Skip to main content
The Review of Scientific Instruments logoLink to The Review of Scientific Instruments
. 2014 Aug 5;85(8):083101. doi: 10.1063/1.4890199

Design and characterization of a novel multimodal fiber-optic probe and spectroscopy system for skin cancer applications

Manu Sharma 1, Eric Marple 2, Jason Reichenberg 3, James W Tunnell 1
PMCID: PMC4137875  PMID: 25173240

Abstract

The design and characterization of an instrument combining Raman, fluorescence, and reflectance spectroscopic modalities is presented. Instrument development has targeted skin cancer applications as a novel fiber-optic probe has been specially designed to interrogate cutaneous lesions. The instrument is modular and both its software and hardware components are described in depth. Characterization of the fiber-optic probe is also presented, which details the probe's ability to measure diagnostically important parameters such as intrinsic fluorescence and absorption and reduced scattering coefficients along with critical performance metrics such as high Raman signal-to-noise ratios at clinically practical exposure times. Validation results using liquid phantoms show that the probe and system can extract absorption and scattering coefficients with less than 10% error. As the goal is to use the instrument for the clinical early detection of skin cancer, preliminary clinical data are also presented, which indicates our system's ability to measure physiological quantities such as relative collagen and nicotinamide adenine dinucleotide concentration, oxygen saturation, blood volume fraction, and mean vessel diameter.

I. INTRODUCTION

Spectroscopic techniques utilize the interaction of light with biological tissue to study tissue optical properties, which change with disease progression and can be used for diagnosis. Deployment of spectroscopic-based devices has the potential to significantly augment clinical diagnosis. Three common spectroscopic techniques are Raman spectroscopy (RS), diffuse reflectance spectroscopy (DRS), and laser-induced fluorescence spectroscopy (LIFS). These techniques have been applied – either individually or in combinations – throughout the entire human body to investigate a wide range of pathologies including: atherosclerosis,1,2 osteoporosis,3 brain edema,4 cataract formation,5 kidney stones,6,7 and diabetes8 and cancer of the breast,9–11 cervix,12 esophagus,13,14 gastrointestinal tract,15,16 brain,17 lungs,18,19 ovaries,20 and bladder.21 Here, we present a novel multi-modal spectroscopy (MMS) device, combining RS, LIFS, and DRS, for the purpose of fast and non-invasive early detection of skin cancer which uses a variety of instrumentation and a custom contact probe capable of delivering and collecting light for all three modalities.

A. Skin cancer

Skin cancer, in both its melanoma and non-melanoma forms, has the largest reported incidence of all cancers in the United States. North America has the 2nd highest age-standardized melanoma incidence rate in the world.22 Early identification of skin cancer is paramount for its effect upon patient survival; the five-year “regional” survival rate of melanoma is 62%, dropping to 16% for “distant” detection. Significant ambiguity exists for the clinical distinction via visual inspection between dysplastic nevi and melanoma, raising issues of considerable practical and financial importance. Currently, the only reliable method of distinguishing between dysplastic nevi and melanoma is via stained biopsies, which is invasive and expensive. Clinician specificity for melanoma is approximately 4%,23,24 which means that roughly 25 more times biopsies are performed than required, translating to an estimated cost of $6 × 109 to the US health care system. In 2007, the mean wait time for new patients in urban areas to see a dermatologist was 33 days; in rural areas, the number rises to 46 days.25 Furthermore, delay in diagnosis due to “off site” analysis imparts an emotional cost to the patient.

Skin cancer patient care is currently limited by the invasive nature of biopsies and the high cost associated with biopsies due to inadequate screening techniques. Motivated by improving this degree of patient care, we present a noninvasive, real-time spectroscopic-based technology to significantly improve lesion pathology specificity for the early detection of skin cancer. Ultimately, this device will save lives through early detection and its improved differentiation between skin pathologies will translate directly to lower costs and morbidity.

B. MMS

In short, MMS characterizes the tissue microenvironment via morphological changes observed through DRS and bio-chemical information via RS and LIFS. The DRS measurement is a function of tissue scattering and absorption properties, which in turn are dependent upon tissue morphological changes. Hence, analysis yields information about tissue blood fraction, oxygen saturation, tissue scattering coefficient, nuclear morphology, and collagen structure. LIFS is bio-chemically sensitive as it interrogates endogenous fluorophores such as nicotinamide adenine dinucleotide (NADH), flavin adenine dinucleotide (FAD), and collagen. Their fluorescence levels change with neoplastic progression that is associated with altered cellular metabolic pathways (NADH, FAD)26 or an altered structural tissue matrix (collagen).26–29 Raman spectroscopy exploits the inelastic scattering (so-called “Raman” scattering) phenomena to detect spectral signatures of important disease progression biomarkers, including lipids, proteins, and amino acids. Raman spectroscopy is more constituent-specific than fluorescence and is capable of spectrally “breaking down” the biochemical composition; however, the two techniques are complementary as they probe different bio-molecular species.

As different techniques provide complementary information, it has been shown that an instrument combining multiple techniques offers a more precise description of disease status. Specifically for skin cancer, it has been demonstrated that a modal combination improves malignant melanoma30 and non-melanoma (basal and squamous cell carcinomas) diagnosis.31 Furthermore, MMS has been successfully applied for the early detection of atherosclerotic plaque.32,33 The successful application of MMS in non-cancer related pathologies indicates that it has considerable potential and its efficacy would be further tested by applying it to skin cancer.

C. Fiber-optic probes

Biomedical optical spectroscopy measurements very commonly use fiber-optic probes, which serve as the optical interface between the sample and the spectroscopic equipment. The fiber-optic probes contain fiber bundles that are responsible for both delivering and collecting light from the sample. Many types of probes have been used within the research community and cannot all be summarized here; however, comprehensive reviews of DRS, LIFS, and RS probe technology are available.34 Briefly, combination LIFS-DRS probes have been demonstrated by past studies,35,36 as well as a probe combining all three techniques.32

There are several important, and challenging, design and functional considerations for a probe that combines reflectance, fluorescence, and Raman signals. First, due to the very low probability of Raman scattering (one inelastically scattered photon for every 109 elastically scattered photons), the optical design must be fine-tuned to maximize RS signal-to-noise ratios. Second, the traditional source-detector geometry employed for reflectance measurements needs to be preserved such that tissue scattering and absorption properties can be separated and quantified.

II. SYSTEM DESCRIPTION

Our MMS system design is modular, consisting of 2 main “sub-units”: hardware to control the LIFS and the DRS measurements and separate hardware for RS acquisition.

A. Sources

The MMS system has three sources for each of the three modalities: a pulsed Xenon flash lamp (L7684, Hamamatsu Photonics, Bridgewater, NJ), which provides broadband 375–700 nm illumination for DRS; a pulsed 337-nm nitrogen laser (NL-100, Stanford Research Systems, Sunnyvale, CA) to induce NADH and collagen fluorescence; and a 830-nm diode laser (Lynx, Germany) for Raman excitation. The Raman diode laser is gated by a mechanical shutter which is controlled by triggering software written in MATLAB and LabVIEW. In order to prevent second-order dispersion contaminating the reflectance spectra, the Xenon white light is first passed through a 340-nm long-pass filter (Asahi Spectra, Torrance, CA) and then coupled into a fiber. The Xenon lamp provides a pulse of full width half maximum (FWHM) 2.9 μs. For LIFS, the nitrogen laser and has been configured to provide approximately 160 μJ per pulse for a pulse FWHM of 3.5 ns. The LIFS signal-to-noise ratio could be increased by increasing the pulse power; however, the value of 160 μJ strikes an effective balance between sufficient signal strength and laser cartridge lifetime. The output power of the Raman diode laser can be controlled by adjusting the supplied current through the custom software; for this application, 56 mW output power (0.198 A supplied current) at 830-nm is delivered from the laser engine. The Raman laser is housed inside the Raman module, which is completely shielded by specially constructed blackened material (Thorlabs, NJ, USA) to prevent any stray light getting in or leaking out.

B. MMS probe

1. Probe design

The multi-spectroscopy probe is similar in design to the probe described in detail elsewhere (US patent no. 8,175,423 and US patent application no. 20120236303). The probe is designed to be used in contact with the skin. The distal end of the probe is a polished, flat surface to ensure that the contact is as uniform as possible across the probe diameter to prevent measurement irregularities arising from gaps between the skin and probe and non-zero contact angles. Illustrations of the probe design are shown in Figure 1, where Figure 1(a) is a front-on view of the probe distal end and Figure 1(b) is an exploded assembly view to show all the components. As seen in Figure 1(a), seven 300 μm Raman collection fibers and a DRS/LIFS “triangle” concentrically surround a 200 μm core Raman delivery fiber.

FIG. 1.

FIG. 1.

Schematics of the MMS probe. The left panel, (a), shows the probe distal end front view. The right panel, (b), shows the assembly exploded view with optical elements such as the filters and front lens identified along with the collection and delivery fibers for all three modalities.

The DRS/LIFS triangle contains two low OH 200 μm core visible light collection fibers and a high OH 200 μm core DRS/LIFS delivery fiber. The low and high OH cores are chosen for collection and delivery, respectively, because of the wavelength dependent attenuation characteristics of silica fiber optic cables: high OH content fibers have lower losses in the UV (hence, the selection for the LIFS delivery fiber) while low OH content fibers have lower losses in the visible (hence, the selection for the collection fibers where all the collected light is in the visible). The triangle of fibers pass through holes drilled in the long pass donut filter and the front lens as illustrated in Figure 1(b) in order to bypass these elements. Without the front lens bypass, it was seen that the source-detector geometry, necessary for extracting optical properties from the reflectance signal, was not sufficiently preserved due spectral aberrations and focusing effects introduced by the front lens. The current design avoids these issues and allows the DRS and LIFS data to be collected in the same fashion as the standard bundle probe configuration. The trade-off is that there is not perfect overlap between the Raman and LIFS/DRS collection spots; however, ray tracing simulations confirm that the overall delivery spot diameter (spanning all three modalities) is approx. 600 μm and this overlap is sufficient.

The Raman portion of the probe uses seven low hydroxyl (OH) content 300-μm core, 0.22 NA collection fibers. A donut shaped 830 nm long pass filter is positioned in front of these seven fibers, which rejects the 830 nm laser light and passes the Raman light from the sample. These seven fibers surround a stainless steel tube inside which is the laser delivery fiber assembly. The laser delivery fiber is a 200-μm core low OH, 0.22 NA fiber which has a small 830 nm band-pass filter positioned in front of it. The choice of fibers and filtering of Raman probes has been discussed by many sources previously.34,37,38 The two-piece converging front lens is made of a plano convex 2 mm diameter curvature sapphire back portion (the high refractive index bends the light sharply) and a flat front portion of 1 mm thick plano Magnesium Fluoride which has virtually no Raman signature. Epoxy is used to bond the required individual components together. The fibers, lens, and other components are placed inside a stainless steel 14 gauge extra thin wall needle tube (0.072 in. ID, 0.083 in. OD 2.1 mm OD).

2. Coupling

The MMS probe has 2 input connections: one for the 830-nm Raman laser and one port for both the N2 laser and the Xenon lamp. Raman and LIFS/DRS ports are separated as near infrared (NIR) light for the Raman modality has different optical design requirements (fiber material, filters, transmission, etc.) than the ultraviolet and visible wavelengths used in DRS and LIFS.

The white light and laser pulses (DRS and LIFS modes) are coupled into optical fibers and guided into a 3 × 1 fiber optic switch (FSM-13, Piezosystems Jena, Germany). The switch is a microelectromechanical (MEMS) device, which uses microprisms to control and open different optical ports to ensure that the 377-nm laser light and broadband Xenon light are separated and coupled sequentially into the MMS probe without any overlap. The switch is controlled via transistor-transistor logic (TTL) pulse trains initiated within the custom software. Light from the switch's output is passed to the dual LIFS/DRS input port of the MMS probe via a subminiature version A (SMA) nipple fitting; roughly 30% loss in signal is measured due to the optical switch and SMA fitting. The 830-nm Raman laser light is delivered without the optical switch and is triggered after the LIFS and DRS pulses.

C. Photodetection

The MMS detection hardware consists of components optimized for visible (LIFS and DRS) and NIR (RS) detection.

1. Reflectance and fluorescence

The LIFS/DRS spectral system comprises a interline CCD camera (CoolSNAP HQ, Princeton Instruments, Trenton, NJ) cooled to −30 °C. For each of the LIFS and DRS pulses, the CCD is gated at 50 μs. The distal ends of the two DRS/LIFS fibers are aligned with the vertical axis of the spectrograph (SpectraPro 2150i, Princeton Instruments, Trenton, NJ) using software provided by the manufacturer (WinSpec, Princeton Instruments, Trenton, NJ). A 150 grooves/mm grating, blazed at 500 nm, is used in order to capture the entire visible spectrum needed for LIFS (385–650 nm) and DRS (375–700 nm). A slit width of 200 μm is used. To improve the SNR, we bin every three pixels for a spectral dispersion of 0.78 nm/pixel and a resulting spectral resolution FWHM of 5.6 pixels (4.32 nm).

2. Raman

The Raman system consists of a 1024 × 1024 camera (IMG, Finger Lakes Instrumentation, NY) cooled to −30 °C which is coupled to an f/1.8 spectrograph (Holospec, Kaiser Optical Systems, MI) utilizing a low-frequency Stokes grating for 830 nm excitation. The longer Raman excitation wavelength reduces tissue autofluorescence by a factor of four compared to 785 nm excitation. A custom 200 μm width slit was installed. This slit width was selected as it strikes an effective balance between ensuring sufficient resolution for spectral analysis and signal strength. The measured spectral dispersion is 1.79 cm−1 per pixel and the spectral resolution is a FWHM of 13.43 cm−1.

D. Data acquisition hardware and software

Custom software has been written in LabVIEW (National Instruments, Austin, TX) for single-click operation of the entire MMS system. The software executes MMS data collection by sequentially capturing DRS, LIFS, and RS spectra. For the DRS and LIFS modalities, the sources are triggered for data acquisition via TTL pulses provided by a timer-counter board (NI 2121, National Instruments, Austin, TX) while for RS this same timer-counter board triggers the mechanical shutter to open (as the diode laser is a continuous source and therefore always on). The DRS and LIFS camera is controlled by a PCI card (PCI-6602, National Instruments, Austin, TX) and operated, in part, by pre-written software (R3 Software, Princeton, NJ). At the basic software architecture level, the Raman instrument components (laser and camera) are controlled via drivers written in C++ and incorporated into a MATLAB code; however, these codes and drivers are called and user inputs implemented within LabVIEW. Spectra are displayed for instant user feedback via onboard binning and background subtraction. Finally, for the DRS and LIFS modalities, an extra step is required for optical switch operation.

E. Calibration

To achieve day-to-day consistency and accurate spectral measurements, calibration procedures must be performed. The Raman and LIFS/DRS wavelength calibrations are performed by measuring known spectral lines from a solid 4-acetamidophenol (Tylenol) capsule and a mercury-argon pencil lamp (Hg-1, Ocean Optics, FL), respectively. To account for how the quantum efficiency of the detectors alter the fluorescence and Raman spectra (system response), the spectrum of a calibrated Tungsten light source (LS1-Cal, Ocean Optics, FL) shone onto a 20% reflectance standard is recorded; the system response is not required for the DRS measurement as it is inherently normalized. To account for white light source day-to-day variations, the reflectance amplitude is measured by recording the spectrum of a solid titanium dioxide standard. This step ensures that all reflectance measurements are calibrated to the LUT model before extraction of optical properties. Background calibration, which accounts for stray light and dark current, is performed for all three modalities by taking spectra with the lights off. With external triggering, the shot-to-shot variation from the N2 laser was measured at 12%. Therefore, to account for these fluctuations, a beam splitter was installed to create a power measurement arm and each fluorescence spectra is then normalized by this measured power (3A-P, Ophir Optics, Israel). These calibration procedures were performed prior to every clinical data collection day (to account for day-to-day variations) and also whenever any optical alignment took place such as distal end re-insertion into a spectrometer, slit width adjustment, and camera or lens alignment.

F. Clinical safety and transportation

As our system is used in the clinic, stringent safety protocols had to be followed. The output energy of the N2 laser is measured at 6.5 μJ, which is considerably lower than the maximum permissible levels (53 μJ) of a Class 1 device.39 The mechanical shutter, which blocks the 830-nm laser diode laser, opens when triggered through the software, closes immediately after the acquisition and remains closed until the next acquisition.

For clinical portability, all MMS components are mounted to a two-level utility cart (4546-10, Rubbermaid, Winchester, VA) as shown in Figure 2. The utility cart was specially outfitted with 6-in. pneumatic caster wheels to prevent vibration and increase transportation ease. To ensure electrical line isolation and prevent electrical damage, all powered MMS components (spectrometers, cameras, etc.) are connected to the main power via an isolation transformer power conditioner (IS250, Tripplite, Chicago, IL). Furthermore, in the event of main power loss and to allow the system to be transported between clinical rooms, the MMS system is also connected to a battery supply (CP1500AVR, CyberPower, Shakopee, MN) which provides approximately 10 min of external power.

FIG. 2.

FIG. 2.

Photo of the MMS clinical system (left) and the MMS probe (right).

III. SPECTRAL MODELING AND ANALYSIS

A. Reflectance

The diffusion approximation to the radiative transport equation, which is most commonly used in order to extract optical property information from reflectance spectra,40,41 is only valid when the separation between the excitation fiber (source) and collection fibers (detector) are less than the mean free path and in absorption-dominated tissues (high albedo absorption ≫ scattering). MMS work conducted by other researchers, who studied atherosclerosis, used this diffusion approximation to calculate reflectance using their MMS probe. However, the diffusion approximation is not valid for our application as skin cancers originate in shallow cutaneous layers, which requires relatively short source-detector separations to confine the light only to these shallow layers, and the lesions of interest (i.e., dysplasia) can have low albedo. As the diffusion approximation is not valid, we use a look-up table (LUT) inverse model approach that is appropriate for our conditions (short source-detector separation and low albedo) and accurately measures tissue optical properties such as the reduced scattering coefficient, μs, and absorption coefficient, μa. It has been shown by previous researchers42 that the diffuse reflectance will be the same for any combination of μS, and g that result in the same μS. We have previously shown in our laboratory – by performing detailed parametric studies – that the extracted absorption and reduced scattering coefficient values have less than 10% error when the anisotropy is greater than 0.7 when using the LUT method.43 The LUT is essentially a database of reflectance values across a range of scattering and absorption values. The database is generated by measuring reflectance spectra from a matrix tissue simulating phantoms with known optical properties and then interpolating between these values to generate a topography in R, μa and μs space. Any reflectance spectra, obtained from a sample with unknown optical properties, can be fit to this database in order to determine its optical properties. The tissue phantoms are created by using polystyrene beads with nominal 1 μm diameter and 2.6% solids by volume (Polysciences, Warrington, PA) and black India ink (Speedball, Statesville, NC) as the scattering and absorption media, respectively. Mie theory was used to calculate the μs of the polystyrene beads (and therefore the amount of volume to add for a desired μs) and a spectrophotometer (DU720, Beckman Coulter, CA) to measure the μa of a stock India ink solution. In total, 21 phantoms were used to generate the LUT, spanning physiological relevant values of μs (0.44–4.74 mm−1) and μa (0–2.5 mm−1). Raw spectra were then collected by the probe and reflectance spectra calculated using the following equation:

R diffuse (λ)=I sample (λ)I background (λ)[I standard (λ)I background (λ)]×100/R standard , (1)

where Isample(λ) is the raw spectrum from the phantom, Ibackground(λ) is the background spectrum, Istandard(λ) is the spectralon standard spectrum and 100/Rstandard is a factor used to account for the calibrated reflectance level of the spectralon standard (throughout this paper all results were obtained with a 20% spectralon reflectance standard). Spectra are presented in terms of wavelength by performing the wavelength calibration procedure discussed above. Validation of the LUT is discussed below. This approach follows previous work conducted in our laboratory43,44 and the successful application of the LUT approach for skin cancer diagnosis.45

A nonlinear optimization fitting routine is employed to minimize the difference (χ2) between the database LUT reflectance spectra and the measured reflectance spectra between 400 and 650 nm. For the LUT validation (see Sec. IV A), the reduced scattering coefficient and absorption coefficient are constrained to the following forms:

μs(λ)=μs(λ0)λλ0B, (2)
μa(λ)=2.303yA(λ)absorberL, (3)

where λ0 is 630 nm, A(λ) is the absorbance spectra of the dye when measured using a spectrophotometer, L is the path length of the spectrophotometer measurement, [absorber] is the concentration of absorber used for the spectrophotometer measurement, and y is a scaling factor to account for the dilution of the dye solution used in the spectrophotometer. For the validation study, the fitting outputs are μs(λ0), B, and [absorber], from which μs(λ) and μa(λ) can be calculated. For clinical fitting (see Sec. IV B), the reduced scattering coefficient is constrained in the same fashion (Eq. (2)), however, the physiological absorption coefficient is calculated using the following equations described by van Veen:46

μa(blood)(λ)=150υ[αɛHbO2(λ)+(1α)ɛHb(λ)], (4)
Cpack=1e(2μa(bl)(λ)Dvessel)2μa(bl)(λ)Dvessel, (5)
μa(tissue)(λ)=Cpackνμa(bl)(λ)+[mel]ɛmel(λ), (6)

where ν is the blood volume fraction sampled by the light (assuming a hemoglobin concentration of 150 mg/ml in the bloodstream), α is the oxygen saturation (ratio of HbO2 to total Hb), εHbO2(λ), εHb(λ), and εmel(λ) are the extinction coefficients of oxygenated hemoglobin, deoxygenated hemoglobin, and melanin, respectively, Dvessel is the mean vessel diameter and [mel] is the concentration of melanin. These equations are used to account for the inhomogeneous distribution of blood vessels in tissue. For clinical data, the fit outputs are μs(λ0), B, [mel], ν, α, and Dvessel.

B. Fluorescence

Fluorescence spectra are first background corrected by subtracting a dark spectrum (lights off, probe pointing upwards) from the raw spectrum. Next, a wavelength calibration is performed by using a peak fitting algorithm to find the pixel locations of HgAr lines, fitting a 3rd order polynomial (as 4 strong lines are seen in the visible) from these pixel locations to the known wavelengths of these lines and then converting the entire pixel array to wavelength space. The intensity calibration is performed by scaling the measured blackbody spectrum to the measured values provided by the manufacturer; the measured spectrum will be altered due to the wavelength dependence of the detector's quantum efficiency and this step is necessary in order to correct for this instrument response.

The turbid nature of raw tissue alters the fluorescence signal such that the measured fluorescence spectral shape is altered and its intensity attenuated. Therefore, the intrinsic fluorescence – the true endogenous fluorescence without scattering or absorption distortion – must be calculated in order to accurately model physiological fluorescence. The intrinsic fluorescence is calculated by using the photon migration model of Zhang et al.,47 which uses the measured reflectance of the sample (with a particular μa and μs) and probe specific parameters in order to correct the fluorescence

IF(λ)=F(λ)R(λx)μs(λx)lR0(λx)R0(λx)R0(λ)ɛ(λx)ɛ(λ)R(λ)R0(λ)+ɛ(λ), (7)

where λx is the excitation wavelength, μs(λx) is the value of the scattering coefficient at the excitation wavelength, IF(λ) is the intrinsic fluorescence spectrum, F(λ) is the measured fluorescence spectrum, R(λ) is the measured reflectance spectrum, R(λx) is the value of the measured reflectance at the excitation wavelength, R0(λ) is the measured reflectance spectrum for no absorption, and R0(λx) is the value of the reflectance at the excitation wavelength for no absorption. F(λ) and R(λ) are directly measured via experiment and R(λx) is a constant determined from the R(λ) spectrum. R0(λ) is calculated by first fitting the measured spectrum, R(λ), to the LUT to obtain μa and μs. The corresponding reflectance spectrum with this μs is then numerically evaluated via two-dimensional interpolation of the LUT space by constraining μa = 0. Additionally, ɛ(λ) is calculated using ɛ = eβ−1, where β = S(1−g) for a probe specific constant, S, and the anisotropy, g; ɛ(λx) is the value of ɛ(λ) at the excitation wavelength. The last remaining term, l, is also a probe specific constant.

As discussed above, S and l are probe specific parameters and they are determined empirically. These were calculated as follows. First, the fluorescence spectra of 0.26 μM molar Stilbene 420 (Exciton, OH) dissolved in ethanol was measured. Stilbene-420 was chosen as its peak emission wavelength is similar to that of NADH. The heavily diluted 0.2 μM solution was chosen in order to ensure that the Stilbene itself was not contributing to the scattering signal while guaranteeing sufficient fluorescence SNR. Without any scattering or absorption, this spectra corresponds to IF(λ). Next, a spectra was taken with polystyrene beads added to the 0.26 μM Stilbene solution to create a solution with μs = 1.5 mm−1. As this spectra is only influenced by scattering (no absorption), it is equal to R0(λ) and from which R0(λx) can be determined. In sequential order, 4 volumes of red dye were added to the solution to create solutions with μa values of 0.29, 0.57, 0.86, and 1.14 mm−1. These spectra corresponded to four different R(λ). An optimization routine was written to calculate S and l by minimizing the difference between the four IF(λ) spectra when calculated using Eq. (2). The values of S and l, specific to our particular MMS probe, are 1.83 and 0.163.

In tissue, 337 nm excites both NADH and collagen.48 We assume that the measured intrinsic fluorescence is a linear combination of fluorescence from NADH and collagen, which can be expressed mathematically as

IF(λ)=A1IFNADH(λ)+A2IFcollagen(λ), (8)

where IFNADH(λ) and IFcollagen(λ) are the intrinsic spectra of NADH and collagen, respectively, and χnadh = A1 / (A1 + A2) and χcollagen = A2 / (A1 + A2) represent the relative concentrations of NADH and collagen, respectively. The values of A1 and A2 are determined by a fitting routine.

C. Raman

Raman spectra is collected in the “fingerprint” region (∼400–1800 cm−1), because it is a rich source of Raman bio-markers useful for skin properties and skin cancer diagnosis.49–55 The first processing step involves background subtraction whereby a dark spectrum is subtracted from the raw spectrum. Second, the wavelength and intensity calibrations are performed in a very similar manner as described for the fluorescence spectra; the only difference is that the peak fitting algorithm is used to find Tylenol pixel locations (instead of HgAr lines as needed for the fluorescence). Raman incident laser light can cause tissue autofluorescence, which has practical implications as the fluorescence swamps the Raman signal. This autofluorescence is most likely due to the morphological make-up of the epidermis, which absorbs light up to 1000 nm and can re-emit it as fluorescence.48 Tissue autofluorescence is removed by fitting the Raman spectra to a 5th order polynomial37,38 and subtracting the fit from the raw spectrum, revealing the desired endogenous Raman signals.

IV. MMS SYSTEM AND PROBE PERFORMANCE

A. LUT validation

The LUT was validated by fitting the LUT topography to spectra obtained from 47 validation phantoms with known optical properties. The validation phantoms were fabricated by using the polystyrene beads and colored food dye (red, blue, and green) to simulate scattering and absorption, respectively. These phantoms spanned ranges of 0.72–4.31 mm−1 and 0–2.42 mm−1 for μs and μa, respectively, which covered approximately 90% of the LUT surface. For each validation phantom, spectra were averaged across 5 measurements. Using the LUT fitting approach described above, the optical properties were extracted with normalized root-mean-square errors of 7.19% and 9.81% for μs and μa, respectively, as shown in Figures 3(a) and 3(b); for a particular optical property, the errors were calculated by averaging across all wavelength and all 47 phantoms.

FIG. 3.

FIG. 3.

Comparison between fitted (LUT) and expected (experimental) for the reduced scattering coefficient (left) and absorption coefficient (right).

B. Raman in vivo performance

To demonstrate the Raman performance, in vivo Raman spectra were obtained at different locations on the body of a healthy Caucasian male volunteer, Figure 4. Each location was measured 5 times. Tissue autofluorescence was removed using the polynomial fitting procedure described previously. Encouragingly, bands identified by previous researchers as being important for skin cancer diagnosis52,53,56 – the Amide I, Amide III, and CH2 scissor – show very good signal to noise ratios.

FIG. 4.

FIG. 4.

Sample in vivo Raman spectra obtained using the MMS probe for various body locations. Exposure time is 4 s.

C. MMS clinical performance

The MMS system is currently being used for clinical testing for the detection of non-melanoma and melanoma skin cancers. Clinical data acquisition times are roughly 4.5 s in total, comprising a 4 s Raman exposure and the 3 Xenon flashes, N2 laser pulse, and optical switching making up the remaining 500 μs. The clinical data acquisition procedure was as follows: (1) Dermatologist identified suspicious lesion, (2) 3 repeat measurements made on each lesion, (3) 3 repeat measurements of corresponding normal skin as close to the lesion as possible, and (4) lesion is biopsied and lesions classified using histopathology. Sample MMS spectra are presented in Figure 5 for a basal cell carcinoma. Fitting to Eqs. (2)–(6) and (8) above gives physiological quantities consistent with a non-melanoma skin cancer clinical study previously conducted by our group.

FIG. 5.

FIG. 5.

MMS clinical data of a basal cell carcinoma for Raman (left), diffuse reflectance (right), and intrinsic fluorescence (right) modalities. Physiological quantities obtained from DRS spectral fitting were μs(λ0)=1.7 mm 1, υ = 2.87, α = 0.46, and Dvessel = 0.48 μm and χnadh = 0.55 and χcollagen = 0.45 from the intrinsic fluorescence spectral fitting.

D. Repeatability

As our final long-term goal is to apply this technology in the clinic, two major evaluations would be the specificity/sensitivity of the instrument and its repeatability. As the clinical data collection is ongoing, specificity/sensitivity data are not presented at this stage. However, bench top and clinical data have been collected to quantify the repeatability of the MMS system. In order to quantify the repeatability of the system, each modality was characterized separately. For Raman, the standard deviation as a percentage of the mean for 50 peak intensity measurements of a major Tylenol peak (at ∼1324 cm−1) was calculated to be 1.6% and the variation in Scissor band peak (∼1450 cm−1) intensity from 5 measurements (from Sec. IV B) was calculated to be 6%. It is expected that the in vivo measurement displays greater variation than the bench top measurement due to patient movement. For DRS, the repeatability was measured to be 3.2% for μs and 3.4% for μa. These values were determined by averaging across 3 sets of 20 measurements each (total of 60 measurements) for the reflectance values at maximum absorption and at 630 nm for the 3 most highly absorbing and highly scattering liquid phantoms (blue, green, and red with μs (630 nm) = 4.31 and μa = 2.32 mm−1). From 5 measurements of the volar forearm, the variation was 7.0% and 8.2% for values at 630 nm and 417 nm (Hemoglobin Soret band peak). For LIFS, the repeatability was assessed by calculating the variance in 100 measurements of the fluorescent peak of a heavily diluted Coumarin sample (Exciton, OH) at 337 nm excitation. Without accounting for the laser fluctuation, the variance is 17% and when normalizing the signal by the power arm measurement (see Sec. II E) this variance dropped to 4%, which clearly demonstrates the need for the power arm measurement in order to account for laser fluctuations.

V. DISCUSSION

A. Validation studies

The addition of an absorber to a phantom involves extra pipetting steps in order to complete the spectrophotometer measurement, which is not needed for a purely scattering phantom. It is proposed that the lower extraction accuracy of μa compared to μs is due to the extra steps which represent more sources of measurement error due to instrument resolution and calibration (spectrophotometer, pipettes) and also human error. At this stage, we have no definitive explanation for the μa trend of increasing divergence between fitted and expected values (as seen in Figure 3) with increasing μa values. However, the extra μa measurement steps could offer an explanation as equipment error (spectrophotometer, pipettes) could be proportional to volume or concentration, which would result in the systemic divergence observed.

B. Clinical feasibility

There is a clear discrepancy between the intrinsic fluorescence fit (used to determine the collagen and NADH relative concentrations) and the extracted measured intrinsic fluorescence, Figure 5. This discrepancy is primarily due to the fact that our intrinsic fluorescence model does not fully capture the physics. We assume linear superposition of the two fluorophores, which implies that their measured fluorescence is independent of each other. In reality, these fluorescence signals are coupled as NADH has non-negligible absorption within the visible and will therefore absorb some of collagen's fluorescence emission – the opposite situation will also occur (collagen absorbing some of NADH's fluorescence). Therefore, the measured flux from both NADH and collagen fluorescence will be attenuated due to this fluorescence emission-absorption overlap and will spectrally manifest itself as attenuated signal and a slightly different peak location. A more accurate model would include second-order, nonlinear terms to describe this fluorescence emission-absorption overlap; however, previous researchers13,33 have successfully applied this linear superposition approach for clinical diagnosis as it still does a very good job of capturing the endogenous fluorescence physics.

ACKNOWLEDGMENTS

The authors thank Dr. Naras Rajaram, Dr. Steve Fulgham, and Brandon Nicols for assistance with software and hardware development, instrumentation, and patience. This research was conducted at the Biophotonics Laboratory at the University of Texas at Austin and made possible via funds from the Community Foundation of North Central Wisconsin (the primary author is the A. Ward Fellowship holder) and the National Institutes of Health (NIH) (R21EB015892).

REFERENCES

  • 1.Buschman H. P.et al. , “Diagnosis of human coronary atherosclerosis by morphology-based Raman spectroscopy,” Cardiovasc. Pathol. 10(2), 59–68 (2001). 10.1016/S1054-8807(01)00063-1 [DOI] [PubMed] [Google Scholar]
  • 2.Nazemi J. H. and Brennan J. F. III, “Lipid concentrations in human coronary artery determined with high wavenumber Raman shifted light,” J. Biomed. Opt. 14, 034009 (2009). 10.1117/1.3130302 [DOI] [PubMed] [Google Scholar]
  • 3.McCreadie B. R.et al. , “Bone tissue compositional differences in women with and without osteoporotic fracture,” Bone 39(6), 1190–1195 (2006). 10.1016/j.bone.2006.06.008 [DOI] [PubMed] [Google Scholar]
  • 4.Wolthuis R.et al. , “Determination of water concentration in brain tissue by Raman spectroscopy,” Anal. Chem. 73(16), 3915–3920 (2001). 10.1021/ac0101306 [DOI] [PubMed] [Google Scholar]
  • 5.Ozaki Y.et al. , “Nondestructive analysis of biological materials by FT-IR-ATR. Method I: Direct evidence for the existence of collagen helix structure in lens capsule,” Appl. Spectrosc. 43(1), 138–141 (1989). 10.1366/0003702894201789 [DOI] [Google Scholar]
  • 6.Kodati V. R., Tu A. T., and Turumin J. L., “Raman spectroscopic identification of uric-acid-type kidney stone,” Appl. Spectrosc. 44(7), 1134–1136 (1990). 10.1366/0003702904086470 [DOI] [Google Scholar]
  • 7.Kodati V. R.et al. , “Raman spectroscopic identification of phosphate-type kidney stones,” Appl. Spectrosc. 45(4), 581–583 (1991). 10.1366/0003702914336859 [DOI] [Google Scholar]
  • 8.Berger A. J.et al. , “Multicomponent blood analysis by near-infrared Raman spectroscopy,” Appl. Opt. 38(13), 2916–2926 (1999). 10.1364/AO.38.002916 [DOI] [PubMed] [Google Scholar]
  • 9.Haka A. S.et al. , “Diagnosing breast cancer by using Raman spectroscopy,” Proc. Natl. Acad. Sci. U.S.A. 102(35), 12371 (2005). 10.1073/pnas.0501390102 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Zhu C.et al. , “Diagnosis of breast cancer using fluorescence and diffuse reflectance spectroscopy: A Monte-Carlo-model-based approach,” J. Biomed. Opt. 13, 034015 (2008). 10.1117/1.2931078 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Volynskaya Z.et al. , “Diagnosing breast cancer using diffuse reflectance spectroscopy and intrinsic fluorescence spectroscopy,” J. Biomed. Opt. 13, 024012 (2008). 10.1117/1.2909672 [DOI] [PubMed] [Google Scholar]
  • 12.Robichaux A.et al. , “In vivo detection of cervical dysplasia with near-infrared Raman spectroscopy,” Proceedings of PSIE 4614, Biomedical Vibrational Spectroscopy II, 145, March 28, 2002.
  • 13.Georgakoudi I.et al. , “Fluorescence, reflectance, and light-scattering spectroscopy for evaluating dysplasia in patients with Barrett's esophagus,” Gastroenterology 120(7), 1620–1629 (2001). 10.1053/gast.2001.24842 [DOI] [PubMed] [Google Scholar]
  • 14.Lovat L. B.et al. , “Elastic scattering spectroscopy accurately detects high grade dysplasia and cancer in Barrett's oesophagus,” Gut 55(8), 1078–1083 (2006). 10.1136/gut.2005.081497 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Stone N.et al. , “Near-infrared Raman spectroscopy for the classification of epithelial pre-cancers and cancers,” J. Raman Spectrosc. 33(7), 564–573 (2002). 10.1002/jrs.882 [DOI] [Google Scholar]
  • 16.Dhar A.et al. , “Elastic scattering spectroscopy for the diagnosis of colonic lesions: Initial results of a novel optical biopsy technique,” Gastrointestinal Endosc. 63(2), 257–261 (2006). 10.1016/j.gie.2005.07.026 [DOI] [PubMed] [Google Scholar]
  • 17.Lin W. C.et al. , “Diffuse reflectance spectroscopy for in vivo pediatric brain tumor detection,” J. Biomed. Opt. 15, 061709 (2010). 10.1117/1.3505012 [DOI] [PubMed] [Google Scholar]
  • 18.Bard M. P. L.et al. , “Optical spectroscopy for the classification of malignant lesions of the bronchial tree,” Chest 129(4), 995–1001 (2006). 10.1378/chest.129.4.995 [DOI] [PubMed] [Google Scholar]
  • 19.Huang Z.et al. , “Near-infrared Raman spectroscopy for optical diagnosis of lung cancer,” Int. J. Cancer 107(6), 1047–1052 (2003). 10.1002/ijc.11500 [DOI] [PubMed] [Google Scholar]
  • 20.Lieber C. A.et al. , “Diagnostic tool for early detection of ovarian cancers using Raman spectroscopy,” Biomedical Spectroscopy: Vibrational Spectroscopy and Other Novel Techniques, Proc. SPIE 3918, 129 (2000). 10.1117/12.384934 [DOI] [Google Scholar]
  • 21.Koenig F.et al. , “Spectroscopic measurement of diffuse reflectance for enhanced detection of bladder carcinoma,” Urology 51(2), 342–345 (1998). 10.1016/S0090-4295(97)00612-2 [DOI] [PubMed] [Google Scholar]
  • 22.Parkin D. M.et al. , “Global cancer statistics, 2002,” Ca-Cancer J. Clin. 55(2), 74–108 (2005). 10.3322/canjclin.55.2.74 [DOI] [PubMed] [Google Scholar]
  • 23.Monheit G.et al. , “The performance of MelaFind: A prospective multicenter study,” Arch. Dermatol. 147(2), 188 (2011). 10.1001/archdermatol.2010.302 [DOI] [PubMed] [Google Scholar]
  • 24.Rigel D. S., Russak J., and Friedman R., “The evolution of melanoma diagnosis: 25 years beyond the ABCDs,” CA: Cancer J. Clin. 60(5), 301–316 (2010). 10.3322/caac.20074 [DOI] [PubMed] [Google Scholar]
  • 25.Kimball A. B. and Resneck J. S. Jr., “The US dermatology workforce: A specialty remains in shortage,” J. Am. Acad. Dermatol. 59(5), 741–745 (2008). 10.1016/j.jaad.2008.06.037 [DOI] [PubMed] [Google Scholar]
  • 26.Pavlova I.et al. , “Microanatomical and biochemical origins of normal and precancerous cervical autofluorescence using laser-scanning fluorescence confocal microscopy,” Photochem. Photobiol. 77(5), 550–555 (2003). [DOI] [PubMed] [Google Scholar]
  • 27.Drezek R.et al. , “Understanding the contributions of NADH and collagen to cervical tissue fluorescence spectra: Modeling, measurements, and implications,” J. Biomed. Opt. 6, 385 (2001). 10.1117/1.1413209 [DOI] [PubMed] [Google Scholar]
  • 28.Pavlova I.et al. , “Understanding the biological basis of autofluorescence imaging for oral cancer detection: High-resolution fluorescence microscopy in viable tissue,” Clin. Cancer Res. 14(8), 2396–2404 (2008). 10.1158/1078-0432.CCR-07-1609 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Georgakoudi I.et al. , “NAD (P) H and collagen as in vivo quantitative fluorescent biomarkers of epithelial precancerous changes,” Cancer Res. 62(3), 682 (2002). [PubMed] [Google Scholar]
  • 30.Borisova E.et al. , “Diagnostics of pigmented skin tumors based on laser-induced autofluorescence and diffuse reflectance spectroscopy,” Quantum Electron. 38, 597 (2008). 10.1070/QE2008v038n06ABEH013891 [DOI] [Google Scholar]
  • 31.Rajaram N.et al. , “Pilot clinical study for quantitative spectral diagnosis of non-melanoma skin cancer,” Lasers Surg. Med. 42(10), 716–727 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Scepanovic O. R.et al. , “A multimodal spectroscopy system for real-time disease diagnosis,” Rev. Sci. Instrum. 80(4), 043103 (2009). 10.1063/1.3117832 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Scepanovic O. R.et al. , “Multimodal spectroscopy detects features of vulnerable atherosclerotic plaque,” J. Biomed. Opt. 16(1), 1–10 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Utzinger U. and Richards-Kortum R. R., “Fiber optic probes for biomedical optical spectroscopy,” J. Biomed. Opt. 8, 121 (2003). 10.1117/1.1528207 [DOI] [PubMed] [Google Scholar]
  • 35.Zuluaga A. F.et al. , “Fluorescence excitation emission matrices of human tissue: A system for in vivo measurement and method of data analysis,” Appl. Spectrosc. 53(3), 302–311 (1999). 10.1366/0003702991946695 [DOI] [Google Scholar]
  • 36.Bigio I. J. and Mourant J. R., “Ultraviolet and visible spectroscopies for tissue diagnostics: Fluorescence spectroscopy and elastic-scattering spectroscopy,” Phys. Med. Biol. 42(5), 803 (1997). 10.1088/0031-9155/42/5/005 [DOI] [PubMed] [Google Scholar]
  • 37.Motz J. T.et al. , “Optical fiber probe for biomedical Raman spectroscopy,” Appl. Opt. 43(3), 542–554 (2004). 10.1364/AO.43.000542 [DOI] [PubMed] [Google Scholar]
  • 38.Mahadevan-Jansen A.et al. , “Development of a fiber optic probe to measure NIR Raman spectra of cervical tissue in vivo,” Photochem. Photobiol. 68(3), 427–431 (1998). 10.1111/j.1751-1097.1998.tb09703.x [DOI] [PubMed] [Google Scholar]
  • 39. ANSI, Z136.1-1993, American National Standard for the Safe Use of Lasers, 1993.
  • 40.Farrell T. J., Patterson M. S., and Wilson B., “A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,” Med. Phys. 19, 879 (1992). 10.1118/1.596777 [DOI] [PubMed] [Google Scholar]
  • 41.Kienle A. and Patterson M. S., “Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium,” J. Opt. Soc. Am. A 14(1), 246–254 (1997). 10.1364/JOSAA.14.000246 [DOI] [PubMed] [Google Scholar]
  • 42.Graaff R.et al. , “Condensed Monte Carlo simulations for the description of light transport,” Appl. Opt. 32(4), 426–434 (1993). 10.1364/AO.32.000426 [DOI] [PubMed] [Google Scholar]
  • 43.Nichols B. S., Rajaram N., and Tunnell J. W., “Performance of a lookup table-based approach for measuring tissue optical properties with diffuse optical spectroscopy,” J. Biomed. Opt. 17(5), 057001 (2012). 10.1117/1.JBO.17.5.057001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Rajaram N., Nguyen T. H., and Tunnell J. W., “Lookup table-based inverse model for determining optical properties of turbid media,” J. Biomed. Opt. 13, 050501 (2008). 10.1117/1.2981797 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Rajaram N.et al. , “Design and validation of a clinical instrument for spectral diagnosis of cutaneous malignancy,” Appl. Opt. 49(2), 142–152 (2010). 10.1364/AO.49.000142 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Van Veen R., Verkruysse W., and Sterenborg H., “Diffuse-reflectance spectroscopy from 500 to 1060 nm by correction for inhomogeneously distributed absorbers,” Opt. Lett. 27(4), 246–248 (2002). 10.1364/OL.27.000246 [DOI] [PubMed] [Google Scholar]
  • 47.Zhang Q.et al. , “Turbidity-free fluorescence spectroscopy of biological tissue,” Opt. Lett. 25(19), 1451–1453 (2000). 10.1364/OL.25.001451 [DOI] [PubMed] [Google Scholar]
  • 48.Vo-Dinh T., “Biomedical photonics handbook,” J. Biomed. Opt. 9, 1110 (2004). 10.1117/1.1776177 [DOI] [Google Scholar]
  • 49.Zhao J.et al. , “Real time raman spectroscopy for noninvasive in vivo skin analysis and diagnosis,” in New Developments in Biomedical Engineering (InTech, 2010). [Google Scholar]
  • 50.Nijssen A.et al. , “Discriminating basal cell carcinoma from its surrounding tissue by Raman spectroscopy,” J. Invest. Dermatol. 119(1), 64–69 (2002). 10.1046/j.1523-1747.2002.01807.x [DOI] [PubMed] [Google Scholar]
  • 51.Cartaxo S. B.et al. , “FT-Raman spectroscopy for the differentiation between cutaneous melanoma and pigmented nevus,” Acta Cirurg. Bras. 25(4), 351–356 (2010). [DOI] [PubMed] [Google Scholar]
  • 52.Gniadecka M.et al. , “Melanoma diagnosis by Raman spectroscopy and neural networks: Structure alterations in proteins and lipids in intact cancer tissue,” J. Invest. Dermatol. 122(2), 443–449 (2004). 10.1046/j.0022-202X.2004.22208.x [DOI] [PubMed] [Google Scholar]
  • 53.Gniadecka M.et al. , “Diagnosis of basal cell carcinoma by Raman spectroscopy,” J. Raman Spectrosc. 28(23), 125–129 (1997). [DOI] [Google Scholar]
  • 54.Huang Z.et al. , “Raman spectroscopy of in vivo cutaneous melanin,” J. Biomed. Opt. 9, 1198 (2004). 10.1117/1.1805553 [DOI] [PubMed] [Google Scholar]
  • 55.Gniadecka M., Nielsen O. F., and Wulf H., “Water content and structure in malignant and benign skin tumours,” J. Mol. Struct. 661–662, 405–410 (2003). 10.1016/j.molstruc.2003.08.030 [DOI] [Google Scholar]
  • 56.Zhao J.et al. , “Real-time Raman spectroscopy for non-invasive skin cancer detection-preliminary results,” Conf. Proc. IEEE Eng. Med. Biol. Soc. 2008, 3107–3109 (2008). [DOI] [PubMed] [Google Scholar]

Articles from The Review of Scientific Instruments are provided here courtesy of American Institute of Physics

RESOURCES