Abstract
Introduction: High levels of sound have several negative effects, such as noise-induced hearing loss and delayed growth and development, on premature infants in neonatal intensive care units (NICUs). In order to reduce sound levels, they should first be measured. This study was performed to assess sound levels and determine sources of noise in the NICU of Alzahra Teaching Hospital (Tabriz, Iran). Methods: In a descriptive study, 24 hours in 4 workdays were randomly selected. Equivalent continuous sound level (Leq), sound level that is exceeded only 10% of the time (L10), maximum sound level (Lmax), and peak instantaneous sound pressure level (Lzpeak) were measured by CEL-440 sound level meter (SLM) at 6 fixed locations in the NICU. Data was collected using a questionnaire. SPSS13 was then used for data analysis. Results: Mean values of Leq, L10, and Lmax were determined as 63.46 dBA, 65.81 dBA, and 71.30 dBA, respectively. They were all higher than standard levels (Leq < 45 dB, L10 ≤50 dB, and Lmax ≤65 dB). The highest Leq was measured at the time of nurse rounds. Leq was directly correlated with the number of staff members present in the ward. Finally, sources of noise were ordered based on their intensity. Conclusion: Considering that sound levels were higher than standard levels in our studied NICU, it is necessary to adopt policies to reduce sound.
Keywords: Noise, Sound level, Premature infants, Neonatal intensive care unit
Introduction
Recent advances in treatment of respiratory distress syndrome has increased the survival rate of premature infants.1 However, some premature infants are faced with neurobehavioral problems during their school years.2 Premature infants spend most of their first months of life in neonatal intensive care units (NICUs) and are exposed to constant, sudden, and loud noise of 57-97 dBA and maximum intensity of 120 dBA which is equivalent to the noise of a tractor.3 Therefore, improvement of quality of life of premature infants with the focus on NICU environment has become one of the research goals.4,5
Gestational age of less than 37 weeks, difficult transition from intrauterine to extra uterine life, physiological limitations, and underdeveloped central nervous system make premature infants vulnerable to negative effects of high level of sound.6 Premature infants exposed to high level of sound suffer from fluctuations in heart rate and respiratory rate, decreased arterial oxygen saturation, changes in blood pressure, and impaired weight gain, sleep, and learning.7-9, Moreover, high level of sound has negative effects on the primary development of sensory neurons which can disrupt the natural development of the sensory nervous system and cause behavioral, speech, and hearing problems.10,11 The American Academy of Pediatrics (AAP) and the Joint Committee on Infant Hearing (JCIH) have introduced noise as a major physical factor causing pollution in NICUs. They thus suggested that the admittance of infants to these wards might be associated with deafness.12,13 Sound levels are reported as A-weighted decibels (dBA). A-weighted sound level measurements are numerically adjusted to reflect the frequency-dependent nature of human hearing at low sound levels.14 AAP (2007) recommended the standard values of equivalent continuous sound level (Leq), the sound level that is exceeded for only 10% of any specific hour (L10), and maximum sound level during data collection (Lmax) as 45, 50, 65 dBA, respectively.13 According to AAP, lack of compliance to these values would result in the risk of negative effects on the health of premature infants.13 The first step in establishing standard sound levels is measurement.15 Owing to suitable modern technology and equipment, it is currently possible to measure sound levels.16 In some NICUs in the United States of America, sound levels are measured annually.16 In Iran however, little attention is paid to the measurement of sound levels in NICUs. In fact, no study has been conducted on measuring sound levels in NICUs. Comparing the measured sound levels in NICUs with standard levels would facilitate the identification of sources of noise and their effects on sound levels. New policies can accordingly be designed to eliminate or mitigate sources of noise and hence reduce sound levels.17,18 Reducing sound level in NICUs will in turn shorten the NICU stay of premature infants and support their growth and development.15 The aim of the current study was to assess sound levels and to determine sources of noise and their effects on sound levels in the NICU of Alzahra Hospital as the largest NICU in northwestern Iran.
Materials and Methods
After being approved by the Ethics Committee of Tabriz University of Medical Sciences, this descriptive study was conducted to assess sound levels in an NICU with an area of 187 m2. Approximately 8850 infants are delivered in this hospital annually and about 2100 are admitted to the NICU. This area was divided into 3 open care sections (1-3) with 2 partial walls. The nomination of third, second, and first levels of NICU was defined according to the condition of the infants cared. The dimensions of the first, second, and third sections were 11 × 5.5 × 3.5 m3 (area: 60.5 m2), 10 × 5.5 × 3.5 m3 (area: 55 m2), and 11 × 5.5 × 3.5 m3 (area: 60.5 m2), respectively. There were two metal sinks for washing hands and equipment at entrance of sections 1 and 3. Nursing station and entrance were along the second level of NICU. Each section had its own cooling system and one ventilation system was in section 2. Moreover, no acoustic material was used in the construction of the ward and the mean Leq outside the ward was 60 dBA. At the time of sound levels measurement, no building repairs or installation of new equipment were in progress and no sound level reduction protocols were implemented.19
A CEL-440 sound level meter (SLM) and a CEL-382 calibration device (England) capable of detecting sound levels of 10-120 dBA were used. The adjustment of the device to dBA made it sensitive to audible sounds.
Data was collected using a self-made questionnaire consisting of 3 parts of demographic characteristics, sources of noise, and mean sound levels. The validity of the questionnaire was confirmed through content validity and feedback from 3 audiologists, 2 occupational health engineers, 2 infant specialists, and 5 professors of nursing and midwifery.
Since many studies have shown similar sound levels during night and day in NICU,13 we determined sound levels at 24 randomly selected hours (from 7 am to 10 pm) on 4 workdays.
It is important to avoid noise resonance, and measure mean sound levels.16 Therefore, two locations in the center of all open care sections (a total number of 6 points) were selected (Figure 1). The SLM microphone was placed upright at the height of 1.20 m.
During every hour of measurement, the demographic characteristics of the ward and the calibration of the SLM device were assessed by the main researcher. The microphone was then covered with a wide screen to prevent the effects of air flow on the measurements.16 At every location, Leq, L10, Lmax, and peak value of instantaneous sound level (Lzpeak) were measured every 2 minutes and the data was recorded in the data collection. During data collection periods, any sources of noise were recorded. Data analysis was performed using one way analysis of variance (ANOVA) in SPSS13(SPSS Inc., Chicago, IL, USA).
Results
The mean numbers of hospitalized infants at the time of measurements in sections 1, 2, and 3 were 10 and 12, respectively. The mean number of people present at the NICU was 13. The mean numbers of medical ventilators used in these sections were 8, 4, and 0, respectively. For all infants in sections 1 and 2, monitors were used. Pulse oximetry was used for infants in section 3. An intravenous fluid infusion pump was employed as required.
Sound levels ranged between 56.10 and 104.80 dBA. The measured values were higher than standard levels of NICU according to the recommendation of AAP (Table 1). One way ANOVA showed no significant differences between sound levels of the 6 locations [f (5, 15) = 1.71; p = 0.135].
Table 1. Measured and standard sound levels in neonatal intensive care units .
Sound pressure level (dBA) | Standard level | Measured level * | 95% CI | Range |
Leq | 45 | 63.46 (2.97) | 62.97-63.95 | 56.10-70.50 |
L10 | 50 | 65.81 (3.02) | 70.50-71.98 | 57.50-75.00 |
Lmax | 65 | 71.24 (4.48) | 65.32-66.31 | 60.50-89.70 |
Lzpeak | Undefined | 91.68 (4.20) | 90.99-92.38 | 78.30-104.80 |
*: Measured values are expressed as mean (SD).
Leq: Equivalent continuous sound level; L10: Sound level that is exceeded for only 10% of any specific hour; Lmax: Maximum soundlevel during data collection; Lzpeak: Peak value of instantaneous sound pressure level
Sources of noise in the NICU in order of intensity are shown in Figure 2. Washing metal dishes in the metal sink caused the greatest Leq in the ward. Leq varied during different hours of the day. The highest Leq was recorded at nursing rounds (7-8, 13-14, and 19-20). The lowest Leq was detected at 10 pm (Figure 3). Sound levels had direct relationship with the number of people present at the ward (p = 0.007) (Figure 4). The greatest mean value of Lzpeak was caused by dropping metal devices on the ground (Figure 5).
Discussion
The measured values were higher than standard levels of NICU. Previous similar studies also reported the same results.17,18 Despite differences in critical conditions and use of different number of devices, sound levels of the 6 selected locations had no statistically significant differences.However, according to previous research, there is a direct relationship between how critical the condition of an infant in the NICU is and noise levels due to the use of more equipment and presence of greater number of individuals.20 Openings of the 3 sections in the NICU were actually sources of noise for the other 2 sections. Therefore, separation of 3 levels of NICU could be an effective strategy for reducing sound levels.21 Chen et al. were able to decrease Leq to 4.5 dBA by closing an open section of an NICU.22 Because dBA is a logarithmic unit, a 3 dBA difference cuts the intensity of the sound levels by half. Therefore, a 4 dBA difference is very noticeable.23 Ramesh et al. could decrease the sound levels of the noisiest room by 9.58 dBA through implementation of sound level reduction protocols such as separation of rooms with glass and aluminum partitions.24 Sound reduction essentially occurs according to the cascading quality of sound, i.e. in a noisy environment, people tend to speak louder and in a quite environment they tend to speak more quietly.22
Nursing rounds were the noisiest times due to the greatest number of people present in the ward, the highest level of inattention to sources of noise such as alarm of devices and crying of infants, and performing noisy activities. Many previous studies have also reported nurse rounds as one of the noisiest times in the ward.20 Krueger et al. considered nurses to have an important effect on noise levels in NICUs.25 The major source of noise in the hospital is the talking of the personnel.26 Researchers believe that the noise of conversations (mean level: 59-90 dBA) and alarm of devices (mean level: 55-85 dBA) are the two main sources of noise in NICUs that can be controlled by nurses.27 Moreover, the inattention of nurses to alarms of devices and the simultaneous noise of more than one device can increase the mean sound level of the environment up to 7.5 dBA.28 Therefore, the quick response of nurses to alarm of devices will decrease the sound levels of the ward.29 Many studies have performed training programs to increase the knowledge of staff in order to reduce sound levels.28 Teaching the harmful effects of high sound levels and implement-
ation of a behavior modification program have been reported to decrease sound levels by 5 dBA.25 Elander and Hellström informed nurses about sources of noise and thus decreased sound levels of an open care section by 10 dBA. While the staff talked for 62% of each shift before education, the rate decreased to 14% after the education.30 Many studies have been successful in decreasing sound pressure levels of wards by improving nurse activities such as limiting bedside teaching rounds on infants, reduction of the sound of monitors and alarms, and removing phones from care spaces.28
The presence of students significantly increased sound levels. Therefore, educating both medical and nursing students before entering an NICU seems necessary.17 Moreover, there was a significant correlation between the number of people present in the ward and sound level. Since the presence of people is associated with noise, implem- enting regulations to limit the number of people attending an NICU at one time is essential.31 During each of our measurements, there was a possibility of falling of metal files on the ground up to 3 times. Falling of metal objects on the ground causes a mean noise level of 90-110 dBA and results in the highest level of physiological instability in infants.32,33 Therefore, it is recommended to replace metal files with plastic ones.24,34 There were also two big metal surgical sinks in the ward. Turning the tap of a metal sink on or off generates 66-77 dBA of noise.28 In the current study, metal objects were also washed in the metal sinks which significantly increased Lzpeak. Therefore, replacing large steel scrub sinks with smaller ones will help reduce noise.20
All sources of noise levels higher than 55 dBA in the NICU need to be eliminated or mitigated.3 Since the wheels of portable equipment were also identified as sources of noise, they should be replaced with low-noise wheels. In addition, alarms of equipments were important sources of noise in the NICU. Therefore, eliminating or adjusting alarms will reduce sound levels.34,35
A limitation of this study was the awareness of nurses of the presence of researchers which may have affected their behavior and hence the results of this study. Therefore, future studies are recommended to hide SLM microphones from the personnel and measure noise at times they are not informed about. Moreover, this study measured sound pressure levels of the ward for an average of 6 hours a day. However, further research with increased time of measurement would be beneficial in enhancing the accuracy of the results.
Conclusion
The measured values were higher than standard levels of NICU. Every NICU requires a sound level assessment system in order to achieve environmental noise limiting guidelines and to get closer to the standard sound levels. By evaluating sound levels, the sources of noise can be identified and their effects on sound levels can be studied. Moreover, the staff, especially nurses, can decrease sound levels by implementing new policies. Infants will thus be faced less sound stimuli and a safe care environment would be created. The results of this study can also be used to create a protocol to reduce sound level.
Ethical issues
None to be declared.
Conflict of Interest
The authors declare no conflict of interest in this study.
Acknowledgments
This article is part of an MSc thesis. We would like to thank all nurses of the NICU ward who assisted us in performing this research. We are also thankful to the health center specialists, Mr. Ahmadi and Mr. Adlibakhsh, who assisted us in measuring sound pressure levels.
References
- 1.Obladen M. History of surfactant up to 1980. Biol Neonate . 2005; 87(4):308–16. doi: 10.1159/000084878. [DOI] [PubMed] [Google Scholar]
- 2.Perlman JM. Neurobehavioral deficits in premature graduates of intensive care--potential medical and neonatal environmental risk factors. Pediatrics . 2001; 108(6):1339–48. doi: 10.1542/peds.108.6.1339. [DOI] [PubMed] [Google Scholar]
- 3.Bremmer P, Byers JF, Kiehl E. Noise and the premature infant: physiological effects and practice implications. J Obstet Gynecol Neonatal Nurs . 2003; 32(4):447–54. doi: 10.1177/0884217503255009. [DOI] [PubMed] [Google Scholar]
- 4.Brown G. NICU noise and the preterm infant. Neonatal Netw . 2009; 28(3):165–73. doi: 10.1891/0730-0832.28.3.165. [DOI] [PubMed] [Google Scholar]
- 5.Floyd Dazé AM. Challenging designs of neonatal intensive care units. [PubMed]
- 6.Crit Care Nurse. Crit Care Nurse 2005; 25(5):59-66Blackburn SEnvironmental impact of the NICU on develop mental outcomes. J Pediatr Nurs . 1998; 13(5):279–89. [Google Scholar]
- 7.Morris BH, Philbin MK, Bose C. Physiological effects of sound on the newborn. J Perinatol . 2000; 20(8 Pt 2):S55–60. doi: 10.1038/sj.jp.7200451. [DOI] [PubMed] [Google Scholar]
- 8.Wachman EM, Lahav A. The effects of noise on preterm infants in the NICU. Arch Dis Child Fetal Neonatal Ed . 2011; 96(4):F305–9. doi: 10.1136/adc.2009.182014. [DOI] [PubMed] [Google Scholar]
- 9.Graven SN, Browne JV. Sleep and brain development: The critical role of sleep in fetal and early neonatal brain development. Newborn and Infant Nursing Reviews . 2008; 8(4):173–9. [Google Scholar]
- 10.Graven SN, Browne JV. Sensory development in the fetus, neonate, and infant: Introduction and overview. Newborn and Infant Nursing Reviews . 2008; 8(4):169–72. [Google Scholar]
- 11.Graven SN, Browne JV. Auditory development in the fetus and infant. Newborn and Infant Nursing.
- 12.Reviews 2008; 8(4): 187-93.
- 13.American Academy of, Commitee on Environmental. Noise: a hazard for the fetus and newborn. Pediatrics . 1997; 100(4):724–7. [PubMed] [Google Scholar]
- 14.Witt CL. Witt CLAddressing noise in the NICUForward. Adv Neonatal Care . 2008; 8(5 Suppl):S2. doi: 10.1097/01.ANC.0000337264.62846.14. [DOI] [PubMed] [Google Scholar]
- 15.Vahidi R. Noise pollution. Iran: National Program Health Mangment Center; 2004. (Persian).
- 16.Graven SN. Sound and the developing infant in the NICU: conclusions and recommendations for care. J Perinatol . 2000; 20(8 Pt 2):S88–93. doi: 10.1038/sj.jp.7200444. [DOI] [PubMed] [Google Scholar]
- 17.Gray L, Philbin MK. Measuring sound in hospital nurseries. J Perinatol . 2000; 20(8 Pt 2):S100–4. doi: 10.1038/sj.jp.7200440. [DOI] [PubMed] [Google Scholar]
- 18.Berg AL, Chavez CT, Serpanos YC. Monitoring noise levels in a tertiary Neonatal Intensive Care Unit. Contemporary Issues in Communication Science and Disorders . 2010; 37:69–72. [Google Scholar]
- 19.Matook SA, Sullivan MC, Salisbury A, Miller RJ, Lester BM. Variations of NICU sound by location and time of day. Neonatal Netw . 2010; 29(2):87–95. doi: 10.1891/0730-0832.29.2.87. [DOI] [PubMed] [Google Scholar]
- 20.Nogueira Mde F, Di Piero KC, Ramos EG, Souza MN, Dutra MV. Noise measurement in NICUs and incubators with newborns: a systematic literature review. Rev Lat Am Enfermagem . 2011; 19(1):212–21. doi: 10.1590/s0104-11692011000100028. [DOI] [PubMed] [Google Scholar]
- 21.Darcy AE, Hancock LE, Ware EJ. Darcy AE, Hancock LE, Ware EJA descriptive study of noise in the neonatal intensive care unitAmbient levels and perceptions of contributing factors. Adv Neonatal Care . 2008; 8(3):165–75. doi: 10.1097/01.ANC.0000324341.24841.6e. [DOI] [PubMed] [Google Scholar]
- 22.Evans JB, Philbin MK. Facility and operations planning for quiet hospital nurseries. J Perinatol . 2000; 20(8 Pt 2):S105–12. doi: 10.1038/sj.jp.7200437. [DOI] [PubMed] [Google Scholar]
- 23.Chen HL, Chen CH, Wu CC, Huang HJ, Wang TM, Hsu CC. The influence of neonatal intensive care unit design on sound level. Pediatr Neonatol . 2009; 50(6):270–4. doi: 10.1016/S1875-9572(09)60076-0. [DOI] [PubMed] [Google Scholar]
- 24.Milette I. Decreasing noise level in our NICU: The impact of a noise awareness educational program. Adv Neonatal Care . 2010; 10(6):343–51. doi: 10.1097/ANC.0b013e3181fc8108. [DOI] [PubMed] [Google Scholar]
- 25.Ramesh A, Suman Rao PN, Sandeep G, Nagapoornima M, Srilakshmi V, Dominic M, Swarnarekha Swarnarekha. Efficacy of a low cost protocol in reducing noise levels in the neonatal intensive care unit. Indian J Pediatr . 2009; 76(5):475–8. doi: 10.1007/s12098-009-0066-5. [DOI] [PubMed] [Google Scholar]
- 26.Krueger C, Wall S, Parker L, Nealis R. Elevated sound levels within a busy NICU. Neonatal Netw . 2005; 24(6):33–7. doi: 10.1891/0730-0832.24.6.33. [DOI] [PubMed] [Google Scholar]
- 27.Joseph A, Ulrich R. Sound control for improved outcomes in healthcare settings. The Center for Health Care Design . 2007; 4:1–15. [Google Scholar]
- 28.Richardson A, Thompson A, Coghill E, Chambers I, Turnock C. Development and implementation of a noise reduction intervention programme: a pre- and postaudit of three hospital wards. J Clin Nurs . 2009; 18(23):3316–24. doi: 10.1111/j.1365-2702.2009.02897.x. [DOI] [PubMed] [Google Scholar]
- 29.Johnson AN. Adapting the neonatal intensive care environment to decrease noise. J Perinat Neonatal Nurs . 2003; 17(4):280–8. doi: 10.1097/00005237-200310000-00006. [DOI] [PubMed] [Google Scholar]
- 30.Nathan LM, Tuomi SK, Müller AM. Noise levels in a neonatal intensive care unit in the Cape metropole. South African Journal of Child Health . 2008; 2(2):50–4. [Google Scholar]
- 31.Elander G, Hellström G. Reduction of noise levels in intensive care units for infants: Evaluation of an intervention program. Heart Lung . 1995; 24(5):376–9. doi: 10.1016/s0147-9563(05)80058-5. [DOI] [PubMed] [Google Scholar]
- 32.Byers JF, Waugh WR, Lowman LB. Sound level exposure of high-risk infants in different environmental conditions. Neonatal Netw . 2006; 25(1):25–32. doi: 10.1891/0730-0832.25.1.25. [DOI] [PubMed] [Google Scholar]
- 33.White-Traut RC, Nelson MN, Silvestri JM, Patel M, Lee H, Cimo S, McReynolds E. Maturation of the cardiac response to sound in high-risk preterm infants. Newborn and Infant Nursing Reviews . 2009; 9(4):193–9. [Google Scholar]
- 34.Johnson AN. Neonatal response to control of noise inside the incubator. Pediatr Nurs . 2001;27(6):600–5. [PubMed] [Google Scholar]
- 35.Nathan L. Noise levels in a neonatal intensive care unit in the cape metropole [Master Thesis]. Stellenbosch: University of Stellenbosch; 2007.
- 36.McLendon D, Check J, Carteaux P, Michael L, Moehring J, Secrest JW, Clark SE, Cohen H, Klein SA, Boyle D, et al. Implementation of potentially better practices for the prevention of brain hemorrhage and ischemic brain injury in very low birth weight infants. Pediatrics . 2003; 111(4 Pt 2):e497–503. [PubMed] [Google Scholar]