We do not select an overall "best", as that would amount to ranking the relative importance of the different tests, which is problematic. For example, the ability to accurately represent VT may be more important relative to CoV in some clinical applications, but not others. Overall, we note that there is a general trend in which a ranking of algorithms by Inline graphic-value performance in various tests are substantially consistent. Based on the images in Fig. 5, we make the following observations: 1) Artefacts on the image boundary are reduced in Inline graphic, Inline graphic and Inline graphic, since these algorithms explicitly compensate for noise close to the electrodes; 2) The lung regions are better separated in Inline graphic which exhibits increased sensitivity in the lung regions "pulling" them apart; 3) Reconstructions using difference – rather than normalized difference – data exhibit strong artefacts, likely due to the variation in gain on individual channels in the Goe-MF II EIT system; 4) Numerous "streak" artefacts are generated by Inline graphic, which are a result of wrong assumptions in the backprojection sensitivity model; 5) Inline graphic, the only algorithm whose regularization is represented in terms of sensitivity matrix singular values – and thus not on the image space – produces strong "speckle" artefacts; 6) Algorithms using a spatial filter (Inline graphic and Inline graphic) produce spatially smoother images; 7) On the other hand, Inline graphic – designed to preserve boundaries – creates an image with implausible shapes; 8) The use of a 3D forward model, Inline graphic vs. Inline graphic, shows little appreciable difference despite the fact that EIT data from a 3D domain is incompatible with a 2D model [64], since the use of a circular shape is incorrect in both cases.

Overall, this means the variation in appearance of images reconstructed from the same data with different algorithms is not negligible. This poses a challenge to efforts to analyse the images at the level of individual pixels to derive diagnostic information. At the same time, the strength of our functional validation is evident in that we were able to detect clinically relevant physiological changes in spite of the varied appearance of the images. This is in contrast to methods based on comparison with images more faithful to anatomy that are primarily sensitive to morphological distortions.

Our results address two important questions:

  1. Do advanced image reconstruction algorithms provide advantages over the original Sheffield backprojection when applied to in vivo data? Yes, some advanced algorithms offer a small, but significant advantage. Of the linear, one-step (i.e. fast) algorithms considered, Inline graphic is the current best choice. Several other algorithms also offer advantages, such as those that consider electrode movement and the background lung conductivity; however, these "ingredients" can – and should – be included in an improved GREIT-like algorithm. Considering both the images and the significance values in our tests, the best approach (but at the cost of significantly higher computational complexity) is Inline graphic, which uses robust error norms. We also note that several algorithms do not perform well.

  2. Do clinically relevant EIT measures characterizing lung ventilation depend on the algorithm used? For ventilation therapy, EITs clinical relevance is driven by its ability to determine information on the changes in volume and its distribution, which we seek to capture in our evaluation framework. A first reading of Table 4 would indicate that the choice of algorithm affects the significance. Many algorithms have significantly worse Inline graphic-values than the original Inline graphic, with which most clinical and experimental results have been analysed (ie. Inline graphic, Inline graphic, Inline graphic, Inline graphic, Inline graphic, Inline graphic, Inline graphic). The remaining algorithms (Inline graphic, Inline graphic, Inline graphic, Inline graphic) show similar, or in some cases improved, Inline graphic-values to Inline graphic. This group of algorithms may be characterized as "advanced in that they model additional image and sensor characteristics above that available in a classic GN regularized formulation. We note that the RSBP algorithm works surprisingly well given its vintage and ad-hoc formulation. Overall, our analysis does not suggest reconsideration of the validity of previous analyses of EIT data.

Our analysis is limited in that we only considered 2D electrode placement and adjacent injection and measurement patterns. These have prevented us from evaluating 3D reconstructions and limited the sensitivity in the centre of the image [49]. This, however, remains the de-facto standard for in vivo studies. We also purposefully used a circular (cylindrical in 3D) forward model for all algorithms, as this is what Inline graphic assumes, although using accurate body shape (e.g. derived from CT data) is known to decrease artefacts and improve positional accuracy [28], [65]. The effects of using a circular model do not affect our analysis of tidal volume changes, since VT is a scalar not affected by the distribution of pixel values within the ROI used for its calculation. Furthermore, because CoV is a gross measure also defined on a ROI, it is unlikely that using an anatomically accurate model would substantially improve the performance of the evaluated algorithms in detecting shifts in CoV. However, the use of circular models is likely to have adversely affected our analysis of background modelling in Inline graphic, since realistic organ shape could not be incorporated into the forward model. Last but not least, our analysis considered the effect of each "ingredient" individually. Clearly, a combination of the more promising ones is likely to yield a better reconstruction algorithm than any of those tested here. However, naïve combinations are unlikely to produce good results. Instead, careful calibration and testing of many combinations will be needed, for which, we hope, the proposed evaluation framework will prove a useful tool.

Conclusions

We present a methodology to experimentally evaluate lung EIT algorithms, and use it to analyse and compare recent work. The methodology is based on testing the algorithms ability to correctly detect physiological changes (or lack thereof) in a bespoke experimental data set. The basic idea behind this methodology allows it to be easily expanded to incorporate additional tests in the future. Several promising algorithm ingredients are identified; and we recommend research to evaluate approaches combining these factors, as well as other, non-linear algorithms. In order to support such investigation, we release the data and software for this study under a free license, so that it can be used and modified by others (http://sf.net/p/eidors3d/code/HEAD/tree/trunk/pubs/papers/compare-algs-2014).

Funding Statement

BG was supported by the Alexander von Humboldt Fellowship. Experiments were supported with departamental funding. No external funding was received for this study. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1. Adler A, Amato MBP, Arnold JH, Bayford R, Bodenstein M, et al. (2012) Whither lung EIT: Where are we, where do we want to go and what do we need to get there? Physiol Meas 33: 679–694. [DOI] [PubMed] [Google Scholar]
  • 2. Frerichs I, Dargaville PA, Dudykevych T, Rimensberger PC (2003) Electrical impedance tomography: a method for monitoring regional lung aeration and tidal volume distribution? Intensive Care Med 29: 2312–2316. [DOI] [PubMed] [Google Scholar]
  • 3. Moerer O, Hahn G, Quintel M (2011) Lung impedance measurements to monitor alveolar ventilation. Curr Opin Crit Care 17: 260–7. [DOI] [PubMed] [Google Scholar]
  • 4. Wolf GK, Gómez-Laberge C, Rettig JS, Vargas SO, Smallwood CD, et al. (2013) Mechanical ventilation guided by electrical impedance tomography in experimental acute lung injury. Crit Care Med 41: 1296–304. [DOI] [PubMed] [Google Scholar]
  • 5. Ricard JD, Dreyfuss D, Saumon G (2002) Ventilator-induced lung injury. Curr Opin Crit Care 8: 12–20. [DOI] [PubMed] [Google Scholar]
  • 6. Amato MB, Barbas CS, Medeiros DM, Magaldi RB, Schettino GP, et al. (1998) Effect of a protective-ventilation strategy on mortality in the acute respiratory distress syndrome. N Engl J Med 338: 347–354. [DOI] [PubMed] [Google Scholar]
  • 7. The Acute Respiratory Distress Syndrome Network (2000) Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 342: 1301–8. [DOI] [PubMed] [Google Scholar]
  • 8. Marini JJ, Gattinoni L (2004) Ventilatory management of acute respiratory distress syndrome: a consensus of two. Crit Care Med 32: 250–255. [DOI] [PubMed] [Google Scholar]
  • 9. Barber DC, Brown BH (1984) Applied potential tomography. J Phys E: Sci Instrum 17: 723–733. [Google Scholar]
  • 10. Meier T, Luepschen H, Karsten J, Leibecke T, Grossherr M, et al. (2008) Assessment of regional lung recruitment and derecruitment during a PEEP trial based on electrical impedance tomography. Intensive Care Med 34: 543–550. [DOI] [PubMed] [Google Scholar]
  • 11. Miedema M, van der Burg PS, Beuger S, de Jongh FH, Frerichs I, et al. (2013) Effect of nasal continuous and biphasic positive airway pressure on lung volume in preterm infants. J Pediatr 162: 691–697. [DOI] [PubMed] [Google Scholar]
  • 12. Pham TMT, Yuill M, Dakin C, Schibler A (2011) Regional ventilation distribution in the first 6 months of life. Eur Respir J 37: 919–924. [DOI] [PubMed] [Google Scholar]
  • 13. Pulletz S, Kott M, Elke G, Schdler D, Vogt B, et al. (2012) Dynamics of regional lung aeration determined by electrical impedance tomography in patients with acute respiratory distress syndrome. Multidiscip Respir Med 7: 44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Riedel T, Kyburz M, Latzin P, Thamrin C, Frey U (2009) Regional and overall ventilation inhomogeneities in preterm and term-born infants. Intensive Care Med 35: 144–151. [DOI] [PubMed] [Google Scholar]
  • 15. Vogt B, Pulletz S, Elke G, Zhao Z, Zabel P, et al. (2012) Spatial and temporal heterogeneity of regional lung ventilation determined by electrical impedance tomography during pulmonary function testing. J Appl Physiol 113: 1154–1161. [DOI] [PubMed] [Google Scholar]
  • 16. Wolf GK, Grychtol B, Boyd TK, Zurakowski D, Arnold JH (2010) Regional overdistension identified with electrical impedance tomography in the perflubron-treated lung. Physiol Meas 31: S85–S95. [DOI] [PubMed] [Google Scholar]
  • 17. Wolf GK, Grychtol B, Frerichs I, Zurakowski D, Arnold JH (2010) Regional lung volume changes during high-frequency oscillatory ventilation. Pediatr Crit Care Med 11: 610–615. [DOI] [PubMed] [Google Scholar]
  • 18. Dai T, Soleimani M, Adler A (2008) EIT image reconstruction with four dimensional regularization. Med Biol Eng Comput 46: 889–99. [DOI] [PubMed] [Google Scholar]
  • 19. de Lima CR, Mello LAM, Lima RG, Silva ECN (2007) Electrical impedance tomography through constrained sequential linear programming: a topology optimization approach. Meas Sci Technol 18: 2847–2858. [Google Scholar]
  • 20. Lionheart WRB (2004) EIT reconstruction algorithms: pitfalls, challenges and recent developments. Physiol Meas 25: 125–142. [DOI] [PubMed] [Google Scholar]
  • 21. Soleimani M, Gómez-Laberge C, Adler A (2006) Imaging of conductivity changes and electrode movement in EIT. Physiol Meas 27: S103–S113. [DOI] [PubMed] [Google Scholar]
  • 22. Victorino JA, Borges JaB, Okamoto VN, Matos GFJ, Tucci MR, et al. (2004) Imbalances in regional lung ventilation: a validation study on electrical impedance tomography. Am J Respir Crit Care Med 169: 791–800. [DOI] [PubMed] [Google Scholar]
  • 23. Wrigge H, Zinserling J, Muders T, Varelmann D, Günther U, et al. (2008) Electrical impedance tomography compared with thoracic computed tomography during a slow inflation maneuver in experimental models of lung injury. Crit Care Med 36: 903–9. [DOI] [PubMed] [Google Scholar]
  • 24. Frerichs I, Hinz J, Herrmann P, Weisser G, Hahn G, et al. (2002) Detection of local lung air content by electrical impedance tomography compared with electron beam CT. J Appl Physiol 93: 660–6. [DOI] [PubMed] [Google Scholar]
  • 25. Hinz J, Neumann P, Dudykevych T, Andersson LG, Wrigge H, et al. (2003) Regional ventilation by electrical impedance tomography: a comparison with ventilation scintigraphy in pigs. Chest 124: 314–22. [DOI] [PubMed] [Google Scholar]
  • 26. Kunst PW, Vonk Noordegraaf A, Hoekstra OS, Postmus PE, de Vries PM (1998) Ventilation and perfusion imaging by electrical impedance tomography: a comparison with radionuclide scanning. Physiol Meas 19: 481–90. [DOI] [PubMed] [Google Scholar]
  • 27. Richard JC, Pouzot C, Gros A, Tourevieille C, Lebars D, et al. (2009) Electrical impedance tomography compared to positron emission tomography for the measurement of regional lung ventilation: an experimental study. Crit Care 13: R82. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Ferrario D, Grychtol B, Adler A, Sola J, Bohm SH, et al. (2012) Toward morphological thoracic EIT: Major signal sources correspond to respective organ locations in CT. IEEE transactions on bio-medical engineering 59: 3000–8. [DOI] [PubMed] [Google Scholar]
  • 29. Adler A, Arnold JH, Bayford R, Borsic A, Brown B, et al. (2009) GREIT: a unified approach to 2D linear EIT reconstruction of lung images. Physiol Meas 30: S35–S55. [DOI] [PubMed] [Google Scholar]
  • 30. Adler A, Lionheart WRB (2006) Uses and abuses of EIDORS: an extensible software base for EIT. Physiol Meas 27: S25–S42. [DOI] [PubMed] [Google Scholar]
  • 31. Kaipio JP, Kolehmainen V, Somersalo E, Vauhkonen M (2000) Statistical inversion and monte carlo sampling methods in electrical impedance tomography. Inverse Probl 16: 1487–1522. [Google Scholar]
  • 32. Aykroyd RG, Soleimani M, Lionheart WRB (2006) Conditional Bayes reconstruction for ERT data using resistance monotonicity information. Meas Sci Technol 17: 2405–2413. [Google Scholar]
  • 33. Kolehmainen V, Arridge SR, Lionheart WRB, Vauhkonen M, Kaipio JP (1999) Recovery of region boundaries of piecewise constant coefficients of an elliptic PDE from boundary data. Inverse Probl 15: 1375–1391. [Google Scholar]
  • 34. Soleimani M, Lionheart WRB, Dorn O (2006) Level set reconstruction of conductivity and permittivity from boundary electrical measurements using experimental data. Inverse Probl Sci Eng 14: 193–210. [Google Scholar]
  • 35. Brühl M, Hanke M (2000) Numerical implementation of two noniterative methods for locating inclusions by impedance tomography. Inverse Probl 16: 1029–1042. [Google Scholar]
  • 36. Siltanen S, Mueller J, Isaacson D (2000) An implementation of the reconstruction algorithm of A Nachman for the 2D inverse conductivity problem. Inverse Probl 16: 681–699. [Google Scholar]
  • 37. Isaacson D, Mueller JL, Newell JC, Siltanen S (2006) Imaging cardiac activity by the D-bar method for electrical impedance tomography. Physiol Meas 27: S43–50. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Rahmati P, Soleimani M, Pulletz S, Frerichs I, Adler A (2012) Level-set-based reconstruction algorithm for EIT lung images: first clinical results. Physiol Meas 33: 739–750. [DOI] [PubMed] [Google Scholar]
  • 39. Joyce CJ, Baker AB, Kennedy RR (1993) Gas uptake from an unventilated area of lung: computer model of absorption atelectasis. J Appl Physiol 74: 1107–16. [DOI] [PubMed] [Google Scholar]
  • 40. Magnusson L, Spahn DR (2003) New concepts of atelectasis during general anaesthesia. Br J Anaesth 91: 61–72. [DOI] [PubMed] [Google Scholar]
  • 41. Neumann P, Rothen H, Berglund J, Valtysson J, Magnusson A, et al. (1999) Positive end-expiratory pressure prevents atelectasisduring general anaesthesia even in the presence of a high inspired oxygen concentration. Acta Anaesth Scand 43: 295–301. [DOI] [PubMed] [Google Scholar]
  • 42. Adler A, Guardo R, Berthiaume Y (1996) Impedance imaging of lung ventilation: do we need to account for chest expansion? IEEE Trans Biomed Eng 43: 414–420. [DOI] [PubMed] [Google Scholar]
  • 43. Barber D, Brown B, Freeston I (1983) Imaging spatial distributions of resistivity using applied potential tomography. Electron Lett 19: 933–935. [Google Scholar]
  • 44. Santosa F, Vogelius M (1990) A backprojection algorithm for electrical impedance imaging. SIAM J Appl Math 50: 216–243. [Google Scholar]
  • 45. Brown BH, Seagar AD (1987) The Sheffield data collection system. Clin Phys and Physiol Meas 8: 91–97. [DOI] [PubMed] [Google Scholar]
  • 46. Plonsey R (1963) Reciprocity applied to volume conductors and the ECG. IEEE Trans Biomed Eng 10: 9–12. [DOI] [PubMed] [Google Scholar]
  • 47. Schöberl J (1997) NETGEN an advancing front 2D/3D-mesh generator based on abstract rules. Comput Visual Sci 1: 41–52. [Google Scholar]
  • 48. Grychtol B, Adler A (2013) Uniform background assumption produces misleading lung EIT images. Physiol Meas 34: 579–93. [DOI] [PubMed] [Google Scholar]
  • 49. Adler A, Lionheart WRB (2011) Minimizing EIT image artefacts from mesh variability in finite element models. Physiol Meas 32: 823–834. [DOI] [PubMed] [Google Scholar]
  • 50. Graham BM, Adler A (2006) Objective selection of hyperparameter for EIT. Physiol Meas 27: S65–S79. [DOI] [PubMed] [Google Scholar]
  • 51. Adler A, Guardo R (1996) Electrical impedance tomography: regularized imaging and contrast detection. IEEE Trans Med Imag 15: 170–179. [DOI] [PubMed] [Google Scholar]
  • 52. Cheney M, Isaacson D, Newell JC, Simske S, Goble J (1990) NOSER: An algorithm for solving the inverse conductivity problem. Int J Imag Sys Technol 2: 66–75. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53. Hartinger AE, Guardo R, Adler A, Gagnon H (2009) Real-time management of faulty electrodes in electrical impedance tomography. IEEE Trans Biomed Eng 56: 369–377. [DOI] [PubMed] [Google Scholar]
  • 54. Borsic A, Adler A (2012) A primal–dual interior-point framework for using the L1 or L2 norm on the data and regularization terms of inverse problems. Inverse Probl 28: 095011. [Google Scholar]
  • 55. Polydorides N, Lionheart WR (2002) A Matlab toolkit for three-dimensional electrical impedance tomography: a contribution to the Electrical Impedance and Diffuse Optical Reconstruction Software project. Meas Sci Technol 13: 1871. [Google Scholar]
  • 56. Vauhkonen M, Vadász D, Karjalainen PA, Somersalo E, Kaipio JP (1998) Tikhonov regularization and prior information in electrical impedance tomography. IEEE Trans Med Imag 17: 285–293. [DOI] [PubMed] [Google Scholar]
  • 57. Murai T, Kagawa Y (1985) Electrical impedance computed tomography based on a finite element model. IEEE Trans Biomed Eng 32: 177–184. [DOI] [PubMed] [Google Scholar]
  • 58. Borsic A, Graham BM, Adler A, Lionheart WRB (2010) In vivo impedance imaging with total variation regularization. IEEE Trans Med Imag 29: 44–54. [DOI] [PubMed] [Google Scholar]
  • 59. Pulletz S, van Genderingen HR, Schmitz G, Zick G, Schädler D, et al. (2006) Comparison of different methods to define regions of interest for evaluation of regional lung ventilation by EIT. Physiological measurement 27: S115–27. [DOI] [PubMed] [Google Scholar]
  • 60. Radke OC, Schneider T, Heller AR, Koch T (2012) Spontaneous breathing during general anesthesia prevents the ventral redistribution of ventilation as detected by electrical impedance tomography: a randomized trial. Anesthesiology 116: 1227–34. [DOI] [PubMed] [Google Scholar]
  • 61. Zick G, Elke G, Becher T, Schädler D, Pulletz S, et al. (2013) Effect of PEEP and tidal volume on ventilation distribution and end-expiratory lung volume: a prospective experimental animal and pilot clinical study. PloS one 8: e72675. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62. Bikker IG, Preis C, Egal M, Bakker J, Gommers D (2011) Electrical impedance tomography measured at two thoracic levels can visualize the ventilation distribution changes at the bedside during a decremental positive end-expiratory lung pressure trial. Critical care (London, England) 15: R193. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63. Mamatjan Y, Böhm S, Gaggero PO, Adler A (2011) Evaluation of EIT system performance. Physiol Meas 32: 851–865. [DOI] [PubMed] [Google Scholar]
  • 64. Lionheart WRB (1998) Boundary shape and electrical impedance tomography. Inverse Probl 14: 139–147. [Google Scholar]
  • 65. Grychtol B, Lionheart WRB, Bodenstein M, Wolf GK, Adler A (2012) Impact of model shape mismatch on reconstruction quality in electrical impedance tomography. IEEE Trans Med Imag 31: 1754–1760. [DOI] [PMC free article] [PubMed] [Google Scholar]