An exponential growth of computational phantom research in radiation protection, imaging, and radiotherapy: A review of the fifty-year history

X George Xu 1
PMCID: PMC4169876  NIHMSID: NIHMS624156  PMID: 25144730

Abstract

Radiation dose calculation using models of the human anatomy has been a subject of great interest to radiation protection, medical imaging, and radiotherapy. However, early pioneers of this field did not foresee the exponential growth of research activity as observed today. This review article walks the reader through the history of the research and development in this field of study which started some 50 years ago. This review identifies a clear progression of computational phantom complexity which can be denoted by three distinct generations. The first generation of stylized phantoms, representing a grouping of less than dozen models, was initially developed in the 1960s at Oak Ridge National Laboratory to calculate internal doses from nuclear medicine procedures. Despite their anatomical simplicity, these computational phantoms were the best tools available at the time for internal/external dosimetry, image evaluation, and treatment dose evaluations. A second generation of a large number of voxelized phantoms arose rapidly in the late 1980s as a result of the increased availability of tomographic medical imaging and computers. Surprisingly, the last decade saw the emergence of the third generation of phantoms which are based on advanced geometries called boundary representation (BREP) in the form of Non-Uniform Rational B-Splines (NURBS) or polygonal meshes. This new class of phantoms now consists of over 287 models including those used for non-ionizing radiation applications. This review article aims to provide the reader with a general understanding of how the field of computational phantoms came about and the technical challenges it faced at different times. This goal is achieved by defining basic geometry modeling techniques and by analyzing selected phantoms in terms of geometrical features and dosimetric problems to be solved. The rich historical information is summarized in four tables that are aided by highlights in the text on how some of the most well-known phantoms were developed and used in practice. Some of the information covered in this review has not been previously reported, for example, the CAM and CAF phantoms developed in 1970s for space radiation applications. The author also clarifies confusion about “population-average” prospective dosimetry needed for radiological protection under the current ICRP radiation protection system and “individualized” retrospective dosimetry often performed for medical physics studies. To illustrate the impact of computational phantoms, a section of this article is devoted to examples from the author’s own research group. Finally the author explains an unexpected finding during the course of preparing for this article that the phantoms from the past 50 years followed a pattern of exponential growth. The review ends on a brief discussion of future research needs (A supplementary file “3DPhantoms.pdf” to Figure 15 is available for download that will allow a reader to interactively visualize the phantoms in 3D).

1. Introduction

For more than 50 years, radiation dose assessment using computational models of the human anatomy has been a subject of great interest to the fields of radiation protection, medical imaging, and radiotherapy. Health physicists often need to understand how radiation interacts with the human body so that they can ensure the safety of workers and members of the public in accordance with complicated regulatory requirements. In diagnostic radiology and nuclear medicine, the imaging process—involving x-ray and gamma-ray photons powerful enough to traverse bodily tissues—must be optimized to achieve necessary image quality while minimizing potentially harmful radiobiological effects. Radiation therapy aims to deposit a lethal dose to the tumor—which may be subjected to organ motion—using focused external beams of x-ray and gamma-ray photons, electrons, protons, and heavy ions, or using internal sources that are less penetrating, while sparing healthy tissues from toxicity and secondary cancer. The anatomical modeling techniques evolved over time and new phantoms would emerge, as expected. However, the early pioneers of this field did not foresee the exponential growth of research activity that this review article has uncovered.

Radiation dosimetry is a basic science that has to do with the determination of the amount and distribution pattern of ionizing energy deposited in the object of interest. Accurate radiation dosimetry in the human body is quite challenging for several reasons: (1) exposure scenarios are diverse, often including complex and unique geometrical relationships between the source and human body; (2) an exposure can involve multiple radiation types, each of which transverse the human body and interact with tissues according to different radiation physics principles; (3) the human body consists of a very large number of anatomical structures that are heterogeneous in density and composition sometimes under in the influence of organ motion. For instance, cardiac and respiratory motion can result in complex 3-dimensional (3D) and 4-dimensional (4D) dose distribution patterns that must be accounted for during medical imaging or radiotherapy. This last point underscores the importance of anatomical models in radiation dosimetry because dose inside a living person cannot usually be directly measured. Instead, one must use computational or physical anatomical models to estimate the dose delivered to a worker or patient exposed to ionizing radiation. The accuracy of the dose estimate critically depends on how well the anatomical models account for the specific geometry and radiation attenuation properties of each individual—a quite daunting task in light of the fact that every person has a unique body shape and size.

It has been known for a long time that dose inside the body can be derived using either a physical phantom or a computational phantom that mimics human anatomical features. Historically, the term phantom was used in the radiological science literature to mean a physical device that mimics the human body. In the radiation protection community, phantom has also been used to refer to a mathematically defined anatomical model instead of a physiologically based model such as a respiration or blood flow model. In this review article, “computational phantom” and “physical phantom” are used to avoid confusion.

Physical phantoms are made of solid materials which are radiologically equivalent to human tissues. Because the human body consists mostly of water, homogenized water or plastic phantoms are widely used for the calibration of radiation detectors and treatment systems. The simpler designs of these phantoms are useful for routine measurements where standardization between laboratories or hospitals is of critical importance (DeWerd and Kissick 2014). Another use of such homogeneous phantoms is to calibrate calculations by measuring the power output from a specific radiation-emitting machine. In contrast, anthropomorphic phantoms are more realistic and better represent the complex heterogeneity of the human body; they often consist of several tissue-equivalent materials that are molded into shapes of organs or bones to represent part or all of the body. For the ease of placing tiny radiation dosimeters, some of the physical phantoms for dose measurements come in slices with cavities in locations that match with organs of interest.

The approach of using such anthropomorphic physical phantoms for organ dose measurements can be expensive and time-consuming due to necessary experimental and radiation safety procedures. Furthermore, commercially available physical phantoms only come in a limited number of body sizes and do not fully reflect the diversity of the human population. Luckily, the advent of first-generation computers and Monte Carlo simulation methods originally designed for nuclear weapons research demonstrated the feasibility to calculate organ doses using computational phantoms. Such computational phantoms include extensive details of the exterior and interior features of the human body such as the shape, volume, and mass of radiosensitive organs. Coupled with information for tissue density and chemical composition, a computational phantom allows a researcher to simulate radiation interactions and energy deposition patterns in the body accurately. Although experimental work involving a whole-body physical phantom is still needed to verify the calculations especially when involving complex irradiation conditions, the computational approach is, in general, advantageous compared with the experimental approach in versatility, efficiency, precision, and safety. Furthermore, internally distributed radiation sources are best handled by the computational approach.

Since the 1960s, the development and application of computational phantoms have evolved into a specialized field of research that is integral to radiation protection, medical imaging, and radiotherapy. For non-ionizing radiation, similar computational phantoms have been developed over the years to study the biological effects caused by the heat produced by radiofrequency-emitting devices such as electric power lines and wireless cellular-phones. In the twenty years after the first computational phantom was developed in the 1960s, less than two dozen computational phantoms were developed and used by a small group of people in national laboratories who had access to computers. Computational phantoms became widely adopted in the 1980s with the advent of personal computers. By then, medical imaging had made it possible to visualize the anatomy in 3D. An international research community soon took shape and, over the years, several workshops were held to disseminate research ideas, facilitate collaboration, and develop roadmap for the future.

In 1996, Peter Dimbylow organized the first workshop on voxelized computational phantoms at the National Board of Radiological Protection (now Health Protection Agency) in the United Kingdom (Dimbylow 1996). In 2000, Keith Eckerman hosted a similar workshop at Oak Ridge National Laboratory in the United States (Eckerman 2000). The interest about computational phantoms was so widespread in the mid of 2000s that many colleagues decided to form the Consortium of Computational Human Phantoms (CCHP) during a dinner meeting at the Monte Carlo 2005 Topical Meeting, American Nuclear Society, Chattanooga, TN, April 17–21, 2005 (http://www.virtualphantoms.org). Under the umbrella of CCHP, George Xu and Keith Eckerman published the Handbook of Anatomical Models for Radiation Dosimetry in 2009 involved 64 authors from 13 countries (Xu and Eckerman 2009). In 2011, George Xu and Junli Li organized a workshop under the name of “The 3rd International Workshop on Computational Phantoms for Radiation Protection, Imaging and Radiotherapy,” in Beijing, China (http://www.virtualphantoms.org/3rdWorkshopInBeijing.html). The Beijing Workshop was the first time that researchers from both ionizing and non-ionizing radiation communities attended. In Beijing, it was decided that the workshop would be held every other year. The 4th workshop (http://cp2013.org/) was then hosted by Niels Kuster of the Foundation for Research on Information Technologies in Society (IT'IS) in Zurich, Switzerland, May 20–22, 2013. At the Zurich workshop, it was decided to hold the next workshop in Seoul, Korea in 2015.

By 2009, approximately 121 computational phantoms had already been reported in the literature for studies involving ionizing and non-ionizing radiation (Xu and Eckerman 2009). Such a large number of computational phantoms was somewhat surprising at the time, given the fact that less than a dozen existed prior to the 1980s. Most of these phantoms were reported during the late 1980s and early 1990s due to the increased availability of advanced medical imaging technologies such as x-ray computed tomography (CT) and magnetic resonance imaging (MRI). It also became clear that organ surfaces could be defined in a variety of solid geometry modeling techniques including those we were familiar with, such as quadric equations and voxels, but also advanced geometries such as B-Splines, Non-Uniform Rational B-Splines (NURBS), and polygonal meshes. In the recent workshops, a number of questions were on people’s minds:

  • What are the fundamental challenges in phantom related research?

  • Why did the computational phantoms evolve the way they did?

  • Will the number of computational phantom stop increasing at some point of time?

  • What are the differences between “population-average” prospective dosimetry needed for radiological protection under the current ICRP radiation protection system and “individualized” retrospective dosimetry often performed for medical physics studies?

  • Is the concept of “Reference Man” in radiation protection obsolete?

  • What are future research directions?

Answers to these questions, and many others, require an appreciation of the rationales and methods responsible for some of the most important computational phantoms. Through a review about when and how various computational phantoms came about in the last 50 years, this article also attempts to learn insight into where this field of study may be heading in the future.

This article is organized into the following sections: (1) Introduction, (2) Solid-geometry modeling methods: CSG and BREP, (3) Monte Carlo Codes used with Computational Phantoms, (4) The Evolution of Computational Phantoms, (5) Physical Phantoms, (6) Examples of Computational Phantom Applications at RPI, (7) Discussion, and (8) Conclusion.

2. Solid-geometry modeling methods: CSG and BREP

It is essential to understand the geometrical shapes—the building blocks—of computational phantoms. The construction of a computational phantom must consider multiple factors such as anatomy, radiosensitivity of specific organs/tissues, computational efficiency, and geometrical compatibility with the Monte Carlo code that carries out the radiation transport calculation. As a first step, a phantom must be generated by explicitly defining the surfaces of an organ in which radiation interactions and energy depositions occur. The computer graphics community has dealt extensively with solid-geometry modeling for computer-aided-design (CAD). Two general modeling methods have been widely used: (1) constructive solid geometry (CSG) and (2) boundary representation (BREP) (Agostinelli et al 2003, Geant4 Team 2007, Leyton 2001, Stroud 2006). The topology—spatial location and relationship of the surfaces—is fundamentally different for these two methods.

CSG allows the modeler to create a solid object using Boolean operators (or the equivalent) to combine very simple shapes called primitives. Examples of these primitives include cuboids, cylinders, prisms, pyramids, spheres, cones and ellipsoids—surfaces that are easily described by quadric equations. CSG representations are easy to adopt and can yield good results when the objects are relatively simple in shape.

Modern CAD software systems, however, are based on the more powerful BREP methods. There are two types of information in BREP: topological and geometric. Topological information provides the relationships among vertices, edges, and faces. In addition to connectivity, topological information also includes orientation of edges and faces. In advanced BREP-based CAD, the exterior of an object can be defined as NURBS, which afford very smooth surfaces. The faces can alternatively be represented as polygons whose vertices are defined by a set of coordinate values x, y and z. A polygon mesh or unstructured grid is a collection of vertices and polygons that define the geometric shape of a polyhedral object in CAD. In principle, NURBS and polygonal meshes are interchangeable BREP data structures; however, unlike the CSG representation, BREP is much more flexible because a richer set of operation tools are available (e.g. extrusion, chamfering, blending, drafting, shelling, and tweaking). These features allow BREP-based models to include very complex anatomical features. Furthermore, the BREP technique is ideally suited for surface deformation—an operation necessary for the adjustment of organ size and for organ motion simulations and for changing the posture of phantoms to better simulate how humans interact with their environment.

For example, the left lung can be represented in the CSG method by “half an ellipsoid with a section removed” (Cristy and Eckerman 1987). The cut-out section, which is not specified by the original authors, can be defined by a Boolean operation subtracting one ellipsoid (B) from the other (A) to create the left lung, as described below:

A:(X8.55)2+(Y7.5)2+(Z43.524)21,Z43.5B:(X2.55)2+(Y7.5)2+(Z43.524)21,ify<0

In Figure 1a and 1b, the 3D shapes of the left lung before and after the Boolean operation are illustrated. These surface equations are computationally efficient and are accepted by nearly all Monte Carlo codes. When using a Monte Carlo code, the geometry of the left lung is often further simplified by replacing the ellipsoid B with several planes. This type of phantoms is commonly referred to as “stylized” or “mathematical” phantoms. However, even with complicated and carefully designed Boolean operations, phantoms based on quadric surfaces are not anatomically accurate. The true shape of a human lung is more amorphous and cannot be described by a simple ellipsoid.

Figure 1.

Figure 1

A model of the left lung defined by different modeling methods. (a) The CSG-type modeling before the Boolean operation (subtraction) is performed involving two ellipsoids A and B. (b) After the subtraction of B from A. (c) A voxel representation of the lung. (d) A BREP-type of modeling of the same lung using polygon mesh.

Using voxels as a CSG modeling technique, Figure 1c defines the left lung as an assembly of 3D cuboids. Medical image data can be converted to voxel geometry that provides a direct way of realistically describing the human anatomy. The geometry of a voxel is very easy for existing Monte Carlo codes to handle. On the other hand, each tomographic image slice needs to be treated by a “segmentation” process, which assigns each pixel to an organ or tissue of interest such as the lung, bone, or skin using a unique identification number. It can take a significant amount of time to prepare a voxel-based phantom because there is no automatic segmentation algorithm that works on all organs. Furthermore, a voxel phantom is based on images for one subject, therefore lacking the anatomical variability associated with organ size, shape, and location that are important in the current paradigm for radiation protection dosimetry. Furthermore, CT images do not generally distinguish between soft tissues well and are typically not whole-body images. Finally, the boundary of an organ is defined by uneven steps instead of a smooth surface, as shown in Figure 1c. As a result, the anatomical fidelity depends on the voxel size, especially for thin and small tissues such as the skin, eye lens, ribs, and bone marrow. An adjustment to the organ shape will likely involve all underlying voxels, which is computationally inefficient. These types of computational human body models are commonly referred to as “voxel” or “tomographic” phantoms.

The lung can also be defined by the advanced BREP modeling techniques involving NURBS or polygon mesh surfaces. The most common technique to create a BREP-based phantom involves the surface contour extraction of each organ from a tomographic image dataset using a commercial software package, followed by the integration of individual organs into a whole body assembly. In essence, the contours convert the voxels into NURBS or mesh surfaces that are smooth and anatomically realistic. These phantoms are commonly referred to as “NURBS,” “mesh” or “BREP” phantoms. A misleading name for this type of phantom is “hybrid” which does not specify what two formats are actually used. Figure 1d shows the triangular meshes of a left lung, which was derived from high-resolution tomographic images.

3. Monte Carlo Codes used with Computational Phantoms

Computational phantoms must be coupled with Monte Carlo codes that simulate radiation transport inside the human body for the purposes of determining the patterns of radiation interaction and energy deposition. Most health and medical physics applications employ photons and electrons with energies up to 20 MeV and protons up to 300 MeV. Health physics dosimetry, however, also considers neutron sources in nuclear reactors and particles with energies in the TeV range in high energy physics research or space radiation environment. Each type of radiation interacts with matter differently. For example, photons (x-rays or gamma-rays) deposit energy primarily via photoelectric effect, Compton scattering, and pair production processes (Attix 1986). The probability of a photon interaction occurring within an organ or tissue is determined by “cross sections” that are associated with the energy, the tissue electron density, and the tissue chemical composition. Mathematically, the differential cross section per electron for a photon undergoing the Compton scattering at angle φ per unit solid angle Ω is analytically determined by using the Klein-Nishina Equation (Attix 1986)

dσdΩφ=ro22(hν'hν)2(hνhν'+hν'hνsin2φ) (Eq. 2)

where ro is the classical electron radius, and hv and hv’ are photon energies before and after the scattering, respectively. Extensive photon cross-section libraries have been developed for these purposes (Hubbell 1969, Storm and Israel 1970).

In general, Boltzmann radiation transport problems described by various differential, integral, and integro-differential equations can be solved by numerical computational methods including finite difference, finite element, discrete ordinates, and Monte Carlo. However, only the Monte Carlo methods are able to account for all aspects of particle interactions within 3D heterogeneous media such as the human body. Monte Carlo methods, which are based on statistical simulations, have a long history, but the real application to radiation transport simulations and the associated software development arose from nuclear weapons research at Los Alamos National Laboratory during World War II (Hammersley and Handscomb 1964).

In a Monte Carlo code, random numbers are used to determine the distance and fate of a particle by comparing interaction probabilities for every geometrical region of interest. This rather tedious process is repeated for an extremely large number of particles (nowadays often exceeding 1 billion), and each particle is tracked in the 3D anatomical model until all its energy is absorbed or the particle escapes from the transport geometry. The inherent statistical uncertainty can be controlled to be less than 1%, which is often more precise than an experimental result performed in a physical phantom using a dosimeter (for quantities such as the absorbed dose). Experiments with physical phantoms are often still needed, however, to validate the Monte Carlo calculations. This creates a peculiar situation where it is not immediately clear whether the direct measurement in a simplistic physical phantom or the Monte Carlo calculation involving a more realistic compuational phantom provides the more accurate dose estimate. One reason for the rise in popularity of the Monte Carlo methods for dose estimation is the improvement in computer affordability and computing power over the last 30 years. The development of major Monte Carlo code packages is supported by national labs as well as by the user community at large. As a result, Monte Carlo codes are used today for many applications in nuclear engineering, health physics, and medical physics.

Most production Monte Carlo codes were originally developed for nuclear engineering and high energy physics research. Although these codes have been vigorously validated for radiation physics, the software packages are often difficult to use without extensive experience. Nearly all existing Monte Carlo codes can handle CSG shapes including the voxels. In the 1990s, some of these codes had trouble handling the very large numbers of voxels required for simulations involving whole-body computational phantoms (e.q., MCNP would limit the voxels to less than 25 millions).

There are many comprehensive reviews or introductory articles about the Monte Carlo methods for health physics and medical physics (Andreo 1991, Raeside 1976, Turner et al 1985, Zaidi 1999, Zaidi and Sgouros 2003, Rogers 2006). Some of the public-domain, general-purpose Monte Carlo codes used for radiation dose calculations include: EGS (NRC 2013), FLUKA (Battistoni et al 2006), GEANT4 (Allison et al 2006), MCNP (Brown 2003), MCNPX (Pelowitz 2005), MCNP6 (Goorley et al 2013), and PENELOPE (Salvat et al 2003). Specific codes for radiation therapy have also been developed (Rogers 2006).

4. The Evolution of Computational Phantoms

Previously published reviews on the historical development of computational phantoms have focused on a certain time period or a particular phantom type (Caon 2004, Zaidi and Xu 2007, Eckerman et al 2009, Zaidi and Tsui 2009). These reviews did not explicitly classify phantom modeling techniques, and since the time of their publication, many new phantoms have been developed using the BREP methods. An understanding of the modeling techniques and when the research community predominantly adopted each technique provides important insight into future directions. Based on chronological and technical information in the literature, this review article divides computational phantoms into three generations: (1) Stylized phantoms that are based on quadratic equations (1960s to 2000s); (2) Voxel phantoms that are based on tomographic images (1980s to present); (3) BREP phantoms that are based on advanced primitives and are deformable (2000s to present). Figure 2 contrasts these phantom generations in terms of their geometric sophistication.

Figure 2.

Figure 2

Three phantom generations: (1) Stylized phantom; (2) Voxel phantom (but displayed in smooth surfaces); (3) BREP phantom.

4.1. Stylized Phantoms (1960s to 2000s)

The first-generation computational phantoms were developed for the purpose of better assessing organ doses from internally deposited radioactive materials for workers and patients (Eckerman et al 2009). Some of the earliest dose assessment techniques were developed in the first third of the 20th century primarily for use with interstitial radiation sources such as radium. According to Loevinger (1965a, 1965b, 1969), the dosimetry of radioactive materials distributed in the body had been under consideration as early as the 1920s. Quimby has provided an excellent historical review of the early development of radiation dosimetry in nuclear medicine (Quimby 1970). The early techniques were adaptations of methods used for external dose assessment with assumptions and corrections applied to account for the different types of radiation used (NCRP 1985). However, rather than being able to measure the exposure or the absorbed dose, an internal dose assessment required a calculation.

Internal dose calculations were performed during early days using the formulation presented by Marinelli and his colleagues in the 1940s (Marinelli 1942, Marinelli et al 1948). These equations considered only the absorbed dose from beta-emitting radionuclides (classified as non-penetrating radiation) and from gamma-rays (penetrating radiation) emitted in the decay of these radiation sources.

In 1959, the International Commission on Radiological Protection (ICRP) used very simple models for the internal dosimetry calculations associated with the Report of ICRP Committee II (ICRP 1959). In these calculations, each organ of the body was represented as a sphere with an “effective radius.” The radionuclide of interest was assumed to be located at the center of the sphere and the “effective absorbed energy” was calculated for each organ. Corrections were made for the photon energy lost from the sphere. In this approach, the total body was represented as a 30-cm radius sphere. It is also interesting to note that the 30-cm radius sphere was used for an organ designated as “muscle” as well as for the small intestine and the entire gastrointestinal tract.

At the time, these approaches provided reasonably accurate estimates of the dose from a distributed radionuclide. However, most dosimetrists and researchers hoped for improved techniques and more accurate dosimetry estimates as technology developed. There was also a need for dose calculations for a number of new radionuclides introduced into nuclear medicine and more was known regarding the distribution and retention of these radionuclides in specific organs. Of course, the next step was to attempt to model individual organs of the body and ultimately the entire human body in a realistic manner. With the increase in the size and speed of computers, some progress occurred during the late 1950s and through the 1960s, eventually leading to the first-generation of stylized anthropomorphic phantoms.

Table 1 summarizes some of the most important and unique stylized phantoms developed since 1960s. This generation of stylized phantoms originated from work performed at Oak Ridge National Laboratory (ORNL) by Fisher and Snyder in the 1960s (Fisher and Snyder 1966, Fisher and Snyder 1967). Using CSG modeling techniques involving shapes such as elliptical cylinders and cones, they developed the so-called Fisher-Snyer adult phantom. The adult phantom was assumed to be standing erect with the arms at the sides of the body. Three specific regions were defined; the head and neck, the trunk including the arms, and the legs. The head and neck were represented by a 14 cm × 20 cm elliptical cylinder with a height of 24 cm. The trunk and arms were modeled as a larger elliptical cylinder, 20 cm × 40 cm with a height of 70 cm. The legs below the buttocks were modeled as a truncated elliptical cone with a height of 80 cm. Regions of little dosimetric importance were not included, e.g., the hands, feet, ears, nose, etc. The composition of the phantom was assumed to be tissue distributed homogeneously throughout. No attempt was made to model the lungs or skeleton or to define the locations of specific organs in the phantom. Approximately 120 sub-regions were defined in the phantom, which were used to assign approximate values of the absorbed doses to organs located within specific regions. In some cases, absorbed dose estimates for large organs required the evaluation of the doses deposited in several of these regions. Even though the original phantom was designed for use with internally- deposited radionuclides, Snyder saw many other applications. For instance, Snyder used the phantom to study the distribution of dose in the body from external, point sources of gamma-rays (Snyder 1967). He studied four photon energies (0.07, 0.15, 0.5 and 1.0 MeV) and four different source locations at distances of one and two meters from the center of the phantom.

Table 1.

Alphabetical listing of developers of stylized computational phantoms including information on the phantom names, phantom data types, the anatomical features modeled, the human subjects they mimic, whether they were designed for ionizing or non-ionizing radiation applications, and literature references.

Developers Phantom
Names
Data Types Human Subjects Anatomical Features Ionizing (I)
or
Non-ionizing
radiation (N)
References

Bhaba Atomic
Research Centre, India
BARC WBC
Phantoms
(4 phantoms)

Quadric equations

Indian Adult Male
Phantoms representing the
BOMAB phantoms BARC
Reference Phantom and a scaled
version of the ICRP Reference
phantom

I

Bhati et al 2011

Catholic University of
Pusan, Bugok

Korean Male

Quadric equations

Korean Male
MIRD type phantom with outer
body and internal organs modified
according to reference values
from the Korean Ministry of
Science and Technology

I

Kim et al 2010

GSF, Germany

ADAM and
EVA

Quadric equations

Caucasian adult male
and female
Gender-specific phantoms revised
from the ORNL MIRD-5
Phantom for external dose
assessment. Several minor
anatomical changes including the
breast size.

I

Kramer et al 1982
Hanyang University,
Korea

KMIRD

Quadric equations

Korean adult male
Outer body and internal organs of
the ORNL adult male phantom
modified according to Korean
anthropometric data.

I

Park et al 2006
ITN, Portugal ITN WBC
Phantom
Quadric equations Caucasian adult male A mathematical simulation of the
reference male BOMAB phantom.
I Bento et al 2012
Johns Hopkins
University, USA
(formerly with the
University of North
Carolina)

MCAT

Quadric equations

Caucasian adult male
3D and 4D cardiac torso phantom
with gated patient organ motion
information for imaging
applications.

I
Pretorius et al 1997,
Tsui et al 1993,
1994
Key Laboratory of
Particle & Radiation
Imaging, Beijing

CMP

Quadric equations

Chinese adult male
An anthropomorphic phantom
constructed from the Reference
Asian Man and the Chinese
Reference Man

I

Qiu et al 2008

NASA, USA

CAM

Quadric equations

Caucasian adult male
A standing U.S. air force adult
male representing 50th-percentile
height and weight. More than
1000 geometric surfaces and 2450
solid regions.

I

Billings and Yucker 1973
ORNL, USA Fisher-Snyder
Phantom
(MIRD-5) and
others
(6 phantoms)

Quadric equations
Caucasian newborn,1-,
5-, 10-, 15-year-old and
the adult
The first anthropomorphic
phantom representing a
hermaphrodite adult for internal
dosimetry. Organ masses, body
weight and body height
correspond to 50th-percentile data
recommended in ICRP 23. Later,
age-specific phantoms were
developed by others.

I
Fisher and Snyder 1966,
Fisher and Snyder 1967
Pediatric
Phantoms

Quadric Equations
Caucasian newborn, 1-
year, 5-year, 10-year,
and 150-year old
Individual phantoms based on the
literature for each age.

I
Deus and Poston 1976,
Hwang et al 1976,
Jones et al 1976
Cristy-
Eckerman
Family
Phantoms
(6 phantoms)

Quadric equations

Caucasian adult
Based on MIRD-5 Phantom and
others from ORNL, including a
15-year male/female phantom

I

Cristy 1980,
Cristy and Eckerman 1987
Pregnant
Women
(3 phantoms)

Quadric equations
Caucasian pregnant
women at three, six and
nine months of gestation
The stylized adult female was
modified by adding uterine
contents including the fetus, fetal
skeleton and placenta at three
different gestational stages.

I

Stabin et al 1995
Nagoya Institute of
Technology, Japan
Japanese
Infants (2
phantoms)

Quadric equations
Japanese 3-year-old
child
Infant represented by a
homogenous spherical model, and
an ellipsoidal model with three
separate muscle tissues.

N

Hirata et al 2008

Radiation Protection
Bureau, Canada
Mathematical
Models of the
Embryo and
Fetus (4
phantoms)

Quadric equations

Caucasian pregnant
women at 8, 13, 26,38
weeks of gestation
Phantoms include an embryo and
fetus at different gestational
periods not included in the ORNL
Pregnant Women Phantoms.
Designed for dosimetry studies
involving commercial flights.

I

Chen 2004

Fisher and Snyder also developed the “similitude” children phantoms which were scaled-down versions of the adult with added assumption that the entire body was a homogenous tissue (i.e. the lungs and skeleton were ignored) (see discussion by Eckerman et al 2009). These phantoms represented children of 0 (newborn), 1, 5, 10, and 15 years of age. These early phantom designs had outer dimensions representing the average height, surface area, and body mass of a children within each particular age group. These phantoms became known as the “similitude phantoms” because of their resemblance to children. This approach had its limitations because children, in general, are not just “little adults.” However, at the time, these phantoms helped answer a real need in the nuclear medicine community (Kereiakes et al 1965).

In 1969, Snyder and his colleagues reported the first heterogeneous phantom that became known as the “MIRD-5 Phantom,” a named derived from the Medical Internal Radiation Dosimetry (MIRD) Committee of the Society of Nuclear Medicine which adopted the phantom (Snyder et al 1969). This phantom was composed of a skeleton, a pair of lungs, and the remainder (soft tissue). The representation of internal organs in this mathematical phantom was crude, as the simple equations captured only the most general description of the position and geometry of each organ. The original model was intended to represent a healthy “average” adult male, the so-called Reference Man, as defined by the International Commission on Radiological Protection (ICRP). The characteristics of the Reference Man were the result of an extensive review of medical and other scientific literature on the European and North American populations (ICRP 1975). The Reference Man was defined as a 20- to 30-year-old Caucasian, 70 kg in weight and 170 cm in height (the height was later changed to 174 cm). In 1978, Snyder et al (1978) published an elaborative set of specific absorbed fractions using an improved version of their heterogeneous phantom which contained more than 20 organs and more detailed anatomical features.

The limitations associated with the approach of applying a set of scaling factors to the adult phantom to create age-dependent similitude phantoms were clear. Significant efforts were undertaken at ORNL during the mid-1970s to develop individual pediatric phantoms based upon a careful review of the existing literature for each particular age group. This effort produced the next generation of mathematical stylized phantoms that, although they appeared to be modeled after the adult, were designed independently. Three “individual phantoms” were designed by Hwang et al (1976). This set consisted of the newborn, the 1-year old, and 5-year old models. A separate effort was undertaken by Jones et al (1976) for the 15-year old model, and Deus and Poston (Deus and Poston 1976) undertook the design of a 10-year old model after the other four designs were complete. The development of the 10-year old was significantly different from those for the other four ages. In fact, this design was intended to point the way to the next generation of more realistic phantoms (see discussion by Eckerman et al 2009). Even though the design was completed and used for a limited number of dose calculations, it was not popular because of the very complex geometry and, after Poston left ORNL, alternative approaches were developed.

Building upon previous work, Cristy reported the development of a new series of stylized phantoms in 1980 and then with Eckerman in 1987 in the report ORNL/TM-8381 (Cristy 1980, Cristy and Eckerman 1987). This series of “family” of phantoms consisted of an adult male, a newborn, and individuals of ages 1, 5, 10 and 15 (also representing an adult female with additional anatomical features). As shown in Figure 3, each phantom consists of three major sections: (1) an elliptical cylinder representing the trunk and arms; (2) two truncated circular cones representing the legs and feet; and (3) a circular cylinder on which sets an elliptical cylinder capped by half an ellipsoid representing the neck and head. Attached to the legs section is a small region with a planar front surface to contain the testes. The female phantom included two ellipsoids attached to the trunk to represent breasts (not shown in Figure 3). The arms are embedded in the trunk, and minor appendages such as fingers, feet, chin, and nose are omitted.

Figure 3.

Figure 3

The adult male phantom and its dimensions. Similar descriptions and diagrams were purposely followed in a series of ORNL technical reports by Snyder et al (1978), Cristy (1980), and Cristy and Eckerman (1987).

Drawings depicting the external features of all the family phantoms are shown in Figure 4. The pediatric phantoms were designed to form a developmentally consistent family with the existing Snyder adult phantom. The exterior of each phantom has approximately the form of the human body; but, as in their adult phantom, there has been no attempt to model for realistic details because these were presumed to have only small effect on the scattering of photons. Similarly, the description of the interior organs, while approximately correct as to size, shape, position, composition and density, are simplified to provide formulas which could be easily modeled on the computers available at the time. Figure 5 shows a schematic view of the principal organs.

Figure 4.

Figure 4

External views of the age-specific phantom phantoms representing an adult male and children at 15-year old (adult female), 10-year old, 5-year old, 1-year old, and 0-year old (newborn) (From Cristy and Eckerman 1987). When used for an adult female, the 15-year old phantom has breasts appropriate for a reference adult female, which are not shown.

Figure 5.

Figure 5

Anterior view of the principal organs in the head and trunk of the adult phantom developed by Snyder et al (1978). Although the heart and head have been modified, this schematic illustrates the crude nature of the geometric modeling by today’s standards. At the time, however, this was important work represented the state of the science.

4.1.1. Pregnant Woman Phantoms

In 1995, Stabin and his colleagues at ORNL adapted the adult female phantom in this family to represent a pregnant woman at the end of each trimester of pregnancy (Stabin et al 1995). This set of three stylized pregnant female phantoms were used for various internal nuclear medicine applications. Figure 6 shows a drawing of the cross sectional view of the uterine region at 9-months in the pregnant female phantom series by Stabin et al (1995).

Figure 6.

Figure 6

Diagram of the uterus of the 9-month gestation model in the Stabin et al (1995) pregnant female phantom series.

4.1.2. GSF Gender-specific ADAM and EVA Phantoms

In parallel with the efforts at ORNL by Cristy and Eckerman to revise the MIRD-5 Phantom, a group at GSF—National Research Center for Environment and Health in Germany (now known as HZM—the German Research Center for Environmental Health) used the anatomical descriptions of the hermaphrodite MIRD-5 phantom to develop a pair of gender-specific adult phantoms later known as the ADAM and EVA for external dosimetry studies (Kramer et al 1982). The EVA phantom was derived by scaling down all relevant volumes of the MIRD-5 phantom with the total whole body mass ratio of 0.83 that was based on the analysis of ICRP reference organ masses. Then, the female organ masses were modified to create space for neighboring organs. Finally, sex-specific organ such as testes, ovaries, uterus and breasts were introduced into the appropriate phantom to yield ADAM and EVA, respectively. The chin was introduced by removing a section of the neck to created a more realistic external irradiation geometry for the thyroid. The female breasts were represented by two ellipsoid sections attached to the trunk of EVA. There are a number of minor anatomical differences, such as breast sizes, from those reported by Cristy and Eckerman (Kramer et al 1982, Cristy and Eckerman 1987).

4.1.3. CAM and CAF Phantoms for Space Radiation Dosimetry

The Computational Anatomical Man (CAM) and CAF (Computerized Anatomical Female) phantoms, developed by Billings and Yucker in 1973 for the National Aeronautics and Space Administration (NASA), demonstrated a very different and aggressive approach in stylized modeling because the phantoms reportedly consisted of 1100 unique geometric surfaces and 2450 solid regions (Billings and Yucker 1973). According to the authors, internal body geometries such as organs, voids, bones, and bone marrow were explicitly modeled using CSG modeling techniques. A computer program called CAMERA was also developed for performing analyses with the CAM and CAF phantoms. The authors state that “extremely detailed geometrical model of the human anatomy, the most detailed yet prepared, has been developed for use in investigations dealing with exposure of astronauts to the natural space radiation environment”. According to the authors, the model was equally applicable to investigations dealing with exposure of humans to radiation associated with nuclear weapon and nuclear power system environments as well as medical applications such as radiotherapy and radiography (Billings and Yucker 1973). Indeed the surface geometry was so detailed that one may wonder how this was possible using computers in the 1970s. Unfortunately, the CAM and CAF phantoms were never adopted for applications outside the aerospace industry and very little information about the work was accessible by the radiation dosimetry community until Tom Jordan who worked for years for NASA as a contractor recently released some of the images (CMPWG 2013). It is interesting to note one unique exterior anatomical feature of these phantoms: the arms are seperated from the trunk, unlike the MIRD-5 phantom and its successors developed around the same time. Two images of the CAM phantom are shown in Figure 7.

Figure 7.

Figure 7

The CAM phantom. (Left) The whole body view showing arms separated from the trunk. (Right)The close-up view of the facial details (http://cmpwg.ans.org).

4.1.4. MIRD Committee Work

Since the publication of the stylized dosimetry model of Snyder et al in MIRD Pamphlet 5 Revised (Snyder et al 1978), SNM’s MIRD Committee has refined several internal organs to support the development of radiopharmaceutical tracers and therapeutic nuclear medicine. Modifications to the MIRD stylized model have been published as MIRD Pamphlets, which include equations of the new geometries, tabulated absorbed fractions of energy for monoenergetic photons and electrons, and tabulated radionuclide S-values. In 1999, the MIRD Committee adopted 6 new age-specific models of the head and brain for a newborn, 1-year old, 5-year old, 10-year old, 15-year old (also representing the average adult female), and adult male (Bouchet et al 1999). Similar to previous stylized models, simplistic geometrical shapes were used to represent the different regions of the head and brain, with volumes derived from published reference masses and shapes from analysis of MRI images. Later, the MIRD Committee also adopted an age-dependent series of stylized kidney models that are used widely in therapy nuclear medicine for renal toxicity predictions (Bouchet et al 2003).

4.1.5. MCAT Phantom for SPECT and PET Imaging

The stylized modeling technique was also adopted by one group for medical applications. The Mathematical Cardiac Torso (MCAT) phantom, which includes the major thoracic structures and organs, was developed by a research group led by Benjamin Tsui (currently with Johns Hopkins University) at the University of North Carolina for use in nuclear medicine imaging research, specifically single-photon emission computed tomography (SPECT) and positron emission tomography (PET) (Pretorius et al 1997, Tsui et al 1993, Tsui et al 1994). The same group, especially Paul Segars (who was a Ph.D. student of Tsui at the time), later developed more advanced phantoms that are discussed later in this article.

4.1.6. Other Stylized Phantom Works

Table 1 also lists several additional efforts related to stylized phantoms. In the early 1990s, it was clear that the research community no longer favored stylized phantom modeling methods. However, several groups continued to develop stylized phantoms for particular methods. Two groups developed computational phantoms of an embryo and fetus for space radiation dosimetry (Chen 2004) and an adult representing the Korean population (Park et al 2006). A group at the Nagoya Institute of Technology developed two new stylized phantoms of a 9-month old Japanese infant (Hirata et al 2008). Around the same time, researchers at Tsinghua University in Beijing created a new mathematical phantom named the Chinese Mathematical Phantom (CMP) using anatomical data for the Reference Asian Man and the Chinese Reference Man (Qiu et al 2008). A new MIRD phantom based on reference data for the standard Korean male was developed at the Catholic University in Pusan, Bugok (Kim et al 2010). This latter phantom was used to model a patient implanted with 192Ir for brachytherapy treatment of prostate cancer. Bento et al (2011) at the Nuclear and Technological Institute (ITN) in Portugal also developed a new mathematical phantom to simulate the reference male BOMAB phantom. This phantom was used to simulate the detection of internal sources of radiation with a whole body counter (WBC). A series of four mathematical phantoms were developed at the Bhaba Atomic Research Centre in India to simulate the calibration of whole body monitoring systems for internal radionuclide contamination using BOMAB phantoms (Bhati et al 2011). More recently, Gardumi et al converted MIRD-type mathematical phantoms to NURBS before voxelizing them to investigate the so-called “voxel effect” which arises because of the stairstepped approximation of smooth surfaces with cubic voxels (Gardumi et al 2013).

For nearly 50 years since the first stylized phantom was reported, these anatomically simplified phantoms have been used as the de facto “standard” representations of the ICRP “Reference Man” methodology which is based on “population-average” 50th-percentile anatomical parameters specified in ICRP-23 (ICRP 1975) and ICRP-89 (ICRP 2002a). Applications of stylized phantoms have eventually included many aspects of radiation protection, radionuclide therapy, and medical imaging (ICRU 1992). In addition, national and international bodies have adopted organ dose estimates derived from these stylized phantoms in guidelines and regulations related to industrial and medical uses of ionizing radiation.

Although stylized phantoms made it possible to carry out Monte Carlo computations during times when computers were much less powerful, the original developers recognized the obvious shortcomings. Human anatomy is too complex to be realistically modeled with a limited set of surface equations. Many anatomical details in these models were compromised that sometimes led to inaccurate results. For example, when such phantoms were applied to nuclear medicine procedures where precise dosimetry is necessary, the calculated average organ and marrow doses did not produced strong correlations with observed marrow toxicity (Lim et al 1997). Most nuclear medicine physicians consequently tend to administer lower-than-optimal amounts of radioactivity to avoid toxicity. For CT dose reporting, most existing commercial software systems were based on the stylized patient models that are known to cause very large errors for low-energy x-rays (Gu et al 2008a). Similar stylized models have also been used to derive dose-response relationships for Japanese atomic bomb survivors and for medical patients in epidemiological studies. In the external-beam radiotherapy community, a stylized homogenous phantom was used by the Radiation Epidemiology Branch of the National Cancer Institute (NCI) in studies related to assessing secondary organ doses of theraputically irradiated patients (Stovall et al 1989). By the late 1980s, a few groups of researchers began to seek new ways to develop anatomically realistic phantoms. The underlying motivation was the belief was that new anatomically realistic phantoms would not only take advantage of improvements in computer modeling technology, but would ultimately lead to improved estimates for assessing the risks of patients or workers exposed to radiation.

4.2. Voxel Phantoms (1980s to Present)

The development of anatomically realistic models was desirable but impossible until early 1980s when powerful computer and tomographic imaging technologies became available. With the advent of medical imaging techniques such as CT and MRI, researchers could, for the first time, visualize the internal structures of the body in 3D and store the images in versatile digital formats. These advantages brought about the exciting and prolific era of the so-called voxel or tomographic phantoms. Table 2 summarizes a total of 84 phantoms that were constructed, typically from one of three types of tomographic images: CT and MR images of live subjects, and cross-sectional photographs of cadavers. In two previously published review articles, a total of 21 voxel phantoms was reported by Caon (2004) and 38 by Zaidi and Xu (2007). The notable increase in the number of phantom due to a more exhaustive literature search, recent developments, and the inclusion of phantoms developed for use solely in non-ionizing radiation applications.

Table 2.

Alphabetical listing of developers of voxel computational phantoms including information on the phantom names, phantom data types, the anatomical features modeled, the human subjects they mimic, whether they were designed for ionizing or non-ionizing radiation applications, and literature references.

Developers Phantom
Names
Data Types Human Subjects Anatomical Features Ionizing (I) or
Non-ionizing
radiation (N)
References
Austrian Institute of
Technology, Austria
MATSIM head
MATSIM torso
CT International Space Station
Astronaut
Based on the
MATROSHKA RANDO
phantom
I Beck et al 2011
Brooks Air Force Base,
USA
Visible Man Color photos Caucasian 39-year old
male cadaver
Visible Human Project.
More than 40 tissues were
identified.
N Mason et al 2000,
Wang et al 2004
Chang Gung
University, Taiwan
Taiwanese
Adult
CT 30 Taiwanese adults Whole body phantom (152
cm, 50 kg, Female)
I Tung et al 2011
China Institute for
Radiation Protection,
China
CNMAN Color Photos Chinese adult male cadaver Chinese Visible Human
Project
I Zhang et al 2007a
Darmstadt University
of Technology,
Germany
HUGO Color photos Caucasian 39-year old
male cadaver
Visible Human Project. A
total of 32 tissues were
identified.
N Gjonaj et al 2002
ENEA, Italy NUDEL CT Caucasian Male Based on a reconditioned
physical phantom named
AMOS
I Ferrari 2010
FCS Department, Italy DAM MRI 34-year old male volunteer Dielectric anatomical
phantom.
N Mazzurana et al 2003
Federal University of
Pernambuco, Brazil
MAX CT Caucasian adult male
patient
Based on VOXTISS8
phantom and adjusted to
ICRP-89 Reference Man.
I Kramer et al 2003
FAX CT Caucasian adult female
patient
Images of the trunk, the
neck and the lower part of
the head were from CT
scan of a 37-year-old
female. Images of the legs
and feet were from CT
scan of a 62-year-old
woman. The head and arms
were from MAX phantom.
I Kramer et al 2004
MAX06 and
FAX06
CT Caucasian adult male and
female patient
Extension of MAX and
FAX phantoms by adding
more details in the skeleton
to better match the ICRP-
89 Reference Man values.
I Kramer et al 2006
Flinders University,
Australia
ADELAIDE CT Caucasian 14-year old
female patient
Torso phantom, without
head and arms.
I Caon et al 1999,
Caon et al 2000
Graz University of
Technology, Austria
SILVY MRI, CT Caucasian pregnant
woman patient at 30 weeks
gestation
Trunk was based on MR
images of a pregnant
women and on modified
CT images of a woman in
the 30th week of
pregnancy developed by
RPI. The brain and spinal
cord were from NORMAN
and fitted into SILVY.
N Cech et al 2007,
Cech et al 2008
GSF-National
Research Center for
Environment and
Health, Germany
BABY CT Caucasian 8-week old
female cadaver
57 cm in height and
weighed 4.2 kg.
I Williams et al 1986,
Zankl et al 1988
CHILD CT Caucasian 7-year old
female leukemia patient
115 cm in height and
weighed 21.7 kg.
I Williams et al 1986,
Zankl et al 1988
DONNA CT Caucasian 40-year old
female patient
Whole body phantom (176
cm, 79 kg)
I Fill et al 2004,
Petoussi et al 2002
FRANK CT Caucasian 48-year old
male patient
Head and torso I Petoussi et al 2002
HELGA CT Caucasian 26-year old
female patient
From mid-thigh upwards I Fill et al 2004,
Petoussi et al 2002
IRENE CT Caucasian 32-year old
female patient
Whole body phantom (163
cm, 51 kg)
I Fill et al 2004,
Zankl et al 2002
GOLEM CT Caucasian 38-year old
male patient
The weight and height are
similar to those of ICRP 23
reference man.
I Zankl et al 2002
GODWIN CT Caucasian 38-year old
male patient
Modification of Golem
phantom to agree with the
ICRP 89 reference values.
I Zankl et al 2005
VISIBLE
HUMAN
CT Caucasian 39-year old
male cadaver
Head to knee. CT data
from the Visible Human
Project.
I Zankl et al 2002
LAURA CT Caucasian 43-year old
female patient
167 cm height and a
weight of 59 kg.
I Zankl et al 2005
KLARA CT Caucasian 43-year old
female patient
The modification of Laura
to agree with ICRP 89
reference values.
I Zankl et al 2005
KATJA MRI Caucasian pregnant
woman patient in her 24th
week of pregnancy
A woman in her 24th week
of pregnancy. Based on the
modified REGINA
phantom and patient MRI
images of the abdominal
and pelvic regions.
I Becker et al 2007
REGINA
(ICRP
Reference
Phantom)
CT Caucasian 43-year-old
female patient
An adjusted LAURA
phantom according to
ICRP 89 reference values.
I ICRP 2007,
Schlattl et al 2007
REX
(ICRP
Reference
Phantom)
CT Caucasian 38-year-old
male leukemia patient
An adjusted GOLEM
phantom according ICRP
89 reference values
I ICRP 2007,
Schlattl et al 2007
Hanyang University,
South Korea
KORMAN MRI Korean 30-year-old
healthy male
Korean male of average
height and weight.
I Lee et al 2004
KORWOMAN MRI Korean 35-year-old female Korean female of average
height and weight. Legs
were modeled from the
Visible Human Project
data.
I Lee et al 2005
KTMAN-1 MRI Korean 25-year-old male
volunteer
Korean Typical Man
(172 cm, 65 kg, without
arms)
I Lee et al 2006
KTMAN-2 PET and CT Korean 35-year-old male
volunteer
Korean Typical Man-2
(172 cm, 68 kg)
I Lee et al 2006
HDRK-Man Color photos Korean 33-year-old adult
male cadaver
High-Definition Reference
Korean Male phantom
from the Visible Korean
Human data.
I Choi et al 2006,
Kim et al 2008
Health Protection
Agency, UK
(formerly National
Radiological
Protection Board)
NORMAN MRI Caucasian adult male Normalized man. Only 10
sets of ribs rather the
traditional 12.
N, I Dimbylow 1996,
Dimbylow 1997,
Jones 1997
NAOMI MRI Caucasian healthy adult
female volunteer
Weight and height were
scaled to the values by
ICRP 89.
N Dimbylow 2005a,
Dimbylow 2005b
NORMAN-05 MRI Caucasian adult male Based on NORMAN with
new tissues recommended
by ICRP.
I Ferrari and Gualdrini 2005
Pregnant
female, hybrid
phantoms
(4 phantoms)
Quadric
equations and
MRI
Pregnant woman at 8, 13,
26, 38 weeks of gestation
Based on NAOMI and
Chen’s stylized fetal
phantoms
N Dimbylow 2006
Huazhong University
of Science and
Technology, China
VCH Color Photos Chinese adult male cadaver Visible Chinese Human
Project
I Zhang et al 2008a,
Zhang et al 2008b,
Zhang et al 2008c
VCH-FA VCH phantom Chinese female astronauts Based on cryosection
images, the VCH phantom
was constructed via
NURBS so the models
could deform to match the
body parameters of
Chinese female astronauts.
I Sun et.al 2013
Institut National de la
Santé et de la
Recherché Médicale
(INSERM), France
WBPM
(4 Phantoms)
CT 27-year-old male, 52-year-
old-female, two 3-year-
old-boys
4 phantoms of different
age and gender for
radiotherapy treatment.
Based on CT images of 4
volunteers
I Alziar et al 2009
Institut de
Radioprotection et de
Sûreté Nucléaire
(IRSN), France
Personalized
Voxel Phantom
CT Hispanic Male Male victim of a
radiological accident in
South America
I Courageot et al 2011
Japan Atomic Energy
Agency (JAEA), Japan
OTOKO CT Japanese adult male
volunteer
Japanese adult male voxel
phantom (170cm, 65kg)
I Saito et al 2001
JM CT Japanese 54-year-old male
volunteer
Japanese adult male voxel
phantom using a CT scan
in supine posture
I Sato et al 2007a
JM2 CT Japanese 54-year-old male
volunteer
The male subject recruited
for the construction of JM
was selected to obtain CT
scan in upright posture.
I Sato et al 2007b
ONAGO CT Japanese adult female
Volunteer
Japanese adult female
phantom (162cm, 57kg)
I Saito et al 2008
JF CT Japanese adult female
volunteer
Japanese adult female
phantom
(152cm, 44kg)
I Sato et al 2009
Korea Atomic Energy
Research Institute,
South Korea
Photographic
voxel phantom
Color Photos Korean adult volunteers A phantom was
constructed from
photographic images of
subjects. Phantoms were
modeled as homogenous
N Kim et al 2010
National Institute of
Information and
Communications
Technology (NIICT),
Japan
TARO MRI Japanese 22-year-old male
volunteer
Adult male phantom
(171.4cm, 65.0kg)
representing average
anatomical values of
Japanese 18-year-old male
N, I Lee et al 2006
HANAKO MRI Japanese 22-year-old
female volunteer
Adult female phantom
(159.1cm, 52.6kg)
representing average
anatomical values of
Japanese 30-year-old
female.
N, I Lee et al 2006,
Nagaoka et al 2004
Pregnant
woman
MRI Japanese 26-week-
pregnant woman volunteer
Based on the HANAKO
phantom and the
abdominal phantom of a
26-week pregnant woman.
N Nagaoka et al 2006,
Nagaoka et al 2007
Deformed
Children (3
phantoms)
MRI/FFD Japanese 3-, 5-, and 7-year
children
Transformed from the
TARO phantom into
children models using the
FFD algorithm.
N Nagaoka et al 2008
Oak Ridge National
Laboratory (ORNL),
USA
VOXMAT CT and
quadric
equations
Caucasian adult male Voxelized head and torso
phantom with stylized
arms and legs.
I Akkurt 2008
Rensselaer Polytechnic
Institute (RPI), USA
VIP-Man Color photos Caucasian 39-year-old
male cadaver
High resolution images
from Visible Human
Project
I Xu et al 2000
Pregnant
Woman
CT 30-week pregnant woman
patient
A pregnant woman
phantom covering the
abdominal region
I Shi and Xu 2004
RANDO CT
Phantom
CT Adult male A whole-body model of
the RANDO physical
phantom
I Wang et al 2004
Tsinghua University,
China
CVP MRI Chinese adult male
volunteer
Chinese Voxel Phantom
(170cm, 70kg).
I Li et al 2008,
Zeng et al 2006
University Hospitals of
Leuven, Belgium
Phantom 1 MRI 33-week old stillborn male
baby cadaver
Baby phantom (50cm,
1.91kg)
I Smans et al 2008
Phantom 2 CT 22-week old stillborn male
baby cadavers
Baby phantom (30.4cm,
0.59kg)
I Smans et al 2008
University of Florida,
USA
UF 2 month CT Caucasian 6-month old
male cadaver
A voxel phantom
equivalent to a 2 month old
male newborn,
representing a critically ill
child.
I Nipper et al 2002
UF Newborn CT Caucasian 6-day old
female newborn cadaver
A normal 6-day female
newborn phantom; The
lungs were created using
CT images of a 1-month
old patient and the adrenal
glands created using CT
images of a 2-month old
male patient.
I Nipper et al 2002
UF Series A
(5 phantoms)
CT 9-month, 11-, and 14-year
old males; 4- and 8-year
old females patients
UF pediatric phantom
series without arms and
legs.
I Lee et al 2005
UF Series B
(5 phantoms)
CT 9-month, 11-, and 14-year
males; 4- and 8-year
female patients
Based on the UF Series A
phantoms with arms and
legs from CT images of a
healthy Korean adult
attached. The organ masses
were adjusted to ICRP 89
reference values.
I Lee et al 2006
University of Houston,
USA
10 Year Old
Boy
CT 10-year old male Pediatric phantom
developed for craniospinal
proton irradiation.
I Taddei et al 2009
University of
Karlsruhe, Germany
MEET Man Color photos Caucasian 38-year-old
adult male cadaver
Models for simulation of
Electromagnetic,
Elastomechanic and
Thermic behavior of Man,
developed from the Visible
Human Project.
N, I Doerfel and Heide 2007,
Sachse et al 1997
University of Utah,
USA
Anatomically
based model
MRI Caucasian adult male
volunteer
Anatomic phantom. The
outer parts of the arms are
missing.
N Tinniswood et al 1998
Vanderbilt University,
USA
Gibbs
Phantoms
Radiography Caucasian representative
female cadaver
Phantom includes head,
trunk and proximal
extremities developed from
x-ray images
I Pujol and Gibbs 1982,
Gibbs et al 1987,
Gibbs et al 1984
Yale University, USA Zubal CT Caucasian adult male
patient
Head and torso I Zubal et al 1994
MANTISSUE3
-6
CT Caucasian adult male
patient
Arms and legs from the
Visible Human Project
were attached to the Zubal
phantom
N Dawson et al 1997
VOXTISS8 CT Caucasian adult male
patient
Arms and legs were
attached to the Zubal
phantom and the arms
were straightened along the
phantom side
I Sjogreen et al 2001

In terms of solid-geometry modeling techniques, a cubic voxel—one of the basic CSG primatives—is simply a 3D representation of a pixel; however, compared with the medical applications such as radiation treatment planning, the task of developing reference human phantoms presented some unique and intractable challenges: (1) to construct a whole-body phantom, image slices should ideally cover the entire body—a process not normally carried out in rountine medical examinations because of the need to minimize x-ray exposures or the lengthy time required for MRI procedures; (2) a large amount of internal organs/tissues must be identified and segmented for organ dose calculations, whereas, in radiotherapy, only the tumor volume and adjacent regions are routinely outlined; (3) the image data size of a whole-body model, especially when high-resolution images are used, can be potentially too great for a computer to handle; and (4) a standardized patient phantom is often used to study diverse radiation types such as photons, electrons, neutrons, and protons, thus requiring considerable Monte Carlo simulation capabilities.

In terms of the developmental process, tomographic phantoms are fundamentally different from the stylized ones. A tomographic image data set is composed of many slices, each displaying a 2-dimensional (2D) pixel map of the anatomy. The 3D volume of a voxel is measured by multiplying the pixel size by the thickness of an image slice. Unlike stylized phantoms, which are based on quadric surface equations, a voxel phantom contains a huge number of tiny cubes grouped to represent various anatomical structures. However, both quadric surface equations and cubic voxels belong to the same class of CGS geometries.

The creation of a tomographic phantom involves four general steps: (1) acquire a set of tomographic images (e.g. CT, MR, or anatomical photography) that cover the entire volume of the body; (2) identify (or segment) organs or tissues of interest (e.g. lungs, liver, skin, etc.) from the original image slice by assigning every pixel with an identification number; (3) specify the density (e.g., soft tissue, hard bone, air, etc.) and chemical composition of organs or tissues; and (4) Register the segmented image slices into a 3D volume that can be used for 3D visualization (for checking anatomical structures) and for Monte Carlo calculations. Figure 8 illsutrates these steps using the VIP-Man phantom (Xu et al 2000).

Figure 8.

Figure 8

Steps to create a voxel phantom using the Visible Human cadaver image dataset as an example.

The earliest effort to create an image-based phantom for radiation dosimetry is believed to have been reported by S. Julian Gibbs, a radiology professor at Vanderbilt University (Pujol and Gibbs 1982, Gibbs et al 1984, Gibbs et al 1987). In these pioneering studies, Gibbs and her co-workers explored the use of 2D x-ray images as the basis to form an anatomically realistic model of the patient. They used this information in Monte Carlo calculations to assess doses received by patients who underwent dental radiologic procedures.

Zankl and her colleagues at the GSF, Germany decided in the late 1980s to use 3D CT imaging on healthy volunteers and patients to develop what eventually became a family of 12 voxel phantoms: BABY, CHILD, DONNA, FRANK, HELGA, IRENE, GOLEM, GODWIN, VISIBLE HUMAN, LAURA, KLARA, and KATJA (Williams et al 1986, Zankl et al 1988, Smith et al 2000, Petoussi-Henss et al 2002, Zankl et al 2002, Fill et al 2004, Becker et al 2007, Zankl et al 2005). The adult male phantoms were developed first, followed by the adult female, pediatric, and pregnant-woman phantoms.

In its annual report for 2002, ICRP (2002b) states: “An important issue for Committee 2 is the substitution of an anatomically realistic voxel phantom, obtained digitally in magnetic resonance tomography and/or computed tomography, for the MIRD phantom which is a mathematical representation of a human body.” The ICRP Committee 2 has a task group on Dose Calculations (DOCAL) that was directly responsible for the development of a set of standard voxel phantoms. DOCAL is made of active researchers as members and consultants on internal and external dosimetry. The GOLEM and LAURA phantoms later underwent significant revision by the group led by Zankl at GSF, to yield the REX and REGINA phantoms which were released to the public as the ICRP adult Reference Male and Reference Female shown in Figure 9 (ICRP 2009, Schlattl et al 2007). Interestingly, ICRP did not endorse the first-generation stylized phantoms. Figure 9 shows the ICRP adult Reference Male and Female phantoms (ICRP 2009).

Figure 9.

Figure 9

ICRP adult Reference Male and Female that are based on earlier work at the GSF (ICRP 2009).

Several processes were considered to develop the ICRP reference phantoms: (1) CT image datasets of individuals close to the Reference Man and Woman (height and weight) were needed, (2) the datasets were segmented, (3) the body heights were adjusted to reference values by scaling the voxels, (4) the skeletal masses were adjusted to the reference values, and (5) individual organs were adjusted to reference values by adding and subtracting voxels. These processes were extremely time-consuming as the voxel data format is difficult to deform, unlike more advanced BREP geometries. While the ICRP reference phantoms filled a blank in the standardization of phantom-based radiation protection dosimetry, these phantoms had a relatively large slice thickness (up to 8 mm) compared to many phantoms reported later. At these voxel sizes, small organs cannot be realistically defined and the skin and walled organs reportedly contain small holes. To address this problem, A Korean group Yeom et al (2013) recently developed a polygon-surface version of the ICRP Reference Male by converting voxels to polygon-surfaces.

In 1994, Zubal et al (1994) from Yale University published a head-torso model called VoxelMan, which was developed from CT images. The original phantom was used for optimizing nuclear medicine imaging. Improvements to the original phantom were made with an MRI scan data of a human brain. This phantom is commonly known as the “Zubal Phantom” and registered users can freely download the original data from the internet. Two early users later revised the original data to report what are known as the MANTISSUE3-6 and VOXTISS8 phantoms by attaching arms and legs in two different positions to the original torso phantom (Dawson et al 1997, Sjögreen et al 2001). Adopting this publically available data, Kramer et al from Brazil developed an adult male phantom named MAX (Male Adult voXel) in 2003 (Kramer et al 2003) and later an adult female phantom named FAX in 2004 (Kramer et al 2004), both adjusted in accordance with ICRP-89 reference body height and organ masses. Kramer et al revised the skeletons (cortical bone, spongiosa, medullary yellow bone marrow, and cartilage) of MAX and FAX in 2006 to improve their compatibility with the latest ICRP-103 recommendations. These revised phantoms are known as MAX06 and FAX06. The work by Kramer et al is one of the earliest efforts to create a ICRP-89 compatible voxel phantoms for radiation protection dosimetry.

In 1996, Dimbylow from the National Radiological Protection Board (NRPB) (now known as the Health Protection Agency) in the United Kingdom reported the development of an adult male phantom known as NORMAN from MR images (Dimbylow 1996). NORMAN, which has a body height similar to the ICRP Reference Man, was first used by Dimbylow in a finite-element simulation code to determine the specific energy absorption rate from exposures to non-ionizing electromagnetic fields (Dimbylow 1997). In 1997, his colleague Jones adopted NORMAN to estimate organ doses from external and internal photon sources (Jones 1997). In 2005, Dimbylow developed an adult female phantom, NAOMI, also from MRI scans (Dimbylow 2005a). The phantom was rescaled to a height of 1.63 m and a mass of 60 kg, the dimensions of the ICRP Reference Woman. However, to date, the NAOMI phantom has been used only in non-ionizing radiation calculations. In 2005, a revised version of the NORMAN phantom, called NORMAN-5, was created by Ferrari & Gualdrini from ENEA-ION Istituto di Radioprotezione in Italy to derive external photon dose data (Ferrari and Gualdrini 2005). One year later, Dimbylow merged the NAOMI with the stylized fetal phantoms developed by Chen from Canada to create a series of hybrid phantoms for pregnant women (Dimbylow 2006). The process of adjusting two types of geometrical information was reported to be cumbersome.

In 1999, Caon et al from Flinders University in Australia reported a torso phantom named ADELAIDE created from CT images of a 14-year old girl (Caon et al 1999, Caon et al 2000). This phantom was interesting because, for some time, it was the only set of voxel data for a non-adult, and at the time, their studies likely provided the most reliable CT dose estimates for this patient group. Caon later reviewed his and other researchers’ experiences on voxel phantoms (Caon 2004).

The VIP-Man (Visible Photographic Man) voxel phantom was reported in 2000 by Xu and two of his students at Rensselaer Polytechnic Institute (RPI) in the U.S. (Xu et al 2000). VIP-Man was the first phantom based on cross-sectional color photographic images of a cadaver — a 39-year-old male through the National Library of Medicine’s famous Visible Human Project (VHP). The color transversal photos digitally captured at the 0.33 mm×0.33 mm pixel resolution and each photograph was taken after the removal (by shaving) of each successive 1 mm layer of the frozen cadaver by a cryomacrotome (Xu et al 2000). Although the original images were segmented to yield more than 1400 organs and tissues for the purposes of teaching anatomy (Spitzer and Whitlock 1998), only approximately 80 organs and tissues were adopted at RPI for radiation dosimetry purposes. Ultra-fine and color image analysis allowed the RPI group to explicitly segment a number of small and radiosensitive tissues including the stomach mucosa, skin, and red bone marrow. Given the extremely small voxel size, the VIP-Man phantom consists of more than 3.7 billion voxels—the most of any phantom reported at the time. The finalized VIP-Man phantom was unique because it represented an individual with a heavy body mass of 103 kg. Keith Eckerman from ORNL, who headed the ICRP DOCAL Committee at that time, encouraged this effort at RPI because VIP-Man could serve as an interesting variation from the ICRP reference value. However, because the phantom was developed from a cadaver, the lungs of the VIP-Man are deflated and smaller than might normally be expected for a living, breathing individual. Figure 10 is an image that highlights the anatomical differences between the stylized and the voxelized VIP-Man. The VIP-Man was used for a large number of studies in health and medical physics which are discussed on more detail later in this article. It is worth noting that VISIBLE HUMAN developed at GSF was based on CT images at 2–4 mm resolution of the same individual before the body was frozen. The RPI group later also reported a pregnant patient phantom using CT images of a 30-week pregnant female and compared internal dose data with those derived from a stylized phantom (Shi and Xu 2004; Shi et al 2008).

Figure 10.

Figure 10

Comparison of stylized adult phantom (left) and VIP-Man phantom (Xu et al 2000) (right) showing profound differecnes in anatomical detail. Such anatomical differences were believed effect the accuracy of radiation dose estimates.

Realizing the need for additional phantoms representing children of various ages, Bolch and colleagues from the University of Florida (UF) developed a series of pediatric voxel phantoms that appeared between 2002 and 2006, representing children with ages ranging from newborn to 15 years old (Lee et al 2005, Lee et al 2006, Nipper et al 2002). This approach was later extended to two groups (Groups A and B) of phantoms. Group A is composed of male and female voxel phantoms of a newborn, a 1-year old, a 5-year old, a 10-year old, and a 15-year old for whom the phantom stature, total weight, and individual organ masses are within 1% of ICRP-89 reference values. Group B phantoms are constructed by scaling the Group A phantoms up and down to yield phantom at each 1-year age interval, from newborn to 15-years old. The intent of the UF pediatric series was to provide a reference library of phantoms that could be matched to individual patients for age-specific organ dose assessment.

Two Japanese groups were noted in Table 2 for their independent efforts to develop voxel phantoms since 2001. Saito et al (2001) from the Japanese Atomic Energy Research Institute (JAERI) developed an adult male model named Otoko (the first Asian phantom) and an adult female phantom named Onago. More recently, Sato et al developed the JM, JM2, and JF phantoms which have a refined vertical slice thickness (Sato et al 2007a, Sato et al 2007b, Saito et al 2008). These phantoms were used mainly for radiation dosimetry applications in Japan were influenced by earlier projects at the GSF. The other group, Nagaoka et al, from the National Institute of Information and Communications Technology (NIICT) in Japan reported an adult male model, named TARO, and an adult female model, named HANAKO, developed from MR images for radio-frequency electromagnetic-field studies (Nagaoka et al 2004). Later Nagaoka et al (2008) would use a free-form deformation (FFD) to change the exterior features of the adult male phantom to develop Deformed Children phantoms of 3-, 5- and 7-year olds. The authors reported that it was difficult to develop these phantoms with the FFD algorithm and the internal organs are not adjusted to age-dependent values. The Otoko phantom was recently used in a study to calculate dose conversion coefficients for the Japanese population (Takahashi et al 2011).

Several Korean phantoms have been developed by researchers at Hanyang University in Korea from various image sources: Korean Man (KORMAN), Korean Typical MAN-1 (KTMAN-1), Korean Typical Man-2 (KTMAN-2), High-Definition Reference Korean (HDRK), and Korean WOMAN (KORWOMAN). The HDRK phantom was based on sectioned color photographs of an adult male cadaver that has high image resolution (Choi et al 2006, Kim et al 2008). The KTMAN-2 phantom has been used by Lee et al (2011) to measure the effects of selective collimation in cephalography. Kim et al (2010) from the Korea Atomic Energy Institute have developed a series of voxel phantoms of different body shapes in order to better calculate counting efficiencies for whole-body counters instead of using physical BOMAB phantom that has a fixed size.

Zhang et al (2009) summarized three Chinese voxel phantoms developed by three separate groups in China: Chinese Man (CNMAN) produced from color photographs of a cadaver by the China Institute for Radiation Protection (Zhang et al 2007a), Visible Chinese Human (VCH) produced from a different set of cadaver color photographs by the Huazhong University of Science and Technology (HUST) (Zhang et al 2007b, Zhang et al 2008a, Zhang et al 2008b), and Chinese Voxel Phantom (CVP) produced from MRI images by Tsinghua University (Li et al 2009, Zeng et al 2006). The lead developer of the CNMAN phantom, Binquan Zhang, served as a visiting scholar at RPI during 2007–2008. The Chinese government undertook the Chinese version of the Visible Human Project through the so-called Xiang Shan Conference in early 2000s that resulted in multiple cadaver image datasets, some with a slice thickness as fine as 0.1 mm. Figure 11 depicts the VCH phantom which is based on extremely high-resolution cadaver images (Zhang et al 2008b). The HUST group also reported a rat model (Xie et al 2010). Recently, Sun et al (2013) adopted NURBS to construct the VCH-F Astronaut (VCH-FA) phantom for dose calculations in the space radiation environment, incorporating statistical body characteristics of Chinese female astronauts as well as ICRP reference organ mass data. Tung et al (2011) from the Chang Gung University of Taiwan developed a voxel phantom of the Reference Taiwanese Adult using CT images from thirty Taiwanese adults.

Figure 11.

Figure 11

One of the Chinese phantoms—VCH phantom showing (left) internal organs, (middle) whole-body skeleton, and (right) vascular system (Zhang et al 2008b).

A voxel phantom named NUDEL (NUmerical moDEL) was developed by Ferrari in Italy (2010) for use in radiation protection studies. A computational phantom was constructed from CT data of the plastic physical AMOS (Anthropomorphic MOdel for dosimetric Studies) phantom. Dose calculations for several types of nuclide exposure were run in MCNPX code and were compared to experimentally measured values with the physical AMOS phantom. The calculations were also compared to values obtained from the NORMAN-05, GOLEM, DONNA, VOXELMAN, VIP-MAN, REX, and REGINA voxel phantoms.

A radiological accident that occurred in South America in 2009 prompted the construction of a personalized voxel phantom to numerically calculate the dose the victim received. Courageot et al (2011) of the Institute for Radiological Protection and Nuclear Safety (IRSN) converted CT scans into a voxel phantom using the Simulation of External Source Accident with Medical Images (SESAME) tool. Courageot et al (2010) reported the Simulation of External Source Accident with Medical Images (SESAME) tool that allows the use of NURBS to model a victim’s morphology and posture.

Researchers at the French National Institute of Health and Medical Research (INSERM), a public institute focused on biomedical research, designed a series of virtual whole-body patient models (WBPM) for usage in radiotherapy (Alziar et al 2009). They used CT data and the software tools IMAgo and ISOgray to model the phantoms. The phantoms accommodate different radiotherapy treatment positions, genders, and age groups. Alzier et al (2009) developed a software tool to take patient data and adjust the phantom’s anatomy in order to better match that of a specific individual.

Members of the MATSIM Project (MATROSHKA Simulation) at the Austrian Institute of Technology coordinated research to numerically simulate the effects of irradiation under reference radiation fields in outer space (Beck et al 2011). They created a two part voxel phantom using the FLUKA Monte Carlo code and CT images from a physical RANDO phantom. The voxel phantom was split into the MATSIM torso and head. The results of the simulations were within one standard deviation of experimental values. Taddei et al (2009) at the M.D. Anderson Cancer Center developed a voxel phantom in MCNPX code to assess the radiation does to pediatric patients receiving craniospinal irradiation with proton beams. An Iranian and Japanese team (Mofrad et al 2010a) developed a race-specified voxelized organ, specifically a Japanese male liver that contains statistical parameters, for nuclear medicine and internal dosimetry purposes. Patni et al (2011) of the Bhabha Atomic Research Centre in India published dose conversion coefficients obtained from the ICRP adult voxel phantoms. In 2008, Akkurt et al (2008) from ORNL reported their work involving a hybrid of voxel and stylized geometries.

4.3 BREP phantoms (2000s to present)

The past decade has seen a surprising surge in BREP phantom development. The number of such phantoms is increasingly each month. A non-inclusive list of 10 groups, reporting a total of 183 BREP-based phantoms, are summarized in Table 3.

Table 3.

Alphabetical listing of developers of BREP computational phantoms including information on the phantom names, phantom data types, the anatomical features modeled, the human subjects they mimic, whether they were designed for ionizing or non-ionizing radiation applications, and literature references.

Developers Phantom
Names
Data Types Human Subjects Anatomical Features Ionizing (I) or
Non-ionizing
radiation (N)
References
Duke University, USA
NCAT

NURBS

Caucasian 39-year-old male and 59-
year-old female
NURBS-based Cardiac Torso
phantom including organs from the
Visible Human Project CT data of the
male and female. Gated MRI dataset
of a normal patient and 3D angiogram
data are used for motion modeling.

I

Segars 2001
XCAT (47
phantoms)

NURBS
Caucasian 39-year-old male and 59-
year-old female
4D eXtended Cardiac Torso phantom
based on the NCAT, including more
detailed and realistic anatomy and
physiology.

I
Fung et al 2011

XCAT Library

NURBS
58 (35 male and 23 female)
anatomically variable phantoms
extend the XCAT beyond reference
anatomies by developing a series of
anatomically variable 4D XCAT adult
phantoms

I
Segars et al 2013
MOBY NURBS 16-week-old male C57bl/6 mouse Mouse phantom from MR images I Segars and Tsui 2007, Segars et al 2004

Federal University of
Pernambuco, Brazil
FASH and
MASH
Polygon
meshes

Anatomical atlases and models
Adult male and female phantoms
based off of anatomical atlases and
models. Organ masses were adjusted
to ICRP89 reference values.

I
Cassola et al 2010
Kramer et al 2010
FASH and
MASH series
(18 phantoms)
Polygon
meshes

Anatomical atlases and models
Male and Female phantoms for the
10th, 50th, and 90th mass and height
percentiles from the ICRP89 reference
values.

I
Cassola et al 2011
5 and 10 year
old Reference
Males and
Females
Polygon
meshes

Anatomical Atlases and models
Produced by the same methodology as
FASH and MASH. No patient images
were used.

I
de Melo Lima et al 2011
Hanyang University,
Korea
PSRK-Man Polygon
meshes
Korean Male Polygon mesh phantom based on the
VKH-MAN. Simulations can be run
without voxelization
I Kim et al 2011

Institut de
Radioprotection et de
Sûreté Nucléaire (IRSN),
France
Thoracic Female
Torsos
(34 Phantoms)
Polygon
meshes and
NURBS

ICRP Reference Female
Computational Phantom
34 phantoms of varying girth, cup
size, breast tissue, and internal organ
volume modified off of the ICRP
Reference Female Computational
Phantom

I
Farah et al 2010
Farah et al 2011
Adult Male
Whole Body
(25 phantoms)
Polygon
meshes and
NURBS
CAESAR database 25 whole body male phantoms
representing Caucasians of various
body types.
I Broggio et al. 2011
Foundation for Research
on Information
Technologies in Society
(IT’IS), Switzerland
The Virtual
Family (4
phantoms)
Polygon
meshes
Caucasian volunteers of different
gender and ages
Duke: 34-year-old male (174cm,70kg)
Ella: 26-year-old female
(160cm,58kg)
Billie: 11-year-old female
(148cm,34kg)
Thelonious: 6-year-old male
(107cm,17kg)
N Christ et al 2010
The Virtual
Classroom (4
phantoms)
Polygon
meshes
Caucasian volunteers of different
gender and ages
Roberta: 5-year-old female (109 cm,
17.8 kg)
Dizzy: 8-year-old male (140 cm, 26
kg)
Eartha: 8-year-old female (136 cm,
30.7 kg)
Louis 14-year-old male (169 cm, 50.4
kg)
N IT’IS 2011
Fats Caucasian volunteers of different
gender and ages
37-year-old male (182 cm, 120 kg) N IT’IS 2011
Glen Caucasian volunteers of different
gender and ages
84-year-old male (173 cm, 65 kg) N IT’IS 2011
John Hopkins University,
USA
Pediatric XCAT
phantoms (24
phantoms)
NURBS Caucasian 39-year-old male and 59-
year-old female
Used large deformation diffeomorphic
metric mapping to adjust the XCAT
phantom to math pediatric reference
data

I
Tward et al 2011

Rensselaer Polytechnic
Institute (RPI), USA
4D VIP-Man
Chest

NURBS

Caucasian 39-year-old male cadaver
Organ surfaces were extracted from
the VIP-Man phantom and then
extended to 4D by adding the
respiration of the NCAT phantom.

I
Xu and Shi 2005, Zhang et al 2008
RPI-Pregnant
Females
(3-, 6- and 9-
month)
Polygon
meshes

3-, 6-, and 9- month pregnant female
Based on a mixture of anatomical
data. Organs of the mother and fetus
were adjusted to match ICRP-89
references

I

Xu et al 2007
RPI-AM and
RPI-AF (2
phantoms)
Polygon
meshes

Adult Male and Female
Based on mesh anatomical models
that are adjusted to match with ICRP-
89 references. Software supports
deformation and posture change.

I
Zhang 2009,
Na et al 2010
Adult
deformable
Female breast
phantoms (8
phantoms)

Polygon
meshes

Adult deformable female breast
phantoms
Deformable phantoms based on the
RPI-AF phantom. Modified
according to ICRP reference data.

I

Hegenbart et al 2008
Obese Phantoms
(10 phantoms)
Polygon
meshes
Based on population surveys for class
1, 2, and 3 obese patients.
I Ding et al 2012

University of Florida,
USA
UFH-NURBS
phantoms (4
phantoms)

NURBS
Caucasian 6-day-old female newborn
cadaver, 14-year male patient and two
14-year female patients.
UF Hybrid NURBS based on previous
voxel phantoms, add 16 sites of
lymphatic nodes in 8 hybrid phantoms.

I
Lee et al 2007,
Lee et al 2008
Lee et al 2013
UFH-NURBS
fetal phantoms (18 phantoms)
NURBS 11 week and 21.5 week fetal
specimens
18 fetal phantoms of varying age and
weight with detailed soft tissue orga and developing skeletons ns
I Maynard et al 2011
University of Houston,
USA
Pregnant Female
(9 phantoms)
BREP/STL
formatted CAD
A pregnant woman in the 34th
gestational week and a non-pregnant
female
Nine phantoms of limited organs
covering 1- to 9-month pregnant
females from MRI of a non-pregnant
female and pregnant woman.

N

Wu et al 2006
Vanderbilt University,
USA
adult and
pediatric
phantom series
(7 phantoms)

NURBS
Caucasian adult male and female,
newborn, 1-, 5-, 10- and 15-year-old
Derived from the NCAT phantom
with organ and body masses adjusted
to match ICRP-89 references.

I
Stabin et al 2008

4.3.1. Work at UCSU, JHU, and Duke

Segars and Tsui (2009) summarized their work in a book chapter. Paul Segars’s Ph.D. thesis at the University of North Carolina was the first publication that systematically described the anatomical modeling using the NURBS-based techniques (Segars 2001). The well-known NCAT phantom was developed from the Visible Human CT image data set and the 3D anatomy was later extended into the 4th dimension to model cardiac and respiratory motions. The beating heart model of the 4D NCAT was based on 4D tagged MRI data from a real patient. The 4D NCAT phantom offers a vast improvement over the stylized MCAT phantom—a stylized version experimented on earlier by the same group—with more realistic models of the anatomy and the cardiac system, and the respiratory motion (Segars and Tsui 2009). The 4D NCAT has gained a widespread use particularly in nuclear medicine imaging research for evaluating and improving myocardial SPECT imaging. The conceptual design of the NCAT phantom also served as basis for the development of a 4D digital mouse phantom named MOBY (Segars et al 2004, Segars and Tsui 2007, Segars et al 2009). Figure 12 shows the original MIRD phantom together with MCAT, NCAT, XCAT, MOBY and ROBY phantoms provided by Paul Segars.

Figure 12.

Figure 12

Original MIRD phantom is shown with MCAT, NCAT, XCAT, MOBY and ROBY phantoms (Courtesy of Paul Segars).

Figure 13 shows the 4D Extended Cardiac-torso (XCAT) phantom family recently developed as the next version of the 4D NCAT provided by Paul Segars. The XCAT phantom family includes whole-body male and female anatomies based on the high-resolution Visible Male and Female anatomical datasets. In addition to the basic anatomy, the cardiac and respiratory motions were also updated in the XCAT phantom. The series includes 47 phantoms based on of the XCAT phantom representing the cardiac and repiratory motions of multiple patients. The XCAT phantom was mapped to patient CT data to produce the series, Segars ran simulations of PET, SPECT, and CT to demonstrate the applicability of the phantoms. Mishra et al used a modified XCAT phantom to evaluate 3D fluoroscopic image generation from a single planar treatment image (Mishra et al 2013). Then Segars et al (2013) extended the XCAT beyond these reference anatomies to a library of 35 male and 23 female 4D computational phantoms by developing a series of anatomically variable 4D XCAT adult phantoms for imaging research. The NCAT and XCAT phantoms have been used by other research groups to simulate radiation dose from radiography (Tabary et al 2009, Niu et al 2010) and radiotherapy (McGurk et al 2010). A research group constructed a version of the XCAT heart to enhance the range of cardiac disorders that can be studied using the phantom (Veress et al 2011). Tward et al (2011) developed a series of pediatric phantoms from a base adult XCAT phantom. An algorithm to modify the XCAT phantom was developed and used to generate 24 male pediatric patients with 8 organs each.

Figure 13.

Figure 13

A portion of the XCAT Phantom Family representing ages between newborn and 12 years-old. The phantoms can be adjusted to patient-specific information (Courtesy of Paul Segars).

4.3.2. Work at RPI

In 2005, the research group led by Xu at RPI used the VIP-Man phantom to simulate respiratory motions by adopting the gated respiratory motion data of the NCAT phantom (Xu and Shi 2005). The 4D VIP-Man Chest phantom was used to study external-beam treatment planning for a lung cancer patient (Zhang et al 2008c).

From 2005 to 2007, using the BREP modeling technique, Xu et al from RPI reported a set of phantoms at the end of three gestational periods of 3, 6 and 9 months—called RPI-P3, RPI-P6 and RPI-P9 (Xu et al 2007). Unlike the stylized models and the voxel models, these BREP-based models were found to be more flexible. These features allow BREP models to realistically change the size and shape for geometrically complex organs. Figure 14 shows the polygonal mesh model of the 9-month old fetus. Organs in the un-pregnant female are also individually adjusted to the ICRP-89 values and then deformed to allow for the fetus to be inserted using reference information and the help from an experienced anatomist.

Figure 14.

Figure 14

RPI-P phantoms for pregnant women. (a) BREP-type geometry of a 9-month old fetus in mesh format. (b) The mother and fetus after assembly showing the 3-, 6- and 9-month gestational periods (from left to right).

Continuing their BREP technique involving triangular meshes, the RPI group reported in 2008 the development of a pair of phantoms called RPI Adult Male and Adult Female (Zhang et al 2009b). Shown in Figure 15, this pair of adult phantoms was carefully adjusted to match the ICRP-89 reference values for more than 70 organs and 45 bones (including cortical bone, spongiosa, and cavities) as well as muscles. Several software algorithms were systematically developed to automate the deformation and organ overlap detection that were based entirely on about 126 sets of triangle meshes (Download a supplementary file “3DPhantoms.pdf” to this figure to interactively visualize the phantoms in 3D). In a subsequent work, as shown in Figure 16, the RPI Adult Male and Adult Female phantoms were extended into weight-specific phantoms representing the 5th, 25th, 50th, 75th, and 95th weight percentiles (Na et al 2010).

Figure 15.

Figure 15

The triangle mesh based RPI Adult Male and Adult Female phantoms (A supplementary file “3DPhantoms.pdf” to this figure is available for download that will allow a reader to interactively visualize the phantoms in 3D).

Figure 16.

Figure 16

The RPI Adult Male (top) and Adult Female (bottom) phantoms representing the 5th, 25th, 50th, 75th, and 95th weight percentiles (from left to right) (Na et al 2010).

The RPI Adult Female phantom was used to create phantoms of female workers with different breast sizes for the purpose of studying the effect of this parameter on the lung counting of internally deposited radionuclides (Hegenbart et al 2008). This was one of the first efforts to perform the so-called “virtual calibration” for bioassay measurement of internally deposited radionuclides in workers. Existing physical phantoms for this purpose come in a limited number of body sizes. In comparison, computational methods detector efficiency calculation provide a more convenient means of for determining the internal radioactivity content in people across a spectrum of different body shapes and sizes.

Using the same BREP deformable modeling methods, Ding et al (2012) modified the RPI Adult Male and Female phantoms to produce 10 phantoms representing overweight and obese individuals with a Body Mass Indices from 26 to 48. These phantoms consist of more than 100 deformable organs defined in the mesh-geometry format. Two main classes of adipose tissue in the human body were considered: (1) subcutaneous adipose tissue (SAT) located beneath the skin and (2) visceral adipose tissue (VAT) which surrounds the abdominal organs as illustrated in Figure 17.

Figure 17.

Figure 17

Illustration of the method to develop overweight and obese individuals by adding adipose tissues: (a) abdominal organs (surface rendering mode) and VAT (wireframe rendering mode) which surrounds the abdominal organs, (b) SAT layer beneath the skin, defined as the region between the body surface and internal body cavity.

Figure 18 shows 3D views of the RPI BMI-adjustable male and female phantoms. As the first ever set of phantoms for overweight individuals, these phantoms were applied in an interesting study of radiation dose received by patients undergoing CT examinations (Ding et al 2012).

Figure 18.

Figure 18

Phantoms for overweight and obese individuals. (Left) males, and (right) females. The phantoms have the same height (1.76 m for the male and 1.63 m for the female) but differ in weight. From left to right, the weight classifications are, normal-weight, overweight, obese level-I, obese level-II and morbidly obese (Ding et al 2012).

Posture-specific phantom is relatively rare, but is important to more realistically model how people interact with radiation in real-world environments. Han et al (2010) reported a pair of walking phantoms to represent individuals walking on a contaminated ground. Using the same method, Su et al (2012) then changed the posture of these phantoms to sitting. These are some of the earliest effort to design posture-specific phantoms, but the deformation was based on simple and unrealistic postures, as depicted in Figure 19.

Figure 19.

Figure 19

(Left) Walking adult male and adult female phantoms on contaminated ground with a step size of 70 cm and 45 cm, respectively. (Right) Sitting phantoms on a floor above nuclear medicine clinic.

To improve the posture-specific phantoms reported earlier at RPI, Vazquez et al (2014a; 2014b) developed two phantoms, called CHAD, with adjusted postures defined by a motion capture system. They applied the phantoms to simulate unique human postures found in a criticality accident that took place in 1997 in Sarov, Russia (Vazquez et al 2014a) and then a criticality accident at a JCO facility in Tokai-mura, Japan (2014b). Figure 20 shows the process to use motion capture to create a realistic sequence of worker movements.

Figure 20.

Figure 20

Motion capture technology was used to develop realistic posture sequence for a criticality accident. (a) A worker was exposed to criticality excursion and died 66 hours later. (b) An actor reconstructs the postures using motion capture. (c) The postures are recorded sequentially. (d) The CHAD phantom recreates the same sequential postures. (e) A total of 9 postures used for Monte Carlo dose calculations.

4.3.3. Work at the University of Florida

In a series of papers, the UF group led by Wesley Bolch reported their work on “hybrid” family phantoms of both genders and children at various ages (Lee et al 2007, 2008, 2010; Bolch et al 2010). They created the BREP phantom series, called UFH-NURBS phantoms using the following steps. First, they segmented patient-specific CT image data from which they then generated polygonal meshes. These meshes were then converted to the NURBS format using commercial software. They then extracted several contours from the polygonal meshes and generated the NURBS surfaces by a software tool called “lofting”. It was then in the NURBS geometrical domain they carried out organ adjustment to match the ICRP-89 reference values. In the final step, the NURBS–based phantoms were voxelized so that they could be implemented in Monte Carlo calculations. However, in order to voxelize the smooth NURBS models, they transferred the NURBS surfaces back to the polygonal meshes. In September 2008, ICRP established that its future reference phantoms for pediatric individuals would be based upon the UF series of hybrid phantoms. Recently, Geyer et al (2014) summarized their family phantoms and application to CT dose calculations at the Zurich workshop. Figure 21 shows the UF family phantoms developed using BREP methods (Bolch et al 2010).

Figure 21.

Figure 21

UF family phantoms developed from the BREP methods (Bolch et al 2010).

In 2011, the UF group (Maynard et al 2011) reported a family of NURBS based fetal phantoms. The phantoms were based on CT and MR images of fetus specimens of various ages between 10 and 30 weeks and were modified to conform to reference values. Tissues and organs were segmented using the modeling software 3D-DOCTOR and turned into polygon mesh surfaces. The models were then imported into Rhinoceros® 3D modeling software to incorporate NURBS surfaces and correctly orient the models.

One important contribution to the literature by the UF group is in the area of bone marrow dosimetry. For radiation protection purposes, photon or neutron dose response functions can be used to report active marrow and endosteum dose by tallying photon or neutron fluence in spongiosa regions of the skeleton (Eckerman 1985). With anatomically realistic phantoms developed at UF, Johnson et al. (2011) reported information to use the 3-factor method as an alternative to the dose response function for photon skeletal dose. The approach was also applied to neutrons by Bahadori et al. (2011).

The UF phantoms have seen wide use. The UF hybrid adult male phantom was used in a study by Johnson et al (2009) to calculate the effects of patient size on dose conversion coefficients. A model of electron dosimetry on infants based on the UF hybrid newborn phantom and an earlier developed skeleton tissue model (Pafundi et al 2009) was released by Pafundi et al (2010) from the UF. Hough et al (2011) released a model for skeletal based electron dosimetry in the ICRP reference male. CT scans of a cadaver were implemented in the Rhinoceros® 3D software to modify the UF hybrid male reference phantom to include segmented skeletal tissue. Dimbylow et al (2010) published a study that used the UF’s newborn NURBS based voxel phantom to calculate SAR for exposure to electromagnetic fields in the 20 MHz to 6 GHz region. Bahadori et al (2011) from the UF released a publication studying dose estimates from space radiation on astronauts. They modeled the astronauts by adapting the UF family of hybrid phantoms to the 5th, 50th, and 95th percentiles for 40 year old American Males and 40 year old Japanese females. Recently, a group led by Zaidi has used the UF phantoms for a number of nuclear medicine dose calculations (Xie et al 2013, Xie and Zaidi 2014).

4.3.4. Work at Vanderbilt University

The Vanderbilt group led by Michael Stabin, in collaboration with Segars from Duke University, reported a “family” of adult and pediatric phantoms by adapting the NURBS-based NCAT adult male and female phantoms (Stabin et al 2008; Stabin et al 2012). ICRP-89 reference body and organ values were used to adjust NURBS surfaces. The authors state several advantages of this approach: (1) NURBS-based phantoms can be developed much more quickly than working with voxels and manually segmenting individual patient image data sets; (2) The phantoms have a higher level of internal consistency; and (3) The phantoms are complete from head to toe, thus avoiding the problem of missing organs in some of the medical images. It should be noted that the groups at RPI, UF, and Vanderbilt (and Duke) developed these BREP phantoms as part of the joint Virtual Patients Project funded by the National Cancer Institute as well as other individual projects.

4.3.5 Work at the Federal University of Pernambuco

Led by Richard Kramer, a group at the Federal University of Pernambuco (UFPE) in Brazil has been active in developing BREP phantoms. Cassola et al (2010) reported two phantoms based on polygon mesh surfaces. The phantoms, FASH (Female Adult meSH) and MASH (Male Adult meSH), were constructed using software, including Blender, ImageJ, Binvox, and MakeHuman. The researchers based their phantoms on anatomical models and atlases, and showed that whole-body CT scans are unnecessary for phantom design. The organ masses were based on the values recommended for the male and female reference adult outlined in ICRP-89. Cassola compared FASH and MASH to the RPI-AF and RPI-AM phantoms and noted significant differences in anatomy. The UFPE group (Kramer et al 2010) made a series of calculations on the FASH and MASH phantoms. Large differences were observed compared to calculations done on the RPI-AM and RPI -AF mesh phantoms. Cassola et al (2011) continued the work on the FASH and MASH phantoms and published a library of 18 phantoms in 2011. The phantoms were adjusted based on reference values for the 10th, 50th, and 90th height and mass percentiles for Caucasian members of each gender. The reference values were obtained from the PeopleSize software package, which obtained the values from over 100 publications in North America, Asia, Australia, and Europe. In 2011, the group published 5 and 10 year old pediatric phantoms based on the same methodology that created FASH and MASH (Lima 2011). The phantoms were developed with polygon mesh surfaces in the modeling programs BLENDER and MAKEHUMAN and were edited in the programs DIP (Digital Imaging Processing) and QtVoxel. The researchers used ICRP data for the 5- and 10-year old reference children. Figure 22 shows the MASH phantoms organized by weight and height at the 10th, 50th, and 90th percentiles (Cassola et al 2011)

Figure 22.

Figure 22

The anthropomorphic MASH phantom organized by weight and height percentiles (Cassola et al 2011)

4.3.6. Work at IRSN

The group at IRSN developed a series of female torso phantoms in the Rhinoceros® 3D modeling software (Farah et al 2010a). A thoracic torso phantom was produced from mesh surfaces and NURBS, and was based on reference data from the ICRP adult female reference computational phantom. A series of 34 phantoms of differing girth, cup size, breast tissue composition, and internal organ volumes were created from the base phantom. They used the phantoms to ascertain the morphological dependence of counting efficiency curves from in vivo lung monitoring of workers (Farah et al 2010b). In 2011 they released a thoracic male phantom and a mesh equivalent to the physical Livermore phantom for the purposes of simulating in vivo measurements (Farah et al 2011). The phantoms were modeled with mesh and NURBS geometries. Data from CT and MRI scans were used to delineate organs. The data was then imported to Rhinoceros® 3D, where it was assembled into the two phantoms. Simulations of the two phantoms yielded comparable data to those done with voxel phantoms. The phantoms will be the basis for a new library of phantoms in a future study. A separate project at the IRSN produced a library of 25 whole body male phantoms in 2011 (Broggio et al 2011). The phantoms were produced from data in the CAESAR database, a compilation of male and female 3D models constructed from full body optical imaging. A total of 22 male Caucasian optical models were used as the basis for the phantoms. The phantom’s organs were constructed from ICRP reference data and added to the optical models. The phantoms possess a total of 109 segmented organs. The phantoms occupy a range of different body types, organ masses, and organ volumes. Figure 23 are the IRSN male phantoms with different body types from the CAESAR database, a compilation of surface models from full body optical images (Broggio et al 2011).

Figure 23.

Figure 23

Male phantoms of different body types based on the CAESAR database (Broggio et al 2011).

4.4. Non-ionizing radiation applications

Although not the focus of this article, voxel phantoms have also been used for non-ionizing radiation applications are listed in Table 2. Most of this work was neglected in the previous review articles by Caon (2004) and by Zaidi and Xu (2007), although the methods and approaches to phantom design are very similar. In fact, some phantoms, such as the NORMAN phantom, have been used for both ionizing and non-ionizing radiation applications. Table 2 clearly specifies whether the voxel phantoms that have been used for non-ionizing applications: the Visible Man from the VHP color photographs by the Brooks Air Force (Mason et al 2000, Wang et al 2004), the DAM adult male phantom from MR images by a group in Italy (Mazzurana et al 2003), the SILVY 30-week pregnant woman phantom from hybrid CT (originally obtained by RPI) and MR images by the Graz University of Technology, Austria (Cech et al 2007, Cech et al 2008), the MEET Man from VHP color photographs by University of Karlsruhe, Germany (Doerfel and Heide 2007, Sachse et al 1997), and the Anatomically Based Model from MR images by University of Utah (Tinniswood et al 1998).

Findlay and Dimbylow (2009) reported the specific absorption rate (SAR) for exposure to electromagnetic fields using the NORMAN phantom. Findlay and Dimbylow (2010) continued their work on SAR measurements and conducted a study of SAR in children due to Wi-Fi. He rescaled the sitting posture NORMAN phantom so that it matched ICRP reference values for a 10 year old child. The effects of electromagnetic fields from Wi-Fi devices operating at 2.4 and 5 GHz were modeled using a FDTD method.

Uusitupa et al (2010) measured SAR in the 300 to 5000 MHz region utilizing 15 voxel phantoms, including NORMAN, the Japan MALE/FEMALE, the VHP Male, and the VF series. The simulations were run with FDTD code on a HP supercluster at the Helsinki University of Technology in Finland. This study modeled the effects of different postures, human body models, and incoming direction of the electromagnetic field.

4.4.1. Work at IT’IS

Christ et al (2010) and Gosselin et al (2014) report the Virtual Family—a series of BREP-based phantoms developed for electromagnetic exposure calculations by the Foundation for Research on Information Technologies in Society (IT’IS). As shown in Figure 24, the Virtual Family consists of a 34-year old adult male, 26-year old adult female, 11-year old girl, and a 6-year old boy (Christ et al 2010). MRI images from volunteers were analyzed and segmented into 80 different tissues and organs using the imaging processing software iSEG. The boundaries between the tissues and organs were then remodeled using the software tool Amira. The Virtual Family is a part of the larger Virtual Population project at IT’IS. The Virtual Population project has developed 6 additional anatomic models using the same methods that were used with the Virtual Family (Gosselin 2014). The additional models consist of the Virtual Classroom, a series of four child models, and two individually developed models: an obese 37-year old male model, and an aged 84-year old male model.

Figure 24.

Figure 24

The Virtual Family: Duke, Ella, Billie, Thelonious (from left to right) by IT’IS (Christ et al 2010).

The second to last item in Table 3 is a series of 9 phantoms representing a pregnant female in each gestational month developed by a group led by Ji Chen from the University of Houston in collaboration with Wolfgang Kaine of the U.S. Food Drug Administration (FDA) (who was previously with IT’IS) for studying the effects of radiofrequencies emitted from various electronic devices (Wu et al 2006). These phantoms only include a limited number of organs such as the body, placenta, embryonic fluid, bladder, bone, fetus and the uterus. They used patient-specific MRI images and CAD software to specify the organ shapes.

Researchers do not consider the heart models in current phantoms to have enough clinical details. Gu et al (2011) of the Center for Devices and Radiological Health (CDRH) developed a series of high resolution heart phantoms for the purposes of accurate dosimetric calculations. The computational heart models were generated through a nearly automated algorithm created by the researchers that will allow the creation of new heart phantoms. The heart phantoms were inserted into the mesh based Virtual Family of phantoms for simulation. Aubert et al (2013) also built new hybrid computational phantoms (HCPs) with an inserted detailed heart model. The use of a detailed heart model eliminates the problem of identifying the coronaries on the patient's CT.

4.4.2. Work at Hanyang University

Current permutations of hybrid phantoms must be voxelized so that they may be used in Monte Carlo dose calculations. Voxelizing a BREP phantom reintroduces the majority of the limitations of the voxel phantoms. Researchers at Hanyang University in Korea have converted the voxel phantom VKH-Man into a polygon surface phantom using 3D-DOCTOR and directly implemented the phantom into Geant4 code in order to circumvent this limitation (Kim et al 2011). Calculational speed and accuracy on their new phantom, PSRK-Man (Polygon Surface Reference Korean Man) have been compared to the HDRK-Man phantom, which was also based on VKH-Man. The PSRK-Man phantom has overcome many of the limitations of a voxel phantom; however, the calculation speed for the phantom is 70–150 times slower than for its voxel counterpart HDRK-Man. The speed was significantly improved later when this group developed a method to calculate polygon-mesh geometry in GEANT4 code directly (Han et al 2013).

5. Physical Phantoms

Table 4 summarizes selected physical phantoms that are often used to benchmark calculations performed on computational phantoms. These phantoms are typically used for three different applications: external radiation dosimetry, internal radiation dosimetry, and imaging quality assurance. For external radiation dosimetry, a physical phantom is designed so that small thermoluminescent dosimeters (TLDs) (or ion chambers or solid-state detectors) can be inserted in different locations of the phantom to measure doses from external irradiation. Examples of this type of phantom include the RANDO® phantom by the Phantom Laboratory and the ATOM® phantom by the CIRS Inc., which contain tissue equivalent slices that have anatomical maps and cavities for organ dose measurements (Anderson et al 1962, CIRS 2013, Phantom Laboratory 2013). Phantoms for calibrating radiobioassay detectors or nuclear medicine imaging equipment are designed to contain either removable organs that are doped with long-lived radioactive materials or hollow body regions that are filled with short-lived radioactive liquids.

Table 4.

Alphabetical listing of developers of physical phantoms including information on the phantom names, phantom data types, the anatomical features modeled, the human subjects they mimic, whether they were designed for ionizing or non-ionizing radiation applications, and literature references.

Developers Phantom
Names
Anatomical Features Human Subjects Ionizing (I) or
Non-ionizing radiation
(N)
References
Alderson Research
Laboratories (acquired by
the Phantom Laboratory),
USA
RANDO® Lungs, soft tissue and breasts are included;
Natural human skeletons were used.
Caucasian adult male
and female
I Anderson et al 1962
Phantom Laboratory 2013)
Computerized Imaging
Reference Systems, Inc.
(CIRS), USA
ATOM® Bone, lung and soft tissue are included.
Standard phantom includes head, torso,
upper femur and genitalia. Legs and arms
are included with the newborn and 1 year
pediatric phantoms. Breasts can be added.
Caucasian newborn, 1-,
5-, and 10-year-old
children, adult male and
female
I CIRS 2013
3D Sectional
Torso Phantom
Removable lungs, heart, liver, pancreas,
kidney and spleen are included.
Caucasian adult male
torso
I CIRS 2013
Kyoto Kagaku Co. LTD,
Japan
PBU-50 The skeleton, lungs, liver, mediastinum
and kidney models are embedded in soft
tissue substitute.
Japanese adult male I Kyoto Kagaku 2013
CTU-41 One-piece anthropomorphic torso phantom
with anatomical structures.
Japanese adult male I Kyoto Kagaku 2013
Chest phantom
N1 “lungman”
The inner components consists of
mediastinum, pulmonary vasculature and
an abdomen block
Japanese I Kyoto Kagaku 2013
Lawrence Livermore
National Laboratory
(LLNL), USA
LLNL
(commercially
available through
Radiology Support
Devices)
Removable organs such as the lungs, heart,
liver, kidneys, spleen, tracheo-bronchial
lymph nodes are included. Chest plates
simulate different chest wall thicknesses.
The first generation of LLNL phantom
contains a real human male rib cage.
Caucasian adult male I Griffith et al 1978
Radiation Health
Research Institute of
Korea Hydro and Nuclear
Power, South Korea
Typical Korean
Male
Based on CT images of a human subject.
The phantom contains bone, lungs and soft
tissue without arms and legs. The rapid
prototyping and manufacturing technique
were used.
Korean adult male I Kim et al 2006
Radiation Protection
Bureau, Canada
BOMAB
phantom family
Each phantom is composed of 10 elliptical
containers representing head, neck, chest,
gut, arm, thigh, and calf.
Canada 5- and 10-year-
old, reference man,
reference woman, 5 and
95 percentile men
I Kramer et al 1991
University of Florida,
USA
MIRD stylized
Newborn
Based on stylized computational phantom. Newborn I Tresser and Hintenlang 1999
UF 1-year old CT images 1-year-old child I Tresser and Hintenlang 1999
UF Adult Male CT images Adult male Tresser and Hintenlang 1999
UF Newborn
physical
phantom
Based on UF Newborn voxel phantom
including soft tissue, bone and lungs.
Newborn cadaver I Jones et al 2006,
Staton et al 2006

These designs allow the phantoms to mimic internally contaminated individuals. The physical torso phantom by Lawrence Livermore National Laboratory (LLNL) and the Bottle Manikin Absorption (BOMAB) phantom family by the Radiation Protection Bureau, Canada are important examples of radiobioassay calibration phantoms. There are many phantoms that are used for medical image quality assurance purposes. Most of these phantoms, such as the NEMA image quality phantom, cover only partial body and some are anatomically very simple. Table 4 lists examples of several such phantoms by the CIRS and Kyoto Kagaku Co. that are used for image analysis. With anatomically realistic computational phantoms, the UF group led by David Hintenlang has fabricated several physical phantoms representing a newborn, 1-year old, and adult male (Hintenlang et al 2009). Rapid prototyping processes were also used to quickly produce physical phantoms from patient-specific data (Mille and Xu 2008). A detailed review of physical phantoms can be found in a newly released book edited by DeWerd and Kissick (2014).

6. Examples of Computational Phantom Applications by Students at RPI

A potential benefit of this review article is the opportunity to illustrate how computational phantoms have been used for radiation dosimetry. For expedience, projects carried out by my students at RPI since 2000 are used as examples here. The topics cover health physics, diagnostic imaging and radiotherapy. Collaborative projects involving non-radiation related research such as surgical planning is not included (Jin et al 2005).

6.1. Health Physics

Health physics dosimetry typically involves organ dose and effective dose quantities for external and internal sources under standard irradiation conditions. The VIP-Man model was used to compare dosimetry data from VIP-Man — a large sized individual of 40 years old — with other voxel phantoms including the ICRP Computational Phantoms. Details of these studies have been reported for different radiation types including photons (Chao et al 2001a, Xu et al 2005), electrons (Chao and Xu 2001, Chao et al 2001b), neutrons (Bozkurt et al 2000, Bozkurt et al 2001) and protons (Bozkurt and Xu 2004).

6.1.1. External Photon Dosimetry

Using the VIP-Man phantom, Chao et al reported a new set of conversion coefficients from kerma free-in-air to absorbed dose and kerma free-in-air to “effective VIP-Man dose” for external monoenergetic photon beams from 10 keV to 10 MeV (Chao et al 2001a, and a later correction by Chao et al 2003). This study noted that kerma approximation, which assumes secondary electrons from photon interactions deposit their energies at an interaction site, could lead to potential uncertainty for high-energy photons incident on shallow tissues (such as breast, skin, eye lenses, or gonads). The study concluded that the size of the model, kerma approximation, and the anatomical difference were three main factors in causing dosimetric discrepancies. These comparisons also suggested possible ways to improve the stylized models. For example, the stomach is situated too close to the left side of the body when compared to VIP-Man.

Han et al (2010) used the walking phantoms to calculate environmental exposures involving parallel and isotropic planar sources of 137Cs and 60Co with concentrations of 30 kBq/m2. For the parallel plane source case, the organ doses were found to be up to 78% greater for walking phantoms than those for the stationary phantoms with legs together. The dose difference is due to that fact that widely open legs during walking provide less shielding to several organs, especially the kidneys, ovaries, and liver, from parallel sources on the ground. The effective doses of the walking phantoms were on average 15% higher than standing phantoms. On the other hand, when isotropic surface contamination sources were considered, no significant dose difference was observed between phantoms with different postures. This study demonstrated the feasible to use deformable phantoms to represent realistic postures for organ dose calculations in environmental dosimetry studies. Similar findings were reported by Su et al (2012) for sitting phantoms above a nuclear medicine clinic where positron emission tomography (PET) imaging is used.

6.1.2. External Electron Dosimetry

Chao et al (2001b) presented organ doses from the VIP-Man phantom from external electron beams using EGS4-VLSI Monte Carlo code and compared data with those reported for the ADAM phantom using MCNP4 code by Schultz and Zoetelief (1996), a hermaphrodite mathematical model using FLUKA by Ferrari et al (1997), and the MIRD-5 mathematical model using EGS4 by Katagiri et al (2000). These comparisons suggested, that at least for electron dosimetry, a single standard body model does a poor job in representing individuals of diverse anatomy. The study further concluded that a large number of voxel phantoms would need to be investigated before the degree of dose variation is understood.

6.1.3. External Neutron Dosimetry

Using the VIP-Man phantom, Bozkurt and his co-authors reported a new set of fluence-to-absorbed dose and fluence-to-effective dose conversion coefficients calculated for both low-energy (10−9 – 20 MeV) and high-energy (20 – 10000 MeV) neutrons (Bozkurt et al 2000, Bozkurt et al 2001). The absorbed dose for 24 major organs and effective dose results based on the realistic VIP-Man were presented and compared with those based on the simplified MIRD-based phantoms reported in literature. The authors noted discrepancies between the doses calculated on the two phantoms and concluded that several factors may have contributed to the discrepancies. The differences in anatomical models, which cause approximately a 10% difference in effective dose, are because the VIP-Man is heavier and taller and how the Monte Carlo codes treat the transport of high-energy particles, including the use of evaluated data and theoretical models.

6.1.4. External Proton Dosimetry

Bozkurt and Xu (2004) applied the VIP-Man phantom to calculate fluence-to-absorbed dose and fluence-to-effective dose conversion coefficients under high-energy proton environment. The absorbed dose results were presented for 24 major organs of VIP-Man, and the calculated data were compared with those based on mathematical phantoms reported in literature. Some discrepancies in organ dose and effective dose, within 40%, were observed due to the use of different transport models employed by different Monte Carlo codes. Taranenko and Xu (2009) used the RPI-P phantom series to calculate conversion coefficients for fetuses from whole body irradiation with monoenergetic proton beams. The simulation was run in MCNPX for 12 different source energies ranging from 100 MeV to 100 GeV, and for 6 different configurations.

6.1.5. External Dosimetry for Red Bone Marrow

Caracappa and co-workers used two sets of Visible Human images for the identical anatomy to gain insight to the external dose to the red bone marrow, which is the most radiosensitive tissue (Caracappa et al 2009). A Monte Carlo computational model was constructed in this study from the CT images of the Visible Human Project, and compared to the VIP-Man phantom derived from color photographs of the same individual (Xu et al 2000). These two data sets for the same individual offered interesting information that was not previously available. Dose to the red bone marrow was calculated for the CT model using the uniform mixture assumptions and using the cellularity factors adopted by ICRP to test the previous assumptions and evaluate the accuracy of the computed dose to the red bone marrow in Monte Carlo simulations. Based on the newly developed algorithms, three dosimetry applications were investigated and tested. Broad beam photon irradiation in occupational exposure results in similar doses for high energies, but differences as great as 40% for low energies. An electron total-body irradiation procedure for treating skin cancer was also studied, with a 39% difference in red bone marrow dose between the existing method and the proposed revised method. These results demonstrate the advantage of the new algorithms by accounting for marrow cellularity and distribution various bone sites in the anatomical and dosimetry models.

6.1.6. Internal Electron Dosimetry

Based on the VIP-Man phantom, Chao and Xu calculated complete sets of specific absorbed fractions for internal electron emitters (Chao and Xu 2001). This was also the first report of internal electron data for walled organs such as the esophagus, lower large intestine, stomach, and upper large intestine. Although electrons are considered as weakly penetrating radiation and researchers have usually ignored the dose to organs other than the source organ, results from this study show that doses to neighbor organs and nearby organs can be too great to be neglected. This study provided convincing evidence that internal electrons do affect organs beyond the source organ.

6.1.7. Internal Photons Dosimetry for GI tract

In this study, the VIP-Man phantom was used to calculate SAFs for the gastrointestinal (GI) tract (Xu et al 2005). SAFs for sources in GI tract have been previously studied based on stylized phantoms. Using the VIP-Man phantom, the authors compared SAFs for stomach wall from VIP-Man to those previously published by Cristy and Eckerman for photon sources in the stomach content. The stylized models have been widely utilized by the nuclear medicine dosimetry community. However, SAFs derived from this phantom can have considerable uncertainties when compared to the realistic VIP-Man under certain conditions. This study clearly demonstrated the advantage of the VIP-Man phantom whose small voxel size allows the dosimetry to be performed on small tissues structures such as the mucosal layer in the GI tract.

6.1.8. Dynamic phantoms for criticality accident dose reconstruction

Using the CHAD phantoms developed motion-capture data, Vazquez et al (2014a; 2014b) simulated how workers were fatally exposed to extremely high levels of radiation. Implementation of the emergent techniques produced more accurate and more detailed dose estimates for the workers than were reported in previous studies. In Vazquez et al (2014b), a total-body dose of 6.43 and 26.38 Gy was estimated for the two workers, who assumed a crouching and a standing posture, respectively. Additionally, organ-specific dose estimates were determined, including a 7.93 Gy dose to the thyroid and 6.11 Gy dose to the stomach for the crouching worker and a 41.71 Gy dose to the liver and a 37.26 Gy dose to the stomach for the standing worker. Implications for the medical prognosis of the workers are discussed, and the results of this study were found to correlate better with the patient outcome than previous estimates, suggesting potential future applications of such methods for improved epidemiological studies involving next-generation computational phantom tools.

6.2. Radiological Imaging

6.2.1. Organ doses from SPECT and PET brain imaging

To estimate internal dosimetry for brain imaging, a head and brain portion of the VIP-Man was used to implement into the Monte Carlo code, EGS4-VLSI (Chao and Xu 2004). Fifteen sub-regions were modeled including caudate nucleus, cerebellum, cerebral cortex, cerebral white matter, corpus callosum, eyes, lateral ventricles, lenses, lentiform nucleus, optic chiasm, optic nerve, pons and middle cerebellar peduncle, skull CSF, thalamus, and thyroid. S-values were calculated for the most important sources and targets encountered in SPECT and PET brain imaging. These results were then compared to those from the stylized head/brain model recommended by the MIRD (Bouchet et al 1996). Although heavier individuals (e.g. VIP-Man) will usually receive lower radiation doses, the stylized head/brain model underestimates the S-values by 15% on average for a patient similar to the VIP-Man model. More tomographic head/brain models are needed in order to compare various brain sizes and anatomical variations. Before such an inter-comparison is performed, the results presented in this paper are useful for patients who are similar to VIP-Man in body size or weight.

6.2.2. Organ doses from X-ray Radiographs

VIP-Man was used by Mark Winslow, a Ph.D. student at RPI, in collaboration with Walter Huda from University of Syracuse, to calculate values of energy imparted (ε) and effective dose (E) for monoenergetic photons (30–150 keV) in radiographic examinations. Energy deposition in the organs and tissues of the human phantom were obtained using Monte Carlo simulations. These monoenergetic E/ε values can generate values of E/ε for any x-ray spectrum and can be used to convert values of energy imparted into effective dose for patients undergoing common head and body radiological examinations (Winslow et al 2005). Later, in his doctoral research, Mark Winslow, in collaboration with Birsen Yazici of RPI, also studied the image quality by analyzing approximately 2000 simulated chest x-ray images for the VIP-Man using the ROC/AUC analysis (Son et al 2006, Winslow et al 2006).

6.2.3. Organ Dose from CT

Gu et al (2009) used the RPI Pregnant Female phantoms to run dose calculations for multi detector CT (MDCT) scans. The MDCT scanner and the phantoms were implemented in MCNPX code. The dose profiles showed that there was little risk to the patient or the fetus from the MDCT scans. Ding et al (2012) developed 10 obese phantoms for the purpose of optimizing image quality and CT dose in obese patients. It was found that calculated dose for obese patients differed significantly from the dose calculated for normal weight phantoms.

6.3. Radiotherapy

6.3.1. Adjoint Monte Carlo method for external-beam prostate radiation treatment

The abdominal portion of the VIP-Man model was used in the doctoral research by Brian Wang, a Ph.D. student at RPI, to develop and demonstrate an Adjoint Monte Carlo (AMC) method for optimizing the external beam directions in the so-called 3D Conformal Radiation Treatment of prostate cancer (Wang et al 2005a). The AMC method had been widely used in nuclear reactor physics research but was never demonstrated for treatment planning in realistic 3D patient anatomy. With the VIP-Man model, which was already implemented in the MCNP code with multi-group adjoint cross-sections, it was possible to test the theory in clinically relevant scenarios. This work was in collaboration with Moshe Goldstein, a nuclear engineer from Israel who first proposed the possibility during a sabbatical at ORNL, and Narayan Sahoo, who was a clinical therapeutic physicist at Albany Medical Center at the time. In this application, the adjoint fluxes for the prostate (PTV) and the rectum and bladder (OARs) in the VIP-Man phantom were calculated on a spherical surface of approximately 1-m radius, centered at the center of gravity of PTV (Wang et al 2005a). An “importance” ratio, defined as the PTV dose divided by the weighted OAR doses, was calculated for each of the available beamlets to select the best beam angles. Finally, the doses in PTV and OAR were calculated using the forward Monte Carlo method. This study demonstrated the feasibility of the AMC method in optimizing external beam directions based on anatomical information in a 3D and realistic patient anatomy. The study also identified issues to be further addressed before this method could become clinically useful (Wang et al 2005b).

6.3.2. Non-target organ doses from proton radiation treatments

In a separate study, doctoral student Brian Wang, worked with Harald Paganetti of Massachusetts General Hospital to adopt the VIP-Man phantom to assess organ doses and the associated risk for developing secondary cancer after proton radiation treatment (Jiang et al 2005).

6.3.3. Respiration management in IGRT

After graduating from RPI, Chengyu Shi worked with RPI doctoral student Juying Zhang, to apply the respiration-simulating 4D VIP-Man phantom for Image-Guide Radiotherapy (IGRT) of lung cancer (Zhang et al 2008). To extend the geometry-based modeling to physics-based modeling, Eom et al (2010) introduced the use of finite element analysis to develop patient-specific phantoms that simulate respiratory motions in a predictive manner.

6.3.4. Brachytherapy dosimetry

Using the RPI Adult Female phantom, graduate student Matt Mille collaborated with Mark Rivard of Tufts University (Mille et al 2010) to simulate patients undergoing balloon brachytherapy of the breast. Monte Carlo simulations were performed to compare doses from treatments involving 192Ir or electronic sources using Monte Carlo simulations. This work helped demonstrate that the recently developed miniature x-ray sources may offer a more optimal treatment for the patient because of the lower radiation doses received by organs and tissues far from the treatment site.

6.3.5. Imaging doses in IGRT for Radiotherapy

Mr. Jianwei Gu and workers used the VIP-Man phantom to calculate organ doses from image-guided radiation treatment (IGRT) (Gu et al 2008b). Two imaging procedures were considered: kV Cone Beam CT and MV Cone Beam CT. The results indicate that thyroid received the highest dose in head and neck scans for both kV and MV CBCT, and the bladder receives the highest dose in prostate scan for both kV and MV CBCT. The effective doses for H&N scan and for prostate scan are at the same level in both kV CBCT and MV CBCT. This study provided a method to compute organ doses and effective dose that are useful in treatment planning and risk assessment. A second study performed calculations on the prostate, simulating kV CBCT and MDCT (Ding et al 2010). The calculations indicated that the imaging dose from standard IGRT procedures is high enough to warrant modifications to the procedure.

6.3.6. Proton Radiography

In his Ph.D. work, Bin Han collaborated with George Chen from MGH to use Monte Carlo simulations to evaluate the performance of a time-resolved proton range telescope (TRRT) (Han et al. 2011). This was done by tracking 230-MeV protons as they passed through position detectors, a patient 4DCT phantom, and scintillation detectors. The proton water equivalent length (WEL) was deduced using a reconstruction algorithm that incorporated the linear proton track to improve image quality, and three patients’ 4DCT images were used to measure WEL variations and tumor motions. The results from the simulations showed that the tumor trajectories from the WEL map agreed with direct 4DCT measurements within one millimeter.

7. Discussion

It is possible that Tables 13 do not include all phantoms ever reported in the literature. The grouping of these phantoms into three types may not be always accurate since it was done using author supplied descriptions. Nevertheless, these tables allow us to plot and analyze the trends, as plotted in Figure 25. It can be observed that, interestingly, only a total of 38 stylized phantoms were reported in the 50 years since the phantom was first developed in the 1960s. The work on stylized phantoms peaked in the 1980s with the publication of the so-called Cristy and Eckerman stylized family phantoms (Cristy and Eckerman 1987) which were widely adopted as the de facto standard in radiation protection dosimetry. The 2nd generation voxel phantoms surfaced in the late 1980s and gradually reached the peak in the middle of 2000s with a total of 85 voxel phantoms as of 2014. In comparison, the BREP phantoms did not emerge until early 2000s and, out of the total of 287 BREP phantoms as of 2014, the majority were reported within the past several years. When data for three phantom types were shown together in Figure 25, a surprising pattern of exponential growth in research activity becomes apparent. Given the fact that the radiation protection dosimetry community relied on a dozen stylized phantoms for decades, one might have probably predicted a similar trend for voxel or BREP phantoms. In this case, we are wrong about the phantoms as we are often wrong about the general trend for technology, according to Ray Kurzweil, the author of the best-seller “The Singularity Is Near: When Humans Transcend Biology.”

Figure 25.

Figure 25

The number of phantoms in existence since 1966, showing a somewhat surprising exponential growth due to the rapid increase in voxel and BREP phantoms in recent decades (Note: once a phantom is reported in the literature, it is counted in subsequent years when plotting this figure).

An important question is: Why computational phantoms have evolved the way they did? Mathematical formulations of organs and tissues of the body used in the dosimetry of internally distributed radionuclides existed as early as the 1940s, although the first anthropomorphic phantom was not reported until the 1960s. In the 1970s and 1980s, the sophistication of these stylized phantoms was increased significantly. This evolution began with the specification of a single organ mass, followed by the use of simple shapes to simulate organs or the entire body of an adult human. The desire to model the entire body of the “Reference Man” and to specify the location, shape, volume and mass of organs in the body as realistically as possible has remained the same to this day. The climax for stylized phantoms was reached in the 1980s when the gender- and age-specific family phantoms were systematically documented (Cristy and Eckerman 1987) and widely adopted for various studies in internal and external radiation dosimetry, as well as in medical imaging and radiotherapy. By that time, Monte Carlo codes and personal computers had become accessible to a large number of researchers. I did my PhD research at Texas A&M University using two dozen PCs with Intel 486 processors and MCNP Version 3 (Reece et al 1994, Xu et al 1995, Xu et al 1996, Reece and Xu 1997).

The research on stylized human models at ORNL up to the 1980s played an essential role in the history of computational phantoms. The sex-specific adult phantoms at GSF in the early 1980s were revisions of the MIRD-5 phantom originally developed at ORNL. Major extensions in the 1990s, on the pregnant women and brain/head models, were also closely tied to the earlier work at ORNL. The direct involvement of ORNL’s scientists in the Society of Nuclear Medicine (SNM)’s MIRD Committee facilitated the necessary standardization process. It is clear that close collaborations between leading developers were a key factor contributing to the success of first-generation computational phantoms. Not all phantoms of this generation enjoyed the same recognition in the history, however. In fact, CAM and CAF phantoms were practically unknown to people outside the NASA community for decades. The early 1990s marks the beginning of an exciting new era of voxel phantoms. With easy access to rapidly advancing computers and medical imaging technologies, computational phantom research was no longer dominated by a few groups.

The shift from stylized phantoms to voxel phantoms in the late 1980s was initially motivated by the desire to improve upon anatomical realism. The advent of modern computers and medical imaging fueled the researcher-initiated efforts. During the 1990s, however, it was unclear to the research community what roles voxel phantoms would play. If voxel phantoms were to replace stylized phantoms as the standard in radiation protection, how much improvement in dose estimates should be expected? There were already strong indications that the methods used in voxel phantoms were not ideal, as observed by Caon (2004) and Zaidi and Xu (2007). For example, the segmentation of original images into organs and tissues often required a laborious and tedious manual process, requiring months or years to complete. Some of the voxel phantoms were based on relatively large image slice thickness, thus the anatomical accuracy of such phantoms was inevitably compromised. Even today, there is no consensus as to what constitutes a faithful segmentation procedure in creating these whole-body voxel phantoms because the process often required assumption about the anatomy during the image analysis. Certain organs in the stomach and GI tract have poor image contrast and, as a result, the segmentation is nearly impossible in CT without the use of contrast agent. Then, there is also a question of how small the voxel size should be. While voxels at 2 mm × 2 mm × 2 mm seem to do a good job representing most organs, they are not fine enough for some small and radiosensitive organs to be delineated. For this reason, the skin of most existing voxel phantoms is defined artificially as the outermost voxel layer in a phantom. The segmentation of two radiosensitive tissues—the red bone marrow and bone surface that are explicitly recommended for radiation protection dose calculations—is especially difficult. Consequently, doses to the red bone marrow and bone surface are always calculated using algorithms involving other parts of the bone. It is worth noting that phantoms based on cadaver images provided an opportunity to segment the red bone marrow explicitly. For example, Xu et al (2000) reported for the first time the whole-body red bone marrow distribution by harvesting color pixels of 0.33 mm × 0.33 mm resolution. Later the method was adopted in the development of several cadaver-based phantoms (Zhang et al 2009).

The lack of standardized procedures contributed to the current situation that although many phantoms and dosimetry data are reported, the accuracy may be impossible to compare. Voxel phantoms were realistic in depicting the anatomy, but they are also tied to a specific individual (thus conceivably not representative for a large group of people). The degree of anatomical differences between two equally realistic voxel phantoms surprised many developers who were used to the idea that a single radiation protection phantom could represent the average population. The “Reference Man” methodology required a computational phantom to match the 50th percentile values in terms of body height and weight for a specific gender and age group. From the literature review, it was found that many developers later rushed to revise the original voxel phantoms by adjusting the organ sizes in the original image data to match with the ICRP-89 recommended anatomical data. In doing so, these voxel phantoms in fact lost their individual realism — a shortcoming that was associated previously with the stylized phantoms.

It is important to note the differences between “population-averaged” prospective dosimetry needed for radiological protection under the current ICRP radiation protection system and “individualized” retrospective dosimetry needed for accident dose reconstruction, medical dose tracking, or epidemiological study1. The ICRP system assumes that the workers and members of the public are adequately protected as long as the organ doses and effective doses estimated for the Reference Man — a hypothetical gender-averaged adult person whose relevant anatomical and physiological parameters are at the 50th percentile of the population defined by ICRP 89 — are kept below the ICRP dose limits. In other words, the ICRP system of radiological protection does not require the assessment of the “real dose” to every exposed person. To follow ICRP recommendations for radiation protection dosimetry, one would need to (1) assess external exposures (e.g., by measuring quantities that can be related to a physical quantity such as fluence) and internal intakes (e.g., by performing bioassay for the amount of radioactivity inside the body) and then (2) convert these quantifies using ICRP reference dose coefficients (e.g., effective dose per fluence or per Bq inhaled) to yield estimates of organ dose or effective dose (ICRP 2007). Similarly, the use of a personnel dosimeter would require the comparison of the “deep dose equivalent” measured at 1-cm in tissue equivalent materials with “effective dose”. These dose coefficients, but design, should be calculated using only ICRP Computational Phantoms (e.g., 50th percentile reference phantoms) and reference biokinetic models (ICRP 2009). This means that non-reference phantoms reported in this review article should not be used to assign values of effective dose according to the design of the ICRP radiation protection system. So what is the usage of these phantoms reported in this article? To ICRP, these non-reference phantoms only helped to understand potential uncertainties in using the ICRP Computational Phantoms as the standard models. However, when the effective dose to the Reference Worker is deemed to be high compared to the dose limits such as in the case of an accident or in the case of assessing the dose-response functions during an epidemiological study, one should be encouraged to assess the actual and true organ doses to the exposed individual. At this point, any individual-specific computational phantoms representing an individual of both genders at different ages, heights, and weights can and should be used. Similarly, the ICRP radiation protection system does not apply to medical radiation exposures. A good example is the need to report CT doses to patients who are not subjected to the ICRP occupational dose limits. In that and other potential medical dosimetry applications, the researcher is no longer obligated to restrict their calculation to ICRP Computational Phantoms or reference biokinetic models. In fact, non-reference families of phantoms have been used for CT dose reporting (see www.virtualphantoms.com). Therefore, many voxel and BREP phantoms covered in this article are ideally suited for this type of “individualized” retrospective applications that benefit from person-specific information in the calculations.

Looking into the future, is it necessary or feasible to bring about a change in the ICRP radiation protection concept described above? If so, how should the research community participate? The BREP phantoms have demonstrated the feasibility to develop new-generation phantoms that represent a much broader range of individuals in terms of body height and weight, as well as organ topology—features that were impossible even 10 years ago. So, should we move beyond the “Reference Man” paradigm even for radiation protection purposes? Why cannot we adopt a method that quantifies “uncertainty” in every dose calculation using a range of phantoms? There is no doubt that we are at crossroads now, perhaps as we were 20 years ago when voxel phantoms were about to emerge. The diversity of computational human phantoms developed by various groups within the community reflects the true variations amongst people and, ultimately, the true underlying uncertainty in radiation dose estimates.

This review also revealed a human factor that we do not normally see in the scientific discourse. In the history of voxel phantom development, Zubal was one of the first to share the original image data freely with other users based on a mutual agreement. Heated debate continued for some time at workshops regarding the intellectual property associated with phantoms that are revised by phantom users. It is often a technical necessity for a researcher to name a phantom created by him or her. However, it is not clear who should own such a copyright because each of the four steps of developing a phantom can be carried out by a different group. One scenario is when the original images were acquired and segmented by one individual and a different individual performed additional image processing and modification before implementing the data into a specific Monte Carlo code. Such changes produce a practically unique phantom, and proper naming is often useful for research purposes, even though individuals involved in the process do not seem to always agree upon the ownership of such a product. Some have chosen not to share phantoms partially due to this concern. Others are afraid that sharing may cost an advantage in research in a time when too many voxel phantoms exist.

The history of computational phantom development has shown that it is the need for applications, not the need for policy-making, that determines the course of technological advancement. The need for simulating organ motions for cardiac imaging, for example, resulted in the developments of MCAT phantom by Tsui et al (2004) using quadric and superquadric surface equations as well as more recently the NCAT models by Segars (2001) using the NURBS technology. Xu and Shi (2005) adopted “the geometry-based” respiration algorithm in the NCAT phantom for radiation treatment. Later Eom et al (2009) developed a “physics-based” respiration-simulating 4D phantoms for the need to understand and “predict” the effects of respiration on radiation treatment. Using the same approach, Lee et al (2007) developed the size-adjustable pediatric models. The BREP-based pregnant females by Xu et al (2007) and those by Stabin et al (2008) are also examples of application-driven research that will likely continue to dominate the research horizon in the future.

As shown in Figure 25, the BREP-based phantoms are clearly going to be the future. NURBS geometries are flexible and computationally efficient, but fine details may be lost on certain organs that have complex topology. On the other hand, polygonal models can be used to create very smooth surfaces with impressive anatomical detail by paying the price of having too many vertices. Geometrical modeling of the human body is a challenge for that it consists of organ surfaces of complex and unique shapes. For cardiac and respiratory motions at the frequency range of 10–100 cycles per second, the mesh models may still be acceptable. However, previous work has also shown that the NURBS primitives were easy to adopt for both real-time and non-real-time applications. Therefore, the specific strategy will likely be based on the applications and user preference. Regardless of the specific BREP data structure, there is currently an urgent need for application-based software that can streamline the process.

Looking forward, the following issues should be addressed by the research community in the next 5–10 years:

(1) Intercomparison of ICRP-89 compatible phantoms

More than a dozen groups worldwide have developed anatomically realistic and ICRP-89 compatible phantoms. However, organ locations and shapes differ from phantom to phantom. In particular, the dosimetric differences between these phantoms — some of which represent Asian populations instead of Caucasians —and the ICRP Reference Computational Phantoms are not well understood. What is needed is a systematic intercomparison of these phantoms and quantification of dosimetric differences related to variations in organ topology.

(2) A shift away from the Reference Man-based paradigm in radiation protection

Research on radiation protection phantoms has been influenced by the “Reference Man” paradigm. This approach required a computational phantom to match the 50th percentile population-average values in terms of body height and weight. However, anatomical variations associated with body size and organ shape can cause as much as 100% difference in the estimated organ doses. So the use of ICRP Reference Computational Phantoms carries a very large uncertainty. Future radiation protection systems may require this uncertainty to be quantified and reduced (say, to a level of 30% or less). To achieve that future goal, a shift away from the ICRP “Reference Man” methodology must take place by expanding the 50th-percentile phantoms into a much larger set of phantoms ranging from the 10th percentile to the 90th percentile, as well as people who are overweight, in each of the gender and age groups. Anthropometric data from population surveys can be used to evaluate trends in body and organ distributions. Both body weight index and trunk height can be used to match a specific individual with this expanded library of reference phantoms. Such phantoms are already used today for retrospective patient dose tracking and accidental dose reconstruction. It is anticipated that ICRP will begin to consider adopting such phantoms for prospective dosimetry in the future.

(3) Physics-based methods for deformation modeling

Motion-simulating 4D computational phantoms will play an increasingly vital role in the understanding and management of organ motion in radiotherapy and medical imaging. Work has been done to model cardiac and respiratory motions using BREP-based methods. These phantoms contain patient-specific imaging data such as 4D CT and therefore provide a realistic simulation of the motion. A limitation to these models is that they provide only one realization of the motion, specifically that observed in the patient images. They do not have the ability to realistically predict motion variations that may occur inside the same individual. To model variations in motion, one would require accurate handling of the interactions at the organ interfaces. Finite element analysis methods have the advantage of being physics-based and can be used in single- and multiple-organ deformable registration. Furthermore, internal dose models are currently based on a static 3D lung anatomy. Phantoms that account for aerodynamic flow of radioactive particulates in various parts of a deformable respiratory system will be developed in the next decade.

(4) Monte Carlo simulations with advanced geometries

Monte Carlo codes for radiation transport simulations were originally designed for nuclear engineering and high-energy physics applications. Although these codes contain excellent radiation physics algorithms, they suffer from poor software engineering design and are only able to handle simple geometries. These deficiencies have resulted in three problems that compromise current efforts in phantom research: (1) The implementation of various anatomically complex phantoms in these codes requires cumbersome manual processes. (2) Currently the NURBS-based or mesh-based phantoms must be converted to voxels before using with Monte Carlo programs (but a few groups have begun to address this issue as discussed earlier in this article). (3) Existing Monte Carlo codes are unable to handle a “moving” target such as the dynamic heart or lung. Research to convert an object defined in CAD software to those acceptable by a Monte Carlo code has been on-going in the nuclear engineering community for years. Future research is needed to develop new and computationally efficient ray-tracing algorithms that can directly process NURBS- and mesh-based geometric objects for Monte Carlo radiation transport calculations. Open-source Monte Carlo codes will be widely adopted in the next decade. With new technologies such as cloud-computing and hardware accelerated Monte Carlo methods involving the GPUs and coprocessors, one expects that dose calculation involving a whole-body voxel and BREP phantom can be carried out in seconds in the near future (for examples, see our own work by Liu et al 2014 and Su et al 2014). Such near-real-time Monte Carlo methods will likely further increase the rate of computational phantom research toward the 4th generation of “patient-specific” phantoms.

8. Conclusion

A paper which appeared in 2000 on the VIP-Man phantom predicted that the advantages afforded by both the voxels and BREP-type of surface geometries would be eventually combined (Xu et al 2000):

For the purposes of setting radiation protection standards, it may be possible to eventually bridge these two types of models, leading to a new generation of hybrid “standard” model(s) that will be acceptable to the radiation protection community. Such a new generation of models for radiation protection should be realistic enough to accurately represent major radiosensitive tissues and organs, and flexible enough to represent different populations by scaling. Computers are going to be so powerful that very complex models can be handled without a problem.”

These so-called “hybrid” computational phantoms were indeed realized in less than a decade as shown in Figure 25. In the next 10 years and beyond, advances in computational phantom research will be once again exponential. Increasingly personalized whole-body computational phantoms will be developed and applied for various clinical applications. Such phantoms will contain deformable anatomies that are physics-based and are, therefore, biomechanically realistic in depicting real-time and multi-organ deformation associated with cardiac and respiratory motions. These phantoms will also possess physiological and functional information of the human body at the organ and cellular levels obtained from multiple scanners. Breakthroughs in computational radiobiology, in the context of cancer radiotherapy, are expected to bring a new horizon to the personalized radiation medicine by understanding and harnessing the massive power of genomic data. At the same time, the power of computers will reach the exascale by the end of 2020s owing to highly efficient hardware designs (such as NVIDIA GPUs and INTEL coprocessors), thus making real-time Monte Carlo calculations for next-generation computational phantoms possible. The 50-year history reviewed in this article shows clearly that coordinated and cooperative efforts among radiological engineers, computer scientists, biologists, and clinicians will always be the key to the success of future research endeavors.

Supplementary Material

pmb303293suppdata.pdf

Acknowledgments

This article benefitted from discussions in the past decade with many colleagues, especially Keith Eckerman, John Poston, Dan Reece, Wes Bolch, Mike Stabin, Randy Brill, Paul Segars, Ben Tsui, Harald Paganetti, Mike McNitt-Gray, George Zubal, Maria Zankl, Nina Petoussi-Henss, Richard Kramer, Qian Liu, Junli Li, Chan Kim, and Habib Zaidi. This article adopted several chapters of the Handbook of Anatomical Models For Radiation Dosimetry for which I was a co-editor. RPI research activities highlighted in the article were based on the following Ph.D. or M.S. projects (the year of graduation indicated): Ahmet Bozkurt (2000), TC Ephraim Chao (2001), Chengyu Shi (2004), Mark Winslow, (2005), Brian Wang (2005), Peter Caracappa (2006), Bryan Bednarz (2008), Juying Zhang (2009), Yong Hum Na (2009), Jianwei Gu (2010), Bin Han (2011), Aiping Ding (2012), Matthew Mille (2013), and Justin Vazquez (2013). Grants Funding for these projects include: National Science Foundation (BES-9875532), National Library of Medicine (R03LM007964, R01LM009362, and R01LM009362-03S1), National Cancer Institute (R01CA116743 and R42CA115122), National Institute of Biomedical Imaging and Bioengineering (R42EB010404), and National Institute of Standards and Technology (70NANB9H9198). Several RPI students and postdoc researchers helped me with the preparation and submission of this article: Xining Du, Michael Pinkert, Ashley Rhodes, Tianyu Liu, Haikuan Liu, and Matthew Mille.

Footnotes

1

This paragraph about the distinction between ICRP “population-averaged” prospective dosimetry and “individualized” retrospective dosimetry is based on extensive discussion with Prof Wesley Bolch.

References

  1. Akkurt H, Bekar KB, Eckerman KF. VOXMAT: Phantom model with combination of voxel and mathematical geometry. Health Phys. 2008;95(1):S100. [Google Scholar]
  2. Alderson SW, Lanzl LH, Rollins M, Spira J. An instrumented phantom system for analog computation of treatment plans. Am. J. Roentgenol. Radium. Ther. Nucl. Med. 1962;87:185–195. [PubMed] [Google Scholar]
  3. Allison J, et al. Geant4 developments and applications. IEEE Trans. Nucl. Sci. 2006;53:270–278. [Google Scholar]
  4. Alziar I, et al. Individual radiation therapy patient whole-body phantoms for peripheral dose evaluations: method and specific software. Phys. Med. Biol. 2009;54:N375–N383. doi: 10.1088/0031-9155/54/17/N01. [DOI] [PubMed] [Google Scholar]
  5. Andreo P. Monte Carlo techniques in medical radiation physics. Phys. Med. Biol. 1991;36:861–920. doi: 10.1088/0031-9155/36/7/001. [DOI] [PubMed] [Google Scholar]
  6. Attix FH. Introduction to Radiological Physics and Radiation Dosimetry. New York, NY: John Wiley & Sons, Inc.; 1986. [Google Scholar]
  7. Bahadori AA, Johnson PB, Jokisch DW, Eckerman KF, Bolch WE. Response functions for computing absorbed dose to skeletal tissues from neutron irradiation. Phys. Med. Biol. 2011;56:6873–6897. doi: 10.1088/0031-9155/56/21/008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Bahadori AA, Van Baalen M, Shavers MR, Dodge C, Semones EJ, Bolch WE. The effect of anatomical modeling on space radiation dose estimates: A comparison of doses for NASA phantoms and the 5th, 50th, and 95th percentile male and female astronauts. Phys. Med. Biol. 2011;56:1671–1694. doi: 10.1088/0031-9155/56/6/010. [DOI] [PubMed] [Google Scholar]
  9. Battistoni G, Cerutti F, Fassò A, Ferrari A, Muraroi S, Ranft J, Roesler S, Sala PR. The FLUKA code: Description and benchmarking. AIP Conf. Proc. 2007;896:31–49. [Google Scholar]
  10. Beck P, et al. MATSIM: Development of a voxel model of the MATROSHKA astronaut dosimetric phantom. IEEE Trans. Nuc. Sci. 2011;58:1921–1926. [Google Scholar]
  11. Becker J, Zankl M, Petoussi-Henss N. A software tool for modification of human voxel models used for application in radiation protection. Phys. Med. Biol. 2007;52:N195–N205. doi: 10.1088/0031-9155/52/9/N01. [DOI] [PubMed] [Google Scholar]
  12. Bento J, Barros S, Teles P, Neves M, Gonçalves I, Corisco J, Vaz P. Monte Carlo simulation of the movement and detection efficiency of a whole-body counting system using a BOMAB phantom. Radiat. Prot. Dosimetry. 2012;148:403–413. doi: 10.1093/rpd/ncr201. [DOI] [PubMed] [Google Scholar]
  13. Bhati S, Patni HK, Ghare VP, Singh IS, Nadar MY. Monte Carlo calculations for efficiency calibration of a whole-body monitor using BOMAB phantoms of different sizes. Radiat. Prot. Dosimetry. 2012;148:414–419. doi: 10.1093/rpd/ncr203. [DOI] [PubMed] [Google Scholar]
  14. Billings MP, Yucker WR. The Computerized Anatomical Man (CAM) model. Houston, TX: National Aeronautics and Space Administration; 1973. NASA CR-134043. [Google Scholar]
  15. Bolch W, Lee C, Wayson M, Johnson P. Hybrid computational phantoms for medical dose reconstruction. Radiat Environ Biophys. 2010;49(2):155–168. doi: 10.1007/s00411-009-0260-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Bouchet LG, Bolch WE, Weber DA, Atkins HL, Poston JW., Sr A revised dosimetric model of the adult head and brain. J. Nucl. Med. 1996;37:1226–1236. [PubMed] [Google Scholar]
  17. Bouchet LG, Bolch WE, Weber DA, Atkins HL, Poston JW., Sr MIRD Pamphlet No. 15: Radionuclide S values in a revised dosimetric model of the adult head and brain. J. Nucl. Med. 1999;40:62S–101S. [PubMed] [Google Scholar]
  18. Bouchet LG, Bolch WE, Blanco HP, Rajon DA, Clairand I, Sgouros G, Wessels BW. MIRD Pamphlet No. 19: Absorbed fractions and radionuclide S values for six age-dependent multi-region models of the kidney. J. Nucl Med. 2003;44(7):1113–1147. [PubMed] [Google Scholar]
  19. Bozkurt A, Chao TC, Xu XG. Fluence-to-dose conversion coefficients from monoenergetic neutron beams below 20 MeV based on the VIP-Man anatomical model. Phys. Med. Biol. 2000;45:3059–3079. doi: 10.1088/0031-9155/45/10/321. [DOI] [PubMed] [Google Scholar]
  20. Bozkurt A, Chao TC, Xu XG. Fluence-to-dose conversion coefficients based on the VIP-Man anatomical model and MCNPX code for monoenergetic neutrons above 20 MeV. Health Phys. 2001;81:184–202. doi: 10.1097/00004032-200108000-00011. [DOI] [PubMed] [Google Scholar]
  21. Bozkurt A, Xu XG. Fluence-to-dose conversion coefficients for monoenergetic proton beams based on the VIP-Man anatomical model. Radiat. Prot. Dosimetry. 2004;112:219–235. doi: 10.1093/rpd/nch393. [DOI] [PubMed] [Google Scholar]
  22. Bozkurt A, Bor D. Simultaneous determination of equivalent dose to organs and tissues of the patient and of the physician in interventional radiology using the Monte Carlo method. Phys. Med. Biol. 2007;52:317–330. doi: 10.1088/0031-9155/52/2/001. [DOI] [PubMed] [Google Scholar]
  23. Broggio D, Beurrier J, Bremaud M, Desbrée A, Farah J, Huet C, Franck D. Construction of an extended library of adult male 3D models: rationale and results. Phys. Med. Biol. 2011;56:7659–7692. doi: 10.1088/0031-9155/56/23/020. [DOI] [PubMed] [Google Scholar]
  24. Brown FB, editor. Report LA-UR-03-1987. Los Alamos, NM: Los Alamos National Laboratory; 2003. MCNP—A general Monte Carlo N-Particle transport code, version 5. [Google Scholar]
  25. Caon M, Bibbo G, Pattison J. Monte Carlo calculated effective dose to teenage girls from computed tomography examinations. Radiat. Prot. Dosimetry. 2000;90:445–448. [Google Scholar]
  26. Caon M. Voxel-based computational models of real human anatomy: A review. Radiat. Environ. Biophys. 2004;42:229–235. doi: 10.1007/s00411-003-0221-8. [DOI] [PubMed] [Google Scholar]
  27. Caon M, Bibbo G, Pattison J. An EGS4-ready tomographic computational model of a 14-year-old female torso for calculating organ doses from CT examinations. Phys. Med. Biol. 1999;44:2213–2225. doi: 10.1088/0031-9155/44/9/309. [DOI] [PubMed] [Google Scholar]
  28. Caracappa PF. Ph.D. Dissertation. Troy, NY: Rensselaer Polytechnic Institute; 2006. Development and evaluation of a new algorithm for determining radiation dose to the red bone marrow. [Google Scholar]
  29. Caracappa PF, Chao TCE, Xu XG. A study of predicted bone marrow distribution on calculated marrow dose from external radiation exposures using two sets of image data for the same individual. Health Phys. 2009;96:661–674. doi: 10.1097/01.HP.0000346304.45813.36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Cassola VF, Lima VJM, Kramer R, Khoury HJ. FASH and MASH: Female and male adult human phantoms based on polygon mesh surfaces: I. Development of the anatomy. Phys. Med. Biol. 2010;55:133–162. doi: 10.1088/0031-9155/55/1/009. [DOI] [PubMed] [Google Scholar]
  31. Cassola VF, Milian FM, Kramer R, de Oliveira Lira CAB, Khoury HJ. Standing adult human phantoms based on 10th, 50th and 90th mass and height percentiles of male and female Caucasian populations. Phys. Med. Biol. 2011;56:3749–3772. doi: 10.1088/0031-9155/56/13/002. [DOI] [PubMed] [Google Scholar]
  32. CCHP. Tomographic models for radiation protection dosimetry session in The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World; Proceedings of the Monte Carlo 2005 Topical Meeting held in Chattanooga; April 17–21; TN. LaGrange, Park IL: American Nuclear Society; 2005. on CD-ROM (ISBN:0-89448-695-0) [Google Scholar]
  33. Cech R, Leitgeb N, Pediaditis M. Current densities in a pregnant woman model induced by simultaneous ELF electric and magnetic field exposure. Phys. Med. Biol. 2008;53:177–186. doi: 10.1088/0031-9155/53/1/012. [DOI] [PubMed] [Google Scholar]
  34. Cech R, Leitgeb N, Pediaditis M. Fetal exposure to low frequency electric and magnetic fields. Phys. Med. Biol. 2007;52:879–888. doi: 10.1088/0031-9155/52/4/001. [DOI] [PubMed] [Google Scholar]
  35. Chao TC, Bozkurt A, Xu XG. Conversion coefficients based on the VIP-Man anatomical model and EGS4-VLSI code for external monoenergetic photons from 10 keV to 10 MeV. Health Phys. 2001;81:163–183. doi: 10.1097/00004032-200108000-00010. [DOI] [PubMed] [Google Scholar]
  36. Chao TC, Bozkurt A, Xu XG. Organ dose conversion coefficients for 0.1–10 MeV electrons calculated for the VIP-Man tomographic model. Health Phys. 2001;81:203–214. doi: 10.1097/00004032-200108000-00012. [DOI] [PubMed] [Google Scholar]
  37. Chao TC, Xu XG. Specific absorbed fractions from the image-based VIP-Man body model and EGS4-VLSI Monte Carlo code: Internal electron emitters. Phys. Med. Biol. 2001;46:901–927. doi: 10.1088/0031-9155/46/4/301. [DOI] [PubMed] [Google Scholar]
  38. Chao TC, Bozkurt A, Xu XG. (correction) Conversion coefficients based on the VIP-Man anatomical model and EGS4-VLSI code for external monoenergetic photons from 10 keV to 10 MeV. Health Phys. 2003;84(3):390. [Google Scholar]
  39. Chao TC, Xu XG. S-values calculated from a tomographic head/brain model for brain imaging. Phys. Med. Biol. 2004;49:4971–4984. doi: 10.1088/0031-9155/49/21/009. [DOI] [PubMed] [Google Scholar]
  40. Chen J. Mathematical models of the embryo and fetus for use in radiological protection. Health Phys. 2004;86:285–295. doi: 10.1097/00004032-200403000-00005. [DOI] [PubMed] [Google Scholar]
  41. Choi SH, Lee C-S, Cho SK, Chung MS, Na SH, Kim CH. Construction of a high-definition ‘Reference Korean’ voxel phantom for organ and tissue radiation dose calculation. IFMBE Proc. 2006;14:4204–4207. [Google Scholar]
  42. Christ A, et al. The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulations. Phys. Med. Biol. 2010;55:N23–N38. doi: 10.1088/0031-9155/55/2/N01. [DOI] [PubMed] [Google Scholar]
  43. CIRS. Tissue Simulation and Phantom Technology. [last access 02/27/2014];2013 http://www.cirsinc.com.
  44. CMPWG. Phantoms. [last access 02/27/2014];2013 http://cmpwg.ans.org/phantoms/camera.pdf.
  45. Courageot E, Huet C, Clairand I, Bottollier-Depois JF, Gourmelon P. Numerical dosimetric reconstruction of a radiological accident in South America in April 2009. Radiat. Prot. Dosimetry. 2011;144:540–542. doi: 10.1093/rpd/ncq338. [DOI] [PubMed] [Google Scholar]
  46. Courageot E, Sayah R, Huet C. Development of modified voxel phantoms for the numerical dosimetric reconstruction of radiological accidents involving external sources: Implementation in SESAME tool. Phys. Med. Biol. 2010;55:N231–N241. doi: 10.1088/0031-9155/55/9/N02. [DOI] [PubMed] [Google Scholar]
  47. Cristy M. Mathematical phantoms representing children of various ages for use in estimates of internal dose. Oak Ridge, TN: Oak Ridge National Laboratory; 1980. [last access 02/27/2014]. U.S. Nuclear Regulatory Commission Report NUREG/CR-1159 http://web.ornl.gov/info/reports/1980/3445605812234.pdf. [Google Scholar]
  48. Cristy M, Eckerman KF. Specific absorbed fractions of energy at various ages from internal photon sources I: Methods. Oak Ridge, TN: Oak Ridge National Laboratory; 1987. [last access 02/27/2014]. Oak Ridge National Laboratory Report ORNL/TM-8381/V1 http://ordose.ornl.gov/documents/tm8381V1.pdf. [Google Scholar]
  49. Dawson TW, Caputa K, Stuchly MA. A comparison of 60 Hz uniform magnetic and electric induction in the human body. Phys. Med. Biol. 1997;42:2319–2329. doi: 10.1088/0031-9155/42/12/001. [DOI] [PubMed] [Google Scholar]
  50. Deus SF, Poston JW. Proceedings of the Symposium on Radiopharmaceutical Dosimetry. Rockville, MD: U.S. Food and Drug Administration; 1976. The development of a mathematical phantom representing a 10-year-old for use in internal dose calculations. HEW Publication 76-8044. [Google Scholar]
  51. DeWerd LA, Kissick M. In: The Phantoms of Medical and Health Physics, editor. Berlin, Germany: Springer; 2014. [Google Scholar]
  52. Dimbylow P. Development of pregnant female, hybrid voxel-mathematical models and their application to the dosimetry of applied magnetic and electric fields at 50 Hz. Phys. Med. Biol. 2006;51:2383–2394. doi: 10.1088/0031-9155/51/10/003. [DOI] [PubMed] [Google Scholar]
  53. Dimbylow P. Development of the female voxel phantom, NAOMI, and its application to calculations of induced current densities and electric fields from applied low frequency magnetic and electric fields. Phys. Med. Biol. 2005a;50:1047–1070. doi: 10.1088/0031-9155/50/6/002. [DOI] [PubMed] [Google Scholar]
  54. Dimbylow P. Resonance behaviour of whole-body averaged specific energy absorption rate (SAR) in the female voxel model, NAOMI. Phys. Med. Biol. 2005b;50:4053–4063. doi: 10.1088/0031-9155/50/17/009. [DOI] [PubMed] [Google Scholar]
  55. Dimbylow PJ. Proc. Workshop on Voxel Phantom Development. Chilton, UK: 1996. The development of realistic voxel phantoms for electromagnetic field dosimetry. [Google Scholar]
  56. Dimbylow PJ. FDTD calculations of the whole-body averaged SAR in an anatomically realistic voxel model of the human body from 1 MHz to 1 GHz. Phys. Med. Biol. 1997;42:479–490. doi: 10.1088/0031-9155/42/3/003. [DOI] [PubMed] [Google Scholar]
  57. Dimbylow P, Bolch W, Lee C. SAR calculations from 20 MHz to 6 GHz in the University of Florida newborn voxel phantom and their implications for dosimetry. Phys. Med. Biol. 2010;55:1519–1530. doi: 10.1088/0031-9155/55/5/017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  58. Ding A, Gu J, Trofimov AV, Xu XG. Monte Carlo calculation of imaging doses from diagnostic multidetector CT and kilovoltage cone-beam CT as part of prostate treatment plans. Med. Phys. 2010;37:6199–6204. doi: 10.1118/1.3512791. [DOI] [PMC free article] [PubMed] [Google Scholar]
  59. Ding A, Mille MM, Liu T, Caracappa PF, Xu XG. Extension of RPI-adult male and female computational phantoms to obese patients and a Monte Carlo study on the effects on CT imaging dose. Phys. Med. Biol. 2012;57:2441–2459. doi: 10.1088/0031-9155/57/9/2441. [DOI] [PMC free article] [PubMed] [Google Scholar]
  60. Doerfel H, Heide B. Calibration of a phoswich type partial body counter by Monte Carlo simulation of low-energy photon transport. Radiat. Prot. Dosimetry. 2007;123:464–472. doi: 10.1093/rpd/ncl503. [DOI] [PubMed] [Google Scholar]
  61. Eckerman KF. Aspects of the dosimetry of radionuclides within the skeleton with particular emphasis on the active marrow Proc. In: Schlafke-Stelson AT, Watson EE, editors. 4th Int. Radiopharmaceutical Dosimetry Symp. Oak Ridge, TN: Oak Ridge Associated Universities; 1985. pp. 514–534. [Google Scholar]
  62. Eckerman KF, Stabin MG. Electron absorbed fractions and dose conversion factors for marrow and bone by skeletal regions. Health Phys. 2000;78:199–214. doi: 10.1097/00004032-200002000-00009. [DOI] [PubMed] [Google Scholar]
  63. Eckerman KF, Poston JW, Sr, Bolch WE, Xu XG. Handbook of Anatomical Models for Radiation Dosimetry. New York, NY: Taylor & Francis; 2009. The stylized computational phantoms developed at ORNL and elsewhere; pp. 43–64. [Google Scholar]
  64. Eom J, Xu XG, De S, Shi C. Predictive modeling of lung motion over the entire respiratory cycle using measured pressure-volume data, 4DCT images, and finite element analysis. Med. Phys. 2010;37:4389–4400. doi: 10.1118/1.3455276. [DOI] [PMC free article] [PubMed] [Google Scholar]
  65. Farah J, Broggio D, Franck D. Creation and use of adjustable 3D phantoms: Application for the lung monitoring of female workers. Health Phys. 2010a;99:649–661. doi: 10.1097/HP.0b013e3181dc4f58. [DOI] [PubMed] [Google Scholar]
  66. Farah J, Broggio D, Franck D. Female workers and in vivo lung monitoring: A simple model for morphological dependence of counting efficiency curves. Phys. Med. Biol. 2010b;55:7377–7395. doi: 10.1088/0031-9155/55/23/013. [DOI] [PubMed] [Google Scholar]
  67. Farah J, Broggio D, Franck D. Examples of Mesh and NURBS modelling for in vivo lung counting studies. Radiat. Prot. Dosimetry. 2011;144:344–348. doi: 10.1093/rpd/ncq313. [DOI] [PubMed] [Google Scholar]
  68. Ferrari A, Pelliccioni M, Pillon M. Fluence to effective dose and effective dose equivalent conversion coefficients for electrons from 5 MeV to 10 GeV. Radiat. Prot. Dosimetry. 1997;69:97–104. [Google Scholar]
  69. Ferrari P. Development of an integrated couple of anthropomorphic models for dosimetric studies. Radiat. Prot. Dosimetry. 2010;142:191–200. doi: 10.1093/rpd/ncq194. [DOI] [PubMed] [Google Scholar]
  70. Ferrari P, Gualdrini G. An improved MCNP version of the NORMAN voxel phantom for dosimetry studies. Phys. Med. Biol. 2005;50:4299–4316. doi: 10.1088/0031-9155/50/18/005. [DOI] [PubMed] [Google Scholar]
  71. Fill UA, Zankl M, Petoussi-Henss N, Siebert M, Regulla D. Adult female voxel models of different stature and photon conversion coefficients for radiation protection. Health Phys. 2004;86:253–272. doi: 10.1097/00004032-200403000-00003. [DOI] [PubMed] [Google Scholar]
  72. Findlay RP, Dimbylow PJ. SAR in a child voxel phantom from exposure to wireless computer networks (Wi-Fi) Phys. Med. Biol. 2010;55:N405–N411. doi: 10.1088/0031-9155/55/15/N01. [DOI] [PubMed] [Google Scholar]
  73. Findlay RP, Dimbylow PJ. Spatial averaging of fields from half-wave dipole antennas and corresponding SAR calculations in the NORMAN human voxel model between 65 MHz and 2 GHz. Phys. Med. Biol. 2009;54:2437–2447. doi: 10.1088/0031-9155/54/8/012. [DOI] [PubMed] [Google Scholar]
  74. Fisher HLJ, Snyder WS. Health Physics Division Annual Progress Report for Period Ending July 31, 1966. Oak Ridge, TN: Oak Ridge National Laboratory; 1966. Variation of dose delivered by 137Cs as a function of body size from infancy to adulthood; pp. 221–228. Report ORNL-4007. [Google Scholar]
  75. Fisher HLJ, Snyder WS. Health Physics Division Annual Progress report for Period Ending July 31, 1967. Oak Ridge, TN: Oak Ridge National Laboratory; 1967. Distribution of dose in the body from a source of gamma rays distributed uniformly in an organ. Report ORNL-4168. [Google Scholar]
  76. Fung GS, Segars WP, Gullberg GT, Tsui BM. Development of a model of the coronary arterial tree for the 4D XCAT phantom. Phys. Med. Biol. 2011;56:5651–5663. doi: 10.1088/0031-9155/56/17/012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  77. Furler M. M.S. Thesis. Troy, NY: Rensselaer Polytechnic Institute; 2007. Methods Of Converting Geometry In CAD To MCNP Code. [Google Scholar]
  78. Gardumi A, Farah J, Desbrée A. Creation of ORNL NURBS-based phantoms: Evaluation of the voxel effect on absorbed doses from radiopharmaceuticals. Radiat. Prot. Dosimetry. 2013;153:273–281. doi: 10.1093/rpd/ncs103. [DOI] [PubMed] [Google Scholar]
  79. Geyer AM, O’Reilly S, Lee C, Long DJ, Bolch WE. The UF/NCI family of hybrid computational phantoms representing the current US population of male and female children and adolescents—applications to CT dosimetry. Phys. Med. Biol. 2014 doi: 10.1088/0031-9155/59/18/5225. (in press) [DOI] [PMC free article] [PubMed] [Google Scholar]
  80. Gibbs SJ, Pujol A, Chen TS, Malcolm AW, James AE. Patient risk from interproximal radiography. Oral Surg. Oral Med. O. 1984;58:347–354. doi: 10.1016/0030-4220(84)90066-5. [DOI] [PubMed] [Google Scholar]
  81. Gibbs SJ, Pujol A, Jr, Chen TS, Carlton JC, Dosmann MA, Malcom AW, James AE., Jr Radiation doses to sensitive organs from intraoral dental radiography. Dentomaxillofac Radiol. 1987;16:67–77. doi: 10.1259/dmfr.1987.0010. [DOI] [PubMed] [Google Scholar]
  82. Gjonaj E, Bartsch M, Clemens M, Schupp S, Weiland T. High-resolution human anatomy models for advanced electromagnetic field computations. IEEE Trans. Magn. 2002;38:357–360. [Google Scholar]
  83. Goorley JT, et al. Initial MCNP6 Release Overview - MCNP6 version 1.0. Los Alamos, NM: Los Alamos National Laboratory; 2013. Report LA-UR-13-22934. [Google Scholar]
  84. Gosselin M-C, et al. Development of a new generation of high-resolution anatomical models for medical device evaluation: The Virtual Population 3.0. Phys. Med. Biol. 2014 doi: 10.1088/0031-9155/59/18/5287. (in press) [DOI] [PubMed] [Google Scholar]
  85. Griffith RV, Dean PN, Anderson AL, Fisher JC. Fabrication of a tissue-equivalent torso phantom for intercalibration of in-vivo transuranic-nuclide counting facilities. Proceedings of International Symposium on Advances in Radiation Protection Monitoring; International Atomic Energy Agency; Stockholm, Sweden. 1978. also as Lawrence Livermore National Laboratory Report UCRL-80343. [Google Scholar]
  86. Gu J, Dorgu A, Xu XG. Comparison of main software packages for CT dose reporting. Health Phys. 2008a;95(1):S50. [Google Scholar]
  87. Gu J, Bednarz B, Xu XG, Jiang SB. Assessment of patient organ doses and effective doses using the VIP-Man adult male phantom for selected cone-beam CT imaging procedures during image guided radiation therapy. Radiat. Prot. Dosimetry. 2008b;131:431–443. doi: 10.1093/rpd/ncn200. [DOI] [PubMed] [Google Scholar]
  88. Gu J, Bednarz B, Caracappa PF, Xu XG. The development, validation and application of a multi-detector CT (MDCT) scanner model for assessing organ doses to the pregnant patient and the fetus using Monte Carlo simulations. Phys. Med. Biol. 2009;54:2699–2717. doi: 10.1088/0031-9155/54/9/007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  89. Gu J. Ph.D. Dissertation. Troy, NY: Rensselaer Polytechnic Institute; 2010. Development of CT scanner models for patient organ dose calculations using Monte Carlo methods. [Google Scholar]
  90. Gu S, Gupta R, Kyprianou I. Computational high-resolution heart phantoms for medical imaging and dosimetry simulations. Phys. Med. Biol. 2011;56:5845–5864. doi: 10.1088/0031-9155/56/18/005. [DOI] [PubMed] [Google Scholar]
  91. Hammersley JM, Handscomb DC. Monte Carlo Methods. London, U.K: Metheun & Co. Ltd.; 1964. [Google Scholar]
  92. Han B, Zhang J, Na YH, Caracappa PF, Xu XG. Modeling and Monte Carlo organ dose calculations for workers walking on ground contaminated with Cs-137 and Co-60 gamma sources. Radiation Prot. Dosimetry. 2010;141:299–304. doi: 10.1093/rpd/ncq184. [DOI] [PMC free article] [PubMed] [Google Scholar]
  93. Han B, Xu XG, Chen GTY. Proton radiography and fluoroscopy of lung tumors: A Monte Carlo study using patient-specific 4DCT phantoms. Med. Phys. 2011;38:1903–1911. doi: 10.1118/1.3555039. [DOI] [PMC free article] [PubMed] [Google Scholar]
  94. Han MC, Kim CH, Jeong JH, Yeom YS, Kim S, Wilson PH, Apostolakis J. DagSolid: a new Geant4 solid class for fast simulation in polygon-mesh geometry. Phys. Med. Biol. 2013;58(13):4595–4609. doi: 10.1088/0031-9155/58/13/4595. [DOI] [PubMed] [Google Scholar]
  95. Hegenbart L, Na YH, Zhang JY, Urban M, Xu XG. A Monte Carlo study of lung counting efficiency for female workers of different breast sizes using deformable phantoms. Phys. Med. Biol. 2008;53:5527–5538. doi: 10.1088/0031-9155/53/19/017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  96. Hintenlang D, Moloney W, Winslow J. Handbook of Anatomical Models for Radiation Dosimetry. New York, NY: Taylor & Francis; 2009. Physical Phantoms for Experimental Radiation Dosimetry. [Google Scholar]
  97. Hirata A, Ito N, Fujiwara O, Nagaoka T, Watanabe S. Conservative estimation of whole-body-averaged SARs in infants with a homogeneous and simple-shaped phantom in the GHz region. Phys. Med. Biol. 2008;53:7215–7223. doi: 10.1088/0031-9155/53/24/014. [DOI] [PubMed] [Google Scholar]
  98. Hough M, Johnson P, Rajon D, Jokisch D, Lee C, Bolch W. An image-based skeletal dosimetry model for the ICRP reference adult male—internal electron sources. Phys. Med. Biol. 2011;56:2309–2346. doi: 10.1088/0031-9155/56/8/001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  99. Hubbell JH. Photon Cross Sections, Attenuation Coefficients and Energy Absorption Coefficients from 10 keV to 100 GeV. Washington, DC: National Bureau of Standards; 1969. Report NSRDS-NBS 29. [Google Scholar]
  100. Hwang JML, Shoup RL, Poston JW. Mathematical description of a newborn human for use in dosimetry calculations. Oak Ridge, TN: Oak Ridge National Laboratory; 1976. Report ORNL/TM-5453. [Google Scholar]
  101. ICRP. Report of Committee II on Permissible Dose for Internal Radiation. Oxford, UK: Pergamon Press Ltd.; 1959. [Google Scholar]
  102. ICRP. Report of the Task Group on Reference Man. Oxford, UK: Pergamon Press Ltd.; 1975. ICRP Publication 23. [Google Scholar]
  103. ICRP. Basic anatomical and physiological data for use in radiological protection reference values. Oxford, UK: Pergamon Press Ltd.; 2002a. ICRP Publication 89. [Google Scholar]
  104. ICRP. [last access 03/02/14];2002 Annual Report of the International Commission on Radiological Protection. 2002b http://www.icrp.org/docs/2002_ann_rep_52_429_03.pdf.
  105. ICRP. Recommendations of the International Commission on Radiological Protection. Oxford, UK: Pergamon Press Ltd.; 2007. ICRP Publication 103. [Google Scholar]
  106. ICRP. Adult reference computational phantoms. Oxford, UK: Pergamon Press Ltd.; 2009. ICRP Publication 110. [Google Scholar]
  107. ICRU. Phantoms and computational models in therapy, diagnosis and protection. Bethesda, MD: International Commission on Radiation Units and Measurements; 1992. ICRU Report 48. [Google Scholar]
  108. Jiang H, Wang B, Xu XG, Suit HD, Paganetti H. Simulation of organ-specific patient effective dose due to secondary neutrons in proton radiation treatment. Phys. Med. Biol. 2005;50:4337–4353. doi: 10.1088/0031-9155/50/18/007. [DOI] [PubMed] [Google Scholar]
  109. Jin W, Lim Y-J, Xu XG, Singh TP, De S. Improving the visual realism of virtual surgery. In: Westood JD, et al., editors. Proceedings of Medicine Meets Virtual Reality 13: The Magical Next Becomes the Medical Now. Amsterdam, Netherlands: IOS Press Inc.; 2005. pp. 227–233. [Google Scholar]
  110. Johnson P, Lee C, Johnson K, Siragusa D, Bolch WE. The influence of patient size on dose conversion coefficients: A hybrid phantom study for adult cardiac catheterization. Phys. Med. Biol. 2009;54:3613–3629. doi: 10.1088/0031-9155/54/12/001. [DOI] [PubMed] [Google Scholar]
  111. Johnson PB, Bahadori AA, Eckerman KF, Lee C, Bolch WE. Response functions for computing absorbed dose to skeletal tissues from photon irradiation – an update. Phys Med Biol. 2011;56:2347–2366. doi: 10.1088/0031-9155/56/8/002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  112. Jones AK, Simon TA, Bolch WE, Holman MM, Hintenlang DE. Tomographic physical phantom of the newborn child with real-time dosimetry I. Methods and techniques for construction. Med. Phys. 2006;33:3274–3282. doi: 10.1118/1.2256686. [DOI] [PubMed] [Google Scholar]
  113. Jones DG. A realistic anthropomorphic phantom for calculating organ doses arising from external photon irradiation. Radiat. Prot. Dosimetry. 1997;72:21–29. [Google Scholar]
  114. Jones RM, Poston JW, Hwang JL, Jones TD, Warner GG. The development and use of a fifteen-year-old equivalent mathematical phantom for internal dose calculations. Oak Ridge, TN: Oak Ridge National Laboratory; 1976. Report ORNL/TM-5278. [Google Scholar]
  115. Katagiri M, Hikoji M, Kitaichi M, Sawamura S, Aoki Y. Effective dose and organ doses per unit fluence calculated for monoenergetic 0.1 MeV to 100 MeV electrons by the MIRD-5 phantom. Radiat. Prot. Dosimetry. 2000;90(4):393–401. doi: 10.1093/oxfordjournals.rpd.a006535. [DOI] [PubMed] [Google Scholar]
  116. Kereiakes JG, Seltzer RA, Blackburn B, Saenger EL. Radionuclide doses to infants and children: A plea for a standard child. Health Phys. 1965;11:999–1004. doi: 10.1097/00004032-196510000-00002. [DOI] [PubMed] [Google Scholar]
  117. Kim CH, Choi SH, Jeong JH, Lee C, Chung MS. HDRK-Man: A whole-body voxel model based on high-resolution color slice images of a Korean adult male cadaver. Phys. Med. Biol. 2008;53:4093–4106. doi: 10.1088/0031-9155/53/15/006. [DOI] [PubMed] [Google Scholar]
  118. Kim CH, Jeong JH, Bolch WE, Cho K-W, Hwang SB. A polygon-surface reference Korean male phantom (PSRK-Man) and its direct implementation in Geant4 Monte Carlo simulation. Phys. Med. Biol. 2011;56:3137–3161. doi: 10.1088/0031-9155/56/10/016. [DOI] [PubMed] [Google Scholar]
  119. Kim JH, Kim CS, Whang JH. Assessment of radiation dose for surrounding organs and persons approaching implanted patients upon brachytherapy of prostate cancer with Iridium-192. Radiat. Prot. Dosimetry. 2010;141:283–288. doi: 10.1093/rpd/ncq181. [DOI] [PubMed] [Google Scholar]
  120. Kim JI, Choi H, Lee BI, Lim YK, Kim CS, Lee JK, Lee C. Physical phantom of typical Korean male for radiation protection purpose. Radiat. Prot. Dosimetry. 2006;118:131–136. doi: 10.1093/rpd/nci338. [DOI] [PubMed] [Google Scholar]
  121. Kim JS, Ha WH, Jeong JH, Cho K-W, Lee JK. Use of photographic images to construct voxel phantoms for use in whole-body counting. Radiat. Prot. Dosimetry. 2010;138:119–122. doi: 10.1093/rpd/ncp215. [DOI] [PubMed] [Google Scholar]
  122. Kramer GH, Burns L, Noel L. The BRMD BOMAB phantom family. Health Phys. 1991;61:895–902. doi: 10.1097/00004032-199112000-00025. [DOI] [PubMed] [Google Scholar]
  123. Kramer R, Zankl M, Williams G, Drexler G. The male (ADAM) and female (EVA) adult mathematical phantoms. Neuherberg, Germany: Institut fuer Strahlenschutz, GSF-Forschungszentrum fuer Umwelt und Gesundheit; 1982. The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods: Part I. GSF-Report S-885. [Google Scholar]
  124. Kramer R, Cassola VF, Khoury HJ, Vieira JW, Lima VJM, Brown KR. FASH and MASH: Female and male adult human phantoms based on polygon mesh surfaces: II. Dosimetric calculations. 2010;55:163–189. doi: 10.1088/0031-9155/55/1/010. [DOI] [PubMed] [Google Scholar]
  125. Kramer R, Khoury HJ, Vieira JW, Lima VJM. MAX06 and FAX06: Update of two adult human phantoms for radiation protection dosimetry. Phys. Med. Biol. 2006;51:3331–3346. doi: 10.1088/0031-9155/51/14/003. [DOI] [PubMed] [Google Scholar]
  126. Kramer R, Khoury HJ, Vieira JW, Loureiro ECM, Lima VJM, Lima FRA, Hoff G. All about FAX: A Female Adult voXel phantom for Monte Carlo calculation in radiation protection dosimetry. Phys. Med. Biol. 2004;49:5203–5216. doi: 10.1088/0031-9155/49/23/001. [DOI] [PubMed] [Google Scholar]
  127. Kramer R, Vieira JW, Khoury HJ, Lima FRA, Fuelle D. All about MAX: A male adult voxel phantom for Monte Carlo calculations in radiation protection dosimetry. Phys. Med. Biol. 2003;48:1239–1262. doi: 10.1088/0031-9155/48/10/301. [DOI] [PubMed] [Google Scholar]
  128. Kyoto Kagaku co., LTD. Patient Simulators, Imaging Phantoms for Skill Training. [last access 02/28/14];2013 http://www.kyotokagaku.com/
  129. Lee C, Nagaoka T, Lee J-K. Implementation of Japanese male and female tomographic phantoms to multi-particle Monte Carlo mode for ionizing radiation dosimetry. J. Nucl. Sci. Technol. 2006;43:937–945. [Google Scholar]
  130. Lee B, et al. Dose evaluation of selective collimation effect in cephalography by measurement and Monte Carlo simulation. Radiat. Prot. Dosimetry. 2011;148:58–64. doi: 10.1093/rpd/ncq596. [DOI] [PubMed] [Google Scholar]
  131. Lee C, Lee C, Williams JL, Bolch WE. Whole-body voxel phantoms of paediatric patients—UF Series B. Phys. Med. Biol. 2006;51:4649–4661. doi: 10.1088/0031-9155/51/18/013. [DOI] [PubMed] [Google Scholar]
  132. Lee C, Williams JL, Lee C, Bolch WE. The UF series of tomographic computational phantoms of pediatric patients. Med. Phys. 2005;32:3537–3548. doi: 10.1118/1.2107067. [DOI] [PubMed] [Google Scholar]
  133. Lee C, Lee J, Lee C. Korean adult male voxel model KORMAN segmented from magnetic resonance images. Med. Phys. 2004;31:1017–1022. doi: 10.1118/1.1689013. [DOI] [PubMed] [Google Scholar]
  134. Lee C, Lee C, Park SH, Lee JK. Development of the two Korean adult tomographic computational phantoms for organ dosimetry. Med. Phys. 2006;33:380–390. doi: 10.1118/1.2161405. [DOI] [PubMed] [Google Scholar]
  135. Lee C, Lodwick D, Hasenauer D, Williams JL, Lee C, Bolch WE. Hybrid computational phantoms of the male and female newborn patient: NURBS-based whole-body models. Phys. Med. Biol. 2007;52:3309–3333. doi: 10.1088/0031-9155/52/12/001. [DOI] [PubMed] [Google Scholar]
  136. Lee C, Lodwick D, Williams JL, Bolch WE. Hybrid computational phantoms of the 15-year male and female adolescent: Applications to CT organ dosimetry for patients of variable morphometry. Med. Phys. 2008;35:2366–2382. doi: 10.1118/1.2912178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  137. Lee C, Lodwick D, Hurtado J, Pafundi D, Williams JL, Bolch WE. The UF family of reference hybrid phantoms for computational radiation dosimetry. Phys. Med. Biol. 2010;55:339–363. doi: 10.1088/0031-9155/55/2/002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  138. Leyton M. A Generative Theory of Shape. Berlin, Germany: Springer-Verlag; 2001. [Google Scholar]
  139. Li J, Qiu R, Zhang Z, Liu L, Zeng Z, Bi L, Li W. Organ dose conversion coefficients for external photon irradiation using the Chinese voxel phantom (CVP) Radiat. Prot. Dosimetry. 2009;135:33–42. doi: 10.1093/rpd/ncp087. [DOI] [PubMed] [Google Scholar]
  140. Lim SM, DeNardo GL, DeNardo DA, Shen S, Yuan A, O’Donnell TO, DeNardo SJ. Prediction of myelotoxicity using radiation doses to marrow from body, blood, and marrow sources. J. Nucl. Med. 1997;38:1374–1378. [PubMed] [Google Scholar]
  141. Lima VJM, Cassola VF, Kramer R, Lira CABO, Khoury HJ, Vieira JW. Development of 5- and 10-year-old pediatric phantoms based on polygon mesh surfaces. Med. Phys. 2011;38:4723–4736. doi: 10.1118/1.3615623. [DOI] [PubMed] [Google Scholar]
  142. Liu T, Du X, Su L, Ji W, Carothers C, Shephard M, Liu B, Kalra MK, Brown FB, Fitzegerald P, Xu XG. ARCHER-CT, an extremely fast Monte Carlo code for patient-specific CT dose calculations using Nvidia GPU and Intel coprocessor technologies: Part I—software development and testing. Phys. Med. Biol. 2014 (submitted) [Google Scholar]
  143. Loevinger R, Japha EM, Brownell GL. In: Discrete radioisotope sources Radiation Dosimetry. Hine GJ, Brownell GL, editors. New York, NY: Academic Press; 1965a. pp. 693–799. [Google Scholar]
  144. Loevinger R, Hiolt JG, Hine GJ. In: Internally administered radionuclides Radiation Dosimetry. Hine GJ, Brownell GL, editors. New York, NY: Academic Press; 1965b. pp. 801–873. [Google Scholar]
  145. Loevinger R. In: Distributed radionuclide sources Radiation Dosimetry Volume III. 2nd Edition. Attix FH, Tochilin, editors. New York, NY: Academic Press; 1969. p. 51. [Google Scholar]
  146. Marinelli LD. Dosage determination with radioactive isotopes Am. J. Roentgenol. Rad. Therapy. 1942;47:210–216. [Google Scholar]
  147. Marinelli LD, Quimby EH, Hine GJ. Dosage determination with radioactive isotopes II. Practical considerations in therapy and protection. Am. J. Roentgenol. Rad. Therapy. 1948;59:260–281. [PubMed] [Google Scholar]
  148. Mason PA, Ziriax JM, Hurt WD, Walters TJ, Ryan KL, Nelson DA, Smith KI, D’andrea JA. Recent advancements in dosimetry measurements and modeling. In: Klauenberg BJ, Miklavčič D, editors. Radio Frequency Radiation Dosimetry and Its Relationship to the Biological Effects of Electromagnetic Fields. Dordrecht, Netherlands: Kluwer Academic Publishers; 2000. pp. 141–155. [Google Scholar]
  149. Maynard MR, Geyer JW, Aris JP, Shifrin RY, Bolch W. The UF family of hybrid phantoms of the developing human fetus for computational radiation dosimetry. Phys. Med. Biol. 2011;56:4839–4879. doi: 10.1088/0031-9155/56/15/014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  150. Mazzurana M, Sandrini L, Vaccari A, Malacarne C, Cristoforetti L, Pontalti R. A semi-automatic method for developing an anthropomorphic numerical model of dielectric anatomy by MRI. Phys. Med. Biol. 2003;48:3157–3170. doi: 10.1088/0031-9155/48/19/005. [DOI] [PubMed] [Google Scholar]
  151. McGurk R, Seco J, Riboldi M, Wolfgang J, Segars P, Paganetti H. Extension of the NCAT phantom for the investigation of intra-fraction respiratory motion in IMRT using 4D Monte Carlo. Phys. Med. Biol. 2010;55:1475–1490. doi: 10.1088/0031-9155/55/5/014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  152. Mille M, Xu XG. Fabrication of Human Organs for Realistic Calibration Phantoms by Rapid Prototyping. Health Phys. 2008;95(1):S11. [Google Scholar]
  153. Mille M, Xu XG, Rivard M. Comparison of organ doses for patients undergoing balloon brachytherapy of the breast with HDR 192Ir or electronic sources using Monte Carlo simulations in a heterogeneous human phantom. Med. Phys. 2010;37:662–671. doi: 10.1118/1.3292292. [DOI] [PMC free article] [PubMed] [Google Scholar]
  154. Mille M. Ph.D. Dissertation. Troy, NY: Rensselaer Polytechnic Institute; 2013. A study of shape-dependent partial volume correction in PET imaging using ellipsoidal phantoms fabricated via rapid prototyping. [Google Scholar]
  155. Mishra P, Li R, James SS, Mak RH, Williams CL, Yue Y, Berbeco RI, Lewis JH. Evaluation of 3D fluoroscopic image generation from a single planar treatment image on patient data with a modified XCAT phantom. Phys. Med. Biol. 2013;58:841–858. doi: 10.1088/0031-9155/58/4/841. [DOI] [PMC free article] [PubMed] [Google Scholar]
  156. Mofrad FB, Zoroofi RA, Tehrani-Fard AA, Akhlahpoor S, Hori M, Chen YW, Sato Y. Statistical construction of a Japanese male liver phantom for internal radionuclide dosimetry. Radiat. Prot. Dosimetry. 2010;141:140–148. doi: 10.1093/rpd/ncq164. [DOI] [PubMed] [Google Scholar]
  157. Moignier A, Derreumaux S, Broggio D, Beurrier J, Chea M, Boisserie G, Franck D, Aubert B, Mazeron JJ. Potential of hybrid computational phantoms for retrospective heart dosimetry after breast radiation therapy: A feasibility study. Int. J. Radiat. Oncol. Biol. Phys. 2013;85:492–499. doi: 10.1016/j.ijrobp.2012.03.037. [DOI] [PubMed] [Google Scholar]
  158. Na YH, Zhang B, Zhang JY, Caracappa PF, Xu XG. Deformable adult human phantoms for radiation protection dosimetry: Anthropometric data representing size distributions of adult worker populations and software algorithms. Phys. Med. Biol. 2010;55:3789–3811. doi: 10.1088/0031-9155/55/13/015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  159. Na YH. Ph.D. Dissertation. Troy, NY: Rensselaer Polytechnic Institute; 2009. Deformable adult human phantoms for radiation protection dosimetry: methods for adjusting body and organ sizes to match population-based percentile data. [Google Scholar]
  160. Nagaoka T, Togashi T, Saito K, Takahashi M, Ito K, Ueda T, Osada H, Ito H, Watanabe S. An anatomically realistic voxel model of the pregnant woman and numerical dosimetry for a whole-body exposure to RF electromagnetic fields. Conf. Proc. IEEE Eng. Med. Biol. Soc. 2006;1:5463–5467. doi: 10.1109/IEMBS.2006.260807. [DOI] [PubMed] [Google Scholar]
  161. Nagaoka T, Kunieda E, Watanabe S. Proportion-corrected scaled voxel models for Japanese children and their application to the numerical dosimetry of specific absorption rate for frequencies from 30 MHz to 3 GHz. Phys. Med. Biol. 2008;53:6695–6711. doi: 10.1088/0031-9155/53/23/004. [DOI] [PubMed] [Google Scholar]
  162. Nagaoka T, Togashi T, Saito K, Takahashi M, Ito K, Watanabe S. An anatomically realistic whole-body pregnant-woman model and specific absorption rates for pregnant-woman exposure to electromagnetic plane waves from 10 MHz to 2 GHz. Phys. Med. Biol. 2007;52:6731–6745. doi: 10.1088/0031-9155/52/22/012. [DOI] [PubMed] [Google Scholar]
  163. Nagaoka T, Watanabe S, Sakurai K, Kunieda E, Watanabe S, Taki M, Yamanaka Y. Development of realistic high-resolution whole-body voxel models of Japanese adult males and females of average height and weight, and application of models to radio-frequency electromagnetic-field dosimetry. Phys. Med. Biol. 2004;49:1–15. doi: 10.1088/0031-9155/49/1/001. [DOI] [PubMed] [Google Scholar]
  164. NCRP. The experimental basis for absorbed-dose calculations in medical uses of radionuclides. Besthesda, MD: National Council on Radiation Protection and Measurements; 1985. NCRP Report No. 83. [Google Scholar]
  165. Nipper JC, Williams JL, Bolch WE. Creation of two tomographic voxel models of paediatric patients in the first year of life. Phys. Med. Biol. 2002;47:3143–3164. doi: 10.1088/0031-9155/47/17/307. [DOI] [PubMed] [Google Scholar]
  166. Niu X, Yang Y, Jin M, Wernick MN, King MA. Regularized fully 5D reconstruction of cardiac gated dynamic SPECT images. IEEE Trans. Nucl. Sci. 2010;57:1085–1095. doi: 10.1109/TNS.2010.2047731. [DOI] [PMC free article] [PubMed] [Google Scholar]
  167. NRC. EGSnrc. [last access 03/02/14];2013 http://irs.inms.nrc.ca/software/egsnrc/
  168. Pafundi D, Lee C, Watchman C, Bourke V, Aris J, Shagina N, Harrison J, Fell T, Bolch W. An image-based skeletal tissue model for the ICRP reference newborn. Phys. Med. Biol. 2009;54:4497–4531. doi: 10.1088/0031-9155/54/14/009. [DOI] [PubMed] [Google Scholar]
  169. Pafundi D, Rajon D, Jokisch D, Lee C, Bolch W. An image-based skeletal dosimetry model for the ICRP reference newborn—internal electron sources. Phys. Med. Biol. 2010;55:1785–1814. doi: 10.1088/0031-9155/55/7/002. [DOI] [PubMed] [Google Scholar]
  170. Park S, Lee JK, Lee C. Development of a Korean adult male computational phantom for internal dosimetry calculation. Radiat. Prot. Dosimetry. 2006;121:257–264. doi: 10.1093/rpd/ncl042. [DOI] [PubMed] [Google Scholar]
  171. Patni HK, Nadar MY, Akar DK, Bhati S, Sarkar PK. Selected organ dose conversion coefficients for external photons calculated using ICRP adult voxel phantoms and Monte Carlo code Fluka. Radiat. Prot. Dosimetry. 2011;147:406–416. doi: 10.1093/rpd/ncq462. [DOI] [PubMed] [Google Scholar]
  172. Pelowitz DB, editor. MCNPX User’s Manual Version 2.5.0. Los Alamos, NM: Los Alamos National Laboratory; 2005. Report LA-CP-05-0369. [Google Scholar]
  173. Petoussi-Henss N, Zankl M, Fill U, Regulla D. The GSF family of voxel phantoms. Phys. Med. Biol. 2002;47:89–106. doi: 10.1088/0031-9155/47/1/307. [DOI] [PubMed] [Google Scholar]
  174. Phantom Laboratory. RANDO Phantoms. [last access 03/02/14];2013 http://www.phantomlab.com/rando.html.
  175. Pretorius PH, Xia W, King MA, Tsui BM, Pan TS, Villegas BJ. Evaluation of right and left ventricular volume and ejection fraction using a mathematical cardiac torso phantom. J. Nucl. Med. 1997;38:1528–1535. [PubMed] [Google Scholar]
  176. Pretorius PH, King MA, Tsui BM, LaCroix KJ, Xia W. A mathematical model of motion of the heart for use in generating source and attenuation maps for simulating emission imaging. Med. Phys. 1999;26:2323–2332. doi: 10.1118/1.598746. [DOI] [PubMed] [Google Scholar]
  177. Pujol A, Gibbs SJ. A Monte Carlo method for patient dosimetry from dental x-ray. Dentomaxillofac Radiol. 1982;11:25–33. doi: 10.1259/dmfr.1982.0003. [DOI] [PubMed] [Google Scholar]
  178. Qiu R, Li J, Zhang Z, Wu Z, Zeng Z, Fan J. Photon SAF calculation based on the Chinese mathematical phantom and comparison with the ORNL phantoms. Health Phys. 2008;95:716–724. doi: 10.1097/01.HP.0000318889.50519.56. [DOI] [PubMed] [Google Scholar]
  179. Quimby EH. The development of radiation dosimetry in nuclear medicine. In: Cloutier RJ, Edwards CL, Snyder WS, editors. Medical Radionuclides: Radiation Dose and Effects; Proceedings of AEC Symposium Series 20 (CONF-691212); Washington, DC: U.S. Atomic Energy Commission; 1970. pp. 7–15. [Google Scholar]
  180. Raeside DE. Monte Carlo principles and applications. Phys. Med. Biol. 1976;21:181–197. doi: 10.1088/0031-9155/21/2/001. [DOI] [PubMed] [Google Scholar]
  181. Reece WD, Poston JW, Sr, Xu XG. Determining the effective dose equivalent for external photon radiation: Calculational results for beam and point source geometries. Radiat. Prot. Dosimetry. 1994;55:5–21. [Google Scholar]
  182. Reece WD, Xu XG. Determining the effective dose equivalent for external photon radiation: Assessing effective dose equivalent from personal dosemeter readings. Radiat. Prot. Dosimetry. 1997;69:167–178. [Google Scholar]
  183. Rogers DW. Fifty years of Monte Carlo simulations for medical physics. Phys. Med Biol. 2006;51:R287–R301. doi: 10.1088/0031-9155/51/13/R17. [DOI] [PubMed] [Google Scholar]
  184. Sachse FB, et al. MEET Man–Models for Simulation of Electromagnetic, Elastomechanic and Thermic Behavior of Man. Karlsruhe, Germany: Universität Karlsruhe; 1997. [last access 03/02/14]. available online http://www.ibt.kit.edu/english/973.php. [Google Scholar]
  185. Saito K, Wittmann A, Koga S, Ida Y, Kamei T, Funabiki J, Zankl M. Construction of a computed tomographic phantom for a Japanese male adult and dose calculation system. Radiat. Environ. Biophys. 2001;40:69–75. doi: 10.1007/s004110000082. [DOI] [PubMed] [Google Scholar]
  186. Saito K, Koga S, Ida Y, Kamei T, Funabiki J. Construction of a voxel phantom based on CT data for a Japanese female adult and its use for calculation of organ doses from external electrons. Jpn. J. Health Phys. 2008;43:122–130. [Google Scholar]
  187. Salvat F, Fernández-Varea JM, Sempau J. [last access 03/02/14];PENELOPE—A code system for Monte Carlo simulation of electron and photon transport. 2003 ISBN: 92-64-02145-0, available online http://www2.cose.isu.edu/~tforest/Classes/NucSim/penelope-2003.pdf.
  188. Sato K, Noguchi H, Emoto Y, Koga S, Saito K. Japanese adult male voxel phantom constructed on the basis of CT images. Radiat. Prot. Dosimetry. 2007;123:337–344. doi: 10.1093/rpd/ncl101. [DOI] [PubMed] [Google Scholar]
  189. Sato K, Noguchi H, Endo A, Emoto Y, Koga S, Saito K. Development of a voxel phantom of Japanese adult male in upright posture. Radiat. Prot. Dosimetry. 2007;127:205–208. doi: 10.1093/rpd/ncm272. [DOI] [PubMed] [Google Scholar]
  190. Sato K, Noguchi H, Emoto Y, Koga S, Saito K. Development of a Japanese adult female voxel phantom. J. Nucl. Sci. Technol. 2009;46:907–913. [Google Scholar]
  191. Schlattl H, Zankl M, Petoussi-Henss N. Organ dose conversion coefficients for voxel models of the reference male and female from idealized photon exposures. Phys. Med. Biol. 2007;52:2123–2145. doi: 10.1088/0031-9155/52/8/006. [DOI] [PubMed] [Google Scholar]
  192. Schultz FW, Zoetelief J. Organ and effective doses in the male phantom Adam exposed in AP direction to broad unidirectional beams of monoenergetic electrons. Health Phys. 1996;70:498–504. doi: 10.1097/00004032-199604000-00006. [DOI] [PubMed] [Google Scholar]
  193. Segars WP. Ph.D. Dissertation. Chapel Hill, NC: University of North Carolina; 2001. Development and application of the new dynamic NURBS-based cardiac-torso (NCAT) phantom. [Google Scholar]
  194. Segars WP, Tsui BMW, Frey EC, Johnson GA, Berr SS. Development of a 4-D digital mouse phantom for molecular imaging research. Mol. Imaging Biol. 2004;6:149–159. doi: 10.1016/j.mibio.2004.03.002. [DOI] [PubMed] [Google Scholar]
  195. Segars WP, Tsui B. 4D MOBY and NCAT phantoms for medical imaging simulation of mice and men. J. Nucl. Med. 2007;48(Suppl. 2):203P. [Google Scholar]
  196. Segars WP, Lalush DS, Frey EC, Manocha D, King M, Tsui BMW. Improved dynamic cardiac phantom based on 4D NURBS and tagged MRI. IEEE Trans. Nucl. Sci. 2009;56:2728–2738. doi: 10.1109/TNS.2009.2016196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  197. Segars WP, Tsui BMW. The MCAT, NCAT, XCAT, and MOBY computational human and mouse phantoms. In: Xu XG, EckermanK F, editors. Handbook of Anatomical Models for Radiation Dosimetry. Boca Raton, FL: Taylor & Francis; 2009. pp. 105–134. [Google Scholar]
  198. Segars WP, Sturgeon GM, Ward DJ, Ratnanather JT, Miller MI, Tsui BMW. The new XCAT series of digital phantoms for multi-modality imaging. IEEE Nucl. Sci. Symp. & Med. Imaging Conf. 2010:2392–2395. [Google Scholar]
  199. Segars WP, et al. Population of anatomically variable 4D XCAT adult phantoms for imaging research and optimization. Med. Phys. 2013;40:043701. doi: 10.1118/1.4794178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  200. Shi C, Xu XG. Development of a 30-week-pregnant female tomographic model from computed tomography (CT) images for Monte Carlo organ dose calculations. Med. Phys. 2004;31:2491–2497. doi: 10.1118/1.1778836. [DOI] [PubMed] [Google Scholar]
  201. Shi C, Xu XG, Stabin MG. SAF values for internal photon emitters calculated for the RPI-P pregnant-female models using Monte Carlo methods. Med. Phys. 2008;35:3215–3224. doi: 10.1118/1.2936414. [DOI] [PMC free article] [PubMed] [Google Scholar]
  202. Shi CY. Ph.D. Dissertation. Troy, NY: Rensselaer Polytechnic Institute; 2004. Development and application of a tomographic model from CT images for calculating internal dose to a pregnant woman. [Google Scholar]
  203. Sjögreen K, Ljungberg M, Wingårdh K, Erlandsson K, Strand SE. Registration of emission and transmission whole-body scintillation-camera images. J. Nucl. Med. 2001;42:1563–1570. [PubMed] [Google Scholar]
  204. Smans K, Tapiovaara M, Cannie M, Struelens L, Vanhavere F, Smet M, Bosmans H. Calculation of organ doses in x-ray examinations of premature babies. Med. Phys. 2008;35:556–568. doi: 10.1118/1.2828183. [DOI] [PubMed] [Google Scholar]
  205. Smith TJ, Petouissi N, Zankl M. Comparison of internal radiation doses estimated by MIRD and voxel techniques for a “family” of phantoms. Eur. J. Nucl. Med. 2000;27:1387–1398. doi: 10.1007/s002590000294. [DOI] [PubMed] [Google Scholar]
  206. Snyder WS. Health Physics Division Annual Progress report for Period Ending July 31, 1967. Oak Rdige, TN: Oak Ridge National Laboratory; 1967. Variation of dose in man from exposure to a point source of gamma rays; p. 257. Report ORNL-4168. [Google Scholar]
  207. Snyder WS, Ford MR, Warner GG, Fisher HL., Jr Estimates of absorbed fractions for monoenergetic photon sources uniformly distributed in various organs of a heterogeneous phantom. J. Nucl. Med. 1969;10(Suppl 3):5–52. [PubMed] [Google Scholar]
  208. Snyder WS, Ford MR, Warner GG. Estimates of specific absorbed fractions for monoenergetic photon sources uniformly distributed in various organs of a heterogeneous phantom. Oak Ridge, TN: Oak Ridge National Laboratory; 1978. MIRD Pamphlet 5, Revised. [PubMed] [Google Scholar]
  209. Son IY, Winslow M, Yazici B, Xu XG. X-ray imaging optimization using virtual phantoms and computerized observer modelling. Phys. Med. Biol. 2006;51:4289–4310. doi: 10.1088/0031-9155/51/17/011. [DOI] [PubMed] [Google Scholar]
  210. Spitzer VM, Whitlock DG. Atlas of the Visible Human Male: Reverse Engineering of the Human Body. Sudbury, MA: Jones and Bartlett Publishers; 1998. [Google Scholar]
  211. Stabin MG, Watson EE, Cristy M, Ryman JC, Eckerman KF, Davis JL, Marshall D, Gehlen MK. Mathematical models and specific absorbed fractions of photon energy in the nonpregnant adult female and at the end of each trimester of pregnancy. Oak Ridge, TN: Oak Ridge National Laboratory; 1995. Report ORNL/TM-12907. [Google Scholar]
  212. Stabin M, Emmons MA, Segars WP, Fernald M, Brill AB. ICRP-89 based adult and pediatric phantom series. J. Nucl. Med. 2008;49(Supp. 1):14P. [Google Scholar]
  213. Stabin MG, Xu XG, Emmons MA, Segars WP, Shi C, Fernald MJ. RADAR reference adult, pediatric and pregnant female phantom series for internal and external dosimetry. J. Nucl. Med. 2012;53:1807–1813. doi: 10.2967/jnumed.112.106138. [DOI] [PMC free article] [PubMed] [Google Scholar]
  214. Staton RJ, Jones A Kyle, Lee C, Hintenlang David E, Arreola MM, Williams Jonathon L, Bolch Wesley E. A tomographic physical phantom of the newborn child with real-time dosimetry II. Scaling factors for calculation of mean organ dose in pediatric radiography. Med. Phys. 2006;33:3283–3289. doi: 10.1118/1.2256687. [DOI] [PubMed] [Google Scholar]
  215. Storm L, Israel HI. Photon cross sections from 1 keV to 100 MeV for elements Z=1 to Z=100. Atomic Data and Nuclear Data Tables 7. 1970:565–681. [Google Scholar]
  216. Stovall M, Smith SA, Rosenstein M. Tissue doses from radiotherapy of cancer of the uterine cervix. Med. Phys. 1989;16:726–733. doi: 10.1118/1.596331. [DOI] [PubMed] [Google Scholar]
  217. Stroud I. Boundary Representation Modeling Techniques. London, UK: Springer-Verlag; 2006. [Google Scholar]
  218. Su L, Han B, Xu XG. Calculated organ doses for individuals in a sitting posture above a contaminated ground and a PET imaging room. Radiat. Prot. Dosimetry. 2012;148:439–443. doi: 10.1093/rpd/ncr194. [DOI] [PMC free article] [PubMed] [Google Scholar]
  219. Su L, Yang YM, Bednarz B, Sterpin E, Du X, Liu T, Ji W, Xu XG. ARCHER—A photon-electron coupled Monte Carlo dose computing engine for GPU: software development and application to helical tomotherapy. Med. Phys. 2014 doi: 10.1118/1.4884229. (submitted) [DOI] [PMC free article] [PubMed] [Google Scholar]
  220. Sun W, Jia X, Xie T, Xu F, Liu Q. Construction of boundary-surface-based Chinese female astronaut computational phantom and proton dose estimation. J. Radiat. Res. 2013;54:383–397. doi: 10.1093/jrr/rrs100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  221. Tabary J, Marache-Francisco S, Valette S, Segars WP, Lartizien C. Realistic x-ray CT simulation of the XCAT phantom with SINDBAD; 2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC); 2009. pp. 3980–3983. [Google Scholar]
  222. Taddei PJ, Mirkovic D, Fontenot JD, Giebeler A, Zheng Y, Kornguth D, Mohan R, Newhauser WD. Stray radiation dose and second cancer risk for a pediatric patient receiving craniospinal irradiation with proton beams. Phys. Med. Biol. 2009;54:2259–2275. doi: 10.1088/0031-9155/54/8/001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  223. Takahashi M, Kinase S, Kramer R. Evaluation of counting efficiencies of a whole-body counter using Monte Carlo simulation with voxel phantoms. Radiat. Prot. Dosimetry. 2011;144:407–410. doi: 10.1093/rpd/ncq417. [DOI] [PubMed] [Google Scholar]
  224. Taranenko V, Xu XG. Foetal dose conversion coefficients for ICRP-compliant pregnant models from idealised proton exposures. Radiat. Prot. Dosimetry. 2009;133:65–72. doi: 10.1093/rpd/ncp020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  225. Tinniswood AD, Furse CM, Gandhi OP. Power deposition in the head and neck of an anatomically based human body model for plane wave exposures. Phys. Med. Biol. 1998;43:2361–2378. doi: 10.1088/0031-9155/43/8/026. [DOI] [PubMed] [Google Scholar]
  226. Tresser MA, Hintenlang DE. Construction of a newborn dosimetry phantom for measurement of effective dose. Health Phys. 1999;76:S190. [Google Scholar]
  227. Tsui BMW, Zhao XD, Gregoriou GK, Lalush DS, Frey EC, Johnston RE, Mccartney WH. Quantitative cardiac SPECT reconstruction with reduced image degradation due to patient anatomy. IEEE Trans. Nucl. Sci. 1994;41:2838–2844. [Google Scholar]
  228. Tsui BM, Terry JA, Gullberg GT. Evaluation of cardiac cone-beam single photon emission computed tomography using observer performance experiments and receiver operating characteristic analysis. Inv. Radiol. 1993;28:1101–1112. doi: 10.1097/00004424-199312000-00004. [DOI] [PubMed] [Google Scholar]
  229. Tung CJ, Tsai SF, Tsai HY, Chen IJ. Determination of voxel phantom for reference Taiwanese adult from CT image analyses. Radiat. Prot. Dosimetry. 2011;146:186–190. doi: 10.1093/rpd/ncr144. [DOI] [PubMed] [Google Scholar]
  230. Turner JE, Wright HA, Hamm RN. A Monte Carlo primer for health physicists. Health Phys. 1985;48:717–733. doi: 10.1097/00004032-198506000-00001. [DOI] [PubMed] [Google Scholar]
  231. Tward DJ, Ceritoglu C, Sturgeon G, Segars WP, Miller MI, Ratnanather JT. Generating patient-specific dosimetry phantoms with whole-body diffeomorphic image registration; 2011 IEEE 37th Annual Northeast Bioengineering Conference (NEBEC); 2011. pp. 1–2. [Google Scholar]
  232. Uusitupa T, Laakso I, Ilvonen S, Nikoskinen K. SAR variation study from 300 to 5000 MHz for 15 voxel models including different postures. Phys. Med. Biol. 2010;55:1157–1176. doi: 10.1088/0031-9155/55/4/017. [DOI] [PubMed] [Google Scholar]
  233. Vazquez JA, Ding A, Haley T, Caracappa PF, Xu XG. Development of a dynamic computational human phantom for animated dosimetry simulation using motion capture data and a dose-reconstruction demonstration considering the 1997 Sarov criticality accident. Health Phys. 2014a doi: 10.1097/HP.0000000000000019. (in press) [DOI] [PubMed] [Google Scholar]
  234. Vazquez JA, Caracappa PF, Xu XG. Development of posture-specific computational phantoms using motion capture technology and application to radiation dose reconstruction for the 1999 Tokaimura Nuclear Criticality Accident. Phys. Med. Biol. 2014b doi: 10.1088/0031-9155/59/18/5277. (in press) [DOI] [PubMed] [Google Scholar]
  235. Veress AI, Segars WP, Tsui BMW, Gullberg GT. Incorporation of a left ventricle finite element model defining infarction into the XCAT imaging phantom. IEEE Trans. Med. Imaging. 2011;30:915–927. doi: 10.1109/TMI.2010.2089801. [DOI] [PMC free article] [PubMed] [Google Scholar]
  236. Wang B, Xu XG, Kim CH. A Monte Carlo CT model of the Rando Phantom. Trans. Am. Nucl. Soc. 2004;90:473–474. [Google Scholar]
  237. Wang J, Fujiwara O, Watanabe S, Yamanaka Y. Computation with a parallel FDTD system of human-body effect on electromagnetic absorption for portable telephones. IEEE Trans. Microw. Theory Tech. 2004;52:53–58. [Google Scholar]
  238. Wang B, Xu XG, Goorley JT, Bozkurt A. The use Of MCNP code for an extremely large voxel model VIP-Man. The Monte Carlo Method: Versatility 43 Unbounded In A Dynamic Computing World; Proceedings of the Monte Carlo 2005 Topical Meeting; April 17–21; Chattanooga, TN. LaGrange, Park IL: American Nuclear Society; 2005a. [last access 03/02/14]. on CD-ROM (ISBN:0-89448-695-0) also available online http://mathematicsandcomputation.cowhosting.net/MonteCarlo05/FullPapers/usr-georgexu-3-paper.pdf. [Google Scholar]
  239. Wang B, Xu XG, Goldstein M, Sahoo N. Adjoint Monte Carlo method for prostate external photon beam treatment planning: An application to 3-D patient anatomy. Phys. Med. Biol. 2005b;50:923–935. doi: 10.1088/0031-9155/50/5/015. [DOI] [PubMed] [Google Scholar]
  240. Williams G, Zankl M, Abmayr W, Veit R, Drexler G. The calculation of dose from external photon exposures using reference and realistic human phantoms and Monte Carlo methods. Phys. Med. Biol. 1986;31:449–452. doi: 10.1088/0031-9155/31/4/010. [DOI] [PubMed] [Google Scholar]
  241. Winslow M, Huda W, Xu XG, Chao TC, Shi CY, Ogden KM, Scalzetti EM. Use of the VIP-Man model to calculate energy imparted and effective dose for x-ray examinations. Health Phys. 2004;86:174–182. doi: 10.1097/00004032-200402000-00008. [DOI] [PubMed] [Google Scholar]
  242. Winslow M, Xu XG, Yazici B. Development of a simulator for radiographic image optimization. Comput. Methods Programs Biomed. 2005;78:179–190. doi: 10.1016/j.cmpb.2005.02.004. [DOI] [PubMed] [Google Scholar]
  243. Wu D, Shamsi S, Chen J, Kainz W. Evaluations of specific absorption rate and temperature increase within pregnant female models in magnetic resonance imaging birdcage coils. IEEE Trans. Microw. Theory Tech. 2006;54:4472–4478. [Google Scholar]
  244. Xie T, Zaidi H. Evaluation of radiation dose to anthropomorphic paediatric models from positron-emitting labelled tracers. Phys. Med. Biol. 2014;59:1165–1187. doi: 10.1088/0031-9155/59/5/1165. [DOI] [PubMed] [Google Scholar]
  245. Xie T, Bolch WE, Lee C, Zaidi H. Pediatric radiation dosimetry for positron-emitting radionuclides using anthropomorphic phantoms. Med. Phys. 2013;40:102502–102514. doi: 10.1118/1.4819939. [DOI] [PubMed] [Google Scholar]
  246. Xie T, Zhang G, Li Y, et al. Comparison of absorbed fractions of electrons and photons using three kinds of computational phantoms of rat. Applied Physics Letters. 2010;97(3) [Google Scholar]
  247. Xu XG, Reece WD, Poston JW. A Study of the angular dependence problem in effective dose equivalent assessment. Health Phys. 1995;68:214–224. doi: 10.1097/00004032-199502000-00007. [DOI] [PubMed] [Google Scholar]
  248. Xu XG, Reece WD. Sex-specific tissue weighting factors for effective dose equivalent calculations. Health Phys. 1996;70:81–86. doi: 10.1097/00004032-199601000-00012. [DOI] [PubMed] [Google Scholar]
  249. Xu XG, Chao TC, Bozkurt A. VIP-Man: An image-based whole-body adult male model constructed from color photographs of the Visible Human Project for multi-particle Monte Carlo calculations. Health Phys. 2000;78:476–486. doi: 10.1097/00004032-200005000-00003. [DOI] [PubMed] [Google Scholar]
  250. Xu XG, Chao TC, Bozkurt A. Comparison of effective doses from various monoenergetic particles based on the stylised and the VIP-Man tomographic models. Radiat. Prot. Dosimetry. 2005;115:530–535. doi: 10.1093/rpd/nci116. [DOI] [PubMed] [Google Scholar]
  251. Xu XG, Taranenko V, Zhang J, Shi C. A boundary-representation method for designing whole-body radiation dosimetry models: pregnant females at the ends of three gestational periods—RPI-P3, -P6 and -P9. Phys. Med. Biol. 2007;52:7023–7044. doi: 10.1088/0031-9155/52/23/017. [DOI] [PubMed] [Google Scholar]
  252. Xu XG. Computational phantoms for radiation dosimetry: A 40-year history of evolution. In: Xu XG, Eckerman KF, editors. Handbook of Anatomical Models for Radiation Dosimetry. Boca Raton, FL: Taylor & Francis; 2009. pp. 3–42. [Google Scholar]
  253. Xu XG, Eckerman KF. Handbook of Anatomical Models for Radiation Dosimetry. Boca Raton, FL: Taylor & Francis; 2009. [Google Scholar]
  254. Yeom YS, Han MC, Kim CH, Jeong JH. Conversion of ICRP male reference phantom to polygon-surface phantom. Phys. Med. Biol. 2013;58:6985–7007. doi: 10.1088/0031-9155/58/19/6985. [DOI] [PubMed] [Google Scholar]
  255. Zaidi H. Relevance of accurate Monte Carlo modeling in nuclear medical imaging. Med. Phys. 1999;26:574–608. doi: 10.1118/1.598559. [DOI] [PubMed] [Google Scholar]
  256. Zaidi H, Sgouros G, editors. Therapeutic Applications of Monte Carlo Calculations in Nuclear Medicine. London, UK: IOP Publishing; 2003. [Google Scholar]
  257. Zaidi H, Xu XG. Computational anthropomorphic models of the human anatomy: the path to realistic Monte Carlo modeling in radiological sciences. Annu. Rev. Biomed. Eng. 2007;9:471–500. doi: 10.1146/annurev.bioeng.9.060906.151934. [DOI] [PubMed] [Google Scholar]
  258. Zaidi H, Tsui BMW. Review of computational anthropomorphic anatomical and physiological models. Proc. IEEE. 2009;97:1938–1953. [Google Scholar]
  259. Zankl M, Fill U, Petoussi-Henss N, Regulla D. Organ dose conversion coefficients for external photon irradiation of male and female voxel models. Phys. Med. Biol. 2002;47:2367–2385. doi: 10.1088/0031-9155/47/14/301. [DOI] [PubMed] [Google Scholar]
  260. Zankl M, Veit R, Williams G, Schneider K, Fendel H, Petoussi N, Drexler G. The construction of computer tomographic phantoms and their application in radiology and radiation protection. Radiat. Environ. Biophys. 1988;27:153–164. doi: 10.1007/BF01214605. [DOI] [PubMed] [Google Scholar]
  261. Zankl M, Becker J, Fill U, Petoussi-Henss N, Eckerman KF. GSF male and female adult voxel models representing ICRP Reference Man—the present status. The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World; Proceedings of the Monte Carlo 2005 Topical Meeting held; April 17–21; Chattanooga, TN. LaGrange, Park IL: American Nuclear Society; 2005. [last access 03/02/14]. on CD-ROM (ISBN:0-89448-695-0) also available online http://mathematicsandcomputation.cowhosting.net/MonteCarlo05/FullPapers/usr-zankl-1-paper.pdf. [Google Scholar]
  262. Zeng Z, Li J, Qiu R. Dose assessment for space radiation using a proton differential dose spectrum. Journal of Tsinghua University (Science and Technology) 2006;46:374–376. [Google Scholar]
  263. Zhang B, Ma J, Liu L, Cheng J. CNMAN: A Chinese adult male voxel phantom constructed from color photographs of a visible anatomical data set. Radiat. Prot. Dosimetry. 2007a;124:130–136. doi: 10.1093/rpd/ncm184. [DOI] [PubMed] [Google Scholar]
  264. Zhang G, Liu Q, Luo Q. Monte Carlo simulations for external neutron dosimetry based on the visible Chinese human phantom. Phys. Med. Biol. 2007b;52:7367–7383. doi: 10.1088/0031-9155/52/24/011. [DOI] [PubMed] [Google Scholar]
  265. Zhang G, Liu Q, Zeng S, Luo Q. Organ dose calculations by Monte Carlo modeling of the updated VCH adult male phantom against idealized external proton exposure. Phys. Med. Biol. 2008a;53:3697–3722. doi: 10.1088/0031-9155/53/14/001. [DOI] [PubMed] [Google Scholar]
  266. Zhang G, Luo Q, Zeng S, Liu Q. The development and application of the visible Chinese human model for Monte Carlo dose calculations. Health Phys. 2008b;94:118–125. doi: 10.1097/01.HP.0000285256.48498.b4. [DOI] [PubMed] [Google Scholar]
  267. Zhang JY, Xu XG, Shi C, Fuss M. Development of a geometry-based respiratory motion-simulating patient model for radiation treatment dosimetry. J. Appl. Clin. Med. Phys. 2008c;9:2700. doi: 10.1120/jacmp.v9i1.2700. [DOI] [PMC free article] [PubMed] [Google Scholar]
  268. Zhang B, Ma J, Zhang G, Liu Q, Qiu R, Li J. The Chinese computational phantoms: CNMAN, VCH, and CVP. In: Xu XG, Eckerman KF, editors. Handbook of Anatomical Models for Radiation Dosimetry. Boca Raton, FL: Taylor & Francis; 2009a. pp. 279–304. [Google Scholar]
  269. Zhang J, Na YH, Caracappa PF, Xu XG. RPI-AM and RPI-AF, a pair of mesh-based, size-adjustable adult male and female computational phantoms using ICRP-89 parameters and their calculations for organ doses from monoenergetic photon beams. Phys. Med. Biol. 2009b;54:5885–5908. doi: 10.1088/0031-9155/54/19/015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  270. Ziriax JM, Smith KI, Nelson DA, Ryan KL, Gajsek P, D’Andrea JA, Walters TJ, Hurt WD, Mason PA. Effects of frequency, permittivity, and voxel size on predicted specific absorption rate values in biological tissue during electromagnetic-field exposure. IEEE Trans. Microw. Theory Tech. 2000;48:2050–2058. [Google Scholar]
  271. Zubal IG, Harrell CR, Smith EO, Rattner Z, Gindi G, Hoffer PB. Computerized three-dimensional segmented human anatomy. Med. Phys. 1994;21:299–302. doi: 10.1118/1.597290. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

pmb303293suppdata.pdf

RESOURCES