Abstract
Purpose
In this work we present a dual-phase diffusion tensor imaging (DTI) technique that incorporates a correction scheme for the cardiac material strain, based on 3D myocardial tagging.
Methods
In vivo dual-phase cardiac DTI with a stimulated echo approach and 3D tagging was performed in 10 healthy volunteers. The time course of material strain was estimated from the tagging data and used to correct for strain effects in the diffusion weighted acquisition. Mean diffusivity, fractional anisotropy, helix, transverse and sheet angles were calculated and compared between systole and diastole, with and without strain correction. Data acquired at the systolic sweet spot, where the effects of strain are eliminated, served as a reference.
Results
The impact of strain correction on helix angle was small. However, large differences were observed in the transverse and sheet angle values, with and without strain correction. The standard deviation of systolic transverse angles was significantly reduced from 35.9±3.9° to 27.8°±3.5° (p<0.001) upon strain-correction indicating more coherent fiber tracks after correction. Myocyte aggregate structure was aligned more longitudinally in systole compared to diastole as reflected by an increased transmural range of helix angles (71.8°±3.9° systole vs. 55.6°±5.6°, p<0.001 diastole). While diastolic sheet angle histograms had dominant counts at high sheet angle values, systolic histograms showed lower sheet angle values indicating a reorientation of myocyte sheets during contraction.
Conclusion
An approach for dual-phase cardiac DTI with correction for material strain has been successfully implemented. This technique allows assessing dynamic changes in myofiber architecture between systole and diastole, and emphasizes the need for strain correction when sheet architecture in the heart is imaged with a stimulated echo approach.
Introduction
The influence of myocardial fiber architecture on cardiac morphology and mechanics is of significant interest. The helical organization of myocardial fibers [1]–[4] and the formation of myocytes into sheets [5]–[7] by branching and interconnection have been well described. While microscopy provides high-resolution images [8]–[10], MR diffusion weighted imaging enables investigation of the intact organ. Despite its lower spatial resolution compared to microscopy, MR has found application in ex vivo [4], [6], [11]–[20] and also in a small number of in vivo [21]–[27] studies of the heart. More recently, diffusion tensor imaging (DTI) has enabled tractography of the myocardium both ex vivo [28]–[30] and in vivo [25], [31], [32].
In order to investigate differences in fiber configuration during the cardiac cycle, Chen et al. [33] presented a comparison of ex vivo pig hearts first arrested in diastole and later fixated in systole. These ex vivo findings were confirmed by Hales et al. [34]. Significant differences in helix angle and sheet angle distributions between systole and diastole were found. Further investigations involved the study of sheet rearrangement in myocardial pathologies [35], [36].
In vivo DTI of the human and animal heart has been performed using diffusion weighted stimulated echo acquisition modes (STEAM) [21], [27], [37]–[40] and first and second order motion compensated diffusion weighted spin-echoes [24]–[26], [31], [32] in combination with echo-planar imaging readouts. While spin-echo diffusion weighted imaging requires strong gradient systems in order to be applied in the in vivo heart, STEAM based sequence can be performed with standard gradient hardware.
Early reports suggested that the impact of material strain is of significance in cardiac DTI [21], [22], [38], [39]. To this end, it was proposed to trigger STEAM encoding and decoding to the so-called “sweet spots” of myocardial strain at which the temporal mean of strain approaches zero [21]. Despite this insight, cardiac diffusion-weighted STEAM at various time points in the cardiac cycle has been reported recently [27], [41]–[44].
The objective of the present work is to address the impact of material strain on the diffusion tensor when imaging the in-vivo heart using the STEAM sequence. To this end, a tensor correction scheme based on cine 3D tagging data is presented. In addition, the diffusion weighted STEAM sequence is modified to allow for dual-phase and slice-interleaved imaging thereby accelerating cardiac diffusion tensor imaging by a factor of two relative to previous single-phase approaches. Differences in fiber and sheet architecture between systole and diastole, without and with strain correction of the in-vivo human heart are presented.
Methods
Study protocol
Ten subjects without any history of cardiac disease (4 male/6 female, age 27±8years, weight 68±7 kg, heart rate 66±11 bpm) were imaged on a clinical 1.5T scanner (Achieva system, Philips Healthcare, Best, The Netherlands). The scanner was equipped with a gradient system delivering 40 mT/m maximum strength and 200 mT/m/ms maximum slew rate per physical gradient axis. A 5-channel cardiac array coil was used for signal detection.
Written informed consent was obtained from each subject prior to imaging, and the study protocol was approved by the ethics committee of the canton of Zurich. Obtained informed consent included imaging as well as publication of anonymized data.
Short-axis balanced steady state free precession cine data with a temporal resolution of 7 ms were obtained in the 2-chamber and short-axis planes of the left ventricle (LV) to identify the ventricular systolic (trigger delay: 277±19 ms) and diastolic (trigger delay: 627±85 ms) standstill periods. For image-based shimming, a B0 field map was acquired in the short-axis view covering the entire LV [45].
Myocardial tagging sequence
Three orthogonally orientated line tagged cine image volumes covering the entire LV were acquired within three consecutive breath holds [46]. To compensate for differences in breath hold levels a gating respiratory navigator was applied prior to the acquisition of each stack (acceptance window 15 mm). Resulting navigator offsets were used for stack alignment during image reconstruction. In order to avoid tag line fading during the cardiac cycle, complementary spatial modulation of magnetization (CSPAMM) was applied requiring two signal averages, with inverted tagging modulation [47]. Imaging parameters were as follows: FOV: 108×108×108 mm3, spatial resolution: 3.5×7.7×7.7 mm3 (tagging/readout × phase encoding × phase encoding) reconstructed to 0.96×0.96×0.96 mm3, temporal resolution: 18 ms and 7 mm tag line distance. To achieve a temporal resolution of 18 ms, the maximum slew rate of the gradient system (200 mT/m/ms) was used for the segmented echo planar imaging-readout (EPI factor 7, 3 excitations per heart phase).
Diffusion sequence
Dual-phase cardiac STEAM was implemented using a reduced field-of-view technique [48]. To avoid saturation in adjacent slices a tilted local-look pulse scheme was incorporated (Figure 1) [49]. Residual signal from the edge of the field-of-view (black triangles in Figure 1c) was suppressed using regional saturation (rest) slabs. A scheme to interleave slices (SL1 and SL2) and heart phases (SYS and DIA) was implemented by applying the STEAM encoding block of slice 1 in systole (SL1 SYS) and slice 2 in diastole (SL2 DIA) of the first R-R interval. The corresponding STEAM decoding block including the readout was applied in the second R-R interval, respectively. The two paired slices had a gap of 25 mm, thus avoiding cross talk from the angulated excitation due to contraction of the heart. After acquiring all signal averages of all diffusion encoding directions, the slice order was switched to complete the acquisition of both heart phases for each slice. Thereby two slices were acquired in two heart phases within a single scan, hence reducing scan time relative to sequential single-phase, single-slice acquisition by a factor of two.
The acquisition of diffusion-weighted images was divided into multiple navigator-gated breath holds (acceptance window of 5 mm). Parameters of the diffusion sequence were: 224×100 mm2 field-of-view, 2×2 mm2 in-plane resolution, 8 mm slice thickness, number of slices 6, TE/TR 18 ms/2R-R intervals, partial Fourier factor 0.62. A single shot EPI readout was used. The two heart phases in the diffusion protocol were triggered to quiescent phases of systole and mid-diastole as defined on the cine images. Unipolar diffusion gradients were played out in 10 directions on a unit-sphere [50] with a b-value of 500 s/mm2. To reduce echo time, FID crushers necessary in STEAM were removed for b = 500 s/mm2 acquisitions, but kept for the b = 0 s/mm2 acquisition.
All eight signal averages of a diffusion weighted image were acquired within one breath hold. Different diffusion weightings were obtained in consecutive breath holds. The volunteers were allowed sufficient time to recover in-between breath holds to ensure consistent heart rates among the data series. A total of 11 breath holds of 14–16 s duration each were acquired per slice and two heart phases, resulting in a 15–18 min total net acquisition time for six slices at two heart phases.
Tensor reconstruction
To compensate for residual slice mismatch due to inconsistency in breath hold levels within the 5 mm gating window and to account for eddy-current induced geometrical distortions, all diffusion weighted images were registered to the b = 0 image by means of affine image registration (elastix toolbox [51]).
Systolic and diastolic diffusion tensors were estimated based on the modified Stejskal-Tanner equations. To account for non-zero diffusion weighting of the “b = 0 s/mm2” scan due to diffusion weighting introduced by the FID crusher gradients present in STEAM, the signal equation was modified as:
with B being the modified b-matrix containing the b-values of the diffusion weighted images b and the “b = 0” image :
denotes the negative logarithm of the measured signal per pixel including the signal of the “b = 0” image :
and the vector containing the unknown tensor elements and the true b = 0 signal :
where x, y and z are the coordinates of the normalized diffusion direction, T the transpose and † the Moore-Penrose pseudo inverse.
Correction for material strain
Material strain effects were compensated for based on the stretch history of tissue [38], [39]. From 3D tagging data, three-dimensional displacement fields were calculated with a custom-made software utilizing the 3D SinMod algorithm [52]. The LV was manually masked on the tagging data as well as on the acquired b = 0 s/mm2 image. To compensate for mismatch and different spatial resolution of the acquired data, the shape of the DTI mask was mapped onto the re-sliced tagging mask by means of coherent point drift registration [53]. Having identified the position of each diffusion tensor estimated from the DTI acquisition within the displacement field, a cube parallel to the canonical basis in which the displacement fields and the position of the tensors are represented, was defined at each point and tracked over the R-R interval. The time course of the right stretch tensor was calculated from the deformation gradient field obtained from the tracked cubes as described by Hess et al. [54]. Stretch tensors were calculated relative to the systolic and the diastolic time points of diffusion imaging. The strain effect on the diffusion measurement is described by
with U(τ) being the time course of the right stretch tensor and Δ the duration of the R-R interval. In accordance with [39] the equation is expanded as:
For validation purposes, additional DTI data were acquired during the systolic “sweet-spot” (trigger delay: 160 ms) in one of the volunteers and compared to data in end systole (trigger delay: 305 ms) and diastole (trigger delay: 620 ms).
Data analysis
From the six slices pairs of two were grouped for the basal, medial and apical level and mean diffusivity (MD) and fractional anisotropy (FA) in systole and diastole were compared for each volunteer. Additionally, helix, transverse and sheet angles were calculated using projections of the first and third eigenvectors as described in Figure 2 (a–b) [3]. To allow tracking the transmural course of the helix angle, a local anatomical basis was defined. To this end, the shape of the LV was mapped onto an “ideal” circular ring by means of coherent point drift mapping [53]. This procedure allowed the definition of a locally normalized transmural position independent of variation of local thickness of the myocardium (Figure 2c). The helix angle alignment was analyzed on a slice-by-slice basis. Therefore the myocardium was separated into five layers: epicardial, sub-epicardial, mid-wall, sub-endocardial and endocardial similarly to previously reported helix angle analysis [27]. The transmural helix gradient from the linear fit as well as the range of the transmural course of the helix angle are reported for basal, medial and apical levels (Figure 2d). Transverse and sheet angle distributions are analyzed by means of histograms for each slice similar to the analysis of Hales et al. [34]. The standard deviation of the transverse angle distribution is reported as measure of coherence. Sheet angle distributions were fitted with a quadratic function (Figure 2e) and the coefficient of the quadratic component is presented as a measure of sheet realignment during systolic contraction. Additionally the mean of the absolute value of the sheet angle was calculated for basal, medial and apical levels in systole and diastole.
The tensors shown in this study have not been interpolated as in previous works [25], [31]. Statistical differences between systolic and diastolic values were tested using a two-tailed paired student's t-test. A p-value<0.05 was considered statistically significant. All tests were Bonferroni-corrected for multiple testing.
Results
Dual-phase cardiac DTI data was successfully acquired in all subjects. Total exam time including subject preparation was 1.5–2 hours.
The raw data images acquired at the systolic sweet spot, in peak systole and in diastole are presented in Figure 3. Data are shown for the “b = 0”, the first three and the last diffusion- encoding direction as well as the average of all diffusion directions. Tagging data from the three orthogonally line-tagged stacks are given alongside. The temporally averaged stretch tensors as calculated from the tagging data allow radial, circumferential and longitudinal stretch components, which are presented as stretch maps, to be assessed.
Figure 4a,b shows the time course of the radial, circumferential and longitudinal stretch calculated from the right stretch tensors at the apical, medial and basal levels for systole and diastole. The time points of acquisition of systolic and diastolic DTI data as well as the systolic sweet spot are indicated by vertical lines in Figure 4a,b. The transmural course of the helix angles and transverse and sheet angle histograms in systole (trigger delay: 305 ms) and diastole (trigger delay: 620 ms) (without and with strain correction) are shown for the medial level in Figure 4c,d. In addition, DTI data acquired in the systolic sweet spot (trigger delay: 160 ms) is overlaid. Strain correction results in changes in the transverse and sheet angle distributions in systole. Values obtained upon strain correction approach data acquired in the sweet spot. Differences in diastole are found to be smaller with and without strain correction.
In Figure 5 systolic and diastolic tensor fields at a medial level with and without strain correction are compared. The superquadric representation of the diffusion tensor [15] was employed while glyphs were color-coded by the helix angle. It is observed that systolic diffusion tensors have been rearranged into the natural helical alignment after strain correction. In diastole, however, correction effects were subtler, mainly illustrated by small changes in the main diffusivities.
MD and FA for both heart phases, with and without strain correction, are reported in Table 1. After strain correction, the MD was increased in systole and decreased in diastole, both with statistical significance. The FA was significantly increased upon strain correction in systole, but remained unchanged in diastole.
Table 1. MD, FA at basal, medial and apical level.
systole | diastole | ||||
w/o correction | with correction | w/o correction | with correction | ||
MD [10−4 mm2/s] | base | 8.6±1.2 * | 9.5±1.3 | 8.5±1.0 * | 8.2±1.0 † |
medial | 8.8±1.4 * | 10.1±1.8 | 9.2±1.1 * | 8.7±1.1 † | |
apex | 9.6±0.8 * | 11.2±1.2 | 10.1±1.1 * | 9.4±0.9 † | |
FA | base | 0.52±0.03 * | 0.61±0.02 | 0.61±0.05 | 0.61±0.04 |
medial | 0.52±0.05 * | 0.60±0.03 | 0.61±0.04 | 0.61±0.04 | |
apex | 0.48±0.02 * | 0.55±±0.02 | 0.57±0.03 | 0.57±0.03 † |
* indicates statistical significance (p-value<0.05) between uncorrected and corrected data and
indicates statistical significance between systole and diastole.
Figure 6 displays helix angle maps and the dependency of helix angle on the transmural depth for systole and diastole, with and without strain correction. Data are given as mean ± one standard deviation across the study population. While in diastole only little change in helix angles is observed upon strain correction, helix angles at basal level are significantly different with strain correction in systole. The mean transmural helix angle range in diastole across the volunteers was reduced by 2.2±4.4° at the basal level and increased by 2.2±6.1° at the medial level and 1.0±5.0° at the apical level after strain correction. For systole, the transmural helix angle range was decreased by 9.4±9.9° at the basal level, 1.6±5.3° at the medial level and 6.7±9.9° at the apical level after strain correction. Differences in diastole were mostly not statistically significant. In systole, statistically significant differences at the medial and basal levels were found when comparing data without and with strain correction.
Figure 7 shows transverse angle maps and histograms of the study population. Significant differences were observed before and after strain correction for both systole and diastole, at each cardiac levels. Negative transverse angles were found at the posterior RV-LV intersection. The distribution of the systolic transverse angle has a lower variance after strain correction suggesting a more coherent fiber track. It is noted that transverse angle distributions in systole and diastole show a similar variance after strain correction.
Sheet angle maps and histograms are shown in Figure 8. The characteristic distribution of sheet angle into two populations is well seen, particularly with strain correction. Strain correction is observed to change systolic sheet distributions markedly, in some cases producing almost inverted distributions of those obtained without correction.
A comparison of helix, transverse and sheet angles for the systolic and diastolic heart phase with and without strain-correction is provided in Table 2. Significant differences in helix angle range between systole and diastole were seen in the medial and apical levels while differences for transverse and sheet angles between systole and diastole were found at the medial and basal levels of the heart. Sheet angle histograms are broadened in systole compared to diastole.
Table 2. Angulation analysis.
systole | diastole | |||||
w/o correction | with correction | w/o correction | with correction | |||
helix angle | range | base | 72.8°±15.4° * | 63.9°±11.5° | 55.3°±9.3° | 53.0°±9.6° |
medial | 77.1°±7.7° | 75.2°±6.2° | 55.9°±3.7° | 58.1°±6.1° † | ||
apex | 81.1°±12.3° | 76.0°±10.6° | 55.0°±11.9° | 56.0°±13.0° † | ||
gradient [ ° /%depth] | base | −0.93±0.18 * | −0.83±0.13 | −0.74±0.11 | −0.72±0.11 | |
medial | −1.00±0.11 | −0.98±0.10 | −0.73±0.06 | −0.75±0.07 † | ||
apex | −1.06±0.15 | −1.00±0.12 | −0.71±0.14 | −0.71±0.16 † | ||
transverse angle | mean | base | −3.0°±4.5° * | 3.5°±4.8° | −0.9°±2.3° | −1.8°±2.3° † |
medial | 0.2°±4.3° | −1.7°±3.1° | −0.4°±1.1° | −0.4°±1.1° | ||
apex | 0.2°±3.05° * | −6.3±5.1° | 1.3°±2.3° | 1.3°±2.3° † | ||
SD | base | 36.0°±3.8° * | 29.5°±5.1° | 27.7°±3.6° * | 30.2°±3.4° | |
medial | 34.0°±5.5° * | 24.6°±4.1° | 24.4°±3.1° * | 27.2°±3.5° | ||
apex | 35.8°±3.4° * | 27.3°±2.4° | 25.2°±4.3° * | 28.9°±3.8° | ||
sheet angle | quadratic fit [×10−6] | base | −5.33±3.9 * | 3.7±4.2 | 14.9±4.9 * | 10.6±3.9 † |
medial | −1.44±4.7 * | 9.3±4.5 | 18.8±3.7 * | 14.0±4.3 | ||
apex | 3.5±5.5 * | 12.6±5.7 | 19.2±4.8 * | 14.2±6.7 | ||
mean |γ| | base | 38.0°±4.3° * | 50.4°±3.5° | 61.2°±5.1° * | 56.9°±4.2° † | |
medial | 41.6°±3.7° * | 55.1°±4.8° | 65.0°±3.6° * | 60.4°±4.3° † | ||
apex | 58.7°±5.9° * | 60.5°±4.1° | 66.0°±4.6° * | 60.9°±6.6° |
Helix transverse and sheet angle analysis at basal, medial and apical level is shown prior and after strain correction.
* indicates statistical significance (p-value <0.05) between uncorrected and corrected data and
indicates statistical significance between systole and diastole.
Discussion
In this study, dual-heart phase cardiac DTI with strain correction was successfully implemented and applied on 10 healthy volunteers to study differences in myofiber architecture between systole and diastole.
The slice and phase interleaving scheme permitted a reduction in scan time by a factor of two relative to a single-phase DTI protocol, which would need to be repeated in systole and diastole. Given that angular diffusion resolution was encoded in separate breath held scans with the current implementation, the number of breath holds required per dual-slice set is dictated by the number of diffusion directions. While breath hold durations were short (14–16 sec), free-breathing acquisition is nevertheless preferred to increase acceptance in practice. To this end, respiratory navigation in conjunction with patient feedback could be incorporated into our approach [27], or alternatively a modified respiratory navigation scheme to increase gating efficiency without the need for patient feedback [40] may be applied.
Local-look excitation was used to reduce the field-of-view in phase-encode direction by a factor of 2.5 to 3 depending on slice angulation and patient size. Alternatively, undersampling strategies may be employed including parallel imaging [55]–[57] or compressed sensing [58], [59]. Further reduction of scan time could also be achieved by combining the proposed method with simultaneous excitation of multiple slices and subsequent unfolding using parallel imaging principles [60]–[62].
DTI of the heart with the stimulated echo approach has previously been performed at sweet spots in the cardiac cycle, where the effects of strain are eliminated [21], [22]. The exact locations of these sweet spots is a function of the heart rate of each volunteer, but generally falls within mid-systole and mid-diastole. More recently, DTI of the myocardium has been described at end-systole, where the heart reaches a quiescent or stand-still phase [27], [41], [44]. With the present work, it has, however, demonstrated that DTI of the myocardium at end-systole is significantly influenced by strain. The effect of strain on helix angle measurements is small but its impact on measures of sheet architecture, such as sheet angle, is extremely large. Our results confirm those of Tseng and colleagues, who likewise showed that imaging away from the systolic/diastolic sweet spots produced small differences in helix angle but very large differences in sheet angle [21], [22].
The validity of strain correction was verified by comparing data acquired in systole and diastole to data obtained in the systolic sweet spot, for which actual material strain equals the average strain across the cardiac cycle. While no major difference was found between the diastolic data obtained without strain correction, and the sweet spot data, significant change was seen for systolic data upon strain correction.
In systole, longitudinal and circumferential diffusion components are underestimated while radial diffusion components are overestimated without strain correction. This effect leads to stretching of the diffusion tensor in radial direction and compression in the two orthogonal directions. Consequently, after estimating the tensors' eigenbasis, the second and third eigenvectors are swapped leading to higher bin counts for the sheet angle around 0°. The first eigenvector, which is predominantly aligned within the circumferential and longitudinal plane, is rotated out of plane. Histograms of the transverse angle demonstrate a wider spread prior to strain correction. Without strain correction, the change in sheet angulation between systole and diastole is significantly overestimated and fiber tracks appear less coherent. Here, the need for strain correction in systole was clearly demonstrated.
The changes in fiber configuration between systole and diastole seen in this study indicate a greater longitudinal alignment of myofibers during contraction. Similar results have been described in excised rat hearts arrested in systole and diastole [33]. Likewise, histological and MR findings from excised porcine [63] and goat [12] heart revealed that the helical fiber structure from epicardium at the apex crosses to endocardium at a medial level and back to epicedium at the base, hypothesizing that the presence of non-zero transverse angles are responsible for wall thickening during contraction [63]. These findings are in agreement with the data reported here.
In the present work sheet angle histograms were generated for basal, medial and apical levels similar to work by Hales et al. [34]. The in vivo results presented here show a significant change in sheet angle from systole to diastole. In the contracted state, fewer counts of larger angulation were found, while the counts of intermediate angles were increased. These results are in accordance with those of Dou et al. al [22], who showed that sheet orientation becomes more radial in systole. Similar to prior reports on isolated hearts [33]–[35], the changes in sheet angle histograms were most pronounced at the basal level and less at the apical level.
A potential study limitation lies in the intrinsic coupling of the b-value of the STEAM approach with the subject's heart rate. While the standard deviation of heart rate between acquisitions was only 2.6±1.1 bpm, the heart rate during a breath hold maneuver may have changed significantly. To minimize the impact of heart rate variation on the tensor directionality, all averages of a single diffusion encoding direction were acquired within a single breath hold.
Besides material strain, base SNR is of critical importance regarding systematic errors in determining apparent diffusion. Magnitude averaging of the low SNR DWI data was performed resulting in a Rician noise distribution of the averaged data. Accordingly, signal attenuation by diffusion is biased by the SNR dependent noise floor [64]. Since the SNR at the apex is considerably higher as compared to the base of the heart due to its proximity to the receive coil array, the relative underestimation of apparent diffusion measured in the apex is less compared to the value at the basal level.
Conclusion
An approach for dual-phase cardiac DTI with correction for myocardial strain has been successfully implemented and has allowed changes in myofiber architecture between systole and diastole to be studied in the human heart in vivo. The results obtained with strain correction are in agreement with experimental ex vivo data and prior in vivo data in healthy volunteers. The potential of DTI to characterize myocardial anatomy in the heart is high, but strain correction at phases other than the sweet spots will be crucial for the accurate characterization of myocyte architecture.
Acknowledgments
The authors thank Dr. Gérard Crelier for providing the processing software for 3D tagged data.
Data Availability
The authors confirm that all data underlying the findings are fully available without restriction. Data used in this study will be made available via ETH Institutional Data Access. The link will be sent upon email request by Dr. Roger Lüchinger: luechinger@biomed.ee.ethz.ch or the corresponding author: kozerke@biomed.ee.ethz.ch.
Funding Statement
This work was supported by the Swiss National Science Foundation, grant #CR3213_132671/1; UK EPSRC (EP/I018808/1), Adult Congenital Heart Disease Service GSTT and the National Institute for Health Research (NIHR) Biomedical Research Centre at Guy's and St Thomas' NHS Foundation Trust and King's College London; and the National Institutes of Health (5R01HL112831). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Streeter DD Jr, Spotnitz HM, Patel DP, Ross J Jr, Sonnenblick EH (1969) Fiber orientation in the canine left ventricle during diastole and systole. Circ Res 24: 339–347. [DOI] [PubMed] [Google Scholar]
- 2. Greenbaum RA, Ho SY, Gibson DG, Becker AE, Anderson RH (1981) Left ventricular fibre architecture in man. Br Heart J 45: 248–263. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Scollan DF, Holmes A, Winslow R, Forder J (1998) Histological validation of myocardial microstructure obtained from diffusion tensor magnetic resonance imaging. Am J Physiol 275: H2308–2318. [DOI] [PubMed] [Google Scholar]
- 4. Schmid P, Jaermann T, Boesiger P, Niederer PF, Lunkenheimer PP, et al. (2005) Ventricular myocardial architecture as visualised in postmortem swine hearts using magnetic resonance diffusion tensor imaging. Eur J Cardiothorac Surg 27: 468–472. [DOI] [PubMed] [Google Scholar]
- 5. LeGrice IJ, Smaill BH, Chai LZ, Edgar SG, Gavin JB, et al. (1995) Laminar structure of the heart: ventricular myocyte arrangement and connective tissue architecture in the dog. Am J Physiol 269: H571–582. [DOI] [PubMed] [Google Scholar]
- 6. Helm PA, Tseng HJ, Younes L, McVeigh ER, Winslow RL (2005) Ex vivo 3D diffusion tensor imaging and quantification of cardiac laminar structure. Magn Reson Med 54: 850–859. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Kung GL, Nguyen TC, Itoh A, Skare S, Ingels NB Jr, et al. (2011) The presence of two local myocardial sheet populations confirmed by diffusion tensor MRI and histological validation. J Magn Reson Imaging 34: 1080–1091. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Young AA, Legrice IJ, Young MA, Smaill BH (1998) Extended confocal microscopy of myocardial laminae and collagen network. J Microsc 192: 139–150. [DOI] [PubMed] [Google Scholar]
- 9. Costa KD, Takayama Y, McCulloch AD, Covell JW (1999) Laminar fiber architecture and three-dimensional systolic mechanics in canine ventricular myocardium. Am J Physiol 276: H595–607. [DOI] [PubMed] [Google Scholar]
- 10. Pope AJ, Sands GB, Smaill BH, LeGrice IJ (2008) Three-dimensional transmural organization of perimysial collagen in the heart. Am J Physiol Heart Circ Physiol 295: H1243–H1252. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Hsu EW, Muzikant AL, Matulevicius SA, Penland RC, Henriquez CS (1998) Magnetic resonance myocardial fiber-orientation mapping with direct histological correlation. Am J Physiol 274: H1627–1634. [DOI] [PubMed] [Google Scholar]
- 12. Geerts L, Bovendeerd P, Nicolay K, Arts T (2002) Characterization of the normal cardiac myofiber field in goat measured with MR-diffusion tensor imaging. Am J Physiol Heart Circ Physiol 283: H139–145. [DOI] [PubMed] [Google Scholar]
- 13. Tseng WY, Wedeen VJ, Reese TG, Smith RN, Halpern EF (2003) Diffusion tensor MRI of myocardial fibers and sheets: correspondence with visible cut-face texture. J Magn Reson Imaging 17: 31–42. [DOI] [PubMed] [Google Scholar]
- 14. Chen J, Song SK, Liu W, McLean M, Allen JS, et al. (2003) Remodeling of cardiac fiber structure after infarction in rats quantified with diffusion tensor MRI. Am J Physiol Heart Circ Physiol 285: H946–954. [DOI] [PubMed] [Google Scholar]
- 15. Ennis DB, Kindlman G, Rodriguez I, Helm PA, McVeigh ER (2005) Visualization of tensor fields using superquadric glyphs. Magn Reson Med 53: 169–176. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Wu EX, Wu Y, Nicholls JM, Wang J, Liao S, et al. (2007) MR diffusion tensor imaging study of postinfarct myocardium structural remodeling in a porcine model. Magn Reson Med 58: 687–695. [DOI] [PubMed] [Google Scholar]
- 17. Wu Y, Chan CW, Nicholls JM, Liao S, Tse HF, et al. (2009) MR study of the effect of infarct size and location on left ventricular functional and microstructural alterations in porcine models. J Magn Reson Imaging 29: 305–312. [DOI] [PubMed] [Google Scholar]
- 18. Strijkers GJ, Bouts A, Blankesteijn WM, Peeters TH, Vilanova A, et al. (2009) Diffusion tensor imaging of left ventricular remodeling in response to myocardial infarction in the mouse. NMR Biomed 22: 182–190. [DOI] [PubMed] [Google Scholar]
- 19. Hales PW, Burton RA, Bollensdorff C, Mason F, Bishop M, et al. (2011) Progressive changes in T(1), T(2) and left-ventricular histo-architecture in the fixed and embedded rat heart. NMR Biomed 24: 836–843. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Eggen MD, Swingen CM, Iaizzo PA (2012) Ex vivo diffusion tensor MRI of human hearts: relative effects of specimen decomposition. Magn Reson Med 67: 1703–1709. [DOI] [PubMed] [Google Scholar]
- 21. Tseng WY, Reese TG, Weisskoff RM, Wedeen VJ (1999) Cardiac diffusion tensor MRI in vivo without strain correction. Magn Reson Med 42: 393–403. [DOI] [PubMed] [Google Scholar]
- 22. Dou J, Reese TG, Tseng WY, Wedeen VJ (2002) Cardiac diffusion MRI without motion effects. Magn Reson Med 48: 105–114. [DOI] [PubMed] [Google Scholar]
- 23. Wu MT, Tseng WY, Su MY, Liu CP, Chiou KR, et al. (2006) Diffusion tensor magnetic resonance imaging mapping the fiber architecture remodeling in human myocardium after infarction: correlation with viability and wall motion. Circulation 114: 1036–1045. [DOI] [PubMed] [Google Scholar]
- 24. Gamper U, Boesiger P, Kozerke S (2007) Diffusion imaging of the in vivo heart using spin echoes-considerations on bulk motion sensitivity. Magn Reson Med 57: 331–337. [DOI] [PubMed] [Google Scholar]
- 25. Toussaint N, Sermesant M, Stoeck CT, Kozerke S, Batchelor PG (2010) In vivo human 3D cardiac fibre architecture: reconstruction using curvilinear interpolation of diffusion tensor images. Med Image Comput Comput Assist Interv 13: 418–425. [DOI] [PubMed] [Google Scholar]
- 26. Nguyen C, Fan Z, Sharif B, He Y, Dharmakumar R, et al. (2013) In vivo three-dimensional high resolution cardiac diffusion-weighted MRI: A motion compensated diffusion-prepared balanced steady-state free precession approach. Magn Reson Med [DOI] [PubMed] [Google Scholar]
- 27. Nielles-Vallespin S, Mekkaoui C, Gatehouse P, Reese TG, Keegan J, et al. (2013) In vivo diffusion tensor MRI of the human heart: Reproducibility of breath-hold and navigator-based approaches. Magn Reson Med 70: 454–465. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Sosnovik DE, Wang R, Dai G, Wang T, Aikawa E, et al. (2009) Diffusion spectrum MRI tractography reveals the presence of a complex network of residual myofibers in infarcted myocardium. Circ Cardiovasc Imaging 2: 206–212. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Mekkaoui C, Huang S, Chen HH, Dai G, Reese TG, et al. (2012) Fiber architecture in remodeled myocardium revealed with a quantitative diffusion CMR tractography framework and histological validation. J Cardiovasc Magn Reson 14: 70. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Mekkaoui C, Porayette P, Jackowski MP, Kostis WJ, Dai G, et al. (2013) Diffusion MRI tractography of the developing human fetal heart. PLoS One 8: e72795. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Toussaint N, Stoeck CT, Sermesant M, Schaeffter T, Kozerke S, et al. (2013) In vivo human cardiac fibre architecture estimation using shape-based diffusion tensor processing. Med Image Anal [DOI] [PubMed] [Google Scholar]
- 32. Sosnovik DE, Mekkaoui C, Huang S, Chen HH, Dai G, et al. (2014) Microstructural impact of ischemia and bone marrow-derived cell therapy revealed with diffusion tensor magnetic resonance imaging tractography of the heart in vivo. Circulation 129: 1731–1741. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Chen J, Liu W, Zhang H, Lacy L, Yang X, et al. (2005) Regional ventricular wall thickening reflects changes in cardiac fiber and sheet structure during contraction: quantification with diffusion tensor MRI. Am J Physiol Heart Circ Physiol 289: H1898–1907. [DOI] [PubMed] [Google Scholar]
- 34. Hales PW, Schneider JE, Burton RA, Wright BJ, Bollensdorff C, et al. (2012) Histo-anatomical structure of the living isolated rat heart in two contraction states assessed by diffusion tensor MRI. Prog Biophys Mol Biol 110: 319–330. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Cheng YJ, Lang D, Caruthers SD, Efimov IR, Chen J, et al. (2012) Focal but reversible diastolic sheet dysfunction reflects regional calcium mishandling in dystrophic mdx mouse hearts. Am J Physiol Heart Circ Physiol 303: H559–568. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Chen Y, Chen J, Caruthers S, Wickline SA (2013) Progressive Myocardial Sheet Dysfunction from 3 to 16 months in Duchene Muscular Dystrophy Mice (mdx) Defined by Diffusion Tensor MRI (DTI); Salt Lake City, Utah, USA. pp. 3077.
- 37. Edelman RR, Gaa J, Wedeen VJ, Loh E, Hare JM, et al. (1994) In vivo measurement of water diffusion in the human heart. Magn Reson Med 32: 423–428. [DOI] [PubMed] [Google Scholar]
- 38. Reese TG, Weisskoff RM, Smith RN, Rosen BR, Dinsmore RE, et al. (1995) Imaging myocardial fiber architecture in vivo with magnetic resonance. Magnetic Resonance in Medicine 34: 786–791. [DOI] [PubMed] [Google Scholar]
- 39. Reese TG, Wedeen VJ, Weisskoff RM (1996) Measuring diffusion in the presence of material strain. Journal of Magnetic Resonance, Series B 112: 253–258. [DOI] [PubMed] [Google Scholar]
- 40. Harmer J, Toussaint N, Pushparajah K, Stoeck CT, Chan RW, et al. (2013) In-vivo Diffusion Tensor Imaging of the Systemic Right Ventricle at 3T; Salt Lake City, Utah, USA. [Google Scholar]
- 41. McGill L-A, Ismail TF, Nielles-Vallespin S, Ferreira P, Scott AD, et al. (2012) Reproducibility of in-vivo diffusion tensor cardiovascular magnetic resonance in hypertrophic cardiomyopathy. J Cardiovasc Magn Reson 14: 86. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Nielles-Vallespin S, Ferreira PF, Gatehouse P, Keegan J, de Silva R, et al.. Time-resolved In Vivo Cardiac Diffusion Tensor MRI of the Human Heart; 2013; Salt Lake City, Utah, USA. pp. 0479.
- 43. Ferreira P, Kilner PJ, McGill L-A, Nielles-Vallespin S, Scott AD, et al. (2014) Aberrant myocardial sheetlet mobility in hypertrophic cardiomyopathy detected using in vivo cardiovascular magnetic resonance diffusion tensor imaging. Journal of Cardiovascular Magnetic Resonance 16: P338. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. McGill L-A, Ferreira P, Scott AD, Nielles-Vallespin S, Silva R, et al. (2014) Comparison of cardiac DTI parameters between systole and diastole. Journal of Cardiovascular Magnetic Resonance 16: P39. [Google Scholar]
- 45. Schar M, Kozerke S, Boesiger P (2003) Considerations on Shimming for Cardiac Applications at 1.5 and 3.0T; Toronto. [Google Scholar]
- 46. Rutz AK, Ryf S, Plein S, Boesiger P, Kozerke S (2008) Accelerated whole-heart 3D CSPAMM for myocardial motion quantification. Magn Reson Med 59: 755–763. [DOI] [PubMed] [Google Scholar]
- 47. Fischer SE, McKinnon GC, Maier SE, Boesiger P (1993) Improved myocardial tagging contrast. Magn Reson Med 30: 191–200. [DOI] [PubMed] [Google Scholar]
- 48. Feinberg DA, Hoenninger JC, Crooks LE, Kaufman L, Watts JC, et al. (1985) Inner volume MR imaging: technical concepts and their application. Radiology 156: 743–747. [DOI] [PubMed] [Google Scholar]
- 49. Reischauer C, Wilm BJ, Froehlich JM, Gutzeit A, Prikler L, et al. (2011) High-resolution diffusion tensor imaging of prostate cancer using a reduced FOV technique. Eur J Radiol 80: e34–41. [DOI] [PubMed] [Google Scholar]
- 50. Jones DK, Horsfield MA, Simmons A (1999) Optimal strategies for measuring diffusion in anisotropic systems by magnetic resonance imaging. Magn Reson Med 42: 515–525. [PubMed] [Google Scholar]
- 51. Klein S, Staring M, Murphy K, Viergever MA, Pluim JP (2010) elastix: a toolbox for intensity-based medical image registration. IEEE Trans Med Imaging 29: 196–205. [DOI] [PubMed] [Google Scholar]
- 52. Wang H, Stoeck CT, Kozerke S, Amini AA (2013) Analysis of 3D cardiac deformations with 3D SinMod. Conf Proc IEEE Eng Med Biol Soc 2013: 4386–4389. [DOI] [PubMed] [Google Scholar]
- 53. Myronenko A, Song X (2010) Point set registration: coherent point drift. IEEE Trans Pattern Anal Mach Intell 32: 2262–2275. [DOI] [PubMed] [Google Scholar]
- 54. Hess AT, Zhong X, Spottiswoode BS, Epstein FH, Meintjes EM (2009) Myocardial 3D strain calculation by combining cine displacement encoding with stimulated echoes (DENSE) and cine strain encoding (SENC) imaging. Magn Reson Med 62: 77–84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Bammer R, Keeling SL, Augustin M, Pruessmann KP, Wolf R, et al. (2001) Improved diffusion-weighted single-shot echo-planar imaging (EPI) in stroke using sensitivity encoding (SENSE). Magn Reson Med 46: 548–554. [DOI] [PubMed] [Google Scholar]
- 56. Bammer R, Auer M, Keeling SL, Augustin M, Stables LA, et al. (2002) Diffusion tensor imaging using single-shot SENSE-EPI. Magn Reson Med 48: 128–136. [DOI] [PubMed] [Google Scholar]
- 57. Alexander AL, Lee JE, Wu YC, Field AS (2006) Comparison of diffusion tensor imaging measurements at 3.0 T versus 1.5 T with and without parallel imaging. Neuroimaging Clin N Am 16: 299–309, xi. [DOI] [PubMed] [Google Scholar]
- 58. Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med 58: 1182–1195. [DOI] [PubMed] [Google Scholar]
- 59. Menzel MI, Tan ET, Khare K, Sperl JI, King KF, et al. (2011) Accelerated diffusion spectrum imaging in the human brain using compressed sensing. Magn Reson Med 66: 1226–1233. [DOI] [PubMed] [Google Scholar]
- 60. Breuer FA, Blaimer M, Heidemann RM, Mueller MF, Griswold MA, et al. (2005) Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging. Magn Reson Med 53: 684–691. [DOI] [PubMed] [Google Scholar]
- 61. von Deuster C, Stoeck CT, Giese D, Harmer J, Chan RW, et al. Concurrent dual-slice cardiac DTI of the in-vivo human heart; 2013; Salt Lake City, Utah, USA. [Google Scholar]
- 62. Lau AZ, Tunnicliffe EM, Frost R, Koopmans PJ, Tyler DJ, et al. (2014) Accelerated human cardiac diffusion tensor imaging using simultaneous multislice imaging. Magn Reson Med [DOI] [PubMed] [Google Scholar]
- 63. Smerup M, Agger P, Nielsen EA, Ringgaard S, Pedersen M, et al. (2013) Regional and epi- to endocardial differences in transmural angles of left ventricular cardiomyocytes measured in ex vivo pig hearts: functional implications. Anat Rec (Hoboken) 296: 1724–1734. [DOI] [PubMed] [Google Scholar]
- 64. Kristoffersen A (2007) Optimal estimation of the diffusion coefficient from non-averaged and averaged noisy magnitude data. J Magn Reson 187: 293–305. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The authors confirm that all data underlying the findings are fully available without restriction. Data used in this study will be made available via ETH Institutional Data Access. The link will be sent upon email request by Dr. Roger Lüchinger: luechinger@biomed.ee.ethz.ch or the corresponding author: kozerke@biomed.ee.ethz.ch.