Abstract
Background
Measuring esophageal pressure (Pes) using an air-filled balloon catheter (BC) is the common approach to estimate pleural pressure and related parameters. However, Pes is not routinely measured in mechanically ventilated patients, partly due to technical and practical limitations and difficulties. This study aimed at comparing the conventional BC with two alternative methods for Pes measurement, liquid-filled and air-filled catheters without balloon (LFC and AFC), during mechanical ventilation with and without spontaneous breathing activity. Seven female juvenile pigs (32–42 kg) were anesthetized, orotracheally intubated, and a bundle of an AFC, LFC, and BC was inserted in the esophagus. Controlled and assisted mechanical ventilation were applied with positive end-expiratory pressures of 5 and 15 cmH2O, and driving pressures of 10 and 20 cmH2O, in supine and lateral decubitus.
Main Results
Cardiogenic noise in BC tracings was much larger (up to 25% of total power of Pes signal) than in AFC and LFC (<3%). Lung and chest wall elastance, pressure-time product, inspiratory work of breathing, inspiratory change and end-expiratory value of transpulmonary pressure were estimated. The three catheters allowed detecting similar changes in these parameters between different ventilation settings. However, a non-negligible and significant bias between estimates from BC and those from AFC and LFC was observed in several instances.
Conclusions
In anesthetized and mechanically ventilated pigs, the three catheters are equivalent when the aim is to detect changes in Pes and related parameters between different conditions, but possibly not when the absolute value of the estimated parameters is of paramount importance. Due to a better signal-to-noise ratio, and considering its practical advantages in terms of easier calibration and simpler acquisition setup, LFC may prove interesting for clinical use.
Introduction
The assessment of several mechanical and functional properties of the lungs and chest wall depends on the estimation of pleural pressure (Ppl). The end-expiratory transpulmonary pressure (Ptrans, the difference between alveolar pressure and Ppl) has been used for positive end-expiratory pressure (PEEP) adjustment in mechanically ventilated patients [1]. Estimates of Ppl are also necessary to evaluate respiratory effort and work of breathing, for example during assisted mechanical ventilation (MV) and weaning from MV, in order to detect the occurrence of respiratory fatigue [2]–[4]. Furthermore, Ppl can be used to split estimates of total respiratory system elastance into its pulmonary and chest wall components [2].
In clinical practice, Ppl is estimated from the esophageal pressure (Pes). Usually, Pes is measured with an air-filled balloon catheter inserted in the middle portion of the esophagus [5], [6]. Several investigations reported a good agreement between measurement of the changes in Ppl and Pes [7]–[9]. Noteworthily, other studies reported that an offset exists between absolute values of Ppl and Pes, and that the agreement between the changes in those two pressures is reduced in mechanically ventilated subjects, supine position, and injured (i.e. more heterogeneous) lungs [10]–[14].
Despite its potential benefits, Pes is not routinely measured in mechanically ventilated patients [4]. This is partly related to the following factors: 1) complex calibration, involving balloon filling and maneuvers on the patient; 2) contamination of the measurements by cardiogenic and movement artifacts; and 3) need of repositioning and recalibration due to signal quality loss over time [4], [8], [15], [16]. Thus, alternative technologies that improve the signal-to-noise ratio of Pes measurement and simplify its use at the bedside are desirable.
The main aim of this study was to compare the traditional air-filled balloon catheter technique with liquid-filled and air-filled catheters without balloon for measurement of Pes during controlled MV, as well as MV with spontaneous breathing, in juvenile pigs without lung injury. We calculated the partitioned elastance (EL for lung and Ecw for chest wall), transpulmonary pressure (PL), esophageal pressure time product (PTP), and the inspiratory work of breathing (WOBi) with each of those catheter techniques and assessed the differences. We hypothesized that both the liquid-filled and air-filled catheters without balloon can be used interchangeably with the traditional air-filled balloon catheter technique for monitoring of Pes, and estimating related parameters.
Materials and Methods
Experimental protocol
After ethical approval by the local authorities (Landesdirektion Sachsen, Dresden, Germany, reg. nr. 24-9168.11-1/2009-27), seven healthy female juvenile pigs (32–42 kg) were intravenously anesthetized (propofol 2–7 mg/kg/h, sufentanil 0.3–1.5 µg/kg/h) and tracheally intubated with a cuffed tube (8.0 mm inner diameter).
After intubation, the animals were mechanically ventilated using an EVITA XL ventilator (Dräger, Germany) with biphasic positive airway pressure/airway pressure release ventilation (BIPAP/APRV) mode (FiO2 = 1), which consists in transitions between two levels of airways pressure (Plow and Phigh) at a fixed rate. After the administration of a bolus of atracurium bromide (1 mg/kg) to abolish spontaneous breaths, the animals underwent a sequence of changes in the levels of PEEP ( = Plow) and driving pressure (ΔP = Phigh–Plow), resulting in the following combinations of PEEP/ΔP: 5/10, 5/20, 15/10, and 15/20 cmH2O. Inspiration and expiration durations were fixed to last 1 s in all conditions. All steps of the sequence lasted 2 minutes and were performed first in supine and then in right lateral decubitus. The absence of spontaneous breathing in this phase was monitored by visual inspection of Pes tracing. When muscle relaxation ceased and consistent inspiratory efforts appeared in the Pes tracings, the animals were kept in right lateral decubitus under BIPAP/APRV ventilation (PEEP = 5 cmH2O, ΔP = 10 cmH2O), but allowing unsupported spontaneous breaths, with inspiratory time of 1 s and expiratory time adjusted manually to permit three or more spontaneous breaths after each controlled cycle. This step lasted approximately 4 minutes and was then repeated in the supine decubitus. The animals were then euthanized with an i.v. injection of thiopental (2 g) and KCl 1 M (50 mL).
Three catheters were adopted for esophageal pressure measurement: a conventional commercially available balloon catheter (Cardinal Health, USA); an air-filled catheter, obtained by removing the balloon portion from the previous catheter; a liquid-filled catheter, adapted from the disposable tubing commonly used for invasive blood pressure measurement (taken from the DTXPlus kit, Becton Dickinson, USA) by removing its distal Luer-lock adapter and creating four holes near the tip, as illustrated in Figure 1. A constant flow of saline solution of approximately 3 ml/h was maintained in the catheter. The three catheters were bundled in a combined catheter, as illustrated in Figure 1, and introduced in the esophagus. The inner diameter was 2.0 mm for the liquid-filled catheter and 2.6 mm for the others. Due to their mechanical properties, all catheters can be considered as rigid tubes for what concerns the propagation of pressure waves in the range of measured Pes. Bench tests were performed by placing the bundle of catheters horizontally at the bottom of a recipient and increasing the water level from 0 to 5, 15, and 20 cm. Pressure measurement errors were always <5%, and the differences between catheters were always <1%.
Data acquisition and processing
The correct positioning of the catheters in the esophagus was achieved using the signal of the air-filled balloon catheter as a reference, according to the procedure previously described [5]. Briefly, after inserting the empty balloon catheter in the stomach and injecting 0.5 mL of air into the balloon, the catheter was withdrawn slowly until Pes oscillations appeared and were in phase with airways pressure oscillations (i.e. balloon in the esophagus). Following that, the catheter was drawn a further 10 cm, in order to position the balloon midway between the apex and the base of the lungs. Occlusion and chest compression maneuvers were then used to correct the final position of the balloon and the amount of air in the balloon catheter in order to obtain a ratio between Pes and airways opening pressure of ∼1.
These maneuvers were performed at the beginning of the experiment during controlled MV in the paralyzed animals. Occlusion maneuvers during spontaneous breathing efforts, as previously described [7], were performed before and after the assisted ventilation task, which confirmed in all cases that repositioning and recalibration of the catheters were not necessary. The balloon catheter, air-filled catheter, and airway opening were connected to gas pressure transducers (163PC01D48-PCB, Sensortechnics GmbH, Germany). The liquid-filled catheter was connected to a conventional disposable blood pressure transducer (DTXPlus, Becton Dickinson, USA), positioned at the height of the midaxillary line (expected height of the tip of the catheter in the esophagus). The outputs of all transducers were linearly related to pressure in the range of interest, and were calibrated before the experiments adopting a linear regression approach, using several pressure levels obtained with gas pressure generators or water columns. These signal, together with one derivation of EKG obtained with a biosignal amplifier (LP511, Grass Technologies, USA), were synchronously acquired using a data acquisition card (NI USB-6210, National Instruments, USA) with a sampling frequency of 1000 Hz.
Airflow and airway opening pressure signals were continuously acquired from the ventilator and fed into the PC through a serial interface (sampling frequency of 125 Hz), and synchronized off-line with the other signals (by means of automatic time alignment, using maximal covariance, between the airway pressure signal acquired using the data acquisition system and the airway pressure signal that could also be retrieved from the ventilator), after resampling all signals at 100 Hz.
The onset of inspiration and expiration were automatically detected in the airflow signal, and then manually edited. The cycles were manually labeled as controlled or spontaneous, and for the latter the onset of the inspiratory effort was identified in the esophageal pressure tracings (onset of a drop larger than 2 cmH2O occurring right before the onset of inspiration). Ptrans was computed as the difference between the airways opening pressure and Pes measured with each catheter.
In each respiratory cycle, several indexes were estimated using the Pes signal obtained from each catheter. From the Ptrans tracings, the end-expiratory value and the inspiratory change were computed. The dynamic lung and chest wall elastance (EL and Ecw, respectively) were estimated from pressure, volume, and flow signals using a least-squares identification of the coefficients of the conventional linear unicompartmental models of lung mechanics shown in Eqs. 1 and 2:
(Eq.1) |
(Eq.2) |
where RL and Rcw represent resistance to airflow, F is airflow, V is volume, and Pes,0 and Ptrans,0 are the values of Pes and Ptrans when F = 0 and V = 0, respectively. Ecw estimates were not considered for the spontaneous breaths, since Ecw possess a physiological meaning only when the chest wall is a ‘passive’ structure, as in controlled ventilation, which is not the case for spontaneous breaths (resulting in misleading lower Ecw estimates). For the spontaneous breaths only, the pressure time product (PTP) was estimated for each cycle, as shown in Eq. 3:
(Eq.3) |
where t1 corresponds to the onset of the inspiratory effort, and t2 corresponds to the time when Pes returns to its value at t1 (as illustrated in Figure 2-b). The inspiratory work of breathing (WOBi) was estimated for each inspiration based on the Campbell diagram [2], [3]. Average values of the parameters were computed for each subject in the following conditions: controlled breaths during BIPAP/APRV without spontaneous breathing (average of 60 breaths); controlled breaths during BIPAP/APRV with spontaneous breathing (average of 20 breaths); spontaneous unsupported breaths during BIPAP/APRV with spontaneous breathing (average of 80 breaths).
Statistical analysis
During BIPAP/APRV without spontaneous breathing, a general linear model with Greenhouse-Geisser correction was used to test the effect on the average value of each estimated parameter of the method to measure Pes, considering position (lateral/supine), PEEP (5/15 cmH2O), and ΔP (10/20 cmH2O). During BIPAP/APRV with spontaneous breaths, the same analysis was repeated separately for controlled and spontaneous cycles (and disregarding the PEEP and ΔP factors, which were fixed). The amount of cardiogenic noise in each Pes tracing was quantified as the percentage of power of the signal in the frequency band [HR-10 HR+10], where HR is the average heart rate (in bpm) computed from the EKG (power spectrum was computed using Welch’s modified periodogram method [17], using multiple segments corresponding to 40 s of data, a 90% overlap between adjacent segments, and a Hanning windowing). The agreement between estimates of the parameters obtained using different catheters was assessed through the graphical method introduced by Bland and Altman [18], and quantified by means of bias and precision (mean and SD of difference between two methods, respectively). Additionally, the difference between the parameters estimated using different methods to measure Pes was tested using paired t-tests. All tests were performed with the software SPSS (IBM, USA), and statistical significance was accepted at p<0.05.
Results
Figure 2 shows a representative example of Pes tracings obtained with the three catheters, during BIPAP/APRV with and without spontaneous breaths. While the inspiratory/expiratory changes are roughly similar in all instances, the measurement obtained with the air-filled balloon catheter is more contaminated by cardiogenic noise than the other methods. Such phenomenon is also highlighted by the spectra of EKG and balloon signals, showing a coincident large peak at the heart frequency. This was a consistent feature in the tracings of all subjects. In fact, while for the catheters without balloon the fraction of Pes power related to cardiogenic oscillations was only <3%, for the balloon catheter it was>25% during controlled ventilation, and>15% during BIPAP.
Figure 3 shows Bland-Altman plots comparing the estimate of each parameter obtained using the balloon measurements with those obtained using the other two catheters. A considerable dispersion can be noted in all instances, which does not seem to be affected by the magnitude of the estimated parameter. Significant bias was detected in several cases, as reported in Table 1.
Table 1. Bias (precision) of parameters estimated using the liquid- or air-filled catheters compared to those obtained with the balloon catheter, in different conditions.
balloon Vs. | EL | Ecw | ΔPtrans | Ptrans,endexp | PTP | WOBi | |||||||||||
[cmH2O/L] | [cmH2O/L] | [cmH2O] | [cmH2O] | [cmH2O•s] | [cmH2O•L] | ||||||||||||
spont BIPAP | 7.9 | (8.2) | ** | −1.3 | (1.5) | ** | −0.3 | (2.2) | 0.8 | (0.6) | *** | 0.3 | (0.3) | ||||
contr BIPAP | −2.7 | (5.3) | 1.8 | (5.3) | −1.0 | (2.4) | 0.1 | (2.2) | |||||||||
contr MV | −2.0 | (5.0) | ** | 2.0 | (5.0) | ** | −0.6 | (1.4) | ** | −1.1 | (1.4) | *** | |||||
spont BIPAP | 7.7 | (7.6) | ** | 0.3 | (0.9) | −0.9 | (2.4) | 0.4 | (1.0) | 0.2 | (0.5) | ||||||
contr BIPAP | 0.8 | (7.4) | −1.6 | (7.0) | −0.5 | (3.0) | −0.9 | (2.5) | |||||||||
contr MV | 2.4 | (5.5) | ** | −2.4 | (5.5) | ** | 0.6 | (1.6) | * | −1.2 | (2.0) | *** |
BIPAP: biphasic positive airway pressure; contr MV: controlled mechanical ventilation (with animal paralyzed using atracurium bromide); contr BIPAP: assisted ventilation (allowing unsupported spontaneous breaths), considering only the controlled breaths; spont BIPAP: assisted ventilation, considering only the unsupported spontaneous breaths. EL, Ecw: lung and chest wall elastance; ΔPtrans: inspiratory change in transpulmonary pressure; Ptrans,end-exp: end-expiratory value of transpulmonary pressure; PTP: pressure-time product; WOBi: inspiratory work of breathing. Significance of t-test:
*p<0.05;
**p<0.01;
***p<0.001.
Figure 4 illustrates the average estimate of each parameter using the three different measurements of Pes, for each combination of ventilation settings and position tested. The effect of the method to measure Pes was not significant when only controlled breaths were considered (independently of spontaneous breaths being allowed or not). Conversely, a significant effect was found when solely the unsupported spontaneous breaths were employed. In particular, the balloon catheter measurement produced somewhat lower EL, PTP and WOBi estimates during spontaneous breaths. Also, for BIPAP without spontaneous breathing a significant influence of PEEP, ΔP, and position was found for all parameters (except for a non-significant effect of ΔP on EL). For BIPAP allowing unsupported spontaneous breaths a consistent position-dependent effect was not detected.
Discussion
The main findings of this study were that in anesthetized and mechanically ventilated pigs: 1) Pes measured with the balloon catheter showed more contamination by cardiogenic noise than the methods without balloon; 2) in particular, such a noise resulted in consistently lower estimates of EL, PTP and WOBi during spontaneous breathing when using the balloon catheter, as compared to the other catheters; 3) differences between Pes derived parameters showed considerable dispersion, independently of the catheter considered; 4) despite dispersion and bias, all three methods allowed to detect similar changes in EL, PTP, WOBi, inspiratory change and end-expiratory value of transpulmonary pressure between different ventilation settings and positions.
Our primary objective was to assess the viability of techniques alternative to the balloon for the measurement of esophageal pressure in a clinical routine scenario. Two alternatives were considered: liquid-filled and air-filled catheters, which have been previously proposed [6], [9], [19]–[22], but have not found a widespread clinical acceptance, possibly due to the lack of a comprehensive validation study. To compare the techniques, we considered not only the differences in the recorded tracings, but also in several clinically relevant derived indexes, and in the practical difficulties in positioning, calibrating, and managing that can discourage their adoption in the clinical practice.
Overall, the agreement of parameters estimated using the balloon catheter with those measured with air- or liquid-filled catheters is far from ideal. A considerable dispersion was found in all instances, associated in some cases with non-negligible bias. This indicates that the parameters estimated with the air-filled and liquid-filled catheter might not be interchangeable with the balloon catheters when comparing the absolute value of the parameters estimated. However, it is debatable which of the three methods, if any, should be considered the “gold standard”. One notable finding should be mentioned. Even if the phenomenon does not reach statistical significance in all cases, absolute end-expiratory transpulmonary pressure estimated using the balloon catheter appears to be somehow larger in several cases. This is likely related to the fact that the balloon technique tends to measure the lowest pressure found among the several holes present in the catheter along the length of the balloon [23], resulting in a possibly small but negative offset in Pes estimation compared to the other two methods, in which the holes are only near or at the tip the catheter. Adding side holes along the length of the other two catheters should lead to a similar behavior of absolute Pes measurement, if such behavior is desired. However, the results of several investigation question the use of Pes for estimating absolute values of Ppl, since a bias exists between the two at the same location, which can vary with posture and catheter positioning [11], [12], [24]. Also Ppl changes considerably between different locations due to gravitational effects and inhomogeneity of the lungs [4], [25]. Such topology of the absolute value of Ppl cannot be estimated by Pes. In this context, the limitations regarding the possibility of estimating Ppl absolute values are intrinsic to the use of Pes and hence common to all the possible methods/catheters to measure it. Hence, as summarized by Drummond et al., "great attention to absolute pressure measurements in the esophagus is not justified", while the changes in Ppl can be estimated satisfactorily using Pes, at least for the case of a healthy homogenous lung [24]. However, the offset in absolute measurement among methods might be relevant for other applications, such as the PEEP tuning strategy introduced by Talmor et al. [1], which is based on the simplifying assumption that Ppl is the same for the whole lung and is 5 cmH2O smaller than Pes in all conditions.
One source of the dispersion among estimates was the cardiogenic artifacts, which in our tests were generally much larger in the balloon tracings, and represent the most persistent and notable difference among the methods considered.
We speculate that this finding is related to the fact that the balloon is a closed and deformable system. Each ventricular contraction produces the following sequence of events: 1) transient reduction in the volume of the heart; 2) mechanical propagation of part of such deformation to the esophagus; 3) very small increase in the esophageal lumen, with a negligible decrease in Pes; 4) very small transient increase in the internal volume of the balloon, which adheres to part of the internal surface of the esophagus (possibly "glued" to it by a layer of secretion) and consequently follows somewhat its deformation; 5) transient decrease of the pressure within the balloon (since the system is closed) but not in Pes. In this context, the balloon acts partly as a transducer/amplifier of the changes in the geometry of the esophagus unrelated to esophageal pressure, rather than simply transmitting Pes oscillations. Considering that the balloon catheter is a closed system, changes in its internal pressure and volume (ΔP and ΔV) from a given state (P0,V0) must obey the equation (P0+ΔP)•(V0+ΔV) = P0•V0, and consequently ΔP = P0•ΔV/•(V0+ΔV). If a small change in volume is considered (ΔV<<V0), then ΔP≈-P0•ΔV/V0. Considering the reasonable simplification that P0 is approximately the atmospheric pressure ( = 1031 cmH2O), the previous equation shows that changes in the internal volume of the balloon as small as 0.1% (i.e. ΔV/V = 0.001) are capable of generating swings of about 1cmH20. Such swings are compatible with our results, and are unrelated to those in the pressure outside the balloon (Pes, in our case), since they reflect solely the mechanical deformations of the internal volume of the balloon, originated by cardiogenic mechanical excitation. Conversely, being the air- and liquid-filled catheters undeformable and open systems, this effect cannot be observed.
The cardiogenic artifacts are likely to result in larger estimation errors of the parameters using the balloon catheter, which is considerably more affected by this problem according to our results. One example is the estimation of PTP and WOBi. This estimation is based on the area below/above the Pes curve (expressed as function of time or volume, respectively), which is distorted by cardiogenic noise (as exemplified in Figure 2). Even if complex algorithms can be applied in an attempt to reduce (but not cancel) cardiogenic oscillations [26], [27], adopting air- or liquid-filled catheters seems a more practical option in this context. However, we cannot exclude that the discrepancy was exacerbated by the fact that in piglets the heart is relatively larger and nearer to the esophagus than in adult humans.
In spite of that, all the three measuring techniques allowed detecting changes of the same magnitude and direction when MV settings are modified. The results are compatible with the existing literature regarding the effect of PEEP and VT on lung and chest wall compliances [21] and of posture on transpulmonary pressure [28], [29]. Thus, all methods can be used interchangeably when the focus is to detect changes of the parameters, rather than absolute values, among different ventilation conditions.
Finally, several practical limitations and advantages of each of the Pes measurement methods should be considered. While for air- or liquid-filled catheters only a correct positioning is necessary, for the balloon an appropriate choice and testing of the amount of air to be injected is required for a reliable measurement [5]. Also, while for the balloon measurements a specific catheter is required, a common nasogastric catheter might be used for liquid- or air-filled measurements [21]. Furthermore, while for balloon catheters the measurement is affected by several factors such as the amount of injected air, thickness, and dimensions of the balloon [30] − possibly resulting in differences between manufacturers and operators − for systems without balloon the only requirement to guarantee repeatability seems to be a sufficient rigidity of the catheter.
The major limitation encountered with air-filled catheter was the frequent worsening of the signal quality and possible occlusion of the catheter. Reversing this situation (flushing repeatedly the catheter using an air-filled syringe) was, in our experience, a cumbersome procedure in most cases. However, the results found suggest that this limitation did not affect significantly the estimation of the parameters of interest.
Liquid-filled catheters possess the appealing feature that measurements can be readily performed in most intensive care units and operating rooms with the monitors and pre-calibrated disposable transducers routinely used for invasive blood pressure monitoring [20]. The present technology of these transducers grants a very linear response to pressure changes, low noise, and flat frequency response in the range of interest (far superior than what achievable with air-filled or balloon catheters). Also, the built-in system of flushing and continuous flow of liquid (of approximately 3 mL/h) permits easy removal of air-bubbles in the measuring system. It also allows to successfully flush the catheter when signal quality worsen or occlusion is suspected, which occurred only once in two animals during the experiments. Nevertheless, the limitations of liquid-filled catheters, which are known since the 1950s [22], and possibly prevented the adoption of this technique in applications other than pediatric, should not be underestimated. Firstly, movement of the subject (e.g. shivering) can introduce non-negligible artifacts in the measurement. Secondly, large breaths might result in vertical displacement of the esophagus and consequently of the catheter, generating a water column of variable height between the transducer and the tip of the catheter, which affects the measurement of the respiratory swings of Pes. Thirdly, placing the transducer at exactly the same height of the tip of the catheter is not trivial (since the latter is not visible), which likely results in a water-column between the two, and consequently a fixed offset in absolute Pes measurement. However, while this limitation is clearly relevant for standing or sitting subjects, in supine, prone, and lateral decubitus patients we speculate that the bias can be limited using anatomical landmarks and is likely to be small compared to the unknown offset between Ppl and Pes.
Conclusions
In anesthetized and mechanically ventilated pigs, air-filled and liquid-filled esophageal catheters without balloon can be used interchangeably with the traditional catheter with balloon to estimate Pes related parameters when the aim is to detect changes between different conditions, but possibly not when the absolute value of the estimated parameters is of paramount importance. Due to a better signal-to-noise ratio, and considering its practical advantages in terms of easier calibration and simpler acquisition setup, liquid-filled esophageal catheters without balloon may prove interesting for clinical use.
Funding Statement
The study was partially funded by the MedDrive Program of the Faculty of Medicine of the Technical University Dresden (Germany). Alessandro Beda's work is supported by FAPEMIG (Programa Pesquisador Mineiro, Brazil) and CNPq (Bolsista de Produtividade em Pesquisa, Brazil). Nadja C. Carvalho is supported by CAPES (Brazil). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1. Talmor D, Sarge T, Malhotra A, O’Donnell CR, Ritz R, et al. (2008) Mechanical ventilation guided by esophageal pressure in acute lung injury. N Engl J Med 359: 2095–2104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Benditt JO (2005) Esophageal and gastric pressure measurements. Respir Care 50: 68–75. [PubMed] [Google Scholar]
- 3. Jubran A (1998) Monitoring patient mechanics during mechanical ventilation. Crit Care Clin 14: 629–53. [DOI] [PubMed] [Google Scholar]
- 4. Akoumianaki E, Maggiore SM, Valenza F, Bellani G, Jubran A, et al. (2014) The Application of Esophageal Pressure Measurement in Patients with Respiratory Failure. Am J Respir Crit Care Med 189: 520–531. [DOI] [PubMed] [Google Scholar]
- 5.Zin WA, Milic-Emili J (2005) Esophageal Pressure Mesurement. In: Hamid Q, Joanne Shannon MD, Martin JG, editors. Physiologic Basis of Respiratory Disease. BC Decker, Incorporated. pp. 639–647. [Google Scholar]
- 6. Milic-Emili J, Mead J, Turner JM, Glauser EM (1964) Improved technique for estimating pleural pressure from esophageal balloons. J Appl Physiol 19: 207–211. [DOI] [PubMed] [Google Scholar]
- 7. Baydur A, Behrakis PK, Zin WA, Jaeger M, Milic-Emili J (1982) A simple method for assessing the validity of the esophageal balloon technique. Am Rev Respir Dis 126: 788–791. [DOI] [PubMed] [Google Scholar]
- 8. Higgs BD, Behrakis PK, Bevan DR, Milic-Emili J (1983) Measurement of pleural pressure with esophageal balloon in anesthetized humans. Anesthesiology 59: 340–343. [DOI] [PubMed] [Google Scholar]
- 9. Asher MI, Coates AL, Collinge JM, Milic-Emili J (1982) Measurement of pleural pressure in neonates. J Appl Physiol 52: 491–494. [DOI] [PubMed] [Google Scholar]
- 10. Mead J, Gaensler EA (1959) Esophageal and pleural pressures in man, upright and supine. J Appl Physiol 14: 81–83. [DOI] [PubMed] [Google Scholar]
- 11. Pelosi P, Goldner M, McKibben A, Adams A, Eccher G, et al. (2001) Recruitment and derecruitment during acute respiratory failure: an experimental study. Am J Respir Crit Care Med 164: 122–130. [DOI] [PubMed] [Google Scholar]
- 12. De Chazal I, Hubmayr RD (2003) Novel aspects of pulmonary mechanics in intensive care. Br J Anaesth 91: 81–91. [DOI] [PubMed] [Google Scholar]
- 13. Hurewitz AN, Sidhu U, Bergofsky EH, Chanana AD (1984) How alterations in pleural pressure influence esophageal pressure. J Appl Physiol 56: 1162–1169. [DOI] [PubMed] [Google Scholar]
- 14. Dechman G, Sato J, Bates JHT (1992) Factors affecting the accuracy of esophageal balloon measurement of pleural pressure in dogs. J Appl Physiol 72: 383–388. [DOI] [PubMed] [Google Scholar]
- 15. Bernard GR (2008) PEEP Guided by Esophageal Pressure - Any Added Value? N Engl J Med 359: 2166–2168. [DOI] [PubMed] [Google Scholar]
- 16. Baydur A, Cha EJ, Sassoon CS (1987) Validation of esophageal balloon technique at different lung volumes and postures. J Appl Physiol 62: 315–321. [DOI] [PubMed] [Google Scholar]
- 17. Welch P (1967) The use of fast Fourier transform for the estimation of power spectra: a method based on time averaging over short, modified periodograms. Audio Electroacoust IEEE Trans 15: 70–73. [Google Scholar]
- 18. Bland JM, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 327: 307–310. [PubMed] [Google Scholar]
- 19. Coates AL, Davis GM, Vallinis P, Outerbridge EW (1989) Liquid-filled esophageal catheter for measuring pleural pressure in preterm neonates. J Appl Physiol 67: 889–893. [DOI] [PubMed] [Google Scholar]
- 20. Maxted KJ, Shaw A, Macdonald TH (1977) Choosing a catheter system for measuring intra-oesophageal pressure. Med Biol Eng Comput 15: 398–401. [DOI] [PubMed] [Google Scholar]
- 21. Karason S, Karlsen KL, Lundin S, Stenqvist O (1999) A simplified method for separate measurements of lung and chest wall mechanics in ventilator-treated patients. Acta Anaesthesiol Scand 43: 308–315. [DOI] [PubMed] [Google Scholar]
- 22. Mead J, McIlroy MB, Selverstone NJ, Kriete BC (1955) Measurement of Intraesophageal Pressure. J Appl Physiol 7: 491–495. [DOI] [PubMed] [Google Scholar]
- 23. Panizza JA, Finucane KE (1992) Comparison of balloon and transducer catheters for estimating lung elasticity. J Appl Physiol 72: 231–235. [DOI] [PubMed] [Google Scholar]
- 24. Drummond GB, Wright ADG (1983) Inaccuracy of oesophageal pressure for pleural pressure estimation in supine anaesthetized subjects. Br J Anaesth 55: 585–593. [DOI] [PubMed] [Google Scholar]
- 25. Hubmayr RD (2010) Is there a place for esophageal manometry in the care of patients with injured lungs? J Appl Physiol 108: 481–482. [DOI] [PubMed] [Google Scholar]
- 26. Cheng Y-PP, Wu H-DD, Jan G-JJ, Wang C-YY (2001) Removal of cardiac beat artifact in esophageal pressure measurement via a modified adaptive noise cancellation scheme. Ann Biomed Eng 29: 236–243. [DOI] [PubMed] [Google Scholar]
- 27. Schuessler TF, Gottfried SB, Goldberg P, Kearney RE, Bates JHT (1998) An adaptive filter to reduce cardiogenic oscillations on esophageal pressure signals. Ann Biomed Eng 26: 260–267. [DOI] [PubMed] [Google Scholar]
- 28. Washko GR, O’Donnell CR, Loring SH (2006) Volume-related and volume-independent effects of posture on esophageal and transpulmonary pressures in healthy subjects. J Appl Physiol 100: 753–758. [DOI] [PubMed] [Google Scholar]
- 29. Ferris BG, Mead J, Frank NR (1959) Effect of body position on esophageal pressure and measurement of pulmonary compliance. J Appl Physiol 14: 521–524. [Google Scholar]
- 30. Beardsmore CS, Helms P, Stocks J, Hatch DJ, Silverman M (1980) Improved esophageal balloon technique for use in infants. J Appl Physiol 49: 735–742. [DOI] [PubMed] [Google Scholar]